Abstract
Luminescence is a powerful method for detecting trace analytes and monitoring biological processes. However, most bioluminescence reagents, including luciferase and its substrates, are sensitive to temperature, limiting their usable shelf lives, and resulting in inconsistent performance. Enhancing the stability of these reagents could reduce signal variability between experiments, simplify workflows, and address cold chain storage issues. In this study, we demonstrate the application of the platform stabilization technology, capillary-assisted vitrification (CAV), as a tool to stabilize different luciferases and their substrates, and the application of the stabilized reagents in both in vitro and in vivo bioluminescent assays. We demonstrate that CAV-stabilized reagents can be stored and shipped ambiently, maintain consistent performance over time, and are suitable for use in cell viability quantification, tumor monitoring, in vivo imaging, microbial detection, and immunoassays. Additionally, different reagents can be co-formulated to make ready-to-use assay kits that can also be shipped and stored ambiently. Our results demonstrate that CAV stabilization is a viable alternative to traditional storage methods, with broad potential to improve bioluminescence workflows.
Citation: Radford S, Peris-Taverner Y, Dibble M, Corn K, Zhu T, Martello S, et al. (2026) Development of shelf-stable reagents and assay kits for bioluminescence applications using the capillary-assisted vitrification platform stabilization technology. PLoS One 21(11): e0354533. https://doi.org/10.1371/journal.pone.0354533
Editor: Tianwen Wang, Anhui Polytechnic University, CHINA
Received: July 9, 2026; Accepted: October 18, 2026; Published: November 5, 2026
Copyright: © 2026 Radford et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Data Availability: The data that support the findings of this study are available through FigShare and can be accessed at the following link: https://doi.org/10.6084/m9.figshare.32796183.
Funding: Ambient Biosciences and Promega Corporation are commercial entities. Ambient Biosciences and Promega Corp. provided support in the form of salaries for authors SR, YPT, AL, SR, TC, AJ, MSR, LB, MD, and MD, but did not have any additional role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript. The specific roles of these authors are articulated in the ‘author contributions’ section.
Competing interests: SR, YPT, AL, SR, TC, AJ, and LB are employees of Ambient Biosciences and hold a beneficial ownership stake in the company. MSR is a contract employee of Ambient Biosciences and holds a beneficial ownership stake in the company. MD and MD are employees of Promega Corporation. Both Ambient Biosciences and Promega Corp. hold patents related to the technology discussed in this manuscript, and both companies have products in development related to this research. This does not alter our adherence to PLOS ONE policies on sharing data and materials.
Introduction
Luminescence has emerged as a powerful tool for biological and biochemical research [1–5]. Luminescence is defined as the production and emission of visible light resulting from a chemical reaction that releases energy in the form of light instead of heat. The term bioluminescence describes luminescent reactions that occur in nature and are typically the result of an enzymatic reaction. The enzyme reacts with a small molecule substrate in the presence of specific co-factors which may include oxygen, adenosine triphosphate (ATP) and calcium or other metals. Bioluminescence-based imaging has been widely adopted as a tool for identifying and monitoring tumor growth and monitoring viral infection in animal models [6–8]. Bioluminescence can also be used to assess the efficacy and toxicity of drug candidates, evaluate cell viability, and detect microbial contamination in hospital and food production settings [9–12].
Despite the broad utility of bioluminescence-based assays, the limited thermal stability of the enzymes and substrates used to generate the assay signals continues to present logistical challenges [13]. The reagents require long-term storage at ultra-cold temperatures and the performance of the reagents can degrade over time, adversely impacting sensitivity and assay reproducibility and limiting their use in point-of-care applications. Although significant effort has been made to enhance the stability of bioluminescent reagents through genetic engineering, chemical modification and specialized formulations [14–20], the components remain highly temperature dependent.
Techniques such as lyophilization, and the more modern lyobead technology, have been successfully used to enhance stability of a broad range of reagents and samples and have occasionally led to improved assay performance [21–23]. However, these processes can be both cost prohibitive and damaging to sensitive biomolecules. Lyophilization exposes biomolecules to a range of chemical and physical stresses including cold denaturation, cryoconcentration, pH shifts, unfavorable protein-excipient interactions, and exposure to solid-liquid interfaces as ice crystals form [24–26]. Together, these stresses can compromise biomolecule integrity, reduce functional activity, and ultimately lead to inconsistent performance.
Luminescent substrates present an additional stabilization challenge. Most luminescent substrates are hydrophobic and may exhibit limited solubility making them challenging to lyophilize on their own and particularly difficult to lyophilize when incorporated into a mixture. Lyophilization of highly hydrophobic materials generally requires extensive, molecule-specific optimization, and may or may not produce satisfactory results [27].
Novel stabilization technologies that can be applied by researchers to stabilize diverse bioluminescent reagents are therefore still needed. The ideal solution should be easy to deploy in a laboratory setting, require minimal capital investment, require minimal molecule-specific optimization and facilitate production of assay master mixes and kits.
We previously described the development of a novel stabilization technology called Capillary-Assisted Vitrification (CAV, referred to as Capillary Mediated Vitrification or CMV in previously published work) [28–31]. Like lyophilization, CAV achieves stabilization by transitioning sensitive biomolecules into an amorphous glass through a process referred to as vitrification. In the glassy state, both intra- and inter-molecular motion is restricted preventing collisions and the molecular conformational changes required for many common degradation pathways [32]. Stability is further achieved by removing bulk water and thereby slowing or halting water-mediated degradation processes such as deamidation.
Compared to lyophilization, CAV is a relatively gentle process and does not require a freezing step. In the CAV process, the reagent of interest is mixed with a pre-formulated stabilization buffer, referred to as BioFix™ or BioFix+™, and applied to a solid, porous material referred to as a scaffold. The formulated liquid embeds into the pores of the scaffold due to capillary action. The reagent-loaded scaffold is then placed into a purpose-built vacuum chamber, referred to as the Ambient Stabilizer, and dried under coarse vacuum. A combination of the high surface area provided by the capillary network within the scaffold and a controlled evacuation of the vacuum chamber allows for gentle desiccation of the sample matrix avoiding both boiling and freezing the liquid.
Although the CAV technology has been applied successfully for the stabilization of antibody reagents and PCR master mixes [30,31], only limited work has been done using bioluminescent reagents. In a previous study, we demonstrated that stabilized luciferase could be stored for up to 6 weeks at room temperature with minimal impact on the enzyme activity but did not evaluate long-term stability or the suitability of the stabilized material in real-world analytical methods [28]. Here, we expand on our previous work by demonstrating that the CAV technology can be applied not only to luciferase, but also to luciferin and furimazine, two commonly used bioluminescent substrates. We demonstrate long-term stability of luciferase, luciferin, and furimazine and evaluate batch-to-batch consistency for stabilized luciferase. We establish that the reagents can be co-formulated into a variety of ready-to-use assay master mixes using a common stabilization protocol. Finally, we evaluate the suitability of the CAV stabilized reagents in a wide range of in vitro, in vivo, and biochemical methods including cell viability assays, tumor imaging, and mix-and-read bioluminescent immunoassays. Overall, the results of this investigation indicate that CAV-stabilization is an attractive platform stabilization technology for stabilizing bioluminescent reagents.
Results
Performance and stability of luciferase and luciferin
Implementation of bioluminescence applications under real-world conditions requires the ability to store reagents for periods of 6 months or more and to withstand short-term exposures to extreme temperature resulting from temperature excursions during shipping or storage. Thus, we initiated long-term stability studies of luciferase and luciferin at room temperature and a short-term stability study at 55 °C. At each time point, the activity of the CAV-stabilized reagents was evaluated by comparing the performance of the CAV-stabilized luciferase or luciferin to a control preparation made from reagents that were stored frozen and subsequently thawed and prepared on the day of use (Fig 1).
Reprinted from Bronsart, L. 2026 https://BioRender.com/k45a805, under a CC BY license, with permission from Biordender original copyright 2026.
At the start of the stability study, we first evaluated the recovery of the two reagents immediately post-stabilization. As shown in Fig 2A, the signals obtained for the CAV-stabilized luciferin were comparable to the control preparation. The dose response curves generated using the CAV-stabilized and control reagents were evaluated graphically and analyzed using a two-sided t-test; the results were not statistically different between the stabilized and control material (p-value = 0.31). When the results for the CAV-stabilized luciferase were compared to the control preparation of luciferase (Fig 2B), a statistically significant difference between CAV-stabilized and the control preparation of luciferase was observed (p-value = 0.009). The CAV-stabilized material provided slightly higher signals and lower variability across replicate results compared to the frozen control. Despite the small difference in signal, the data indicates that the stabilization buffer does not interfere with the enzymatic reaction that produces the luminescent signal and that the process results in little or no loss of material.
(B) Comparison of luminescent signal generated using CAV-stabilized luciferase (orange diamonds) compared to the control material prepared from freshly thawed luciferase (blue circles). Error bars show standard deviation across the replicate results (n = 3).
We further evaluated the performance of the CAV-stabilized reagent by calculating the assay limit of quantitation (LOQ) for stabilized reagents and control preparations. Across experiments, the estimated LOQ ranged from 0.7-2.5 µg/mL using control preparations of luciferin and luciferase. Using CAV-stabilized luciferin or luciferase resulted in an LOQ of 2.3 and 1.1 µg/mL, respectively. This is well within the range observed for the reference preparations. In all cases, the dose response curves exhibited coefficient of determination (R2) of >0.99. Thus, we conclude that the stabilization process had no adverse impact on linearity or sensitivity.
Next, aliquots of CAV-stabilized reagents and controls were placed on stability. At each time point, the activity of the CAV-stabilized reagent and the liquid controls were again compared to a reference standard prepared from freshly thawed luciferase and luciferin. At the end of the study, stability data were evaluated graphically to estimate the degradation rates at each temperature. As shown in S1 Fig and S2 Fig, the CAV-stabilized materials show a linear trend when the natural log of activity is plotted versus time, indicating the degradation process can be described by a first-order reaction rate. According to the differential rate law, the slope of this line is equal to -k, where k is the rate constant. The half-life (t1/2) can then be calculated using the equation:
The t1/2 values are shown in Table 1. The t1/2 for the CAV-stabilized luciferin was estimated to be 39 months at room temperature and 2.1 months at 55 °C. The t1/2 value for CAV-stabilized luciferase was estimated to be 19 months at 55 °C. Although a linear trend was observed for luciferase stored in the CAV-form at 25 °C, the half-life could not be accurately measured. The activity of the CAV-stabilized luciferase was essentially flat across the nearly 2 year study, showing little or no downward trend and resulting in a correlation coefficient of 0.13. Even after 22 months of storage, the enzyme maintained 97% activity (S2 Fig). Based on this data, we conclude that the t1/2 will significantly exceed 2 years at room temperature.
It is important to note that stability was assessed based on activity within the target assay only. No structural testing was performed. While activity is typically correlated with structure and is considered a key indicator of reagent stability, further testing is required to determine whether any structural changes occurred during storage.
Except for luciferin stored at room temperature (t1/2 = 0.1 months), we were unable to determine t1/2 for the liquid formulations. When stored as a liquid at 55 °C, both luciferase and luciferin lost all activity in less than 1 day. Test results for the luciferase formulation stored as a liquid at 25 °C were not reproducible between time points or between repeated experiments. Due to this variability, the performance of the liquid luciferase stored under this condition is considered unacceptable for an analytical reagent.
Overall, our stability evaluation confirms the following: 1) the CAV process significantly enhances the stability of both luciferase and luciferin, 2) the CAV-stabilized reagents can be stored for greater than 6 months at room temperature, and 3) the CAV-stabilized reagents can withstand short-term temperature excursions of up to 55 °C. Together, the data suggests that the CAV-stabilized reagents are suitable for ambient distribution and storage.
To further evaluate the performance of the CAV-luciferase with respect to reagent consistency over time and between stabilization cycles, a scale-up study was performed. In each run, a total of 512 scaffolds were prepared and stabilized in a single stabilization cycle. A minimum of 6 scaffolds were tested immediately post-stabilization and again after the scaffolds had been stored for 9 months at room temperature. The mean activity across all four batches was 103% both post-stabilization and after 9 months of storage (Table 2). The relative standard deviation (RSD) across scaffolds produced within a batch ranged from 1.9% to 7.8%. The RSD across the batches was 7.2% at time 0 and 5.1% after 9 months of storage. This data demonstrates that the CAV-process is highly consistent and further confirms the observation that the stabilized reagents are stable for months or years at room temperature.
Experimental applications
Suitability for imaging and reporter gene assays.
We next evaluated the suitability of the CAV-stabilized luciferin for use in an in vivo imaging application and compared results to those generated with luciferin that was stored lyophilized or frozen. We quantified tumor burden using a 4T1 mouse mammary carcinoma model. Tumor size was monitored for up to 21 days after 4T1 tumor inoculation (Fig 3), and the results obtained with bioluminescence imaging were compared to measurements made with calipers.
We observed no statistical difference in the relationship between caliper measurements and bioluminescence signal of tumor growth between the luciferin groups (Fig 3A–C). Although the CAV-stabilized luciferin experienced temperatures exceeding 41 °C during shipment as measured using a temperature tracker, we found that both in vivo bioluminescence on day 16 post-inoculation and ex vivo tumor measurements were comparable in mean signal and the CAV-stabilized luciferin exhibited a smaller variation in signal compared to the frozen control luciferin (Fig 3C-D). This side-by-side analysis suggests that the stabilization process results in a functional and stable product equivalent to the current gold standard in bioluminescence imaging of tumors in mice.
In vivo tumor imaging reveals consistent bioluminescence measurements between frozen reference (blue) or CAV-stabilized (orange) luciferin. (A) Mice implanted with 4T1 breast cancer cells in the #4 right mammary fat pad and imaged with the frozen reference luciferin (2 of 15 mice shown) (left) or CAV-stabilized luciferin (right). (B) Relationship between bioluminescence intensity and tumor volume based on caliper measurements. (C) Comparison of in vivo bioluminescence on day 16 post-inoculation (n = 7 reference, n = 8 CAV-stabilized). (D) Comparison of ex vivo tumor bioluminescence (n = 7 per group). The boxplots in C and D show the mean (x), median, the interquartile range (IQR), and ± 1.5 x IQR.
Bioluminescence is also frequently used to monitor the expression of reporter genes in cell culture applications. Cells containing the reporter gene are cultured, subjected to the test condition, and treated with luciferin or another substrate to produce a luminescent signal that is proportional to the level of reporter gene expression. To assess whether vitrified luciferin could be used in this type of application, luciferase-expressing 4T1 mouse breast cancer cells were prepared as a suspension culture and then serially diluted. Fig 4 shows that the frozen and CAV-vitrified luciferin signal are not statistically different based on ANCOVA results.
(B) Comparison of 4T1 cell viability data obtained using reference reagent master mix (blue circles) or CAV-stabilized master mix (orange triangles) (n = 4). Error bars represent standard deviation.
Co-formulated master mixes for cell viability and microbial contamination.
As described above, many bioluminescent assays require multiple reagents that, under traditional storage paradigms, must be individually stored. On the day of use, each reagent must be located in the cold storage unit, thawed, diluted and co-formulated to produce a master mix. With the goal of developing thermally stable kits for bioluminescence testing, we prepared stabilized, assay master mixes by co-formulating all of the reagents required to run a specific test method prior to stabilization. On the day of the test, a single scaffold was removed from the packaging, eluted, and immediately used in the appropriate assay. No further reagent preparation or dilution steps were required prior to use. For each experiment, reagents stored in the traditional manner (lyophilized or frozen, with each reaction component stored in a separate vial as shown in Fig 1), were used to prepare a control reagent mixture.
To test the performance of the CAV-stabilized master mix for cell viability applications, freshly cultured 4T1 cells were placed in a 96-well plate and diluted to a range of concentrations ranging from 1 to 6 X 104 cells/well and treated with the reagent master mix. The results obtained with the frozen and CAV-stabilized mixture are visually indistinguishable and not statistically different based on an ANCOVA (Fig 4B). The assay LOQ is estimated to be approximately 13 cells/well using the control master mix and 15 cells/well using the CAV-stabilized master mix.
Given the linear response and sensitivity observed using the 4T1 cells, we evaluated a microbial contamination application. To test for surface contamination (Fig 5), slides were prepared by applying a buffer control (negative condition) or freshly cultured E. coli (1x105 cells for the positive condition). Swabs from the slides were placed into a tube containing either the traditional or CAV-stabilized reagent master mix. The slide was considered positive if the measured signal was above the detection limit. All samples were correctly identified as positive or negative using both the traditional and CAV-stabilized master mix (S1 Table). Although this data successfully demonstrates proof-of-concept, additional testing would be required to verify the limit of detection and confirm comparable sensitivity.
Glass slides were treated with E. coli or sterile buffer and dried. The slides were swabbed, and the swab was placed into a tube containing the reaction mixture prepared fresh or from CAV-stabilized reagents. The resulting luminescence was analyzed using a plate reader. Reprinted from Bronsart, L. (2026) https://BioRender.com/x47p386 under a CC BY license, with permission from Biordender original copyright 2026.
Homogeneous immunoassay kits.
The Lumit® Flex technology is a homogeneous immunoassay that utilizes a ternary split NanoLuc® luciferase complementation system [33–35]. Two small peptide tags (Peptide α and Peptide β) are chemically or genetically attached to analyte-specific antibodies or binding proteins. A detection polypeptide is then added, and remains inactive until it is brought into proximity to the peptide tags. When the ternary components assemble upon analyte binding, the fully reconstituted NanoLuc enzyme catalyzes oxidation of the furimazine substrate, generating a bright, stable bioluminescent signal. The assay formats show a dose-dependent increase in luminescent signal in response to increasing concentrations of recombinant human IL-6 (rhL-6). While each of the individual reaction components can typically be stored frozen or lyophilized, co-formulation of the components into a ready-to-use mixture can be challenging due to solubility and stability issues, particularly with the furimazine substrate.
Because the CAV-technology has been successfully used to stabilize both small and large molecule reagents, we explored whether CAV stabilization could be used to prepare an interleukin-6 (IL-6) Lumit Flex master mix (Fig 6A). The performance of the standard solution-based Lumit Flex IL-6 immunoassay was compared to CAV-stabilized furimazine alone, a mixture of the two peptide tags and the detection polypeptide, or a master mix in which all of the assay components were co-stabilized. On the day of use, the CAV-reagents were reconstituted in PBS and an rhIL-6 standard curve generated and compared to a standard curve generated with the standard, solution-based reagents. For the furimazine and peptide/polypeptide mixtures, the data was analyzed graphically by plotting the signal-to-noise ratio as a function of rhIL-6 concentration. Using the standard stabilization process with no optimization, the liquid and CAV-stabilized furimazine produced similar signal-to-noise ratios across the range of IL-6 concentrations tested. This suggests that the vitrification did not adversely impact the activity of the substrate or the signal generation process (Fig 6B). Stabilization of the peptide/polypeptide mixture did result in a decrease in signal-to-noise at low IL-6 concentrations relative to the control preparation (Fig 6C). Testing on real-world samples would be required to fully evaluate the impact of the signal shift on the assay sensitivity. However, for most immunoassays, a signal/noise ratio of 10 or more is preferred to ensure adequate precision and accuracy. Using the liquid preparation, IL-6 concentration ranging from 0.07-40 ng/mL met this criterion. For the CAV-stabilized mixture, the range of IL-6 concentrations that yielded an acceptable signal/noise ratio shifted upwards to 0.4-250 ng/mL. The shift in assay response and the reduced signal intensity persisted when all of the assay components including the substrate, peptide tags, and polypeptide were co-formulated and stabilized (Fig 6D).
(B) Comparison of the signal/noise ratio observed using a traditional liquid preparation of furimazine versus a CAV-stabilized preparation. (C) Comparison of the signal/noise ratio observed using a liquid mixture of the two peptide tags and the large polypeptide compared to a CAV-stabilized mixture of the three components. (D) Comparison of the performance of Lumit Flex immunoassay using traditional reagent formulation (blue circles) and a co-formulated master mix containing all assay components and stabilized using the CAV process (orange triangles). The CAV stabilized material shows a slight loss in absolute signal and a reduction in sensitivity but good linearity and a broad dynamic range. Error bars indicate the maximum and minimum signals observed across duplicate runs.
Although the solution-based assay exhibits higher absolute signal intensity, the performance of the CAV-stabilized immunoassay may be sufficient for many applications, including studying multiple pathologies, response to therapy, and inflammation [36–38]. Additionally, this data demonstrates the performance that can be achieved with no optimization of either the stabilization protocol or the assay conditions. With optimization, it may be feasible to enhance the sensitivity further
Scanning electron microscopy
For characterization purposes, scanning electron microscopy (SEM) images were collected of the scaffold material both prior to application of the formulated reagent and after vitrification. For this experiment, the BioFix buffer was used alone. Due to the low concentration of the reagents used in these experiments, it is unlikely that their presence would impact the appearance of the scaffold post-vitrification. Representative images are shown in S3 Fig. As expected, the scaffold has a fibrous appearance with voids, or pores, visible between the fibers (dark regions of the image). After vitrification, a hazy film or glass is visible. The glass appears to coat fibers, resulting in an open, porous structure. This structure is critical for ensuring rapid dissolution of the glass upon rehydration.
Discussion
Bioluminescence reagents are highly valued for their exceptional sensitivity and robust signal generation and have been successfully implemented across a broad range of analytical applications. Despite their widespread use, thermostability limitations and current storage paradigms continue to present logistical challenges that can hinder successful deployment. Today, most bioluminescent reagents continue to require ultracold storage conditions. This adds significant operational costs and risks loss of valuable materials should the freezer fail. Additionally, the use of frozen reagents adds complexity to the day-to-day execution of the analytical methods; each time an assay is executed, multiple materials must be located, thawed, and diluted, which introduces the opportunity for error and run-to-run variability.
To address these challenges, researchers have explored various stabilization approaches. While genetic engineering may provide limited improvements in luciferase thermostability [16], it is not a solution for small molecule substrates. If carefully optimized and executed, lyophilization and lyobead technologies can in some cases enhance reagent stability as previously noted [22]. However, both methods expose biomolecules to multiple types of stress [24,25] and they are rarely readily available to the bench scientist. Like lyophilization, vacuum foam drying requires complex capital equipment and is not suitable for single-use volumes [39]. While readily accessible, conventional air drying often requires elevated temperatures that can thermally inactivate sensitive enzymes like luciferase and the process is difficult to scale [39].
Our findings demonstrate that CAV offers a distinct and complementary platform. By eliminating the freezing step, the CAV process reduces the risk of cold damage and solubility concerns, allowing researchers to stabilize diverse bioluminescent reagents, both individually and in complex mixtures. The process can be performed with small volumes, making it amenable to storage in single-use quantities. And, as demonstrated here, a single, standardized process can be applied to both large and small molecule reagents with little or no optimization, allowing researchers to easily produce custom assay kits.
The CAV process relies on a unique physical process and the synergistic relationship between the stabilization matrix and the scaffold. The CAV process utilizes a carefully optimized, proprietary mixture of excipients, which like most lyophilization formulations, includes a saccharide component. The saccharide promotes the formation of an amorphous glass matrix, replaces the hydrogen bonds lost when water is removed (water replacement), and furhter enhances stability through preferential exclusion [40].
Unlike lyophilization, where the goal is to produce a freestanding porous glass cake [41], the glass that forms during the CAV process is inherently porous, due to the architecture of the scaffold. Because we utilize the physical architecture of the scaffold to dictate the macroscopic porosity of the final glass, we do not need to rely on the complex physics of freezing to engineer a freestanding porous structure. This reliance on the scaffold’s inherent architecture is a primary reason why CAV requires significantly less formulation optimization compared to traditional lyophilization. Ultimately, this scaffold-driven porosity ensures rapid dissolution of the glass upon rehydration and eliminates the challenges associated with preventing lyo cake collapse.
The choice of the scaffold material is an important consideration as in some cases, there may be an interaction between the material being stabilized and the scaffold material of construction. The most common interaction observed to date is non-specific adsorption or weak binding due to charge-charge and/or hydrophobic interactions between the scaffolds and the biomolecule. In those cases, the issue can generally be resolved by utilizing alternative scaffold materials or modifying the formulation with the addition of surfactants or salts. Scaffolds can be produced from a variety of natural and synthetic polymers provided they offer high porosity and hydrophilicity. The pore size of the materials may vary but attention must be paid to the resultant capillary forces such that liquid absorption capacity is not negatively impacted.
Despite its advantages, the CAV platform has potential molecule- and formulation-specific limitations. While most materials can be stabilized using the standard stabilization protocols, some reagents or assays may require further optimization. In this study, the Lumit Flex immunoassay exhibited a reduction in absolute signal intensity and assay sensitivity compared to the traditional liquid format when stabilized using the standard stabilization protocol. Additionally, although the process is compatible with a broad range of detergents, buffers, salts, chelating agents and other common assay co-factors, the process is not compatible with all excipients. For example, the process can routinely tolerate up to 1% glycerol in the final formulated material, meaning some commercial reagents may require buffer exchange. Finally, the physical characteristics of the scaffold dictate processing limits; each scaffold has a maximum loading volume and a minimum residual hold-up volume, necessitating careful matching of the scaffold size to specific assay requirements to minimize mass loss.
Ultimately, the CAV process offers distinct operational advantages by decoupling reagent stabilization from cumbersome and expensive infrastructure, such as heavy-duty compressors or liquid nitrogen. The compact Ambient Stabilizer can be seamlessly integrated into standard laboratory environments or biological safety cabinets, giving researchers the ability to prepare custom, shelf-stable materials on demand. Eliminating the need to store and ship reagents using cold or ultra-cold storage can have an immediate impact on laboratory carbon footprints and improve global access to research, clinical products, and testing supplies [42]. The thermally stable microbial contamination kit and Lumit Flex immunoassay demonstrate robust analyte detection in a simplified format, highlighting the potential of CAV stabilization to facilitate rapid, on-site testing in decentralized settings such as rural clinics, field hospitals, and outbreak monitoring sites.
Materials and methods
Sample preparation
QuantiLum® Recombinant Luciferase, Beetle Luciferin, VivoGlo® Luciferin, and Lumit® Flex human IL-6 assay reagents were obtained from Promega Corp. (Madison, WI). Upon receipt, the luciferase was stored at −80 °C as per the manufacturer’s recommendations. The lyophilized beetle luciferin was reconstituted at 10 mg/mL in phosphate buffered saline (PBS, pH 7.4; Cytiva, Logan UT), aliquoted and stored at −80 °C. VivoGlo Luciferin was reconstituted in a 1:1 mixture of PBS and BioFix stabilization buffer, and aliquots stored at −20 °C.
CAV stabilized samples were prepared in the lab using the Ambient Stabilizer (Ambient Biosciences, Ann Arbor MI). The CAV process was described previously [28], and a schematic of the process is provided in Fig 1. Briefly, the reagent of interest is prepared at a 2X concentration and mixed with an equal volume of BioFix or BioFix+ stabilization buffer (Ambient Biosciences, Ann Arbor MI). The formulated material is dispensed onto a porous material referred to as the scaffold. The quantity of reagent is determined based on the assay requirements and generally each scaffold is designed to hold enough reagent for a single assay run.
The BioFix and BioFix+ buffers are commercially available, preformulated stabilization reagents. The BioFix+buffer contains the same base stabilizers with the addition of a carrier protein which prevents non-specific binding of proteins to surfaces. The BioFix+ is recommended when working with low concentration protein reagents but can be used for other reagents including small molecules.
Scaffolds were selected from a catalog of commercially available options that differ primarily in terms of the volume of reagent that can be applied and stabilized on each scaffold. For each experiment, the smallest scaffold that would hold a suitable volume of the stock reagent was selected. The VivoGlo luciferin, luciferase, and microbial contamination kits were therefore stabilized using BioFix 175 Scaffolds (Ambient Biosciences, Ann Arbor, MI). All other materials were stabilized using BioFix Classic Scaffolds (Ambient Biosciences, Ann Arbor, MI).
On the day of use, the CAV-stabilized samples were rehydrated by incubating the CAV-stabilized samples in elution buffer. The choice of elution buffer was determined separately for each assay based on the assay requirements. Incubation times of 1 minute or less were sufficient to fully rehydrate the CAV-stabilized material. Unless otherwise noted, the eluted material was used in the assay with no further manipulation. Fig 1 provides a schematic of the rehydration process and for comparison purposes, the workflow used to prepare liquid controls.
Luciferase and luciferin activity assay
Luciferase and luciferin activity were measured using a microtiter plate-based assay. A luciferin reaction mixture containing 1.5 mM beetle luciferin, 4.5 mM magnesium sulfate (Sigma, St. Louis, MO), and 0.5 mM ATP (Sigma, St. Louis MO) was prepared in 25 mM EPPS buffer (Sigma.,St. Louis MO), pH 7.8. Separately, serial dilutions of luciferase ranging in concentration from 0 to 23 µg/mL were prepared. To initiate the assay, a 50 µL aliquot of each luciferase dilution was placed in a 96-well flat-bottom plate with a clear bottom and black walls (Corning, Corning NY) and mixed with a 50 µL aliquot of the luciferin reaction mixture. The plates were incubated for 8 minutes after the addition of all reagents and luminescence was measured using a Synergy H1 microplate reader (BioTek, Winooski, VT) and the endpoint luminescence protocol.
The activity of the test articles was determined by comparing the response of the test article with a reference standard. The reference standard was made using a luciferin mixture and luciferase dilutions that were freshly prepared from frozen stocks on the day of use. The relative activity of the test article was calculated by direct interpolation for each sample dilution that fell within the linear range of the assay and averaging across the dilution adjusted results.
The test articles consisted of one of the following: 1) CAV-stabilized luciferase, 2) luciferase stored in a liquid formulation, 3) CAV-stabilized luciferin, or 4) luciferin stored in a liquid formulation. When luciferase was tested in either the CAV-stabilized or liquid formulation, freshly thawed and diluted luciferin was used. Likewise, when testing luciferin test articles, a freshly prepared luciferase solution was used.
The response curves generated using CAV-stabilized and control reagents were evaluated using a two-side t-test. The Limit of Quantification (LOQ) was determined based on the standard deviation of the response and the slope of the calibration curve. The standard error of the y-intercept generated from the linear regression analysis was utilized as the standard deviation of the response (). The LOQ was calculated using the equations
, where
is the slope of the calibration curve.
Stability assessments
To evaluate the stability of CAV-stabilized luciferase and luciferin, the reagents were initially stabilized individually. CAV-stabilized luciferin samples were prepared so that each scaffold contained 239 µg luciferin in BioFix+ stabilization buffer. At the time of testing, the luciferin scaffolds were eluted with 0.5 mL of a solution containing 4.5 mM magnesium sulfate, 0.5 mM ATP in 25 mM EPPS buffer, pH 7.8. A liquid control was also prepared. The composition of the liquid controls was matched to the CAV-stabilized samples post elution.
CAV-stabilized luciferase samples were prepared so that each scaffold contained 92 µg of luciferase in BioFix+ stabilization buffer. At the time of the test, the luciferase scaffolds were eluted with 4 mL of PBS, producing a solution containing 23 µg/mL luciferase which was then used to prepare serial dilutions for testing. Again, a liquid control was prepared that matched the composition of the luciferase eluted from the BioFix scaffolds, but did not include the BioFix+ stabilization buffer.
The CAV-stabilized samples and liquid controls were stored on the bench (room temperature condition) or in a temperature-controlled incubator set at 55 °C. For the CAV-stabilized samples, three scaffolds were removed from storage at each time point. Triplicate scaffolds were eluted and independently tested, generating three test results per time point. For the liquid controls, a single aliquot was removed at each time point. The liquid control was sampled and tested three times.
Tumor imaging in mice
All animal experiments were carried out according to a protocol approved by the Vanderbilt University Institutional Animal Care and Use Committee and adhered to all relevant regulations and guidelines. Bioluminescence was evaluated in mice using luciferase-expressing 4T1 cell lines. The CAV-stabilized luciferin samples used in this experiment contained 9 mg of VivoGlo® Luciferin per scaffold. A reference solution was prepared using lyophilized VivoGlo® luciferin reconstituted in PBS. Materials for the tumor imaging studies were shipped to the testing site ambiently via ground transportation and stored at room temperature until use. Ground transportation required 6 days, and during this time the material was exposed to temperatures ranging from 71-107 °F (21–42 °C). The corresponding liquid controls were shipped on dry ice.
Seven-week-old BALB/c female mice (Charles River Laboratories) were implanted with 5 × 104 4T1-luciferase-expressing cells, suspended in 50 μL of PBS, in the #4 right mammary gland. On the day of use, scaffolds were eluted in 1 mL per scaffold resulting in a final post-elution concentration of 9 mg/mL. Excess reference solution was stored as frozen aliquots between imaging sessions. Bioluminescence imaging 2x/week followed 30 min after a 150 mg/kg i.p. injection of CAV-stabilized luciferin or reference luciferin (n = 7 reference, n = 8 CAV-stabilized). To minimize pain and distress, discomfort produced from tumor implantation was controlled using inhaled isoflurane anesthetic by nose cone (2% isoflurane in 100% O2, flow rate 1 L/min). During imaging, each subject was anesthetized on a heated platform using isoflurane delivered via an induction chamber and 2% isoflurane lines. Imaging is non-invasive, and anesthesia was introduced only to keep the subject still during the experiments. We monitored all mice for potential signs of distress over the course of the experiment. Tumors were imaged using an IVIS Lumina Series III imaging system (Caliper LifeSciences). Tumor length and width were measured using digital calipers 2x/week up to 21 days post-inoculation or when humane endpoint (tumor volume>1,000 mm3) was reached. Tumor volume was calculated using the formula: Volume = (D12 x D2)/2, where D1 is the minimum diameter and D2 is the maximum diameter. Tumors reached an average of 650 ± 238 mm3 (mean ± SD) and one mouse died before meeting the criteria for euthanasia.
4T1 reporter gene expression
A luciferase-expressing mouse mammary carcinoma cell line 4T1/Puro Dual Labeled Stable Cell Line was obtained from GeneCopoeia (SL020). The cell line is stably transfected with both a firefly luciferase gene and eGFP gene. Cells were propagated and maintained in RPMI selection medium (BioWhittaker/Lonza, Walkersville MD and Gibco, Grand Island NY) containing 1% GlutaMAX (Gibco, Grand Island NY), 1% Ab/Am, puromycin (Gibco, Grand Island NY) and 10% FBS (Gibco, Grand Island NY). Cells were cultured at 37 °C in a 5% CO2 atmosphere.
On the day of testing, cells were suspended in medium containing 450 µg/mL luciferin and titrated from 7.8 X 103 to 106 cells per well in 2-fold serial dilutions in an opaque white 96-well cell culture plate (Corning, Corning NY). Prior to use, the luciferin was either stored frozen and thawed and diluted in media on the day-of-use (reference solutions) or in the CAV stabilized form. Each CAV-stabilized aliquot contained 2.7 mg of beetle luciferin and was eluted in 6 mL of media. After a 15-minute exposure, bioluminescence intensity was read using a Synergy H1 microplate reader. The experiment was carried out on each of 3 days for n = 3 replicate runs. The relative activity for the CAV-stabilized luciferin was determined by calculating the average of all dilutions that fell within the linear range of the assay and reporting that result as a percentage relative to the reference standard. ANCOVA analysis was performed to test for statistical differences in the data set. The limit of quantitation (LOQ) was defined as the mean blank signal plus 10 times the standard deviation of the blanks, respectively. These signal thresholds were subsequently converted to concentrations by interpolation against the log-log linear regression equation of the raw, uncorrected standard curve.
Cell viability assessment
Bioluminescence was evaluated in CRL-2539 4T1 cells (ATCC, Manassas VA) to evaluate whether luciferase and luciferin could be co-formulated into an easy-to-use assay master mix or kit for cell viability testing. This cell line does not contain a luciferase reporter gene. Cells were titrated from 1 to 6 × 104 cells/well in an opaque, white 96-well cell culture plate, centrifuged, and washed with Dulbecco’s Phosphate Buffered Saline (DPBS, Lonza, Walkersville MD). A 20 µL aliquot of lysis buffer containing 25 mM tris-phosphate (pH 7.8), 2 mM dithiothreitol, 2 mM EDTA, 10% glycerol, 1% Triton X-100 was added to each well and the plate was incubated on a shaker for 20 minutes. Scaffolds containing 4.4 µg luciferase and 840 µg luciferin co-vitrified in BioFix+ stabilization buffer were prepared. Six scaffolds were eluted together in 5.28 mL buffer containing 4.5 mM magnesium sulfate and 25 mM EPPS, pH 7.8, for a final concentration of 5 µg/mL luciferase and 3 mM luciferin. A reference standard was prepared from freshly thawed luciferase and luciferin diluted to 5 µg/mL luciferase and 3 mM luciferin in buffer containing 4.5 mM magnesium sulfate, 0.28% bovine serum albumin (BSA), 25 mM EPPS, pH 7.8. Subsequently, plate wells were filled with 100 µL master mix with sample or standard. After a 15-minute exposure, bioluminescence intensity was read using a Synergy H1 microplate reader. The relative activity for the CAV-stabilized master mix was determined by calculating the average across all dilutions that fell within the linear range of the assay and reporting that result as a percentage relative to the reference standard. ANCOVA analysis was performed to test for statistical differences in the data set. Using only data that fell within the linear range of the assay (10–2222 cells/well) the LOQ was defined as the mean blank signal plus 10 times the standard deviation of the blanks. These signals were subsequently converted to concentrations by interpolation against the log-log linear regression equation of the standard curve.
Microbial contamination testing
A microbial contamination kit was prepared by vitrifying 0.875 µg luciferase, 139 µg luciferin, 9 µg DTT, and 0.035 µg Triton X-100 in BioFix+ stabilization buffer on BioFix 175 scaffolds. This master mix contained the enzyme, substrate and all co-factors required for testing in a single CAV-stabilized preparation. Sterile broth or broth containing 1 X 105 E. coli cells were applied to sterile microscope slides and allowed to dry for 2 hours at room temperature. On the day of the test, a liquid reference control mixture was prepared from 4.16 µg/mL luciferase, 2.5 mM luciferin, 0.333 mM DTT, 2% glycerol, 0.2% Triton X-100, 3.5 mM MgSO4 in 25 mM Tris buffer. CAV scaffolds were eluted in 0.5 mL buffer containing 2% glycerol, 3.5 mM MgSO4 in 25 mM Tris. A FLOQSwab (Copan Cat #534CS01) was soaked in 0.5 mL reaction mixture, used to separately swab the surface of each glass slide, and then returned to a tube containing the reference master mix prepared from frozen stocks or from the CAV-stabilized master mix. A 100 µL quantity of the reaction mixture was then transferred to a 96-well plate and read for luminescence in a Synergy H1 plate reader. Limit of detection was defined as the mean of the negative control plus three times the standard deviation of those six negative control samples. In this dataset, LOD was defined as 798 RLU. Anything above this threshold was considered positive. The microbial contamination assay format is shown schematically in Fig 5.
Lumit® Flex IL-6 immunoassay
Anti-IL-6 antibody clones 5IL-6 and 505E (Thermo Fisher) were chemically conjugated with the Lumit® Flex alpha and beta peptide (Promega Corp), respectively, and optimized for working concentration as previously reported [33]. Briefly, each antibody was incubated with an amine-reactive form of the peptide labels for 1 hour at room temperature, quenched with 2 M Tris buffer, and then diluted in 80% glycerol for storage. A solution-based assay was established using a master mix containing 2 µM Lumit® Flex Detection Protein, 5IL-6-Lumit® Flex Peptide Alpha labeled antibody, 505E-Lumit® Flex Peptide Beta labeled antibody, and 10 µM of Nano-Glo Live Cell Substrate (Promega Corp). This master mix/coformulation was mixed with recombinant IL-6 (R&D Systems) titration prepared in assay buffer. After 1.5 h, the luminescence was read on a GloMax Discover Luminometer (Promega Corp).
The CAV-stabilization of the Lumit Flex IL-6 immunoassay was performed using the BioFix+ buffer. Scaffolds were prepared containing the substrate only, the two labeled peptides plus the polypeptide detection protein, or all of the assay components co-formulated together. Post-stabilization, scaffolds were eluted with 500 µL of PBS + 0.01% BSA to collect the assay components, which were then added to a non-binding white 96-well plate (Corning) containing a dilution of IL-6.
Scanning Electron Microscopy (SEM)
SEM images of a BioFix scaffold before application of any liquid and post-vitrification were collected without sputter coating using a Thermo Fisher Quattro ESEM in low vacuum mode at a pressure of 0.25 Torr. Images were collected with an accelerating voltage of 10.00 kV, current of 0.37 nA, and a horizontal field width of 300 µm.
Supporting information
S1 Fig. Luciferin stability.
Activity of luciferin as a function of time when stored in the CAV-stabilized form either at room temperature (orange triangles) or 55 °C (pink diamonds) or as a liquid at room temperature (blue circles). Data are consistent with a first-order reaction model and are therefore plotted as the natural log of the measured activity versus time. The positive control served as the assay reference and was diluted from 10 to 2.1 µg/mL in triplicate. For each vitrified test article, we eluted 3 scaffolds, and a single dilution series spanning the same concentration range was prepared from each scaffold.
https://doi.org/10.1371/journal.pone.0354533.s001
(DOCX)
S2 Fig. Luciferase stability.
Activity of CAV-stabilized luciferase as a function of time when stored at room temperature (orange triangles) or 55 °C (pink diamonds). Stability trends appear to follow first-order kinetics, and the results are therefore plotted as the natural log of the activity versus time. Data for liquid luciferase formulations not shown. At 55 °C, the luciferase stored as a liquid lost all activity in less than 1 day. At 25 °C, the test results for the liquid luciferase were not reproducible across time points or repeat stability studies. The positive control served as the assay reference and was diluted from 10 to 2.1 µg/mL in triplicate. For each vitrified test article, we eluted 3 scaffolds, and a single dilution series spanning the same concentration range was prepared from each scaffold.
https://doi.org/10.1371/journal.pone.0354533.s002
(DOCX)
S1 Table. Detection of microbial contamination.
Raw signals and reportable results obtained for the microbial detection assay using traditional, liquid reagents versus the CAV-stabilized master mix.
https://doi.org/10.1371/journal.pone.0354533.s003
(DOCX)
S3 Fig. Representative SEM images of the scaffold material (left) and of a vitrified sample (right).
The images show the fibrous nature of the scaffold material with the darker areas representing voids or pores in the material. Post-vitrification, an amorphous glass forms and coats the fibers. This produces a porous glass with high surface area, enabling rapid dissolution.
https://doi.org/10.1371/journal.pone.0354533.s004
(DOCX)
Acknowledgments
The authors thank Dr. Pravansu Mohanty for technical support, Dr. Rachel Elias for SEM imaging, and Dr. Craig L. Duvall for IVIS use. The authors acknowledge technical support from the Michigan Center for Materials Characterization. Schematics were prepared using BioRender.
References
- 1. Deshpande SS. Principles and applications of luminescence spectroscopy. Crit Rev Food Sci Nutr. 2001;41(3):155–224. pmid:11310779
- 2. Dunuweera AN, Dunuweera SP, Ranganathan K. A Comprehensive Exploration of Bioluminescence Systems, Mechanisms, and Advanced Assays for Versatile Applications. Biochem Res Int. 2024;2024:8273237. pmid:38347947
- 3. Kulmala S, Suomi J. Current status of modern analytical luminescence methods. Anal Chim Acta. 2003;500(1–2):21–69.
- 4. Roda A, Pasini P, Mirasoli M, Michelini E, Guardigli M. Biotechnological applications of bioluminescence and chemiluminescence. Trends Biotechnol. 2004;22(6):295–303. pmid:15158059
- 5. Syed AJ, Anderson JC. Applications of bioluminescence in biotechnology and beyond. Chem Soc Rev. 2021;50(9):5668–705. pmid:33735357
- 6. Gleneadie HJ, Dimond A, Fisher AG. Harnessing bioluminescence for drug discovery and epigenetic research. Front Drug Discov. 2023;3:1249507.
- 7. Mali SB. Bioluminescence in cancer research - Applications and challenges. Oral Oncol Rep. 2024;9:100127.
- 8. Coleman SM, McGregor A. A bright future for bioluminescent imaging in viral research. Future Virol. 2015;10(2):169–83. pmid:26413138
- 9. Baljinnyam B, Ronzetti M, Simeonov A. Advances in luminescence-based technologies for drug discovery. Expert Opin Drug Discov. 2023;18(1):25–35. pmid:36562206
- 10. Nante N, Ceriale E, Messina G, Lenzi D, Manzi P. Effectiveness of ATP bioluminescence to assess hospital cleaning: a review. J Prev Med Hyg. 2017;58(2):E177–83. pmid:28900359
- 11. Whitehead KA, Smith LA, Verran J. The detection of food soils and cells on stainless steel using industrial methods: UV illumination and ATP bioluminescence. Int J Food Microbiol. 2008;127(1–2):121–8. pmid:18678428
- 12.
Deininger RA, Lee J. Rapid detection of bacteria in drinking water. In: Omelchenko A, Pivovarov AA, Swindall WJ. Modern tools and methods of water treatment for improving living standards. Dordrecht: Springer. 2005. p. 71–8.
- 13. Reyes S, Rodriguez R, Dikici E, Daunert S, Deo S. Bioluminescence in Clinical and Point-of-Care Testing. Biosensors (Basel). 2025;15(7):422. pmid:40710072
- 14. Mehrabi M, Hosseinkhani S, Ghobadi S. Stabilization of firefly luciferase against thermal stress by osmolytes. Int J Biol Macromol. 2008;43(2):187–91. pmid:18555523
- 15. Ganjalikhany MR, Ranjbar B, Hosseinkhani S, Khalifeh K, Hassani L. Roles of trehalose and magnesium sulfate on structural and functional stability of firefly luciferase. J Mol Catal B-Enzym. 2010;62(2):127–32.
- 16. Koksharov MI, Ugarova NN. Thermostabilization of firefly luciferase by in vivo directed evolution. Protein Eng Des Sel. 2011;24(11):835–44. pmid:21900306
- 17. Lohrasbi-Nejad A, Torkzadeh-Mahani M, Hosseinkhani S. Hydrophobin-1 promotes thermostability of firefly luciferase. FEBS J. 2016;283(13):2494–507. pmid:27191938
- 18. Jazayeri FS, Amininasab M, Hosseinkhani S. Structural and dynamical insight into thermally induced functional inactivation of firefly luciferase. PLoS One. 2017;12(7):e0180667. pmid:28672033
- 19. Pozzo T, Akter F, Nomura Y, Louie AY, Yokobayashi Y. Firefly luciferase mutant with enhanced activity and thermostability. ACS Omega. 2018;3(3):2628–33.
- 20. Shi C, Killoran MP, Hall MP, Otto P, Wood MG, Strauss E, et al. 5,5-Dialkylluciferins are thermal stable substrates for bioluminescence-based detection systems. PLoS One. 2020;15(12):e0243747. pmid:33315907
- 21. Emrich T, Stracke JO, Guo X, Damhjell K, Moelleken J, Vogel R, et al. Increasing robustness, reliability and storage stability of critical reagents by freeze-drying. Bioanalysis. 2021;13(10):829–40. pmid:33890493
- 22. Bhambhani A, Verch T, Jones MC, Retzlaff M, Sinacola JR, Bouaraphan SL. Lyosphere critical reagent kit. 2021. US Patent Application US20210223262A1. https://patents.google.com/patent/US20210223262A1/en
- 23. Molnar A, Lakat T, Hosszu A, Szebeni B, Balogh A, Orfi L, et al. Lyophilization and homogenization of biological samples improves reproducibility and reduces standard deviation in molecular biology techniques. Amino Acids. 2021;53(6):917–28. pmid:34002278
- 24. Privalov PL. Cold denaturation of proteins. Crit Rev Biochem Mol Biol. 1990;25(4):281–305. pmid:2225910
- 25. Bhatnagar BS, Bogner RH, Pikal MJ. Protein stability during freezing: separation of stresses and mechanisms of protein stabilization. Pharm Dev Technol. 2007;12(5):505–23. pmid:17963151
- 26. Mahler H-C, Friess W, Grauschopf U, Kiese S. Protein aggregation: pathways, induction factors and analysis. J Pharm Sci. 2009;98(9):2909–34. pmid:18823031
- 27. Jakubowska E, Lulek J. The application of freeze-drying as a production method of drug nanocrystals and solid dispersions – A review. J Drug Deliv Sci and Technol. 2021;62:102357.
- 28. Shank-Retzlaff M, Taverner YP, Joshi P, Renu S, Chitikela A, Koneru A, et al. Capillary-Mediated Vitrification: A Novel Approach for Improving Thermal Stability of Enzymes and Proteins. J Pharm Sci. 2022;111(8):2280–7. pmid:35235844
- 29. Renu S, Shank-Retzlaff M, Sharpe J, Bronsart L, Mohanty P. Capillary-Mediated Vitrification: Preservation of mRNA at Elevated Temperatures. AAPS J. 2022;24(4):75. pmid:35710853
- 30. Amle S, Radford S, Wang Z, Bronsart L, Mohanty P, Renu S, et al. Use of capillary-mediated vitrification to produce thermostable, single-use antibody conjugates as immunoassay reagents. J Immunol Methods. 2023;516:113460. pmid:36967060
- 31. Shank-Retzlaff M, Abdeen SJ, Bronsart L, Cieslak AN, Cruse JK, Kinne AS, et al. Capillary mediated vitrification is a novel technique that enables storage of antibody critical reagents at ambient temperature: Impact on binding, structure, and laboratory sustainability. J Pharm Biomed Anal. 2024;251:116409. pmid:39208649
- 32. Olgenblum GI, Hutcheson BO, Pielak GJ, Harries D. Protecting Proteins from Desiccation Stress Using Molecular Glasses and Gels. Chem Rev. 2024;124(9):5668–94. pmid:38635951
- 33. Kincaid VA, Wang H, Sondgeroth CA, Torio EA, Ressler VT, Fitzgerald C, et al. Simple, Rapid Chemical Labeling and Screening of Antibodies with Luminescent Peptides. ACS Chem Biol. 2022;17(8):2179–87. pmid:35862857
- 34. Torio EA, Ressler VT, Kincaid VA, Hurst R, Hall MP, Encell LP, et al. Development of a rapid, simple, and sensitive point-of-care technology platform utilizing ternary NanoLuc. Front Microbiol. 2022;13:970233. pmid:36386626
- 35. Hall MP, Unch J, Binkowski BF, Valley MP, Butler BL, Wood MG, et al. Engineered luciferase reporter from a deep sea shrimp utilizing a novel imidazopyrazinone substrate. ACS Chem Biol. 2012;7(11):1848–57. pmid:22894855
- 36. Brabek J, Jakubek M, Vellieux F, Novotny J, Kolar M, Lacina L. Interleukin-6: Molecule in the Intersection of Cancer, Ageing and COVID-19. Int J Mol Sci. 2020;21(21):7937.
- 37. Lin S-W, Shen C-F, Liu C-C, Cheng C-M. A Paper-Based IL-6 Test Strip Coupled With a Spectrum-Based Optical Reader for Differentiating Influenza Severity in Children. Front Bioeng Biotechnol. 2021;9:752681. pmid:34692664
- 38. Grebenciucova E, VanHaerents S. Interleukin 6: at the interface of human health and disease. Front Immunol. 2023;14:1255533. pmid:37841263
- 39. Walters RH, Bhatnagar B, Tchessalov S, Izutsu K-I, Tsumoto K, Ohtake S. Next generation drying technologies for pharmaceutical applications. J Pharm Sci. 2014;103(9):2673–95. pmid:24916125
- 40. Li J, Wang H, Wang L, Yu D, Zhang X. Stabilization effects of saccharides in protein formulations: A review of sucrose, trehalose, cyclodextrins and dextrans. Eur J Pharm Sci. 2024;192:106625. pmid:37918545
- 41. Gaidhani KA, Harwalkar M, Bhambere D, Nirgude PS. Lyophilization/freeze drying - a review. World J Pharm Res. 2015;4(8):516–43.
- 42. Xu Q, Deng B, Wang Y, Liu W, Chen G. Small, affordable, ultra-low-temperature vapor-compression and thermoelectric hybrid freezer for clinical applications. Cell Rep Phys Sci. 2023;4(12):101735.