Abstract
Loosening of the artificial cup and inlay is the most common reasons for total hip replacement failures. Polyethylene wear and aseptic loosening are frequent reasons. Furthermore, over the past few decades, the population of patients receiving total hip replacements has become younger and more active. Hence, a higher level of activity may include an increased risk of implant loosening as a result of friction-induced wear. In this study, an instrumented hip implant was used to measure the contact forces and friction moments in vivo during walking. Subsequently, the three-dimensional coefficient of friction in vivo was calculated over the whole gait cycle. Measurements were collected from ten subjects at several time points between three and twelve months postoperative. No significant change in the average resultant contact force was observed between three and twelve months postoperative. In contrast, a significant decrease of up to 47% was observed in the friction moment. The coefficient of friction also decreased over postoperative time on average. These changes may be caused by ‘running-in’ effects of the gliding components or by the improved lubricating properties of the synovia. Because the walking velocity and contact forces were found to be nearly constant during the observed period, the decrease in friction moment suggests an increase in fluid viscosity. The peak values of the contact force individually varied by 32%-44%. The friction moment individually differed much more, by 110%-129% at three and up to 451% at twelve months postoperative. The maximum coefficient of friction showed the highest individual variability, about 100% at three and up to 914% at twelve months after surgery. These individual variations in the friction parameters were most likely due to different ‘running-in’ effects that were influenced by the individual activity levels and synovia properties.
Citation: Damm P, Bender A, Bergmann G (2015) Postoperative Changes in In Vivo Measured Friction in Total Hip Joint Prosthesis during Walking. PLoS ONE 10(3): e0120438. https://doi.org/10.1371/journal.pone.0120438
Academic Editor: Amitava Mukherjee, VIT University, INDIA
Received: August 25, 2014; Accepted: January 22, 2015; Published: March 25, 2015
Copyright: © 2015 Damm et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited
Data Availability: Data are available from the Orthoload database: http://www.orthoload.com/database/. To view the data, input the following search terms: Implant: Hip Joint III; Activity: Walking; Level Walking; Parameter: none; Patient: all. Relevant video and numerical data files are named as follows: H1L_130710_1_110; H1L_151110_1_51; H2R_151110_1_184; H2R_160211_2_67; H2R_14_03_11_1_186; H3L_040211_1_83; H3L_140311_1_126; H3L_260811_1_87; H4L_250311_2_12; H4L_290411_1_149; H4L_240611_1_170; H4L_111111_2_60; H5L_040711_2_54; H5L_020911_2_8; H5L_060212_2_53; H6R_220212_1_42; H6R_200412_1_126; H6R_280912_1_89; H7R_210212_1_36; H7R_170412_1_109; H7R_130712_1_128; H8L_260712_1_57; H8L_130912_1_153; H9L_191212_1_38; H9L_180213_1_89; H10R_260413_1_81; H10R_260613_1_158; H10R_250913_1_130. Additional relevant data may be found using the following search terms: Implant: Hip Joint III; Activity: Gaitanalysis; Walking; Level Walking; Parameter: none; Patient: all. Relevant files are named as follows: H1L_060511_1_50; H2R_150811_2_98; H3L_141111_1_28; H4L_270112_2_80; H5L_050412_1_75; H6R_201112_2_107; H7R_191112_1_47; H8L_270213_2_55; H8L_240413_1_42; H9L_301013_1_62; H10R_300114_1_82.
Funding: Support was provided by the German Research Society (SFB760 and BE 804/19-1) [http://www.dfg.de] and Deutsche Arthrose-Hilfe e.V. [http://www.arthrose.de]. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
Competing interests: The authors have declared that no competing interests exist.
Introduction
Loosening of the artificial cup and inlay is the most common reason for the failure of total hip replacements [1–3]. Polyethylene (PE) wear and aseptic loosening are frequent reasons for revisions of total hip joint prostheses (THP), account for 26% and 48% of all revisions, respectively [1,4]. One study has indicated that 30% to 40% of all THP revisions require a change of cup or inlay [5], thus making friction in the joint one of the main parameters affecting the life span of total hip prostheses. THP patients have become younger and more active over the past few decades [6,7]. This higher level of activity has resulted in increased wear rates and, consequently, requires earlier revision of THP [8,9].
To determine the parameters of friction in THP, several in vitro studies under different test conditions have been published, using new or explanted prostheses and different head diameters and lubricants [10–21]. In most studies, the coefficient of friction (μ) was determined by a simple pendulum test, which showed that the friction in the THP was primarily influenced by the material of the sliding partners and the lubrication regime.
Previous studies with instrumented implants measured the contact forces in THP in vivo [22–26]. However, in these studies, it was not possible to determine the friction within the implant, with the exception of one study that reported the friction-induced temperature increase during walking [27]. In that publication, it was postulated that the temperature in the THP was mainly caused by the friction parameters between the sliding partners and the lubricating properties of the synovia. Peak temperatures up to 43°C were measured after one hour of walking. Because bone tissues may already be damaged at temperatures exceeding 43°C [28], it can be assumed that repeated high implant temperatures are a risk factor for the long-term stability of the THP.
Using a newly developed instrumented hip implant [29], we were able to measure the joint friction in vivo for the first time. Initial results were reported for three months postoperatively (pOP) [30], and it was shown that the in vivo coefficient of friction is similar to the in vitro value obtained during the extension phase of the hip joint. However, during the flexion phase, μ strongly differs from the in vitro data. The maximum in vivo values of μ, calculated from the contact forces and friction moments, suggest a ‘dry’ friction during the hip flexion phase of walking.
The aim of this study was to measure the changes of contact forces, friction moments and coefficients of friction between three and twelve months pOP. We hypothesized that μ would decline due to ‘running-in’ effects and that both, friction moments and μ, would show great individual variability, depending on the synovial properties.
Methods
Variables
Fres Resultant hip joint contact force
Fres Vector of Fres
Mres Resultant friction moment
Mres Vector of Mres
FFriction Friction force
FFriction Vector of FFriction
R Radius of implant head (16 mm)
R Vector from origin of coordinates to contact point P
D Vector from origin of coordinates to vector H
H Vector vertical to Mres, pointing to P
μ Coefficient of friction
HS Ipsilateral heel strike
CTO Contralateral toe off
CHS Contralateral heel strike
TO Ipsilateral toe off
XPE Cross-linked polyethylene
Measurement equipment
Joint forces and friction moments were measured in vivo using instrumented hip implants. The prosthesis (CTW, Merete Medical, Berlin, Germany) is based on a clinically successful implant with a titanium stem, a 32 mm Al2O3 ceramic head and a XPE inlay. A telemetry circuit, six-strain gauges and an induction coil are arranged in the hollow neck. The implant is powered inductively by a coil around the hip joint. The strain gauge signals are transferred to the external receiver at radio frequency via an antenna inside of the implant head [31]. The telemetric load signals and the patient movements are recorded simultaneously on video. Detailed descriptions of the implant [29] and the external measurement system [31,32] have previously been published.
Using the six strain gauge signals, the three force and three moment components acting on the implant head are calculated with an accuracy of 1–2%. The femur-based coordinate system [33] is located in the head center of a right-sided implant; data from left-sided implants are mirrored. Positive forces Fx, Fy, and Fz act in the lateral, anterior, and superior directions; the measured friction moments (Mx, My and Mz) turn right around the positive coordinate axes. The resultant contact force Fres and the resultant friction moment Mres are calculated from their three components. Because the force component in direction of the femoral axis always acts downwards, Fres points toward the center of the implant head.
Patients and measurements
The study was approved by the ethical committee (EA2/057/09) and registered with the ‘German Clinical Trials Register’ (DRKS00000563). Ten patients (8m/2f) with hip osteoarthritis gave written informed consent to participate in the study and have their images published (Table 1).
Measurements were taken during level walking at several time points (Table 2) between three and thirteen months pOP. Selected trials of each investigated subject are also shown and can be downloaded at the public data base www.OrthoLoad.com.
Data evaluation
All forces and moments are reported as a percentage of patient bodyweight (%BW and %BWm, respectively). Average force- and moment-time patterns were calculated separately for each subject and measurement day using a ‘time warping’ method [34]. This averaging procedure was first performed on repeated trials of the single subjects. The obtained load patterns of all subjects were then averaged again, leading to data which are typical for an ‘average’ subject. All data were analyzed for the time points of three, six, nine and twelve months pOP (Table 2).
Unless stated otherwise, all presented data refers to the obtained ‘average’ subject. Because errors between the single trials were minimized over all loading cycles, the peak values of the average curves can slightly deviate from the averaged numerical values at the 1st and 2nd peak.
Changes of the measured peak values over the pOP time were analyzed for each subject separately and for the ‘average’ subject (Tables 3 and 4). The changes over time were analyzed individually using the Mann-Whitney-U test and across all subjects using the Wilcoxon test. Furthermore a regression analysis was performed, for each subject separately and for the average subject, to determine the correlation between the measured peak values of Fres, Mres and μ (y) and the pOP time in months (m), using the logarithmic relationship (1) The coefficient of correlation (R²) and the residual-standard-error (RSE) were calculated and statistically analyzed with a t-test. For all statistical calculations a type-I-error level was defined with α = 0.05.
Coefficient of friction
In the following equations, the underlined symbols are vectors; all others are scalar values. In the Coulomb approach, the friction force FFriction between the cup and ball is determined using the resultant contact force Fres and the coefficient of friction μ (Fig. 1): (2) The ball turns around the axis of Mres. H is the lever arm between FFriction and the axis of Mres. Because FFriction counteracts the moment Mres, we obtain (3) (2) and (3) deliver (4) R is the radius of the ball and points in direction of Fres: (5) H can be substituted by (6) with D being the orthogonal projection of R on Mres: (7) Applying (5) and (7), equation (6) becomes (8) with (9) because equation (5) holds true.
To obtain μ with an accuracy of 5%, only joint forces Fres ≥ 25%BW and moments Mres ≥ 0.02%BWm were included in the analysis.
In reality, the component Fz is measured negatively, and the direction of Fres is thus towards the ball surface
Gait velocity
The gait velocity was determined individually at three months pOP using the video clips, as described elsewhere [30]. At twelve months pOP, the velocity was determined individually using simultaneously measured 3D gait data (Vicon Nexus, Vicon Motion Systems Ltd., UK). The distance travelled during one walking trial was measured between the first and last HS of the ipsilateral leg for each walking trial separately. It was divided by the elapsed time and the obtained speed was subsequently averaged over all trials.
Results
During gait, the resultant joint force Fres had two extreme values (Fig. 2), labeled Fres1 and Fres2. At the instants of these peak forces the friction moments Mres1 and Mres2 were determined. The coefficient of friction μ was calculated throughout the whole gait cycle and the values μ1 and μ2 at the instants of Fres1 and Fres2 plus the absolute maximum value μmax were analyzed.
Average time courses of Fres, Mres and μ during one gait cycle at three and twelve months pOP. The indicated values Mres1, Mres2, μ1, μ2 and μmax were determined at the instants of the peak forces Fres1 and Fres2.
Joint contact force
The resultant joint force Fres was calculated using the measured force components in lateral, anterior, and superior directions. In Fig. 2, the patterns of Fres (bold lines) of the ‘average’ subject are shown for two different time points, three and twelve months pOP. Both patterns are similar and show two typical peak values. The first maximum (Fres1) occurs at the time of contralateral toe off (CTO), and the second maximum (Fres2) at the contralateral heel strike (CHS).
At three months pOP, peak values of 255%BW (Fres1) and 238%BW (Fres2) were determined for the ‘average’ subject (Figs. 2 and 3, Table 3). The values of Fres1 individually ranged between 209%BW (H3L) and 301%BW (H8L) (Fig. 4), which is a variation by 44% Fres2 laid between 217%BW (H9L) and 287 (H8L); it had a variation of 32%.
Load pattern during one gait cycle at three (left) and twelve (right) months postoperatively.
At twelve months pOP (Figs. 2 and 3, Table 3), peak values of 252%BW (Fres1) and 245%BW (Fres2) were determined for the ‘average’ subject. Fres1 individually varied (Fig. 4) between 220%BW (H3L) and 294%BW (H7R) by 34%. Fres2 ranged between 215%BW (H9L) and 283%BW (H8L) with a variation of 32%.
Over the three- to twelve-month pOP period (Figs. 2, 3 and 4), Fres1 declined by -1% on average (Table 4). It increased in four subjects, but decreased in five subjects and remained constant in one (H7R). The individual changes laid between +6% (H1L) and -13% (H8L). The change of Fres2 during the pOP time was +4%. Fres2 increased in seven and decreased in three subjects (Table 4). The individual changes laid between -8% (H10R) and +13% (H6R). For the ‘average’ subject a logarithmic correlation was determined between the pOP month and the peak values Fres1 and Fres2 (Fig. 3) with R² = 0.03 and R² = 0.68 (Table 5).
Friction moment
During walking, Mres increased in all patients between heel strike (HS) and CHS (Fig. 2, dashed lines), with a first maximum around CHS. The local minimum of the Mres curve between CHS and TO represents the instant when the rotational direction of the joint changes from extension to flexion. It is followed by a second, smaller maximum during the flexion phase, shortly before TO.
At three months pOP, friction moments of 0.174%BWm (Mres1) and 0.234%BWm (Mres2) were determined for the ‘average’ subject (Figs. 2 and 3, Table 3). After the local minimum, a smaller second maximum of 0.186%BWm was determined during the flexion phase of the gait cycle. Mres1 individually ranged between 0.096%BWm (H9L) and 0.220%BWm (H7R), a variation by 129%. Mres2 laid between 0.137%BWm (H9L) and 0.301%BWm (H10R) (Fig. 4) and varied by 119%. The smaller second maximum ranged from 0.124%BWm (H5L) to 0.261%BWm (H2R) with a variation of 110%.
At twelve months pOP (Figs. 2 and 3, Table 3), average friction moments of 0.107%BWm (Mres1) and 0.174%BWm (Mres2) were found. Mres1 ranged between 0.027%BWm (H2R) and 0.143%BWm (H1L); it highly varied by 430% (Fig. 4). Mres2 laid between 0.069%BWm (H9L) and 0.402%BWm (H10R), which is an even higher variation of 451%. The second smaller maximum had an average of 0.130%BWm. It ranged from 0.045%BWm (H4L) to 0.247%BWm (H10R) and varied by 449%.
From three to twelve months pOP, Mres1 decreased significant on average by -47% (Table 4). But this change was individually very different, with values between +13% (H10R) and -72% (H2R). Mres2 sunk on average by -34%, with individual changes between +30% (H10R) and -60% (H8L). The smaller second maximum of Mres was decreased by -28%. With exception of H10R, Mres1 and Mres2 decreased in all subjects over the first twelve months pOP.
A logarithmic correlation was observed between the pOP month and the peak values Mres1 and Mres2 (Fig. 3) with R² = 0.994 and R² = 0.917 for the ‘average’ subject (Table 5). The slope of Mres from HS to CTO decreased by approximately -50% during the pOP time.
Coefficient of friction
The coefficient of friction (μ) increased throughout the whole load phase from HS to TO (Fig. 2, dotted lines). This increase was approximately linear during the stance phase between CTO and CHS. Shortly after the CHS, at approximately 70% of the gait cycle, μ increased sharply and always reached its absolute maximum (μmax) shortly before TO.
At three months pOP (Fig. 2, Table 3), values of μ1 = 0.045, μ2 = 0.062 and μmax = 0.175 were determined for the ‘average’ subject. As with Fres and Mres, the coefficient of friction individually varied much (Fig. 4). μ1 ranged from 0.032 (H8L) to 0.063 (H1L), and varied by 97%. μ2 laid between 0.04 (H9L) and 0.081 (H10R), which is a variation by 103%. μmax had values between 0.095 (H8L) and 0.229 (H2R); it varied by 141%.
At twelve months pOP (Fig. 2, Table 3), values of μ1 = 0.028, μ2 = 0.047 and μmax = 0.179 were determined for the ‘average’ subject (Fig. 2). μ1 varied extremely between 0.007 (H2R) to 0.071 (H10R), which is a variation of 914%. μ2 was individually less different; it had values between 0.02 (H9L) to 0.116 (H10R) and thus varied by 480%. μmax ranged from 0.079 (H8L) to 0.442 (H10R) with a variation of 459%.
From three to twelve months pOP (Figs. 2, 3 and 4), the coefficient of friction decreased on average (Table 4). The reduction of μ1 was -34%, with individual changes between +18% (H10R) and -72% (H2R). μ2 sunk by -46% on average, with individual changes between +48% (H10R) and -59% (H8L). The values of μ1 and μ2 decreased in all subjects, with the exception of H10R. μmax decreased by -8% on average; it rose in seven subjects but sunk in three of them. The individual changes laid between +101% (H10R) and -42% (H9L).
A logarithmic correlation between the pOP month and μ was found for the ‘average’ subject, with R² = 0.992 (μ1), R² = 0.993 (μ2) and R² = 0.251 (μmax) (Table 5). Furthermore, the correlation was also calculated for each subject separately. The coefficients of correlation for μ1 ranged between 0.999 (H2R) and 0.226 (H10R), those for μ2 between 0.999 (H6R) and 0.745 (H8L) and for μmax between 0.954 (H9L) and 0.007 (H1L), respectively.
Discussion
In this study, it was possible for the first time to simultaneously measure joint contact forces and friction moments in vivo during walking. High friction moments may endanger the stability of cup fixation and cause fretting between the prosthesis head and the taper. Loosening moments of acetabular cups depend on the quality of the cup-bone interface, which is influenced, among others, by the fixation technique and the type of coating. For cementless press-fit cups, cyclic loosening moments as low as 8Nm were determine in vitro [35]. The maximal measured friction moment (intra-individual average!) was 0.402%BWm in H10R (Fig. 4), which corresponds to 3.99Nm or 50% of the critical value. This shows that the real friction moments, which act during higher demanding activities than walking or when extreme force occur during stumbling [36], may reach dangerous values, especially shortly after implantation.
Joint loads and gliding parameters
Simulator tests on friction and wear in hip implants have been performed under varying conditions. In some cases, the joint was moved in one plane only, using sine wave loads and movements, while the movements around the other joint axes were ignored [11,18,21,37]. However, during walking, the hip joint moves not only around the x-axis (flexion/extension) but also around the y-axis (abduction/adduction) and the z-axis (internal/external rotation). An earlier study [30] showed that friction also depends on the moments around the other two axes. Furthermore, the joint movement is not sinusoidal, and the contact loads typically have two peaks at CTO and CHS. We have shown that the friction moment Mres increases in vivo between HS and CHS, when it reaches its maximum. This finding is contrary to the in vitro situation, where the measured friction moment showed a plateau phase during the loading of the joint [11,21].
In the current in vivo study, μ has been calculated in 3D and has been proven not to be constant throughout the gait cycle (Fig. 2). During each step it increased (Table 3, 4) from 0.045 to 0.175 (three months pOP) respectively from 0.028 to 0.179 (twelve months pOP). μ reaches its absolute maximum shortly before TO, when the movement changes from extension to flexion. At that instant, the relative velocity between ball and cup is either zero or very low. Earlier studies reported that μ differed when the lubrication conditions were changed [19,20,38]. The high peak value of μ is probably best explained by the assumption that the synovia is squeezed out under high loads and transported back into the joint when the load is low during the swing phase [21].
Postoperative changes
No significant change of the resultant contact force Fres in the ‘average’ subject was observed between three and twelve months pOP. However, the joint loads during the first three months are not reported here. In an earlier study, Fres showed during that time an average increase of 18% (Fres1) and 21% (Fres2) when walking with crutches [39]. This data was only based on subjects H1L to H7R, who also participated in this study. Future simulations of bone remodeling around hip implants by finite element modeling should simulate increasing contact forces during the first three months while maintaining constant contact force values afterwards.
A significant decrease of the friction moment Mres in the postoperative course was observed throughout the whole gait cycle for the ‘average’ subject. At the instants of the two peaks of Fres, Mres decreased significantly by 47% (Mres1) and 34% (Mres2). These changes may be caused by ‘running-in’ effects of the gliding components or by improved lubricating properties of the synovia [40]. A fluid with a high viscosity needs more time to flow out of the intra-articular joint space, provided that the contact force and sliding speed are the same. Synovia is a non-Newton fluid [40], and its viscosity therefore depends on contact force, sliding velocity and shear stress [13,19,40,41]. Because the walking velocity and contact forces were found to be nearly constant during the observed period, the decrease in Mres suggests an increase in fluid viscosity by about 50%, an effect which had not be investigated in the current study. The logarithmic decrease of the coefficient of friction over the postoperative time suggests a change in synovia viscosity and, possibly, an additional smoothening of the gliding surfaces.
Typical simulator studies, which investigated the friction of soft-on-hard parings as in our study (XPE/Al2O3), delivered μ-values between 0.04 and 0.09 [11,13,42] for a lubricated regime and for a ‘dry’ sliding condition values between 0.13 and 0.14 were determined [43]. The sharp increase of μ between CHS and TO did not change significantly between three and twelve months pOP. This finding suggests that during early flexion, hip joint friction does not depend much on synovial properties. This finding, along with the high absolute peak values μmax, indicates the presence of ‘dry’ friction [43] in the joint during the early flexion phase. However, several studies have shown that the fluid film thickness in a soft-on-hard pairing is much lower than the roughness of the polyethylene bearing surface [44,45]. This suggests that ‘mixed’ or ‘boundary lubrication’ conditions occurring for this tribological pairing [42].
A time-dependent logarithmic trend of μmax was also observed, but only with R² = 0.251 and a high inter-individual range (0.013 to 0.886). The significance of postoperative changes of μmax in only some of the subjects may be due to different activity levels and therefore different running in effects [46].
Inter-individual differences
The observed peak values of Fres varied individually by 44% (three months pOP) and 34% (twelve months pOP). In contrast to these relatively small deviations, Mres individually differed by 129% (three months pOP) and even 451% (twelve months pOP). For μ, the individual variation laid between 141% (three months pOP) and 914% (twelve months pOP). These strong individual variations of Mres are most likely due to different ‘running-in’ effects, which are caused by the individual activity levels and synovial properties.
In subject H10R, the peak μmax was extremely high and was greater than that observed in all of other subjects. This observation may possibly be explained if one assumes that in H10R the movement in the joint really fell to zero at the instant when the movement changed from extension to flexion. In the other subjects, however, the joint may still have been slightly rotating in the frontal and horizontal planes. The fluid film may thus have been broken down completely in the joint of H10R only. To clarify this hypothesis, future study will focus on individually different movements in the joint and on deviating areas of load transfer, which will be analyzed using individual gait data and anatomical conditions.
It was shown that the peak values of the in vivo acting friction moments during walking were smaller than critical torsion torque at the cup-bone interface [35]. However, higher friction moments can be expected during high demanding activities, which might be critical for the cup fixation, especially shortly after implantation.
This study has reported for the first time post operative changes and the individual differences of the in vivo forces and friction moments acting in total hip joint replacements. These in vivo loads can now be used as realistic input data for friction and wear simulator studies.
Examples of the in vivo measurements are published at the public data base www.orthoload.com.
Limitations of the study
There are some limitations to this study. The number of investigated subjects was small and only one implant type was investigated only during level walking. The peak values of the friction moment and friction coefficient and their time-dependent changes individually varied extremely. Possible factors, causing these variations, were not evaluated. Such factors as implant orientation, sliding speed and contact area in the joint, as well as friction during other activities are currently investigated in an additional study.
Author Contributions
Conceived and designed the experiments: PD GB. Performed the experiments: PD. Analyzed the data: PD AB. Contributed reagents/materials/analysis tools: PD AB. Wrote the paper: PD GB.
References
- 1.
Bergen H. Annual Report 2010—Norwegian Arthroplasty Register. 2010.
- 2.
David D, Graves S, Tomkins A. Australien Orthopaedic Association National Joint Replacement Registry. Anual Report. Adelaide:AOA; 2013. 2013.
- 3.
Garellick G, Karrholm J, Rogmark C, Rolfson O, Herberts P. Swedish Hip Arthroplasty Register—Annual Report 2011. 2011.
- 4.
CJRR. CJRR report: Total hip and total knee replacements in Canada. Canadian Institute for Health Information. 2008.
- 5. Havelin LI, Fenstad AM, Salomonsson R, Mehnert F, Furnes O, Overgaard S, et al. The Nordic Arthroplasty Register Association: a unique collaboration between 3 national hip arthroplasty registries with 280,201 THRs. Acta Orthop. 2009;80: 393–401. pmid:19513887
- 6. Huch K, Müller KAC, Stürmer T, Brenner H, Puhl W, Günther K- P. Sports activities 5 years after total knee or hip arthroplasty: the Ulm Osteoarthritis Study. Ann Rheum Dis. 2005;64: 1715–1720. pmid:15843453
- 7. Chatterji U, Ashworth MJ, Lewis PL, Dobson PJ. Effect of total hip arthroplasty on recreational and sporting activity. ANZ J Surg. 2004;74: 446–449. pmid:15191479
- 8. Flugsrud GB, Nordsletten L, Espehaug B, Havelin LI, Meyer HE. The effect of middle-age body weight and physical activity on the risk of early revision hip arthroplasty. Acta Orthop. 2007;78: 99–107. pmid:17453400
- 9. Malchau H, Herberts P, Söderman P, Odén A. Prognosis of Total Hip Replacement. swedish Natl hip Arthroplast Regist. 2000;
- 10. Liao Y. The effect of frictional heating and forced cooling on the serum lubricant and wear of UHMW polyethylene cups against cobalt—chromium and zirconia balls. Biomaterials. 2003;24: 3047–3059. pmid:12895577
- 11. Brockett C, Williams S, Jin Z, Isaac G, Fisher J. Friction of Total Hip Replacements With Different Bearings and Loading Conditions. J Biomed Mater Res. 2006; 508–515.
- 12. Affatato S, Spinelli M, Zavalloni M, Mazzega-Fabbro C, Viceconti M. Tribology and total hip joint replacement: current concepts in mechanical simulation. Med Eng Phys. 2008;30: 1305–17. pmid:18774742
- 13. Hall RM, Unsworth A. Review Friction in hip prostheses. Biomaterials. 1997;18: 1017–1026. pmid:9239463
- 14. Scholes SC, Unsworth A. The effects of proteins on the friction and lubrication of artificial joints. Proc Inst Mech Eng Part H J Eng Med. 2006;220: 687–693.
- 15. Scholes SC, Unsworth A, Hall RM, Scott R. The effects of material combination and lubricant on the friction of total hip prostheses. Wear. 2000;241: 209–213.
- 16. Fialho JC, Fernandes PR, Eça L, Folgado J. Computational hip joint simulator for wear and heat generation. J Biomech. 2007;40: 2358–2366. pmid:17270192
- 17. Mattei L, Di Puccio F, Piccigallo B, Ciulli E. Lubrication and wear modelling of artificial hip joints: A review. Tribol Int. Elsevier; 2011;44: 532–549.
- 18. Saikko V. A simulator study of friction in total replacement hip joints. Proc Inst Mech Eng Part H J Eng Med. 1992;206: 201–211. pmid:1300112
- 19. Scholes SC, Unsworth A. Comparison of friction and lubrication of different hip prostheses. Proc Inst Mech Eng Part H J Eng Med. 2000;214: 49–57.
- 20. Unsworth A. The effects of lubrication in hip joint prostheses. Phys Med Biol. 1978;23: 253–68. pmid:643921
- 21. Williams S, Jalali-Vahid D, Brockett C, Jin Z, Stone MH, Ingham E, et al. Effect of swing phase load on metal-on-metal hip lubrication, friction and wear. J Biomech. 2006;39: 2274–81. pmid:16143337
- 22. Bergmann G, Rohlmann , Graichen F. In vivo Messung der Hüftgelenkbelastung 1. Teil: Krankengymnastik. Z Orthop. 1989;127: 672–679. pmid:2618148
- 23. Stansfield B, Nicol AC, Paul JP, Kelly IG, Graichen F, Bergmann G. Direct comparison of calculated hip joint contact forces with those measured using instrumented implants. An evaluation of a three-dimensional mathematical model of the lower limb. J Biomech. 2003;36: 929–936. pmid:12757801
- 24. Brand RA, Crowninshield RD, Pedersen DR. Forces on the femoral head during acitivties of daily living. Iova Orthop J. 1982;2: 43–49.
- 25. Paul JP. Force actions transmitted by joints in the human body. Proc R Soc Lond B Biol Sci. 1976;192: 163–72. pmid:3785
- 26. Taylor SJ, Perry JS, Meswania JM, Donaldson N, Walker PS, Cannon SR. Telemetry of forces from proximal femoral replacements and relevance to fixation. J Biomech. 1997;30: 225–34. pmid:9119821
- 27. Bergmann G, Graichen F, Rohlmann A, Verdonschot N, van Lenthe GH. Frictional heating of total hip implants. Part 1: measurements in patients. J Biomech. 2001;34: 421–8. pmid:11266664
- 28. Li S, Chien S, Branemark P- I. Heat shock-induced necrosis and apoptosis in osteoblasts. J Orthop Res. 1999;17: 891–9. pmid:10632456
- 29. Damm P, Graichen F, Rohlmann A, Bender A, Bergmann G. Total hip joint prosthesis for in vivo measurement of forces and moments. Med Eng Phys. 2010;32: 95–100. pmid:19889565
- 30. Damm P, Dymke J, Ackermann R, Bender A, Graichen F, Halder A, et al. Friction in total hip joint prosthesis measured in vivo during walking. PLoS One. 2013;8: e78373. pmid:24260114
- 31. Graichen F, Bergmann G, Rohlmann A. Telemetric transmission system for in vivo measurement of the stress load of an internal spinal fixator. Biomed Tech Biomed Eng. 1994;39: 251–258. pmid:7811908
- 32. Graichen F, Arnold R, Rohlmann A, Bergmann G. Implantable 9-channel telemetry system for in vivo load measurements with orthopedic implants. IEEE Trans Biomed Eng. 2007;54: 253–61. pmid:17278582
- 33. Wu G, Siegler S, Allard P, Kirtley C, Leardini A, Rosenbaum D, et al. ISB recommendation on definitions of joint coordinate system of various joints for the reporting of human motion—part I: ankle, hip, and spine. J Biomech. 2002;35: 543–548. pmid:11934426
- 34. Bender A, Bergmann G. Determination of Typical Patterns from Strongly Varying Signals. Comput Methods Biomech Biomed Engin. 2011;iFirst: 1–9. pmid:21905293
- 35. Curtis MJ, Jinnah RH, Valerie WD, Hungerford DS. The initial stability of uncemented acetabular components. Br Editor Sov Bone Jt Surg. 1992;74-B: 372–376.
- 36. Bergmann G, Graichen F, Rohlmann . Hip joint contact forces during stumbling. Langenbeck’s Arch Surg. 2004;389: 53–59.
- 37. Scholes SC, Unsworth A, Goldsmith AAJ. A frictional study of total hip joint replacements. Phys Med Biol. 2000;45: 3721–35. pmid:11131195
- 38. Hall RM, Unsworth A, Wroblewski BM, Siney P, Powell NJ. The friction of explanted hip prostheses. Br J Rheumatol. 1997;36: 20–6. pmid:9117168
- 39. Damm P, Schwachmeyer V, Dymke J, Bender A, Bergmann G. In vivo hip joint loads during three methods of walking with forearm crutches. Clin Biomech. Elsevier Ltd; 2013;28: 530–5. pmid:23643290
- 40. Fam H, Bryant JT, Kontopoulou M. Rheological properties of synovial fluids. Biorheology. 2007;44: 59–74. Available: http://www.ncbi.nlm.nih.gov/pubmed/17538199 pmid:17538199
- 41. Balazs EA. The Physical Properties of synovial Fluid and the special role of hyaluronic Acid. Disord Knee. 1974; 63–75.
- 42. Jin Z, Stone M, Ingham E, Fisher J. (v) Biotribology. Curr Orthop. 2006;20: 32–40.
- 43. Xiong D, Ge S. Friction and wear properties of UHMWPE/Al2O3 ceramic under different lubricating conditions. Wear. 2001;250: 242–245.
- 44. Jalali-Vahid D, Jagatia M, Jin ZM, Dowson D. Prediction of lubricating film thickness in UHMWPE hip joint replacements. J Biomech. 2001;34: 261–6. pmid:11165292
- 45. Fisher J, Dowson D. Tribology of total artificial joints. Proc Inst Mech Eng H. 1991;205: 73–9. Available: http://www.ncbi.nlm.nih.gov/pubmed/1840723 pmid:1840723
- 46. Unsworth A. Tribology of artifical hip joints. Proc Inst Mech Eng Part J J Eng Tribol. 2006;220: 711–718.