Abstract
Background
High-resolution optical coherence tomography (OCT) allows the detection of macular pathology and involvement of the optic nerve in a wide spectrum of diseases. For the differentiation of diseased and healthy status, normal values of retinal layer segmentation are critical. Yet, normative values mostly cover adult populations with only sparse data for paediatric cohorts. We present data of retinal layer characteristics via OCT in a healthy paediatric cohort.
Methods
This prospective cross-sectional study screened 75 healthy children (male = 42, female = 33, range 4–17 years) without visual problems. OCT was performed with a peripapillary ring and macula scan protocol to determine paediatric normative values for routine parameters (peripapillary retinal nerve fibre layer thickness (pRNFL), total macular volume (TMV), macular retinal thickness (RT)). The macula scan (6mm grid) was segmented using the device-inherent automated segmentation software (Heidelberg Eye Explorer) for retinal layers: RNFL, ganglion cell layer (GCL), inner plexiform layer (IPL), inner nuclear layer (INL), outer plexiform layer (OPL), outer nuclear layer (ONL) in 9 segments each and mean of the 9 segments.
Results
We obtained OCT data of 72 children with mean age 12.49 years (standard deviation, SD, 2.18; minimum 3.93). Mean global pRNFL was 102.20 μm (SD 8.24), mean TMV 8.81 mm3 (0.30) and mean RT (all segments) 318.22 μm (10.19). Segmented macular retinal layer thicknesses (mean of all segments) were: RNFL 27.67 μm (2.14), GCL 41.94 μm (2.50), IPL 34.97 μm (2.10), INL 35.18 μm (2.15), OPL 29.06 μm (2.24), ONL 68.35 μm (6.20).
Conclusion
The OCT is a useful non-invasive imaging technique for the examination of the retina in children with short duration, high imaging resolution and no known adverse effects. Normative values may serve as a comparator for different neuropaediatric disorders and are first presented with this study using an up-to-date and standardized OCT imaging technique.
Citation: Runge A-K, Remlinger J, Abegg M, Ferrazzini T, Brügger D, Weigt-Usinger K, et al. (2022) Retinal layer segmentation in a cohort of healthy children via optical coherence tomography. PLoS ONE 17(11): e0276958. https://doi.org/10.1371/journal.pone.0276958
Editor: Friedemann Paul, Charite Universitatsmedizin Berlin, GERMANY
Received: March 26, 2022; Accepted: October 18, 2022; Published: November 3, 2022
Copyright: © 2022 Runge et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Data Availability: The data are accessible via the Bern Open Repository and Information System (https://doi.org/10.48620/68).
Funding: The author(s) received no specific funding for this work.
Competing interests: RG has received speaker and board honoraria from Baxter, Bayer Schering, Biogen Idec, Bristol Meyer Squibb, CSL Behring, Eisai, Genzyme, Janssen, Merck Serono, Novartis, Stendhal, Talecris and TEVA. His department has received grant support from Bayer Schering, BiogenIdec, Genzyme, Merck Serono, Novartis and TEVA. All of RG’s declarations are unrelated to the content of this manuscript. AS has received speaker honoraria and/or travel compensation for activities with Bristol Myers Squibb, Novartis, and Roche, and research support by the Swiss MS Society and Baasch Medicus Foundation, not related to this manuscript. AKR, MA, JR, TF, DB, KWU, TL, RG, AS have no competing interests to declare that are relevant to the content of this article.
List of abbreviations: ART, automatic real time; ETDRS, early treatment diabetic retinopathy study; GCL, ganglion cell layer; G: global; HIV, human immunodeficiency virus; INL, inner nuclear layer; IPL, inner plexiform layer; N, nasal; NI, nasal inferior; NS, nasal superior; OCT, optical coherence tomography; ONL, outer nuclear layer; OPL, outer plexiform layer; PMB, papillo-macular bundle; pRNFL, peripapillary retinal nerve fibre layer thickness; RT, macular retinal thickness; SD, standard deviation; SD-OCT, spectral domain optical coherence tomography; T, temporal; TI, temporal-inferior; TMV, total macular volume; TS, temporal-superior
Introduction
In the course of the last 30 years, optical coherence tomography (OCT) has become the single most important imaging modality in ophthalmology [1]. The more recent advent of high resolution OCT and automatic segmentation of retinal layers further increased the sensitivity and specificity to detect important diseases such as macular pathology and diseases of the optic nerve. For the latter, the analysis of macular ganglion cell layer thickness shows benefits as compared to the more standard analysis of the peripapillary nerve fibre layer [2]. The separation of healthy status from disease may be on a morphological basis, but is often defined by a deviation of parametric values from normal values. In the context of OCT, the thickness of peripapillary retinal nerve fibre layer thickness (pRNFL) for example allows to identify on a statistical basis patients with optic neuropathy or disc swelling. Such normative data may thus be critical in the detection of diseases. However, normative data, especially for macular layer segmentation, are not available for the population of persons younger than 18 years. As OCT is easy to use and has no adverse effects such as radiation, it is a feasible tool in the examination of children with (neuro-) ophthalmological conditions.
As retinal thickness decreases with age in healthy adults [3–7], it is uncertain whether adult normative values can be extrapolated to children. The evolution of retinal thickness during childhood is so far unclear. Previous studies have only examined standard OCT measures (pRNFL) and/or macular volume/thickness) in healthy children with varying OCT techniques [8–12]. Only one study describes normative values for different retinal layers after segmentation, yet, using a custom-made software and may thus be difficult to transfer to other devices [13].
This study aims to examine a healthy paediatric population to establish normative values of standard OCT measures (pRNFL, total macular volume (TMV), macular retinal thickness (RT)) and the segmented macular retinal layers using the device-inherent standardized software.
Methods
Study design and participants
Seventy-five children were recruited from school classes and nurseries via the Neuropaediatric department of the Clinic for paediatric and adolescent medicine in Bochum in a cooperation with the department of Neurology, St. Josef-Hospital, Ruhr-University Bochum. Informed consent was obtained from at least one parent/legal guardian and the children themselves. The parents were given a questionnaire to detect co-morbidities. Boys and girls under 18 years of age and without visual/ocular diseases were included in the study. Axial length measurement or other ophthalmological evaluations were not performed.
In- and exclusion criteria
Of the 75 screened, 3 children had to be excluded from the study: 2 children due to premature birth and 1 child because of high myopia. In addition, 2 children were found with concomitant medication. This was not deemed to influence the results by the principal investigators. The exclusion criteria are shown in Table 1.
OCT
The examination was done using a spectral domain OCT (SD-OCT) (Spectralis®-OCT, Heidelberg Engineering, Heidelberg, Germany).
Two types of scans were used within this study: a peripapillary scan and a macular scan.
The peripapillary scan can be analysed using the thickness profile, which gives the cross-sectional view of the retina in a circle around the papilla (Fig 1). The resolution mode for the peripapillary scan was high resolution. The scan pattern used was 1 B-scan with a diameter of 3.6mm. The quality parameters were set to be automatic real-time (ART) ≥ 30 (max. 100) and a quality ≥ 25. Seven sectors are given for this ring scan: global (G), nasal (N), temporal (T), nasal-superior (NS), nasal-inferior (NI), temporal-superior (TS) and temporal-inferior (TI). The papillo-macular bundle (PMB) is separately measured. N/T gives the ratio of the nasal and temporal sector. The automated pRNFL measurement compares the measured values to deposited adult normative values.
Upper left: fundus image, upper right: OCT scan of retinal layers, lower left: RNFL values per sector (in μm), lower right: RNFL values over the whole OCT scan with integration in normative percentiles (as set by the device for adults) and coloured areas to indicate values outside the normal range. Abbreviations: OCT–optical coherence tomography, RNFL–retinal nerve fibre layer.
The macular scan is shown in Fig 2. The resolution mode for the macular scan was a high resolution mode. The scan pattern used was 61 B-scans in a distance of 124µm and with an early treatment diabetic retinopathy study (ETDRS) grid size of 1, 3 and 6mm (radius). The quality parameters were set to be ART ≥ 11 and a quality ≥ 25. The fovea is centred (centre point) with two surrounding circles. Each circle is divided into 4 sectors: nasal (N1 / N2), temporal (T1 / T2), superior (S1 / S2), inferior (I1 / I2), 1 representing the inner, 2 representing the outer circle. The sector surrounding the centre point with 1mm diameter is called C0.
Upper left: fundus image with colours coding for GCL, lower left: OCT scan of retinal layers, right: GCL values per sector and coded by colours. Abbreviations: OCT–optical coherence tomography, GCL–ganglion cell layer.
pRNFL, TMV and RT were provided by the software. For macular retinal layer segmentation, the segmentation software (Segmentation Batch, Heidelberg Eye Explorer) was used segmenting all layers of the retina automatically (Fig 3). The internal scan quality measure was defined to be ≥25 to be included in the analysis. Scans which did not fully display the retina were excluded by the software and did not undergo the analysis.
From top to bottom: Retinal nerve fibre layer (RNFL), ganglion cell layer (GCL), inner plexiform layer (IPL), inner nuclear layer (INL), outer plexiform layer (OPL), outer nuclear layer (ONL); retinal thickness (RT).
Manual correction was deliberately abstained from in our study to only include results that the segmentation software determined using a segmentation algorithm and to avoid potential bias caused by manual adaptation. Scans were checked for errors occurring during the segmentation process: Scans with minor inaccuracies not remarkably affecting the results were not manually adapted, but still included; scans with major inaccuracies were excluded from the study.
Normative values for the inner retinal layers were analysed for a total volume of the 6mm grid in mm3, for all 9 sectors (C0, N1/2, T1/2, S1/2, I1/2) and a mean of these sectors in µm: 1) RNFL, 2) GCL, 3) IPL, 4) INL, 5) OPL, 6) ONL.
Statistics
All data were exported from the Heidelberg Eye Explorer to Microsoft Excel, Version 16.35 for Mac. All statistical analyses were performed using GraphPad Prism Software, Version 8.0.1 for Windows, GraphPad Software, Inc., La Jolla, California, USA. To account for a potential bias due to inter-eye correlation in individuals with two eyes included in the analyse, an outlier identification was performed using the ROUT method, based on the False Discovery Rate (FDR), with a Q set to 1%. Outliers were present in 37 individuals vs. 35 individuals without any outlier (102 detected outliers in total of 14’740 values). Using Fisher’s exact test, no different distribution of participants with one or both eyes included in the analysis for the presence of any outlier was detected (p = 0.57). Results are thus presented using values of both eyes whenever available. The full outlier analysis is given in the open data repository (https://doi.org/10.48620/68).
Approval of ethics committee
The study has been approved by the ethics committee of the medical faculty of Ruhr-University Bochum with the registration number 3952–11 and complies with the tenets of the Declaration of Helsinki. Written informed consent was obtained from at least one parent/legal guardian and the individual participants included in the study.
Due to the change of the principal investigator, the anonymised evaluation of the data was continued in a new centre as deemed appropriate by the responsible committee (Cantonal Ethics Committee Bern). All data were anonymized prior to any analysis.
Results
Cohort characteristics
Demographic data of the cohort are shown in Table 2. The age distribution of the cohort is presented in detail in Table 3.
OCT measurements
Peripapillary retinal nerve fibre layer.
Data of the pRNFL measurement are shown in Table 4.
Macular scans.
Automated TMV and RT as well as segmentation data were acquired on the same scans (mean quality 31.85, SD 3.23; mean ART 14.64, SD 6.44).
Total macular volume and retinal thickness.
TMV and RT data are shown in Table 5 as generated by the standard OCT algorithm of the macular scan. Corresponding values gathered with the segmentation software differ only slightly (data given in https://doi.org/10.48620/68).
Segmentation data.
Inner retinal layer volumes of the 6mm grid and thicknesses for RNFL, GCL, IPL, INL, OPL and ONL are shown in Tables 6–11.
Discussion
The OCT has become an important diagnostic tool to acquire in vivo images of the retina at a spatial resolution in the range of microns [14]. We set out to determine normative values in a healthy paediatric cohort for future use in different paediatric disorders.
In our study, the examination process was quick, and no major difficulties occurred. One main restriction of the OCT that we hypothesized was the age of the children, because a young child might not sit quiet during the required time. However, we observed an overall good compliance. This will alleviate early detection of ocular diseases such as glaucoma by OCT with consecutive earlier treatment initiation and limitation of disease progression [15–17].
In other paediatric populations with healthy eyes, OCT proved good reproducibility [13, 18]. Our results additionally demonstrate feasibility in a routine clinical setting.
Our study provides normative values of healthy children that can be used as a reference in studies of specific disorders or in clinical practice. The OCT protocols used here are standard pre-set protocols as detailed above. This allows the use of our data in different settings with good comparability.
Comparison of previous studies evaluating children to our study is limited. Only few evaluate healthy children and report standard OCT parameters (pRNFL, TMV, RT) [8–11]. Some studies only focus on few retinal layers [18, 19]. To our knowledge, only one study with complete segmentation data has been published before, but used a custom software which may hamper broader clinical applicability and comparability [13]. We assume that different segmentation algorithms define the border between layers slightly differently and the thickness values derived from distinct layers may not be directly comparable. This is a limitation of the current study, as our findings are most likely only valid for examinations done with a Heidelberg Spectralis system, one of the most commonly used devices, but still a clear issue in the comparison of different OCT studies.
Within previous studies, different OCT protocols have been used or are sometimes not precisely described which may lead to different results [3, 12]. Comparability of the data is thus limited due to a lack of precise information of the machine used, the areas analysed or the diameters of the sectors [12].
Most and best researched in children are values for pRNFL. The mean global pRNFL measured during our study is comparable to other cohorts including adults [4, 5, 20] and children [21, 22]. Half of the other studies report slightly higher mean pRNFL values [10, 13, 23–25] and the other half slightly lower mean pRNFL values [8, 9, 11, 20, 26], respectively. Most of these studies were performed using SD-OCT as in our study, some using Stratus OCT [9, 10, 24].
TMV could not be compared due to previously mentioned discrepancies (as for example the use of OCT-3 [16], Cirrus OCT [27] or a grid of 3.5mm [21]).
The age distribution in our analysis with most participants between 10 and 16 years might limit transferability of the data to very young children. We did not detect a correlation of age and pRNFL or TMV, yet, our cohort was not powered to examine age effects and this might be attributed to a narrow age range and not hold true for younger children. This is why we refrained from additional correlation analyses for all retinal layers with age or other parameters. Still, in one study, the influence of an age-related decrease in RNFL/retinal thickness seemed to be negligible from the age of 4 years and older [3] whereas another study reported an increase of the macular thickness with age for the central subfield [13] in a cohort of comparable age to ours. Correlations of some OCT parameters with age seem to be biologically reasonable as a study found evidence of progressive foveal development in children [7]. It might be interesting to longitudinally observe the development of the eyes of children from birth on as the current data remains inconclusive.
OCT measurement reproducibility in children is reported to be similar to adults. Additional comparisons of this study to ours are limited because children were not healthy and no segmentation data was provided [28].
Some studies suggest differences between Caucasian, Black and Asian populations [12, 14, 23, 29]. Such differences between ethnicities could not be analysed in our study since all children were Caucasian.
Comparability of our study to the one investigating segmentation data using a custom software and manual segmentation of the layers is further limited because scan parameters differed to the commonly used EDTRS grids. With these settings, the inner retinal layers, which are the focus of our work and seem relevant in both ophthalmologic and neurologic conditions, could not be measured as the area around the fovea investigated was too narrow [13].
Our segmentation data shows that RNFL and IPL in children seem rather comparable to normative values for young adults reported previously [6]. However, the GCL values in our cohort of children are slightly higher. It is described that GCL and IPL decrease with age in adults [6], yet, we cannot determine whether this difference is rather of methodological or biological origin and will need further longitudinal investigations.
Conclusions
Our study shows that SD-OCT is an effective imaging technique for the examination of the RNFL, TMV and RT, as well as the analysis of individual layers of the retina. We demonstrate feasibility in children from the age of approximately 4 years on.
Normative values comparable to the data presented here do thus far not exist and will be helpful in the clinical use of the standard OCT settings and the system-inherent software. Our data might facilitate diagnosis of ophthalmological and neurological disorders in children in clinical practice and in future studies.
Supporting information
S1 File. Anonymized dataset link: https://doi.org/10.48620/68.
https://doi.org/10.1371/journal.pone.0276958.s001
(TXT)
References
- 1. Baker-Schena L, Edmond JC, El-Dairi M, Hartnett ME, Maldonado RS, Stout JT: OCT Gains Ground in Pediatric Diagnosis and Treatment. In: American Academy of Ophthalmology, EyeNet Magazine, Paediatric Ophthalmology Clinical Update, March 2016, p35–36. 2016.
- 2. Mühlemann F, Grabe H, Fok A, Wagner F, Brügger D, Sheldon CA, et al: Homonymous hemiatrophy of ganglion cell layer from retrochiasmal lesions in the visual pathway. Neurology 2020, 94(3):e323–e329.
- 3. Alamouti B, Funk J: Retinal thickness decreases with age: an OCT study. The British Journal of Ophthalmology 2003, 87(7):899–901.
- 4. Bendschneider D, Tornow RP, Horn FK, Laemmer R, Roessler CW, Juenemann AG, et al. Retinal Nerve Fiber Layer Thickness in Normals Measured by Spectral Domain Oct. Journal of Glaucoma 2010, 19(7):475–482.
- 5. Budenz DL, Anderson DR, Varma R, Schuman J, Cantor L, Savell J, et al. Determinants of Normal Retinal Nerve Fiber Layer Thickness Measured by Stratus OCT. Ophthalmology 2007, 114(6):1046–1052.
- 6. Nieves-Moreno M, Martínez-de-la-Casa JM, Cifuentes-Canorea P, Sastre-Ibáñez M, Santos-Bueso E, Sáenz-Francés F, et al. Normative database for separate inner retinal layers thickness using spectral domain optical coherence tomography in Caucasian population. PLOS ONE 2017, 12(7):e0180450.
- 7. Tariq YM, Li H, Burlutsky G, Mitchell P: Retinal nerve fiber layer and optic disc measurements by spectral domain OCT: normative values and associations in young adults. Eye 2012, 26(12):1563–1570.
- 8. Al-Haddad C, Barikian A, Jaroudi M, Massoud V, Tamim H, Noureddin B: Spectral domain optical coherence tomography in children: normative data and biometric correlations. BMC Ophthalmology 2014, 14:53. pmid:24755354
- 9. Larsson E, Eriksson U, Alm A: Retinal nerve fibre layer thickness in full-term children assessed with Heidelberg retinal tomography and optical coherence tomography: normal values and interocular asymmetry. Acta Ophthalmologica 2011, 89(2):151–158.
- 10. Salchow DJ, Oleynikov YS, Chiang MF, Kennedy-Salchow SE, Langton K, Tsai JC, et al. Retinal Nerve Fiber Layer Thickness in Normal Children Measured with Optical Coherence Tomography. Ophthalmology 2006, 113(5):786–791. pmid:16650674
- 11. Wang C- Y, Zheng Y- F, Liu B, Meng Z- W, Hong F, Wang X- X, et al. Retinal Nerve Fiber Layer Thickness in Children: The Gobi Desert Children Eye Study. Investigative Ophthalmology & Visual Science 2018, 59(12):5285–5291.
- 12. Banc A, Ungureanu MI: Normative data for optical coherence tomography in children: a systematic review. Eye (Lond) 2021, 35(3):714–738.
- 13. Yanni SE, Wang J, Cheng CS, Locke KI, Wen Y, Birch DG, et al. Normative Reference Ranges for the Retinal Nerve Fiber Layer, Macula, and Retinal Layer Thicknesses in Children. American journal of ophthalmology 2013, 155(2):354–360.e351.
- 14. Grover S, Murthy RK, Brar VS, Chalam KV: Normative Data for Macular Thickness by High-Definition Spectral-Domain Optical Coherence Tomography (Spectralis). American Journal of Ophthalmology 2009, 148(2):266–271. pmid:19427616
- 15. Forte R, Cennamo GL, Finelli ML, de Crecchio G: Comparison of time domain Stratus OCT and spectral domain SLO/OCT for assessment of macular thickness and volume. Eye 2008, 23(11):2071–2078.
- 16. Hess DB, Asrani SG, Bhide MG, Enyedi LB, Stinnett SS, Freedman SF: Macular and retinal nerve fiber layer analysis of normal and glaucomatous eyes in children using optical coherence tomography. American Journal of Ophthalmology 2005, 139(3):509–517.
- 17. Pierro L, Zampedri E, Milani P, Gagliardi M, Isola V, Pece A: Spectral domain OCT versus time domain OCT in the evaluation of macular features related to wet age-related macular degeneration. Clinical Ophthalmology (Auckland, NZ) 2012, 6:219–223.
- 18. Lee YP, Ju YS, Choi DG: Ganglion cell-inner plexiform layer thickness by swept-source optical coherence tomography in healthy Korean children: Normative data and biometric correlations. Sci Rep 2018, 8(1):10605. pmid:30006534
- 19. Cheng L, Wang M, Deng J, Lv M, Jiang W, Xiong S, et al. Macular Ganglion Cell-Inner Plexiform Layer, Ganglion Cell Complex, and Outer Retinal Layer Thicknesses in a Large Cohort of Chinese Children. Investigative Ophthalmology & Visual Science 2019, 60(14):4792–4802.
- 20. Barrio-Barrio J, Noval S, Galdós M, Ruiz-Canela M, Bonet E, Capote M, et al. Multicenter Spanish study of spectral-domain optical coherence tomography in normal children. Acta Ophthalmologica 2013, 91(1):e56–e63.
- 21. Lee JWY, Yau GSK, Woo TTY, Yick DWF, Tam VTY, Lai JSM: Retinal Nerve Fiber Layer Thickness in Myopic, Emmetropic, and Hyperopic Children. Medicine 2015, 94(12).
- 22. Zhu B- D, Li S- M, Li H, Liu L- R, Wang Y, Yang Z, et al. Retinal Nerve Fiber Layer Thickness in a Population of 12-Year-Old Children in Central China Measured by iVue-100 Spectral-Domain Optical Coherence Tomography: The Anyang Childhood Eye StudyRetinal Nerve Fiber Layer Thickness in Children. Investigative Ophthalmology & Visual Science 2013, 54(13):8104–8111.
- 23. Huynh SC, Wang XY, Rochtchina E, Mitchell P: Peripapillary Retinal Nerve Fiber Layer Thickness in a Population of 6-Year-Old Children. Ophthalmology 2006, 113(9):1583–1592.
- 24. Pawar N, Maheshwari D, Ravindran M, Ramakrishnan R: Retinal nerve fiber layer thickness in normal Indian pediatric population measured with optical coherence tomography. Indian Journal of Ophthalmology 2014, 62(4):412–418.
- 25. Turk A, Ceylan OM, Arici C, Keskin S, Erdurman C, Durukan AH, et al. Evaluation of the Nerve Fiber Layer and Macula in the Eyes of Healthy Children Using Spectral-Domain Optical Coherence Tomography. American Journal of Ophthalmology 2012, 153(3):552–559.e551.
- 26. Rao A, Sahoo B, Kumar M, Varshney G, Kumar R: Retinal Nerve Fiber Layer Thickness in Children <18 Years by Spectral-domain Optical Coherence Tomography. Seminars in Ophthalmology 2013, 28(2):97–102.
- 27. Altemir I, Pueyo V, Elía N, Polo V, Larrosa JM, Oros D: Reproducibility of Optical Coherence Tomography Measurements in Children. American Journal of Ophthalmology 2013, 155(1):171–176.e171.
- 28. Chung HK, Han YK, Oh S, Kim SH: Comparison of Optical Coherence Tomography Measurement Reproducibility between Children and Adults. PLoS One 2016, 11(1):e0147448. pmid:26808961
- 29. El-Dairi MA, Asrani SG, Enyedi LB, Freedman SF: Optical coherence tomography in the eyes of normal children. Arch Ophthalmol 2009, 127(1):50–58.