Abstract
Background
High-flow nasal oxygenation is increasingly used during sedation procedures and general anesthesia in apneic patients. Transcutaneous CO2 (ptcCO2)-monitoring is used to monitor hypercapnia. This study investigated ptcCO2-monitoring during apneic oxygenation.
Methods
We included 100 patients scheduled for elective surgery under general anesthesia in this secondary analysis of a randomized controlled trial. Before surgery, we collected ptcCO2 measured by TCM4 and TCM5 monitors and arterial blood gas (ABG) measurements every two minutes during 15 minutes of apnea. Bland-Altman plots analyzed agreement between measurement slopes; linear mixed models estimated the different measuring method effect, and outlined differences in slope and offset between transcutaneous and arterial CO2 partial pressures.
Results
Bland-Altman plots showed a bias in slope (95% confidence intervals) between ABG and TCM4-measurements of 0.05mmHg/min (-0.05 to 0.15), and limits of agreement were -0.88mmHg/min (-1.06 to -0.70) and 0.98mmHg/min (0.81 to 1.16). Bias between ABG and TCM5 was -0.14mmHg/min (-0.23 to -0.04), and limits of agreement were -0.98mmHg/min (-1.14 to -0.83) and 0.71mmHg/min (0.55 to 0.87). A linear mixed model (predicting the CO2-values) showed an offset between arterial and transcutaneous measurements of TCM4 (-15.2mmHg, 95%CI: -16.3 to -14.2) and TCM5 (-19.1mmHg, -20.1 to -18.0). Differences between the two transcutaneous measurements were statistically significant.
Conclusions
Substantial differences were found between the two transcutaneous measurement systems, and between them and ABG. Transcutaneous CO2 monitoring cannot replace arterial CO2-monitoring during apneic oxygenation. In clinical settings with rapidly changing CO2-values, arterial blood gas measurements are needed to reliably assess the CO2-partial pressure in blood.
Citation: Schweizer T, Hartwich V, Riva T, Kaiser H, Theiler L, Greif R, et al. (2023) Limitations of transcutaneous carbon dioxide monitoring in apneic oxygenation. PLoS ONE 18(6): e0286038. https://doi.org/10.1371/journal.pone.0286038
Editor: Steven E. Wolf, University of Texas Medical Branch at Galveston, UNITED STATES
Received: September 8, 2022; Accepted: January 24, 2023; Published: June 1, 2023
Copyright: © 2023 Schweizer et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Data Availability: The Swiss Human Research Act required for any new study on existing data a new ethic committee approval, therefore data sharing is only possible with such an approval. https://www.swissethics.ch/ Cantonal Ethics Committee Bern, Murtenstrasse 31, Hörsaaltrakt Pathologie, Eingang 43A, Büro H372, 3010 Bern, info.kek@be.ch, https://www.gsi.be.ch/de/start/ueber-uns/kommissionen-gsi/ethikkommission.html Direction of Division of Research, Department of Anaesthesiology and Pina Medicine, Bern University Hospital, Bern, Switzerland, anaesthesie@insel.ch http://www.anaesthesiologie.insel.ch/de/; phone: +41 31 632 39 65.
Funding: This study was supported only by an institutional research grant assigned to Dr. Lorenz Theiler (KAS THLD 1-18) of the Department of Anaesthesiology and Pain Medicine, Bern University Hospital, Bern, Switzerland. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript. There was no additional external funding received for this study.
Competing interests: The authors have declared that no competing interests exist.
Introduction
High-flow nasal oxygenation (HFNO) delivers heated, humidified oxygen via nasal cannulas at high rates of gas flow [1]. It is used for procedural interventions during spontaneous ventilation and during apneic oxygenation under general anesthesia [2–4]. The factor limiting the use of HFNO is the increase in carbon dioxide (CO2), not the drop in blood oxygen level [5, 6]. During apnea, oxygenation is maintained by a constant influx of 100% oxygen into the lungs, but elimination of CO2 is insufficient [7]. While monitoring of end-tidal carbon dioxide (petCO2) from the expiratory arm of the breathing circuit via capnography is standard care during anesthesia in ventilated patients, it cannot be used during HFNO in apneic patients due to the open system.
Following its successful implementation in neonatology, pediatrics, pulmonology and sleep medicine, transcutaneous CO2 (ptcCO2) monitoring is increasingly used in adult anesthesia and critical care [8–10]. Using transcutaneous CO2 for non-invasive monitoring of carbon dioxide seems suitable in apneic patients receiving HFNO [11]. However, there is not yet agreement how accurate these measurements in reality are, as ptcCO2 and petCO2 did not correlate in one out of three patients [12].
Currently, there is lack of evidence about the quality of ptcCO2 measurements in anesthetized apneic patients with rapidly changing arterial partial pressures of CO2 (paCO2).
Outcomes
This secondary analysis of a randomized controlled trial published before [13, 14] tested the hypothesis that the slope of ptcCO2 measured with two similar transcutaneous ptcCO2 measurement monitors adequately reflects the arterial CO2 ascent during nasal oxygen for apneic oxygenation under general anesthesia. Therefore, the primary outcome was the transcutaneous CO2 measurements obtained from the transcutaneous CO2 monitors TCM4 and TCM5.
Methods
The Cantonal Ethics Committee of Bern approved the study (2018–00293), which was registered at ClinicalTrials.gov (NCT03478774) and the Swiss Trial Registry KOFAM (SNCTP000002861).
Trial design
This manuscript reports a side protocol of a randomized controlled non-inferiority trial and was carried out according to the STROBE statement [13, 14].
Participants, eligibility criteria and settings
The detailed study intervention and inclusion and exclusion criteria were published previously [14]. Operating room lists were screened for eligible patients daily. Included were adult patients younger than 80 years with an American Society of Anesthesiologists (ASA) physical status of I to III. Excluded were patients with expected difficult mask ventilation, obstructive sleep apnea syndrome requiring therapy, necessity of rapid sequence induction due to aspiration risk, known chronic obstructive pulmonary disease GOLD classification 2 or higher, cervical spine instability, the need for flexible optic intubation, nasal obstruction, pregnancy, known coronary heart disease, arrythmias requiring therapy, known heart failure higher than New York Heart Association classification I, treatment with beta-receptor antagonists, known stenosis of the carotid or vertebral arteries, peripheral arterial disease with a Fontaine classification higher than 2b, hyperkalemia (potassium level higher than 5.5 mmol/l), pulmonary arterial hypertension with an pressure higher than 35mmHg, body mass index of less than 16 kg/m2 or more than 35 kg/m2, known increased intracranial pressure, patients scheduled for intracranial surgery, anemia with hemoglobin less than 100g/l, known muscular disorders and allergies or contraindications to the anesthesia agents used in the study [13, 14].
Written informed consent was obtained from all study participants, and the study was conducted according to the guidelines of the Declaration of Helsinki. Data was obtained at the Bern University Hospital in Bern, Switzerland, as part of an investigation into a possible ventilatory effect of different flow rates of trans nasal oxygen delivery between March 2018 and December 2019.
Interventions
Anesthetized patients under neuromuscular blockade were evaluated during apnea over a 15-minute period before elective surgery. During this study, CO2 partial pressure was measured continuously via transcutaneous monitors and in arterial blood gases every 2 minutes [14].
We used TCM4 and TCM5, two transcutaneous CO2 monitoring devices produced by the same manufacturer (Radiometer, Krefeld, Germany). Both are using the same measurement algorithm. The TCM sensors were placed as recommended by the manufacturer in the sub clavicular area on the patients’ chest. The probe temperature of the TCM4 was set at 44°C for all measurements. The probe temperature of the TCM5 was set at 44°C for the first 39 patients and at 42°C in the 57 following patients to compare a possible effect of temperature on the measurement accuracy. With both monitors, the Sensor 84 (Radiometer, Krefeld, Germany) was used to determine O2 and CO2 partial pressures at the same time. To differentiate between the combined O2/CO2 sensor and a CO2-only sensor, we used the Sensor 54 (Radiometer, Krefeld, Germany) with the TCM5 in five patients as a control.
The gold standard reference method for measuring CO2 partial pressure is arterial blood gas (ABG) analysis [15]. Arterial blood samples were drawn every 2 minutes using a safePICO syringe (Radiometer, Krefeld, Germany) and were analyzed with an ABL800 flex blood gas analyser (Radiometer, Krefeld, Germany) according to the manufacturer’s manual at the certified central laboratory of the Bern University Hospital.
After arrival of the patient in the operating theatre, the probes of the two TCM monitors were applied on the patient’s chest and an arterial Flowswitch cannula (BD, Franklin Lakes, USA) was placed ultrasound-guided into the radial artery to measure blood pressure and to draw arterial blood samples. Induction and maintenance of general anesthesia were standardized as described earlier [14]. After induction of anesthesia with neuromuscular blockade and successful bag-mask ventilation, apneic oxygenation was commenced as randomized. Four groups received oxygen via nasal cannulas. One group received 2 l/min O2 via standard nasal cannula and the airway was continuously kept patent applying jaw thrust. Another group received 10 l/min O2 via standard nasal cannula plus jaw thrust. Two groups received 70 l/min O2 via a high-flow nasal cannula, the airway was kept patent in one of them with jaw trust, and in the other group with a video laryngoscope. The fifth group received 0.25 l/min O2 via a tracheal tube, therefore, we did not include this group into the current analysis, as only apneic oxygenation via nasal cannula was within the scope of this sub-study [14]. Absence of muscular twitch during electrical stimulation showed complete neuromuscular blockade (TOF-Watch; Organon Ltd, Dublin, Ireland). Absence of diaphragmatic movements was assessed with continuous electrical impedance tomography (PulmoVista 500; Draeger, Luebeck, Germany). Norepinephrine was used to maintain normotension, which was defined as values within 20% of pre-operative values measured on the ward before surgery.
Before start of anesthesia, and again from the start of apneic oxygenation, arterial blood gas samples were drawn every two minutes and ptcCO2 values were recorded every minute. After completion of the study, airway management and surgery were performed as planned.
Sample size
Sample size calculation was performed for the main study, where a difference between paCO2 group means of 0.3mmHg/min was assumed as clinically relevant, using a two-sample t-test, assuming a non-inferiority margin of 0.3, a common standard deviation of 0.35, a power of 80% and a one-sided alpha of 0.025. Twenty-five patients were found to be necessary per group (including 3 patients per group as a safety margin) [14].
Randomization
Patients were stratified according to body mass index and smoking status, then randomly allocated to one of the five study groups. Computer-generated randomization (www.randomisation.com) was used by the research staff of the department to allocate study participants to the respective groups and was kept in sealed opaque envelopes, which were opened after induction of anesthesia and successful bag-mask ventilation by a study nurse not involved in the randomization process.
Blinding
As patients were under anaesthesia they were blinded to their allocated group, but due to the different delivery of oxygen and the different procedures to keep the airway open blinding the study staff was not possible [14].
Interim analysis and stopping rules
There was no interim analysis planned and performed. The predefined clinical stopping rules for study patients were SpO2 less than 92%, transcutaneous carbon dioxide greater than 100 mmHg, pH less than 7.1, potassium greater than 6 mmol l–1, or an apneic period of 15 minutes. Reaching one of these criteria terminated the apnea, and bag-mask ventilation was started.
Statistical analysis
All recorded study data were transferred into an electronic database (REDCap, Research Electronic Data Capture, Vanderbilt University, 2004). There were no differences of increase in paCO2 detected between flow rate groups in the main study, therefore we analysed the groups together [14] We defined filter criteria for the data to transparently exclude technical errors and other biases. For example, an averaged ptcCO2 increase of less than one mmHg/min during the apneic period would reflect insufficient attachment of the transcutaneous sensor or lack of a proper seal against CO2 volatilization into room air (for transparency, S1 Appendix displays statistics and graphs based on the unfiltered raw data). In the first minutes of apnea, paCO2 increases faster than afterwards, because CO2 in arterial and venous blood is equilibrating. As a result, measurements of the first three apneic minutes were excluded from analysis (Fig 1) [16]. According to the literature, an increase in CO2 of more than four mmHg/min after the first minutes of apnea in anesthetized patients under neuromuscular blockade is unlikely, therefore we excluded higher values [7, 16]. We also excluded transcutaneous measured values if they were lower than the measurements one minute before, or ABG measurements if they were lower or equal to the previous results. Such lower readings are not plausible during apneic oxygenation and were considered to be technical errors or sampling dilution.
We used Bland-Altmann plots (presenting mean slopes, mean difference in slopes and 95% limits of agreement) to test our hypothesis, the agreement between the increase of paCO2 and ptcCO2 over time [17]. Linear mixed models were fitted to the data in order to estimate the effect of the measuring method on CO2 values and to outline differences in slope and offset between the two transcutaneous monitors. Method, time, and their interaction were used as fixed covariates, together with random intercepts and slopes for patients. The models were fitted with restricted maximum likelihood and 95% confidence intervals (CI), and p-values were calculated using Satterthwaite’s approximation for the degrees of freedom. Normality of residuals and random effects were assessed visually using Q-Q-plots; variance homogeneity using residuals-vs-fitted plots. Subgroup analysis was performed for sensor temperature.
Results are presented as numbers (%) or mean ± SD. A p-value <0.05 was considered statistically significant. Analyses were performed using Stata version 16.1 (StataCorp LT, Texas, USA) or R version 4.0.3. (R Core Team (2020), R Foundation for Statistical Computing, Vienna, Austria).
Results
The CONSORT flow diagram is displayed as Fig 1. One hundred twenty-five patients were enrolled in the main study, 100 received nasal oxygen and were included in this secondary analysis [14]. Measurements of four patients were excluded due to major technical problems. (Fig 1). Patients’ characteristics are displayed in Table 1.
Fig 2 depicts the graphical analysis of all CO2 measurements during the observation period.
Fig 3 shows the Bland-Altman plots of the total slopes (i.e., the change per minute in CO2 from first to last measurement) of the two transcutaneous measurements compared to the arterial measurements. Bias in slopes (95% CI) between the ABG measurements and the TCM4 measurements were 0.05 mmHg/min (-0.05 to 0.15), and limits of agreement were -0.88 mmHg/min (-1.06 to -0.70) and 0.98 mmHg/min (0.81 to 1.16). Bias between ABG measurements and TCM5 measurements were -0.14 mmHg/min (-0.23 to -0.04), and limits of agreement were -0.98 mmHg/min (-1.14 to -0.83) and 0.71 mmHg/min (0.55 to 0.87).
A positive difference indicates a steeper slope for the transcutaneous measurement. The colored area indicates 95% confidence intervals for bias (grey) and limits of agreement (red, blue).
Table 2. shows the model coefficients from a linear mixed model for CO2. The offset between ABG and TCM4, ABG and TCM5, and TCM5 and TCM4 was significant, as were the differences in slope between ABG and TCM5, and TCM5 and TCM4 (Table 2). This is also displayed in Figs 4 and 5 shows box plots of CO2 measurements every minute using the two transcutaneous monitors and ABG.
The offset between ABG and TCM4, ABG and TCM5, and TCM5 and TCM4 was significant, as was the difference in slope between ABG and TCM5, and TCM5 and TCM4.
Boxes indicate lower to upper quartiles and whiskers show the most extreme point within 1.5 times the interquartile range from the upper and lower quartile, respectively. Points beyond that range are indicated with circles.
The offset of the measurements with the Sensor 54 (CO2-only sensor) to the ABG as reference method was -10.0mmHg (-12.2 to -7.7, p<0.001). The mean difference in slope for ABG measurements was 0.20 mmHg/min (-0.02 to 0.43, p = 0.08). An interaction model showed a difference between TCM5 measurements with the Sensor 84 (combined O2/CO2 sensor) to the measurements with the Sensor 54 of 9.3mmHg (6.2 to 12.5) in offset and 0.39 mmHg/min (0.07 to 0.71) in slope.
Discussion
We found substantial differences between transcutaneous and arterial CO2 measurements, as well as between the two transcutaneous monitors investigated. From a clinical perspective, the slope of increase in paCO2 can be regarded as linear after the initial fast increase in anesthetized patients with complete neuromuscular blockade.
So far, the dynamics of the CO2 increase in anesthetized, paralyzed, and apneic, but oxygenated humans are not entirely understood. Well documented is the initial fast increase in paCO2 in the first minute, due to the equilibration of arterial and venous pCO2 as a result of hyperoxic apnea when minimal CO2 is exhaled [18]. The subsequent increase is less clear. Research showed that the total increase of paCO2 in patients with a blocked airway can best be described as a logarithmic function, but also simplified as a piecewise linear model separating the first minute of apnea from the time following it [19]. Researchers demonstrated in five patients that the initial fast increase is followed by an almost linear increase in CO2 [20]. Figs 2 and 5 support the assumption that the increase of paCO2 after the equilibration of arterial and venous pCO2 may be regarded as linear within the apneic period of 15 minutes. Of note, we ensured maintenance of arterial blood pressure, and the deep anesthetic state possibly counteracted sympathetic stimulation caused by rising CO2 levels. Previous research showed a larger flattening, which could be caused by the diminishing neuromuscular blockade over time, with consecutive small movements of the diaphragm and thereby a ventilatory effect [21].
Although the bias shown in the Bland-Altman plots in Fig 3 lies close to 0, limits of agreements differ substantially. Previous studies defined a maximum difference of 7.5 mmHg between pCO2 measurements as clinically acceptable in a steady-state measurement [22, 23]. If this maximum difference is also acknowledged in measurements with rising pCO2 after 15 minutes (regardless of the offset), limits of agreement in slope have to be within 7.5mmHg/15min, which is equal to ±0.5mmHg/min, and this seems to be clinically acceptable. The larger levels of agreement and the number of measurements outside this range indicate that the measurement techniques cannot be used interchangeably in this population and setting (Fig 3). Further research is required to understand why in some patients the slope led to a difference of more than 7.5mmHg over 15min.
Researchers tested the accuracy of ptcCO2 in patients admitted to the intensive care unit and found mean ptcCO2 to be 2.2mmHg higher than paCO2 (limits of agreement: -9.2 to 13.6mmHg) [24]. In contrast, in our standardized anesthesia setting, we found a far greater difference in offset at time 0 (-15.2 mmHg between ABG and TCM4, and -19.1 mmHg between ABG and TCM5). One hypothesis for the reasons for these differences is that the sensors provided by the manufacturers were probably very new in these studies [9, 22–24]. Response time of transcutaneous CO2 electrodes lies usually between 20-80s, but with increasing age, it slows down over time [25, 26]. This might play a more important role during the monitoring of CO2 in apneic oxygenation with fast-changing CO2 values rather than in steady-state measurements in the intensive care setting [27]. Unfortunately, manufacturers do not provide information on the sensors’ time-function association.
Reliability can be limited by some technical pitfalls that can lead to inaccurate measurements: trapped air bubbles between the skin and the sensor, incorrect sensor placement, and incorrect sensor maintenance [15]. In addition, a propofol-induced peripheral vasodilatation-mediated drop in core body temperature might lead to hypothermia which could impair trending ability [28].
It is not completely clear why the CO2-only sensor (sensor 54) showed a smaller offset to the ABGA measurements than the combined O2/CO2 sensor (sensor 84). Due to the small number of measurements performed with the sensor 54, this result should be re-evaluated in future studies.
In clinical practice, it might be useful to compare transcutaneous CO2 values with an arterial CO2 measurement at the beginning of apnea in order to determine the offset. Subsequent arterial CO2-measurements might then be used to determine the slope, which would improve the reliability on which clinical decisions could be made. Therefore, we recommend an arterial blood gas analysis at the beginning of apnea, and further blood gas analysis every 10 minutes later to allow for better therapeutic decisions under apneic oxygenation.
In healthy patients without pre-existing diseases, transcutaneous CO2 measurements might be suitable for CO2 trend monitoring. In those patients, comparison with petCO2 prior to apnea might be sufficient. However, arterial blood gas analysis should be performed if there is uncertainty about the presence of hypercarbia to ensure patient safety.
Potential limitations of our study are the single-centre study design and the fact that all measurements were made in apneic patients of our specific patient population (compare inclusion/exclusion criteria). Further studies of other users in different settings and including different patient populations are needed to confirm and generalize our observations and their clinical implications. As our observation period was only 15 minutes, we cannot make conclusive statements for longer periods of time and if these devices might be used as trend monitors. Approximately 7.3% of data measured after the first minutes had to be excluded from the analysis (based on previously published data) to verify correct measurements, e.g., a rise of CO2 between 1-4mmHg/min [7]. We cannot comment on why this happened and if more reliable CO2 measurements would have provided different results.
Conclusions
In conclusion, transcutaneous CO2 monitoring cannot replace arterial CO2 measurements during apneic oxygenation as commercially available transcutaneous CO2 monitors showed inconsistent deviations from the gold standard, which remains arterial blood gas analysis. The offset between arterial CO2 partial pressure and the transcutaneous CO2 partial pressure measurements, as well as the increase in the CO2 slopes varied significantly over time.
Supporting information
S1 Checklist. CONSORT 2010 checklist of information to include when reporting a randomised trial*.
https://doi.org/10.1371/journal.pone.0286038.s001
(DOC)
S1 Appendix. Statistics and graphs based on the unfiltered data.
https://doi.org/10.1371/journal.pone.0286038.s002
(PDF)
Acknowledgments
The authors would like to thank Maren Loosli, anesthesia study nurse; the undergraduate students Thora Ottenhausen, Carl Conrad, Sophie Hunger, Julian Meyer, Julian Lennertz, Paulina Kaluza, and Mirco Lareida; as well as the personnel of the Department of Anaesthesiology and Pain Medicine at the Bern University Hospital in Switzerland for their help in conducting the study. We also want to thank Jeannie Wurz for review of the English in the manuscript.
References
- 1. Riva T, Meyer J, Theiler L, Obrist D, Bütikofer L, Greif R, et al. Measurement of airway pressure during high-flow nasal therapy in apnoeic oxygenation: a randomised controlled crossover trial*. Anaesthesia. 2020;(November 2019):27–35. pmid:32776518
- 2. Lau J, Loizou P, Riffat F, Stokan M, Palme CE. The use of THRIVE in otolaryngology: our experiences in two Australian tertiary facilities. Australian Journal of Otolaryngology. 2019;2(1):22–22.
- 3. Huang L, Athanasiadis T, Woods C, Dharmawardana N, Ooi EH. The use of transnasal humidified rapid insufflation ventilatory exchange in laryngeal and pharyngeal surgery: Flinders case series. Australian Journal of Otolaryngology. 2019;2:17–17.
- 4. Lyons C, Callaghan M. Apnoeic oxygenation with high-flow nasal oxygen for laryngeal surgery: a case series. Anaesthesia. 2017;72(11):1379–87. pmid:29047136
- 5. Gustafsson IM, Lodenius , Tunelli J, Ullman J, Jonsson Fagerlund M. Apnoeic oxygenation in adults under general anaesthesia using Transnasal Humidified Rapid-Insufflation Ventilatory Exchange (THRIVE)- A physiological study. Br J Anaesth [Internet]. 2017;118(4):610–7. Available from: pmid:28403407
- 6. Riva T, Theiler L, Jaquet Y, Giger R, Nisa L. Early experience with high-flow nasal oxygen therapy (HFNOT) in pediatric endoscopic airway surgery. Int J Pediatr Otorhinolaryngol [Internet]. 2018;108(February):151–4. Available from: pmid:29605345
- 7. Patel A, Nouraei SAR. Transnasal Humidified Rapid-Insufflation Ventilatory Exchange (THRIVE): A physiological method of increasing apnoea time in patients with difficult airways. Anaesthesia. 2015;70(3):323–9. pmid:25388828
- 8. Eberhard P. The design, use, and results of transcutaneous carbon dioxide analysis: current and future directions. Anesth Analg. 2007/12/06. 2007;105(6 Suppl):S48–52. pmid:18048898
- 9. Rohling R, Biro P. Clinical investigation of a new combined pulse oximetry and carbon dioxide tension sensor in adult anaesthesia. J Clin Monit Comput. 1999;15(1):23–7. pmid:12578058
- 10. Riva T, Préel N, Theiler L, Greif R, Bütikofer L, Ulmer F, et al. Evaluating the ventilatory effect of transnasal humidified rapid insufflation ventilatory exchange in apnoeic small children with two different oxygen flow rates: a randomised controlled trial. Anaesthesia. 2020;(November 2019):1–9.
- 11. Huang L, Dharmawardana N, Badenoch A, Ooi EH. A review of the use of transnasal humidified rapid insufflation ventilatory exchange for patients undergoing surgery in the shared airway setting. J Anesth [Internet]. 2020;34(1):134–43. Available from: pmid:31612348
- 12. Ebeling CG, Riccio CA. Apneic Oxygenation With High-Flow Nasal Cannula and Transcutaneous Carbon Dioxide Monitoring During Airway Surgery: A Case Series. A Pract [Internet]. 2019;12(10):366–8. Available from: https://journals.lww.com/aacr/Fulltext/2019/05150/Apneic_Oxygenation_With_High_Flow_Nasal_Cannula.7.aspx pmid:30475239
- 13. Theiler L, Schneeberg F, Riedel T, Kaiser H, Riva T, Greif R. Apnoeic oxygenation with nasal cannula oxygen at different flow rates in anaesthetised patients: a study protocol for a non-inferiority randomised controlled trial. BMJ Open [Internet]. 2019/07/14. 2019;9(7):e025442. Available from: https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6629420/pdf/bmjopen-2018-025442.pdf pmid:31300494
- 14. Riva T, Greif R, Kaiser H, Riedel T, Huber M, Theiler L, et al. Carbon Dioxide Changes during High-flow Nasal Oxygenation in Apneic Patients: A Single-center Randomized Controlled Noninferiority Trial. Anesthesiology. 2022;136(1):82–92. pmid:34758057
- 15. Huttmann SE, Windisch W, Storre JH. Techniques for the measurement and monitoring of carbon dioxide in the blood. Ann Am Thorac Soc. 2014/04/08. 2014;11(4):645–52. pmid:24701974
- 16. Stock MC. Carbon dioxide and apnea: Common knowledge and common sense. J Clin Anesth. 1998;10(3):181–3. pmid:9603585
- 17. Bland JM, Altman DG. Agreement between methods of measurement with multiple observations per individual. J Biopharm Stat. 2007;17(4):571–82. pmid:17613642
- 18. Mertzlufft FO, Brandt L, Stanton-Hicks M, Dick W. Arterial and mixed venous blood gas status durig apnoea of intubation—Proof of the Christiansen-Douglas-Haldane effect in vivo. Anaesth Intensive Care. 1989;17(3):325–31.
- 19. Stock MC, Schisler JQ, McSweeney TD. The PaCO2 rate of rise in anesthetized patients with airway obstruction. J Clin Anesth [Internet]. 1989;1(5):328–32. Available from: http://www.sciencedirect.com/science/article/pii/0952818089900706 pmid:2516732
- 20. Eger EI, Severinghaus JW. The rate of rise of PaCO2 in the apneic anesthetized patient. Anesthesiology: The Journal of the American Society of Anesthesiologists. 1961;22(3):419–25.
- 21. Piosik ZM, Dirks J, Rasmussen LS, Kristensen CM, Kristensen MS. Exploring the limits of prolonged apnoea with high‐flow nasal oxygen: an observational study. Anaesthesia [Internet]. 2020 Nov 11;anae.15277. Available from: https://onlinelibrary.wiley.com/doi/10.1111/anae.15277 pmid:33179248
- 22. Bendjelid K, Schütz N, Stotz M, Gerard I, Suter PM, Romand JA. Transcutaneous Pco2 monitoring in critically ill adults: Clinical evaluation of a new sensor. Crit Care Med. 2005;33(10):2203–6. pmid:16215371
- 23. Bolliger D, Steiner LA, Kasper J, Aziz OA, Filipovic M, Seeberger MD. The accuracy of non-invasive carbon dioxide monitoring: A clinical evaluation of two transcutaneous systems. Anaesthesia. 2007;62(4):394–9. pmid:17381578
- 24. Spelten O, Fiedler F, Schier R, Wetsch WA, Hinkelbein J. Transcutaneous PtcCO2 measurement in combination with arterial blood gas analysis provides superior accuracy and reliability in ICU patients. J Clin Monit Comput. 2017;31(1):153–8. pmid:26628269
- 25. McMillan G, Baril R, Assn O, Lozowski D. pH measurement and control. Chemical Engineering. 2010;117(8).
- 26. van Weteringen W, Goos TG, van Essen T, Ellenberger C, Hayoz J, de Jonge RCJ, et al. Novel transcutaneous sensor combining optical tcPO2 and electrochemical tcPCO2 monitoring with reflectance pulse oximetry. Med Biol Eng Comput. 2020;58(2):239–47. pmid:31741291
- 27. Bromley I. Transcutaneous monitoring—understanding the principles. Infant [Internet]. 2008;4(3):95–8. Available from: http://search.ebscohost.com/login.aspx?direct=true&db=cin20&AN=105793627&site=ehost-live
- 28. Rodriguez P, Lellouche F, Aboab J, Buisson CB, Brochard L. Transcutaneous arterial carbon dioxide pressure monitoring in critically ill adult patients. Intensive Care Med [Internet]. 2006/02/02. 2006;32(2):309–12. Available from: https://link.springer.com/content/pdf/10.1007%2Fs00134-005-0006-4.pdf pmid:16450093