Characteristics of ciliary muscle profile in high myopes measured by swept-source anterior segment optical coherence tomography

  • Hiu Yin Kwok ,

    Contributed equally to this work with: Hiu Yin Kwok, Hoi Yee Kwok, Tsz Nok Ng

    Roles Conceptualization, Data curation, Formal analysis, Investigation, Project administration, Validation

    Affiliation School of Optometry, The Hong Kong Polytechnic University, Kowloon, Hong Kong, SAR, China

  • Hoi Yee Kwok ,

    Contributed equally to this work with: Hiu Yin Kwok, Hoi Yee Kwok, Tsz Nok Ng

    Roles Conceptualization, Data curation, Formal analysis, Investigation, Methodology

    Affiliation School of Optometry, The Hong Kong Polytechnic University, Kowloon, Hong Kong, SAR, China

  • Tsz Nok Ng ,

    Contributed equally to this work with: Hiu Yin Kwok, Hoi Yee Kwok, Tsz Nok Ng

    Roles Conceptualization, Data curation, Formal analysis, Investigation, Methodology

    Affiliation School of Optometry, The Hong Kong Polytechnic University, Kowloon, Hong Kong, SAR, China

  • Tsz-wing Leung,

    Roles Validation, Visualization, Writing – review & editing

    Affiliations School of Optometry, The Hong Kong Polytechnic University, Kowloon, Hong Kong, SAR, China, Research Centre for Sharp Vision, The Hong Kong Polytechnic University, Kowloon, Hong Kong, SAR, China, Centre for Eye and Vision Research Limited, Hong Kong, SAR, China

  • Byung Soo Kang ,

    Roles Supervision, Writing – original draft, Writing – review & editing

    bs.kang@cevr.hk

    Affiliation Centre for Eye and Vision Research Limited, Hong Kong, SAR, China

  • Chea-su Kee

    Roles Conceptualization, Funding acquisition, Project administration, Writing – review & editing

    Affiliations School of Optometry, The Hong Kong Polytechnic University, Kowloon, Hong Kong, SAR, China, Research Centre for Sharp Vision, The Hong Kong Polytechnic University, Kowloon, Hong Kong, SAR, China, Centre for Eye and Vision Research Limited, Hong Kong, SAR, China

Abstract

Objective

To characterize and compare the ciliary muscle thickness (CMT) between low and high myopes using swept-source anterior segment optical coherence tomography (AS-OCT).

Methods

Forty visually healthy young Chinese adults aged 18–25 years were divided into two groups based on refractive errors: low myopia (n = 20, spherical-equivalent refractive error (SER) between −0.50 D to −3.00 D) and high myopia (n = 20, SER ≤ -6.00 D). Cycloplegic refractions were performed before axial length (AL) and CMT were measured using a partial coherence laser interferometer and an AS-OCT respectively. CMT was measured perpendicularly to the sclera-ciliary muscle interface at 1 mm (CMT_1), 2 mm (CMT_2), and 3 mm (CMT_3) posterior to the scleral spur, and at the location with maximal thickness (CMT_MAX).

Results

High myopes demonstrated thicker CMT at 2 mm (CMT_2, p = 0.035) and 3 mm (CMT_3, p = 0.003) posterior to the scleral spur, but thinner maximal CMT (CMT_MAX, p = 0.005) than low myopes. The apical CMT_1 and CMT_MAX were also thinner in high myopes than in low myopes (both p< 0.001). CMT_MAX, apical CMT_1, and apical CMT_MAX correlated directly with SER and inversely with AL; in contrast, CMT_2 and CMT_3 showed inverse correlations with SER but direct correlations with AL.

Conclusion

Our findings revealed significant differences in CMT between low and high myopes, with high myopes showing thicker CMT at 2 mm and 3 mm posterior to the scleral spur, but thinner maximal CMT. These results provide new evidence of the potential structural differences in ciliary muscles during myopia development and progression.

Introduction

The ciliary muscle, a smooth muscle of the ciliary body, governs ocular accommodation by adjusting the shape and power of the crystalline lens through the contraction and relaxation of muscle tension [1]. A resilient ciliary muscle is crucial for maintaining clear vision across various viewing distances. Recently, there has been increased interest in ciliary muscle morphology due to its potential involvement in myopia (nearsightedness) development [24]. Myopia, a common, widespread, and progressive refractive error, is projected to affect half of the global population by 2050 [5]. Although no unified theory has fully elucidated the etiology of myopia, excessive near work has long been considered a primary risk factor [68]. In addition, myopic eyes typically exhibit a significant lag of accommodation [911], potentially resulting from inadequate ciliary muscle contraction control. The accommodation lag leads to hyperopic defocus, in which distant images are projected behind the retina. Such optical defocus could act as a visual error signal that accelerates myopic eye growth [12]. Additionally, the ciliary muscle’s anatomical proximity to the sclera suggests that biomechanical forces from ciliary muscle contractions may restructure the local arrangement of scleral collagen [13, 14], predisposing the eye to axial elongation and subsequent myopia development.

Morphology of the ciliary muscle is commonly quantified by its thickness, either as a whole [15] or at various locations posterior to the scleral spur [1618]. The association between ciliary muscle thickness (CMT) and myopia remains an active area of research, as some studies have reported that increased myopia is correlated with thicker CMT, while others found no significant association (Table 1). These discrepancies may be attributable to differences in study design, CMT measurement methods, variations in subjects’ age range, and ethnicity [1523]. The inconsistent findings underscore the necessity for further research on CMT in myopic eyes, particularly within specific population subgroups.

thumbnail
Table 1. Key findings of morphological changes of ciliary muscle in ametropia.

https://doi.org/10.1371/journal.pone.0289135.t001

Previously, anatomical changes in the ciliary muscle of ametropic eyes were measured using ultrasound biomicroscopy (UBM) [15, 17, 19]. This technology, however, requires direct contact with the ocular surface using a probe and yields relatively low-resolution images of the anterior ocular segment. Furthermore, anterior segment examinations using UBM often employ a fluid-filled scleral shell, necessitating patients to lie face-up, which may affect anterior segment geometry due to gravity [24, 25]. In contrast, anterior segment optical coherence tomography (AS-OCT), a non-invasive imaging technique with high-resolution capabilities for anterior eye structures, overcomes these limitations can better characterizes ciliary muscle morphology.

Despite the growing volume of research on ciliary muscles in myopic eyes, CMT in highly myopic eyes (–6.00 D or greater) remains underexplored. Previous studies, particularly those using AS-OCT, were limited to participants with low to moderate myopia [18, 20, 23] or included only a small number of highly myopic eyes (n < 8) [16, 17]. It is important to note that high myopia significantly increases the risk of multiple sight-threatening retinal diseases, such as glaucoma, cataract, myopic maculopathy, and retinal detachment [26]. In addition to its potential link to high myopia development, a comprehensive understanding of ciliary muscle anatomical changes holds significant clinical implications for presbyopia [2729], glaucoma [30, 31], and intraocular lens (IOL) implantation [3234]. Consequently, this study aims to characterize CMT in highly myopic eyes by comparing it with low myopes using a non-contact, high-resolution Fourier-domain swept-source AS-OCT [35].

Methods

Subjects

Forty visually healthy young Chinese adults (aged 18 to 25 years) participated in this study between March 2021 and December 2021. The restricted age range aimed to avoid the potential effects of age in either developing or ageing eyes on CMTs [29, 36]. Participants were equally divided into two groups according to their refractive errors: 1) Low myopia group (n = 20): spherical-equivalent refractive error (SER) between –0.50 D to –3.00 D, astigmatism < 1.00 D; 2) High myopia group (n = 20): SER ≤ –6.00 D, astigmatism < 2.00 D. All participants had best-corrected distance visual acuity of 20/25 or better and anisometropia ≤ 1.00 D. Exclusion criteria included ocular or systemic diseases, strabismus, and a history of receiving ocular surgery or myopia interventions. A flowchart illustrating the study design is provided in Fig 1. The study adhered to the tenets of the Declaration of Helsinki, and the study design and protocols described below were reviewed and approved by the Human Subjects Ethics Sub-committee (HSEARS20210302005) of the Hong Kong Polytechnic University. Informed written consent was obtained from each participant before the study started.

Procedures

Refractive errors and ocular biometric parameters were measured after cycloplegia. Participants underwent slit-lamp examination and non-contact tonometry (NT-530P, Nidek, Japan) measurements to determine suitability of cycloplegia administration. None of the participants had contraindications for cycloplegic agents, such as van Herick ratio < 0.25 or intraocular pressure > 21 mmHg. Gonioscopy was planned for individuals with a van Herick ratio < 0.5; however, all participants in the study had a van Herick ratio of 0.8 or larger. Two drops of 1% Tropicamide (Mydriacyl 1%, Alcon, USA), five minutes apart, were instilled to paralyze the ciliary muscle [37, 38]. This drug administration protocol has a comparable cycloplegic effect to 1% cyclopentolate [3941]. Objective refraction was performed using an autorefractor (ARK-510A, Nidek, Japan) 30 minutes after administering the first eye drop of cycloplegia. The refractive error was then confirmed with subjective refraction using the maximum plus maximum acuity as the endpoint [42]. Spherical-equivalent refractive error (SER) was calculated according to the spherical (S) and cylindrical (C) errors:

Axial length (AL) was measured from the anterior corneal apex to the retinal pigment epithelium layer using partial coherence laser interferometry (IOLMaster, Zeiss, Germany). Five consecutive readings were acquired and averaged for each participant.

Ciliary muscle thickness (CMT) measurement.

A swept-source Anterior Segment Optical Coherence Tomographer (AS-OCT, Casia SS-1000, Tomey Corporation, Japan) was used to characterize the ciliary muscle profile following the completion of cycloplegic refraction (approximately 40 minutes after the first drop of cycloplegia instilment). Participants, who had undergone cycloplegia, were instructed to fixate on an internal target located at the temporal visual field inside the instrument to enable locating and scanning the anterior segment at the nasal side (about 1 mm below the pupil center and 3 mm nasally from the limbus; Fig 2, area circled in red). The nasal CMT was chosen for measurement because the scleral spur could be more easily identified than the temporal side [43]. Images were acquired under the 2D Angle HD mode, with a sampling resolution of 2048-line A-scan per image and a cross-sectional scan range of 8 mm x 8 mm. Three consecutive images were captured and analyzed.

thumbnail
Fig 2. A representative image of nasal ciliary muscle captured by the AS-OCT.

Ciliary muscle thicknesses at the nasal side were measured at the location (red circle) 1 mm below the pupil center (yellow star) and 3 mm nasally from the limbus.

https://doi.org/10.1371/journal.pone.0289135.g002

The Casia SS-1000 AS-OCT has been previously validated and demonstrated to have excellent intra-device repeatability for measuring anterior chamber depth, angle-to-angle distance, pupil diameter, and crystalline lens rise [44]. However, the repeatability of CMT measurements has not been assessed. To confirm the repeatability of CMT measurements, an additional set of images were collected within the same day of the visit. After the first set of image acquisition, participants were asked to lean back from the device. The same examiner repeated the procedures described above to realign the eye position with AS-OCT and re-measure the CMT.

The AS-OCT images obtained were subjected to processing using a protocol similar to that employed by previous studies [18, 21, 22]. The image analysts were masked to the refractive states of the subjects to mitigate the effect of subjective measurement bias. First, the raw AS-OCT images were processed using the “Sharpen” function in ImageJ (National Health Institute, USA) [45], enhancing contrast and accentuated details to facilitate clear differentiation of the ciliary muscle-sclera boundary (Fig 3A). Subsequently, the ImageJ scale was calibrated as follows: (a) distance in pixels = 100; (b) known distance = 1; (c) unit of length = mm; (d) decimal place = 5. This calibration ensured 1:1 correspondence between the AS-OCT image and physical measurements, thereby aiding in accurate determination of CMT at a later stage. Afterwards, a non-destructive grid of lines, covering an area of 1 mm², was created using an ImageJ plugin. The AS-OCT images were rotated to optimize alignment with the scleral contour. Manual measurements of CMT were then performed perpendicular to the grid at various locations, with each measurement repeated thrice to ensure accuracy. The position of the scleral spur was first defined in the AS-OCT images by tracking the boundary between the ciliary muscle fibers and the sclera until it reaches the anterior chamber (Fig 3B) [46]. The CMT was then measured perpendicularly to the sclera-ciliary muscle interface at 1 mm (CMT_1), 2 mm (CMT_2), and 3 mm (CMT_3) posterior to the scleral spur and at the location with the maximal thickness (CMT_MAX). The thickness of the ciliary muscle’s apical portion, which is mainly composed of circular fibers (and partially radial fibers), was calculated at 1 mm posterior to the scleral spur and at the location demonstrating maximal CMT:

The CMT and apical fiber thickness obtained from the three consecutive captured images were averaged for each measurement for further analyses.

thumbnail
Fig 3. Representative images of the nasal ciliary muscle captured by AS-OCT.

(A) The raw AS-OCT image. (B) The processed AS-OCT image. The dashed line encloses the region of interest, while the solid lines indicate the CMT measured at different locations: 1 mm (CMT_1), 2 mm (CMT_2) and 3 mm (CMT_3) posterior to the scleral spur or at the point of maximal thickness (CMT_MAX). Color-shaded areas indicate individual CMT regions: blue (CMT_3), yellow (CMT_2), and pink (CMT_1 and CMT_MAX).

https://doi.org/10.1371/journal.pone.0289135.g003

Statistical analysis

All statistical analyses were performed using IBM SPSS (Version 26, IBM, Armonk, NY, USA). To evaluate intra-session repeatability of AS-OCT in CMT measurement, paired t-test with Bland-Altman analysis (95% limits of agreement: bias ± (1.96 x SD of differences)) were applied to compare the differences between the two sets of images collected from the each participant [47]. The differences in CMT between the low and high myopic groups were compared by unpaired t-tests or Mann-Whitney U-tests after verifying data normality with the Shapiro-Wilk test. Spearman’s correlation coefficient analysis was performed to test the relationship between CMTs and ocular biometric parameters (SER and AL). The significance level for all tests was set at 5%.

Results

Demographic information

Table 2 summarizes the demographic information for the low and high myopic groups. As expected, high myopes had SER −6 D and axial length 2.7 mm longer than the low myopic group (unpaired t-test, all p< 0.001). However, there were no significant differences in age (Mann-Whitney U-test, p = 0.904) and gender proportion (Chi-squared χ2 test, p = 0.204) between the groups.

thumbnail
Table 2. Demographic characteristics of low and high myopes.

https://doi.org/10.1371/journal.pone.0289135.t002

Repeatability of AS-OCT

All AS-OCT image acquisitions were repeated twice in the same visit and the intra-session repeatability in CMT measurement was satisfactory (Table 3). There were no significant differences between the two sets of CMTs measurements (paired t-tests, all p≥ 0.26), except for CMT_2 (mean difference: −6.55 μm, paired t-test, p = 0.009) and apical CMT_1(mean difference: −9.42 μm, paired t-test, p = 0.014). The Bland-Altman analysis showed no systematic trend for all regional CMTs (Fig 4).

thumbnail
Fig 4. Bland-Altman plots showing the agreement between the two sets of regional CMTs.

In each plot, the solid blue line and dotted red lines represent the mean difference and the 95% limits of agreement, respectively.

https://doi.org/10.1371/journal.pone.0289135.g004

thumbnail
Table 3. Mean differences and 95% limits of agreements (LoA) between the two sets of CMT measurements.

https://doi.org/10.1371/journal.pone.0289135.t003

Ciliary muscle thickness in high myopes

As shown in Table 4, all regional CMT readings except CMT_1 (unpaired t-test, p = 0.108) showed significant differences between low and high myopes. The high myopic group had thicker CMT at 2 mm (CMT_2) and 3 mm (CMT_3) posterior to the scleral spur by 34.74±68.50 μm (p = 0.035) and 40.47±51.89 μm (p = 0.003), but thinner maximum CMT (CMT_MAX) by 31.82±56.10 μm (p = 0.005) than the low myopic group. The apical CMT_1 and CMT_MAX were also thinner in high myopes than in low myopes, by 53.24±44.72 μm and 66.55±51.80 μm (both p< 0.001), respectively.

thumbnail
Table 4. Comparisons of CMTs between low and high myopes at different locations.

https://doi.org/10.1371/journal.pone.0289135.t004

The regional CMTs, except CMT_1 (Pearson’s r< 0.215, p> 0.08), were significantly correlated with SER and axial length (Fig 5A and 5B). CMT_MAX, apical CMT_1, and apical CMT_MAX were correlated directly with SER (all Pearson’s r≥ 0.346, p< 0.05) and inversely with axial length (all Pearson’s r≥ −0.408, p< 0.01), indicating that eyes with higher myopia and longer axial length were associated with thinner maximal ciliary muscle thickness and apical fiber thickness. In contrast, CMT_2 and CMT_3 were inversely correlated with SER (Pearson’s r = −0.340 & −0.502, p = 0.03 & 0.001) and axial length (Pearson’s r = 0.310 & 0.469, p = 0.051 & 0.002).

thumbnail
Fig 5. Correlations between CMTs and biometric parameters at different locations.

CMTs (A&B) and apical CMTs (C&D) were plotted against SER (Left) and AL (Right). Circle and cross symbols represent low and high myopes, respectively. Red: CMT_1, blue: CMT_2, green: CMT_3, black: CMT_MAX. Pearson’s r is presented. *p<0.05, **p<0.01, ***p<0.001.

https://doi.org/10.1371/journal.pone.0289135.g005

Discussion

This study characterized and compared regional ciliary muscle thicknesses between low and high myopes using a high-resolution AS-OCT. The findings revealed that compared to low myopes, the ciliary muscle in highly myopic eyes exhibited reduced thickness at the apical and maximum thickness regions (CMT_MAX, apical CMT_1, and apical CMT_MAX), and increased thickness posteriorly (CMT_2 and CMT_3) (see Table 4). An earlier study employing UBM also reported a generally thicker ciliary muscle in highly myopic eyes: however, due to the limited image resolution, the analysis focused on the ciliary muscle as a whole rather than conducting a comprehensive assessment of regional morphology [15]. Consistent with a subsequent study employing a higher resolution AS-OCT [16, 18, 2023], our results identified a thicker CMT_2 and CMT_3 in high myopes than low myopes. The significant correlations observed between CMT_2 and CMT_3 with SER and axial length confirm that an increase in ciliary muscle thickness was associated with higher myopia and increased axial length, which are generally in line with other AS-OCT studies that primarily examined low and moderate myopes (Bailey et al. [20], Sheppard and Davies (only CMT2) [16], Jeon et al. (only CMT3) [17], Pucker et al. [18], Buckhurst et al. [21], and Kaphle et al. [23]).

While myopic eyes consistently exhibit a thickening of the posterior ciliary muscle, the variations in maximum ciliary muscle thickness and apical ciliary muscle thickness in association with myopia have revealed inconsistencies across studies. In concordance with the current findings, Pucker et al. [18] and Kuchem et al. [22] found that apical CMT_1 and CMT_MAX were thinner, rather than thicker, in eyes with higher myopia. Conversely, some studies found no significant correlation between apical ciliary muscle thickness and myopia or axial length. It is essential to recognize the differences in the study populations of these studies. Pucker et al. [18] focused exclusively on children (aged 6–14 years) with low to moderate myopia, while Kuchem et al. [22]. investigated anisometropic young adults, including only eight participants with high myopia in one eye. Our research contributes new evidence that shows the apical ciliary muscle was thinner while the posterior ciliary muscle was thickened in highly bilaterally myopic young adults, compared to age- and gender-matched low myopes.

Myopic eyes consistently display morphological changes in the ciliary muscle; however, the causal relationship with myopia development remains uncertain. It was initially anticipated that ocular expansions in myopic eyes would stretch the ciliary muscle, resulting in a more elongated and thinner configuration [48]. Notably, the current findings revealed that the thinning of the ciliary muscle occurred only at the apical or maximum thickness regions, which are less likely to be subjected to stretching. In contrast, the ciliary muscle at the posterior segment was thickened, as corroborated by other studies. The consistency of these observations suggests that mechanical stretching during myopia development may not directly influence ciliary muscle morphology, particularly in the posterior segment. Alternatively, the morphological changes in the ciliary muscle of high myopes may be attributable to relocation of the ciliary muscle to a more posterior position due to axial elongation [16]. This process could result in a relatively thinner apical and thicker posterior ciliary muscle.

It has also been hypothesized that an eye with thicker posterior ciliary muscles may be more susceptible to myopia development. The thickened ciliary muscle could generate higher mechanical resistance, constraining equatorial expansion during emmetropization. The disproportionate axial elongation [49] may account for the relatively prolate posterior retinal contour frequently observed in myopic eyes [50]. Such a prolate retinal contour leads to relative hyperopic defocus in the peripheral visual field [49, 50]. The relatively peripheral hyperopic defocus, which is recognized as a visual error signal for axial myopic eye growth, may increase the risk for myopia development [12, 5153] (note, however, contradictory findings in Mutti et al. [49]). Furthermore, ciliary muscle hypertrophy may somewhat impair contractile function, leading to a more significant lag of accommodation. As a result, the hyperopic defocus during near work due to accommodative lag may also instigate and accelerate myopia development [20]. However, other studies have reported contradictory findings [9, 35, 54, 55].

To the best of our knowledge, this is the first study to have examined the regional anatomical variations in ciliary muscles of highly myopic patients by employing state-of-the-art imaging modalities. Previously applied techniques, such as UBM, exhibit commendable repeatability [15, 56]; however, the contemporary swept-source AS-OCT has garnered interest due to its capacity to generate high-resolution cross-sectional anterior segment images without touching the cornea [57]. The earlier AS-OCT model (time-domain Visante AS-OCT, Zeiss, Germany) significantly contributed to understanding the correlation between ciliary muscle thickness and myopia [21], but suffered from drawbacks, including a sluggish image acquisition rate [58]. In contrast, the Casia SS-1000 AS-OCT, a first-generation swept-source Fourier-domain OCT instrument, facilitated rapid anterior segment imaging in a single session [35]. The findings of the current study indicate that the Casia SS-1000 AS-OCT could measure ciliary muscle thickness with high repeatability (Table 3 and Fig 4). Despite a significant difference of 6.55 μm between the two sets of repeated measurement for CMT_2, this discrepancy seemingly exerts no influence on identifying substantial differences in ciliary muscle thicknesses between high and low myopes, which were > 18.51 μm (Table 4).

In conclusion, our results support the notion that ciliary muscle thicknesses display variations between low and high myopes. However, certain limitations in the current study warrant consideration for future experiments. First, although the repeatability of AS-OCT is adequate for discerning CMT differences between low and high myopes, it should be noted that the range of 95% LoA was considerable in some instances (e.g., CMT_2: −35.81 μm to 22.71 μm, Table 3). This may affect the reliability of CMT comparisons for small between-group differences (i.e., the difference in CMT between low and high myopes: 18.51 μm to 40.47 μm, see Table 4). Second, the resolution of Casia AS-OCT may introduce variability in CMT measurements. It was observed that image resolution varied across ciliary muscle regions, with the scleral spur and the ciliary muscle apex appearing more distinct in the posterior regions compared to the anterior regions. Thus, determining the region for CMT_1 and CMT_MAX measurements posed challenges. Third, this study did not assess inter-examiner repeatability. The accuracy of manual pixel coordinate selection for the scleral spur and the ciliary muscle margin could potentially influence the repeatability of CMT readings. Further evaluation of inter-session, inter-examiner, and intra-examiner repeatability is necessary. Fourth, the methodology employed for segmenting regions for CMT measurements does not account for the possibility of varying ciliary muscle length depending on refractive errors [16]. Incorporating proportional CMT measures may be required to ensure comparisons are made in similar regions of ciliary muscles. Fifth, a recent study observed alternations in CMT after administering 1% atropine [59]. However, considering the varying accommodative demands between the low and high myopic groups when fixating on the fixation target during measurements, the current study employed two drops of 1% tropicamide to minimize the potential accommodation fluctuations and mitigate their impact on the comparison of CMT between groups [2, 3]. Although the current study utilized a different cycloplegic agent, the ciliary muscle morphology presented may still not fully represent its natural state. It underscores the importance of a comprehensive evaluation of the potential influence of various cycloplegic agents on ciliary muscle morphology. Finally, this study offers only cross-sectional data on the ciliary muscle morphology in high myopes with a limited sample size derived from a Chinese population. Thus, cautions should be paid when extrapolating these findings to other highly myopic populations. Longitudinal studies are also warranted to investigate the mechanism and role of ciliary muscles in myopia development.

Acknowledgments

Authors acknowledge Dr Maureen Boost for her constructive advice in the preparation of the manuscript.

References

  1. 1. Remington LA. Chapter 6 ‐ Aqueous and Vitreous Chambers. In: Remington LA, editor. Clinical Anatomy of the Visual System (Second Edition). Saint Louis: Butterworth-Heinemann; 2005. p. 103–15.
  2. 2. Mutti DO. Hereditary and environmental contributions to emmetropization and myopia. Optometry and vision science: official publication of the American Academy of Optometry. 2010;87(4):255–9. pmid:20019643
  3. 3. Chen JC, Schmid KL, Brown B. The autonomic control of accommodation and implications for human myopia development: a review. Ophthalmic & physiological optics: the journal of the British College of Ophthalmic Opticians (Optometrists). 2003;23(5):401–22. pmid:12950887
  4. 4. Morgan IG, Rose KA. Myopia and international educational performance. Ophthalmic & physiological optics: the journal of the British College of Ophthalmic Opticians (Optometrists). 2013;33(3):329–38. pmid:23662964
  5. 5. Holden BA, Fricke TR, Wilson DA, Jong M, Naidoo KS, Sankaridurg P, et al. Global Prevalence of Myopia and High Myopia and Temporal Trends from 2000 through 2050. Ophthalmology. 2016;123(5):1036–42. pmid:26875007
  6. 6. Saw SM, Chua WH, Hong CY, Wu HM, Chan WY, Chia KS, et al. Nearwork in early-onset myopia. Investigative ophthalmology & visual science. 2002;43(2):332–9. pmid:11818374
  7. 7. Huang H-M, Chang DS-T, Wu P-C. The association between near work activities and myopia in children—a systematic review and meta-analysis. PloS one. 2015;10(10):e0140419. pmid:26485393
  8. 8. Ip JM, Saw SM, Rose KA, Morgan IG, Kifley A, Wang JJ, et al. Role of near work in myopia: findings in a sample of Australian school children. Investigative ophthalmology & visual science. 2008;49(7):2903–10.
  9. 9. Mutti DO, Mitchell GL, Hayes JR, Jones LA, Moeschberger ML, Cotter SA, et al. Accommodative Lag before and after the Onset of Myopia. Investigative ophthalmology & visual science. 2006;47(3):837–46. pmid:16505015
  10. 10. Kaphle D, Varnas SR, Schmid KL, Suheimat M, Leube A, Atchison DA. Accommodation lags are higher in myopia than in emmetropia: Measurement methods and metrics matter. Ophthalmic and Physiological Optics. 2022;42(5):1103–14. pmid:35775299
  11. 11. Gwiazda J, Thorn F, Bauer J, Held R. Myopic children show insufficient accommodative response to blur. Investigative ophthalmology & visual science. 1993;34(3):690–4. pmid:8449687
  12. 12. Schaeffel F, Glasser A, Howland HC. Accommodation, refractive error and eye growth in chickens. Vision Research. 1988;28(5):639–57. pmid:3195068
  13. 13. Summers Rada JA, Shelton S, Norton TT. The sclera and myopia. Experimental eye research. 2006;82(2):185–200. pmid:16202407
  14. 14. Harper AR, Summers JA. The dynamic sclera: extracellular matrix remodeling in normal ocular growth and myopia development. Experimental eye research. 2015;133:100–11. pmid:25819458
  15. 15. Muftuoglu O, Hosal BM, Zilelioglu G. Ciliary body thickness in unilateral high axial myopia. Eye (Lond). 2009;23(5):1176–81. pmid:18551140
  16. 16. Sheppard AL, Davies LN. In vivo analysis of ciliary muscle morphologic changes with accommodation and axial ametropia. Invest Ophthalmol Vis Sci. 2010;51(12):6882–9. pmid:20671285
  17. 17. Jeon S, Lee WK, Lee K, Moon NJ. Diminished ciliary muscle movement on accommodation in myopia. Exp Eye Res. 2012;105:9–14. pmid:23068564
  18. 18. Pucker AD, Sinnott LT, Kao CY, Bailey MD. Region-Specific Relationships Between Refractive Error and Ciliary Muscle Thickness in Children. Invest Ophth Vis Sci. 2013;54(7):4710–6. pmid:23761093
  19. 19. Oliveira C, Tello C, Liebmann JM, Ritch R. Ciliary body thickness increases with increasing axial myopia. American Journal of Ophthalmology. 2005;140(2):324–5. pmid:16086961
  20. 20. Bailey MD, Sinnott LT, Mutti DO. Ciliary body thickness and refractive error in children. Invest Ophthalmol Vis Sci. 2008;49(10):4353–60. pmid:18566470
  21. 21. Buckhurst H, Gilmartin B, Cubbidge RP, Nagra M, Logan NS. Ocular biometric correlates of ciliary muscle thickness in human myopia. Ophthal Physl Opt. 2013;33(3):294–304. pmid:23452002
  22. 22. Kuchem MK, Sinnott LT, Kao CY, Bailey MD. Ciliary Muscle Thickness in Anisometropia. Optometry Vision Sci. 2013;90(11):1312–20. pmid:24100479
  23. 23. Kaphle D, Schmid KL, Davies LN, Suheimat M, Atchison DA. Ciliary Muscle Dimension Changes With Accommodation Vary in Myopia and Emmetropia. Invest Ophth Vis Sci. 2022;63(6):24-. pmid:35749128
  24. 24. Rosen DB, Conway MD, Ingram CP, Ross RD, Montilla LG, Peyman GAJNDMiO. A brief overview of ophthalmic ultrasound imaging. 2019.
  25. 25. Wagner S, Zrenner E, Strasser T. Ciliary muscle thickness profiles derived from optical coherence tomography images. Biomed Opt Express. 2018;9(10):5100–14. pmid:30319924
  26. 26. Ohno-Matsui K, Wu PC, Yamashiro K, Vutipongsatorn K, Fang Y, Cheung CMG, et al. IMI Pathologic Myopia. Investigative ophthalmology & visual science. 2021;62(5):5. pmid:33909033
  27. 27. Ruggeri M, de Freitas C, Williams S, Hernandez VM, Cabot F, Yesilirmak N, et al. Quantification of the ciliary muscle and crystalline lens interaction during accommodation with synchronous OCT imaging. Biomed Opt Express. 2016;7(4):1351–64. pmid:27446660
  28. 28. Pardue MT, Sivak JG. Age-related changes in human ciliary muscle. Optom Vis Sci. 2000;77(4):204–10. pmid:10795804
  29. 29. Sheppard AL, Davies LN. The Effect of Ageing on In Vivo Human Ciliary Muscle Morphology and Contractility. Invest Ophth Vis Sci. 2011;52(3):1809–16. pmid:21071738
  30. 30. Gohdo T, Tsumura T, Iijima H, Kashiwagi K, Tsukahara S. Ultrasound biomicroscopic study of ciliary body thickness in eyes with narrow angles. Am J Ophthalmol. 2000;129(3):342–6. pmid:10704550
  31. 31. Ku JY, Nongpiur ME, Park J, Narayanaswamy AK, Perera SA, Tun TA, et al. Qualitative evaluation of the iris and ciliary body by ultrasound biomicroscopy in subjects with angle closure. J Glaucoma. 2014;23(9):583–8. pmid:23429617
  32. 32. Wang L, Wolfe P, Chernosky A, Paliwal S, Tjia K, Lane S. In vitro delivery performance assessment of a new preloaded intraocular lens delivery system. J Cataract Refract Surg. 2016;42(12):1814–20.
  33. 33. Sugiura T, Kaji Y, Tanaka Y. Anatomy of the ciliary sulcus and the optimum site of needle passage for intraocular lens suture fixation in the living eye. J Cataract Refract Surg. 2018;44(10):1247–53. pmid:30172566
  34. 34. Okamoto Y, Okamoto F, Nakano S, Oshika T. Morphometric assessment of normal human ciliary body using ultrasound biomicroscopy. Graef Arch Clin Exp. 2017;255(12):2437–42. pmid:28971252
  35. 35. Wagner S, Schaeffel F, Zrenner E, Strasser T. Prolonged nearwork affects the ciliary muscle morphology. Exp Eye Res. 2019;186.
  36. 36. Strenk SA, Semmlow JL, Strenk LM, Munoz P, Gronlund-Jacob J, DeMarco JK. Age-related changes in human ciliary muscle and lens: A magnetic resonance imaging study. Invest Ophth Vis Sci. 1999;40(6):1162–9.
  37. 37. Egashira SM, Kish LL, Twelker JD, Mutti DO, Zadnik K, Adams AJ. Comparison of Cyclopentolate Versus Tropicamide Cycloplegia in Children. Optometry Vision Sci. 1993;70(12):1019–26. pmid:8115124
  38. 38. Mutti DO, Zadnik K, Egashira S, Kish L, Twelker JD, Adams AJ. The Effect of Cycloplegia on Measurement of the Ocular Components. Invest Ophth Vis Sci. 1994;35(2):515–27. pmid:8113002
  39. 39. Hofmeister EM, Kaupp SE, Schallhorn SC. Comparison of tropicamide and cyclopentolate for cycloplegic refractions in myopic adult refractive surgery patients. J Cataract Refr Surg. 2005;31(4):694–700. pmid:15899444
  40. 40. Yazdani N, Sadeghi R, Momeni-Moghaddam H, Zarifmahmoudi L, Ehsaei A. Comparison of cyclopentolate versus tropicamide cycloplegia: A systematic review and meta-analysis. J Optom. 2018;11(3):135–43. pmid:29132914
  41. 41. Lin LL, Shih YF, Hsiao CH, Su TC, Chen CJ, Hung PT. The cycloplegic effects of cyclopentolate and tropicamide on myopic children. Journal of ocular pharmacology and therapeutics: the official journal of the Association for Ocular Pharmacology and Therapeutics. 1998;14(4):331–5. pmid:9715436
  42. 42. Grosvenor TP. Primary care optometry: Elsevier Health Sciences; 2007.
  43. 43. Sakata LM, Lavanya R, Friedman DS, Aung HT, Seah SK, Foster PJ, et al. Assessment of the scleral spur in anterior segment optical coherence tomography images. Archives of ophthalmology. 2008;126(2):181–5. pmid:18268207
  44. 44. Zhang XF, Li M, Shi Y, Wan XH, Wang HZ. Repeatability and agreement of two anterior segment OCT in myopic patients before implantable collamer lenses implantation. Int J Ophthalmol-Chi. 2020;13(4):625–31. pmid:32399415
  45. 45. Schneider CA, Rasband WS, Eliceiri KW. NIH Image to ImageJ: 25 years of image analysis. Nature Methods. 2012;9(7):671–5. pmid:22930834
  46. 46. Li M, Luo Z, Yan X, Zhang H. Diagnostic power of scleral spur length in primary open-angle glaucoma. Graefes Arch Clin Exp Ophthalmol. 2020;258(6):1253–60. pmid:32146558
  47. 47. Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Int J Nurs Stud. 2010;47(8):931–6.
  48. 48. Vanalphen GWHM. Choroidal Stress and Emmetropization. Vision Res. 1986;26(5):723-&. pmid:3750852
  49. 49. Mutti DO, Hayes JR, Mitchell GL, Jones LA, Moeschberger ML, Cotter SA, et al. Refractive error, axial length, and relative peripheral refractive error before and after the onset of myopia. Invest Ophth Vis Sci. 2007;48(6):2510–9.
  50. 50. Atchison DA, Pritchard N, Schmid KL, Scott DH, Jones CE, Pope JM. Shape of the retinal surface in emmetropia and myopia. Invest Ophth Vis Sci. 2005;46(8):2698–707. pmid:16043841
  51. 51. Charman WN, Radhakrishnan H. Peripheral refraction and the development of refractive error: a review. Ophthal Physl Opt. 2010;30(4):321–38. pmid:20629956
  52. 52. Smith EL, Kee CS, Ramamirtham R, Qiao-Grider Y, Hung LF. Peripheral vision can influence eye growth and refractive development in infant monkeys. Invest Ophth Vis Sci. 2005;46(11):3965–72. pmid:16249469
  53. 53. Smith EL, Hung LF, Huang J. Relative peripheral hyperopic defocus alters central refractive development in infant monkeys. Vision Res. 2009;49(19):2386–92. pmid:19632261
  54. 54. Gwiazda J, Thorn F, Held R. Accommodation, accommodative convergence, and response AC/A ratios before and at the onset of myopia in children. Optometry Vision Sci. 2005;82(4):273–8.
  55. 55. Weizhong L, Zhikuan Y, Wen L, Xiang C, Jian G. A longitudinal study on the relationship between myopia development and near accommodation lag in myopic children. Ophthalmic & physiological optics: the journal of the British College of Ophthalmic Opticians (Optometrists). 2008;28(1):57–61. pmid:18201336
  56. 56. Li J, Drechsler J, Lin AN, Widlus M, Qureshi A, Stoleru G, et al. Repeatability and Reliability of Quantified Ultrasound Biomicroscopy Image Analysis of the Ciliary Body at the Pars Plicata. Ultrasound Med Biol. 2021;47(7):1949–56. pmid:33858721
  57. 57. Ang M, Baskaran M, Werkmeister RM, Chua J, Schmidl D, dos Santos VA, et al. Anterior segment optical coherence tomography. Prog Retin Eye Res. 2018;66:132–56. pmid:29635068