Abstract
Background
Despite the development of several methods for diagnosing COVID-19, long-term validation of such methods remains limited. In the early phase of the COVID-19 pandemic, we developed a rapid and sensitive diagnostic method based on reverse transcription loop-mediated isothermal amplification (RT-LAMP) methodology, which is suitable for point-of-care application or for use in resource-limited settings to detect SARS-CoV-2. To assess the applicability of the RT-LAMP assay technique to resource-limited regions, such as rural areas in Africa, and to verify the usability of the method against various SARS-CoV-2 variants, the method was validated using clinical samples collected longitudinally during the pandemic.
Methodology/Principal findings
First, the sensitivity of the RT-LAMP assay for detecting 10 SARS-CoV-2 variants was evaluated using viral RNA samples extracted from cell culture with a portable battery-supported device, resulting in the successful detection of 20–50 copies of the viral genome within 15 min, regardless of the variant. COVID-19 positive samples collected in Gabon between March 2020 and October 2021 were used to evaluate the sensitivity of the assay and to calculate the copy number of the SARS-CoV-2 genome. More than 292 copies of the viral genome were detected with 100% probability within 15 min in almost all tests.
Conclusions
This long-term validation study clearly demonstrated the applicability of the RT-LAMP assay for the clinical diagnosis of COVID-19 in resource-limited settings of Africa, such as rural areas in Gabon. The results show the potential of the assay as a promising COVID-19 diagnostic method, especially in rural and remote regions located far from the official diagnosis facilities in urban or semi-urban areas.
Author summary
RT-LAMP, a rapid and sensitive method for detecting pathogens, can be used in resource-limited settings. Although several RT-LAMP assays have been developed for the diagnosis of infectious diseases, reports on the long-term validation of these assays are limited. To demonstrate the applicability of the assay and to verify the usability of the method against various SARS-CoV-2 variants, it is important to conduct long-term validation studies using clinical samples collected in the region where the assay may be introduced for diagnoses. This study revealed the high sensitivity of the RT-LAMP assay for detecting viral RNAs of 10 SARS-CoV-2 variants, including the Omicron variant. Moreover, more than 292 copies of the viral genome were detected with 100% probability within 15 min using clinical samples collected between March 2020 and October 2021 in Gabon. The results demonstrate the suitability of the assay for use in rural areas of Africa where it is difficult for residents to access official COVID-19 diagnosis facilities.
Citation: Abe H, Ushijima Y, Bikangui R, Ondo GN, Moure A, Yali-Assy-Oyamli Y, et al. (2022) Long-term validation of a reverse transcription loop-mediated isothermal amplification (RT-LAMP) assay for the rapid detection of SARS-CoV-2 from March 2020 to October 2021 in Central Africa, Gabon. PLoS Negl Trop Dis 16(12): e0010964. https://doi.org/10.1371/journal.pntd.0010964
Editor: Jonas Klingström, Karolinska Institutet, SWEDEN
Received: June 9, 2022; Accepted: November 19, 2022; Published: December 1, 2022
Copyright: © 2022 Abe et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Data Availability: All relevant data are within the manuscript and its Supporting Information files.
Funding: This study was supported by the Science and Technology Research Partnership for Sustainable Development (SATREPS) from Japan International Cooperation Agency (JICA) and Japan Agency for Medical Research and Development (AMED), Grant number SATREPS, JP20jm0110013; AMED, Grant number JP20wm0225003; a Grant-in-Aid for Scientific Research (KAKENHI) from the Japan Society for the Promotion of Science (JSPS), Grant number JP19K24679; and a RONPAKU (Dissertation Ph.D.) program from JSPS, Grant number R12110. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
Competing interests: The authors have declared that no competing interests exist.
Introduction
During the past two years since the novel coronavirus disease (COVID-19) was declared a pandemic by the World Health Organization (WHO), each country has developed and implemented appropriate countermeasures suitable for controlling the spread of COVID-19. Despite the worldwide administration of vaccines, repeated waves of COVID-19 infection have been reported globally owing to the frequent emergence of novel variants of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2), which show higher infectivity and/or an ability to escape from the effects of neutralizing antibodies [1,2].
Several techniques have been developed for detecting SARS-CoV-2, such as nucleic acid amplification tests (NAATs) and lateral flow devices (LFDs). NAATs remain common methods of COVID-19 diagnosis owing to their high sensitivity and specificity [3–5]. Several reverse transcription-quantitative PCR (RT-qPCR) methods developed by research institutes worldwide have been considered gold standards [6,7]; however, RT-qPCR requires a detection instrument and stable power. Hence, the utilization of RT-qPCR in resource-limited settings or for on-site diagnosis may be difficult. LFDs are also routinely used for COVID-19 diagnosis worldwide due to their excellent usability and portability, although the lower sensitivity of LFDs than NAATs remains a potential issue [8,9]. Reverse transcription loop-mediated isothermal amplification (RT-LAMP) is a rapid and sensitive NAAT that involves a simple isothermal step and can be performed with a portable battery-driven device. It shows comparable sensitivity to RT-qPCR [10], indicating that RT-LAMP is suitable for COVID-19 molecular diagnosis, which can be completed in 15–20 min, even in settings where RT-qPCR implementation would be difficult.
COVID-19 diagnostic systems have been established in Africa, and most countries have started free vaccination programs. For gaining an appropriate understanding of COVID-19, genetic surveillance of SARS-CoV-2 is being actively conducted, with countries such as South Africa, Nigeria, and Senegal leading in surveillance. This extensive surveillance has resulted in the identification of a novel variant, Omicron, which shows high infectivity and the ability to evade the immune system [11–13]. Nonetheless, the overall pattern of spread and infectivity of COVID-19 remains unclear in Africa, especially in remote areas, possibly due to the difficulty faced by residents in accessing diagnosis facilities, which are typically set in urban or semi-urban areas. In Gabon, Central Africa, COVID-19 cases have been well-monitored since the early phase of the pandemic, and the genetic surveillance of SARS-CoV-2 is being actively conducted with continuous efforts of the government [14–17]. However, COVID-19 is diagnosed using RT-qPCR in urban or semi-urban settings in the country. Upon observation of COVID-19 symptoms in residents, diagnostic tests are difficult to perform in rural areas.
We have developed a variety of RT-LAMP assays to detect viruses that pose a concern to public health, such as Ebola and Zika viruses, for field surveillance [18–20]. Taking advantage of our experience, we developed a rapid and sensitive RT-LAMP assay to detect SARS-CoV-2, which was approved as an official diagnostic method by the government of Japan in March 2020 [21]. The RT-LAMP assay targeted the Orf1ab region and detected more than 23.7 copies of the SARS-CoV-2 genome with a 100% probability within 15 min, showing comparable sensitivity to RT-qPCR [21]. Although several other reports have been published examining the development of RT-LAMP assays against SARS-CoV-2, most assays have been evaluated using only samples derived from patients infected with the original Wuhan-type or D614G variants [22–25]. Hence, it remains unknown whether these RT-LAMP assays demonstrate similar sensitivity and specificity for the detection of other variants. Nevertheless, reports validating novel diagnostic methods are limited compared to those examining the development of these methods.
To examine the continuous usability of the RT-LAMP assay against various SARS-CoV-2 variants that have emerged since the beginning of the pandemic, we conducted a long-term validation study using clinical samples collected in Gabon, considering the future introduction of the assay to resource-limited areas. Whole-genome sequencing of SARS-CoV-2 detected in clinical samples was also attempted to accurately determine the variant, which might affect the sensitivity of the assay. As a result, the RT-LAMP assay showed high sensitivity for detecting all SARS-CoV-2 variants tested in this study. Our results demonstrate the suitability of the diagnostic assay in rural and remote areas and may help provide a precise understanding of the spread and infectivity of COVID-19 in Africa.
Methods
Ethics statement
This study was approved by the Institutional Ethical Committee of Centre de Recherches Médicales de Lambaréné (CERMEL) and Institute of Tropical Medicine at Nagasaki University (approval numbers CEI-007 and 170921177, respectively). Informed consent was not required as the study used anonymous samples with the total confidentiality of each sample.
Viral RNA extracted from clinical samples and synthesis of standard RNA
Viral RNA was routinely extracted from nasopharyngeal swab samples using the QIAamp Viral RNA Mini Kit (Qiagen, Hilden, Germany) during routine COVID-19 diagnosis process at CERMEL. Viral RNAs extracted from COVID-19 positive samples were used in this study. To calculate the copy numbers of viral genomes, synthesized RNAs were prepared using the T7 RiboMAX Express Large Scale RNA Production System (Promega, Madison, WI, USA) with artificially synthesized DNA of the viral target sequence conjugated to the T7 promoter sequence, as described previously [21].
Number of COVID-19 infection cases in Gabon
The number of new cases of infection in Gabon was obtained from the WHO COVID-19 dashboard [1]. We defined the first, second, and third waves of COVID-19 in Gabon as follows: April 27–August 17, 2020; January 11–May 31, 2021; and September 6–November 22, 2021 [17].
Viral strains of SARS-CoV-2 variants
To obtain viral RNA, SARS-CoV-2 variants were propagated in VeroE6 cells, as previously described [26]. Culture supernatants were collected four days after infection, centrifuged at 2,000 × g for 15 min, and stored at −80°C until further use. The reference strain names used in this study are listed as follows: B.1.1 (D614G), hCoV-19/Japan/NGS-SC-1/2020 (GISAID accession number, EPI_ISL_481254); Alpha (B.1.1.7), hCoV-19/Japan/QK002/2020 (EPI_ISL_768526); Beta (B.1.351), hCoV-19/Japan/TY8-612-P1/2021 (EPI_ISL_1123289); Gamma (P.1), hCoV-19/Japan/TY7-503/2021 (EPI_ISL_877769); Delta (B.1.617.2), hCoV-19/Japan/TY11-927-P1/2021 (EPI_ISL_2158617); Lambda (C.37), hCoV-19/Japan/TY33-456/2021 (EPI_ISL_4204973); Mu (B.1.621.1), hCoV-19/Japan/TY26-717/2021 (EPI_ISL_4470503); Kappa (B.1.617.1), hCoV-19/Japan/TY11-330-P1/2021 (EPI_ISL_2158613); Omicron (BA.1), hCoV-19/Japan/TY38-873P0/2021 (EPI_ISL_7418017); Omicron (BA.2), hCoV-19/Japan/TY40-385/2022 (EPI_ISL_9595859). The virus strains were provided by the National Institute of Infectious Diseases, Japan, except for the hCoV-19/Japan/NGS-SC-1/2020 (B.1.1) strain, which was isolated from a clinical sample of a patient detected positive for COVID-19 at the Institute of Tropical Medicine, Nagasaki University. For the selection of BA.2 variant sequences to count mutations in the primer binding region, “complete” and “low coverage excluded” sequences submitted from Europe, Americas, Asia, Africa, and Oceania in 2022 were downloaded from GISAID (accessed on April 5, 2022). The data corresponding to 5,924 strains were obtained, and the sequences were aligned using MAFFT (https://mafft.cbrc.jp/alignment/server/). For the BA.3 variant, all “complete” and “low coverage excluded” sequences were downloaded from GISAID (accessed on April 5, 2022) containing data corresponding to 1,751 strains. After eliminating sequences outside the primer binding sites, SARS-CoV-2 strains that showed ambiguous bases or Ns in the primer binding region were removed manually.
RT-LAMP assay
RT-LAMP assay was performed with the isothermal master mix reagent (ISO-004; Canon Medical Systems, Tochigi, Japan) using the Genelyzer FIII real-time fluorescence detection platform (Canon Medical Systems), as described previously [21]. The reaction mixture (total volume, 25 μL) contained 15 μL of isothermal master mix, 1 U of AMV reverse transcriptase (Nippon gene, Tokyo, Japan), 20 pmol (each) of FIP and BIP primers, 5 pmol (each) of F3 and B3 outer primers, 10 pmol (each) of LF and LB loop primers, and 1 μL of RNA sample (template). The primer sequences used are listed in Table 1. The reaction was performed under the following conditions: 68°C for 20 min, followed by a dissociation analysis at 95–75°C with a temperature change rate of 0.1°C/s. SARS-CoV-2 detection by RT-LAMP was performed twice for each clinical sample, and the average time to positivity (Tp) was calculated. Tp was defined as time period (minutes) from the beginning of the assay until the fluorescent signal reached 20,000. The synthesized RNAs containing the target sequence of the RT-LAMP assay were used as positive controls. Non-specific amplification was excluded by comparing the melting temperature with that of the positive control [20].
RT-qPCR assay
RT-qPCR was performed using the One Step PrimeScript III RT-qPCR Mix (Takara Bio, Shiga, Japan) as reported previously [27]. The reaction mixture (total volume = 20 μL) comprised 10 μL of 2× One Step PrimeScript III RT-qPCR Mix, 0.4 μL of ROX Reference Dye, 2 μL of 10× primer-probe mixture, 1 μL of RNA sample, and 6.6 μL of RNase-free water. The primers and probes were developed by the National Institute of Infectious Diseases, Japan (NIID). Forward primer, reverse primer, and FAM-labeled probes were used at final concentrations of 500 nM, 700 nM, and 200 nM, respectively [28]. Primer and probe sequences are listed in Table 1. The RT-qPCR reaction was performed using the StepOnePlus instrument (Thermo Fisher Scientific, Waltham, MA, USA) with a thermal cycle program conducted using the following conditions: 52°C for 5 min and 95°C for 10 s, followed by 45 cycles of 95°C for 5 s, and 60°C for 35 s. A threshold cycle (Ct) value of 40 was set as the cut-off value. To quantify viral genome copies, a standard curve was generated using ten-fold serial dilutions of the synthesized standard RNA derived from RT-qPCR target sequences. Each clinical sample was analyzed using RT-qPCR to quantify the copy number of the viral genome.
Whole-genome sequencing of SARS-CoV-2
COVID-19 positive samples assessed by RT-qPCR were subjected to whole-genome sequencing using a method similar to that reported previously [29]. Briefly, multiplex PCR was performed using primer sets designed by the online program Primal Scheme (https://primalscheme.com/) with an amplicon size of 450 bp. The template sequence was derived from the sequence of the Wuhan-Hu-1 strain (GISAID accession No. EPI_ISL_402125). Extracted viral RNA was reverse transcribed using SuperScript IV Reverse Transcriptase (Thermo Fisher Scientific) combined with random hexamers, according to the protocol of the manufacturer. Reverse-transcribed complementary DNA was used to perform multiplex PCR with Q5 High-Fidelity DNA Polymerase (New England Biolabs, Ipswich, MA, USA) as previously described [30]. Libraries were prepared using 50–500 ng of multiplex PCR products and the NEBNext Ultra II FS DNA Library Prep kit (New England Biolabs) in combination with NEBNext Multiplex Oligos for Illumina (New England Biolabs), according to the instructions of the manufacturer. After quantitative inspection of each library using the NEBNext Library Quant Kit for Illumina (New England Biolabs), we used the MiniSeq High Output Reagent Kit (Illumina, San Diego, CA, USA) to obtain complete genome sequences of SARS-CoV-2. Mapping of the paired-end reads was performed using CLC Genomics Workbench 11.0.1 software (Qiagen) using a whole-genome sequence of the Wuhan-Hu-1 strain as a template. Consensus sequences were extracted and aligned with reference strains using the BioEdit 7.0.5.3 software (http://www.mbio.ncsu.edu/BioEdit/bioedit.html). Sanger sequencing was performed to obtain complete genome sequences of several strains that showed short gaps or ambiguous sequences. After whole-genome sequencing, the lineage of each strain was analyzed using Pangolin v3.0 (https://github.com/cov-lineages/pangolin).
Results
Collection of samples positive for COVID-19 and determination of the lineage of SARS-CoV-2 in Gabon
Continuous genetic surveillance of SARS-CoV-2 has been conducted since the beginning of the COVID-19 epidemic in Gabon in March 2020. A total of 336 samples were selected for validation of the RT-LAMP assay, including at least one sample collected every month. Although the lineages of SARS-CoV-2 were determined by whole-genome sequencing during the surveillance period [17], SARS-CoV-2 lineages of 40 samples were still unclear. We attempted to determine SARS-CoV-2 lineages in these samples using next-generation sequencing and Sanger sequencing. In total, the lineages of 310 SARS-CoV-2 strains were determined. Between March 2020 and October 2021, three major waves of COVID-19 infection were recorded in Gabon (Fig 1). The B.1.1 variant that possesses D614G substitution in the spike protein was prevalent in the first wave, and the third wave involved infection primarily by the delta variant, as observed in the corresponding wave in the African continent. A variety of variants were detected in the second wave in Gabon: B.1 variant, 9.7%; B.1.1, 7.1%; Alpha, 35.7%; Beta, 5.6%; Eta, 12.2%; B.1.1.318, 17.3%; and other minor variants, 12.2% (Fig 1 and S1 Table).
The number of samples collected every month is shown on the graph. Bar colors depict the lineages of SARS-CoV-2 detected in each sample. Monthly COVID-19 cases in Gabon are represented on the graph with a dotted line. The first, second, and third waves of COVID-19 are highlighted in orange, and the lineages of SARS-CoV-2 detected in samples collected during each COVID-19 wave are shown in the upper pie graph.
Sensitivity of RT-LAMP assay for detection of viral RNA of SARS-CoV-2 variants
We previously developed an RT-LAMP assay using the sequence of the Wuhan-Hu-1 strain (GISAID accession number EPI_ISL_402125) as a reference [21]. To examine the sensitivity of the RT-LAMP assay for the detection of other variants of SARS-CoV-2, a dilution series of purified viral RNAs was prepared for B.1.1 (D614G), Alpha, Beta, Gamma, Delta, Kappa, Lambda, and Mu variants, corresponding to copy numbers ranging from 2 to 200,000 per reaction. The RT-LAMP assay showed a high sensitivity for all tested variants, detecting 50 copies of viral RNAs with 100% probability (Tables 2 and S2). In 20 copies of viral RNAs, the RT-LAMP assay detected B.1.1, Alpha, and Beta variants with 100% probability, whereas the detection probability of the Delta variant was 83.3%, that of Gamma, Kappa, and Mu variants was 66.7%, and that of the Lambda variant was 50.0%. The sensitivity to SARS-CoV-2 variants was consistent with previous data that showed a 100% probability of detecting 50 copies of the target RNA [21]. Particularly, the maximum average time to positivity (Tp) was 13.08 minutes (20 copies of the Lambda variant) in all tests, with Tp values lower than 15 minutes for any variant. These results indicate that the RT-LAMP assay demonstrated similar sensitivity for detecting other variants and can be considered suitable as a rapid diagnostic method for COVID-19.
Validation of RT-LAMP assay using clinical samples collected between March 2020 and October 2021 in Gabon
To evaluate the accuracy of the rapid diagnostic assay, 336 clinical samples collected in Gabon were tested using the RT-LAMP assay. The results showed a significant correlation between copies per reaction and Tp (Pearson’s correlation coefficient, r = −0.809; p < 0.001) (Fig 2). The minimum copy number was 69.2 copies per reaction, which was detected using the RT-LAMP assay in this study. The results are consistent with those of a previous report [21]. To validate the accuracy of the RT-LAMP assay for each variant detected in this study, variant-specific detection rates were calculated (Table 3). The assay detected SARS-CoV-2 with 100% probability in samples containing more than 102 copies of the Delta or B.1.1.318 variant genome per reaction. For other variants, several samples that were analyzed as positive by RT-qPCR were deemed negative by RT-LAMP for copies ranging from 102–103, and very few samples with <102 copies were detected as positive. Overall, the RT-LAMP assay could detect SARS-CoV-2 genomes with 100% probability in clinical samples including more than 292 copies of the target. When considering the copy range of 102–103, the capacity of SARS-CoV-2 detection markedly decreased when using clinical samples with less than 292 copies per reaction. Several samples that were analyzed negative by RT-qPCR were included, all of which were also detected as negative by RT-LAMP, indicating no false-positive reactions using the RT-LAMP assay. This result would support the high specificity of the primers to SARS-CoV-2 as reported in our previous study that revealed no cross-reactivity of the primers with SARS-CoV [21].
The time to positivity (Tp) determined by RT-LAMP was plotted against viral genome copies per reaction for each sample assessed in this study. Colors depict the lineages of SARS-CoV-2. The correlation curve is shown on the plots.
Validation of the application of RT-LAMP assay for the detection of Omicron variants
To validate the ability of the assay to detect Omicron variants, which were not included in the clinical samples assessed in this study, we evaluated the sensitivity of the RT-LAMP assay for the detection of viral RNA extracted from the culture supernatants of Omicron variants BA.1 and BA.2. The RT-LAMP assay could detect no more than 20 copies of the BA.1 genome with 100% probability, whereas it could detect 50 copies of the BA.2 genome with 100% probability and 20 copies with 83.3% probability (Table 4). The results clearly showed comparable sensitivity of the assay for the detection of Omicron and other variants. To estimate the potential ability for the detection of currently spreading BA.2 and BA.3 variants, we examined the nucleotide identity between sequences of RT-LAMP primers and Omicron variant genomes. With a selection to find “complete” and “low coverage excluded” sequences in GISAID, genomic data of 5,920 BA.2 strains submitted from regions across the globe and all deposited sequences of BA.3 strains comprising data corresponding to 1,678 strains were used in this study. Comparison between primer sequences and genomic data revealed mutations at 27 out of 186 sites in BA.2 strains (Fig 3). However, the minimum nucleotide identity was 99.85% at the sixth nucleotide of the B1 primer, and the average nucleotide identity in the entire primer binding region was more than 99.99% for BA.2 strains (Fig 3). Among 61 strains that showed mutations out of 5,920 BA.2 strains, 58 strains showed one mutation per strain, and only three strains showed two mutations in the region. Concerning the BA.3 strains, the mutations were observed at only eight sites within the primer binding region, and the minimum nucleotide identity was 99.88% (Fig 3). Among six BA.3 strains that possessed the mutation, one Indian strain (GISAID accession number, EPI_ISL_10716808) showed five mutations; however, only one mutation per strain was observed in the other five strains. As a very small population of each sublineage possessed mutations in the primer binding site, these results strongly indicated that the RT-LAMP assay would be useful for detecting Omicron variants.
Nucleotide identity was plotted against the nucleotide position of each RT-LAMP primer sequence. The arrows depict the length and direction of each primer. Colors indicate each primer set. A total of 5,920 and 1,678 strains were used for evaluating BA.2 and BA.3 variant sequences, respectively.
Discussion
Since the declaration of the pandemic in March 2020, systematic diagnostic methods for COVID-19 have been established in each country throughout the world. NAAT is one of the most sensitive and reliable methods for detecting SARS-CoV-2 in various types of clinical samples. Several RT-LAMP assays have been developed for COVID-19 diagnosis with the advantage of rapid and sensitive detection in a simple isothermal amplification step [22–25]. However, reports validating novel diagnostic methods are limited compared to those showing development of these methods, even though it is crucial to examine the long-term applicability of these methods.
RT-LAMP is a rapid and sensitive method for detecting target sequences using a portable detector that can be powered by a battery. Moreover, RT-LAMP is highly specific for targeted pathogens because of the use of 4–6 primers targeting 6–8 primer binding sites [31]. The advantages of the assay include diagnoses of infectious diseases on-site or in rural areas where resources and stable electricity are largely limited. Before the emergence of COVID-19, we had developed a variety of RT-LAMP assays to detect viruses that are a concern to public health, such as Ebola virus, Marburg virus, Lassa virus, and Zika virus [18–20,32–35]. In particular, RT-LAMP assays targeting Ebola and Zika viruses were tested on-site during the outbreaks, detecting target viruses within 30 min. Other groups have also actively developed several LAMP diagnostic assays for the detection of not only viruses but also other pathogens, such as malarial parasites and bacteria, for use in resource-limited settings [36–40]. Thus, RT-LAMP is recognized for its applications in on-site diagnosis or use in rural areas.
In the diagnosis of COVID-19, amplification failures, such as dropout or unusual delay, are occasionally reported with RT-qPCR. During the period when Alpha variant (B.1.1.7) was prevalent in Europe, several research groups reported S or N gene dropouts using commercial diagnostic kits based on RT-qPCR, which critically affected accurate diagnosis [41–43]. Subsequently, several cases of gene dropouts were reported in other variants, especially with S or N gene, indicating a potential risk with the use of RT-qPCR in diagnosis [44,45]. However, any point mutation in the primer binding site minimally affects amplification time and sensitivity, and is acceptable for rapid and accurate molecular diagnosis by RT-LAMP [46]. Moreover, our RT-LAMP assay could function well during the current spread of BA.2 and BA.3 Omicron variants due to very few mutations in the primer binding site (Fig 3). Fortunately, we selected the primer binding site at the nucleotide position of 21100–21355 in the orf1b region, which might show lower diversity than the S or N gene with remarkable sensitivity for use in diagnosis (Tables 2 and 4) [47]. Thus, our RT-LAMP assay demonstrated the potential for long-term usage, even in future waves of novel variants.
To date, several RT-LAMP assays have been developed to detect SARS-CoV-2, including our previous study [21–25,48,49]. Most of these assays were developed soon after the declaration of the COVID-19 pandemic and could detect 1.0 × 104–2.0 × 101 copies/reaction of SARS-CoV-2 genomes, indicating enough sensitivity to use in diagnosis for COVID-19. Using clinical specimens, several studies that developed novel RT-LAMP assays for COVID-19 calculated the Ct values of the detection limit by RT-qPCR, which ranged from 32 to 35 [23,26,48,49]. The detection limit of the current RT-LAMP assay was ranged within the Ct value of 33–36 using clinical specimens (Table 3), showing good consistency with previous studies. As the primer binding regions of these established RT-LAMP assays were largely distinct among each study, sensitivity and specificity seemed to be dependent simply on each primer sequence but not on specific target genes. Several studies attempted to detect SARS-CoV-2 directly from clinical specimens. Although sensitivity was 10–100-fold lower in direct RT-LAMP than that using extracted RNA samples, recent availability of reagents for the direct RT-LAMP assay may help the assay become more simple and convenient. Established RT-LAMP assays were validated using the original Wuhan-type or D614G variant, whereas a number of variants emerged after the development of these assays. Thus, it has been required to validate the established assays using newly emerged variants.
Antigen-detecting rapid diagnostic tests (Ag-RDTs) are powerful tools for clinical diagnosis at the point of care. A limit of detection (LOD) of 106 genomic copies/ml (corresponding to approximately 102 pfu/ml) was considered acceptable by WHO for Ag-RDTs used in clinical diagnosis [50–52], corresponding to 2.5 × 104 copies/reaction in our diagnostic system (total 25 μl/reaction). The RT-LAMP assay evaluated in this study showed an LOD of 20–50 copies of targeted viral RNAs during the analysis of a variety of SARS-CoV-2 variants, and the assay detected more than 292 copies with a 100% probability in clinical diagnosis (Tables 2–4). Although there was a difference of the LOD between purified viral RNAs and clinical samples, it might be general to observe such a case due to existing inhibitors of the RT-LAMP reaction, remaining salts and ethanol, and interference of annealing or extension steps by a large amount of unnecessary specimen-derived RNAs. These results clearly show that the sensitivity of the RT-LAMP assay was high enough to detect COVID-19 infection, even if the viral titer was too low for successful detection using Ag-RDTs. Moreover, in comparison with the nucleic acid lateral-flow assays (NLFAs; e.g. dipsticks), the RT-LAMP assay shows an advantage in time to detect viruses as NLFAs usually require pre-amplification of viral genome by RT-qPCR or RT-LAMP [53]. Therefore, the RT-LAMP assay has the potential to be used as a powerful diagnostic tool at the point of care, such as Ag-RDTs.
In the African continent, four major waves of COVID-19 have been reported so far [1]. Each African country has aimed to understand the spread and infectivity of COVID-19 via systematic diagnosis and genetic surveillance that promote the submission of whole-genome sequences of SARS-CoV-2 to public databases, identifying novel variants of concern (VOCs) such as the Omicron variants [11–13]. However, more than 50% of African SARS-CoV-2 sequences were submitted from a small number of countries, including South Africa, Kenya, Nigeria, and Senegal, which are conducting extensive genetic surveillance, possibly indicating the imbalance in the capacity of diagnosis and surveillance of COVID-19 among African countries. In Central Africa, where continuous surveillance is still challenging, the promotion of the use of simple and reliable diagnostic methods is urgently needed to elucidate the spread of COVID-19 in remote areas, as it is difficult to access the diagnosis facility, which is usually set in urban or semi-urban cities. Based on this perspective, we examined the accuracy of our RT-LAMP assay in Gabon, expecting a wide distribution of the assay in rural areas of Central Africa in the future. The assay was validated for detecting a number of variants with sufficient sensitivity for COVID-19 diagnosis. It is still necessary to provide sustainable support to understand the entire situation of COVID-19 in Africa.
The limitation of the present study is the difficulty in prediction of the applicability of the RT-LAMP assay against the novel unknown SARS-CoV-2 variants that would be prevalent in the near future. RT-LAMP is more suitable for the use in resource-limited settings than other molecular diagnostic methods, whereas the assay requires several basic resources such as laboratory equipment (e.g. micropipettes and tubes) and fully-charged butteries. The procurement cost for reagents is also a critical issue to be concerned to establish a sustainable diagnostic system. The present assay was approved as an official diagnostic method in Japan, indicating the sufficient reproducibility of the result using the assay. Further validation of the reproducibility by local researchers/clinicians will strengthen the advantage of the assay. For the worldwide use of the current assay including the resource-limited regions, cooperation must be required among researchers, companies, international public health organizations, and the government of each country to appropriately overcome those challenges.
Supporting information
S1 Table. Number of collected samples and SARS-CoV-2 lineage determination during COVID-19 waves in Gabon.
https://doi.org/10.1371/journal.pntd.0010964.s001
(XLSX)
S2 Table. Evaluation of the sensitivity of RT-LAMP assay using viral RNA of SARS-CoV-2 variants.
https://doi.org/10.1371/journal.pntd.0010964.s002
(XLSX)
Acknowledgments
We thank the COVID-19 diagnosis team at the Centre de Recherches Médicales de Lambaréné (CERMEL) for daily sampling and conducting RT-qPCR tests. We also thank Mayuko Kimura for technical assistance and Miku Takano for logistical management.
References
- 1.
World Health Organization. WHO Coronavirus (COVID-19) Dashboard. Available from: https://covid19.who.int/
- 2. Koelle K, Martin MA, Antia R, Lopman B, Dean NE. The changing epidemiology of SARS-CoV-2. Science. 2022;375(6585):1116–1121. pmid:35271324
- 3. Au WY, Cheung PPH. Diagnostic performances of common nucleic acid tests for SARS-CoV-2 in hospitals and clinics: a systematic review and meta-analysis. Lancet Microbe. 2021;2(12):e704–e714. pmid:34661181
- 4. Eftekhari A, Alipour M, Chodari L, Maleki Dizaj S, Ardalan M, Samiei M, et al. A comprehensive review of detection methods for SARS-CoV-2. Microorganisms. 2021;9(2):232. pmid:33499379
- 5.
Foundation for Innovative New Diagnostics. COVID-19. Test directory. Available from: https://www.finddx.org/covid-19/test-directory/
- 6. Corman VM, Landt O, Kaiser M, Molenkamp R, Meijer A, Chu DK, et al. Detection of 2019 novel coronavirus (2019-nCoV) by real-time RT-PCR. Euro Surveill. 2020;25(3):2000045. pmid:31992387
- 7. Vogels CBF, Brito AF, Wyllie AL, Fauver JR, Ott IM, Kalinich CC, et al. Analytical sensitivity and efficiency comparisons of SARS-CoV-2 RT-qPCR primer-probe sets. Nat Microbiol. 2020;5(10):1299–1305. pmid:32651556
- 8. Peto T; UK COVID-19 Lateral Flow Oversight Team. COVID-19: Rapid antigen detection for SARS-CoV-2 by lateral flow assay: A national systematic evaluation of sensitivity and specificity for mass-testing. EClinicalMedicine. 2021;36:100924. pmid:34101770
- 9. Mistry DA, Wang JY, Moeser ME, Starkey T, Lee LYW. A systematic review of the sensitivity and specificity of lateral flow devices in the detection of SARS-CoV-2. BMC Infect Dis. 2021;21(1):828. pmid:34407759
- 10. Soroka M, Wasowicz B, Rymaszewska A. Loop-Mediated Isothermal Amplification (LAMP): The Better Sibling of PCR? Cells. 2021;10(8):1931. pmid:34440699
- 11. Cui Z, Liu P, Wang N, Wang L, Fan K, Zhu Q, et al. Structural and functional characterizations of infectivity and immune evasion of SARS-CoV-2 Omicron. Cell. 2022;185(5):860–871.e13. pmid:35120603
- 12. Cao Y, Wang J, Jian F, Xiao T, Song W, Yisimayi A, et al. Omicron escapes the majority of existing SARS-CoV-2 neutralizing antibodies. Nature. 2022;602(7898):657–663. pmid:35016194
- 13. Suzuki R, Yamasoba D, Kimura I, Wang L, Kishimoto M, Ito J, et al. Attenuated fusogenicity and pathogenicity of SARS-CoV-2 Omicron variant. Nature. 2022;603(7902):700–705. pmid:35104835
- 14. Abe H, Ushijima Y, Bikangui R, Zoa-Assoumou S, Ondo GN, Manouana GP, et al. Unrecognized introduction of SARS-CoV-2 variants of concern to Central Africa: Import and local transmission of B.1.1.7 in Gabon in the very early stage of the variant spread to the African continent. J Med Virol. 2021;93(10):6054–6058. pmid:34185327
- 15. Zoa-Assoumou S, Ndeboko B, Manouana GP, Houechenou RMA, Bikangui R, Mveang-Nzoghe A, et al. SARS-CoV-2 emerging variants in Africa: view from Gabon. Lancet Microbe. 2021;2(8):e349. pmid:34124702
- 16. Manouana GP, Nzamba Maloum M, Bikangui R, Oye Bingono SO, Ondo Nguema G, Honkpehedji JY, et al. Emergence of B.1.1.318 SARS-CoV-2 viral lineage and high incidence of alpha B.1.1.7 variant of concern in the Republic of Gabon. Int J Infect Dis. 2022;114:151–154. pmid:34742926
- 17. Abe H, Ushijima Y, Bikangui R, Ondo Nguema G, Lell B, Adegnika AA, et al. Delays in the arrival of the waves of COVID-19: a comparison between Gabon and the African continent. Lancet Microbe. 2022 Apr 21. pmid:35779562
- 18. Kurosaki Y, Magassouba N, Oloniniyi OK, Cherif MS, Sakabe S, Takada A, et al. Development and evaluation of reverse transcription-loop-mediated isothermal amplification (RT-LAMP) assay coupled with a portable device for rapid diagnosis of Ebola virus disease in Guinea. PLoS Negl Trop Dis. 2016;10(2):e0004472. pmid:26900929
- 19. Kurosaki Y, Magassouba N, Bah HA, Soropogui B, Doré A, Kourouma F, et al. Deployment of a reverse transcription-loop-mediated isothermal amplification test for Ebola virus surveillance in remote areas in Guinea. J Infect Dis. 2016: 214 (suppl 3): S229–S233. pmid:27481863
- 20. Kurosaki Y, Martins DBG, Kimura M, Catena ADS, Borba MACSM, Mattos SDS, et al. Development and evaluation of a rapid molecular diagnostic test for Zika virus infection by reverse transcription loop-mediated isothermal amplification. Sci Rep. 2017;7(1):13503. pmid:29044149
- 21. Yoshikawa R, Abe H, Igasaki Y, Negishi S, Goto H, Yasuda J. Development and evaluation of a rapid and simple diagnostic assay for COVID-19 based on loop-mediated isothermal amplification. PLoS Negl Trop Dis. 2020;14(11):e0008855. pmid:33147214
- 22. Ganguli A, Mostafa A, Berger J, Aydin MY, Sun F, Ramirez SAS, Valera E, Cunningham BT, King WP, Bashir R. Rapid isothermal amplification and portable detection system for SARS-CoV-2. Proc Natl Acad Sci U S A. 2020;117(37):22727–22735. pmid:32868442
- 23. Fowler VL, Armson B, Gonzales JL, Wise EL, Howson ELA, Vincent-Mistiaen Z, et al. A highly effective reverse-transcription loop-mediated isothermal amplification (RT-LAMP) assay for the rapid detection of SARS-CoV-2 infection. J Infect. 2021;82(1):117–125. pmid:33271166
- 24. Lu R, Wu X, Wan Z, Li Y, Jin X, Zhang C. A Novel reverse transcription loop-mediated isothermal amplification method for rapid detection of SARS-CoV-2. Int J Mol Sci. 2020;21(8):2826. pmid:32325642
- 25. Baek YH, Um J, Antigua KJC, Park JH, Kim Y, Oh S, et al. Development of a reverse transcription-loop-mediated isothermal amplification as a rapid early-detection method for novel SARS-CoV-2. Emerg Microbes Infect. 2020;9(1):998–1007. pmid:32306853
- 26. Sakurai Y, Ngwe Tun MM, Kurosaki Y, Sakura T, Inaoka DK, Fujine K, et al. 5-amino levulinic acid inhibits SARS-CoV-2 infection in vitro. Biochem Biophys Res Commun. 2021;545:203–207. pmid:33571909
- 27. Abe H, Ushijima Y, Loembe MM, Bikangui R, Nguema-Ondo G, Mpingabo PI, et al. Re-emergence of dengue virus serotype 3 infections in Gabon in 2016–2017, and evidence for the risk of repeated dengue virus infections. Int J Infect Dis. 2020;91:129–136. pmid:31821892
- 28. Shirato K, Nao N, Katano H, Takayama I, Saito S, Kato F, et al. Development of genetic diagnostic methods for detection for novel coronavirus 2019 (nCoV-2019) in Japan. Jpn J Infect Dis. 2020;73(4):304–307. pmid:32074516
- 29. Abe H, Ushijima Y, Amano M, Sakurai Y, Yoshikawa R, Kinoshita T, et al. Unique evolution of SARS-CoV-2 in the second large cruise ship cluster in Japan. Microorganisms. 2022;10(1):99. pmid:35056548
- 30. Quick J, Grubaugh ND, Pullan ST, Claro IM, Smith AD, Gangavarapu K, et al. Multiplex PCR method for MinION and Illumina sequencing of Zika and other virus genomes directly from clinical samples. Nat. Protoc. 2017;12:1261–1276. pmid:28538739
- 31. Notomi T, Okayama H, Masubuchi H, Yonekawa T, Watanabe K, Amino N, et al. Loop-mediated isothermal amplification of DNA. Nucleic Acids Res. 2000;28:E63. pmid:10871386
- 32. Kurosaki Y, Grolla A, Fukuma A, Feldmann H, Yasuda J. Development and evaluation of a simple assay for Marburg virus detection using a reverse transcription-loop-mediated isothermal amplification method. J Clin Microbiol. 2010;48(7):2330–6. pmid:20421440
- 33. Fukuma A, Kurosaki Y, Morikawa Y, Grolla A, Feldmann H, Yasuda J. Rapid detection of Lassa virus by reverse transcription-loop-mediated isothermal amplification. Microbiol Immunol. 2011; 55(1):44–50. pmid:21175773
- 34. Oloniniyi OK, Kurosaki Y, Miyamoto H, Takada A, Yasuda J. Rapid detection of all known ebolavirus species by reverse transcription-loop-mediated isothermal amplification (RT-LAMP). J Virol Methods. 2017;246:8–14. pmid:28356221
- 35. Pemba CM, Kurosaki Y, Yoshikawa R, Oloniniyi OK, Urata S, Sueyoshi M, Zadeh VR, et al. Development of an RT-LAMP assay for the detection of Lassa viruses in southeast and south-central Nigeria. J Virol Methods. 2019;269:30–37. pmid:30974179
- 36. Paris DH, Imwong M, Faiz AM, Hasan M, Yunus EB, Silamut K, et al. Loop-mediated isothermal PCR (LAMP) for the diagnosis of falciparum malaria. Am J Trop Med Hyg. 2007;77(5):972–6. pmid:17984362.
- 37. Polley SD, Mori Y, Watson J, Perkins MD, González IJ, Notomi T, et al. Mitochondrial DNA targets increase sensitivity of malaria detection using loop-mediated isothermal amplification. J Clin Microbiol. 2010;48(8):2866–71. pmid:20554824
- 38. Seki M, Kilgore PE, Kim EJ, Ohnishi M, Hayakawa S, Kim DW. Loop-mediated isothermal amplification methods for diagnosis of bacterial meningitis. Front Pediatr. 2018;6:57. pmid:29594087
- 39. Vergara A, Boutal H, Ceccato A, López M, Cruells A, Bueno-Freire L, et al. Assessment of a loop-mediated isothermal amplification (LAMP) assay for the rapid detection of pathogenic bacteria from respiratory samples in patients with hospital-acquired pneumonia. Microorganisms. 2020;8(1):103. pmid:31940771
- 40. World Health Organization. The use of loop-mediated isothermal amplification (TB-LAMP) for the diagnosis of pulmonary tuberculosis: policy guidance. 2016. Available from: https://apps.who.int/iris/handle/10665/249154
- 41. Bal A, Destras G, Gaymard A, Stefic K, Marlet J, Eymieux S, et al. Two-step strategy for the identification of SARS-CoV-2 variant of concern 202012/01 and other variants with spike deletion H69-V70, France, August to December 2020. Euro Surveill. 2021;26(3):2100008. pmid:33478625
- 42. Wollschläger P, Todt D, Gerlitz N, Pfaender S, Bollinger T, Sing A, et al. SARS-CoV-2 N gene dropout and N gene Ct value shift as indicator for the presence of B.1.1.7 lineage in a commercial multiplex PCR assay. Clin Microbiol Infect. 2021;27(9):1353.e1–1353.e5. pmid:34044153
- 43. Sánchez-Calvo JM, Alados Arboledas JC, Ros Vidal L, de Francisco JL, López Prieto MD. Diagnostic pre-screening method based on N-gene dropout or delay to increase feasibility of SARS-CoV-2 VOC B.1.1.7 detection. Diagn Microbiol Infect Dis. 2021;101(4):115491. pmid:34464903
- 44. So MK, Park S, Lee K, Kim SK, Chung HS, Lee M. Variant prediction by analyzing RdRp/S gene double or low amplification pattern in Allplex SARS-CoV-2 assay. Diagnostics (Basel). 2021;11(10):1854. pmid:34679552
- 45. Bozidis P, Tsaousi ET, Kostoulas C, Sakaloglou P, Gouni A, Koumpouli D, et al. Unusual N gene dropout and Ct value shift in commercial multiplex PCR assays caused by mutated SARS-CoV-2 strain. Diagnostics (Basel). 2022;12(4):973. pmid:35454022
- 46. Tamanaha E, Zhang Y, Tanner NA. Profiling RT-LAMP tolerance of sequence variation for SARS-CoV-2 RNA detection. PLoS One. 2022;17(3):e0259610. pmid:35324900
- 47. Kaushal N, Gupta Y, Goyal M, Khaiboullina SF, Baranwal M, Verma SC. Mutational frequencies of SARS-CoV-2 genome during the beginning months of the outbreak in USA. Pathogens. 2020;9(7):565. pmid:32668692
- 48. de Oliveira Coelho B, Sanchuki HBS, Zanette DL, Nardin JM, Morales HMP, Fornazari B, Aoki MN, Blanes L. Essential properties and pitfalls of colorimetric Reverse Transcription Loop-mediated Isothermal Amplification as a point-of-care test for SARS-CoV-2 diagnosis. Mol Med. 2021;27(1):30. pmid:33771097
- 49. Lee JYH, Best N, McAuley J, Porter JL, Seemann T, Schultz MB, Sait M, Orlando N, Mercoulia K, Ballard SA, Druce J, Tran T, Catton MG, Pryor MJ, Cui HL, Luttick A, McDonald S, Greenhalgh A, Kwong JC, Sherry NL, Graham M, Hoang T, Herisse M, Pidot SJ, Williamson DA, Howden BP, Monk IR, Stinear TP. Validation of a single-step, single-tube reverse transcription loop-mediated isothermal amplification assay for rapid detection of SARS-CoV-2 RNA. J Med Microbiol. 2020;69(9):1169–1178. pmid:32755529
- 50. World Health Organization. COVID-19 Target product profiles for priority diagnostics to support response to the COVID-19 pandemic v.1.0. 2020. Available from: https://cdn.who.int/media/docs/default-source/blue-print/who-rd-blueprint-diagnostics-tpp-final-v1-0-28-09-jc-ppc-final-cmp92616a80172344e4be0edf315b582021.pdf?sfvrsn=e3747f20_1&download=true
- 51. Department of Health and Social Care. Protocol for evaluation of rapid diagnostic assays for specific SARS-CoV-2 antigens (lateral flow devices) GOV.UK. 2020. Available from: https://www.gov.uk/government/publications/assessment-and-procurement-of-coronavirus-covid-19-tests/protocol-for-evaluation-of-rapid-diagnostic-assays-for-specific-sars-cov-2-antigens-lateral-flow-devices
- 52. Cubas-Atienzar AI, Kontogianni K, Edwards T, Wooding D, Buist K, Thompson CR, et al. Limit of detection in different matrices of 19 commercially available rapid antigen tests for the detection of SARS-CoV-2. Sci Rep. 2021;11(1):18313. pmid:34526517
- 53. Hsiao WW, Le TN, Pham DM, Ko HH, Chang HC, Lee CC, et al. Recent Advances in Novel Lateral Flow Technologies for Detection of COVID-19. Biosensors. 2021;11(9):295. pmid:34562885