Impact of occluder device configurations in in-silico left atrial hemodynamics for the analysis of device-related thrombus

Fig 7

Quantification of particle attachment to the left atrial appendage wall at the end of the cardiac cycle in the 20 patients cohort.

Blue represents the post-left atrial appendage occlusion (LAAO) configuration and, red represents the proposed configuration covering the pulmonary ridge. PR: pulmonary ridge. * denotes patients with device-related thrombus (DRT).

Fig 7

doi: https://doi.org/10.1371/journal.pcbi.1011546.g007