Figures
Abstract
Background
GammaKnife (GK) and CyberKnife (CK) have been the mainstay stereotactic radiosurgery (SRS) solution for multiple brain metastases (MBM) for several years. Recent technological advancement has seen an increase in single-isocentre C-arm linac-based SRS. This systematic review focuses on dosimetric and geometric insights into contemporary MBM SRS and thereby establish if linac-based SRS has matured to match the mainstay SRS delivery systems.
Methods
The PubMed, Web of Science and Scopus databases were interrogated which yielded 891 relevant articles that narrowed to 20 articles after removing duplicates and applying the inclusion and exclusion criteria. Primary studies which reported the use of SRS for treatment of MBM SRS and reported the technical aspects including dosimetry were included. The review was limited to English language publications from January 2015 to August 2023. Only full-length papers were included in the final analysis. Opinion papers, commentary pieces, letters to the editor, abstracts, conference proceedings and editorials were excluded. The Preferred Reporting Items for Systematic Reviews and Meta-Analyses guidelines were followed. The reporting of conformity indices (CI) and gradient indices, V12Gy, monitor units and the impact of translational and rotational shifts were extracted and analysed.
Results
The single-isocentre technique for MBM dominated recent SRS studies and the most studied delivery platforms were Varian. The C-arm linac-based SRS plan quality and normal brain tissue sparing was comparable to GK and CK and in some cases better. The most used nominal beam energy was 6FFF, and optimised couch and collimator angles could reduce mean normal brain dose by 11.3%. Reduction in volume of the healthy brain receiving a certain dose was dependent on the number and size of the metastases and the relative geometric location. GK and CK required 4.5–8.4 times treatment time compared with linac-based SRS. Rotational shifts caused larger changes in CI in C-arm linac-based single-isocentre SRS.
Citation: Mukwada G, Chamunyonga C, Rowshanfarzad P, Gill S, Ebert MA (2024) Insights into the dosimetric and geometric characteristics of stereotactic radiosurgery for multiple brain metastases: A systematic review. PLoS ONE 19(8): e0307088. https://doi.org/10.1371/journal.pone.0307088
Editor: Huidong Wang, The First Hospital of Jilin University, CHINA
Received: April 21, 2024; Accepted: June 30, 2024; Published: August 9, 2024
Copyright: © 2024 Mukwada et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Data Availability: All relevant data are within the manuscript and its Supporting information files.
Funding: The author(s) received no specific funding for this work.
Competing interests: The authors have declared that no competing interests exist.
Introduction
Radiation therapy has been a mainstay modality for the management of many types of cancer. However, despite efforts to eradicate the disease, approximately 20–40% of cancer patients develop brain metastases [1–10]. Depending on the number of brain metastases, available technical resources and professional competency, a choice of whole brain radiotherapy (WBRT), stereotactic radiosurgery (SRS), stereotactic radiation therapy (SRT) or a combination of WBRT, and SRS can be administered. A large body of evidence demonstrates that both SRS (high dose in a single fraction) and SRT (high dose delivered in 1–5 fractions) result in less neurocognitive decline compared with WBRT [1, 6, 11, 12]. Due to their neurocognitive sparing, and positive impact on quality of life, SRS or SRT is now the standard of care for oligometastases and is increasingly preferred for all brain metastases [12, 13]. In a case-matched study comparing the results of 2–9 brain metastases against more than 10 brain metastases, Yamamoto et al., [14] demonstrated that SRS for more than 10 metastases was not unfavourable. This study along with several others [15–19] has resulted in increased clinical demand for SRS for multiple brain metastases (MBM).
SRS and SRT treatment planning and delivery systems for MBM are varied and include Gamma Knife (GK), CyberKnife (CK), Tomotherapy, C-arm linear accelerator (linac) based systems and the recently developed Zap-X radiosurgery system. GK was the first SRS-dedicated solution, and its current design makes use of 192 Cobalt-60 sources that produce precisely focused radiation beams at isocentre through fixed collimators. The CK utilises a linear accelerator that is mounted on a robotic arm guided by real-time imaging. CK SRS delivery is through one of the following beam collimations; cones, Iris (hexagon shaped aperture) and InCise multileaf collimators (MLCs) [20]. Tomotherapy is a general-purpose radiation therapy system that delivers an MLC-modulated fan radiation beam from a bore-mounted rotating source coupled with helical couch movement [21, 22]. C-arm linac-based SRS has several variations that are linked to the individual vendor’s ingenuity or a combination with other vendors’ solution such as Brain Elements Treatment Planning solution (BETPS) also referred to as Brainlab Multiple Mets Elements (Brainlab MME) (BrainLab Inc., Munich, Germany) delivered via a Varian linear accelerator. Delivery is via cones or small-width MLCs (= <5mm at isocentre) through dynamic arc therapy (DCAT) or intensity modulated beams coupled with advanced imaging. Zap-X is a self-shield 3 MV linear accelerator mounted on a gimbal system that is capable of delivering non-coplanar beams and has inbuilt kilovoltage imaging capability allowing real time imaging [23]. Of these planning and delivery solutions, GK and CK are dedicated SRS solutions and are generally regarded as the gold standard in the treatment of MBM [24–28]. Although Zap-X is a dedicated SRS system, its first clinical use was in 2019 so published clinical experience is limited, and is evolving as a technique as demonstrated by frequent software updates [29–31].
Traditional C-arm linac-based SRS approaches for MBM treatment involve multiple-isocentre planning and delivery is time consuming as the metastases are treated one by one, with the patient shifted and isocentre verified before each lesion is treated. Technological advancements in treatment planning systems, high-definition MLCs, frameless masks (immobilisation), automated 6 degrees of freedom (6DoF) couch corrections and improved imaging capabilities have transformed MBM SRS. These recent developments favour a single-isocentre technique [32] with studies demonstrating non-inferiority to the traditional multiple-isocentre technique, except for reduced dose coverage due to rotation shifts which can be mitigated [13, 33, 34]. Although BETPS pioneered single isocentre MBM delivered via conformal arcs, VMAT based MBM SRS is flourishing more recently due to availability.
A systematic review by Rozati et al.,[35] on overall survival following SRS for 10 or more brain metastases showed GK to have less necrosis (12.2%) compared to C-arm linac based SRS (35%). Several overall survival studies conducted by De La Lena et al.,[36] on patients with MBM treated with CK between 2011 and 2017 reported no late complications such as necrosis, with local control exceeding 90% at 1 year.
The variation in radio-necrosis rates is likely attributed to the larger planning target volume (PTV) margins typically used in C-arm linac-based SRS compared to GK and CK. Advances in imaging, immobilisation and the development of fine isocentre SRS machines, is contributing to the reduction in PTV margins for C-arm linac based SRS [6, 13, 32] potentially reducing toxicity and overall survival across different SRS technologies. Furthermore, there are many studies on overall survival for MBM GK and CK SRS deliveries [14, 16, 35–41] and there is a need for such studies for C-arm linac based for a fair comparison [42].
Most SRS techniques are frameless and require thermoplastic masks for immobilisation. Initial translational and rotational setup errors are corrected through couch translations and 6DoF couch rotations respectively. Despite accurate initial patient positioning, several studies have demonstrated intra-fraction motion in SRS [43–47]. Dosimetry for the single-isocentre technique can be adversely impacted by the resulting positioning errors [47, 48]. Nakano et al., [49] recommend a maximum PTV to isocentre distance of 5.5cm for a 1.5mm diameter target and 0.5° rotation, beyond which loss of dose coverage is clinically significant. Prentou et al., [48] recommend a maximum distance of 4cm from isocentre.
The quality of SRS treatment plans is generally evaluated using the following dosimetric parameters; Paddick conformity index (CIp), RTOG conformity index (CIRTOG), Paddick gradient index (GIp), V12Gy, number of MUs, among others. CIp, CIRTOG and GIp are defined as per Clark et al., [50] and Paddick [50, 51] as shown in Eqs 1–3. (1) (2) (3) Here, TV is the target volume, TVPV is the target volume covered by the prescription isodose, PV is the volume of the prescription isodose, PV50% is the volume of the isodose line at 50% of the prescription.
The modified Paddick conformity index (nCI) is also reported, being essentially the inverse of the CIp [52, 53]. Understanding the current use of these indices in the context of recent technical advances is of paramount importance for standardising their utilisation in clinical departments. Homogeneity indices (HI) are not directly relevant to SRS as the GTV or PTV dose tends to be inherently inhomogeneous. The limitations of the current plan metrics are acknowledged and a novel efficiency index that depends on the integral dose has been proposed [54].
Contemporary SRS treatments are prescribed to the isodose surface that covers the optimal percentage of the PTV while maintaining an optimally reduced dose to organs at risk (OAR) [55]. The prescription isodose surface depends on the SRS technique; for example, GK prescriptions are usually to the 50% isodose surface, CK prescriptions are made at around the 70–85% isodose surface while the C-arm linac-based approach uses approximately the 75–90% isodose surface covering the optimal PTV. Several authors demonstrated that the prescription isodose diameter influences the GI [56].
Relative affordability, fast delivery and comparable treatment outcomes have fuelled the increase in C-arm linac-based SRS relative to GK and CK [8, 57]. Volumetric arc therapy (VMAT) or dynamic conformal arc therapy (DCA) approaches have enabled innovation in single-isocentre SRS. This review aims to summarise and evaluate plan quality indices and the impact of geometric shifts on dose coverage in MBM SRS. Furthermore, the radiation dose to the brain healthy tissue (Brain minus PTV) will be evaluated. Ultimately, the review seeks to establish whether the contemporary MBM SRS techniques compare with the mainstay MBM SRS techniques in terms of achievable quality and deliverability.
Methods
Search and screening
A systematic review of the literature was performed using the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) guidelines [58]. The associated flow chart is shown in Fig 1. Peer-reviewed articles were sourced from PubMed, SCOPUS, and Web of Science databases, and manual search of the principal relevant journals and the last search was on 30 August 2023. According to Haddaway et al., [59] searching 5–10 databases offers no significant advantage over searching 2–3 relevant databases. The selected databases for this review are highly reputable scientific resources pertinent to the study, mitigating selection bias. To reduce publication bias [59, 60], additional searches were undertaken manually via Google Search and Google Scholar. Relevant search terms and the records identified through primary and secondary search are shown in Table 1. Relevant syntax operators including Boolean operators, additional filters and limiters were used. The articles retrieved were exported to EndNote version 20. To avoid bias, two authors independently assessed the remaining articles for eligibility. The final articles were double-checked for eligibility by two separate authors. Furthermore, the Critical Appraisal Skills Programme (CASP) checklist forms were completed (S1 Checklist). CASP comprises 10 specific questions suitable for assessing potential bias in articles for systematic reviews.
Inclusion and exclusion criteria
Primary studies which reported the use of SRS for treatment of MBM SRS including SRT and reported the technical aspects including dosimetry were evaluated based on the selection criteria. To eliminate screening bias, the inclusion and exclusion criteria were determined beforehand. In addition, bias towards particular research groups or regions were addressed by screening the abstracts and manuscripts based on content and not authors or participating centres. The review was limited to English language publications from January 2015 to August 2023. Only full-length papers were included in the final analysis. Opinion papers, commentary pieces, letters to the editor, abstracts, conference proceedings and editorials were excluded. Primary studies that had a maximum of 1 metastasis were excluded. Some relevant records that did not meet the inclusion criteria are included in the discussion.
Descriptive analysis
Treatment technical aspects including the treatment modalities and the quality indices were extracted manually from the articles under review to a Microsoft excel spreadsheet and are shown in Table 2. Also, reported information on geometric shifts and their impact on dose coverage were extracted. The extracted data were insufficient for rigorous statistical analysis, so was not used. Instead, graphs and plots were employed for easier data visualization and interpretation.
CIp>1 is nCI. SI-, MI-, DI-, RSI-, and C- stands for single-isocentre, multiple isocentre, dual isocentre, restricted isocentre, and conventional, respectively. A dash indicates missing information in the respective study.
Dosimetric analysis
It was assumed that the quality indices reported in the studies are comparable within the confines of minor differences in treatment planning systems (TPSs) index calculations, with calculation grid sizes assumed small enough for consistency [61]. The extracted mean CIp, nCI, CIRTOG, GIp and normal brain tissue irradiated by a dose of at least 12 Gy (V12Gy) were presented in chart format for easy visualisation and interpretation. Among other factors such as the beam quality, V12Gy is also influenced by the tumour size and total tumour volume (all tumours) [25, 57, 62] and hence the mean total volumes we also extracted. There were large variations in treatment times (including set-up) across techniques. For easy analysis and interpretation, MU ratios were calculated by dividing the MUs for an established technique by those for the comparator technique in the respective article or study. For example, in the study by Homfaier et al., [63] in Table 2, the DCA technique is considered established and is being compared with the SI-VMAT technique, and thus MUs for the DCA provide the normalisation. The range of MUs were also quoted for context.
Results
Descriptive analysis
The search strategy identified a total of 891 records which were narrowed down to 20 articles after applying the inclusion and exclusion criteria. Three studies included a comparison with GK, and another three studies made a comparison with CK. Among linac-based SRS studies there were 19 VMAT, 6 DCA, and only 1 intensity modulated radiation therapy (IMRT) delivery. These machines were coupled with different TPSs namely, 4 Iplan MME, 14 Eclipse, 3 Multiplan (CK), 3 Gamma plan (GK), 3 Monaco and 1 RayStation. Sixteen out of the 20 studies considered more than 4 metastases and the highest number of metastases per patient was 21. Some authors reported GTV total volume while others reported PTV total volume. The GTV total volume ranged from 0.02–10.5 (1.2–11.1) cc and PTV total volume ranged from 0.1–15.6 (0.5–38.6) cc and two outliers PTV total of 70.6 and 74.3 cc. PTV margin ranged from 0–3 mm. Where the PTV margin was not explicitly stated, the authors assumed the same PTV margin was used for all techniques under comparison. Nine studies reported MLCs with 2.5 mm width at machine isocentre, 5 with 5mm width, one reported use of an m3 micro-MLC (Brainlab Inc., Munich, Germany), and one use of a Versa HD MLC. Nominal energies included 16 x 6FFF, 1 x 6FF, 3 x 10FFF, and 3 x 192 Co-60 (GK). Eighteen of the 20 studies reported a single-isocentre technique study by itself or in comparison to a multiple-isocentre technique. Eclipse AAA was the most used dose calculation algorithm, and the most frequently used calculation grid sizes were in the range 1–1.25 mm.
Dosimetric analysis
Across the studies there were variations in the reporting of plan quality indices, tumour volumes and the dose to which the volume should be reported. Also, some studies reported mean values while others reported the median. In this review, the plans’ qualities were essentially evaluated per study and hence median was taken as mean to simplify data presentation. The variations of plan indices are shown in Figs 2–4. The ideal CIp, nCI, CIRTOG should be 1. From Fig 2, the mean CIp ranged from 0.532–0.97. The study with the largest CIp difference was CK vs HyperArc, 0.27 (0.87–0.60) in favour of CK [26]. Mean nCI ranged from 1.10–1.60. The study with the largest nCI difference was CK vs SI-VMAT, 0.4 (1.60–1.20) in favour of SI-VMAT [52]. Mean CIRTOG ranged 0.85–1.78. The study with the largest CIRTOG difference was VMAT vs DCA, 0.57 (1.78–1.21) in favour of VMAT [64].
For each column the first written technique is the reference/established one. For example, for MI-DCA vs RSI-DCA, MI-DCA is the reference. Some of the error bars were clipped to allow better visualisation. CI with no error bars had no standard deviation (stdev) in the respective studies. To simplify the figure, CIp above 1.0 refers to nCI.
For each column the first written technique is the reference/established one. For example, for MI-DCA vs RSI-DCA, MI-DCA is the reference. GI with no error bars had no stdev in the respective studies.
For each column the first written technique is the reference/established. For example, for MI-DCA vs RSI-DCA, MI-DCA is the reference.
The ideal GIp should be 3 or less [56]. The mean GIp was 4.30. From Fig 3, the mean GIp ranged from 3.06–8.01. The study with the largest GIp difference was VMAT vs DCA, 2.15 (8.01–5.86) in favour of VMAT [64]. SI-VMAT 6 vs 10FFF and VMAT opt vs VMAT non-opt showed similar GIp as shown in Fig 3. Overall, all studies showed the reference or established techniques to have better GI except GK vs DCA which favoured DCA [25].
Other studies reported V2Gy up to V16Gy, but only V12Gy was reported in Table 2 as many studies consider V12Gy > 10cm3 to be associated with radiation necrosis. From Fig 4, the mean V12Gy ranged from 2.00–55.30. The study with the largest V12Gy difference was CK vs SI-VMAT in favour of CK 30.5cm3 (37.0–6.50) [52]. In a study by Wu et al., [65] optimised in terms of couch and collimator angles reduced the V12Gy and mean normal brain dose by 3.1 ± 1.6% and 8.4 ± 2.9% respectively compared to an unoptimized VMAT technique. The reduction in the volume of the healthy brain receiving a certain dose was depended on the number and size of the metastases, the relative geometric location and amount of overlap. There was no uniform reporting of the PTV to isocentre distance impact on plan quality indices. In agreement, De Camargo et al., [66] reported and Tsui et al., [46] recommended a maximum of 3.6–3.7cm PTV to isocentre distance for ±2° rotational shift assuming a maximum allowable 0.2 change in CI. In Fig 5c. Prentou et al., showed a change in CI with distance from isocentre for 0.5°, 1.0° and 2.0° rational shift. In the worst rotational shift of ±2°, a 7.2% per cm change in CI was observed.
There was pronounced variation in treatment time or MUs across the SRS techniques. Liu et al., [27] found GK treatment times to be 4.5 times those of VMAT (72 mins for GK with new radiation sources and 16 mins for VMAT including setup and imaging). El Shafie et al., [52] reported median treatment time of 130 mins (CK) vs 13.7 mins (VMAT) including set up and imaging. Also, a study by Zhang et al., [28] found CK MUs to be 8.4 times compared to linac-based techniques. Overall, GK and CK require much more treatment time compared to linac-based SRS techniques. There is also significant variation in MUs for the linac-based SRS techniques as shown in Fig 5. HyperArc vs MI-VMAT showed the largest MU difference with the M-VMAT technique requiring 3.5 times MUs to compared to HyperArc [67].
Analysis of geometric shifts
The change in CIp due to a translational shift of between ±2 mm was found to be similar for multiple and single-isocentre techniques as shown in Fig 5a. However, rotational shifts caused large changes in CIp in single-isocentre compared to multiple-isocentre techniques as shown in Fig 6. Tsui et al., [46] showed that on average the conformity index decreases by a factor of 2.6 when rational shifts change between 0 and ±2°. Decreases in dose conformity of a factor of 3 were observed by Prentou et al., [48] for the same rational shifts.
For each column the first written technique is the Reference. For example, for DCA vs SI-VMAT, DCA is reference.
Discussion
Descriptive analysis
There has been a growing interest in linac-based SRS over the past 8 years, especially because of the widespread availability of VMAT technology, 6DoF couch correction, and the increased uptake of large doses per fraction for body stereotactic radiotherapy for oligometastases and inoperable primary tumours. This may also be attributed to the greater accessibility of linacs in comparison to dedicated SRS delivery systems such as CK and GK. Rapid developments of linac associated capabilities and complimentary TPS capabilities seem to be the catalyst, transforming SRS delivery. Capitalising on VMAT capabilities and optimising treatment parameters such as couch and collimator angles has seen the rise of single-isocentre SRS technique. The availability of dedicated linac based SRS treatment planning solutions such as BrainLab MME BETPS and HyperArc have dominated SRS treatment techniques in recent years. The results show that conventional, optimised VMAT techniques and HyperArc hosted on Eclipse as the most studied SRS techniques in recent articles.
This review showed increased usage of FFF beams relative to the traditional FF beams. The FFF beam provides similar dosimetric coverage compared to FF beams but also offers sharper penumbra, lower dose spillage and peripheral dose, and increased dose rates [73–75]. Use of 6FFF was reported more than 10FFF—this could be due to energy availability, or preference due to the slightly sharper dose fall off exhibited by 6FFF, as reported by Laoui et al., [70]. MLC width plays a role in dose conformity and thinner MLCs are recommended especially for smaller tumours [6, 13, 52]. Nowadays, 5mm MLC width is a minimum standard on conventional C-arm linacs and the comparable quality of treatment plans achieved with thinner MLCs has contributed to the increase in linac-based SRS. This was demonstrated by a survey conducted across Australia and New Zealand that showed 60% of the centres offering MLC linac-based SRS were using 5mm MLC width [76]. The use of MLCs for CK SRS has also been reported [77] which makes the treatment delivery faster compared to cones.
In the studies reviewed, PTV margins ranged from 0 – 2mm except for one linac-based technique planned on RayStation TPS that used 3mm. This was a retrospective comparative study involving Multiplan with 1mm PTV margin for CK and then an additional 2mm for linac based SRS. GK and CK had 0 or 1mm PTV margin at most. The PTV margins are influenced by imaging, immobilisation, and availability of setup error correction methods such as via 6DoF. Such PTV margins are supported by the multicentre benchmarking planning studies conducted by Eaton et al., [24]. Vergalasova et al., [10] in their multi-institutional dosimetric SRS evaluation, started with 2mm PTV margin and are reducing to 1mm margin due to improved confidence in their technique. In a randomised trial by Kirkpatrick et al., [78] the prospective study found no difference in local control between 1mm and 3mm PTV margin groups. However, the study found increased incidences of radiation necrosis for the 3mm PTV margin. In conclusion the study recommends a 1mm PTV margin for linac-based SRS and this support by other studies [6, 10].
Although the individual GTV sizes are usually small (<1 cc), the GTV total volume and PTV total volume can be large. Several studies have indicated similar treatment outcomes for patients with less than 10 metastases compared to those with 10 or more [14, 41, 79, 80]. Yamamoto et al., [14] demonstrated that small tumour volume was a predictor of overall survival and not the number of metastases. Grandhi et al., [80] also showed that the treatment volume was a significant variable to survival and that the number of metastases was not a determinant of survival or local control. A case study of a patient with 37 metastases by Hyde et al., [81] cemented the above point because at 18 months the patient had 100% local control. Generally, as the number of metastases increases, the GTV total volume also increases. Unfortunately, there is no consensus on the limit of the GTV total volume suitable for MBM SRS and this is also demonstrated by the variation in the reviewed GTV and PTV total volume. The different GTV sizes and shapes and GTV total volume makes treatment planning complex. To reduce toxicity in treatments with large GTV total volumes, the strategies include fractionating SRS and reducing the total dose [13, 82].
More than 90% of the reviewed articles involved a single-isocentre SRS technique. Different C-arm linac-based single-isocentre SRS techniques have been compared against each other such as DCA vs HyperArc, Conventional VMAT vs HyperArc, and DCA vs Conventional VMAT. This trend reflects a significant interest and increased usage of the single-isocentre technique, likely due to its availability and affordability. At the core of radiation therapy is the precise and accurate delivery of radiation to the target while sparing healthy tissues. Comparative studies on SRS technology have shown a reduced gap between technologies in terms of delivery accuracy. Emerging SRS technologies, such as ZAP-X, are not only meeting the requirements for precise and accurate dose delivery but are also addressing other considerations, such as the elimination of the need for radiation-shielded bunkers [23, 31].
While it is standard in healthcare for informed consent to be provided before treatment [83], with such emerging SRS technologies, it is paramount to provide patients with adequate information about these alternatives that offer similar outcomes [84]. For example, a study by Mizuno et al., [16] compared SRS to WBRT as the initial treatment for 10–20 brain metastases (BMs) and demonstrated that there were no significant differences in overall survival (OS) and neurological survival (NS) between treatment with SRS and WBRT for BMs. A similar systematic review and meta-analysis by Gaebe et al., [85] also provides evidence that SRS may achieve survival outcomes compared with WBRT in patients with small cell lung cancer, suggesting that some of these patients may benefit from first-line SRS. Given this evidence on quality of life and the shorter treatment duration offered by SRS, it is crucial that these options are discussed with patients.
Dosimetric analysis
A conformity index of 1 reflects an ideal PTV coverage and a decrease from this indicates a decrease in plan quality. There were variations in the CIp and nCI across the studies, and in some cases this evidence was conflicting. In the GK vs DCA study by Chea et al., [25]. GK had better CIp and in a separate GK vs SI-VMAT study by Liu et al., SI-VMAT had better nCI. El Shafie et al., [52] and Kadoya et al., [26] also reported opposing findings on the CIp for CK vs VMAT based SRS. El Shafie et al., [52] used Multiplan (for CK) and RayStationV8 (for SI-VMAT) TPSs and found CK CIp, 0.83 (nCI = 1.2) to be more conformal than SI-VMAT SRS which had CIp of 0.63 (nCI = 1.6). On the other hand, Kadoya et al., [26] used Multiplan (CK) and Eclipse (for SI-VMAT) TPSs and found CK CIp of 0.60 and less conformal to HyperArc SRS CIp of 0.87. The low CIp for SI-VMAT could be linked to the limitation or ability to plan VMAT SRS by RayStationV8 as also demonstrated by limited publications involving this TPS. The lower CIp on Multiplan (CK) was attributed to the number of tumours by the authors. According to Kadoya et al., [26] the lower CIp could be explained from the limitation of cone delivery system on CK, which may not perfectly fit the target especially for multiple targets compared to HyperArc that uses MLCs which can conform to the target [26]. This argument can be easily counteracted since CK can deliver dose from many angles thereby conforming to the target. This demonstrates the complexity of SRS and that the factors influencing the plan quality are not only the complex SRS delivery system.
In a study comparing CK, HyperArc and a standard VMAT SRS, Slosarek et al., [86] found minor differences in CIp between the three techniques i.e CK (0.87), HyperArc (0.86) and 0.81 for standard VMAT. In a separate study, Zhang et al., [87] assessed plan quality indices for CK, IMRT and coplanar VMAT and non-coplanar VMAT (standard VMAT). They found standard VMAT to have a slightly better CIp (0.9) than CK (0.86). By design and ability to treat from several beam angles it is expected that CK will have a better conformity than VMAT based SRS. However, a potentially significant influence of the planners’ ability and bias on plan quality indices has been suggested by other studies [53].
In general, C-arm linac-based SRS showed CIp and nCI were closer to 1 demonstrating comparable conformality across the techniques. HyperArc consistently performed better than any C-arm linac-based SRS it was compared with except DCA VMAT which marginally performed better (1.25 vs 1.24). The above discussion also highlighted the influence of the planner’s ability in these studies.
CIRTOG was only recorded for studies involving CK. In all cases, the CK had better conformity than the C-arm linac-based SRS. Zhang et al., [28] demonstrated that SI-VMAT SRS was more comparable to CK conformality than C-VMAT and IMRT SRS. The ability to deliver non-coplanar multiple beams might be the reason for the SI-VMAT SRS performing slightly better than the latter techniques.
A GI of 3 or less is considered achievable and reasonable [33, 56, 70]. The mean and median GIp were 1.73 and 1.71 higher than the expected Gradient index. This reflects a general difficulty in achieving the ideal gradient index across the techniques. The VMAT vs DCA study by Turkkan et al., [64] had the largest GIp difference and yet it’s using the same TPS and algorithm. This could be attributed to the TPS not being fully optimised for that SRS technique. This view is supported by the conflicting findings in DCA vs SI-VMAT. Velten et al., [53] found DCA to have better GIp. SI-VMAT 6 vs 10FFF and VMAT opt vs VMAT non-opt showed similar GIp. Chea et al., [25] found the DCA to have better GIp (3.22) compared to GK (4.09). Overall, more studies demonstrated that GK and CK had better dose fall off than C-arm linac-based SRS [26–28, 52]. The sharp dose fall off in GK and CK can be attributed to the ability to deliver dose from many optimised non-coplanar beam angles.
Other studies reported brain V2Gy up to V16Gy, but only brain V12Gy was reported in Table 2 and Fig 6 as many studies consider brain V12Gy < 10cm3 to be a technically achievable planning objective with SRS, and a parameter where the risk of radiation necrosis has been collected and reported across various studies [24, 27, 88–91]. Fractionated SRS (SRT) is recommended for brain V12Gy >10 cm3 to reduce the risk of radionecrosis [82]. A shortcoming is that some of these studies have often included the GTV tumour in the “brain” volume, when in fact the larger the PTV margin, the more normal brain is treated to high dose. Because some studies report on GK with no margin, others who do add a PTV margin treat normal brain to a higher dose and so we have different rates of radionecrosis reported for the same planning objective in different studies.
A combination of the direct beam and out-of-field (OOF) dose impacts the brain V12Gy. The OOF dose comes from head leakage, collimator, and patient scatter. In addition, the V12Gy depends on the number or and size of tumours. The CK vs SI-VMAT study by Kadoya et al., [26] showed CK V12Gy to be approximately 6 times better than SI-VMAT. This huge difference was not observed in any other study and can’t be fully explained. Pokhrel et al., [71] found HyperArc to have more V12Gy than DCA-VMAT. Otherwise, the other studies showed comparable V12Gy.
GK and CK demand longer treatment times in comparison to C-arm linac-based SRS. This substantially affects efficiency and serves as a motivating factor to explore other techniques. CK mostly use cones and technically require more MUs. There were also MU variations among the C-arm linac-based SRS. MI-VMAT technique required 3 times MUs to deliver the same dose compared to the HyperArc technique. Furthermore, the VMAT or DCA uses dynamic MLCs and that reduces treatment time. The reviewed papers did not address the time taken per patient from CT simulation to treatment. This time depends on many facets especially the clinical workflows of individual departments and is very important for bench marking especially for centres wanting to start an SRS program or those considering quality improvement.
Geometric shifts analysis
Studies show that intra-fraction motion for SRS can be up to 1.5mm and 2° [32], therefore the impact on dose coverage is normally studied up to 2mm translational and 2° rational shifts [13]. The compromise in CIp due to translational shifts was similar for both multiple isocentres and single-isocentre. However, the change in CIp was different between multiple-isocentre and single-isocentre for rational shifts. Single-isocentre treatments were greatly impacted compared to multiple isocentre treatments and this was observed by several authors [49, 92, 93]. Selvan et al., [94] observed a decrease in PTV coverage for 1–5° rotational error as the tumour radial distance increases from the isocentre in a single-isocentre SRS technique. Nakano et al., [49] performed extensive simulation of setup errors involving 6DoF couch and found that target coverage is more compromised by rotational setup errors in single-isocentre treatments. SRS delivery systems that cannot provide set-up accuracy within 0.5mm and 0.5° should not be used for targets that are smaller than 1.5cm in diameter. The study also concluded that there was increased risk of missing the target with increase in separation distance and a decrease in GTV size. This poses a potential risk for poor local control in C-arm linac based single isocentre MBM SRS compared to GK and CK. Although the C-arm linac treatment plan may appear optimal, treatment delivery might be less accurate for distal metastases. Additionally, C-arm linac MBM SRS techniques tend to use larger PTV margins which increases the risk of radionecrosis. C-arm linacs are more widely available compared to GK and CK, which are dedicated SRS systems. The dosimetric and geometric accuracies discussed are comparable and there is a huge benefit from shorter treatment times and more throughput with C-arm linac based SRS.
Limitations
Most of the treatment plans generated for comparison were “in silico” and no patient specific QA measurements were performed to compare with the plans. Some of the studies did not indicate if PTV margin was the same for different techniques in the same study and hence the same PTV margin per study was assumed. The planners across all the studies may have varied levels of experience and skills, contributing to planner subjectivity in the findings. There is scarcity of reports or studies on overall survival for C-arm linac based MBM SRS as monotherapy. Conducting a randomized trial to investigate overall survival for MBM patients treated with C-arm linac-based single-isocentre SRS compared to GK or CK would be valuable. Some studies have compared patient setup, pre-treatment and during-treatment imaging as well as the actual beam-on time across two or three techniques. However, it would be valuable to conduct studies that include the required time for the whole process from CT simulation to treatment.
Conclusion
A review of SRS treatment techniques over the past 8 years demonstrated that C-arm linac-based SRS is gaining more traction due to improved planning and delivery mechanisms. There is a growing number of studies comparing single-isocentre SRS technique with traditional SRS techniques. These studies are showing comparable conformity indices between GK, CK, and C-arm linac-based SRS techniques. Overall GK and CK showed slightly better gradient index and sparing of normal brain. However, in some cases, HyperArc and DCA have better sparing of normal brain, especially in cases with multiple targets. GK and CK SRS treatments take much longer to deliver compared to C-arm linac-based SRS. Rotational errors are a major source of dosimetric error in single-isocentre SRS techniques and 6DoF geometric corrections are highly recommend. Care should be taken for large PTV to isocentre distance as it increases the healthy brain dose. As centres with C-arm linac based SRS gain confidence with their techniques, there is potential for reducing PTV margins, thereby reducing the risk of radionecrosis. Developing software capable of real time calculation of geometric accuracy real time calculation during MBM SRS delivery would be beneficial. Overall, the maturity of C-arm linac-based SRS has been demonstrated and should be rolled out in departments with capable linac-based facilities.
References
- 1. Blomain E.S., et al., Stereotactic radiosurgery practice patterns for brain metastases in the United States: a national survey. Journal of Radiation Oncology, 2018. 7(3): p. 241–246. pmid:30416676
- 2. Brown P.D., et al., Effect of Radiosurgery Alone vs Radiosurgery With Whole Brain Radiation Therapy on Cognitive Function in Patients With 1 to 3 Brain Metastases A Randomized Clinical Trial. Jama-Journal of the American Medical Association, 2016. 316(4): p. 401–409. pmid:27458945
- 3. Kocher M., et al., Stereotactic radiosurgery for treatment of brain metastases. Strahlentherapie und Onkologie, 2014. 190(6): p. 521–532.
- 4. van der Meer P.B., et al., Individual changes in neurocognitive functioning and health-related quality of life in patients with brain oligometastases treated with stereotactic radiotherapy. J Neurooncol, 2018. 139(2): p. 359–368. pmid:29663172
- 5. Barbour A.B., et al., Radiation Therapy Practice Patterns for Brain Metastases in the United States in the Stereotactic Radiosurgery Era. Advances in Radiation Oncology, 2020. 5(1): p. 43–52. pmid:32051889
- 6. Grishchuk D., et al., ISRS Technical Guidelines for Stereotactic Radiosurgery: Treatment of Small Brain Metastases (≤1 cm in Diameter). Practical Radiation Oncology, 2023. 13(3): p. 183–194.
- 7. Slagowski J.M. and Wen Z., Selection of single-isocenter for multiple-target stereotactic brain radiosurgery to minimize total margin volume. Phys Med Biol, 2020. 65(18): p. 185012.
- 8. Palmer J.D., et al., Single-Isocenter Multitarget Stereotactic Radiosurgery Is Safe and Effective in the Treatment of Multiple Brain Metastases. Advances in Radiation Oncology, 2020. 5(1): p. 70–76. pmid:32051892
- 9. Gondi V., et al., Radiation Therapy for Brain Metastases: An ASTRO Clinical Practice Guideline. Practical Radiation Oncology, 2022. 12(4): p. 265–282. pmid:35534352
- 10. Vergalasova I., et al., Multi-Institutional Dosimetric Evaluation of Modern Day Stereotactic Radiosurgery (SRS) Treatment Options for Multiple Brain Metastases. Frontiers in Oncology, 2019. 9. pmid:31231614
- 11. Soliman H., et al., Stereotactic radiosurgery (SRS) in the modern management of patients with brain metastases. Oncotarget, 2016. 7(11): p. 12318–12330. pmid:26848525
- 12. Chaung K.V., Sloan A.E., and Choi S., Limited brain metastases: a narrative review. Ann Palliat Med, 2021. 10(5): p. 6016–6027. pmid:33977758
- 13. Hartgerink D., et al., LINAC based stereotactic radiosurgery for multiple brain metastases: guidance for clinical implementation. Acta Oncol, 2019. 58(9): p. 1275–1282. pmid:31257960
- 14. Yamamoto M., et al., Stereotactic radiosurgery for patients with multiple brain metastases: a case-matched study comparing treatment results for patients with 2–9 versus 10 or more tumors: Clinical article. Journal of Neurosurgery JNS, 2014. 121(Suppl_2): p. 16–25.
- 15. Kraft J., et al., Stereotactic Radiosurgery for Multiple Brain Metastases. Current Treatment Options in Neurology, 2019. 21(2). pmid:30758726
- 16. Mizuno T., et al., Comparison between stereotactic radiosurgery and whole-brain radiotherapy for 10–20 brain metastases from non-small cell lung cancer. Mol Clin Oncol, 2019. 10(5): p. 560–566. pmid:30967951
- 17. Shuto T., et al., Multiinstitutional prospective observational study of stereotactic radiosurgery for patients with multiple brain metastases from non–small cell lung cancer (JLGK0901 study–NSCLC). Journal of Neurosurgery JNS, 2018. 129(Suppl1): p. 86–94. pmid:30544291
- 18. Zindler J.D., et al., Whole brain radiotherapy versus stereotactic radiosurgery for 4–10 brain metastases: a phase III randomised multicentre trial. BMC Cancer, 2017. 17(1): p. 500. pmid:28743240
- 19. Nichol A., et al., Volumetric Radiosurgery for 1 to 10 Brain Metastases: A Multicenter, Single-Arm, Phase 2 Study. International Journal of Radiation Oncology Biology Physics, 2016. 94(2): p. 312–321. pmid:26678660
- 20. Jang S.Y., et al., Dosimetric comparison between cone/Iris-based and InCise MLC-based CyberKnife plans for single and multiple brain metastases. Journal of Applied Clinical Medical Physics, 2016. 17(5): p. 184–199. pmid:27685124
- 21. Agostinelli S., et al., Dosimetric analysis of Tomotherapy-based intracranial stereotactic radiosurgery of brain metastasis. Physica Medica, 2018. 52: p. 48–55.
- 22. Barra S., et al., Radiosurgery with Helical Tomotherapy:Outcomes for Patients with One or Multifocal Brain Metastasis. Technology in Cancer Research & Treatment, 2015. 14(6): p. 693–699. pmid:24750003
- 23. Srivastava S.P., et al., Machine performance and stability of the first clinical self-shielded stereotactic radiosurgery system: Initial 2-year experience. Journal of Applied Clinical Medical Physics, 2023. 24(3): p. e13857. pmid:36519493
- 24. Eaton D.J., Lee J., and Paddick I., Stereotactic radiosurgery for multiple brain metastases: Results of multicenter benchmark planning studies. Practical Radiation Oncology, 2018. 8(4): p. e212–e220. pmid:29452865
- 25. Chea M., et al., Dosimetric study between a single isocenter dynamic conformal arc therapy technique and Gamma Knife radiosurgery for multiple brain metastases treatment: impact of target volume geometrical characteristics. Radiation Oncology, 2021. 16(1). pmid:33639959
- 26. Kadoya N., et al., Automated noncoplanar treatment planning strategy in stereotactic radiosurgery of multiple cranial metastases: HyperArc and CyberKnife dose distributions. Medical Dosimetry, 2019. 44(4): p. 394–400. pmid:30827765
- 27. Liu H., et al., Plan quality and treatment efficiency for radiosurgery to multiple brain metastases: Non-coplanar rapidarc vs. Gamma Knife. Frontiers in Oncology, 2016. 6(FEB). pmid:26904504
- 28. Zhang S., et al., Noncoplanar VMAT for Brain Metastases: A Plan Quality and Delivery Efficiency Comparison With Coplanar VMAT, IMRT, and CyberKnife. Technol Cancer Res Treat, 2019. 18: p. 1533033819871621. pmid:31451059
- 29. Hendricks B.K., et al., ZAP-X Gyroscopic Radiosurgery System: A Preliminary Analysis of Clinical Applications within a Retrospective Case Series. Stereotactic and Functional Neurosurgery, 2021. 100(2): p. 99–107. pmid:34942629
- 30. Srivastava S.P., et al., Treatment planning system and beam data validation for the ZAP-X: A novel self-shielded stereotactic radiosurgery system. Medical Physics, 2021. 48(5): p. 2494–2510. pmid:33506520
- 31. Pan L., et al., The Zap-X Radiosurgical System in the Treatment of Intracranial Tumors: A Technical Case Report. Neurosurgery, 2021. 88(4): p. E351–E355. pmid:33442727
- 32. Meeks S.L., et al., Practical Considerations for Single Isocenter LINAC Radiosurgery of Multiple Brain Metastases. Practical Radiation Oncology, 2022. 12(3): p. 195–199. pmid:34619373
- 33. Ruggieri R., et al., Linac-based VMAT radiosurgery for multiple brain lesions: Comparison between a conventional multi-isocenter approach and a new dedicated mono-isocenter technique. Radiation Oncology, 2018. 13(1). pmid:29506539
- 34. O’Toole J., et al., Improving efficiency in the radiation management of multiple brain metastases using a knowledge-based planning solution for single-isocentre volumetric modulated arc therapy (VMAT) technique. Journal of Medical Radiation Sciences, 2021. 68(4): p. 364–370. pmid:34310846
- 35. Rozati H., Chen J., and Williams M., Overall survival following stereotactic radiosurgery for ten or more brain metastases: a systematic review and meta-analysis. BMC Cancer, 2023. 23(1): p. 1004. pmid:37858075
- 36. de la Peña C., et al., CyberKnife Stereotactic Radiosurgery in brain metastases: A report from Latin America with literature review. Reports of Practical Oncology & Radiotherapy, 2018. 23(3): p. 161–167. pmid:29760591
- 37. Feng C.H., et al., Automated contouring and planning pipeline for hippocampal-avoidant whole-brain radiotherapy. Radiat Oncol, 2020. 15(1): p. 251. pmid:33126894
- 38. Ferini G., et al., Whole Brain Irradiation or Stereotactic RadioSurgery for five or more brain metastases (WHOBI-STER): A prospective comparative study of neurocognitive outcomes, level of autonomy in daily activities and quality of life. Clinical and Translational Radiation Oncology, 2022. 32: p. 52–58. pmid:34926839
- 39. Greto D., et al., Gamma Knife Radiosurgery in the management of single and multiple brain metastases. Clin Neurol Neurosurg, 2016. 141: p. 43–7. pmid:26731463
- 40. Mengue L., et al., Brain metastases treated with hypofractionated stereotactic radiotherapy: 8 years experience after Cyberknife installation. Radiation Oncology, 2020. 15(1): p. 11.
- 41. Yamamoto M., et al., A case-matched study of stereotactic radiosurgery for patients with multiple brain metastases: comparing treatment results for 1–4 vs ≥ 5 tumors: Clinical article. Journal of Neurosurgery JNS, 2013. 118(6): p. 1258–1268.
- 42. Kalash R., et al., Exceptional Eight-year Response to Stereotactic Radiosurgery Monotherapy for Multiple Brain Metastases. Cureus, 2017. 9(12): p. e2001. pmid:29507849
- 43. Mangesius J., et al., Intrafractional 6D head movement increases with time of mask fixation during stereotactic intracranial RT-sessions. Radiation Oncology, 2019. 14(1): p. 231. pmid:31852497
- 44. Tsuruta Y., et al., Evaluation of intrafractional head motion for intracranial stereotactic radiosurgery with a thermoplastic frameless mask and ceiling-floor-mounted image guidance device. Physica Medica, 2021. 81: p. 245–252. pmid:33485142
- 45. Lewis B.C., et al., Monitoring frequency of intra-fraction patient motion using the ExacTrac system for LINAC-based SRS treatments. J Appl Clin Med Phys, 2018. 19(3): p. 58–63. pmid:29577592
- 46. Tsui S.S.W., Wu V.W.C., and Cheung J.S.C., Comparison of dosimetric impact of intra-fractional setup discrepancy between multiple- and single-isocenter approaches in linac-based stereotactic radiotherapy of multiple brain metastases. Journal of Applied Clinical Medical Physics, 2022. 23(2). pmid:34931447
- 47. Eder M.M., et al., Single-isocenter stereotactic radiosurgery for multiple brain metastases: Impact of patient misalignments on target coverage in non-coplanar treatments. Zeitschrift fur Medizinische Physik, 2022. 32(3): p. 296–311. pmid:35504799
- 48. Prentou G., et al., Dosimetric impact of rotational errors on the quality of VMAT-SRS for multiple brain metastases: Comparison between single- and two-isocenter treatment planning techniques. Journal of Applied Clinical Medical Physics, 2020. 21(3): p. 32–44. pmid:32022447
- 49. Nakano H., et al., Effect of setup error in the single-isocenter technique on stereotactic radiosurgery for multiple brain metastases. Journal of Applied Clinical Medical Physics, 2020. 21(12): p. 155–165. pmid:33119953
- 50. Clark G.M., et al., Feasibility of Single-Isocenter Volumetric Modulated Arc Radiosurgery for Treatment of Multiple Brain Metastases. International Journal of Radiation Oncology*Biology*Physics, 2010. 76(1): p. 296–302. pmid:19836151
- 51. Paddick I., A simple scoring ratio to index the conformity of radiosurgical treatment plans. Technical note. J Neurosurg, 2000. 93 Suppl 3: p. 219–22. pmid:11143252
- 52. El Shafie R.A., et al., Single-Isocenter Volumetric Modulated Arc Therapy vs. CyberKnife M6 for the Stereotactic Radiosurgery of Multiple Brain Metastases. Frontiers in Oncology, 2020. 10. pmid:32457829
- 53. Velten C., et al., Single isocenter treatment planning techniques for stereotactic radiosurgery of multiple cranial metastases. Physics and Imaging in Radiation Oncology, 2021. 17: p. 47–52. pmid:33898778
- 54. Dimitriadis A., and Paddick Ian, A novel index for assessing treatment plan quality in stereotactic radiosurgery. J Neurosurg, 2018. 129: p. 118–124. pmid:30544322
- 55. Wilke L., et al., ICRU report 91 on prescribing, recording, and reporting of stereotactic treatments with small photon beams. Strahlentherapie und Onkologie, 2019. 195(3): p. 193–198.
- 56. Paddick I. and Lippitz B., A simple dose gradient measurement tool to complement the conformity index. Journal of Neurosurgery JNS, 2006. 105(Supplement): p. 194–201. pmid:18503356
- 57. Palmiero A.N., et al., Single-Isocenter Volumetric Modulated Arc Therapy (VMAT) Radiosurgery for Multiple Brain Metastases: Potential Loss of Target(s) Coverage Due to Isocenter Misalignment. Cureus, 2020. 12(10): p. e11267. pmid:33274143
- 58. Moher D., et al., Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement. PLoS Med, 2009. 6(7): p. e1000097. pmid:19621072
- 59. Haddaway N.R., et al., Making literature reviews more reliable through application of lessons from systematic reviews. Conservation Biology, 2015. 29(6): p. 1596–1605. pmid:26032263
- 60. Tricco A.C., et al., Few systematic reviews exist documenting the extent of bias: a systematic review. Journal of clinical epidemiology, 2008. 61(5): p. 422–434. pmid:18394534
- 61. Menon S.V., et al., Evaluation of Plan Quality Metrics in Stereotactic Radiosurgery/Radiotherapy in the Treatment Plans of Arteriovenous Malformations. J Med Phys, 2018. 43(4): p. 214–220. pmid:30636846
- 62. Wu Q.R., et al., Quality of coverage: conformity measures for stereotactic radiosurgery. J Appl Clin Med Phys, 2003. 4(4): p. 374–81. pmid:14604427
- 63. Hofmaier J., et al., Single isocenter stereotactic radiosurgery for patients with multiple brain metastases: Dosimetric comparison of VMAT and a dedicated DCAT planning tool. Radiation Oncology, 2019. 14(1). pmid:31186023
- 64. Türkkan G., et al., Dosimetric Comparison of Volumetric-Modulated Arc Therapy and Dynamic Conformal Arc Therapy in Threefraction Single-isocenter Stereotactic Radiosurgery for Multiple Brain Metastases. Turk Onkoloji Dergisi, 2022. 37(1): p. 67–73.
- 65. Wu Q.X., et al., Optimization of Treatment Geometry to Reduce Normal Brain Dose in Radiosurgery of Multiple Brain Metastases with Single-Isocenter Volumetric Modulated Arc Therapy. Scientific Reports, 2016. 6. pmid:27688047
- 66. de Camargo A.V., et al., Evaluation of the correlation between dosimetric, geometric, and technical parameters of radiosurgery planning for multiple brain metastases. Journal of Applied Clinical Medical Physics, 2021. 22(8): p. 83–92. pmid:34212482
- 67. Ohira S., et al., HyperArc VMAT planning for single and multiple brain metastases stereotactic radiosurgery: A new treatment planning approach. Radiation Oncology, 2018. 13(1). pmid:29378610
- 68. Chang J., Wernicke A.G., and Pannullo S.C., Restricted single isocenter for multiple targets dynamic conformal arc (RSIMT DCA) technique for brain stereotactic radiosurgery (SRS) planning. Journal of Radiosurgery and SBRT, 2018. 5(2): p. 145–156. pmid:29657895
- 69. Hossain S., et al., Normal Brain Sparing With Increasing Number of Beams and Isocenters in Volumetric-Modulated Arc Beam Radiosurgery of Multiple Brain Metastases. Technology in Cancer Research & Treatment, 2016. 15(6): p. 766–771. pmid:26596914
- 70. Laoui S., et al., Flattening filter free beam energy selection and its impact in multitarget intracranial stereotactic radiosurgery treatments. Medical Dosimetry, 2020. 45(4): p. 363–367. pmid:32553505
- 71. Pokhrel D., et al., Dynamic conformal arcs-based single-isocenter VMAT planning technique for radiosurgery of multiple brain metastases. Medical Dosimetry, 2021. 46(2): p. 195–200. pmid:33303353
- 72. Priyadarsini V.H., et al., A Comprehensive Evaluation of Single Isocenter Multiple Target SRS Plans and the Analytical Relationship between Plan Quality Indices with the Number and Volume of Targets. Asian Pacific Journal of Cancer Prevention, 2022. 23(9): p. 3103–3112. pmid:36172673
- 73. Ghemiş D.M. and Marcu L.G., Progress and prospects of flattening filter free beam technology in radiosurgery and stereotactic body radiotherapy. Critical Reviews in Oncology/Hematology, 2021. 163: p. 103396. pmid:34146680
- 74. Lai Y., et al., Dosimetric superiority of flattening filter free beams for single-fraction stereotactic radiosurgery in single brain metastasis. Oncotarget, 2017. 8(21): p. 35272–35279. pmid:27823985
- 75. Rieber J., et al., Radiosurgery with flattening-filter-free techniques in the treatment of brain metastases: Plan comparison and early clinical evaluation. Strahlenther Onkol, 2016. 192(11): p. 789–796.
- 76. Pudsey L., et al., Current status of intra-cranial stereotactic radiotherapy and stereotactic radiosurgery in Australia and New Zealand: key considerations from a workshop and surveys. Physical and Engineering Sciences in Medicine, 2022. 45(1): p. 251–259. pmid:35113342
- 77. Jang Si Young, L. R., Ozhasoglu Cihat, Burton Steven, and H M.S. Heron Dwight, Dosimetric comparison between cone/Iris-based and InCise MLC-based CyberKnife plans for single and multiple brain metastases. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS,, 2016. VOLUME 17(NUMBER 5). pmid:27685124
- 78. Kirkpatrick J.P., et al., Defining the Optimal Planning Target Volume in Image-Guided Stereotactic Radiosurgery of Brain Metastases: Results of a Randomized Trial. International Journal of Radiation Oncology*Biology*Physics, 2015. 91(1): p. 100–108. pmid:25442342
- 79. Karlsson B., et al., Thirty years’ experience with Gamma Knife surgery for metastases to the brain: Clinical article. Journal of Neurosurgery JNS, 2009. 111(3): p. 449–457.
- 80. Grandhi R., et al., Stereotactic radiosurgery using the Leksell Gamma Knife Perfexion unit in the management of patients with 10 or more brain metastases: Clinical article. Journal of Neurosurgery JNS, 2012. 117(2): p. 237–245.
- 81. Hyde C., et al., Fractionated Radiosurgery Alone for Thirty-seven Brain Metastases: Not Everything that can be Counted Counts. Cureus, 2017. 9(12): p. e1985. pmid:29503779
- 82. Milano M.T., et al., Single- and Multifraction Stereotactic Radiosurgery Dose/Volume Tolerances of the Brain. Int J Radiat Oncol Biol Phys, 2021. 110(1): p. 68–86. pmid:32921513
- 83. Satyanarayana Rao K.H., Informed consent: an ethical obligation or legal compulsion? J Cutan Aesthet Surg, 2008. 1(1): p. 33–5. pmid:20300341
- 84. Pietrzykowski T. and Smilowska K., The reality of informed consent: empirical studies on patient comprehension—systematic review. Trials, 2021. 22(1): p. 57. pmid:33446265
- 85. Gaebe K., et al., Stereotactic radiosurgery versus whole brain radiotherapy in patients with intracranial metastatic disease and small-cell lung cancer: a systematic review and meta-analysis. Lancet Oncol, 2022. 23(7): p. 931–939. pmid:35644163
- 86. Slosarek K., et al., In silico assessment of the dosimetric quality of a novel, automated radiation treatment planning strategy for linac-based radiosurgery of multiple brain metastases and a comparison with robotic methods. Radiation Oncology, 2018. 13(1). pmid:29544504
- 87. Zhang S., Yang R., and Wang X., Dosimetric quality and delivery efficiency of robotic radiosurgery for brain metastases: Comparison with C-arm linear accelerator based plans. J Appl Clin Med Phys, 2019. 20(11): p. 104–110. pmid:31580532
- 88. Milano M.T., et al., Executive summary from American Radium Society’s appropriate use criteria on neurocognition after stereotactic radiosurgery for multiple brain metastases. Neuro-Oncology, 2020. 22(12): p. 1728–1741. pmid:32780818
- 89. Mori Y., et al., Dosimetric Study of Automatic Brain Metastases Planning in Comparison with Conventional Multi-Isocenter Dynamic Conformal Arc Therapy and Gamma Knife Radiosurgery for Multiple Brain Metastases. Cureus, 2016. 8(11): p. e882. pmid:28003946
- 90. Yoon J.W., et al., Combined effect of dose gradient and rotational error on prescribed dose coverage for single isocenter multiple brain metastases in frameless stereotactic radiotherapy. Radiation Oncology, 2021. 16(1). pmid:34465331
- 91. Minniti G., et al., Stereotactic radiosurgery for brain metastases: analysis of outcome and risk of brain radionecrosis. Radiat Oncol, 2011. 6: p. 48. pmid:21575163
- 92. Bossart E., et al., Assessment of single isocenter linear accelerator radiosurgery for metastases and base of skull lesions. Phys Med, 2021. 81: p. 1–8. pmid:33278764
- 93. Roper J., et al., Single-Isocenter Multiple-Target Stereotactic Radiosurgery: Risk of Compromised Coverage. Int J Radiat Oncol Biol Phys, 2015. 93(3): p. 540–6. pmid:26460996
- 94. Selvan K.T., et al., Dosimetric Effect of Rotational Setup Errors in Single-Isocenter Volumetric-Modulated Arc Therapy of Multiple Brain Metastases. Journal of Medical Physics, 2019. 44(2): p. 84–90. pmid:31359925