Decontamination of Geobacillus Stearothermophilus using the Arca Aerosolized Hydrogen Peroxide decontamination system

  • Loren Benjamin Mead ,

    Roles Conceptualization, Investigation, Methodology, Project administration, Supervision, Validation, Writing – original draft

    lmead@abaton.care

    Affiliations Abaton, Board of Directors, Washington, D.C, United States of America, Department of Emergency Medicine, University of Texas Health Science Center at Houston, Houston, TX, United States of America

  • Tanner Mathison,

    Roles Conceptualization, Data curation, Funding acquisition, Investigation, Methodology, Project administration, Supervision, Writing – review & editing

    Affiliation Abaton, Board of Directors, Washington, D.C, United States of America

  • Garry Osborne,

    Roles Data curation, Methodology

    Affiliation Abaton, Board of Directors, Washington, D.C, United States of America

  • Anne Marie Richards

    Roles Data curation, Methodology, Writing – review & editing

    Affiliation Abaton, Board of Directors, Washington, D.C, United States of America

Abstract

Introduction

In response to the limited supply of personal protective equipment during the pandemic caused by SARS-CoV-2, recent studies demonstrate that gaseous H2O2 is an effective decontaminant of N95 filtering facepiece respirators to enable reuse of these items in a clinical setting. This paper evaluates the efficacy of the Arca Aerosolized Hydrogen Peroxide Decontamination System (Arca), a novel aerosolized H2O2 decontamination system, using biologic indicator testing.

Materials and methods

The Arca produces and circulates H2O2 aerosol inside of a sealed stainless steel chamber. The Arca’s decontamination efficacy was evaluated in 8 decontamination trials with 2 H2O2 concentrations (3% and 12%) and 4 decontamination cycle durations (45, 60, 90, and 120 minutes). Efficacy was evaluated by testing: 1) the concentration in parts per million (ppm) of H2O2 produced inside the chamber and the concentration in ppm of H2O2 vented from the chamber, and 2) the decontamination of Mesa Biologic Indicator filter strips (BI) inoculated with Geobacillus Stearothermophilus. Control tests were conducted by submerging BI strips in 3mL of 3% and 12% H2O2 for 120 minutes (negative controls) and by not exposing one BI strip to H2O2 (positive control).

Results

Greater than 5000 ppm of H2O2 was detected on the concentration strips inside the chamber for each of the eight decontamination trials. No vented H2O2 was detected on the external concentration strips after any decontamination trial. No growth was observed for any of the negative controls after seven days. The positive control was positive for growth.

Conclusion

The Arca Aerosolized Hydrogen Peroxide Decontamination System is effective at decontaminating bacterial G. Stearothermophilus at a cycle time of 45 minutes utilizing 6mL of 3% H2O2 solution.

Introduction

The SARS-CoV-2 coronavirus (COVID-19) pandemic has caused a worldwide shortage in personal protective equipment (PPE) including N95 filtering facepiece respirator (FFR) and surgical masks. Both mask types have been shown to dramatically reduce the rate of infection by airborne viruses such as influenza, but the masks are designed for single use before disposal [1]. Due to a shortage of FFRs and the critical role FFRs play in protecting health care workers at the outset of the COVID-19 pandemic, the Center for Disease Control recommended limited reuse of decontaminated FFRs to extend supplies of PPE [1,2]. The Occupational Safety and Health Administration has similarly produced guidelines for the limited reuse of FFRs [3].

Aerosolized or vaporized hydrogen peroxide (H2O2) is an established reactant for decontaminating surfaces inoculated with resistant bacterial and viral pathogens [4]. Recent studies demonstrate that H2O2 delivered in an aerosol or vapor as low as 500 ppm has been shown to achieve 1000 times reductions in viral organism activity on inoculated FFRs while maintaining adequate mask fitment through multiple decontamination cycles [58]. Previous work has demonstrated that scalable, proof-of-concept H2O2 decontamination systems could achieve adequate minimum concentrations reported in the literature to eradicate Coronavirus and other pathogens [9,10].

To mitigate infection risk with reuse of FFRs, the Food and Drug Administration granted emergency use authorizations (EUAs) to several commercially available hydrogen peroxide FFR decontamination procedures [11]. However, the commercially available methods of FFR decontamination are costly and limited in availability [1214]. The H2O2 decontamination methods which received an EUA include those listed in Table 1. The number of maximum decontamination cycles per FFR varies by method with a range of two to 20 cycles [15]. The price per institution for these systems can exceed $50,000 (Table 1). Low-resource institutions are often unable to access these services and purchase adequate supplies of PPE [16]. Therefore, the development of a smaller-scale, less-costly decontamination devices is warranted.

thumbnail
Table 1. Value Comparison of H2O2 decontamination systems with EUAs.

https://doi.org/10.1371/journal.pone.0273937.t001

This study seeks to evaluate the efficacy of a low-cost design engineered for use in low-resource settings to expand the supply of decontaminated PPE for safe re-use by frontline workers. Our objectives were to test the device’s efficacy at different H2O2 concentrations and Arca device cycle durations by evaluating: 1) the concentration in parts per million (ppm) of H2O2 produced inside the chamber and the concentration in ppm of H2O2 vented from the chamber, and 2) the decontamination of Mesa Biologic Indicator filter strips (BI) inoculated with 106 Geobacillus Stearothermophilus.

Materials and methods

The Arca Aerosolized Hydrogen Peroxide Decontamination System (Arca) produces H2O2 inside of a sealed stainless steel chamber and circulates the aerosol with four CG IP67 personal computing fans. The Arca uses a Venturi tube design to generate an aerosol: an air compressor forces high velocity air through a tube where the cross-sectional area is reduced with the result of lowering the air pressure (Fig 1). The design lowers the pressure to below 2300 Pa, the vaporization pressure of 10% H2O2 in water solution [20]. The H2O2 is fed to the low pressure opening of the tube, where the low pressure induces partially vaporization, forming small droplets of solution (i.e. an aerosol). A catch is included at the low pressure point to trap large droplets and recirculate them back to the low-pressure area. FFRs are placed on a wire mesh rack inside the chamber and exposed to the H2O2 for decontamination (Fig 2). Circulated H2O2 aerosol is removed from the chamber interior using a Speedair 4ZL07 condenser unit to avoid operator exposure when removing decontaminated FFRs (Fig 3). The device’s manufacturing and component costs totaled less than $2,000.

thumbnail
Fig 2. Arca interior demonstrating mesh rack, circulating fans, and Venturi tube nozzle.

https://doi.org/10.1371/journal.pone.0273937.g002

thumbnail
Fig 3. Arca exterior showing nebulizer (top), condenser unit (bottom), and electronics housing (left).

https://doi.org/10.1371/journal.pone.0273937.g003

The Arca’s decontamination efficacy was evaluated by a testing protocol exposing MesaStrip BI filter strips inoculated with 106 G. Stearothermophilus [22]. We performed a total of 8 trials varying the H2O2 concentration (3% and 12%) and decontamination cycle duration (45, 60, 90, and 120 minutes). Decontamination efficacy was evaluated by H2O2 concentration and biologic testing within the device chamber and venting system.

Concentration testing was performed using Bartovartion Very High Level Peroxide Test Strips, which change color to indicate exposure to H2O2 at concentrations from 0 to 5000 ppm [23]. For each trial, one BI was secured with Scotch MagicTM Tape to an FFR located centrally inside the chamber. An additional concentration strip was placed on the exterior of the door and evaluated for vented H2O2 from the chamber after each cycle was completed and the door to the chamber opened. All cycles were performed in a well-ventilated area.

Additionally, one BI was submerged in 3 mL of 3% and one in 3 mL of 12% H2O2 for 120 minutes each to serve as negative controls. One BI was not exposed to H2O2 to serve as a positive control.

The BIs were handled using sterile technique and after testing were shipped in sealed plastic bags to STERIS Laboratories. Sterility testing was performed according to STERIS Laboratories protocol ST/01, in which the BIs were incubated in 10 mL of Tryptic Soy Broth at 55–60 degrees Celsius for seven days. The cultures were evaluated daily and interpreted as “No growth” only if the medium remained clear without color change or turbidity after seven days [24].

Results

Concentration testing

The concentration strips for all decontamination cycles showed internal concentrations of H2O2 exceeding 5000 ppm. The exterior concentration strips monitoring for vented H2O2 all showed undetectable levels of H2O2 (Tables 2 and 3).

thumbnail
Table 2. Concentration strip results for 3% H2O2 solution testing.

https://doi.org/10.1371/journal.pone.0273937.t002

thumbnail
Table 3. Concentration strip results for 12% H2O2 solution testing.

https://doi.org/10.1371/journal.pone.0273937.t003

Decontamination testing

All BIs exposed to H2O2 had no growth at seven days of incubation (Tables 4 and 5). The positive control that was not exposed to H2O2 demonstrated bacterial growth.

thumbnail
Table 4. BI growth at 7 days following exposure to 3% H2O2 solution.

https://doi.org/10.1371/journal.pone.0273937.t004

thumbnail
Table 5. BI growth at 7 days following exposure to 12% H2O2 solution.

https://doi.org/10.1371/journal.pone.0273937.t005

Discussion

The limited supply of FFRs during the SARS-Cov-2 pandemic endangers health care and other frontline workers. While the FDA and CDC have encouraged the decontamination of FFRs for reuse, existing decontamination systems are inaccessible in low-resource settings. Development of systems designed for deployment in low-resource settings will increase the supply of PPE for at-risk personnel.

This study attempted to evaluate a novel device’s ability to eradicate bacterial pathogens from the surface of an FFR without exposing users to unsafe levels of H2O2. Our testing demonstrates that the novel Arca device can decontaminate bacterial pathogens without exposing users to high concentrations of H2O2. These results are consistent with the literature on existing gaseous H2O2 technologies. The eradication of bacteria achieved at every trial suggests that the Arca could achieve adequate decontamination with a lower concentration H2O2 solution, a shorter cycle duration, or a smaller volume of H2O2. Based on previously reported H2O2 testing, the bactericidal properties of the device suggest virucidal efficacy as well [9,25].

The Arca has a throughput of 16 FFRs per cycle, with a daily output of up to 512 FFRs. The marginal cost to clean each additional FFR is significantly less than $1, as operating expenses include small volumes of widely available concentrations of H2O2 and the variable cost of electricity to power the device. The Arca’s production cost at less than $2,000 per unit is significantly lower than the price of existing commercially available solutions. At scale, this manufacturing cost should decrease. These devices are intended to be donated at no cost to recipient institutions, therefore the operating expenses are the only costs passed on to end users. Future iterations of the device could reduce costs further by using more off-the-shelf components in its manufacturing process. The decreased cost of the Arca improves its accessibility in low-resource settings compared with existing FFR decontamination systems.

Limitations

The sterility testing performed on the BI strips resulted in a binary “No growth” or “Growth” outcome, and did not demonstrate log kill rates. Therefore, it is possible that small populations of bacteria survived decontamination cycles but did not change the TSB medium’s turbidity. However, the same testing methodology is used as part of quality assurance testing to assess sterility achieved by FDA-approved devices, with an industry standard sterility assurance level (SAL) of 10−6. The FDA requires only an SAL of 10−3 for products that contact intact skin [26].

Testing was performed under optimal conditions using a single FFR. Real world use may lead to decreased efficacy due to variations in power supply to the Arca, increased pathogen burden, or greater surface area of multiple FFRs. Additionally, while bacteria are considered more resistant to H2O2 than viruses, the eradication of this study’s BIs may not be generalizable to all mutations of SARS-CoV-2.

Future directions

Experimental derivation of the absolute minimum cycle duration, volume of H2O2, and percent H2O2 solution to achieve decontamination will allow for greater throughput of the Arca device. Testing in a BSL-3 lab will evaluate the Arca’s efficacy at decontamination of SARS-CoV-2 and other pathogens directly inoculated on PPE.

Fitment testing of FFRs after decontamination will allow for experimental derivation of the maximum decontamination cycles per FFR. Additional testing of other PPE and various equipment may expand the use cases for the Arca.

Additionally, “real world” field testing will help to determine any potential causes of diminished aerosol concentration, inadequate decontamination, or non-optimal user experience.

Conclusion

The Arca Aerosolized Hydrogen Peroxide Decontamination System can safely decontaminate FFRs inoculated with bacteria using a minimum cycle time of 45 minutes with 6 mL of H2O2 solution.

Acknowledgments

We would like to acknowledge the tireless work of our systems engineers, Garry Osborne and Keith Crowder, as well as our head of business development, Michael Mazza. Without their constant support, the Arca would have been neither built nor tested. Our acknowledgements would not be complete without thanking Dr. Sarah Huepenbecker and Natalie Punchak for their sage advice and emotional support.

References

  1. 1. Apr 03 RR|, 2012. Lab study supports use of N95 respirators for flu protection [Internet]. CIDRAP. [cited 2020 Jun 14]. Available from: https://www.cidrap.umn.edu/news-perspective/2012/04/lab-study-supports-use-n95-respirators-flu-protection.
  2. 2. Recommended Guidance for Extended Use and Limited Reuse of N95 Filtering Facepiece Respirators in Healthcare Settings | NIOSH | CDC [Internet]. 2020 [cited 2020 Jun 28]. Available from: https://www.cdc.gov/niosh/topics/hcwcontrols/recommendedguidanceextuse.html.
  3. 3. Enforcement Guidance for Respiratory Protection and the N95 Shortage Due to the Coronavirus Disease 2019 (COVID-19) Pandemic | Occupational Safety and Health Administration [Internet]. [cited 2021 Jan 11]. Available from: https://www.osha.gov/memos/2020-04-03/enforcement-guidance-respiratory-protection-and-n95-shortage-due-coronavirus.
  4. 4. Chemical Disinfectants | Disinfection & Sterilization Guidelines | Guidelines Library | Infection Control | CDC [Internet]. 2019 [cited 2021 Jan 11]. Available from: https://www.cdc.gov/infectioncontrol/guidelines/disinfection/disinfection-methods/chemical.html.
  5. 5. Technical Report for Hydrogen Peroxide Methods for Decontaminating N95 Respirators [Internet]. [cited 2020 Jun 28]. Available from: https://static1.squarespace.com/static/5e8126f89327941b9453eeef/t/5ed525e43bb86f3745a1ded6/1591027172069/2020-05-29_N95DECON_HydrogenPeroxide_V2-1_TechnicalReport_FINAL.pdf.
  6. 6. Viscusi DJ, Bergman MS, Eimer BC, Shaffer RE. Evaluation of Five Decontamination Methods for Filtering Facepiece Respirators. Ann Occup Hyg. 2009 Nov;53(8):815–27. pmid:19805391
  7. 7. Polkinghorne A, Branley J. Evidence for decontamination of single-use filtering facepiece respirators. J Hosp Infect [Internet]. 2020 May 27 [cited 2020 Jun 30]; Available from: https://www.ncbi.nlm.nih.gov/pmc/articles/PMC7251398/. pmid:32473179
  8. 8. Schwartz A, Stiegel M, Greeson N, Vogel A, Thomann W, Brown M, et al. Decontamination and Reuse of N95 Respirators with Hydrogen Peroxide Vapor to Address Worldwide Personal Protective Equipment Shortages During the SARS-CoV-2 (COVID-19) Pandemic. Appl Biosaf. 2020 Jun;25(2):67–70. pmid:36035079
  9. 9. Cheng VCC, Wong SC, Kwan GSW, Hui WT, Yuen KY. Disinfection of N95 respirators by ionized hydrogen peroxide during pandemic coronavirus disease 2019 (COVID-19) due to SARS-CoV-2. J Hosp Infect. 2020 Jun 1;105(2):358–9. pmid:32277965
  10. 10. Grossman J, Pierce A, Mody J, Gagne J, Sykora C, Sayood S, et al. Institution of a Novel Process for N95 Respirator Disinfection with Vaporized Hydrogen Peroxide in the Setting of the COVID-19 Pandemic at a Large Academic Medical Center. J Am Coll Surg. 2020 Aug 1;231(2):275–80. pmid:32353399
  11. 11. Health C for D and R. Emergency Use Authorizations. FDA [Internet]. 2020 Jun 7 [cited 2020 Jun 7]; Available from: https://www.fda.gov/medical-devices/emergency-situations-medical-devices/emergency-use-authorizations.
  12. 12. Battelle CCDS for COVID19—Satellite Locations [Internet]. Battelle. [cited 2020 Jun 28]. Available from: https://www.battelle.org/inb/battelle-ccds-for-covid19-satellite-locations.
  13. 13. SteraMist [Internet]. Engineering For Change. [cited 2020 Jun 28]. Available from: https://www.engineeringforchange.org/solutions/product/steramist/.
  14. 14. DOD Contract for 60 N95 Critical Care Decontamination Units: $415M Contract, Each Unit Can [Internet]. U.S. DEPARTMENT OF DEFENSE. [cited 2020 Jun 30]. Available from: https://www.defense.gov/Newsroom/Releases/Release/Article/2148352/dod-contract-for-60-n95-critical-care-decontamination-units-415m-contract-each/.
  15. 15. Guide to decontamination methods for filtering facepiece respirators [Internet]. [cited 2021 Jan 11]. Available from: https://www.vizientinc.com/covid-19/covid-19-ffr-decontamination-methods#summary.
  16. 16. McMahon DE, Peters GA, Ivers LC, Freeman EE. Global resource shortages during COVID-19: Bad news for low-income countries. PLoS Negl Trop Dis. 2020 Jul 6;14(7):e0008412. pmid:32628664
  17. 17. Portable Ambulance Decontamination Systems Market Survey Report. 24.
  18. 18. McCreanor V, Graves N. An economic analysis of the benefits of sterilizing medical instruments in low-temperature systems instead of steam. Am J Infect Control. 2017 Jul 1;45(7):756–60. pmid:28365144
  19. 19. Battelle under fire for decontamination system performance [Internet]. MassDevice. 2020 [cited 2021 Jan 11]. Available from: https://www.massdevice.com/battelle-under-fire-for-decontamination-system-performance/.
  20. 20. Vapor Pressures | USP Technologies [Internet]. [cited 2022 Mar 26]. Available from: https://www.h2o2.com/technical-library/physical-chemical-properties/physical-properties/default.aspx?pid=25&name=Vapor-Pressures.
  21. 21. O’Callaghan C, Barry PW. The science of nebulised drug delivery. Thorax. 1997 Apr 1;52(Supplement 2):S31–44. pmid:9155849
  22. 22. Spore Strip Steam—Technical Report.pdf [Internet]. [cited 2021 Jan 11]. Available from: https://info.mesalabs.com/hubfs/1.%20Sterilization%20and%20Disinfection%20Control/Website/Technical%20Reports/Spore%20Strip%20Steam%20-%20Technical%20Report.pdf.
  23. 23. Very High Level Peroxide Test Strips, 0–5000 ppm [Vial of 50 Strips] | Bartovation Test Strips [Internet]. [cited 2020 Jun 30]. Available from: https://bartovation.com/product/disinfectants-sanitizers/peroxide/high-level-peroxide-500-5000ppm-for-concentrated-h2o2-solutions-50-strips/.
  24. 24. Form-14-Guide-to-Testing-Services.pdf [Internet]. [cited 2022 Jul 30]. Available from: https://www.steris-ast.com/wp-content/uploads/2020/02/Form-14-Guide-to-Testing-Services.pdf.
  25. 25. Tuladhar E, Terpstra P, Koopmans M, Duizer E. Virucidal efficacy of hydrogen peroxide vapour disinfection. J Hosp Infect. 2012 Feb 1;80(2):110–5. pmid:22153909
  26. 26. Sterility Review of 510(k) Submissions: Common Microbiology Issues Found in 510(k) Submissions [Internet]. [cited 2022 Jul 30]. Available from: https://www.fda.gov/media/141277/download.