Abstract
This study investigates whether examining the shape of arterial pulse waves and fitting to them a physiology-based mathematical model of pulse wave propagation can provide additional insights into the state of the cardiovascular system in patients with severe traumatic brain injury (sTBI), potentially enhancing vasopressor dosing strategies. We conducted a longitudinal study on 25 sTBI patients in an intensive care unit. Arterial pulse waves were recorded non-invasively from wrists and ankles using an oscillometric method and were used to inform a 0-1D model of the arterial blood flow dynamics. Model-estimated, patient-specific cardiovascular parameters were then used in a statistical model to predict changes in the administered dose of vasopressor (norepinephrine) in the next 24 hours. The model fits to the recorded pulse waves were satisfactory, with the coefficients of determination () of approximately 0.9 and the differences between the measured and model-estimated mean arterial pressure of 0.1 ± 1.0 mmHg (
=0.99). Except for a few patients, we found no clear association between the model-estimated parameters and norepinephrine dose at the time of pulse wave recording. Nevertheless, our predictive model achieved a balanced accuracy of 0.85 when trained and tested on the entire dataset and 0.76 when using the leave-one-out cross-validation, with 8 misclassifications among the total of 77 observations. Thus, despite the known inter-patient variability of hemodynamic response to vasopressors, the proposed method allowed predicting the direction of norepinephrine dose changes in the next 24 hours with satisfactory accuracy. Subject to further studies and extensive validation, our approach could inform a decision-support tool for optimizing vasopressor dosing on a per-patient basis.
Authors summary
Hypotension is a dangerous complication in patients with severe traumatic brain injury (sTBI), making precise control of systemic blood pressure of utmost importance. To maintain blood pressure at prescribed levels, physicians often administer vasopressors (for example, norepinephrine). In current clinical practice, the vasopressor dose is frequently adjusted by trial and error, guided by continuous monitoring of the patient’s hemodynamic state. This approach is time-consuming and likely sub-optimal, highlighting the need for more efficient methods to guide vasopressor therapy. In this study, we investigated whether non-invasive peripheral pulse wave recordings, combined with mathematical modeling of cardiovascular dynamics, could help predict changes in vasopressor dosing. Our results show that by incorporating personalized cardiovascular parameters into a statistical predictive model, it is possible to predict - with satisfactory accuracy - the direction of change (or lack thereof) in the administered norepinephrine dose within the next 24 hours. With further research and rigorous validation, this approach may support more effective vasopressor management using easily available, non-invasive clinical data.
Citation: Wolos K, Pstras L, Bialonczyk U, Debowska M, Dabrowski W, Siwicka-Gieroba D, et al. (2025) Personalized pulse wave propagation modeling to improve vasopressor dosing management in patients with severe traumatic brain injury. PLoS Comput Biol 21(9): e1013501. https://doi.org/10.1371/journal.pcbi.1013501
Editor: Stacey D. Finley, University of Southern California, UNITED STATES OF AMERICA
Received: April 6, 2025; Accepted: September 8, 2025; Published: September 15, 2025
Copyright: © 2025 Wolos et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Data Availability: The GitHub repository, which contains an implementation of the mathematical model presented in this study and the anonymized data, can be found at: https://github.com/kwolos/STBI.
Funding: JP, MD, WD, DSG and KW were partly supported by the National Science Centre (Poland), grant No. 2018/31/D/ST7/03472. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
Competing interests: The authors have declared that no competing interests exist.
Introduction
Severe traumatic brain injury (sTBI) can lead to permanent disability or death, and its treatment remains a challenge [1,2]. Among the many complications associated with sTBI, hemodynamic instability, particularly hypotension, is of utmost importance [3]. An abnormally low blood pressure poses a substantial risk to patients with sTBI, exacerbating cerebral ischemia and further compromising neurological function [4,5]. To avoid hypotension in sTBI patients, clinicians commonly use vasopressors, particularly norepinephrine (NE) [6] which, through various mechanisms, mainly peripheral vasoconstriction [7], leads to an increase in arterial blood pressure. Currently, adjusting vasopressor doses in critically ill patients is an iterative process based on observing mid- to long-term changes in vital physiological parameters such as mean arterial pressure (MAP) or systolic pressure (SP). According to the guidelines for sTBI patients, to decrease mortality and improve outcomes, MAP should be maintained above 80 mmHg [6], whereas SP should be maintained above 100 mmHg in patients aged 50–69 years, and above 110 mmHg in other patients [8]. However, such an iterative, often trial-and-error process might be far from optimal, carrying the risk of transient sub-optimal dose levels, which may lead to inadequate perfusion followed by hypoperfusion-related complications or to inadvertent overdosing, resulting in other adverse effects [9]. For instance, a study on 1093 ICU patients showed that their MAP was within the target range only in 43% of time (on average), while being below and above the target in 9% and 48% of time, respectively [10]. Therefore, there is a clear need to develop new tools that would help guide the process of vasopressor dosing in sTBI patients, to ensure the best possible treatment outcomes and minimize mortality.
Several approaches to the problem of guiding vasopressor dose adjustments in critically ill patients using mathematical modeling can be found in the literature. Bighamian et al. introduced a latency-dose-response cardiovascular model to predict MAP and other cardiovascular parameters in response to epinephrine dosing [11,12], employing phenomenological equations with parameters personalized for both normotensive and hypotensive patients, as well as piglets. Yapps et al. employed a logistic regression model to predict hypotension events using data on MAP and vasopressor dosing, either taken from a public database (MIMIC II) or collected in surgical ICU patients [13]. Tang et al. demonstrated a model for MAP prediction based on current MAP, heart period, NE infusion rate, and respiratory rate [14], which was later extended and validated using data from septic patients [15]. Recently, Kao et al. proposed a lumped-parameter cardiovascular model that incorporates baroreflex feedback and a dynamic dose-response model of vasopressors, validated using data from piglets receiving phenylephrine [16]. Also, machine-learning approaches have been explored to predict vasopressor requirements in critically ill patients [17–20]. However, none of the above studies addressed patients with sTBI.
Interestingly, Johnston et al., who investigated the pharmacodynamics and pharmacokinetics of dopamine and norepinephrine in a group of eight patients with moderate to severe head injury, found that the pharmacodynamics of NE seem to be unpredictable in these patients, showing no significant correlations between NE dose levels or plasma concentrations and MAP, cardiac index (CI), or systemic vascular resistance index (SVRI) [21]. While higher NE doses were generally associated with higher MAP and higher SVRI, they found no correlations between changes in plasma NE and changes in MAP, CI, or SVRI, which suggests inter-patient variability of hemodynamic response to NE. Therefore, it appears difficult to quantitatively predict the exact hemodynamic effects of increasing the dose of NE.
This apparent unpredictability of cardiovascular response to NE may stem from the fact that current methods primarily rely on analyzing only a few standard hemodynamic parameters, which may not be sufficient. The arterial pulse waveform can provide more detailed information on the status of the cardiovascular system, as it is the sum of the forward and reflected pulse waves, with reflections occurring wherever there is a change in blood flow, i.e., at vessel bifurcations or where the stiffness of the arteries changes [22]. Analysis of the pulse waveform can therefore provide additional information about the state of the arterial system [22]. For example, it can help assess arterial stiffness and evaluate the effects of vasoactive drugs [23,24]. Our goal was to determine whether information obtained from personalized pulse wave propagation modeling [25–29] could be useful in predicting the direction of change (or lack thereof) in the administered NE dose in sTBI patients.
Materials and methods
Ethics statement
The study was approved by the Bioethical Committee at the Medical University of Lublin, Poland (KE-0254/253/2020) and was performed in accordance with the Declaration of Helsinki and all applicable regulations. Written informed consent was obtained from each patient or their legal representative in case the patient was unconscious and/or under sedation at the time of enrolment in the study.
Study subjects
The study involved a cohort of 25 sTBI patients routinely treated at the intensive care unit; see Table 1 for group characteristics. Each patient was continuously monitored using a hemodynamic monitor, and based on its readings, especially MAP, and the patient’s overall clinical condition, clinicians manually adjusted the vasopressor doses. All dose changes, along with the times of those changes, were documented. All patients received NE. Five patients received additionally dobutamine. Due to missing data on NE dosage, we had to exclude data from 3 patients. Data from further 2 patients were excluded due to poor quality of the recorded pulse waves. Thus, the final analysis is based on data from 20 patients.
Pulse wave recordings
Arterial pulse waves were recorded once per day simultaneously on both wrists and ankles (four sites in total) using the AngE device (SOT Medical, Austria), which can perform simultaneous oscillometric recordings of pulse waves at multiple sites along with electrocardiogram (ECG); see Fig 1A for an overview of the measurement protocol. The device initially inflates the cuffs to 180 mmHg and records the pressure oscillations in all cuffs for 5 seconds. The pressure in all cuffs is then reduced by 10 mmHg and the 5-second recording is repeated. This process continues until the cuff pressure reaches 40 mmHg. Thus, the total duration of a single recording is about 2 minutes; see Fig 1B for exemplary pulse wave recording. The recorded oscillations represent changes in the cuff pressure caused by the pulsatile changes of arterial volume under the cuff (attenuated to some extent), and hence for cuff pressure close to the diastolic pressure the shape of the obtained wave may be treated as relatively closely resembling the shape of the arterial volume wave [30,31]. After the recording, for each cuff, the device selects the 5-second fragment of the wave with the highest amplitude, potentially recorded at different pressure levels in different cuffs (hence at different times). For our analysis, we took from all cuffs the pulse wave fragments corresponding to the same cuff pressure level (the closest to the average level selected by the device) to analyze the waves recorded simultaneously at all sites. For each recording site, we then averaged all beats from the selected fragment of the recorded wave to obtain a single arterial volume waveform. We did not observe any significant wave profile changes following the averaging procedure. Such average waveforms from all four recording sites were used in the model optimization step (see further sections).
(a) During the treatment (up to 5 days) one recording of pulse waves was taken each day (at various times of day) using the AngE device, while the vasopressor dose was routinely adjusted. (b) Exemplary pulse waves recorded by the AngE device.
Note that in order to obtain the waveform that would most closely reflect the shape of the true arterial volume waveform, one should perform the recording with the cuff pressure kept at a level slightly below the local diastolic pressure (DP), since cuff pressures above DP lead to temporary occlusion of the arteries and hence lead to some distortion of the recorded waves [30,31]. However, at low cuff pressures, the amplitude and the overall quality of the recorded wave are much lower, which in some cases makes it difficult to identify individual beats and ultimately makes the averaged shape of the waveform less reliable (especially in patients with low pressure amplitude). Therefore, in this study, we decided to use the waves with the highest amplitude (as mentioned earlier), which were recorded at cuff pressure close to MAP, thus somewhat higher than DP. In other words, we used waveforms whose shape may differ slightly from the shape of actual arterial volume waves, but this enabled us to efficiently identify individual beats and hence obtain the averaged waveforms that could be used in our framework. Moreover, even if there was no occlusion of the arteries below the cuff, the waveform observed at the cuff level could still be slightly different from the actual arterial volume waveform due to the fact that the tissues between the arteries and the cuff (including the artery wall itself) may transmit the volume changes to a different extent at different pressure levels (in all cases largely attenuated).
Note also that in this study, we fit the model-simulated volume waveforms from a specific artery under the cuff (e.g., the radial artery) to the cuff-recorded waveforms, which in fact, reflect blood volume changes not in one specific artery but in all arteries below the cuff (e.g., also in the ulnar artery). This is another simplification of our approach, in which we assumed that the shapes of the volume waveforms in the radial and ulnar arteries are relatively similar to each other, and hence we treated the waveform observed at the cuff level as corresponding to the volume waveform of the radial artery. Moreover, given the relatively low changes in the diameter of the radial artery along its length, we assume that the volume waveform of the fragment of the radial artery under the cuff is equivalent to volume waveforms in other parts of the radial artery, and hence, for simplicity, we analyze model-simulated volume waveform of the whole radial artery (similarly for the anterior tibial artery for the recordings at the ankle cuffs).
Other data obtained from the AngE device
In addition to recording the oscillometric waves, the AngE device also provides a number of cardiovascular parameters, some of which were used in our analysis. These include information on heart rate (HR) and blood pressure, i.e., systolic pressure (SP), diastolic pressure (DP), and mean arterial pressure (MAP), as well as parameters related to the recorded pulse waves, i.e., the duration of the rise in the pulse waveform from its foot to peak (‘rise time’), and the ratio of the duration of the rise in the pulse waveform and the duration of its fall from peak to the foot of the next wave (‘rise to fall’), calculated for the waveforms from each limb separately.
Pulse wave propagation model
We utilized a one-dimensional arterial network model composed of 71 main compliant arteries connected to zero-dimensional boundary conditions. For a detailed description of the model, please refer to our previous publications [32–34] and S1 File. The model simulates pulsatile blood flow and the corresponding blood pressure and volume waves in the whole modeled arterial tree, taking into account patient-specific characteristics, such as terminal vascular resistances, stiffness of arteries, and left-ventricular time-varying elastance function (LVTVE). We used the Levenberg-Marquardt optimization algorithm to minimize the objective (error) function by adjusting selected subject-specific model parameters. The objective function included the differences (errors) between simulated and recorded pulse waveforms (arterial volume waveforms) in four considered locations, the differences between recorded (on the cuff on the left wrist) and simulated (on the radial artery) SP and DP, and a penalty term to penalize abnormal (unphysiological) values of the simulated stroke volume (SV), according to the following equation:
where ,
and
denote normalized, simulated and measured values, respectively,
is a
-norm,
=1000 is the number of analyzed time points in the normalized simulated and measured arterial volume waveforms (
; see S1 File and Fig 2 for details (including the explanation with regard to the weights of the individual terms in our objective function).
Fitting of the model-simulated arterial volume waveforms at four sites (wrists and ankles) to recorded waveforms by tuning selected model parameters.
In order to select model parameters to be optimized during the above minimization procedure, we performed a sensitivity analysis combined with an identification analysis. To this end, we first performed the Sobol’ sensitivity analysis [35,36] and calculated the first-order sensitivity indices (for each studied parameter, at multiple time points of the cardiac cycle), which quantify the direct contributions of individual parameters to the given model output variability (see Sensitivity analysis section in S1 File for details). Then we performed a parameter identification analysis using the procedure described by Olufsen et al. [37], which determines the possible pairwise correlations between the selected model parameters by calculating the relative local sensitivity matrix followed by the calculation of the inverse of the Hessian matrix (see Parameter identification section in S1 File for details). Following the above calculations, we selected five model parameters for which the analyzed outputs of the model (i.e. arterial volume waveforms at the considered locations as well as SP and DP) showed the greatest sensitivity and which were not highly correlated with each other. In particular, we selected the following model parameters: (time to the onset of constant LVTVE),
(maximal value of the LVTVE function),
(scaling factor for the resistances of small arteries and arterioles),
(pressure in the left atrium), and
(a parameter describing the stiffness of the large arteries). The values of all other parameters of the model were fixed at the levels taken from the literature as typical values for a 45-year-old man weighing 75 kg. Specifically, the values of parameters describing the stiffness of the arteries were based on a formula derived from the work of Olufsen et.al. [38], while the resistance and compliance parameters of small arteries and arterioles were taken from the work of Alastruey et al. [39]. In addition, some of the fixed parameter values were also scaled according to patient’s height. More details on the selection of individual parameters and their exact values, along with references, can be found in S1 File.
Predictive model
To investigate the clinical utility of the model-derived, patient-specific cardiovascular parameters, we incorporated them into a statistical model to predict whether the next NE dose adjustment within 24 hours will increase the dose or not. More precisely, for each case (i.e., each time of pulse wave recording), we assigned one of two labels as follows:
- 0 (negative class) - if within 24 hours from the given pulse wave measurement, there was no dose adjustment or the first adjustment lowered the NE dose,
- 1 (positive class) - if the first dose adjustment within 24 hours from the pulse wave measurement increased the NE dose.
We used a generalized mixed linear model with a binomial link function. This model can be viewed as a logistic regression classifier suitable for data where the assumption of observation independence is violated. A detailed description of the method can be found in [40] . We considered two versions of the predictive model – a full model and a simplified model. The features considered in our full model included five patient-specific cardiovascular parameters obtained from fitting the 0-1D model to recorded pulse waves and blood pressure data ,
,
, and
), patient characteristics (age, height, and weight), standard cardiovascular parameters measured by the AngE device (HR, SP, DP, MAP), pulse wave-related parameters provided by the AngE device (rise time of the recorded waveform from all four cuffs, and ‘rise to fall’ ratio of the recorded waveform from all four cuffs), current NE dose, and pulse wave velocity (PWV) calculated using our pulse wave propagation model (between the beginning of the ascending aorta and the end of the femoral artery, using the foot-to-foot method). To assess the utility of the parameters related to pulse waveforms (both obtained from the AngE device and estimated from our model), we also analyzed a simplified model that excluded these parameters and was based solely on patient characteristics, HR and blood pressure data, and NE dose. All features were standardized to avoid numerical problems in the process of optimizing model parameters. For both full and simplified models, the feature selection for the final model was performed using the step-wise procedure based on the Akaike Information Criterion [41]. The model performance was validated using the leave-one-out cross-validation, a suitable approach for small datasets [42]. A probability threshold of 0.5 was used, meaning that patients with a predicted probability of having an increased vasopressor dose above 0.5 were classified as belonging to the positive class.
Results
Model fits to recorded pulse waveforms
The quality of the model fits to the recorded arterial volume waveforms was satisfactory, with the mean coefficients of determination () of 0.92, 0.91, 0.88, and 0.90 for the left arm, right arm, left leg, and right leg, respectively; Fig 3A shows the plots comparing all normalized model-estimated and recorded arterial volume waveforms. The model accurately estimated MAP with the mean difference between the model-derived and measured values of 0.1 ± 1.0 mmHg (
), with no dependence on the MAP level (Pearson’s r = 0.11, p-value = 0.35), see Fig 3B. An exemplary model fit to the recorded waveform is presented in Fig 3C; see Figs A – D in S2 File for all model fits.
(a) Comparison of the recorded arterial volume waveforms and corresponding in time model-simulated volume waveforms, with both the recorded and simulated waveforms normalized in amplitude (n = 77). The perfect match of the two waveforms corresponds to a diagonal line. (b) Bland-Altman plot comparing the measured and model-estimated mean arterial pressure for all analyzed cases (n = 77). (c) Typical model-simulated arterial volume waveforms (black lines) compared to the recorded (averaged) waveforms (light blue lines).
Correlations between vasopressor doses and estimated parameters
After fitting the model to the recorded waveforms and blood pressure data, we compared the estimated patient-specific cardiovascular parameters with the NE doses at the times of pulse wave recordings. We found no correlations between the model parameters and NE doses neither for the entire cohort, nor for individual patients. Only in a few patients (ID =1, 2, 4) we observed some qualitative associations between the NE doses and parameters and
, see Fig 4. Similarly, we compared SP and DP (both estimated by the model and measured by AngE). In this case, we also did not observe any correlations, except for patient ID = 6, in whom SP and DP values also increased with increasing NE dose during treatment (see Figs A and B in S3 File).
Shown are values of the parameters (blue dots) and norepinephrine doses (red lines).
Prediction of vasopressor dose change
The dataset used in our analysis consisted of 88 observations (i.e., separate pulse wave recordings) from a group of 20 patients, among which for 77 observations we were able to assign a label 0 or 1 (the remaining observations did not have sufficient follow-up time of 24 hours). During the step-wise feature selection for our predictive model (full model), 10 features were found to be associated with the target outcome; see Table 2. Notably, these features included two model-estimated cardiovascular parameters: , and
, that is left-atrial pressure, and maximal value of the LVTVE function, respectively. When trained and tested on the entire dataset, the model achieved a balanced accuracy of 0.85 (accuracy 0.90). When validated using the leave-one-out cross-validation (LOOCV), the model’s balanced accuracy was 0.76 (accuracy 0.82). The confusion matrices for the full model trained and tested on the entire dataset or tested using LOOCV are presented in Table 3.
Table 2 also shows the features selected for the simplified model, which excluded pulse wave-related parameters (either derived from the AngE device or obtained from the pulse wave propagation model). This simplified model exhibits lower balanced accuracy compared to the full model, both when trained and tested on the entire dataset (0.75) and when validated using LOOCV (0.67). A detailed comparison of the performance metrics of the two models is provided in Table 4.
In S4 File, we illustrated the predictions of the NE dose adjustments corresponding to all pulse wave recordings in all analyzed patients (the results from the full model trained and tested on the entire dataset). In some cases, despite the incorrect prediction, the model correctly anticipated the overall trend in dose adjustments (see patients with ID 7 and 18). Interestingly, in the case of the patient with ID = 18, the prediction in Day 1 was a false negative (see Fig A in S4 File) with a very low probability value (0.02). This prediction was false because at the next dose adjustment the NE dose was increased; however, this dose increase was only transient, and soon the dose was decreased to much lower levels.
For some observations from the last day, we did not have information about the potential changes in NE dose in the following 24 hours. These cases were excluded from our main analysis, but we proceeded with predictions for these cases anyway, even if their accuracy could not be assessed. In each of these cases, the model predicted that the dose would remain constant or be decreased within the next 24 hours, which seems to be expected, as one would expect the patient’s condition to improve after 4–5 days of treatment.
Discussion
In our study on patients with severe traumatic brain injury, we explored the potential of using information extracted from arterial pulse waveforms (either directly or through pulse wave propagation modeling) to predict changes in NE dose administration that will likely be needed within the next 24 hours. We showed that knowledge of selected patient-specific cardiovascular parameters derived from the pulse waveforms, which are not measured during standard treatment procedures, appears to be useful for predicting changes in the administered NE dose in these patients possibly providing some insights into the mid- to long-term effects of the current NE dose considering the current state of patient’s cardiovascular system, thus providing the basis for a tool that could be potentially used to guide vasopressor dosing, in particular with regard to adjusting the doses sooner.
Key results
The proposed statistical model predicting NE dose adjustments (full model) is based on 10 features, including three features derived from the 0-1D cardiovascular model (,
, and PWV) and two features computed directly from the recorded pulse waveform (‘rise time’ and ‘rise to fall’ ratio in the right arm). Integrating the parameters available from the AngE measurements with parameters obtained from the cardiovascular model optimization enhances the performance of our predictive model. Specifically, the model that included pulse-wave-derived features (full model), exhibited a 33.3% higher sensitivity (i.e., the rate of correct predictions of NE dose increases) and 13.4% higher balanced accuracy compared to the model without those features (simplified model) when trained and tested on the entire dataset; see Table 4 for more details. The markedly higher sensitivity while maintaining similar specificity is essential from our perspective, as accurately predicting that the vasopressor dose is likely to be increased within the next 24 hours seems to be particularly important when focused on preventing episodes of hypotension in sTBI patients.
All analyzed patients were treated with NE; four patients were also given dobutamine. The effects of NE include, among others, vasoconstriction, which leads to increased arterial stiffness [43]. In line with this, the proposed predictive model accounts for the model-estimated PWV, a standard marker of arterial stiffness [44]. NE also improves end-systolic elastance [45] and increases the heart rate [46] and systemic diastolic blood pressure [47], all of which are features of our full model (end-systolic elastance is described by ). According to our predictive model, higher values of
and PWV are associated with a decreased likelihood of requiring an increased NE dose within the next 24 h. This is not unexpected, given that these parameters are related to blood pressure. The fact that they were selected as predictors for our predictive model suggests that they may carry information on the current state of the cardiovascular system that may not be captured by standard cardiovascular parameters. Interestingly, our full predictive model does not include MAP as a predictor, although it is the main parameter used by clinicians to determine vasopressor dosage. However, MAP can be estimated from systolic blood pressure (SP) and diastolic pressure (DP) using a linear relationship, such as MAP = 2/3DP + 1/3SP [48], which may explain the absence of MAP and presence of DP in our full model. Indeed, during our various tests, when we removed SP and DP from the set of potential predictors, MAP emerged as an important predictor.
Secondary findings
We also compared the patient-specific cardiovascular parameters estimated during model optimization (i.e., ,
,
,
and
) with the NE doses. In general, we did not find any relationship between NE doses and the estimated parameter values. As shown in previous studies [21,49] the pharmacodynamics of NE appear unpredictable and may be due to patient variability. Such variability may explain the observed lack of relationship between the values of the estimated parameters and the administered NE doses in our cohort. However, in three patients (ID =1, 2, 4) we observed relatively similar patterns between NE doses and the values of two estimated parameters:
and
. We believe that this was not coincidental and may be due to the fact that NE is responsible, among others, for the increase in peripheral resistance (
is the scaling factor for the resistances of small arteries and arterioles) and arterial stiffness (
describes the stiffness of large arteries). The fact that such relationships were not observed in other patients may be due to the effects of other vasoactive substances, e.g., endogenous catecholamines, which can disrupt the relationship between NE dose and the discussed parameters [50], or due to impact of other drugs not considered in our study, or due to neural mechanisms controlling vascular tone (also not considered in our study).
Other studies
The problem of optimizing the dosing of vasopressors in critically ill patients has been the subject of multiple studies. For predicting a patient’s condition or the required vasopressor dose, researchers mainly use standard hemodynamic parameters that are routinely measured during standard treatment procedures and validate their methods using data from ICU patients [17–20]. In the case of mathematical models of the hemodynamic response to vasopressors, the models are mainly validated using data from either healthy individuals [11,12], septic patients [14,15], or animals [16]. To our knowledge, however, no such predictive or modelling studies have focused on sTBI patients, and hence our study seems to be the first to use personalized cardiovascular parameters estimated from a pulse wave propagation model (based on non-invasive oscillometric measurements of arterial pulse waveforms) to be fully conducted on data from a cohort of sTBI patients treated with NE.
Recently, a novel approach to vasopressor dosing has been proposed by Rinehart et. al. who developed a closed-loop controller for automatized vasopressor administration (targeting MAP) in perioperative or intensive care settings [51,52].This controller has been subsequently tested by Joostens et. al. in a randomized controlled study on titrating NE in 18 sTBI patients [53], showing that in patients in whom NE was titrated in an automatically controlled manner, MAP remained within the target range 96% of time on average (over the four-hour study period), as compared to only 43% of time in the group with standard, manual NE titration. The approach by Rinehart and Joostens, i.e., automatized NE titration, is entirely different to that considered in our study, and while it offers a great perspective for a fully automatized process of vasopressor dosing (not only in sTBI patients but in other critically ill patients or in patients undergoing surgeries), it requires special or modified equipment for drug titration. On the other hand, our approach could potentially lead to improved vasopressor dosing strategies within standard care using the existing equipment. Moreover, it is possible that integrating pulse wave features in a controller targeting MAP could potentially lead to even better performance of such a controller.
Limitations
Our study has certain limitations. First, our findings are based on relatively limited data from a small cohort of patients, with only one pulse wave recording per day at various, arbitrary times, unrelated to vasopressor dose adjustments (although despite this, our full model was able to predict with relatively high balanced accuracy (0.85) whether the NE dose would be increased within the next 24 hours). Second, the data to which the model is fitted is subject to error measurement and natural variability (especially blood pressure), and therefore it may not necessarily reflect accurately the current (average) state of the cardiovascular system. Moreover, although the model parameters that were fixed rather than adjusted appear to have little effect on the arterial volume waveforms and blood pressure, possible changes in many of these parameters could collectively have a non-negligible effect on the behavior of the modeled system and thus on the values of the estimated (adjusted) parameters, which in turn could affect the predictions in our statistical model. Furthermore, by adjusting the value of parameter , we assume that the resistance of all small arteries and arterioles (the terminal 0D elements in our model) changes to the same extent (in percentage terms, from their baseline values), which is a simplification, given that vessels in different tissues and organs may respond differently to vasoactive substances (not only in terms of the magnitude of the change in their tone but even in terms of the direction of this change) and may be also subject to varying degrees of vascular autoregulation (e.g., in cerebral vessels, although in patients with sTBI cerebrovascular autoregulation is frequently impaired [54]). Also, for simplicity, in all our simulations we assumed intact cerebral vasculature, thus ignoring the fact that patients with sTBI will likely experience alterations in cerebrovascular function, which could affect the total cerebral resistance and compliance and thus could influence central and peripheral hemodynamics, potentially affecting the results of our study. Finally, the presented approach is based on recording pulse waveforms at multiple locations (to account for the possible differences in peripheral pulse waveforms between different limbs, we chose to fit our cardiovascular model to pulse waveforms from four sites, recorded oscillometrically). We want to highlight, however, that the proposed methodology could be potentially used with pulse waveform recorded only at a single site, thus not necessarily requiring recordings on four limbs. Moreover, one could use arterial volume waveforms recorded using other methods, or use arterial pressure waveforms (e.g., recorded using applanation tonometry).
Conclusions and future work
This study provides a preliminary demonstration of the potential of pulse wave propagation modeling for sTBI patient management. Using the data collected in sTBI patients, we developed a statistical model to predict vasopressor dose increases within 24 hours with relatively high balanced accuracy (0.85). Including in this predictive model the patient-specific parameters related to arterial stiffness and heart function, estimated from the pulse wave propagation model, turned out to provide a higher model sensitivity compared to the model without pulse wave-related features. The proposed framework could be potentially used to adjust vasopressor doses not only based on the current condition of the patient (as is currently done) but also (to some extent) taking into account predictions of vasopressor dose adjustments that are likely to be needed in the future. However, it is not obvious yet how exactly the vasopressor doses could be adjusted taking into account the predictions of the proposed model and whether this would lead to improved patient outcomes. Moreover, here we proposed a model for predicting whether the vasopressor dose will be increased by clinicians within 24 hours from the given time (the time of recording the pulse wave and blood pressure), which can be treated only as indirect information of the likely future state of the patient. Ideally, a similar framework should be developed with a model predicting directly the future state of the patient (e.g., MAP). Future studies should investigate these aspects using more comprehensive models (with various cardiovascular regulatory mechanisms) and more data, including more frequent pulse wave recordings, continuous vital parameter monitoring, and data on endogenous catecholamines.
Supporting information
S1 File. Detailed description of the mathematical model and sensitivity analysis.
https://doi.org/10.1371/journal.pcbi.1013501.s001
(PDF)
S2 File. Fits of the model-simulated arterial volume waveforms to the recorded waveforms.
https://doi.org/10.1371/journal.pcbi.1013501.s002
(PDF)
S3 File. Visualization of changes of systolic and diastolic pressure during the treatment.
https://doi.org/10.1371/journal.pcbi.1013501.s003
(PDF)
S4 File. Visualization of the vasopressor doses and their adjustments during the observation period, along with the predictions of the statistical model (full model) with regard to the changes in vasopressor dose within the next 24 hours from the time of pulse wave recording.
https://doi.org/10.1371/journal.pcbi.1013501.s004
(PDF)
References
- 1. Yan A, Torpey A, Morrisroe E, Andraous W, Costa A, Bergese S. Clinical Management in Traumatic Brain Injury. Biomedicines. 2024;12(4):781. pmid:38672137
- 2. Rakhit S, Nordness MF, Lombardo SR, Cook M, Smith L, Patel MB. Management and Challenges of Severe Traumatic Brain Injury. Semin Respir Crit Care Med. 2021;42(1):127–44. pmid:32916746
- 3. Robbins B, Almassalkhi L, Baum R, Blackburn M, Davis J, Edwards L. Pharmacotherapy adjuncts for traumatic brain injury: A narrative review of evidence and considerations in the emergency department. Am J Emerg Med. 2025;89:78–84.
- 4. Sookplung P, Siriussawakul A, Malakouti A, Sharma D, Wang J, Souter MJ, et al. Vasopressor use and effect on blood pressure after severe adult traumatic brain injury. Neurocrit Care. 2011;15(1):46–54. pmid:20878264
- 5. Toro C, Ohnuma T, Komisarow J, Vavilala MS, Laskowitz DT, James ML. Early vasopressor utilization strategies and outcomes in critically ill patients with severe traumatic brain injury. Anesth Analg. 2022;135(6):1245.
- 6. Rossaint R, Afshari A, Bouillon B, Cerny V, Cimpoesu D, Curry N, et al. The European guideline on management of major bleeding and coagulopathy following trauma: sixth edition. Crit Care. 2023;27(1):80.
- 7. Stampfl M, DeBlieux P. A clinical review of vasopressors in emergency medicine. J Emerg Med. 2024;67(1):e31-41.
- 8. Carney N, Totten AM, O’Reilly C, Ullman JS, Hawryluk GWJ, Bell MJ, et al. Guidelines for the Management of Severe Traumatic Brain Injury, Fourth Edition. Neurosurgery. 2017;80(1):6–15. pmid:27654000
- 9. Russell JA, Gordon AC, Williams MD, Boyd JH, Walley KR, Kissoon N. Vasopressor Therapy in the Intensive Care Unit. Semin Respir Crit Care Med. 2021;42(1):59–77. pmid:32820475
- 10. Rinehart J, Ma M, Calderon MD, Bardaji A, Hafiane R, Van der Linden P, et al. Blood pressure variability in surgical and intensive care patients: Is there a potential for closed-loop vasopressor administration?. Anaesth Crit Care Pain Med. 2019;38(1):69–71. pmid:30513357
- 11. Bighamian R, Reisner AT, Jin-Oh Hahn. An analytic tool for prediction of hemodynamic responses to vasopressors. IEEE Trans Biomed Eng. 2014;61(1):109–18. pmid:23955691
- 12. Bighamian R, Soleymani S, Reisner AT, Seri I, Hahn J-O. Prediction of Hemodynamic Response to Epinephrine via Model-Based System Identification. IEEE J Biomed Health Inform. 2016;20(1):416–23. pmid:25420273
- 13. Yapps B, Shin S, Bighamian R, Thorsen J, Arsenault C, Quraishi SA, et al. Hypotension in ICU Patients Receiving Vasopressor Therapy. Sci Rep. 2017;7(1):8551. pmid:28819101
- 14. Tang Y, Brown S, Sorensen J, Harley JB. Reduced Rank Least Squares for Real-Time Short Term Estimation of Mean Arterial Blood Pressure in Septic Patients Receiving Norepinephrine. IEEE J Transl Eng Health Med. 2019;7:4100209. pmid:31475080
- 15. Tang Y, Brown SM, Sorensen J, Harley JB. Physiology-Informed Real-Time Mean Arterial Blood Pressure Learning and Prediction for Septic Patients Receiving Norepinephrine. IEEE Trans Biomed Eng. 2021;68(1):181–91. pmid:32746013
- 16. Kao YM, Sampson CM, Shah SA, Salsbury JR, Tivay A, Bighamian R. A Mathematical Model for Simulation of Vasoplegic Shock and Vasopressor Therapy. IEEE Transactions on Biomedical Engineering. 2023;70(5):1565–74.
- 17. Holder AL, Shashikumar SP, Wardi G, Buchman TG, Nemati S. A Locally Optimized Data-Driven Tool to Predict Sepsis-Associated Vasopressor Use in the ICU. Crit Care Med. 2021;49(12):e1196–205. pmid:34259450
- 18. Kwak GH, Ling L, Hui P. Predicting the Need For Vasopressors in the Intensive Care Unit Using an Attention Based Deep Learning Model. Shock. 2021;56(1):73–9. pmid:33177372
- 19. Choi Y, Kim KH, Kim Y, Choi DH, Joo YH, Choi SW, et al. Prediction of vasopressor needs in hypotensive emergency department patients using serial arterial blood pressure data with deep learning. Hong Kong J Emerg Med. 2024;31(5):233–41.
- 20. Prasad V, Reisner AT, Lynch JC, Filbin MR, Heldt T. Modeling of Usual Care: Vasopressor Initiation for Sepsis With Hypotension. Front Med (Lausanne). 2022;9:715856. pmid:35360743
- 21. Johnston AJ, Steiner LA, O’Connell M, Chatfield DA, Gupta AK, Menon DK. Pharmacokinetics and pharmacodynamics of dopamine and norepinephrine in critically ill head-injured patients. Intensive Care Med. 2004;30(1):45–50. pmid:14586494
- 22.
Westerhof N, Stergiopulos N, Noble MIM, Westerhof BE. Wave Travel and Reflection. Snapshots of Hemodynamics. Springer International Publishing. 2018. 75–88. doi: https://doi.org/10.1007/978-3-319-91932-4_12
- 23.
O’Rourke M, Adji A. Contour of pressure and flow waves in arteries. McDonald’s blood flow in arteries. 7th ed. CRC Press. 2022.
- 24. Reavette RM, Ramakrishnan A, Rowland EM, Tang MX, Mayet J, Weinberg PD. Detecting heart failure from B-mode ultrasound characterization of arterial pulse waves. Am J Physiol-Heart Circ Physiol. 2024;327(1):H80-8.
- 25. Bikia V, Rovas G, Stergiopulos N. Cardiac output estimated from an uncalibrated radial blood pressure waveform: validation in an in-silico-generated population. Front Bioeng Biotechnol. 2023.
- 26. Bikia V, Papaioannou TG, Pagoulatou S, Rovas G, Oikonomou E, Siasos G, et al. Noninvasive estimation of aortic hemodynamics and cardiac contractility using machine learning. Sci Rep. 2020;10(1):15015. pmid:32929108
- 27. Bikia V, Pagoulatou S, Trachet B, Soulis D, Protogerou AD, Papaioannou TG, et al. Noninvasive Cardiac Output and Central Systolic Pressure From Cuff-Pressure and Pulse Wave Velocity. IEEE J Biomed Health Inform. 2020;24(7):1968–81. pmid:31796418
- 28. Zhang X, Wu D, Miao F, Liu H, Li Y. Personalized Hemodynamic Modeling of the Human Cardiovascular System: A Reduced-Order Computing Model. IEEE Trans Biomed Eng. 2020;67(10):2754–64. pmid:32142412
- 29. Zhang X, Liu J, Cheng Z, Wu B, Xie J, Zhang L, et al. Personalized 0D‐1D multiscale hemodynamic modeling and wave dynamics analysis of cerebral circulation for an elderly patient with dementia. Int J Numer Methods Biomed Eng. 2021.
- 30.
Stork M, Jilek J, Stork M, Jilek J. Cuff pressure pulse waveforms: their current and prospective applications in biomedical instrumentation. Biomedical engineering, trends in electronics, communications and software. IntechOpen. 2011.
- 31.
Jilek J, Stork M. The contours of arterial pulsations in the blood pressure cuff are hemodynamic waveforms rather than oscillations. In: Proceedings of the 2nd international conference on Circuits, systems, control, signals, Stevens Point, Wisconsin, USA, 2011. 140–3.
- 32. Poleszczuk J, Debowska M, Dabrowski W, Wojcik-Zaluska A, Zaluska W, Waniewski J. Patient-specific pulse wave propagation model identifies cardiovascular risk characteristics in hemodialysis patients. PLoS Comput Biol. 2018;14(9):e1006417. pmid:30216341
- 33. Poleszczuk J, Debowska M, Dabrowski W, Wojcik-Zaluska A, Zaluska W, Waniewski J. Subject-specific pulse wave propagation modeling: Towards enhancement of cardiovascular assessment methods. PLoS One. 2018;13(1):e0190972. pmid:29324835
- 34. Wołos K, Pstras L, Debowska M, Dabrowski W, Siwicka-Gieroba D, Poleszczuk J. Non-invasive assessment of stroke volume and cardiovascular parameters based on peripheral pressure waveform. PLoS Comput Biol. 2024;20(4):e1012013. pmid:38635856
- 35. Sobol′ IM. Global sensitivity indices for nonlinear mathematical models and their Monte Carlo estimates. Math Comput Simul. 2001;55(1):271–80.
- 36. Saltelli A, Annoni P, Azzini I, Campolongo F, Ratto M, Tarantola S. Variance based sensitivity analysis of model output. Design and estimator for the total sensitivity index. Comput Phys Commun. 2010;181(2):259–70.
- 37. Olufsen MS, Ottesen JT. A practical approach to parameter estimation applied to model predicting heart rate regulation. J Math Biol. 2013;67(1):39–68.
- 38. Olufsen MS, Peskin CS, Kim WY, Pedersen EM, Nadim A, Larsen J. Numerical simulation and experimental validation of blood flow in arteries with structured-tree outflow conditions. Ann Biomed Eng. 2000;28(11):1281–99. pmid:11212947
- 39. Alastruey J, Khir AW, Matthys KS, Segers P, Sherwin SJ, Verdonck PR. Pulse wave propagation in a model human arterial network: assessment of 1-D visco-elastic simulations against in vitro measurements. J Biomech. 2011;44(12):2250–8.
- 40.
Gałecki A, Burzykowski T. Linear Mixed-Effects Models Using R. Springer New York. 2013. doi: https://doi.org/10.1007/978-1-4614-3900-4
- 41. Akaike H. A new look at the statistical model identification. IEEE Trans Autom Control. 1974;19(6):716–23.
- 42.
James G, Witten D, Hastie T, Tibshirani R. An Introduction to Statistical Learning. Springer US. 2021. doi: https://doi.org/10.1007/978-1-0716-1418-1
- 43. Vallée F, Passouant O, Le Gall A, Joachim J, Mateo J, Mebazaa A, et al. Norepinephrine reduces arterial compliance less than phenylephrine when treating general anesthesia-induced arterial hypotension. Acta Anaesthesiol Scand. 2017;61(6):590–600. pmid:28543052
- 44. Pilz N, Heinz V, Ax T, Fesseler L, Patzak A, Bothe TL. Pulse Wave Velocity: Methodology, Clinical Applications, and Interplay with Heart Rate Variability. Rev Cardiovasc Med. 2024;25(7):266. pmid:39139426
- 45. Zhou X, Pan J, Wang Y, Wang H, Xu Z, Zhuo W. Left ventricular-arterial coupling as a predictor of stroke volume response to norepinephrine in septic shock - a prospective cohort study. BMC Anesthesiol. 2021;21(1):56. pmid:33596822
- 46. Tank AW, Lee Wong D. Peripheral and central effects of circulating catecholamines. Compr Physiol. 2015;5(1):1–15. pmid:25589262
- 47.
Procaccini DE, Sawyer JE, Watt KM. 19 - Pharmacology of Cardiovascular Drugs. In: Ungerleider RM, Meliones JN, Nelson McMillan K, Cooper DS, Jacobs JP. Critical Heart Disease in Infants and Children. Third ed. Philadelphia: Elsevier. 2019. 192-212.e6.
- 48.
DeMers D, Wachs D. Physiology, Mean Arterial Pressure. StatPearls. Treasure Island (FL): StatPearls Publishing. 2024.
- 49. Beloeil H, Mazoit JX, Benhamou D, Duranteau J. Norepinephrine kinetics and dynamics in septic shock and trauma patients. Br J Anaesth. 2005;95(6):782–8.
- 50. Kinoshita K. Traumatic brain injury: pathophysiology for neurocritical care. J Intensive Care. 2016;4:29. pmid:27123305
- 51. Rinehart J, Ma M, Calderon MD, Cannesson M. Feasibility of automated titration of vasopressor infusions using a novel closed-loop controller. J Clin Monit Comput. 2018;32(1):5–11.
- 52. Rinehart J, Joosten A, Ma M, Calderon MD, Cannesson M. Closed-loop vasopressor control: in-silico study of robustness against pharmacodynamic variability. J Clin Monit Comput. 2019;33(5):795–802.
- 53. Joosten A, Rinehart J, Cannesson M, Coeckelenbergh S, Pochard J, Vicaut E. Control of mean arterial pressure using a closed-loop system for norepinephrine infusion in severe brain injury patients: the COMAT randomized controlled trial. J Clin Monit Comput. 2024;38(1):25–30.
- 54. Sviri GE, Aaslid R, Douville CM, Moore A, Newell DW. Time course for autoregulation recovery following severe traumatic brain injury. J Neurosurg. 2009;111(4):695–700. pmid:19392589