Effect of perfusion index on oxygen reserve index accuracy in estimating arterial oxygen tension in anesthetized dogs: Data reanalysis

Abstract

Multi-wave CO-oximetry, utilizing the oxygen reserve index (ORi), estimates arterial partial pressure of oxygen (PaO2) in mild hyperoxemia, between 100 and 200 mmHg, and concurrently quantifies local perfusion at the measurement site using the perfusion index (PI). This study explores how variations in PI influence the accuracy of ORi in estimating PaO2 in anesthetized dogs. Data from 37 mechanically ventilated dogs were retrospectively reanalyzed using a different approach. ORi and PI values were collected using a CO-oximeter. The data were categorized into four groups based on PI quartiles. In each group, the relationship between ORi and PaO2 was assessed using linear regression analysis, and the area under the receiver operating characteristic curve (AUROC) investigated the diagnostic performance of ORi in detecting PaO2 >  150 mmHg. Strong relationships between ORi and PaO2 were observed in groups with PI values <  2 (r2 ≥  0.63). The AUROC of ORi for identifying PaO2 > 150 mmHg decreased with PI >  2 compared to lower values (0.76 vs >  0.88). In this study, PI values >  2 negatively impacted ORi’s ability to estimate PaO2, likely due to fluctuations in blood flow perfusing the measurement site. The results of this study suggests that consideration of the PI value is essential when titrating oxygen therapy using ORi in anesthetized dogs.

Introduction

Continuous assessment of oxygenation in anesthetized veterinary patients is essential to detect potentially life-threatening conditions [1]. While arterial blood gas analysis is the gold standard for measuring arterial oxygen content, its invasiveness and delayed results make it less practical, especially in small to medium-sized animals [2]. Pulse oximetry offers a non-invasive method to estimate oxygenation by measuring the percentage of hemoglobin saturated with oxygen in peripheral arterial blood (SpO2). This method is based on the relationship between arterial partial pressure of oxygen (PaO2) and oxygen-saturated hemoglobin (SaO2). Conventional pulse oximetry utilizes light wavelengths of 660 nm (red) and 940 nm (infra-red) to differentiate between oxyhemoglobin and deoxygenated hemoglobin [3]. However, the accuracy of pulse oximetry is influenced by the quality of arterial blood flow at the sensor site [4, 5], as measured by the perfusion index (PI) [6]. PI reflects the strength of peripheral pulse and varies with factors such as stroke volume and vascular tone [7]. Masimo’s Rainbow SET technology enhances pulse oximetry by incorporating CO-oximetry, which provides continuous feedback on perfusion status and signal quality. Studies in healthy humans have reported mean PI values ranging from 2.2 to 3.5 [8,9], though values can vary widely. In dogs under general anesthesia, PI values ranged from 0.3 to 1.9, influenced by factors such as tongue width and ambient light [10,11]. Ventilation mode and passive leg raising test also affects PI measurements [12,13]. Despite challenges like poor perfusion, modern pulse oximeters generally provide accurate SaO2 estimates even under conditions of hypoperfusion, hypotension, hypothermia, or vasopressors use [14]. In veterinary settings, probe placement can influence PI values and SpO2 readings [11].

Masimo Rainbow SET technology broadens measurement capabilities by assessing various hemoglobin forms and oxygen saturation in both arterial and venous blood, providing a more comprehensive view of oxygenation status. The oxygen reserve index (ORi), developed for human use, estimates the oxygen reserve by evaluating PaO2 levels from 100 to 200 mmHg [1520]. Human studies have demonstrated a strong correlation between ORi and PaO2, though veterinary research shows varying correlations across species. A study on anesthetized donkeys found a mild correlation [21], whereas studies on dogs demonstrated a moderate correlation with ORi being effective in monitoring oxygen therapy [22,23,24].

A linear mixed-effects regression model showed that PI does not influence the correlation between PaO2 and ORi. However, the diagnostic performance of ORi in detecting PaO2 ≥  150 mmHg revealed high uncertainty at elevated PI values [22], suggesting a need for further research. The current study aims to reanalyze previously collected data to explore ORi’s effectiveness in estimating PaO2 changes across different PI ranges in anesthetized dogs and to identify the optimal PI value range for accurate ORi cut-off values for PaO2 ≥  150 mmHg. We hypothesized that local perfusion, as measured by PI, could have a significant impact on the correlation between ORi and PaO2, potentially influencing the accuracy and reliability of this relationship.

Materials and methods

Animals

The Ethics Committee of the University of Padova approved the study (OPBA 75/2021), and the dog owner gave consent before any procedures. In this retrospective analysis, simultaneous values of ORi, PaO2, and PI were analyzed from data collected by the authors in a previous study [22]. The study enrolled thirty-seven American Society of Anesthesiologists physical status I and II dogs anesthetized between October 2022 and May 2023, of both sexes (20 males and 17 females). Median (1st quartile–3rd quartile) age was 98 (73-122) months old and weight was 29.0 (20.0-33.5) kg, with a median body condition score of 6 (5-6) out of 9. The animals were admitted to the Veterinary Teaching Hospital, University of Padova, and underwent soft tissue surgeries or diagnostic imaging procedures under general anesthesia.

Anesthesia

Dogs scheduled for diagnostic procedures received 0.1–0.2 mg/kg butorphanol (Dolorex; AnimalHealth Intervet Italia Srl, Segrate, Italy) alone or with 2–4 μg/kg dexmedetomidine (Dexdomitor; OrionCorporation, Espoo, Finland) intramuscularly as premedication before intravenous catheter placement. Methadone 0.1–0.2 mg/kg (Semfortan, DechraVeterinary Products Srl, Torino, Italy) was used for preemptive analgesia in animals undergoing surgical procedures. General anesthesia was induced with propofol (PropoVet; Zoetis, Roma, Italy) intravenously to effect, until orotracheal intubation was possible, followed by maintenance with sevoflurane (Sevorane; AbbVie 148 S.r.l., Campoverde di Aprilia, Italy) or isoflurane (Isoflo; Zoetis, Roma, Italy) carried in a mixture of oxygen and air to obtain a fraction of inspired oxygen (FiO2) between 0.21 and 0.50. After anesthesia induction, a 22G arterial catheter (Delta Med; Viadana, Italy) was inserted for blood pressure measurement and sampling. Mechanical ventilation with a pressure-control mode was started immediately after induction, and the settings were adjusted to maintain the end-tidal carbon dioxide pressure (EtCO2) between 35 and 45 mmHg. Parameters monitored during anesthesia and recorded at the time of arterial blood sampling included direct systemic arterial blood pressure (systolic [SAP], mean [MAP], and diastolic [DAP] pressure), EtCO2, FiO2, and temperature (Temp). All data were displayed on a multi-parameter monitor (Datex S/5; GE Healthcare; Helsinki, Finland).

Measurements

Pulse trace and values of SpO2, ORi and PI were measured using a multiwavelength pulse CO-oximeter (Rad-97, Masimo Corp., Irvine, CA, USA), with the adhesive probe (RD Rainbow Lite SET-1 Neo; Masimo Corp., Irvine, CA, USA) wrapped circumferentially around the folded tongue and connected to the CO-oximeter. Values of PaO2 were measured by a blood gas analyzer (EDAN i15, EDAN, Shenzhen, China) using a multi-parameter cartridge (Test Cartridge BG10, EDAN, Shenzhen, China) after collecting a 1 ml blood sample collected through the arterial catheter into a pre-heparinized syringe (Pulset; Westmed Inc., AZ, USA). Prior to sampling, 2 ml of blood were drawn and discarded to prevent sample contamination or dilution. After sampling, the artery was flushed with 1 ml of heparinized solution (10 IU/ml). Coupled values of ORi and PaO2 were divided into four groups according to the value of PI. Since a physiologic range for PI is unavailable in both human and veterinary studies, in this study we arbitrary stratified the observation based on the four quartile ranges for the PI values. This approach relies on the median and interquartile range as a method to describe how data are distributed within a sample representative of the population. Values associated with PI values ≤  the first interquartile were included in group P1. Those with PI>  the first interquartile and ≤  the median were categorized into group P2. Values with PI>  the median and ≤  the third interquartile were allocated to group P3, while those with PI values>  the third interquartile were placed in group P4.

Statistical analysis

The normal distribution of continuous variables was investigated using a Shapiro-Wilk test. Normally distributed values were presented as mean ±  standard deviation; otherwise, they were reported as median and interquartile range (IQR: 1st quartile–3rd quartile). Values of PaO2, ORi, hematocrit (Hct), PR (pulse rate), SAP, MAP, DAP, and Temp were compared among groups using a one-way ANOVA or Kruskal-Wallis test, as appropriate. Linear regression analysis, reporting the coefficient of determination (r2), was performed to investigate the linear relationship between all values of ORi and PaO2 in the entire dataset or within each PI group. The goodness of fit for the linear regression model in each PI group was evaluated using the F-statistic, as calculated by the lm() function, to determine the explanatory power of the model within different ranges of PI [25]. Receiver operating characteristic (ROC) curve analysis and the Youden Index were used to determine the optimal cut-off ORi value that identified a PaO2 >  150 mmHg within each PI group. For each cut-off, the associated sensitivity, specificity, positive predictive value, negative predictive value, the area under the curve (AUROC), and the 95% confidence interval (CI) were calculated. All statistical analyses were performed using RStudio (RStudio, PBC, Boston, 196 MA, US) as the interface for R (The R Foundation for Statistical Computing, Vienna, Austria). Significance was set at p <  0.05.

Results

Among all paired values of ORi and PaO2 collected in the previous study [22], two were excluded because they did not have corresponding PI values. A total of 99 paired ORi, PaO2, PI, PR, SAP, MAP, DAP and Hct measurements were collected. Median ORi value was 0.53 (0.33-0.75), and mean PaO2 was 150 ±  37 mmHg. Median PI was 1.30 (0.80-1.90), and groups P1, P2, P3, and P4 included values of PI between 0.23 and 0.79, between 0.80 and 1.30, between 1.40 and 1.90, and between 2.00 and 3.40, respectively. Matched measurements of ORi and PaO2 were 25, 25, 26, and 23 in groups P1, P2, P3, and P4, respectively. Of the 37 dogs included in the study, 19 had ORi-PaO2 values that consistently corresponded to PI values within the same PI range. In contrast, for the remaining 18 dogs, the ORi-PaO2 values were paired with PI values that spanned different PI ranges. The distribution of ORi values varied slightly among the three groups (P1, P2, P3), displaying an almost bell-shaped pattern. In contrast, a flat distribution was observed in group P4 (Fig 1). No statistically significant difference was found in PaO2, ORi, Hct, PR, SAP, MAP, DAP, and Temp among groups (Table 1).

thumbnail
Table 1. Clinical data of dogs and comparisons among groups. Values are distributed into 4 groups based on the value of the perfusion index (PI). Groups P1, P2, P3, and P4 included measurements matching PI values between 0.23 and 0.79, 0.80 and 1.30, 1.40 and 1.90, and 2.00 and 3.40, respectively. Data are reported as mean ±  standard deviation or median and interquartile range (IQR). Statistical Significance was set for p <  0.05.

https://doi.org/10.1371/journal.pone.0319823.t001

thumbnail
Fig 1. Dot plot and paired density plot of oxygen reserve index (ORi) and arterial partial pressure of oxygen (PaO2) in dogs.

Groups P1, P2, P3, and P4 included measurements matching PI values between 0.23 and 0.79, 0.80 and 1.30, 1.40 and 1.90, and 2.00 and 3.40, respectively.

https://doi.org/10.1371/journal.pone.0319823.g001

A significant strong linear relationship between ORi and PaO2 values was found in groups P1, P2 and P3 (Fig 2). The relationship between ORi and PaO2 values in group P4 was very weak (r2 =  0.18, p =  0.04).

thumbnail
Fig 2. Coefficient of determination (r2) were assessed using linear regression analysis between paired values of oxygen reserve index (ORi) and arterial partial pressure of oxygen (PaO2) in groups P1 (A), P2 (B), P3 (C), and P4 (D).

Groups P1, P2, P3, and P4 included measurements matching PI values between 0.23 and 0.79, 0.80 and 1.30, 1.40 and 1.90, and 2.00 and 3.40, respectively. F-stat: F-statistic.

https://doi.org/10.1371/journal.pone.0319823.g002

The Youden index identified ORi cut-off values of 0.53, 0.56, and 0.67, indicating PaO2 >  150 mmHg in groups P1, P2, and P3, respectively, with AUROC >  0.88, sensitivity >  0.85, and specificity >  0.83 (Table 2). In group P4, the ORi cut-off value indicating PaO2 >  150 mmHg was 0.85, with an AUROC (0.76) and a sensitivity (0.60) lower than that observed in the other groups.

thumbnail
Table 2. Diagnostic performance of oxygen reserve index (ORi) to detect arterial partial pressure of oxygen above 150 mmHg. Groups P1, P2, P3, and P4 included measurements matching PI values between 0.23 and 0.79, 0.80 and 1.30, 1.40 and 1.90, and 2.00 and 3.40, respectively.

https://doi.org/10.1371/journal.pone.0319823.t002

Discussion

The main finding of this study indicates that measurements of ORi associated with a PI greater than 2 result in less accuracy in estimating PaO2, whereas higher accuracy is observed for PI range of 0.80-1.90. Within this range, the diagnostic capability of the oxygen reserve index in detecting arterial partial pressure of oxygen above 150 mmHg is maximized. The findings of the current study provide additional clinical insights and can complement the information presented by the authors in their initial assessment of ORi as a novel index that could enhance safety during anesthesia and in the early postoperative phase. Based on these results, the recommendation is to monitor PaO2 using an alternative device, such as a blood gas analyzer, if the CO-oximeter displays a PI outside of the specified range.

In dogs, ORi is an innovative and attractive parameter that has been investigated for non-invasively estimating oxygenation status in hyperoxemic patients [22,23]. Compared to standard pulse oximetry, which constantly registers SpO2 levels above 98% when PaO2 exceeds 100 mmHg, ORi was able to anticipate the decline in SpO2 during hyperoxemia in dogs [26]. Moreover, ORi demonstrated a good correlation with oxygen flow rate in sedated dogs [24], making it an appealing tool for monitoring and optimizing oxygen treatment. However, this also emphasizes the paramount importance of having a good accuracy for this device, as excessive hyperoxemia could lead to significant oxygen wastage and potentially result in atelectasis in dogs [27,28]. Moreover, in a study involving humans undergoing laparoscopic gastrectomy, the combination of ORi and SpO2 guided FiO2 adjustment reduced hyperoxemia compared to using SpO2 alone [29]. The clinical utility of ORi relies on its ability to estimate PaO2 within a reference range of 100 and 200 mmHg. Paired measurements of ORi and PaO2 exhibited a stronger correlation when the displayed PI was lower than 2. The increase in PI may reduce the accuracy of ORi in estimating PaO2, though the exact reasons remain unclear, based on the data collected in this study. PI reflects the ratio of pulsatile to non-pulsatile blood flow, so increases in pulsatile flow or decreases in non-pulsatile flow elevate PI. Vasodilation, often induced by inhalation anesthesia [30], raises PI levels, as seen in both humans and animals [3135]. Additionally, external pressure from soft tissue or pulse oximeter probes can reduce non-pulsatile flow [36], further elevating PI and potentially affecting ORi reliability. In a study involving anesthetized dogs, inserting material between the tongue and the pulse oximeter probe resulted in significantly increased PI and a concurrent change in the recorded SpO2 [10]. In our study, using a pulse oximeter adhesive tape probe around the folded tongue may have led to tissue compression in some dogs. Furthermore, the probe position might have varied slightly among patients. In a study involving anesthetized dogs, PI values were higher at distances of 0.5 cm and 1 cm rostrally compared to the root of the tongue, significantly affecting the SpO2 output [11]. The difference in PI value could be attributed to the varying vascularization across different regions of the canine tongue, each with its distinct distribution of arteries, veins, and arteriovenous anastomoses. At the apex of the tongue, arteriovenous anastomoses are abundant, and external pressure applied to the tongue may compress them, potentially affecting the pulse oximeter signal. Mair and colleagues [11] observed higher SpO2 values with lower PI in anesthetized dogs using a Masimo Rainbow SET Pulse CO-oximeter. In contrast, in another study, the same authors [10] observed higher SpO2 values with higher PI using a Masimo SET Rad 5 pulse oximeter, but the lack of SaO2 measurements limits the interpretation of SpO2 accuracy based on PI in dogs. In humans, a recent study in healthy patients found a significant positive correlation between PI and SpO2, but not with ORi [37]. However, in this study, neither SaO2 nor PaO2 were measured to assess the accuracy of the device. These findings suggest only that there might be a relationship between the PI and signal acquisition by the pulse oximeter. It is important to note that the accuracy of SpO2 in estimating oxygen content may differ from ORi. While SpO2 readings primarily reflect arterial blood flow, ORi measurement is influenced by both arterial and venous blood. Consequently, PI, which accounts for both pulsatile and non-pulsatile blood flow, could exert a more significant impact.

In our study, PI also demonstrated an effect on the diagnostic ability of ORi in detecting mild hyperoxemia. Conversely, in a study involving critically ill patients, the diagnostic performance of ORi in detecting PaO2 levels above 100 mmHg appeared to be unaffected by the PI [38]. However, the authors reported a weak overall correlation (r =  0.13) between all ORi and PaO2 values, with a mean PI value of 2.9, notably higher than in our study. Additionally, the correlation between ORi and PaO2 across different PI values was not evaluated. In our investigation, we assessed the diagnostic capability of ORi in identifying PaO2 levels above 150 mmHg. We found that ORi’s diagnostic performance remains robust for PI values below 2, but it slightly diminishes for higher values, with a specificity of about 60%. This suggests that ORi is less effective in identifying individuals with hyperoxemia beyond 150 mmHg when the PI is above 2, while the sensitivity exceeds 83% across all PI values in our study. This suggests that ORi accurately identify patients with PaO2 levels below 150 mmHg, even when the PI exceeds 2. These findings underscore the utility of ORi in maintaining safe oxygen levels and promptly detecting critical deterioration in oxygenation status, as observed in human studies [39,40].

Our study showed some limitations. One was the method of stratifying paired data based on quartile ranges. Given the arbitrary nature of this choice, we cannot rule out the possibility that using a different grouping method, it could impact the outcomes, underscoring the need, in future studies, for a more accurate stratification approach. No information was available on the clinically relevant range of PI at which ORi estimates PaO2 with the highest accuracy. The authors decided to preliminarily investigate the diagnostic performance of ORi by dividing the sample according to the median and interquartile range. This is a descriptive data method used for reporting distribution of a variable in a population or sample. In future studies, narrower subgroups could more accurately determine the optimal interval of PI associated with the highest ORi accuracy in estimating PaO2. However, in the current study, increasing the number of PI ranges would result in subgroups with a limited number of observations, significantly decreasing statistical power.

Moreover, the sample size was not calculated, as the study is a reanalysis of previously collected data. However, in the retrospective sample size estimation, assuming an alpha of 0.05, a power of 0.9, and a coefficient of determination of 0.52 [22], the minimum number of paired measurements was estimated to be 12. Therefore, despite being a retrospective analysis, the number of values assigned to all PI groups exceeds the minimum sample size, ensuring sufficient statistical power.

In this study, several factors may have influenced the variability in PI readings. Inadequate probe contact or alignment and the use of a single-sized probe across dogs of varying body sizes could contribute to differences in PI, especially since tongue width was not measured. Over 50% of the dogs showed minimal PI variation, limiting their classification to a single PI group, which may have introduced subjective biases. The study focused on dogs with stable systemic hemodynamic conditions, but peripheral hypoperfusion at the probe site cannot be ruled out. In humans, PI values below 1.4 may indicate peripheral hypoperfusion [8], but no such cut-off exists for dogs, and the relationship between PI and poor peripheral perfusion in dogs is poorly understood. Additionally, the study only assessed local perfusion using PI values, neglecting other variables such as the core-to-toe temperature difference, which is correlated with perfusion [8]. The retrospective nature of the study introduces limitations, such as potential biases in data collection. A repeated measurement model was previously used to analyze the ORi-PaO2 correlation, but this approach did not account for individual animal effects. Despite these limitations, the study confirms that over 50% of the variability in ORi can be explained by PaO2, particularly when PI readings are below 1.9.

Conclusions

This study highlights the pivotal role of PI values in influencing the accuracy of ORi readings for estimating PaO2 in anesthetized dogs. Elevated PI values (>2) can significantly decrease the ORi accuracy in estimating arterial oxygen levels. Clinicians should take into consideration the PI measurements displayed by the CO-oximeter when adjusting oxygen therapy based on ORi readings to optimize oxygen administration. Due to the retrospective nature of this study, further researches are warranted to confirm the ability of ORi to estimate PaO2 over a wide range of PI and to investigate how this may impact anesthesia outcomes.

References

  1. 1. Grubb T, Sager J, Gaynor JS, Montgomery E, Parker JA, Shafford H, et al. 2020 AAHA anesthesia and monitoring guidelines for dogs and cats. J Am Anim Hosp Assoc. 2020;56(2):59–82. pmid:32078360
  2. 2. Hagley MJ, Hopper K, Epstein SE. Characteristics of arterial catheter use and related complications in dogs and cats in an intensive care unit. J Vet Emerg Crit Care (San Antonio). 2021;31(4):469–75. pmid:34124836
  3. 3. Jubran A. Pulse oximetry. Crit Care. 1999;3(2):R11–7. pmid:11094477
  4. 4. Jubran A. Pulse oximetry. Crit Care. 2015;19(1):272. pmid:26179876
  5. 5. Schallom L, Sona C, McSweeney M, Mazuski J. Comparison of forehead and digit oximetry in surgical/trauma patients at risk for decreased peripheral perfusion. Heart Lung. 2007;36(3):188–94. pmid:17509425
  6. 6. Goldman JM, Petterson MT, Kopotic RJ, Barker SJ. Masimo signal extraction pulse oximetry. J Clin Monit Comput. 2000;16(7):475–83. pmid:12580205
  7. 7. Coutrot M, Dudoignon E, Joachim J, Gayat E, Vallée F, Dépret F. Perfusion index: Physical principles, physiological meanings and clinical implications in anaesthesia and critical care. Anaesth Crit Care Pain Med. 2021;40(6):100964. pmid:34687923
  8. 8. Lima AP, Beelen P, Bakker J. Use of a peripheral perfusion index derived from the pulse oximetry signal as a noninvasive indicator of perfusion. Crit Care Med. 2002;30(6):1210–3. pmid:12072670
  9. 9. Keller G, Cassar E, Desebbe O, Lehot J-J, Cannesson M. Ability of pleth variability index to detect hemodynamic changes induced by passive leg raising in spontaneously breathing volunteers. Crit Care. 2008;12(2):R37. pmid:18325089
  10. 10. Mair A, Martinez-Taboada F, Nitzan M. Effect of lingual gauze swab placement on pulse oximeter readings in anaesthetised dogs and cats. Vet Rec. 2017;180(2):49. pmid:27815366
  11. 11. Mair A, Ferreira J, Ricco C, Nitzan M. Appraisal of the “penumbra effect” using lingual pulse oximetry in anaesthetized dogs and cats. Vet Anaesth Analg. 2020;47(2):177–82. pmid:31982340
  12. 12. Shelley KH. Photoplethysmography: beyond the calculation of arterial oxygen saturation and heart rate. Anesth Analg. 2007;105(6 Suppl):S31–6. pmid:18048895
  13. 13. Beurton A, Gavelli F, Teboul J-L, De Vita N, Monnet X. Changes in the Plethysmographic Perfusion Index During an End-Expiratory Occlusion Detect a Positive Passive Leg Raising Test. Crit Care Med. 2021;49(2):e151–60. pmid:33332814
  14. 14. Poorzargar K, Pham C, Ariaratnam J, Lee K, Parotto M, Englesakis M, et al. Accuracy of pulse oximeters in measuring oxygen saturation in patients with poor peripheral perfusion: a systematic review. J Clin Monit Comput. 2022;36(4):961–73. pmid:35119597
  15. 15. Scheeren TWL, Belda FJ, Perel A. The oxygen reserve index (ORI): a new tool to monitor oxygen therapy. J Clin Monit Comput. 2018;32(3):379–89. pmid:28791567
  16. 16. Applegate RL 2nd, Dorotta IL, Wells B, Juma D, Applegate PM. The Relationship Between Oxygen Reserve Index and Arterial Partial Pressure of Oxygen During Surgery. Anesth Analg. 2016;123(3):626–33. pmid:27007078
  17. 17. Vos JJ, Willems CH, van Amsterdam K, van den Berg JP, Spanjersberg R, Struys MMRF, et al. Oxygen Reserve Index: validation of a new variable. Anesth Analg. 2019;129(2):409–15. pmid:30138170
  18. 18. Bathe J, Malik S, Pinnschmidt HO, Zitzmann A, Behem CR, Trepte CC, et al. Effectiveness of preoxygenation by conventional face mask versus non-invasive ventilation in morbidly obese patients: measurable by the oxygen-reserve index?. J Clin Monit Comput. 2022;36(6):1767–74. pmid:35167036
  19. 19. Cheng H-W, Yeh C-Y, Chang M-Y, Ting C-K, Chang P-L. How early warning with the Oxygen Reserve Index (ORi™) can improve the detection of desaturation during induction of general anesthesia?. J Clin Monit Comput. 2022;36(5):1379–85. pmid:34953137
  20. 20. Fadel ME, Shangab MO, Walley HE, Al Taher H, Lobo FA. Oxygen Reserve Index and Arterial Partial Pressure of Oxygen: Relationship in Open Heart Surgery. J Clin Monit Comput. 2023;37(6):1435–40. pmid:37024751
  21. 21. Bellini L, Dzikiti BT, De Benedictis GM, Algarin Sepulveda FR, Maney JK. Oxygen reserve index as a noninvasive indicator of arterial partial pressure of oxygen in anaesthetized donkeys: a preliminary study. Vet Anaesth Analg. 2021;48(3):388–92. pmid:33632615
  22. 22. Zanusso F, De Benedictis GM, Zemko P, Bellini L. Non-invasive assessment of oxygenation status using the oxygen reserve index in dogs. BMC Vet Res. 2023;19(1):241. pmid:37980491
  23. 23. Sureiman O, Mangera C. F-test of overall significance in regression analysis simplified. J Pract Cardiovasc Sci. 2020;6(2):116.
  24. 24. Watanabe T, Kanno N, Suzuki S, Yogo T, Harada Y, Hara Y. Investigation of the association between oxygen reserve index and arterial partial oxygen pressure in anesthetized dogs. Vet Anaesth Analg. 2024;51(1):10–5. pmid:37949743
  25. 25. Bellini L, De Benedictis GM. Oxygen Reserve Index as a Tool to Monitor Four Techniques of Oxygen Supplementation at Different Flow Rates in Dogs Sedated with Dexmedetomidine and an Opioid. Animals (Basel). 2023;13(19):3077. pmid:37835683
  26. 26. Zanusso F, De Benedictis GM, Bellini L. Exploring oxygen reserve index for timely detection of deoxygenation in canine patients recovering from anesthesia. Res Vet Sci. 2024;173:105268. pmid:38631076
  27. 27. Staffieri F, Franchini D, Carella GL, Montanaro MG, Valentini V, Driessen B, et al. Computed tomographic analysis of the effects of two inspired oxygen concentrations on pulmonary aeration in anesthetized and mechanically ventilated dogs. Am J Vet Res. 2007;68(9):925–31. pmid:17764405
  28. 28. De Monte V, Grasso S, De Marzo C, Crovace A, Staffieri F. Effects of reduction of inspired oxygen fraction or application of positive end-expiratory pressure after an alveolar recruitment maneuver on respiratory mechanics, gas exchange, and lung aeration in dogs during anesthesia and neuromuscular blockade. Am J Vet Res. 2013;74(1):25–33. pmid:23270342
  29. 29. Ahn JH, Shim J-G, Park J, Lee SH, Ryu K-H, Cho E-A. Oxygen reserve index guided fraction of inspired oxygen titration to reduce hyperoxemia during laparoscopic gastrectomy: A randomized controlled trial. Medicine (Baltimore). 2022;101(46):e31592. pmid:36401493
  30. 30. Schwinn DA, McIntyre RW, Reves JG. Isoflurane-induced vasodilation: role of the alpha-adrenergic nervous system. Anesth Analg. 1990;71(5):451–9. pmid:1977331
  31. 31. Galvin EM, Niehof S, Verbrugge SJ, Maissan I, Jahn A, Klein J, et al. Peripheral flow index is a reliable and early indicator of regional block success. Anesth Analg. 2006;103(1):239–43, table of contents. pmid:16790660
  32. 32. Ginosar Y, Weiniger CF, Meroz Y, Kurz V, Bdolah-Abram T, Babchenko A, et al. Pulse oximeter perfusion index as an early indicator of sympathectomy after epidural anesthesia. Acta Anaesthesiol Scand. 2009;53(8):1018–26. pmid:19397502
  33. 33. Kus A, Gurkan Y, Gormus SK, Solak M, Toker K. Usefulness of perfusion index to detect the effect of brachial plexus block. J Clin Monit Comput. 2013;27(3):325–8. pmid:23397432
  34. 34. Gatson BJ, Garcia-Pereira FL, James M, Carrera-Justiz S, Lewis DD. Use of a perfusion index to confirm the presence of sciatic nerve blockade in dogs. Vet Anaesth Analg. 2016;43(6):662–9. pmid:27109568
  35. 35. Endo Y, Miyasho T, Imahase H, Kawamura Y, Sakamoto Y, Yamashita K. Use of perfusion index to detect hemodynamic changes in endotoxemic pigs. J Vet Emerg Crit Care (San Antonio). 2020;30(5):534–42. pmid:32652875
  36. 36. Tardy Y, Meister JJ, Perret F, Brunner HR, Arditi M. Non-invasive estimate of the mechanical properties of peripheral arteries from ultrasonic and photoplethysmographic measurements. Clin Phys Physiol Meas. 1991;12(1):39–54. pmid:2036772
  37. 37. Gökdemİr GŞ. Evaluation of the pleth variability index, perfusion index, and other physiological parameters after COVID-19. Eur Rev Med Pharmacol Sci. 2023;27(22):11185–91. pmid:38039050
  38. 38. de Courson H, Julien-Laferrière T, Georges D, Boyer P, Verchère E, Biais M. The ability of Oxygen Reserve Index® to detect hyperoxia in critically ill patients. Ann Intensive Care. 2022;12(1):40. pmid:35576087
  39. 39. Szmuk P, Steiner JW, Olomu PN, Ploski RP, Sessler DI, Ezri T. Oxygen Reserve Index: a novel noninvasive measure of oxygen reserve--a pilot study. Anesthesiology. 2016;124(4):779–84. pmid:26978143
  40. 40. Tsymbal E, Ayala S, Singh A, Applegate RL 2nd, Fleming NW. Study of early warning for desaturation provided by Oxygen Reserve Index in obese patients. J Clin Monit Comput. 2021;35(4):749–56. pmid:32424516