Abstract
This study aimed to evaluate the dosimetric benefits of magnetic resonance (MR)-guided radiotherapy (MRgRT) compared to computed tomography (CT)-guided RT for partial breast treatment. First, treatment plan quality was compared between MRgRT with step-and-shoot (S&S) IMRT and CT-guided RT with volumetric-modulated arc therapy (VMAT). Second, the dosimetric benefit of online adaptation in MRgRT was assessed by comparing the initial planned doses with daily MR-defined target geometry to those with planning CT-defined target geometry. Plan quality for MRgRT was comparable to, or slightly lower than, CT-guided RT. The V95% of the planning target volume (PTV) was lower by 1.72% and 1.94%, respectively, with a lower minimum dose and a higher maximum dose. For the OARs, the dosimetric parameters were generally similar, but for the skin, the Dmax, D0.5mL, and D1mL were slightly higher in MRgRT. Evaluation of the initial planned doses with the daily MR-based target definition resulted in a maximum deviation of 46% for the PTV V95%, demonstrating uncertainty in CT-based target definition. Although the MRgRT planning quality may be slightly compromised due to the S&S IMRT technique, the dosimetric benefits from online adaptation can be substantial when CT-based target definitions need to be modified based on daily MR imaging guidance.
Citation: Sung J, Song G, Jeong I, Ha J, Cho Y, Lee J, et al. (2026) Dosimetric benefits of 1.5 T MR-guided radiotherapy in partial breast treatment. PLoS One 21(3): e0342973. https://doi.org/10.1371/journal.pone.0342973
Editor: Minsoo Chun, Chung-Ang University Gwangmyeong Hospital, KOREA, REPUBLIC OF
Received: October 2, 2025; Accepted: January 29, 2026; Published: March 12, 2026
Copyright: © 2026 Sung et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Data Availability: All relevant data are within the manuscript.
Funding: This study was supported by the Innovative Research Grant from Gangnam Severance Hospital, Yonsei University College of Medicine in the form of a grant awarded to JK, and by the Basic Science Research Program through the National Research Foundation of Korea (NRF), funded by the Ministry of Science and ICT in the form of a grant awarded to JS (2022R1A2C1092888). The specific roles of these authors are articulated in the ‘author contributions’ section. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
Competing interests: The authors have declared that no competing interests exist.
1. Introduction
Partial breast irradiation (PBI) has been increasingly utilized in postoperative breast radiotherapy for patients with early-stage breast cancer [1,2]. This treatment has shown comparable efficacy to whole breast irradiation consisting of 5 weeks of external beam radiotherapy with or without an additional 1–2 weeks of boost irradiation [3,4], with superior outcomes in terms of acute and late treatment-related toxicity and cosmesis [5,6]. As radiation beams in PBI are targeted only at a part of the breast volume, accurately identifying the target volume is essential. While PBI target identification is typically achieved by locating surgical clips inserted during breast surgery or detecting a seroma, challenges arise because of potential clip migration or variations in seroma volume throughout radiation treatment [7–9].
Magnetic resonance (MR)-guided RT (MRgRT) can be an optimal treatment technique for PBI because of high soft tissue contrast during daily MR imaging, potentially leading to accurate target definition [10]. In MRgRT, daily MR images are acquired before treatment beam irradiation, whereas in conventional RT, cone-beam computed tomography (CT) or megavoltage CT images are obtained. Detailed visualization of soft tissues, particularly seroma, is a notable advantage of MR imaging [11–13]. When changes in anatomic shape or seroma volume are detected in daily MR images, MRgRT can modify the initial structure based on the image findings and can establish online adaptive plans accordingly. Because of these advantages, recent studies have focused on determining the appropriate planning target volume (PTV) margin size and evaluating the PTV margin in MR-Linac systems [14–16]. For PBI, reducing the PTV margin to 3 mm or less in MRgRT using 0.35 T MR-Linac may offer better normal tissue protection than that with volumetric-modulated arc therapy (VMAT) with a 10 mm margin [17].
However, treatment machines currently used for MRgRT have some limitations when compared to conventional Linacs. The Elekta Unity (Elekta AB, Stockholm, Sweden) and ViewRay MRIdian (ViewRay, Inc, Oakwood Village, OH), which are currently available commercial MR-Linacs, can only conduct step-and-shoot (S&S) intensity-modulated radiation therapy (IMRT). Because of this limitation, the resulting treatment plan quality of S&S IMRT using MR-Linac may be lower than that of VMAT using conventional Linacs. A comparative study of treatment plans for patients with prostate and rectal cancer showed a decrease in plan homogeneity and increase in the PTV Dmean, D1%, and doses to organs at risk (OARs) with Elekta Unity’s S&S compared to the conventional Linac’s VMAT [18]. In cases of anal cancer, while doses for target and OARs were comparable, the PTV45Gy D98% coverage was compromised in 3 out of 10 MR-Linac plans [19]. For APBI treatment, the ViewRay MRIdian’s S&S IMRT results were similar to or slightly worse than those of VMAT [17]. However, for PBI using MR-Linac, plan quality evaluations have only been conducted for ViewRay MRIdian, with no studies addressing the dosimetric benefits related to the changes in target position and shape during treatment.
Therefore, this study aimed to comprehensively evaluate the dosimetric benefits of MRgRT over CT-guided RT for partial breast radiotherapy. First, the plan quality of S&S IMRT plans generated on the MRgRT system and VMAT plans created on a conventional linac was compared using the same CT images and identical contour structures. In the second, the MR-based target structures modified at the treatment were applied to the dose distribution of the original CT-based MRgRT initial plan. This enabled a quantitative assessment of the dose differences that would occur if each patient were treated with the initial non-adapted plan, thereby allowing an evaluation of the dosimetric advantages of online adaptation.
2. Materials and methods
2.1. Patient data
This study was a retrospective analysis approved by the Institutional Review Board (IRB) of the Yonsei University Health System (IRB No. 3-2024-0233). The study was conducted through a retrospective review of medical records, and access to the data for research purposes was granted on August 14, 2024. All patient data were fully anonymized during and after data collection, and the investigators had no access to any information that could directly identify individual participants. The study included 20 breast cancer patients who underwent MRgRT using the adaptive-to-shape (ATS) workflow on an Elekta Unity system [20]. For all patients, the PTV was adjusted through the ATS process based on the visibility of the tumor bed on the daily MR image acquired on the first treatment day. Table 1 shows patient clinical information as well as the differences between the initial PTV defined on the planning CT and the modified PTV revised based on the daily MR image acquired at the first treatment fraction.
To evaluate the dosimetric benefit of online adaptation in MRgRT associated with PTV changes, the 20 patients were classified into three groups according to patterns of PTV positional and volumetric changes based on the results shown in Table 1. First, three patients (Patients 12, 16, and 17) who showed prominent positional changes between the initial PTV and the modified PTV in both visual assessment and quantitative centroid differences were separately classified as Group 1 (positional-change group). Group 2 (volumetric-change group) consisted of seven patients whose absolute PTV volume change exceeded 10 mL. Finally, Group 3 (minimal-change group) included the remaining ten patients who showed relatively small changes in both PTV volume and position.
2.2. Acquisition of planning CT and delineation of structures
Planning CT images were acquired in the supine position using a head cushion, wing board, and knee step provided by Elekta, with a 2 mm slice thickness. Target volumes were defined based on the planning CT image according to the departmental standard protocols using MIM Maestro v7.2 (MIM Software Inc., Cleveland, OH). Initially, the tumor bed was defined as the surgical cavity, including surgical clips and any changes in the surrounding tissue architecture. A 1-cm margin around the tumor bed was used to create the clinical target volume (CTV). Subsequently, it was modified to exclude the chest wall and a 5-mm strip of skin [21]. As in cases where the PTV was set equal to the CTV during PBI treatment with CyberKnife [22], the PTV was defined the same as the CTV in this study because MRgRT allows for robust visualization of the tumor bed, target modification in every treatment, and motion monitoring to detect intrafractional variation. OARs (ipsilateral breast, contralateral breast, lung, heart, and skin) were also contoured according to standard departmental guidelines (Fig 1). All structures were contoured and verified by the board-certified radiation oncologist responsible for each patient.
PTV, planning target volume; OARS, organs at risk.
2.3. Treatment planning
VMAT plans for Versa HD (Elekta AB, Stockholm, Sweden) were created with the beam isocenter set at the center of the target. To facilitate a fair comparison between the MRgRT and CT-guided RT plans, all the treatment plans were generated to ensure that both plans for the same patient were of the highest possible quality and consistency, following the departmental dosimetric criteria (Table 2). Since each patient had a different target size or location, optimization objectives were individually fine-tuned to minimize OAR doses while maintaining PTV dose constraints. Single-arc VMAT plans for the CT-guided RT were generated with gantry angle intervals of approximately 180–200°, with a minimum segment of 5 mm at the isocenter for delivery on Versa HD (Fig 2(a)). All plans were calculated using a collapsed cone convolution algorithm with a 3 mm calculation grid spacing in RayStation v5.0.3 (RaySearch Lab, Stockholm, Sweden).
The number of red arrows indicates the number of beams, and the red arrow direction indicates beam direction (Blue: PTV).
MRgRT plans were generated using Monaco 5.51 (Elekta AB, Stockholm, Sweden). The S&S IMRT plans were generated using 4–6 beams and fewer than 150 segments with 7 MV flattened filter-free beams (Fig 2(b)). Optimization objectives were adjusted for individual patient plans to achieve the departmental dosimetric criteria summarized in Table 2. All plans were calculated using a GPU-accelerated Monte Carlo dose calculation algorithm with 1% dose uncertainty and a 3 mm calculation grid spacing. Plans were considered acceptable when all objectives were met.
2.4. Comparison of plan quality: S&S IMRT vs. VMAT
Using the contoured structures based on the planning CT, the S&S IMRT plans for MRgRT were dosimetrically compared to VMAT plans retrospectively generated for CT-guided RT using Elekta Versa HD. All plans were compared using the evaluation parameters shown in Table 2. The conformity index (CI) and homogeneity index (HI) were calculated using the following formulas: [23]
where total body V95% represents the volume receiving 95% of the prescribed dose in the patient body.
where D5% and D95% represent the minimum doses to 5% and 95% of the PTV, respectively. Statistical analysis for all measures used the Student’s t-test, with statistical significance set at values of p < 0.05.
2.5. Evaluation of dosimetric benefits of online adaptation in MRgRT
To evaluate the dosimetric benefits of online adaptation in MRgRT, we analyzed the impact of uncertainties in CT-based target definition. This analysis was designed to simulate a clinical scenario in which treatment is delivered without MR-guided adaptation, relying solely on CT-defined target structures. For each of the 20 patients, a conventional CT-based treatment plan was created using the RayStation treatment planning system for delivery on a Versa HD linear accelerator. The initial dose distribution from this plan was then mapped to the modified PTV structure, which was delineated on the MR image acquired on the first day of treatment (Fig 3). The PTV location and shape were redefined by board-certified radiation oncologists based on daily MR imaging during the online adaptation process. Rigid registration between the CT and MR images was performed using the sternum and the breast on the treatment side as anatomical landmarks, following standard CT and CBCT alignment protocols commonly used in conventional radiotherapy. Several dose volume histogram metrics for the PTV, including the volume receiving at least 95 percent of the prescribed dose (V95), the maximum dose (Dmax), the minimum dose (Dmin), the homogeneity index (HI), and the conformity index (CI), were calculated based on the modified PTV structure and compared with the corresponding values from the initial PTV in the reference plan.
The CT-based dose distribution was rigidly registered to the daily MR image acquired at the first treatment fraction and mapped onto the MR-based recontoured PTV. Red and white isodose lines represent 95% and 10% of the prescribed dose, respectively. The CT-based PTV is shown in red, and the MR-based recontoured PTV is shown in blue.
3. Results
Clinically acceptable treatment plans were achieved for all the patients with MRgRT and CT-guided RT. Fig 4 compares the qualitative dose distributions of Patient 1 and 18, who received 30 Gy and 10 Gy, respectively, and Table 3 provides a quantitative analysis across all 20 patients. For the PTV coverage and uniformity, the VMAT plans of CT-guided RT showed slightly better results compared to the S&S IMRT plan of the MRgRT for both PBI and the boost treatment. The V95% of MRgRT were 1.72% and 1.94% lower than those of CT-guided RT for PBI and the boost treatment, respectively (p = 0.003*, 0.002*). The Dmin of the MRgRT were 7.89% and 1.27% lower than those of CT-guided RT, whereas the Dmax of MRgRT were 0.83% and 5.31% higher than those of CT-guided RT for the PBI and boost treatment, respectively. Overall, the HI and CI of the MRgRT plans were similar or slightly worse compared to the CT-guided RT plans. For OARs, the dosimetric evaluation parameters of MRgRT are similar to those of CT-guided RT, except for the skin receiving PBI treatment. The Dmax, D0.5mL and D1mL to the skin were 3.6, 3.4, and 3.2 Gy higher in MRgRT than in CT-guided RT, respectively (p < 0.001*, p = 0.004*, 0.006*, respectively).
(a) MR-guided S&S IMRT plan for the 30 Gy PBI treatment of Patient 1. (b) CT-guided VMAT plan for the 30 Gy PBI treatment of Patient 1. (c) MR-guided S&S IMRT plan for the 10 Gy boost treatment of Patient 18. (d) CT-guided VMAT plan for the 10 Gy boost treatment of Patient 18.
We simulated how much PTV doses changed compared to the initial planned doses during the treatment period. As described in Section 2.1 (Patient data), patients in this study were classified into three groups based on the differences between the CT-based PTV and the MR-based PTV modified using the daily MR image acquired at the first treatment fraction: Group 1, which exhibited pronounced positional changes; Group 2, which showed substantial volumetric changes; and Group 3, which demonstrated relatively minimal changes in both position and volume. Table 4 represents the results for Group 2 and 3 to evaluate the dose delivered to the PTV according to changes in volume. V95%, Dmax, Dmin, HI, and CI of the PTV that decreased more than 10 mL were similar to those of the initial PTV except for Dmin and CI which increased by 20.14% and 2.95. For the PTV that increased more than 10 mL than the initial PTV, the V95% and Dmin were reduced by 16.72% and 71.52%, respectively. HI increased by 3.74 while the remaining dosimetric parameters were similar to those of the initial target. For PTVs whose changed volume ranged from −10 mL to 10 mL, the V95% and Dmin were reduced by 12.25% and 42.49%, respectively, while the HI increased by 0.25.
As shown in Fig 5, Group 1 (Patient 12, 16, and 17) was noticeably altered in the daily MR images, substantial reductions in target coverage were observed when the initial CT-based dose distribution was mapped onto the modified PTVs. As shown in Table 5, while Dmax remained comparable between the initial and modified PTVs, the coverage of the target volume was significantly reduced. For Patient 12, the V95% and Dmin of the modified PTV dropped to 53.3% and 23.6%, respectively, compared to 99.4% and 88.0% in the original CT-based PTV. Similar trends were observed in Patients 16 and 17. The V95% decreased from 100.0% and 99.7% to 55.1% and 59.8%, respectively, and the Dmin decreased from 94.9% and 85.9% to 16.3% and 3.8%, respectively.
For Patient 12: reference CT (a), daily MR (b), and corresponding PTV structures. For Patient 16: reference CT (c), daily MR (d), and corresponding PTV structures. For Patient 17: reference CT (e), daily MR (f); no PTV is visible in the CT slice shown. Blue contours are the initial PTV structures based on reference CT, and yellow contours are the modified PTV structures based on daily MR.
4. Discussion
MRgRT enables the delivery of online adaptive radiotherapy based on daily MR imaging and is therefore considered suitable for PBI. Compared with CT imaging, MR imaging provides superior soft-tissue contrast, allowing clearer visualization of the surgical cavity and seroma. In addition, anatomical variations such as changes in seroma that may occur between CT simulation and the actual treatment course can be identified on daily MR images and addressed through online adaptation, representing an important advantage in PBI treatment. However, to the best of the authors’ knowledge, the dosimetric benefit of MRgRT-based online adaptation has not yet been evaluated for PBI. Therefore, this study aimed to assess the dosimetric advantages provided by MRgRT-based online adaptation in patients undergoing PBI.
First, plan quality between S&S IMRT and VMAT was compared to assess whether MRgRT is disadvantaged by limitations of the planning technique. The two techniques showed generally comparable plan quality with no clinically meaningful differences, consistent with previous studies [24–26]. Subsequently, this study evaluated the impact of image-based target definition on dosimetric outcomes. The PTV recontoured on the daily MR image acquired on the first treatment day was assumed to best represent the actual target shape and position. The dose delivered to this modified PTV was quantitatively evaluated by applying the same CT-based dose distribution from the initial treatment plan and compared with the dose to the original CT-based PTV in that plan.
We used only the first fraction MR based PTV because previous studies have shown that the largest anatomical changes occur between CT simulation and the start of treatment, while changes during later fractions are minimal [27,28]. In addition, both the PBI and boost treatments in this study consisted of five fractions delivered consecutively over a short treatment course, and meaningful anatomical changes were rarely observed after the first fraction.
Among the 20 patients, excluding the 3 with the most significant changes, 4 of the remaining 17 patients experienced a reduction in target volume by more than 10 mL, resulting in increased target coverage. However, for the remaining 13 patients, the PTV was mostly expanded, which included areas that were not initially covered by the original PTV, leading to a decrease in target coverage and minimum dose.
In the three cases, not only the shape and volume but also the position of the PTV showed changes. As shown in Table 1, the differences in the center positions between the CT-based initial PTV and the PTV redefined on the first-day MR image were 22.6 mm for Patient 12, 9.6 mm for Patient 16, and 11.6 mm for Patient 17. These positional differences can be explained by two factors. First, the actual target location may have changed due to seroma changes during the interval between CT simulation and the first treatment day. Second, the superior soft-tissue contrast of MR imaging allows the surgical cavity to be visualized much more clearly than on CT, enabling a more accurate redefinition of the PTV. As shown in Fig 5, Patient 12 (Fig 5(a), (b)) appears to exhibit a positional difference due to seroma change and the unclear seroma boundary on CT. For patients 16 and 17 (Fig 5C–(f)), the surgical cavity was visualized much more clearly on MR than on CT, allowing the postoperative structures to be identified more accurately and the PTV to be redefined accordingly.
When comparing the initial planned doses to the initial PTV with those to the modified PTV obtained after the online adaptive planning process, V95% decreased by 46.1%, 45.0%, and 40.0%, respectively. While conventional linac systems typically use CBCT to acquire images for daily anatomical positioning, the lower resolution of CBCT makes it difficult to confirm the PTV’s position. In such situations, MRgRT can precisely identify the seroma and tumor bed positions using MR images and perform contour adjustments and replanning through online adaptive processes, making it specialized for partial breast treatment.
This study has several limitations. First, the dose distribution was not recalculated on the MR images. Instead, the dose from the CT-based treatment plan was transferred to the MR-defined structures using rigid registration. Therefore, the resulting dose distribution does not represent a physically accurate distribution based on the actual MR anatomy, but should be interpreted as a conceptual dosimetric estimation. Second, this study focused on evaluating PTV dose differences caused by visualization discrepancies between CT and MR images. As such, dose evaluations for OARs and cumulative anatomical changes over the treatment course were not included in the analysis. Therefore, we are planning a study that will incorporate physically accurate dose recalculation on MR images, deformable image registration, and quantitative dose evaluation for OARs.
5. Conclusion
While MRgRT using S&S IMRT may compromise plan quality compared to CT-guided RT using VMAT, it offers substantial advantages by allowing precise adjustments based on daily MR imaging. This capability to accurately visualize and adapt to changes in the target’s position and anatomy throughout the treatment course significantly enhances treatment accuracy and efficacy.
References
- 1. Ott OJ, Stillkrieg W, Lambrecht U, Schweizer C, Lamrani A, Sauer T-O, et al. External-Beam-Accelerated Partial-Breast Irradiation Reduces Organ-at-Risk Doses Compared to Whole-Breast Irradiation after Breast-Conserving Surgery. Cancers (Basel). 2023;15(12):3128. pmid:37370738
- 2. Polgár C, Major T, Takácsi-Nagy Z, Fodor J. Breast-Conserving Surgery Followed by Partial or Whole Breast Irradiation: Twenty-Year Results of a Phase 3 Clinical Study. Int J Radiat Oncol Biol Phys. 2021;109(4):998–1006. pmid:33186620
- 3. Polgár C, Fodor J, Major T, Sulyok Z, Kásler M. Breast-conserving therapy with partial or whole breast irradiation: ten-year results of the Budapest randomized trial. Radiother Oncol. 2013;108(2):197–202. pmid:23742961
- 4. Chang JS, Chang JH, Kim N, Kim YB, Shin KH, Kim K. Intensity Modulated Radiotherapy and Volumetric Modulated Arc Therapy in the Treatment of Breast Cancer: An Updated Review. J Breast Cancer. 2022;25(5):349–65. pmid:36265885
- 5. Goldberg M, Bridhikitti J, Khan AJ, McGale P, Whelan TJ. A Meta-Analysis of Trials of Partial Breast Irradiation. Int J Radiat Oncol Biol Phys. 2023;115(1):60–72. pmid:36155214
- 6. Meattini I, Marrazzo L, Saieva C, Desideri I, Scotti V, Simontacchi G, et al. Accelerated Partial-Breast Irradiation Compared With Whole-Breast Irradiation for Early Breast Cancer: Long-Term Results of the Randomized Phase III APBI-IMRT-Florence Trial. J Clin Oncol. 2020;38(35):4175–83. pmid:32840419
- 7. Sung S, Lee JH, Lee JH, Kim SH, Kwak Y-K, Lee S-W, et al. Displacement of Surgical Clips during Postoperative Radiotherapy in Breast Cancer Patients Who Received Breast-Conserving Surgery. J Breast Cancer. 2016;19(4):417–22. pmid:28053630
- 8. Jeon SH, Shin KH, Park S-Y, Kim J-I, Park JM, Kim JH, et al. Seroma change during magnetic resonance imaging-guided partial breast irradiation and its clinical implications. Radiat Oncol. 2017;12(1):103. pmid:28633637
- 9. Kader HA, Truong PT, Pai R, Panades M, Jones S, Ansbacher W, et al. When is CT-based postoperative seroma most useful to plan partial breast radiotherapy? Evaluation of clinical factors affecting seroma volume and clarity. Int J Radiat Oncol Biol Phys. 2008;72(4):1064–9. pmid:18411005
- 10. Berlangieri A, Elliott S, Wasiak J, Chao M, Foroudi F. Use of magnetic resonance image-guided radiotherapy for breast cancer: a scoping review. J Med Radiat Sci. 2022;69(1):122–33. pmid:34523823
- 11. Lowrey N, Koch CA, Purdie T, Simeonov A, Conroy L, Han K. Magnetic Resonance Imaging for Breast Tumor Bed Delineation: Computed Tomography Comparison and Sequence Variation. Adv Radiat Oncol. 2021;6(4):100727. pmid:34409213
- 12. Al-Hammadi N, Caparrotti P, Divakar S, Riyas M, Chandramouli SH, Hammoud R, et al. MRI Reduces Variation of Contouring for Boost Clinical Target Volume in Breast Cancer Patients Without Surgical Clips in the Tumour Bed. Radiol Oncol. 2017;51(2):160–8. pmid:28740451
- 13. Giezen M, Kouwenhoven E, Scholten AN, Coerkamp EG, Heijenbrok M, Jansen WPA, et al. MRI- versus CT-based volume delineation of lumpectomy cavity in supine position in breast-conserving therapy: an exploratory study. Int J Radiat Oncol Biol Phys. 2012;82(4):1332–40. pmid:21708426
- 14. Eijkelenkamp H, Boekhoff MR, Verweij ME, Peters FP, Meijer GJ, Intven MPW. Planning target volume margin assessment for online adaptive MR-guided dose-escalation in rectal cancer on a 1.5 T MR-Linac. Radiother Oncol. 2021;162: 150–5. pmid:34280404
- 15. Kim J, Sung J, Lee SJ, Cho KS, Chung BH, Yang D, et al. Optimal planning target margin for prostate radiotherapy based on interfractional and intrafractional variability assessment during 1.5T MRI-guided radiotherapy. Front Oncol. 2023;13:1337626. pmid:38173837
- 16. de Muinck Keizer DM, Kerkmeijer LGW, Willigenburg T, van Lier ALHMW, Hartogh MDd, van der Voort van Zyp JRN, et al. Prostate intrafraction motion during the preparation and delivery of MR-guided radiotherapy sessions on a 1.5T MR-Linac. Radiother Oncol. 2020;151:88–94. pmid:32622779
- 17. Musunuru HB, Yadav P, Olson SJ, Anderson BM. Improved Ipsilateral Breast and Chest Wall Sparing With MR-Guided 3-fraction Accelerated Partial Breast Irradiation: A Dosimetric Study Comparing MR-Linac and CT-Linac Plans. Adv Radiat Oncol. 2021;6(3):100654. pmid:34195491
- 18. van de Schoot AJAJ, van den Wollenberg W, Carbaat C, de Ruiter P, Nowee ME, Pos F, et al. Evaluation of plan quality in radiotherapy planning with an MR-linac. Phys Imaging Radiat Oncol. 2019;10:19–24. pmid:33458263
- 19. Smith D, Knight K, Sim J, Lim Joon D, Foroudi F, Khoo V. A planning-based feasibility study of MR-Linac treatment for anal cancer radiation therapy. Med Dosim. 2023;48(4):267–72. pmid:37507334
- 20. Winkel D, Bol GH, Kroon PS, van Asselen B, Hackett SS, Werensteijn-Honingh AM, et al. Adaptive radiotherapy: The Elekta Unity MR-linac concept. Clin Transl Radiat Oncol. 2019;18:54–9. pmid:31341976
- 21. Shaitelman SF, Anderson BM, Arthur DW, Bazan JG, Bellon JR, Bradfield L, et al. Partial Breast Irradiation for Patients With Early-Stage Invasive Breast Cancer or Ductal Carcinoma In Situ: An ASTRO Clinical Practice Guideline. Pract Radiat Oncol. 2024;14(2):112–32. pmid:37977261
- 22. Lee WH, Chang JS, Kim MJ, Park VY, Yoon JH, Kim SY, et al. First Experience in Korea of Stereotactic Partial Breast Irradiation for Low-Risk Early-Stage Breast Cancer. Front Oncol. 2020;10:672. pmid:32411612
- 23. Cho Y, Cho YJ, Chang WS, Kim JW, Choi WH, Lee IJ. Evaluation of optimal treatment planning for radiotherapy of synchronous bilateral breast cancer including regional lymph node irradiation. Radiat Oncol. 2019;14(1):56. pmid:30935400
- 24. van de Schoot AJAJ, van den Wollenberg W, Carbaat C, de Ruiter P, Nowee ME, Pos F, et al. Evaluation of plan quality in radiotherapy planning with an MR-linac. Phys Imaging Radiat Oncol. 2019;10:19–24. pmid:33458263
- 25. Da Silva Mendes V, Nierer L, Li M, Corradini S, Reiner M, Kamp F, et al. Dosimetric comparison of MR-linac-based IMRT and conventional VMAT treatment plans for prostate cancer. Radiat Oncol. 2021;16(1):133. pmid:34289868
- 26. Nachbar M, Mönnich D, Kalwa P, Zips D, Thorwarth D, Gani C. Comparison of treatment plans for a high-field MRI-linac and a conventional linac for esophageal cancer. Strahlentherapie und Onkologie. 2019;195: 327–334.
- 27. Jeon SH, Shin KH, Park S-Y, Kim J-I, Park JM, Kim JH, et al. Seroma change during magnetic resonance imaging-guided partial breast irradiation and its clinical implications. Radiat Oncol. 2017;12(1):103. pmid:28633637
- 28. Truong MT, Hirsch AE, Kovalchuk N, Qureshi MM, Damato A, Schuller B, et al. Cone-beam computed tomography image guided therapy to evaluate lumpectomy cavity variation before and during breast radiotherapy. J Appl Clin Med Phys. 2013;14(2):4243. pmid:23470944