Abstract
Background
Anatomy-based fitting (ABF) for cochlear implant users is a new era that seeks improved outcomes. Recently, different imaging modalities, such as plain X-rays, have been proposed to build the ABF as an alternative to the computed tomography (CT) scan. This study aimed to assess the feasibility and validity of OTOPLAN® software in building ABF using plain X-ray imaging.
Patients and methods
A retrospective evaluation of postoperative CT scans and plain X-ray post-op images of 54 patients was analyzed using the OTOPLAN® software. The post-op analysis was done for the angular insertion depth (AID) and center frequency of each electrode contact using both imaging modalities. Moreover, inter-rater reliability was assessed for measurements obtained from CT scans and plain X-ray images.
Results
Non-significant statistical and clinical mismatches were detected when comparing the AID and center frequency measurements assessed using CT and X-rays. The absolute difference between CT and X-ray approaches ranged from 0.0 to 4.6 degrees for AID and 0.2 to 0.5 semitone for frequency. Moreover, the AID and the frequency measurements from CT and X-ray images demonstrated almost perfect agreement between the raters. The inter-observer reliability for CT scans showed that the intraclass correlation coefficient (ICC) exceeded 0.97 for AID and 0.95 for the frequency across all electrode contacts.
Citation: Alahmadi A, Abdelsamad Y, Hafez A, Hagr A (2024) X-ray guided anatomy-based fitting: The validity of OTOPLAN. PLoS ONE 19(11): e0313567. https://doi.org/10.1371/journal.pone.0313567
Editor: Renato S. Melo, UFPE: Universidade Federal de Pernambuco, BRAZIL
Received: July 19, 2024; Accepted: October 26, 2024; Published: November 15, 2024
Copyright: © 2024 Alahmadi et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Data Availability: After consulting with the ethics committee, we are unable to share the data without first submitting an official request to them. This request will undergo a process of evaluation and may or may not be approved. Therefore, All data used in the analysis of this study can be made available upon request from the corresponding author or from our organization through this link. https://x.com/kaesc1?lang=ar.
Funding: The author(s) received no specific funding for this work.
Competing interests: The authors have declared that no competing interests exist.
Introduction
Cochlear implantation (CI) stands as an effective treatment option for patients with severe to profound hearing loss that can significantly improve speech perception and communication [1]. The human cochleae exhibit a wide range of variations among individuals, mainly in the cochlear duct’s shape, length, and coiling pattern and even at the intra-cochlear level [2, 3]. These variations lead to the distinction of the tonotopic organization of sound frequency along the cochlea’s basilar membrane, which impacts the individuals’ sensitivity to specific frequencies and auditory perception [4]. Understanding the intricate anatomy of the cochlea plays an essential role in different aspects of the CI, such as surgical planning, electrode insertion and positioning, and post-operative programming.
The performance of CI recipients is markedly influenced by various factors such as the age at implantation, residual hearing and auditory function, device and surgical factors, and the quality of the CI processor programming [5]. One crucial aspect of the processor programming is the frequency allocation for each electrode contact. In current clinical practice, programming software typically relies on the default frequency fitting based on general population averages or standardized parameters [6]. However, this generalized default fitting approach does not account for the interindividual variation in cochlear anatomy, potentially resulting in a frequency-to-place mismatch [7].
Recent research has highlighted that incorporating anatomical variations in CI programming could potentially improve the outcome of the CI. It has been reported that implanted post-lingually deaf patients demonstrated better speech discrimination and rapid adaptation after employing anatomical mapping with subsequent frequency allocation [8].
OTOPLAN® is a computer-based otological surgical planning software introduced by MED-EL (Innsbruck, Austria) in collaboration with CASCINATION AG (Bern, Switzerland). This software offers an easy way to analyze the anatomy of the temporal bone, estimating various cochlear metrics such as cochlear duct length (CDL) and angular insertion depth (AID) [9, 10]. This software plays a pivotal role in anatomy-based fitting (ABF) for CI, which involves tailoring frequency allocation to the individual’s unique cochlear anatomy, aiming to maximize the overall auditory performance [11]. To guide personalized programming using ABF, post-operative radiological imaging data should be imported to the OTOPALN® and then to the programming software to assign a specific frequency range for each electrode contact based on its position inside the cochlea utilizing knowledge of the patient’s anatomy [12].
Computed tomography (CT) is considered a valuable tool for the preoperative assessment of CI and postoperative assessment of selected cases such as suspected malfunction of the device [13, 14]. However, routine post-operative CT scanning may be restricted by specific considerations, such as exposure to a hefty dose of ionizing radiation, which may be associated with potential risks for CI patients [15]. Additionally, it is costly, especially if not covered by health insurance. Furthermore, it is time-consuming, and children may move, which may cause artifacts. Fourth, there is difficulty in localizing the center point in each electrode. Fifth, it is done in the radiology department, which requires an appointment in most centers. These concerns about using post-operative CT put limitations on using ABF in routine clinical settings.
In contrast, post-insertion plain X-ray is routinely used to visualize the CI electrode position in the cochlea in many centers [16]. The first published results on ABF mapping using the post-insertion X-ray images are promising as they demonstrated the validity of this approach in AID measurements for frequency allocation in Comparison to CT-based measurements [17]. It is worth noting that in this study, the X-ray images were manually examined using the RadiAnt DICOM Viewer [17].
To date, no research has assessed the feasibility of analyzing the X-ray images with OTOPLAN® software to obtain the AID and its corresponding center frequency. To address this gap, the current study aimed to assess the application of OTOPLAN® software in performing post-op analysis on plain X-ray imaging by evaluating the AID and corresponding frequency for each electrode contact that would be used for ABF.
Subjects and methods
This retrospective study commenced on 1 November 2023 after obtaining Institutional Review Board approval (IRB No. E-23-7997). Convenience sampling was used to select patients who fulfilled all the inclusion criteria, and data was retrieved from CI patients’ medical records that were presented to our tertiary cochlear implant center. Patient data was extracted anonymously; therefore, authors had no access to information that could identify individual participants.
Patients’ selection
Based on reviewing the CI patients’ records, 50 patients from our Center, were enrolled in the present study. Our inclusion criteria were as follows: ears with normal cochlea that underwent cochlear implant surgery, with complete CI insertion, and those who performed postoperative temporal bone CT scans due to exceptional circumstances such as mastoiditis, swelling over the magnet, or exposure to trauma. The study included all patients who underwent post-operative plain X-ray imaging as a part of their clinical routine. Patients with cochlear malformation, no post-operative CT, insufficient CT imaging resolution, inner ear anomalies, and partial electrode insertions were excluded.
Image analysis
Using both post-op CT scans and plain X-ray images (cochlear view), the manual approach of OTOPLAN® 3.1.0 (V5) was used to calculate the AID and estimate the corresponding frequency for each electrode contact. These measurements were performed by two independent observers with the same level of training on the software, used the software professionally for at least three years, and have a solid understanding of the ear anatomy.
For CT images, the analysis was conducted as follows:
- The post-op CT image was uploaded into OTOPLAN® data management.
- The observers manually defined the cochlear view, cochlear diameter (A value), width (B value), and height (H value).
- Then, the software automatically calculated the cochlear duct length (CDL), the AID, and the corresponding center frequency for each contact.
As a next step, for plain X-ray images, the analysis was conducted as follows:
- The post-op X-ray image was uploaded into OTOPLAN® data management.
- The observers manually defined the cochlear coordinates–round window and cochlear center points.
- Once the electrode contact points were set manually, the software automatically calculated each contact’s AID and corresponding center frequency.
The mean of both raters’ measurements for each sample was used for further analysis.
Sample size calculation
Assuming that the detected effect size between Ct and X-ray measured angular insertion depth (Cohen’s d for paired t-test, 1982) was 0.58, estimated from Alahmadi et al. study [17]. The required optimal sample size for paired CI patients at a 5% level of significance to achieve 90% power of the study will be 34 cases, adding to a 20% dropout rate (7 more cases) to get a total of 41 patients’ records required to conduct the study.
Statistical analysis
Data collected for cochlear parameters, AID, and frequencies for each electrode contact were analyzed using R software version 4.2.2, "Innocent and Trusting." Descriptive statistics such as means, standard deviations (SD), and ranges were used for quantitative variables, whereas frequency and percentage were applied for categorical variables.
Intra-class correlation coefficient (ICC) was used to assess the inter-observer reliability between two independent readers for AID and frequency measurements obtained using CT scan and X-ray imaging. ICC values were interpreted based on the following: poor (<0.40), fair (0.40–0.59), good (0.60–0.74) and excellent (0.75–1.00). Furthermore, comparative analyses between measured AID and frequency using CT scan and X-ray were carried out using paired t-test, or Wilcoxon signed rank test, according to the normality assumption tested using the Shapiro–Wilk test. Significance was determined by P values less than 0.05 and a Confidence Interval (CI) of 95%.
Results
The study was conducted on 50 cochlear implanted ears; 56.0% were right-sided, and 44.0% were left-sided. The mean age was 16.9 ± 12.5 years, and the mean age at implantation was 9.0 ± 12.6 years old, with males representing about 60.0% and females 40.0%. Notably, 72.0% of cases were implanted with FLEX28, and about 28.0% used the FORM24 electrode array (MED-EL, Innsbruck, Austria). The average cochlear parameters measured values were 9.4 ± 0.6 mm for A-value, 7.0 ± 0.5 mm for B-value, 3.7 ± 0.3 mm for H-value, and 36.9 ± 2.4 mm for CDL as detailed in Table 1.
Inter-observer reliability for AID and frequency measurements obtained from CT scan and X-ray
The AID and the frequency measurements, assessed via CT by two independent readers, demonstrated almost perfect agreement. The inter-observer reliability while using CT scans evaluated by the ICC exceeded 0.96 for AID and 0.94 for the frequency across all twelve electrode contacts—similarly, the AID and frequency measured using X-ray images displayed almost perfect agreement between the two raters. While using X-ray images, the inter-observer reliability demonstrated ICC of at least 0.92 for the AID and 0.9 for the frequency across all electrode contacts (Table 2).
Comparative analysis of AID and frequency measurements obtained from CT scan and X-ray
The mean AIDs measured by CT and X-ray across all electrode contacts are summarized in Table 3 and depicted in Figs 1 and 2. The absolute difference between both approaches ranged from 0.47 to 4.74 degrees, indicating non-significant mismatching (p values > 0.05) across all electrode contacts.
Represented as means, and error bars represent the standard deviations.
Represented as means and error bars represented the 95% Confidence Intervals.
When assessing the frequency across the twelve-electrode contacts, the average frequency errors between measurements obtained from CT and X-ray ranged from -0.29 to 0.65 semitones, indicating non-significant mismatching (p values > 0.05) across all electrode contacts as presented in Table 3, Figs 3 and 4.
Represented as means, and error bars represent the standard deviations.
Showing the best-fitted line with a 95% Confidence Interval.
Discussion
OTOPLAN® is a valuable tool for preoperative planning and post-operative electrode placement analysis of CI [18]. One of the principal uses of this planning software is to facilitate ABF, which aims to reduce frequency-to-place mismatch by ensuring precise electrode selection based on individual cochlear anatomy [19]. ABF holds promise for improving outcomes across a broad range of CI recipients, such as patients with cochlear malformation, residual hearing, or asymmetric hearing loss [18, 20]. Recent studies have reported better speech performance in patients with a more minor frequency-to-place mismatch [21, 22]. ABF for CI necessitates post-operative radiological evaluation, which is done using CT or X-ray imaging modalities. However, many considerations must be addressed to ensure that the radiological evaluation needed for ABF will benefit the CI recipients without compromising their health. The data presented herein support the validity and reliability of OTOPLAN® software for first time the in analyzing post-operative X-ray images and in building ABF.
In the present study, OTOPLAN® software was used for postoperative assessment of AID and frequency across the twelve-electrode contacts, using routine post-insertion X-ray images in Comparison to post-operative CT scans. Our results demonstrated that the inter-observer reliability of the AID based on post-operative CT and X-ray measurements was almost perfect, with an ICC exceeding 0.97 for CT and 0.92 for X-ray across all twelve electrode contacts. The reliability of this planning software in determining the AID of the array has been reported in an earlier study using post-operative CT scans [23]. Of note is that no previous study evaluated the feasibility of using this software in ABF using post-insertion X-ray images. The results of one previous study conducted by Alahmadi et al. were consistent with our results. However, their results depended on manually analyzing the post-insertion X-ray images using a DICOM viewer [17]. Our results demonstrated perfect agreement between the two readers in all twelve electrode contacts, which is closer to previously published findings [17].
In the present study, the absolute difference of AIDs measured by CT and X-ray ranged from 0.0 to 4.6 degrees, indicating non-significant mismatching across all electrode contacts. Interestingly, the C6 electrode contact exhibited the most remarkable absolute difference in AID measurements, with a value of 4.55 degrees between CT and X-ray approaches. These results agree with a previous study, which revealed that the absolute difference in AID measurements between the X-ray and CT was minimal for all electrodes (absolute mean difference of 4.7 degrees), with the highest difference at the C6 electrode [17]. This might be due to the slightly different appearance of electrode contacts when the electrode array approaches the first-to-second turn. However, further research might be suggested to gain valuable insights into possible reasons for the finding. In addition, Yoshimura et al. demonstrated a strong linear correlation between the AID measurements obtained from preoperative CT and those measured on post-operative X-ray [24]. Similarly, Fernandes et al. reported that X-rays and CTs analyzed manually provided an equivalent mean insertion depth with a mean difference of− 0.9 degrees [25]. Another study based on histological analysis of the cadaveric temporal bone model supported our findings, demonstrating that both radiographs and CTs can provide reasonably accurate angular depth of insertion measurements [26].
Next, OTOPLAN® used the retrieved AID to calculate the corresponding frequency automatically. When assessing frequency, our results revealed that the frequency measurements assessed via CT and X-ray by two independent readers demonstrated almost perfect agreement with ICC = 0.95 for CT and 0.9 for X-ray across all twelve electrode contacts. In addition, we detected that the average frequency errors between measurements obtained from CT and X-ray ranged from -0.2 to 0.5 semitone, indicating neither statistically nor clinically significant mismatching across all electrode contacts. These findings align with the results of a previous study, which reported that the mean deviation in the absolute semitone between the mean of each electrode contact central frequency on CT and X-ray was (0.6 ± 0.4) with a median of 0.5 semitones [17].
Post-operative imaging of CI recipients is fundamental to providing information about the electrode position [27]. Although CT is highly accurate, it is associated with exposing patients to a considerable amount of ionizing radiation and has higher levels of metal artifacts [28]. It has been recommended that extensive postoperative radiological evaluation using CT for children should be avoided except for cases of malformation or any other complications [29]. On the other hand, using post-operative X-ray to determine electrode position is currently considered routine practice in clinics for uncomplicated cases, as it provides an accurate position inside the cochlea and reliable measurements for CI insertion depth [16, 27]. X-rays provide a lower-cost, lower-radiation option, which may be especially beneficial for pediatric patients more sensitive to ionizing radiation. However, X-rays have significant limitations when considering this approach in a clinical setting, including lower spatial resolution than CT scans, which may limit the capacity to detect some electrode-related problems. In addition, to ensure accurate measurements of the AIDs and corresponding frequencies, particular skull X-ray views have to be used, such as a modified Stenvers view (cochlear view) and Caldwell view [30]. Although these X-ray views are recognized as sufficient for evaluating the electrode insertion depth, the interpretations’ accuracy depends on the experience level of the radiology technicians and the readers [25, 31].
To our knowledge, this study is the first to assess the application of OTOPLAN® software in ABF utilizing routine plain X-ray imaging. Our findings supported the idea that analyzing post-operative X-ray images with this software is a valid and reliable approach for ABF. However, it is essential to acknowledge that there are some limitations in the present study; one of them is that the discussed approach of ABF is only compatible with the electrode array of MED-EL. So, further research is needed to evaluate the applicability of ABF with other cochlear implant systems. In addition, only the manual option of OTOPLAN® was used to analyze postoperative CT scans and X-ray images.
Acknowledgments
The authors thank Dr. Nikola Ivanovic for reviewing the protocol and Rahma Sweedy for doing the statistical analysis.
References
- 1. Rasmussen KMB, West NC, Bille M, Sandvej MG, Cayé-Thomasen P. Cochlear Implantation Improves Both Speech Perception and Patient-Reported Outcomes: A Prospective Follow-Up Study of Treatment Benefits among Adult Cochlear Implant Recipients. J Clin Med. 2022;11: 2257. pmid:35456353
- 2. Liu Y-K, Qi C-L, Tang J, Jiang M-L, Du L, Li Z-H, et al. The diagnostic value of measurement of cochlear length and height in temporal bone CT multiplanar reconstruction of inner ear malformation. Acta Otolaryngol. 2016;137: 119–126. pmid:27577263
- 3. Avci E, Nauwelaers T, Lenarz T, Hamacher V, Kral A. Variations in microanatomy of the human cochlea. J Comp Neurol. 2014/04/12. 2014;522: 3245–3261. pmid:24668424
- 4. Rutherford MA, von Gersdorff H, Goutman JD. Encoding sound in the cochlea: from receptor potential to afferent discharge. J Physiol. 2021/03/29. 2021;599: 2527–2557. pmid:33644871
- 5. Holden LK, Firszt JB, Reeder RM, Uchanski RM, Dwyer NY, Holden TA. Factors Affecting Outcomes in Cochlear Implant Recipients Implanted With a Perimodiolar Electrode Array Located in Scala Tympani. Otology & Neurotology. 2016;37: 1662–1668. pmid:27755365
- 6. Vaerenberg B, Smits C, De Ceulaer G, Zir E, Harman S, Jaspers N, et al. Cochlear Implant Programming: A Global Survey on the State of the Art. The Scientific World Journal. 2014;2014: 1–12. pmid:24688394
- 7. Canfarotta MW, Dillon MT, Buss E, Pillsbury HC, Brown KD, O’Connell BP. Frequency-to-Place Mismatch: Characterizing Variability and the Influence on Speech Perception Outcomes in Cochlear Implant Recipients. Ear Hear. 2020;41: 1349–1361. pmid:32205726
- 8. Di Maro F, Carner M, Sacchetto A, Soloperto D, Marchioni D. Frequency reallocation based on cochlear place frequencies in cochlear implants: a pilot study. Eur Arch Otorhinolaryngol. 2022/01/15. 2022;279: 4719–4725. pmid:35032205
- 9. George-Jones NA, Tolisano AM, Kutz JW Jr., Isaacson B, Hunter JB. Comparing Cochlear Duct Lengths Between CT and MR Images Using an Otological Surgical Planning Software. Otol Neurotol. 2020;41: e1118–e1121. pmid:32925847
- 10. Iyaniwura JE, Elfarnawany M, Ladak HM, Agrawal SK. An automated A-value measurement tool for accurate cochlear duct length estimation. J Otolaryngol Head Neck Surg. 2018;47: 5. pmid:29357924
- 11. Lassaletta L, Calvino M, Sanchez-Cuadrado I, Gavilán J. Does it make any sense to fit cochlear implants according to the anatomy-based fitting? Our experience with the first series of patients. Frontiers in Audiology and Otology. 2023;1.
- 12. Kurz A, Müller-Graff F-T, Hagen R, Rak K. One Click Is Not Enough: Anatomy-Based Fitting in Experienced Cochlear Implant Users. Otol Neurotol. 2022;43: 1176–1180. pmid:36351223
- 13. PL LR Júnior, Rocha MD, Walsh PV, Antunes CA, DF Calhau CM. Evaluation by imaging methods of cochlear implant candidates: radiological and surgical correlation. Braz J Otorhinolaryngol. 2008;74: 395–400. pmid:18661014
- 14. Widmann G, Dejaco D, Luger A, Schmutzhard J. Pre- and post-operative imaging of cochlear implants: a pictorial review. Insights Imaging. 2020;11. pmid:32803542
- 15. Cao C-F, Ma K-L, Shan H, Liu T-F, Zhao S-Q, Wan Y, et al. CT Scans and Cancer Risks: A Systematic Review and Dose-response Meta-analysis. BMC Cancer. 2022;22: 1238. pmid:36451138
- 16. Anne S, Juarez JM, Shaffer A, Eleff D, Kitsko D, Sydlowski S, et al. Utility of intraoperative and postoperative radiographs in pediatric cochlear implant surgery. Int J Pediatr Otorhinolaryngol. 2017;99: 44–48. pmid:28688564
- 17. Alahmadi A, Abdelsamad Y, Thabet EM, Hafez A, Alghamdi F, Badr KM, et al. Advancing Cochlear Implant Programming: X-ray Guided Anatomy-Based Fitting. Otol Neurotol. 2024;45: 107–113. pmid:38206059
- 18. Gatto A, Tofanelli M, Costariol L, Rizzo S, Borsetto D, Gardenal N, et al. Otological Planning Software-OTOPLAN: A Narrative Literature Review. Audiol Res. 2023;13: 791–801. pmid:37887851
- 19. Kurz A, Herrmann D, Hagen R, Rak K. Using Anatomy-Based Fitting to Reduce Frequency-to-Place Mismatch in Experienced Bilateral Cochlear Implant Users: A Promising Concept. J Pers Med. 20230708th ed. 2023;13. pmid:37511722
- 20. Rader T, Schrank L, Spiegel JL, Nachtigäller P, Spiro JE, Hempel J-M, et al. Comparison of speech perception in bimodal cochlear implant patients with respect to the cochlear coverage. HNO. 2023;72: 17–24. pmid:37608133
- 21. Lambriks L, van Hoof M, Debruyne J, Janssen M, Chalupper J, van der Heijden K, et al. Imaging-based frequency mapping for cochlear implants—Evaluated using a daily randomized controlled trial. Front Neurosci. 2023;17: 1119933. pmid:37123376
- 22. Mertens G, Van de Heyning P, Vanderveken O, Topsakal V, Van Rompaey V. The smaller the frequency-to-place mismatch the better the hearing outcomes in cochlear implant recipients? European Archives of Oto-Rhino-Laryngology. 2021;279: 1875–1883. pmid:34131770
- 23. Canfarotta MW, Dillon MT, Buss E, Pillsbury HC, Brown KD, O’Connell BP. Validating a New Tablet-based Tool in the Determination of Cochlear Implant Angular Insertion Depth. Otol Neurotol. 2019;40: 1006–1010. pmid:31290802
- 24. Yoshimura H, Watanabe K, Nishio SY, Takumi Y, Usami SI. Determining optimal cochlear implant electrode array with OTOPLAN. Acta Otolaryngol. 20231108th ed. 2023;143: 748–752. pmid:37737708
- 25. Fernandes V, Wang Y, Yeung R, Symons S, Lin V. Effectiveness of skull X-RAY to determine cochlear implant insertion depth. J Otolaryngol Head Neck Surg. 2018;47: 50. pmid:30176926
- 26. Gallant S, Friedmann DR, Hagiwara M, Roland JT, Svirsky MA, Jethanamest D. Comparison of Skull Radiograph and Computed Tomography Measurements of Cochlear Implant Insertion Angles. Otology & Neurotology. 2019;40: e298–e303. pmid:30741910
- 27. Vogl T, Tawfik A, Emam A, Naguib N, Nour-Eldin A, Burck I, et al. Pre-, Intra- and Post-Operative Imaging of Cochlear Implants. RöFo—Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren. 2015;187: 980–989. pmid:26327670
- 28. Wei F, Li J, Zhou C, Li Y, Wang X, Huang B, et al. Combined application of single-energy metal artifact reduction and reconstruction techniques in patients with Cochlear implants. J Otolaryngol Head Neck Surg. 2020;49: 65. pmid:32907631
- 29. Aschendorff A. Imaging in cochlear implant patients. GMS Curr Top Otorhinolaryngol Head Neck Surg. 2011;10: Doc07. pmid:22558057
- 30. XU J, XU S, COHEN L, CLARK G. Cochlear view: Postoperative radiography for cochlear implantation. Am J Otolaryngol. 2000;21: 49–56. pmid:10651435
- 31. McClenaghan F, Nash R. The Modified Stenver’s View for Cochlear Implants—What do the Surgeons Want to Know? J Belg Soc Radiol. 2020;104: 37. pmid:32676547