Determining the reliability of the Vitalight CO2 Monitor via the assessment of the carbon dioxide profile on city transit buses

  • Courtney E. Lessel ,

    Roles Conceptualization, Data curation, Formal analysis, Funding acquisition, Investigation, Methodology, Visualization, Writing – original draft, Writing – review & editing

    clessel@laurentian.ca

    Affiliation Centre for Research in Occupational Safety and Health, Laurentian University, Greater Sudbury, Ontario, Canada

  • Katie A. Goggins ,

    Contributed equally to this work with: Katie A. Goggins, Alison A. Godwin

    Roles Conceptualization, Funding acquisition, Supervision, Validation, Writing – review & editing

    Affiliations Centre for Research in Occupational Safety and Health, Laurentian University, Greater Sudbury, Ontario, Canada, Bharti School of Engineering and Computation, Laurentian University, Greater Sudbury, Ontario, Canada

  • Alison A. Godwin

    Contributed equally to this work with: Katie A. Goggins, Alison A. Godwin

    Roles Conceptualization, Funding acquisition, Resources, Supervision, Validation, Writing – review & editing

    Affiliations Centre for Research in Occupational Safety and Health, Laurentian University, Greater Sudbury, Ontario, Canada, School of Kinesiology and Health Sciences, Laurentian University, Greater Sudbury, Ontario, Canada

Abstract

Public transit vehicles represent small, dynamic, indoor spaces, with conditions that can be favourable for the development of poor air quality. The use of CO2 monitors is recommended as a potential strategy for improving ventilation; however, information around how non-experts can use them for personal monitoring is limited. The purpose of this field study was to assess the real-world usability of a low-cost monitor, and to provide general recommendations around personal risk assessment using a monitor in combination with the evaluation of environmental factors. Measurements of CO2 and surveying of commute conditions were carried out with two CO2 monitors, the Vitalight and the Aranet, during 250 public transit trips between October 17th, 2022 and July 18th, 2023. Results show that CO2 measurements taken by the Vitalight and the Aranet were highly correlated (Time 1 (T1): r = .948; Time 2 (T2): r = .966). Additionally, the Vitalight paralleled the Aranet with respect to its responses to: temperature; relative humidity; occupancy level; seating position; how often the bus doors opened; and whether the windows were open or closed. However, measurements of CO2 taken by the Vitalight were significantly lower than those taken by the Aranet (T1: t(249) = −22.52, p < .001; T2: t(249) = −32.44, p < .001). Based on these results, we make three recommendations to provide guidance to non-experts around personal monitoring for ventilation on transit buses, including use and interpretation of the Vitalight monitor, and environmental conditions to assess, to inform actions to take to improve ventilation.

Introduction

Indoor air quality (IAQ) is the air quality in and around buildings or structures [1]. Ventilation, a component of IAQ, is the replacement of indoor air with ambient air [2,3], which helps with the dilution of contaminants [4], including: chemical and biological contaminants, and particulate matter [2]. Managing ventilation is key to maintaining good IAQ, and is a control method for reducing the transmission of airborne infections [3,58].

Low IAQ has been correlated with negative health effects [2,4]. For example, high concentrations of carbon dioxide (CO2) have been linked to headaches, sleepiness, fatigue, and reduced cognitive function [1,4,8]. Additionally, it has been positively correlated with the transmission of viruses, such as COVID-19 [9]. Recent work indicates that CO2 concentrations above 800 ppm increase the stability of COVID-19 in the air [10], increasing it’s chance of transmissibility between humans. Therefore, improving ventilation can decrease the concentration of airborne biological contaminants; potentially helping to destabilize them and, ultimately, improve health outcomes.

There are existing difficulties related to real-time monitoring, and direct measurement, of airborne biological contaminants [11]. Therefore, standards for IAQ generally regulate ventilation and filtration, relative to quantities of non-biological contaminants [12] (e.g., American Society of Heating, Refrigerating and Air-Conditioning Engineers (ASHRAE) standard 62.1–2022 [13]). In indoor spaces, most CO2 is produced by the occupants, so it is considered representative of the fraction of air that is composed of exhaled breath [1416]. Based on this assumption, CO2 monitoring is used to indicate whether the level of ventilation is sufficient to dilute airborne contaminants [13,14,1719]. It can also provide information around occupancy, which can be used to control ventilation in real-time (e.g., demand-control ventilation) [2,14].

In 2023, largely as an outcome of concerns raised during the COVID-19 pandemic, ASHRAE published standard 241, which outlines the control of infectious aerosols with ventilation, and air filtration and cleaning [20]. Additionally, greater focus has been placed on the use of personal CO2 monitors to assess ventilation as a proxy for infection risk in indoor air [10,11]. The average concentration of CO2 in ambient air is approximately 400 ppm [1,3,15,17,21], however, in exhaled breath, the concentration of CO2 is approximately 40,000 ppm [3,15,16]. Based on equations calculating the fraction of rebreathed air, at CO2 concentrations of 600 ppm, ~ 0.50–0.53% of the air is rebreathed, ~ 0.92–1.00% at 800 ppm, ~ 1.32–1.50% at 1000 ppm, ~ 1.58–2.10% at 1200 ppm, and ~2.89–4.20% at 2000ppm [22,23]. Therefore, when ventilation is insufficient, exhaled CO2 and airborne contaminants can build up [4,24]. The level of build-up is affected by: occupancy level [3,14,25]; occupant characteristics [17]; duration of time people are in the space; and activities taking place in the space [3,14].

Real-time CO2 monitoring can be used to help define high risk times for transmission, where additional control measures should be implemented [7]. Values from real-time monitoring can also be used with modelling to estimate the relative risk of infection from airborne illnesses [9,16]. Moreover, personal monitoring provides area specific information about short-term exposures that individuals can use for personal risk assessment [4]. In general, CO2 limits between 800 and 1000 ppm are recommended as indicators of insufficient ventilation [3,14,26,27]. However, during pandemic conditions, CO2 limits should be kept as close to outdoor air conditions as possible [28].

Public transit vehicles represent small, dynamic, indoor spaces [21]. These environments can pose a health risk for transit operators, and passengers, as contacts between people facilitate exposures to airborne biological contaminants [29], such as bacteria and viruses. The urban transit system in Canada recorded an average number of 126 million passenger trips per month between May 2023 and April 2024 [30]. According to data from the Canadian Human Activity Pattern Survey (CHAPS) 2, individuals spend approximately five-percent of their day in vehicles, with urban Canadians who reported using buses [31], spending more than one-hour a day on them [32].

Much of the research around IAQ in vehicles has focused on exposure to traffic-related contaminants, such as particulate matter, rather than airborne biological contaminants [4,32,33]. However, CO2 in vehicles can build up due to both vehicle emissions, and human breathing [4]. How fast CO2 builds up inside a vehicle depends on several factors, including the vehicle’s: level of occupation, air volume, and ventilation rate [4]. In urban transit vehicles, although trips are generally short, conditions (e.g., high occupation, high passenger exchange, low ventilation) can favour the propagation of airborne pathogens [4,12,26,34,35], including SARS, influenza, and RSV. Most existing standards for IAQ, such as ASHRAE standards 62.1 and 241 [13,20], apply to the control of air quality in buildings spaces. In contrast, there is a dearth of information around standards related to IAQ in ground transit vehicles, though ventilation on public transit vehicles favours natural ventilation [11].

On urban transit vehicles (i.e., buses and trams), CO2 monitors have been used to track temperature, relative humidity, CO2 concentration, and particulate matter levels across routes [21]; to assess the generation and accumulation of CO2 in the passenger space [11]; and to assess ventilation [26] and the factors that affect ventilation [36] in the vehicles. These studies focused on finding factors that affected the concentration of CO2. However, their purpose was to provide information to officials about when to implement ventilation controls, not to inform individual control strategies.

Additional studies have used CO2 monitors carried by individuals to assess air quality during different types of commuting [37], and during travel in different locations [25]. While these studies provided some information around personal monitoring, their intention was to inform peoples’ choices about general pollution exposure during commuting [37], or risks associated with specific travel areas [25]. The studies did not seek to empower people to take their own measurements with personal monitoring devices.

Currently, both the Canadian [38] and Ontarian [39] government recommendations for protecting against COVID-19, and other respiratory illnesses, include improving air quality and ventilation. The use of CO2 monitors is recommended as a strategy for improving indoor ventilation [40]; however, information around how non-experts can use them for personal risk assessment is limited [25]. Specifically, in contrast to reference (gold-standard) sensors, low cost CO2 monitors are not subject to evaluation and certification based on regulatory standards, so the monitor specifications, how to calibrate them, or how well they need to perform is not necessarily known [41]. Therefore, concerns exist regarding the accuracy, reliability, calibration requirements, and required correction of low-cost CO2 monitors [15,17,28,42]. Unreliable data can mislead users, which may contribute to the misinterpretation of contaminant levels as being acceptable or unacceptable [42].

To date, although CO2 monitoring has been recommended as a strategy to assess ventilation, and for risk assessment, previous studies have targeted large scale application for official use, not individual monitoring for personal risk assessment. There is some information that individual monitoring can be used to provide information related to individual exposure to particles, for example black carbon [43]. However, clear guidance is required to give non-experts confidence in using personal monitors, and to enable them to effectively use the monitors to make personal risk assessments [14,17,44].

To inform the use of CO2 monitors for personal risk assessment, we took measures across 250 transit trips, between October 17th, 2022 and July 18th, 2023. Using these measures, we aimed to: 1) Determine whether measurements from a low-cost CO2 monitor (Vitalight Mini) were consistent with an Aranet4Home CO2 monitor; 2) Identify factors related to the concentration of CO2 on urban transit buses, and determine whether they were consistent across the two CO2 monitors tested; and 3) Discuss the usability of a low-cost monitor, and provide some general recommendations around personal risk assessment using a monitor in combination with the evaluation of environmental factors.

Materials and methods

Transit system and bus fleet

For this field case study, data collection occurred on buses from the public transit authority in Greater Sudbury, the largest city in Northern Ontario, with a population of approximately 166,000. The public transit system includes ~59 standard buses, that operate over a network of 23 routes [45] and cover 1,077 stops [46]. In 2022, approximately 2.6 million trips were taken on public transit in Greater Sudbury, averaging out to about 22.7 trips per person in the service area [47]. Transit routes operate on a staggered schedule, with different routes starting and ending at different times. Operation times range between 5:30am and 12:40am on weekdays, and 6:00am and 12:30am on weekends.

The bus fleet is comprised entirely of Nova LFS diesel buses, acquired between approximately 2006 and 2021. All of the standard buses are 12.2 metres long (Fig 1a), and have a maximum capacity of 80 passengers; though the interior layout of the fleet varies, one of the most common layouts is illustrated in Fig 1b. The buses have two doors on the passenger side, and 10 windows total that can be opened in addition to the driver’s window. Transit buses are equipped with an air-conditioning system (AC) that is operational in the summer at high ambient air temperature (≥25 °C), and may also be used in the spring and autumn. In the winter and the autumn, the AC is turned off, and the heating systems turned on. In general, if the AC or the heating system is turned on, the windows of the bus are closed, however, this is up to the driver’s discretion. While less likely with the heating system, it is not uncommon for windows in the passenger area to be open in combination with the AC being on.

thumbnail
Fig 1. Transit bus schematic.

Schematic representation of the a) exterior and b) the most common interior layout, of the GOVA transit buses (NOVA LFS diesel bus).

https://doi.org/10.1371/journal.pone.0342566.g001

Ethics statement

Ethical approval and specific work permits or approvals were not required for this study as it took place in public spaces. It only involved environmental measurements, or observations of factors that did not require researchers to interact with the people in those spaces. Additionally, none of the information collected is personal information, or could be used to identify any people in those spaces.

CO2 monitors

Two types of CO2 monitors were used, the Vitalight Mini CO2 Detector (Vitalight: Fig 2a) (Guangdong Bioall Medical Technology Co., LTD.: Guangdong Province, People’s Republic of China), and the Aranet4 Home (Aranet: Fig 2b) (Aranet: Riga, Latvia). The technical characteristics of the Aranet and the Vitalight are summarized in Table 1. For cost comparison, in 2022, the Vitalight was $59.99+ tax CAD on Amazon and the Aranet was $349.99+ tax CAD at Canadian Tire. Combining the cost and device characteristics (Table 1), the Vitalight was selected as it was 1/7th of the cost of the Aranet, included the non-dispersive infrared (NDIR) sensor for increased accuracy of CO2 measurements [15,48], operated within the desired environmental range, and had a high sampling frequency with a simple visual output similar to the Aranet. Furthermore, at the time of the study, the Vitalight was more readily available in North America, relative to other low cost monitors.

thumbnail
Table 1. Technical characteristics of the Aranet4Home detector and the Vitalight Mini CO2 detector.

https://doi.org/10.1371/journal.pone.0342566.t001

thumbnail
Fig 2. CO2 detectors.

The a) Vitalight Mini CO2 detector, and the b) Aranet4 Home CO2 detector.

https://doi.org/10.1371/journal.pone.0342566.g002

For device calibration, the Vitalight performs an auto-calibration where it takes the three lowest CO2 readings from its previous seven days of use, and averages the value to 400 ppm [49]. Given this method of auto-calibration, for the duration of this study, the Vitalight was always turned on outside, and remained outside for approximately 30 minutes during calibration. The Aranet was used as the reference monitor in this study, as it has been found to correlate well with reference monitors in the literature [15,50]. Additionally, given that this was a field study with a long data collection period, an Aranet was considered to be more robust to constant use in a dynamic environment relative to a standard reference sensor. To compensate for this, the Aranet was calibrated by the research team in the lab against an IAQ-CALC Model 7545 (TSI Incorporated, Shoreview MN, USA) over the course of the study period. The Aranet calibrates to a baseline value of 420 ppm.

Measurements

The Aranet and the Vitalight were both used to measure CO2 concentration (ppm), while the Aranet was also used to measure ambient temperature (°C), and relative humidity (%). Continuous measurement provides a more accurate assessment over time; however, while the Aranet can be used for continuous data logging, the Vitalight can only provide point-to-point measures. Therefore, a snapshot approach was used in this field study. Measurements were taken at two time points per trip, with the first measurement occurring approximately one-minute after boarding the bus (T1), and the second occurring approximately one-minute prior to disembarking (T2). Sampling took place at a height of ~0.5m (seated knee height). The date, time of day, route number, occupancy level, seating position, whether or not windows were open, number of windows open, and the number of times the bus stopped and opened the doors, were also recorded for each transit trip (Table 2).

thumbnail
Table 2. Variables measured using the ride survey, including the category and frequency of measurement per transit ride.

https://doi.org/10.1371/journal.pone.0342566.t002

Bus routes have two phases: 1) Outbound – covers when the bus leaves the terminal to when it reaches its half-way point and turns around to return; and 2) Inbound – covers when the bus turns around at the half-way point to when it returns to the terminal. Most of the 250 recorded trips were from stops at the outbound midpoint to its end, or from the start of the inbound route to its midpoint, so trips did not consist of measurements along the entire length of the route. The majority of measurements were performed during morning and afternoon commutes, and there were variable passenger loads.

Statistical analysis

Descriptive statistics were performed on all collected variables. A two-tailed paired samples t-test was used to assess differences in CO2 concentration between the two monitors. Pearson correlation coefficient was used to assess the relationships between the CO2 concentration measured across the two monitors, and between temperature, relative humidity, and number of bus stops. With respect to correlation strength, r2 values < 0.5 indicated weak agreement, r2 values between 0.5 and 0.7 indicated moderate agreement, and r2 values > 0.7 indicated strong agreement [51]. Two-tailed two sample (Mann-Whitney U test) and three sample (Kruskal-Wallis H test) non-parametric tests were used to assess differences in the concentration of CO2 by occupancy level, seating position, and window status. For significant Kruskal-Wallis H tests, post hoc comparisons were conducted using Mann-Whitney U tests with a Bonferroni correction. All tests were performed at an alpha of.05, with the adjusted alpha for the Bonferroni correction at.0167. All statistical analyses of the data were carried out using SPSS v.29.0.1.0 (IBM Corp: Armonk, New York).

Results

Descriptive statistics

Between October 17th, 2022 and July 18th, 2023 measurements were taken across 250 transit trips, with readings most commonly taken from the front driver’s side (DS; 35.2%), the front passenger’s side (PS; 23.2%), the middle PS (22.4%), and the middle DS (12.0%) of the bus. The trips were approximately evenly split between the morning (48.8%) and the afternoon (51.2%); were predominantly carried out on route 3 (89.6%); and generally had low (34.8%), medium-low (26.0%), or medium (21.2%) levels of occupancy. Windows were only open for 15.6% of trips with an average of 2.0 (±1.3) windows observed to be open on those trips. For the 250 trips, buses made an average of 7.3 (±2.4) stops where the doors opened, with a minimum of one stop, and a maximum of 13 stops made. See Tables 3 and 4 for the complete descriptive summary, and Fig 3 for a spatial representation of the inbound and outbound portions of route 3.

thumbnail
Table 3. Descriptive summary of observations taken on the 250 transit trips between October 2022 and July 2023.

https://doi.org/10.1371/journal.pone.0342566.t003

thumbnail
Table 4. Descriptive summary of measurements taken on the 250 transit trips between October 2022 and July 2023.

https://doi.org/10.1371/journal.pone.0342566.t004

thumbnail
Fig 3. Route 3 map.

Spatial map representing transit route 3, which represents 89.6% of the route data. The blue line, with the blue circle markers, represents the outbound portion of the route and its stops; the orange line, with the orange triangular markers, represents the inbound portion of the route and its stops. Spatial map was reprinted, and modified, from the original version from NASA Earth Observer [52]. The map data is licensed under the Open Data Commons Open Database License (ODbL) by OpenStreetMap and its contributors [53].

https://doi.org/10.1371/journal.pone.0342566.g003

Fig 4 shows the daily average CO2 concentrations, recorded by each monitor, across the study period. At T1, CO2 concentrations ranged between 400–1214 ppm ((Mean ± SD) 683.1 ± 185.9) for the Vitalight, and 482–1370 ppm (769.6 ± 189.1) for the Aranet; at T2, the range was 400–1273 ppm (644.1 ± 166.7) for the Vitalight, and 462–1386 ppm (735.5 ± 172.5) for the Aranet (Table 4). Temperatures ranged from 4.6–28.5°C (18.02 ± 5.00) at T1, and 7.5–33.9°C (19.66 ± 3.93) at T2, while relative humidity was 12–65% (32.3 ± 11.4) at T1, and 12–61% (30.2 ± 10.1) at T2 (Table 4).

thumbnail
Fig 4. Daily average CO2 concentrations.

Time-series plot of the daily average CO2 concentrations, recorded by the Vitalight and the Aranet, across the nine-month study period. Seasons are separated by dashed lines.

https://doi.org/10.1371/journal.pone.0342566.g004

For the purpose of further analyses, all route data were grouped, due to the predominance of trips on route 3, compared to routes 1, 14, and 29. The occupancy and seating position variables were also re-grouped to have three categories each. New categories for occupancy were set to: 1. Low occupancy = Very low and low; 2. Medium occupancy = Medium-low and medium; and 3. High occupancy = Medium-high, high, and very high. New categories for seating position were set to: 1. Front = Front DS and front PS; 2. Middle = Middle DS and middle PS; and 3. Back = Back DS, back centre, and back PS.

CO2 concentration measurements differed by type of CO2 monitor

Pearson correlations were performed to evaluate the relationship between the concentrations of CO2 measured by the Vitalight and the Aranet. There was a significant strong positive relationship between the measured concentrations of CO2 for the Vitalight and the Aranet, at T1 (r(248) =.95, p < .001) (Fig 5a), and T2 (r(248) =.97, p < .001) (Table 5; Fig 5b).

thumbnail
Table 5. Pearson correlations (r) between concentrations of CO2 measured with the Vitalight and the Aranet, and temperature, relative humidity, and number of stops with doors open, at time 1 and time 2. r2 values are in brackets.

https://doi.org/10.1371/journal.pone.0342566.t005

thumbnail
Fig 5. CO2 monitor correlation.

Using Pearson’s correlation, there was a strong positive relationship between CO2 concentration (ppm) as measured by the Vitalight and the Aranet at a) T1 and b) T2.

https://doi.org/10.1371/journal.pone.0342566.g005

While measures of CO2 concentration were strongly correlated for the two CO2 monitors, two-tailed paired samples t-tests were performed to evaluate whether the measured concentration of CO2 differed between the Aranet and the Vitalight. At T1, the Vitalight (M = 683.07, SD = 185.862) had a significantly lower CO2 measurement than the Aranet (M = 769.57, SD = 189.078), t(249) = −22.517, p < .001 (Fig 6a). The same trend was observed at T2, where the Vitalight (M = 644.09, SD = 166.656) had a significantly lower CO2 measurement than the Aranet (M = 735.49, SD = 172.525), t(249) = −32.436, p < .001 (Fig 6b).

thumbnail
Fig 6. CO2 monitor comparison.

CO2 concentration (ppm) as measured by the Aranet was significantly higher than the Vitalight at a) T1 and b) T2.

https://doi.org/10.1371/journal.pone.0342566.g006

CO2 concentration is negatively correlated to temperature at T1

Pearson correlations were performed to evaluate the relationship between the concentration of CO2 and temperature. There was a significant weak negative relationship between the concentration of CO2 and temperature at T1, as measured by the Vitalight (r(248) = −.37, p < .001), and the Aranet (r(248) = −.38, p < .001) (Table 5). The relationship between the concentration of CO2 and temperature at T2 was not significant for the Vitalight (r(248) = −.03, p = .606), or the Aranet (r(248) = −.02, p = .704) (Table 5).

CO2 concentration is positively correlated to relative humidity at T1 and T2

Pearson correlations were performed to evaluate the relationship between the concentration of CO2 and relative humidity. There was a significant weak positive relationship between the concentration of CO2 and relative humidity at T1 for the Vitalight (r(248) =.52, p < .001) and the Aranet (r(248) =.50, p < .001), and at T2 for the Vitalight (r(248) =.19, p = .003), and the Aranet (r(248) =.16, p = .012) (Table 5).

CO2 concentration is positively correlated to the number of times the bus stopped

Pearson correlations were performed to evaluate the relationship between the concentration of CO2 and the number of stops with doors open that the bus made. The relationship between the concentration of CO2 and the numbers of stops was only assessed at T2, as only at this time point had the maximum number of stops been made. There was a significant weak positive relationship between the concentration of CO2 and the number of stops the bus made for both the Vitalight (r(248) =.25, p < .001), and the Aranet (r(248) =.27, p < .001) (Table 5).

CO2 concentration differed by occupancy level at T2

Kruskal-Wallis tests were performed to evaluate whether the concentration of CO2 differed by occupancy level. The results indicated that there was a significant difference in the concentration of CO2 measured by the Vitalight at T2 at low, medium, and high occupancy, H(2) = 50.345, p < .001. Mann-Whitney U post hoc comparisons with a Bonferroni adjustment indicated that the concentration of CO2 measured by the Vitalight at T2 was significantly lower at low occupancy (Mdn = 540.5, IQR = 453.8–629.3) compared to medium occupancy (Mdn = 668.0, IQR = 577.8–765.8) (U = 2860.00, z = −6.22, p < .001), and at low occupancy (Mdn = 540.5, IQR = 453.8–629.3) compared to high occupancy (Mdn = 736.0, IQR = 657.0–831.0) (U = 676.00, z = −5.224, p < .001) (Fig 7a). The other comparison was not statistically significant after the Bonferroni adjustment (p > .0167). Similarly, there was a significant difference in the concentration of CO2 measured by the Aranet at T2 at low, medium, and high occupancy, H(2) = 46.570, p < .001. Mann-Whitney U post hoc comparisons with a Bonferroni adjustment indicate that the concentration of CO2 measured by the Aranet at T2 was significantly lower at low occupancy (Mdn = 620.0, IQR = 551.0–710.0) compared to medium occupancy (Mdn = 747.5, IQR = 659.3–850.8) (U = 2985.00, z = −5.95, p < .001), and at low occupancy (Mdn = 620.0, IQR = 551.0–710.0) compared to high occupancy (Mdn = 849.0, IQR = 708.0–947.0) (U = 712.50, z = −5.033, p < .001) (Fig 7b). The other comparison was not statistically significant after the Bonferroni adjustment (p > .0167). At T1, there was no significant difference in the concentration of CO2 measured by the Vitalight at low (Mdn = 678.0, IQR = 544.8–796.3), medium (Mdn = 665.0, IQR = 524.5–815.0), or high (Mdn = 638.0, IQR = 503.0–841.0) occupancy, H(2) = 0.163, p = .922, or by the Aranet at low (Mdn = 736.0, IQR = 607.0–888.8), medium (Mdn = 734.5, IQR = 604.5–916.3), or high (Mdn = 796.0, IQR = 613.0–910.0) occupancy, H(2) = 0.108, p = .947.

thumbnail
Fig 7. CO2 concentration by occupancy.

CO2 concentration (ppm) was significantly affected by occupancy level at T2, according to the Kruskal-Wallis H test. Mann-Whitney U post hoc tests showed CO2 was lower at low compared to medium and high occupancy, as measured by a) the Vitalight and b) the Aranet.

https://doi.org/10.1371/journal.pone.0342566.g007

CO2 concentration differed by seating position at T1

Kruskal-Wallis tests were performed to evaluate whether the concentration of CO2 differed by seating position. There was a significant difference in the concentration of CO2 measured by the Vitalight at T1 between the front, middle, and back seating positions, H(2) = 17.130, p < .001. Mann-Whitney U post hoc comparisons with a Bonferroni adjustment indicate that the concentration of CO2 was significantly lower in the front of the bus (Mdn = 622.0, IQR = 513.8–771.0) compared to the back of the bus (Mdn = 816.0, IQR = 686.0–1026.3) (U = 591.00, z = −3.804, p < .001), and in the middle of the bus (Mdn = 713.0, IQR = 569.3–866.3) compared to the back of the bus (Mdn = 816.0, IQR = 686.0–1026.3) (U = 491.00, z = −2.428, p = .015) (Fig 8a). The other comparison was not statistically significant after the Bonferroni adjustment (p > .0167). Likewise, there was a significant difference in the concentration of CO2 measured by the Aranet at T1 between the front, middle, and back seating positions, H(2) = 17.466, p < .001. Mann-Whitney U post hoc comparisons with a Bonferroni adjustment indicate that the concentration of CO2 was significantly lower in the front of the bus (Mdn = 689.5, IQR = 591.8–849.8) compared to the middle of the bus (Mdn = 796.5, IQR = 657.3–911.5) (U = 4988.50, z = −2.612, p = .009), and in the front of the bus (Mdn = 689.5, IQR = 591.8–849.8) compared to the back of the bus (Mdn = 940.0, IQR = 737.5–1143.8) (U = 617.50, z = −3.664, p < .001) (Fig 8b). The other comparison was not statistically significant after the Bonferroni adjustment (p > .0167). At T2, there was not a significant difference in the concentration of CO2 measured by the Vitalight between the front (Mdn = 612.0, IQR = 511.3–718.3), middle (Mdn = 652.5, IQR = 511.0–740.5), or back (Mdn = 699.0, IQR = 568.5–964.5) seating positions, H(2) = 4.330, p = .115, or by the Aranet between the front (Mdn = 692.5, IQR = 597.3–810.8), middle (Mdn = 723.0, IQR = 614.8–818.0), or back (Mdn = 772.5, IQR = 650.5–1006.3) seating positions, H(2) = 3.563, p = .168.

thumbnail
Fig 8. CO2 concentration by seating position.

CO2 concentration (ppm) was significantly affected by seating position at T1, according to the Kruskal-Wallis H test. Mann-Whitney U post hoc tests showed CO2 was higher at the back of the bus compared to the front and the middle, as measured by a) the Vitalight, and lower at the front of the bus compared to the middle and the back, as measure by b) the Aranet.

https://doi.org/10.1371/journal.pone.0342566.g008

CO2 concentration differed by window status at T1 and T2

Two-tailed Mann-Whitney U tests were performed to evaluate whether the concentration of CO2 differed when the windows on the bus were open or closed. At T1, the concentration of CO2 measured by the Vitalight was significantly lower when windows were open (Mdn = 509.0, IQR = 435.0–596.0) compared to closed (Mdn = 712.0, IQR = 562.0–819.0), U = 1710.50, z = −5.80, p < .001 (Fig 9a). Likewise, the concentration of CO2 measured by the Aranet was significantly lower when windows were open (Mdn = 594.0, IQR = 533.0–647.0) compared to closed (Mdn = 786.0, IQR = 660.0–920.0), U = 1710.50, z = −5.80, p < .001 (Fig 9b). The same trend was observed at T2, where the concentration of CO2 as measured by the Vitalight was significantly lower when windows were open (Mdn = 479.0, IQR = 419.0–586.0) compared to closed (Mdn = 657.0, IQR = 555.0–754.0), U = 1648.50, z = −5.94, p < .001 (Fig 9c). Similarly, the concentration of CO2 measured by the Aranet was significantly lower when windows were open (Mdn = 579.0, IQR = 523.0–655.0) compared to closed (Mdn = 730.0, IQR = 639.0–862.0), U = 1557.00, z = −6.17, p < .001 (Fig 9d).

thumbnail
Fig 9. CO2 concentration by window status.

CO2 concentration (ppm) was significantly higher when the bus windows were closed compared to open, as measured by a) the Vitalight, and b) the Aranet, at T1, and c) the Vitalight, and d) the Aranet, at T2, according to Mann-Whitney U tests.

https://doi.org/10.1371/journal.pone.0342566.g009

Discussion

Rebreathed air

Ongoing concerns related to airborne infectious illnesses have increased the focus on using CO2 monitors as a proxy for infection risk [10,11], as both CO2 and airborne contaminants can build up if there’s insufficient ventilation [4,24]. Public transit vehicles, can represent a health risk to drivers, and passengers, as contacts between people facilitate exposures to airborne biological contaminants [29].

In this work, CO2 measurements were collected across 250 transit trips, with an average trip duration of 15−20 minutes. The average concentration of CO2 measured with the Aranet was 769.6 ± 189.1 ppm (Vitalight: 683.1 ± 185.9) at T1, and 735.5 ± 172.5 (Vitalight: 644.1 ± 166.7) at T2. Carbon dioxide is predominantly produced by the occupants in indoor spaces, so its concentration is representative of the fraction of rebreathed air [1416]. Therefore, based on the Aranet values, on average the proportion of rebreathed air during these commutes was ~ 0.7–1.0% (Vitalight: ~ 0.5–0.7%) [22]. However, with measures of CO2 ranging between 482−1370 ppm for the Aranet (Vitalight: 400−1214) at T1, and 462−1386 (Vitalight: 400−1273) at T2, the proportion of rebreathed air could range from ~0%−2.6% (Vitalight: ~ 0–2.3%), such that up to every one in 39 breaths were considered rebreathed [22]. Combining the typical adult respiratory rate of 12−20 breaths per minute [54], with 1/39 rebreathed breaths at the documented CO2 concentrations indicated in a 15−20 minute trip, an occupant could rebreathe between 6.2 to 10.3 contaminated breaths. As concentrations of CO2 above 800 ppm increase SARS-CoV-2 aerostability [10], this study indicates that, at times, the measures of CO2 were high enough to suggest increased concern for transmissibility of this airborne hazard.

Are measures of CO2 concentration consistent between the Vitalight and the Aranet?

Low-cost CO2 monitors are recommended as a potential strategy for improving indoor ventilation [40]; however, information around how non-experts can use them for personal risk assessment is limited [25]. Therefore, the first objective of this paper was to compare the performance of a Vitalight CO2 monitor against an Aranet, as the Vitalight is more budget friendly to the general public, however, it lacks the published performance validation of the Aranet [15,50].

Ranges of CO2 concentration as measured by the Vitalight (400–1273 ppm) and the Aranet (462–1386 ppm) were consistent with values measured on urban transit vehicles in Brunswick and Hanover, Germany (Bus: 400–1161 ppm; Tram: 400–1197 ppm; Train: 400–1105 ppm) [21]; Cleveland, United States (Articulated bus peak CO2 at peak and non-peak occupancy: 760 and 790 ppm; Standard bus peak CO2 at peak and non-peak occupancy: 1600 and 920 ppm; Shuttle van peak CO2 at peak and non-peak occupancy: 1870 and 1360 ppm) [26], Zaragoza, Spain (Tram: 572 ± 75–835 ± 232 ppm) [11]; and Barcelona, Spain (Three types of buses: 555–2112 ppm) [36]. However, with respect to this study, compared to the Aranet, the Vitalight tended to underrepresent the CO2 concentration, though the monitor measures were highly correlated. These observations are consistent with the literature, where both Ishigaki and colleagues [15] and Demanega and colleagues [48] found that low-cost CO2 monitors that used NDIR sensors, correlated well with reference monitors, and could be used to indicate trends in CO2, if they were properly calibrated. Therefore, although the absolute values measured by the Vitalight and the Aranet are not exactly the same, the Vitalight is capable of dynamically tracking changes in CO2 concentration, in a trend consistent with the Aranet.

Factors related to the concentration of CO2 across monitors

In addition to determining whether CO2 measures were consistent between the Vitalight and the Aranet, we wanted to identify other factors that could affect CO2 concentration. Therefore, the second objective of this paper was to identify factors related to the concentration of CO2 on transit buses, and determine if those factors were consistent across monitors.

Factors that were previously found to affect the measurements of CO2 include: relative humidity and temperature [21,41,55,56]; open windows or doors [6,3436,57]; level of occupancy [6,11,21,35,37]; and seating position or room location [6,36,58,59]. In this study, these factors were assessed to determine if responses to them were consistent across monitors. For all of the factors assessed, the Vitalight paralleled the Aranet in response direction.

NDIR sensors can respond to environmental factors such as temperature and relative humidity, as they can change the properties of the gas molecules that the sensors assess [55,60]. In the current study, both the Vitalight and the Aranet showed a weak negative correlation with temperature at T1, indicating that at this time point as temperature increased, the measured concentration of CO2 decreased. At T2, no significant correlation with temperature was found with either monitor. For relative humidity, there was a weak positive correlation for both monitors, indicating that as relative humidity increased, the measured concentration of CO2 increased. The correlation was significant at both time points, however, the r values for both monitors decreased from T1 to T2. Consistent with the current study, a simulation study done to identify environmental factors that affect CO2 sensor accuracy and precision found that when temperature increases, the measure of CO2 decreases, and when relative humidity increases, the measure of CO2 increases [61]. These observations are also consistent with a study assessing field calibration methods, that found CO2 measures were negatively correlated with temperature (r = −.61) and positively correlated with relative humidity (r = .62) [56]. In the current study, higher correlations of temperature and relative humidity were observed at T1 compared to T2, which may indicate a holdover effect on the monitors in changing from outdoor ambient conditions to indoor conditions. In-line with this hypothesis, Ródenas García and colleagues [41] observed that a warm-up time may be required to ensure a sensor’s optical system is thermally stable in order to improve their accuracy. In general, field calibration, or calibration of the monitors to the conditions they are going to be used at, can help to improve their performance accuracy [62,63].

Ventilation on public transit vehicles favours natural ventilation [11], which includes opening the windows and doors. In the current study, for both the Vitalight and the Aranet, the concentration of CO2 was significantly lower when the bus windows were open, compared to closed, and this was observed at both T1 and T2. Consistent with these observations, opening the windows in buildings [6], and driving with the windows open in commercial trucks [64], school buses [65], tour buses [58], and public transit buses [34,36,57], has been shown to reduce the measured concentration of CO2. For example, an assessment of air change rates found that in buses with open windows, the number of air changes per hour was more than 2.5 times higher than in buses with closed windows [57]. Similarly, both Baselga and colleagues [11] and Zhu and colleagues [35] found that wind speed affects the rate of natural ventilation, with higher wind speeds leading to higher ventilation. However, as wind speed was not assessed in the current study, it cannot be determined whether this trend holds true with these data. In contrast to opening the windows, for both the Vitalight and the Aranet, there was a weak positive correlation with the number of times the doors on the bus opened, indicating that higher levels of CO2 tended to occur when the doors opened more. These observations are somewhat opposed to studies in the literature that show opening vehicle doors helped to reduce levels of CO2 [24,35,36]. This discrepancy may reflect that, in the current study, when the doors on the bus opened more, the bus occupancy did not necessarily change sufficiently to decrease CO2 concentrations. Additionally, when the bus stopped, both sets of doors did not always open, which could also have limited their role in air exchange during stops. It is possible that outdoor CO2 concentrations could have been elevated while the bus doors opened, but given the rural study location, this is unlikely to have affected the concentration gradient in short bursts.

In indoor spaces, CO2 is largely produced by the occupants of the space, so CO2 concentration can provide information around how crowded a space is [2,14]. On crowded buses, CO2 levels have been observed to be three times higher than ambient outdoor concentrations (i.e., 1200 ppm vs. 400 ppm) [37]. In the current study, for both the Vitalight and the Aranet, at T2 the concentration of CO2 was lower at low occupancy compared to both medium and high occupancies. These data are consistent with observations in other types of passenger vehicles, including cars [66], buses, such as shuttle buses, articulated buses, and standard buses [21,26,35,37], trams [11,21], and trains [21], as well as in buildings [6]. In all these cases, high levels of occupancy have been related to higher concentrations of CO2, as well as a reduced amount of time needed to reach high CO2 concentrations [66]. The concentration of CO2 did not differ across levels of occupancy for either monitor at T1, however, this is likely a limitation in the methodology. An estimate of average occupancy was only taken once, as immediately following boarding, both in the morning and the afternoon, there were large fluctuations in occupancy as there are multiple close placed stops at these points in the route, so accurate estimates were difficult to make. In contrast, closer to T2, stops tend to be spaced further apart so occupancy is more static. Therefore, the occupancy estimates were ultimately more reflective of occupancy at T2 rather than T1.

The distribution of gases, such as CO2, within the passenger space of transit vehicles is heterogeneous, and dependent on patterns of airflow within the space [59,67]. Observations with the Vitalight and the Aranet indicate that there were differences in the dispersion of CO2 along the length of the bus noted at T1. For the Vitalight, concentrations of CO2 were significantly higher at the back of the bus, relative to the middle section and the front section; for the Aranet, concentrations of CO2 were significantly lower in the front of the bus, relative to the middle section and the back sections. Though there was a slight discrepancy between the Vitalight and the Aranet, in general, measured concentrations were lowest at the front of the bus, and highest at the back of the bus. Similar dispersions have been noted in tour buses [58], and transit buses, as drivers tend to keep their windows open, contributing to lower CO2 concentrations in the front of the bus, while the absence of doors in the far back of the buses contributes to higher concentrations [36]. The concentration of CO2 did not differ across seating position for either monitor at T2. The different influence of seating position on concentrations of CO2 between T1 and T2 is not specifically known, but different levels of air flow at the two time points could contribute. For example, at T1 there are multiple closely placed stops, relative to T2, where stops tend to be spaced further apart leading to more uninterrupted driving time. Longer spans of uninterrupted driving time, could facilitate more airflow, through open windows, as well as through “leaks” in door gaps or poorly sealed windows. This could contribute to a more homogenous CO2 concentration throughout the length of the bus, in contrast to stop and go driving.

Usability of the Vitalight monitor

Building on the previous two objectives, the final objective of this study was to assess the usability of the Vitalight. Additionally, we provide some general recommendations around personal risk assessment using the Vitalight in combination with the evaluation of environmental factors.

Overall, the data in this study indicate that measurements taken with the Vitalight were highly correlated to those taken by the Aranet at both T1 and T2. Furthermore, the Vitalight paralleled the Aranet in measurement responses to: temperature, relative humidity, number of stops made, occupancy level, seating position, and whether windows were open or closed. The largest discrepancy between the two monitors was the absolute measured values of CO2, such that the Vitalight tended to underrepresent the concentration relative to the Aranet. Taken together, these data indicate that the Vitalight is capable of dynamically tracking changes in CO2 concentration, so increases and decreases in concentration can be evaluated in real-time.

Consistent with the observations in this study, Ishigaki and colleagues [15] assessed the accuracies of 12 low cost, and seven high cost, CO2 sensors, and they found that the NDIR sensors that correlated well with a reference sensor, could be calibrated and used to identify trends in CO2 concentration. Although multiple methods of calibration exist (e.g., [9,41,55,56,68]), given that the purpose of the current study is to outline usability of a low-cost CO2 monitor for non-experts, we have suggested recommendations that are simple, and useable by a wide audience, without a specific correction factor for the Vitalight.

Taken together, there are three main recommendations (Fig 10) around personal risk assessment that can be extracted from these results:

thumbnail
Fig 10. Recommendation summary.

Summary of the three recommendations for personal risk assessment using the Vitalight to assess ventilation.

https://doi.org/10.1371/journal.pone.0342566.g010

  1. To reduce error on readings when moving from outdoor ambient conditions to indoor conditions, the Vitalight should be allowed a brief “warm-up” time to allow it to equilibrate to the new temperature and humidity conditions. The holdover effect of external temperature and humidity was observable at T1, or approximately 1-minute post boarding, so a warm-up time of several minutes is recommended;
  2. The Vitalight can dynamically track changes in CO2 concentration, and large increases (e.g., 300–500 ppm) indicate that actions should be taken to improve ventilation. For example, if the initial measurement is 500 ppm, or approximately that of outdoor air, an increase between 300–500 ppm would put the measurement in the range of 800–1000 ppm which is high, indicating that ventilation should be improved;
  3. When CO2 readings are high (800 ppm or above [3,14,26]), or to keep them from getting high, transit riders can:
    • Open the windows on the bus;
    • Sit away from other passengers, or, if possible, take buses with lower levels of occupancy or at less busy times of day;
    • Sit closer to the front of the bus, preferentially within the front third.

Overall, these recommendations can help people to use CO2 readings from the Vitalight to estimate whether ventilation on transit buses is sufficient, as well as providing people with actions they can take to keep readings low.

Limitations

This field study is not intended to provide exhaustive information on the average concentration of CO2 on transit buses in Sudbury. Instead, it was meant to assess the real-world usability of a non-validated low-cost CO2 monitor for personal monitoring, and to determine additional environmental factors that could be controlled to reduce readings, so not all factors in the transit space were controlled. A small number of bus routes out of the entire network were evaluated, and measurements were taken during prime commuting times in the morning and the afternoon. Additionally, aside from whether the bus windows were open or closed, no data was collected about the ventilation settings on the buses, such as whether the air was being recirculated.

Air quality measures are not precise, and characteristics like temperature, humidity, airflow, the composition of contaminants in the air, etc., are fluid, and change over time [14,69]. Therefore, any measures for air quality are relative to the time and conditions when the measure was taken [69]. Additionally, information from CO2 monitors is dependent on the accuracy of the monitor used, the location of the monitor, and specific monitor calibration requirements [14,17]. For this data, a limitation was the snapshot approach, which was used due to the Vitalight monitor’s inability to continuously data log. Consequently, the data here reflect the specific CO2 monitors used, the conditions of the commuting trips taken, and the time points when the data was collected.

Ideally, CO2 monitors should be placed ~ 0.5m away from windows, doors, ventilation grilles, and people [3,14,28], however, on a bus this is not reasonably feasible. Placing CO2 monitors in individual’s breathing zones can increase readings relative to those that are further away from occupants [24], which may have impacted CO2 measurements taken with both monitors in this study.

Another limitation is that the Vitalight performs an auto-calibration where it takes the three lowest CO2 readings from its previous seven days of use, and averages the value to 400 ppm [49]. Given this method, the Vitalight monitor was always turned on outside, to ensure calibration to the outdoor ambient concentration of CO2. However, using 400 ppm as the baseline concentration can be inaccurate, because, even outside, the concentration of CO2 varies and can be higher than 400 ppm. For example, average CO2 concentrations measured outside of four schools in Ottawa, Ontario, in the fall of 2013, ranged between 389 ppm (95% CI 374–403 ppm) and 842 ppm (95% CI 386–1298 ppm) [70]. Calibration to the incorrect baseline, will affect the reliability of all measured data, which could limit the usability of the Vitalight in locales where the ambient CO2 concentration is significantly different from 400 ppm.

While the above factors may have limited the accuracy of the collected data, the paralleled responses of two different CO2 monitors across field conditions demonstrate that low-cost CO2 monitors can dynamically assess changes in CO2, indicating that they can be used as a component of a personal monitoring strategy regarding the sufficiency of ventilation in transit vehicles.

Conclusions

Measurements of CO2 and surveying of environmental factors were carried out with two CO2 monitors during commutes on 250 public transit trips, with the aim of assessing the usability of a low-cost CO2 monitor, and providing general recommendations around personal risk assessment using a low-cost monitor in combination with the evaluation of environmental factors. The results show that CO2 measurements taken with the Vitalight were highly correlated with those taken by the Aranet, and the Vitalight paralleled the Aranet with respect to its responses to: temperature; relative humidity; occupancy level; seating position; how often the bus doors opened; and whether the windows were open or closed. However, measurements of CO2 taken by the Vitalight, on average, underrepresented measures taken by the Aranet. Based on these observations, a list of recommendations and actions related to the use of the Vitalight for personal risk assessment was developed.

Future research should look at evaluating additional Vitalight monitors to assess the precision of the CO2 sensors across individual devices. Additionally, as the purpose of this study was to provide recommendations to non-experts to facilitate their use of low-cost monitors in personal risk assessment, future studies should assess the usability of these recommendations by non-experts, and whether there are associated behaviour changes, such as increased confidence and self-efficacy, with their application in personal monitoring.

Supporting information

S2 Table. Extreme Trip Data.

Subset of data for the trips with peak CO2 and occupancy levels.

https://doi.org/10.1371/journal.pone.0342566.s002

(DOCX)

Acknowledgments

We would like to thank the students and staff at the Centre for Research in Occupational Safety and Health for their support and feedback on this project.

References

  1. 1. Tran VV, Park D, Lee Y-C. Indoor air pollution, related human diseases, and recent trends in the control and improvement of indoor air quality. Int J Environ Res Public Health. 2020;17(8):2927. pmid:32340311
  2. 2. Kakoulli C, Kyriacou A, Michaelides MP. A review of field measurement studies on thermal comfort, indoor air quality and virus risk. Atmosphere. 2022;13(2):191.
  3. 3. SAGE, EMG. Role of ventilation in controlling SARS-CoV-2 transmission. United Kingdom: Scientific Advisory Group for Emergencies [SAGE]; 2020 pp. 32. Available from: https://assets.publishing.service.gov.uk/government/uploads/system/uploads/attachment_data/file/928720/S0789_EMG_Role_of_Ventilation_in_Controlling_SARS-CoV-2_Transmission.pdf
  4. 4. Almeida SM, Martins V. Exposure to air pollutants in ground transport microenvironments. Handbook of Indoor Air Quality. Singapore: Springer; 2022. pp. 2023–2055. https://link.springer.com/referenceworkentry/10.1007/978-981-16-7680-2_54#DOI
  5. 5. Braggion A, Dugerdil A, Wilson O, Hovagemyan F, Flahault A. Indoor air quality and COVID-19: a scoping review. Public Health Rev. 2024;44:1605803. pmid:38273885
  6. 6. Burridge HC, Fan S, Jones RL, Noakes CJ, Linden PF. Predictive and retrospective modelling of airborne infection risk using monitored carbon dioxide. Indoor Built Environ. 2021;31(5):1363–80.
  7. 7. Nair AN, Anand P, George A, Mondal N. A review of strategies and their effectiveness in reducing indoor airborne transmission and improving indoor air quality. Environ Res. 2022;213:113579. pmid:35714688
  8. 8. Wolkoff P, Azuma K, Carrer P. Health, work performance, and risk of infection in office-like environments: The role of indoor temperature, air humidity, and ventilation. Int J Hyg Environ Health. 2021;233:113709. pmid:33601136
  9. 9. Peng Z, Jimenez JL. Exhaled CO2 as a COVID-19 infection risk proxy for different indoor environments and activities. Environ Sci Technol Lett. 2021;8(5):392–7. pmid:37566374
  10. 10. Haddrell A, Oswin H, Otero-Fernandez M, Robinson JF, Cogan T, Alexander R, et al. Ambient carbon dioxide concentration correlates with SARS-CoV-2 aerostability and infection risk. Nat Commun. 2024;15(1):3487. pmid:38664424
  11. 11. Baselga M, Alba JJ, Schuhmacher AJ. The control of metabolic CO2 in public transport as a strategy to reduce the transmission of respiratory infectious diseases. Int J Environ Res Public Health. 2022;19(11):6605. pmid:35682191
  12. 12. Carrazana E, Ruiz-Gil T, Fujiyoshi S, Tanaka D, Noda J, Maruyama F, et al. Potential airborne human pathogens: A relevant inhabitant in built environments but not considered in indoor air quality standards. Sci Total Environ. 2023;901:165879. pmid:37517716
  13. 13. ASHRAE. Ventilation and Acceptable Indoor Air Quality. Peachtree Corners, Georgia, United States: ASHRAE; 2022. Available from: www.ashrae.org
  14. 14. EMG, SPI-B. Application of CO2 monitoring as an approach to managing ventilation to mitigate SARS-CoV-2 transmission. United Kingdom: Scientific Advisory Group for Emergencies [SAGE]; 2021 pp. 20. Available from: https://assets.publishing.service.gov.uk/government/uploads/system/uploads/attachment_data/file/992966/S1256_EMG_SPI-B_Application_of_CO2_monitoring_as_an_approach_to_managing_ventilation_to_mitigate_SARS-CoV-2_transmission.pdf
  15. 15. Ishigaki Y, Enoki K, Yokogawa S. Accuracy Verification of Low-Cost Co2 Sensors Used for Covid-19 Risk Mitigation. Rochester, NY; 2023.
  16. 16. Rudnick SN, Milton DK. Risk of indoor airborne infection transmission estimated from carbon dioxide concentration. Indoor Air. 2003;13(3):237–45. pmid:12950586
  17. 17. ASHRAE. ASHRAE Position Document on Indoor Carbon Dioxide. ASHRAE; 2022. Available from: https://www.ashrae.org/file%20library/about/position%20documents/pd_indoorcarbondioxide_2022.pdf
  18. 18. Lv M, Liu S, Cao Q, Zhang T, Liu J. Influence of ventilation on indoor air quality. Handbook of Indoor Air Quality. Singapore: Springer Nature Singapore; 2022. pp. 1637–73.
  19. 19. Persily A, de Jonge L. Carbon dioxide generation rates for building occupants. Indoor Air. 2017;27(5):868–79. pmid:28321932
  20. 20. ASHRAE. Control of Infectious Aerosols. Peachtree Corners, Georgia, United States: ASHRAE; 2023. Available from: www.ashrae.org
  21. 21. Salthammer T, Fauck C, Omelan A, Wientzek S, Uhde E. Time and spatially resolved tracking of the air quality in local public transport. Sci Rep. 2022;12(1):3262. pmid:35228615
  22. 22. Elfstrom D. Rebreathed fraction of air from CO2. In: Google Sheets [Internet]. 2020 [cited 26 Jun 2024]. Available: https://docs.google.com/spreadsheets/d/1AjFzhqM_NILYvZjgE8n0CvGZzYh04JpF_DO0phrOcFw/edit?pli=1&gid=0#gid=0
  23. 23. Rodríguez-Vidal I, Martín-Garín A, González-Quintial F, Rico-Martínez JM, Hernández-Minguillón RJ, Otaegi J. Response to the COVID-19 pandemic in classrooms at the university of the basque country through a user-informed natural ventilation demonstrator. Int J Environ Res Public Health. 2022;19(21):14560. pmid:36361439
  24. 24. Gangrade S, Vamsi B, Prasannaa , Saran R, Dhariwal J. Empirical Studies Assessing the CO2 Levels in Indoor Spaces. In: Chakrabarti A, Singh V, editors. Design in the Era of Industry 40, Volume 1. Singapore: Springer Nature; 2023. pp. 197–206.
  25. 25. Cadnum JL, Alhmidi H, Donskey CJ. Planes, trains, and automobiles: use of carbon dioxide monitoring to assess ventilation during travel. Pathog Immun. 2022;7(1):31–40. pmid:35316971
  26. 26. Greentree DH, Wilson BM, Donskey CJ. Carbon dioxide monitoring demonstrates variations in the quality of ventilation on public transportation buses and university student shuttle vans and identifies effective interventions. Pathog Immun. 2023;8(1):148–60. pmid:38035133
  27. 27. Li Y, Cheng P, Jia W. Poor ventilation worsens short-range airborne transmission of respiratory infection. Indoor Air. 2022;32(1):e12946. pmid:34704625
  28. 28. Eykelbosh A. Indoor CO2 sensors for COVID-19 risk mitigation: Current guidance and limitations. Vancouver, British Columbia: National Collaborating Centre for Environmental Health; 2021 pp. 16. Available from: https://ncceh.ca/sites/default/files/FINAL%20-%20Using%20Indoor%20CO2%20Sensors%20for%20COVID%20MAY%2017%202021.pdf
  29. 29. Nowakowicz-Dębek B, Pawlak H, Wlazło Ł, Maksym P, Kapica J, Chmielowiec-Korzeniowska A, et al. Evaluating bioaerosol exposure among bus drivers in the public transport sector. J Occup Environ Hyg. 2017;14(11):D169–72. pmid:28650742
  30. 30. Statistics Canada. Table 23-10-0251-01 - Passenger bus and urban transit statistics, by the North American Industry Classification System (NAICS). In: Statistics Canada [Internet]. 2024 [cited 24 Jun 2024]. Available from: https://www150.statcan.gc.ca/t1/tbl1/en/cv.action?pid=2310025101
  31. 31. Matz CJ, Stieb DM, Davis K, Egyed M, Rose A, Chou B, et al. Effects of age, season, gender and urban-rural status on time-activity: CanadianHuman Activity Pattern Survey 2 (CHAPS 2). Int J Environ Res Public Health. 2014;11(2):2108–24. pmid:24557523
  32. 32. Matz CJ, Stieb DM, Egyed M, Brion O, Johnson M. Evaluation of daily time spent in transportation and traffic-influenced microenvironments by urban Canadians. Air Qual Atmos Health. 2018;11(2):209–20. pmid:29568337
  33. 33. Van Ryswyk K, Evans GJ, Kulka R, Sun L, Sabaliauskas K, Rouleau M, et al. Personal exposures to traffic-related air pollution in three Canadian bus transit systems: the Urban Transportation Exposure Study. J Expo Sci Environ Epidemiol. 2021;31(4):628–40. pmid:32678304
  34. 34. Corzo SF, Godino DM, Ramajo DE. Air circulation study inside and outside of urban buses induced by the opening of windows. Environ Sci Pollut Res Int. 2023;30(8):20821–32. pmid:36260226
  35. 35. Zhu S, Demokritou P, Spengler J. Experimental and numerical investigation of micro-environmental conditions in public transportation buses. Build Environ. 2010;45(10):2077–88. pmid:32288006
  36. 36. Querol X, Alastuey A, Moreno N, Minguillón MC, Moreno T, Karanasiou A, et al. How can ventilation be improved on public transportation buses? Insights from CO2 measurements. Environ Res. 2022;205:112451. pmid:34848209
  37. 37. Moreno T, Reche C, Rivas I, Cruz Minguillón M, Martins V, Vargas C, et al. Urban air quality comparison for bus, tram, subway and pedestrian commutes in Barcelona. Environ Res. 2015;142:495–510. pmid:26277386
  38. 38. Government of Canada. Respiratory infectious diseases: How to reduce the spread with personal protective measures. In: Canada.ca [Internet]. 2024 [cited 27 May 2024]. Available from: https://www.canada.ca/en/public-health/services/diseases/respiratory-infectious-diseases-reduce-spread-personal-protective-measures.html
  39. 39. Government of Ontario. Protection from COVID‑19 and other respiratory illnesses | ontario.ca. In: ontario.ca [Internet]. 2024 [cited 27 May 2024]. Available from: http://www.ontario.ca/page/protection-covid-19-and-other-respiratory-illnesses
  40. 40. Government of Canada. COVID-19: Improving indoor ventilation. 2022 [cited 5 June 2024]. Available from: https://www.canada.ca/en/public-health/services/diseases/2019-novel-coronavirus-infection/prevention-risks/covid-19-improving-indoor-ventilation.html
  41. 41. Ródenas García M, Spinazzé A, Branco PTBS, Borghi F, Villena G, Cattaneo A, et al. Review of low-cost sensors for indoor air quality: Features and applications. Appl Spectrosc Rev. 2022;57(9–10):747–79.
  42. 42. Rai AC, Kumar P, Pilla F, Skouloudis AN, Di Sabatino S, Ratti C, et al. End-user perspective of low-cost sensors for outdoor air pollution monitoring. Sci Total Environ. 2017;607–608:691–705. pmid:28709103
  43. 43. Nieuwenhuijsen MJ, Donaire-Gonzalez D, Rivas I, de Castro M, Cirach M, Hoek G, et al. Variability in and agreement between modeled and personal continuously measured black carbon levels using novel smartphone and sensor technologies. Environ Sci Technol. 2015;49(5):2977–82. pmid:25621420
  44. 44. Public Health Ontario. Heating, Ventilation, and Air Conditioning (HVAC) Systems in Buildings and COVID-19. Toronto, Ontario: King’s Printer; 2023 pp. 20. Available from: https://www.publichealthontario.ca/-/media/documents/ncov/ipac/2020/09/covid-19-hvac-systems-in-buildings.pdf?la=en
  45. 45. City of Greater Sudbury. GOVA Routes and Schedules. In: Greater Sudbury [Internet]. 2023 [cited 1 Aug 2023]. Available from: https://www.greatersudbury.ca/live/transit/gova-routes-and-schedules/
  46. 46. City of Greater Sudbury. TransitStopsGTFS. In: City of Greater Sudbury Open Data Portal [Internet]. 2023 [cited 1 Aug 2023]. Available from: https://opendata-sudbury.opendata.arcgis.com/datasets/Sudbury::transitstopsgtfs-view/explore?location=46.466383%2C-80.856429%2C12.79
  47. 47. City of Greater Sudbury. 2022 Annual Report: For the year ending December 31, 2022. City of Greater Sudbury; 2023. Available from: https://www.greatersudbury.ca/city-hall/reports-studies-policies-and-plans/pdfs/annual-report-2022/
  48. 48. Demanega I, Mujan I, Singer BC, Anđelković AS, Babich F, Licina D. Performance assessment of low-cost environmental monitors and single sensors under variable indoor air quality and thermal conditions. Build Environ. 2021;187:107415.
  49. 49. Brooke E. Vitalight Mini CO2 Detector Review – An Affordable Alternative? In: BreatheSafeAir: Air Quality Monitor Reviews [Internet]. 2022 [cited 16 May 2023]. Available from: https://breathesafeair.com/vitalight-mini-co2-detector-review/
  50. 50. Villanueva F, Jiménez E, Felisi JM, Garrido T, Jimenez JL, Ródenas M, et al. Guide about affordable CO2 detectors for COVID-19 Prevention, version 1.6. Aireamos; 2021. Available from: https://bit.ly/monitorsCO2
  51. 51. Moore DS, Notz WI, Flinger MA. The Basic Practice of Statistics. New York: WH Freeman; 2013.
  52. 52. NASA. EO Explorer. In: NASA Earth Observer [Internet]. 2025 [cited 17 Dec 2025]. Available from: https://science.nasa.gov/earth/earth-observatory/explorer/
  53. 53. OpenStreetMap Contributors. Copyright and License. In: OpenStreetMap [Internet]. [cited 17 Dec 2025]. Available from: https://www.openstreetmap.org/copyright
  54. 54. Whited L, Hashmi MF, Graham DD. Abnormal Respirations. Treasure Island (FL): StatPearls Publishing. 2025.
  55. 55. Marinov MB, Djermanova N, Ganev B, Nikolov G, Janchevska E. Performance evaluation of low-cost carbon dioxide sensors. Proc XXVII International Scientific Conference Electronics. Sozopol, Bulgaria: IEEE; 2018. pp. 1–4.
  56. 56. Spinelle L, Gerboles M, Villani MG, Aleixandre M, Bonavitacola F. Field calibration of a cluster of low-cost commercially available sensors for air quality monitoring. Part B: NO, CO and CO2. Sens Actuators B: Chem. 2017;238:706–15.
  57. 57. Feng Y, Zhang Y, Ding X, Fan Y, Ge J. Multi-scale risk assessment and mitigations comparison for COVID-19 in urban public transport: A combined field measurement and modeling approach. Build Environ. 2023;242:110489. pmid:37333517
  58. 58. Chiu C-F, Chen M-H, Chang F-H. Carbon dioxide concentrations and temperatures within tour buses under real-time traffic conditions. PLoS One. 2015;10(4):e0125117. pmid:25923722
  59. 59. Woodward H, Fan S, Bhagat RK, Dadonau M, Wykes MD, Martin E, et al. Air flow experiments on a train carriage—towards understanding the risk of airborne transmission. Atmosphere. 2021;12(10):1267.
  60. 60. Vafaei M, Amini A. Chamberless NDIR CO2 sensor robust against environmental fluctuations. ACS Sens. 2021;6(4):1536–42. pmid:33784085
  61. 61. Singh OP, Malarvili M. Effect of temperature, pressure and humidity on carbon dioxide concentration—simulation study. IOP Conf Ser: Mater Sci Eng. 2020;884(1):012007.
  62. 62. Petersen J, Kristensen J, Elarga H, Andersen RK, Midtstraum A. Accuracy and Air Temperature Dependency of Commercial Low-cost NDIR CO2 Sensors: An Experimental Investigation. Proceedings of COBEE2018. Melbourne, Australia; 2018. pp. 203–207. Available from: https://orbit.dtu.dk/en/publications/accuracy-and-air-temperature-dependency-of-commercial-low-cost-nd
  63. 63. Shrestha SS, Maxwell GM. An experimental evaluation of HVAC-grade carbon-dioxide sensors--Part 3: humidity, temperature, and pressure sensitivity test results. ASHRAE Transac. 2010;116:271–84.
  64. 64. Cheng X, Tan Z, Wang X, Tay R. Air quality in a commercial truck cabin. Transp Res Part D: Transp Environ. 2006;11(6):389–95.
  65. 65. Van Dyke M, King B, Esswein E, Adgate J, Dally M, Kosnett M. Investigating dilution ventilation control strategies in a modern U.S. school bus in the context of the COVID-19 pandemic. J Occup Environ Hyg. 2022;19(5):271–80. pmid:35311626
  66. 66. Angelova RA, Markov DG, Simova I, Velichkova R, Stankov P. Accumulation of metabolic carbon dioxide (CO2) in a vehicle cabin. IOP Conf Ser: Mater Sci Eng. 2019;664(1):012010.
  67. 67. Yoo S-J, Kurokawa A, Matsunaga K, Ito K. Spatial distributions of airborne transmission risk on commuter buses: Numerical case study using computational fluid and particle dynamics with computer-simulated persons. Exp Comput Multiph Flow. 2023;5(3):304–18. pmid:36788805
  68. 68. Kang Y, Aye L, Ngo TD, Zhou J. Performance evaluation of low-cost air quality sensors: A review. Sci Total Environ. 2022;818:151769. pmid:34801495
  69. 69. Edwards NJ, Colder B, Sullivan J, Naramore L. A practical approach to indoor air quality for municipal public health and safety. Open J Polit Sci. 2021;11(01):176–91.
  70. 70. MacNeill M, Dobbin N, St-Jean M, Wallace L, Marro L, Shin T, et al. Can changing the timing of outdoor air intake reduce indoor concentrations of traffic-related pollutants in schools? Indoor Air. 2016;26(5):687–701. pmid:26340686