Abstract
Purpose
To evaluate the impact of monitor units (MUs) optimization on volumetric modulated arc therapy (VMAT) plan for nasopharyngeal carcinoma (NPC).
Methods
Twenty-one NPC patients were retrospectively analyzed. Dual-arc VMAT plan were designed using photon optimization algorithms without the monitor unit objective (MUO) tool, denoted as the base plan. Each base plan was re-optimized with the MUO tool with the Maximum MU parameter set to 30% of the base plans’ total MUs and Strength parameters set to 50, 80, and 100, generating plans S50, S80, and S100. Target and organ-at-risk (OAR) dose distributions, MUs, beam delivery time, and gamma passing rates were compared between re-optimized and base plans. Statistical analysis was performed using SPSS 17.0 (paired t-tests; significance: P < 0.05).
Results
Plan S100 reduced target PCTV2 D98% by >4% (relative to the base plan) in four patients. Plan S80 reduced target PGTV and PGTVnd Dmax and target PCTV2 D98% for >3% but <4% in two patients, while other target dose parameters changed by <2%. Compared to the base plan, all re-optimized plans increased the brainstem Dmax (P < 0.05), though the maximum increase was < 1.5%. Plan S50 reduced both parotid glands D50% and Dmean (P < 0.001), while plan S80 reduced both parotids Dmean and the left parotid D50% (P < 0.001). Conversely, S100 increased both parotids D50% and Dmean and the spinal cord Dmax (P < 0.05). Plan S80 and S100 increased the thyroid V40 (P < 0.05). MU reductions averaged 5.1% (S50), 21.4% (S80), and 30.9% (S100), with consistent beam delivery times (~2.5 minutes). Gamma passing rates improved sequentially from the base plan to S50, S80, and S100.
Conclusion
MU optimization in NPC VMAT planning effectively reduces MUs and enhances delivery accuracy (improved gamma passing rates). While target coverage and OAR sparing were generally maintained, higher MUO strengths (e.g., S100) may necessitate careful consideration of dosimetric trade-offs. Moderate MUO settings (e.g., S80) offer a favorable balance between MU reduction and plan fidelity.
Citation: Zeng H, Li Z, Chen Z, Tang S, Lin Q, Zhong M, et al. (2025) Impact of monitor unit optimization in volumetric modulated arc therapy planning for nasopharyngeal carcinoma. PLoS One 20(6): e0327153. https://doi.org/10.1371/journal.pone.0327153
Editor: Chung-Ta Chang, Far Eastern Memorial Hospital, TAIWAN
Received: March 24, 2025; Accepted: June 10, 2025; Published: June 25, 2025
Copyright: © 2025 Zeng et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Data Availability: All relevant data are within the paper and its Supporting Information files.
Funding: The author(s) received no specific funding for this work.
Competing interests: The authors have declared that no competing interests exist.
Introduction
Nasopharyngeal carcinoma (NPC) is one of the most common malignant cancers in China, particularly in the southeastern regions, where the incidence rate ranges from 20 to 50 per 100,000 individuals [1]. Given the anatomical complexity of the nasopharynx and the tumor’s high radiosensitivity, radiotherapy remains the primary treatment modality [2]. Radiotherapy techniques have evolved from three-dimensional conformal radiotherapy (3DCRT) to fixed-field intensity-modulated radiotherapy (IMRT), and more recently, to volumetric modulated arc therapy (VMAT) [3]. These advancements have significantly improved the clinical outcomes for NPC patients [4,5]. Notably, VMAT offers superior normal tissue sparing, reduces early adverse effects, and enhances survival rates compared to IMRT [6]. Additionally, VMAT achieves excellent target coverage while minimizing doses to organs at risk (OARs), particularly neural structures and salivary glands [7,8].
VMAT delivers radiation through continuous adjustments of gantry rotation speed, dose rate, and multi-leaf collimator (MLC) movement during treatment [9]. Compared to IMRT, VMAT significantly reduces monitor units (MUs), shortens treatment duration, and improves delivery efficiency [10,11]. Lower MUs also decrease the risk of secondary malignancies induced by scattered radiation [12].
The Eclipse treatment planning system (TPS) provides multiple optimization tools, including generalized equivalent uniform dose optimization, dose-volume histogram (DVH)-based structure optimization, normal tissue objective optimization, and the monitor unit objective (MUO) tool [13–16]. The MUO tool regulates MUs in VMAT plan through three adjustable parameters: Minimum MU, Maximum MU, and Strength. Minimum MU and Maximum MU define the lower and upper bounds for MUs, respectively, while Strength determines the priority of MU reduction (Fig 1). Previous studies have applied the MUO tool in VMAT plan for hypopharyngeal, prostate, cervical, breast, and non-small cell lung cancers [17–20]. These studies demonstrated that judicious MUO use reduces total MUs without compromising target coverage or OAR doses. However, research on MUO optimization in NPC-specific VMAT plans remains limited. NPC poses unique challenges due to its intricate anatomy, high target dose requirements, and close proximity to multiple OARs [21]. This study investigates the impact of MUO-driven optimization on VMAT plan quality for NPC.
The settings include dose-volume objectives, normal tissue objectives, and MU objective, where MU objective are used to control the total MU of the VMAT plan.
Materials and methods
Patient cohort
This retrospective study received approval from the Institutional Review Board at Gaozhou People’s Hospital (GYLLPJ-2023027). We enrolled 21 patients with NPC treated with VMAT from October 1, 2020 to March 1, 2021. The requirement for informed consent was waived as this is a retrospective study for research purposes only. Data collection and analysis occurred from October and December 2024. All patient data were anonymized prior to analysis. The cohort comprised 18 males and 3 females, with ages ranging from 31 to 76 years (median: 45 years; mean ± SD: 51 ± 12 years). Inclusion criteria were: (1) histopathological confirmation of NPC diagnosis, and (2) completion of the prescribed radiotherapy course. We excluded patients who: (1) lacked complete pretreatment imaging or pathological records, or (2) presented with cachexia at diagnosis.
CT simulation and target delineation
Patients were immobilized in a thermoplastic head-neck-shoulder mask (Orfit Industries, Wijnegem, Belgium) with a head rest, maintaining a standardized supine position with arms positioned alongside the body. Simulation CT scans were acquired using a Siemens SOMATOM Confidence large-bore CT simulator (Siemens Healthineers, Forchheim, Germany), with the following scanning settings: scan range extending from the cranial vertex to 5 cm inferior to the aortic arch, 3 mm slice thickness. All CT datasets were then transferred to the Eclipse TPS (version 13.6; Varian Medical Systems, Palo Alto, CA) for target volume and OAR delineation. Target volumes were defined as follows: Gross tumor volume (GTV): Primary tumor delineated based on fused diagnostic MRI and clinical findings, GTVnd: Bilateral metastatic lymph node, CTV1: High-risk subclinical disease volume, CTV2: Low-risk subclinical disease volume. Each target volume received a uniform 3 mm isotropic expansion to generate the corresponding planning target volumes (PGTV, PGTVnd, PCTV1, and PCTV2). OARs were contoured according to established guidelines and included: Central nervous system structures (spinal cord, brainstem, optic chiasm, bilateral optic nerves), Sensory organs (bilateral eyes, lenses), Endocrine structures (pituitary gland), Salivary glands (bilateral parotid glands), Neurological and musculoskeletal structures (temporal lobes, temporomandibular joints).
Dose prescription
Treatment plans employed a simultaneous integrated boost technique with the following prescription doses delivered in 32 fractions (5 fractions weekly): 7000 cGy for PGTV, 6800 cGy for PGTVnd, 6000 cGy for PCTV1, and 5400 cGy for PCTV2. Dose coverage objectives required that ≥95% of each target volume receive 100% of the prescribed dose. OAR constraints were implemented as follows: brainstem ≤ 5400 cGy, spinal cord ≤ 4500 cGy, optic chiasm and nerves ≤ 5400 cGy, lenses ≤ 1000 cGy, eyes ≤ 4500 cGy, For the parotid glands, the mean dose to at least one gland was constrained to ≤ 2600 cGy, with D50% < 3000 cGy. The temporomandibular joint was limited to < 6000 cGy. All target volume definitions and plan evaluations adhered to the Radiation Therapy Oncology Group (RTOG) 0615 protocol [22] and the International Commission on Radiation Units and Measurements (ICRU) Report 83 [23].
VMAT planning
For each patient, we implemented a dual-arc VMAT delivery strategy with the following specifications. The first arc employed counterclockwise gantry rotation from 179° to 181° with a collimator angle of 10°, while the second arc used clockwise rotation from 181° to 179° with a collimator angle of 350°. All plans utilized a 6 MV photon beam with a maximum dose rate of 600 MU/min and were optimized using a photon optimization algorithm in Varian Eclipse v13.6 TPS. The initial optimization created a “base” plan without MUO tool. Subsequently, we generated three additional plans by re-optimizing only the MUO parameters while maintaining identical field geometry, dose-volume objectives, and structure weighting factors. Based on Prado et al.‘s findings [24], we set the Minimum MU to 0 and Maximum MU to 30% of the base plan’s total MUs, while testing Strength parameters of 50, 80, and 100. This approach yielded four distinct plans per patient: (1) the original “base” plan without MUO; (2) “S50” with Strength = 50; (3) “S80” with Strength = 80; and (4) “S100” with Strength = 100. The 30% maximum MU constraint was specifically chosen based on previous evidence demonstrating its effectiveness in MU reduction while preserving plan quality [24]. This systematic evaluation of Strength parameters allowed comprehensive assessment of the MUO tool’s sensitivity in NPC VMAT planning.
Comparison of target and OAR doses, MUs, and beam delivery time
We conducted a comprehensive comparison between the three MUO-optimized plans (S50, S80, and S100) and the base plan using multiple evaluation parameters. For target volumes, we analyzed the D95% coverage for all planning target volumes (PGTV, PGTVnd, PCTV1, and PCTV2), along with maximum dose (Dmax), mean dose (Dmean), and homogeneity index (HI) as defined in ICRU Report 83 [23], OAR evaluation included Dmax assessment for critical serial organs (brainstem, spinal cord, optic chiasm, optic nerves, lenses, eyes, and pituitary gland), Dmean for parallel organs (parotid glands, thyroid gland, and submandibular glands), and specific dose-volume parameters including D50% for parotid glands and V40 (volume receiving ≥40 Gy) for thyroid gland. Treatment efficiency was quantified by monitoring the total MUs and precisely measuring beam delivery time using a calibrated stopwatch. The delivery time measurement specifically captured the beam-on to beam-off duration for each arc, deliberately excluding setup time and beam mode up time, with the total delivery time representing the sum of both arcs’ active treatment durations. This rigorous evaluation protocol enabled systematic comparison of plan quality versus delivery efficiency across all optimization approaches while maintaining consistent measurement standards.
Pretreatment plan verification
For pre-treatment plan quality verification, the 3D dose distributions of all four VMAT plans (base, S50, S80, and S100) were measured for each patient using the Sun Nuclear ArcCheck diode array phantom (Sun Nuclear Inc, Melbourne, FL) mounted on a Varian Trilogy linear accelerator (Varian Medical Systems, Palo Alto, CA). The measured dose distributions were compared with the TPS-calculated doses using gamma index analysis in SNC Patient software (Sun Nuclear Inc, Melbourne, FL). The primary evaluation used a global absolute dose gamma evaluation of a 3%/3 mm with a 10% dose threshold [25]. Prior to measurements, the output dose of the Trilogy linear accelerator was calibrated, following the International Atomic Energy Agency TRS-398 protocol [26], to ensure dosimetric accuracy. To further assess the impact of MU reduction on plan delivery accuracy, additional gamma analyses were performed under stricter criteria: 2%/2 mm, 3%/2 mm, and 1%/1 mm. This multi-criteria evaluation allowed systematic investigation of whether reduced MUs influenced the gamma passing rate (GPR) while maintaining clinically acceptable plan quality.
Statistical analysis
All statistical analysis was performed using SPSS Statistics version 17.0 (IBM Corp., Armonk, NY, USA). We first assessed data normality using the Shapiro-Wilk test. Normally distributed data were presented as mean ± standard deviation (x̄ ± s). while non-normally distributed data were expressed as median (interquartile range). For normally distributed parameters, paired Student’s t-tests were conducted to compare the MUO-optimized plans (S50, S80, and S100) with the base plan, otherwise, the multi-correlation sample non-parametric Friedman test is used. The significance level was set at P < 0.05 (two-tailed).
Results
Target dose comparison
The target dose parameters across the four treatment plans (base, S50, S80, and S100) were compared, with detailed results presented in Tables 1 and 2. Figs 2–4 illustrate the relative differences in target dose parameters between the MUO-optimized plans and the base plan for all 21 patients.
= 21). Among the 21 patients in plan S50, only one showed a decrease in target dose exceeding 2% but remaining below 2.5% compared to the base plan. For the PGTV, dose deviations were within 1% of the base plan.
= 21). Among the 21 patients in Plan S80, only one case showed a deviation of D95% exceeding 1.5% but remaining below 2% compared to the base plan. For PGTV, all cases D95% deviations were within 1% of the base plan.
= 21). In Plan S100, the PCTV2 showed a notable dose reduction compared to the base plan. Among the cases, four patients demonstrated a decrease in D95% exceeding 3%, while another four patients had a D98% reduction surpassing 4%. However, for the PGTV, all dose deviations from the base plan were within 2%.
Plan S50 showed statistically significant differences (P < 0.05) in several key parameters compared to the base plan: the Dmean of PGTV, D95% and HI of PGTVnd, and Dmean and HI of PCTV2. Importantly, all other dose parameters maintained comparable values (P > 0.05), demonstrating that the S50 optimization preserved most dosimetric characteristics while achieving MU reduction.
More extensive variations emerged in Plan S80, with significant differences (P < 0.05) observed in multiple parameters: D95% and HI of PGTV; D95%, Dmean, and HI of PGTVnd; and D95%, Dmax, and HI of PCTV2. However, other dose parameters remained unchanged, suggesting a selective impact of the increased MUO strength.
The most pronounced differences occurred in Plan S100, where only PGTV Dmax and Dmean, PGTVnd Dmax, and PCTV1 Dmax showed no significant variation (P > 0.05). Clinically relevant dosimetric changes included a > 4% decrease in PCTV2 D98% for 4 patients (19%) and 3–4% reductions in PGTV and PGTVnd Dmax and PCTV2 D98% for 1 patient (5%). Most other parameter changed within 2%.
Visual analysis of Fig 5 (cross-sectional dose distribution) revealed progressive extension of the 40 Gy isodose line toward the mandible across S50 → S80 → S100 plans, while the DVH in Fig 6 provided dose-volume relationships. These findings demonstrate a clear trade-off between MU reduction efficacy and target coverage preservation, with S80 appearing to represent an optimal balance between these competing objectives. The progressive changes across the plans suggest that MUO strength requires careful selection to maintain clinical dosimetric goals while achieving efficiency gains.
The 7000 cGy isodose lines (yellow) in all four plans demonstrated adequate coverage of the target PGTV (red). However, in Plan S100, the 7000 cGy isodose line showed inward retraction and failed to completely cover the PGTV. Compared with the base plan, Plan S50, S80 and S100 exhibited progressively expanding low-dose isodose lines (4500 cGy and 4000 cGy), resulting in deteriorated dose conformity.
For the target, plan S100 showed significant deviations in the DVH compared to the base plan. Regarding OARs, plan S100 also exhibited notable DVH variations for the parotid glands and spinal cord. In contrast, plan S50 demonstrated nearly identical DVH curves to the base plan for both target and OARs.
OARs dose comparison
The comparative assessment of OAR doses across the four treatment plans revealed several important findings (Tables 3,4). All MUO-optimized plans (S50, S80, S100) demonstrated significant increases (P < 0.05) in brainstem Dmax compared to the base plan, though the absolute differences remained clinically modest (<1.5%). Notably, plan S100 showed additional elevation in spinal cord Dmax (P < 0.05).
Parotid gland analysis revealed distinct optimization patterns: plan S50 achieved reductions in both D50% and Dmean for bilateral parotids (P < 0.001), while plan S80 maintained Dmean reduction bilaterally (P < 0.001) but only reduced left parotid D50% (P < 0.001 vs P = 0.181 for right). In contrast, plan S100 increased both dose parameters for parotids (P < 0.05). Thyroid gland evaluation showed V40 increases in both S80 and S100 plans (P < 0.05).
Critical structures including bilateral temporal lobes (Dmax), temporomandibular joints (Dmax), and submandibular glands (Dmean) maintained stable dose distributions across all plans without significant variations (P > 0.05). These findings demonstrate that while MUO optimization affects certain OARs differentially based on strength setting, it preserves dose constraints to several critical neural and musculoskeletal structures.
The results suggest a clear trade-off between MU reduction efficacy and OAR sparing, with plan S50 representing the most favorable balance – achieving significant parotid dose reduction without compromising other critical structures. Higher strength optimizations (particularly S100) showed diminishing returns, introducing dose elevations to multiple OARs while providing only marginal additional MU reduction benefits. These findings have important clinical implications for NPC radiotherapy planning, where parotid sparing is particularly crucial for maintaining patient quality of life.
MU and beam delivery time comparison
The analysis of treatment delivery efficiency revealed important findings (Table 5). Compared to the base plan, all MUO-optimized plans demonstrated statistically significant reductions in MUs (P < 0.05). Plan S50 achieved a modest 5.1% average MU reduction, while S80 and S100 showed more substantial decreases of 21.5% and 30.9%, respectively. This progressive reduction pattern correlated directly with increasing MUO strength parameters.
Interestingly, despite these significant MU reductions, the beam-on time remained consistent across all plans. The average delivery time maintained approximately 2.5 minutes for all four treatment plans (base, S50, S80, and S100). This observation suggests that while the MUO tool effectively reduces total MUs - potentially lowering secondary cancer risks from scattered radiation – it does not translate into faster treatment delivery under current operational parameters.
Gamma passing rate comparison
The GPR evaluation of the four treatment plans under different gamma criteria (3%/3 mm, 2%/2 mm, 3%/2 mm, and 1%/1 mm) revealed important insights into plan delivery accuracy (Table 6). Plan S50, with a modest 5.1% MU reduction, demonstrated equivalent GPR performance to the base plan (P > 0.05), indicating that limited MU optimization maintains delivery accuracy. In contrast, both Plan S80 (21.5% MU reduction) and Plan S100 (30.9% MU reduction) showed differences in GPR compared to the base plan (P < 0.001) under all evaluation conditions.
A particularly interesting finding emerged when examining the relationship between MU reduction and delivery precision. Under the most stringent 1%/1 mm criterion, we observed a consistent trend where greater MU reduction correlated with improved GPR. This relationship suggests that the reduced MU variability in optimized plans may enhance delivery precision, particularly for more aggressive optimization settings. However, this potential benefit must be carefully weighed against the previously noted trade-offs in target coverage and OAR sparing observed with higher MUO strength settings.
Discussion
Modern radiotherapy has revolutionized the treatment of NPC, with IMRT establishing itself as the preferred treatment method by achieving 5-year local control rates of approximately 90% [27]. The subsequent development of VMAT has further advanced the field, offering significant improvements over conventional IMRT through its dynamic delivery system. VMAT’s arc-based optimization provides superior dose conformity to complex tumor geometries while simultaneously enhancing protection of surrounding critical structures [28]. These technical advantages have translated into tangible clinical benefits, most notably in reducing the incidence and severity of radiation-induced late complications that impact patients’ quality of life. Clinical studies have consistently demonstrated VMAT’s ability to mitigate treatment-related toxicities, including sensorineural hearing loss, trismus, and temporal lobe injury [29–31]. The technique’s improved efficiency, characterized by shorter treatment delivery times, further enhances its clinical utility without compromising therapeutic outcomes.
The MUO tool has been investigated across various treatment sites with varying results. While Varian established the basic functionality of the MUO tool, clinical implementation strategies have primarily been developed through subsequent research. Jiménez-Puertas et al. demonstrated that setting Maximum MU to 400 could achieve 15–22% MU reduction in prostate, gynecological, and head-neck VMAT plans without compromising dosimetric quality, though their study did not examine the impact of the Strength parameter [15]. Clemente et al.‘s prostate cancer study revealed an important dose-quality trade-off, with Strength = 100 achieving 28% MU reduction but worsening target homogeneity by 23% [18]. Huang et al. applied the MUO tool to non-small cell lung cancer stereotactic body radiation therapy plans, demonstrating successful MU reduction while maintaining OAR doses [19]. Mancosu et al. investigated MUO tool for VMAT breast treatments. They fixed Strength = 100 and Maximum MU used were −50%, −20%, 20% and 50% of the reference MU plans. They found no significant deviations in target PTV coverage while for OARs some unacceptable deviations were observed. The MU reduction obtained range from 18% to 39% [20]. Our findings in NPC show similar trends, with Strength = 100 producing 31% MU reduction but 17% homogeneity degradation in PGTV, suggesting this represents a characteristic effect of high-strength MUO optimization.
The MUO-optimized plans achieved progressively greater MU reductions from S50 (5.1%) to S100 (30.9%), all statistically significant (P < 0.004). However, this efficiency came at variable clinical costs. Most notably, the highest-efficiency S100 plans showed clinically unacceptable target underdosing in 38% of cases (8/21), with PCTV2 D98% deviations up to 6% that could potentially compromise tumor control. Intermediate S80 plans maintained better dosimetric fidelity while still achieving 21.5% MU reduction, with most deviations remaining below clinically relevant thresholds (<1% or <50 cGy total dose). The minimal OAR dose increases (≤1.5% Dmax for critical neural structures) across all plans suggest these may represent inherent system characteristics rather than optimization failures.
These findings carry important clinical implementation recommendations. For centers prioritizing dose fidelity, conservative S50 optimization provides modest MU reduction (5.1%) with minimal clinical impact. The S80 setting offers a balanced compromise (21.5% MU reduction) suitable for most routine cases, while S100 optimization (30.9% reduction) should be reserved for specific clinical scenarios with rigorous quality assurance protocols. The anatomical dependence of these outcomes – particularly the heightened sensitivity of complex head-and-neck target volumes compared to other sites like breast or prostate – underscores the need for site-specific optimization guidelines rather than universal parameter sets. Future development should focus on hybrid algorithms that combine MUO with other optimization strategies to break the current efficiency-fidelity trade-off paradigm.
Our study revealed important insights into how MU reduction impacts various aspects of treatment delivery and plan quality. Notably, the beam-on time remained consistent at approximately 2.5 minutes across all plans regardless of MU reduction level, which can be attributed to the Eclipse TPS photon optimization algorithm’s tendency to maintain a relatively constant gantry rotation speed of 6.0°/s. This finding demonstrates that MU reduction primarily affects dose rate modulation and MLC motion speed rather than overall treatment duration.
We observed that MU reduction was associated with improved GPR, particularly under stringent 1%/1 mm evaluation criteria. The S50, S80, and S100 plans showed progressive GPR improvements of 0.4%, 3.7%, and 5.6% compared to the base plan respectively. This result was contributed from several factors: (1) generation of simpler MLC shapes, (2) more consistent dose rate change patterns, and (3) reduced out-of-field scatter dose. These findings suggest that moderate MU reduction may actually enhance plan quality by creating more reproducible beam delivery sequences.
The clinical implications of these results are significant. For complex treatment plans that initially show marginal GPR (<95% under 3 mm/3% criteria), judicious application of the MUO tool could serve dual purposes – both reducing MUs and improving delivery accuracy. Additionally, the reduction in total MUs directly translates to decreased scatter dose to patients, potentially lowering the risk of secondary radiation-induced malignancies [32]. This benefit is particularly relevant for younger patients or those with long expected survival times.
This study has several important limitations that should be considered when interpreting the results. As a retrospective analysis, our investigation focused specifically on evaluating the dosimetric impacts of MUO optimization in NPC VMAT planning, including effects on target coverage, OAR doses, beam-on time, and GPR. However, we did not assess treatment endpoints such as tumor control probability or normal tissue complication probability, which represent important gaps in understanding the full clinical implications of our findings. Secondly, the advantages of using the MUO tool need to be confirmed through long-term patient follow-up and based on a large number of case practices.
Conclusion
Our comprehensive evaluation of MUO-optimized VMAT plans for NPC provides clear guidance for clinical implementation. The study demonstrates that careful selection of MUO parameters can achieve significant MU reduction while maintaining treatment quality. Among the three optimization levels investigated, Plan S80 emerges as the optimal choice for routine clinical practice, delivering a 21.5% MU reduction (P < 0.001) while preserving target coverage and OAR protection. This moderate optimization level maintained all dosimetric parameters within clinically acceptable ranges and showed equivalent GPR to conventional plans, making it suitable for standard NPC treatment plans.
Supporting information
S1 File. Dose parameters for the target and OARs, as well as dose-volume histogram parameters for each plan of a representative patient.
https://doi.org/10.1371/journal.pone.0327153.s001
(ZIP)
References
- 1. Imano N, Murakami Y, Nakashima T, Takeuchi Y, Takahashi I, Doi Y, et al. Clinical outcomes of concurrent chemoradiotherapy with volumetric modulated arc therapy in patients with locally advanced nasopharyngeal carcinoma. Jpn J Radiol. 2017;35(11):673–80. pmid:28887732
- 2. Akbas U, Koksal C, Kesen ND, Ozkaya K, Bilge H, Altun M. Nasopharyngeal carcinoma radiotherapy with hybrid technique. Med Dosim. 2019;44(3):251–7. pmid:30366620
- 3. Mohan G, T P AH, A J J, K M SD, Narayanasamy A, Vellingiri B. Recent advances in radiotherapy and its associated side effects in cancer—a review. JoBAZ. 2019;80(1).
- 4. Leong YH, Soon YY, Lee KM, Wong LC, Tham IWK, Ho FCH. Long-term outcomes after reirradiation in nasopharyngeal carcinoma with intensity-modulated radiotherapy: A meta-analysis. Head Neck. 2018;40(3):622–31. pmid:29130584
- 5. Sun X-S, Li X-Y, Chen Q-Y, Tang L-Q, Mai H-Q. Future of Radiotherapy in Nasopharyngeal Carcinoma. Br J Radiol. 2019;92(1102):20190209. pmid:31265322
- 6. Guo R, Tang L-L, Mao Y-P, Zhou G-Q, Qi Z-Y, Liu L-Z, et al. Clinical Outcomes of Volume-Modulated Arc Therapy in 205 Patients with Nasopharyngeal Carcinoma: An Analysis of Survival and Treatment Toxicities. PLoS One. 2015;10(7):e0129679. pmid:26146828
- 7. Amaoui B, Semghouli S, Mouhssine D, Bayla A, Bouih A, Bouih N. Dosimetric analysis of VMAT irradiation in nasopharyngeal carcinoma: A Moroccan radio oncology center’s experience. Eur J Mol Clin Med. 2021;8(3):979–88.
- 8. Franzese C, Fogliata A, Loi M, Badalamenti M, Franceschini D, Comito T, et al. Dosimetric impact of volumetric modulated arc therapy for nasopharyngeal cancer treatment. Rep Pract Oncol Radiother. 2021;26(1):101–10. pmid:34046220
- 9. Otto K. Volumetric modulated arc therapy: IMRT in a single gantry arc. Med Phys. 2008;35(1):310–7. pmid:18293586
- 10. Wang Q, Qin J, Cao R, Xu T, Yan J, Zhu S, et al. Comparison of Dosimetric Benefits of Three Precise Radiotherapy Techniques in Nasopharyngeal Carcinoma Patients Using a Priority-Classified Plan Optimization Model. Front Oncol. 2021;11:646584. pmid:34646757
- 11. Amin N, Atiq A, Ikram M, Atiq M, Naeem H, Yousaf M, et al. Dosimetric analysis of Rapid Arc (VMAT) treatment planning in head and neck cancer for quality assurance treatment. J King Saud University - Sci. 2023;35(2):102476.
- 12. Ho H-W, Lee SP, Lin H-M, Chen H-Y, Huang C-C, Wang S-C, et al. Dosimetric comparison between RapidArc and HyperArc techniques in salvage stereotactic body radiation therapy for recurrent nasopharyngeal carcinoma. Radiat Oncol. 2020;15(1):164. pmid:32641082
- 13. Fogliata A, Thompson S, Stravato A, Tomatis S, Scorsetti M, Cozzi L. On the gEUD biological optimization objective for organs at risk in Photon Optimizer of Eclipse treatment planning system. J Appl Clin Med Phys. 2018;19(1):106–14. pmid:29152846
- 14. Bell JP, Patel P, Higgins K, McDonald MW, Roper J. Fine-tuning the normal tissue objective in eclipse for lung stereotactic body radiation therapy. Med Dosim. 2018;43(4):344–50. pmid:29277249
- 15. Jiménez-Puertas S, Sánchez-Artuñedo D, Hermida-López M. Assessment of the Monitor Unit Objective tool for VMAT in the Eclipse treatment planning system. Rep Pract Oncol Radiother. 2018;23(2):121–5. pmid:29556140
- 16. Oliver M, Gagne I, Popescu C, Ansbacher W, Beckham WA. Analysis of RapidArc optimization strategies using objective function values and dose-volume histograms. J Appl Clin Med Phys. 2009;11(1):3114. pmid:20160684
- 17. Ahamed S, Singh N, Gudipudi D, Mulinti S, Talluri A, Soubhagya B, et al. Assessment of monitor unit limiting strategy using volumetric modulated arc therapy for cancer of hypopharynx. Phys Med. 2017;35:73–80. pmid:28228330
- 18. Clemente S, Cozzolino M, Chiumento C, Fiorentino A, Caivano R, Fusco V. Monitor unit optimization in RapidArc plans for prostate cancer. J Appl Clin Med Phys. 2013;14(3):4114. pmid:23652245
- 19. Huang B-T, Lin Z, Lin P-X, Lu J-Y, Chen C-Z. Monitor unit optimization in stereotactic body radiotherapy for small peripheral non-small cell lung cancer patients. Sci Rep. 2015;5:18453. pmid:26679747
- 20. Mancosu P, Reggiori G, Alongi F, Cozzi L, Fogliata A, Lobefalo F, et al. Total monitor units influence on plan quality parameters in volumetric modulated arc therapy for breast case. Phys Med. 2014;30(3):296–300. pmid:24076257
- 21. Zhang A, Gao L, Shi X, Kang W. Research Progress of Radiotherapy for Nasopharyngeal Carcinoma. Int J Front Med. 2022;4(8):10–2. http://doi:10.25236/ijfm.2022.040803
- 22. Lee NY, Zhang Q, Pfister DG, Kim J, Garden AS, Mechalakos J, et al. Addition of bevacizumab to standard chemoradiation for locoregionally advanced nasopharyngeal carcinoma (RTOG 0615): a phase 2 multi-institutional trial. Lancet Oncol. 2012;13(2):172–80. pmid:22178121
- 23. Hodapp N. Verordnung, Dokumentation und Kommunikation der fluenzmodulierten Photonenstrahlentherapie (IMRT) [The ICRU Report 83: prescribing, recording and reporting photon-beam intensity-modulated radiation therapy (IMRT)]. Strahlenther Onkol. 2012;88(1):97-9. http://doi:10.1007/s00066-011-0015-x pmid:22234506
- 24. Prado A, Gaitán Á, Leonor M, Manzano M, Cabello E, Díaz R, et al. Performance of the eclipse monitor unit objective tool utilizing volumetric modulated arc therapy for rectal cancer. Rep Pract Oncol Radiother. 2019;24(2):227–32. pmid:30858766
- 25. Miften M, Olch A, Mihailidis D, Moran J, Pawlicki T, Molineu A, et al. Tolerance limits and methodologies for IMRT measurement-based verification QA: Recommendations of AAPM Task Group No. 218. Med Phys. 2018;45(4):e53–83. pmid:29443390
- 26.
International atomic energy agency, Absorbed Dose Determination in External Beam Radiotherapy, Technical Reports Series No. 398 (Rev. 1), Vienna;IAEA: 2024. https://doi.org/10.61092/iaea.ve7q-y94k
- 27. Gou X, Duan B, Shi H, Qin L, Xiao J, Chen N. The relations of dosimetric parameters with long-term outcomes and late toxicities in advanced T-stage nasopharyngeal carcinoma with IMRT. Head Neck. 2020;42(1):85–92. pmid:31650657
- 28. Lu S, Fan H, Hu X, Li X, Kuang Y, Yu D, et al. Dosimetric comparison of helical tomotherapy, volume-modulated arc therapy, and fixed-field intensity-modulated radiation therapy in locally advanced nasopharyngeal carcinoma. Front Oncol. 2021;11:764946. pmid:34804969
- 29. He L, Xiao J, Wei Z, He Y, Wang J, Guan H, et al. Toxicity and dosimetric analysis of nasopharyngeal carcinoma patients undergoing radiotherapy with IMRT or VMAT: A regional center’s experience. Oral Oncol. 2020;109:104978. pmid:32861986
- 30. Lee H-F, Lan J-H, Chao P-J, Ting H-M, Chen H-C, Hsu H-C, et al. Radiation-induced secondary malignancies for nasopharyngeal carcinoma: a pilot study of patients treated via IMRT or VMAT. Cancer Manag Res. 2018;10:131–41. pmid:29403311
- 31. Huang T-L, Tsai M-H, Chuang H-C, Chien C-Y, Lin Y-T, Tsai W-L, et al. Quality of life and survival outcome for patients with nasopharyngeal carcinoma treated by volumetric-modulated arc therapy versus intensity-modulated radiotherapy. Radiat Oncol. 2020;15(1):84. pmid:32307024
- 32. Patel S, Drodge S, Jacques A, Warkentin H, Powell K, Chafe S. A comparative planning analysis and integral dose of volumetric modulated arc therapy, helical tomotherapy, and three-dimensional conformal craniospinal irradiation for pediatric medulloblastoma. J Med Imaging Radiat Sci. 2015;46(2):134–40. pmid:31052086