Abstract
Objective
The objective of this study was to determine if peri-operative contrast enhanced ultrasound (CEUS) is as accurate as magnetic resonance imaging (MRI) in measuring the immediate effect of microwave ablation (MWA) treatment of uterine fibroids. Secondary outcomes were ablation ratio and proportion of recurrence of treated fibroids at 6 months.
Methods
The study was a prospective observational trial. Seventeen patients with symptomatic uterine fibroids were included and treated with percutaneous ultrasound guided microwave ablation. Before the ablation, the volume of the up to three largest fibroids were recorded with MRI and CEUS. At the end of operation, the ablated volume was recorded with CEUS and compared to MRI within a week. At 6 months, the ablated volume was again recorded with MRI and any sign of regrowth of the treated fibroid(s) was noted.
Results
A total of 19 fibroids were analyzed. The intraclass correlation coefficient (ICC) between ablation volume obtained by CEUS and MRI respectively was calculated to be 0.965 (95% confidence interval 0.909, 0.987). The Bland-Altman graph showed that the mean difference between volumes was –1.68 ml and standard deviation 22.60 ml, giving the limits of agreement (−45.98, 42.61). Regrowth of vascularized tissue was seen in 8 fibroids (47%).
Citation: Beermann M, Jonsdottir G, Cronsioe A, Hasselrot K, Kopp-Kallner H (2026) Evaluation of the accuracy of peri-operative CEUS in measuring the effect of microwave ablation treatment of uterine fibroids, in comparison with MRI. PLoS One 21(8): e0356141. https://doi.org/10.1371/journal.pone.0356141
Editor: Yufeng Zhou, Chongqing Medical University, CHINA
Received: April 6, 2026; Accepted: July 30, 2026; Published: August 25, 2026
Copyright: © 2026 Beermann et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Data Availability: All relevant data are within the paper.
Funding: Karolinska Institutet FoUI-986094.
Competing interests: Marie Beermann reports a relationship with Medtronic that includes non-financial support and travel reimbursement. Helena Kopp-Kallner reports a relationship with Medtronic that includes non-financial support and travel reimbursement. Helena Kopp-Kallner reports a relationship with Mermaid Medical that includes non-financial support. Other authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper. This does not alter our adherence to PLOS ONE policies on sharing data and materials.
Introduction
Uterine fibroids are common in women of reproductive age, reaching a lifetime incidence of up to 70% [1]. Although benign in nature, symptoms of heavy menstrual bleeding, pelvic pressure, pain and infertility may severely decrease quality of life in affected women [2]. Treatment recommendations should be individualized and based on symptoms, the size, number and location of fibroids, and desire for future pregnancy. If medical treatment is not sufficient, hysterectomy is often recommended as curative treatment but may lead to quite severe short- and long-term complications [3]. In recent years, increasing attention has been paid to minimally invasive therapies such as high-intensity focused ultrasound (HIFU), uterine artery embolization (UAE), radiofrequency ablation (RFA) and percutaneous microwave ablation (MWA), with focus on uterine preservation and symptom improvement [4]. HIFU is a totally non-invasive but time-consuming method, and best suited for women with fewer fibroids and smaller uteri [5]. Although non-invasive, bowel or other organs in the way of the ultrasound beam constitutes a contraindication to treatment [6]. UAE is an established method to decrease volume of fibroids and improve clinical symptoms [7], even though post-operative pain and post-embolization syndrome (PES) associated with the procedure have somewhat limited its use [8]. There has also been concern that UAE may increase the risk of earlier menopause by affecting ovarian function, and the risk appears to be greater in women older than 40 years. However, previous studies have yielded conflicting results [9,10]. RFA is a thermal ablation technique inducing coagulative necrosis of tissue around the tip of an antenna placed in the target of interest [11]. For treatment of fibroids, RFA has mainly been used with a laparoscopic or, of late, vaginal approach even though there are studies of percutaneous RFA showing promising results [12–14]. Like RFA, microwave ablation (MWA) generates heat and causes coagulative necrosis. Used correctly, it is safe, fast and easy to perform, and it is well tolerated by patients [15,16]. Compared to the other non- or minimally invasive thermal techniques discussed above, MWA has the advantage of consistently higher tissue temperatures leading to larger ablation volumes in shorter time. MWA is also consistently effective independent of tissue types, leading to a more predictable outcome [17].
To assess uterine fibroids in more detail, magnetic resonance imaging (MRI) has remained the gold standard [18]. However, in daily practice, and as treatment with thermal ablative techniques is increasingly performed by gynecologists, ultrasound is emerging as a method with obvious advantages such as possibility of pre-operative, peri-operative and post-operative evaluation [19]. In addition, MRI is both expensive and often scarce even in high resource countries [20] and it might pose logistical problems in an outpatient setting.
The microwave antenna is easily visible with ultrasound which allows for ultrasound guided placement using established methods for needle guidance. During the emission of microwaves, gray-scale ultrasound will show hyperechoic microbubbles starting from the emission point and spreading to cover the whole ablation zone [21]. In addition to gray-scale ultrasound, contrast enhanced ultrasound (CEUS) will show any residual vascularization of treated fibroids in real time and thus give the opportunity of repeated ablation if needed [21,22]. This is an important factor in improving ablation results.
Accuracy and generalizability of peri-operative CEUS have been established [22] but the aim of this study was to reproduce the results and to investigate if it can be used for accurate estimation of the ablation volume in a clinical setting where MRI is not available. Long-term follow up at 6 months was used to investigate the frequency of recurrence of ablated fibroids as a previous study from the same institution showed regrowth of successfully treated lesions [23], even though this is not very well described in the literature.
Methods
The study was designed as a prospective observational trial. Seventeen consecutive women (willing to comply with protocol of the sub-study) who participated in a larger study at Danderyd Hospital, Sweden, from 01/01/2020 to 31/08/2023 were included. The parent study was aiming to evaluate the effect of ultrasound guided MWA of symptomatic uterine fibroids in a larger population.
All patients suffered from heavy menstrual bleeding, but many also had other symptoms such as pelvic pressure. They were premenopausal and had no wish for future pregnancies and were examined with vaginal ultrasound before inclusion. There was no limitation as to size, number and location of the fibroids. Included participants in the sub-study underwent contrast enhanced MRI before, within a week of ablation and again after 6 months, in addition to the ultrasound evaluation offered per protocol within the main study.
The ablation volume and the ablation ratio (calculated as the non-enhanced volume post-ablation divided by pre-ablation volume) measured by MRI were compared to the peri-operative CEUS (pre- and post-ablation). The ablation ratio was also correlated with the FIGO classification decided by MRI, as this has been shown to affect the treatment effect [5,14]. Long-term results were evaluated with MRI after 6 months, to investigate the frequency of recurrence of ablated fibroids. There was no CEUS performed at 6 months clinical follow-up, as it was not feasible in the gynecology office where the patients were seen.
The number of patients included in the sub-study was chosen to match a previous study from the same institution [24], as it was considered adequate for a feasibility study aimed at exploring the potential of CEUS for pre- and post-ablation assessment. An added inclusion criterium in the sub-study was willingness to comply with protocol including three MRI scans, and exclusion criterium was any condition entailing contraindication to contrast enhanced MRI. Suitable patients were informed orally and in writing by their responsible gynecologist and signed informed consent for the main study and the sub-study.
MRI
For the MRI scans, a series of standard T1-weighted, T2-weighted and contrast enhanced T1-weighted imaging were performed with a 3T SIGNATM from GE HealthCare. On the pre-operative MRI, the individual volume of the up to three largest fibroids was recorded, as well as the vascularization (i.e., contrast enhancement, yes/no). Fibroid volume was calculated according to the formula width x height x length x 0.52. FIGO classification of the fibroids was used to categorize fibroid location into 3 groups: 1) FIGO 1–2: submucosal, 2) FIGO 3–5 including hybrid: intramural 3) FIGO 6: subserosal. Within a week and six months from the day of ablation, the ablated volume was recorded with MRI. At six months, any sign of regrowth of the treated fibroid(s) was noted. The MRI images were read by a senior consultant in radiology with 10 years of experience of gynecological MRI and if needed discussed with a second radiologist until a consensus agreement was reached.
Ultrasound-guided MWA
Ultrasound (LOGIQ 10, GE HealthCare) was used for peri-operative imaging and needle guidance. Peri-operatively, the fibroid volume (width x height x length x 0.52) and vascularization were recorded with abdominal CEUS (2,4 ml SonoVue®, Bracco, Milan, Italy). At the end of the operation, the ablated (i.e., the non-perfused) volume was recorded in the same way and used for comparison with MRI within a week of ablation
With the patient under general anesthesia in the supine position the microwave antenna (Emprint™ Ablation Generator with Thermosphere™ Technology, Medtronic, Minneapolis, USA) was inserted into the fibroid percutaneously until the tip was 0.5–1 cm from the periphery of the fibroid. The distance was chosen depending on surrounding tissue, to avoid damage to heat-sensitive structures outside of the uterus. The microwave effect used was 60–100 Watts, depending on the size and location of the fibroid. The effect was monitored by ultrasound, and the ablation was stopped when the microbubbles, known to roughly represent the ablation zone, covered the whole fibroid or the part of the fibroid that was suitable for ablation. In larger lesions the antenna was pulled back and/or reinserted as necessary to cover the whole lesion. If CEUS showed signs of remaining vascularization in the treated fibroid, the procedure was repeated if safe.
The microwave ablations were carried out by an interventional radiologist with 10+ years of interventional experience.
Statistical analysis
The agreement between ablated volumes per fibroid, obtained from CEUS and contrast enhanced MRI, was analyzed by using intraclass correlation coefficient (ICC) for absolute agreement and Bland-Altman regression. A p-value <0.05 was considered significant. Statistical analyses were performed using SPSS version 29 (IBM corporation, Armonk, New York, US).
Results
A total of 17 women with symptomatic fibroids were included in the sub-study. Two fibroids were treated in two women and as every lesion was considered as independent, a total of 19 fibroids were analyzed. The pre- and post-ablation volumes within a week measured on contrast enhanced MRI were compared with the results of the peri-operative CEUS. The mean time between ablation and post-operative MRI was 5.5 days, median 4 days. Fig 1 shows an example of what the ablation zones might look like with CEUS and MRI respectively. Two patients were lost to follow-up at 6 months. Age at first ablation (mean 45 years), number of ablated fibroids, FIGO classification and pre-ablation volumes are shown in Table 1. All fibroids were interpreted as being well vascularized pre-treatment, with MRI and CEUS alike. Four patients (24%) had fibroids that were not ablated due to small size or difficult location, in addition to the treated fibroid(s). The fibroids were treated with 60–100 Watts during 8–35.5 minutes, depending on size and location. Due to remaining vascularization on immediate post-operative CEUS, 6 fibroids out of 19 (32%) had an additional ablation in the same session. Table 2 shows the ablated volume and ablation ratio measured by CEUS and MRI. The mean of fibroid diameter and the mean volume before ablation was 6.56 + /- 1.63 cm and 156 + /- 102 ml respectively. The mean ablated volume (i.e., the avascular volume) was 104 + /- 60 ml (105 ml measured by MRI and 103 ml measured by CEUS). The mean ablation ratio (= volume of the non-enhanced field in the fibroid after ablation/ volume of the fibroid before ablation x 100) measured by ultrasound and MRI was 76% and 80%, respectively. The mean ablation ratio was 90%, 71% and 78% respectively for FIGO groups 1–3 which was not a significant difference, but the sample size is too small for generalization. For four fibroids the ablation volume was greater than the fibroid treated, due to ablation of some normal myometrium around the fibroid.
For the ultrasound scan a transabdominal probe is used in the axial plane.
The volumes obtained from CEUS and MRI were plotted and a regression line was drawn indicating a linear trend between the two groups of measurements (Fig 2). The intraclass correlation coefficient (ICC) was calculated to be 0.965 (95% confidence interval 0.909, 0.987).
The interclass correlation coefficient (ICC) was 0.965 and the agreement is considered good if it is > 0.75.
The difference between, and the mean of, ablation volumes measured by CEUS (immediately after ablation) and MRI (within a week from ablation) were calculated, and correlation (difference, mean) was 0.008 (p = 0.932), indicating that the difference is independent of the mean and that no proportional bias is present. The data could therefore be used for the Bland-Altman graph plotted in Fig 3. The mean difference between volumes was –1.68 ml and standard deviation 22.60 ml, giving the limits of agreement (−45.98, 42.61). The graph shows that all points but one fall within the 95% limit of agreement. The maximum difference between the groups (ablation volumes measured by CEUS and MRI) was 63 ml. Within the limits of agreement, the largest difference was 38 ml.
The black line constitutes the mean.
The MRI follow-up at 6 months showed that 17 out of 19 (89%) of the ablation zones had decreased in size as expected. Regrowth of vascularized fibroid tissue, defined as recurrent contrast enhancement in treated lesions, was seen in 8 fibroids (47.1%, 95% CI: 26.2%–69.0%).
Discussion
The aim of this study was to investigate if the use of contrast enhanced ultrasound (CEUS) in the treatment of microwave ablation of uterine fibroids can be used as an alternative to MRI for evaluation of the ablation volume, at the time of the procedure. Even though MRI is the gold standard for imaging of fibroids, it is a bottleneck in many countries and hospitals, in addition to its high cost and obvious disadvantages such as an impossibility of using it for intra-operative evaluation. Thus, in the mother study of 123 women, ultrasound alone was used for pre- peri- and post-ablation monitoring to save time and money [25]. Using CEUS for immediate evaluation in a peri-operative setting enables visualization of any remaining vascularization within treated fibroids which can then be complimentary ablated in the same session. In our study, 6 fibroids out of 19 (32%) had an immediate additional ablation after the evaluation with CEUS. This must be considered a major advantage of CEUS over MRI. Our results, as presented in Table 2, are in agreement with previous studies from China. In a study from 2014, Lei et al. [22] found good correlation between ablation volumes measured by CEUS and post-operative contrast enhanced MRI, and Zhang et al. [26] concluded that CEUS could even be advantageous over MRI as it gives reliable real-time information on the perfusion and possible viable fibroid residue within the ablated area. Our study confirms these results in a different setting adding to the generalizability of the method, which is of importance before implementing it in a clinical practice outside of Asia.
The agreement between ablated volumes obtained from CEUS and contrast enhanced MRI was analyzed by the intraclass correlation coefficient (ICC) and Bland-Altman regression.
The intraclass correlation coefficient (ICC) was calculated to be 0.965 (95% confidence interval 0.909, 0.987). The agreement can be considered as good if ICC is higher than 0.75 [27]. The Bland-Altman analysis is used to evaluate the agreement between two measurement techniques. It allows for identification of any systematic difference or possible outliers and can help compare a new measurement method to a reference standard. In our study the mean difference between volumes was –1.68 ml and the standard deviation 22.60 ml, giving the limits of agreement (−45.98, 42.61) as depicted in Fig 3. Within the limits of agreement, the largest absolute difference was 38 ml. It is a larger difference (36.5% of the mean ablated volume) than in a similar study by Lei et al. (11.8%) [22]. This could have many explanations, one being size. The mean diameter of treated fibroids in this study was 6.56 + /- 1.63 cm and mean volume was 156 ml before ablation. This is higher than in the study by Lei et al (mean diameter 5.56 + /-1,26 cm, mean volume not specified). The largest difference between the ablation volume measured by CEUS and MRI in this study was 63 ml in a clear outlier. It was noted in the second largest fibroid with a pre-operative volume of 298 ml (MRI). For large lesions, ultrasound may have the limitation of smaller field of view than MRI. In addition, a characteristic of fibroids is echo-shadowing, which further complicates the measurement of large fibroids and for these lesions MRI is a more precise method. Due to technical reasons, it is also difficult to achieve spherical ablation zones larger than 5–6 cm in diameter [28], and for very large fibroids other treatments can be considered as more appropriate. Also, the bubbles present in the ablation zone immediately after treatment might obscure the ultrasound image and make correct measurements more difficult to obtain. This could be avoided by delaying the post-operative ultrasound, but in clinical practice with a patient under sedation or general anesthesia this is usually not an option and would also prevent the possibility of repeated treatment of residual areas of vascularization within the fibroid in the same session, as discussed above. For logistical reasons, the post-operative MRI could not be performed immediately following the ablation. The mean time between ablation and post-operative MRI was 5.5 days (median 4 days). A study of microwave ablation of liver tumors by Alzubaidi et al. [29] found that the ablation zone expanded during the first 24 hours after treatment, likely due to evolving edema and tissue response. None of our patients had the post-operative MRI performed within the first 48 hours. Thus, remaining post-operative expansion of the non-enhancing zone is not likely to affect results but as discussed previously, large lesions are difficult to evaluate with ultrasound. For most cases (68%), the absolute difference between volume measured by CEUS and MRI was less than 10 ml and in conclusion we argue that the CEUS evaluation can be performed intra-operatively and that the two methods agree sufficiently to support the use of CEUS in our clinical setting of outpatient treatment.
If CEUS replaces MRI for evaluation of MWA treatment of uterine fibroids, adequate training of the provider in use of CEUS is necessary We used a commercially available product for CEUS and a well-established technique for fibroid evaluation. We therefore believe that our results are generalizable to any setting with corresponding experience with ultrasound evaluation with CEUS.
In countries where CEUS is not available, gray-scale ultrasound will be able to assist the ablation treatment but not to reliably assess the avascular ablation zone.
The ablation ratio was high at 76–80% (measured by ultrasound and MRI, respectively). Notably, not all fibroids can be completely ablated safely due to location with heat sensitive tissues in close proximity to the fibroid, but a reduction of vascularized volume may still reduce symptoms, as shown in previous studies [25]. In ablation of malignant liver tumors, minimal ablative margin (MAM) is important for improving local control following thermal ablation [30]. At least 5 mm of healthy tissue surrounding the tumor should be covered by the ablation zone, to reduce the risk of local recurrence. When it comes to benign lesions like fibroids safety comes first, but if there are no heat sensitive structures at risk the aim should be complete ablation. For four fibroids the ablation ratio exceeded 100%. All were classified as FIGO group 1 (submucosal) which, due to location, allowed more aggressive treatment and thus an ablation zone covering more than the actual lesion. More commonly in this study, the outermost rim of the fibroid was left to avoid harm.
Four out of 17 patients (24%) had more fibroids than the 1–2 treated, and depending on the woman’s age at time of treatment these lesions can potentially grow to symptomatic size over time [31]. Also, in almost 50% of ablated patients (47.1%, 95% CI: 26.2%–69.0%) there was regrowth of vascularized fibroid tissue defined as recurrent contrast enhancement in treated lesions, even though the ablation volume (i.e., the avascular part of the fibroid) continued to shrink. The wide confidence interval due to the small sample size makes it difficult to generalize the results, but we proposed that when a rim of the fibroid must be left untreated due to the safety of surrounding structures as discussed above, there is a potential for regrowth. The same results were shown in a previous study from the same institution [23] but has not, to our knowledge, been well described in the literature. For tumors treated with thermal ablation in other organs, re-ablation of local recurrence is an established method [32,33]. We propose that re-ablation could be an option also for uterine fibroids showing signs of regrowth or inadequate ablation results, especially since the patient acceptance of the method is high [24]. Treatment methods like uterine artery embolization (UAE) may be as effective [24] but not as well tolerated as MWA [34]. Our intention is to investigate the efficacy of reablation further in future studies.
A limitation to the study is the lack of blinding and inter-/intra-observer reliability assessment of imaging (CEUS and MRI). However, as the volumes measured by CEUS and MRI were compared after all data were collected and there was an option of discussion among colleagues until consensus was reached, we believe the effect on the results is limited.
Another general limitation to the study is the small number of cases. It is to be considered as a feasibility study to evaluate if intraoperative CEUS is a useful method for pre- and post-evaluation of microwave ablation of fibroids in an outpatient setting, without the support of MRI and outside of Asia. Also, even with the current sample size, the correlation between CEUS and MRI was strong enough to support CEUS for intraoperative estimation and guidance.
In conclusion, we believe that ultrasound and specifically CEUS could replace MRI for evaluation of the ablation zone in a clinical setting. Microwave ablation seems to be an effective method for treatment of symptomatic fibroids with a mean ablation rate of 76–80% in this study. The 6-month follow-up indicates that regrowth of fibroids will occur even when initial treatment was technically successful, suggesting that re-ablation could be considered.
References
- 1. Khan AT, Shehmar M, Gupta JK. Uterine fibroids: current perspectives. Int J Womens Health. 2014;6:95–114. pmid:24511243
- 2. Donnez J, Dolmans M-M. Uterine fibroid management: from the present to the future. Hum Reprod Update. 2016;22(6):665–86. pmid:27466209
- 3. Aarts JWM, Nieboer TE, Johnson N, Tavender E, Garry R, Mol BWJ, et al. Surgical approach to hysterectomy for benign gynaecological disease. Cochrane Database Syst Rev. 2015;2015(8):CD003677. pmid:26264829
- 4. Morris JM, Liang A, Fleckenstein K, Singh B, Segars J. A systematic review of minimally invasive approaches to uterine fibroid treatment for improving quality of life and fibroid-associated symptoms. Reprod Sci. 2023;30(5):1495–505. pmid:36401073
- 5. Mauri G, Nicosia L, Xu Z, Di Pietro S, Monfardini L, Bonomo G, et al. Focused ultrasound: tumour ablation and its potential to enhance immunological therapy to cancer. Br J Radiol. 2018;91(1083):20170641. pmid:29168922
- 6. Quinn SD, Gedroyc WM. Thermal ablative treatment of uterine fibroids. Int J Hyperthermia. 2015;31(3):272–9. pmid:25815582
- 7. Keung JJ, Spies JB, Caridi TM. Uterine artery embolization: a review of current concepts. Best Pract Res Clin Obstet Gynaecol. 2018;46:66–73. pmid:29128204
- 8. Toor SS, Jaberi A, Macdonald DB, McInnes MDF, Schweitzer ME, Rasuli P. Complication rates and effectiveness of uterine artery embolization in the treatment of symptomatic leiomyomas: a systematic review and meta-analysis. AJR Am J Roentgenol. 2012;199(5):1153–63. pmid:23096193
- 9. Kaump GR, Spies JB. The impact of uterine artery embolization on ovarian function. J Vasc Interv Radiol. 2013;24(4):459–67. pmid:23384832
- 10. Kim C-W, Shim HS, Jang H, Song YG. The effects of uterine artery embolization on ovarian reserve. Eur J Obstet Gynecol Reprod Biol. 2016;206:172–6. pmid:27697621
- 11. Arellano RS. What’s new in percutaneous ablative strategies for hepatocellular carcinoma and colorectal hepatic metastases? 2020 update. Curr Oncol Rep. 2020;22(10):105.
- 12. Bradley LD, Pasic RP, Miller LE. Clinical performance of radiofrequency ablation for treatment of uterine fibroids: systematic review and meta-analysis of prospective studies. J Laparoendosc Adv Surg Tech A. 2019;29(12):1507–17. pmid:31702440
- 13. Lim KJ, Yoon DY, Kim J H, Cho YK, Kim HJ, Ahn HS, et al. Percutaneous radiofrequency ablation for symptomatic uterine leiomyomas: a systematic review and meta-analysis. Clin Exp Obstet Gynecol. 2016;43(5):643–9. pmid:30074311
- 14. Iversen H, Lenz S, Dueholm M. Ultrasound-guided radiofrequency ablation of symptomatic uterine fibroids: short-term evaluation of effect of treatment on quality of life and symptom severity. Ultrasound Obstet Gynecol. 2012;40(4):445–51. pmid:22378629
- 15. Liu H, Zhang J, Han Z-Y, Zhang B-S, Zhang W, Qi C-S, et al. Effectiveness of ultrasound-guided percutaneous microwave ablation for symptomatic uterine fibroids: a multicentre study in China. Int J Hyperthermia. 2016;32(8):876–80. pmid:27405972
- 16. Zhang J, Feng L, Zhang B, Ren J, Li Z, Hu D, et al. Ultrasound-guided percutaneous microwave ablation for symptomatic uterine fibroid treatment--a clinical study. Int J Hyperthermia. 2011;27(5):510–6. pmid:21756048
- 17. Zhang Q, Liang X, Chen Z. An updated review of thermal ablation technology for uterine fibroids and adenomyosis: focusing on protecting fertility. Int J Womens Health. 2024;16:1551–63. pmid:39346931
- 18. Rajan DK, Margau R, Kroll RR, Simons ME, Tan KT, Jaskolka JD, et al. Clinical utility of ultrasound versus magnetic resonance imaging for deciding to proceed with uterine artery embolization for presumed symptomatic fibroids. Clin Radiol. 2011;66(1):57–62. pmid:21147300
- 19. Benacerraf BR, Abuhamad AZ, Bromley B, Goldstein SR, Groszmann Y, Shipp TD, et al. Consider ultrasound first for imaging the female pelvis. Am J Obstet Gynecol. 2015;212(4):450–5. pmid:25841638
- 20. Murali S, Ding H, Adedeji F, Qin C, Obungoloch J, Asllani I, et al. Bringing MRI to low- and middle-income countries: directions, challenges and potential solutions. NMR Biomed. 2024;37(7):e4992. pmid:37401341
- 21. Wang F, Zhang J, Han Z-Y, Cheng Z-G, Zhou H-Y, Feng L, et al. Imaging manifestation of conventional and contrast-enhanced ultrasonography in percutaneous microwave ablation for the treatment of uterine fibroids. Eur J Radiol. 2012;81(11):2947–52. pmid:22341698
- 22. Lei F, Jing Z, Bo W, Dongmei H, Zhencai L, Xue J, et al. Uterine myomas treated with microwave ablation: the agreement between ablation volumes obtained from contrast-enhanced sonography and enhanced MRI. Int J Hyperthermia. 2014;30(1):11–8. pmid:24286236
- 23. Beermann M, Jonsdottir G, Cronisoe A, Hasselrot K, Kopp Kallner H. Long term follow-up of uterine fibroids treated with microwave ablation: an up to 3-year observational study of volume, regrowth, and symptoms. Int J Hyperthermia. 2022;39(1):1158–63. pmid:36049885
- 24. Jonsdottir G, Beermann M, Lundgren Cronsioe A, Hasselrot K, Kopp Kallner H. Ultrasound guided microwave ablation compared to uterine artery embolization treatment for uterine fibroids - a randomized controlled trial. Int J Hyperthermia. 2022;39(1):341–7. pmid:35134317
- 25. Jonsdottir G, Beermann M, Lanz E, Nikodell A, Cronsioe A, Hasselrot K, et al. Ultrasound guided microwave ablation treatment of uterine fibroids: clinical response and patient acceptability. Acta Obstet Gynecol Scand. 2025;104(2):350–6. pmid:39697096
- 26. Zhang Y, Zhang M, Fan X, Mao D. Contrast-enhanced ultrasound is better than magnetic resonance imaging in evaluating the short-term results of microwave ablation treatment of uterine fibroids. Exp Ther Med. 2017;14(5):5103–8. pmid:29201222
- 27. Lee J, Koh D, Ong CN. Statistical evaluation of agreement between two methods for measuring a quantitative variable. Comput Biol Med. 1989;19(1):61–70. pmid:2917462
- 28. Hui TCH, How GY, Chim MSM, Pua U. Comparative study of ablation zone of EMPRINT HP microwave device with contemporary 2.4 GHz microwave devices in an ex vivo porcine liver model. Diagnostics (Basel). 2023;13(16):2702. pmid:37627962
- 29. Alzubaidi S, Wallace A, Naidu S, Knuttinen MG, Kriegshauser SJ, Oklu R, et al. Single-arm prospective study comparing ablation zone volume between time zero and 24 h after microwave ablation of liver tumors. Abdom Radiol (NY). 2024;49(9):3136–42.
- 30. Wang X, Sofocleous CT, Erinjeri JP, Petre EN, Gonen M, Do KG, et al. Margin size is an independent predictor of local tumor progression after ablation of colon cancer liver metastases. Cardiovasc Intervent Radiol. 2013;36(1):166–75. pmid:22535243
- 31. Tinelli A, Sparic R, Kadija S, Babovic I, Tinelli R, Mynbaev OA, et al. Myomas: anatomy and related issues. Minerva Ginecol. 2016;68(3):261–73. pmid:26785282
- 32. Camisassi N, Mauri G, Vigna PD, Bonomo G, Varano GM, Maiettini D, et al. Local recurrence of renal cell carcinoma successfully treated with fusion imaging-guided percutaneous thermal ablation. Ecancermedicalscience. 2020;14:1070. pmid:32728386
- 33. McGhana JP, Dodd GD 3rd. Radiofrequency ablation of the liver: current status. AJR Am J Roentgenol. 2001;176(1):3–16. pmid:11133529
- 34. Ganguli S, Faintuch S, Salazar GM, Rabkin DJ. Postembolization syndrome: changes in white blood cell counts immediately after uterine artery embolization. J Vasc Interv Radiol. 2008;19(3):443–5. pmid:18295706