The impact of anodization modification on titanium interaction with human osteoblasts and fibroblasts (in vitro study)

  • Kamonmard Moungthong,

    Roles Formal analysis, Investigation, Software, Writing – original draft

    Affiliation Department of Prosthodontics, Faculty of Dentistry, Chulalongkorn University, 34 Henri-Dunant Road, Wangmai, Patumwan, Bangkok, Thailand

  • Daneeya Na Nan,

    Roles Formal analysis, Software, Validation

    Affiliation Division of Research Affairs and Center of Excellence for Dental Stem Cell Biology, Faculty of Dentistry, Chulalongkorn University, Bangkok, Thailand

  • Pravej Serichetaphongse,

    Roles Methodology, Supervision, Writing – review & editing

    Affiliations Faculty of Dentistry, Siam University, 38 Petchkasem Road, Bangwa, Phasicharoen, Bangkok, Thailand, Dental Center, Bumrungrad International Hospital, Bangkok, Thailand

  • Wareeratn Chengprapakorn

    Roles Formal analysis, Supervision, Writing – review & editing

    wchengprapakorn@gmail.com

    Affiliation Department of Prosthodontics, Faculty of Dentistry, Chulalongkorn University, 34 Henri-Dunant Road, Wangmai, Patumwan, Bangkok, Thailand

Abstract

Title

The impact of new anodization modification on titanium interaction with human osteoblasts and fibroblasts (in vitro)

Purpose

To compare the behavior of osteoblast and fibroblast cells, as well as the surface characteristics, of modified anodized titanium surfaces at different voltages with sandblasted and acid-etched titanium surfaces.

Methods

Twenty titanium discs were fabricated and subjected to 4 types of surface treatments: machined surface, sandblasted and acid-etched surface (SLA), modified anodized A surface at 30V DC and modified anodized B surface at 70V DC. Surface roughness and contact angle were measured. Osteoblast cells were seeded onto titanium discs and evaluated for cell proliferation, Alkaline phosphatase (ALP) activity, Alizarin red staining, and cell morphology with SEM imaging. Immunofluorescent staining was performed to assessed cell adhesion of both osteoblast and fibroblast cells. Statistic analysis was conducted using ANOVA with Tukey post hoc multiple comparisons.

Results

The topography and osteoblast cell spreading on both modified anodized titanium surfaces differed significantly from machined and SLA surfaces (p < 0.05). However, there was no significant difference in osteoblast cell spreading between modified anodized surfaces A and B. In addition, anodized surfaces A and B showed significantly higher ALP activity and Alizarin red staining than both SLA and machined surfaces (p < 0.05). For human gingival fibroblast cells, the modified anodized surfaces exhibited significantly higher cell adherence (p < 0.05), followed by the machined surface, with SLA surfaces showing the least adherence.

Conclusions

The modified anodized surfaces showed superior biocompatibility, enhancing attachment of both osteoblast and fibroblast cells, whereas the SLA surfaces exhibited the least fibroblast cell adherence.

Introduction

Dental implants are widely recognized as a reliable, long-term solution for tooth replacement, with their success largely dependent on effective osseointegration and soft tissue integration [14]. Over recent years, significant advancements in implant surface modification have emerged, aiming to accelerate osseointegration for quicker loading times and to reduce the risk of peri-implantitis by inhibiting biofilm development and attachment [57]. Common techniques for creating microtopography on implant surfaces include sandblasting, acid etching, plasma spraying, grit blasting, hydroxyapatite coating, and anodization [8,9]. Research indicates that an optimal implant surface roughness, typically ranging between 1–1.5 µm, promotes the osteoblast cell response, thereby enhances osseointegration [5,1012]. Despite these advancements, the ongoing challenge remains to identify or develop a surface treatment method that maximizes osteoconductive properties while simultaneously preventing bacterial invasion. Anodization is one such surface modification technique showing promise, as it supports both osseointegration and soft tissue integration through favorable cellular responses [13,14]. However, the impact of anodization on soft tissue behavior, particularly at varying voltages, demands further exploration. Analyzing osteoblast and fibroblast cell adhesion on titanium surface anodized at different voltages can provide valuable insights for selecting the most appropriate implants for specific clinical conditions, ultimately guiding dental implant choices more effectively.

Anodization involves electrochemically treating the implant surface, often following a sandblast large grit and acids etch (SLA) treatment, to create an oxide layer [15]. This oxide layer acts as a corrosion barrier with high surface energy [16]. The thickness of the oxide layer is controlled by the anodizing voltage, which also affects the surface color due to wavelength reflection [1719]. Recent studies have demonstrated that anodization effectively enhances osteoblast adhesion and proliferation, alkaline phosphatase activity, and calcium deposits, thereby improving osseointegration [13,14,20]. For instance, Traini et al. indicated that anodized implant surface with a titanium oxide (TiO2) coating exhibit lower surface tension and a hydrophilic surface, which fosters improved blood clot formation [21]. This hydrophilic quality is crucial as it facilitates osteoblast cell attachment to the implant surface, enhancing the osseointegration process necessary for immediate implant loading and long-term success [22].

Maintaining peri-implant soft tissue health is critical for the long-term success of dental implants, as peri-implant infections are the primary cause of implant failure, leading to bone loss [23,24]. Hemidesmosome attachment plays a vital role in limiting bacterial infiltration, thus reducing the risk of infection [2528]. Anodization enhances both osseointegration and soft tissue attachment through favorable responses from osteoblasts and fibroblasts. In vitro studies have demonstrated better fibroblast cell adherence and reduced biofilm formation on anodized surfaces [29,30]. While SLA treatments improve cellular adhesion, they also inadvertently enhance bacterial adherence, which can result in increased inflammation and bone loss compared to polished surfaces [3133]. In contrast, anodized surfaces exhibit significantly decreased bacterial adherence, presenting a promising alternative for implant surfaces [30,34,35]. Advancements in anodized implant surfaces have shown improved interactions with osteoblasts and gingival fibroblasts. However, there is a lack of comprehensive studies examining cellular responses across different implant surfaces at varying anodization voltages. Our objective is to investigate the effects of different anodization voltages on titanium surfaces, alongside machined and SLA surfaces, focusing on osteoblast and fibroblast cell responses, surface characteristics, and hydrophilicity.

Materials and methods

The study size was calculated from the pilot study. Twenty type IV titanium discs, each measuring 9 mm in diameter and 3 mm in thickness, were utilized according to ISO 5832–2:2018 standards [36]. The discs were divided into four groups, with each group receiving different surface treatment. The control group was machined titanium surface. The first surface modification is sandblasted with large grit and acid etch (SLA). The titanium discs were sandblasted with aluminum oxide (Al2O3), followed by acid etching using a mixture of 60% sulfuric acids and 30% hydrochloric acid. Two distinct anodization processes were applied to the surfaces following sandblasting and acid etching. The first one were anodized in a 5% sulfuric solution using a voltage range of 20-30V DC at 0.1 A to get anodized A (blue) surface. The second one were anodized in a 5% sulfuric solution using a voltage range of 60-70V DC at 0.1 A to get anodized B (yellow) surface. After surfaces modifications were completed, all samples underwent a sterilization process by cleansing with alcohol, deionized water, and phosphate buffer saline in preparation for subsequent analysis.

Surface characterization

The surface roughness of the titanium discs was measured using a surface profilometer (Alicona, infinitefocus, Austria) at x10 magnification. Three different locations on each sample were scanned and measured to obtain representative data. The surface morphology of the titanium discs were assessed using scanning electron microscopy (SEM)(JSM 5410LV, JEOL, Japan) at x5000 magnification. The three-dimensional profile images of different titanium surfaces were measured using a surface profilometer (Alicona, infinitefocus, Austria) at x50 magnification.

The surface wettability was determined using the sessile drop method by measuring the water contact angle [37]. The 10 μL sessile droplet of deionized water was deposited vertically on the surface using a micropipette, ensuring no physical contact with the surface. The contact angle was measured 5 times for each samples to obtain consistent data. The results were reported as mean ± standard deviation.

Behavior of osteoblasts and fibroblasts on titanium discs

Human osteoblast cells were obtained from bone fragments during dental implant procedures. Written informed consent was obtained from volunteers aged 20−50 years, with approval from the ethics board of Faculty of Dentistry, Chulalongkorn University (HREC-DCU 2022−098). Bone fracments were harvested between 16 December 2022 and 29 May 2023 and cultured in high glucose-Dulbecco modified eagle medium (DMEM Gibco BRL, Ca, USA) containing 10% fetal bovine serum, 2 mM of L-glutamine, penicillin (100 U/ml), streptomycin (100 mg/ml, Gibco) and amphotericin B (5 mg/ml). Induction of differentiation to osteoblasts was promoted by culturing with 10 nM dexamethasone, 25 µg/ml L-ascorbic acid, and 10 mM β-glycerophosphate. Cells from the third to the sixth passages were use for experiment after characterization.

Human gingival fibroblast cells were obtained from the molar teeth of 3 patients, aged between 20–30 years, extracted for orthodontic reasons at the Department of Surgery, the Faculty of Dentistry, Chulalongkorn University (HREC-DCU 2024–011) between 6 May 2024 and 20 August 2024. Gingival fibroblasts were scraped from the middle third of the root and cultured for 3–4 weeks in DMEM containing 10% fetal bovine serum, 2 mM of L-glutamine, penicillin (100 U/ml), streptomycin (100 mg/ml) (Gibco) and amphotericin B (5 mg/ml). Cells from the third to the fifth passages were used in the study.

All tissues used to isolate osteoblasts and fibroblasts were obtained from healthy patients. The patients had no history of systemic disease, smoking, or medication that could affect bone or soft tissue healing. Cells were seeded on polystyrene surface and all titanium discs in 48 well plates at a density of 2 × 104 cells/surface. The cultures were incubated for 24 hours at 37 °C. The proliferation rate of osteoblasts and fibroblasts was evaluated using the MTT (methylthiazolytetrazolium) assay. Measurements were taken at 24 hours, 3 days, and 7 days post-seeding. At each time point, the culture media was measured at an absorbance of 570 nm using a microplate reader (Elx800; Biotek, Winooski, VT).

Immunofluorescent staining

Human osteoblast cells were seeded at 2 × 104 cells per well in 24-well plates for 1 hour and 3 hours. Human gingival fibroblast cells were also seeded at the same density and observed after 3 hours. Cells were rinsed twice with 0.1M PBS, fixed in 4% paraformaldehyde and permeabilized in 0.1% Triton X 100 for 10 minutes. Cells were incubated overnight with primary antibody vinculin (1:100; Abcam, Grand Island, NY) to stain focal adhesion points. Rhodamine-conjugated phalloidin (1:200; Molecular Probes-Life Technologies, Grand Island, NY) was used to stain actin filaments. Nuclei were stained with 300 nM 4´,6-diamidino-2-phenylindole, dihydrochloride (DAPI) for 20 min. A glass coverslip was mounted on the titanium surface containing cells using an antifade kit (Vectashield, Vector Laboratories, Burlingame, CA). Fluorescent images were analyzed using a fluorescent microscope (Axiovert 40CFL, Carl Zeiss, Gottingen, Germany). Cell adhesion was quantified by counting the number of cells in 3 fields of view per sample, across three discs for each titanium surface.

Alkaline phosphatase (ALP) activity

Osteogenically differentiated human osteoblast cells were seeded on all titanium discs in 48-well plates at the density of 3x104 cells per well. ALP activity was measured at two time points: after 3 and after 7 days. The cells were incubated with 300 μL of BCIP/NBT (5-Bromo-4-chloro-3-indolyl phosphate/Nitro blue tetrazolium) substrate solution (Sigma Life Science, USA) for 30 minutes at 37°C. For quantification, the stained cells were rinsed with 200 µl of 2M KOH to stop the reaction, followed by 300 µl of DMSO to dissolve the intracellular NBT-formazan crystal [38]. The absorbance was measured using a microplate reader at 620 nm (Elx800; Biotek, Winooski, VT) [39].

Alizarin red S staining

Assessment of mineralization was conducted at two time points: after 14 days and after 21 days. Cells were fixed with cold methanol for 10 minutes. Fixed cells were stained with 1% Alizarin Red S solution (Sigma Life Science, USA) for 5 minutes at room temperature to detect calcium deposits. The quantification was performed by dissolved stained sample in 10% of cetylpyridinium chloride and measured for absorbance at 570 nm.

Statistical analysis

Statistical analysis was performed using Prism (GraphPad, La Jolla, CA, USA). Shapiro-Wilk’s test was used to assess the normality of the data and Levene’s test was used to determine the homogeneity of variance. Analysis of Variance (ANOVA) with Tukey post hoc multiple comparisons were conducted to determine the significance of the results. The data presented for all groups are expressed as standard deviation (SD). Significance levels were categorized as follows: * p < 0.05, ** p < 0.01, *** p < 0.001, **** p < 0.0001

Result

Surface characteristic analysis

The surface roughness (Ra) and root mean square surface roughness (Sa) values for different titanium surface treatments are presented in the Table 1. Statistical analysis indicated that the machined surface (control) had the lowest Ra (0.33 µm) and Sa (0.38 µm) values, which were significantly different from the treated surfaces. Sandblast and acid-etches surface exhibited increased roughness compared to the machined surface, contributing to a more textured appearance. Anodized surfaces (A and B) showed comparable roughness values to each other, with increased roughness compared to the machined surface. However, anodized A displayed slightly higher mean Ra (1.19 µm) and Sa (1.31 µm) values compared to anodized B (1.15 and 1.29 µm) (Fig 1).

thumbnail
Table 1. Surface roughness parameter measurement in Arithmetric Mean Roughness (Ra) and Root mean Square Roughness (Sa), different small letters (alphabet) indicated a significant difference between material types.

https://doi.org/10.1371/journal.pone.0336639.t001

thumbnail
Fig 1. The surface roughness (Ra) and root mean square surface roughness (Sa) values of different titanium surface treatments.

https://doi.org/10.1371/journal.pone.0336639.g001

The SEM images (Fig 2) revealed that machined surface had relatively uniform, smooth texture characterized by distinct machining lines. SLA surface showed particularly irregular and textured surface morphology. Both anodized surfaces (A and B) exhibited similar rough and pore-filled textures. However, the anodized A surface had deeper valleys, which appeared as pink region in Fig 2.

thumbnail
Fig 2. The SEM image of surface topographic of different titanium surfaces (magnification X5000) and Three-dimensional profile images of different titanium surfaces topographic at X50 magnification, Machined surface (A), Sandblasted and acid-etched surface (B), Anodized A surface (C), Anodized B surface (D). Scale bars = 10 µm.

https://doi.org/10.1371/journal.pone.0336639.g002

The surface wettability of different titanium surface treatments was assessed through contact angle measurements, as depicted in Fig 3. Results demonstrated statistically significant differeces in contact angles among all surface treatments (p < 0.05). The machined surface showed the highest contact angle of 65.9 degrees, indicating the lowest wettability. The SLA-treated surface exhibited a moderately lower contact angle at 57.7 degrees in comparison. The anodized A surface had a notably lower contact angle of 20.28 degrees, while the anodized B surface recorded the lowest contact angle of 15.08 degrees, reflecting the highest level of wettability among the samples.

thumbnail
Fig 3. The average measurement of the contact angle on various titanium surfaces.

Machined surface (A), Sandblasted and acid-etched surface (B), Anodized A surface (C), Anodized B surface (D), Signification level at p < 0.05 (*).

https://doi.org/10.1371/journal.pone.0336639.g003

Behavior of osteoblastic and fibroblast cells on titanium surface analysis

Cell cytotoxicity and proliferation

It is observed that both the anodized A and B surfaces showed an increasing proliferation rate when compared to other titanium surfaces (Fig 4). On day 1, the proliferation of the osteoblast cells on the anodized A titanium surface was statistically significantly higher than on all other titanium surfaces (p < 0.05). By day 3, the proliferation rate of both anodized A and B surfaces were statistically significantly higher than those of the machine surface and the sandblasted and acid-etched surface (p < 0.05). However, no statistically significant difference was found between the anodized A and the anodized B groups (p < 0.05). By day 7, there were no significant differences in cell proliferation among test groups, as initial advantages of the anodized A surface diminished overtime. Across time point, significant proliferation changes were found in machined surface titanium between every tested interval (day1 and 3, day 3 and 7, day 1 and 7) The SLA surface exhibited significant differences in proliferation between days 3 and 7 and days 1 and 7 only. Anodized A demonstrated significant changes solely between days 1 and 7, while Anodized B showed significant differences proliferation across all intervals tested.

thumbnail
Fig 4. Osteoblast cell viability and proliferation assay after 1, 3, and 7 days.

Signification level at p < 0.05 (*), p < 0.001 (**), p < 0.005 (***).

https://doi.org/10.1371/journal.pone.0336639.g004

Immunofluorescent staining

At the 1-hour mark (Fig 5), both anodized A and anodized B surfaces recruited a higher number of osteoblast cells compared to machined and SLA surfaces. Osteoblasts on these anodized surfaces displayed larger sizes, and the focal adhesion molecules (indicated by green fluorescence) were more abundant than those on machined and SLA surfaces. Co-localization of actin and vinculin at the cell borders, appearing orange, was the most prominent on anodized B surface. By the 3-hour time point, osteoblast cells had extended their cytoskeletons and transitioned from a round morphology to polygonal shapes. The highest number of cells were found on anodized A and B surfaces, followed by the SLA and machined surfaces. Co-localization of actin and vinculin was again most abundant on the anodized B surface.

thumbnail
Fig 5. Immunofluorescent staining of osteoblast and gingival fibroblast cell for 1 and 3 hours.

(magnification X20), The focal adhesion molecule of the cells was identified by vinculin staining (green color). The actin filaments in the cells were highlighted by the phalloidin staining (red color). The nucleus of the cells was visible through the use of DAPI (blue color). Scale bars = 50 µm.

https://doi.org/10.1371/journal.pone.0336639.g005

From Fig 6, a statistically significant difference in the number of gingival fibroblast cells was observed between the anodized titanium A, B, and the machined surfaces compared to the SLA surface. The SLA surfaces contained the lowest number of fibroblast cells. Both the polystyrene coverslip and machined surface demonstrated increased cytoskeletal structure and fibroblast cell sizes. Although the anodized A and B surfaces had smaller cell sizes compared to the machined surface, the cytoskeletal distribution and polygonal form were still visible, in contrast to the circular shape and lack of cytoskeletal distribution observed on SLA surfaces.

thumbnail
Fig 6. Cell amount of osteoblast and fibroblast cells on each type of titanium surface at 1 hour and 3 hours.

Signification level at p < 0.05 (*), p < 0.001 (**), p < 0.005 (***).

https://doi.org/10.1371/journal.pone.0336639.g006

Alkaline phosphatase

On day 3, there was no statistically significant difference in the alkaline phosphatase levels observed among various surfaces (Fig 7), although the anodized B surface exhibited the highest level among those tested. By day 7, alkaline phosphatase levels on both anodized A and B surfaces were statistically significantly higher than those measured on the machined surface (control). There was no statistically significant difference in alkaline phosphatase levels between the anodized A, anodized B, and the sandblasted/acid-etched surfaces. When comparing ALP activity over time, statistically significant differences were found at both days 3 and 7 for all titanium surface types.

thumbnail
Fig 7. Alkaline phosphatase activity after 3, and 7 days (left), Alizarin Red test after 14, and 21 days (right), Signification level at p < 0.05 (*), p < 0.001 (**), p < 0.005 (***), p < 0.0001(****).

https://doi.org/10.1371/journal.pone.0336639.g007

Alizarin red

At day 14, alizarin red staining did not differ significantly between groups, though the anodized B showed a slightly increase (Fig 7). By day 21, anodized B exhibited the highest alizarin red levels, indicating greater mineralization compared to other surfaces. Alizarin red levels on both anodized A and B surfaces were statistically significantly higher than those on the machined surface (control). Additionally, the anodized B surface showed a statistically significantly higher alizarin red level compared to SLA surfaces but there was no significant difference between anodized A and B.

In time-dependent comparison, Alizarin Red staining demonstrated statistically significant differences at both days 14 and 21 among all examined titanium surface categories.

Discussion

Grade IV titanium was chosen for its higher purity, resulting in fewer allergic reactions and better corrosion resistance than Grade V due to the absence of aluminum and vanadium, which supports long-term biocompatibility [40,41]. Previous implant studies have also used Grade IV, making it suitable choice for this experiment [13,42,43]. The anodization process used for anodized A and B surfaces followed sandblasting and acid etching, resulting in a significant increase in surface roughness. This modification created a more pronounced valley and peak topography compared to SLA surfaces, as demonstrated in Fig 2. The roughness of the anodized A and B surfaces, ranging from 1.12 to 1.19 µm, aligns with previous findings that highlight this range as optimal for promoting osteoblast responses [12]. Research indicates that higher anodizing voltages produce rougher surfaces, characterized by numerous and larger micropores, increased crystallinity and a thicker oxide layer. These features collectively enhance osseointegration, as they provide a more suitable environment for bone cell attachment and growth [4446]. Additionally, the anodized surfaces exhibited the lowest contact angle among different surface treatments tested, indicating superior wettability. The decreased contact angle is directly associated with increased hydrophilicity. Enhanced wettability facilitates better protein adsorption and cell adhesion, thereby promoting cell proliferation and differentiation [47].

This study employed human osteoblast and gingival fibroblast cells to closely replicate realistic cell activities. Previous studies consistently indicate a superior cell response on anodized titanium surfaces compared to machined surfaces [46,48,49]. However, the optimal voltage for promoting osteoblast adhesion, proliferation, and mineralization remains a point of contention in the literature. Some studies have reported that higher voltages yield better cell responses [50,51], whereas others have found no significant differences across various voltages [52,53]. Ercan et al. noted a peak in osteoblast density at 15 volts on day 5 [51]. In this study, we selected lower voltages (20–30 volts) for anodized A and higher voltages (60–70 volts) for anodized B to assess cell proliferation at day 1, 3, 7. Two anodization voltage ranges (20–30 V and 60–70 V) were selected to represent the typical lower and upper limits for forming titanium nanotubular. These ranges, commonly used in research and compatible with available equipment, have also been employed in commercial implant systems (Nobel Biocare and MegaGen) aligning with reported cellular responses [51,54]. The results demonstrated that both anodized A and B surfaces exhibited higher cell growth and cytocompatibility for osteogenesis compared to other titanium surfaces over a 7-day period. These findings align with several studies that have reported enhanced proliferation rates on anodized surfaces compared to machined surface [51,52]. At day 7, cells demonstrated greater adhesion and more extensive spreading on the titanium surface compared to days 1 and 3. Such enhanced attachment and morphology of osteoblasts are favorable for improving implant stability during the healing phase. This finding further supports the clinical concept of immediate loading implants, where rapid and stable cell adhesion to the implant surface is essential for early osseointegration. However, there was no significant difference in osteoblast cells response between the two anodized surfaces. No significant differences in cell adhesion, proliferation, or differentiation were observed between the two anodization voltages for most assays, except for the MTT assay on day 1. This may be explained by a plateau effect, where nanotube morphology and surface chemistry within the tested range were sufficiently similar to elicit comparable cellular responses. It appears that osteoblasts and fibroblasts respond primarily to the presence of nanotubular structures rather than subtle variations in tube diameter. These findings suggest that the general nanotubular architecture is more critical for biological outcomes than precise anodization voltage.

The initial reaction at the tissue-implant interface is cell adhesion. Cell expansion and adherence were detected through the interaction between vinculin (green fluorescence) and actin (red fluorescence), which appeared as an orange color in the immunofluorescent images (Fig 5). The force generated by the expanded actin stress fibers, resulting from the partial coupling of vinculin and talin to actin binding, is known as the “actomyosin force machinery” [55]. Vinculin plays an important role in regulate cell adhesion by connecting with the actin cytoskeleton and stabilizing focal adhesions. Differentiating osteoblasts are often characterized by prominent stress fiber bundles along their borders [56,57]. These findings suggest that vinculin is a key indicator of cellular response and interaction with titanium surfaces over time. Our results correspond with those of previous studies [50,58], showing that the bright orange and green fluorescence (indicating vinculin) on anodized A and B surfaces signifies enhanced cell adherence, extension, and differentiation. Immunofluorescent images demonstrated a superior cell response on the anodized titanium surfaces, which was consistent with SEM observations indicating improved osteoblast cell adhesion on anodized surfaces over time (1–3 hours). The hydrophilic nature of the anodized surfaces (characterized by low contact angles) facilitated better osteoblast proliferation, differentiation, and adhesion as evidenced by the presence of filopodia protrusions and microvilli extension.

Normally, fibroblast cells showed good adherence to the machine surface compared to the rough surface (SLA surface) [59,60]. However, a lot of research suggests a microrough surface to promote gingival fibroblast cell attachment [28,61,62]. In this study, the anodized A and B surfaces demonstrated superior cell attachment and spreading compared to the SLA surfaces. Notably, the anodized A surface exhibited the highest fibroblast cell attachment, followed by anodized B, machine surfaces, and the least cell attachment was SLA surfaces respectively. The modified anodized titanium surface (sandblasted before anodization) employed in this work showed results consistent with those of Xu et al., indicating that anodized titanium surfaces improve fibroblast cell responsiveness [63]. Furthermore, several studies have confirmed that fibroblast cells attach more effectively to anodized surfaces in comparison to the control surfaces [29,53,64,65]. Gulati et al. also showed that fibroblast cells exhibit longer filopodia spread and improved cell alignment on the anodized surfaces treated at higher voltages. Additionally, these studies noted and increase in integrinβ1 expression, a receptor that mediates cellular binding onto the implant surface [53]. Figs 4 and 5 of this study further corroborate these findings through extensive cell spreading and organized filopodia structures observed via immunofluorescent imaging. The enhanced fibroblast cell response on the anodized surfaces, illustrated by the growing filopodia, aligns with findings from prior studies. Specifically, vinculin-actin interaction observed in fibroblast cells appeared as orange color in Fig 5, which was most abundant at the borders of the anodized A surface, indicating the highest fibroblast cell adhesion. Vinculin, a crucial component of focal adhesions and cell mechanosensory mechanisms, plays a key role in mediating integrin functions that influence cell motility and behavior [66]. This detailed cellular interaction is critical for the formation of a strong extracellular matrix, thereby enhancing tissue integration and healing. The anodized titanium surface facilitates the gingival fibroblasts adhesion, deposition, and proliferation, which forms a robust bond potentially capable of preventing microbial invasion [6,53,67]. It is noteworthy that human gingival fibroblast response, as observed through immunofluorescence, is superior on the anodized surface compared to the SLA surface, a finding that aligns with the results of Lima et al [29]. The reason is the anodized surfaces had porosities, the TiO2 oxide layer, and the highest wettability, which is favorable for fibroblast deposition, adhesion, and proliferation [29,68].

ALP activity was used to assess initial osteoblast differentiation, while alizarin red staining was employed to evaluate mineralized nodule formation on titanium surfaces. In this study, ALP activity and alizarin red staining were highest on anodized B surfaces, likely influenced by the applied voltage and the thickness of the oxide layer. The hydrophilicity of the anodized surfaces and the presence of the Ti-OH groups on the surface layer likely promoted this enhanced response [47]. When Ca2+ ions combine with OH groups, the surface gains a positive charge and reacts with phosphate groups to form amorphous calcium phosphate [69]. This observation is supported by Erdogan et al, who found that anodization at the higher voltages resulted in the best alizarin red measurements at 2 weeks, although no significant differences were noted between different anodized voltages [52]. Consequently, increasing anodization voltage appears to enhance oxide layer thickness, improving surface wettability, promoting cell proliferation, spreading, ALP activation, and calcium deposition. In this experiment, although anodized B surfaces exhibited lower contact angles compared to anodized A surfaces, no statistically significant differences were found between ALP activity and alizarin red staining of the anodized A and B surfaces (p < 0.05). Fig 3, 6 and 7 indicate that there is no notable difference in bone formation between machined titanium and SLA surfaces. Typically, fibrin from the blood clots helps osteoblasts adhere under normal physiological conditions. However, because our experiment did not included fibrin, osteoblast attachment to the SLA surface was reduced. On the other hand, the anodized surface’s higher hydrophilicity enhanced cell adhesion, as demonstrated by the immunofluorescent images leading to increase bone formation in line with previous studies [21].

Anodized titanium surfaces show enhanced biocompatibility due to a combination of physicochemical and mechanical properties which enhance surface wettability and promote cell spreading and cell proliferation. From the data present in this study, human osteoblast cells and fibroblast cells respond better on anodized surfaces than on SLA and machined. However, the mechanism of human osteoblast cell and fibroblast cell behavior on the anodized titanium at the molecular level needs further investigation.

Conclusion

The study found that modified anodized titanium surfaces substantially improve osteoblast and fibroblast attachment, proliferation, and differentiation when compared to machined or SLA surfaces. Enhanced surface roughness, porosity and hydrophilicity were identified as key contributors to superior cell responses and stronger tissue integration. These attributes render modified anodized surfaces particularly suitable for applications requiring robust bone and soft tissue integration, such as dental implants. It remains challenging to distinguish the individual effects of surface roughness and hydrophilicity due to the substrates employed; thus, the enhanced cellular response should be regards as a consequence of combined surface features. Notably, high-voltage anodization promotes osteoblast adhesion, whereas low-voltage anodization is more conducive to fibroblast adhesion. Future studies is warrented to optimizie anodization parameters for further improvement in cell responses and to investigate the broader clinical implications of these findings.

Acknowledgments

This study was supported by the Prosthodontic Department, Faculty of Dentistry, Chulalongkorn University. Center of Excellence for Dental Stem Cell Biology, Chulalongkorn University. Oral Biology Research Center, Chulalongkorn University. Dental Material R&D Center, Chulalongkorn University. We thank Dr. Prasit Pavasant for technical assistance.

References

  1. 1. Pandey C, Rokaya D, Bhattarai BP. Contemporary Concepts in Osseointegration of Dental Implants: A Review. Biomed Res Int. 2022;2022:6170452. pmid:35747499
  2. 2. Buser D, Mericske-Stern R, Bernard JP, Behneke A, Behneke N, Hirt HP, et al. Long-term evaluation of non-submerged ITI implants. Part 1: 8-year life table analysis of a prospective multi-center study with 2359 implants. Clin Oral Implants Res. 1997;8(3):161–72. pmid:9586460
  3. 3. Javed F, Ahmed HB, Crespi R, Romanos GE. Role of primary stability for successful osseointegration of dental implants: Factors of influence and evaluation. Interv Med Appl Sci. 2013;5(4):162–7. pmid:24381734
  4. 4. Abrahamsson I, Berglundh T, Linder E, Lang NP, Lindhe J. Early bone formation adjacent to rough and turned endosseous implant surfaces. An experimental study in the dog. Clin Oral Implants Res. 2004;15(4):381–92. pmid:15248872
  5. 5. Rosa MB, Albrektsson T, Francischone CE, Schwartz Filho HO, Wennerberg A. The influence of surface treatment on the implant roughness pattern. J Appl Oral Sci. 2012;20(5):550–5. pmid:23138742
  6. 6. Smeets R, Stadlinger B, Schwarz F, Beck-Broichsitter B, Jung O, Precht C, et al. Impact of Dental Implant Surface Modifications on Osseointegration. Biomed Res Int. 2016;2016:6285620. pmid:27478833
  7. 7. Tarnow DP, Emtiaz S, Classi A. Immediate loading of threaded implants at stage 1 surgery in edentulous arches: ten consecutive case reports with 1- to 5-year data. Int J Oral Maxillofac Implants. 1997;12(3):319–24. pmid:9197096
  8. 8. Wennerberg A, Albrektsson T. On implant surfaces: a review of current knowledge and opinions. Int J Oral Maxillofac Implants. 2010;25(1):63–74. pmid:20209188
  9. 9. Le Guéhennec L, Soueidan A, Layrolle P, Amouriq Y. Surface treatments of titanium dental implants for rapid osseointegration. Dent Mater. 2007;23(7):844–54. pmid:16904738
  10. 10. Andrukhov O, Huber R, Shi B, Berner S, Rausch-Fan X, Moritz A, et al. Proliferation, behavior, and differentiation of osteoblasts on surfaces of different microroughness. Dent Mater. 2016;32(11):1374–84. pmid:27637551
  11. 11. Feller L, Jadwat Y, Khammissa RAG, Meyerov R, Schechter I, Lemmer J. Cellular responses evoked by different surface characteristics of intraosseous titanium implants. Biomed Res Int. 2015;2015:171945. pmid:25767803
  12. 12. Wennerberg A, Albrektsson T. Suggested guidelines for the topographic evaluation of implant surfaces. Int J Oral Maxillofac Implants. 2000;15(3):331–44. pmid:10874798
  13. 13. Zhu X, Chen J, Scheideler L, Reichl R, Geis-Gerstorfer J. Effects of topography and composition of titanium surface oxides on osteoblast responses. Biomaterials. 2004;25(18):4087–103. pmid:15046900
  14. 14. Yang Y, Oh N, Liu Y, Chen W, Oh S, Appleford M, et al. Enhancing osseointegration using surface-modified titanium implants. JOM. 2006;58(7):71–6.
  15. 15. Sul Y-T, Byon E, Wennerberg A. Surface characteristics of electrochemically oxidized implants and acid-etched implants: surface chemistry, morphology, pore configurations, oxide thickness, crystal structure, and roughness. Int J Oral Maxillofac Implants. 2008;23(4):631–40. pmid:18807558
  16. 16. Karambakhsh A, Afshar A, Ghahramani S, Malekinejad P. Pure Commercial Titanium Color Anodizing and Corrosion Resistance. J of Materi Eng and Perform. 2011;20(9):1690–6.
  17. 17. Karambakhsh A, Afshar A, Malekinejad P. Corrosion Resistance and Color Properties of Anodized Ti-6Al-4V. J of Materi Eng and Perform. 2010;21(1):121–7.
  18. 18. Wadhwani C, Brindis M, Kattadiyil MT, O’Brien R, Chung K-H. Colorizing titanium-6aluminum-4vanadium alloy using electrochemical anodization: Developing a color chart. J Prosthet Dent. 2018;119(1):26–8. pmid:28506656
  19. 19. Charrière R, Lacaille G, Pedeferri MP, Faucheu J, Delafosse D. Characterization of the gonioapparent character of colored anodized titanium surfaces. Color Research & Application. 2014;40(5):483–90.
  20. 20. Sharma A, McQuillan AJ, Sharma LA, Waddell JN, Shibata Y, Duncan WJ. Spark anodization of titanium-zirconium alloy: surface characterization and bioactivity assessment. J Mater Sci Mater Med. 2015;26(8):221. pmid:26260697
  21. 21. Traini T, Murmura G, Sinjari B, Perfetti G, Scarano A, D’Arcangelo C, et al. The Surface Anodization of Titanium Dental Implants Improves Blood Clot Formation Followed by Osseointegration. Coatings. 2018;8(7):252.
  22. 22. Wennerberg A, Jimbo R, Stübinger S, Obrecht M, Dard M, Berner S. Nanostructures and hydrophilicity influence osseointegration: a biomechanical study in the rabbit tibia. Clin Oral Implants Res. 2014;25(9):1041–50. pmid:23782316
  23. 23. Zitzmann NU, Berglundh T. Definition and prevalence of peri-implant diseases. J Clin Periodontol. 2008;35(8 Suppl):286–91. pmid:18724856
  24. 24. John V, Shin D, Marlow A, Hamada Y. Peri-Implant Bone Loss and Peri-Implantitis: A Report of Three Cases and Review of the Literature. Case Rep Dent. 2016;2016:2491714. pmid:27833766
  25. 25. Corvino E, Pesce P, Mura R, Marcano E, Canullo L. Influence of Modified Titanium Abutment Surface on Peri-implant Soft Tissue Behavior: A Systematic Review of In Vitro Studies. Int J Oral Maxillofac Implants. 2020;35(3):503–19. pmid:32406646
  26. 26. Kim J-J, Lee J-H, Kim JC, Lee J-B, Yeo I-SL. Biological Responses to the Transitional Area of Dental Implants: Material- and Structure-Dependent Responses of Peri-Implant Tissue to Abutments. Materials (Basel). 2019;13(1):72. pmid:31877848
  27. 27. Schupbach P, Glauser R. The defense architecture of the human periimplant mucosa: a histological study. J Prosthet Dent. 2007;97(6 Suppl):S15-25. pmid:17618929
  28. 28. Canullo L, Annunziata M, Pesce P, Tommasato G, Nastri L, Guida L. Influence of abutment material and modifications on peri-implant soft-tissue attachment: A systematic review and meta-analysis of histological animal studies. J Prosthet Dent. 2021;125(3):426–36. pmid:32197823
  29. 29. Lima JHCD, Robbs PCM, Tude EMO, De Aza PN, Costa EMD, Scarano A, et al. Fibroblasts and osteoblasts behavior after contact with different titanium surfaces used as implant abutment: An in vitro experimental study. Heliyon. 2024;10(3):e25038. pmid:38322837
  30. 30. Genova T, Chinigò G, Munaron L, Rivolo P, Luganini A, Gribaudo G, et al. Bacterial and Cellular Response to Yellow-Shaded Surface Modifications for Dental Implant Abutments. Biomolecules. 2022;12(11):1718. pmid:36421732
  31. 31. Schmidlin PR, Müller P, Attin T, Wieland M, Hofer D, Guggenheim B. Polyspecies biofilm formation on implant surfaces with different surface characteristics. J Appl Oral Sci. 2013;21(1):48–55. pmid:23559112
  32. 32. Crawford RJ, Webb HK, Truong VK, Hasan J, Ivanova EP. Surface topographical factors influencing bacterial attachment. Adv Colloid Interface Sci. 2012;179–182:142–9. pmid:22841530
  33. 33. Berglundh T, Gotfredsen K, Zitzmann NU, Lang NP, Lindhe J. Spontaneous progression of ligature induced peri-implantitis at implants with different surface roughness: an experimental study in dogs. Clin Oral Implants Res. 2007;18(5):655–61. pmid:17608738
  34. 34. Del Curto B, Brunella MF, Giordano C, Pedeferri MP, Valtulina V, Visai L, et al. Decreased bacterial adhesion to surface-treated titanium. Int J Artif Organs. 2005;28(7):718–30. pmid:16049906
  35. 35. Caous JS, Lövenklev M, Fäldt J, Langton M. Adhesion of Streptococcus mitis and Actinomyces oris in co-culture to machined and anodized titanium surfaces as affected by atmosphere and pH. BMC Oral Health. 2013;13:4. pmid:23298213
  36. 36. Standardization IOf. ISO 5832-2:2018(en) Implants for surgery — Metallic materials — Part 2: Unalloyed titanium. 2018.
  37. 37. Schuster JM, Schvezov CE, Rosenberger MR. Influence of Experimental Variables on the Measure of Contact Angle in Metals Using the Sessile Drop Method. Procedia Materials Science. 2015;8:742–51.
  38. 38. Choi HS, Kim JW, Cha Y-N, Kim C. A quantitative nitroblue tetrazolium assay for determining intracellular superoxide anion production in phagocytic cells. J Immunoassay Immunochem. 2006;27(1):31–44. pmid:16450867
  39. 39. Rook GA, Steele J, Umar S, Dockrell HM. A simple method for the solubilisation of reduced NBT, and its use as a colorimetric assay for activation of human macrophages by gamma-interferon. J Immunol Methods. 1985;82(1):161–7. pmid:3928762
  40. 40. Vora HD, Shanker Rajamure R, Dahotre SN, Ho Y-H, Banerjee R, Dahotre NB. Integrated experimental and theoretical approach for corrosion and wear evaluation of laser surface nitrided, Ti-6Al-4V biomaterial in physiological solution. J Mech Behav Biomed Mater. 2014;37:153–64. pmid:24922620
  41. 41. Koike M, Lockwood PE, Wataha JC, Okabe T. Initial cytotoxicity of novel titanium alloys. J Biomed Mater Res B Appl Biomater. 2007;83(2):327–31. pmid:17385227
  42. 42. Cordeiro JM, Barão VAR. Is there scientific evidence favoring the substitution of commercially pure titanium with titanium alloys for the manufacture of dental implants?. Mater Sci Eng C Mater Biol Appl. 2017;71:1201–15. pmid:27987677
  43. 43. Jain S, Williamson RS, Janorkar AV, Griggs JA, Roach MD. Osteoblast response to nanostructured and phosphorus-enhanced titanium anodization surfaces. J Biomater Appl. 2019;34(3):419–30. pmid:31126206
  44. 44. Zhu X, Kim KH, Jeong Y. Anodic oxide films containing Ca and P of titanium biomaterial. Biomaterials. 2001;22(16):2199–206. pmid:11456059
  45. 45. Sul Y-T, Johansson CB, Jeong Y, Wennerberg A, Albrektsson T. Resonance frequency and removal torque analysis of implants with turned and anodized surface oxides. Clin Oral Implants Res. 2002;13(3):252–9. pmid:12010155
  46. 46. Kim M-H, Park K, Choi K-H, Kim S-H, Kim SE, Jeong C-M, et al. Cell adhesion and in vivo osseointegration of sandblasted/acid etched/anodized dental implants. Int J Mol Sci. 2015;16(5):10324–36. pmid:25955650
  47. 47. Zhao G, Schwartz Z, Wieland M, Rupp F, Geis-Gerstorfer J, Cochran DL, et al. High surface energy enhances cell response to titanium substrate microstructure. J Biomed Mater Res A. 2005;74(1):49–58. pmid:15924300
  48. 48. De Angelis E, Ravanetti F, Cacchioli A, Corradi A, Giordano C, Candiani G, et al. Attachment, proliferation and osteogenic response of osteoblast-like cells cultured on titanium treated by a novel multiphase anodic spark deposition process. J Biomed Mater Res B Appl Biomater. 2009;88(1):280–9. pmid:18688796
  49. 49. Sista S, Nouri A, Li Y, Wen C, Hodgson PD, Pande G. Cell biological responses of osteoblasts on anodized nanotubular surface of a titanium-zirconium alloy. J Biomed Mater Res A. 2013;101(12):3416–30. pmid:23559548
  50. 50. Voltrova B, Hybasek V, Blahnova V, Sepitka J, Lukasova V, Vocetkova K, et al. Different diameters of titanium dioxide nanotubes modulate Saos-2 osteoblast-like cell adhesion and osteogenic differentiation and nanomechanical properties of the surface. RSC Adv. 2019;9(20):11341–55. pmid:35520235
  51. 51. Ercan B, Webster TJ. Greater osteoblast proliferation on anodized nanotubular titanium upon electrical stimulation. Int J Nanomedicine. 2008;3(4):477–85. pmid:19337416
  52. 52. Erdogan YK, Ercan B. Anodized Nanostructured 316L Stainless Steel Enhances Osteoblast Functions and Exhibits Anti-Fouling Properties. ACS Biomater Sci Eng. 2023;9(2):693–704. pmid:36692948
  53. 53. Gulati K, Moon H-J, Kumar PTS, Han P, Ivanovski S. Anodized anisotropic titanium surfaces for enhanced guidance of gingival fibroblasts. Mater Sci Eng C Mater Biol Appl. 2020;112:110860. pmid:32409032
  54. 54. Michalska-Domańska M, Nyga P, Czerwiński M. Ethanol-based electrolyte for nanotubular anodic TiO2 formation. Corrosion Science. 2018;134:99–102.
  55. 55. Hu K, Ji L, Applegate KT, Danuser G, Waterman-Storer CM. Differential transmission of actin motion within focal adhesions. Science. 2007;315(5808):111–5. pmid:17204653
  56. 56. Rodríguez JP, González M, Ríos S, Cambiazo V. Cytoskeletal organization of human mesenchymal stem cells (MSC) changes during their osteogenic differentiation. J Cell Biochem. 2004;93(4):721–31. pmid:15660416
  57. 57. Kilian KA, Bugarija B, Lahn BT, Mrksich M. Geometric cues for directing the differentiation of mesenchymal stem cells. Proc Natl Acad Sci U S A. 2010;107(11):4872–7. pmid:20194780
  58. 58. Das K, Bose S, Bandyopadhyay A. Surface modifications and cell-materials interactions with anodized Ti. Acta Biomater. 2007;3(4):573–85. pmid:17320494
  59. 59. Atsuta I, Ayukawa Y, Furuhashi A, Ogino Y, Moriyama Y, Tsukiyama Y, et al. In vivo and in vitro studies of epithelial cell behavior around titanium implants with machined and rough surfaces. Clin Implant Dent Relat Res. 2014;16(5):772–81. pmid:23448501
  60. 60. An N, Rausch-fan X, Wieland M, Matejka M, Andrukhov O, Schedle A. Initial attachment, subsequent cell proliferation/viability and gene expression of epithelial cells related to attachment and wound healing in response to different titanium surfaces. Dent Mater. 2012;28(12):1207–14. pmid:23083807
  61. 61. Teng F-Y, Ko C-L, Kuo H-N, Hu J-J, Lin J-H, Lou C-W, et al. A comparison of epithelial cells, fibroblasts, and osteoblasts in dental implant titanium topographies. Bioinorg Chem Appl. 2012;2012:687291. pmid:22287942
  62. 62. Grössner-Schreiber B, Herzog M, Hedderich J, Dück A, Hannig M, Griepentrog M. Focal adhesion contact formation by fibroblasts cultured on surface-modified dental implants: an in vitro study. Clin Oral Implants Res. 2006;17(6):736–45. pmid:17092235
  63. 63. Xu R, Hu X, Yu X, Wan S, Wu F, Ouyang J, et al. Micro-/nano-topography of selective laser melting titanium enhances adhesion and proliferation and regulates adhesion-related gene expressions of human gingival fibroblasts and human gingival epithelial cells. Int J Nanomedicine. 2018;13:5045–57. pmid:30233172
  64. 64. Ma Q, Wang W, Chu PK, Mei S, Ji K, Jin L, et al. Concentration- and time-dependent response of human gingival fibroblasts to fibroblast growth factor 2 immobilized on titanium dental implants. Int J Nanomedicine. 2012;7:1965–76. pmid:22619534
  65. 65. Ferrà-Cañellas MDM, Llopis-Grimalt MA, Monjo M, Ramis JM. Tuning Nanopore Diameter of Titanium Surfaces to Improve Human Gingival Fibroblast Response. Int J Mol Sci. 2018;19(10):2881. pmid:30249013
  66. 66. Wozniak MA, Modzelewska K, Kwong L, Keely PJ. Focal adhesion regulation of cell behavior. Biochim Biophys Acta. 2004;1692(2–3):103–19. pmid:15246682
  67. 67. Ivanoff C-J, Widmark G, Johansson C, Wennerberg A. Histologic evaluation of bone response to oxidized and turned titanium micro-implants in human jawbone. Int J Oral Maxillofac Implants. 2003;18(3):341–8. pmid:12814308
  68. 68. Miura S, Takebe J. Biological behavior of fibroblast-like cells cultured on anodized-hydrothermally treated titanium with a nanotopographic surface structure. J Prosthodont Res. 2012;56(3):178–86. pmid:22264674
  69. 69. Brunette DM, Tengvall P, Textor M, Thomsen P. Titanium in medicine: material science, surface science, engineering, biological responses and medical applications. Springer. 2001.