Is POC-CCA a truly reliable test for schistosomiasis diagnosis in low endemic areas? The trace results controversy

  • José M. Peralta,

    Roles Conceptualization, Writing – original draft, Writing – review & editing

    Affiliation Departmento de Imunologia, Instituto de Microbiologia Paulo de Góes, Universidade Federal do Rio de Janeiro, Rio de Janeiro, Brazil

  • Marta G. Cavalcanti

    Roles Conceptualization, Writing – original draft, Writing – review & editing

    cavalcanti.marta66@gmail.com

    Affiliations Departmento de Imunologia, Instituto de Microbiologia Paulo de Góes, Universidade Federal do Rio de Janeiro, Rio de Janeiro, Brazil, Serviço de Doenças Infecciosas e Parasitárias, Hospital Universitário Clementino Fraga Filho, Universidade Federal do Rio de Janeiro, Rio de Janeiro, Brazil

Introduction

In the last decade, development of point-of-care platforms (POCs) for diagnosis of infectious diseases or rapid tests (RTs) has increased in order to attend to the need for reliable diagnostic tests for surveillance of endemic areas. Also, preventive chemotherapy in areas without sophisticated resources has become more affordable according to some field studies [1,2]. The advantages of application of RTs such as POC-Schistosoma circulating cathodic antigen (CCA) are undeniable in community settings. But some evidence also pointed to other uses, like in institutional settings, by investigating active infections in susceptible populations like travelers, immigrants, and/or refugees [3,4]. Nonetheless, a great deal of discussion still remains toward the exclusive use of the RTs and/or POCs as tools for monitoring schistosomiasis in controlled areas or areas undergoing elimination. Despite the obvious advantages, the growing application of RTs has also shown some pitfalls.

Use of novel approaches in the diagnosis of schistosomiasis: Controversial results

When the first commercial version launched, POC-CCA acceptance was huge among field researchers. CCA can be detected by a labeled monoclonal antibody immobilized on the sample membrane. The solution then runs over the strip where the antigen–antibody complex attaches to another monoclonal antibody immobilized at the test line. The lateral flow assay is an easy to use test, which the first version worked by applying a single drop of urine with a buffer. Following 20 minutes of incubation, results were determined by visual reading (Schisto POC-CCA cassette based test; Rapid Medical Diagnostics, Pretoria, South Africa). Most data produced came from collaborative foreign research groups working in areas of moderate to high endemicity and populations infected with high parasite loads but also in low endemic areas as well as individuals with very light infection [510]. RTs and/or POCs may surpass conventional methods (such as the Kato-Katz test) as reliable diagnostic tools in moderate and high endemic areas. Yet variable results may also be found on a daily basis in these areas, and very low reactivity reactions in non–egg-excretors might misdiagnose active infection after praziquantel (PZQ) use [9]. Also, POC-CCA accuracy may be questionable in areas of S. haematobium mono infection or multiple Schistosoma infections [1113]. But discussion becomes really intense when RTs are used in low parasite load and/or low endemic areas. Although several works consistently demonstrated that sensitivity, and also specificity, are compromised in low endemic areas, some new evidence claims that POC-CCA reliability and accuracy are high in areas of low prevalence [1416]. Nonetheless, as in areas of high and moderate endemicity, residual reactivity (trace) was fairly demonstrated in individuals without active schistosomiasis both before and after (re)treatment in community settings with low parasite burden and/or low endemicity [16]. Usually, the bottleneck is the use of the low-sensitivity Kato-Katz test as a “gold standard.” Interpretation of trace as positive (presence of infection) or negative (absence of infection) creates lots of discussion about the validation of POC-CCA as a single test to replace Kato-Katz for the estimation of schistosomiasis prevalence in transmission areas. Particularly concerned with POC-CCA performance in low endemic areas, Clements and colleagues [17] evaluated trace results by applying Bayesian latent class analysis to improve the low sensitivity of the Kato-Katz and the so-called absence of a “gold standard.” In the same study, up-converting phosphor lateral flow circulating anodic antigen (UCP-LF CAA), a non-commercially available assay, was used to evaluate preselected trace-positive POC-CCA samples by assuming UCP-LF CAA was 100% specific. The results showed that POC-CCA outperformed Kato-Katz. However, considering trace as negative, Schistosoma active infection was missed in many cases. In contrast, trace as positive overestimated the prevalence. But by running UCP-LF CAA on selected trace-positive samples, the prevalence estimates from POC-CCA trace-positive samples should indeed represent the real prevalence in a lightly infected population. Recently, a new S. mansoni POC-CCA test became available, provided by only one manufacturer in the world. The new assay now uses only two drops of urine straight into a cassette instead of one drop of urine in addition to one drop of buffer as the recently discontinued previous version. The reading is done after 20 minutes like the older version. So far, no information was declared by the company regarding modifications on the test principle.

Despite the successful removal of traces, the elimination of potential false positivity resulted in no detection of true positives. Therefore, individuals with active schistosomiasis were misdiagnosed as negative by the new test version in low endemic areas. In the past, custom-tailored kits were tested, aiming to overcome the debatable accuracy and performance of POC-CCA in a low-prevalence population of school children [18]. Because only limited information was disclosed by the manufacturers, one assumes that changes on test sensitivity were also performed since the last commercially available assay.

One of the main reasons to adopt RTs as a diagnostic strategy is detection of active infection for surveillance purposes, including fast determination of prevalence in transmission areas and drug response. However, the POC-CCA for S. mansoni showed conflicting results depending on the area studied. In low endemic areas where low parasite loads or no egg-excretion predominates, POC-CCA was unsatisfactory as a solitary tool for diagnosis of active infections [15, 19]. Previous epidemiological studies in areas of high and moderate endemicity showed promising preliminary results during pretreatment with PZQ. However, persistence of very weak reactivity (trace) in the absence of egg excretion and/or other evidence of active infection could be seen in individuals treated with one to three rounds of PZQ [9]. Mwinzi and colleagues [20] showed that after the first round of PZQ use, 47% and 34% of POC-CCA reactive individuals ended up responding to second and third rounds of treatment, respectively, by becoming negative on POC-CCA. Nonetheless, more than 60% of retreated individuals still remained reactive. Other diagnostic tests had low agreement between pre- and post-therapy use, therefore the discussion remains about the usefulness of trace results. Recently, Coelho and colleagues [21] proposed optimization of POC-CCA in low–worm-burden samples by using lyophilization of urine as an antigen concentration method, showing that most of trace results turned negative and some negative samples became trace. In short, predefinition of trace as positive or negative may end up misleading test interpretation. In contrast, Prada and colleagues [22] addressed the true meaning of trace post-PZQ use and showed that POC-CCA seems to be a better predictor of post-treatment prevalence. Even trace results could be associated with true active infection. Following changes performed in RT manufacturing, as described above, very weak responses were swept out of the low endemicity areas, and no traces of “true” active infection could be detected. Now, the underscored results point to a possible underestimation of the real prevalence both pre- and post-treatments since traces disappeared. Without any consultation of the scientific community, the only available commercial test used so far in epidemiological studies changed its formulation and discontinued the previous test version (Schisto POC-CCA cassette based test; Rapid Medical Diagnostics, Pretoria, South Africa). No previous validation studies were performed in low endemic areas before commercial release of the test by the company. Also, the actual conflicting results do not permit decisions like the substitution of reference tests for POC-CCA.

Conclusions and authors’ view

RTs showed high accuracy and performance in moderate and high transmission areas. The tests in platforms of point-of-care to diagnose Schistosoma infections are an appealing solution for mapping and surveillance of transmission areas because of their main characteristics: user-friendly and relatively low price. However, consistent evidence showed limited application of RTs in schistosomiasis diagnosis in low endemic areas. Mostly, in those settings, low parasite burden predominates, which results in a decrease of diagnostic test accuracy. Moreover, the presence of trace reactivity is a real pitfall. POC-CCA very low intensity reactions may hide low parasite–load-induced infections. Therefore, removal of trace might result in no detection of Schistosoma infection pre- and post-PZQ use and, as a result, improvement of the test must be encouraged. However, testing should precede the use by national programs. As it is, the diagnosis of Schistosoma infection pre- and post-therapy based solely on RTs and/or POC may be compromised in areas of controlled or eliminated schistosomiasis. Strategies based on a single commercial product without prevalidation in both low endemicity areas and in institutional settings can be worrisome. Blinded changes in RT formulation may end up increasing disaster situations and confusion. Commercially available diagnostic tools should be tested exhaustingly before release for general use in transmission and nonendemic areas worldwide.

References

  1. 1. Zarei M. Infectious pathogens meet poin-of-care diagnostics. Biosens Bioelectron 2018; 106: 193–203. pmid:29428589
  2. 2. Kuupiel D, Bawontuo V, Mashamba-Thompson TP. Improving the accessibility and efficiency of poin-of-care diagnostics services in low- and middle-income countries: lean and agile supply chain management. Diagnostics (Basel). 2017;7(4). pii: E58.
  3. 3. Chernet A, Kling K, Sydow V, Kuenzli E, Hatz C, Utzinger J, van Lieshout L, Marti H, Nickel B, Labhardt ND, Neumayr A. Accuracy of Diagnostic Tests for Schistosoma mansoni Infection in Asymptomatic Eritrean Refugees: Serology and Point-of-Care Circulating Cathodic Antigen Against Stool Microscopy. Clin Infect Dis. 2017;65(4):568–574. pmid:28430889
  4. 4. Becker SL, Marti H, Zimmermann S, Vidacek D, Herrmann M, Utzinger J, Schnabel PA, Bohle RM. Application in Europe of a urine-based rapid diagnostic test for confirmation of Schistosoma mansoni infection in migrants from endemic areas. Euro Surveill. 2015; 20(23). pii: 21151.
  5. 5. Colley DG, Andros TS, Campbell CH Jr. Schistosomiasis is more prevalent than previously thought: what does it mean for public health goals, policies, strategies, guidelines and intervention programs? Infect Dis Poverty. 2017;6 (1):63. pmid:28327187
  6. 6. Danso-Appiah A, Minton J, Boamah D, Otchere J, Asmah RH, Rodgers M, Bosompem KM, Eusebi P, De Vlas SJ. Accuracy of point-of-care testing for circulatory cathodic antigen in the detection of schistosome infection: systematic review and meta-analysis. Bull World Health Organ. 2016; 94(7):522–533A. pmid:27429491
  7. 7. Bezerra FSM, Leal JKF, Sousa MS, Pinheiro MCC, Ramos AN Jr, Silva-Moraes V, Katz N. Evaluating a point-of-care circulating cathodic antigen test (POC-CCA) to detect Schistosomamansoni infections in a low endemic area in north-eastern Brazil. Acta Trop. 2018;182:264–270. pmid:29526480
  8. 8. Erko B, Medhin G, Teklehaymanot T, Degarege A, Legesse M. Evaluation of urine-circulating cathodic antigen (Urine-CCA) cassette test for the detection of Schistosoma mansoni infection in areas of moderate prevalence in Ethiopia. Trop Med Int Health. 2013; 18(8):1029–35. pmid:23590255
  9. 9. Coulibaly JT, N’Gbesso YK, Knopp S, N’Guessan NA, Silué KD, van Dam GJ, N’Goran EK, Utzinger J. Accuracy of urine circulating cathodic antigen test for the diagnosis of Schistosoma mansoni in preschool-aged children before and after treatment. PLoS Negl Trop Dis. 2013; 7(3):e2109. pmid:23556011
  10. 10. Colley DG, Binder S, Campbell C, King CH, Tchuem Tchuenté LA, N’Goran EK, Erko B, Karanja DM, Kabatereine NB, van Lieshout L, Rathbun S. A five-country evaluation of a point-of-care circulating cathodic antigen urine assay for the prevalence of Schistosoma mansoni. Am J Trop Med Hyg. 2013; 88(3):426–32. pmid:23339198
  11. 11. Sanneh B, Joof E, Sanyang AM, Renneker K, Camara Y, Sey AP, Jagne S, Baldeh I, Ceesay SJ, Sambou SM, Ogoussan K. Field evaluation of a schistosome circulating cathodic antigen rapid test kit at point-of-care for mapping of schistosomiasis endemic districts in The Gambia. PLoS ONE. 2017; 12(8):e0182003. pmid:28797128
  12. 12. Obeng BB, Aryeetey YA, de Dood CJ, Amoah AS, Larbi IA, Deelder AM, Yazdanbakhsh M, Hartgers FC, Boakye DA, Verweij JJ, van Dam GJ, van Lieshout L. Application of a circulating-cathodic-antigen (CCA) strip test and real-time PCR, in comparison with microscopy, for the detection of Schistosoma haematobium in urine samples from Ghana. Ann Trop Med Parasitol. 2008 Oct;102(7):625–33. pmid:18817603
  13. 13. Ashton RA, Stewart BT, Petty N, Lado M, Finn T, Brooker S, Kolaczinski JH. Accuracy of circulating cathodic antigen tests for rapid mapping of Schistosoma mansoni and S. haematobium infections in Southern Sudan. Trop Med Int Health. 2011;16(9):1099–103.
  14. 14. Okoyo C, Simiyu E, Njenga SM, Mwandawiro C. Comparing the performance of circulating cathodic antigen and Kato-Katz techniques in evaluating Schistosoma mansoni infection in areas with low prevalence in selected counties of Kenya: a cross-sectional study. BMC Public Health. 2018;18(1):478. pmid:29642875
  15. 15. Oliveira WJ, Magalhães FDC, Elias AMS, de Castro VN, Favero V, Lindholz CG, Oliveira ÁA, Barbosa FS, Gil F, Gomes MA, Graeff-Teixeira C, Enk MJ, Coelho PMZ, Carneiro M, Negrão-Corrêa DA, Geiger SM. Evaluation of diagnostic methods for the detection of intestinal schistosomiasis in endemic areas with low parasite loads: Saline gradient, Helmintex, Kato-Katz and rapid urine test. PLoS Negl Trop Dis. 2018;12(2):e0006232. pmid:29470516
  16. 16. Grenfell RFQ, Taboada D, Coutinho LA, Pedrosa MLC, Assis JV, Oliveira MSP, Cruz RR, Almeida A, Silva-Moraes V, Katz N, Coelho PMZ. Innovative methodology for point-of-care circulating cathodic antigen with rapid urineconcentration for use in the field for detecting low Schistosoma mansoni infection and for control of cure with high accuracy. Trans R Soc Trop Med Hyg. 2018;112(1):1–7. pmid:29522211
  17. 17. Clements MN, Corstjens PLAM, Binder S, Campbell CH Jr, de Dood CJ, Fenwick A, Harrison W, Kayugi D, King CH, Kornelis D, Ndayishimiye O, Ortu G, Lamine MS, Zivieri A, Colley DG, van Dam GJ. Latent class analysis to evaluate performance of point-of-care CCA for low-intensity Schistosoma mansoni infections in Burundi. Parasit Vectors. 2018;11(1):111. pmid:29475457
  18. 18. POC-CCA: Field evaluations of the Point-of-Care (POC) Circulating Cathodic Antigen (CCA) urine assay for detection of S. mansoni infection. Schistosomiasis Consortium for Operational Research and Evaluation website. https://score.uga.edu/projects/poc-cca/. [cited 2018 Oct 12].
  19. 19. Shiff C. Accurate diagnostics for schistosomiasis: a new role for PCR? Reports in Parasitology 2015:4 23–29.
  20. 20. Mwinzi PN, Kittur N, Ochola E, Cooper PJ, Campbell CH Jr, King CH, Colley DG. Additional Evaluation of the Point-of-Contact Circulating Cathodic Antigen Assay for Schistosoma mansoni Infection. Front Public Health. 2015; 3:48 pmid:25853117
  21. 21. Coelho PM, Siqueira LM, Grenfell RF, Almeida NB, Katz N, Almeida Á, Carneiro NF, Oliveira E. Improvement of POC-CCA Interpretation by Using Lyophilization of Urine from Patients with Schistosoma mansoni Low Worm Burden: Towards an Elimination of Doubts about the Concept of Trace. PLoS Negl Trop Dis. 2016;10(6):e0004778. pmid:27326453
  22. 22. Prada JM, Touloupou P, Adriko M, Tukahebwa EM, Lamberton PHL, Hollingsworth TD. Understanding the relationship between egg- and antigen-based diagnostics of Schistosoma mansoni infection pre- and post-treatment in Uganda. Parasit Vectors. 2018; 11(1):21. pmid:29310695