Figures
Abstract
Background
Physiological pacing targeting the cardiac conduction system is increasingly being adopted as an alternative to conventional right ventricular (RV) pacing for the treatment of bradyarrhythmias, although its effects on ventricular repolarization remain underexplored.
Objective
This study evaluates depolarization and repolarization responses to different pacing techniques using ultra-high-frequency electrocardiograms (UHF-ECGs).
Methods
Temporary pacing was performed at different cardiac areas in 178 patients with bradycardia. Depolarization was assessed via QRS duration (QRSd), QRS area (QRSa), ventricular dyssynchrony (e-DYS), and activation time dispersion (dAT computed from leads V1-V6 and dAT4-6 from leads V1-V6). Repolarization was analyzed using the corrected QT interval (QTc), T-wave area (Ta), Periodic Repolarization Dynamics (PRD), and repolarization time dispersion (dRTc and dRTc4-6).
Results
His bundle pacing (HBP) preserved ventricular activation patterns similar to spontaneous rhythm. Left bundle branch pacing (LBBP) induced moderate depolarization changes, primarily due to delayed right ventricular activation, while maintaining left ventricular synchrony. In particular, dAT showed no significant differences between HBP and spontaneous rhythm, while differences between LBBAP and spontaneous rhythm were significant but below 7 ms in median. When restricted to the left ventricle (LV), no significant differences in dAT4-6 were found between LBBAP and spontaneous rhythm. Importantly, e-DYS for HBP showed similar values to spontaneous rhythm, LBBP led to a significant reduction (median differences approximately 20 ms), and RVP was associated with a significant increase (above 15 ms in median). In line with these results, QRSd and QRSa showed the largest values for RVP. In terms of ventricular repolarization, median differences in the QTc interval between pacing modes and spontaneous rhythm were below 3 ms for HBP, above 1 ms for LBBP, and above 20 ms for RVP. All pacing modes led to a reduction in PRD, with the most marked reductions observed for LBBP, particularly for selective LBBP, with median changes with respect to spontaneous rhythm of 4.6 degrees. RT and RTc showed similar trends for all pacing techniques. Ta, however, showed median differences with respect to spontaneous rhythm above 100 and 34 Vs when pacing the RV at the apex and the septum, respectively, whereas such median differences were below 16
Vs for HBP and below 2
Vs for LBBP.
Conclusion
Physiological stimulation via HBP or LBBAP generates ventricular depolarization and repolarization responses that more closely resembles that of spontaneous rhythm, in high contrast to the largely different response induced by RV pacing. HBP and LBBAP have distinct technical characteristics, including differences in capture thresholds, lead stability, and procedural aspects. These techniques serve as alternatives to conventional RV pacing.
Citation: Palacios S, Smisek R, Curila K, Nguyen U, Prinzen FW, Halamek J, et al. (2026) Ventricular activation and repolarization in response to physiological and conventional pacing using ultra-high-frequency electrocardiography. PLoS One 21(4): e0344111. https://doi.org/10.1371/journal.pone.0344111
Editor: Elena G. Tolkacheva, University of Minnesota, UNITED STATES OF AMERICA
Received: August 17, 2025; Accepted: February 16, 2026; Published: April 13, 2026
Copyright: © 2026 Palacios et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Data Availability: The data analyzed in this study were obtained from patient recordings and access to the anonymized datasets used in this work may be granted upon reasonable request. Because the dataset includes sensitive clinical and electrophysiological information from human participants, its sharing is strictly regulated under the European General Data Protection Regulation (GDPR, EU 2016/679). Accordingly, any transfer or secondary use of identifiable or potentially re-identifiable biomedical data requires prior written authorization from the data custodians and approval by the relevant institutional ethics committee. Data inquiries should be directed to Ing. Lukas Poviser, MSc. (lukas.poviser@lf3.cuni.cz) investigator and data custodian at Department of Cardiology Third Faculty of Medicine, Charles University and University Hospital Královské Vinohrady, Prague, Czech Republic.
Funding: This study was supported by Agencia Estatal de Investigación (Ministerio de Ciencia e Innovación) in the form of projects and grant awarded to E.P. (PID2022-140556OB-I00 and TED2021-130459B-I00), J.P.M. (PID2022-140556OB-I00 and TED2021-130459B-I00) and S.P. (BES-2017-080587) and by Gobierno de Aragón in the form of project and reference group for E.P. (LMP94_21 and BSICoS T39_23R), J.P.M. (LMP94_21 and BSICoS T39_23R) and S.P. (BSICoS T39_23R). Computations were performed using ICTS NANBIOSIS (HPC Unit at University of Zaragoza). The specific roles of these authors are articulated in the ‘author contributions’ section. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
Competing interests: F.W.P. reports research grants from Medtronic, Abbott, MicroPort CRM, and Biotronik. K.C. declares consultancy agreements with Medtronic, Abbott, Boston scientific. F.P., R.S., J.H., K.C., and P.J. are founders of the UHF-ECG system (VDI technologies) and are shareholders of the company VDI Technologies, Inc. F.P., J.H., and P.J. are among the inventors of the awarded patent No. US 9,949,655 B2 and European patent No. EP 3 082 589 B1. F.P., R.S., J.H., K.C., P.J. and F.W.P. are among the inventors of the awarded patents No. US 11,517,243 B2 and No. EP 3827743 “Method of electrocardiographic signal processing and apparatus for performing the method”. The remaining authors declare no competing interests.
1 Introduction
Cardiac arrhythmias affect 1.5% to 5% of the general population and account for up to 20% of all deaths worldwide [1,2]. In patients with bradyarrhythmias who have very slow heart rates, pacemakers apply electrical stimulation to restore the heart’s electrical activity to a normal cardiac rhythm. For several decades, the main pacing technique for treating bradyarrhythmias has been right ventricular pacing (RVP). Common locations for RVP stimulation are the right ventricular apex (so-called right ventricular apex pacing, RVAP) and the septum (so-called right ventricular septal pacing, RVSP). Despite the successful results of RVP for the treatment of bradyarrhythmias, numerous studies have shown its detrimental effects in terms of electrical and mechanical dyssynchrony of the ventricles, which predispose patients to develop mitral and tricuspid regurgitation, atrial fibrillation, cardiomyopathy, heart failure, and death [2–5].
As an alternative to the conventional, nonphysiological RVP, physiological pacing techniques have been proposed that stimulate the specialized cardiac conduction system to provide more physiological ventricular activation [6]. These techniques include His-bundle pacing (HBP) and left bundle branch (LBB) area pacing (LBBAP). HBP in patients who required pacemaker implantation was initially described by Deshmuk et al. [7] and, since then, its feasibility and safety have been demonstrated in different investigations. Importantly, HBP has been associated with a significant reduction compared to RVP in the combined endpoint of all-cause mortality, hospitalization for heart failure, and upgrade to biventricular pacing [8]. Recent studies indicate that, when implemented using a standardized and meticulous implantation technique, HBP can provide stable pacing parameters and favorable electrical outcomes [9,10]. Nevertheless, certain technical and procedural considerations have been reported in clinical practice, such as the anatomical complexity of the His bundle, its small size and surrounding fibrous tissue, as well as the variability of pacing capture thresholds and the potential for bundle injury during lead implantation [11–15]. LBBAP, by delivering pacing distal to the His bundle to achieve either left bundle branch capture or left ventricular septal myocardial pacing, has been described as facilitating lower and more consistent capture thresholds, while producing narrow paced QRS complexes and a high degree of left ventricular (LV) synchrony, particularly in patients with LBB block (LBBB) [11,14,15]. Despite the growing evidence supporting both HBP and LBBAP, larger randomized studies are still warranted to further assess feasibility, long-term safety, and clinical effectiveness [16], including direct comparisons of their respective effects on ventricular depolarization and repolarization.
To improve the characterization of ventricular activation in response to physiological pacing compared to conventional pacing, a number of studies have been conducted in recent years based on the standard electrocardiogram (ECG) and the vectorcardiogram (VCG). Most of these studies have evaluated the QRS morphology, duration, and area [17,18]. Additionally, ventricular depolarization patterns have been characterized from ultra-high-frequency ECG (UHF-ECG) recordings [19,20]. By proposing and validating novel measures of local depolarization duration and ventricular electrical dyssynchrony derived from UHF-ECG recordings, these studies have compared interventricular synchrony and the time for LV wall depolarization between various conventional and physiological pacing techniques [19,21–26].
Research on the effects of pacing on ventricular repolarization is much more limited than on ventricular activation. Some studies have quantified the duration of classical repolarization intervals, such as the heart rate-corrected QT interval, the JT interval, and the T-peak-to-T-end interval. These studies have shown a prolongation of the repolarization intervals following RVP but not following HBP or LBBAP [22,26–28]. Other studies have reported changes in the T wave, even if mostly from a qualitative point of view, without providing quantitative measures of T wave features [29–31].
Here, we extend the analysis of ECG ventricular depolarization and repolarization by processing UHF-ECG recordings from patients with physiological ventricular activation undergoing pacemaker implantation for bradycardia therapy. Both spatial and temporal features of the depolarization and repolarization phases of UHF-ECG recordings are quantified during spontaneous rhythm and in response to different types of conventional and physiological pacing. We hypothesize that physiological pacing techniques may lead to QRS and T wave patterns that better resemble those of spontaneous rhythm in patients without ventricular conduction abnormalities and that LBBAP techniques may be an equally effective option as HBP in reducing ventricular activation dyssynchrony, particularly in the LV, and in preserving ventricular repolarization characteristics.
2 Materials and methods
2.1 Study population
The study included mostly patients without bundle branch block and with an indication for pacemaker therapy due to bradycardia. Only 3.4% of the patients presented with left bundle branch block (LBBB), and 3.4% of them presented with right bundle branch block (RBBB). Six hundred and fifty-eight 14-lead UHF-ECG recordings, sampled at 5,000 Hz, were obtained from 178 patients (76 ± 7 years old, 68.5% male) using VDI UHF-ECG, VDI Technologies, Brno, Czech Republic, Cardiocenter of Faculty Hospital Kralovske Vinohrady, and the Third Medical Faculty of Charles University, Prague, Czech Republic. Almost 53% of the patients had atrioventricular block, and more than 38% had sick sinus syndrome. The remaining 9% of patients had other pacing indications, mainly bifascicular or trifascicular block and atrial fibrillation with planned AV junctional ablation. Other clinical characteristics of the study population are presented in Table 1. This research was conducted in accordance with the Helsinki Declaration as revised in 2016, was approved by the Ethics Committee of the Faculty Hospital Kralovske Vinohrady in Prague, and all subjects signed informed consent before enrollment.
The UHF-ECG recordings included 12 standard leads, as well as V7 and V8 leads recorded as described in previous studies [19]. UHF-ECG recordings were acquired for 1–10 minutes in a supine position during the pacemaker implantation procedure.
During the implantation procedure, patients underwent temporary application of various conventional and physiological pacing techniques, after which the final pacing strategy was selected at the discretion of the implanting physicians. Conventional pacing included RVAP and RVSP. Physiological pacing included various subtypes of HBP and LBBAP. HBP subtypes were selective HBP (sHBP), characterized by pure His capture pacing, and nonselective HBP (nsHBP), which involved the capture of both the His bundle and adjacent myocardial tissue. The LBBAP subtypes were: selective LBB pacing (sLBBP), involving exclusive LBB capture; nonselective LBB pacing (nsLBBP), with capture of both the LBB and the adjacent left septal myocardium; and LV septal pacing (LVSP), when the lead failed to capture the LBB but reached the LV subendocardium. In some patients, multiple recordings were collected in spontaneous rhythm and/or under some pacing types. In those cases, the first available recording was used for paired analysis.
The distribution of ECG recordings in spontaneous rhythm and different pacing types throughout the dataset was as follows. 37 recordings (from 37 patients) were recorded for RVAP, 102 (from 87 patients) for RVSP, 50 (from 47 patients) for sHBP, 160 (from 137 patients) for nsHBP, 13 (from 11 patients) for sLBBP, 47 (from 40 patients) for nsLBBP, and 87 (from 50 patients) for LVSP. In addition, 199 recordings (from 170 patients) were acquired during spontaneous heart rhythm. All patients signed an informed consent prior to enrollment.
2.2 Pacemaker implantation
For HBP, the procedure described in previous work was followed [8]. The His bundle region was mapped using a SelectSecureTM lead (model 3830, 69 cm, Medtronic Inc., Minneapolis, MN), delivered through a fixed-curve sheath (C315HIS, Medtronic). sHBP and nsHBP captures were defined as previously described [12,23]. For LBBAP, the lead was moved toward the right ventricle (RV) along a line between the His bundle region and the RV apex and was screwed deep into the septum to obtain a position on the left side of the interventricular septum showing a paced QRS morphology of RBBB/pseudo-RBBB in lead V1 [24]. sLBBP, nsLBBP, and LVSP were described as in previous studies [24,25]: In brief, concomitant myocardial and LBB capture (non-selective LBBP) was defined by a pseudo-RBBB morphology with the terminal r/R in V1 during pacing with an output of 5 V at 0.5 ms, which changed to sLBBP or LVSP when decreasing the pacing output. Selective capture of the LBB (sLBBP) was observed after decreasing the pacing output from nsLBBP, with an unchanged R wave peak time in V5. Pure myocardial capture of the left septum without LBB capture (LVSP) was characterized by a prolonged R wave peak time in V5 (> 10 ms) after decreasing the pacing output from nsLBBP. [24]. For RVAP and RVSP, the right ventricular leads were implanted in a standard form at the apex or septum of the RV [12,23].
2.3 UHF-ECG data analysis
2.3.1 Preprocessing.
ECG signals were high-pass filtered to remove baseline wander. A 50 Hz notch digital filter was applied to attenuate power line interference.
2.3.2 Pacing spike removal.
A semi-automated algorithm was developed to remove pacing artifacts, which departed from the method described in [32]. The algorithm was as follows:
- Orthogonal leads XYZ were obtained from the standard 12-lead ECG using the Kors transformation matrix [33]. The vector magnitude, v(n), was computed as:
(1)
whereis the vector norm of
, with x(n), y(n), and z(n) being the three orthogonal leads in
.
- The magnitude slope, dv/dt, was calculated from the difference between consecutive samples of the time series v(n). The onset and end of the pacing artifact were selected as the first sample no and the last sample ne for each beat of the time series v(n), respectively, that satisfied the following conditions:
(2)
whereand
are thresholds to determine the onset and end of the pacing spike. In this study, the threshold values were set to
mV/ms and
mV/ms.
- The pacing spike was removed by linear interpolation between the signal amplitude at no and ne. Fig 1 shows a segment of an ECG lead before and after removing the pacing spike by applying the described algorithm.
Pacing spike cancellation in a segment of an ECG lead. The signals before and after removing the pacing spike are shown in blue and red, respectively.
2.3.3 Median beat calculation.
The QRS complexes in each signal were detected and classified according to their morphology, following previously proposed methods [34]. Subsequently, the delineation of the peaks, onsets, and ends of all ECG waves was performed using a wavelet-based single-lead automatic delineation software [35].
A representative beat for each ECG recording was defined as the median beat of all beats presenting the dominant morphology. The calculation of the median beat included the following steps:
- RR intervals were computed from the fiducial points of the QRS complex.
- The statistical mode of the RR interval time series was calculated for each recording. An RR histogram was computed using 20-ms bins. The bin containing the RR mode was selected, and the median of the beats corresponding to the RR intervals in that bin was computed.
- All beats within the RR bin were aligned with respect to the computed median beat by maximizing the cross-correlation.
- The rank correlation between the initially calculated median beat and each of the aligned beats was calculated. Beats with rank correlation coefficients below 0.85 were discarded.
- The final median beat was computed from the aligned beats that were not discarded.
2.3.4 Depolarization indices.
The single-lead local depolarization duration (Vxd) and electrical dyssynchrony in the ventricles (e-DYS) were calculated from UHF-ECG by first dividing the spectrum from 150 Hz to 1,000 Hz into 16 frequency bands, following the method described in [19]. For each band and precordial lead (V1-V8), the amplitude envelopes were calculated using the Hilbert transform, with the amplitude envelopes segmented according to the QRS onset and end annotations. The median amplitude envelope in each frequency band was normalized so that its integral was one. e-DYS was calculated as the maximal time difference between the center of mass of the QRS above the 50 percent threshold of the baseline-to-peak amplitude for the leads V1-V8. The sign of e-DYS was associated with the order of ventricular activation. Negative e-DYS values indicated that the LV was activated earlier than the right ventricle.
Additionally, the local activation duration in a given lead, defined between the crossings at half the maximum peak magnitude of the average of the normalized median envelopes in that lead, was defined as Vxd. Low Vxd values were related to high apparent conduction velocities, presumably due to the contribution of the intrinsic conduction system, while high Vxd values were related to a nonhomogeneous substrate or nonphysiological origin of myocardial propagation. Vd was calculated as the mean of all the computed Vxd values and represented the average duration of depolarization propagation.
The duration of the QRS (QRSd) was measured in each median beat (computed as described in section 2.3.3) as the time between the onset and end of the QRS complex, with QRS onset and end marks obtained from the wavelet-based delineator in each lead and the subsequent application of post-processing rules to obtain a unique multilead QRSd. In particular, QRS onset was identified as the earliest delineation mark of the individual leads that had at least 3 neighboring marks within an interval of 12 ms, and QRS end was selected as the latest delineation mark of the individual leads whose 3 nearest neighbors were within an interval of 10 ms [35,36]. The same measurement approach was applied to both paced and intrinsic QRS complexes.
From the 8 independent leads of the median beat, the vectorcardiogram (VCG) was synthesized using the Kors method [37]. The QRS area (QRSa) in each orthogonal lead (X, Y, and Z) was calculated as the integral of the absolute value of the QRS complex from the beginning to the end of the QRS, as illustrated in Fig 2a). The total QRSa was calculated as:
where QRSa,X, QRSa,Y, and QRSa,Z denote the QRS areas in leads X, Y, and Z.
a) QRS and T-wave areas (shaded areas) in orthogonal leads. b) ECG beat for one lead with identified time points used to compute AT (red) and RT (green), calculated from the multilead QRS onset.
In addition, the local activation time (AT) was calculated for each individual precordial lead as the center of mass of the absolute QRS complex of the representative median beat, using the multilead QRS onset as a reference. An example is shown in Fig 2b), where the AT is marked with a red point. The difference between the last and the first AT values within the 6 precordial leads V1-V6 was defined as the dispersion of the activation times. This dispersion was calculated as the absolute value of the difference (denoted dAT) or by including a negative sign when the number of the first activated lead (1–6 for V1 to V6) was greater than that of the last activated lead (denoted dATs).
To specifically assess the activation dyssynchrony within the LV, analogous measurements of AT dispersion were calculated focusing only on leads V4-V6. These variables were denoted as dAT4-6 and dATs4-6 when the absolute value and the signed value of AT dispersion in V4-V6 were measured, respectively.
The results in the following section are presented using X for each calculated marker X.
X denotes the difference between the value of marker X under a given pacing modality and the corresponding value in spontaneous rhythm.
2.3.5 Repolarization indices.
The QT interval, which represents the time needed for ventricular activation and repolarization, was calculated as the interval between the multilead QRS complex onset and the multilead T-wave end of the median beat, determined by applying post-processing selection rules to the delineation marks of individual leads [35]. Taking into account the QT intervals of all recordings in the dataset, a correction for the effects of heart rate was performed. To derive the correction formula, different regression models were first fitted to the QT and RR measurements, with the regression models being linear, hyperbolic, parabolic, logarithmic, shifted logarithmic, exponential, arc tangent, hyperbolic tangent, arc hyperbolic sine, and arc hyperbolic cosine [38]. The regression model leading to the lowest residual from the fitting was selected, and the corresponding correction formula was derived by projecting the QT interval to a standard level of RR equal to one second. The corrected QT intervals calculated according to the derived formula were denoted by QTc. This formula was QTc = QT + 0.196· (1-RR), where QT and RR intervals are measured in seconds.
In each precordial lead, the repolarization time (RT) was determined as the center of mass of the absolute T wave of the representative median beat (represented as a green point in Fig 2b), measured from the multilead QRS onset mark. A corrected RT, denoted as RTc, was calculated by deriving a correction formula analogous to the QT interval correction to remove any RR influence on RT. The correction formula in this case was RTc = RT + 0.123· (1-RR), where RT and RR intervals are measured in seconds. The dispersion of RT, denoted as dRT, was calculated as the difference between the maximum and minimum RTc values in the 6 precordial leads V1-V6. The signed dispersion of RT, denoted as dRTcs, included a negative sign to the value of dRTc when the number of the first repolarized lead (1–6 for V1 to V6) was higher than that of the last repolarized lead.
In a similar manner to dAT4-6 calculation, RT dispersion within the LV was specifically calculated by focusing the analysis on the leads V4-V6 only. The corresponding variables were denoted as dRTc4-6 and dRTcs4-6.
Based on the AT and RT variables, the activation–recovery interval (ARI) was calculated for each recording from the difference between the corresponding RT and AT values. Subsequently, a heart-rate-corrected ARI measurement (ARIc) was calculated using an approach similar to that used to compute QTc and RTc.
The T-wave area (Ta) was calculated from the orthogonal leads X, Y, and Z according to the following formula:
where Ta,X, Ta,Y, and Ta,Z denote the T-wave areas in leads X, Y, and Z. An example of the calculation of the T-wave area is shown in Fig 2a).
Another studied repolarization marker was the Periodic Repolarization Dynamics (PRD), which was proposed to assess sympathetic modulation of ventricular repolarization by measuring low-frequency (below 0.1 Hz) oscillations in the T-wave vector. The PRD index was computed using the method described in [39], which represents a modified version of the method originally published in [40]. The method included the following steps:
- The T wave of each cardiac beat was segmented by defining a window dependent on the QRS fiducial point and the RR interval corresponding to that beat. The onset of the T-wave window, denoted by
, was set at 90 ms after the fiducial mark QRSi:
. The end of the T-wave window, denoted by
, was defined as
for RRi below 720 ms and
in other cases.
- A constant value was subtracted from each T wave in each of the analyzed leads so that the voltage at the end of the T-wave window was set to 0 mV.
- The average electrical vector was calculated for each T-wave window. The angle dT° between two consecutive T-wave windows was calculated from the dot product of the normalized average vectors.
- A 10th-order median filter was applied to attenuate outliers and artifacts in the dT° time series.
Fig 3 illustrates steps (i) to (iv) and the resulting dT° time series over 100 beats.
From the calculated dT° time series, a method based on phase rectified signal averaging (PRSA) [41] was applied to evaluate the magnitude of its low-frequency oscillations, which are measured by the PRD index [42]. This evaluation involved the following steps: - Anchor points were defined by comparing the averages of the M = 9 values of the series dT° before and after each candidate anchor point
. A point
was considered to be an anchor point if:
(5)
The value of M = 9 was established because it allows for detection of frequencies in the range of interest (from 0.025 to 0.1 Hz), as fully described in [41]. - Windows of 2L values were defined around each anchor point. Anchor points in the first and last L samples of the dT° series were discarded, as windows of length 2L could not be defined around them. In this study, L = 20 was chosen, as it allowed for the detection of frequencies in the range of interest.
- The PRSA series was obtained by averaging the dT° values over all 2L-sample windows around anchor points contained in each recording.
a) T waves for four consecutive beats. b) Three-dimensional representation of each pair of T-wave vectors. c) Angle between two consecutive T-wave vectors, dT°, for the represented beats. d) dT° time series along 100 consecutive beats.
PRD was defined as the difference between the maximum and minimum values of the PRSA series.
Analogous to the calculation of X for depolarization markers, differences were calculated between the value of the marker X under a given pacing modality and the corresponding value in spontaneous rhythm.
2.4 Statistical analysis
Continuous and discrete variables are presented as median [interquartile range (IQR)] and counts (percentages), respectively. The Wilcoxon signed-rank test was used for paired comparison of continuous variables, such as those performed to assess the statistical significance of the differences between a marker measured under a pacing modality and the same marker measured in spontaneous rhythm. For unpaired comparisons, the Mann–Whitney U test (or Wilcoxon rank-sum test) was applied when evaluating differences in the indices between a pacing modality and RVAP. All statistical analyses were performed with MATLAB R2020a (9.8). Differences were considered statistically significant if the associated p-value < 0.05. NS is used to denote nonsignificant.
3 Results
3.1 Depolarization indices
3.1.1 QRS duration.
Fig 4 shows the median and IQR of for all the pacing types analyzed in this study. QRS complexes in spontaneous rhythm had a median duration of 103.6 [18.4] ms. The widest QRS complexes were observed after RVP, both RVSP (median [IQR] of
equal to 35.0 [25.5] ms, p < 0.001 compared to spontaneous rhythm) and RVAP (35.8 [24.8] ms, p < 0.001). LBBAP led to shorter QRS complexes than RVP, although still significantly longer than those for spontaneous rhythm (with increases of 16.0 [21.7] ms, 20.1 [29.2] ms and 14.8 [28.1] ms for sLBBP, nsLBBP, and LVSP, respectively, with respect to spontaneous rhythm, p < 0.001). sHBP stimulation induced median QRSd values similar to those in spontaneous rhythm (with changes of −2.3 [15.0] ms, NS), while nsHBP pacing was associated with the lowest QRSd values (changes of −6.6 [27.1] ms, p-value<0.001 compared to spontaneous rhythm). All physiological pacing techniques showed significant differences (p < 0.001) compared to RVAP.
*p<0.05 with respect to spontaneous rhythm, †p<0.05 with respect to RVAP.
3.1.2 Electrical dyssynchrony and activation duration.
Figs 5 and 6 show box plots for e-DYS and
Vd. Similarly to the observations for
, RVP techniques (RVSP and RVAP) showed the largest dyssynchrony and activation duration (
e-DYS: 20.8 [22.0] ms;
Vd: 13.7 [15.2] ms for RVSP, and e-DYS: 16.0 [39.2] ms, Vd: 25.7 [11.8] ms for RVAP).
*p<0.05 with respect to spontaneous rhythm, †p<0.05 with respect to RVAP.
*p<0.05 with respect to spontaneous rhythm, †p<0.05 with respect to RVAP.
LBBAP techniques (sLBBP, nsLBBP, and LVSP) induced negative e-DYS values, indicative of higher ventricular synchrony than for spontaneous rhythm (−14.3 [9.6] ms, −18.6 [13.8] ms, and −12.9 [10.1] ms, p < 0.001 in the three cases). HBP techniques (sHBP and nsHBP) generated small positive
e-DYS values (1.7 [10.1] ms and 6.3 [13.7] ms, respectively, p = 0.02 and p < 0.001). Both HBP and LBBAP techniques were associated with small
Vd values, indicating that Vd was similar to that of spontaneous rhythm.
Statistically significant differences were found for all physiological techniques with respect to RVAP, in terms of both e-DYS and
Vd.
3.1.3 QRS area.
The values of QRSa during spontaneous rhythm were 50.06 [27.23] Vs. The highest differences in QRSa with respect to spontaneous rhythm were found for RVAP and RVSP, which presented increases of 28.8 [27.3]
Vs and 92.0 [54.9]
Vs (p-values < 0.001 with respect to spontaneous rhythm). sLBBP, nsLBBP, and sHBP showed the lowest
values, which were close to zero (−0.9 [13.0]
Vs, 7.8 [20.4]
Vs, and 0.4 [9.6]
Vs) Box plots for
are found in Fig 7.
*p<0.05 with respect to spontaneous rhythm, †p<0.05 with respect to RVAP.
All physiological techniques yielded significantly lower values of than RVAP.
3.1.4 Activation time patterns.
Fig 8 shows the average AT values in the precordial leads V1-V6 for spontaneous rhythm and each cardiac pacing type. As can be seen in the figure, leads V1, V5, and V6 present the earliest activation in spontaneous rhythm (54.5 ms in lead V6), while V2-V4 present the latest activation (59.5 ms in lead V3). Similarly, sHBP and nsHBP show activation patterns in which the regions corresponding to leads V1, V5, and V6 are activated earlier than those corresponding to leads V2-V4. LBBAP pacing types, as expected, present the earliest activations in leads V4-V6 (earliest activation at 60.0 ms for sLBBP, 66.5 ms for nsLBBP, and 57.4 ms for LVSP), while the latest activations are found in leads V1-V3 (latest activation at 70.9 ms, 76.2 ms, and 61.5 ms, respectively). The AT values for RVP types are the largest in practically all leads compared to other pacing types and spontaneous rhythm. For RVAP, the activation starts in V4 and ends in V1 and V6. For RVSP, the activation starts in V1 and ends in V4 and V6.
3.1.5 Dispersion of activation time.
The median dAT for spontaneous rhythm was 8.80 [5.00] ms. As shown in Fig 9, sHBP and nsHBP rendered the closest dAT values to the intrinsic rhythm (dAT: 0.5 [8.0] ms and 0.6 [7.1] ms, respectively, NS for both cases). LBBAP and RVP techniques presented significantly higher dAT values, with
dAT: 6.6 [9.9] ms, 5.8 [11.6] ms, 2.5 [8.4] ms, 5.0 [9.0] ms, and 6.4 [9.0] ms for sLBBP, nsLBBP, LVSP, RVSP, and RVAP, respectively, p < 0.001 with respect to the intrinsic rhythm.
*p<0.05 with respect to spontaneous rhythm, †p<0.05 with respect to RVAP.
When considering the sign of dispersion, sHBP and nsHBP exhibited median values of dATs close to zero (0.0 [17.6] ms and 0.6 [23.4] ms, respectively), indicating high similarity to spontaneous rhythm. sLBBP, nsLBBP, and LVSP rendered slightly negative
dATs values, representative of lower dATs (−6.8 [13.5] ms, −7.0 [21.1] ms, and −2.8 [22.8] ms, p < 0.05, respectively). RVSP was associated with the highest
dATs values (8.2 [26.4] ms, p < 0.05). The corresponding box plots are presented in S1 Fig.
The results for the dispersion of activation times within the LV (dAT4-6) are shown in Fig 10. Both sHBP and nsHBP led to lower , which was not significantly different from zero (0.6 [8.2] ms and −1.4 [6.2] ms, respectively). sLBBP and nsLBBP also had median values close to zero (−0.8 [5.6] ms and −1.2 [5.0] ms, NS). Conventional pacing types had the highest values (1.4 [9.0] ms for RVSP and 2.0 [8.9] ms for RVAP, p < 0.05 in both cases). The corresponding results for
are presented in S2 Fig.
*p<0.05 with respect to spontaneous rhythm, †p<0.05 with respect to RVAP.
For dAT, sHBP, nsHBP, and LVSP led to values significantly lower than RVAP. For
, significantly lower values than those for RVAP were found for nsHBP, nsLBBP, and LVSP.
3.2 Repolarization indices
3.2.1 Corrected QT.
The median QTc under spontaneous rhythm was 429 [43] ms. As shown in Fig 11, sHBP and nsHBP led to QTc values lower than zero (−3.0 [49.0] ms, NS, and −9.4 [44.3] ms, p < 0.05, respectively). Of the LBBAP techniques, nsLBBP led to small positive
QTc (1.1 [49.4] ms, NS), while sLBBP led to slightly higher values (11.8 [85.9] ms, NS). RVSP and RVAP induced larger QTc prolongations (
QTc: 24.4 [29.8] ms and 31.1 [49.8] ms, p < 0.05). QTc was significantly shorter under sHBP, nsHBP, nsLBBP, and LVSP than under RVAP.
*p<0.05 with respect to spontaneous rhythm, †p<0.05 with respect to RVAP.
3.2.2 T-wave area.
Fig 12 shows the values of for each type of cardiac pacing. Interestingly, the observed trend is very similar to that observed for
, shown in Fig 7. The value of Ta for spontaneous rhythm was 30.50 [27.83]
Vs. The highest
values were found for RVSP (34.5 [43.9]
Vs, p < 0.001) and RVAP (109.4 [42.5]
Vs, p < 0.001). Physiological pacing approaches, particularly sHBP, sLBBP, and nsLBBP, presented
values close to zero, which means similar T-wave areas to those found for spontaneous rhythm (sHBP: −4.2 [9.4]
Vs, NS; sLBBP: −0.9 [20.1]
Vs, NS; nsLBBP: 1.3 [21.3]
Vs, p < 0.05). All physiological pacing techniques rendered smaller T-wave areas than RVAP.
*p<0.05 with respect to spontaneous rhythm, †p<0.05 with respect to RVAP.
3.2.3 Periodic repolarization dynamics.
For spontaneous rhythm, the PRD median value was 4.94 [3.93] degrees. As illustrated in Fig 13, PRD took negative values for all cardiac pacing techniques. The lowest
PRD values were found for sLBBP and RVAP (−4.6 [4.7] and −4.1 [5.9] degrees, respectively). In the case of sHBP, its
PRD value was close to zero (−0.5 [4.0] degrees, NS). Of the evaluated physiological pacing modes, statistically significant differences with respect to RVAP were found for sHBP, nsHBP, and RVSP.
*p<0.05 with respect to spontaneous rhythm, †p<0.05 with respect to RVAP.
3.2.4 Repolarization time patterns.
Fig 14 shows the average RT values corrected for the RR interval in the precordial leads V1-V6 for spontaneous rhythm and each cardiac pacing type. For spontaneous rhythm, the first lead to repolarize was V2 (321.6 ms), with lead V5 being the last to repolarize (336.5 ms). Similar RT patterns were found for the two types of selective physiological pacing, i.e., sHBP and sLBBP, with the earliest repolarization occurring in V1-V2 and the latest in V5. The RT patterns for nsHBP, nsLBBP, and LVSP were similar to the corresponding AT patterns, with V4, V5, and V6 being the first leads to repolarize in each case. RVAP showed RT patterns similar to those of AT; however, in this case, lead V6 was the last to repolarize. For RVSP, the repolarization took longer to complete in lead V3.
3.2.5 Dispersion of repolarization time.
The median value of dRTc for spontaneous recordings was 31.2 [22.2] ms. Fig 15 shows that dRTc was comparable for all pacing types, with the lowest values observed for RVAP (−14.4 [28.1] ms).
*p<0.05 with respect to spontaneous rhythm, †p<0.05 with respect to RVAP.
When conducting the analysis for , all pacing techniques showed small values, with the lowest median value found for LVSP (−5.4 [28] ms), as shown in Fig 16.
*p<0.05 with respect to spontaneous rhythm, †p<0.05 with respect to RVAP.
The corresponding results for dRTcs and
are presented in S3 and S4 Figs.
In addition to activation and repolarization times, the results for ARIc are presented in S5 Fig.
A disaggregated analysis for patients with atrioventricular block and those with sick sinus syndrome was performed to assess the response to each pacing technique in these two subpopulations. Despite minor quantitative differences, the results were qualitatively consistent across both groups.
4 Discussion
This study provides a detailed characterization of the response of ventricular depolarization and repolarization to seven different types of cardiac pacing in patients without abnormalities in ventricular conduction and with an indication for pacemaker implantation due to bradycardia. In addition to quantifying depolarization and repolarization indices commonly used in other studies, we propose additional markers to describe the spatial dispersion of the activation and repolarization times from UHF-ECG recordings. Properties related to the duration, area, and spatial heterogeneity of ECG activation and repolarization waves are shown to differentiate the effects of each of the tested physiological and conventional cardiac pacing techniques, with implications for the development of deleterious effects on cardiac function.
4.1 Pacing-induced effects on ventricular depolarization
For the characterization of cardiac pacing effects on ventricular depolarization, this study first examined classical ECG markers reported in the literature, such as the duration and area of the QRS complex. Other more recently proposed markers quantified from UHF-ECG recordings were also analyzed, including Vd (measuring local activation duration) and e-DYS (measuring activation dyssynchrony). Finally, novel markers that can be quantified from standard ECGs (not necessarily sampled at very high frequency) were proposed based on ATs in different ECG leads and their spatial dispersion across the entire ventricles and across the LV only.
From the analysis of all the described markers, we found that the physiological pacing techniques sHBP and nsHBP exhibit an electrical response comparable to that observed in the spontaneous rhythm in patients without any conduction disorders. LBBAP techniques, including sLBBP, nsLBBP, and LVSP, despite presenting larger differences with respect to spontaneous rhythm than HBP-based techniques, are able to preserve the synchrony in the activation of the LV, with only a delayed activation in the RV. The conventional pacing techniques RVSP and RVAP show markedly different responses compared to those found in spontaneous rhythm, with high dispersion in the ventricular ATs. This behavior was also observed in previous studies [23–25].
QRSd is a measure commonly used to assess ventricular synchrony. Changes in QRSd have been associated with the occurrence of major adverse cardiovascular events [43,44]. Here, we showed that sHBP and nsHBP had values close to zero, indicating similarities to spontaneous rhythm. The longest QRSd values were found for RVSP and RVAP. The three LBBAP techniques, i.e., sLBBP, nsLBBP, and LVSP, were associated with similar values of
, which were higher than those for HBP. Our results are in line with those reported in previous studies that compared the ventricular activation synchrony of sLBBP and nsLBBP with that of LVSP [45] and RVAP [46]. The larger QRSd values found for sLBBP, nsLBBP, and LVSP compared to spontaneous or His-paced activations can be attributed to the fact that the electrical pacing impulse is rapidly transmitted to the LV through the LBB or the myocardium close to it, while the RV is activated with some delay.
Activation indices derived from the VCG have been suggested to predict the response to cardiac resynchronization therapy more accurately than QRSd [47]. Here, QRSa was measured from the VCG. QRSa has been reported to be less sensitive to uncertainty in the determination of the QRS boundaries, as the extremes of the QRS complex scarcely contribute to the area [48,49], and to present less variability in its measurement than QRSd [50]. Previous works have reported QRSa to be an index of dyssynchrony in electrical activation, with a large QRSa corresponding to the delayed activation of the posterolateral wall of the LV, independently of QRS morphology [51]. Our results indicate that sHBP and sLBBP have QRSa values similar to those in spontaneous rhythm, whereas nsHBP, nsLBBP, and LVSP have somewhat higher values that are still far from the large values associated with RVAP. Considering that a decrease in QRSa has been shown to be an independent predictor of survival and reverse cardiac remodeling [52], the results of the present study indicate that physiological pacing, especially through selective HBP or LBBP techniques, entails significantly lower risk than conventional pacing techniques.
Using markers specifically derived from UHF-ECG recordings, such as e-DYS and Vd, the ventricular response to different pacing techniques has been investigated in previous studies [19,21,24]. These studies have shown that HBP produces physiological activation times and patterns, preserving fast activation of the LV lateral wall and avoiding any deterioration in interventricular synchrony. No significant differences between sHBP and nsHBP in terms of ventricular activation patterns were reported when tested in patients without conduction disorders. Our results are in line with those previous findings. The results for LBBAP indicated a reduction in e-DYS values for LBBAP with respect to intrinsic rhythm, in association with lower activation times in leads V4-V6. The e-DYS values for HBP were very close to those of intrinsic rhythm, in agreement with previously reported results [53–55].
Furthermore, we measured the local ATs from the precordial leads V1-V6 and quantified their spatial differences with the aim of characterizing the sequence of activation and its spatial dispersion [56]. Our results indicate that HBP techniques preserve the activation synchrony of spontaneous rhythm, as manifested by similar dAT values. For LBBAP techniques, the synchrony is reduced, but only because the RV activates later, whereas the synchrony in LV activation, quantified by dAT4-6, is higher than that for spontaneous rhythm. Conventional pacing through RVSP or RVAP leads to increased dispersion in the activation of the two ventricles or the LV alone. Increases in the dispersion of both ventricular activation and repolarization have been shown to be important in the genesis and maintenance of ventricular arrhythmias and sudden cardiac death [57,58].
4.2 Pacing-induced effects on ventricular repolarization
Differences in ventricular repolarization as a function of pacing techniques were investigated by measuring the QTc interval, the T-wave area, the PRD index reflecting the magnitude of the low-frequency oscillations in the T wave, and the RTs across the precordial leads V1-V6, together with their dispersion.
The QTc interval has been widely used in the literature to provide an overall measure of the duration of repolarization after compensating for the effects of heart rate. In previous works, patients with significantly prolonged QTc intervals who underwent permanent pacemaker implantation have been reported to face a higher risk of new-onset LV systolic dysfunction, cardiac death, ventricular arrhythmias, and sudden cardiac death [59–61]. Other studies have shown that the pacing-induced increase in QTc, with respect to intrinsic rhythm, was less pronounced for LBBP techniques than for RVSP [22]. A study including 55 patients with LBBB and an indication for cardiac resynchronization therapy (CRT) [62] showed that HBP did not lead to immediate significant changes in corrected QT and Tpeak–Tend intervals with respect to baseline. Our results are in line with such findings. In particular, we show that HBP and LBBAP (sLBBP, nsLBBP, LVSP) are associated with similar QTc values to spontaneous rhythm (represented by values of QTc close to zero), or even lower values in the case of nsHBP, whereas RVSP and RVAP present significantly higher
QTc values, indicative of QTc prolongation.
Ta has been postulated to carry long-term prognostic information on cardiovascular mortality in apparently healthy populations [63–65]. Furthermore, Ta has been suggested to be a sensitive feature that characterizes the ventricular repolarization response to pacing, with the capacity to track minor ST-segment and T-wave abnormalities [65] and to serve as a marker that can predict the response to CRT [48,66]. The results of our work indicate that conventional pacing techniques, especially RVAP, lead to high values of Ta, while physiological pacing techniques lead to Ta values similar to those measured under intrinsic rhythm. These findings extend those of previous work by adding the characterization of Ta in response to a large number of conventional and physiological pacing techniques [67–69] and confirm the suitability of HBP and LBBAP techniques in terms of their effects on ventricular repolarization.
PRD is an arrhythmic risk marker that measures the magnitude of low-frequency oscillations in the T-wave vector modulated by sympathetic nervous system activity. Increased PRD values have been shown to be a strong predictor of all-cause mortality, cardiac mortality, and ventricular arrhythmias in various cardiac diseases and conditions [39,40,70–74]. Our results show that all cardiac pacing techniques induce a decrease in the median PRD values found in spontaneous rhythm, with the decrease reaching approximately 5 degrees, in median, for sLBBP and RVAP. The basis for such results deserves further investigation. Although no significant correlation was found between PRD and heart rate, in accordance with previous reports [75], additional studies are warranted to fully rule out the impact of heart rate on pacing-induced reduction in PRD.
To further characterize the response of ventricular repolarization to cardiac pacing, we evaluated the dispersion of RTs across the six precordial leads V1-V6 and focused only on V4-V6. No significant differences are present between most cardiac pacing techniques and spontaneous rhythm. In general, the median values of dRTc and
for all pacing modalities, except for RVAP, are close to zero. Even if enhanced dispersion of ventricular repolarization has been suggested as indicative of reverse remodeling [76,77] and a greater susceptibility to arrhythmias such as Torsades de Pointes [78], our results support the suitability of all physiological pacing techniques, including LBBAP techniques, based on their ability to reproduce the basal dispersion of LV repolarization. Such suitability is also supported by the findings achieved for
ARIc.
In summary, our results confirm previous evidence of the large similarities between HBP, particularly sHBP, and spontaneous rhythm in patients without ventricular conduction disorders [21]. The LBBAP techniques, despite differing from spontaneous rhythm in some global measurements of ventricular AT and RT dispersion, have been confirmed to lead to high synchrony in LV activation and repolarization. These results provide additional evidence of the validity of both HBP and LBBAP [79]. Conventional RVP techniques, however, are associated with the highest global and LV-specific dyssynchrony characteristics, which could have long-term adverse effects on ventricular function [80]. Our research thus supports the use of LBBAP modalities as physiological pacing techniques that overcome the shortcomings associated with conventional RVP.
5 Study limitations and future work
This work included ECGs of patients who underwent different types of cardiac pacing. However, there is a large variability in the number of patients subjected to each type of pacing. In particular, only 13 recordings were available for sLBBP, while more than 150 recordings were available for nsHBP or spontaneous rhythm. Another limitation is the lack of access to follow-up information. Long-term follow-up studies with a large representative number of patients for each type of pacing are required to corroborate the present results.
It is also important to note that most of the recordings were obtained by applying one stimulation technique after another. Although there was some waiting time between the application of a type of cardiac stimulation and the next one, the influence of the so-called cardiac memory phenomenon [81], particularly on ventricular repolarization, cannot be ruled out. Therefore, studies with longer time intervals between the application of different types of stimulation or the use of recordings without sequential pacing would be required to confirm the present findings.
Currently, the number of hospitals and medical centers where UHF-ECG signals are recorded is limited. Future studies should confirm the clinical value of the proposed markers dAT, dRTc, dAT4-6 and dRTc4-6, which can be measured from standard ECGs, as an adjunct or substitute for the marker e-DYS quantified from UHF-ECG signals.
6 Conclusions
This study analyzed a set of markers to characterize ventricular depolarization and repolarization in response to seven different types of cardiac stimulation. Physiological stimulation using both HBP and LBBAP produced functional behaviors that were close to those of the intrinsic rhythm in patients without cardiac disorders, whereas conventional RV stimulation differed markedly. The differences between HBP and LBBAP were minor for most markers, particularly those related to LV depolarization and repolarization. These findings support the use of LBBAP and HBP as alternative physiological pacing options in patients with narrow QRS complexes. Selection between these approaches may reasonably depend on operator experience, technical feasibility, and individual patient characteristics.
Supporting information
S1 Fig. Box plots of
dATs for each of the pacing types respect to spontaneous rhythm.
*p<0.05 respect to spontaneous rhythm, †p<0.05 respect to RVAP.
https://doi.org/10.1371/journal.pone.0344111.s001
(EPS)
S2 Fig. Box plots of
for each of the pacing types respect to spontaneous rhythm.
*p<0.05 respect to spontaneous rhythm, †p<0.05 respect to RVAP.
https://doi.org/10.1371/journal.pone.0344111.s002
(EPS)
S3 Fig. Box plots of
dRTs for each of the pacing types respect to spontaneous rhythm.
*p<0.05 respect to spontaneous rhythm, †p<0.05 respect to RVAP.
https://doi.org/10.1371/journal.pone.0344111.s003
(EPS)
S4 Fig. Box plots of
for each of the pacing types respect to spontaneous rhythm.
*p<0.05 respect to spontaneous rhythm, †p<0.05 respect to RVAP.
https://doi.org/10.1371/journal.pone.0344111.s004
(EPS)
S5 Fig. Box plots of
for each of the pacing types respect to spontaneous rhythm.
*p<0.05 respect to spontaneous rhythm, †p<0.05 respect to RVAP.
https://doi.org/10.1371/journal.pone.0344111.s005
(EPS)
References
- 1.
Desai DS, Hajouli S. Arrhythmias. In: StatPearls. StatPearls Publishing. 2023.
- 2. Patel MH, Sampath S, Kapoor A, Damani DN, Chellapuram N, Challa AB. Advances in Cardiac Pacing: Arrhythmia Prediction, Prevention and Control Strategies. Frontiers in Physiology. 2021;12.
- 3. Bank AJ, Gage RM, Burns KV. Right ventricular pacing, mechanical dyssynchrony, and heart failure. J Cardiovasc Transl Res. 2012;5(2):219–31. pmid:22189959
- 4. Akerström F, Pachón M, Puchol A, Jiménez-López J, Segovia D, Rodríguez-Padial L, et al. Chronic right ventricular apical pacing: adverse effects and current therapeutic strategies to minimize them. Int J Cardiol. 2014;173(3):351–60. pmid:24721486
- 5. Naqvi TZ, Chao C-J. Adverse effects of right ventricular pacing on cardiac function: prevalence, prevention and treatment with physiologic pacing. Trends Cardiovasc Med. 2023;33(2):109–22. pmid:34742888
- 6. Arora V, Suri P. Physiological pacing: a new road to future. Indian Journal of Clinical Cardiology. 2021;2(1):32–43.
- 7. Deshmukh P, Casavant DA, Romanyshyn M, Anderson K. Permanent, Direct His-Bundle Pacing. Circulation. 2000;101(8):869–77.
- 8. Abdelrahman M, Subzposh FA, Beer D, Durr B, Naperkowski A, Sun H, et al. Clinical Outcomes of His Bundle Pacing Compared to Right Ventricular Pacing. J Am Coll Cardiol. 2018;71(20):2319–30. pmid:29535066
- 9. Kaddour M, Burri H. Should We Pace the His Bundle or the Left Bundle Branch Area?. Cardiovasc Med. 2024.
- 10. Jastrzębski M, Huang W, Moriña-Vazquez P, Vijayaraman P. His Bundle Pacing in the Era of Left Bundle Branch Pacing. Arrhythmia and Electrophysiology Review. 2025;14(e06).
- 11. Zanon F, Ellenbogen KA, Dandamudi G, Sharma PS, Huang W, Lustgarten DL, et al. Permanent His-bundle pacing: a systematic literature review and meta-analysis. Europace. 2018;20(11):1819–26. pmid:29701822
- 12. Vijayaraman P, Dandamudi G, Zanon F, Sharma PS, Tung R, Huang W, et al. Permanent His bundle pacing: Recommendations from a Multicenter His Bundle Pacing Collaborative Working Group for standardization of definitions, implant measurements, and follow-up. Heart Rhythm. 2018;15(3):460–8. pmid:29107697
- 13. Zhang S, Zhou X, Gold MR. Left Bundle Branch Pacing: JACC Review Topic of the Week. J Am Coll Cardiol. 2019;74(24):3039–49. pmid:31865972
- 14. Palmisano P, Ziacchi M, Dell’Era G, Donateo P, Ammendola E, Coluccia G, et al. Rate and nature of complications of conduction system pacing compared with right ventricular pacing: Results of a propensity score-matched analysis from a multicenter registry. Heart Rhythm. 2023;20(7):984–91. pmid:36906165
- 15. Parlavecchio A, Vetta G, Coluccia G, Pistelli L, Caminiti R, Crea P, et al. Success and complication rates of conduction system pacing: a meta-analytical observational comparison of left bundle branch area pacing and His bundle pacing. J Interv Card Electrophysiol. 2024;67(4):719–29. pmid:37642801
- 16. Huang W, Chen X, Su L, Wu S, Xia X, Vijayaraman P. A beginner’s guide to permanent left bundle branch pacing. Heart Rhythm. 2019;16(12):1791–6. pmid:31233818
- 17. Lin J, Hu Q, Chen K, Dai Y, Chen R, Sun Q, et al. Relationship of paced left bundle branch pacing morphology with anatomic location and physiological outcomes. Heart Rhythm. 2021;18(6):946–53. pmid:33781981
- 18. Cano O Ó, Jover P, Vijayaraman P. An Evidence-Based Update on Physiological Pacing. Current Treatment Options in Cardiovascular Medicine. 2023;25(10):415–39.
- 19. Jurak P, Curila K, Leinveber P, Prinzen FW, Viscor I, Plesinger F, et al. Novel ultra-high-frequency electrocardiogram tool for the description of the ventricular depolarization pattern before and during cardiac resynchronization. J Cardiovasc Electrophysiol. 2020;31(1):300–7. pmid:31788894
- 20. Mizner J, Jurak P, Linkova H, Smisek R, Curila K. Ventricular Dyssynchrony and Pacing-induced Cardiomyopathy in Patients with Pacemakers, the Utility of Ultra-high-frequency ECG and Other Dyssynchrony Assessment Tools. Arrhythm Electrophysiol Rev. 2022;11:e17. pmid:35990106
- 21. Curila K, Prochazkova R, Jurak P, Jastrzebski M, Halamek J, Moskal P, et al. Both selective and nonselective His bundle, but not myocardial, pacing preserve ventricular electrical synchrony assessed by ultra-high-frequency ECG. Heart Rhythm. 2020;17(4):607–14. pmid:31805370
- 22. Wang J, Liang Y, Wang W, Chen X, Bai J, Chen H, et al. Left bundle branch area pacing is superior to right ventricular septum pacing concerning depolarization-repolarization reserve. J Cardiovasc Electrophysiol. 2020;31(1):313–22. pmid:31778249
- 23. Curila K, Jurak P, Halamek J, Prinzen F, Waldauf P, Karch J, et al. Ventricular activation pattern assessment during right ventricular pacing: Ultra-high-frequency ECG study. J Cardiovasc Electrophysiol. 2021;32(5):1385–94. pmid:33682277
- 24. Curila K, Jurak P, Jastrzebski M, Prinzen F, Waldauf P, Halamek J, et al. Left bundle branch pacing compared to left ventricular septal myocardial pacing increases interventricular dyssynchrony but accelerates left ventricular lateral wall depolarization. Heart Rhythm. 2021;18(8):1281–9. pmid:33930549
- 25. Curila K, Jurak P, Vernooy K, Jastrzebski M, Waldauf P, Prinzen F, et al. Left Ventricular Myocardial Septal Pacing in Close Proximity to LBB Does Not Prolong the Duration of the Left Ventricular Lateral Wall Depolarization Compared to LBB Pacing. Front Cardiovasc Med. 2021;8:787414. pmid:34950718
- 26. Yanagisawa S, Inden Y, Watanabe R, Tsurumi N, Suzuki N, Nakagomi T, et al. Depolarization and repolarization dynamics after His-bundle pacing: Comparison with right ventricular pacing and native ventricular conduction. Ann Noninvasive Electrocardiol. 2022;27(5):e12991. pmid:35802829
- 27. Pestrea C, Cicala E, Ivascu M, Gherghina A, Ortan F, Pop D. Comparison of depolarization and repolarization parameters in left vs. right ventricular septal pacing——an intraprocedural electrocardiographic study. Journal of Cardiovascular Development and Disease. 2023;10(3):108.
- 28. Ponnusamy SS, Vijayaraman P. Pacing for atrioventricular block with preserved left ventricular function: On-treatment comparison between his bundle, left bundle branch, and right ventricular pacing. Indian Pacing Electrophysiol J. 2023;23(6):196–202. pmid:37776973
- 29. Ortega DF, Barja LD, Logarzo E, Mangani N, Paolucci A, Bonomini MP. Non-selective His bundle pacing with a biphasic waveform: enhancing septal resynchronization. Europace. 2018;20(5):816–22. pmid:28520951
- 30. Vijayaraman P, Herweg B, Dandamudi G, Mittal S, Bhatt AG, Marcantoni L, et al. Outcomes of His-bundle pacing upgrade after long-term right ventricular pacing and/or pacing-induced cardiomyopathy: Insights into disease progression. Heart Rhythm. 2019;16(10):1554–61. pmid:30930330
- 31. Zhong C, Jiang Z, Zhou X, Shan Q. Reversion of cardiac memory during left bundle branch pacing. J Electrocardiol. 2020;59:81–3. pmid:32023497
- 32. Haq KT, Javadekar N, Tereshchenko LG. Detection and removal of pacing artifacts prior to automated analysis of 12-lead ECG. Comput Biol Med. 2021;133:104396. pmid:33872969
- 33. Kors JA, van Herpen G, Sittig AC, van Bemmel JH. Reconstruction of the Frank vectorcardiogram from standard electrocardiographic leads: diagnostic comparison of different methods. Eur Heart J. 1990;11(12):1083–92. pmid:2292255
- 34.
Plesinger F, Jurco J, Jurak P, Halamek J. Robust multichannel QRS detection. In: Computing in Cardiology, 2014. 557–60.
- 35. Martínez JP, Almeida R, Olmos S, Rocha AP, Laguna P. A wavelet-based ECG delineator: evaluation on standard databases. IEEE Trans Biomed Eng. 2004;51(4):570–81. pmid:15072211
- 36. Almeida R, Martínez JP, Rocha AP, Laguna P. Multilead ECG delineation using spatially projected leads from wavelet transform loops. IEEE Trans Biomed Eng. 2009;56(8):1996–2005. pmid:19447698
- 37. Man S, Algra AM, Schreurs CA, Borleffs CJW, Scherptong RWC, van Erven L, et al. Influence of the vectorcardiogram synthesis matrix on the power of the electrocardiogram-derived spatial QRS-T angle to predict arrhythmias in patients with ischemic heart disease and systolic left ventricular dysfunction. J Electrocardiol. 2011;44(4):410–5. pmid:21704219
- 38. Pueyo E, Smetana P, Caminal P, de Luna AB, Malik M, Laguna P. Characterization of QT interval adaptation to RR interval changes and its use as a risk-stratifier of arrhythmic mortality in amiodarone-treated survivors of acute myocardial infarction. IEEE Trans Biomed Eng. 2004;51(9):1511–20. pmid:15376499
- 39. Palacios S, Cygankiewicz I, Bayés de Luna A, Pueyo E, Martínez JP. Periodic repolarization dynamics as predictor of risk for sudden cardiac death in chronic heart failure patients. Sci Rep. 2021;11(1):20546. pmid:34654872
- 40. Rizas KD, Nieminen T, Barthel P, Zürn CS, Kähönen M, Viik J, et al. Sympathetic activity-associated periodic repolarization dynamics predict mortality following myocardial infarction. J Clin Invest. 2014;124(4):1770–80. pmid:24642467
- 41. Bauer A, Kantelhardt JW, Bunde A, Barthel P, Schneider R, Malik M, et al. Phase-rectified signal averaging detects quasi-periodicities in non-stationary data. Physica A: Statistical Mechanics and its Applications. 2006;364:423–34.
- 42. Rizas KD, Hamm W, Kääb S, Schmidt G, Bauer A. Periodic Repolarisation Dynamics: A Natural Probe of the Ventricular Response to Sympathetic Activation. Arrhythm Electrophysiol Rev. 2016;5(1):31–6. pmid:27403291
- 43. Małecka B, Ząbek A, Lelakowski J. Shortening of paced QRS complex and clinical improvement following upgrading from apical right ventricular pacing to bifocal right ventricular or biventricular pacing in patients with permanent atrial fibrillation. Kardiol Pol. 2010;68(11):1234–41. pmid:21108200
- 44. Sakatani T, Sakamoto A, Kawamura K, Tanigaki T, Tsubakimoto Y, Isodono K, et al. Clinical Outcome After Permanent Pacemaker Implantation in Patients With a High Percentage of Ventricular Pacing. Int Heart J. 2015;56(6):622–5. pmid:26549389
- 45. Heckman LIB, Luermans JGLM, Curila K, Van Stipdonk AMW, Westra S, Smisek R, et al. Comparing Ventricular Synchrony in Left Bundle Branch and Left Ventricular Septal Pacing in Pacemaker Patients. J Clin Med. 2021;10(4):822. pmid:33671420
- 46. Tokavanich N, Prasitlumkum N, Mongkonsritragoon W, Cheungpasitporn W, Thongprayoon C, Vallabhajosyula S, et al. A network meta-analysis and systematic review of change in QRS duration after left bundle branch pacing, His bundle pacing, biventricular pacing, or right ventricular pacing in patients requiring permanent pacemaker. Sci Rep. 2021;11(1):12200. pmid:34108548
- 47. Maass AH, Vernooy K, Wijers SC, van ’t Sant J, Cramer MJ, Meine M, et al. Refining success of cardiac resynchronization therapy using a simple score predicting the amount of reverse ventricular remodelling: results from the Markers and Response to CRT (MARC) study. Europace. 2018;20(2):e1–10. pmid:28339818
- 48. Engels EB, Végh EM, Van Deursen CJM, Vernooy K, Singh JP, Prinzen FW. T-wave area predicts response to cardiac resynchronization therapy in patients with left bundle branch block. J Cardiovasc Electrophysiol. 2015;26(2):176–83. pmid:25230363
- 49. Plesinger F, van Stipdonk AMW, Smisek R, Halamek J, Jurak P, Maass AH, et al. Fully automated QRS area measurement for predicting response to cardiac resynchronization therapy. J Electrocardiol. 2020;63:159–63. pmid:31324399
- 50. van Stipdonk AMW, Ter Horst I, Kloosterman M, Engels EB, Rienstra M, Crijns HJGM, et al. QRS Area Is a Strong Determinant of Outcome in Cardiac Resynchronization Therapy. Circ Arrhythm Electrophysiol. 2018;11(12):e006497. pmid:30541356
- 51. Mafi Rad M, Wijntjens GWM, Engels EB, Blaauw Y, Luermans JGLM, Pison L, et al. Vectorcardiographic QRS area identifies delayed left ventricular lateral wall activation determined by electroanatomic mapping in candidates for cardiac resynchronization therapy. Heart Rhythm. 2016;13(1):217–25. pmid:26232766
- 52. Ghossein MA, van Stipdonk AMW, Plesinger F, Kloosterman M, Wouters PC, Salden OAE, et al. Reduction in the QRS area after cardiac resynchronization therapy is associated with survival and echocardiographic response. J Cardiovasc Electrophysiol. 2021;32(3):813–22. pmid:33476467
- 53. Catanzariti D, Maines M, Cemin C, Broso G, Marotta T, Vergara G. Permanent direct his bundle pacing does not induce ventricular dyssynchrony unlike conventional right ventricular apical pacing. J Interv Card Electrophysiol. 2006;16(2):81–92.
- 54. Arnold AD, Shun-Shin MJ, Ali N, Keene D, Howard JP, Chow J-J, et al. Left ventricular activation time and pattern are preserved with both selective and nonselective His bundle pacing. Heart Rhythm O2. 2021;2(5):439–45. pmid:34667958
- 55. Zhang J, Guo J, Hou X, Wang Y, Qian Z, Li K, et al. Comparison of the effects of selective and non-selective His bundle pacing on cardiac electrical and mechanical synchrony. Europace. 2018;20(6):1010–7. pmid:28575215
- 56. Schaufelberger M, Schuler S, Bear L, Cluitmans M, Coll-Font J, Onak ÖN, et al. Comparison of Activation Times Estimation for Potential-Based ECG Imaging. Comput Cardiol (2010). 2019;46:10.22489/cinc.2019.379. pmid:32190705
- 57. Vassallo JA, Cassidy DM, Kindwall KE, Marchlinski FE, Josephson ME. Nonuniform recovery of excitability in the left ventricle. Circulation. 1988;78(6):1365–72. pmid:3191591
- 58. Antzelevitch C. Role of spatial dispersion of repolarization in inherited and acquired sudden cardiac death syndromes. Am J Physiol Heart Circ Physiol. 2007;293(4):H2024-38. pmid:17586620
- 59. Cho EJ, Park S-J, Park KM, On YK, Kim JS. Paced QT interval as a risk factor for new-onset left ventricular systolic dysfunction and cardiac death after permanent pacemaker implantation. Int J Cardiol. 2016;203:158–63. pmid:26523358
- 60. Del-Carpio Munoz F, Gharacholou SM, Scott CG, Nkomo VT, Lopez-Jimenez F, Cha YM, et al. Prolonged ventricular conduction and repolarization during right ventricular stimulation predicts ventricular arrhythmias and death in patients with cardiomyopathy. JACC: Clinical Electrophysiology. 2017;3(13):1580–91.
- 61. Schurr JW, Grewal PK, Fan R, Rashba E. QT interval measurement in ventricular pacing: Implications for assessment of drug effects and pro-arrhythmia risk. J Electrocardiol. 2022;70:13–8. pmid:34826635
- 62. Moriña-Vázquez P, Moraleda-Salas MT, López-Masjuan-Ríos Á, Esteve-Ruiz I, Arce-León Á, Lluch-Requerey C, et al. Improvement in electrocardiographic parameters of repolarization related to sudden death in patients with ventricular dysfunction and left bundle branch block after cardiac resynchronization through His bundle pacing. J Interv Card Electrophysiol. 2023;66(9):2003–10. pmid:36930350
- 63. Greenland P, Xie X, Liu K, Colangelo L, Liao Y, Daviglus ML, et al. Impact of minor electrocardiographic ST-segment and/or T-wave abnormalities on cardiovascular mortality during long-term follow-up. Am J Cardiol. 2003;91(9):1068–74. pmid:12714148
- 64. Anttila I, Nikus K, Kähönen M, Jula A, Reunanen A, Salomaa V, et al. Prognostic implications of quantitative ST-segment characteristics and T-wave amplitude for cardiovascular mortality in a general population from the Health 2000 Survey. Ann Med. 2010;42(7):502–11. pmid:20854212
- 65. Kenttä TV, Sinner MF, Nearing BD, Freudling R, Porthan K, Tikkanen JT, et al. Repolarization Heterogeneity Measured With T-Wave Area Dispersion in Standard 12-Lead ECG Predicts Sudden Cardiac Death in General Population. Circ Arrhythm Electrophysiol. 2018;11(2):e005762. pmid:29440187
- 66. Végh EM, Engels EB, van Deursen CJM, Merkely B, Vernooy K, Singh JP, et al. T-wave area as biomarker of clinical response to cardiac resynchronization therapy. Europace. 2016;18(7):1077–85. pmid:26462704
- 67. Nguyên UC, Claridge S, Vernooy K, Engels EB, Razavi R, Rinaldi CA, et al. Relationship between vectorcardiographic QRSarea, myocardial scar quantification, and response to cardiac resynchronization therapy. J Electrocardiol. 2018;51(3):457–63. pmid:29454649
- 68. Verzaal NJ, van Deursen CJM, Pezzuto S, Wecke L, van Everdingen WM, Vernooy K, et al. Synchronization of repolarization after cardiac resynchronization therapy: A combined clinical and modeling study. J Cardiovasc Electrophysiol. 2022;33(8):1837–46. pmid:35662306
- 69. Dural M, Ghossein MA, Gerrits W, Daniels F, Meine M, Maass AH, et al. Association of vectorcardiographic T-wave area with clinical and echocardiographic outcomes in cardiac resynchronization therapy. Europace. 2023;26(1):euad370. pmid:38146837
- 70. Rizas KD, McNitt S, Hamm W, Massberg S, Kääb S, Zareba W, et al. Prediction of sudden and non-sudden cardiac death in post-infarction patients with reduced left ventricular ejection fraction by periodic repolarization dynamics: MADIT-II substudy. Eur Heart J. 2017;38(27):2110–8. pmid:28431133
- 71. Hamm W, Stülpnagel L, Vdovin N, Schmidt G, Rizas KD, Bauer A. Risk prediction in post-infarction patients with moderately reduced left ventricular ejection fraction by combined assessment of the sympathetic and vagal cardiac autonomic nervous system. Int J Cardiol. 2017;249:1–5. pmid:29121716
- 72. Schüttler D, Hamm W, Bauer A, Brunner S. Routine heart rate-based and novel ECG-based biomarkers of autonomicnervous system in sports medicine. Dtsch Z Sportmed. 2020;71(6):141–50.
- 73.
Boas R, Sappler N, von Stuülpnagel L, Klemm M, Dixen U, Thune JJ. Periodic repolarization dynamics identifies ICD responders in nonischemic cardiomyopathy: a DANISH substudy. Circulation. 2022.
- 74. Liao D, Nedergaard RB, Unnisa M, Mahapatra SJ, Faghih M, Phillips AE, et al. Electrocardiography Assessment of Sympatico-Vagal Balance during Resting and Pain Using the Texas Instruments ADS1299. Bioengineering (Basel). 2023;10(2):205. pmid:36829699
- 75. Palacios S, Caiani EG, Landreani F, Martínez JP, Pueyo E. Long-term microgravity exposure increases ECG repolarization instability manifested by low-frequency oscillations of T-wave vector. Frontiers in Physiology. 2019;10:1510.
- 76. Lellouche N, De Diego C, Boyle NG, Wiener I, Akopyan G, Child JS, et al. Relationship between mechanical and electrical remodelling in patients with cardiac resynchronization implanted defibrillators. Europace. 2011;13(8):1180–7. pmid:21486911
- 77. Banavalikar B, Thajudeen A, Namboodiri N, Nair KKM, Pushpangadhan AS, Valaparambil AK. Long-term effects of cardiac resynchronization therapy on electrical remodeling in heart failure—A prospective study. Pacing and Clinical Electrophysiology. 2017;40(11):1279–85.
- 78. Dunnink A, Stams TRG, Bossu A, Meijborg VMF, Beekman JDM, Wijers SC, et al. Torsade de pointes arrhythmias arise at the site of maximal heterogeneity of repolarization in the chronic complete atrioventricular block dog. Europace. 2017;19(5):858–65. pmid:28525920
- 79. Palmisano P, Collantoni M, Bonanno S, Giuranna M, Santoro A, Dattilo G, et al. Long-term outcome of His bundle pacing with a current of injury compared with left bundle branch area pacing. Heart Rhythm. 2025;22(10):e1032–3. pmid:40541685
- 80. Hussain MA, Furuya-Kanamori L, Kaye G, Clark J, Doi SAR. The Effect of Right Ventricular Apical and Nonapical Pacing on the Short- and Long-Term Changes in Left Ventricular Ejection Fraction: A Systematic Review and Meta-Analysis of Randomized-Controlled Trials. Pacing Clin Electrophysiol. 2015;38(9):1121–36. pmid:26096902
- 81. Rosenbaum MB, Blanco HH, Elizari MV, Lázzari JO, Davidenko JM. Electrotonic modulation of the T wave and cardiac memory. Am J Cardiol. 1982;50(2):213–22. pmid:7102553