Abstract
Background
Partial range of field (PROF) intraocular lenses (IOLs) provide good distance visual acuity with enhanced intermediate visual acuity. Neuroadaptation to new intraocular lenses can take several months. This study investigates visual and patient-reported outcomes (PROMs) in patients implanted with the PROF IOL ICB00 (Johnson & Johnson Vision) 18 months post-surgery. A particular focus was put on spectacle independence while performing activities of daily life.
Methods
This ambispective, single-center study included 41 patients (aged 48–84). Visual acuities (distance, intermediate), refraction, and PROMs were assessed ≥18 months post-surgery. Spectacle independence was evaluated using the PRSIQ questionnaire. Patients also self-reported on visual quality, task performance, and photic phenomena.
Results
Binocular visual acuity of 0.20 logMAR or better was achieved by 100.0% of patients for uncorrected or corrected distance, by 73.2% for uncorrected intermediate, and by 79.5% for distance-corrected intermediate vision. Photic phenomena were reported by <10%. Mean visual quality scores were 1.68 ± 0.72 for distance and 2.05 ± 0.92 for intermediate vision, where 1 equaled very good and 6 equaled very poor. Complete spectacle independence was reported by 87.8% for distance and 53.7% for intermediate vision. Satisfaction rates were 90.2% (distance), 87.8% (intermediate), and 51.2% (near).
Citation: Latz C, Licht A, Ponto KA, Menzel-Severing J, Piñero DP, Mirshahi A (2025) Functional vision in daily life: Clinical and patient-reported outcomes 18 months after enhanced partial range of field IOL implantation. PLoS One 20(10): e0333174. https://doi.org/10.1371/journal.pone.0333174
Editor: Andrzej Grzybowski,, University of Warmia, POLAND
Received: July 4, 2025; Accepted: September 9, 2025; Published: October 7, 2025
Copyright: © 2025 Latz et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Data Availability: All relevant data are within the paper and its Supporting Information files.
Funding: This study was conducted as an Investigator Initiated Study and received support through an unrestricted research grant provided by Johnson & Johnson. The funder had no role in the design, execution, analysis, or interpretation of the study, nor any influence on the content of this article.
Competing interests: AM has received consulting and lecture fees from Ziemer Ophthalmic Systems AG and Alcon Deutschland GmbH. AL has received consulting fees from Bayer AG KP, JM-S, DP and CL have no financial interest to disclose. The authors have no proprietary or commercial interest in the medical devices that are described in this manuscript. This does not alter our adherence to PLOS ONE policies on sharing data and materials.
Background
In 2024, the ESCRS Functional Vision Working Group proposed a functional classification for presbyopia-correcting IOLs depending on two metrics:
- (1). the range of field (RoF) achieved in the monocular defocus curve with best correction at distance at 0.2 logMAR visual acuity (VA) level and
- (2). the improvement of VA from intermediate to near distance visual acuity (DVA) [1].
According to this classification the ICB00 (Johnson & Johnson Vision), is an enhanced PROF IOL [2]. This lens features an aspheric anterior and an aspherical posterior surface, with the anterior surface designed to have a continuous increase in power from the periphery to the center [3]. It extends the RoF from distance to > 1.2 to <1.58 diopters (D) (Fig 1). These types of IOLs are also referred as monofocal plus or enhanced monofocal IOLs. A limitation of this functional classification is that it solely describes the VA achieved along the RoF, omitting other significant functional outcomes such as dysphotopsia.
Although excellent DVA and enhanced uncorrected intermediate visual acuity (UIVA), and strong patient acceptance of enhanced monofocal IOLs have been extensively studied [2,4,5], – along with low rates of postoperative photic phenomena comparable to standard monofocal IOLs-, there is limited research on patient-reported outcomes (PROMs) in particular spectacle independence, which reflect patients’ real-world experiences and perceptions [3,6–8].
Given that neuroadaptation and habituation- key factors in patient satisfaction- can take six months or longer, it is particularly important to analyze these at a longer time interval [9]. Morlock et al. identified an important difference in the classification of calling oneself spectacle independent yet using correction for certain activities, which was also related to being able to function better with the correction. These perceived differences were classified as `need´, `wear´, and `function´ and led to the Patient Reported Spectacle Independence Questionnaire (PRSIQ) [10]. To understand spectacle independence and patient satisfaction after ICB00 implantation, visual acuity was tested at different distances, the PRSIQ as well as additional questions regarding photic phenomena and functionality in daily activities were evaluated 18 months after cataract surgery.
Methods
Aim, design and setting of the study
This ambispective, non-comparative, single-center study enrolled 41 patients who had undergone sequential, bilateral, uncomplicated cataract surgery with implantation of the ICB00 IOL and had a follow up period of 18 months or longer. Recruitment took place from October 17, 2023, to November 30, 2023, by contacting patients by phone or email. 95 consecutive patients were contacted of which 41 agreed to return to the clinic for additional testing. The last patient was examined on December 08, 2023. Inclusion criteria were corneal astigmatism < 0.75D, age of 45 years or older, and visually significant cataract. Exclusion criteria included systemic diseases that could alter study outcomes, prior ocular or refractive surgery, irregular astigmatism, zonular alterations affecting IOL position or stability, active ocular disease, retinal pathologies and severe glaucoma (defined as a mean deviation deficit > 12 dB on visual fields). Axial length independent of very short or very long eyes was not an exclusion criterion.
Before inclusion, all patients were thoroughly informed about the study´s nature, and written informed consent was obtained in accordance with the tenets of the Declaration of Helsinki. This study was approved by the medical ethics committee of the Medical Chamber of North-Rhine, Germany (CE # 2124770CE01).
Preoperative examination
The preoperative evaluation included uncorrected distance visual acuity (UDVA) and corrected distance visual acuity (CDVA), objective refraction by autorefractometry, optical biometry and keratometry (IOLMaster 700, Carl Zeiss Meditec), non-contact tonometry, slit lamp biomicroscopy, optical coherence tomography of the macula and optic disc (Carl Zeiss Meditec), and dilated fundus evaluation. The Barrett TK Universal II formula was used for IOL power calculation in all cases with a target of emmetropia, specifically a target value of plano or the first positive value were used to choose the IOL power.
Long-term follow-up visit
Eighteen months after cataract surgery, all patients were contacted by phone or email and scheduled for a long-term follow-up visit. At this visit, a complete visual evaluation was conducted, including monocular and binocular measurement of UDVA and CDVA, subjective refraction, uncorrected intermediate visual acuity (UIVA) and distance-corrected intermediate visual acuity (DCIVA) (measured at 66 cm). Near visual acuity was not measured. Since EDOF or monofocal plus IOLs tend to produce false negative values on autorefractometry only subjective refraction was used.
Spectacle independence Assessment was evaluated using the PRSIQ [10]. Patients rated the frequency of spectacle need, wear and strain without correction for various activities at near, intermediate and distance on a Likert scale (e.g., always, sometimes, never), enabling the calculation of a PRSIQ score, which is calculated as the average of all individual scores.
Quality of vision and photic phenomena were assessed using a self-developed questionnaire which is depicted in Fig 1. Patients were asked to quantify their distance and intermediate visual quality on a scale from 1 (very good) to 6 (very poor).
Additionally, patients were asked about the perception of halos, glare, blurring, and starbursts at distance and intermediate vision. Daily functionality without spectacles was assessed by asking if the dashboard while driving was clearly visible, and if screenwork and reading were possible without correction. `Yes´ was transposed to 1, `no´ to 6 and modifications like increasing font size to 3, in order to calculate a score similar to the PRSIQ score.
Surgery
All surgeries were performed by two experienced surgeons using a standard technique of suture-less microincision phacoemulsification [11]. A clear corneal incision with a width of 2.4 mm was manually placed, as were two paracenteses (1.0 mm). Care was taken to achieve a capsulorhexis diameter of approximately 5 mm to ensure complete coverage of the IOL optic with the anterior capsule.
Nuclear disassembly and cortical aspiration were performed using the Centurion vision systems by Alcon (Fort Worth, TX, USA). The IOL was delivered either under irrigation or viscoelastic protection with a shooter.
Statistical analysis
Data analysis was performed using the software SPSS version 22.0 for Windows (SPSS). Normality of all data distributions was initially evaluated by means of the Kolmogorov-Smirnov test. A descriptive analysis of all continuous variables was carried out, calculating the average values with their corresponding standard deviations and the ranges of maximum and minimum values. For categorical variables, frequencies of different conditions or aspects were determined. For bilateral comparisons, a p-value below 0.05 was considered statistically significant.
Results
Preoperative patient data are depicted in Table 1. The mean age of the study group was 69.4 years (SD: 9.0, median: 70.0, range: 48–84 years). 24 male patients (58.5%) and 17 female patients (41.5%). were enrolled. Dense cataracts were indicated by a LOCS III score of 4 and 3.5 for nuclear colour and opacity, respectively.
Table 2 summarizes the visual and refractive outcomes 18 months post-surgery: Mean binocular logMAR values were 0.05 ± 0.07 for UDVA, 0.18 ± 0.12 for UIVA, and 0.17 ± 0.11 for DCIVA. Binocular visual acuity of 0.20 logMAR or better was achieved by 100.0% of patients for UDVA or DCVA, by 73.2% for UIVA, and by 79.5% for DCIVA (Fig 1) Postoperative spherical equivalent (SE) was within ±0.50 D in 80.5% of right eyes and 82.9% of left eyes. Likewise, SE was within ±1.00 D in 95.1% of right eyes and 92.7% of left eyes. The PRSIQ and an additional questionnaire were used to assess spectacle independence and satisfaction. The distribution of these answers is depicted in Table 3. At distance, 4 out of 41 (9.8%) patients reported the perception of halos or glare; and 1 out of 41 (2.4%) blurry vision (Fig 2). At intermediate distance these perceptions decreased to 1 in 41 (2.4%) for halos and blurry vision, and to 2 in 41 (4.9%) for glare. Refractive error in this group had a higher spherical equivalent than in the non-halo seeing group, but this difference did not reach statistical significance (Table 4). Patients subjectively graded their visual quality at distance and intermediate vision on a scale from 1 (very good) to 6 (very poor). The mean distance and intermediate visual quality satisfaction scores were 1.68 (SD: 0.72; Median: 2.00; Range: 1–3) and 2.05 (SD: 0.92; Median: 2.00; Range: 1–4), respectively. No patient rated their satisfaction with distance or intermediate visual quality as 4 or worse. (Table 3). 95.1% of patients reported a clearly visible dashboard when driving, increasing to 97.5, when the dashboard was brightly illuminated. 45% of patients were able to perform screen work without spectacles, while 40% had to enlarge the font size. 63.4% of patients achieved spectacle-free reading when the font size was large enough. For distance vision, 87.8% of patients were spectacle free, while 53.7% did not need glasses for intermediate vision (Table 3 and Fig 3A). This was confirmed by the fact, that 82.5% of patients did not wear glasses for distance at any time, while 36.6% did not wear glasses for intermediate vision at any time. (Fig 3B). In contrast, 97.6% of patients required glasses for near vision activities (Fig 3A). In total, 90.2% of patients were completely, mostly, or moderately satisfied with their unaided distance vision. 87.8% were satisfied with intermediate vision, 51.2% with near vision and 90.0% with overall vision (Fig 3D). No adverse events were recorded during the follow-up period, and none of the patients developed posterior capsular opacification (PCO) requiring YAG capsulotomy (Fig 4).
Patients reported visual quality, photic perceptions and daily functionality. Scale from 1 (excellent) to 6 (extremely poor). `Yes´ was transposed to 1, `No´ to 6 and modifications like increasing font size to 3. English translation of prompts in red.
Yes, I perceive halos, glare or blurry vision at intermediate or distance vision.
`Glasses´ were specified as glasses, magnifiers or contacts.
Discussion
To the best of our knowledge this is the first study to analyze spectacle independence and patient satisfaction at long term after implantation of the ICB00. Our findings regarding excellent distant visual acuity and improved intermediate visual acuity align with those of other studies evaluating the same IOL [3,7,8,12–14]. as well as other models of enhanced monofocal IOLs [15,16]. For instance, Goslings et al. reported mean binocular UDVA of 0.11 ± 0.11, and UIVA of 0.12 ± 0.11 at three months post-implantation [7]. These results are slightly inferior to those observed in our study, which may be attributed to differences in the IOL calculation formula used, though this was not specified in their publication. Similarly, Giglio et al. reported mean postoperative binocular UDVA, UIVA and DCIVA values of −0.03 ± 0.07, 0.17 ± 0.12 and 0.13 ± 0.11, in 30 eyes using the Barrett TK Universal II formula [3]. Mencucci et al. also reported comparable mean binocular UDVA, UIVA and DCIVA values (0.03 ± 0.05 vs. 0.16 ± 0.10 vs. 0.15 ± 0.08) in 80 eyes of 40 patients using the Holladay 1 formula for axial lengths between 22.0 mm and 25.0 mm and Hoffer Q formula for axial lengths < 22.0 mm. Axial length measurements in their study were obtained using the IOL master 500 (Carl Zeiss Meditec AG) [14]. However, other studies have reported worse UDVA and better UIVA values in eyes implanted with the ICB00, particularly when a micro-monovision approach had been chosen or a trend toward a low myopic residual refractive error was observed [17]. Micro monovision is a common strategy to increase the range of field into the near zone, where the dominant eye is targeted towards emmetropia, and the other eye is targeted towards a myopic refractive error. Similarly, with-the-rule astigmatism can lead to a pseudo-accommodation and thereby leading to improved UNVA with enhanced monofocal lenses.
The PRSIQ questionnaire is a validated tool for assessing spectacle independence following cataract surgery [10]. In our study the PRSIQ was complemented with a questionnaire to assess not only satisfaction with uncorrected visual quality, but also ability to perform common intermediate distance tasks such as viewing the dashboard clearly while driving or reading from a computer screen or a handheld device. The grading scale was adapted from the German school system, which is widely recognized and familiar to patients. The scores for visual quality obtained from our questionnaire (1.68 and 2.05) were comparable to those calculated from the PRSIQ (1.71 and 2.22).
Using the PRSIQ, Stodulka and Pracharova investigated a similar IOL with a unique geometry that creates a power gradient from the centre to the periphery [18]. They found that over 80% of patients achieved spectacle independence for distance and intermediate vision with this EDOF IOL.
Similarly, in our study, intermediate vision with the ICB00 achieved scores indicating satisfactory spectacle independence, ranging between 2.05 and 2.85 on both the PRSIQ and the self-developed questionnaire. These outcomes are consistent with results from the validated Catquest 9SF questionnaire used in other studies evaluating the ICB00, where patients consistently perceived a benefit in intermediate vision compared to conventional monofocal IOLs [3,7,8]. Regarding near vision, despite its limitations, approximately half of the patients reported satisfaction with their near visual functionality. This may be attributed to the level of visual acuity provided, which allows for various near-vision activities without spectacle correction. Indeed, 63.4% of patients were able to read without correction if the font size was sufficiently large. This finding is supported by studies by Goslings et al. [7] and Giglio et al. [3], who used the Catquest 9SF questionnaire to assess difficulties in performing vision-related activities after ICB00 implantation and noted improvements in Rasch-calibrated scores for near vision tasks. Lopes et al. demonstrated significant differences between eyes implanted with a conventional monofocal IOL and those with the ICB00, with the latter group reporting less difficulty in tasks such as reading newspaper print and price-tags while shopping [8].
Interestingly, overall satisfaction with uncorrected vision was relatively high with a score of 1.83. This indicates that patients perceive excellent distance vision combined with satisfactory intermediate vision as a highly favourable outcome. This is particularly noteworthy given that our patient cohort included not only previously emmetropic or hyperopic individuals, but also myopic patients, as reflected in the pre-operative refractive errors detailed in Table 1. When correlating axial length and PROMs, there was a trend toward better outcomes in shorter eyes, which was statistically significant when exclusively looking at intermediate vision tasks (S1 File). Given a wide age range from 48 to 84 years in our group, the early presbyopic patients could form a bias in view of higher satisfaction rates and better neuroadaptation.
Photic phenomena were reported by less than 10% of the study group. The refractive error in this group was higher than in the non-reporting group. Since the reporting group was relatively small this difference in refractive error did not reach statistical significance. The fact that the perception of halos and glare decreased with intermediate vision in comparison to distance vision leads in our view to the explanation that a significant reason for the perception of halos and glare is the refractive error. The high number of patients who denied photic phenomena aligns with optical simulations showing that the enhanced monofocal IOL evaluated in this study produces fewer halos compared to other extended range of vision IOLs [19]. Lee and colleagues compared the ICB00 with a diffractive EDOF IOL and found that while spectacle independence was higher in the diffractive group, this came at the expense of more glare and halos [20]. Similarly, Corbelli et al. reported that the enhanced monofocal IOL had the advantage of reduced perception of halos and glare compared to a diffractive IOL [21].
How to measure and rank the effectiveness of presbyopia-correcting IOLs remains a very controversial topic. An important outcome is not only spectacle independence but also side effects such as photic phenomena and the cost of it. Presbyopia-correcting IOLs are more expensive than monofocal IOLs. Since prices vary regionally, it is impossible to give a valid answer to which IOL system offers the best cost-effectiveness ratio. Lan et al. have addressed this question by measuring on 194 patients in two Chinese centres the objective spectacle independence rate – as the proportion of patients with binocular UDVA, UIVA and UVNA all better than 0.1 logMAR-, costs, average cost-effectiveness ratios (ACERs, $/1% rate) [22].
In terms of effectiveness, full RoF IOLs (trifocal (93%) and bifocal (75%) were followed by EDOF with 67.9%. Monovision (14.3%) and monofocal strategies (7.4%) had the least favourable outcomes. In terms of ACERs the EDOF strategy [$72.85/1% rate (95% CI 52.02–93.70)] ranked between the diffractive bifocal [$69.06/1% rate (95% CI 30.89–107.21)], and the trifocal ($93.01/1% rate [95% CI 83.23–102.79]), while monovision [$136.83/1% rate (95% CI – 55.40 to 329.14)] showed less cost effectiveness [22]. Daka et al. have given an overview of seven reviews comparing the effectiveness of different presbyopia correcting IOLs. None of the reports analyzed cost-effectiveness. They conclude, that while spectacle independence in particular for UNVA is best with multifocal IOLs, a mini-monovision approach with nondiffractive EDOF or enhanced monofocal IOLs can be an option for patients concerned about visual disturbances from diffractive designs of trifocal IOLs [23].
Our study has several limitations, including the absence of a control group with a monofocal IOL, the lack of data on corneal spherical aberration in the study patients and post operative contrast sensitivity and defocus curves. We did not measure DCNVA nor UNVA, since the ICB00 is not designed to provide a clear focus in the near range. Additionally, the self-developed questionnaire was not validated through focus groups, cognitive interviews or pretesting, which limits its reliability and validity. However, the similarity on scores between the PRSIQ and the self-developed questionnaire for the question on visual quality serves as a partial validation. Given that neuroadaptation and habituation can take six months or longer [9], this study provides valuable insights into long-term patient-perceived outcomes. As the array of FROF and PROF IOLs continues to expand, long-term data are crucial for improving patient counselling.
In conclusion, the ICB00 provides excellent distance vision and satisfactory intermediate visual quality, resulting in high levels of spectacle independence and excellent long- term patient satisfaction with overall vision. Near visual outcomes, however, varied significantly among subjects and typically required spectacle correction. These findings underscore the ICB00´s potential as a reliable option for patients seeking improved distance. and intermediate vision.
Supporting information
S1 File. Analysis of axial length and PRSIQ and questionnaire-score.
P = PRSIQ; Q = questionnaire; intermediate includes all answers relevant to intermediate vision (i.e., dashboard visibility, computer); near includes all answers relevant to near vision (i.e., reading a book).
https://doi.org/10.1371/journal.pone.0333174.s001
(TIF)
S3 File. Conversion table decimal and LogMAR visual acuity.
https://doi.org/10.1371/journal.pone.0333174.s003
(TIF)
References
- 1. Ribeiro F, Dick HB, Kohnen T, Findl O, Nuijts R, Cochener B, et al. Evidence-based functional classification of simultaneous vision intraocular lenses: seeking a global consensus by the ESCRS Functional Vision Working Group. J Cataract Refract Surg. 2024;50(8):794–8. pmid:39083407
- 2. Fernández J, Rocha-de-Lossada C, Zamorano-Martín F, Rodríguez-Calvo-de-Mora M, Rodríguez-Vallejo M. Positioning of enhanced monofocal intraocular lenses between conventional monofocal and extended depth of focus lenses: a scoping review. BMC Ophthalmol. 2023;23(1):101. pmid:36918799
- 3. Giglio R, Vinciguerra AL, Presotto M, Jonak K, Rejdak R, Toro MD, et al. Visual outcomes and patient satisfaction after bilateral implantation of an enhanced monofocal intraocular lens: a single-masked prospective randomized study. Int Ophthalmol. 2024;44(1):112. pmid:38407686
- 4. Redruello-Guerrero P, Rivera-Izquierdo M, Jiménez-Gutiérrez C, Láinez-Ramos-Bossini AJ, Yela R, López-Marín I. Improvement of intermediate vision with new monofocal intraocular lenses: A systematic review and meta-analysis. Eur J Ophthalmol. 2023;33(3):1308–19. pmid:36124373
- 5. Wan KH, Au ACK, Kua WN, Ng ALK, Cheng GPM, Lam NM, et al. Enhanced Monofocal Versus Conventional Monofocal Intraocular Lens in Cataract Surgery: A Meta-analysis. J Refract Surg. 2022;38(8):538–46. pmid:35947003
- 6. Nanavaty MA, Ashena Z, Gallagher S, Borkum S, Frattaroli P, Barbon E. Visual Acuity, Wavefront Aberrations, and Defocus Curves With an Enhanced Monofocal and a Monofocal Intraocular Lens: A Prospective, Randomized Study. J Refract Surg. 2022;38(1):10–20. pmid:35020542
- 7. Goslings O, Veraart H, van de Laar-Muskens J, Piñero DP. Clinical outcomes with an aspheric monofocal and a new enhanced monofocal intraocular lens with modified optical profile. Graefes Arch Clin Exp Ophthalmol. 2023;261(8):2315–26. pmid:37256375
- 8. Lopes D, Loureiro T, Carreira R, Rodrigues Barros S, Cardoso JN, Campos P, et al. Comparative evaluation of visual outcomes after bilateral implantation of an advanced or conventional monofocal intraocular lens. Eur J Ophthalmol. 2022;32(1):229–34. pmid:33579171
- 9. de Vries NE, Webers CAB, Touwslager WRH, Bauer NJC, de Brabander J, Berendschot TT, et al. Dissatisfaction after implantation of multifocal intraocular lenses. J Cataract Refract Surg. 2011;37(5):859–65. pmid:21397457
- 10. Morlock R, Wirth RJ, Tally SR, Garufis C, Heichel CWD. Patient-Reported Spectacle Independence Questionnaire (PRSIQ): Development and Validation. Am J Ophthalmol. 2017;178:101–14. pmid:28341605
- 11. Latz C, Licht A, Bayati P, Taya K, Mirshahi A. Cystoid macular edema after low-energy femtosecond-assisted cataract surgery. Graefes Arch Clin Exp Ophthalmol. 2024;263(3):829–35. pmid:39404898
- 12. Gigon E, Bouthour W, Panos GD, Pajic B, Massa H. Real world outcomes of the new Tecnis Eyhance IOL. Eur J Ophthalmol. 2023;33(3):1390–7. pmid:36567614
- 13. de Luis Eguileor B, Martínez-Indart L, Martínez Alday N. Diferencias en la visión intermedia: lente intraocular monofocal vs. lente intraocular monofocal de rango extendido. Arch Soc Esp Oftalmol. 2020;95:523–7.
- 14. Mencucci R, Cennamo M, Venturi D, Vignapiano R, Favuzza E. Visual outcome, optical quality, and patient satisfaction with a new monofocal IOL, enhanced for intermediate vision: preliminary results. J Cataract Refract Surg. 2020;46(3):378–87. pmid:32050218
- 15. Ang RET, Stodulka P, Poyales F. Prospective Randomized Single-Masked Study of Bilateral Isofocal Optic-Design or Monofocal Intraocular Lenses. Clin Ophthalmol. 2023;17:2231–42. pmid:37559778
- 16. Bova A, Vita S. Clinical and Aberrometric Evaluation of a New Monofocal IOL with Intermediate Vision Improvement. J Ophthalmol. 2022;2022:4119698. pmid:35847352
- 17. Park ES, Ahn H, Han SU, Jun I, Seo KY, Kim EK, et al. Visual outcomes, spectacle independence, and patient satisfaction of pseudophakic mini-monovision using a new monofocal intraocular lens. Sci Rep. 2022;12(1):21716. pmid:36522397
- 18. Stodulka P, Pracharova Z. Evaluation of extended depth-of-focus hydrophobic intraocular lens with an optic concept based on combination of high-order aberrations. Ther Adv Ophthalmol. 2023;15:25158414231200108. pmid:37799815
- 19. Azor JA, Vega F, Armengol J, Millan MS. Optical Assessment and Expected Visual Quality of Four Extended Range of Vision Intraocular Lenses. J Refract Surg. 2022;38(11):688–97. pmid:36367260
- 20. Lee JH, Moon SY, Chung HS, Park SY, Lee H, Kim JY, et al. Clinical outcomes of a monofocal intraocular lens with enhanced intermediate function compared with an extended depth-of-focus intraocular lens. J Cataract Refract Surg. 2022;48(1):61–6. pmid:34117177
- 21. Corbelli E, Iuliano L, Bandello F, Fasce F. Comparative analysis of visual outcome with 3 intraocular lenses: monofocal, enhanced monofocal, and extended depth of focus. J Cataract Refract Surg. 2022;48(1):67–74. pmid:34054077
- 22. Lan Q, Liu Y, Xu F, Li M, Li Y, Yang T, et al. Cost-Effectiveness of Presbyopia Correction Among Seven Strategies of Bilateral Cataract Surgery Based on a Prospective Single-Blind Two-Center Trial in China. Ophthalmol Ther. 2022;11(6):2067–82. pmid:36071311
- 23. Daka Q, Henein C, Fang CEH, Mustafa R, Cocaj E, Azuara-Blanco A, et al. Effectiveness of intraocular lenses designed to correct presbyopia after cataract surgery: an overview of systematic reviews. Br J Ophthalmol. 2025;bjo-2025-327363. pmid:40623803