Biomechanical assessment of upright birthing positions using marker-based and markerless motion capture systems

  • Lauren Haworth ,

    Contributed equally to this work with: Lauren Haworth

    Roles Conceptualization, Data curation, Formal analysis, Funding acquisition, Investigation, Methodology, Project administration, Resources, Software, Validation, Visualization, Writing – original draft, Writing – review & editing

    Affiliation Allied Health Research Unit, University of Lancashire, Preston, United Kingdom

  • Inderjeet Kaur,

    Roles Conceptualization, Funding acquisition, Methodology, Validation, Writing – original draft, Writing – review & editing

    Affiliation Fernadez Hospitals, Fernandez Foundation, Hyderabad, Telangana, India

  • Maimoona Ahmed,

    Roles Conceptualization, Funding acquisition, Methodology, Writing – original draft, Writing – review & editing

    Affiliation Fernadez Hospitals, Fernandez Foundation, Hyderabad, Telangana, India

  • Anastasia Topalidou

    Roles Conceptualization, Data curation, Formal analysis, Funding acquisition, Investigation, Methodology, Project administration, Resources, Software, Supervision, Validation, Visualization, Writing – original draft, Writing – review & editing

    ATopalidou@uclan.ac.uk

    Affiliation School of Nursing and Midwifery, University of Lancashire, Preston, United Kingdom

Abstract

Understanding the effects of birthing positions on labour is crucial for optimising maternal and foetal outcomes. Upright positions are encouraged but their biomechanics are not fully understood. Biomechanical changes during labour can make certain positions more or less favourable depending on individual physical characteristics. Understanding these factors is essential for tailoring strategies that enhance maternal comfort and facilitate labour. This study aimed to quantify the biomechanics of seven common upright birthing positions, comparing their biomechanical characteristics and evaluating the sensitivity and accuracy of marker-based and markerless motion capture systems. Fifteen healthy, non-pregnant women performed seven upright birthing positions. Hip, pelvis, and trunk kinematics were assessed using a 9-camera marker-based system and an 8-camera markerless system. Significant biomechanical differences were found between birthing positions. The “squat” position showed the most hip flexion and abduction, “B-Ball” had the greatest anterior pelvic tilt, and “all-fours” exhibited the most posterior tilt. For the trunk, “upright” led to the most extension, and “elbows bent knees” showed the most flexion. Agreement between systems varied by joint and plane: it was moderate to strong for hip angles in the coronal plane and for pelvic angles in the transverse plane, but limited in the sagittal plane and inconsistent for trunk measures. This study highlights the biomechanical differences in upright birthing positions and emphasises the need for personalised birthing strategies. Understanding labour biomechanics is crucial for improving maternal and foetal well-being and reducing complications. By providing comprehensive and evidence-based information, women can make informed decisions about their birthing positions, enhancing outcomes and lowering the risk of maternal and neonatal complications globally.

Introduction

Childbirth is a pivotal event in a woman’s life. Understanding the effects of birthing positions on the labour process and optimising these positions is crucial for improving maternal and foetal outcomes [1]. The World Health Organisation (WHO) encourages the adoption of upright positions (flexible sacrum positions) during labour due to their association with favourable childbirth outcomes [2]. However, the scientific understanding of the biomechanics of various birthing positions – recumbent, semi-recumbent, and upright – remains limited [3]. This limitation is primarily due to the constraints of methodologies available to collect and analyse biomechanical data and correlate them with maternal and neonatal outcomes. Methods such as pelvimetry [4,5], magnetic resonance imaging [6,7], and marker-based motion capture technology [8,9] have inherent limitations that prevent their use during labour. Additionally, these methods cannot effectively capture the dynamic and continuous nature of the birthing process [49]. This gap in knowledge impedes a comprehensive understanding of how different positions affect the birthing process. While computational modelling studies provide important insights, they cannot address this knowledge gap due to their inherent limitations, simplifications, and assumptions [10,11].

Several postural approaches have been proposed as alternatives for women’s positions during delivery. While qualitative studies suggest these methods enhance maternal satisfaction [12,13], their effectiveness in promoting spontaneous vaginal births requires further investigation [3,14]. Theoretically, the optimal birthing position aligns the axis of foetal progression perpendicularly to the superior pelvic inlet, minimising obstacles by adjusting the lumbar dorsal hinge [8]. However, individual anatomical variations and maternal comfort may influence the feasibility of this alignment in clinical practice. Different birthing positions exert unique biomechanical effects, influencing both labour progression and maternal experience. For instance, pelvic tilt and sagittal plane hip angle play significant roles in foetal descent and overall labour mechanics [3,8].

Research into the biomechanics of labour primarily concentrates on the angles of the pelvis and hips [10,11], often neglecting equally crucial elements such as the spine, knees, and other vital factors, including body shape and weight. It is well understood that the human body functions in a domino-like fashion, where a modification in one area directly affects another in a kinetic chain. However, limited research has focused on the comprehensive biomechanical aspects of birthing positions, considering the global posture of the labouring woman [9,15] rather than just specific parts. Understanding this is vital for creating optimal conditions for foetal progression and improving childbirth outcomes [3]. The spine and pelvis are closely interconnected through the skeletal and muscular systems, and alterations in one area can influence the other. For instance, a change in the lumbar spine’s curvature can alter the tilt of the pelvis [16,17]. This is because the pelvis and the lower spine work together to support and balance the body, especially during movement and weight-bearing activities. Changes in spinal alignment, therefore, can lead to compensatory changes in the pelvic angle to maintain overall balance and function [18]. The hips, knees, ankles, feet, and lower back form an interconnected kinetic chain, crucial for movement and stability during childbirth. Dysfunction or limited motion in any part of this chain, such as the knees, can necessitate compensatory adjustments in adjacent areas, like the hips. For example, restricted knee mobility or angulations may alter the optimal hip positioning [19] and affect the choice of childbirth positions. Understanding these connections is essential for selecting positions that accommodate individual biomechanical needs and facilitate a smoother labour process. Among other factors, the shape and size of the pelvis [20], along with individual variations in the physical characteristics of lower extremities [21], play a crucial role in determining the effectiveness and comfort of various labour positions. Additionally, variations in abdominal weight and overall body mass result in shifts in the centre of gravity [22,23] and increased pressure on the spine and pelvis [24]. These biomechanical changes influence which labour positions may be more or less favourable for a given individual. Understanding these factors is essential for tailoring labour strategies to enhance comfort and facilitate the birthing process. These biomechanical variations highlight the necessity of tailoring and personalising positioning recommendations and support by maternal health providers to accommodate individual maternal needs and preferences. However, without the biomechanical knowledge to investigate these positions and understand their effects, due to the aforementioned limitations, we cannot provide biomechanically evidenced recommendations.

Recently, biomechanists around the world have started to investigate recumbent and semi-recumbent positions due to the practicalities in using motion capture techniques, and some upright position such as squatting [3,8,9,15]. To overcome the issues with marker-based systems that do not allow investigation in a real clinical environment, some attention has shifted towards newly developed markerless systems [8], although there is not yet a comprehensive understanding of their capabilities.

To advance understanding of available motion capture methods and their feasibility in real-world, non-laboratory settings, this study compares two systems: the reference standard marker-based system and a novel markerless system. The aim is to evaluate their effectiveness and potential advantages in assessing whole-body biomechanics across commonly used upright birthing positions. To the best of our knowledge, this is the first study to apply markerless motion capture technology for this purpose, providing a more comprehensive and nuanced perspective on the biomechanics of different birthing positions. The study was guided by the following hypotheses and questions:

  1. Hypothesis 1: Different upright birthing positions will exhibit distinctly different biomechanical characteristics of the spine, torso, pelvis, and hips.

Research Question 1: What are the specific biomechanics of upright birthing positions on the characteristics of the spine, torso, pelvis, and hips?

  1. Hypothesis 2: The markerless motion capture system will provide comparable sensitivity to change among different upright birthing positions compared to the marker-based motion capture system.

Research Question 2: To what extent does the markerless motion capture system, compared to the marker-based, accurately capture and detect changes among different upright birthing positions?

Materials and methods

Design

A within-subjects, repeated measures design was used to analyse changes in postural kinematic parameters in seven upright birthing positions. The research was developed in collaboration with the public and stakeholders, ensuring it was conducted ‘with’ and ‘by’ them rather than merely ‘to,’ ‘about,’ or ‘for’ them. The stakeholders and the Patient and Public Involvement and Engagement (PPIE) group influenced the development of the study’s methodological approach and the choice of birthing positions to evaluate, focusing on those most widely used worldwide and the practicality of gathering data in a timely manner. Specifically, stakeholders contributed insights into the clinical relevance of positions commonly used in high-, middle-, and low-income countries, while the PPIE group supported protocol refinement through discussion sessions and lab-based mock-ups, offering feedback on comfort, acceptability, and feasibility. By adopting this cooperative strategy, the study was made relevant and reflective of the needs and viewpoints of the intended audience [25,26].

Participants

Healthy non-pregnant females (biological sex) of reproductive age (18–49 years) [27], who were free from any injury, pain, illness, or medical condition that would limit their ability to perform the specific positions were recruited through campus-based advertisements and social media posts. In accordance with previously published rules of thumb for pilot studies [2830], a sample size of n = 15 participants was determined for this exploratory study.

Equipment

Data were collected synchronously using two motion capture systems: a 9-Oqus-camera marker-based motion capture system (Qualisys Medical AB, Sweden) and an 8-Micus-camera markerless motion capture system (Qualisys, Medical AB, Sweden) [31]. Data from both systems were collected simultaneously in Qualisys Track Manager v2021.2 (Qualisys, Medical AB, Sweden), ensuring the systems were synchronised and calibrated in space and time by applying the same wand calibration [32]. A target calibration residual error below 0.5 mm was set, and all calibrations achieved residuals under this threshold. Footage from the markerless system was recorded at 50 Hz and data from the marker-based system was recorded at 100 Hz, as per previous comparative studies [3133]. Sampling frequencies vary due to the different capabilities of the two systems. Markerless systems typically use lower sampling frequencies due to the constraints of the frame rate of the RGB video cameras within the system [31]. Reducing the sampling frequency of the marker-based system to match the markerless system may affect the precision and accuracy of the marker-based data [34]. A higher frame rate in the markerless system would increase the computational load during data processing, impacting real-time application [31,34].

Data collection

Data was collected between 6th July and 24th October 2023. Participants attended a 45-minute data collection session in the University’s Motion Analysis Laboratory. Following participant consent, age, height, and weight were recorded. To maximise the quality of the data, all participants wore athletic style shorts and an adapted t-shirt which afforded access to the spine.

Data collection using the marker-based system requires the placement of passive retro-reflective markers using the calibrated anatomical system technique to afford tracking of segmental kinematics in 6 degrees of freedom [35]. Markers were attached using dermatological friendly double-sided adhesive tape, and were applied to the acromions, sternal notch, 7th cervical vertebra (C7), anterior superior iliac spines (ASIS), posterior superior iliac spines (PSIS), greater trochanters, and the medial and lateral epicondyles. Clusters of four non-collinear markers were attached to the thighs and spinal clusters were applied level with the 7th thoracic vertebra, 3rd and 5th lumbar vertebra [36,37]. Using this configuration of markers, the study effectively delineated anatomical segments including the Lower Thoracic (LT), Upper Lumbar (UL) and Lower Lumbar (LL) spinal segments, a thorax (defined using the anatomical locations of the acromions, sternal notch, 7th cervical vertebra and ASIS), a pelvis (defined using the anatomical locations of the ASIS and the PSIS), and hips (Fig 1).

thumbnail
Fig 1. Marker set used for the marker-based motion capture system.

https://doi.org/10.1371/journal.pone.0322970.g001

The seven upright birthing positions investigated were:

  1. Standing upright [Upright] (Fig 2A): Participants were instructed to stand up straight with feet shoulder-width apart. They were asked to keep their palms facing their legs and look straight ahead at a specific point indicated to them, maintaining a natural and relaxed posture without stiffening their shoulders or arms.
  2. Leaning forward – support on the palms [Palms] (Fig 2B): Participants were instructed to stand beside the bed and lean forward, placing their palms on the bed at a width that felt comfortable to them, with their elbows extended. They were asked to keep their knees straight and their feet at a width that was comfortable for them.
  3. Leaning forward – support on elbows [Elbows, extended knees] (Fig 2C): As above, participants were instructed to stand beside the bed and lean forward, supporting their weight on their elbows and forearms, which were placed on the bed. They were asked to cross their palms, forming a triangle. The feet were positioned as described previously, with the knees kept straight.
  4. Leaning forward – support on elbows – knees bent [Elbows, bent knees] (Fig 2D): Participants were instructed to stand beside the bed and lean forward, supporting their weight on their elbows and forearms, which were placed on the bed. They were asked to cross their palms, forming a triangle. Their feet were placed at a width that was comfortable for them, and they were to bend their knees to an angle of 90–120 degrees, according to their own comfort and preference.
  5. Deep squat [Squat] (Fig 2E): Participants were instructed to stand to the side of the bed. With their palms placed on the bed for support, but not holding onto the bed, participants were asked to perform a deep squat within a comfortable range.
  6. Hands and knees [All-4s] (Fig 2F): Participants were instructed to climb onto the bed and assume an all-fours position, with the spine in a neutral alignment and their palms placed directly under their shoulders, ensuring proper support and balance. Additionally, they were asked to look at a specific point to keep their neck straight, ensuring that the C7 marker and the sternal notch marker were not obscured. This alignment was crucial for accurate data collection and to maintain the integrity of the marker placements.
  7. Sitting on a birthing ball [B-Ball] (Fig 2G): Participants were instructed to sit on a birthing ball with their legs spread wide to ensure balance and comfort. They were asked to place their hands on their thighs in a position that did not obstruct any markers. Additionally, they were directed to focus their gaze on a specific point that was indicated to them. The size of the ball was chosen according to the participant’s height. To maintain stability, participants ensured their feet were fully planted on the floor.
thumbnail
Fig 2. Birthing positions: 2A. Upright, 2B. Palms, 2C. Elbows with knees extended, 2D Elbows with knees bent, 2E. Squat, 2F. All-4s, 2G. B-Ball.

https://doi.org/10.1371/journal.pone.0322970.g002

During protocol development, and in combination with PPIE consultation, a formula was developed to set the height of the bed according to each participant’s height. This provided a practical solution, enabling standardisation relative to each participant’s stature, whilst also ensuring comfort and consistency during data collection. Maintaining a consistent reference point relative to body height is crucial for obtaining reliable kinematic data. The following formula was used: (Participant’s height in inches) + 1 = height of physio bed in cm.

Participants maintained each of the seven birthing positions for six seconds and assumed each position three times. As each position is considered as a static hold exercise, participants were required to remain motionless, counteracting the effects of gravity throughout the duration of the task. Considering the characteristics of participants who were ‘healthy’ but not necessarily athletically trained, and the number of required repetitions per task, six seconds was deemed a suitable amount of time to ensure good data was collected without causing potential muscle fatigue amongst participants. To reduce bias and control for the potential influence of confounding factors, such as the effects of any extraneous variables that might affect the results on the outcome measures being studied, the order of the seven positions was randomised individually for each participant using the online Research Randomizer Tool (http://www.randomization.com). Participants were given a minimum of 30-second intervals between positions, with additional time provided upon request. After completing all seven birthing positions once, the sequence was repeated twice more, maintaining the same randomisation order for each cycle, thus ensuring consistency across the three rounds of data collection. The data collection session concluded once all three sequences were completed.

Data processing and analysis

The time domain of the measurements for both systems was time normalised to 101 data points [31]. As the marker-based system is considered the reference standard, the most stable two seconds of data were used for each trial, which was determined using manual event determination. Events for the markerless data were then synchronised with the events from the marker-based data.

Marker-based data.

Anatomical frames were defined by landmarks positioned at the medial and lateral borders of the joint, from these right-handed segment co-ordinate systems were defined. The kinematics were calculated based on the Cardan sequence of XYZ equivalent to the joint coordinate system [38]. Raw kinematic data were exported from Qualisys Track Manager v2021.2 (Qualisys, AB, Sweden) to Visual3D (C-Motion Inc, USA) and filtered using a 6 Hz Low Pass filter. Intersegmental spinal position (LT segment relative to UL segment, UL segment relative to LL segment and LL segment relative to the pelvis), trunk (trunk segment relative to the pelvis), pelvis (pelvis segment relative to the lab) and hip (thigh segment relative to the pelvis) angles were exported into Microsoft Excel, and the mean position for each trial were found.

Markerless data.

Theia3D (Theia Markerless Inc., Kingston, ON, Canada) is a deep learning algorithm-based approach to markerless motion capture that uses synchronised video data to perform 3D human pose estimation [39,40]. The raw video footage from 8 video cameras was processed using Theia3D (v2023.1.0.3161), from which 4x4 pose matrices of each body segment were exported for analysis into Visual3D (C-Motion Inc, USA). Theia3D embeds a multibody kinematic model consisting of two separate kinematic chains: one for the lower extremity and one for the upper extremity, and a separate head segment (https://www.theiamarkerless.ca/docs/model.html). For this study, from the upper extremity chain, the torso (trunk) segment was used, and from the lower extremity chain, the pelvis and hip joints were used. Trunk, pelvis and hip angles were exported from Visual3D (C-Motion Inc, USA) into Microsoft Office Excel 365 (Microsoft Corp, USA), and the mean position for each trial were found.

Statistical analysis

All statistical analyses were performed using SPSS v28 (IBM Corp., NY, USA). Descriptive statistics were used to describe the characteristics of participants.

To investigate the differences between birthing positions, one-way Repeated Measures Analysis of Variance (ANOVA) tests were performed for each motion capture system separately. This was also undertaken to determine the sensitivity of the markerless system, i.e., whether statistically significant differences between birthing positions were the same. Mean joint angles (°), standard deviations (SD), main effects and effect sizes (partial eta squared, ηp2) were reported. Where a main effect was observed, Least Significant Difference post-hoc pairwise comparisons were performed. Pairwise comparisons, mean differences (MD), and 95% confidence intervals of the mean differences (CI) were reported.

As a further step, to examine the sensitivity of the markerless system, and determine the extent to which data from the markerless motion capture systems matched that from the marker-based system, linear regression models were used for each outcome measurement. As it is considered the reference standard, variables from the marker-based system were selected as dependent variables and those from the markerless system as the independent/ predictor variables. This approach was chosen to reflect our aim of assessing how well the markerless system could replicate the reference-standard outputs of the marker-based system. By defining the marker-based data as the dependent variable, we directly evaluated how closely the markerless system aligned with the reference standard. The R square (R2) was reported as a goodness-of-fit measure (0.0–1.0); greater R2 values represent smaller differences between the two systems. Significance values (p value) were also presented for each linear regression model. Statistical significance was set at the p < 0.05 level throughout.

Ethical considerations and data protection

The study was approved by the University of Central Lancashire – Health Ethics Committee (HEALTH 01203). Data collection conformed to the Declaration of Helsinki [41], and all information collected was kept strictly confidential, with participant anonymity maintained throughout. Data were collected, managed, analysed, and stored in accordance with General Data Protection Regulations (GDPR). Volunteers provided written informed consent prior to participation. During the consent process, participants were briefed on data protection measures and their rights concerning their data.

Results

Participant characteristics

Fifteen healthy non-pregnant participants were included in the study; their characteristics are described in Table 1.

Differences between upright birthing positions: Markerless motion capture system

Hip.

At the hip, significant main effects were observed in the sagittal, coronal and transverse planes (p < 0.001; Table 2). In the sagittal plane, participants assumed the most amount of hip flexion in the “squat” position, and the most amount of hip extension in the “upright” position (Table 9). Apart from when comparing between “B-Ball” and “elbows, bent knees” (p = 0.973), all other positions produced significantly different hip flexion-extension angles compared to each other (p < 0.016; Table 3).

thumbnail
Table 2. Mean (SD), main effects (Sig) and effect sizes (ηp2) for data from the markerless and marker-based motion capture systems.

https://doi.org/10.1371/journal.pone.0322970.t002

thumbnail
Table 3. Markerless system variables with main effects for hip and pelvis positions in the sagittal plane.

https://doi.org/10.1371/journal.pone.0322970.t003

In the coronal plane, the “squat” position produced the greatest amount of hip abduction, while the “upright” position produced the least. With the exception of the comparisons between “all-4s” and “elbows, extended knees” (p = 0.777), “all-4s” and “palms” (p = 0.254), and “B-Ball” and “elbows, bent knees” (p = 0.231; Table 4), all other positions were significantly different from each other (p < 0.006).

thumbnail
Table 4. Markerless system variables with main effects for hip positions in the coronal and transverse planes.

https://doi.org/10.1371/journal.pone.0322970.t004

In the transverse plane, the “squat” position produced the greatest external rotation, while “elbows, extended knees” produced the greatest internal rotation. Non-significant differences were observed when comparing “upright” to “palms” (p = 0.059), “palms” to “elbows, extended knees” (p = 0.074), and “B-Ball” to “all-4s” (p = 0.586; Table 4). All other positions were significantly different from each other (p < 0.018).

Pelvis.

At the pelvis, significant main effects were observed in the sagittal plane (p < 0.001; Table 2). “B-Ball” produced the greatest amount of anterior tilt, while “all-4s” produced the greatest amount of posterior tilt. Except when comparing “all-4s” and “elbows, extended knees” (p < 0.239; Table 3), all other positions were significantly different from each other (p < 0.037).

Trunk.

For the trunk, significant main effects were observed in the sagittal and coronal planes (p < 0.001; Table 2). In the sagittal plane, the “upright” position produced the greatest amount of trunk extension while the “squat” produced the most trunk flexion. All positions were significantly different from each other (p < 0.002), except when comparing “all-4s” and “palms” (p = 0.055), “all-4s” and “B-Ball” (p = 0.194), “elbows, bent knees” and “elbows, extended knees” (p = 0.088), “squat” and “elbows, extended knees” (p = 0.760), and “squat” and “elbows, bent knees” (p = 0.317; Table 5).

thumbnail
Table 5. Markerless system variables with main effects for trunk position in the sagittal and coronal planes.

https://doi.org/10.1371/journal.pone.0322970.t005

In the coronal plane, “all-4s” produced the least side flexion whilst “elbows, extended knees” produced the most side flexion. Except when comparing between “upright” and “B-Ball” (p = 0.069), “upright” and “all-4s” (p = 0.802), “all-4s” and “squat” (p = 0.121), and “all-4s” and “B-Ball” (p = 0.646; Table 5), all other positions were significantly different from each other (p < 0.049).

Differences between upright birthing positions: Marker-based motion capture system

Hip.

At the hip, significant main effects were observed in the sagittal, coronal and transverse planes (p < 0.001; Table 2). In the sagittal plane, the squat position produced the most hip flexion and the upright position produced the most hip extension. Except when comparing between “all-4s” and “elbows, extended knees” (p = 0.207), and “all-4s” and “B-Ball” (p = 0.173; Table 6), all other positions were significantly different from each other (p < 0.007).

thumbnail
Table 6. Marker-based system variables with main effects for hip positions in the sagittal and coronal planes.

https://doi.org/10.1371/journal.pone.0322970.t006

In the coronal plane, the “squat” position produced the greatest amount of hip abduction, and the “upright” position produced the least amount. All positions were significantly different from each other (p < 0.041), except when comparing between “all-4s” and “palms” (p = 0.505), “all-4s” and “elbows, extended knees” (p = 0.505), “B-Ball” and “elbows, bent knees” (p = 0.225), and “palms” and “elbows, extended knees” (p = 0.093; Table 6).

In the transverse plane the “squat” position produced the most external rotation and the “upright” position produced the most internal rotation. Significant differences were observed when comparing the “upright” position with “palms” (p = 0.018), “elbows, extended knees” (p = 0.014), “squat” (p < 0.001) and “B-Ball” (p = 0.012; Table 7). “Palms” was also significantly different from “elbows, bent knees” (p = 0.010) and “squat” (p = 0.002). “Elbows, extended knees” was significantly different from “elbows, bent knees” (p = 0.002) and “squat” (p = 0.004). “Elbows, bent knees” was significantly different from the “squat” (p < 0.001) and “B-Ball” (p = 0.026). The “squat” position was significantly different from “B-Ball” (p = 0.005) and “all-4s” (p < 0.001).

thumbnail
Table 7. Marker-based system variables with main effects for pelvic and hip positions in the sagittal and transverse planes.

https://doi.org/10.1371/journal.pone.0322970.t007

Pelvis.

At the pelvis, significant main effects were observed in the sagittal plane (p < 0.001; Table 2). “B-Ball” produced the most anterior tilt and “all-4s” produced the most posterior tilt. All positions were significantly different from each other (p < 0.015; Table 7).

Trunk.

For the trunk, significant main effects were observed in the sagittal plane (p = 0.002; Table 2). The “upright” position produced the most trunk extension and “elbows, bent knees” produced the most trunk flexion. The “upright” position was significantly different from all other positions (p < 0.005) except “all-4s” (p = 0.855; Table 8). “All-4s” was significantly different from “elbows, extended knees” (p < 0.001), “elbows, bent knees” (p = 0.005), “squat” (p = 0.006) and “B-Ball” (p < 0.001).

thumbnail
Table 8. Marker-based system variables with main effects for trunk and upper lumbar relative to lower lumbar positions in the sagittal plane.

https://doi.org/10.1371/journal.pone.0322970.t008

Upper lumbar segment relative to lower lumbar segment.

Significant main effects were observed in the sagittal plane (p = 0.043; Table 2). The “upright” position produced the most extension and the “elbows, extended knees” produced the most flexion. The “upright” position was significantly different from all other positions (p < 0.036) except “all-4s” (p = 0.353). “Palms” was significantly different from “squat” and “B-Ball” (p < 0.026), and “B-Ball” was significantly different from “all-4s” (p = 0.025; Table 8).

Lower thoracic segment relative to upper lumbar segment.

Significant main effects were observed in the sagittal plane (p < 0.001; Table 2). The “upright” position was significantly different from all other positions (p < 0.026) except “all-4s” (p = 0.893; Table 10). “Palms” was significantly different from “all-4s” (p = 0.010). “Elbows, extended knees” was significantly different from “elbows, bent knees”, “squat”, and “all-4s” (p < 0.019; Table 9). “Elbows, bent knees” was significantly different from “all-4s” (p = 0.012), and “B-Ball” was significantly different from “all-4s” (p = 0.005). The “upright” position produced the most extension and the “B-Ball” position produced the most flexion (Table 10).

thumbnail
Table 9. Marker-based system variables with main effects for lower thoracic relative to upper lumbar positioning in the sagittal plane.

https://doi.org/10.1371/journal.pone.0322970.t009

thumbnail
Table 10. Comparative summary of hip, pelvis, and trunk biomechanics across upright birthing positions.

https://doi.org/10.1371/journal.pone.0322970.t010

Differences between motion capture systems

Hip.

The linear regression models revealed that the data from the markerless motion capture system matched the data from the marker-based system in only one out of seven positions for the hip joint variables in the sagittal plane. However, the markerless system performed better in the coronal plane, matching the data in six out of seven positions (Table 11). In the sagittal plane, the markerless system corresponded with the marker-based system only in the “all-4s” position (R2 0.575, p = 0.002). In the coronal plane, the markerless system matched the marker-based system in the “upright” (R2 0.799, p < 0.001), “palms” (R2 0.454, p < 0.008), “elbows, extended knees” (R2 0.639, p < 0.001), “elbows, bent knees” (R2 0.714, p < 0.001), “squat” (R2 0.653, p < 0.015), and “B-Ball” (R2 0.630, p < 0.001) positions, with a trend towards significance in “all-4s” (R2 0.270, p < 0.057). Importantly, in the transverse plane, no statistically significant relationships were found between the markerless and marker-based systems for any of the positions (all p > 0.05), suggesting limited alignment in this plane for hip joint angles.

thumbnail
Table 11. Predictive relationships between markerless and marker-based joint angle data across birthing positions.

https://doi.org/10.1371/journal.pone.0322970.t011

Pelvis.

The linear regression models revealed that the data from the markerless motion capture system matched the data from the marker-based system in only two out of seven positions for pelvis variables in the sagittal plane. The markerless system performed better in the coronal and transverse planes, matching the data in four (coronal) and five (transverse) out of seven positions (Table 11). In the sagittal plane, the markerless system corresponded with the marker-based system in the “squat” (R2 0.680, p = 0.003) and “B-Ball” positions (R2 0.510, p = 0.004). In the coronal plane, the markerless system agreed with the marker-based system in “elbows, extended knees” (R2 0.829, p < 0.001), “elbows, bent knees” (R2 0.875, p < 0.001) and “B-Ball” (R2 0.510, p = 0.004) positions. In the transverse plane, the markerless system matched the marker-based system in “palms” (R2 0.714, p < 0.001), “elbows, extended knees” (R2 0.661, p < 0.001), “elbows, bent knees” (R2 0.829, p < 0.001), “squat” (R2 0.670, p < 0.004) and “B-Ball” (R2 0.701, p < 0.001) positions.

Trunk.

The linear regression models revealed that the data from the markerless motion capture system matched the data from the marker-based system for variables in only two out of seven positions for trunk variables in the sagittal and transverse planes (Table 11). In the sagittal plane, the markerless system corresponded with the marker-based system in the “upright” (R2 0.425, p = 0.008) and “B-Ball” positions (R2 0.663, p < 0.001). In the transverse plane, the markerless system matched the marker-based system in the “palms” (R2 0.614, p < 0.001) and “all-4s” (R2 0.382, p = 0.014) positions.

Discussion

This study served a dual purpose, representing one of the first attempts to quantify and compare the biomechanical characteristics of common upright birthing positions – providing new insights into pelvic, hip, and trunk configurations that have not been comprehensively compared before – and to establish the baseline accuracy and sensitivity of a markerless motion capture system against the established marker-based reference standard. Conducting this validation under controlled static conditions is a necessary first step in biomechanics, as it allows evaluation of a new system’s performance while minimising sources of motion artefact before extending analyses to dynamic or clinical environments. While the findings support our first hypothesis—that different upright birthing positions exhibit distinctly different biomechanical characteristics in the spine, torso, pelvis, and hips—they reject our second hypothesis. The markerless motion capture system did not provide comparable sensitivity to the marker-based system across all upright birthing positions.

Markerless motion capture sensitivity

While the markerless system showed promise in providing kinematic data in certain positions, particularly in the coronal (hips and pelvis) and transverse (pelvis) planes, it did not consistently match the sensitivity of the marker-based system across all planes and positions. Strong correlations between the two systems were observed in some positions, indicating potential for markerless technology as a viable alternative in some clinical applications. However, significant discrepancies were noted in the sagittal plane, particularly for the hip joint, where the markerless system was less reliable. Previous validation studies [31,42,43] have generally reported higher sagittal-plane agreement in dynamic gait tasks. However, the lower sagittal correspondence in our study likely reflects the static nature of the postures, and the increased self-occlusion of key landmarks from body tissues, such as the abdomen or quadriceps, which can reduce depth perception and joint visibility in certain positions. They may also reflect the system’s inherent limitations in accurately recognising joint positions during complex postures, such as ‘all-4s’ and ‘elbows extended knees.’ Additionally, beyond potential issues with occlusion, limitations in algorithm accuracy [31,42,43] may further contribute to the system’s shortcomings. These findings, while diverging from sagittal plane results in previous gait-based studies [31,42,43] underscore the need for further refinement of the markerless system to ensure precision and reliability across all planes and positions. Despite its potential, the current limitations of markerless technology suggest that it may not yet be reliable enough for assessing the biomechanics of certain static positions, particularly for the joints and postures we analysed in this study.

Biomechanical variations between positions

Based on the above, we will discuss only the variations from the marker-based system. The results highlight that the biomechanics of each birthing position have unique effects on the hip in all planes, and in the sagittal plane for the pelvis, and trunk. Notably, the “squat” position exhibited the highest levels of hip flexion in the sagittal plane, hip abduction in the coronal plane, and external rotation in the transverse plane. This combination suggests that “squat” produces greater openness of the pelvic outlet, which could facilitate the descent of the foetus and aid in labour progression [44,45]. However, a recent systematic review and meta-analysis does not show the squatting position during childbirth to be definitively beneficial [46] The review indicated that, although the squatting position may decrease the need for instrumental delivery, it is also associated with an increased risk of caesarean section [46]. There was no clear evidence of a reduction in the duration of the second stage of labour or significant improvements in maternal or foetal outcomes. As there is no strong evidence either for or against squatting, women should be encouraged to choose the position they find most comfortable during labour. Furthermore, more studies are needed to better understand childbirth positions, especially in Western countries, as the majority of studies reviewed were conducted in Asia [46]. The applicability of findings from Asian studies to Western populations may be limited due to differences in cultural practices, lifestyle, and anatomical variations. In many Asian countries, squatting is a common daily activity from an early age, which may lead to greater hip mobility and muscular adaptation to the position. In contrast, in Western populations, women may find it more difficult to adopt and sustain a deep squatting position due to differences in habitual movement patterns and musculoskeletal conditioning [4749]. Additionally, anatomical variations in pelvic morphology between racial and ethnic groups have been documented. For instance, research indicates that White women tend to have a wider pelvic inlet, wider outlet, and shallower anteroposterior outlet compared to African-American women [50]. These structural differences could influence the ease of adopting certain birthing positions and the extent to which they impact labour outcomes. Therefore, further research is needed to explore the effectiveness and feasibility of squatting during labour in diverse populations.

In contrast to squatting, the “upright” position produced the greatest hip extension in the sagittal plane and the least amount of hip abduction in the coronal plane. While this may slightly reduce the dimensions of the pelvic outlet, it could provide a more comfortable and less physically demanding posture for some women. The “upright” position also promoted significant trunk extension in the sagittal plane, which increases the activation of the trunk extensor muscles to maintain stability by counteracting anterior loads. [51]. Additionally, this posture encourages greater spinal extension, which could reduce discomfort for women who prefer an upright position. On the other hand, positions that induce greater trunk flexion, such as “elbows, bent knees,” may serve to redistribute weight and create a more forward-leaning posture, which could be advantageous during certain stages of labour.

The significant variations observed in pelvic tilt between positions are particularly important. The “B-Ball” position, which showed the greatest anterior pelvic tilt in the sagittal plane, could offer increased space in the pelvic inlet, facilitating the early stages of labour by allowing more room for the foetal head to engage. However, anterior pelvic tilt also increases lumbar lordosis (the inward curve of the lower back), which can strain the lower back [52,53], potentially causing discomfort or misaligning the pelvis during labour. On the other hand, positions such as “all-fours” exhibited the greatest posterior pelvic tilt. This reduces the curvature of the lumbar spine, leading to a more neutral or slightly flexed lumbar position, which may help ease back pain and improve spinal alignment [53]. This posterior tilt could be particularly advantageous in later stages of labour when comfort and pain relief are priorities. These variations in pelvic tilt not only influence pelvic openness but also directly impact maternal comfort, especially in later stages of labour when comfort and pain relief are priorities. Positions like “all-fours,” which decrease lumbar lordosis, may be particularly beneficial for labour progression by creating more space in the birth canal and reducing lower back pain.

Limitations of motion capture systems in labour settings

While both the marker-based and markerless motion capture systems provide valuable biomechanical data, significant limitations exist when considering their application in real-world labouring environments. The marker-based system, which requires the placement of retro-reflective markers on the participant’s body, poses several challenges for use in a clinical setting, particularly during labour. The presence of markers can be uncomfortable for a labouring woman, as they will disturb her from lying on her back or side or adopting other positions, causing pressure, discomfort, and pain. Furthermore, markers can dislodge or fall off during movement, active positions, and handling by healthcare professionals or labour companions. In cases of interventions, emergencies and/or transfers, the time required to remove the markers would introduce unnecessary delays and complexities, which, apart from being impractical in critical situations, might also have a negative effect on safety. Additionally, the marker-based system relies on multiple cameras, cables, and processing units that would be challenging to set up in a labouring room and would add the risk of tripping hazards. Also, healthcare professionals and labour companions assist and touch the labouring woman, so their hands or bodies could inadvertently obscure the markers, making it impossible to gather reliable data in such a dynamic and cluttered environment [8,9,5456]. The markerless motion capture system, although eliminating the need for physical markers, still faces similar practical limitations in a labouring room, including set up and ensuring unobstructed views [57,58]. Additionally, our findings indicate that the markerless system demonstrated lower sensitivity across multiple positions, including in the sagittal plane. This poses a specific challenge in clinical settings, as the sagittal plane is critical for assessing trunk flexion, pelvic tilt, and foetal descent during labour. This finding further underscores the current inadequacy of existing technologies for capturing labour biomechanics in real time. Despite these current hardware-driven limitations, establishing the baseline performance of markerless motion capture under controlled, static conditions is a critical prerequisite for future application in dynamic tasks and/or pregnant cohorts. The present study provides this foundational technical validation. Once system sensitivity and reliability are confirmed, the next necessary step is to evaluate the protocol in late-pregnant cohorts within the laboratory, where hardware limitations can be accommodated. These staged developments will ultimately support the long-term goal of applying markerless techniques during labour, when the technology becomes sufficiently unobtrusive for clinical use.

Although not part of this study, it is worth mentioning that both motion capture systems face additional challenges when considering water births. For the marker-based system, the use of water would likely cause the markers to detach from the skin, rendering the system ineffective. For both systems, the presence of water would introduce additional complexity due to refraction and surface ripples, which could incorrectly track the position of markers and/or create substantial visual noise, affecting the system’s ability to generate reliable data [59,60]. Although it is possible to combine underwater cameras with a land-based camera system into a single motion capture setup, such an extremely sophisticated arrangement would further interfere with the environment, the practice, and the labouring woman’s comfort [61,62].

From the above, it is evident that no currently available motion capture system can effectively measure the biomechanics of labour. This is due to the complexity of current equipment and methodologies, the dynamic nature of birth, and the practical constrains of the labouring environment. These limitations underscore the urgent need to develop new methods capable of accurately capturing labour biomechanics. Understanding the biomechanics of labour remains a critical need—not only because biomechanical complications are a major contributor to maternal and neonatal mortality, especially in low-income countries, [10] but also because it is unacceptable in 2025 that we still lack even a basic understanding of the mechanics involved in the onset of human life. Addressing this knowledge gap is essential for improving childbirth outcomes globally.

Study limitations and future directions

As this was the first study assessing different birthing positions using two motion capture techniques, healthy non-pregnant women were selected as the study population. If pregnant women had been included, we would have needed to assess participants in the third trimester, as close as possible to their due date, due to the significant physiological [63,64] and anatomical changes that occur during pregnancy, particularly toward the latter stages of the third trimester [65,66]. This population would be considered vulnerable due to the potential physical exertion [67], raising ethical concerns. Ensuring the safety of both the mother and foetus, and taking into account the risk-benefit ratio, is paramount [68]. Early-stage research using non-pregnant women allows for the exploration of biomechanical principles without exposing pregnant women to unnecessary risks [68].

Additionally, for an initial study providing essential groundwork, non-pregnant women offer a more controlled group for studying basic biomechanics. Their body weight, centre of gravity, and musculoskeletal function remain relatively stable compared to pregnant women, whose bodies undergo rapid changes [65,69]. This stability allows researchers to isolate and study the effects of specific birthing positions without the added variables introduced by pregnancy. This also helps establish baseline biomechanical data, testing the accuracy and sensitivity of motion capture systems before applying them to the more complex and variable biomechanics of pregnant women. Such a step is crucial for refining equipment and study methods. While using a non-pregnant population allowed for the development and validation of this protocol for the first time and provided valuable insights into general joint biomechanics across upright birthing positions, the findings should be interpreted with caution when translating them to labour scenarios. Key factors such as uterine contractions, foetal descent and rotation, and the dynamic interaction between the foetus and maternal pelvis, which significantly influence labour progression, are not represented in this study. Therefore, while this work establishes foundational biomechanical patterns and supports future methodological development, further studies in term-pregnant populations are necessary to ensure ecological validity and clinical relevance.

In particular, such studies should account for changes in spinal curvature, especially the lumbar curve, which increases by approximately 41% between 12 and 32 weeks of pregnancy [65]. They should also consider hormonal influences, such as relaxin, and their effects on pelvic mobility. However, experimental studies have not shown a direct relationship between high levels of relaxin and increased pelvic mobility or peripheral joint mobility in pregnant women [70]. Shifts in the centre of gravity should also be examined [71]. Furthermore, understanding the biomechanics of labour is crucial not only to enhance maternal and foetal well-being but also to reduce adverse outcomes and complications [72]. Future work could integrate empirical findings into computational models to simulate the influence of specific postures on pelvic dynamics, foetal descent, and maternal effort, offering an additional lens through which to understand the biomechanical optimisation of labour progression. Researchers and clinicians should collaborate to create a method that allows for this analysis and knowledge, as current techniques show significant limitations for use during childbirth in clinical settings [56]. Understanding childbirth should be a top priority on the research agenda, not only because maternal and neonatal mortality remains unacceptably high [7274] but also because, among all the Sustainable Development Goals (SDG) for 2030, the SDG 3.1 target to reduce global maternal mortality is the only one that has failed [75]. Moreover, it is unacceptable to have such limited knowledge and understanding of the biomechanics of the beginning of human life.

Conclusion

This study provides valuable insights into the biomechanics of various upright birthing positions, highlighting both the capabilities and limitations of current motion capture techniques. Future research should focus on refining these technologies and applying them in real-world labour environments. The findings demonstrate distinct biomechanical differences between positions, underscoring the need for personalised birthing strategies that consider the unique biomechanical effects of each posture. Personalising support during childbirth is essential for optimising maternal comfort, well-being, and successful delivery. Maternal healthcare providers must have a solid understanding of anatomy and biomechanics, recognising how changes in one part of the body can affect others. For instance, variations in bone length, such as in the thighs, can significantly impact the mechanics of positions like squatting. A deep understanding of these relationships is crucial for guiding women toward positions that support labour progress while accommodating their unique body structures and preferences. Furthermore, different birthing positions may offer varying benefits depending on an individual’s characteristics, preferences, and the stage of labour. Women should receive comprehensive information and guidance during the antenatal period to make informed decisions about their birthing positions. These findings support the development of personalised, evidence-based birthing strategies that address each woman’s biomechanical needs. Tailoring care to individual circumstances can improve maternal comfort, reduce complications, and enhance maternal and neonatal outcomes. Advancing research in childbirth biomechanics should remain a research priority, as it has the potential to significantly reduce maternal and neonatal mortality, a global challenge that persists today.

Acknowledgments

The authors would like to thank Rebekah McCrimmon, Lecturer of Midwifery, for her clinical insights, and Dr Jonathan Sinclair, Reader in Sport & Exercise Sciences and Nutrition & Exercise Sciences, for his invaluable contribution to the statistical analysis.

References

  1. 1. Priddis H, Dahlen H, Schmied V. What are the facilitators, inhibitors, and implications of birth positioning? A review of the literature. Women Birth. 2012;25(3):100–6. pmid:21664208
  2. 2. World Health Organization. WHO recommendations intrapartum care for a positive childbirth experience executive summary. 2018. https://iris.who.int/bitstream/handle/10665/272447/WHO-RHR-18.12-eng.pdf
  3. 3. Desseauve D, Fradet L, Lacouture P, Pierre F. Position for labor and birth: State of knowledge and biomechanical perspectives. Eur J Obstet Gynecol Reprod Biol. 2017;208:46–54. pmid:27888706
  4. 4. Michel SCA, Rake A, Treiber K, Seifert B, Chaoui R, Huch R, et al. MR obstetric pelvimetry: effect of birthing position on pelvic bony dimensions. AJR Am J Roentgenol. 2002;179(4):1063–7. pmid:12239066
  5. 5. Anderson N. X-ray pelvimetry: helpful or harmful?. J Fam Pract. 1983;17:405–12.
  6. 6. Reitter A, Daviss B-A, Bisits A, Schollenberger A, Vogl T, Herrmann E, et al. Does pregnancy and/or shifting positions create more room in a woman’s pelvis?. Am J Obstet Gynecol. 2014;211(6):662.e1-9. pmid:24949546
  7. 7. Ami O, Maran J-C, Musset D, Dubray C, Mage G, Boyer L. Using Magnetic Resonance Imaging During Childbirth to Demonstrate Fetal Head Moldability and Brain Compression: Prospective Cohort Study. JMIR Form Res. 2022;6(11):e27421. pmid:36322921
  8. 8. Desseauve D, Pierre F, Gachon B, Decatoire A, Lacouture P, Fradet L. New approaches for assessing childbirth positions. J Gynecol Obstet Hum Reprod. 2017;46(2):189–95. pmid:28403977
  9. 9. Desseauve D, Fradet L, Lacouture P, Pierre F. Is there an impact of feet position on squatting birth position? An innovative biomechanical pilot study. BMC Pregnancy Childbirth. 2019;19(1):251. pmid:31324160
  10. 10. Borges M, Moura R, Oliveira D, Parente M, Mascarenhas T, Natal R. Effect of the birthing position on its evolution from a biomechanical point of view. Comput Methods Programs Biomed. 2021;200:105921. pmid:33422852
  11. 11. Hemmerich A, Bandrowska T, Dumas GA. The effects of squatting while pregnant on pelvic dimensions: A computational simulation to understand childbirth. J Biomech. 2019;87:64–74. pmid:30851977
  12. 12. Kjeldsen LL, Dahlen HG, Maimburg RD. Expectations of the upcoming birth - A survey of women’s self-efficacy and birth positions. Sex Reprod Healthc. 2022;34:100783. pmid:36244077
  13. 13. Al Aryani Z, Orabi A, Fouly H. Examining the impact of upright and recumbent positions on labor outcomes in Saudi Arabia: A quasi-experiment. Belitung Nurs J. 2022;8(4):316–24.
  14. 14. Gupta JK, Sood A, Hofmeyr GJ, Vogel JP. Position in the second stage of labour for women without epidural anaesthesia. Cochrane Database Syst Rev. 2017;5(5):CD002006. pmid:28539008
  15. 15. Desseauve D, Pierre F, Fernandez A, Panjo H, Decatoire A, Lacouture P, et al. Assessment of Pelvic-Lumbar-Thigh Biomechanics to Optimize The Childbirth Position: An “In Vivo” Innovative Biomechanical Study. Sci Rep. 2019;9(1):15968. pmid:31685875
  16. 16. Topalidou A, Tzagarakis G, Souvatzis X, Kontakis G, Katonis P. Evaluation of the reliability of a new non-invasive method for assessing the functionality and mobility of the spine. Acta Bioeng Biomech. 2014;16:117–24.
  17. 17. Topalidou A, Tzagarakis G, Balalis K, Papaioannou A. Posterior Decompression and Fusion: Whole-Spine Functional and Clinical Outcomes. PLoS One. 2016;11(8):e0160213. pmid:27513643
  18. 18. Cho Y, Jo DJ, Hyun S-J, Park JH, Yang NR. From the Spinopelvic Parameters to Global Alignment and Proportion Scores in Adult Spinal Deformity. Neurospine. 2023;20(2):467–77. pmid:37401065
  19. 19. Illmeier G, Rechberger JS. The Limitations of Anterior Knee Displacement during Different Barbell Squat Techniques: A Comprehensive Review. J Clin Med. 2023;12(8):2955. pmid:37109294
  20. 20. Gorman J, Roberts CA, Newsham S, Bentley GR. Squatting, pelvic morphology and a reconsideration of childbirth difficulties. Evol Med Public Health. 2022;10(1):243–55. pmid:35663511
  21. 21. Kim S, Miller M, Tallarico A, Helder S, Liu Y, Lee S. Relationships between physical characteristics and biomechanics of lower extremity during the squat. J Exerc Sci Fit. 2021;19(4):269–77. pmid:34712337
  22. 22. Son SM. Influence of Obesity on Postural Stability in Young Adults. Osong Public Health Res Perspect. 2016;7(6):378–81. pmid:28053843
  23. 23. Takeda K, Shimizu K, Imura M. Changes in balance strategy in the third trimester. J Phys Ther Sci. 2015;27(6):1813–7. pmid:26180327
  24. 24. Zawojska K, Wnuk-Scardaccione A, Bilski J, Nitecka E. Correlation of Body Mass Index with Pelvis and Lumbar Spine Alignment in Sagittal Plane in Hemophilia Patients. Medicina (Kaunas). 2019;55(10):627. pmid:31554178
  25. 25. INVOLVE. Briefing notes for researchers: involving the public in NHS, public health and social care research. Eastleigh: INvolve. 2012.
  26. 26. Arumugam A, Phillips LR, Moore A, Kumaran SD, Sampath KK, Migliorini F, et al. Patient and public involvement in research: a review of practical resources for young investigators. BMC Rheumatol. 2023;7(1):2. pmid:36895053
  27. 27. World Health Organization. Maternal, Newborn, Child and Adolescent Health and Ageing. 2023. https://www.who.int/data/maternal-newborn-child-adolescent-ageing/indicator-explorer-new/mca/women-of-reproductive-age-(15-49-years)-population-(thousands)#:~:text=Women%20of%20reproductive%20age%20(15,49%20years)%20population%20(thousands)
  28. 28. Julious SA. Sample size of 12 per group rule of thumb for a pilot study. Pharmaceutical Statistics. 2005;4(4):287–91.
  29. 29. Lancaster GA, Dodd S, Williamson PR. Design and analysis of pilot studies: recommendations for good practice. J Eval Clin Pract. 2004;10(2):307–12. pmid:15189396
  30. 30. Whitehead AL, Julious SA, Cooper CL, Campbell MJ. Estimating the sample size for a pilot randomised trial to minimise the overall trial sample size for the external pilot and main trial for a continuous outcome variable. Stat Methods Med Res. 2016;25(3):1057–73. pmid:26092476
  31. 31. Das K, de Paula Oliveira T, Newell J. Comparison of markerless and marker-based motion capture systems using 95% functional limits of agreement in a linear mixed-effects modelling framework. Sci Rep. 2023;13(1):22880. pmid:38129434
  32. 32. Ceseracciu E, Sawacha Z, Cobelli C. Comparison of markerless and marker-based motion capture technologies through simultaneous data collection during gait: proof of concept. PLoS One. 2014;9(3):e87640. pmid:24595273
  33. 33. Lahkar BK, Muller A, Dumas R, Reveret L, Robert T. Accuracy of a markerless motion capture system in estimating upper extremity kinematics during boxing. Front Sports Act Living. 2022;4:939980. pmid:35958668
  34. 34. Nakano N, Sakura T, Ueda K, Omura L, Kimura A, Iino Y, et al. Evaluation of 3D Markerless Motion Capture Accuracy Using OpenPose With Multiple Video Cameras. Front Sports Act Living. 2020;2:50. pmid:33345042
  35. 35. Cappozzo A, Catani F, Croce UD, Leardini A. Position and orientation in space of bones during movement: anatomical frame definition and determination. Clin Biomech (Bristol). 1995;10(4):171–8. pmid:11415549
  36. 36. Chohan A, Payne K, Selfe J, Richards J. Biomechanical testing of a spinal alignment cushion (“Rophi cushion”) in adults with lower back pain. Gait Posture. 2013.
  37. 37. Haworth L, May K, Janssen J, Selfe J, Chohan A. The impact of breast support garments on fit, support and posture of larger breasted women. Appl Ergon. 2022;101:103701. pmid:35151120
  38. 38. Grood ES, Suntay WJ. A joint coordinate system for the clinical description of three-dimensional motions: application to the knee. J Biomech Eng. 1983;105(2):136–44. pmid:6865355
  39. 39. Kanko RM, Laende EK, Davis EM, Selbie WS, Deluzio KJ. Concurrent assessment of gait kinematics using marker-based and markerless motion capture. J Biomech. 2021;127:110665. pmid:34380101
  40. 40. Kanko RM, Laende EK, Strutzenberger G, Brown M, Selbie WS, DePaul V, et al. Assessment of spatiotemporal gait parameters using a deep learning algorithm-based markerless motion capture system. J Biomech. 2021;122:110414. pmid:33915475
  41. 41. World Medical Association. World Medical Association Declaration of Helsinki: ethical principles for medical research involving human subjects. JAMA. 2013;310(20):2191–4. pmid:24141714
  42. 42. Heitzmann D, Rist V, Block J, Alimusaj M, Wolf S. Markerless versus marker-based motion analysis in subjects with lower limb amputation: A case series. Gait & Posture. 2022;97:S95–6.
  43. 43. Wren TAL, Isakov P, Rethlefsen SA. Comparison of kinematics between Theia markerless and conventional marker-based gait analysis in clinical patients. Gait Posture. 2023;104:9–14. pmid:37285635
  44. 44. Russell JG. Moulding of the pelvic outlet. J Obstet Gynaecol Br Commonw. 1969;76(9):817–20. pmid:5823681
  45. 45. Gupta JK, Glanville JN, Johnson N, Lilford RJ, Dunham RJ, Watters JK. The effect of squatting on pelvic dimensions. Eur J Obstet Gynecol Reprod Biol. 1991;42(1):19–22. pmid:1778286
  46. 46. Dokmak F, Michalek IM, Boulvain M, Desseauve D. Squatting position in the second stage of labor: A systematic review and meta-analysis. Eur J Obstet Gynecol Reprod Biol. 2020;254:147–52. pmid:32966958
  47. 47. Savla R, Mullerpatan R, Agarwal B, Kuttan V, Kumar S. Influence of Physical Activity Including Squat Exposure on Trunk Muscle Strength and Labour Outcome in Pregnant Women. Int J Exerc Sci. 2024;17(1):504–16. pmid:38665166
  48. 48. Ganokroj P, Chaowalitwong J, Kerdsomnuek P, Sudjai N, Lertwanich P, Vanadurongwan B. Three-dimensional motion analysis of ten common Asian sitting positions in daily living and factors affect range of hip motions. BMC Musculoskelet Disord. 2021;22(1):618. pmid:34253220
  49. 49. Gupta JK, Brayshaw EM, Lilford RJ. An experiment of squatting birth. Eur J Obstet Gynecol Reprod Biol. 1989;30(3):217–20. pmid:2714504
  50. 50. Handa VL, Lockhart ME, Fielding JR, Bradley CS, Brubaker L, Cundiff GW, et al. Racial differences in pelvic anatomy by magnetic resonance imaging. Obstet Gynecol. 2008;111(4):914–20. pmid:18378751
  51. 51. Biviá-Roig G, Lisón JF, Sánchez-Zuriaga D. Changes in trunk posture and muscle responses in standing during pregnancy and postpartum. PLoS One. 2018;13(3):e0194853. pmid:29584774
  52. 52. Levine D, Whittle MW. The effects of pelvic movement on lumbar lordosis in the standing position. J Orthop Sports Phys Ther. 1996;24(3):130–5. pmid:8866271
  53. 53. Król A, Polak M, Szczygieł E, Wójcik P, Gleb K. Relationship between mechanical factors and pelvic tilt in adults with and without low back pain. J Back Musculoskelet Rehabil. 2017;30(4):699–705. pmid:28372304
  54. 54. Szczerbik E, Kalinowska M. The influence of knee marker placement error on evaluation of gait kinematic parameters. Acta Bioeng Biomech. 2011;13:43–6.
  55. 55. Desseauve D, Fradet L, Gherman RB, Cherni Y, Gachon B, Pierre F. Does the McRoberts’ manoeuvre need to start with thigh abduction? An innovative biomechanical study. BMC Pregnancy Childbirth. 2020;20(1):264. pmid:32366292
  56. 56. Topalidou A. Biomechanics of childbirth. British Journal of Midwifery. 2024;32(8):446–8.
  57. 57. Tang H. Comparison of marker-based and markerless motion capture systems in gait biomechanics during running. Georgia Southern University. 2023.
  58. 58. Topley M, Richards JG. A comparison of currently available optoelectronic motion capture systems. J Biomech. 2020;106:109820. pmid:32517978
  59. 59. Thompson I, Banks J, Warner M, Hudson D. Assessment of error levels across the domain of a three dimensional underwater motion capture methodology. ISBS Proceedings Archive. 2022;40.
  60. 60. Kwon Y-H, Casebolt JB. Effects of light refraction on the accuracy of camera calibration and reconstruction in underwater motion analysis. Sports Biomech. 2006;5(2):315–40. pmid:16939159
  61. 61. Silvatti AP, Cerveri P, Telles T, Dias FAS, Baroni G, Barros RML. Quantitative underwater 3D motion analysis using submerged video cameras: accuracy analysis and trajectory reconstruction. Comput Methods Biomech Biomed Engin. 2013;16(11):1240–8. pmid:22435960
  62. 62. Bernardina GRD, de Andrade AGP, Monnet T, Cerveri P, Silvatti AP. Simultaneous in-air and underwater 3D kinematic analysis of swimmers: Feasibility and reliability of action sport cameras. J Biomech. 2024;168:112078. pmid:38663110
  63. 63. Soma-Pillay P, Nelson-Piercy C, Tolppanen H, Mebazaa A. Physiological changes in pregnancy. Cardiovasc J Afr. 2016;27(2):89–94. pmid:27213856
  64. 64. Tan EK, Tan EL. Alterations in physiology and anatomy during pregnancy. Best Pract Res Clin Obstet Gynaecol. 2013;27(6):791–802. pmid:24012425
  65. 65. Yoo H, Shin D, Song C. Changes in the spinal curvature, degree of pain, balance ability, and gait ability according to pregnancy period in pregnant and nonpregnant women. J Phys Ther Sci. 2015;27(1):279–84. pmid:25642091
  66. 66. Franklin ME, Conner-Kerr T. An analysis of posture and back pain in the first and third trimesters of pregnancy. J Orthop Sports Phys Ther. 1998;28(3):133–8. pmid:9742469
  67. 67. Okafor UB, Goon DT. Uncovering Barriers to Prenatal Physical Activity and Exercise Among South African Pregnant Women: A Cross-Sectional, Mixed-Method Analysis. Front Public Health. 2022;10:697386. pmid:35433615
  68. 68. McKiever M, Frey H, Costantine MM. Challenges in conducting clinical research studies in pregnant women. J Pharmacokinet Pharmacodyn. 2020;47(4):287–93. pmid:32306165
  69. 69. Conder R, Zamani R, Akrami M. The Biomechanics of Pregnancy: A Systematic Review. J Funct Morphol Kinesiol. 2019;4(4):72. pmid:33467386
  70. 70. Aldabe D, Ribeiro DC, Milosavljevic S, Dawn Bussey M. Pregnancy-related pelvic girdle pain and its relationship with relaxin levels during pregnancy: a systematic review. Eur Spine J. 2012;21(9):1769–76. pmid:22310881
  71. 71. Casagrande D, Gugala Z, Clark SM, Lindsey RW. Low Back Pain and Pelvic Girdle Pain in Pregnancy. J Am Acad Orthop Surg. 2015;23(9):539–49. pmid:26271756
  72. 72. Topalidou A, Haworth L, Kaur I, Ahmed M, Chohan A. Assessment of the pelvic and body interface pressure during different recumbent and semi-recumbent birthing positions. Clin Biomech (Bristol). 2024;119:106328. pmid:39191043
  73. 73. World Health Organization. Global progress in tackling maternal and newborn deaths stalls since 2015: UN. 2023. https://www.who.int/news/item/09-05-2023-global-progress-in-tackling-maternal-and-newborn-deaths-stalls-since-2015--un
  74. 74. Vallely LM, Calvert B, De Silva M, Panisi L, Babona D, Bolnga J, et al. Improving maternal and newborn health and reducing stillbirths in the Western Pacific Region - current situation and the way forward. Lancet Reg Health West Pac. 2022;32:100653. pmid:36785855
  75. 75. Khalil A, Samara A, O’Brien P, Coutinho CM, Quintana SM, Ladhani SN. A call to action: the global failure to effectively tackle maternal mortality rates. Lancet Glob Health. 2023;11(8):e1165–7. pmid:37474218