Abstract
Background
For primary angle-closure and angle-closure glaucoma, the fact that refractive error sometimes deviates from predictions after intraocular lens (IOL) implantation is familiar to cataract surgeons. Since controversy remains in the accuracy of IOL power calculation formulas, both traditional and network meta-analysis on formula accuracy were conducted in patients with primary angle-closure conditions.
Methods
A comprehensive literature search was conducted through Aug 2022, focusing on studies on intraocular lens power calculation in primary angle-closure (PAC) and primary angle-closure glaucoma (PACG). A systemic review and network meta-analysis was performed. Quality of studies were assessed. Primary outcomes were the mean absolute errors (MAE) and the percentages of eyes with a prediction error within ±0.50 diopiters (D) or ±1.00 D (% ±0.50/1.00 D) by different formulas.
Results
Six retrospective studies involving 419 eyes and 8 formulas (Barrett Universal II, Kane, SRK/T, Hoffer Q, Haigis, Holladay I, RBF 3.0 and LSF) were included. SRK/T was used as a reference as it had been investigated in all the studies included. Direct comparison showed that none of the involved formula outperformed or was defeated by SRK/T significantly in terms of either MAE or % ±0.50/1.00 D (all P>0.05). Network comparison and ranking possibilities disclosed BUII, Kane, RBF 3.0 with statistically insignificant advantage. No significant publication bias was detected by network funnel plot.
Citation: Lu W, Hou Y, Yang H, Sun X (2022) A systemic review and network meta-analysis of accuracy of intraocular lens power calculation formulas in primary angle-closure conditions. PLoS ONE 17(10): e0276286. https://doi.org/10.1371/journal.pone.0276286
Editor: Andrzej Grzybowski, University of Warmia, POLAND
Received: August 31, 2022; Accepted: October 4, 2022; Published: October 14, 2022
Copyright: © 2022 Lu et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Data Availability: All relevant data are within the manuscript and its Supporting Information files.
Funding: This research project was supported by Shanghai Municipal Health Commission (20214Y0073) and Clinical Research Plan of SHDC (SHDC2020CR6029). The authors were also supported by grants from National Key Research and Development Program of China (2020YFA0112700); the State Key Program of National Natural Science Foundation of China (82030027), and the subject of major projects of National Natural Science Foundation of China (81790641). The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
Competing interests: The authors have declared that no competing interests exist.
Introduction
Primary angle-closure disease (PACD), as the leading cause of irreversible blindness in East Asia, has been projected to affect 32 million adults aged between 40 and 80 years old by 2040 [1]. Given the growing popularity of physical examination and people’s concerns about their ocular health in past decades, ophthalmologists are constantly moving towards early detection and treatment of PACD. Though laser peripheral iridotomy (LPI) prophylaxis is a widely performed measure in patients at risk of angle-closure, it has little effect in treating existent peripheral anterior synechia (PAS) at intermediate and advanced stages [2]. Nowadays, cataract surgery (or clear-lens extraction) has been a preferred option for eyes with PACD due to its multiple benefits. i.e., improving eyesight, lowering intraocular pressure (IOP), and widening the narrowed anterior chamber angle [3, 4], yet the clinical benefits might be hindered by the dissatisfactory post-operative visual acuity.
Compared with cataract surgery in normal population, a range of publications have suggested that similar procedures in PACD patients often resulted in inaccurate refractive outcomes [3, 5]. According to current studies, for simple cataract patients, the percentage of post-operative refractive error within ±0.50 D varied between 73% and 88%, while the percentage dropped to 50%~68% for PACD patients [5, 6], indicating extra inaccuracy in terms of IOL power prediction in PACD patients, which had drawn ophthalmologists’ attention. Along with well-accepted precision of ocular biometry, accuracy of manufacturer IOL power quality control and the now-mature surgical techniques, the accuracy of IOL power calculation becomes the most essential factor for achieving the target refraction post-operatively. A dozen formulas for IOL power calculation have been established with diverse core principles on the basis of normal eyes since 1980s, yet the accuracy varies [7–9]. Theoretically, it could be well imagined that discrepancies in IOL prediction would occur when these formulas, especially regressive ones [10], were applied to special population with unusual ocular biometric parameters, e.g., significantly shallow anterior chamber (AC), thick lens and short axial length (AL) [11–13]. Therefore, the final issue boils down to choosing or even developing a proper IOL calculation formula for PACD eyes to achieve optimal postoperative visual acuity after cataract surgeries [14, 15].
Hence, we performed a systemic review together with network meta-analysis of the accuracy of IOL power calculation formulas in patients with primary angle-closure conditions, hoping to figure out which formula currently available provides the highest accuracy for this target cohort by comparing and ranking those commonly used ones on their performance.
Methods
Prior to analysis, a detailed protocol was made in accordance with the standard systemic review guidelines outlined by the Cochrane Reviewers’ Handbook. The PRISMA checklist was reviewed, strictly following the items required for systemic review and network meta-analysis (S1 Checklist) [16, 17]. Our research adhered to the tenets of the Declaration of Helsinki, and was registered on PROSPERO (CRD42022326541).
Literature search
Two independent investigators (W.L. and Y.H.) conducted the comprehensive literature search in the following databases for relevant English language literature: PubMed, Cochrane Data Base of Systematic Reviews, the Cochrane Central Register of Controlled Trials (CENTRAL) and Web of Science through Aug 2022. The search terms were (“angle closure” OR “angle-closure”) AND (“cataract” OR “IOL” OR “intraocular lens”). The detailed search results of that search process are provided in S1 Table. In order to ensure that no relevant study would be omitted, two independent authors (W.L. and Y.H.) also checked the reference lists for relevant studies.
Inclusion and exclusion criteria
Inclusion criteria for studies were: 1) randomized controlled trials, cross-sectional studies, retrospective or prospective observational studies; 2) eyes diagnosed with PACS (primary angle-closure suspect, defined as eyes in which at least 180° of the posterior pigmented trabecular meshwork was invisible on gonioscopy in the primary position of gaze without indentation but with neither increased IOP nor glaucomatous neuropathy), PAC (defined as eyes with evidence of PAS or increased IOP on top of gonioscopic criteria for PACS), or PACG (defined as those PAC who additionally demonstrated glaucomatous damage of the optic nerve) [18]; 3) eyes undergoing uncomplicated cataract surgery or clear lens extraction with one-piece IOL implantation; 4) comparison of at least two types of IOL power calculation formulas available and 5) pre- and postoperative data collected with no less than 1-month interval, providing at least one outcome variable (i.e., mean absolute error (MAE) or percentages of predicted refractive error within a certain range) or one that can be calculated.
Exclusion criteria for studies were: 1) eyes with prior intraocular surgeries other than laser peripheral iridotomy (LPI), given that cataract surgery in patients with prior trabeculectomy had significantly greater refractive surprise than those in the control groups [19, 20]; 2) eyes with other conditions ineligible for inclusion (e.g. severe pterygium, severe corneal or vitreous opacity, macular degeneration, and retinal detachment, etc. [21]); 3) preoperative ocular biometric measurements performed with instruments other than IOLMaster; 4) studies that did not provide targeted outcome data or that were computationally unavailable.
Data extraction and quality assessment
Two authors (W.L. and Y.H.) independently extracted the following data from each study: first author, year of publication and journal, country where the study was conducted, study period, study design, original inclusion criteria and exclusion criteria, total number of eyes included, demographic and biometry data of patients, IOLs types for each patient, patients’ prior surgical history, formulas being evaluated, time intervals between pre- and postoperative biometric evaluation, and both outcomes measurements and values. Disagreements were resolved by discussion until consensus was reached or by consulting a third author (H.Y.).
The quality of studies were assessed using a modified Quality Assessment of Diagnostic Accuracy Studies-2 (QUADAS-2) tool designed for diagnostic tests [22], and the level of evidence were evaluated as recommended by the Oxford Centre for Evidence-based Medicine. Risk of publication bias were measured using network funnel plot.
Outcome measurements
Mean absolute error (MAE), % ±0.50 D and % ±1.00 D were used as major outcomes of the review. MAE was defined as mean of the absolute value of the differences between the actual and the predicted postoperative spherical equivalent (SE) of refraction using each formula. % ±0.50 D (or % ±1.00 D) was defined as the percentage of eyes with prediction error ranging from -0.50 D (or -1.00 D) to + 0.50 D (or +1.00 D) of the target refraction after IOL implantation, as previously reported [15]. If any outcome value was not reported, we calculated it from the raw data provided by the studies. If the standard deviation (SD) data could not be retrieved from the text, this study was not eligible for MAE pooling and were therefore excluded.
Statistical analysis
For traditional meta-analysis, STATA v 14.0 (StataCorp LP) was applied to generate head-to-head comparisons between SRK/T formula and the others. MAE was analyzed by weighted mean difference (WMD), and % ±0.50 D and % ±1.00 D were evaluated by the estimated risk ratios (RRs) with 95% credible intervals (CIs). An WMD less than 0 and RR greater than 1 indicated that the alternative formula was more accurate than SRK/T. I2 value, together with the P value of Cochran’s Q test (for MAE) or Mantel Haenszel Q test (for % ±0.50/1.00 D), were reported for estimating the statistical heterogeneity in case of the bias introduced by small study effect [23]. An I2 value greater than 50% with the P value less than 0.05 was considered substantial heterogeneity, where random-effect model analysis was performed. Otherwise the fixed effect model was used. The risk for publication bias was measured by network funnel plots, and a symmetrical funnel plot indicated the relatively low risk of bias.
For network meta-analysis, pairwise analysis was conducted based on direct and indirect evidence by R v 3.5.2 (RStudio, Inc.). WMD of MAE, RRs with 95% credible intervals (CIs) for % ±0.50 D and % ±1.00 D were pooled. If no evidence of inconsistency was detected, the consistency model was adopted to calculate the relative effects of all investigated formulas. The Markov Chain Monte Carlo (MCMC) methods was applied to obtain the pooled effect sizes, with 4 chains calculated. Each chain comprised 20 000 simulation iterations, 5000 adaptation iterations and a thinning interval of 1. Statistically significant findings were recognized when bars of CIs did not include the value of 0 for the mean difference (MD) or 1 for the RR. Different formulas were ranked according to their performances of accuracy, i.e., the calculated possibilities of each formula at a specific rank based on their MAE, % ±0.50 D and % ±1.00 D, and the results were summarized and illustrated by bar plots.
Results
Literature search
Initial literature search identified 2752 articles (Fig 1). After comprehensive screening of the articles, 6 eligible studies were included in this systemic review and network meta-analysis, involving 419 eyes and 8 IOL calculation formulas. The demographic and other main characteristics extracted were summarized in Tables 1 and 2.
Shown is the process from database searching to the studies selected for this meta-analysis. After literature search and comprehensive selection, 6 articles were identified eligible for network meta-analysis. LPI: laser peripheral iridotomy; IOL: intraocular lens.
All studies included were designed as retrospective observational studies and mostly with a female-predominate pattern. Patients, if necessary, underwent prior surgeries but no more than LPI. One-piece IOL were implanted and all the ocular biometries were performed by IOLMaster preoperatively, with a follow-up period of no less than 1 month. For formulas, the overall network of the original direct comparisons of the 8 methods involved was illustrated in Fig 2. SRK/T had been investigated in all the studies included, therefore showing strongest links with the other formulas (Fig 2).
Each node represents 1 formula, with the size of the node proportional to the numbers of eyes enrolled using the corresponding formula. The edges that connect nodes suggest direct comparisons extracted from original studies, with the width proportionate to the number of studies. SRK/T: Sanders, Retzlaff, Kraff/theoretical; BUII: Barrett Universal II; RBF: Hill-Radial Basis Function; LSF: Ladas Super Formula.
Quality identification of included studies
The quality of each study evaluated by using the QUADAS-2 tool was plotted in Fig 3. Generally, all studies were of acceptable risk of bias. Further identification of risk of bias of included studies was summarized in S2 Table, with level of evidence recommended by the Oxford Centre for Evidence-based Medicine, indicating that all the studies included were qualified for this systemic review and meta-analysis.
All of the studies were rated as ‘unclear’ in terms of the potential bias in patient selection, since they were all retrospective and did not report the enrollment strategy of patients (consecutively or randomly). As the exact meanings of ‘diagnostic tests’, ‘index tests’ or ‘reference standard’ differed from their original interpretations in this research field, the risk for bringing ‘index tests’ bias was considered low after our careful evaluation. Since all studies used SRK/T as a relative reference, the risk of bias was regarded as low for ‘reference standard’. Regarding time flow, all studies were of high quality. QUADAS-2: Quality Assessment of Diagnostic Accuracy Studies-2.
Traditional meta-analysis
Traditional meta-analysis was conducted for direct comparison between the most commonly used SRK/T formula (which was evaluated by all the studies included) and the others. However, the single-study investigation of BUII, Kane and Holladay I formula for MAE comparison, and RBF 3.0, LSF formula for % ±0.50 D and % ±1.00 D comparison made them ineligible for pooled analysis. All pairs exhibited little heterogeneity as I2 < 50% with P value greater than 0.05 so that fixed model was applied. In terms of MAE, neither Haigis nor Hoffer Q showed significant superiority over SRK/T by direct comparison (Fig 4A, P = 0.140 between SRK/T and Haigis; P = 0.990 between SRK/T and Hoffer Q). With regard to % ±0.50 D and % ±1.00 D in comparison to SRK/T, Hoffer Q and Kane showed insignificantly higher percentage (Fig 4B, RR 1.03, P = 0.434 for % ±0.50 D by SRK/T-Hoffer Q; RR 1.05, P = 0.250 for % ±1.00 D by SRK/T-Hoffer Q; RR 1.03, P = 0.731 for % ±0.50 D by SRK/T-Kane and RR 1.02, P = 0.779 for % ±1.00 D by SRK/T-Kane); Haigis exhibited nonsignificant inferiority (Fig 4B, RR 0.99, P = 0.859 for % ±0.50 D by SRK/T-Haigis; RR 0.97, P = 0.306 for % ±1.00 D by SRK/T-Haigis), while the performance of BUII relative to SRK/T turned out to be contradictory for the predicted refractive error within different ranges, yet both being insignificant (RR 0.92, P = 0.427 for % ±0.50 D and RR 1.02, P = 0.671 for % ±1.00 D).
Direct comparisons of MAE between SRK/T and Haigis/Hoffer Q (A) and % ±0.50/1.00 D between SRK/T and Haigis/Hoffer Q/BUII/Kane (B) were displayed in the forest plots. P value of Cochran’s Q test (for MAE) or Mantel-Haenszel Q test (for % ±0.50/1.00 D) were reported for testing study heterogeneity. SRK/T: Sanders, Retzlaff, Kraff/theoretical; BUII: Barrett Universal II.
Network meta-analysis
A network meta-analysis was further conducted concerning the accuracy of IOL power calculation formulas in PAC/PACG eyes based on the studies included. Pairwise comparisons were realized according to both direct and indirect evidence, with the results exhibited as forest plots in Fig 5 using the widely used SRK/T as a reference for all the MAE, % ±0.50 D and % ±1.00 D comparisons. 5 formulas were compared for MAE and 6 for % ±0.50 D and % ±1.00 D after data eligibility evaluation. All the studies passed the consistency test (S1 and S2 Figs), and the trace and density plot of network comparison indicated satisfactory convergence of the model. Risk of publication bias of MAE, % ±0.50 D and % ±1.00 D was measured using network funnel plot (S3A–S3C Fig for MAE, % ±0.50 D and % ±1.00 D, respectively), which appeared symmetrical and therefore indicated relatively small publication bias. Generally, there were no significant difference disclosed by any pairwise comparison, yet statistical analysis revealed potential superiority or inferiority. Specifically, BUII (MD -0.12, 95% CI -0.29 to 0.044) and Kane (MD -0.046, 95% CI -0.20 to 0.12) appeared insignificantly superior to SRK/T regarding MAE (Fig 5A). Additionally, pooled results showed both RBF 3.0, Kane and Hoffer Q with probably higher percentage of prediction error within % ±0.50 D, while all formulas except for Haigis seemed to beat SRK/T concerning % ±1.00 D (Fig 5B and 5C). In either MAE or % ±0.50/1.00 D, Haigis might underperform SRK/T, and LSF was likely to exhibit worse prediction accuracy in terms of % ±0.50 D.
The SRK/T formula was set as a reference and was compared with other listed methods. The pooled results for comparison of MAE, % ±0.50 D and % ±1.00 D were illustrated in A, B and C, respectively. SRK/T: Sanders, Retzlaff, Kraff/theoretical; BUII: Barrett Universal II; RBF: Hill-Radial Basis Function; LSF: Ladas Super Formula.
To rank these formulas, we further calculated the rank possibilities of 8 formulas at each position based on their performance as measured by MAE, % ±0.50 D and % ±1.00 D. The weighed overall ranks of 6 formulas according to the MAE and of 7 formulas according to % ±0.50 D and % ±1.00 D were plotted in Fig 6A–6C, respectively, showing the potential efficacy of these methods in predicting IOL power in patients with PAC and PACG. Specifically, BUII outperformed in controlling MAE, followed by Kane (71.1% and 14.7% probabilities for being 1st, respectively). However, the ranks of % ±0.50 D and % ±1.00 D showed variability. For % ±0.50 D, the RBF 3.0, Kane and Hoffer Q formulas showed 77.4%, 80.1% and 56.1% of the possibilities to rank the top 3, respectively; while in terms of % ±1.00 D, the possibilities were 69.0% for Hoffer Q, 53.9% for both BUII and LSF for being top 3. For further confirmation of study consistency, node-splitting analysis was conducted to identify discrepancies between the direct and indirect effect of comparisons (S4 Fig), and the result showed that direct and indirect evidence were in agreement with each other and with the results of the consistency model.
Ranking possibility showed the probabilities of each formula at different positions when considering their performance of prediction accuracy, which was assessed by MAE (A) and prediction error (% ±0.50 D (B) and % ±1.00 D (C)). All the formulas were plotted in descending order. SRK/T: Sanders, Retzlaff, Kraff/theoretical; BUII: Barrett Universal II; RBF: Hill-Radial Basis Function; LSF: Ladas Super Formula.
Discussion
Accuracy of IOL power calculation formulas applied in PACD patients has long been controversial. Nowadays, more than ten formulas have been developed and undoubtedly there are more on the way, into which various parameters are incorporated to improve calculation accuracy. Every formula is reasonably suggested to be used with respect to its specific range of application [27]. However, the fact that almost all the formulas were derived from normal population resulted in their application in PACD tantamount to external validation based on ill populations. Consequently, discrepancies are unsurprising in IOL prediction in eyes with PACD. Additionally, currently available studies conducted among PACD eyes showed divergent accuracies of different formulas [7, 9, 21, 24–26]. All these facts stress the importance of our study.
Principal findings
In this network meta-analysis, we had undergone thorough literature search, enrolling 6 qualified studies with 419 eyes. Though with the intention of targeting PACD patients, our preliminary search revealed that all studies that were eligible mainly focused on conditions of synechial angle closure without/with glaucomatous neuropathy in the PACD spectrum (i.e., PAC, PACG). Though PACS was also mentioned in some studies [9], the fact that it did not strictly follow the ISGEO definition led to our exclusion of the data in case any bias were introduced. Furthermore, given that less than 1% of PACS would progress into PAC/PACG per year as stated by He et al. [28] and that only 2.5~3.75% PACS ultimately develop PAC/PACG [29], PACS might be an independent subgroup different from PAC/PACG in view of pathogenesis and prognosis. Consequently, we only included studies focusing on PAC and PACG for further analysis. QUADAS-2 were applied (Fig 3) with further identification of risk of bias of included studies (S2 Table), and in general, all the studies included showed acceptable risk of bias.
Somewhat pitifully, our results of both traditional and network meta-analysis showed no significant differences among all the included formulas concerning the refractive errors being introduced. However, there were signs that some formulas might outperform the others. For traditional meta-analysis, direct comparisons of eligible formulas with SRK/T indicated that Kane, a new-generation formula [30], showed slight but nonsignificant superiority over SRK/T for % ±0.50 D and % ±1.00 D. The outperformance of new-generation formulas could be attributed to the incorporation of various essential ocular parameters involved in pathogenesis of primary angle-closure conditions [30, 31], yet more were to be discovered of the exact reasons behind the potential error. The contradictory performance of BUII, one of the most accurate and widely-used new-generation formulas, could be explained by its mere evaluation in only 2 enrolled studies, for which bias might exist. We were also surprised by the mismatch between clinical practice and our findings, i.e., the underperformance of Haigis and the outperformance of rarely-used Hoffer Q. But all these conclusions should be drawn with care as none of the comparison appeared statistically significant. For network meta-analysis, BUII achieved better postoperative outcomes than other formulae based on MAE, consistent with previous studies and current clinical experience [8]. However, BUII and some other seemingly preferable formulas like RBF 3.0 and LSF had relatively wide confidence intervals, indicating their instability when applied for PAC/PACG eyes. Altogether with the analysis of ranking possibilities, IOL calculation formulas varied in their performance, thereby further demonstrating the nonexistence of their absolute advantages over each other in PAC/PACG eyes.
Limitations of this study
The major limitation of this study was the limited studies eligible for this meta-analysis, which on the one hand hindered the revealing of clinically significant findings and on the other hand made further subgroup analyses less possible. As for the reasons, firstly, inadequate clinical researches were conducted thus far in this field, owing to the easily-misinterpreted dissatisfactory postoperative visual quality for PACD patients as a result of glaucomatous neuropathy rather than caused by predicted refractive error of implanted IOL. Moreover, we found a greater number of studies focusing on the assessment of IOL power calculation formulas with respect to short AL or shallow AC [13, 30, 32–34], which were seemingly associated with angle-closure conditions. Nevertheless, short AL or shallow AC were not necessarily the major or only factors that contribute to PACD. Any abnormality in iris configurations, lens thickness, ciliary body conformation, zonular fiber tension or choroidal volume could contribute to the occurrence of PACD [1]. In a word, PACD, especially PAC/PACG, is a spectrum of diseases with complex pathogenesis. Hence, we chose to conduct this meta-analysis of formula accuracy on PACD patients instead of on those with solely shorter AL or shallow AC.
In terms of the outcome measurements being evaluated and statistically analyzed, MAE was not necessarily the best choice. As a continuous data of averaged absolute error (AE) brought by different formulas, MAE were to be easily influenced by the distribution pattern of the original data. Since AE usually does not conform to Gaussian distribution, the MAE with its standard deviation is probably not the appropriate measure of central tendency and variability of the data. Nevertheless, as MAE had been introduced as a major outcome measurement by almost all the studies, it was still analyzed in this meta-analysis for a preliminary look at whether it was consistent with the percentages of cases within the refractive limits. For cases of asymmetric distributions of outcome measurements, the median absolute error (MedAE) would be a better representative of the central tendency, yet only a few studies provided the MedAE by formulas with even less information on the range of the original data [7, 24, 26], resulting in between-study comparisons of MedAE infeasible. Additionally, Holladay et al. recommended that SD of prediction error (PE) has shown to be perfect and tightly correlated with other parameters of assessing the accuracy of a IOL power calculation formula [35], however as SD is a random variable, it is not applicable in network meta-analysis. Furthermore, % ±0.50 and % ±1.00 D, as being available in most studies, were also recommended for revealing central and dispersion tendency regardless of the data distribution, which was thereby considered as being more reliable when evaluating formula accuracy.
Excluded studies
As a systemic review and network meta-analysis, some relevant studies carried out in PACD populations were excluded for quantitative analysis for certain considerations. One of the situations was two studies with PACD patients implanted with 3-piece IOL, one showing unexpected and obvious discrepancies with 1-piece IOL-implantation [36], and the other providing unavailable outcome variables [11]. Though in clinical practice nowadays, not a few specialists hold that 3-piece IOLs would result in better clinical outcomes in terms of deepening and stabilizing ACD, the limited researches made it impossible for us to draw a conclusion as to which type of IOL outperforms with regard to postoperative refraction when applied in PACD patients. Meanwhile, the unclear degree of discrepancies brought by the IOL type led to exclusion of this study. Further evaluation and subgroup analysis could be done in the future to demonstrate different behaviors between one-piece and three-piece IOL in the capsular bag only if more researches were to be conducted. Another scenario was that some studies enrolled PACD patients with prior anti-glaucoma surgeries other than LPI (mainly trabeculectomy). Given it had already been demonstrated that cataract surgery in patients with prior trabeculectomy had significantly greater refractive surprise than those in the control groups [19, 20], these studies were also excluded. We had also ruled out patients with ongoing acute angle-closure episodes in case that corneal edema and intraocular inflammation would pose unexpected effects on IOL power prediction, though those with signs of previous acute attack were deemed as acceptable after further evaluation.
Implications for future clinical practice and research
The clinical significance of this systemic review and meta-analysis was firstly to identify the potential preferable IOL power calculation formulas with higher accuracy in PACD eyes; and secondly to give advice on future studies in this field. For potential preferable formulas, we suggested that BUII, Kane and RBF 3.0 might exert better clinical outcomes, yet the modification of existing formulas or development of new formulas specifically for PACD eyes were the real way to get to the root of this unmet clinical need. To promote high-quality researches in the future, here are a few recommendations: 1) Researchers should mind that 1-piece and 3-piece IOL might exert distinct clinical outcomes, and dividing them apart when conducting analysis could be beneficial for evaluating both the formula accuracy and IOL stability in angle-closure eyes; 2) It is necessary to only include patients with no prior anti-glaucoma surgeries other than LPI, for no one could be sure about the exact degree of potential zonular fiber injuries, capsular disposition and irregular astigmatism resulted from prior surgical interventions; 3) MedAE, SD of PE and the ratio of cases within certain limits of AE are considered as being more preferable, for they are more reliable and rational than MAE. Besides, as Holladay et al. suggested, attention should be paid if unequal variance exists among outcome variables derived from different formulas, where heteroscedastic method for data analysis should be performed [35]; 4) For measurement accuracy, unified application of new-generation optical biometry like IOLMaster or Lenstar have been recommended for ocular biometry, and subjective manifest refraction is recommended for refractive measurements; 5) IOL-specific constants ought to be optimized according to Langenbucher et al. [37, 38], and a cross-validation strategy is advised if possible. If not, downloading constants from the User Group for Laser Interference Biometry (ULIB) or using the Excel Query/What IF function [7] are other options; 6) Eyes with a postoperative visual acuity less than 0.4 should not be enrolled for formula constant optimization or accuracy assessment. All these aspects send expectations for the future development of both high-quality studies and more appropriate formulas for PACD patients, the breakthrough of which lying in the identification of essential factors that significantly lead to postoperative refractive errors in PACD patients, e.g., the lens vault (LV) [7], the ACD/AL, the relative position of ciliary process, etc. On the basis of these factors, new formulas might be generated, bringing hope and brighter future for those who expect to be rescued from the otherwise possibly dissatisfactory visual outcomes.
Conclusion
This systemic review and network meta-analysis of limited evidence suggested that no significant superiority was revealed among the commonly used SRK/T, Hoffer Q, Holladay 1, Haigis, BUII, Kane, RBF 3.0 and LSF formulas in eyes with PAC/PACG. Further well-designed studies are warranted to validate the efficacy of various formulas among PACD, yet modified or even new formulas derived specifically for PACD patients should come into being if none of these current formulas are stably satisfactory.
Supporting information
S1 Checklist. Adoption of the PRISMA 2009 checklist for reporting of systematic reviews and meta-analyses.
https://doi.org/10.1371/journal.pone.0276286.s001
(DOC)
S1 Fig. Consistency test for MAE.
Between-study heterogeneity of MAE was measured using mtc.anohe package in R, with the results showing consistency in both pair-wise and network comparison.
https://doi.org/10.1371/journal.pone.0276286.s002
(TIFF)
S2 Fig. Consistency test for % ±0.50 D and % ±1.00 D.
Between-study heterogeneity of A. % ±0.50 D and B. % ±1.00 D was measured using mtc.anohe package in R, with the results showing consistency in all the network comparison.
https://doi.org/10.1371/journal.pone.0276286.s003
(TIFF)
S3 Fig. Network funnel plot.
Risk of publication bias of A. MAE, B. % ±0.50 D and C. % ±1.00 D was measured using network funnel plot showing symmetry, indicating relatively small publication bias.
https://doi.org/10.1371/journal.pone.0276286.s004
(TIFF)
S4 Fig. The plotted node-splitting analysis.
The decision rule selected Haigis-Hoffer Q comparisons for MAE (A), Haigis-BUII, Haigis-Hoffer Q and Haigis-Kane comparisons for % ±0.50 D (B) and % ±1.00 D (C) because of presence of both direct and indirect comparisons in these pairs. As reflected by P-value, the direct and indirect comparisons were in agreement of each other. MAE: mean absolute error; BUII: Barrett Universal II.
https://doi.org/10.1371/journal.pone.0276286.s005
(TIFF)
S1 Table. Detailed search results from different sources.
https://doi.org/10.1371/journal.pone.0276286.s006
(DOCX)
S2 Table. Risk of bias of included studies.
Level of evidence as recommended by the Oxford Centre for Evidence-based Medicine. *Retrospective Observational Studies.
https://doi.org/10.1371/journal.pone.0276286.s007
(PDF)
References
- 1. Sun X, Dai Y, Chen Y, Yu DY, Cringle SJ, Chen J, et al. Primary angle closure glaucoma: What we know and what we don’t know. Prog Retin Eye Res. 2017;57:26–45. Epub 2017/01/01. pmid:28039061.
- 2. Lin Z, Liang Y, Wang N, Li S, Mou D, Fan S, et al. Peripheral anterior synechia reduce extent of angle widening after laser peripheral iridotomy in eyes with primary angle closure. J Glaucoma. 2013;22(5):374–9. Epub 2013/05/21. pmid:23685912.
- 3. Razeghinejad MR, Myers JS. Contemporary approach to the diagnosis and management of primary angle-closure disease. Surv Ophthalmol. 2018;63(6):754–68. Epub 2018/05/20. pmid:29777727.
- 4. Azuara-Blanco A, Burr J, Ramsay C, Cooper D, Foster PJ, Friedman DS, et al. Effectiveness of early lens extraction for the treatment of primary angle-closure glaucoma (EAGLE): a randomised controlled trial. The Lancet. 2016;388(10052):1389–97. pmid:27707497
- 5. Kang SY, Hong S, Won JB, Seong GJ, Kim CY. Inaccuracy of intraocular lens power prediction for cataract surgery in angle-closure glaucoma. Yonsei Med J. 2009;50(2):206–10. Epub 2009/05/12. pmid:19430552; PubMed Central PMCID: PMC2678694.
- 6. Khoramnia R, Auffarth G, Labuz G, Pettit G, Suryakumar R. Refractive Outcomes after Cataract Surgery. Diagnostics (Basel). 2022;12(2). Epub 2022/02/26. pmid:35204334; PubMed Central PMCID: PMC8870878.
- 7. Seo S, Lee CE, Kim YK, Lee SY, Jeoung JW, Park KH. Factors affecting refractive outcome after cataract surgery in primary angle-closure glaucoma. Clin Exp Ophthalmol. 2016;44(8):693–700. Epub 2016/04/16. pmid:27082207.
- 8. Wanichwecharungruang B, Amornpetchsathaporn A, Wongwijitsook W, Kongsomboon K, Chantra S. Evaluation of ocular biometry in primary angle-closure disease with two swept source optical coherence tomography devices. PLoS One. 2022;17(3):e0265844. Epub 2022/03/22. pmid:35312733; PubMed Central PMCID: PMC8936461.
- 9. Lee TE, Yoo C, Kim YY. The effects of peripheral anterior synechiae on refractive outcomes after cataract surgery in eyes with primary angle-closure disease. Medicine (Baltimore). 2021;100(14):e24673. Epub 2021/04/10. pmid:33832065; PubMed Central PMCID: PMC8036052.
- 10. Hoffer KJ. The Hoffer Q formula: A comparison of theoretic and regression formulas. Journal of Cataract and Refractive Surgery. 1993;19(6):700–12. pmid:8271165
- 11. Rhiu S, Lee ES, Kim TI, Lee HS, Kim CY. Power prediction for one-piece and three-piece intraocular lens implantation after cataract surgery in patients with chronic angle-closure glaucoma: a prospective, randomized clinical trial. Acta Ophthalmol. 2012;90(8):e580–5. Epub 2012/09/14. pmid:22971184.
- 12. Kim YC, Sung MS, Heo H, Park SW. Anterior segment configuration as a predictive factor for refractive outcome after cataract surgery in patients with glaucoma. BMC Ophthalmol. 2016;16(1):179. Epub 2016/10/21. pmid:27756264; PubMed Central PMCID: PMC5070221.
- 13. Gokce SE, Montes De Oca I, Cooke DL, Wang L, Koch DD, Al-Mohtaseb Z. Accuracy of 8 intraocular lens calculation formulas in relation to anterior chamber depth in patients with normal axial lengths. J Cataract Refract Surg. 2018;44(3):362–8. Epub 2018/04/02. pmid:29605282.
- 14. Wang Q, Jiang W, Lin T, Wu X, Lin H, Chen W. Meta-analysis of accuracy of intraocular lens power calculation formulas in short eyes. Clin Exp Ophthalmol. 2018;46(4):356–63. Epub 2017/09/10. pmid:28887901.
- 15. Wei L, Meng J, Qi J, Lu Y, Zhu X. Comparisons of intraocular lens power calculation methods for eyes with previous myopic laser refractive surgery: Bayesian network meta-analysis. J Cataract Refract Surg. 2021;47(8):1011–8. Epub 2021/07/23. pmid:34290197.
- 16. Liberati A, Altman DG, Tetzlaff J, Mulrow C, Gotzsche PC, Ioannidis JP, et al. The PRISMA statement for reporting systematic reviews and meta-analyses of studies that evaluate healthcare interventions: explanation and elaboration. BMJ. 2009;339:b2700. Epub 2009/07/23. pmid:19622552; PubMed Central PMCID: PMC2714672.
- 17. Hutton B, Salanti G, Caldwell DM, Chaimani A, Schmid CH, Cameron C, et al. The PRISMA extension statement for reporting of systematic reviews incorporating network meta-analyses of health care interventions: checklist and explanations. Ann Intern Med. 2015;162(11):777–84. Epub 2015/06/02. pmid:26030634.
- 18. Paul J Foster Ralf Buhrmann, Harry A Quigley Johnson GJ. The definition and classification of glaucoma in prevalence surveys. Br J Ophthalmol. 2002;86:238–42. pmid:11815354
- 19. Shah J, Vijaya L, Shantha B, George R. The effect of prior trabeculectomy on refractive outcomes of cataract surgery. Am J Ophthalmol. 2013;156(5):1070–1. Epub 2013/10/22. pmid:24138756.
- 20. Zhang N, Tsai P, Catoira-Boyle Y, Morgan L, Hoop J, Cantor L, et al. The effect of prior trabeculectomy on refractive outcomes of cataract surgery. Am J Ophthalmol. 2013;155(5):858–63. pmid:23398980
- 21. Li Y, Guo C, Huang C, Jing L, Huang Y, Zhou R, et al. Development and Evaluation of the Prognostic Nomogram to Predict Refractive Error in Patients With Primary Angle-Closure Glaucoma Who Underwent Cataract Surgery Combined With Goniosynechialysis. Front Med (Lausanne). 2021;8:749903. Epub 2022/01/04. pmid:34977061; PubMed Central PMCID: PMC8714900.
- 22. Whiting P, Rutjes A, Westwood M, Mallett S, Deeks J, Reitsma J, et al. QUADAS-2: A Revised Tool for the Quality Assessment of Diagnostic Accuracy Studies. Ann Intern Med. 2011;155(8):529–36. pmid:22007046
- 23. von Hippel PT. The heterogeneity statistic I2 can be biased in small meta-analyses. BMC Medical Research Methodology. 2015;15(1). pmid:25880989
- 24. Joo J, Whang WJ, Oh TH, Kang KD, Kim HS, Moon JI. Accuracy of intraocular lens power calculation formulas in primary angle closure glaucoma. Korean J Ophthalmol. 2011;25(6):375–9. Epub 2011/12/02. pmid:22131773; PubMed Central PMCID: PMC3223703.
- 25. Song WK, Sung KR, Shin JW, Kwon J. Effects of Choroidal Thickness on Refractive Outcome Following Cataract Surgery in Primary Angle Closure. Korean J Ophthalmol. 2018;32(5):382–90. Epub 2018/10/13. pmid:30311461; PubMed Central PMCID: PMC6182207.
- 26. Hou M, Ding Y, Liu L, Li J, Liu X, Wu M. Accuracy of intraocular lens power calculation in primary angle-closure disease: comparison of 7 formulas. Graefes Arch Clin Exp Ophthalmol. 2021;259(12):3739–47. Epub 2021/07/15. pmid:34258655.
- 27. Aristodemou P, Knox Cartwright NE, Sparrow JM, Johnston RL. Formula choice: Hoffer Q, Holladay 1, or SRK/T and refractive outcomes in 8108 eyes after cataract surgery with biometry by partial coherence interferometry. J Cataract Refract Surg. 2011;37(1):63–71. Epub 2010/12/25. pmid:21183100.
- 28. He M, Jiang Y, Huang S, Chang DS, Munoz B, Aung T, et al. Laser peripheral iridotomy for the prevention of angle closure: a single-centre, randomised controlled trial. The Lancet. 2019;393(10181):1609–18. pmid:30878226
- 29. Mou DP, Liang YB, Fan SJ, Peng Y, Wang NL, Thomas R. Progression rate to primary angle closure following laser peripheral iridotomy in primary angle-closure suspects: a randomised study. Int J Ophthalmol. 2021;14(8):1179–84. Epub 2021/08/21. pmid:34414081; PubMed Central PMCID: PMC8342280.
- 30. Darcy K, Gunn D, Tavassoli S, Sparrow J, Kane JX. Assessment of the accuracy of new and updated intraocular lens power calculation formulas in 10 930 eyes from the UK National Health Service. J Cataract Refract Surg. 2020;46(1):2–7. Epub 2020/02/13. pmid:32050225.
- 31. Xia T, Martinez CE, Tsai LM. Update on Intraocular Lens Formulas and Calculations. Asia Pac J Ophthalmol (Phila). 2020;9(3):186–93. Epub 2020/06/06. pmid:32501896; PubMed Central PMCID: PMC7299214.
- 32. Hoffer KJ, Savini G. IOL Power Calculation in Short and Long Eyes. Asia Pac J Ophthalmol (Phila). 2017;6(4):330–1. Epub 2017/08/07. pmid:28780778.
- 33. Yang S, Whang WJ, Joo CK. Effect of anterior chamber depth on the choice of intraocular lens calculation formula. PLoS One. 2017;12(12):e0189868. Epub 2017/12/19. pmid:29253884; PubMed Central PMCID: PMC5734766.
- 34. Melles RB, Holladay JT, Chang WJ. Accuracy of Intraocular Lens Calculation Formulas. Ophthalmology. 2018;125(2):169–78. Epub 2017/09/28. pmid:28951074.
- 35. Holladay JT, Wilcox RR, Koch DD, Wang L. Review and recommendations for univariate statistical analysis of spherical equivalent prediction error for IOL power calculations. J Cataract Refract Surg. 2021;47(1):65–77. Epub 2020/08/10. pmid:32769751.
- 36. Wu Y, Zhang S, Zhong Y, Bian A, Zhang Y, Wang Z. Prediction of effective Lens position using anterior segment optical coherence tomography in Chinese subjects with angle closure. BMC Ophthalmol. 2021;21(1):454. Epub 2021/12/29. pmid:34961542; PubMed Central PMCID: PMC8711180.
- 37. Langenbucher A, Haigis W, Seitz B. Difficult lens power calculations. Curr Opin Ophthalmol. 2004;15(1):1–9. pmid:14743011
- 38. Langenbucher A, Szentmary N, Cayless A, Muller M, Eppig T, Schroder S, et al. IOL Formula Constants: Strategies for Optimization and Defining Standards for Presenting Data. Ophthalmic Res. 2021;64(6):1055–67. Epub 2021/02/03. pmid:33530082; PubMed Central PMCID: PMC8743903.