Longitudinal ultrasound assessment of jugular venous pressure reliably detects hypervolemia: An observational study in healthy volunteers

  • Karin Vogt,

    Roles Conceptualization, Data curation, Investigation, Methodology, Project administration, Software, Validation, Writing – original draft

    Affiliation Department of Endocrinology, Diabetology and Metabolism, University Hospital Basel, Basel, Switzerland

  • David Widmer,

    Roles Data curation, Investigation, Methodology, Validation, Writing – review & editing

    Affiliation Intensive Care Unit, University Hospital Basel, Basel, Switzerland

  • Mark Kirsch,

    Roles Conceptualization, Investigation, Methodology, Supervision, Validation, Writing – review & editing

    Affiliation Division of Internal Medicine, University Hospital Basel, Basel, Switzerland

  • Laura Potasso

    Roles Conceptualization, Data curation, Formal analysis, Funding acquisition, Investigation, Resources, Software, Supervision, Validation, Visualization, Writing – review & editing

    laura.potasso@usb.ch

    Affiliations Department of Endocrinology, Diabetology and Metabolism, University Hospital Basel, Basel, Switzerland, University of Basel, Basel, Switzerland

Abstract

Background

Ultrasound (US) assessment of the internal jugular vein (IJV) to measure ultrasound jugular venous pressure (uJVP) has been proposed as a promising non-invasive tool for evaluating hydration status. However, a standardized method for uJVP measurement is currently lacking.

Methods

This cross-sectional study compared 4 previously described IJV-US methods in healthy, euvolemic volunteers. Methods 1 and 2 use transverse view, measuring uJVP where IJV is (1)smaller than common carotid artery or (2)collapsed throughout the respiratory cycle. Methods 3a/b use longitudinal view, measuring at the tip(3a) or base(3b) of the IJV tapering portion. US assessment was conducted by 2 independent, blinded investigators at individualized head-of-bed elevation angle according to standardized procedures. Endpoints were: a)percentage of participants with measurements differing >1 standard deviation(SD) from the previously described mean (6.5 ± 1.5 cmH₂O) for euvolemic patients, b)reproducibility and c)feasibility.

Results

30 participants(50% females) were included. Median uJVP(IQR) was 3.8 cmH₂O(3.0–4.0), 4.5 cmH₂O(3.8–5.3), 3.9 cmH₂O(3.3–4.5), and 3.3 cmH₂O(2.8–3.5) for methods 1-3b. uJVP differed >1SD in 17/30(56.7%), 3/14(21.4%), 14/29(51.7%) and 25/30available assessment pairs(83.3%), respectively. Reproducibility for methods 1-3b was 18/30(60.0%), 7/30(23.3%), 19/30(63.3%) and 25/30 participants(83.3%). Assessment was feasible in 30/30 participants(100%) for methods 1 and 3b, 29/30(97%) for method 3a and 14/30(47%) for method 2.

Conclusion

The longitudinal method measuring uJVP at the base of tapering portion showed the highest reproducibility and feasibility. Overall, measured uJVP values were consistently within the upper threshold for normal central venous pressure, but lower than the previously described values, supporting the use of uJVP for detecting hypervolemia, but not hypovolemia.

1. Introduction

Hydration status assessment is essential in many clinical conditions, such as dyspnea, electrolyte imbalances, hypotension, sepsis, heart, liver and kidney failure as well as in the perioperative setting. However, it represents a challenge for the physician, as to date no validated, effective, non-invasive global hydration status assessment protocol exist [13]. Ultrasound (US) is often employed for non-invasive hydration status evaluation [1,2,411]. Inferior vena cava (IVC) US [12,13] is the most widely used method, supplemented by lung US [14,15], portal/hepatic/renal vein flow assessment (venous excess US, VExUS) [16,17], and cardiac US, depending on clinical suspicion and technical applicability [1,10,11]. Yet, abdominal US assessments can be challenging in the presence of different factors such as obesity, wound dressings or bowel gas [18]. Furthermore, VExUS and cardiac US are time-consuming and complex, limiting their feasibility in ward or emergency settings because they require specially trained physicians [1921].

Clinical examination of the neck veins has long been a fundamental component of bedside hydration status assessment [22], though the external jugular veins are not always reliably visualized [23,24]. Internal jugular vein (IJV) US and the estimation of US jugular venous pressure (uJVP) as a surrogate for the invasively measured central venous pressure (CVP) and right atrial pressure (RAP) in use for hydration status assessment, have emerged as promising non-invasive alternatives, given their feasibility and ease of learning [2527]. However, despite a growing body of literature on IJV-US, there is no consensus on the methodology or standardized definition of uJVP.

To date, there are 4 commonly used approaches to visualize and measure uJVP, of which 2 use transverse and 2 longitudinal view [2836]. The primary objective of this methodological, praxis-oriented study was therefore to evaluate the accuracy of the four previously described uJVP measurement methods. Additionally, we analyzed the reproducibility and feasibility of these methods.

2. Methods

2.1. Trial design and oversight

This investigator-initiated, single-center, observational cross-sectional study with blinded assessment was conducted in healthy adult volunteers. The study took place from 22nd October 2024–3rd December 2024 at University Hospital of Basel, Switzerland, and was funded by a research grant from the Margot and Erich Goldschmidt & Peter René Jacobson-Foundation. Ethical approval was obtained from the Ethics Committee Northwest Switzerland (EKNZ 2024−01640), and the study was registered on ClinicalTrials.gov (Identifier: NCT06706960). The trial steering committee was composed of the authors who designed and supervised the trial. Data quality for each visit was reviewed by an investigator who had not participated in that particular visit. All participants provided written informed consent before enrollment. Expenses related to the study visit were reimbursed, no financial compensation was offered.

2.2. Inclusion and exclusion criteria

Thirty healthy adults aged 18 years or older, 15 males and 15 females, were eligible. Exclusion criteria were selected to ensure an euvolemic study population. Individuals with conditions affecting body water homeostasis, those presenting with abnormal vital signs, or with possible confounding factors were excluded. For detailed information see Table 1.

2.3. Trial procedures

At the start of the visit, vital signs and fluid intake data were recorded. All participants were independently examined on the same day by two blinded examiners trained in Point-of-Care Ultrasound (POCUS). The examiners consisted of 2 internal medicine residents, 1 senior internal medicine physician, and 1 endocrinologist, all following a standardized protocol. To ensure consistency, the examiners completed two 2-hour training sessions on the protocol before the study. US assessments were conducted using the linear transducer of a handheld GE® VScan Air CL probe, connected to an Apple® iPad. The right IJV was scanned with participants positioned on a bed in a semi-recumbent posture (Fig 1), head held in a neutral or slightly leftward-turned position (no more than 30°).

thumbnail
Fig 1. Internal jugular vein assessment steps.

A: graphic showing participant and probe positioning (transverse assessments in green, longitudinal assessments in yellow); B: graphic showing the technique applied to measure internal jugular vein (IJV) height in cmH₂O.

https://doi.org/10.1371/journal.pone.0346162.g001

The head-of-bed elevation angle was individually adjusted using a standardized procedure under US guidance before initiating the examination and was kept consistent for both investigators. The goal was to visualize uJVP in the mid-third of the neck to minimize confounding factors such as proximity to the confluence with the subclavian vein or the presence of venous valves, as previously described in the literature [28,37,38].

uJVP was measured in centimeters of water (cmH₂O). For each method, IJV height was marked on the skin using a skin marker. Sternal angle served as the reference point for measurements (above= positive values, below= negative values), a horizontal reference line was projected onto the body surface. Measurements were taken vertically and rounded to the nearest 0.5 cm. In accordance with standard practice, 5 cmH₂O (the assumed constant distance from right atrium to sternal angle) were added to the measured IJV height to calculate uJVP [23]. The time from the start of the examination to IJV height skin mark was measured and recorded for each method. Acceptance of each US protocol by the participant was assessed using a visual analogue scale (VAS, 10 = best). Predefined US images were stored and used by the data safety monitoring board members to verify the results. To minimize potential bias, the order of measurement methods was alternated between participants and kept in the same sequence for both investigators. All methods began with the ultrasound probe positioned at the clavicle, scanning upwards.

The ultrasound methods for measuring IJV height were as follows (Fig 2):

thumbnail
Fig 2. Ultrasound jugular venous pressure assessment methods.

Ultrasound images for methods 1 to 3b. IJV = internal jugular vein, A = artery, V = vein, CCA = common carotid artery.

https://doi.org/10.1371/journal.pone.0346162.g002

  1. 1) Transverse view, measurement at the point where IJV is smaller than CCA throughout the entire respiratory cycle (called method 1)
  2. 2) Transverse view, measurement at the point where IJV is completely collapsed throughout the entire respiratory cycle (method 2)
  3. 3) Longitudinal view, measurement at either the tip (method 3a) or the base (method 3b) of the IJV tapering portion

Because identical probe positioning was used for the last 2 methods, they were summarized as “method 3, a + b”.

2.4. Endpoints

The primary aim of the study was to assess the accuracy and reliability of the current used uJVP assessment methods by comparing the assessment results with the existing normal values reported in the literature. Therefore, the primary endpoint was set at the proportion of participants whose measurements differed >1 standard deviation (SD) from the expected uJVP value, defined as 6.5 + /- 1.5 cmH₂O independent from sex for each of the 4 applied methods. This value based on common clinical practice [39] as well as previous studies. Those studies included a) a study by Socransky et al in supposedly euvolemic emergency department patients, defining normal uJVP using method 3b [35], b) a systematic review by Wang et al comparing different uJVP and other IJV-US cut-off values for hypo- and hypervolemia [25], and c) a study by Wang et al applying method 1 for uJVP assessment in patients scheduled for right heart catheterization [36]. The null hypothesis stated that there would be no difference compared to the expected value, i.e., that all methods will be equally accurate in measuring uJVP. The percentage was calculated by dividing the number of participants whose measurements differed by > 1 SD from the expected uJVP value by the number of participants for whom both investigators were able to measure uJVP (= number of available assessment pairs) and then multiplying the result by 100.

We performed a sensitivity analysis by investigating possible sex-related differences in the uJVP measurements.

Secondary endpoints included:

  1. a) Reproducibility: inter-observer agreement for measurements of each method
  2. b) Analysis of feasibility across methods

2.5. Statistical analysis

The sample size was determined assuming an expected uJVP mean ± SD of 6.5 ± 1.5 cmH₂O independent from sex [25,35,36]. To avoid bias due to measurement deviations, the study population was calculated assuming a mean in the study group of 5.5 or below or 7.5 cmH₂O or above. Using the equation

with alpha error of 0.05 and a power of 95%, 29 participants were needed to verify the hypothesis. We rounded it at 30 to preserve the 50% sex distribution.

Data was collected in the online database Redcap®. Analysis was based on full analysis set (FAS).

We conducted a descriptive analysis. Mean and standard deviation (SD) were calculated for age, while median and interquartile range (IQR) were used for all other continuous variables. Percentages were used for binary variables. We compared uJVP values to the predefined threshold using a one-sample Wilcoxon signed-rank test. Differences between methods within the same participants were assessed using a paired Wilcoxon signed-rank test. Sex differences in uJVP values were evaluated using the Wilcoxon rank-sum test (Mann–Whitney U test). Results were visualized using box plots. The percentage of participants differing >1 SD from the expected value was calculated. Reproducibility was calculated in percentage of participants having the same evaluation from the two independent investigators with a tolerance of maximum 1 cm, divided by the total number of participants. Additionally, Bland-Altman analysis was performed to compare the measurements obtained by the two investigators, allowing assessment of bias and limits of agreement. Two-way Intraclass Correlation Coefficient (ICC) with absolute agreement, 95% confidence intervals (CI) and p-value were reported for all methods. Feasibility was assessed on the base of a) percentage of participants in which the IJV-height visualization was possible as defined by the method, divided by the total number of participants; b) duration of each available assessment in seconds and c) acceptance of the different US-assessment protocols by all the participants based on VAS. All analyses were performed using R statistical software [40].

2.6. Writing process/ use of artificial intelligence

ChatGPT (GPT-4o) was used to verify grammatical accuracy. Fig 1 was generated using artificial intelligence (DALL·E 3 via ChatGPT (GPT-4o)), based on photos of the study setting.

3. Results

3.1. Participant enrollment and baseline characteristics

Thirty volunteers (15 men, 15 women) were recruited. The mean age was 33.0 years (SD ± 10.2), and the median BMI was 22.9 kg/m2 (IQR 21.6–25.7). In total, 11 participants (36.7%) were taking medications. Most were females (n = 9), with eight using oral contraceptives and one taking levothyroxine. Of the two males, one was taking ibuprofen for a minor injury, and the other a low-dose selective serotonin reuptake inhibitor (SSRI). Detailed baseline characteristics are depicted in Table 2. The head-of-bed elevation angle during assessment was 20° in 33%, 15° in 30%, 10° in 20% and 25–45° in 17% of participants.

3.2. Primary endpoint

Median uJVP was 3.8 cmH₂O (IQR 3.0–4.0) for method 1, 4.5 cmH₂O (IQR 3.8–5.3) for method 2, 3.9 cmH₂O (IQR 3.3–4.5) for method 3a and 3.3 cmH₂O (IQR 2.8–3.5) for method 3b. The percentage of participants whose measurements differed by >1 SD from the expected value was 56.7%, 21.4%, 51.7% and 83.3% for methods 1-3b, respectively (Table 3, Fig 3), all below the expected value.

thumbnail
Fig 3. Ultrasound based Jugular Venous Pressure (uJVP) results.

Box-plot showing uJVP-results according to method. y-axis: in cmH₂O; dotted lines = expected mean 6.5 cmH₂O (red) ± 1 SD (orange) based on previous studies.

https://doi.org/10.1371/journal.pone.0346162.g003

3.3. Sensitivity analysis

Although altogether male participants showed slightly higher uJVP values than females, there was no relevant differences in the longitudinal methods 3a (median 4.1 vs 3.7, p = 0.37) or 3b (median 3.4 vs 3.0 p = 0.28), whereas both transversal methods 1 (median 4.0 vs 3.1, p = 0.02) and 2 (median 5.4 vs 4.0, p = 0.01) showed a sex-related difference.

3.4. Secondary endpoints

Method 2 yielded significantly higher uJVP values compared to method 1 (uJVP 4.5 cmH₂O vs 3.8 cmH₂O, p = 0.01) and method 3b (uJVP 4.5 cmH₂O vs 3.3 cmH₂O, p = 0.0007) (Fig 3).

Bland-Altman analysis showed that Method 3b had the smallest bias (−0.25 vs −0.30 for method 1, −0.83 for method 2, and −0.42 for method 3a) and the narrowest limits of agreement (−2.26 to 1.76), indicating the highest interrater reproducibility (Fig 4,Table 4).

thumbnail
Fig 4. Bland-Altmann analysis.

Bland-Altman plots for methods 1 to 3b. Orange vertical line identifies 5 cmH2O. uJVP = ultrasound jugular venous pressure, LoA = limits of agreement.

https://doi.org/10.1371/journal.pone.0346162.g004

ICC with absolute agreement was 0.32 (95% CI −0.031 to 0.61, p = 0.04) for method 1; 0.45 (95% CI −0.12 to 0.83, p = 0.06) for method 2; 0.130 (95% CI −0.24 to 0.46, p = 0.24) for method 3a; and 0.49 (95% CI 0.17 to 0.72, p = 0.002) for method 3b.

The number of participants whose measurement differences between the two investigators fell within 1 cm was 18/30 (60.0%), 7/30 (23.3%), 19/30 (63.3%), and 25/30 (83.3%) for Methods 1 through 3b, respectively (Table 3).

IJV-height visualization was possible in 100% of cases for methods 1 and 3b, 97% for method 3a and 47% for method 2. The median duration was 29.5 s (IQR 20.8–41.3) for method 1, 32.5 s (IQR 17.0–45.0) for method 2, and 38.5 s (23.5–62.0) for method 3a/b. The median tolerability (VAS, 10 = best) was 9.8 (IQR 8.6–10.0) for method 1, 9.5 (IQR 8.6–10.0) for method 2, and 9.5 (IQR 9.0–10.0) for method 3 a/b (Table 3).

3.5. Serious events

No serious events occurred during the study.

3.6. Summary of the results

In our observational, methodological study of 30 healthy volunteers (50% female) comparing four uJVP methods, we found:

  • uJVP measurements: All values were <8 cmH₂O, correctly below the hypervolemia threshold, whereas the hypovolemia threshold was exceeded in 56.7%, 21.4%, 51.7%, and 83.3% for methods 1–3b, respectively.
  • Best-performing method: Longitudinal assessment at the base of the IJV tapering portion (method 3b) showed the highest interrater agreement (25/30, 83.3%) and could be visualized in 30/30 participants. Bland-Altmann analysis supported the use of Method 3b for consistent uJVP assessment between investigators (bias −0.25, limits of agreement −2.26 to 1.76). Acquisition time was 38.5 seconds, and tolerability (VAS 9.5) was comparable to other methods.
  • Practical recommendation: Individualized head-of-bed elevation, visualizing uJVP in the mid-third of the neck, can help minimize confounding factors such as proximity to the subclavian vein or venous valves.

4. Discussion

Our study aimed to systematically compare different IJV-US methods for hydration status assessment, yielding three main findings. First, the uJVP values measured in our cohort of healthy volunteers were lower than those previously reported in the literature as normal values. Second, reproducibility varies substantially between the four applied methods, with the best results for the longitudinal assessment at the base of the tapering portion (method 3b). Third, ultrasound assessment for JVP is well tolerated by the examined person and the required duration short enough to represent an easy and applicable point-of-care test.

The first finding of our study regards the uJVP values in euvolemic patients: All the assessed methods consistently yielded uJVP measurements below the expected normal values. Those values were based on previous studies investigating and defining uJVP, amongst which there is a great heterogeneity regarding the applied ultrasound method and reference standard [25,30,35,36]. Socransky et al [35] used longitudinal US view, measuring at the base of IJV taper (equivalent to method 3b), to examine 77 supposedly euvolemic emergency department patients, and found mean uJVP to be 6.35 cm. Siva et al [30] used longitudinal and transverse US view, measuring at the tip of the tapering portion (equivalent to a combination of methods 3a/2), to correlate uJVP with invasively measured CVP in a collective of 44 patients. Wang et al [36] used transverse US view, measuring as soon as IJV was smaller than CCA (equivalent to method 1), to correlate uJVP with invasively measured RAP in a sample of 100 patients. All these studies differed from our patient per age (mean age above 45 years old), BMI (mean BMI above 25 kg/m2) and head of bed elevation (fixed at 45°); and in most cases, uJVP was observed just above the clavicle. This is the site of confluence into the subclavian vein, where the visualized tapering of IJV may be overestimated due to a) negative intrathoracic pressure splinting the vein open as it enters the chest cavity, and b) venous valves present in this region [28,37]. In our study, all participants had uJVP values <8 cmH₂O and were thus correctly identified as non-hypervolemic according to the described cut-offs. Therefore, we support the conclusion of a metanalysis published by Wang et al in 2022 [25] that uJVP is useful for assessing hypervolemia, but not suitable for identifying hypovolemia.

One could argue that our lower values were obtained because we did not use a fixed head-to-bed elevation angle. However, we advocate for our approach using individualized head-of-bed elevation angle, as it reduces confounding factors through anatomical influences; reason why it has recently also been applied in other studies [38]. In addition, this is a pragmatic, practice-oriented approach, as maintaining a fixed head-of-bed elevation angle is not always feasible in routine clinical settings.

The second result of our study was that longitudinal assessment at the base of the tapering portion showed the highest reproducibility, both in the Bland-Altman analysis and when using a narrower limit of agreement of 1 cm. Moreover, it showed the highest ICC when looking at absolute agreement. One possible explanation is the reduced susceptibility of this method to probe pressure, a known confounding factor for IJV-US assessment [41], in contrast to the transversal methods 1 and 2. Further, IJV layering (i.e., the effect that a non-collapsed IJV is visualized due to enhanced venous return from cranial veins, independent of right atrial pressure [37]) may alter the location of the tip (method 3a) but not the base of the tapering portion. One could argue that a reproducibility of about 83% may still be insufficient for reliable use in clinical practice. However, IVC-US, which is currently most used for US hydration status assessment, has also shown only a moderate reproducibility [38,42,43]. Importantly, longitudinal uJVP measurements showed robust results with no significant differences between male and female participants, in contrast to transverse measurements. As uJVP is an indirect measure of CVP, which shows no sex-related differences [44], no differences were expected. This again underlines the greater reliability of longitudinal measurements compared to transverse ones.

The third finding of our study concerns another key feature of a standardized method: feasibility. In our study, Method 3b was well tolerated and yielded results in 100% of participants. Additionally, the assessment duration had a median time of less than one minute, supporting previous evidence [41] suggesting IJV-US as a highly feasible tool for clinical practice. In contrast, method 2, using a transverse visualization, despite showing similar results about tolerability and assessment duration, produced results in only 47% of participants, as no complete IJV collapse could be visualized until the mandibular angle in the remaining cases.

This study has several limitations, beginning with its small sample size and observational design. However, to the best of our knowledge, it represents the largest cohort in which all four previously described methods were systematically assessed by two independent, blinded investigators. Further, we did not compare our uJVP measurements to invasive volume status assessment methods (i.e., CVP, RAP) or other proposed ultrasound evaluations (i.e., IVC-US, echocardiography), to confirm euvolemic status. However, this was consistent with the study’s primary objective which was methodological in nature. Additionally, our cohort consisted of young, healthy volunteers without known medical conditions affecting volume status and without clinical signs of hypo- or hypervolemia, making it likely that participants were euvolemic. Another potential limitation is that the investigators conducting the ultrasound assessments had only basic ultrasound skills. However, all investigators underwent two structured training sessions and adhered strictly to a standardized examination protocol. Moreover, this in fact enhances the study’s generalizability.

5. Conclusion

In conclusion, we advocate for a standardized approach to IJV-US in future studies and clinical applications. Our findings support the use of a longitudinal IJV view, with uJVP measurement taken at the base of the tapering portion. Head-of-bed elevation angle should be adjusted individually to minimize confounding factors. Further studies are needed to correlate uJVP measured with this method to invasively assessed CVP or RAP, allowing for consistent definition of hypovolemia.

Acknowledgments

We sincerely thank all the participants for their valuable contribution to this study. We are also deeply grateful to the dedicated staff of the outpatient day clinic for their support and assistance throughout the assessment process.

References

  1. 1. Pellicori P, Platz E, Dauw J, Ter Maaten JM, Martens P, Pivetta E, et al. Ultrasound imaging of congestion in heart failure: examinations beyond the heart. Eur J Heart Fail. 2021;23(5):703–12. pmid:33118672
  2. 2. Prager R, Arntfield R, Wong MYS, Ball I, Lewis K, Rochwerg B, et al. Venous congestion in septic shock quantified with point-of-care ultrasound: a pilot prospective multicentre cohort study. Can J Anaesth. 2024.
  3. 3. Deißler L, Wirth R, Frilling B, Janneck M, Rösler A. Hydration status assessment in older patients. Dtsch Arztebl Int. 2023;120(40):663–9.
  4. 4. Kim D-W, Chung S, Kang W-S, Kim J. Diagnostic Accuracy of Ultrasonographic Respiratory Variation in the Inferior Vena Cava, Subclavian Vein, Internal Jugular Vein, and Femoral Vein Diameter to Predict Fluid Responsiveness: A Systematic Review and Meta-Analysis. Diagnostics (Basel). 2021;12(1):49. pmid:35054215
  5. 5. Longino A, Martin K, Leyba K, Siegel G, Thai TN, Riscinti M, et al. Prospective evaluation of venous excess ultrasound for estimation of venous congestion. Chest. 2024;165(3):590–600. pmid:37813180
  6. 6. Singla D, Gupta B, Varshney P, Mangla M, Walikar BN, Jamir T. Role of carotid corrected flow time and peak velocity variation in predicting fluid responsiveness: a systematic review and meta-analysis. Korean J Anesthesiol. 2023;76(3):183–93. pmid:36377330
  7. 7. Argaiz ER, Koratala A, Reisinger N. Comprehensive assessment of fluid status by point-of-care ultrasonography. Kidney360. 2021;2(8):1326–38. pmid:35369665
  8. 8. Orso D, Guglielmo N, Federici N, Cugini F, Ban A, Mearelli F, et al. Accuracy of the caval index and the expiratory diameter of the inferior vena cava for the diagnosis of dehydration in elderly. J Ultrasound. 2016;19(3):203–9.
  9. 9. Wong A, Olusanya O, Watchorn J, Bramham K, Hutchings S. Utility of the Venous Excess Ultrasound (VEXUS) score to track dynamic change in volume status in patients undergoing fluid removal during haemodialysis - the ACUVEX study. Ultrasound J. 2024;16(1):23. pmid:38538806
  10. 10. Kaptein EM, Kaptein MJ. Inferior vena cava ultrasound and other techniques for assessment of intravascular and extravascular volume: an update. Clin Kidney J. 2023;16(11):1861–77. pmid:37915939
  11. 11. Elhassan MG, Chao PW, Curiel A. The conundrum of volume status assessment: revisiting current and future tools available for physicians at the bedside. Cureus. 2021;13(5):e15253. pmid:34188992
  12. 12. Squizzato A, Maroni L, Marrazzo C, Riva N, Guasti L. Diagnostic accuracy of inferior vena cava ultrasound for heart failure in patients with acute dyspnoea: a systematic review and meta-analysis. Emerg Med J. 2021;38(3):232–9. pmid:33298601
  13. 13. Orso D, Paoli I, Piani T, Cilenti FL, Cristiani L, Guglielmo N. Accuracy of ultrasonographic measurements of inferior vena cava to determine fluid responsiveness: a systematic review and meta-analysis. J Intensive Care Med. 2020;35(4):354–63. pmid:29343170
  14. 14. Volpicelli G, Elbarbary M, Blaivas M, Lichtenstein DA, Mathis G, Kirkpatrick AW, et al. International evidence-based recommendations for point-of-care lung ultrasound. Intensive Care Med. 2012;38(4):577–91. pmid:22392031
  15. 15. Platz E, Jhund PS, Girerd N, Pivetta E, McMurray JJV, Peacock WF, et al. Expert consensus document: Reporting checklist for quantification of pulmonary congestion by lung ultrasound in heart failure. Eur J Heart Fail. 2019;21(7):844–51. pmid:31218825
  16. 16. Beaubien-Souligny W, Rola P, Haycock K, Bouchard J, Lamarche Y, Spiegel R, et al. Quantifying systemic congestion with Point-Of-Care ultrasound: development of the venous excess ultrasound grading system. Ultrasound J. 2020;12(1):16. pmid:32270297
  17. 17. Longino AA, Martin KC, Leyba KR, McCormack L, Siegel G, Sharma VM, et al. Reliability and reproducibility of the venous excess ultrasound (VExUS) score, a multi-site prospective study: validating a novel ultrasound technique for comprehensive assessment of venous congestion. Crit Care. 2024;28(1):197. pmid:38858766
  18. 18. Di Nicolò P, Tavazzi G, Nannoni L, Corradi F. Inferior vena cava ultrasonography for volume status evaluation: an intriguing promise never fulfilled. J Clin Med. 2023;12(6):2217. pmid:36983218
  19. 19. Koratala A, Romero-González G, Soliman-Aboumarie H, Kazory A. Unlocking the potential of VExUS in Assessing Venous Congestion: The Art of Doing It Right. Cardiorenal Med. 2024;14(1):350–74.
  20. 20. Koratala A, Reisinger N. Venous excess doppler ultrasound for the nephrologist: pearls and pitfalls. Kidney Med. 2022;4(7):100482. pmid:35707749
  21. 21. Wiegers SE, Ryan T, Arrighi JA, Brown SM, Canaday B, Damp JB, et al. 2019 ACC/AHA/ASE advanced training statement on echocardiography (revision of the 2003 ACC/AHA clinical competence statement on echocardiography): a report of the ACC competency management Committee. Circ: Cardiovascular Imaging. 2019;12(7).
  22. 22. Rizkallah J, Jack M, Saeed M, Shafer LA, Vo M, Tam J. Non-invasive bedside assessment of central venous pressure: scanning into the future. PLoS One. 2014;9(10):e109215. pmid:25279995
  23. 23. McGee SR. Physical examination of venous pressure: a critical review. Am Heart J. 1998;136(1):10–8.
  24. 24. Breidthardt T, Moreno-Weidmann Z, Uthoff H, Sabti Z, Aeppli S, Puelacher C, et al. How accurate is clinical assessment of neck veins in the estimation of central venous pressure in acute heart failure? Insights from a prospective study. Eur J Heart Fail. 2018;20(7):1160–2. pmid:29314487
  25. 25. Wang MK, Piticaru J, Kappel C, Mikhaeil M, Mbuagbaw L, Rochwerg B. Internal jugular vein ultrasound for the diagnosis of hypovolemia and hypervolemia in acutely ill adults: a systematic review and meta-analysis. Intern Emerg Med. 2022;17(5):1521–32. pmid:35718838
  26. 26. Ammirati E, Marchetti D, Colombo G, Pellicori P, Gentile P, D’Angelo L, et al. Estimation of right atrial pressure by ultrasound-assessed jugular vein distensibility in patients with heart failure. Circ Heart Fail. 2024;17(2):e010973. pmid:38299348
  27. 27. Pellicori P, Kallvikbacka-Bennett A, Dierckx R, Zhang J, Putzu P, Cuthbert J, et al. Prognostic significance of ultrasound-assessed jugular vein distensibility in heart failure. Heart. 2015;101(14):1149–58. pmid:26006717
  28. 28. Lipton B. Estimation of central venous pressure by ultrasound of the internal jugular vein. Am J Emerg Med. 2000;18(4):432–4. pmid:10919533
  29. 29. Deol GR, Collett N, Ashby A, Schmidt GA. Ultrasound accurately reflects the jugular venous examination but underestimates central venous pressure. Chest. 2011;139(1):95–100. pmid:20798190
  30. 30. Siva B, Hunt A, Boudville N. The sensitivity and specificity of ultrasound estimation of central venous pressure using the internal jugular vein. J Crit Care. 2012;27(3):315.e7-11. pmid:22137379
  31. 31. Avcil M, Kapci M, Dagli B, Omurlu IK, Ozluer E, Karaman K, et al. Comparision of ultrasound-based methods of jugular vein and inferior vena cava for estimating central venous pressure. Int J Clin Exp Med. 2015;8(7):10586–94. pmid:26379848
  32. 32. Nik Muhamad NA, Safferi RS, Robertson CE. Internal jugular vein height and inferior vena cava diameter measurement using ultrasound to determine central venous pressure: a correlation study. Med J Malaysia. 2015;70(2):63–6. pmid:26162379
  33. 33. Kerleroux B, Pasco J, Dupuis M, Eustache M, Lemrabott A, Jouzel C, et al. Ultrasonographic assessment of the internal jugular vein for the estimation of central venous pressure in hemodialysis patients: A preliminary study. J Clin Ultrasound. 2018;46(4):253–8. pmid:29314087
  34. 34. Henriot B, Scanff A, Sebillot M, Lautredoux F, Jego P. Jugular vein ultrasound for estimating changes in blood volume: a new tool for all physicians. Intern Emerg Med. 2021;16(7):2005–7. pmid:34196887
  35. 35. Socransky SJ, Wiss R, Robins R, Anawati A, Roy M-A, Yeung IC. Defining normal jugular venous pressure with ultrasonography. CJEM. 2010;12(4):320–4. pmid:20650024
  36. 36. Wang L, Harrison J, Dranow E, Aliyev N, Khor L. Accuracy of ultrasound jugular venous pressure height in predicting central venous congestion. Ann Intern Med. 2022;175(3):344–51. pmid:34958600
  37. 37. Arishenkoff D. Evaluation of the JVP with PoCUS - An In-Depth Guide. 2023.
  38. 38. Klangthamneam S, Meemook K, Petnak T, Sonkaew A, Assavapokee T. Correlation between right atrial pressure measured via right heart catheterization and venous excess ultrasound, inferior vena cava diameter, and ultrasound-measured jugular venous pressure: a prospective observational study. Ultrasound J. 2024;16(1):50. pmid:39612127
  39. 39. Assavapokee T, Thadanipon K. Examination of the Neck Veins. N Engl J Med. 2020;383(24):e132. pmid:33296562
  40. 40. R: The R Project for Statistical Computing.
  41. 41. Wang L, Harrison J, Khor L. Volume status: a preload assessment by ultrasound of the inferior vena cava and jugular venous pulsation. Med Clin North Am. 2025;109(1):121–35. pmid:39567089
  42. 42. Akkaya A, Yesilaras M, Aksay E, Sever M, Atilla OD. The interrater reliability of ultrasound imaging of the inferior vena cava performed by emergency residents. Am J Emerg Med. 2013;31(10):1509–11. pmid:24012423
  43. 43. Bowra J, Uwagboe V, Goudie A, Reid C, Gillett M. Interrater agreement between expert and novice in measuring inferior vena cava diameter and collapsibility index. Emerg Med Australas. 2015;27(4):295–9. pmid:26072675
  44. 44. Franke WD, Johnson CP, Steinkamp JA, Renwei W, Halliwill JR. Cardiovascular and autonomic responses to lower body negative pressure. Clin Auton Res. 2003;36–44.