Abstract
Mechanical heart valves (MHVs) remain the most durable option for valve replacement, but they are highly thrombogenic and therefore necessitate life-long anticoagulation. We designed a novel MHV assembly based on “Surface Anticoagulation by Electrically induced Biomimetic Glycocalyx” (SAEBG) that imposes a weak negative potential across the valve’s blood-contacting surfaces using an implantable pacemaker as an energy source. The concept is inspired by the native vascular endothelium, where the luminal glycocalyx carries a net negative surface charge that repels platelets and proteins. The present study aims to assess hemocompatibility and thromboresistance in our electrically activated MHV. Bileaflet MHVs (n = 9) were immersed in human platelet-rich plasma or whole blood, of which 5 valves were connected to a programmable pulse generator. Control valves (0 V) (n = 4) were compared to those activated at 0.25 (n = 3), and 0.5 V(n = 2). The control/activated valves were immersed in PRP/blood for 30 min under gentle agitation at 35 ± 2 °C. Hemolysis and blood cell integrity in the supernatant fluid were quantified using standard hematology analyses. Valves immersed in PRP were assessed by scanning electron microscopy (SEM). The cell/deposit free area as assessed by SEM was highest with 0.5 V (~96%), while the corresponding free areas for controls and 0.25 V were ~86% and ~58%, respectively. The platelet counts, platelet and coagulation markers were close to the measurement uncertainty ranges for all the experiments. Hemolysis was < 0.1% for all conditions, and the leucocyte and red-cell counts changed by <2%. Imposing a mild electrical potential (~0.5 V) on an MHV reproduces the anti-thrombotic behaviour of vascular endothelium. The electrical field did not injure blood cells or activate coagulation pathways, establishing a safe and effective voltage window for the SAEBG valve to enable further swine model experiments.
Citation: Sajja LR, Koppula A, Mathew T, Bhatt A (2026) Surface anticoagulation of mechanical heart valves using electrically induced biomimetic glycocalyx: An in-vitro study to assess hemocompatibility and optimal voltage. PLoS One 21(5): e0336760. https://doi.org/10.1371/journal.pone.0336760
Editor: Jeevithan Elango, UCAM: Universidad Catolica San Antonio de Murcia, CHINA
Received: November 7, 2025; Accepted: April 8, 2026; Published: May 6, 2026
Copyright: © 2026 Sajja et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Data Availability: All the data all the fully available without restriction. The raw hematological and coagulation studies for each assessment are shown in Tables 2-5. All the SEM images used in the manuscript have been uploaded as a Supplementary information ZIP file. Information on the US patent granted to valve assembly (including the design details) on which this work is based is available in the public domain (https://patents.google.com/patent/US10736726B2/en?oq=US10736726B2).
Funding: The author(s) received no specific funding for this work.
Competing interests: NO authors have competing interests.
Introduction
Heart valve prostheses were reported to be implanted in more than ~280000 patients worldwide in the year 2009, and about 50% of them were mechanical heart valves (MHV) [1]. Since then, there has been only a marginal increase in the usage of MHV due to a preference for bioprosthetic valves. Nevertheless, MHVs are critically important for younger patients because of their durability and superior mechanical performance [2]. Although they are mechanically durable, their synthetic surfaces trigger platelet adhesion and coagulation, so patients must take long-term anticoagulants such as warfarin [3]. The need for life-long anticoagulation increases the risk of bleeding and interferes with activities of daily living; suboptimal therapy carries a risk of thromboembolism, while over-anticoagulation (higher prothrombin time and international normalized ratio) leads to hemorrhage [4]. Approaches to improve hemocompatibility, such as pyrolytic carbon surfaces or heparin and hydrophilic polymer coatings, reduce but do not abolish thrombogenicity, and coatings may degrade over time [5–7]. A breakthrough would be a durable MHV that resists thrombus formation without systemic anticoagulation.
The vascular endothelium naturally maintains blood fluidity through a carbohydrate-rich glycocalyx on its luminal surface. This layer contains negatively charged glycosaminoglycans, including hyaluronic acid, chondroitin sulphate, and heparan sulphate. The negative surface charge of endothelial cells arises from the glycocalyx and specialised membrane lipids; this electrostatic barrier repels platelets and limits protein adsorption, contributing to the anti-adhesive nature of the endothelium. Several reviews have elaborated on the structure and functions of glycocalyx [8–11]. Artificial surfaces, by contrast, typically become coated with plasma proteins that trigger platelet adhesion and activation [12]. Sawyer and colleagues demonstrated that cathodically polarized vascular grafts resisted thrombosis, suggesting that an imposed electrical field could restore hemocompatibility [13–17]. Building on this concept, we designed an assembly of a bileaflet MHV made of pyrolytic carbon with a pacemaker lead connected to the housing of the valve, which in turn is connected to the pulse generator that delivers a programmable potential to the blood-contacting surfaces.
The present study was conducted to evaluate the hemocompatibility and thromboresistance of the electrically activated mechanical heart valve. For a proof of concept, we explore the antithrombotic effect at two specific voltages, i.e., 0.25 and 0.5 V. These voltages were chosen for the following reasons: 1. The normal vascular endothelium in humans is antithrombotic at a potential of 0.3 V due to negatively charged residues in glycocalyx. The choice of 0.25 & 0.5 V is motivated in part by the proximity of these values to the physiological endothelial potential in health. 2. The present work is based on the assembly of a prosthetic mechanical heart valve and the commercially available pacemakers (with leads). The lowest voltage of the commonly used pacemakers is 0.25 V (St. Jude pacemaker, now Abbott) or 0.5 V (Medtronic). 3. Higher voltage at the valves is likely to increase the risk of myocardial capture (inadvertent action potential genesis in myocardium due to pacemaker activation of the valve) as the threshold pacing voltage is approached, potentially affecting rhythm genesis or propagation. 4. Higher voltage also increases the rate of battery discharge, shortening the duration of antithrombotic efficacy when used in a clinical context, requiring frequent pulse generator replacement. This concept of mechanical prosthetic heart valve assembly was granted a United States of America Patent (US10736726B2) in the year 2020 [18].
Materials and methods
The in vitro studies were conducted at the Division of Thrombosis Research, Biomedical Technology wing, Sree Chitra Tirunal Institute for Medical Sciences and Technology, Thiruvanthapuram, India, between 14/02/2017 and 29/06/2019.
Prototype valve assembly
The SAEBG prototype consisted of a standard 25 mm bileaflet mechanical heart valve (Standard St. Jude Medical mechanical heart valve, St. Paul, Minnesota, USA) with pyrolytic carbon discs and a titanium housing. A Cobalt-Chromium-Molybdenum (CoCrMo) alloy electrode (St. Jude Medical Tendril Pacing Lead, Sylmar, California, USA) was welded to the valve housing through a fenestration made in the sewing ring of valve and the lead was covered with a PTFE tube graft, which was sutured to the margins of the fenestration of the sewing ring, the lead was connected to an external pacemaker-derived generator (St. Jude Medical ENDURITY CORE, Sylmar, California, USA). The prosthetic valve-lead-pacemaker circuitry is shown in Fig 1 (A and B). The generator delivered monophasic square pulses at pre-programmed amplitudes (0.25 V and 0.5 V @ 0.4 ms pulse duration) and a frequency of 70 pulses min ⁻ ¹. Control heart valves were identical, but not connected to the lead.
A-Electrical connectivity of the prosthetic valve to the pacemaker with a lead electrode. B-Expanded view of prosthetic valve-lead junction. C-Prosthetic valve with lead was placed in a 50 ml polystyrene test tube for exposure to blood/PRP, while the pacemaker was placed outside the test tube, in a plastic container (shown in D). D-Experimental session with blood/PRP exposure to the valve in an environ shaker and an incubator at 35 ± 2 °C for 30 min.**.
Blood collection and preparation
Fresh human whole blood was obtained from healthy volunteers under institutional approval (First approval: SCT/IEC/594/April-2014, Re-approval: SCT/IEC/1366/April-2019) and anticoagulated with citrate phosphate dextrose. Platelet-rich plasma (PRP) was prepared by centrifugation at 2500 rpm for 5 min; platelet-poor plasma (PPP) was obtained by centrifugation at 4000 rpm for 15 min. Whole blood for hemolysis assays was stored at room temperature and processed within one hour.
In-vitro exposure and voltage titration
All the in vitro testing procedures conformed to the International Organization for Standardization (ISO) guidelines laid down in the document on Biological evaluation of medical devices – Part 4: Selection of tests for interactions with blood (ISO 10993–4:2017) [19]. The in vitro experiments exposed the test/control valves to blood products for 30 minutes, and the supernatant blood/plasma was tested for cell counts and coagulation markers at two time points. The first (initial) sample is the supernatant fluid collected immediately following the first contact of the valve with blood/plasma, and a second (final) sample was aliquoted 30 minutes after the immersion. The three in vitro experiments were done on different days as follows: 1) Uncharged control valve (V = 0) with PRP, 2) Two test valves (one each at 0.25 and 0.5 V) and one uncharged control valve with PRP, 3) Three test valves (two at 0.25 V and one at 0.5 V) and 2 uncharged control valves with whole blood. Valve discs were placed in 50-ml polystyrene test tubes containing 5 ml phosphate-buffered saline and gently agitated for 5 min to remove surface contaminants. After decanting the saline, 20 ml of PRP (or 40 ml of whole blood for hemolysis assays) was added to the tubes containing the valve discs. Around 5mL of PRP (or 10 mL of blood) was aspirated immediately for the initial analysis (initial sample). The remaining 15 ml PRP (or 30 ml blood) was agitated at 70 ± 5 rpm with an Environ shaker in an incubator at 35 ± 2 °C for 30 min, after which the PRP (or blood) was aspirated and analyzed (final sample). Empty polystyrene tubes filled with PRP (or blood) were used as a reference, and measurements were made from the reference samples at similar time points, i.e., initial and 30 minutes.
Hematologic and coagulation assays
The initial and final Blood/PRP samples were analyzed for cell counts and coagulation profile before (initial) and after exposure (30 minutes) and the percentage change (initial to 30-minutes) was recorded for test, control, and reference samples. Uncertainty of measurement was reported in percentage (%) units to assess the significance of the experiment-associated changes in the variables.
The Platelet/ Leucocyte/RBC counts, Hematocrit (Hct), and hemoglobin concentration were assessed by Sysmex-K 4500 automated hematology analyzer. The free hemoglobin in the plasma was measured using a Diode array Spectrophotometer. Using the hemoglobin (total Hb), free plasma hemoglobin (free Hb), and Hematocrit (Hct) determined from the methods outlined above, percentage hemolysis was calculated using the following Equation 1 [20]:
Platelet activation was assessed using activation markers like P-selectin and platelet factor 4 (PF4) for PRP and blood samples, respectively. For P-selectin, samples were diluted with ACD-PBS to get a final Platelet count of 150−300 × 104 ml-1. Fluorescent-tagged CD26P antibody was added to the sample, incubated for 1 hour and analyzed by flow cytometry using Epics XL, Beckman Coulter. The PF4 was analyzed with commercial ELISA kits (Asserachrom PF4, Diagnostica Stago, France). Absorbance was measured, and ELISA data were processed using the iMark microplate reader and MPM6 software, respectively.
Partial thromboplastin time (PTT) and Fibrinogen levels was measured in the PRP/blood samples using Diagnostica Stago reagent kits on STart 4 Hemostasis Analyzer. Activated complement (C3a) was measured in the samples using commercial ELISA kits (Quidel, USA). Absorbance was measured using BioRad iMark ELISA microplate reader.
Scanning electron microscopic imaging (SEM)
One PRP-exposed valve for each voltage condition (control, 0.25 V, and 0.5 V) was subjected to SEM imaging. The valves (test and control) were rinsed in buffered saline, fixed in 2% glutaraldehyde, and dehydrated in graded ethanol. The valves were later sputter-coated with gold and subjected to scanning electron microscopy (SEM) at magnifications of up to 5000 × . Multiple images were acquired for each valve. The SEM images were analyzed using the Fiji distribution of ImageJ software Version 1.54p. Calibration was performed using a linetool and the scale provided on the raw SEM image. Following the calibration, the total visible valve area was measured using a rectangular selection tool on each image, and values were summed over all the images of a particular valve. The RBCs, platelets, and amorphous acellular deposits in each image were manually marked with a polygon tool, fitted with a spline, and the corresponding areas were measured. The absolute areas are reported in μm2, and the relative areas are reported as a percentage (%) of the total area. The total area occupied by deposits (cellular/amorphous) and deposit-free areas in each SEM image was computed. The numerical values of areas from all the images in a voltage condition were aggregated and tabulated.
Data analysis
The raw variable values (initial and 30 minutes) and the percentage change between initial and final values for test, control, and reference samples were computed. Qualitative SEM findings were summarised descriptively. Numerical values of the total valve surface area analyzed, absolute/relative areas of different cellular and amorphous components are tabulated. Because the study objective was to qualitatively characterize and numerically summarize the effect of voltage on a small number of valves, formal statistical comparisons were not performed.
Results
The number of valves tested for each condition are shown in the Table 1 below:
Hematologic and coagulation assays
Platelet counts and activation Platelet counts changed by less than 10% (Tables 2–5) across all conditions and replicates. Flow-cytometric analysis of P-selectin (CD62P) (Table 3), marker of platelet activation with PRP experiments, showed a similar change in electrically activated valves and the reference sample, indicating no additional risk of platelet activation due to the electrical stimulation of the valves.
Platelet factor 4 (PF4) secretion PF4 is a positively charged protein stored in platelet α-granules and is released upon platelet activation. It was used as a marker of platelet activation in blood exposure experiments. At 0.25 V, PF4 concentrations increased from the control valve (0 V) to that at 0.5 V (Table 4), while similar trend was not seen in the replicate experiment (Table 5). The data does not indicate a consistent difference in the platelet activation between the different voltage conditions. Nevertheless, the voltage activation of the valve (0.25 & 0.5 V) does not appear to fully suppress the platelet activation processes indexed by PF4, an implication that is also corroborated by SEM findings (discussed later).
Partial thromboplastin time (PTT) and fibrinogen Coagulation parameters remained largely unchanged by the electric field. PTT changes are within or close to the measurement uncertainty for both PRP and blood exposure (Tables 2–5-5), except for the PRP experiment at 0.5 V where the PTT was higher. Similarly, the fibrinogen levels were close to the measurement uncertainty for all the experiments. (Tables 2–5-5).
Complement activation (C3a) Complement activation assays did not reveal a consistent trend across the experiments. With whole blood, C3a levels decreased with valves at 0.25 V, 0.5 V, and controls (Tables 4–5). PRP experiments showed an increase (Table 3) or a decrease of C3a (Table 2). These conflicting trends could indicate variations in donor complement activity, as opposed to being attributable to the influence of the electrical field.
Leukocyte counts Across all voltages and controls, leucocyte counts changed by less than 5% from baseline (Tables 4–5).
RBC counts, Hemoglobin concentration and hemolysis Red-cell counts were stable within the measurement uncertainty margin for all the experimental conditions (Tables 4–5). Hemoglobin concentrations (Tables 4–5) remained within ±2% of baseline for all samples with no consistent trend across voltage groups. Plasma haemoglobin (free Hb) increased after exposure (Table 4), possibly due to minor hemolysis, but the large percentage change seen is reflective of low baseline plasma hemoglobin levels. The estimated average percentage hemolysis in all samples was ≤ 0.1%. Typical values ranged from 0.05 to 0.06% at 0.25 & 0.5 V, indicating that electrical activation of the valve is non-hemolytic.
From the coagulation and hematologic assays (summarized in Figs 2 and 3), we infer the following: applying a potential at 0.25 V or 0.5 V does not injure blood cells or trigger coagulation. The small changes observed are within the reference variability and are similar to those seen in uncharged controls.
% change from initial to 30-minutes of immersion of the prosthetic valve is shown in the plot. PTT: Partial thromboplastin time, C3a: Activated complement C3. The height of the bar (numerical value displayed above the bar) for control (0 V) corresponds to the average of the two control valve measurements (Tables 2–3), while it depicts the individual valve values for 0.25 and 0.5 V.**.
% change from initial to 30 minutes of immersion of the prosthetic valve is shown in the plot. The height of the bars (numerical value displayed above the bar) for control (0 V) and 0.25 V corresponds to the average of two replicate measurements (Tables 4–5), while it depicts the individual values for valve at 0.5 V. PTT: Partial thromboplastin time, C3a: Activated complement C3, Hb: Hemoglobin, PF4: Platelet factor 4. **.
Morphological surface characterization of valves for Platelets, RBC, and acellular deposits with SEM
SEM images revealed a decrease in adsorption of cells and cell-free amorphous material as the voltage increased from 0.25 to 0.5 V (Figs 4 and 5, Table 6). On valves activated at 0.25 V, amorphous deposits can be seen to cover a substantial portion of the valve surface (Figs 4A & 4B). These deposits are likely adsorbed plasma proteins and/or fibrin-rich clots. Fig 4C shows a large cluster of adherent deformed RBCs. Many cells have collapsed, irregular surfaces, and lack the smooth biconcave contour. Some are shrunken, reflecting membrane damage or dehydration during the processing for SEM. The area of the valve covered by cells/proteinaceous deposits is more than valve at 0.5 V, but paradoxically less than the control valve (Table 6).
The scanning parameters of the images A, B, C are – A: Mag = 2700 × , WD = 13.1 mm; B: Mag = 5000 × , WD = 13.1 mm; C: Mag = 3000 × , WD = 13.4 mm. Mag = Magnification, WD = working distance.**.
The scanning parameters of the images A, B, C are – A: Mag = 1800 × , WD = 12.4 mm; B: Mag = 2500 × , WD = 12.4 mm; C: Mag = 5000 × , WD = 13.7 mm. Mag = Magnification, WD = working distance.**.
At low magnification (Fig 5A), the valve at 0.5 V shows a small number of scattered RBC, but no obvious proteinaceous/fibrin deposits are seen. At higher magnification, some amorphous deposits (Fig 5B) and isolated cells are seen (Figs 5B and 5C). The area covered by cells or protein deposits is the least, with 0.5 V (~4%).
The low magnification SEM image of the control valve (Fig 6A) shows a single cluster of RBCs, and at least three small clusters of platelets at the spreading-dendritic stage indicating activation of platelets [21, 22]. The control valve also shows adherent RBCs and proteinaceous deposits covering ~95% the scanned valve area (Fig 6B and 6C). The SEM images of an unexposed reference valve are shown for comparison in Fig 7.
We show the measured values of absolute and relative areas of different cellular and amorphous deposits on the valve surface at different voltages in Table 6. Although three images per voltage are shown in the figures (Figs 4-6), the values in Table 6 represent the aggregate of all the images for each voltage condition. All images and their corresponding area measurements are shown in the supplementary information (SEM images at 0.25 V: S1-S10 Figs; 0.5 V: S11-S15 Figs, Control: S16-S25 Figs). It can be seen that the area occupied by cells and/or amorphous deposits decreased as the valve voltage is increased from 0.25 to 0.5 V. In the valves examined, there is a decline in the absolute area occupied by activated platelets with 0.25 and 0.5 V compared to the control valve. The numerical values of absolute/relative areas seem to indicate that the diminished deposition with higher voltage is due to lower adherence of cells (RBC & platelet).
Discussion
We investigated the hemocompatibility and the potential anti-thrombotic properties of electrically activated prosthetic mechanical heart valves using an invitro approach. We found that the overall cell-adherence, particularly platelets, diminished with the surface voltage at 0.5 V. The deposit-free area increased from 58.38% at 0.25 V to 95.96% at 0.5 V, mostly due to reduced platelet adherence. The voltage-activated valves (0.25 & 0.5 V) were relatively free of activated platelets, while the control (0 V) valve surface had several platelet aggregates in the spreading-dendritic form. The acellular material adsorption was the lowest with the control, highest with 0.25 V, and intermediate with 0.5 V. The Electrical biomimetic glycocalyx (at 0.25 V & 0.5 V) did not eliminate the platelet activation as indicated by the PF4 assay (with whole blood immersion). No consistent differences in coagulation parameters (platelet markers, Partial thromboplastin time, and fibrinogen), cell (platelet, RBC, and leucocyte) counts, and complement activation were found between test and control conditions in the in vitro immersion experiments. Plasma hemoglobin and %hemolysis also did not show a notable difference during the 30-minutes of immersion between the test and valve conditions. These results indicate good hemocompatibility with electrically activated valves, and a minimal tendency to injure blood cells, activate platelets, or trigger coagulation.
Stable platelet counts indicate that the electric field did not cause notable platelet consumption. Lack of consistent difference in platelet and coagulation pathway markers between different voltage conditions implies that the electrical activation of the valve with negative charge does not confer a higher risk of activating platelets and/or coagulation pathways, supporting the safety of this intervention. The shear and agitation-driven platelet degranulation could account for the marginal changes in platelet markers (P-selectin and PF4) [23]. Platelets undergo characteristic morphological transformations as they are progressively activated, viz., round (stage I), dendritic (stage II), spreading dendritic (stage III), spread (stage IV), and fully spread (stage V) forms [21,22]. The round is the quiescent circulating form, while the fully spread is the completely activated form. The presence of spreading dendritic stage platelet clusters on the control valves, and their virtual absence on the voltage-activated valves, supports the potential effectiveness of the electrical glycocalyx in conferring thromboresistance to the valve and warrants further investigation in larger studies to replicate these results. The monotonous and steep decline in platelet (and RBC) adhesion and activation with the voltage can be explained by greater electrostatic repulsion between the negative zeta potential of the blood cell surface and the more negatively charged valve surface [24]. This is similar to the repulsive electrostatic interaction between the sialic-acid-rich blood cell surface and the endothelial glycocalyx, which is endowed with negatively charged glycosaminoglycans (GAG) [8–11]. The charge-based interaction is known to keep the blood cells away from the endothelium in health. By imposing an electrical potential of ~0.5 V across the valve surface, we reproduced this electrostatic barrier and minimized the tendency for cell adhesion, platelet activation, and cell-based coagulation-driven fibrin deposition [25–27]. In contrast to the effects on cell adhesion/activation, it is also well known that electronegative surfaces increase the contact-based activation of coagulation pathways [28–31]. This could account for intermediate levels of acellular deposition (possibly protein/coagulum) on valves at 0.5 V, reflecting the counteracting effects of diminished cell adhesion, (therefore reduced cell-based coagulation), and increased tendency for direct contact-based coagulation. The trade-off between the contrasting effects of cell adhesion and contact activation at higher surface electronegativity needs to be clarified in future studies.
These findings build on earlier reports that electrically polarised surfaces reduce thrombosis. Sawyer and Pate (1953) [13] made the earliest observations on the role of bioelectric factors in intravascular thrombosis. They studied the effect of electric current applied to electrodes touching the blood vessel wall. The thrombus deposition occurred near the anode (positively charged) but not near the cathode (negatively charged). In subsequent studies [14–17], Sawyer et al. documented a plethora of bioelectric effects viz., a) negative charge on blood vessel wall and blood cells b) reduction and reversal of negative charge after vascular injury at the site of thrombosis c) decrease in negative charge with a fall in pH d) increase in negative charge density with anti-thrombotic drugs and an reversal of polarity with prothrombotic drugs e) enhanced thrombogenecity with positively charged prosthetic materials in contrast to nonthrombogenic nature of negatively charged prosthetic materials f) better nonthrombogenic behaviour with a greater magnitude and more uniform distribution of negative charge. Sawyer’s work used cathodic DC [13–16] (direct current at 200μA)- or high-frequency AC [17] (10μA alternating current at 100 KHz) polarization of metallic grafts, while our study applies a programmable cathodic potential (0.25/0.5 V square pulse train at 70/min) generated by a pacemaker to the pyrolytic carbon mechanical heart valve. This study demonstrates that applying a modest electrical potential to the surfaces of an MHV can emulate the anti-thrombotic properties of the vascular endothelium. Our study indicates that a voltage of 0.5 V ensures thromboresistance and hemocompatibility. The energy requirement for maintaining a 0.5 V potential at 70 bpm is minimal and could potentially offer safe long-term non-pharmacological anticoagulation, free patients from life-long systemic anticoagulant drug usage, improve quality of life, and expand access to durable valve replacement.
Our study has the following limitations. Sample sizes were very small (single valve for some conditions as shown in Table 1), and analysis was descriptive; formal statistical testing was not performed. Findings from the single valve assessment (particularly SEM imaging) need to be verified in larger studies. Hence, our findings are preliminary in nature as the study does not have the statistical power to warrant a generalization of the results, particularly the superiority of 0.5 V over 0.25 V for the antithrombotic effect. It is also possible that there could be individual differences in the optimal voltage (0.25 vs. 0.5 V) for antithrombosis. Nevertheless, a greater anti-thrombotic effect with a higher voltage is a physiologically plausible conjecture that could be assessed in future studies. The optimal valve surface voltage that achieves maximum antithrombosis, minimum or no risk myocardial electrophysiological capture, while having a long battery life (for sustained effect) is still an open question. Differences in the manufacturing characteristics of the valves (possible surface irregularities) could potentially explain the results obtained with respect to the effect of different voltages. However, the valves used in the study are approved and widely used for valve replacement in humans. The occurrence of thrombosis after implantation due to surface defects has not been reported to our knowledge. Furthermore, SEM image assessment did not reveal any gross surface defect that is likely to induce thrombosis in any of the fields examined, making the possibility of a manufacturing defect explaining the study findings very unlikely.
This ex vivo study used human PRP and whole blood under static agitation; the complex hemodynamics of the cardiac outflow tract may influence platelet interactions differently in vivo. Biomarkers such as PF4 and complement C3a are surrogate indicators; the definitive endpoint remains prevention of clinical valve thrombosis. Only a single disc geometry and two voltages above the threshold were tested; other materials or pulse waveforms may alter the optimal potential. The in vitro paradigm precludes the assessment of the electrophysiological effects of voltage applied to the valve on the heart. However, the voltages tested in the study (0.25 & 0.5 V) are unlikely to significantly impact the rhythm genesis or propagation in the cardiac muscle for the following reasons: 1. The threshold voltage for myocardial capture for pacing applications is ~ 1.5 V @ 0.4 ms pulse width, while we have used the pacemaker in a non-pacing mode at 0.25 & 0.5 V for electrical valve activation. The normal vascular endothelium that lines the inside of the heart/blood vessels has a potential difference of ~0.3 V 2. The valve prosthesis will be implanted in the mitral position, which is surrounded by an electrically insulating annulus fibrosus (the physiological fibrous skeleton of the heart). The annulus fibrosis prevents the direct propagation of electrical activity from atria to ventricles in health, and similar behaviour is expected to protect the heart from the electrophysiological spillover effects. 3. The valve prosthesis has a cuff of sewing ring, made up of polyester fabric, which helps suture the valve to the annulus. This fabric material has electrical insulator properties. 4. Peri-prosthetic valve fibrosis, known to occur after valve implantation, further walls off the prosthetic valve electrically from the rest of the heart. A schematic image indicating the potential method of deployment in the clinical context for which the valve is proposed to be used is shown in Fig 8, including pacemaker connectivity.
A- Prosthetic mechanical heart valve. B- Valve-lead junction. C-Lead electrode. D-Pacemaker.
Further work will focus on in vivo validation and long-term safety. Swine animal models will be used to assess valve patency over extended periods without systemic anticoagulation. Device miniaturization and integration into implantable pacemakers will be explored to create a fully implantable self-anticoagulant mechanical heart valve.
Supporting information
S1 Fig. Cells and other deposits are annotated by yellow colored outlines.
Total area = 112.749 μm2, platelet area = 31.212 μm2 (single platelet in dendritic spreading stage of platelet activation), RBC area = 0, Acellular deposit area = 0, Total area occupied by masses (cells and deposits)=31.212 μm2 (27.68%), Mass free area = 81.537 μm2 (72.32%), Magnification = 10000x, working distance = 14 mm. The scale bar is shown in the image for reference.
https://doi.org/10.1371/journal.pone.0336760.s001
(TIFF)
S2 Fig. Cells and other deposits are annotated by yellow colored outlines.
Mass of shrunken RBCs are seen along with amorphous material deposits. Total area = 444.587 μm2, platelet area = 0, RBC area = 62.06 μm2, Acellular deposit area = 21.662 μm2, Total area occupied by masses (cells and deposits)=83.722 μm2 (18.83%), Mass free area = 360.865 μm2 (81.17%), Magnification = 5000x, working distance = 13.1 mm. The scale bar is shown in the image for reference.
https://doi.org/10.1371/journal.pone.0336760.s002
(TIFF)
S3 Fig. Cells and other deposits are annotated by yellow colored outlines.
Single deformed RBC is seen along with amorphous material deposits. Total area = 452.001 μm2, platelet area = 0, RBC area = 21.61 μm2, Acellular deposit area = 25.811 μm2, Total area occupied by masses (cells and deposits)=47.421 μm2 (10.49%), Mass free area = 404.58 μm2 (89.51%), Magnification = 5000x, working distance = 13.2 mm. The scale bar is shown in the image for reference.
https://doi.org/10.1371/journal.pone.0336760.s003
(JPG)
S4 Fig. Cells and other deposits are annotated by yellow colored outlines.
Single RBC is seen along with amorphous material deposits. Total area = 450.614 μm2, platelet area = 0, RBC area = 20.986 μm2, Acellular deposit area = 28.163 μm2, Total area occupied by masses (cells and deposits)=49.149 μm2 (10.91%), Mass free area = 401.465 μm2 (89.09%), Magnification = 5000x, working distance = 12.6 mm. The scale bar is shown in the image for reference.
https://doi.org/10.1371/journal.pone.0336760.s004
(TIFF)
S5 Fig. Cells and other deposits are annotated by yellow colored outlines.
A single RBC, along with large areas of amorphous deposits covering the valve, is seen. Total area = 451.497 μm2, platelet area = 0, RBC area = 19.736 μm2, Acellular deposit area = 298.594 μm2, Total area occupied by masses (cells and deposits)=318.33 μm2 (70.51%), Mass free area = 133.167 μm2 (29.49%), Magnification = 5000x, working distance = 13.1 mm. The scale bar is shown in the image for reference.
https://doi.org/10.1371/journal.pone.0336760.s005
(TIFF)
S6 Fig. Cells and other deposits are annotated by yellow colored outlines.
Large areas of amorphous deposits resembling fibrin coagulum covering the valve are seen. Total area = 1553.502 μm2, platelet area = 0, RBC area = 0 μm2, Acellular deposit area = 1138.716 μm2, Total area occupied by masses (cells and deposits)=1138.716 μm2 (73.3%), Mass free area = 414.786 μm2 (26.7%), Magnification = 2700x, working distance = 13.1 mm. The scale bar is shown in the image for reference.
https://doi.org/10.1371/journal.pone.0336760.s006
(TIFF)
S7 Fig. Cells and other deposits are annotated by yellow colored outlines.
Large numbers of RBCs forming aggregates are seen covering the valve. Total area = 1267.215 μm2, platelet area = 0, RBC area = 570.823 μm2, Acellular deposit area = 5.531 μm2, Total area occupied by masses (cells and deposits)=576.354 μm2 (45.48%), Mass free area = 690.861 μm2 (54.52%), Magnification = 3000x, working distance = 13.4 mm. The scale bar is shown in the image for reference.
https://doi.org/10.1371/journal.pone.0336760.s007
(TIFF)
S8 Fig. Cells and other deposits are annotated by yellow colored outlines.
Isolated RBCs and small focal deposits are seen on the valve. Total area = 451.65 μm2, platelet area = 0, RBC area = 56.026 μm2, Acellular deposit area = 98.267 μm2, Total area occupied by masses (cells and deposits)=154.293 μm2 (34.16%), Mass free area = 297.357 μm2 (65.84%), Magnification = 5000x, working distance = 12.9 mm. The scale bar is shown in the image for reference.
https://doi.org/10.1371/journal.pone.0336760.s008
(TIFF)
S9 Fig. Cells and other deposits are annotated by yellow colored outlines.
Large amorphous material seen on the valve. Total area = 451.499 μm2, platelet area = 0, RBC area = 0 μm2, Acellular deposit area = 270.202 μm2, Total area occupied by masses (cells and deposits)=270.202 μm2 (59.85%), Mass free area = 181.297 μm2 (40.15%), Magnification = 5000x, working distance = 12.8 mm. The scale bar is shown in the image for reference.
https://doi.org/10.1371/journal.pone.0336760.s009
(TIFF)
S10 Fig. Cells and other deposits are annotated by yellow colored outlines.
Clusters of RBCs and small focal deposits are seen on the valve. Total area = 1804.586 μm2, platelet area = 0, RBC area = 351.496 μm2, Acellular deposit area = 75.685 μm2, Total area occupied by masses (cells and deposits)=427.181 μm2 (23.67%), Mass free area = 1377.405 μm2 (76.33%), Magnification = 2500x, working distance = 12.3 mm. The scale bar is shown in the image for reference.
https://doi.org/10.1371/journal.pone.0336760.s010
(TIFF)
S11 Fig. Cells and other deposits are annotated by yellow colored outlines.
Isolated RBC, smaller platelet aggregate, and amorphous deposits are seen on the valve. Total area = 1800.529 μm2, platelet area = 21.542, RBC area = 21.64 μm2, Acellular deposit area = 529.43 μm2, Total area occupied by masses (cells and deposits)=573.393 μm2 (31.85%), Mass free area = 1227.136 μm2 (68.15%), Magnification = 2500x, working distance = 12.9 mm. The scale bar is shown in the image for reference.
https://doi.org/10.1371/journal.pone.0336760.s011
(TIFF)
S12 Fig. Cells and other deposits are annotated by yellow colored outlines.
Amorphous deposits resembling RBC ghosts were seen on the valve. Total area = 452.173 μm2, platelet area = 0, RBC area = 0, Acellular deposit area = 219.492 μm2, Total area occupied by masses (cells and deposits)=219.492 μm2 (48.54%), Mass free area = 232.681 μm2 (51.46%), Magnification = 5000x, working distance = 12.9 mm. The scale bar is shown in the image for reference.
https://doi.org/10.1371/journal.pone.0336760.s012
(TIFF)
S13 Fig. Cells and other deposits are annotated by yellow colored outlines.
A single RBC with finger like cytoplasmic projections, presumably due to dehydration, is seen. Total area = 265.16 μm2, platelet area = 0, RBC area = 16.573 μm2, Acellular deposit area = 0, Total area occupied by masses (cells and deposits)=16.573 μm2 (6.25%), Mass free area = 248.587 μm2 (93.75%), Magnification = 6500x, working distance = 13 mm. The scale bar is shown in the image for reference.
https://doi.org/10.1371/journal.pone.0336760.s013
(TIFF)
S14 Fig. Cells and other deposits are annotated by yellow colored outlines.
Two isolated RBCs with wrinkled membranes are seen. Total area = 450.215 μm2, platelet area = 0, RBC area = 51.721 μm2, Acellular deposit area = 0.807, Total area occupied by masses (cells and deposits)=52.528 μm2 (11.67%), Mass free area = 397.687 μm2 (88.33%), Magnification = 5000x, working distance = 13 mm. The scale bar is shown in the image for reference.
https://doi.org/10.1371/journal.pone.0336760.s014
(JPG)
S15 Fig. Cells and other deposits are annotated by yellow colored outlines.
Few solitary RBCs and small deposits are seen, but much of the valve area is virtually deposit-free. Total area = 23034.515 μm2, platelet area = 0, RBC area = 169.504 μm2, Acellular deposit area = 18.334, Total area occupied by masses (cells and deposits)=187.838 μm2 (0.82%), Mass free area = 22846.68 μm2 (99.18%), Magnification = 700x, working distance = 12.5 mm. The scale bar is shown in the image for reference.
https://doi.org/10.1371/journal.pone.0336760.s015
(TIFF)
S16 Fig. Cells and other deposits are annotated by yellow colored outlines.
A solitary RBC and a cluster of RBCs are seen. One or more platelets in the spreading dendritic stage are seen. Total area = 561.36 μm2, platelet area = 20.105, RBC area = 68.779 μm2, Acellular deposit area = 0, Total area occupied by masses (cells and deposits)=88.88 μm2 (15.83%), Mass free area = 472.476 μm2 (84.17%), Magnification = 4500x, working distance = 12.7 mm. The scale bar is shown in the image for reference.
https://doi.org/10.1371/journal.pone.0336760.s016
(TIFF)
S17 Fig. Cells and other deposits are annotated by yellow colored outlines.
A cluster of activated platelets in the spreading dendritic stage is seen in the centre of the field. Total area = 447.435 μm2, platelet area = 44.362, RBC area = 0 μm2, Acellular deposit area = 19.548, Total area occupied by masses (cells and deposits)=63.91 μm2 (14.28%), Mass free area = 383.525 μm2 (85.72%), Magnification = 5000x, working distance = 12.6 mm. The scale bar is shown in the image for reference.
https://doi.org/10.1371/journal.pone.0336760.s017
(TIFF)
S18 Fig. Cells and other deposits are annotated by yellow colored outlines.
A solitary RBC is seen in the center of the field. Total area = 447.921 μm2, platelet area = 0, RBC area = 18.494 μm2, Acellular deposit area = 0, Total area occupied by masses (cells and deposits)=18.494 μm2 (4.13%), Mass free area = 429.427 μm2 (95.87%), Magnification = 5000x, working distance = 12.6 mm. The scale bar is shown in the image for reference.
https://doi.org/10.1371/journal.pone.0336760.s018
(TIFF)
S19 Fig. Cells and other deposits are annotated by yellow colored outlines.
A large cluster of deformed RBCs is seen in the lower half of the field. At least three clusters of adherent and activated platelets in the spreading dendritic stage are seen in addition to focal amorphous deposits. Total area = 3491 μm2, platelet area = 508.738 μm2, RBC area = 149.973 μm2, Acellular deposit area = 85.18 μm2, Total area occupied by masses (cells and deposits)=743.891 μm2 (21.31%), Mass free area = 2747.109 μm2 (78.69%), Magnification = 1800x, working distance = 12.4 mm. The scale bar is shown in the image for reference.
https://doi.org/10.1371/journal.pone.0336760.s019
(TIFF)
S20 Fig. Cells and other deposits are annotated by yellow colored outlines.
A larger activated platelet-rich deposit is seen in the field. Total area = 449.697 μm2, platelet area = 205.294 μm2, RBC area = 0, Acellular deposit area = 7.403 μm2, Total area occupied by masses (cells and deposits)=213.327 μm2 (47.44%), Mass free area = 236.37 μm2 (52.56%), Magnification = 5000x, working distance = 12.5 mm. The scale bar is shown in the image for reference.
https://doi.org/10.1371/journal.pone.0336760.s020
(TIFF)
S21 Fig. Cells and other deposits are annotated by yellow colored outlines.
Two isolated swollen RBC are seen with focal amorphous deposits. Total area = 1793.741 μm2, platelet area = 0 μm2, RBC area = 39.984 μm2, Acellular deposit area = 7.977 μm2, Total area occupied by masses (cells and deposits)=47.961 μm2 (2.67%), Mass free area = 1745.78 μm2 (97.33%), Magnification = 2500, working distance = 12.4 mm. The scale bar is shown in the image for reference.
https://doi.org/10.1371/journal.pone.0336760.s021
(TIFF)
S22 Fig. Cells and other deposits are annotated by yellow colored outlines.
Total area = 448.409 μm2, platelet area = 0 μm2, RBC area = 0 μm2, Acellular deposit area = 0 μm2, Total area occupied by masses (cells and deposits)=0 (0%), Mass free area = 448.409 μm2 (100%), Magnification = 5000x, working distance = 13.7 mm. The scale bar is shown in the image for reference.
https://doi.org/10.1371/journal.pone.0336760.s022
(TIFF)
S23 Fig. Cells and other deposits are annotated by yellow colored outlines.
A single platelet extending its dendritic process, indicating the incipient stage of platelet activation. Total area = 454.589 μm2, platelet area = 2.379 μm2, RBC area = 0, Acellular deposit area = 0, Total area occupied by masses (cells and deposits)=2.379 μm2 (0.52%), Mass free area = 452.21 μm2 (99.48%), Magnification = 5000x, working distance = 13.7 mm. The scale bar is shown in the image for reference.
https://doi.org/10.1371/journal.pone.0336760.s023
(TIFF)
S24 Fig. Cells and other deposits are annotated by yellow colored outlines.
Focal amorphous deposits are seen. Total area = 451.483 μm2, platelet area = 0, RBC area = 0, Acellular deposit area = 14.404 μm2, Total area occupied by masses (cells and deposits)=14.404 μm2 (3.19%), Mass free area = 437.079 μm2 (96.81%), Magnification = 5000x, working distance = 13.8 mm. The scale bar is shown in the image for reference.
https://doi.org/10.1371/journal.pone.0336760.s024
(TIFF)
S25 Fig. Cells and other deposits are annotated by yellow colored outlines.
Total area = 77.924 μm2, platelet area = 0, RBC area = 0, Acellular deposit area = 14.404 μm2, Total area occupied by masses (cells and deposits)=0 (0%), Mass free area = 77.924 μm2 (100%), Magnification = 12000x, working distance = 13.8 mm. The scale bar is shown in the image for reference.
https://doi.org/10.1371/journal.pone.0336760.s025
(TIFF)
Acknowledgments
We thank the volunteers who donated blood for this study and the staff of the Biomedical Technology Wing, Sree Chitra Tirunal Institute for Medical Sciences and Technology, Trivandrum, for performing the hemocompatibility assays. This work was supported by the Sajja Heart Foundation.
References
- 1. Pibarot P, Dumesnil JG. Prosthetic heart valves: selection of the optimal prosthesis and long-term management. Circulation. 2009;119(7):1034–48. pmid:19237674
- 2. Wang Y, Fu Y, Wang Q, Kong D, Wang Z, Liu J. Recent advancements in polymeric heart valves: From basic research to clinical trials. Mater Today Bio. 2024;28:101194. pmid:39221196
- 3. Leviner DB, Abraham D, Ronai T, Sharoni E. Mechanical Valves: Past, Present, and Future—A Review. J Clinical Medicine. 2024;13.
- 4. Palumbo IM, Menichelli D, Biccirè FG, Pannunzio A, Pignatelli P, Pastori D. Long-Term Risk of Hospitalization and Death in Patients With Mechanical Prosthetic Heart Valves. Am J Cardiol. 2024;228:16–23. pmid:39097153
- 5. Forti S, Lunelli L, Della Volpe C, Siboni S, Pasquardini L, Lui A. Hemocompatibility of pyrolytic carbon in comparison with other biomaterials. Diam Relat Mater. 2011;20:762–9.
- 6. Zawidlak-Węgrzyńska B, Rydz J, Musioł M, Radziwon-Balicka A. Polymer-drug anti-thrombogenic and hemocompatible coatings as surface modifications. Pharmaceutics. 2024;16(3).
- 7. Bhattacharjee A, Savargaonkar AV, Tahir M, Sionkowska A, Popat KC. Surface modification strategies for improved hemocompatibility of polymeric materials: a comprehensive review. RSC Adv. 2024;14(11):7440–58. pmid:38433935
- 8. Milusev A, Rieben R, Sorvillo N. The Endothelial Glycocalyx: A Possible Therapeutic Target in Cardiovascular Disorders. Front Cardiovasc Med. 2022;9:897087. pmid:35647072
- 9. Nieuwdorp M, Meuwese MC, Vink H, Hoekstra JBL, Kastelein JJP, Stroes ESG. The endothelial glycocalyx: a potential barrier between health and vascular disease. Curr Opin Lipidol. 2005;16(5):507–11. pmid:16148534
- 10. Pries AR, Secomb TW, Gaehtgens P. The endothelial surface layer. Pflugers Arch. 2000;440(5):653–66. pmid:11007304
- 11. Reitsma S, Slaaf DW, Vink H, van Zandvoort MAMJ, oude Egbrink MGA. The endothelial glycocalyx: composition, functions, and visualization. Pflugers Arch. 2007;454(3):345–59. pmid:17256154
- 12. Kuchinka J, Willems C, Telyshev DV, Groth T. Control of Blood Coagulation by Hemocompatible Material Surfaces-A Review. Bioengineering (Basel). 2021;8(12):215. pmid:34940368
- 13. Sawyer PN, Pate JW. Bio-electric phenomena as an etiologic factor in intravascular thrombosis. Am J Physiol. 1953;175(1):103–7.
- 14. Gileadi E, Stanczewski B, Parmeggiani A, Lucas T, Ranganathan M, Srinivasan S. Antithrombogenic characteristics of cathodically polarized copper prostheses. J Biomed Mater Res. 1972;6(6):489–98.
- 15. Srinivasan S, Sawyer PN. Role of surface charge of the blood vessel wall, blood cells, and prosthetic materials in intravascular thrombosis. J Colloid Interface Sci. 1970;32(3):456–63. pmid:5417531
- 16. Chopra PS, Srinivasan S, Lucas T, Sawyer PN. Relation between thrombosis on metal electrodes and the position of metal in the electromotive series. Nature. 1967;215(5109):1494. pmid:6059571
- 17. Lucas TR, Stanczewski B, Ramasamy N, Srinivasan S, Kammlott GW, Sawyer PN. Nonthrombogenic ac polarized copper prosthesis. Biomater Med Devices Artif Organs. 1975;3(2):215–32. pmid:1182252
- 18.
Sajja LR. Mechanical prosthetic heart valve assembly for the provision of surface anticoagulation thereon. The United States Patent and Trademark Office (USPTO). 2020. 11.
- 19.
Biological evaluation of medical devices - Part 4: Selection of tests for interactions with blood. International Organization for Standardization. 2017. https://www.iso.org/standard/70936.html
- 20. de Jonge G, Dos Santos TL, Cruz BR, Simionatto M, Bittencourt JIM, Krum EA, et al. Interference of in vitro hemolysis complete blood count. J Clin Lab Anal. 2018;32(5):e22396. pmid:29396875
- 21. Modic M, Junkar I, Stana‐Kleinschek K, Kostanjšek R, Mozetič M. Morphology Transformations of Platelets on Plasma Activated Surfaces. Plasma Processes & Polymers. 2014;11(6):596–605.
- 22. Allen RD, Zacharski LR, Widirstky ST, Rosenstein R, Zaitlin LM, Burgess DR. Transformation and motility of human platelets: details of the shape change and release reaction observed by optical and electron microscopy. J Cell Biol. 1979;83(1):126–42. pmid:511936
- 23. Rana A, Westein E, Niego B, Hagemeyer CE. Shear-Dependent Platelet Aggregation: Mechanisms and Therapeutic Opportunities. Front Cardiovasc Med. 2019;6:141. pmid:31620451
- 24. Hughes MP. The cellular zeta potential: cell electrophysiology beyond the membrane. Integr Biol. 2024.
- 25. Riddel JP Jr, Aouizerat BE, Miaskowski C, Lillicrap DP. Theories of blood coagulation. J Pediatr Oncol Nurs. 2007;24(3):123–31. pmid:17475978
- 26. Hoffman M, Monroe DM. Coagulation 2006: a modern view of hemostasis. Hematol Oncol Clin North Am. 2007;21(1):1–11. pmid:17258114
- 27. Ferreira N a t á l i a C, de Oliveira SM, Sant’Ana DLM, das Graças CM. A cell-based model of coagulation and its implications. Rev Bras Hematol Hemoter. 2010;55(31):416–21.
- 28. Grover SP, Mackman N. Intrinsic Pathway of Coagulation and Thrombosis. Arterioscler Thromb Vasc Biol. 2019;39(3):331–8. pmid:30700128
- 29. van der Graaf F, Keus FJ, Vlooswijk RA, Bouma BN. The contact activation mechanism in human plasma: activation induced by dextran sulfate. Blood. 1982;59(6):1225–33. pmid:6177360
- 30. Park S, Park JK. Back to basics: the coagulation pathway. Blood Res. 2024;59(1):35. pmid:39466528
- 31. Vogler EA, Siedlecki CA. Contact activation of blood-plasma coagulation. Biomaterials. 2009;30(10):1857–69. pmid:19168215