Precision and accuracy of FEV1 measurements from the Vitalograph copd-6 mini-spirometer in a healthy Ugandan population

  • Wajd Abbas Hassan Hansen,

    Roles Formal analysis, Writing – original draft

    Affiliations Lægerne i Hirtshals APS, Hirtshals, Denmark, Research Unit for Environment, Work and Health, Danish Ramazzini Center, Department of Public Health, Aarhus University, Aarhus, Denmark

  • Vivi Schlünssen,

    Roles Conceptualization, Funding acquisition, Methodology, Supervision, Writing – review & editing

    Affiliations Research Unit for Environment, Work and Health, Danish Ramazzini Center, Department of Public Health, Aarhus University, Aarhus, Denmark, National Research Center for the Working Environment, Copenhagen, Denmark

  • Erik Jørs,

    Roles Conceptualization, Methodology, Supervision, Writing – review & editing

    Affiliations Department of Occupational and Environmental Medicine, Odense University Hospital, Odense, Denmark, Occupational and Environmental Medicine, Department of Clinical Research, University of Southern Denmark, Odense, Denmark

  • Daniel Sekabojja,

    Roles Conceptualization, Methodology, Resources, Writing – review & editing

    Affiliation Uganda National Association of Community and Occupational Health, Kampala, Uganda

  • John C. Ssempebwa,

    Roles Conceptualization, Methodology, Writing – review & editing

    Affiliation School of Public Health, Makerere University, Kampala, Uganda

  • Ruth Mubeezi,

    Roles Conceptualization, Methodology, Writing – review & editing

    Affiliation School of Public Health, Makerere University, Kampala, Uganda

  • Philipp Staudacher,

    Roles Conceptualization, Methodology, Writing – review & editing

    Affiliations Eawag, Swiss Federal Institute of Aquatic Science and Technology, Dübendorf, Switzerland, Institute of Biogeochemistry and Pollutant Dynamics, ETH Zurich, Zurich, Switzerland

  • Samuel Fuhrimann,

    Roles Conceptualization, Methodology, Writing – review & editing

    Affiliation Institute for Risk Assessment Sciences, Utrecht University, Utrecht, Netherlands

  • Martin Rune Hassan Hansen

    Roles Conceptualization, Data curation, Formal analysis, Funding acquisition, Investigation, Methodology, Project administration, Writing – original draft

    martinrunehassanhansen@ph.au.dk

    Affiliations Research Unit for Environment, Work and Health, Danish Ramazzini Center, Department of Public Health, Aarhus University, Aarhus, Denmark, National Research Center for the Working Environment, Copenhagen, Denmark

Abstract

Objective

Evaluate the accuracy and precision of the copd-6 mini-spirometer for FEV1 in a rural Ugandan population.

Methods

In a cross-sectional study, 171 smallholder farmers performed spirometry with copd-6, and a diagnostic-quality spirometer.

Results and discussion

The copd-6 underestimated FEV1 at low flows and overestimated FEV1 at high flows. Across all participants, the device slightly overestimated FEV1 by 0.04 [0.02; 0.06] L. Calibration data showed similar patterns.

Conclusion

The copd-6 could be considered as an affordable tool for research on lung function impairment in resource-constrained settings. However, further validation in a study population with obstructive lung disease is needed.

Introduction

Chronic obstructive pulmonary disease (COPD) causes over 3 million deaths annually; more than 80% in low and middle-income countries [1]. In Uganda, the rural prevalence of COPD among adults > 35 years is 6.1% [2]. Spirometry is necessary for diagnosis, yet often unavailable or prohibitively expensive [3]—a typical situation in many low- and middle-income countries [3].

The copd-6 mini-spirometer (Vitalograph, Ennis, Ireland) is relatively inexpensive, costing around 125 USD–compared to more advanced spirometers that often cost upward of 600 USD. The copd-6 has been shown to have reasonable precision for the diagnosis of COPD [4], and holds promise as a diagnostic tool in resource-constrained settings. But the accuracy of the device remains insufficiently described, limiting its potential for use by local researchers investigating risk factors for lung function impairment, since inaccurate measurements have the potential to bias exposure-response relationships. The purpose of this study is therefore to evaluate both the accuracy and precision of the copd-6 mini-spirometer for the measurement of FEV1 in a Ugandan population.

Methods

Study population

Data was collected as a part of the “Pesticide Exposure, Asthma and Diabetes in Uganda” (PEXADU) project, which was a study on the possible associations between pesticide exposure, pulmonary function and diabetes in a cohort of smallholder farmers in Uganda [57]. Details regarding participant recruitment and inclusion/exclusion have been presented elsewhere [57]. In brief, we recruited 364 farmers from the Wakiso District in central Uganda with the help of two local farmer’s organizations. We visited a number of smaller farmer’s groups affiliated with these two organizations and invited all members aged > 18 years to participate, except pregnant women. Participants came to the project examination center at baseline in September-October 2018, with two rounds of follow-up in November-December 2018 and January-February 2019, respectively. All data analyzed in this paper are from the baseline examination.

Spirometry testing

Out of 364 participants, 304 performed spirometry. The majority of the remaining participants were excluded because they self-reported one of the following: Myocardial infarction in the last 3 months, angina pectoris, hemoptysis, any surgery in the last 3 months, aortic aneurism, history of pulmonary embolism, active tuberculosis or other current respiratory infection. We also excluded one individual with severe hypertension (defined as blood pressure > 200 mmHg systolic or > 120 mmHg diastolic).

Participants underwent spirometry with both a diagnostic-quality spirometer (MicroDL, Micro Medical, Rochester, Kent, England) and a copd-6 mini-spirometer (lot number 0317/2018). Tests were conducted between 7 AM and 5 PM. A pseudo-random number generator randomized the order of the devices in the lung function test. Only a few minutes passed between testing with each of the two devices. For the MicroDL device, participants first blew five times into the device. If their results did not fulfill standard quality criteria [8], they got four additional attempts. To avoid fatigue, participants only blew three times with the copd-6.

Ethics approval and consent to participate

The project was conducted in accordance with the Declaration of Helsinki. All participants gave written informed consent before inclusion and were financially compensated for lost earnings on the examination day. Ethical approval was granted by the Higher Degrees Research and Ethics Committee at Makerere University School of Public Health, Kampala, Uganda (MakSPH-HDREC, registration number 577) and the Uganda National Council for Science and Technology, Kampala, Uganda (registration number HS234ES).

Calibration checks

The calibration of the copd-6 devices was checked daily using a 3-liter calibration syringe (MIR 919000, Medical International Research Inc., Rome, Italy). The syringe was emptied three times with each of three speeds: Slow (as slowly as possible, while finishing within 6 seconds), medium, and fast (as fast as possible). The calibration of the MicroDL devices was checked in a similar way, except that there was no time limit for the slow plunges of the calibration syringe. Only copd-6 calibration check data will be presented here. Calibration check data for the MicroDL devices have been reported elsewhere [6].

Data entry

Data from the MicroDL were extracted in digital format using Spida 5 PC software (MicroDL, Micro Medical, Rochester, Kent, England). Data from the copd-6 were entered directly in a structured database at the time of testing or calibration, using the ODK Collect app [9].

Quality control

The copd-6 device gives a warning in case the subject coughs during a blow, or in case of slow starts, and such blows were excluded. We excluded all copd-6 results from participants with less than two accepted blows. Furthermore, we excluded non-repeatable copd-6 results, defined as a difference in the best and second-best FEV1 and FEV6 > 0.25 liters.

A medical doctor assessed the quality of MicroDL spirometry according to modified ATS criteria [8], as previously reported elsewhere [6, 7]. All MicroDL results were excluded if the participant had performed less than two acceptable blows, or if the results were non-repeatable, defined as a difference in the best and second-best FEV1 or FVC > 0.25 liters.

Statistical analyses

Participant FEV1 was analyzed in Bland-Altman plots. Trends in d = (FEV1,copd6FEV1,MicroDL) as a function of m = (FEV1,copd6+FEV1,MicroDL)/2 was analyzed in a mixed effect model with a fixed effect for m and a random effect for participant family. To account for non-linearity, m was modelled using restricted cubic splines with four knots. Since the MicroDL reported FVC and not FEV6, analyses of FEV6 and FEV1/FEV6 were limited to 21 participants whose MicroDL spirograms showed that their FEV6 and FVC were equal. For copd-6 calibration check data, reported FEV6 vs. speed of pushing the piston was analyzed using Spearman’s rank correlation.

Data management and analyses were performed in Stata 15 (StataCorp, College Station, Texas, United States). The statistical analyses were specified a priori and the analysis protocol published in an online repository before analysis [10]. All deviations between the protocol and the final analyses are listed in S1 File.

Results

Out of 364 participants in the PEXADU study, 304 (84%) performed spirometry, and 171 (47%) fulfilled quality criteria for both the MicroDL and copd-6 devices and were included in the analyses. Demographic information on the study population is shown in Table 1. An overview of participant inclusion is provided in Fig 1.

The mean FEV1 from the MicroDL was 2.48 L (Table 2). Overall, the copd-6 slightly overestimated FEV1, with a mean difference (defined as FEV1,copd6 –FEV1,MicroDL) of 0.04 [0.02; 0.06] L. Most participants had an absolute difference in FEV1 < 0.25 L between the two devices, although there were some outliers (Fig 2). Across participants, the average of participants’ FEV1 from the copd-6 and MicroDL devices (m) ranged from 0.98 L to 4.68 L. Based on the trend in the mixed effect model (i.e., not individual participants), the difference between the two devices was -0.06 [-0.17; 0.05] L at the minimum m (0.98 L), and 0.21 [0.08; 0.33] L at the maximum m (4.68 L). We found similar results in a number of sensitivity analyses that were stratified by the specific devices used for testing, stratified by the order of testing (copd-6 before MicroDL or vice versa), only included the first three blows with the MicroDL, or used stricter repeatability criteria (S2 File).

thumbnail
Fig 2. Bland-Altman plot of copd-6 and MicroDL participant results for FEV1.

Each dot represents one participant. Solid gray line = 0.00 L. Dashed gray lines = ± 0.25 L. Black lines = trend with 95% confidence interval.

https://doi.org/10.1371/journal.pone.0253319.g002

thumbnail
Table 2. Summary metrics for results from MicroDL and copd-6 spirometers.

https://doi.org/10.1371/journal.pone.0253319.t002

For the 21 subjects where FEV6 could be directly compared between the copd-6 and MicroDL, the overall difference in mean FEV1/FEV6 was -0.02 [-0.03; 0.00]. The difference seemed to depend on the value of FEV1/FEV6; the copd-6 overestimated ratios close to 1 and underestimated lower ratios (S2 File).

Calibration data showed the same pattern: Underestimation of volumes at low flow and overestimation at high flow (Fig 3), but most results for low and medium flow were within 3.00 liters ± 3% (the accuracy required by ATS guidelines) [8].

thumbnail
Fig 3. Scatterplots of calibration check data for copd-6, stratified by the speed of pushing the piston of the calibration syringe.

Spearman’s ϱ = 0.82, p < 0.001. Each dot represents one plunge of the calibration syringe. Arrows pointing downward are plunges with reported FEV6 < 2.8 liters. Gray lines located at 3.00 liters ± 3%.

https://doi.org/10.1371/journal.pone.0253319.g003

Discussion

Overall, the copd-6 measures FEV1 reasonably well in our relatively healthy study population. However, the device slightly underestimates volume at low flows and overestimates volume at high flows. In addition, it may underestimate FEV1/FEV6, but as FEV1/FEV6 could be analyzed for only few subjects, we have limited confidence in this latter finding.

A recent systematic review of 14 original papers found that measurements from the copd-6 had an Area Under the Curve (AUC) of 0.90 for the diagnosis of COPD [4]. It should be noted that AUC reflects only precision, and not accuracy. As the number of persons with airway obstruction in our study was low, we do not have statistical power to calculate sensitivity and specificity of the copd-6 for obstruction; hence, our results cannot be directly compared with the systematic review. However, some authors of original studies have reported numerical results from the copd-6 that allow direct comparisons. In accordance with our results, one study showed a small but statistically significant overestimation of FEV1 by the copd-6, with underestimation at low flows and overestimation at high flows [11]. But on the other hand, seven studies indicated that the copd-6 underestimated FEV1 [1218], with estimated differences from diagnostic-quality spirometers ranging from 0.01 to 0.17 L. In two of the latter studies, Bland-Altman plots indicated that the copd-6 underestimates FEV1 more when the FEV1 is high [12, 18], while one study did not show a clear trend in FEV1 bias for the copd-6 [16]. We do not have a clear explanation for the differences between previous studies and the present one. Since we saw the same trends in participant and calibration check data, our results are not explained by any insufficiently calibrated MicroDL devices. Given that calibration check data from two different copd-6 devices gave very similar results (S3 File), our results are also not explained by a defective copd-6. A possible explanation for the between-study discrepancies could be the use of copd-6 devices from different lots that therefore differed in calibration, but that is purely speculative.

Out of the 304 participants who underwent spirometry in the current study, only 179 (59%) fulfilled quality criteria with the copd-6, while 290 (95%) fulfilled quality criteria with the MicroDL device. Since the same technicians were responsible for testing with the copd-6 and MicroDL devices, this difference is unlikely to be due to problems with coaching, as poor coaching would have influenced both copd-6 and MicroDL results. Furthermore, the corresponding author supervised the spirometry technicians during data collection, and it was his impression that the technicians vigorously coached the participants during each blow. Out of the 125 persons whose copd-6 results were excluded, 98 (78%) were excluded because they had performed less than two blows without warnings. It could be speculated that the copd-6 is perhaps too conservative when assessing the quality of individual blows (and labelling them as “unacceptable” by giving a warning), but since the device does not export spirograms for manual review, we cannot investigate this further.

A number of sensitivity analyses have been performed to investigate the robustness of our findings. Participants blew up to twelve times into the copd-6 and MicroDL devices, which could theoretically have biased our results due to fatigue after repeated forced exhalations. However, the order of testing (copd-6 before MicroDL or vice versa) was randomized, and sensitivity analyses stratified by order of testing showed similar results as the main analysis (S2 File), meaning that our findings are unlikely to be considerably influenced by patient fatigue. We also do not think that the different number of blows allowed with the two devices (three blows with the copd-6, up to nine blows with the MicroDL) poses a problem, as a sensitivity analysis limited to the first three blows with the MicroDL gave similar results as the main analysis (S2 File). To maximize the amount of available data and avoid selection bias, our main analysis used less strict criteria for the reproducibility of FEV1, FVC and FEV6 than recommended by the ATS [8]. However, while using less strict reproducibility criteria is expected to introduce imprecision in our estimates, there is no reason to think that it would lead to bias, and a sensitivity analysis using the official ATS criteria [8] also gave similar results as the main analysis (S2 File).

Study participants were recruited in a non-random manner by inviting members of a number of farmer’s groups in one specific district of Uganda, meaning that the study population may not be representative of Ugandan smallholder farmers overall. However, as the goal of the present study was not to assess the prevalence of lung function impairment among Ugandan smallholder farmers in general, but rather to compare FEV1 measurements made with the copd-6 and MicroDL spirometers in a specific study population, we do not think that this poses a threat to the validity of our findings regarding FEV1. As participants were relatively healthy, and the number of participants with airway obstruction was limited, it was difficult for us to assess the validity of FEV1/FEV6 measured by the copd-6. To further investigate the validity of the FEV6 and FEV1/FEV6 measurements, a new study would be needed, with a larger proportion of participants with airway obstruction, and results from the copd-6 should be compared to a diagnostic-quality spirometer that reports FEV6 in addition to FVC.

The tendency of the copd-6 to slightly underestimate volume at low flows and overestimate volume at high flows means that if it is used to measure FEV1 in studies of risk factors for pulmonary function impairment, exposure-response relationships may be biased away from the null. Researchers using FEV1 measurements from the copd-6 need to check the calibration of their devices and account for any inaccuracies.

Conclusion

Overall, in our relatively health study population the copd-6 slightly overestimates FEV1 by 0.04 L, but the bias is flow-dependent. The copd-6 could be considered as an affordable way to conduct research on pulmonary function impairment in resource-constrained settings, but careful calibration checks are necessary. Further validation in a study population with obstructive diseases is needed, and the copd-6 should be compared to a gold standard spirometer that reports FEV6 in addition to FVC.

Supporting information

S1 File. Deviations between analysis protocol and final analyses.

https://doi.org/10.1371/journal.pone.0253319.s001

(PDF)

S2 File. Supplementary results for participant data.

https://doi.org/10.1371/journal.pone.0253319.s002

(PDF)

S3 File. Supplementary results for copd-6 calibration check data.

https://doi.org/10.1371/journal.pone.0253319.s003

(PDF)

Acknowledgments

We wish to thank the participants and our collaborators from the Diálogos Foundation, the Uganda National Association of Community and Occupational Health, the Agency For Integrated Rural Development, Wakiso District Farmers Association and Caritas Uganda. We are thankful for the efforts of the field team: Amusa Wamawobe, Betty Kateregga, Brenda Wagaba, Evans Twin, Grace Lubega, Jonathan Mugweri, Imelda Namatovu, Joviah Gonza, Lydia Yariwo and Timothy Masaba.

Parts of Tables 1 and 2 have previously been published elsewhere and are reproduced here with permission [7].

References

  1. 1. Global, regional, and national deaths, prevalence, disability-adjusted life years, and years lived with disability for chronic obstructive pulmonary disease and asthma, 1990–2015: a systematic analysis for the Global Burden of Disease Study 2015. Lancet Respir Med. 2017;5(9):691–706. Epub 2017/08/22. pmid:28822787; PubMed Central PMCID: PMC5573769.
  2. 2. Siddharthan T, Grigsby M, Morgan B, Kalyesubula R, Wise RA, Kirenga B, et al. Prevalence of chronic respiratory disease in urban and rural Uganda. Bull World Health Organ. 2019;97(5):318–27. Epub 2019/09/26. pmid:31551628; PubMed Central PMCID: PMC6747035.
  3. 3. Kibirige D, Sanya RE, Nantanda R, Worodria W, Kirenga B. Availability and affordability of medicines and diagnostic tests recommended for management of asthma and chronic obstructive pulmonary disease in sub-Saharan Africa: a systematic review. Allergy Asthma Clin Immunol. 2019;15:14. Epub 2019/03/23. pmid:30899279; PubMed Central PMCID: PMC6407228.
  4. 4. Rosero Arenas MdlÁ García García MÁ, Briones Urtiaga MdM, Martínez Cornejo A. Utilidad del miniespirómetro COPD-6 en el diagnóstico precoz de EPOC. Open Respiratory Archives. 2020;2(3):132–40.
  5. 5. Hansen MRH, Jørs E, Sandbæk A, Sekabojja D, Ssempebwa JC, Mubeezi R, et al. Exposure to cholinesterase inhibiting insecticides and blood glucose level in a population of Ugandan smallholder farmers. Occup Environ Med. 2020. Epub 2020/07/08. pmid:32632029.
  6. 6. Hansen MRH. Exposure to pesticides in present-day use, diabetes mellitus and lung function impairment. PhD thesis. Aarhus, Denmark: Aarhus University; 2020. https://doi.org/10.5281/zenodo.3775222
  7. 7. Hansen MRH, Jørs E, Sandbæk A, Sekabojja D, Ssempebwa JC, Mubeezi R, et al. Organophosphate and carbamate insecticide exposure is related to lung function change among smallholder farmers: a prospective study. Thorax. 2021. Epub 2021/01/23. pmid:33479045.
  8. 8. Miller MR, Hankinson J, Brusasco V, Burgos F, Casaburi R, Coates A, et al. Standardisation of spirometry. European Respiratory Journal. 2005;26(2):319–38. pmid:16055882
  9. 9. Hartung C, Lerer A, Anokwa Y, Tseng C, Brunette W, Borriello G. Open data kit: tools to build information services for developing regions. Proceedings of the 4th ACM/IEEE International Conference on Information and Communication Technologies and Development; London, United Kingdom. 2369236: ACM; 2010. p. 1–12.
  10. 10. Hansen MRH, Jørs E, Sandbæk A, Sekabojja D, Ssempebwa J, Mubeezi R, et al. Protocol for statistical analyses in the study entitled "Pesticide exposure, asthma and diabetes in Uganda (PEXADU)". Zenodo. 2019.
  11. 11. Sami R, Omidi A, Sadegh R. Validity and Reliability of COPD-6 Device for Detecting Chronic Obstructive Pulmonary Disease in High-Risk Individuals. Tanaffos. 2020;19(3):201–7. Epub 2021/04/06. pmid:33815540; PubMed Central PMCID: PMC8008408.
  12. 12. Represas CR, Rial MB, Fernández VL, Silva AIG, del Campo Pérez V, Fernández-Villar A. Assessment of the Portable COPD-6 Device for Detecting Obstructive Airway Diseases. Archivos de Bronconeumología ((English Edition)). 2010;46(8):426–32.
  13. 13. Miravitlles M, Llor C, Calvo E, Diaz S, Díaz-Cuervo H, Gonzalez-Rojas N. Validación de la versión traducida del Chronic Obstructive Pulmonary Disease-Population Screener (COPD-PS). Su utilidad y la del FEV1/FEV6 para el diagnóstico de enfermedad pulmonar obstructiva crónica. Medicina Clínica. 2012;139(12):522–30. pmid:22015009
  14. 14. Thorn J, Tilling B, Lisspers K, Jorgensen L, Stenling A, Stratelis G. Improved prediction of COPD in at-risk patients using lung function pre-screening in primary care: a real-life study and cost-effectiveness analysis. Prim Care Respir J. 2012;21(2):159–66. Epub 2012/01/25. pmid:22270480; PubMed Central PMCID: PMC6548026.
  15. 15. Franco-Marina F, Fernandez-Plata R, Torre-Bouscoulet L, García-Sancho C, Sanchez-Gallen E, Martinez D, et al. Efficient screening for COPD using three steps: a cross-sectional study in Mexico City. NPJ primary care respiratory medicine. 2014;24:14002. Epub 2014/05/21. pmid:24841708; PubMed Central PMCID: PMC4373258.
  16. 16. Labor M, Vrbica Ž, Gudelj I, Labor S, Plavec D. Diagnostic accuracy of a pocket screening spirometer in diagnosing chronic obstructive pulmonary disease in general practice: a cross sectional validation study using tertiary care as a reference. BMC Family Practice. 2016;17(1). pmid:27542843
  17. 17. Llordés M, Zurdo E, Jaén Á, Vázquez I, Pastrana L, Miravitlles M. Which is the Best Screening Strategy for COPD among Smokers in Primary Care? COPD: Journal of Chronic Obstructive Pulmonary Disease. 2017;14(1):43–51. pmid:27797591
  18. 18. Thorat YT, Salvi SS, Kodgule RR. Peak flow meter with a questionnaire and mini-spirometer to help detect asthma and COPD in real-life clinical practice: a cross-sectional study. NPJ Prim Care Respir Med. 2017;27(1):32. Epub 2017/05/11. pmid:28487516; PubMed Central PMCID: PMC5435090.