Skip to main content
Advertisement
Browse Subject Areas
?

Click through the PLOS taxonomy to find articles in your field.

For more information about PLOS Subject Areas, click here.

  • Loading metrics

The efficacy of Netrod Six-electrode Radiofrequency Renal Denervation system: mathematical modeling, phantom experiments and animal studies

  • Xiaohua Song ,

    Roles Software, Writing – original draft

    ‡ These authors are joint senior authors on this work.

    Affiliations College of Chemistry and Life Science, Beijing University of Technology, Beijing, China, Industrial Informatization and Science and Technology Bureau of Changqing District, Jinan, China

  • Jiulin Guo ,

    Roles Data curation

    ‡ These authors are joint senior authors on this work.

    Affiliation Shanghai Golden Leaf MedTech CO., Ltd, Shanghai, China

  • Hongguang Cao,

    Roles Data curation

    Affiliation Shanghai Golden Leaf MedTech CO., Ltd, Shanghai, China

  • Qun Nan

    Roles Funding acquisition, Writing – review & editing

    nanqun@bjut.edu.cn

    Affiliation College of Chemistry and Life Science, Beijing University of Technology, Beijing, China

Abstract

The principle of NetrodTM Renal Denervation RF Generator is based on the heating of radiofrequency electrodes to achieve the therapeutic temperature for peripheral nerve ablation with minimal impact on intra-arterial blood flow. The extravascular heat distribution and continuous ablation were evaluated using COMSOL Multiphysics with an idealized 3D model and validated in an ex vivo experiment. The safety of ablation was verified by H&E staining of a hypertensive dog model in the treatment group of animal experiments, and changes in blood pressure before and after the procedure(assessed at 7, 30, and 60 days in animal experiments) were used to characterize the efficacy. The simulation results showed that the maximum injury depth range of the NetrodTM Six-electrode is 5.08mm and the depth of the ablation zone obtained in the Phantom experiment was 3–6mm. Circumferential coverage was up to 91.99% and the ablation zone was continuous. In animal experiments, eight animals in the treatment group had significantly lower blood pressure (88.9%). Nerve cells in nerve bundles showed vacuolar changes and thickening of peripheral sheaths after peripheral nerve ablation of renal arteries.The NetrodTM Six-electrode presented here has potential to effectively and safely ablate renal nerves. The demonstrated efficacy and safety in animal models provide a foundation for clinical trials in humans.

Introduction

Hypertension is a cardiovascular syndrome with elevated arterial pressure in the body circulation as the main clinical manifestation and is an important risk factor for cardiovascular and cerebrovascular diseases. Hypertension is a common chronic disease that continues to increase in prevalence worldwide and is expected to reach 1.56 billion people by 2025, with refractory hypertension(RH) being particularly challenging to treat. [1]. Refractory hypertension(RH) is defined as hypertension in which blood pressure cannot be achieved even after adequate and reasonable combination of at least 3 antihypertensive drugs (including diuretics) based on dietary modification and lifestyle improvement [2]. RH accounts for about 5–15% of hypertensive patients, and some sources report that it can reach up to 20%-30%. There is still a clinical need for new easy, safe and effective treatments for hypertension.

Excessive sympathetic activation plays a significant role in the development and maintenance of hypertension [35]. Several studies have indicated that blocking sympathetic nerves may serve as a viable target for hypertension treatment [68]. Renal denervation (RDN) is a minimally invasive procedure that reduces blood pressure by modulating sympathetic hyperexcitability [9]. This technique is gaining acceptance and widespread application in clinical settings across various countries [10].Other studies [11] have also shown that RDN has the potential to extend to other autonomic nervous system disorders such as atrial fibrillation [12,13], heart failure [14,15] and tachycardia [16].

Although research related to RDN is still in its early stages, several new devices have been developed in recent years [17]. Common radiofrequency (RF) catheter ablation devices include the Symplicity System, Symplicity Spyral System, EnligHTN System, FlashPoint™ System, and Iberis System. We have developed the NetrodTM Six-electrode Radiofrequency Renal Denervation system and have investigated the safety and efficacy of the system in simulations, phantom experiment, and animal studies.

In the following sections, we will detail the numerical simulations and validation experiments conducted with the Netrod™ Six-electrode. This research aims to provide a scientific basis for assessing the clinical efficacy of renal sympathetic nerve ablation using the Netrod™ Six-electrode.

Methods

Equipment introduction

NetrodTM Six-electrode Radiofrequency Renal Denervation system (Shanghai Golden Leaf MedTech Co. Ltd) includes NetrodTM Renal Denervation RF Generator and NetrodTM Six-electrode Radiofrequency Renal Denervation catheter (Fig 1).

thumbnail
Fig 1. Radiofrequency ablation mainframe and NetrodTM Six-electrode structure diagram.

https://doi.org/10.1371/journal.pone.0323336.g001

The generator primarily consists of a microcontroller, a signal source circuit, a power drive circuit, a power output circuit, a detection circuit, and a display operation control unit. The detection circuit mainly detects monitors voltage, current, and temperature levels (Fig 2).

thumbnail
Fig 2. NetrodTM Renal Denervation RF Generator components.

https://doi.org/10.1371/journal.pone.0323336.g002

The microcontroller is the core component of the generator. It not only processes the voltage, current, and temperature data transmitted by the detection circuit, but also outputs impedance, temperature, and power data to the display operation control unit. Furthermore, it intelligently adjusts the power output of the signal sources based on temperature and impedance to maintain the electrode surface temperature at a constant value, thereby ensuring safety during operation. The display operation control unit real-time records and displays the temperature, impedance, and power of each electrode during the radiofrequency ablation, and draws relevant curves (Fig 3), allowing medical staff to monitor the working status of the system in real-time through the display.

thumbnail
Fig 3. 120 seconds of radiofrequency ablation, the power(P), temperature(T), and impedance(Z) of electrode 4 change over time.

https://doi.org/10.1371/journal.pone.0323336.g003

The radiofrequency ablation catheter is composed of a central wire drawing with a developing head, a mesh tube connector, an electrode, a mesh basket bracket, a catheter sleeve, a seven-hole tube, and a handle assembly (Fig 4).The simulated schematic of the radiofrequency ablation catheter entering the vessel is shown in Fig 5.

thumbnail
Fig 4. NetrodTM Six-electrode Basket Radiofrequency Renal Denervation Catheter.

1: Central wire drawing with a developing head; 2: Mesh basket bracket; 3: Electrode; 4: Seven-hole tube; 5: Adjustment knob; 6: Handle.

https://doi.org/10.1371/journal.pone.0323336.g004

thumbnail
Fig 5. Basket catheter ablation of renal sympathetic nerves.

A. Renal sympathetic nerve distribution; B,C. The basket catheter enters the branch for ablation; D. The basket catheter expands the basket stent to perform ablation in the main trunk of the renal artery.

https://doi.org/10.1371/journal.pone.0323336.g005

Finite element method modeling

We use COMSOL Multiphysics software (COMSOL, Burlington MA, USA), a finite element simulation software, to perform electromagnetic field and temperature field coupling calculations.

In radiofrequency ablation therapy, thermal energy is generated by converting the electrical energy that is applied as the application of radiofrequency current to the target renal artery tissue. The electric field control equation [18] can be calculated by solving the following Laplace equation. The frequency of the RF current is set to 470 KHz [19], and the biological medium can be considered resistive because the displacement current is much smaller than the conduction current and thus can be neglected [20]. The distributed heat source Qr is given by Eq. (1)

(1)

Where J is the current density, measured in A/m2; Qr is the distributed heat source, measured in J. The voltage is obtained from the solved Laplace equation, which is the controlling equation for solving the electric field problem in the presence of the internal power supply; see Eq. (2) and (3).

(2)(3)

Where V is the voltage, measured in V; E is the electric field strength, measured in V/m; is the electrical conductivity, measured in S/m. In the model, the boundary conditions at the current interface are set as follows:

All external surfaces of the abdominal cavity

(4)

The initial potential is set to 0 V, the outer surface of the electrodes is set to a voltage value with temperature control. In the entire model, all the outer surfaces of the tissue are grounded with zero current density.

In the simulation, an idealized three-dimensional vascular geometry model of the renal artery was constructed, as shown in Fig 6. The tissue included two major parts, the renal artery (6mm diameter) and the abdominal cavity (26mm × 26mm × 30mm), which was constructed as the external environment for the simulation. The arterial inlet flow rate was set to 0.5 m/s, and the ablation time was set to 120s.The material property parameters of the abdominal portion, blood, and electrodes are shown in Table 1.

thumbnail
Table 1. shows material property parameters of the abdominal part, blood and electrodes [2125].

https://doi.org/10.1371/journal.pone.0323336.t001

thumbnail
Fig 6. Three-dimensional renal artery tissue and net basket-based RF electrode model.

https://doi.org/10.1371/journal.pone.0323336.g006

Where is the density, kg/m3; σ is the conductivity, S/m; C is the specific heat capacity, J/kg; is the thermal conductivity, W/(m°C); is the relative dielectric constant.

Phantom experiment validation

Tissue-mimicking phantoms play a crucial role in the development, performance testing, quality assurance, and formulation of effective treatment plans for RF ablation devices [26]. Given that the walls of real human renal vessels are very thin, the wall thickness is often neglected in experimental models. In this study, we utilized a cylindrical mold with a vascular channel, where the surrounding area simulates human abdominal tissue. The entrance diameter of the vascular channel is 6 mm, corresponding to the dimensions of the simulation model. Polyacrylamide is particularly advantageous due to its stability at high temperatures. Many researchers have reported successful applications of polyacrylamide-based transparent tissue-mimicking molds, incorporating protein as a temperature indicator for thermal ablation studies [27]. The raw materials and formulation details for the specific mold used in this study are presented in Table 2. [28]

thumbnail
Table 2. Proportioning table for the preparation of 100 ml solution [29].

https://doi.org/10.1371/journal.pone.0323336.t002

The target temperature for the electrode was set at 60°C, and the experiment was conducted for a duration of 120 seconds. For thermal damage analysis, we evaluated the depth of thermal damage in the body mold through post-experimental assessment. The body mold changes color from blue to white when the temperature exceeds 45°C. Therefore, the color change on the surface of the body mold served as an indicator of the ablation effect.

Animal study

Dogs subjected to animal testing and experimental protocols were treated in strict accordance with protocols approved by the Institutional Animal Care Committee of Shanghai Harborside Medical Technology Co., Ltd., based on the Institutional Animal Care and Use Committees (IACUC) Animal Care and Use Program on Animal Experimentation (approval No. IACUC-2021–003).

A total of 18 female experimental dogs (treatment group 101–109 and control group 201–209, 9 animals each) that had successfully established hypertension models were included in this experiment and divided into three points of time, with 6 dogs in each point of time (3 animals each in the treatment group and the control group). The three points of time were 7 days (101–103 for the treatment group and 201–203 for the control group), 30 days (104–106 for the treatment group and 204–206 for the control group) and 60 days (107–109 for the treatment group and 207–209 for the control group). Under X-ray fluoroscopy, the left and right renal arteries of the animals in the treatment group were ablated with the system, and the operational performance of the experimental dogs in the treatment group was evaluated, and the Digital Subtraction Angiography (DSA) images are shown in Fig 7(a)7(d). The duration of ablation was set to 120s. As a measure of efficacy, blood pressure was measured at regular intervals before and during the post-test observation period, and we measured the change in blood pressure at three points of time after RDN. During survival, animals were observed clinically. To investigate the short-term, medium-term, and long-term safety of the experimental dogs in the treatment group, three points of time, 7, 30, and 60 days, were included in this experiment. On postoperative days 7 (±1), 30 (±3) and 60 (±5) (with the day of surgery as 0 day), renal arteriograms were performed in animals scheduled for dissection to visualize the test site and renal arteries with nerves were harvested for immuno-histochemical evaluation. Animals were euthanized by deep anesthesia and an overdose of potassium chloride for autopsy and histopathological analysis. Quantification of protein denaturation and necrosis of renal sympathetic nerves around renal arteries was performed using hematoxylin-eosin (H&E) staining.

Results

Simulation results of NetrodTM Six-electrode

As the ablation time increases, the ablation area expands, with data detailed in Table 3.

With prolonged ablation time, the boundary of the region where the temperature reaches 45°C expands. Since temperature-controlled RF ablation was employed, the maximum temperature was confined to the vicinity of the electrodes, as shown in Table 3. The Netrod™ Six-electrode device can reach temperatures up to 45°C at distances of approximately 5.08 mm (>4 mm) from the electrode.

As shown in Fig 8, The NetrodTM Six-electrode can work simultaneously, and the mesh basket structure allows the electrodes to deliver RF energy close to the vessel wall. The NetrodTM Six-electrode can achieve 360° total ablation of sympathetic nerves in 30s.

thumbnail
Fig 8. The effective ablation angle of NetrodTM Six-electrode(t

 = 30s).

https://doi.org/10.1371/journal.pone.0323336.g008

The continuity of the NetrodTM Six-electrode ablation zone at an ablation time of 30s is shown in Fig 9. Each of the NetrodTM Six-electrode sits on a separate mesh basket of wires, which better conforms to the vessel wall as the lumen changes, keeping the electrode arrangement consistent.

thumbnail
Fig 9. The continuity of the ablation area between the NetrodTM Six-electrode(t 

= 30s).

https://doi.org/10.1371/journal.pone.0323336.g009

Results of phantom experiments

As shown in Fig 10, during the ablation process, the six electrodes discharge to form a continuous 360° spiral ribbon, with the ablation ranges of the first and sixth electrodes overlapping radially. The RF ablation experiments resulted in an ablation zone with a width of 6–8 mm and a depth of 3–6 mm, as depicted in Fig 10.

The results from the phantom experiments validate the continuous ablation zone of the Netrod™ Six-electrode observed in the simulation. These experiments demonstrate that, despite the discrepancies in depth, the width of the ablation zone remains consistent, further supporting the accuracy of the simulation model. By comparing the body model experiments with the simulation results, we can confirm the effectiveness of the six electrodes in achieving a 360° spiral ablation. This consistency not only enhances confidence in the simulation model but also provides a reliable basis for clinical applications, indicating that the electrodes can effectively control ablation range and depth during actual procedures.

Animal experiment results

The surgeon performed radiofrequency ablation of the left and right renal arteries of the hypertensive model animals according to the protocol (9 animals in the treatment group) or a sham procedure with renal artery angiography only (9 animals in the control group). There were no intraoperative abnormalities or serious adverse events related to animal vital signs, and the operability of the system was rated as excellent by the surgeon.

The group of 18 experimental animals was mentioned in the previous section. The blood pressure of the animals was measured before and periodically after the procedure, and the blood pressure reduction data at different points of time are shown in Fig 11. Of the three animals in the test group at the point of time of 7 (±1) days, post-ablation diastolic and mean arterial pressures in animal 101 may not be statistically different relative to pre-ablation due to individual differences (P ≥ 0.05). In the other two animals, systolic, diastolic, and mean arterial pressure were significantly lower than before ablation (P ≤ 0.05). Blood pressure was significantly lower than before ablation (P ≤ 0.05) in all six animals in the test group at the 30 (±3) and 60 (±5) day point of time. Pre-ablation systolic, diastolic, and mean arterial pressures in the nine control animals during this procedure were only marginally reduced or increased relative to pre-sham procedure, with no significant differences (P ≥ 0.05).

thumbnail
Fig 11. Values of blood pressure reduction in animals of the treatment group at different point of time.

https://doi.org/10.1371/journal.pone.0323336.g011

The body weights of the experimental animals were not significantly abnormal during the experimental period, and the clinicopathological indices of the animals did not show any significant health-related abnormalities in the preoperative and pre-dissection test results. During the postoperative observation period, the experimental animals did not show any abnormalities related to the treatment. No abnormalities such as stenosis or dissection were observed in the renal arteries of the animals when the angiography was performed before dissection, and no abnormalities were observed when the animals were dissected for observation and sampling. The above experiments prove the safety of the device. The results of the validity analysis in the histopathological analysis showed that nerve fiber damage around bilateral renal artery vessels in animals of the treatment group at each point of time, showing different degrees of vacuolar degeneration, hyaline degeneration and coagulative necrosis. Degeneration of nerve fibers around the renal artery vessels was also seen in pathological section samples from most of the control animals, which was considered to be a nerve injury reaction caused by the sham operation itself. In the histopathological analysis, we measured the maximum distance from the point of ablation to the nerve fiber injury in animals in the treatment group. The ablation depth measurement was performed by a professional animal pathologist to identify the damaged nerve fibers furthest from the ablation location, which was automatically measured by pathology image software, and the measured depth is shown schematically in Fig 12. The distribution of ablation depth data is shown in Fig 13.

thumbnail
Fig 13. Distance of nerve damage at different points of time.

https://doi.org/10.1371/journal.pone.0323336.g013

The distance of nerve injury at the three points of time was not particularly different, and it was inferred that reversible repair of the injured nerve would not occur over time. The results of H&E staining of the ablation sites showed that the intima of the renal artery vessels in all groups of animals showed varying degrees of thickening and inflammatory infiltration. None of them showed desquamation, dissection, calcification or organization. Occasionally, there is thrombus formation in the lumen, which was considered to be caused by blood clotting in the blood vessels during animal dissection. Vascular injury and inflammatory infiltration of the tunica media were present to varying degrees in all groups of animals and no contractures or pseudoaneurysm were present. Occasional hyaline degeneration of the muscularis was seen in both groups, which was considered to be caused by ablation or the procedure itself. Different degrees of necrosis of perivascular tissue, inflammatory infiltration, fibrosis and fatty infiltration were seen in the adventitia and perivascular tissue of renal arteries in both groups, and the overall pathological irritant response was more pronounced in the test group than in the control group. An overview of the efficacy and safety index pathological results is shown in Fig 14 for representative samples at different time periods, with animal number 101 (7 days), 104 (30 days) and 107 (60 days) for the treatment group and 201 (7 days), 204 (30 days) and 207 (60 days) for the control group.

thumbnail
Fig 14. H&E staining of nerves (inflammatory cells shown by yellow arrows; tissue necrosis shown by blue arrows; fatty infiltration shown by green arrows; intimal thickening shown by red arrows).

https://doi.org/10.1371/journal.pone.0323336.g014

Fig 15A and 15B show the peripheral nerve of the renal artery after ablation, and a large number of vacuolar changes of nerve cells in the nerve bundle are seen; C and D are the peripheral nerve of the renal artery in the control group, and no significant abnormal changes were seen in the nerve bundles. The H&E staining pathological pictures of the canine renal artery peripheral nerve 42 days after the controlled experimental radiofrequency ablation of the renal artery are shown in Fig 16. The peripheral nerve bundle of the renal artery in the test group showed morphological changes and thickening of the peripheral sheath. No significant abnormal changes were seen in the peripheral nerve bundle of the renal artery in the control group. Combining the pathological results of the above efficacy and safety indicators, it was inferred that the treatment could effectively ablate perivascular nerve tissue and cause irreversible damage to nerve fibers. The degree of local inflammation and other pathological reactions in the treatment group was significantly greater than that in the control group at all points of time, which was considered to be caused by intraoperative ablation damage to tissues, but did not show significant safety problems, and all of them decreased with the prolongation of postoperative time.

thumbnail
Fig 15. H&E staining of the peripheral nerve of the renal artery.

https://doi.org/10.1371/journal.pone.0323336.g015

thumbnail
Fig 16. H&E staining of the peripheral nerve of the renal artery after 42 days.

https://doi.org/10.1371/journal.pone.0323336.g016

Discussion

The simulation results showed that the maximum injury depth range of the NetrodTM Six-electrode is 5.08 mm and the depth of the ablation zone obtained in the phantom experiment was 3–6 mm. Circumferential coverage was up to 91.99% and the ablation zone was continuous. In animal experiments, eight animals in the treatment group had significantly lower blood pressure (88.9%). Nerve cells in nerve bundles showed vacuolar changes and thickening of peripheral sheaths after peripheral nerve ablation of renal arteries. The NetrodTM Six-electrode presented here has potential to effectively and safely ablate renal nerves.

The study reports a significant reduction in blood pressure in 88.9% of the treated animals, highlighting the potential efficacy of the NetrodTM system in managing hypertension. Renal denervation has emerged as a promising therapeutic option. The findings suggest that the NetrodTM system could offer a safe and effective treatment option for patients with resistant hypertension who do not respond adequately to pharmacological therapies. The demonstrated efficacy and safety in animal models provide a strong foundation for clinical trials in humans. If successful, this technology could expand treatment options for hypertension and potentially other conditions influenced by sympathetic overactivity, such as heart failure or chronic kidney disease. The research findings on the NetrodTM Renal Denervation RF Generator and its Six-electrode system highlight its potential as a safe and effective tool for renal denervation. Further clinical studies are warranted to validate these findings in human patients and explore broader applications in cardiovascular and renal diseases.

AI Raisi et al. [30] reported a circumferential coverage angle of 54.90° ± 4.36° for a single-electrode RF application, encompassing approximately 15% of the arterial wall circumference. In contrast, the circumferential coverage angle that can be achieved in the Netrod™ six-electrode simulation experiment is 360°. Yanyan Cheng et al. noted that the placement of spiral electrodes could cause blood to flow spirally along the catheter, potentially resulting in more significant blood flow disturbances. Following balloon electrode placement in the renal artery, normal blood flow is obstructed, resulting in maximum blood flow velocity at the front of the electrode [31]. The mesh basket design of the Netrod™ Six-electrode effectively mitigates blood flow obstruction caused by both the spiral structure and the balloon.

The relevant parameters (including size and material) of the electrodes used in the ex vivo and animal experiments are the same as those of the electrodes used in the simulation. However, our preliminary research has some limitations. First, we utilized a simplified model of the Netrod™ Six-electrode and the renal arteries, omitting the full structure of the Netrod™ Six-electrode and only simulating the electrode positions. Although the electrode positions can vary and adhere to the vessel wall according to the lumen diameter, the contact force between the electrodes and tissue was not considered in our model. While simulation studies provide valuable insights, there remains a significant gap between these findings and their clinical application. In a clinical setting, factors such as patient-specific anatomy, variability in tissue properties, and real-time physiological responses can significantly influence the outcomes of RF ablation procedures. Secondly, the dynamic nature of blood flow and its interaction with the ablation device must be carefully assessed to optimize treatment effectiveness and minimize complications. Additionally, the animal experiments were selected from 18 female dogs, and it was not possible to determine whether gender had an effect on the results of the experiment. And experimental dogs may not reflect the characteristics of human arteries with hypertension. Depth data on the canine model does not fully reflect ablation depths in other animals and in humans. Finally, in this study, we verified the effectiveness of this system through a modeling study, and the animal experiment was a further step to verify the effectiveness and safety of this system, as the data in the modeling study was set up in an ideal state, and there was a gap between the depth of the animal experiment and the animal experiment due to the real arterial structure of the animals and the shape of the arteries.The depth measured in animal studies was based on the furthest nerve damaged, rather than soft tissue involved. Hence, the depth of ablation could be underestimated inherently, if the distribution of nerves surrounding the renal artery segment was scarce beyond 3 mm and this as one of the reasons why there were discrepancies between the depths observed in different settings.

To bridge this gap, future studies should focus on integrating more complex and realistic models that account for these variables, including variations in blood flow dynamics and tissue response during ablation. Clinical trials will also be essential to validate the efficacy and safety of the Netrod™ Six-electrode in diverse patient population demographics, ensuring that laboratory findings translate effectively into real-world applications.

Conclusion

Through mathematical modeling, phantom experiments and animal study, we established a novel renal denervation system based on the Netrod™ Six-electrode. This system utilizes automatic temperature control and a mesh basket structure to target the nerves around the renal arteries. Simulation studies of the Netrod™ Six-electrode demonstrated that heat distribution effectively achieved denervation of the renal sympathetic nerves while maintaining a constant electrode temperature of 60°C.

This paper focuses on mathematical modeling, phantom experiments and animal studies to demonstrate the feasibility of the proposed new technique. We performed renal denervation in hypertensive dogs model and in an open surgical setting to show the efficacy and safety of the system in vivo. Further clinical studies are needed to validate the effectiveness and safety of the NetrodTM Six-electrode denervation system for blood pressure control in vivo.

References

  1. 1. Kearney P, Whelton M, Reynolds K. Global burden of hypertension: analysis of worldwide data. Lancet. 2005;365(9455):217–23.
  2. 2. Calhoun DA, Jones D, Textor S, Goff DC, Murphy TP, Toto RD, et al. Resistant hypertension: diagnosis, evaluation, and treatment: a scientific statement from the American Heart Association Professional Education Committee of the Council for High Blood Pressure Research. Circulation. 2008;117(25):e510-26. pmid:18574054
  3. 3. Yang X, Lin L, Zhang Z, Chen X. Effects of catheter-based renal denervation on renin-aldosterone system, catecholamines, and electrolytes: A systematic review and meta-analysis. J Clin Hypertens (Greenwich). 2022;24(12):1537–46. pmid:36321724
  4. 4. Sharp ASP, Tunev S, Schlaich M, Lee DP, Finn AV, Trudel J, et al. Histological evidence supporting the durability of successful radiofrequency renal denervation in a normotensive porcine model. J Hypertens. 2022;40(10):2068–75. pmid:35866489
  5. 5. Díaz-Morales N, Baranda-Alonso EM, Martínez-Salgado C, López-Hernández FJ. Renal sympathetic activity: A key modulator of pressure natriuresis in hypertension. Biochem Pharmacol. 2023;208:115386. pmid:36535529
  6. 6. Esler MD, Krum H, Schlaich M, Schmieder RE, Böhm M, Sobotka PA, et al. Renal sympathetic denervation for treatment of drug-resistant hypertension: one-year results from the Symplicity HTN-2 randomized, controlled trial. Circulation. 2012;126(25):2976–82. pmid:23248063
  7. 7. Esler MD, Böhm M, Sievert H, Rump CL, Schmieder RE, Krum H, et al. Catheter-based renal denervation for treatment of patients with treatment-resistant hypertension: 36 month results from the SYMPLICITY HTN-2 randomized clinical trial. Eur Heart J. 2014;35(26):1752–9. pmid:24898552
  8. 8. Krum H, Schlaich MP, Sobotka PA, Böhm M, Mahfoud F, Rocha-Singh K, et al. Percutaneous renal denervation in patients with treatment-resistant hypertension: final 3-year report of the Symplicity HTN-1 study. Lancet. 2014;383(9917):622–9. pmid:24210779
  9. 9. Linz D, Hohl M, Elliott AD, Lau DH, Mahfoud F, Esler MD, et al. Modulation of renal sympathetic innervation: recent insights beyond blood pressure control. Clin Auton Res. 2018;28(4):375–84. pmid:29429026
  10. 10. Götzinger F, Kunz M, Lauder L, Mahfoud F, Böhm M. Radio frequency-based renal denervation: a story of simplicity?. Future Cardiol. 2023;19(9):431–40. pmid:37791469
  11. 11. Goyal A, Jain H, Verma A, Jain J, Shamim U, Kanagala SG, et al. The role of renal denervation in cardiology and beyond: An updated comprehensive review and future directives. Curr Probl Cardiol. 2024;49(2):102196. pmid:37952794
  12. 12. Zuhair M, Keene D, Kanagaratnam P, Lim PB. Percutaneous Neuromodulation for Atrial Fibrillation. Card Electrophysiol Clin. 2024;16(3):281–96. pmid:39084721
  13. 13. Doiron JE, Li Z, Yu X, LaPenna KB, Quiriarte H, Allerton TD, et al. Early Renal Denervation Attenuates Cardiac Dysfunction in Heart Failure With Preserved Ejection Fraction. J Am Heart Assoc. 2024;13(4):e032646. pmid:38353216
  14. 14. Wu Y, Song M, Wu M, Lin L. Advances in device-based treatment of heart failure with preserved ejection fraction: evidence from clinical trials. ESC Heart Fail. 2024;11(1):13–27. pmid:37986663
  15. 15. Liang X, Shang S, Bai Z, et al. Effects of Renal Denervation on Gap Junction in High-Pacing-Induced Heart Failure Dogs. Anatol J Cardiol. 2024 Jul 16;28(9):429–36.
  16. 16. Takahashi K, Egami Y, Nishino M, Tanouchi J. Clinical impact of stellate ganglion phototherapy on ventricular tachycardia storm requiring mechanical circulatory support devices: a case report. Eur Heart J Case Rep. 2024;8(4):ytae177. pmid:38690554
  17. 17. White WB, Galis ZS, Henegar J, Kandzari DE, Victor R, Sica D, et al. Renal denervation therapy for hypertension: pathways for moving development forward. J Am Soc Hypertens. 2015;9(5):341–50. pmid:25979410
  18. 18. Fang Z, Moser M, Zhang E. Design of a novel electrode of radiofrequency ablation for large tumors: In vitro validation and evaluation. J Biomech Eng. 2019;141(3).
  19. 19. Pop M, Molckovsky A, Chin L, Kolios MC, Jewett MAS, Sherar MD. Changes in dielectric properties at 460 kHz of kidney and fat during heating: importance for radio-frequency thermal therapy. Phys Med Biol. 2003;48(15):2509–25. pmid:12953912
  20. 20. González-Suárez A, Berjano E. Comparative Analysis of Different Methods of Modeling the Thermal Effect of Circulating Blood Flow During RF Cardiac Ablation. IEEE Trans Biomed Eng. 2016;63(2):250–9. pmid:26258928
  21. 21. Plourde B, Vallez L, Sun B. The effect of plaque removal on pressure drop and flow rate through an idealized stenotic lesion. Biol Med. 2015;8(1).
  22. 22. Dong T, Cheng Y, Tian Z. Ex-vivo experimental study in hepatic vascular to simulate the radiofrequency ablation of renal sympathetic nerve. Beijing Biomed Eng. 2018;37(03):265–72.
  23. 23. Dong T, Nan Q, Tian Z, Cheng Y. Radiofrequency ablation of renal sympathetic nerve: Numerical simulation and ex vivo experiments. Advances in Mechanical Engineering. 2018;10(6).
  24. 24. Shao Y L, Leo H L, Chua K J. Studying the thermal performance of a bipolar radiofrequency ablation with an improved electrode matrix system: In vitro experiments and modelling J. Applied Thermal Engineering. 2017:116.
  25. 25. Zheng F, Bing Z, Michael M. Design of a novel electrode of radiofrequency ablation for large tumors: A finite element study. J Eng Sci Med Diagn Ther. 2018;1(1).
  26. 26. Negussie AH, Partanen A, Mikhail AS, Xu S, Abi-Jaoudeh N, Maruvada S, et al. Thermochromic tissue-mimicking phantom for optimisation of thermal tumour ablation. Int J Hyperthermia. 2016;32(3):239–43. pmid:27099078
  27. 27. Chen Y, Youn P, Furgeson DY. Thermo-targeted drug delivery of geldanamycin to hyperthermic tumor margins with diblock elastin-based biopolymers. J Control Release. 2011;155(2):175–83. pmid:21846483
  28. 28. Huang Y, Hu B, Shen E. Experimental study of interaction between radiofrequency ablation and realtime temperature monitoring. Chin J Interv Imaging Ther. 2010;7(3):296–300.
  29. 29. Zhang BL, Hu B, Kuang SL, Huang Y, Wu R, Jiang LX. A tissue-mimicking phantom for visualizing coagulation temperature distribution at the same range of biological tissue during thermal ablation. Chin J Med Imaging Technol. 2007;2007(09):1405–8.
  30. 30. Al Raisi SI, Pouliopoulos J, Barry MT, Qian P, Thiagalingam A, Swinnen J, et al. Renal Artery Branch Denervation: Evaluation of Lesion Characteristics Using a Thermochromic Liquid Crystal Phantom Model. Heart Lung Circ. 2020;29(3):445–51. pmid:30967336
  31. 31. Cheng Y, Liu H, Tian Z, Zhang M, Liu Y, Nan Q. Evaluating the thermal performance of a balloon-based renal sympathetic denervation system with array electrodes: a finite element study. Electromagn Biol Med. 2021;40(4):488–501. pmid:34352188