A machine learning approach in the non-invasive prediction of intracranial pressure using Modified Photoplethysmography

  • Anmar Abdul-Rahman ,

    Roles Conceptualization, Data curation, Formal analysis, Investigation, Methodology, Project administration, Resources, Software, Validation, Visualization, Writing – original draft, Writing – review & editing

    anmar_rahman@hotmail.com

    Affiliation Department of Ophthalmology, Counties Manukau District Health Board, Auckland, New Zealand

  • William Morgan,

    Roles Conceptualization, Data curation, Investigation, Methodology, Project administration, Resources, Software, Supervision, Writing – review & editing

    Affiliations Centre for Ophthalmology and Visual Science, The University of Western Australia, Perth, Australia, Lions Eye Institute, University of Western Australia, Perth, Australia

  • Dao-Yi Yu

    Roles Conceptualization, Data curation, Investigation, Methodology, Project administration, Supervision, Writing – review & editing

    Affiliations Centre for Ophthalmology and Visual Science, The University of Western Australia, Perth, Australia, Lions Eye Institute, University of Western Australia, Perth, Australia

Abstract

The ideal Intracranial pressure (ICP) estimation method should be accurate, reliable, cost-effective, compact, and associated with minimal morbidity/mortality. To this end several described non-invasive methods in ICP estimation have yielded promising results, however the reliability of these techniques have yet to supersede invasive methods of ICP measurement. Over several publications, we described a novel imaging method of Modified Photoplethysmography in the evaluation of the retinal vascular pulse parameters decomposed in the Fourier domain, which enables computationally efficient information filtering of the retinal vascular pulse wave. We applied this method in a population of 21 subjects undergoing lumbar puncture manometry. A regression model was derived by applying an Extreme Gradient Boost (XGB) machine learning algorithm using retinal vascular pulse harmonic regression waveform amplitude (HRWa), first and second harmonic cosine and sine coefficients (an1,2, bn1,2) among other features. Gain and SHapley Additive exPlanation (SHAP) values ranked feature importance in the model. Agreement between the predicted ICP mean, median and peak density with measured ICP was assessed using Bland-Altman bias±standard error. Feature gain of intraocular pressure (IOPi) (arterial = 0.6092, venous = 0.5476), and of the Fourier coefficients, an1 (arterial = 0.1000, venous = 0.1024) ranked highest in the XGB model for both vascular systems. The arterial model SHAP values demonstrated the importance of the laterality of the tested eye (1.2477), which was less prominent in the venous model (0.8710). External validation was achieved using seven hold-out test cases, where the median venous predicted ICP showed better agreement with measured ICP. Although the Bland-Altman bias from the venous model (0.034±1.8013 cm water (p<0.99)) was lower compared to that of the arterial model (0.139±1.6545 cm water (p<0.94)), the arterial model provided a potential avenue for internal validation of the prediction. This approach can potentially be integrated into a neurological clinical decision algorithm to evaluate the indication for lumbar puncture.

Introduction

Investigation of the physiological properties of the human cerebrospinal fluid was first described in 1891 when Quinke published his studies on the diagnostic and therapeutic applications of lumbar puncture. He standardized the technique of intracranial pressure (ICP) measurement by connecting the lumbar puncture needle with a fine glass pipette in which cerebrospinal fluid was allowed to rise, a method still currently applied [1]. Whereas data from a recent large international, multi-center study on lumbar puncture feasibility that included 3,868 patients confirmed the procedures’ safety [2], complications from lumbar puncture including iatrogenic meningitis, hemorrhage, post-lumbar puncture headache are well recognized [36]. In addition to the risks stated above, continuous ICP monitoring includes ventricular catheter-related problems such as cerebral infections, catheter occlusion, or malposition. [79]. Financial costs incurred at a single tertiary care institution due to hospitalization for post-lumbar puncture complications were estimated at $20,000 USD/year [10]. To mitigate these risks, various modalities of non-invasive ICP estimation have been described. Of these studies ophthalmodynamometry provided the earliest attempts at non-invasive ICP estimation, they were based on earlier observations by Deyl in 1898 who postulated that papilledema was due to central retinal vein compression where it emerged from the optic nerve into the subarachnoid space in the optic nerve’s dural sheath [11]. To further prove this mechanism, Cushing and Borley experimentally induced papilledema and described the loss of spontaneous venous pulsation due to collapse of the central retinal vein when ICP was raised [12]. Their findings were supported by a canine model, in which the temporal succession of events demonstrated that loss of spontaneous venous pulsation and venous dilation preceded optic disc enlargement was demonstrated [13]. In 1927 Baurmann provided the first evidence of a strong linear correlation between ICP and retinal venous pulse pressure measured by ophthalmodynamometry [14]. His results were replicated by other investigators in a series of animal studies [1519]. These findings were also reported in human studies by Firsching [20, 21], Motschmann [22] and co-workers. To improve ICP predictive accuracy, Querfurth et al. combined ophthalmodynamometry with simultaneous color doppler of the central retinal and ophthalmic arterial flow velocities. Although the linear correlation between retinal venous pulse pressure and ICP was strong. The combination of ophthalmodynamometry and color doppler parameters was shown to improve the correlation compared to either parameter alone [23]. Intuitively, a strong linear correlation would imply strong predictive accuracy, yet paradoxically these linear models did not achieve a strong predictive power expected from the linear correlation and none of these methods have superseded invasive methods of ICP measurement to date [2426].

The technique and instrumentation for ophthalmodynamometry was developed by Bajardi around 1906 for the indirect estimation of ocular perfusion [27]. Although this device and its several iterations was superseded by carotid doppler, we have over a series of publications [2834] described a novel imaging method of Modified Photoplethysmography where a combination of ophthalmodynamometry as a means of generating a range of induced intraocular pressures and slit-lamp imaging of the optic nerve enables modelling of the retinal venous and arterial pulse amplitude and timing characteristics. The physiological basis for the choice of a harmonic regression model in image analysis is to emulate the photoplethysmography wave periodic and non-periodic components [35]. The model consists of a linear spline, which represents the mean of the signal and adjusts for inter-frame image displacement. The first two harmonics of the Fourier series are fitted to the periodic component and a first-order autoregressive error component accounts for the error process. Frequency domain decomposition is performed using custom software, where heat maps of the retinal vascular pulse amplitude distribution are generated and Fourier coefficients of the first two harmonics are extracted. Additional to modelling the retinal arterial and venous systems separately using this method, uniquely, frequency domain analysis allows computationally efficient information filtering and comparative processing of the retinal vascular pulse characteristics [30, 36]. The trade-off in signal resolution between the temporal (Δt) and frequency (Δf) domains is defined by the Heisenberg-Gabor uncertainty principle. It is mathematically expressed as an inequality equation (Δf ⋅ ΔtC), where (C) is a constant with a value dependent on how the frequency is measured and conveys the general concept that a non-zero function and its Fourier decomposition cannot be localised to arbitrary precision [37]. Although information between these domains is conserved, unique hemodynamic phenomena can exist solely as a property of the frequency domain such as dispersion (frequency dependent velocity of a wave), impedance (frequency dependent resistance), and the harmonic amplitude distribution. Modified photoplethysmography requires imaging at a slit-lamp, therefore is best applied in an ophthalmology outpatient setting, this excludes the applicability of the method for critically ill patients in the emergency or intensive care setting. A sophisticated handheld system is currently under development by our group, which may make future iterations of the clinical system more versatile for ICP measurement. Although it is possible to perform the test by a single operator, we have used an observer to vocalize the force readings from the ophthalmodynamometer, this allows editing the videos as further detailed in the materials and methods section. Moreover, the technique requires patient cooperation to maintain fixation and remain seated as is required for slit-lamp ophthalmoscopy, this excludes patients with cognitive impairment. A reduced model of the pulsation event is possible from the output of Modified Photoplethysmography it cannot provide information on the hemodynamic pressure-flow wave and although it provides heat maps of the pulse amplitude and timing characteristics, further work is required to analyze vascular geometry specific changes of the pulse characteristics in the Fourier domain. The interaction between ICP, IOP, and the retinal vascular pulse may share characteristics of chaotic systems. In general, systems with at least two of the following properties are considered to be chaotic: bifurcation and period doubling, period three, transitivity and dense orbit, sensitive dependence to initial conditions, and expansivity [38]. Even though Modified Photoplethysmography, unlike other non-invasive methods of ophthalmic ICP estimation, provides continuous vascular observations to potentially resolve this question. The limitation in our work is date is due to the lack of concurrent continuous ICP and IOP measurements during retinal imaging, which particularly the latter, may not be possible out of an experimental / physical model setting. In a recently published study, we described the interaction of the harmonic regression amplitude (HRWa) distribution in the retinal vascular system with intraocular and intracranial pressure using a linear mixed effects model. This approach enabled the computation of the variance estimated by these variables in linear space. It was demonstrated that linear interactions of IOP, ICP and the retinal vascular pulse accounted for less than 10% of the variance. This poor explanatory power of a linear model precludes it as a predictor of ICP. This was due to a non-constant variance of the error term of the predictors (heteroscedasticity), which indicated that the correlation was linear within individuals but non-linear between individuals [28]. Therefore, addressing this non-linear component is crucial to achieving high ICP predictive accuracy. In this study, we applied an Extreme Gradient Boost (XGB) supervised machine learning decision tree algorithm in non-invasive ICP estimation from features extracted from the retinal arteries and veins separately. This approach addresses the non-linear component in the data structure and model fit was adjusted through tuning of the hyperparameters. Additionally, parallelization and distributed computing allowed for rapid data computation run times. We hypothesize that the combination of Fourier-domain decomposition of the retinal vascular pulse wave parameters and a regression model derived from a decision tree machine learning algorithm could provide an avenue to non-invasive ICP prediction.

Materials and methods

Subject recruitment

Twenty-eight participants were recruited prospectively from the Lions Eye Institute over five years (2015–2020) from referrals made to the clinic for ophthalmic assessment before lumbar puncture for suspicion of idiopathic intracranial hypertension. Study approval was obtained from the University of Western Australia Human Ethics Committee adhering to the tenets of the Declaration of Helsinki. Participants were required to have clear ocular media, no prior history of co-existing retina or optic nerve disease, and were needed to be able to cooperate with the imaging protocol. Written consent was obtained from each of the participants. Lumbar puncture was performed in the lateral decubitus position and measured in centimeter (cm) water. The ophthalmic examination consisted of measurement of visual acuity, Goldmann tonometry, slit-lamp examination, color fundus photography, and modified photoplethysmography. The latter test consists of contact lens ophthalmodynamometry, the purpose was to vary induced intraocular pressure (IOPi), with concomitant video imaging of the optic disc.

An ICP of 25 cm water was considered the upper normal limit [39], this threshold classified the twenty-one patients in the training and test study groups into ten cases in the high intracranial pressure group (ICPh>25cm water) and eight in the normal intracranial pressure group (ICPn≤25cm water). Three cases overlapped both groups as a result of interchanging between the ICPn to the ICPh groups over the observation period (Fig 1). A total of 129,600 data points were sampled from the images, 56,932 arterial and 72,668 venous data points (Fig 2). Three eyes were excluded from the analysis due to poor image quality. Additionally, seven subjects were recruited for model validation. data from these cases were not subjected to the training or testing phases of the analysis.

thumbnail
Fig 1. Classification of the study population.

Training and test groups study groups, patients with normal intracranial pressure ICPn (ICP≤25cm water) and high intracranial pressure ICPh (ICP>25cm water).

https://doi.org/10.1371/journal.pone.0275417.g001

thumbnail
Fig 2. Data points from training and test study groups.

The data points are sub-classified into arterial and venous points, in patients with normal intracranial pressure ICPn (ICP≤25cm water) and high intracranial pressure ICPh (ICP>25cm water).

https://doi.org/10.1371/journal.pone.0275417.g002

Image acquisition

Data capture was performed by an operator and an observer in an outpatient setting. While the operator concentrated on the imaging task, the observer vocalized the ophthalmodynamometric force, this was required to create the splice points for video editing. The test requires patient cooperation to sit upright, remain stationary and fixate on a target as required for slit-lamp ophthalmoscopy. Details for the image acquisition workflow is published in earlier work [30]. The optic nerve was imaged under a dynamic range of intraocular pressures using a Meditron ophthalmodynamometer (Meditron GmbH, Poststrasse, Völklingen, Germany). This device consists of a sensor ring, which measures the compression force on the eye. The sensor surrounds a central Goldmann three-mirror fundus contact lens. The ophthalmodynamometric force (ODF) displayed as Meditron units (mu), which were then converted to induced intraocular pressure (IOPi) as described by Morgan et al. [40] using the following formula: (1) Where IOPb is the baseline intraocular pressure (IOP) in millimeters mercury (mmHg). Video of the optic nerve was captured with an imaging slit-lamp (Carl Zeiss, Germany) with a mounted digital camera (Canon 5D Mark III, Japan). Several sequences of at least three cardiac cycles in length were taken, each at a rate of 25 frames/second. When possible recordings were taken from both eyes. A range of induced intraocular pressure values was between 7–73 mmHg were obtained from each subject. Videos showing motion artifact, reflection from optical media, or decentration of the optic nerve in the image sequence for less than three consecutive cardiac cycles were rejected from the analysis. A pulse oximeter (Nellcor N65, Covidien, Mansfield, MA) was applied to the right index finger; the audio signal from the pulse oximeter was recorded with the video sequence of the optic nerve. This allowed synchronization of the retinal vascular pulse with the cardiac cycle. Timing of the cardiac cycle was generated from the audio signal from the subject’s pulse oximetry recorded on the audio trace of the video segment, which in turn enabled the mathematical analysis of the periodic component from green channel transmittance. A Single high-quality three-cardiac cycle length video recording was extracted from each recording session.

Image analysis

Image processing was done in Adobe Photoshop CS6, Individual image frames were extracted from each video sequence and saved as Tagged Image File Format (TIFF) files. Each of these images was cropped to an array of pixels. All images from three cardiac cycles were analyzed in R statistical package using custom software [41]. Each data point was represented by the mean of the green channel intensity at time measured as a fraction of the cardiac cycle, rather than in seconds. The periodic trend component was modelled separately for the arteries and veins as a harmonic regression waveform expansion: (2) f(t)p = The periodic component of the time series.

a0 = Coefficient representing the mean of f(t)p.

an = Coefficient of the cosine function of f(t)p.

bn = Coefficient of the sine function of f(t)p.

n = Integer 0,1,2… etc representing the harmonic component.

ϵ = error term

Higher harmonic frequency model comparisons were conducted using Akaike Information Criterion (AIC). In most eyes AIC preferred models with first and second-order frequencies, therefore the final analysis was limited to the first and second harmonics. A harmonic regression model was fitted to each pixel in the time series and used to quantify the retinal vascular pulse wave parameters including the harmonic regression wave amplitude (HRWa). The model includes a Fourier series representation using the first and second harmonics, linear spline non-periodic component, and a first-order autoregressive error component. Timing attributes captured by the custom software include the cardiac cycle time and time to the minimum point of the harmonic regression wave (time to trough) measured in fractions of the cardiac cycle as indicated from the audio pulse oximetry signal. Image analysis and model fit are detailed in previous publications [30, 33, 36].

Machine learning algorithm-Extreme Gradient Boost (XGB)

Extreme gradient boost algorithm is an ensemble machine learning regression method based on decision trees that use the gradient descent architecture to boost weak learners [42]. Boosting builds decision trees sequentially such that each subsequent tree aims to reduce the errors of the previous tree and the residual errors are then updated. R statistical package [41] was used to generate the model for each vascular system independently. Bayesian optimization was used to tune seven of the model hyperparameters aimed at regulating the model fit. Five-fold cross-validation with ten early stopping rounds was applied in this step. Tuned hyperparameters included lambda (λ) L2 regularization term on weights (analogous to Ridge regression), this parameter reduces the influence of outliers by factoring in the denominator of the similarity score (defined as the ratio of the sum of the residuals squared and the number of the residuals). Gamma (γ) is the minimum loss reduction required to make a further partition on a leaf node of the tree. The larger gamma is, the more conservative the algorithm. Eta (η), the learning rate, is a scalar that determines step size in gradient descent and shrinks the weights on each step. Alpha (α) L1 regularization term on weights (analogous to Lasso regression). The maximum depth of the tree (max_depth) is the maximum number of nodes allowed from the root to the farthest leaf of a decision tree, deeper trees can model more complex relationships by adding more nodes. Minimum child weight (min_child_weight) is the minimum number of samples required to create a new node in the tree. Subsample corresponds to the fraction of observations (the rows) to subsample at each step, and nrounds the number of model iterations [42, 43]. Hyperparameters values for each model are listed in Table 1. If parameters are not set, default values are chosen by the XGB algorithm.

thumbnail
Table 1. Bayesian optimised Extreme Gradient Boost hyperparameters.

https://doi.org/10.1371/journal.pone.0275417.t001

In our study the training data consisted of nine numerical features (IOPi, HRWa, the cosine and sine coefficients of the first and second harmonic waves (an1,2), bn1,2), hemiretinal location of the vessel (superior or inferior retina), and laterality (right or left eye)). The training labels which were the ICP measured by lumbar puncture in the lateral decubitus position in centimeter (cm) water, were ultimately used to generate the ICP predictions. The training set consisted of 80% of randomly selected vascular pulsation points, the test set consisted of the remaining 20% of the data points. Model parameters were assessed using feature importance, which is a ranking score representing the contribution from the selected feature to the model prediction. It is calculated for a single decision tree by the amount that each attribute split point improves the performance measure, weighted by the number of observations for which the node is responsible. There are three methods for measuring feature importance in XGB, frequency (weight), which is the number of times a feature is used to split the data across all trees. Cover is the number of times a feature is used to split the data across all trees weighted by the number of training data points that go through those splits, and gain is the average training loss gained when using a particular feature at a branching point. Gain, therefore, represents the refinement in accuracy brought by a feature to the branches of the decision tree. XGB divides feature importance by default into two clusters, cluster one contains features of the highest importance to the model, and other features are aggregated in cluster two. To identify the main features driving model prediction, SHAP (SHapley Additive exPlanations) values were calculated, this is an additive feature attribution method that provides a quantitative evaluation of the tree ensemble’s overall impact in the form of particular feature contributions [44, 45]. External validation was achieved using seven hold-out test cases, which were evaluated in a blinded manner. From the hold-out test set mean, median and peak density (defined as the maximum point generated from the distribution of predicted ICPs from each model) of the predicted ICP were compared to the measured ICP. Both Bland-Altman plots and the t-test were used to measure the mean difference between predicted and measured ICP for both the arteries and the veins separately. A flow chart of image processing, analysis, and XGB model application is summarised in Fig 3.

thumbnail
Fig 3. Flow chart of image processing, analysis, and XGB model application.

1) Alignment and segmentation of image frames from a video sequence spanning three consecutive cardiac cycles captured from 21 subjects using Modified Photoplethysmography. 2) Image analysis was performed by fitting a harmonic regression model to each pixel cluster. The periodic component was represented by the first two harmonics of a Fourier series. 3) The Fourier coefficients, harmonic regression amplitude, together with the distance along the vessel, induced intraocular pressure (IOPi), ocular laterality (right/left), and hemiretinal locus of the blood vessel (superior/inferior) constituted the model features. 4) Separate retinal arterial (XGBA) and venous (XGBV) Extreme gradient Boost models were constructed, where an 80/20% split was chosen for the training/test set for each vascular model. 5) Seven hold-out test cases were used for external model validation and intracranial pressure (ICP) prediction.

https://doi.org/10.1371/journal.pone.0275417.g003

Statistical analysis

The distribution of the HRWa and the majority of the Fourier coefficients was non-normal, therefore the median was used as a measure of central tendency and the interquartile range (IQR) was used to assess the dispersion of this measure. Where appropriate the mean and standard deviation were reported. The range, minimum, and maximum of these parameters were also computed. When the Levene test was applied to assess the multifactorial homogeneity of variance of the predictors, heteroscedasticity was demonstrated as the assumption of homogeneity of variance was violated (p<0.0001). The Kruskal-Wallis test was used in the hypothesis test of the differences in the medians, and the paired Wilcoxon test with Bonferroni–Holm correction was used for posthoc analysis. Model fit was assessed using R2 square, which is a comparison of the residual sum of squares with the total sum of squares. Model Prediction accuracy was estimated by calculating the Mean Squared Error (MSE), defined as the square of the difference between the predicted and actual values of the test set, it assigns more weight to larger errors. Root mean square error (RMSE), which is the standard deviation of the residuals (prediction errors), the higher the number the greater the standard deviation σ of the distribution of errors. MSE and RMSE are used to evaluate the influence of outliers on predictions. The mean absolute error (MAE) calculated by the magnitude average difference between the predicted and actual values of the test set was also reported.

Results

Descriptive statistics

There were a total of twenty females (95.2%) and one male (4.8%) in the study population. The age demonstrated a bimodal distribution with a mean of 32 years (sd 8.32, range 17–47 years). In the ICPn group median ICP was 18.50 cm water (range 9.50 to 24, IQR = 6), the corresponding values in ICPh group were 31 cm water (range 25.50 to 68, IQR = 10). Table 2 demonstrates the Fourier wave amplitude descriptive parameters in both study groups. Hypothesis tests within ICP group differences were contrasted by vessel type, statistical significance (p<0.0001) was achieved for all Fourier parameters except the bn2 coefficient in the ICPh group, which demonstrated no statistically significant difference between the retinal arteries and the retinal veins. Both venous and arterial Fourier wave amplitude descriptive parameters stratified by ICP (between-group differences) demonstrated statistical significance (p<0.001) for all parameters except the retinal venous bn1, and arterial an2 coefficients. The ICPh group showed a lower median retinal venous (4.743 vs 5.314) and a higher arterial HRWa (4.559 vs 4.139) compared to the ICPn group (Fig 4).

thumbnail
Fig 4. Violin plots of the harmonic regression waveform amplitude (HRWa).

The central marker in the violin plots indicate the median and interquartile range. Noted are the reduction of the difference in maximum and median retinal vascular pulsation amplitudes within the ICP groups as a consequence of a reduction of the venous and increase in the arterial pulsation amplitudes within the ICPh group. Between group differences indicate a ICPh and ICPn groups.

https://doi.org/10.1371/journal.pone.0275417.g004

thumbnail
Table 2. Summary descriptive statistics for the high intracranial pressure (ICPh) and normal intracranial pressure (ICPn) groups.

https://doi.org/10.1371/journal.pone.0275417.t002

Machine learning model

Model fit and feature importance.

The model fit was was comparable for the arterial and venous data as indicated by an R2=0.89 and R2=0.91 respectively. Other accuracy parameters were similar for the arterial (MSE = 10.99, MAE = 2.03, RSME = 3.32) and the venous model (MSE = 11.85, MAE = 2.11, RSME = 3.44). The venous decision tree was more complex than the arterial, whereas the venous model was composed of a total of 451 nodes, 450 edges, and 35,589 leaves, the arterial model consisted of a total of 137 nodes, 136 edges, and 7,951 leaves. Model complexity and fit of the arterial and venous models are demonstrated in (Figs 5 and 6). The weighted cover in the figures represents the distribution of the average weighted number of residuals clustered in leaves at a certain depth of the decision tree.

thumbnail
Fig 5. Retinal arterial model fit.

The arterial model consisted of a total of 137 nodes, 136 edges, and 7,951 leaves. The model had an R2 of 0.89, and other accuracy parameters were: MSE = 10.99, MAE = 2.03, RSME = 3.32.

https://doi.org/10.1371/journal.pone.0275417.g005

thumbnail
Fig 6. Retinal venous model fit.

The venous model was composed of a total of 451 nodes, 450 edges, and 35,589 leaves. The model had a higher R2 of 0.91, other accuracy parameters were: MSE = 11.85, MAE = 2.11, RSME = 3.44.

https://doi.org/10.1371/journal.pone.0275417.g006

Global feature importance ranks the nine features of each vascular model by the feature gain, cover and frequency used in the prediction of ICP (Figs 7 and 8). For the arterial model these were IOPi (0.6092), an1 (0.1000) and HRWa (0.0804), similarly for the venous model IOPi (0.5476) and an1 (0.1024), dominated the feature importance, however unlike the arterial model HRWa (0.124) showed higher importance compared to an1 coefficient. When feature frequencies of the arterial and venous models were compared IOPi (0.2547 vs 0.2028), an1 (0.1357 vs 0.1362) and HRWa (0.1228 vs 0.1291), accounted for approximately 51–47% of each model’s feature importance respectively (Table 3).

thumbnail
Fig 7. Importance plot retinal arterial model.

IOPi, an1 and HRWa were the most important features in this model. IOPi = Induced intraocular pressure, HRWa=Harmonic regression wave amplitude, an1,2 = the cosine coefficient of the first and second harmonics, bn1,2 = the sine coefficient of the first and second harmonics, laterality = left / right eye, Distance = distance along the retinal vessel measured in mm, hemiretina = superior / inferior retina.

https://doi.org/10.1371/journal.pone.0275417.g007

thumbnail
Fig 8. Importance plot retinal venous model.

IOPi, HRWa, and an1 were the most important features in this model. IOPi = Induced intraocular pressure, HRWa=Harmonic regression wave amplitude, an1,2 = the cosine coefficient of the first and second harmonics, bn1,2 = the sine coefficient of the first and second harmonics, laterality = left / right eye, Distance = distance along the retinal vessel measured in mm, hemiretina = superior / inferior retina.

https://doi.org/10.1371/journal.pone.0275417.g008

thumbnail
Table 3. Feature importance of the machine learning models.

https://doi.org/10.1371/journal.pone.0275417.t003

Model SHAP values.

The SHAP summary plot (Figs 9 and 10) combines feature importance with feature effects, therefore it allows to explore interactions between features for the predicted variable. It is important to consider that SHAP values do not identify causality [45, 46]. Four properties can be derived from the SHAP summary plot:

  1. Feature importance: features are ranked on the y-axis in descending order according to their importance in the prediction of ICP from each vascular model.
  2. Impact: SHAP measures the impact of variables taking into account the interaction with other variables of the model. The horizontal location shows whether the effect of that value is associated with a higher or lower prediction. This is accomplished by calculating the importance of a feature by comparing what a model predicts with and without the feature in every possible combination. The x-axis measures the SHAP value, which indicates the change in model output in log-odds.
  3. Value: The color scale indicates whether that variable is high (blue) or low (yellow) for that observation where every point represents a row from the original dataset. Overlapping points are jittered in the x-axis direction. It should be noted that categorical variables were numerically encoded for the analysis (left/inferior = 0, right/superior = 1) for the laterality and hemiretinal location respectively giving a binary nature to the color scale.
  4. Correlation: Indicated by the relation between impact and color scale, a positive correlation occurs when high values of the feature have a positive predictive impact.
thumbnail
Fig 9. SHAP summary plot retinal arterial model demonstrating the feature contribution of the XGB model predicting ICP from the arterial model.

Induced intraocular pressure (IOPi) was the most important feature in the model (mean SHAP = 5.3884), approximately four times the value of the cosine coefficient of the first harmonic (an1 mean = 1.4689). Laterality = Right/Left eye, Distance = Retinal vascular pulsation amplitude as a function of distance from the center of the optic disc in mm, HRWa=Harmonic regression wave amplitude, an1,2, bn1,2=cosine and sine coefficients of the first and second harmonics, hemiretina = Superior/Inferior retina.

https://doi.org/10.1371/journal.pone.0275417.g009

thumbnail
Fig 10. SHAP summary plot retinal venous model demonstrating the feature contribution of the XGB model predicting ICP from the venous model.

Induced intraocular pressure (IOPi) was the most important feature in the model (mean SHAP = 5.723), approximately four times the value of the harmonic regression wave amplitude (HRWa mean = 1.425). HRWa=Harmonic regression wave amplitude, an1,2, bn1,2=cosine and sine coefficients of the first and second harmonics, laterality = Right/Left eye, Distance = Retinal vascular pulsation amplitude as a function of distance from the center of the optic disc in mm, hemiretina = Superior/Inferior retina.

https://doi.org/10.1371/journal.pone.0275417.g010

From both plots, it can be observed that the impact of IOPi on the model prediction was dependent on the value and showed a positive correlation, low IOPi values had a low predictive impact, and higher IOPi values had a positive impact on prediction, particularly for the venous model. This feature also showed higher dispersion than any of the tested features. The hemiretinal location of the tested vessel showed the lowest impact on model predictability. Laterality and distance of the data point along the vessel measured from the center of the optic disc attained more significance compared to the feature importance without variable interactions (Table 3), particularly for the arterial model, where there was a correlation with laterality, right eyes had negative and left eyes a positive impact on prediction, the venous model did not demonstrate this effect. Pulsation values obtained from vascular points in proximity to the optic disc had a less predictive impact and pulsation values from more peripheral locations in the vessel had a higher impact on model prediction (positive correlation).

The arterial model showed no correlation with HRWa values, furthermore, this feature was less significant when interactions were considered. In contrast, the venous model demonstrated a negative correlation, and it retained its significance both with and without variable interactions (Table 3).

The an1 Fourier coefficient had the highest feature importance of all coefficients in both models, and other coefficients had a low rank. The correlation of the coefficients was different for each model. Whereas the arterial model demonstrated that the an1,2, and bn1,2 showed low positive correlation on model predictability, the venous model on the other hand showed that an1,2 and bn1,2 had opposite correlation on predictability, where an1, bn1,2 values had a negative correlation and an2 had a positive correlation.

Mean SHAP values are listed in Table 4, where IOPi demonstrates the highest mean SHAP values in both vascular models (arterial = 5.3884 and venous = 5.6375), this was approximately four times the mean value of an1 (arterial = 1.4689 and venous = 1.3856) and the others among the three most significant features (arterial laterality = 1.2477, venous HRWa=1.7024).

thumbnail
Table 4. Mean SHAP values for the arterial and venous models.

https://doi.org/10.1371/journal.pone.0275417.t004

Model external validation

A group of seven cases was used as the holdout test set, in this group all cases were females. Four cases (57.1%) had an ICP > 25cm water. The median ICP in this group was 29.5cm water (range 26 to 32, IQR = 5.25). The remaining three cases (42.9%) had a median ICP of 20cm water (range 17 to 22, IQR = 2.5). Table 5 summarises the mean, median, and peak density of the predicted ICP from each vascular model. When the mean, median, and peak density for predicted ICP from the arterial and venous models are compared using the t-test and Bland-Altman bias statistic (Table 6), it is clear that the predicted ICP estimated from the venous median had the best agreement with measured ICP as indicated by the lowest Bland-Altman bias. A comparison of measured and median estimated ICP is demonstrated graphically in Bland-Altman plots (Figs 11 and 12).

thumbnail
Fig 11. Bland-Altman plot predicted median predicted intracranial pressure of the arterial Extreme Gradient Boost model.

The intervals of two standard deviations are considered as the concordance limits between the two measurements, accounting for 95% of the observed differences. The Bland-Altman bias was 0.139±1.6545 cm water (p<0.94), the arterial model provided a potential avenue for internal validation of the prediction.

https://doi.org/10.1371/journal.pone.0275417.g011

thumbnail
Fig 12. Bland-Altman plot predicted median predicted intracranial pressure of the venous Extreme Gradient Boost model.

The intervals of two standard deviations are considered as the concordance limits between the two measurements, accounting for 95% of the observed differences. The Bland-Altman bias from the venous model (0.034±1.8013 cm water (p<0.99)) was lower compared to that of the arterial model.

https://doi.org/10.1371/journal.pone.0275417.g012

thumbnail
Table 5. Hold-out test set comparing measured and predicted ICP (cm water) using the XGB models for both the arteries and veins.

https://doi.org/10.1371/journal.pone.0275417.t005

thumbnail
Table 6. Bland-Altman analysis and t-test, comparing between the measured and predicted ICP using XGB.

https://doi.org/10.1371/journal.pone.0275417.t006

Discussion

Extreme Gradient Boost demonstrated favorable accuracy in the non-invasive prediction of ICP applied to Modified Photoplethysmography data. Quantitative interference due to optical reflections, shadowing, blink, and motion artifact secondary to saccadic movements render ophthalmic imaging artifact prone. Modified Photoplethysmography controls these multiple sources of interference. The Goldmann contact lens used for optic nerve observation and imaging eliminates blinking and reduces motion artifact providing optical continuity, field stability, and reducing information degradation. Induced intraocular pressure generates a range vascular pulse amplitude responses, therefore, enabling comparative analysis of the pulse wave under a range of transmural pressures. Moreover, retinal vascular pulse wave decomposition in the Fourier domain allows for computationally efficient information filtering. The harmonic regression approach applied in image analysis not only adjusts for motion artifacts through its linear spline; it applies a statistical approach to evaluate the fit of the Fourier harmonics to the non-periodic component of the vascular pulse. Hence it facilitates the decision of rejecting an analysis output where the signal model fit does not achieve statistical significance at each image pixel cluster [30, 36].

Current clinical ophthalmic literature cites other approaches to non-invasive ICP prediction, these can be classified into clinical/retinal imaging (pupillometry, IOP, optical coherence tomography (OCT), fundus photography, ophthalmodynamometry), radiological (Ultrasonography (U/S)/Computed Tomography (CT)/Magnetic Resonance Imaging (MRI)), and electrophysiological methods (Flash Visual Evoked Potentials (VEP)). The accessibility of the pupil and the development of objective quantitative hand-held modalities to measure pupillary diameter and function have made it an appealing option for the non-invasive estimation of ICP in an intensive care unit (ICU) setting. In the largest multi-center study to date, Chen et al. used the NeurOptics pupillometer to correlate the pupillary function with ICP in a population of 134 ICU patients. They described the temporal inverse relationship of pupillary reactivity with raised ICP, which they termed the Neurological Pupil index, which is derived by algorithmically transforming parameters involved in the pupillary light reflex. Interestingly, this approach had forecasting capability as the authors reported that an abnormal Neurological Pupil index preceded an ICP spike on average by 15.9 hours [47]. However, this approach yielded a course range of ICP values correlated with pupillary function. In a prospective observational study, Stevens et al. found a weak but statistically insignificant relationship between the Neurological Pupil index and ICP [48]. Hence, further research is required to establish the role of automated pupillometry in ICP estimation.

Tonometry has demonstrated inconsistent results in ICP prediction. Sajjadi et al. measured IOP using a Schiotz tonometer in 50 subjects who underwent lumbar puncture. They reported a strong correlation (R = 0.955, p<0.001) independent of body mass index, age, and neurological diagnosis [49]. Other investigators failed to replicate these results using different techniques of IOP estimation [5054].

Over the last two decades, OCT has been central to the diagnosis and management of optic nerve disorders, the earliest report was by Borchert et al. who patented a method to estimate ICP using OCT measurements of RNFL thickness; however, the authors do not provide the correlations between these variables necessary generate a prediction [55]. A multitude of parameters have been evaluated for potential estimation of ICP [5660]. In a multicenter study, Vijay et al. reported that optic nerve head central thickness was found to be the most closely associated parameter with ICP (R = 0.60–0.73) among a variety of macular and optic nerve protocols [61]. However, the association between the many OCT parameters, papilledema severity, and ICP is complex and remains undefined. Moreover, due to the need for subject cooperation, this test cannot be applied to patients with severe neurological disorders. To address this limitation, Andersen et al. used the retinal arterio-venous ratio as a biomarker of elevated ICP. Images were recorded using an Epicam portable camera, this method achieved a 94% (85–98%) sensitivity and 50% (34–66%) specificity in detecting patients with an ICP ≥ 20mmHg. Indicating that although there was a 94% probability of correctly identifying individuals with ICP ≥ 20 mmHg, this was mitigated by the 50% probability of misclassifying healthy individuals [62], thereby limiting the practical applicability of this approach.

Radiological imaging of retinal vascular parameters has been explored as a substitute for clinical methods especially since practicality demands the ability to predict ICP with minimal patient cooperation in the ICU setting. In a prospective case-control study, Jeub et al. used transbulbar sonography for the measurement of vascular flow in the central retinal artery. At a threshold value of 11.0 cm/s, the peak systolic velocity predicted pathological ICP levels with a 70% sensitivity and 69% specificity [63]. Using spectral Doppler imaging, Miller et al. reported the reduction in blood flow velocity in both the central retinal artery and central retinal vein in 18 children with elevated ICP (p<0.02), however, the limited number of recruited subjects for this study precluded the calculation of statistical test accuracy parameters [64]. Ragauskas et al compared a two-depth transorbital Doppler technique to measure blood flow velocity in the ophthalmic artery with ultrasonographic measurement of optic nerve sheath diameter [65], this method was reported to have better diagnostic reliability for detecting elevated ICP compared to optic nerve sheath diameter measurement [6669]. An independent clinical validation study determined that this technique had a fair agreement to ICP measured using lumbar puncture [70]. The theoretical basis establishing a correlation between optic nerve sheath diameter and ICP was first suggested in 1968 by Hayreh et al. in a primate model [71]. Subsequent studies have imaged this parameter using U/S, CT, and MRI making it a favorable option for subjects with significant neurological impairment. In the largest study to date, Rajajee et al. found the optimal optic nerve sheath diameter for detection of ICP > 20 mmHg was >0.48 cm as measured by U/S, where a sensitivity of 96% and specificity of 94% were achieved [72]. Most studies indicate an optic nerve sheath diameter of >5 mm as the threshold for determining elevated ICP [7375]. Measurement of the optic nerve sheath diameter using CT or MRI can overcome this limitation. High agreement and reproducibility have been reported between these imaging modalities [7678]. However, radiological methods have high operational costs, and image interpretability is operator dependent particularly in the case of U/S. Some disorders such as subarachnoid hemorrhage may not be suited for this modality [79]. Moreover, the range of pressure correlation with optic nerve sheath diameter can be narrow, the iCOP study reported a favorable correlation between optic nerve sheath diameter and ICP between 3.7 mm Hg and 26.5 mm Hg [80]. The precision and accuracy of MRI measurements of optic nerve sheath diameter are yet to be defined, as well as the optimal measurement technique and the influence of the time course of ICP fluctuations on changes in optic nerve sheath dimension [81].

Flash Visual Evoked Potentials (FVEPs) can demonstrate the integrity of the visual pathway from the retina to the occipital cortex. A longer FVEP wave crest latency, a reduction in amplitude, and an increase in wave width have been observed with elevated ICP. These findings were reported from two early studies where a relatively strong linear relationship (R2 ≈ 0.7) between FVEP N2 wave latency and ICP was observed. Using this method high correlation was particularly demonstrated at ICP levels >300 mm water [82, 83]. These findings were replicated using either a single or combined modality with transcranial doppler [84, 85]. There are significant limitations with FEVPs, a significant level of expertise is required to administer the test, it is unsuitable in patients with bifrontal lobe pathology, retinal damage, or optic neuropathy [84]. Moreover, factors such as blood glucose concentration, the patient’s nerve conduction rate, and electrolytes levels can result in high variance in the FVEP waveform properties [86].

Early studies which applied ophthalmodynamometry to this research question correlated retinal venous pulse pressure as a single parameter with ICP [2022]. The wide range of linear correlations (R = 0.69–0.968) did not, however, yield a clear conclusion regarding the applicability of their findings. Further attempts to improve predictability such as seeking an optimal reliability cutoff point or excluding patients with papilledema either did not produce the intended practical outcome or further restricted its applicability [21]. Querfurth et al. introduced simultaneous color doppler monitoring of the central retinal and ophthalmic arterial flow velocities to improve predictability. They reported a more significant correlation (R = 0.95, p < 0.005) with combined parameter approach [23]. Modified photoplethysmography provides a quantitative ICP prediction along a continuous scale. Fourier domain decomposition of the retinal vascular pulse amplitude enables a selective inclusion of pulse wave harmonics hence, increasing the signal-to-noise ratio in the data set. In a recent study, we described a physiological model to estimate ICP where a plot of the induced intraocular pressure (x-axis) against the retinal venous pulse (y-axis) was measured using Modified Photoplethysmography. Predicted ICP was plotted from the x-axis intersection extrapolated from the peak retinal venous pulsation amplitude. A mean absolute error of 3.0 mmHg was achieved using this technique [29]. However, the XGB approach adds unique advantages: 1) The machine learning decision tree model addresses the heteroscedastic data structure directly and allows the model fit to be optimized as cases are added to the dataset thereby futureproofing the model’s performance. 2) An improved predictive analysis by a reduction in the mean bias to 0.034±1.80 cm water (0.025±1.32 mmHg). 3) The ability to generate an independent prediction from the retinal arteries and veins and from all points from the retinal vessels in the image field enables visualization of the distribution of the predictions in the form of a peak density plot and to draw comparisons from the independent outputs. Therefore the XGB approach provides a method of internally validating the prediction from each case. However, papilledema may reduce the predictive accuracy when the renormalization of the ICP is followed by a lag in regression of the optic nerve changes [87], this may impact diagnostic accuracy of post-treatment serial Modified Photoplethysmography imaging tests.

Artificial intelligence classification methods that use clinical, electrophysiologic, or radiographic data to discriminate between normal and high ICP have been described [8890]. Neural network classification models have yielded a total accuracy ranging between 70.2±4.5% to 92.05±2.25%. Golzan et al. estimated ICP using a neural network regression model derived from retinal venous pulse amplitude measured using the Dynamic Vessel Analyser, which is the only commercially available device used to quantitatively measure the retinal vascular pulse parameters. This device uses arbitrary units rather than frequency domain decomposition of the pulse wave. They reported a mean square error of ±1.27 mmHg in ICP prediction. However, they did not validate the model predictions on external cases, therefore, only internal validation results were reported. Moreover, only the venous pulse amplitude was used in the model prediction the arterial pulse characteristics was not taken into consideration in the model. In contrast to our method, the pulse amplitude was quantified using empirical units rather than a frequency domain decomposition strategy [91]. In our study, the XGB approach provided unique advantages over other machine learning options, hyperparameter tuning using Bayesian optimization (Table 1), mitigated model overfitting, this is the case where model output is not generalizable due to close fit to the sample data, a known limitation in decision tree machine learning algorithms. The wide normal ICP range (2–25 cm water) [39, 92] and the predominance of non-linear dynamics in the interactions between ICP, IOP, and the retinal vascular pulse [28] may both contribute to data heteroscedasticity, which is the non-constant distribution of the error term of the predictors. This was addressed by the unique ability of a decision tree regression to generate a prediction without specifying an average structure for the model. The specific pruning algorithm is a significant factor in addressing the heteroscedasticity [93]. Moreover, the XGB analysis approach requires minimal data pre-processing and neither normalization nor scaling are necessary [94, 95].

Intracranial pressure predictions derived from the venous model demonstrated better agreement with measured ICP compared to that provided by the arterial model. This may be due to the higher amplitude pulsation in the retinal veins compared to the arteries and the difference in venous pulsation amplitude between the ICPn and ICPh groups. Functional and structural differences between the arterial and venous systems, which in turn result in differences in wall tension and compliance may have played a role. Blood vessel walls consist mainly of water (70%), which is inelastic and incompressible, the remaining structure consists of a mesh of fibers with elastic properties. The fibrillary material consists of collagen, elastin, and smooth muscle cell layer in various percentages depending on the type and location of the vessel [96, 97]. Heterogeneity of structural building blocks of the vessel wall results in a radius-tension curve with non-linear characteristics. Whereas collagen fibers demonstrate a steep length-tension relationship and are closer to the linear relationship as expected from Hooke’s law, elastin has a flat length-tension relationship [98], and that of smooth muscle is modified by the level of contraction [96]. Additionally, there are geometric factors that augment the venous response to changes in transmural pressure, at physiological ranges, a relatively small increase of transmural pressure from 0 to 10 mm Hg increases the venous volume by ∼ 200%, which reflects a change in geometry from ellipsoidal to circular associated with an increased cross-sectional area. At supra-physiological venous transmural pressure (>40mmHg) there is a minimal increase in the relative volume as a result of the increase in compliance. In contrast, the arterial wall shows a smaller slope of the gradient between relative volume and transmural pressure accounting for the higher tolerance to transmural pressure [98]. Therefore, the arterial wall has curvilinear compliance over a range of transmural pressures, whereas venous walls have a semi-sigmoidal range of compliance. These structural and consequently functional differences may underlie the differences in vascular response to raised ICP and consequently may impact model predictive accuracy. Models derived from each vascular system misclassified one of the hold-out test cases with high ICP (case-2 in the arterial model and case-7 for the venous model). Additionally, both models predicted raised ICP for case 1 when the measured ICP was normal. The agreement between the predictions (≥25 cm water) derived from the XGB models from both vascular systems in case-1 (Table 5) raises the possibility that the predicted ICP may represent the actual ICP. This scenario highlights the challenges of lumbar puncture manometry. Studies based on continuous monitoring report that ICP generally measures between 5 -10 mm Hg above atmospheric pressure [99]. However, this value may fluctuate within a range of 30 cm water over 12 hours [100]. Therefore, a time delay between the lumbar puncture and the ophthalmic assessment may have resulted in a spurious outcome in this case. Furthermore, a precise ICP estimate requires a consistent manometry technique where the needle entry point needs to be on the same level as the midline of the spine, which should also be at the same level as the patient’s head. To complicate matters leakage around the needle can result in an erroneous estimate [6]. Further consideration needs to be given to the fact that ICP is estimated and measured from different levels. A historic assumption intracranially measured ICP (EVD-ICP) is equal to opening pressure measured at the lumbar spine (LP-ICP) [101, 102]. In recent years, this matter has been a subject of debate where some studies have documented agreement between intracranial and lumbar measurements [103], correlated [104, 105], and others have disputed both agreement and correlation [106, 107]. Among these studies, Lenfeldt et al. experimentally raised the ICP using computerized lumbar infusion in subjects with communicating hydrocephalus. Simultaneously measured LP-ICP and Brain tissue ICP demonstrated a strong correlation (R2 = 0.98) between the two ICP measurements throughout a pressure range of 0–600 mm water, the mean±standard deviation was -10±29 mm water. Therefore, LP-ICP correlates strongly with brain tissue-ICP in the absence of pathological obstruction to the cerebrospinal fluid flow [105].

Both feature importance (Figs 7 and 8) and SHAP summary plots (Figs 9 and 10) demonstrated that IOPi was the most significant parameter in both the arterial and venous models. Moreover, there was a positive correlation between IOPi and the impact on both XGB models. This indicates that ophthalmodynamometry is an important component of our image acquisition methodology. Moreover, the findings are consistent with the high correlation between retinal venous opening pressure and ICP described in ophthalmodynamometric studies in human [2023] and animal studies [18, 19, 108110]. Changes in the venous system from raised ICP arises from elevated downstream cerebral venous pressures and external pressure on the subarachnoid segment of the central retinal vein, which increases vascular resistance [111]. The literature has been less clear on the changes in retinal arterial circulation in this context. Moreover, it is intriguing that arterial pulsation parameters provided information sufficient to generate a prediction in spite of the less clinically recognised changes in the retinal arterial system with raised ICP. Our study demonstrated an increase in the arterial wall pulsation amplitude in the ICPh group albeit at a narrower range than that of the venous system (Fig 4). In a recently published study, in addition to the retinal vascular pulse amplitude antipodal effect in response to abnormally elevated intracranial pressure, where a decrease in retinal venous pulse was accompanied by an increase in the arterial pulse proportionate to raised ICP [28]. We postulated that this effect may represent functional differences in the transmural pressure gradient between the retinal arteries and the retinal veins, where the former exhibits a higher range than the latter, therefore the retinal artery demonstrates pulsations at higher induced IOPi values, this combined with raised venous pressure and reduced venous compliance and possibly augmentation of the retinal arterial pressure wave in association with elevated ICP. Other compensatory mechanisms known to modify the cerebrovascular dynamics through auto-regulatory control have been described in the retinal arterial system as well [112115]. However, the role of these mechanisms in our observations remain unresolved. We found that the SHAP values suggested that ocular laterality was a prominent factor in the predictive model, particularly that derived from the arterial system, it could be hypothesized that hemodynamic phenomena depend on physiologically inherent asymmetry in vascular dynamics play an important role in the optimization of cardiovascular functions [116, 117]. Recent studies have demonstrated the association between ocular laterality on retinal vascular occlusions [118, 119]. Although there are anatomic reasons for retinal arterial occlusions pertaining to embolic etiologies, the reason why asymmetry plays a role in the arterial rather than the venous system remains unclear.

A major limitation of machine learning algorithms is the black-box problem, given that the nature of machine learning is based on accuracy-driven performance metrics, these models will likely continue to become even more opaque in the future, especially machine learning-generated ensembles of decision trees [120123]. More recently, software packages like DALEX [124], breakdown [125] and XGBExplainer [126] have made some gains in terms of XGB model interpretability. A fundamental restriction inherent in all tree-based models is the inability to extrapolate target values beyond the range of the training data when making predictions, this is in contrast to linear models that can extrapolate predictions beyond the range of the training dataset [127]. Our study included a small number of participants, a larger dataset would likely offer improved generalizability. Image acquisition requires subject cooperation, which hinders the technique’s applicability in patients with cognitive or significant neurological deficits, therefore future developments would include a compact imaging system.

Conclusion

The venous XGB model showed higher predictive accuracy compared to the arterial. Among the nine evaluated features IOPi was the most important model feature in both vascular systems. Although the venous median predicted ICP showed the highest agreement with measured ICP using lumbar puncture, an independent prediction derived from the retinal arteries can provide a system of internal validation for the result. Previous studies have not considered the retinal arterial system for this purpose. The prediction model can potentially be integrated into a neurological clinical decision algorithm to evaluate the indication for lumbar puncture.

References

  1. 1. Sonig A, Jumah F, Raju B, Patel NV, Gupta G, Nanda A. The Historical Evolution of Intracranial Pressure Monitoring. World Neurosurgery. 2020;. pmid:32179192
  2. 2. Duits FH, Martinez-Lage P, Paquet C, Engelborghs S, Lleo A, Hausner L, et al. Performance and complications of lumbar puncture in memory clinics: results of the multicenter lumbar puncture feasibility study. Alzheimer’s & Dementia. 2016;12(2):154–163. pmid:26368321
  3. 3. Engelborghs S, Niemantsverdriet E, Struyfs H, Blennow K, Brouns R, Comabella M, et al. Consensus guidelines for lumbar puncture in patients with neurological diseases. Alzheimer’s & Dementia: Diagnosis, Assessment & Disease Monitoring. 2017;8:111–126. pmid:28603768
  4. 4. Randolph W, Carmel A, Frohman D, et al. Prevention of post–lumbar puncture headaches Report of the Therapeutics and Technology Assessment Subcommittee of the American Academy of Neurology. Neurology. 2000;55:909–914.
  5. 5. Pandian J, Sarada C, Radhakrishnan V, Kishore A. Iatrogenic meningitis after lumbar puncture a preventable health hazard. Journal of Hospital Infection. 2004;56(2):119–124. pmid:15019223
  6. 6. Doherty CM, Forbes RB. Diagnostic lumbar puncture. The Ulster medical journal. 2014;83(2):93. pmid:25075138
  7. 7. Robba C, Cardim D, Tajsic T, Pietersen J, Bulman M, Rasulo F, et al. Non-invasive intracranial pressure assessment in brain injured patients using ultrasound-based methods. In: Intracranial Pressure & Neuromonitoring XVI. Springer; 2018. p. 69–73.
  8. 8. Holloway KL, Barnes T, Choi S, Bullock R, Marshall LF, Eisenberg HM, et al. Ventriculostomy infections: the effect of monitoring duration and catheter exchange in 584 patients. Journal of neurosurgery. 1996;85(3):419–424. pmid:8751626
  9. 9. Hoefnagel D, Dammers R, Ter Laak-Poort M, Avezaat C. Risk factors for infections related to external ventricular drainage. Acta neurochirurgica. 2008;150(3):209–214. pmid:18278575
  10. 10. Dakka Y, Warra N, Albadareen R, Jankowski M, Silver B. Headache rate and cost of care following lumbar puncture at a single tertiary care hospital. Neurology. 2011;77(1):71–74. pmid:21593442
  11. 11. Hayreh SS. Pathogenesis of optic disc edema in raised intracranial pressure. Progress in retinal and eye research. 2016;50:108–144. pmid:26453995
  12. 12. Cushing H, Bordley J. Observations on experimentally induced choked disc. Johns Hopkins Hospital; 1909.
  13. 13. Kornder LH. Hydrocephalus and choked disk in dogs. Archives of Internal Medicine. 1919;23(2):197–219.
  14. 14. Baurmann M. Weitere Ergebnisse der intrakraniellen Druckmessung mit Hilfe der Netzhautvenenpuls-Beobachtung. Ber Dtsch Ophthalmol Ges. 1927;46:96–117.
  15. 15. Berens C, Smith HT, Cornwall LH. Changes in the fundus and in the blood pressure in the retinal arteries in increased intracranial pressure: papilledema and optic atrophy. Archives of Neurology & Psychiatry. 1928;20(6):1151–1171.
  16. 16. Gibbs FA. Relationship between the pressure in the veins on the nerve head and the cerebrospinal fluid pressure. Archives of Neurology & Psychiatry. 1936;35(2):292–295.
  17. 17. Hedges TR Jr. A correlative study of orbital vascular and intracranial pressure in the rhesus monkey. Transactions of the American Ophthalmological Society. 1963;61:589.
  18. 18. Rios-Montenegro EN, Anderson DR, David NJ. Intracranial pressure and ocular hemodynamics. Arch Ophthalmol. 1973;89(1):52–58. pmid:4630889
  19. 19. Hayreh SS, Edwards J. Ophthalmic arterial and venous pressures. Effects of acute intracranial hypertension. The British journal of ophthalmology. 1971;55(10):649. pmid:5124843
  20. 20. Firsching R, Schütze M, Motschmann M, Behrens-Baumann W. Venous ophthalmodynamometry: a noninvasive method for assessment of intracranial pressure. J Neurosurg. 2000;93(1):33–36. pmid:10883902
  21. 21. Firsching R, Müller C, Pauli SU, Voellger B, Röhl FW, Behrens-Baumann W. Noninvasive assessment of intracranial pressure with venous ophthalmodynamometry. Journal of neurosurgery. 2011;115(2):371–374. pmid:21529131
  22. 22. Motschmann M, Müller C, Kuchenbecker J, Walter S, Schmitz K, Schütze M, et al. Ophthalmodynamometry: a reliable method for measuring intracranial pressure. Strabismus. 2001;9(1):13–16. pmid:11262696
  23. 23. Querfurth HW, Arms SW, Lichy CM, Irwin WT, Steiner T. Prediction of intracranial pressure from noninvasive transocular venous and arterial hemodynamic measurements. Neurocrit Care. 2004;1(2):183–194. pmid:16174913
  24. 24. Canac N, Jalaleddini K, Thorpe SG, Thibeault CM, Hamilton RB. pathophysiology of intracranial hypertension and noninvasive intracranial pressure monitoring. Fluids and Barriers of the CNS. 2020;17(1):1–21. pmid:32576216
  25. 25. Khan MN, Shallwani H, Khan MU, Shamim MS. Noninvasive monitoring intracranial pressure–A review of available modalities. Surg Neurol Int. 2017;8. pmid:28480113
  26. 26. Czosnyka M, Pickard JD. Monitoring and interpretation of intracranial pressure. Journal of Neurology, Neurosurgery & Psychiatry. 2004;75(6):813–821. pmid:15145991
  27. 27. Sanborn GE, Miller NR, Maguire M, Kumar AJ. Clinical-Angiographic Correlation of Ophthalmodynamometry in Suspected Carotid Artery Disease: Prospective Study. Archives of Ophthalmology. 1981;99(10):1811–1813. pmid:7295132
  28. 28. Abdul-Rahman A, Morgan W, Jo Khoo Y, Lind C, Kermode A, Carroll W, et al. Linear interactions between intraocular, intracranial pressure, and retinal vascular pulse amplitude in the fourier domain. Plos one. 2022;17(6):e0270557. pmid:35763528
  29. 29. Morgan W, Vukmirovic A, Abdul-Rahman A, Khoo Y, Kermode A, Lind C, et al. Zero retinal vein pulsation amplitude extrapolated model in non-invasive intracranial pressure estimation. Scientific Reports. 2022;12(1):1–9. pmid:35338201
  30. 30. Abdul-Rahman A, Morgan W, Yu DY. Measurement of normal retinal vascular pulse wave attenuation using modified photoplethysmography. PloS one. 2020;15(5):e0232523. pmid:32379837
  31. 31. Morgan WH, Hazelton ML, Yu DY. Retinal venous pulsation: Expanding our understanding and use of this enigmatic phenomenon. Prog Retin Eye Res. 2016;55:82–107. pmid:27417037
  32. 32. Morgan WH, Khoo YJ, Kermode AG, Lind CR, Hazelton ML, Parsons KE, et al. Utilisation of retinal vein photoplethysmography to measure intracranial pressure. Journal of Neurology, Neurosurgery & Psychiatry. 2020;. pmid:32732385
  33. 33. Morgan WH, Abdul-Rahman A, Yu DY, Hazelton ML, Betz-Stablein B, Lind CR. Objective detection of retinal vessel pulsation. PLoS One. 2015;10(2):e0116475. pmid:25643350
  34. 34. Morgan WH, Hazelton ML, Betz-Stablein BD, Yu DY, Lind CR, Ravichandran V, et al. Photoplethysmographic measurement of various retinal vascular pulsation parameters and measurement of the venous phase delay. Investig Ophthalmol Vis Sci. 2014;55(9):5998–6006. pmid:25183767
  35. 35. Allen J. Photoplethysmography and its application in clinical physiological measurement. Physiol Meas. 2007;28(3):1–59. pmid:17322588
  36. 36. Betz-Stablein B, Hazelton ML, Morgan WH. Modelling retinal pulsatile blood flow from video data. Stat Methods Med Res. 2016; p. 1575–1584. pmid:27587593
  37. 37. Folland GB, Sitaram A. The uncertainty principle: a mathematical survey. Journal of Fourier analysis and applications. 1997;3(3):207–238.
  38. 38. Effah-Poku S, Obeng-Denteh W, Dontwi I. A study of chaos in dynamical systems. Journal of Mathematics. 2018;2018.
  39. 39. Fleischman D, Berdahl JP, Zaydlarova J, Stinnett S, Fautsch MP, Allingham RR. Cerebrospinal fluid pressure decreases with older age. PLoS One. 2012;7(12). pmid:23300737
  40. 40. Morgan WH, Cringle SJ, Kang MH, Pandav S, Balaratnasingam C, Ezekial D, et al. Optimizing the calibration and interpretation of dynamic ocular force measurements. Graefes Arch Clin Exp Ophthalmol. 2010;248(3):401–407. pmid:20107829
  41. 41. R Core Team. R: A Language and Environment for Statistical Computing; 2020. Available from: https://www.R-project.org/.
  42. 42. Budholiya K, Shrivastava SK, Sharma V. An optimized XGBoost based diagnostic system for effective prediction of heart disease. Journal of King Saud University-Computer and Information Sciences. 2020;.
  43. 43. Chen T, He T, Benesty M, Khotilovich V, Tang Y, Cho H, et al. xgboost: Extreme Gradient Boosting; 2021. Available from: https://CRAN.R-project.org/package=xgboost.
  44. 44. Zoabi Y, Deri-Rozov S, Shomron N. Machine learning-based prediction of COVID-19 diagnosis based on symptoms. npj Digital Medicine. 2021;4(1):1–5. pmid:33398013
  45. 45. Lundberg S, Lee SI. A unified approach to interpreting model predictions. arXiv preprint arXiv:170507874. 2017;.
  46. 46. Lundberg SM, Erion GG, Lee SI. Consistent individualized feature attribution for tree ensembles. arXiv preprint arXiv:180203888. 2018;.
  47. 47. Chen JW, Gombart ZJ, Rogers S, Gardiner SK, Cecil S, Bullock RM. Pupillary reactivity as an early indicator of increased intracranial pressure: The introduction of the Neurological Pupil index. Surgical neurology international. 2011;2. pmid:21748035
  48. 48. Stevens A, Su Z, Toman E, Belli A, Davies D. Optical pupillometry in traumatic brain injury: neurological pupil index and its relationship with intracranial pressure through significant event analysis. Brain Injury. 2019;33(8):1032–1038. pmid:31021683
  49. 49. Sajjadi SA, Harirchian MH, Sheikhbahaei N, Mohebbi MR, Malekmadani MH, Saberi H. The relation between intracranial and intraocular pressures: study of 50 patients. Ann Neurol. 2006;59(5):867–870. pmid:16634008
  50. 50. Czarnik T, Gawda R, Latka D, Kolodziej W, Sznajd-Weron K, Weron R. Noninvasive measurement of intracranial pressure: is it possible? Journal of Trauma and Acute Care Surgery. 2007;62(1):207–211.
  51. 51. Golan S, Kurtz S, Mezad-Koursh D, Waisbourd M, Kesler A, Halpern P. Poor correlation between intracranial pressure and intraocular pressure by hand-held tonometry. Clinical Ophthalmology (Auckland, NZ). 2013;7:1083. pmid:23785230
  52. 52. Sheeran P, Bland J, Hall G. Intraocular pressure changes and alterations in intracranial pressure. The Lancet. 2000;355(9207):899. pmid:10752710
  53. 53. Li Z, Yang Y, Lu Y, Liu D, Xu E, Jia J, et al. Intraocular pressure vs intracranial pressure in disease conditions: a prospective cohort study (Beijing iCOP study). BMC Neurol. 2012;12(1):66. pmid:22862817
  54. 54. Chunyu T, Xiujun P, Li N, Qin L, Tian Z. The correlation between intracranial pressure and intraocular pressure after brain surgery. Int J Ophthalmol Eye Res. 2014;2(5):54–58.
  55. 55. Non-Invasive Method Of Measuring Cerebral Spinal Fluid Pressure. Patent Number: 6,129,682; 2000. Available from: https://ntrs.nasa.gov/api/citations/20080004031/downloads/20080004031.pdf?attachment=true.
  56. 56. Patel MD, Malhotra K, Shirazi Z, Moss HE. Methods for quantifying optic disc volume and peripapillary deflection volume using radial optical coherence tomography scans and association with intracranial pressure. Frontiers in neurology. 2019;10:798. pmid:31396151
  57. 57. Auinger P, Durbin M, Feldon S, Garvin M, Kardon R, Keltner J, et al. Baseline OCT measurements in the idiopathic intracranial hypertension treatment trial, part II: correlations and relationship to clinical features. Investigative ophthalmology & visual science. 2014;55(12):8173–8179. pmid:25370513
  58. 58. Albrecht P, Blasberg C, Ringelstein M, Müller AK, Finis D, Guthoff R, et al. Optical coherence tomography for the diagnosis and monitoring of idiopathic intracranial hypertension. Journal of neurology. 2017;264(7):1370–1380. pmid:28584914
  59. 59. Kaufhold F, Kadas EM, Schmidt C, Kunte H, Hoffmann J, Zimmermann H, et al. Optic nerve head quantification in idiopathic intracranial hypertension by spectral domain OCT. PLoS One. 2012;7(5):e36965. pmid:22615858
  60. 60. Swanson JW, Aleman TS, Xu W, Ying GS, Pan W, Liu GT, et al. Evaluation of optical coherence tomography to detect elevated intracranial pressure in children. JAMA ophthalmology. 2017;135(4):320–328. pmid:28241164
  61. 61. Vijay V, Mollan SP, Mitchell JL, Bilton E, Alimajstorovic Z, Markey KA, et al. Using optical coherence tomography as a surrogate of measurements of intracranial pressure in idiopathic intracranial hypertension. JAMA ophthalmology. 2020;138(12):1264–1271. pmid:33090189
  62. 62. Andersen MS, Pedersen CB, Poulsen FR. A new novel method for assessing intracranial pressure using non-invasive fundus images: a pilot study. Scientific RepoRtS. 2020;10(1):1–7. pmid:32747697
  63. 63. Jeub M, Schlapakow E, Ratz M, Kindler C, Schievelkamp AH, Wabbels B, et al. Sonographic assessment of the optic nerve and the central retinal artery in idiopathic intracranial hypertension. Journal of Clinical Neuroscience. 2020;72:292–297. pmid:31540860
  64. 64. Miller MM, Chang T, Keating R, Crouch E, Sable C. Blood flow velocities are reduced in the optic nerve of children with elevated intracranial pressure. Journal of child neurology. 2009;24(1):30–35. pmid:19168816
  65. 65. Ragauskas A, Daubaris G, Dziugys A, Azelis V, Gedrimas V. Innovative non-invasive method for absolute intracranial pressure measurement without calibration. In: Intracranial Pressure and Brain Monitoring XII. Springer; 2005. p. 357–361.
  66. 66. Ragauskas A, Matijosaitis V, Zakelis R, Petrikonis K, Rastenyte D, Piper I, et al. Clinical assessment of noninvasive intracranial pressure absolute value measurement method. Neurology. 2012;78(21):1684–1691. pmid:22573638
  67. 67. Koskinen LOD, Malm J, Zakelis R, Bartusis L, Ragauskas A, Eklund A. Can intracranial pressure be measured non-invasively bedside using a two-depth Doppler-technique? J Clin Monit Comput. 2017;31(2):459–467. pmid:26971794
  68. 68. Krakauskaite S, Petkus V, Bartusis L, Zakelis R, Chomskis R, Preiksaitis A, et al. Accuracy, precision, sensitivity, and specificity of noninvasive ICP absolute value measurements. In: Intracranial Pressure and Brain Monitoring XV. Springer; 2016. p. 317–321.
  69. 69. Ragauskas A, Bartusis L, Piper I, Zakelis R, Matijosaitis V, Petrikonis K, et al. Improved diagnostic value of a TCD-based non-invasive ICP measurement method compared with the sonographic ONSD method for detecting elevated intracranial pressure. Neurological research. 2014;36(7):607–614. pmid:24620972
  70. 70. Bershad EM, Anand A, DeSantis SM, Yang M, Tang RA, Calvillo E, et al. Clinical validation of a transcranial Doppler-based noninvasive intracranial pressure meter: a prospective cross-sectional study. World neurosurgery. 2016;89:647–653. pmid:26724629
  71. 71. Hayreh SS. Pathogenesis of oedema of the optic disc. Documenta Ophthalmologica. 1968;24(2):289–411. pmid:4972870
  72. 72. Rajajee V, Vanaman M, Fletcher JJ, Jacobs TL. Optic nerve ultrasound for the detection of raised intracranial pressure. Neurocritical care. 2011;15(3):506–515. pmid:21769456
  73. 73. Kim SE, Hong EP, Kim HC, Lee SU, Jeon JP. Ultrasonographic optic nerve sheath diameter to detect increased intracranial pressure in adults: a meta-analysis. Acta Radiologica. 2019;60(2):221–229. pmid:29768927
  74. 74. Koziarz A, Sne N, Kegel F, Nath S, Badhiwala JH, Nassiri F, et al. Bedside optic nerve ultrasonography for diagnosing increased intracranial pressure: a systematic review and meta-analysis. Annals of internal medicine. 2019;171(12):896–905. pmid:31739316
  75. 75. Kishk NA, Ebraheim AM, Ashour AS, Badr NM, Eshra MA. Optic nerve sonographic examination to predict raised intracranial pressure in idiopathic intracranial hypertension: the cut-off points. The neuroradiology journal. 2018;31(5):490–495. pmid:30024291
  76. 76. Kim DY, Kim SY, Hong DY, Sung BY, Lee S, Paik JH, et al. Comparison of ultrasonography and computed tomography for measuring optic nerve sheath diameter for the detection of elevated intracranial pressure. Clinical Neurology and Neurosurgery. 2021;204:106609. pmid:33813371
  77. 77. Patterson DF, Ho ML, Leavitt JA, Smischney NJ, Hocker SE, Wijdicks EF, et al. Comparison of ocular ultrasonography and magnetic resonance imaging for detection of increased intracranial pressure. Frontiers in neurology. 2018;9:278. pmid:29740393
  78. 78. Sekhon MS, Griesdale DE, Robba C, McGlashan N, Needham E, Walland K, et al. Optic nerve sheath diameter on computed tomography is correlated with simultaneously measured intracranial pressure in patients with severe traumatic brain injury. Intensive care medicine. 2014;40(9):1267–1274. pmid:25034476
  79. 79. Zoerle T, Caccioppola A, D’Angelo E, Carbonara M, Conte G, Avignone S, et al. Optic nerve sheath diameter is not related to intracranial pressure in subarachnoid hemorrhage patients. Neurocritical care. 2020;33(2):491–498. pmid:32314244
  80. 80. Xie X, Zhang X, Fu J, Wang H, Jonas JB, Peng X, et al. Noninvasive intracranial pressure estimation by orbital subarachnoid space measurement: the Beijing Intracranial and Intraocular Pressure (iCOP) study. Crit Care. 2013;17(4):R162. pmid:23883736
  81. 81. Kimberly HH, Noble VE. Using MRI of the optic nerve sheath to detect elevated intracranial pressure. Critical care. 2008;12(5):1–2. pmid:18831721
  82. 82. York DH, Pulliam MW, Rosenfeld JG, Watts C. Relationship between visual evoked potentials and intracranial pressure. Journal of neurosurgery. 1981;55(6):909–916. pmid:7299465
  83. 83. Donald Y, Mark L, Steve B, Clark W. Further studies with a noninvasive method of intracranial pressure estimation. Neurosurgery. 1984;14(4):456–461.
  84. 84. Zhao Y, Zhou J, Zhu G. Clinical experience with the noninvasive ICP monitoring system. In: Intracranial Pressure and Brain Monitoring XII. Springer; 2005. p. 351–355.
  85. 85. Zhong J, Li Y, Minhui X, Yihua Z. Realization of a comprehensive non-invasive detection of intracranial pressure analyzer based upon FVEP and TCD. In: Intracranial Pressure and Brain Monitoring XIV. Springer; 2012. p. 127–129.
  86. 86. Andersson L, Sjölund J, Nilsson J. Flash visual evoked potentials are unreliable as markers of ICP due to high variability in normal subjects; 2012.
  87. 87. Dong J, Li Q, Wang X, Fan Y. A Review of the Methods of Non-Invasive Assessment of Intracranial Pressure through Ocular Measurement. Bioengineering. 2022;9(7):304. pmid:35877355
  88. 88. Quachtran B, Hamilton R, Scalzo F. Detection of intracranial hypertension using deep learning. In: 2016 23rd International Conference on Pattern Recognition (ICPR). IEEE; 2016. p. 2491–2496.
  89. 89. Chen W, Cockrell CH, Ward K, Najarian K. Predictability of intracranial pressure level in traumatic brain injury: features extraction, statistical analysis and machine learning-based evaluation. International journal of data mining and bioinformatics. 2013;8(4):480–494. pmid:24400523
  90. 90. Miyagawa T, Sasaki M, Yamaura A. Intracranial pressure based decision making: Prediction of suspected increased intracranial pressure with machine learning. Plos one. 2020;15(10):e0240845. pmid:33085690
  91. 91. Golzan SM, Avolio A, Graham SL. Non-invasive cerebrospinal fluid pressure estimation using multi-layer perceptron neural networks. In: 2012 Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE; 2012. p. 5278–5281.
  92. 92. Lee SC, Lueck CJ. Cerebrospinal fluid pressure in adults. Journal of Neuro-ophthalmology. 2014;34(3):278–283. pmid:25133881
  93. 93. Ruth W, Loughin T. The effect of heteroscedasticity on regression trees. arXiv preprint arXiv:160605273. 2016;.
  94. 94. Chen T, Guestrin C. Xgboost: A scalable tree boosting system. In: Proceedings of the 22nd acm sigkdd international conference on knowledge discovery and data mining; 2016. p. 785–794.
  95. 95. Wang F, Ross CL. Machine learning travel mode choices: Comparing the performance of an extreme gradient boosting model with a multinomial logit model. Transportation Research Record. 2018;2672(47):35–45.
  96. 96. Caro C, Pedley T, Schroter R, Seed W, KH P. Solid mechanics and the properties of blood vessel walls. In: The mechanics of the circulation. Cambridge:, United States: Cambridge University Press; 2012. p. 86–104.
  97. 97. Camasão DB, Mantovani D. The mechanical characterization of blood vessels and their substitutes in the continuous quest for physiological-relevant performances. A critical review. Materials Today Bio. 2021; p. 100106. pmid:33889837
  98. 98. Boron WF, Boulpaep EL. The Cardiovascular System. In: Medical physiology E-book. Elsevier Health Sciences; 2016. p. 447–871.
  99. 99. Winn HR, et al. Physiology of the Cerebrospinal Fluid and Intracranial Pressure. In: Youmans neurological surgery. vol. 1. Elsevier/Saunders; 2011. p. e407–423.
  100. 100. Gücer G, Viernstein L. Long-term intracranial pressure recording in the management of pseudotumor cerebri. Journal of neurosurgery. 1978;49(2):256–263. pmid:671078
  101. 101. Hodgson JS. Combined ventricular and lumbar puncture in the diagnosis of brain tumor: Further studies. Journal of the American Medical Association. 1928;90(19):1524–1526.
  102. 102. Smyth G, Henderson W. Observations on the cerebrospinal fluid pressure on simultaneous ventricular and lumbar punctures. Journal of Neurology and Psychiatry. 1938;1(3):226. pmid:21610930
  103. 103. Speck V, Staykov D, Huttner HB, Sauer R, Schwab S, Bardutzky J. Lumbar catheter for monitoring of intracranial pressure in patients with post-hemorrhagic communicating hydrocephalus. Neurocritical care. 2011;14(2):208–215. pmid:20953846
  104. 104. Eide P, Brean A. Lumbar cerebrospinal fluid pressure waves versus intracranial pressure waves in idiopathic normal pressure hydrocephalus. British journal of neurosurgery. 2006;20(6):407–414. pmid:17439094
  105. 105. Lenfeldt N, Koskinen LO, Bergenheim AT, Malm J, Eklund A. CSF pressure assessed by lumbar puncture agrees with intracranial pressure. Neurology. 2007;68(2):155–158. pmid:17210899
  106. 106. Warden KF, Alizai AM, Trobe JD, Hoff JT. Short-term continuous intraparenchymal intracranial pressure monitoring in presumed idiopathic intracranial hypertension. Journal of Neuro-Ophthalmology. 2011;31(3):202–205. pmid:21483268
  107. 107. Eide P, Due-Tønnessen B, Helseth E, Lundar T. Assessment of intracranial pressure volume relationships in childhood: the lumbar infusion test versus intracranial pressure monitoring. Child’s Nervous System. 2001;17(7):382–390. pmid:11465790
  108. 108. Morgan WH, Yu DY, Cooper RL, Alder VA, Cringle SJ, Constable IJ. The influence of cerebrospinal fluid pressure on the lamina cribrosa tissue pressure gradient. Investigative ophthalmology & visual science. 1995;36(6):1163–1172. pmid:7730025
  109. 109. Morgan WH, Yu DY, Cooper RL, Alder VA, Cringle SJ, Constable IJ. Retinal artery and vein pressures in the dog and their relationship to aortic, intraocular, and cerebrospinal fluid pressures. Microvascular research. 1997;53(3):211–221. pmid:9211399
  110. 110. Morgan WH, Yu DY, Alder VA, Cringle SJ, Cooper RL, House PH, et al. The correlation between cerebrospinal fluid pressure and retrolaminar tissue pressure. Investigative ophthalmology & visual science. 1998;39(8):1419–1428. pmid:9660490
  111. 111. Moss HE. Retinal Vein Changes as a Biomarker to Guide Diagnosis and Management of Elevated Intracranial Pressure. Frontiers in neurology. 2021; p. 1841. pmid:34733231
  112. 112. Armstead WM. Cerebral blood flow autoregulation and dysautoregulation. Anesthesiology clinics. 2016;34(3):465–477. pmid:27521192
  113. 113. Ursino M, Lodi CA. A simple mathematical model of the interaction between intracranial pressure and cerebral hemodynamics. Journal of Applied Physiology. 1997;82(4):1256–1269. pmid:9104864
  114. 114. Tachibana H, Gotoh F, Ishikawa Y. Retinal vascular autoregulation in normal subjects. Stroke. 1982;13(2):149–155. pmid:7064186
  115. 115. Querfurth H, Lagrèze WD, Hedges T, Heggerick P. Flow velocity and pulsatility of the ocular circulation in chronic intracranial hypertension. Acta neurologica scandinavica. 2002;105(6):431–440. pmid:12027831
  116. 116. Gabrys E, Rybaczuk M, Kedzia A. Fractal models of circulatory system. Symmetrical and asymmetrical approach comparison. Chaos, Solitons & Fractals. 2005;24(3):707–715.
  117. 117. Kenner T. Biological asymmetry and cardiovascular blood transport. Cardiovascular Engineering: An International Journal. 2004;4(2):209–218.
  118. 118. Li Y, Hall NE, Pershing S, Hyman L, Haller JA, Lee AY, et al. Age, gender, and laterality of retinal vascular occlusion: a retrospective study from the IRIS® Registry. Ophthalmology Retina. 2022;6(2):161–171. pmid:33991710
  119. 119. Yangjiani L, Wang M, Wong RC, Li D, Wang H, Jin Q, et al. The laterality difference of retinal vascular occlusion. Investigative Ophthalmology & Visual Science. 2020;61(7):1332–1332.
  120. 120. Hu CA, Chen CM, Fang YC, Liang SJ, Wang HC, Fang WF, et al. Using a machine learning approach to predict mortality in critically ill influenza patients: a cross-sectional retrospective multicentre study in Taiwan. BMJ open. 2020;10(2):e033898. pmid:32102816
  121. 121. Cabitza F, Rasoini R, Gensini GF. Unintended consequences of machine learning in medicine. Jama. 2017;318(6):517–518. pmid:28727867
  122. 122. Petkovic D, Kobzik L, Re C. Machine learning and deep analytics for biocomputing: call for better explainability. In: PSB. World Scientific; 2018. p. 623–627.
  123. 123. Zhang Z, Beck MW, Winkler DA, Huang B, Sibanda W, Goyal H, et al. Opening the black box of neural networks: methods for interpreting neural network models in clinical applications. Annals of translational medicine. 2018;6(11). pmid:30023379
  124. 124. Biecek P. DALEX: Explainers for Complex Predictive Models in R. Journal of Machine Learning Research. 2018;19(84):1–5.
  125. 125. Staniak M, Biecek P. Explanations of Model Predictions with live and breakDown Packages. The R Journal. 2018.
  126. 126. Foster D. xgboostExplainer: XGBoost Model Explainer; 2021.
  127. 127. Hooker G. Diagnostics and extrapolation in machine learning. stanford university; 2004.