Comparison of clip and needle electrodes for multifrequency bioimpedance analysis in anesthetized dogs

  • Jun Tamura ,

    Roles Conceptualization, Formal analysis, Investigation, Validation, Writing – original draft

    j-tamura@vetmed.hokudai.ac.jp

    Affiliation Faculty of Veterinary Medicine, Veterinary Teaching Hospital, Hokkaido University, Sapporo, Hokkaido, Japan

  • Takaharu Itami,

    Roles Conceptualization, Investigation, Resources

    Affiliation Department of Veterinary Science, School of Veterinary Medicine, Rakuno Gakuen University, Ebetsu, Hokkaido, Japan

  • Keiko Kato,

    Roles Investigation

    Affiliation Department of Veterinary Science, School of Veterinary Medicine, Rakuno Gakuen University, Ebetsu, Hokkaido, Japan

  • Chihiro Sugita,

    Roles Investigation

    Affiliation Department of Veterinary Science, School of Veterinary Medicine, Rakuno Gakuen University, Ebetsu, Hokkaido, Japan

  • Norihiko Oyama,

    Roles Conceptualization, Formal analysis, Writing – review & editing

    Affiliation Faculty of Veterinary Medicine, Veterinary Teaching Hospital, Hokkaido University, Sapporo, Hokkaido, Japan

  • Kazuto Yamashita

    Roles Conceptualization, Resources, Supervision

    Affiliation Department of Veterinary Science, School of Veterinary Medicine, Rakuno Gakuen University, Ebetsu, Hokkaido, Japan

Abstract

Multifrequency bioimpedance analysis provides resistance, reactance, and phase angle measurements that are used to characterize body fluid distribution and cellular electrical properties. However, studies reporting on body composition assessment using multifrequency bioimpedance analysis in dogs remain limited. Therefore, we aimed to evaluate the measurement characteristics of trunk multifrequency bioimpedance analysis in dogs and assess the influence of body position and electrode type and placement. We anesthetized seven adult Beagle dogs and performed trunk multifrequency bioimpedance analysis at 5, 50, and 250 kHz using clip and needle electrodes in sternal and left lateral recumbency at three electrode paths. For each condition, 150 repeated measurements were obtained and summarized as a mean value. Repeatability and inter-individual variability were assessed using coefficients of variation. Agreement between electrode types was assessed using Bland–Altman analysis and correlation analysis, and the effects of electrode type and body position were evaluated using linear mixed-effects models with dogs as a random effect. Measurements were feasible under all conditions and showed low intra-individual variability. Inter-individual variability was generally 10–15%, but higher variability was observed for reactance and phase angle at low frequencies. Mean biases between clip and needle electrodes were small for most parameters; however, reactance and phase angle at 5 kHz showed weaker correlations and wider limits of agreement. Across models, between-dog differences accounted for most variability, whereas body position influenced several parameters when the stifle was included in the electrode path. These data indicate that clip electrodes yield measurements with small mean biases relative to those obtained using needle electrodes under controlled conditions in anesthetized dogs; however, the limits of agreement varied among parameters, and wider differences observed at low frequencies suggest that the two electrode types should not be considered fully interchangeable. Accordingly, electrode placement and body position should be standardized for experimental comparison and carefully controlled when monitoring longitudinal changes within individual dogs.

Introduction

Bioimpedance analysis is a minimally invasive technique that evaluates the electrical impedance of biological tissues and provides parameters such as resistance, reactance, and phase angle [1,2]. Resistance reflects opposition to electrical current flow through conductive body fluids and is influenced by water amount and distribution of water. Reactance and phase angle reflect capacitive properties related to cell membranes and are influenced by cellular integrity and tissue composition.

Compared with single-frequency bioimpedance analysis, multifrequency bioimpedance analysis (MF-BIA), which measures impedance across a range of frequencies, provides more comprehensive information regarding body fluid distribution and cellular function [35]. At low frequencies, electrical current primarily travels through the extracellular space because cell membranes act as insulators. Therefore, resistance at low frequencies mainly reflects extracellular fluid status. At higher frequencies, the current penetrates cell membranes, enabling the assessment of both extracellular and intracellular compartments.

In human medicine, MF-BIA using clip electrodes is widely applied as a non-invasive and cost-effective tool for assessing body composition, fluid balance, nutritional status, and prognosis in various clinical settings, including chronic kidney disease [6,7], heart failure [7], sepsis [8], cancer [9], and malnutrition [5]. In domestic dogs (Canis lupus familiaris), single-frequency bioimpedance analysis using various types of electrodes, including needles, clips, adhesives, and contact electrodes, is commonly used to estimate total body water, body fat composition, and fluid redistribution after blood donation [1016]. However, studies reporting on body composition assessment using MF-BIA in dogs remain limited [17,18]. Clinically evident peripheral oedema has been reported in critically ill dogs with systemic inflammatory conditions, highlighting the potential clinical importance of objectively assessing alterations in body fluid distribution [19]. Nevertheless, fundamental data on the characteristics of MF-BIA measurements in dogs and the influence of measurement conditions, such as electrode placement site, electrode type, and body position are lacking.

The objectives of this study were (i) to characterize trunk MF-BIA measurements in anesthetized dogs and quantify repeatability and inter-individual variability across various body positions and electrode placement sites, and (ii) to evaluate clip electrodes as a non-invasive attachment method in comparison with needle electrodes, (which are commonly used in experimental settings), to support future clinical applications. We hypothesized that clip electrodes would demonstrate acceptable repeatability and generally good agreement with needle electrodes under controlled anesthetized conditions and that electrode placement and body position would contribute to between-condition differences.

Materials and methods

Ethics statement

This study was approved by the Animal Care and Use Committee of Rakuno Gakuen University (approval No.: VH23B9) and was reported in accordance with the ARRIVE guidelines 2.0.

Animals

Seven intact adult beagle dogs (4 males and 3 females), aged 7–9 years and weighing 8.9–15.5 kg [12.9 ± 2.2 kg (mean ± standard deviation)], were included. The dogs were obtained from a licensed vendor (Hokudo Co., Ltd., Japan) and were considered clinically healthy based on physical examination, complete blood cell count, and biochemistry panel, except for mild hypoalbuminemia in two dogs and elevated alanine aminotransferase in one dog. Food was withheld for at least 12 h before drug administration; however, the dogs had free access to water 30 min before treatment.

Anesthesia and monitoring

Anesthesia was induced by mask induction using sevoflurane in oxygen, followed by endotracheal intubation. Anesthesia was maintained with sevoflurane in oxygen (2.5%). The dogs were positioned in sternal recumbency on a conductive mat and left undisturbed for 30 min to allow stabilization and static electricity discharge.

Rocuronium was administered (0.5 mg/kg IV) followed by constant rate infusion (0.5 mg/kg/h) via an intravenous catheter placed in the left cephalic vein. Physiologic variables were monitored using a multiparameter anesthetic monitor. Mechanical ventilation was adjusted to maintain end-tidal carbon dioxide at 35–40 mmHg (4.7–5.3 kPa). Lactated Ringer’s solution (3 mL/kg/h) was administered intravenously throughout anesthesia. After completion of measurements, sugammadex (4 mg/kg IV) was administered, and the dogs recovered uneventfully.

MF-BIA measurement protocol

MF-BIA was performed using a tetrapolar bioimpedance spectrometer (InBody M20, InBody Japan Inc.), which measured resistance, reactance, and phase angle at 5, 50, and 250 kHz. Measurements were conducted in two body positions (sternal and left lateral recumbency) on a non-conductive silicone mat using clip electrodes and needle electrodes.

To minimize limb contact in lateral recumbency, a cotton towel was placed between the left and right limbs to avoid contact and crossing. Measurements followed a fixed sequence: sternal recumbency with clip electrodes, lateral recumbency with clip electrodes, sternal recumbency with needle electrodes, and lateral recumbency with needle electrodes. A 15-min stabilization period was allowed after each change in body position. Hair at the electrode sites was clipped closely and cleaned with alcohol. For clip electrodes, conductive electrocardiography cream (Keratin Cream; Fukuda Denshi) was applied to improve contact. Needle electrodes consisted of 25-gauge needles inserted subcutaneously to a length of 1 cm.

For each condition, measurements were obtained sequentially across three electrode paths: from the right elbow to the right stifle (ESt), from the level of the first thoracic spinous process to the right stifle (TSt), and from the vertex of the head to the level of the sacral spinous process (HSc). For each path, MF-BIA was measured 150 times at 50 ms intervals and summarized as the mean value for analysis.

Cole–Cole modeling and data quality control

Based on the Cole–Cole model, resistance at zero frequency (R0), resistance at infinite frequency (Rinf), maximum reactance (Xmax), and maximum phase angle (PAmax) were calculated [5]. Measurements were excluded if the resistance at the center of the Cole–Cole circle was not within the range defined by resistance values at 5 and 250 kHz [5]. Randomization and blinding were not applied because all animals underwent measurements in a fixed sequence under identical conditions, and outcomes were recorded directly from instrument output.

Statistical analysis

Descriptive data are presented as mean ± standard deviation. Repeatability was assessed by calculating the coefficient of variation from each set of 150 repeated measurements (standard deviation/mean). Inter-individual variability was evaluated using the coefficient of variation calculated across dogs for each condition. Agreement between clip and needle electrodes was assessed using Bland–Altman analysis. For each dog, body position, and electrode path, the mean value obtained with clip electrodes was paired with the corresponding mean value obtained with needle electrodes. Differences were calculated as clip minus needle measurements. The bias, 95% limits of agreement, their corresponding 95% confidence intervals (CI), and percentage limits of agreement were calculated. Correlation coefficients and associated P values were estimated using regression analysis. Linear mixed-effects models were used to evaluate fixed effects (electrode type and body position) on MF-BIA variables, with dogs included as a random effect. A P value <0.05 was considered statistically significant. Analyses were performed using JMP Pro version 18.2.0 (SAS Institute Inc., Cary, NC, USA).

Results

A total of 1800 measurements were completed for each dog (150 measurements × 3 electrode paths × 4 conditions). MF-BIA measurements were performed at 53.3 ± 10.5, 69.0 ± 10.7, 84.0 ± 11.5, and 103.0 ± 13.9 min after endotracheal intubation in the sternal recumbency with clip electrodes, lateral recumbency with clip electrodes, sternal recumbency with needle electrodes, and lateral recumbency with needle electrodes groups, respectively. Because the resistance at the center of the Cole–Cole circle fell outside the range defined by the resistance values at 5 and 250 kHz, the ESt and TSt data from one dog in the sternal recumbency with clip electrodes group and the TSt data from a different dog in the lateral recumbency with clip electrodes group were excluded from the analysis. MF-BIA parameters and analysis status for individual dogs are provided in S1 Table.

Repeatability and inter-individual variability are summarized in Tables 1 and 2.

thumbnail
Table 1. Repeatability of MF-BIA measurements expressed as coefficients of variation (%).

https://doi.org/10.1371/journal.pone.0355338.t001

thumbnail
Table 2. Inter-individual variability of MF-BIA measurements expressed as coefficients of variation (%).

https://doi.org/10.1371/journal.pone.0355338.t002

Intra-individual coefficients of variation were <1% for resistance, reactance, and phase angle under all conditions. Inter-individual coefficients of variation for ESt and HSc were generally around 10–15%, with higher values observed for reactance at 5 and 50 kHz and for phase angle at 5 kHz. For TSt, inter-individual coefficients of variation for resistance and reactance frequently exceeded 15%, particularly in lateral recumbency.

MF-BIA measurements across body position, electrode path, and electrode type are shown in Table 3.

thumbnail
Table 3. MF-BIA measurements obtained in anesthetized dogs.

https://doi.org/10.1371/journal.pone.0355338.t003

Electrical characteristics derived from Cole–Cole modeling are provided in S2 Table. Across conditions, Xmax and PAmax were consistently observed between 50 and 250 kHz.

Agreement between electrode types is summarized in Table 4, and the corresponding Bland–Altman plots are presented in S1S3 Figs. A total of 39 paired observations were included for each frequency after exclusion of three measurements that did not meet the predefined Cole–Cole quality criterion.

thumbnail
Table 4. Agreement between clip and needle electrode measurements assessed by Bland–Altman analysis.

https://doi.org/10.1371/journal.pone.0355338.t004

Reactance and phase angle at 5 kHz showed the weakest agreement, with wide limits of agreement. For resistance at 5, 50, 250 kHz and for reactance and phase angle at 50 and 250 kHz, mean bias was small and correlations were generally strong, although the percentage limits of agreement for reactance at 50 and 250 kHz were wider than those for resistance and extended beyond 27% on the positive side.

Results of linear mixed-effects models are provided in S3 Table. In HSc, between-dog variation accounted for most of the variance (86.4–93.0%) for most variables, and neither electrode type nor body position showed significant effects except for low-frequency reactance and phase angle. In ESt and TSt, between-dog variation also accounted for most of the variance; however, body position affected several MF-BIA parameters, whereas electrode type did not show significant effects in most models.

Discussion

This study characterized trunk MF-BIA measurements in anesthetized dogs and compared measurements obtained using clip and needle electrodes under controlled conditions. Across various electrode paths and body positions, both electrode types produced feasible measurements with low intra-individual variability. For most frequencies and parameters, clip and needle electrodes showed small mean biases, although the limits of agreement varied among parameters. These findings support the potential use of clip electrodes as a non-invasive alternative under standardized measurement conditions, but the two electrode types should not be considered fully interchangeable.

Several measurements were excluded during Cole–Cole processing, indicating that quality control based on the impedance spectrum can identify datasets that may reflect unstable electrode contact or other artifacts [5]. In the present study, excluded datasets were observed in both sternal and lateral recumbency when clip electrodes were used, consistent with the possibility that contact variability can affect spectral consistency. Careful attention to electrode attachment and verification of spectral plausibility may therefore improve data quality. Electrical property analysis based on the Cole–Cole model showed that the maximum reactance and phase angle consistently appeared between 50 and 250 kHz under all conditions. In contrast, reactance and phase angle at 5 kHz were relatively low and exhibited substantial variability. Although these low frequency measurements are generally considered to have limited physiological relevance, our findings further suggest that they may not reliably reflect the capacitive behavior of cell membranes in dogs. Moreover, although mean biases were small for most impedance parameters, reactance and phase angle at 5 kHz showed weak correlations and wide limits of agreement.

In this study, reactance and phase angle at 250 kHz were higher than those at 50 kHz. Reactance and phase angle are commonly evaluated at 50 kHz, and their values tend to be close to the maximum at this frequency in humans [1,5,20]. An in vitro study of canine skeletal muscle also showed that reactance exhibited near-maximum values at 50 kHz [1]. Consistently, in vivo canine studies on body composition using a single-frequency bioimpedance analysis have typically been conducted at 50 kHz [1115]. Furthermore, previous studies on bioimpedance analysis in donor dogs have employed measurements of reactance and phase angle only at 50 kHz, following protocols commonly applied in human research [16]. The frequency range at which reactance or phase angle reaches its maximum value, which reflects the electrical properties of cell membranes, has been reported to vary according to species or objects [2,2123]. The findings of the present study may reflect species-specific characteristics of MF-BIA in dogs, suggesting that when evaluating reactance or phase angle in dogs, it may be preferable to perform measurements at 250 kHz rather than at 50 kHz in the current setting. In humans, trunk phase angle tends to be higher at 250 kHz than at 50 kHz when compared with hemi-body measurements, including the limbs [5,20], suggesting that this may also reflect characteristics related to electrode placement during the present study setting.

While previous studies in dogs have commonly used measurements between the elbow and stifle [13,14,1618], the present study evaluated three different electrode paths on the trunk: ESt, TSt, and HSc. Measurements were feasible at all electrode paths; however, values obtained at TSt in lateral recumbency tended to show greater inter-individual variability. Moreover, measurements at ESt and TSt may be susceptible to positional effects. In this study, a cotton towel was placed between the limbs to prevent contact and crossing, thereby minimizing any potential influence on the measured values. These differences may be partially explained by variations in segment length, cross-sectional area, and body composition at each site [1,2]. We speculate that electrode paths involving the stifle may lead to higher variability in measurements in lateral recumbency than in sternal recumbency. Therefore, sternal recumbency may be preferable to lateral recumbency for trunk MF-BIA in dogs, particularly when the stifle is included in the electrode path, to minimize measurement variability.

Resistance at 50 kHz, as well as at R0 and at Rinf, especially at the ESt and HSc sites, was consistent with previous reports [16,18], regardless of electrode type or body position. However, other studies measuring resistance in the trunk region have reported lower values, with a mean of approximately 130 Ω [17], possibly due to differences in measurement protocols or dog populations. In contrast, reactance and phase angle at 50 kHz obtained in this study were higher than those reported in previous studies [18]. At the TSt site, resistance tended to be lower than at the ESt and HSc sites, suggesting that the electrode path can affect the MF-BIA results. Factors such as breed, age, segment geometry, and body composition may contribute to inter-individual variability, making it difficult to define absolute reference values [12]. The present study demonstrated relatively high inter-individual variability, with most of the variation in measurements attributable to individual-specific factors.

Despite these sources of variability, MF-BIA showed excellent repeatability, with intra-individual coefficients of variation below 1% under controlled conditions. The effect of movement at the measurement site due to mechanical ventilation on the measured values in anesthetized dogs was considered minimal. In the present study, no significant effect of electrode type was detected on the measured values, except for reactance at 50 kHz in the TSt group. Strong correlations were observed between clip and needle electrodes for resistance (5, 50, and 250 kHz), as well as for reactance and phase angle (50 and 250 kHz). Mean biases were small, although the percentage limits of agreement commonly extended to approximately 10–20% and exceeded this range for some reactance measurements. MF-BIA using clip electrodes, which are non-invasive, may provide a useful clinical tool for case-by-case monitoring of body fluid balance and cell membrane stability, especially when evaluating trends over time in individual dogs.

This study has some limitations. The sample size was small and limited to Beagle dogs, and sex-related effects could not be evaluated. Measurements were obtained under general anesthesia with neuromuscular blockade to minimize motion; therefore, outcomes under conscious conditions remain to be evaluated for clinical translation. In addition, mild laboratory abnormalities in some dogs may have contributed to biological heterogeneity. Despite these limitations, the controlled design allowed assessment of repeatability and agreement across electrode types and conditions.

Conclusions

Trunk MF-BIA measurements obtained using clip electrodes were highly repeatable and showed small mean biases relative to those obtained using needle electrodes in anesthetized dogs, although the limits of agreement varied among parameters and were particularly wide for reactance and phase angle at 5 kHz. MF-BIA using clip electrodes may be useful for non-invasive monitoring of longitudinal trends within individual dogs; however, electrode placement site and body position influenced MF-BIA measurements and should be standardized or carefully accounted for, particularly in lateral recumbency when the stifle is included in the electrode path. Together, these findings support the potential clinical utility of MF-BIA for monitoring body fluid status and membrane integrity in dogs, while highlighting the importance of standardized measurement conditions.

Supporting information

S1 Table. Multifrequency bioimpedance parameters and analysis status for individual dogs.

Values represent the mean of 150 repeated measurements obtained at 50-ms intervals for each dog and measurement condition. R0, resistance at zero frequency; Rinf, resistance at infinite frequency; Xmax, maximum reactance; PAmax, maximum phase angle; ESt, from the elbow to the stifle; TSt, from the level of first thoracic spinous process to the stifle; HSc, from the vertex of the head to the level of sacral spinous process. NA, not available. Measurements marked “Excluded” did not meet the predefined Cole–Cole quality criterion because the resistance at the center of the Cole–Cole circle was outside the range defined by the resistance values at 5 and 250 kHz; corresponding Cole–Cole-derived parameters are shown as NA.

https://doi.org/10.1371/journal.pone.0355338.s001

(XLSX)

S2 Table. Electrical characteristics derived from multifrequency bioimpedance analysis using the Cole–Cole model.

Values are presented as mean ± standard deviation. R0, resistance at zero frequency; Rinf, resistance at infinite frequency; Xmax, maximum reactance; PAmax, maximum phase angle; ESt, from the elbow to the stifle; TSt, from the level of first thoracic spinous process to the stifle; HSc, from the vertex of the head to the level of sacral spinous process.

https://doi.org/10.1371/journal.pone.0355338.s002

(DOCX)

S3 Table. Results of linear mixed-effects model for multifrequency bioimpedance parameters.

Linear mixed-effects models were used to evaluate the effects of electrode type and body position. Effect of individual dogs (ID) was included as a random effect. ESt, from the elbow to the stifle; TSt, from the level of first thoracic spinous process to the stifle; HSc, from the vertex of the head to the level of sacral spinous process; Effect of ID, individual contribution to the total variance.

https://doi.org/10.1371/journal.pone.0355338.s003

(XLSX)

S1 Fig. Bland–Altman plots comparing resistance measurements obtained using clip and needle electrodes.

Panels A, B, and C show measurements obtained at 5, 50, and 250 kHz, respectively. Differences were calculated as clip minus needle measurements. The central solid line indicates the mean difference (bias), the dashed lines indicate the 95% limits of agreement, and the shaded areas indicate the corresponding 95% confidence intervals.

https://doi.org/10.1371/journal.pone.0355338.s004

(PNG)

S2 Fig. Bland–Altman plots comparing reactance measurements obtained using clip and needle electrodes.

Panels A, B, and C show measurements obtained at 5, 50, and 250 kHz, respectively. Differences were calculated as clip minus needle measurements. The central solid line indicates the mean difference (bias), the dashed lines indicate the 95% limits of agreement, and the shaded areas indicate the corresponding 95% confidence intervals.

https://doi.org/10.1371/journal.pone.0355338.s005

(PNG)

S3 Fig. Bland–Altman plots comparing phase angle measurements obtained using clip and needle electrodes.

Panels A, B, and C show measurements obtained at 5, 50, and 250 kHz, respectively. Differences were calculated as clip minus needle measurements. The central solid line indicates the mean difference (bias), the dashed lines indicate the 95% limits of agreement, and the shaded areas indicate the corresponding 95% confidence intervals.

https://doi.org/10.1371/journal.pone.0355338.s006

(PNG)

Acknowledgments

The authors sincerely thank the students of the veterinary anesthesiology laboratory at Rakuno Gakuen University for their assistance with the experiments.

References

  1. 1. Foster KR, Lukaski HC. Whole-body impedance--what does it measure?. Am J Clin Nutr. 1996;64(3 Suppl):388S-396S. pmid:8780354
  2. 2. Lukaski HC. Biological indexes considered in the derivation of the bioelectrical impedance analysis. Am J Clin Nutr. 1996;64(3 Suppl):397S-404S. pmid:8780355
  3. 3. Jaffrin MY, Morel H. Body fluid volumes measurements by impedance: A review of bioimpedance spectroscopy (BIS) and bioimpedance analysis (BIA) methods. Med Eng Phys. 2008;30(10):1257–69. pmid:18676172
  4. 4. Norman K, Stobäus N, Pirlich M, Bosy-Westphal A. Bioelectrical phase angle and impedance vector analysis--clinical relevance and applicability of impedance parameters. Clin Nutr. 2012;31(6):854–61. pmid:22698802
  5. 5. Jun M-H, Kim S, Ku B, Cho J, Kim K, Yoo H-R, et al. Glucose-independent segmental phase angles from multi-frequency bioimpedance analysis to discriminate diabetes mellitus. Sci Rep. 2018;8(1):648. pmid:29330426
  6. 6. Bansal N, Zelnick LR, Himmelfarb J, Chertow GM. Bioelectrical impedance analysis measures and clinical outcomes in CKD. Am J Kidney Dis. 2018;72(5):662–72. pmid:29885923
  7. 7. Mayne KJ, Shemilt R, Keane DF, Lees JS, Mark PB, Herrington WG. Bioimpedance indices of fluid overload and cardiorenal outcomes in heart failure and chronic kidney disease: A systematic review. J Card Fail. 2022;28(11):1628–41. pmid:36038013
  8. 8. Berbigier MC, Pasinato VF, Rubin B de A, Moraes RB, Perry IDS. Bioelectrical impedance phase angle in septic patients admitted to intensive care units. Rev Bras Ter Intensiva. 2013;25(1):25–31. pmid:23887756
  9. 9. Arab A, Karimi E, Vingrys K, Shirani F. Is phase angle a valuable prognostic tool in cancer patients’ survival? A systematic review and meta-analysis of available literature. Clin Nutr. 2021;40(5):3182–90. pmid:33581951
  10. 10. Spinale FG, Reines HD, Cook MC, Crawford FA. Noninvasive estimation of extravascular lung water using bioimpedance. J Surg Res. 1989;47(6):535–40. pmid:2685470
  11. 11. Scheltinga MR, Helton WS, Rounds J, Jacobs DO, Wilmore DW. Impedance electrodes positioned on proximal portions of limbs quantify fluid compartments in dogs. J Appl Physiol (1985). 1991;70(5):2039–44. pmid:1864785
  12. 12. German AJ, Holden SL, Morris PJ, Biourge V. Comparison of a bioimpedance monitor with dual-energy x-ray absorptiometry for noninvasive estimation of percentage body fat in dogs. Am J Vet Res. 2010;71(4):393–8. pmid:20367047
  13. 13. Yaguiyan-Colliard L, Daumas C, Bousbiat S, Jaffrin M, Cardot P, Grandjean D, et al. Indirect prediction of total body water content in healthy adult Beagles by single-frequency bioelectrical impedance analysis. Am J Vet Res. 2015;76(6):547–53. pmid:26000602
  14. 14. Yaguiyan-Colliard L, Daumas C, Nguyen P, Grandjean D, Cardot P, Priymenko N, et al. Evaluation of total body water in canine breeds by single-frequency bioelectrical impedance analysis method: Specific equations are needed for accuracy. BMC Res Notes. 2015;8:336. pmid:26245326
  15. 15. Arreesrisom P, Phichitrasilp T, Naimon N, Charoensin S, Rerkamnuaychoke W, Panpiansil J, et al. Using bioelectrical impedance analysis for modeling regression and predicting body fat accumulation in dogs in East Thailand. Vet World. 2022;15(6):1566–74. pmid:35993073
  16. 16. Nisini N, Corda A, Birettoni F, Miglio A, Antognoni MT. Bioelectrical Impedance Analysis (BIA) detects body resistance increase in dogs undergoing blood donation. Vet Res Commun. 2024;48(6):3889–97. pmid:39331344
  17. 17. Rae LS, Vankan DM, Rand JS, Flickinger EA, Ward LC. Measuring body composition in dogs using multifrequency bioelectrical impedance analysis and dual energy X-ray absorptiometry. Vet J. 2016;212:65–70. pmid:27256027
  18. 18. Rae LS, Rand JS, Ward LC. Measuring body composition in dogs using bioelectrical impedance spectroscopy. Vet J. 2024;304:106067. pmid:38266810
  19. 19. Silverstein DC, Montealegre C, Shofer FS, Otto CM. The association between vascular endothelial growth factor levels and clinically evident peripheral edema in dogs with systemic inflammatory response syndrome. J Vet Emerg Crit Care (San Antonio). 2009;19(5):459–66. pmid:19821887
  20. 20. Jun M-H, Ku B, Kim J, Kim KH, Kim JU. Mediation effect of the duration of diabetes mellitus on the decrease in bioimpedance phase angles in ethnically Korean people: A multicenter clinical study. J Diabetes Investig. 2021;12(5):790–802. pmid:32902171
  21. 21. Seoane F, Lindecrantz K, Olsson T, Kjellmer I, Flisberg A, Bågenholm R. Spectroscopy study of the dynamics of the transencephalic electrical impedance in the perinatal brain during hypoxia. Physiol Meas. 2005;26(5):849–63. pmid:16088073
  22. 22. Li J, Yim S, Pacheck A, Sanchez B, Rutkove SB. Electrical impedance myography to detect the effects of electrical muscle stimulation in wild type and Mdx Mice. PLoS One. 2016;11(3):e0151415. pmid:26986564
  23. 23. Mortreux M, Semple C, Riveros D, Nagy JA, Rutkove SB. Electrical impedance myography for the detection of muscle inflammation induced by λ-carrageenan. PLoS One. 2019;14(10):e0223265. pmid:31574117