Enhancing malaria detection in resource-limited areas: A high-performance colorimetric LAMP assay for Plasmodium falciparum screening

  • Tuyet Kha Nguyen,

    Roles Data curation, Formal analysis, Investigation, Methodology, Writing – original draft

    Affiliation Department of Medical Environmental Biology and Tropical Medicine, School of Medicine, Kangwon National University, Chuncheon, Gangwon-do, Republic of Korea

  • Hojong Jun,

    Roles Data curation, Formal analysis, Methodology

    Affiliation Department of Medical Environmental Biology and Tropical Medicine, School of Medicine, Kangwon National University, Chuncheon, Gangwon-do, Republic of Korea

  • Johnsy Mary Louis,

    Roles Formal analysis, Validation, Writing – review & editing

    Affiliation Department of Medical Environmental Biology and Tropical Medicine, School of Medicine, Kangwon National University, Chuncheon, Gangwon-do, Republic of Korea

  • Ernest Mazigo,

    Roles Formal analysis, Validation, Writing – review & editing

    Affiliation Department of Medical Environmental Biology and Tropical Medicine, School of Medicine, Kangwon National University, Chuncheon, Gangwon-do, Republic of Korea

  • Wang-Jong Lee,

    Roles Formal analysis, Validation

    Affiliation Department of Medical Environmental Biology and Tropical Medicine, School of Medicine, Kangwon National University, Chuncheon, Gangwon-do, Republic of Korea

  • Hyun Cher Youm,

    Roles Resources, Validation

    Affiliation Noul Co., Ltd., Yongin, Gyeonggi-do, Republic of Korea

  • Jieun Shin,

    Roles Resources, Validation

    Affiliation Noul Co., Ltd., Yongin, Gyeonggi-do, Republic of Korea

  • Douglas K. Lungu,

    Roles Investigation, Resources, Validation, Visualization

    Affiliation Wezi Medical Centre, Mzuzu, Malawi

  • Creto Kanyemba,

    Roles Investigation, Resources, Validation, Visualization

    Affiliation Wezi Medical Centre, Mzuzu, Malawi

  • Md Atique Ahmed,

    Roles Formal analysis, Resources, Validation, Visualization, Writing – review & editing

    Affiliation ICMR-Regional Medical Research Centre, NER, Dibrugarh, Assam, India

  • Fauzi Muh,

    Roles Resources, Validation, Visualization, Writing – review & editing

    Affiliation Faculty of Public Health, Department of Epidemiology and Tropical Diseases, Universitas Diponegoro, Semarang, Indonesia

  • Se Jin Lee,

    Roles Formal analysis, Investigation, Validation

    Affiliation Department of Obstetrics and Gynecology, Kangwon National University Hospital, Chuncheon, Gangwon-do, Republic of Korea

  • Sunghun Na,

    Roles Investigation, Resources, Validation

    Affiliation Department of Obstetrics and Gynecology, Kangwon National University Hospital, Chuncheon, Gangwon-do, Republic of Korea

  • Wanjoo Chun,

    Roles Data curation, Validation, Visualization

    Affiliation Department of Pharmacology, School of Medicine, Kangwon National University, Chuncheon, Gangwon-do, Republic of Korea

  • Won Sun Park,

    Roles Data curation, Validation, Visualization

    Affiliation Department of Physiology, School of Medicine, Kangwon National University, Chuncheon, Gangwon-do, Republic of Korea

  • Joo Hwan No,

    Roles Data curation, Resources, Validation, Visualization

    Affiliation Host-Parasite Research Laboratory, Institut Pasteur Korea, Seongnam, Gyeonggi-do, Republic of Korea

  • Min-Jae Kim,

    Roles Resources, Validation, Visualization

    Affiliation Department of Infectious Diseases, Asan Medical Center, University of Ulsan College of Medicine, Seoul, Republic of Korea

  • Eun-Taek Han,

    Roles Conceptualization, Data curation, Validation, Visualization

    Affiliation Department of Medical Environmental Biology and Tropical Medicine, School of Medicine, Kangwon National University, Chuncheon, Gangwon-do, Republic of Korea

  •  [ ... ],
  • Jin-Hee Han

    Roles Conceptualization, Funding acquisition, Project administration, Visualization, Writing – review & editing

    han.han@kangwon.ac.kr

    Affiliation Department of Medical Environmental Biology and Tropical Medicine, School of Medicine, Kangwon National University, Chuncheon, Gangwon-do, Republic of Korea

  • [ view all ]
  • [ view less ]

Abstract

Malaria eradication efforts in resource-limited areas require a rapid, economical, and accurate tool for detecting of the low parasitemia. The malaria rapid diagnostic test (mRDT) is the most suitable for on-site detection of the deadliest form of malaria, Plasmodium falciparum. However, the deletions of histidine rich protein 2 and 3 genes are known to compromise the effectiveness of mRDT. One of the approaches that have been explored intensively for on-site diagnostics is the loop-mediated isothermal amplification (LAMP). LAMP is a one-step amplification that allows the detection of Plasmodium species in less than an hour. Thus, this study aims to present a new primer set to enhance the performance of a colorimetric LAMP (cLAMP) for field application. The primer binding regions were selected within the A-type of P. falciparum 18S rRNA genes, which presents a dual gene locus in the genome. The test result of the newly designed primer indicates that the optimal reaction condition for cLAMP was 30 minutes incubation at 65°C, a shorter incubation time compared to previous LAMP detection methods that typically takes 45 to 60 minutes. The limit of detection (LoD) for the cLAMP using our designed primers and laboratory-grown P. falciparum (3D7) was estimated to be 0.21 parasites/μL which was 1,000-fold higher than referencing primers. Under optimal reaction condition, the new primer sets showed the sensitivity (100%, 95% CI: 80.49–100%) and specificity (100%, 95% CI: 94.64–100%) with 100% (95% CI: 95.70–100%) accuracy on the detection of dried blood spots from Malawi (n = 84). Briefly, the newly designed primer set for P. falciparum detection exhibited high sensitivity and specificity compared to referenced primers. One great advantage of this tool is its ability to be detected by the naked eye, enhancing field approaches. Thus, this tool has the potential to be effective for accurate early parasite detection in resource-limited endemic areas.

Introduction

Malaria is a life-threatening infectious disease caused by Plasmodium species parasites and transmitted through the bites of infected female Anopheles mosquitoes. Early diagnosis is the most critical factor for malaria control and elimination as it actively interrupts transmission and prompts early treatment when patients have relatively mild symptoms. In 2021, malaria incidence was estimated at 247 million cases, with 619,000 malarial deaths in 84 endemic countries [1]. Moreover, over the past two years, the COVID-19 pandemic has hampered malaria control programs, leading to 13.4 million additional malaria cases [1]. Consequently, there is a growing concern for consistent malaria control and early diagnosis.

Since 2010, the World Health Organization (WHO) have been recommending microscopy and rapid diagnostic tests (RDTs) for diagnosis of malaria especially in resource-limited areas to ensure prompt diagnosis and treatment [2]. However, both microscopy and RDTs are prone to error in the reading of final result and exhibit low sensitivity at the low parasitemia compared to molecular-based detection tools [3,4]. Depending on the skills of microscopists, the quality of the reagents, and monitoring systems, microscopy detection has an estimated limit of detection (LoD) exceeding 50–500 parasites/μL in patient whole blood. [3,5]. Although RDTs demonstrate sensitivity around ~100 parasites/μL and high specificity by capturing parasite-specific antigens such as histidine-rich protein 2 (HRP2) for P. falciparum, the long half-life of HRP2 in patient blood circulation after parasite clearance can lead to false positive results [68]. Additionally, the sensitivity of the RDTs is reduced by pfhrp2 and pfhrp3 gene deletions, which leads to false negatives. A meta-analysis has highlighted a global pooled prevalence of 21.3% for pfhrp2 gene deletions and 34.5% for pfhrp3 deletions [9]. Moreover, the false negatives resulting from pfhrp2 and pfhrp3 deletions has been reported as 41.1% globally, predominantly observed in South America and subsequently in African regions [9]. To overcome this limitation, several molecular detection methods have been developed for diagnosis of malaria, including nested polymerase chain reaction (PCR), quantitative PCR, PCR-restriction fragment length polymorphism, multiplex PCR, and droplet digital PCR (ddPCR) [1012]. At the present, PCR is the most sensitive method available, however, its high cost, time-consuming, and a requirement of a well-equipped laboratory limit the field application [13,14]. Thus, novel isothermal amplification techniques such as the isothermal recombinase polymerase amplification (RPA) and loop-mediated isothermal amplification (LAMP) has been intensively developed [15,16].

The LAMP principle hinges on the action of Bacillus stearothermophilus (Bst) DNA polymerase, enabling isothermal amplification through the processes of auto-cycling strand displacement and self-sustained sequence replication. This polymerase is activated at 60°C to 65°C and produces a large quantity of target-specific amplicons within 30 to 60 minutes. The set of four specific primers recognizes six different regions within the target gene sequence, resulting in high specificity and amplify 1010 replications of stem-loop structures [17]. Although conventional LAMP offers certain advantages, including ease of application in the field and higher sensitivity compared to PCR, it encounters limitations when incorporating amplicon indicators [1820]. For instance, the use of polyethyleneimine has been found to hinder amplification efficiency, and ethidium bromide has been associated with potential mutagenic, carcinogenic, or teratogenic risks. Moreover, the process of introducing detection reagents or dyes poses a risk of cross-contamination. Therefore, colorimetric LAMP (cLAMP) has been developed as a promising molecular method for detecting various infectious diseases [2123]. Numerous LAMP assays have been employed for the diagnosis of P. falciparum, targeting conserved genes such as pfhrp2 and mitochondrial DNA [2427]. However, previous studies had detection limits, ranging from 1 to 5 parasites/μL, with incubation times of 40 to 60 minutes [2729]. To enhance the sensitivity and specificity for the early detection of P. falciparum, we developed and demonstrated a newly designed primer set for cLAMP targeting the A-type of 18S rRNA gene located at chromosome 5 and 7 [30]. Then, we applied PCR-based diagnosis methods and real-time cLAMP to evaluate the sensitivity and specificity of the new primer set. The sensitivity and specificity of the designed primer set were compared with the referenced primer set described elsewhere and tested on the extracted clinical samples from Malawi [15].

Materials and methods

Colorimetric LAMP (cLAMP) primers design

Human malaria species 18S rRNA genes were searched on NCBI (https://www.ncbi.nlm.nih.gov/) and PlasmoDB (https://plasmodb.org/plasmo/app) for the design of P. falciparum specific primers. The obtained sequences were aligned using the CLUSTAL-W algorithm of MegAlign software (DNASTAR, Madison, WI) (S1 Fig). The phylogenetic tree was generated using the maximum likelihood method with 1,000 bootstrap pseudo-replicates for accuracy assessment in MEGA11 software [31].

The initial set of four specific primers (F3/B3c and FIP (F1c-F2)/BIP (B1-B2c)) with six binding regions, along with two loop primers (LpB/LpF), were designed using Primer Explorer v5 (Eiken Chemical, Tokyo, Japan) (http://primerexplorer.jp/lampv5e/index.html). The final set of cLAMP primers was manually modified based on the criteria for designing LAMP primers (https://loopamp.eiken.co.jp/en/) and was subsequently verified for specificity using primer-BLAST on NCBI (Table 1 and Fig 1C).

thumbnail
Fig 1. The phylogenetic tree of the Plasmodium 18S rRNA gene and the primer binding regions for cLAMP detection.

(a) A maximum-likelihood phylogenetic tree of the 18S rRNA gene from five Plasmodium species, supported by bootstrap from 1,000 pseudo replications. (b) The reference primer binding regions are used for LAMP detection. (c) The primer binding regions for designed primer in this study targeted the A-type of 18S rRNA genes located on chromosome 5 (XR_002273101.1) and 7 (XR_002273081.2).

https://doi.org/10.1371/journal.pone.0298087.g001

thumbnail
Table 1. Primer sequence for cLAMP detection of P. falciparum 18S rRNA gene.

https://doi.org/10.1371/journal.pone.0298087.t001

Plasmid DNA construction

Plasmid DNA (pDNA) was used to verify the cLAMP primer specificity within Plasmodium species and to determine the limit of detection (LoD). The partial nucleic acid sequences of the 18S rRNA genes from five human malaria parasites were synthesized and cloned into the pTWIST-Amp pDNA (Twist Bioscience, CA) (S1 Table). The pDNA was initially transformed into DH5-alpha competent E. coli cells using the heat shock method and subsequently cultured in 5 mL of LB broth containing 100 mg/mL ampicillin. Then, the pDNA was purified using a Plasmid DNA extraction kit (Macherey-Nagel, Duren, Germany), following the manufacturer’s protocol. The concentration of the extracted pDNA was measured using a spectrometer (Titertek-Berthold, Pforzheim, Germany) and then subjected to ten-fold serial dilution for further analysis, including primer specificity and sensitivity studies.

P. falciparum genomic DNA preparation

To determine the LoD based on the parasite count, the P. falciparum 3D7 strain (ATCC) was cultured using human erythrocytes (2% hematocrit) in RPMI-1640 medium (Invitrogen, CA). The medium was supplemented with 2.3 g/L sodium bicarbonate, 0.05 g/L hypoxanthine, 10% Albumax I solution, and 10 mg/mL gentamicin. The culture was maintained in a gas mixture of 90% N2, 5% O2, and 5% CO2 and incubated at 37°C. Thin blood films were prepared following standard protocols, and parasitemia was measured through Giemsa staining under a microscope at 100X magnification. When the parasitemia reached 5%, genomic DNA was extracted from the cultured samples using a QIAamp DNA mini kit (QIAGEN, Hilden, Germany), following the manufacturer’s protocol. The genomic DNA (gDNA) was utilized for both confirming the LoD analysis by serial dilution of gDNA and verifying the performance of real-time colorimetric LAMP (RT-cLAMP).

Colorimetric LAMP setting

The cLAMP reaction mixture was prepared in a total volume of 25 μL, comprising 12.5 μL of WarmStart® Colorimetric 2X Master Mix (NEB, MA), 0.96 μM of each outer primer (F3 and B3c), 7.68 μM of each inner primer (FIP and BIP), 3.84 μM of each loop primer (LpF and LpB), 1.0 μL of template DNA (pDNA or gDNA), and 5.5 μL of double-distilled water (DDW). The incubation was carried out using the ProFlex PCR System (Applied Biosystems, MA), and the optimal conditions for the cLAMP to detect P. falciparum were determined at 65°C for 30 minutes. Finally, the amplification products were confirmed directly by naked eye observation and electrophoresis with 2% agarose gel.

Polymerase Chain Reaction (PCR)

PCR amplification was performed to confirm the species specificity and sensitivity of the outer primer set (F3-B3c) (S2 Table). PCR mixture was prepared in a total volume of 20 μL comprising AccuPower PCR Premix (Bioneer, Daejeon, Republic of Korea), 1 μL of template DNA (pDNA or gDNA), 16 μL of DDW, and 1.0 μL of each 10 μM outer primers. The PCR condition consisted of an initial denaturation step at 95°C for 5 minutes, followed by 35 cycles (95°C for 30 seconds, 56°C for 30 seconds, and 72°C for 30 seconds), and a final extension step at 72°C for 5 minutes. The amplicons were visualized using electrophoresis with 2% agarose gel.

Nested PCR

To diagnose P. falciparum in field isolates, we employed nested PCR (nPCR) to compare with cLAMP results [10]. The nPCR mixture was prepared in a total volume of 20 μL, which included AccuPower PCR Premix (Bioneer), 1 μL of template DNA (gDNA), 16 μL of DDW, and 1.0 μL of each 10 μM primer. The initial PCR reaction utilized genus-specific primers (rPLU1 and rPLU5) to amplify the 18S rRNA gene (S2 Table). DNA was amplified following conditions: 95°C for 10 minutes, 35 cycles (95°C for 30 seconds, 56°C for 30 seconds, and 72°C for 1.5 minutes), and a final extension at 72°C for 10 minutes. In the second PCR reaction, 1 μL of the first PCR product was combined with P. falciparum-specific primers (rFAL1 and rFAL2) and subjected to a final extension at 72°C for 30 seconds (S2 Table). The amplicons were visualized using electrophoresis with 2% agarose gel.

Quantitative PCR

Quantitative PCR (qPCR) was performed and compared to verify cLAMP performance using cultured P. falciparum and field isolates [11]. Reactions were conducted in a total volume of 20 μL, including 10 μL of 2X PrimeTime Gene Expression Master Mix with reference dye (Integrated DNA Technologies, IA), 4 μL of template DNA (gDNA), 10 μM species-specific primers, and 5 μM of a TaqMan probe (S2 Table). The TaqMan probe assay was performed on an AriaMx Real-Time PCR System (Agilent Technologies, CA) with the following cycling conditions: hot-start polymerase activation at 95°C for 10 min, followed by 40 cycles of denaturation at 95°C for 15 seconds and final annealing for 65 seconds at 65°C.

Real-time colorimetric LAMP (RT-cLAMP)

We adapted the cLAMP assay to verify its real-time performance. This real-time cLAMP (RT-cLAMP) was conducted by adding SYBR Green I fluorescence (Invitrogen, CA) into the cLAMP reaction, and signals were acquired every minute using the AriaMx Real-Time PCR System (Agilent Technologies). The cLAMP reaction mixture was prepared as described above, with the addition of SYBR Green I (1:10,000) for visualizing amplicon amounts and ROX (1:250) as a reference dye. The incubation was carried out at 65°C for 60 minutes, and the reaction data were analyzed using the Agilent AriaMx software (Agilent Technologies).

Diagnosis of clinical isolates and ethical approval

Dried blood spots (DBS) from field isolates were collected during a clinical evaluation study of a malaria diagnostic device called miLab™ (Noul Co., Yongin, Republic of Korea). The clinical isolates were confirmed to have P. falciparum infection without other species co-infection using Bioline MALARIA Ag P.f (Abbott, IL) and microscopy during the collection of isolates. Regardless of their age, informed consent was obtained from patients seeking treatments at Mzuzu Urban Health Center and who were suspected to have malaria infections. All consent were drawn prior to participation, and for children parents/guardians were consented for their involvement. The study received approval on June 6, 2021, from the National Health Sciences Research Committee (NHSRC), a department of the Ministry of Health (MoH) in Malawi (IRB00003905). The samples were collected between June 2021 and May 2022. Clinical isolates were confirmed through microscopic examination at the Mzuzu Urban Health Centre by a trained technician. Whole blood samples were spotted onto Whatman 903 protein saver cards (Whatman International Ltd., Maidstone, England) and allowed to air-dry completely during collection, enclosed into separate plastic bags and transported to Kangwon National University for further molecular analysis. A total of eighty-four dried blood spots (DBS) were randomly selected from the collected samples and used in this study. Genomic DNA was extracted following the manufacturer’s protocol for QIAamp DNA Mini Kits (QIAGEN). The extracted gDNA was subsequently used for nested PCR (nPCR), quantitative PCR (qPCR), and colorimetric LAMP (cLAMP) to compare their performance for molecular diagnosis [10,11]. Parasitemia in the clinical isolates ranged from 1,069 to 285,290 parasites/μL in total 16 samples out of 84 samples, as observed through microscopic examination. All molecular diagnosis were conducted blindly, without prior knowledge of the microscopic examination results. The cLAMP reaction’s endpoint product was visualized with the naked eye under optimal conditions at 65°C for 30 minutes of incubation. All experiments using human specimens adhered to relevant guidelines and regulations and were approved on May 24, 2023, by the Kangwon National University, Republic of Korea (IRB No. KWNUIRB-2023-05-008).

Data visualization and statistical analysis

The qPCR and RT-cLAMP data were obtained from Agilent AriaMx software (Agilent Technologies), and graph visualized by GraphPad Prism 8 software. Comparison of parasite count (parasites/μL) versus qPCR cycles or cLAMP incubation time (min.) and linear regression calculation were visualized by SigmaPlot software v12.

The statistical calculations of clinical specimen diagnosis were performed using the MEDCALP statistical software, which is accessible at https://www.medcalc.org/calc/diagnostic_test.php. Data were compiled into double-entry tables for calculation of sensitivity, specificity, and positive and negative predictive values for each test, along with the determination of 95% confidence intervals.

Results

Colorimetric LAMP primer design

We conducted an analysis of phylogenetic relationships among human malaria parasites to identify the colorimetric LAMP (cLAMP) primer. The A-type 18S rRNA gene sequences of Plasmodium species formed phylogenetic clusters with its respective orthologs (Fig 1A). Thus, the A-type of 18S rRNA gene is suitable for designing a cLAMP primer set with conserved sequences at the intra-species level. We selected the most cited primer set for LAMP assay from previous studies and used as the reference primer set (Fig 1B and Table 1) [15]. This reference primer targeted the P. falciparum S-type 18S rRNA gene, whereas the newly designed primer set targeted the A-type 18S rRNA gene. The A-type 18S rRNA gene exhibits a high copy number (1 x 104 copies per ring stage parasite) at the asexual stage and shares an identical nucleic acid sequence on chromosomes 5 (XR_002273101.1) and 7 (XR_002273081.2) (Fig 1C and Table 1).

Specificity of cLAMP primer between Plasmodium species

To validate the specificity of the colorimetric LAMP (cLAMP) primers, we synthesized partial 18S rRNA gene fragments from human malaria species (S1 Table). The S-type and A-type of P. falciparum 18S rRNA genes (PfS and PfA) were used to verify the reference primer set and the designed primer set, respectively [15]. Furthermore, homologous regions of the A-type 18S rRNA gene in its orthologs, including P. vivax (PvA), P. knowlesi (PkA), P. malariae (PmA), and P. ovale (PoA), were synthesized to confirm primer specificity.

PCR was conducted to confirm the species-specificity of each outer primer set (F3-B3c). The designed and reference outer primers showed specific amplification in the A-type (PfA) and S-type (PfS) 18S rRNA genes of P. falciparum, with estimated amplicon sizes of 229 bp and 220 bp, respectively (Fig 2A). After confirming the specificity of the outer primer set, the cLAMP was conducted under optimal conditions at 65°C for a 30-minute incubation. The presence of the cLAMP amplicon was confirmed by visualizing on a 2% agarose gel electrophoresis, showing a typical ladder pattern of LAMP reaction (Fig 2B). The designed primer showed specific amplification in the A-type of 18S rRNA (PfA). In contrast, the reference primer set amplified both the A-type and S-type of 18S rRNA (PfA and PfS) (Fig 2B). The direct observation by naked eye of the designed primer showed an obvious color change in PfA to yellow, while the reference primer showed intermediate color changes in both PfA and PfS (Fig 2C). Taken together, the cLAMP primers showed no cross-amplification issues with other Plasmodium species.

thumbnail
Fig 2.

The cLAMP specificity of the design (left panel) and reference (right panel) primer set. The plasmid DNA (pDNA) containing 18S rRNA genes from both the P. falciparum S-type (PfS) and A-type (PfA) were utilized to confirm primer specificity. A-type 18S rRNA genes from other Plasmodium species were also employed for confirmation: P. vivax (PvA), P. knowlesi (PkA), P. malariae (PmA), and P. ovale (PoA). (a) The results indicated the specificity of PCR with the outer primer set (F3-B3c) for their corresponding fragments. PfA, with the designed primer, exhibited an expected size of 229 bp, while PfS, with the reference primer, showed an expected size of 220 bp. Lane M represents the DNA ladder mix. (b) The cLAMP results were visualized through 2% agarose gel electrophoresis. (c) The cLAMP results corresponded to (b) and were confirmed through naked eye direct observation. Yellow/orange coloration indicated a positive, whereas pink indicated a negative.

https://doi.org/10.1371/journal.pone.0298087.g002

Analytical Limit of detection (LoD) of cLAMP

The accurate LoD of cLAMP was estimated by two methods utilizing synthesized 18S rRNA genes to calculate the target gene copy number and in vitro cultured P. falciparum.

For the determination of the LoD based on the target gene copy number, synthesized 18S rRNA genes were employed. The copy number of the A-type 18S rRNA was estimated to be 1 x 104 copies per ring stage parasite in a previous study [32]. The PfA and PfS were serially diluted ten-fold from 1.68 x 109 copies/μL which corresponding to 168,000 ring stage parasites/μL. The PCR using outer primers (F3-B3c) detection limit estimated a concentration of 1.68 x 104 copies/μL (corresponding to 1.68 ring stage parasites/μL) when using the designed primer and 1.68 x 108 copies/μL with the reference primer (Fig 3A). The cLAMP results were confirmed through electrophoresis and demonstrated the same sensitivity as the PCR results (Fig 3B). The LoD of the designed primer was 104 times higher in sensitivity compared to the reference primer in both PCR and cLAMP. However, when observed with the naked eye, the designed primer exhibited an intermediate color change at the endpoint of LoD (1.68 x 104 copies/μL) (Fig 3C).

thumbnail
Fig 3. Limit of detection based on plasmid DNA.

(a) PCR detection using the outer primer set with serially diluted plasmid DNA. Ten microliters of the PCR products were visualized on a 2% agarose gel. (b) One microliter of the endpoint products from the cLAMP reaction was observed on a 2% agarose gel. (c) cLAMP results were interpreted by the naked eye; yellow/orange indicates a positive result, whereas pink indicates a negative result. The lane contained P. vivax gDNA (P. vivax) and DDW, which were used as negative controls.

https://doi.org/10.1371/journal.pone.0298087.g003

The LoD based on parasite count (parasites/μL) was performed using genomic DNA extracted from cultured P. falciparum. Initially, PCR was performed to confirm the LoD of both the design and reference outer primers. The designed primer had a 10 times higher amplification efficiency, resulting in a detection limit of 21 parasites/μL, while the reference primer had a detection limit of 210 parasites/μL (Fig 4A). When compared with PCR result, the cLAMP with the designed primer showed a 100-fold higher sensitivity in detecting up to 0.21 parasites/μL (Fig 4B). In contrast, the reference primer showed the same LoD in both PCR and cLAMP (Fig 4B). When comparing the cLAMP results between the designed and reference primers, it revealed a 1,000-fold higher sensitivity in the designed primer (Fig 4C). However, when observed with the naked eye, both the designed and reference primer showed an intermediate color change at each endpoint of LoD 0.21 parasites/μL and 210 parasites/μL, respectively (Fig 4C). This result showed that cLAMP has a higher accuracy and more sensitivity in detecting low parasite numbers compared to PCR.

thumbnail
Fig 4. Limit of detection test based on parasite count.

(a) PCR detection using the outer primer set with serially diluted parasites genomic DNA (parasites/μL; p/μL). Ten microliters of the amplicon were visualized on a 2% agarose gel. (b) One microliter of the cLAMP product was visualized on a 2% agarose gel. Negative control consisting of P. vivax gDNA (P. vivax) and DDW were used to confirm specificity. (c) The cLAMP results corresponding to (b) were visualized with the naked eye; yellow indicated a positive, whereas pink indicated a negative.

https://doi.org/10.1371/journal.pone.0298087.g004

Performance of the cLAMP

To confirm the cLAMP performance, a real-time cLAMP (RT-cLAMP) was conducted and compared to quantitative PCR (qPCR) diagnosis methods [11]. Based on the analytical threshold (ΔR = 50), the qPCR positive read was determined at 21,000 parasites/μL from 23 PCR cycles (Fig 5A). A plot correlating the log threshold cycle with log parasite count (parasites/μL) showed a strong linear regression (r2 = 0.9949) in qPCR detection, with 98.9% amplification efficiency, indicating a 1.989-fold amplification per cycle (Fig 5B).

thumbnail
Fig 5. The performance verification of cLAMP.

(a) Fluorescence (ΔR) detection was based on ten-fold diluted P. falciparum 3D7 gDNA (from 2.1 x 104 to 2.1 x 10−2 parasites/μL) using qPCR. The dotted line indicates the analytical positive threshold line (ΔR = 50). (b) The comparison of parasite count (parasites/μL) versus qPCR cycles shows a linear regression fit with r2 = 0.9949. All error bars represent the standard deviation of triplicate reactions. (c) RT-cLAMP detection results were acquired using SYBR green signal every minute of incubation time. The horizontal dotted line indicates the analytical positive threshold line (ΔR = 730), while the vertical dotted line indicates the time point of positive reading for both 2.1 x 104 and 10−1 parasites/μL. (d) The comparison of parasite count versus RT-cLAMP incubation time (min.) shows a linear regression fit with r2 = 0.9719. All error bars represent the standard deviation of triplicate reactions. (e) Observation of color changes by the naked eye was done every 5 minutes; yellow indicates a positive, whereas pink indicates a negative.

https://doi.org/10.1371/journal.pone.0298087.g005

To estimate the amount of amplicon in RT-cLAMP, fluorescence data were acquired at 1-minute intervals for up to 1 hour. The fluorescence signal surpassed the analytical positive threshold (ΔR = 730) after 19 minutes of incubation at a parasite count of 21,000 parasites/μL. (Fig 5C). The lowest detection was observed at 0.21 parasites/μL after 31 minutes of incubation, which was consistent with the LoD of qPCR (Fig 5A and 5C). A plot correlating the log threshold with time in minutes versus log parasite count revealed a linear regression (r2 = 0.9719) with 179.1% amplification efficiency, indicating a 2.791-fold amplification per minute (Fig 5D).

The naked eye observation of the cLAMP showed distinguishable color changes after 15 minutes of incubation at 21,000 parasites/μL (Fig 5E). The lowest detection of 0.21 parasites/μL showed a color change at the 25-minute time point (Fig 5E). However, parasite counts lower than the lowest detection limit (0.021 parasites/μL) were not detectable by RT-cLAMP and naked eye observation, consistent with the sensitivity of qPCR (Fig 5C and 5E). Taken together, the cLAMP analytical LoD was determined to be 0.21 parasites/μL, and it was detectable within 31 minutes in both the RT-cLAMP and 25 minutes by the naked eye observation.

The sensitivity and specificity of cLAMP diagnosis using clinical specimens

To assess the performance comparison of cLAMP, we conducted and compared three different molecular diagnostic tests: nested PCR (nPCR), quantitative PCR (qPCR), and colorimetric LAMP (cLAMP). The molecular detection was carried out as blind tests, with unknown positivity in microscopic examination. The direct comparisons between cLAMP and nPCR are shown in Fig 6A. The determination of P. falciparum-positive isolates was based on microscopic examination results, allowing us to calculate the sensitivity and specificity of each molecular diagnostic method. Based on the microscopic examination, 16 out of the 84 isolates were positive for P. falciparum, and all three molecular diagnostic methods successfully detected the positive samples (S3 Table). The sixteen microscopy-positive samples were identified with a range of 1,069 to 285,290 parasites/μL (mean = 63,114 parasites/μL) (S3 Table). The results for nPCR, qPCR, and cLAMP revealed sensitivity and positive predictive values of 100% (95% CI: 79.41–100%) (Table 2). Also, specificity and negative predictive values showed to be 100% (95% CI: 94.72–100%) (Table 2). The overall accuracy reached up to 100% (95% CI: 95.70–100%) in all tests when compared with microscopic examination (Table 2).

thumbnail
Fig 6. Validation of cLAMP using 84 clinical specimens.

Comparison of diagnosis performance between cLAMP and nPCR for P. falciparum detection. The cLAMP reaction was read by the naked eye; yellow/orange indicates a positive result, and pink indicates a negative result. The nPCR result was visualized on a 2% agarose gel.

https://doi.org/10.1371/journal.pone.0298087.g006

thumbnail
Table 2. Performance comparison between microscopy versus each molecular diagnosis methods.

Sensitivity, specificity, positive predictive value, negative predictive value, and accuracy were calculated based on the 84 isolates. The number of isolates used was 16 true positive (ranging in 1,069~285,290 paraistes/μL) and 68 true negative based on microscopic examination result.

https://doi.org/10.1371/journal.pone.0298087.t002

Discussion

Early diagnosis is the most important strategy for malaria control in hyperendemic areas. However, the absence of affordable healthcare options and accurate tools for parasite early detection is the challenge to malaria eradication in resource-limited areas [33]. In the early 2000s, LAMP emerged as a novel amplification method with superior properties to be used in field studies. Although most LAMP assays were conducted in a laboratory settings, their applicability in the field is facilitated by cost-effectiveness, easy to performance, and rapidity. Furthermore, LAMP results can be visually interpreted through turbidimetry, fluorescence, or colorimetric changes under a consistent incubation temperature that enhanced their field application [29,3436]. Thus, this study focused on developing a sensitive colorimetric LAMP (cLAMP) primer set and its assay for the early detection of P. falciparum at low parasitemia, potentially enhancing its utilization in the field.

The first study of LAMP assay for P. falciparum achieved 95% sensitivity and 99% specificity when compared to PCR [37]. Since then, several studies have been conducted to enhance the efficiency of LAMP diagnosis. Among those, some studies have targeted the S-type of the 18S rRNA genes for parasite detection [15,38]. However, the S-types of the 18S rRNA were found to be abundant in the gametocyte for S1-type and in sporozoites within the mosquito salivary gland for S2-type [30]. Besides, variation between S1-type and S2-type has been observed in P. falciparum, P. berghei, and P. yoelii [30]. Hence, the primer set designed in this study targeted the A-type of the 18S rRNA gene for dual gene targeting, which shares nearly 100% sequence identity on chromosomes 5 and 7 [30]. One study using patient whole blood demonstrated that 96.8% patients presented higher copy numbers of the A-type 18S rRNA gene than S-type 18S rRNA gene in the erythrocytic stage [39]. Further, the A-type of 18S rRNA gene is stable and is expressed at high levels with approximately 1 x 104 copies in the asexual stage, particularly in the ring stage of parasites [32,40,41]. When collecting peripheral blood from P. falciparum-infected patients for diagnosis, it typically contains only ring stage and gametocyte forms. Therefore, selecting the A-type of the 18S rRNA gene is the most suitable target for early detection, especially in cases of low parasitemia.

In this study, the performance of cLAMP was demonstrated at a LoD of 0.21 parasites/μL, whereas several studies have reported LoDs ranging from 1 to 5 parasites/μL achieved using various mitochondrial DNA targets [25,28,42]. The sensitivity and specificity reported in previous studies ranged from 95% to 98.5% and 96% to 100%, respectively, and all of these studies demonstrated high-quality performance [15,37,42]. However, the reported studies required a reaction incubation time of 45 minutes to 2 hours and the result reading necessitated the use UV illuminator or bioluminescent real-time reporter system, which pose difficulties in field utilization. An important feature of cLAMP assay in this study is its ability to swiftly detected low parasite levels in a shorter incubation time, without the need for specialized equipment [22]. Furthermore, colorimetric detection methods are able to prevent cross-contamination issues when applied in the field condition [23].

Although this study demonstrated many advantages, it is also compromised with certain limitations. Sample preparation needs to be considered, as it can be challenging to conduct in the field. Many methods have been developed to simplify genomic DNA extraction [43,44]. Thus, this cLAMP assay needs to be combined with a simplified genomic DNA extraction system for field application. In this study, we used dried blood spots for the extraction of genomic DNA in the laboratory. The dried blood spot contain various biologically active substances, including cellular components and enzyme inhibitors, which can potentially affect the diagnostic sensitivity when compared to fresh blood samples [45]. The lowest parasite count in the specimen pool of this study was 1,069 parasites/μL (0.02% parasitemia), successfully detected by cLAMP. However, the tested sample pool did not contain an extremely low parasitemia, which can be detected only by molecular detection method and not by microscopy. Thus, this could pose a limitation for statistical analysis. This study used a limited number of clinical isolates, and it is necessary to employ a more varied range of sample pools, including those with fresh specimens and extremely low parasitemia. Thus, it requires confirmation under various sample conditions in the further. In addition, similar to previous studies, interpreting intermediate color changes can be affected by individual differences which is also considered a significant limitation in cLAMP diagnosis for making definitive decisions regarding P. falciparum infections, especially when dealing with extremely low parasitaemia (<0.21 parasites/μL) [22,29]. Such suspected results need to be reconfirmed by the other molecular detection technique.

Despite these limitations, the cLAMP with a newly designed primer set demonstrated higher sensitivity, faster performance, and ease of use compared to other molecular detection methods. These findings offer a time effective technique and simpler result interpretation, making cLAMP a promising choice for the point-of-care (PoC) tool for early malaria detection and large-scale screening.

Supporting information

S1 Fig. The alignment of the primer binding regions within the human malaria parasites.

Two 18S rRNA sequences on chromosome 5 (accession no. XR_002273101.1) and chromosome 7 (accession no. XR_002273081.2) are indicated by different color arrows for each primer binding region (black for F3/B3c pair, green for F2/B2c, yellow for F1c/B1, and red for LpF/LpB).

https://doi.org/10.1371/journal.pone.0298087.s001

(DOCX)

S1 Table. The partial species-specific nucleotide sequences of the 18S rRNA genes from five human malaria parasites: P. falciparum, P. vivax, P. malariae, P. knowlesi, and P. ovale.

https://doi.org/10.1371/journal.pone.0298087.s002

(DOCX)

S2 Table. Primer sequences used in PCR, qPCR, and nPCR for P. falciparum detection.

https://doi.org/10.1371/journal.pone.0298087.s003

(DOCX)

S3 Table. A summary of P. falciparum positive isolates in each diagnosis method.

https://doi.org/10.1371/journal.pone.0298087.s004

(DOCX)

Acknowledgments

The authors are grateful for all the staff and patients associated with the clinics and patients in Malawi.

References

  1. 1. World Health Organization. World Malaria Report 2022. 2022.
  2. 2. World Health Organization. Guidelines for the treatment of malaria.2015.
  3. 3. Amexo M, Tolhurst R, Barnish G, Bates I. Malaria misdiagnosis: effects on the poor and vulnerable. Lancet. 2004;364(9448):1896–8. pmid:15555670.
  4. 4. Orish VN, De-Gaulle VF, Sanyaolu AO. Interpreting rapid diagnostic test (RDT) for Plasmodium falciparum. BMC Res Notes. 2018;11(1):850. Epub 20181204. pmid:30509313; PubMed Central PMCID: PMC6278119.
  5. 5. Berzosa P, de Lucio A, Romay-Barja M, Herrador Z, Gonzalez V, Garcia L, et al. Comparison of three diagnostic methods (microscopy, RDT, and PCR) for the detection of malaria parasites in representative samples from Equatorial Guinea. Malar J. 2018;17(1):333. Epub 20180917. pmid:30223852; PubMed Central PMCID: PMC6142353.
  6. 6. Bell DR, Wilson DW, Martin LB. False-positive results of a Plasmodium falciparum histidine-rich protein 2-detecting malaria rapid diagnostic test due to high sensitivity in a community with fluctuating low parasite density. Am J Trop Med Hyg. 2005;73(1):199–203. pmid:16014858.
  7. 7. Wongsrichanalai C, Barcus MJ, Muth S, Sutamihardja A, Wernsdorfer WH. A review of malaria diagnostic tools: microscopy and rapid diagnostic test (RDT). Am J Trop Med Hyg. 2007;77(6 Suppl):119–27. pmid:18165483.
  8. 8. Murray CK, Gasser RA Jr., Magill AJ, Miller RS. Update on rapid diagnostic testing for malaria. Clin Microbiol Rev. 2008;21(1):97–110. pmid:18202438; PubMed Central PMCID: PMC2223842.
  9. 9. Jejaw Zeleke A, Hailu A, Bayih AG, Kefale M, Amare AT, Tegegne Y, et al. Plasmodium falciparum histidine-rich protein 2 and 3 genes deletion in global settings (2010–2021): a systematic review and meta-analysis. Malar J. 2022;21(1):26. Epub 20220129. pmid:35093092; PubMed Central PMCID: PMC8800273.
  10. 10. Snounou G, Viriyakosol S, Zhu XP, Jarra W, Pinheiro L, do Rosario VE, et al. High sensitivity of detection of human malaria parasites by the use of nested polymerase chain reaction. Mol Biochem Parasitol. 1993;61(2):315–20. pmid:8264734.
  11. 11. Wang B, Han SS, Cho C, Han JH, Cheng Y, Lee SK, et al. Comparison of microscopy, nested-PCR, and Real-Time-PCR assays using high-throughput screening of pooled samples for diagnosis of malaria in asymptomatic carriers from areas of endemicity in Myanmar. J Clin Microbiol. 2014;52(6):1838–45. Epub 20140319. pmid:24648557; PubMed Central PMCID: PMC4042795.
  12. 12. Srisutham S, Saralamba N, Malleret B, Renia L, Dondorp AM, Imwong M. Four human Plasmodium species quantification using droplet digital PCR. PLoS One. 2017;12(4):e0175771. Epub 20170419. pmid:28423028; PubMed Central PMCID: PMC5396971.
  13. 13. Tedla M. A focus on improving molecular diagnostic approaches to malaria control and elimination in low transmission settings: Review. Parasite Epidemiol Control. 2019;6:e00107. Epub 20190425. pmid:31193331; PubMed Central PMCID: PMC6525315.
  14. 14. Britton S, Cheng Q, McCarthy JS. Novel molecular diagnostic tools for malaria elimination: a review of options from the point of view of high-throughput and applicability in resource limited settings. Malar J. 2016;15:88. Epub 20160216. pmid:26879936; PubMed Central PMCID: PMC4754967.
  15. 15. Han ET, Watanabe R, Sattabongkot J, Khuntirat B, Sirichaisinthop J, Iriko H, et al. Detection of four Plasmodium species by genus- and species-specific loop-mediated isothermal amplification for clinical diagnosis. J Clin Microbiol. 2007;45(8):2521–8. Epub 20070613. pmid:17567794; PubMed Central PMCID: PMC1951264.
  16. 16. Kersting S, Rausch V, Bier FF, von Nickisch-Rosenegk M. Rapid detection of Plasmodium falciparum with isothermal recombinase polymerase amplification and lateral flow analysis. Malar J. 2014;13:99. Epub 20140315. pmid:24629133; PubMed Central PMCID: PMC4004163.
  17. 17. Foo PC, Nurul Najian AB, Muhamad NA, Ahamad M, Mohamed M, Yean Yean C, et al. Loop-mediated isothermal amplification (LAMP) reaction as viable PCR substitute for diagnostic applications: a comparative analysis study of LAMP, conventional PCR, nested PCR (nPCR) and real-time PCR (qPCR) based on Entamoeba histolytica DNA derived from faecal sample. BMC Biotechnol. 2020;20(1):34. Epub 20200622. pmid:32571286; PubMed Central PMCID: PMC7310076.
  18. 18. Charpentier E, Benichou E, Pages A, Chauvin P, Fillaux J, Valentin A, et al. Performance evaluation of different strategies based on microscopy techniques, rapid diagnostic test and molecular loop-mediated isothermal amplification assay for the diagnosis of imported malaria. Clin Microbiol Infect. 2020;26(1):115–21. Epub 20190531. pmid:31158521.
  19. 19. Hartmeyer GN, Hoegh SV, Skov MN, Kemp M. Use of Loop-Mediated Isothermal Amplification in a Resource-Saving Strategy for Primary Malaria Screening in a Non-Endemic Setting. Am J Trop Med Hyg. 2019;100(3):566–71. pmid:30675835; PubMed Central PMCID: PMC6402892.
  20. 20. Polley SD, Gonzalez IJ, Mohamed D, Daly R, Bowers K, Watson J, et al. Clinical evaluation of a loop-mediated amplification kit for diagnosis of imported malaria. J Infect Dis. 2013;208(4):637–44. Epub 20130430. pmid:23633403; PubMed Central PMCID: PMC3719897.
  21. 21. Nolasco O, Montoya J, Rosales Rosas AL, Barrientos S, Rosanas-Urgell A, Gamboa D. Multicopy targets for Plasmodium vivax and Plasmodium falciparum detection by colorimetric LAMP. Malar J. 2021;20(1):225. Epub 20210519. pmid:34011373; PubMed Central PMCID: PMC8135177.
  22. 22. Lai MY, Ooi CH, Jaimin JJ, Lau YL. Evaluation of WarmStart Colorimetric Loop-Mediated Isothermal Amplification Assay for Diagnosis of Malaria. Am J Trop Med Hyg. 2020;102(6):1370–2. pmid:32228783; PubMed Central PMCID: PMC7253095.
  23. 23. Dao Thi VL, Herbst K, Boerner K, Meurer M, Kremer LP, Kirrmaier D, et al. A colorimetric RT-LAMP assay and LAMP-sequencing for detecting SARS-CoV-2 RNA in clinical samples. Sci Transl Med. 2020;12(556). Epub 20200727. pmid:32719001; PubMed Central PMCID: PMC7574920.
  24. 24. Paris DH, Imwong M, Faiz AM, Hasan M, Yunus EB, Silamut K, et al. Loop-mediated isothermal PCR (LAMP) for the diagnosis of falciparum malaria. Am J Trop Med Hyg. 2007;77(5):972–6. pmid:17984362.
  25. 25. Polley SD, Mori Y, Watson J, Perkins MD, Gonzalez IJ, Notomi T, et al. Mitochondrial DNA targets increase sensitivity of malaria detection using loop-mediated isothermal amplification. J Clin Microbiol. 2010;48(8):2866–71. Epub 20100616. pmid:20554824; PubMed Central PMCID: PMC2916605.
  26. 26. Mohon AN, Getie S, Jahan N, Alam MS, Pillai DR. Ultrasensitive loop mediated isothermal amplification (US-LAMP) to detect malaria for elimination. Malar J. 2019;18(1):350. Epub 20191016. pmid:31619258; PubMed Central PMCID: PMC6796404.
  27. 27. Aninagyei E, Boakye AA, Tettey CO, Ntiri KA, Ofori SO, Tetteh CD, et al. Utilization of 18s ribosomal RNA LAMP for detecting Plasmodium falciparum in microscopy and rapid diagnostic test negative patients. PLoS One. 2022;17(10):e0275052. Epub 20221006. pmid:36201568; PubMed Central PMCID: PMC9536604.
  28. 28. Aydin-Schmidt B, Xu W, Gonzalez IJ, Polley SD, Bell D, Shakely D, et al. Loop mediated isothermal amplification (LAMP) accurately detects malaria DNA from filter paper blood samples of low density parasitaemias. PLoS One. 2014;9(8):e103905. Epub 20140808. pmid:25105591; PubMed Central PMCID: PMC4126669.
  29. 29. Lucchi NW, Ljolje D, Silva-Flannery L, Udhayakumar V. Use of Malachite Green-Loop Mediated Isothermal Amplification for Detection of Plasmodium spp. Parasites. PLoS One. 2016;11(3):e0151437. Epub 20160311. pmid:26967908; PubMed Central PMCID: PMC4788150.
  30. 30. McGee JP, Armache JP, Lindner SE. Ribosome heterogeneity and specialization of Plasmodium parasites. PLoS Pathog. 2023;19(4):e1011267. Epub 20230413. pmid:37053161; PubMed Central PMCID: PMC10101515.
  31. 31. Tamura K, Stecher G, Kumar S. MEGA11: Molecular Evolutionary Genetics Analysis Version 11. Mol Biol Evol. 2021;38(7):3022–7. pmid:33892491; PubMed Central PMCID: PMC8233496.
  32. 32. Murphy SC, Prentice JL, Williamson K, Wallis CK, Fang FC, Fried M, et al. Real-time quantitative reverse transcription PCR for monitoring of blood-stage Plasmodium falciparum infections in malaria human challenge trials. Am J Trop Med Hyg. 2012;86(3):383–94. pmid:22403305; PubMed Central PMCID: PMC3284350.
  33. 33. Bhatt S, Weiss DJ, Cameron E, Bisanzio D, Mappin B, Dalrymple U, et al. The effect of malaria control on Plasmodium falciparum in Africa between 2000 and 2015. Nature. 2015;526(7572):207–11. Epub 20150916. pmid:26375008; PubMed Central PMCID: PMC4820050.
  34. 34. Notomi T, Okayama H, Masubuchi H, Yonekawa T, Watanabe K, Amino N, et al. Loop-mediated isothermal amplification of DNA. Nucleic Acids Res. 2000;28(12):E63. pmid:10871386; PubMed Central PMCID: PMC102748.
  35. 35. Mori Y, Nagamine K, Tomita N, Notomi T. Detection of loop-mediated isothermal amplification reaction by turbidity derived from magnesium pyrophosphate formation. Biochem Biophys Res Commun. 2001;289(1):150–4. pmid:11708792.
  36. 36. Yamamura M, Makimura K, Ota Y. Evaluation of a new rapid molecular diagnostic system for Plasmodium falciparum combined with DNA filter paper, loop-mediated isothermal amplification, and melting curve analysis. Jpn J Infect Dis. 2009;62(1):20–5. pmid:19168954.
  37. 37. Poon LL, Wong BW, Ma EH, Chan KH, Chow LM, Abeyewickreme W, et al. Sensitive and inexpensive molecular test for falciparum malaria: detecting Plasmodium falciparum DNA directly from heat-treated blood by loop-mediated isothermal amplification. Clin Chem. 2006;52(2):303–6. Epub 20051208. pmid:16339303.
  38. 38. Puri M, Kaur Brar H, Madan E, Srinivasan R, Rawat K, Gorthi SS, et al. Rapid diagnosis of Plasmodium falciparum malaria using a point-of-care loop-mediated isothermal amplification device. Front Cell Infect Microbiol. 2022;12:961832. Epub 20220819. pmid:36061865; PubMed Central PMCID: PMC9437306.
  39. 39. Gruenberg M, Moniz CA, Hofmann NE, Wampfler R, Koepfli C, Mueller I, et al. Plasmodium vivax molecular diagnostics in community surveys: pitfalls and solutions. Malar J. 2018;17(1):55. Epub 20180130. pmid:29378609; PubMed Central PMCID: PMC5789620.
  40. 40. Seilie AM, Chang M, Hanron AE, Billman ZP, Stone BC, Zhou K, et al. Beyond Blood Smears: Qualification of Plasmodium 18S rRNA as a Biomarker for Controlled Human Malaria Infections. Am J Trop Med Hyg. 2019;100(6):1466–76. pmid:31017084; PubMed Central PMCID: PMC6553913.
  41. 41. McCutchan TF, Li J, McConkey GA, Rogers MJ, Waters AP. The cytoplasmic ribosomal RNAs of Plasmodium spp. Parasitol Today. 1995;11(4):134–8. pmid:15275356.
  42. 42. Zainabadi K. Ultrasensitive Diagnostics for Low-Density Asymptomatic Plasmodium falciparum Infections in Low-Transmission Settings. J Clin Microbiol. 2021;59(4). Epub 20210319. pmid:33148707; PubMed Central PMCID: PMC8092727.
  43. 43. Ghatak S, Muthukumaran RB, Nachimuthu SK. A simple method of genomic DNA extraction from human samples for PCR-RFLP analysis. J Biomol Tech. 2013;24(4):224–31. pmid:24294115; PubMed Central PMCID: PMC3792701.
  44. 44. Singh UA, Kumari M, Iyengar S. Method for improving the quality of genomic DNA obtained from minute quantities of tissue and blood samples using Chelex 100 resin. Biol Proced Online. 2018;20:12. Epub 20180601. pmid:29881330; PubMed Central PMCID: PMC5984428.
  45. 45. Visvikis S, Schlenck A, Maurice M. DNA extraction and stability for epidemiological studies. Clin Chem Lab Med. 1998;36(8):551–5. pmid:9806458.