Numerical investigation of LDL nanoparticle collision in coronary artery grafts with porous wall and different implantation angles and two state of inlet velocity

Correction

14 Jun 2024: The PLOS ONE Staff (2024) Correction: Numerical investigation of LDL nanoparticle collision in coronary artery grafts with porous wall and different implantation angles and two state of inlet velocity. PLOS ONE 19(6): e0305829. https://doi.org/10.1371/journal.pone.0305829 View correction

Abstract

This study aimed to reduce the risk of graft occlusion by evaluating the two-phase flow of blood and LDL nanoparticles in coronary artery grafts. The study considered blood as an incompressible Newtonian fluid, with the addition of LDL nanoparticles, and the artery wall as a porous medium. Two scenarios were compared, with constant inlet velocity (CIV) and other with pulsatile inlet velocity (PIV), with LDL nanoparticles experiencing drag, wall-induced lift, and induced Saffman lift forces, or drag force only. The study also evaluated the concentration polarization of LDLs (CP of LDLs) near the walls, by considering the artery wall with and without permeation. To model LDL nanoparticles, the study randomly injected 100, 500, and 1000 nanoparticles in three release states at each time step, using different geometries. Numerical simulations were performed using COMSOL software, and the results were presented as relative collision of nanoparticles to the walls in tables, diagrams, and shear stress contours. The study found that a graft implantation angle of 15° had the most desirable conditions compared to larger angles, in terms of nanoparticle collision with surfaces and occlusion. The nanoparticle release modes behaved similarly in terms of collision with the surfaces. A difference was observed between CIV and PIV. Saffman lift and wall-induced lift forces having no effect, possibly due to the assumption of a porous artery wall and perpendicular outlet flow. In case of permeable artery walls, relative collision of particles with the graft wall was larger, suggesting the effect of CP of LDLs.

1.Introduction

In 2015, coronary artery disease was diagnosed in 110 million people worldwide, resulting in 8.9 million deaths, making it the leading cause of death, accounting for 15.6% of all annual deaths [1]. The conventional treatments for coronary artery disease include percutaneous coronary intervention (PCI) or coronary artery bypass graft (CABG). However, it remains uncertain whether these treatments can effectively reduce the risk of heart attacks, extend patients’ lives, or improve their condition [2]. CABG involves surgically restoring blood flow to a blocked coronary artery in cases with 50–90% stenosis, a condition characterized by the thickening and inflexibility of the artery wall, leading to narrowed blood flow. Stenosis occurs due to the accumulation of cholesterol, lipids, and cellular residuals on the coronary artery wall, which obstruct normal blood flow [3]. Data suggest that within one year of coronary artery bypass grafting, 10–15% of vein grafts experience stenosis, and after 6 to 12 years, more than 70% of vein grafts become occluded [4].

Stenosis is linked to the concentration of lipids near the artery wall. This hypothesis suggests that a higher concentration of lipids at the blood/artery boundary can affect the artery wall’s permeability [1]. Fan et al. investigated the impact of the concentration polarization of LDLs (CP of LDLs) and flow hemodynamics on grafts, examining implantation angles of 30, 45, and 60° and S-type geometry. Their findings revealed that the best flow hemodynamics occurred at a graft implantation angle of 30°, with an increase in angle leading to worsened hemodynamic parameters. Additionally, the CP of LDLs was at its optimum at the 30° implantation angle [2] which is considered the lowest implantation angle. Using numerical simulations, Kabinejadian et al. explored how the graft’s shape can affect the flow hemodynamics and stenosis. Their results showed that the rotational flow formation can increase shear stress on the wall, shorten the remaining time of the particle in the anastomotic region and coronary artery, and promote the duration of artery openness [3]. Fallahi et al. analyzed the hemodynamics of bypass grafts implanted on coronary arteries with elastic walls and varying stenosis levels. They considered blood as a Newtonian fluid with unsteady flow and found that pulsatile flow produced more accurate data. Moreover, the velocity field and shear stress on the wall were lower compared to steady-state flow [4].

Vishesh Kashyap and colleagues studied the curves in coronary arteries. They measured the bend angle, ranging from 0 to 60 degrees. Their findings highlighted that the arteries’ curves impact how blood flows. Generally, higher curvature angles tend to increase the risk of blockages [5].

S. Kenjereš, J.P. van der Krieke, and C. Li conducted a study investigating blood flow and the transport of low-density lipoprotein (LDL) mass in an axisymmetric geometry resembling a stenosis in a diseased coronary artery. Their analysis involved considering the arterial wall as a porous medium comprising multi-layered structures of varying thickness. The primary objective was to develop a mathematical model based on an anatomically realistic representation of the internal structure of the arterial wall, aiming to establish a robust understanding of LDL transport through endothelial cells [6].

Md Foysal Rabbi et al. studied the impact of geometric variations on intravascular and near-wall hemodynamics. Using four patient-specific geometries—featuring branches of 70°, 95°, and 135°, alongside three 90° branches—their findings highlighted the significant influence of geometry, particularly branch angles, on left coronary artery hemodynamics and the risk of occlusion. Notably, they observed substantial effects on near-wall hemodynamics due to varying branch angles [7].

Giuseppe De Nisco et al. explored incorporating blood rheological properties into computational fluid dynamics (CFD) simulations of coronary hemodynamics. They analyzed the right coronary artery (RCA) geometry in 144 patients with varying stenosis levels, considering two models: (i) a Newtonian and (ii) a non-Newtonian blood rheology model. Their findings indicate that blood rheology minimally impacts wall shear stress (WSS) and helical flow profiles. Hence, the use of a Newtonian assumption for blood rheology seems appropriate in coronary hemodynamic simulations [8].

This research aims to identify high-risk areas for re-occlusion in coronary artery grafts implanted at different angles. The study was conducted using two conditions: constant and pulsatile inlet velocities. Permeable and non-permeable walls were used to evaluate fluid outlet and the CP of LDLs. To investigate the impact of nanoparticle numbers on their collision with the graft wall, three release modes were evaluated. Nanoparticles were studied under two force application conditions. In section 2, the problem definition is followed by the presentation of geometry and equations. The grid study and validation of the results are presented in section 3. In section 4 the results are presented and analyzed. Finally, the conclusion is written in section 5.

2. Problem description

Solid spherical LDL nanoparticles were introduced randomly into the flow and analyzed under two conditions: a) considering only the drag force, and b) accounting for the drag, wall-induced lift, and Saffman lift forces. The arterial wall, modeled as a porous medium, was studied in both permeable and non-permeable states to investigate concentration polarization of LDLs (CP of LDLs) near the walls. In the permeable wall scenario, a vertical outflow was simulated at the external vessel wall. The study focused on analyzing two distinct flow regimes: constant and pulsatile. Due to the nanoparticles’ diluted concentration and very low load factor, the impact of the discrete phase on the continuum phase was disregarded [9]. The problem was addressed using a one-way method, initially solving the flow analysis equations and subsequently employing the obtained results to solve the equations concerning nanoparticles.

2.1. Geometry, governing equations and boundary conditions

Fig 1A illustrates the geometry of the problem, with the variable parameter being the θ angle (i.e., the angle of the implanted graft in relation to the main vessel). The θ angles selected were 15, 22.5, 30, 37.5, and 45°. The total vessel length was 100 mm, and both the main artery and graft diameter were 4 mm [10]. The main channel had a diameter of 3.3 mm [11], and the artery wall thickness (porous medium) was 0.35 mm [12]. COMSOL Multiphysics, utilizing the finite element method, was employed for simulating the system dynamics. The simulations were conducted over three cardiac cycles.

thumbnail
Fig 1.

A) geometry B) Pulsatile velocity applied to the inlet [13].

https://doi.org/10.1371/journal.pone.0300326.g001

This study focuses on a single-phase, incompressible fluid with Newtonian[14,15], properties similar to blood, with density of 1060 kg.m-3 and dynamic viscosity of 0.004 Pa.s [15,16]. The continuity and Navier-Stocks equations are expressed as [17]: (1) (2)

Here V, ρ, p, and μ respectively are the velocity, density, pressure, and viscosity of the fluid. The wall was assumed to be inflexible and had no-slip boundary conditions [18]. Two inlet velocity states were used: a constant inlet velocity (CIV) of 25.45 cm. s−1, and a pulsatile inlet velocity (PIV) shown in Fig 1B [13]. The CIV problem was simulated as a steady-state case, while the PIV problem was simulated for three cycles.

The porous medium was simulated by Brinkman equation using Eq (3).

(3)(4)(5)(6)

Porosity (ϵ) and permeability (k) values of the porous medium are provided [19]:

Wilnes et al. measured an outlet velocity from the artery wall of approximately 10−8 m. s−1. In the with-permeation case, the outer surface of the porous medium was assumed to have a normal outlet velocity of 10−8 m. s−1 [2,20,21]. LDL particles were modeled as spherical solid nanoparticles with a mean diameter of 23 nm and a density of 1040 kg.m-3 [22]. Stick boundary conditions were applied to the walls. The impact of the nanoparticle population at the inlet was assessed by considering different numbers of nanoparticles (3000, 15000, and 30000) and injecting 100, 500, and 1000 nanoparticles randomly at the inlet in each 0.1-second time step. The solution time was three cycles, and the governing equation was the second law of Newton (Eq 7).

(7)

Here, mp represents nanoparticle mass, while and indicate velocity and superimposed forces fields, respectively. The forces driving nanoparticle motion in fluid consist of drag, Saffman inductive force, and wall-induced lift [23]. The drag force is expressed as follows: (8)

Here, mp represents nanoparticle mass; () indicates the velocity disparity between the nanoparticle and fluid. And (9)

Here, ρ_p, dp, and μ denote the density, diameter, and viscosity of nanoparticles [24], respectively. The Saffman inductive force can be mathematically represented as [25]: (10)

Where, (11)

Eq 12 represents the wall-induced lift force in the presence of a particle in fluid flow.

(12)

In which, s shows the normalized distance to the first parallel wall. G is the location function of each particle at any time; represents the unit vector from the nearest point in the first parallel wall; D stands for the channel diameter; while β and γ can be respectively determined by: (13) (14)

Here u indicates the velocity component parallel to the surface [25].

(15)

3. Grid study and validation

3.1. Grid study

This study utilized a type-O grid structure with a finer grid near the wall regions for improved accuracy. Five different grids were used in the grid study for each graft implantation angle to determine variations in wall shear stress along the peripheral direction of A cross-section, as shown in Fig 1A. Table 1 lists the maximum difference in shear stress between each grid and the finest grid (Δτmax).

thumbnail
Table 1. Maximum shear stress variation for different graft implantation angles and node numbers.

https://doi.org/10.1371/journal.pone.0300326.t001

Table 1 displays Δτmax for different graft implantation angles and node numbers. Based on this table, 80000 nodes were chosen for each implantation angle based on the identified errors.

3.2. Validation

To validate the results, Sun et al.’s reports were used [26]. Their simulation considered the deposition of LDL nanoparticles in a laminar flow passing through an artery, including drag, Brownian, gravity, Buoyancy, and Van der Waals forces. LDL was simulated as a solid nanoparticle with a density of 1060 kg.m-3, and blood was simulated as a single-phase incompressible fluid with a density of 1030 kg.m-3 and dynamic viscosity of 0.0035 Pa.s. Results included a plot of the number of deposited 20 nm diameter nanoparticles versus inlet velocity, ranging from 0.0005 m.s-1 to 0.5 m.s-1, with 100,000 nanoparticles included. Fig 2 displays the number of nanoparticles trapped in upper and lower walls for various inlet velocities. The comparison between the results of the paper chosen for validation and the simulations in the present study indicates differences of less than 5% in the cases with the maximum discrepancies.

thumbnail
Fig 2. Trapped particle count versus entrance velocities, upper, and lower walls.

https://doi.org/10.1371/journal.pone.0300326.g002

4. Results

The study aims to compare LDL nanoparticles colliding with graft walls at different implantation angles to identify the areas with the highest collision probability. The objective is to determine the ideal geometry by injecting 100–1000 nanoparticles randomly every 0.1-second at the inlet. A dimensionless parameter is used to determine collision distribution among surfaces. Surfaces exhibiting higher collision rates are prone to developing stenosis. In this study, specific surfaces were defined, as illustrated in Fig 3. The study considers two conditions: with/without wall permeation, various force applications, and two inlet velocity modes. The study also explores the wall shear stress, which plays a crucial role in stenosis.

4.1. Surfaces and parameter

Fig 3 shows six bent semi-cylindrical surfaces in the internal boundary of the porous medium. Surfaces 1 and 2 represent the graft inlet, surfaces 3 and 4 represent the middle parts of the graft, and surfaces 5 and 6 represent the graft outlet. The study was conducted in three-dimensional conditions. Results are presented using the dimensionless parameter NCP.

NCP value may vary slightly in simulations due to random nanoparticle injection.

4.2. Comparison of results for permeation and non-permeation from the wall using two different modes of force application

Table 2 compares the effects of wall permeation on nanoparticle collision behavior under two different force application modes. For 500 nanoparticles released in each time step, the non-permeation state exhibited lower values of Normalized Collided Particles (NCP) compared to the permeation state. This was due to the normal flow strengthening the boundary between the fluid and porous medium, directing more nanoparticles towards the walls in the non-permeation state. Both states demonstrated a similar trend in nanoparticle collision behavior with the surfaces. The comparison highlighted the significance of Concentration Polarization of LDLs (CP of LDLs) near the walls. This polarization increases the number of nanoparticles colliding with the surfaces, consequently heightening the risk of stenosis onset and intensification.

thumbnail
Table 2. Effect of implantation angle and surface type on NCP of Grafts.

https://doi.org/10.1371/journal.pone.0300326.t002

Table 2 also includes the NCP values for different graft implantation angles under two modes of force application, where 1000 nanoparticles were released in each time step. The non-permeation state showed a negligible difference in NCP in some cases, due to the normal output flow between the fluid-porous medium boundary, which reduced the impact of wall-induced lift and Saffman forces. In most cases, however, the results were the same. In the permeation state, stronger normal velocity in the fluid-porous medium boundary resulted in completely coinciding results, indicating no impact of Saffman and wall-induced lift forces. The simulations confirmed the results obtained for all cases, and Table 2 is presented as an example of the findings.

4.3. Results based on NCP

Based on Table 3 and Fig 4, an increase in the graft implantation angle resulted in a higher number of nanoparticles colliding with the surface, which increases the risk of stenosis. Surface No. 5 had the highest nanoparticle collision and risk of stenosis, followed by Surface No. 2. In geometries with implanted graft angles of 37.5 and 45°, nanoparticles collided with Surface No. 1, indicating poor performance. The results showed a significant difference between the two inlet velocity conditions, indicating that the effect of pulsatile flow cannot be neglected. Although the overall trend was similar for the number of released nanoparticles, the number of colliding particles varied depending on the release conditions, with different NCP values.

thumbnail
Fig 4. NCP of various surfaces according to the graft’s implantation angle and various number of injection of nanoparticles at the inlet at each time step with permeation from the wall and the application of all forces.

A) CIV and injection of 100 nanoparticle in each step time B) PIV and injection of 100 nanoparticle in each step time C) CIV and injection of 500 nanoparticle in each step time D) PIV and injection of 500 nanoparticle in each step time E) CIV and injection of 1000 nanoparticle in each step time F) PIV and injection of 1000 nanoparticle in each step time.

https://doi.org/10.1371/journal.pone.0300326.g004

thumbnail
Table 3. NCP various surfaces for grafts with different implantation angles with permeation from wall and the application of all forces.

https://doi.org/10.1371/journal.pone.0300326.t003

Table 4 and Fig 5 display the total number of NCPs values across various surfaces in all geometries, indicating that Surface No. 5 experienced the highest overall collisions, followed by Surface No. 1. Specifically, at implantation angles of 37.5° and 45°, Surface No. 1 exhibited relatively high nanoparticle collisions, with Surface No. 2 ranking second overall after Surface No. 5. The study’s findings indicate that the graft’s outlet and inlet registered the highest number of colliding nanoparticles and showed the highest number of points with a heightened risk of stenosis. Furthermore, the results underscore the differences observed in various modes of nanoparticle release.

thumbnail
Fig 5. The total number of NCPs according to various surfaces and geometries with permeation from the wall and with two states of CIV and PIV, with the injection of 100, 500, and 1000 nanoparticles at the inlet at each time step and applying all the forces.

a) CIV on various surfaces b) PIV on various surfaces c) CIV on various geometries d) PIV on various geometries.

https://doi.org/10.1371/journal.pone.0300326.g005

thumbnail
Table 4. The total number of NCPs according various surfaces and geometries.

https://doi.org/10.1371/journal.pone.0300326.t004

Table 4 and Fig 5 also provide information on the total number of NCPs in various geometries. According to the presented data, the graft implanted at an angle of 15° demonstrated the longest lifespan, as it exhibited the lowest number of trapped nanoparticles. A greater number of nanoparticles colliding with the surface increases the risk of stenosis, whereas a smaller number of colliding nanoparticles prolongs the duration of stenosis development. As the graft implantation angle increases, the lifetime of the graft decreases, which is apparent in the different number of released nanoparticles and the applied inlet velocity (constant and pulsatile). These findings also suggest differences between the CIV and PIV regimes.

4.4. Wall shear stress

In Fig 6A, the wall shear stress (WSS) is compared across various geometries. The higher the WSS, the fewer nanoparticles collide with the wall, resulting in less particle build-up due to the stronger shear flow near the wall. As shown in Fig 6A, decreasing the graft implantation angle increases the number of high shear stress surfaces along the graft. Additionally, the WSS on different surfaces decreases as the implanted graft angle increases. Therefore, the graft implanted at a 15° angle exhibited better performance in terms of nanoparticles colliding with surfaces and subsequent stenosis.

thumbnail
Fig 6.

A) WSS in grafts with versus implantation angles B) Axial velocity contour and secondary flow velocity vectors occurred in section A (see Fig 3) in graft with 30° implantation angle.

https://doi.org/10.1371/journal.pone.0300326.g006

The study focused on flow behavior in cross-section A (see Fig 3), with Fig 6B presenting flow parameters for a graft implantation angle of 30°, which also applied to other geometries. A secondary flow arose in the channel’s center due to its curvature and the highest longitudinal velocity being in the cross-section. This caused a secondary rotational velocity field across the channel. Investigating the impact of these secondary velocity fields on nanoparticle movement could be a potential area for future research.

5.Conclusion

This study found that surfaces with more colliding nanoparticles are more likely to develop stenosis. Using one-way simulation of two-phase flow (blood and LDL nanoparticles), smaller implantation angles for grafts resulted in fewer nanoparticles colliding with the graft wall. The graft implanted at 15° showed the best performance, followed by the 22.5° angle. The graft inlet and outlet had the highest collision rate and therefore a higher probability of stenosis. Comparing permeation and non-permeation of walls, a permeable wall increased the NCP, indicating the effect CP of LDLs on nanoparticle collision and stenosis. The Saffman induced lift and wall-induced lift forces had no effect and can be ignored. No difference was found in the collision site of nanoparticles with different injection modes, but NCP values showed dependence on the number of released nanoparticles. Constant and pulsatile inlet velocity modes showed differences in the number of trapped nanoparticles on surfaces. High-stress surfaces decreased with larger implantation angles, resulting in more nanoparticles colliding with the surfaces and intensifying stenosis.

The comparison of simulated geometries indicates that the graft implantation angles of 37.5° and 45° have higher numbers of trapped particles due to their additional surfaces area. Among the graft implantation angles of 15°, 22.5°, and 30°, the 15° angle is preferred as it results in a lower number of colliding nanoparticles, leading to a decrease in stenosis and a longer graft lifetime. Generally, a smaller implantation angle, shorter graft length, and lower curvature decrease the risk of re-occlusion and graft failure. It’s worth noting that only LDL nanoparticles were considered in this study, and the accuracy of the results could be improved by considering other blood components such as plasma, white and red blood cells and etc.

References

  1. 1. Deng X, Marois Y, How T, Merhi Y, King M, Guidoin R. Luminal surface concentration of lipoprotein (LDL) and its effect on the wall uptake of cholesterol by canine carotid arteries. Journal of vascular surgery. 1995;21(1):135–45. pmid:7823352
  2. 2. Fan Z, Sun A, Liu X, Fan Y, Zhao P, Deng X. The accelerated atherogenesis of venous grafts might be attributed to aggravated concentration polarization of low density lipoproteins: a numerical study. Journal of Biomechanics. 2013;46(14):2388–93. pmid:23972425
  3. 3. Kabinejadian F, McElroy M, Ruiz-Soler A, Leo HL, Slevin MA, Badimon L, et al. Numerical assessment of novel helical/spiral grafts with improved hemodynamics for distal graft anastomoses. PLoS One. 2016;11(11):e0165892. pmid:27861485
  4. 4. Fallahi H, Shirani E, Zohravi E. Hemodynamic analysis of coronary artery bypass grafting with elastic walls and different stenoses. Scientia Iranica Transaction B, Mechanical Engineering. 2021;28(2):773–84.
  5. 5. Kashyap V, Arora B, Bhattacharjee S. A computational study of branch-wise curvature in idealized coronary artery bifurcations. Applications in Engineering Science. 2020;4:100027.
  6. 6. Kenjereš S, Van Der Krieke J, Li C. Endothelium resolving simulations of wall shear-stress dependent mass transfer of LDL in diseased coronary arteries. Computers in Biology and Medicine. 2019;114:103453. pmid:31561097
  7. 7. Rabbi MF, Laboni FS, Arafat MT. Computational analysis of the coronary artery hemodynamics with different anatomical variations. Informatics in Medicine Unlocked. 2020;19:100314.
  8. 8. De Nisco G, Rizzini ML, Verardi R, Chiastra C, Candreva A, De Ferrari G, et al. Modelling blood flow in coronary arteries: Newtonian or shear-thinning non-Newtonian rheology? Computer Methods and Programs in Biomedicine. 2023;242:107823. pmid:37757568
  9. 9. Crowe CT, Schwarzkopf JD, Sommerfeld M, Tsuji Y. Multiphase flows with droplets and particles: CRC press; 2011.
  10. 10. Shalman E, Rosenfeld M, Dgany E, Einav S. Numerical modeling of the flow in stenosed coronary artery. The relationship between main hemodynamic parameters. Computers in biology and medicine. 2002;32(5):329–44. pmid:12102752
  11. 11. Murphy JB, Boyle FJ. A full-range, multi-variable, CFD-based methodology to identify abnormal near-wall hemodynamics in a stented coronary artery. Biorheology. 2010;47(2):117–32. pmid:20683155
  12. 12. Wang Z, Liu X, Kang H, Sun A, Fan Y, Deng X. Enhanced accumulation of LDLs within the venous graft wall induced by elevated filtration rate may account for its accelerated atherogenesis. Atherosclerosis. 2014;236(1):198–206. pmid:25075936
  13. 13. Jahangiri M, Saghafian M, Sadeghi MR. Numerical simulation of non-Newtonian models effect on hemodynamic factors of pulsatile blood flow in elastic stenosed artery. Journal of Mechanical Science and Technology. 2017;31:1003–13.
  14. 14. Carvalho V, Pinho D, Lima RA, Teixeira JC, Teixeira S. Blood flow modeling in coronary arteries: A review. Fluids. 2021;6(2):53.
  15. 15. Pandey R, Kumar M, Majdoubi J, Rahimi-Gorji M, Srivastav VK. A review study on blood in human coronary artery: Numerical approach. Computer methods and programs in biomedicine. 2020;187:105243. pmid:31805457
  16. 16. Hutchison JS, Ward RE, Lacroix J, Hébert PC, Barnes MA, Bohn DJ, et al. Hypothermia therapy after traumatic brain injury in children. New England Journal of Medicine. 2008;358(23):2447–56. pmid:18525042
  17. 17. White FM, Majdalani J. Viscous fluid flow: McGraw-Hill New York; 2006.
  18. 18. Diogo B. Computational analysis of the flow in the left coronary arteries: influence of the artery elasticity and the stent presence in blood flow: Thesis, Tecnico Lisboa, Portugal; 2017.
  19. 19. Govindaraju K, Kamangar S, Badruddin IA, Viswanathan GN, Badarudin A, Ahmed NS. Effect of porous media of the stenosed artery wall to the coronary physiological diagnostic parameter: a computational fluid dynamic analysis. Atherosclerosis. 2014;233(2):630–5. pmid:24549189
  20. 20. Wilens S, McCluskey R. The comparative filtration properties of excised arteries and veins. The American journal of the medical sciences. 1952;224(5):540–7. pmid:12985574
  21. 21. Wada S, Karino T. Theoretical prediction of low-density lipoproteins concentration at the luminal surface of an artery with a multiple bend. Annals of biomedical engineering. 2002;30:778–91. pmid:12220078
  22. 22. Krauss RM, Burke DJ. Identification of multiple subclasses of plasma low density lipoproteins in normal humans. Journal of lipid research. 1982;23(1):97–104. pmid:7057116
  23. 23. Crowe C, Sommerfeld M, Tsuji Y. Multiphase flows with: Ž; 1998.
  24. 24. Bagchi P, Balachandar S. Effect of turbulence on the drag and lift of a particle. Physics of fluids. 2003;15(11):3496–513.
  25. 25. Saffman PG. The lift on a small sphere in a slow shear flow. Journal of fluid mechanics. 1965;22(2):385–400.
  26. 26. Sun L, Lin J-z, Bao F-b. Numerical simulation on the deposition of nanoparticles under laminar conditions. Journal of Hydrodynamics. 2006;18(6):676–80.