Development of a loop-mediated isothermal amplification assay combined with a nanoparticle-based lateral flow biosensor for rapid detection of plasmid-mediated colistin resistance gene mcr-1

  • Lin Gong ,

    Contributed equally to this work with: Lin Gong, Fei Tang, Ernan Liu

    Roles Data curation, Methodology, Writing – original draft

    Affiliation Department of Disinfection and Pest Control, Wuhan Center for Disease Control and Prevention, Wuhan, People’s Republic of China

  • Fei Tang ,

    Contributed equally to this work with: Lin Gong, Fei Tang, Ernan Liu

    Roles Conceptualization, Writing – review & editing

    Affiliation MOE Key Laboratory of Environment and Health, Institute of Environmental Medicine, School of Public Health, Tongji Medical College, Huazhong University of Science and Technology, Wuhan, People’s Republic of China

  • Ernan Liu ,

    Contributed equally to this work with: Lin Gong, Fei Tang, Ernan Liu

    Roles Formal analysis, Investigation, Methodology, Project administration

    Affiliation Department of Disinfection and Pest Control, Wuhan Center for Disease Control and Prevention, Wuhan, People’s Republic of China

  • Xiaoli Liu,

    Roles Conceptualization, Project administration

    Affiliation Department of Disinfection and Pest Control, Wuhan Center for Disease Control and Prevention, Wuhan, People’s Republic of China

  • Huiqiong Xu,

    Roles Investigation, Methodology

    Affiliation Department of Disinfection and Pest Control, Wuhan Center for Disease Control and Prevention, Wuhan, People’s Republic of China

  • Yimei Wang,

    Roles Investigation, Methodology

    Affiliation Department of Disinfection and Pest Control, Wuhan Center for Disease Control and Prevention, Wuhan, People’s Republic of China

  • Yadong Song,

    Roles Investigation

    Affiliation Department of Disinfection and Pest Control, Wuhan Center for Disease Control and Prevention, Wuhan, People’s Republic of China

  • Jiansheng Liang

    Roles Conceptualization, Project administration, Supervision, Writing – review & editing

    wh_ljs@sina.com

    Affiliation Department of Disinfection and Pest Control, Wuhan Center for Disease Control and Prevention, Wuhan, People’s Republic of China

Abstract

A loop-mediated isothermal amplification assay combined with a nanoparticle-based lateral flow biosensor (LAMP-LFB) was established for the rapid and accurate detection of the mobilized colistin resistance gene (mcr-1), which causes the loss of colistin antibacterial efficacy in clinical treatments. The amplification stage of the assay was completed in 60 min at 63°C, and the reaction products could be visually detected by employing the LFB, which provided a fast (within 2 min) and objective method to evaluate the amplification results. The LAMP assay amplified the target sequences of mcr-1 with high specificity. In pure strains, the detection limit of the LAMP-LFB assay was 360 fg plasmid DNA/reaction, and in spiked feces samples the value was approximately 6.3×103 CFU/mL (~6.3 CFU/reaction), which was tenfold more sensitive than the PCR assay. The results show that the developed LAMP-LFB assay will be a worthy tool for the simple, rapid, specific, and sensitive detection of mcr-1 gene in clinical settings and resource-limited areas.

Introduction

As one of the last-resort drugs, colistin is an antibiotic that can treat the serious infections caused by carbapenem-resistant Enterobacteriaceae (CRE) [1]. Nevertheless, the global outbreak of CRE has brought about a surged use of colistin, and this phenomenon will inevitably increase the risk of developing colistin resistant bacteria.

In 2015, the understanding of the colistin resistance mechanism involving in chromosomal mutations changed when the plasmid-mediated resistance gene mcr-1 was discovered [2]. The mcr-1 was first discovered in China [2], and afterwards which has been detected in various countries including United States of America, Brazil, and Europe [3, 4]. This gene has been identified in different bacterial species, such as Klebsiella aerogenes [5], Citrobacter freundii [6], Escherichia coli [7], Enterobacter aerogenes [8], Klebsiella pneumoniae [2], Escherichia fergusonii [5], Kluyvera ascorbata [9]; and Salmonella enterica [10]. Not only are mcr-1-positive bacteria present in clinical settings (including veterinary hospitals), livestock and vegetable markets, but they can also be isolated from waterborne sources (sea and river water) [11]. The high transferability of the mcr-1 gene is due to the existence of mcr-1-bearing plasmid reservoirs such as IncFI [12], IncFII [13], IncHI1 [14], IncHI2 [15], IncP [16], IncX4 [3] and IncX3-X4 mosaic version [17]. Additionally, the coexistence with other antibiotic resistance genes (KPC, NDM and ESBL) [1820] broadens the bacteria drug resistance spectrum. The wide dissemination of mcr-1 could lead to fewer antibiotics being able to treat infections caused by multidrug-resistant isolates. Therefore, the rapid and accurate identification of mcr-1 could normalize the use of antibacterial agents and reduce the need for empirical therapy.

Recently, conventional and real-time polymerase chain reaction (PCR) assays were employed in the detection of the mcr-1 gene [21]. Although these common molecular methods possess some merits, such as rapidity and high sensitivity, the requirements of expensive devices and specialized personnel make them unsuitable for primary-care hospitals and “on-site” surveillance [22]. Loop-mediated isothermal amplification (LAMP) is a novel nucleic acid-based technique, known for its reliability, efficiency and rapidity, which has been employed in the detection of genetic material from parasites [23], viruses [24] and bacteria [25]. This approach amplifies DNA (without the denaturing step) at a constant temperature (60–70°C) using Bst DNA polymerase, which triggers the autocycling trait of strand displacement [26]. Due to the isothermal feature, a low-cost water bath can be used to incubate the reaction, instead of an expensive thermocycler [26]. The LAMP-amplified products can be routinely visualized by various means, such as turbidity measurement, electrophoresis, and colorimetric indicator [27]. However, these approaches present some disadvantages. For example, turbidity measurement and electrophoresis require sophisticated equipment, while the colorimetric indicator is unable to detect very low product concentrations. Alternatively, a gold nanoparticles-based lateral flow biosensor (LFB) [28] could be used to detect the amplified target products. Herein, we proposed a LAMP-LFB method to rapidly identify the mcr-1 gene and evaluated its analytical performance in fecal samples. To the best of our knowledge, it is the first reported LAMP assay coupled with an LFB designed for the screening of mcr-1.

Materials and methods

Bacterial strains

A total of 21 mcr-1-positive and 58 mcr-1-negative bacterial strains were used in this study (Table 1). The bacteria carrying the mcr-1 gene consisted of Escherichia coli, Escherichia fergusonii, and Salmonella enteritidis, for which 13, 7 and 1 isolates were acquired, respectively. Five genotype categories of carbapenemase producers (IMP-4, VIM-1, VIM-2, KPC-2, NDM-5) were incorporated in the non-mcr-1 isolates: 8 Escherichia coli, 28 Klebsiella pneumoniae, 2 Enterobacter cloacae, and 2 Pseudomonas aeruginosa. Similarly, the control group also included non-mcr-1/carbapenemase isolates (2 Escherichia coli, 4 Pseudomonas aeruginosa, 3 Acinetobacter baumannii, 2 Serratia marcescens, 3 Staphylococcus aureus, 3 Enterococcus faecalis, and 1 Enterococcus faecium). Conventional PCR and subsequent sequencing were used to identify all resistance genes. Genomic, plasmid, and spiked fecal samples DNA were extracted using bacterial genome, plasmid, and stool DNA extraction kits (Tiangen Biotech Co., Ltd., Beijing, China), respectively, according to the manufacturer’s instructions. DNA concentrations were determined using the NanoDrop 2000 (Thermo Fisher Scientific Inc., Waltham, MA, United States). Plasmid DNA extracted from mcr-1-producing E. coli (WHCDC128) was used as a template in the optimization of the LAMP conditions and determination of the detection limits. The LAMP assay was implemented using isothermal amplification kits, which included Bst DNA polymerase and reaction buffer (BeiJing-HaiTaiZhengYuan Technology Co., Ltd., Beijing, China). Centrifuge tubes loaded with reaction mixtures were set in a simple heating thermostat. Afterwards, the LAMP products could be detected via the disposable LFB (BeiJing-HaiTaiZhengYuan Technology Co., Ltd., Beijing, China), the colorimetric indicator Malachite Green (BeiJing-HaiTaiZhengYuan Technology Co., Ltd., Beijing, China), or electrophoresis, for which an UV transilluminator (Analytik Jena AG, Jena, Germany) was used to analyze the gel.

Primer design for the LAMP assay

We designed LAMP primers targeting the mcr-1 gene sequence (GenBank accession number: MK405590.1) using Primer Explore V5 (version 4) and Primer Premier (version 6.0). A cluster of six primers, including one pair of outer primers (F3 and B3), two loop primers (LF and LB), and two inner primers (FIP and BIP), were designed to target several distinct regions of mcr-1 gene. Moreover, the primers FIP# and LF# originated from the FIP and LF primers, which were labeled with biotin and FITC at the 5’ end, respectively. The primers detailed information is presented in Table 2. All oligomers were purified using HPLC.

thumbnail
Table 2. Primers used in the LAMP assay to identify the mcr-1 gene.

https://doi.org/10.1371/journal.pone.0249582.t002

Preparation of gold nanoparticle-based dipstick biosensor

The development of the LFB (4×6 cm) was based on the previous study of Wang and co-workers [29]. A set of four components–absorbent pad, NC membrane, conjugate pad, and sample pad–were orderly assembled on a support card and held together by a plastic casing. The monitoring reagents biotin-bovine serum albumin (2.5 mg/mL) and anti-FITC antibody (0.15 mg/mL) were immobilized on the detection regions of the dry-reagent strips, and formed the CL and TL, respectively. A 5-mm gap was set between the two lines. Cutting machine was used to cut the assembled cards into 4-mm isometric strips. Finally, the assembled biosensors were packed and stored at room temperature, in dry conditions.

LAMP assay

The LAMP assay was conducted in a 25-μL reaction volume, which contained Bst DNA polymerase (1 μL), colorimetric indicator (1 μL), reaction buffer (12.5 μL), inner primers (1.6 μM), loop primers (0.8 μM), outer primers (0.4 μM), and DNA template (1 μL). The negative controls were preformed using IMP-4-positive K. pneumoniae (WHCDC261) and KPC-2-producing E. coli (WHCDC332), while the blank control was done using nuclease-free water. The LAMP mixtures were incubated in isothermal conditions (60–67°C) for 60 min.

Three methods were used to detect the LAMP reaction products: electrophoresis (2% agarose gel), colorimetric indicator, and LFB. In the electrophoretic analysis, the reaction products (4 μL) were drifted at 100 V for 70 min. When using the colorimetric indicator, the color of the reaction mixture containing amplified products remained the same, while in the negative and blank controls the solution turned transparent [30]. When employing LFB, two red lines (CL and TL) could be visualized for the positive reactions, while only the CL was observed in the negative and blank controls [30].

Specificity and sensitivity of the mcr-1-LAMP-LFB method

DNA templates of 21 mcr-1-positive and 58 non-mcr-1 bacterial isolates were used to determine the specificity of the mcr-1-LAMP-LFB assay. Colorimetric indicator and electrophoresis were used to confirm the LAMP-LFB assay results, and each assay was performed twice. The sensitivity was determined using tenfold serial dilutions (3.6 ng/μL to 36 fg/μL) of plasmid DNA from the reference strain E. coli WHCDC128. Measurements were performed three times.

Evaluation of mcr-1-LAMP-LFB assay in spiked feces samples

The performance of the mcr-1-LAMP-LFB assay in real matrices was assessed using spiked human fecal samples, which were donated by a consenting adult. The samples were previously screened for the resistance gene, using the PCR assay, and were classified as mcr-1-negative. The spike test was carried out as described by Gong and co-workers [30]. Plate counting was performed after the mcr-1-producing reference strain E. coli WHCDC128 was consecutively diluted tenfold (6.3×106–6.3×101 CFU/mL). Further, Afterwards, a fecal sample was inoculated with 100 μL of a diluted culture, this was done for all dilutions. The spiked samples were extracted by a stool DNA extraction kits according to the manufacturer’s specification. The extracted DNA was eluted with 100 μL of buffer solution, and 1 μL of the eluted solution was used in the mcr-1-LAMP-LFB assay. The negative control was performed using a non-spiked fecal sample. Finally, three measurements were used for the determination of LAMP products.

PCR assay

The conventional PCR assay was performed as previously described by Gong and co-workers [30], using 5’-CGGTCAGTCCGTTTGTTC-3’ and 5’-CGGTCAGTCCGTTTGTTC-3’ as the forward and reverse primers, respectively, which were specific for the mcr-1 gene. The detection threshold of the PCR assay was determined using pure cultures and fecal samples.

Ethics statement

We utilized the fecal samples from a healthy volunteer in Wuhan Center for Disease Control and Prevention. The volunteer participated in this study when he had signed informed consent. The study protocol was reviewed and approved by the Institutional Review Board of Wuhan Center for Disease Control and Prevention (WHCDCIRB- SQ-2019012).

Results and discussion

Causing enormous mortality, CRE have become a serious threat in clinical settings. Colistin is recognized as the last line of defense in the treatment of infections caused by those superbugs. However, the antibiotic began to lose its bactericidal effect since the mcr-1 gene emerged, transmitted, and outbroke in various gram-negative bacteria. Moreover, the mcr-1 gene carried by transposons and plasmids could bring about no cure when combining with other antibiotic resistance genes, such as KPC and CTX-M, in a single isolate [31]. Thus, the identification of mcr-1 will contribute to control the gene’s horizontal transmission. Herein, we proposed a loop-mediated isothermal amplification assay coupled with a lateral flow biosensor (LAMP-LFB) to detect mcr-1 gene.

A set of mcr-1 specific primers (Table 2) were designed to target the gene’s conserved region and then screened for the LAMP assay, which was performed under isothermal conditions (62 oC) for 60 min. Amplification products were obtained from mcr-1-positive E. coli (WHCDC128), but not from IMP-4-positive K. pneumoniae (WHCDC261), KPC-2-producing E. coli (WHCDC332) and the blank control (Fig 1). Thus, these results confirmed that the three pairs of primers were suitable for the mcr-1-LAMP-LFB assay.

thumbnail
Fig 1. Confirmation of the mcr-1-LAMP products.

Three detection methods were used to identify the LAMP products: (a) colorimetric indicator; (b) 2% agarose gel electrophoresis; (c) LFB. Four samples were tested: 1. mcr-1-producing E. coli (WHCDC128); 2. IMP-4-positive K. pneumoniae (WHCDC261); 3. KPC-2-producing E. coli (WHCDC332); 4. distilled water. The positive results are shown in tube 1, lane 1, and biosensor 1.

https://doi.org/10.1371/journal.pone.0249582.g001

In the mcr-1-LAMP-LFB system, FIP# and LF# primers (Table 2) were respectively labeled with biotin and fluorescein isothiocyanate (FITC) at the 5’ end. After the LAMP reaction stage, the double-labeled amplicons would be recognized by the anti-fluorescein antibody immobilized in the test line (TL) of the LFB [28]. Further, After adding a small drop (2 μL) of amplification products solution to the biosensor’s test pad, followed by several drops of buffer solution [30], the results could be confirmed visually within 2 min, with the formation of one or two red lines. When compared with electrophoretic monitoring, analysis of LAMP products using LFB is a faster and simpler approach. Additionally, LFB has an accurate judgement when amplicons concentrations are very low, but the color change will be vague in that situation by using colorimetric indicator. Hence, LFB will become a preferable choice for the identification of LAMP products.

The optimal reaction temperature will increase the amplification efficiency of the LAMP assay. Therefore, plasmid DNA templates (3.6 pg/μL) from E. coli WHCDC128 were added into the mcr-1-LAMP reaction system in order to optimize the assay’s temperature conditions. Amplicons of mcr-1 were detected when the reaction was performed at 60–67°C, with 1°C increments for 60 min. According to the electrophoresis results, the best amplification efficiency was obtained in the 62–65°C range (Fig 2), and thus we chose 63°C as the temperature condition to perform further experiments.

thumbnail
Fig 2. Electrophoretic analysis of mcr-1 amplicons obtained under different temperature conditions.

Lanes 1–8: amplification products generated from E. coli WHCDC128 plasmid DNA (3.6 pg/μL) with incremented temperatures (60–67°C, 1°C increments). Lane 9: negative control (3.6 pg of E. coli WHCDC332 genomic DNA). Lane 10: blank control (nuclease-free water).

https://doi.org/10.1371/journal.pone.0249582.g002

The specificity of the LAMP assay was evaluated by employing LFB (Fig 3). Two red detection bands (TL and control line- CL) were observed for the mcr-1-positive strains samples, while only the CL was detected for all the non-mcr-1 samples, including mcr-1/carbapenemase-negative strains, carbapenemase-producing bacteria, and blank control. Therefore, the LAMP assay designed in this study was highly specific for mcr-1.

thumbnail
Fig 3. Specificity of the mcr-1-LAMP-LFB assay.

Biosensors 1–6, mcr-1-positive E. coli WHCDC128, E. coli from WHCDC, E. coli from ICDC, E. fergusonii from ICDC, E. fergusonii from WHCDC, and S. enteritidis from ICDC, respectively. Biosensors 7–14, IMP-4-positive K. pneumoniae, IMP-4-positive E. coli, VIM-1-positive E. cloacae, VIM-2-positive K. pneumoniae, KPC-2-producing K. pneumoniae, KPC-2-producing P. aeruginosa, KPC-2-producing E. coli, NDM-5-positive K. pneumoniae, respectively. Biosensors 15–21, A. baumannii, P. aeruginosa, S. marcescens, E. coli, E. faecalis, E. faecium, S. aureus, respectively (these isolates did not carry the aforementioned genes). Biosensor 22, blank control (nuclease-free water).

https://doi.org/10.1371/journal.pone.0249582.g003

The analytical sensitivity of the mcr-1-LAMP-LFB assay was determined using serial dilutions of DNA templates of E. coli WHCDC128. As shown in Fig 4, LFB analysis indicated that the concentration threshold of the LAMP assay was 360 fg/μL, which was in agreement with the colorimetric indicator and electrophoresis analysis. The practicability of the mcr-1-LAMP-LFB was assessed with stool samples inoculated with mcr-1-positive isolates. Positive results were observed when E. coli WHCDC128 concentration was higher than 6.3×103 CFU/mL (~6.3 CFU/reaction) (Fig 5), while lower concentrations, non-spiked sample, and blank control gave a negative result. Similarly, three different approaches obtained identical outcome. Hence, those manifested the established LAMP-LFB approach was suitable for mcr-1 detection.

thumbnail
Fig 4. Sensitivity of the mcr-1-LAMP-LFB assay.

Tubes (A)/lanes (B)/biosensors (C) 1–6 represented the levels of plasmid DNA (E. coli WHCDC128) 3.6 ng, 360 pg, 36 pg, 3.6 pg, 360 fg and 36 fg per reaction, respectively. Tubes (A)/lanes (B)/biosensors (C) 7 and 8 represented the negative (KPC-2-producing E. coli WHCDC332) and blank controls (nuclease-free water), respectively.

https://doi.org/10.1371/journal.pone.0249582.g004

thumbnail
Fig 5. Sensitivity of the mcr-1-LAMP-LFB assay in spiked fecal samples.

Tubes (A)/lanes (B)/biosensors (C) 1–6 represented the E. coli WHCDC128 DNA levels of 6.3×103, 6.3×102, 6.3×101, 6.3, 6.3×10−1, and 6.3×10−2 CFU per reaction, respectively. Tubes (A)/lanes (B)/biosensors (C) 7 and 8 represented the negative (non-spiked fecal sample) and blank controls (nuclease-free water), respectively.

https://doi.org/10.1371/journal.pone.0249582.g005

The detection limit of the method was 360 fg/L plasmid DNA in pure strains and 6.3×103 CFU/mL in spiked feces samples, which was tenfold more sensitive than the mcr-1-PCR assay (Table 3). Moreover, conventional PCR could be inhibited by various interfering substances present in practical specimens, but such inhibitors did not affect the amplification reaction of the LAMP assay [26]. The mcr-1-LAMP-LFB assay had the equivalent sensitivity and specificity with the mcr-1-MCDA-LFB method described in previously study [30], but the latter was more complex in terms of primers design and screening. Although the detection of mcr-1 was achieved with high sensitivity using other methods, like MALDI-TOF mass spectrometry and real-time PCR [21, 32], their application in fields of resource shortage would be limited by the need of expensive equipment and rigorous experimental conditions. The mcr-1-LAMP-LFB assay overcomes the aforementioned restrictions, requiring only a simple thermostat to supply a constant temperature (63°C), making the assay appropriate for the identification of mcr-1 on the spot. Importantly, the whole process of the trial, including specimen treatment (20 min), LAMP reaction (60 min) and detection (2 min), could be finished within 85 min. The test time was at least 90 min less than the standard mcr-1-PCR method. Therefore, the established method showed distinct merit in the aspect of testing time.

thumbnail
Table 3. Conventional PCR and mcr-1-LAMP-LFB assay detection limits.

https://doi.org/10.1371/journal.pone.0249582.t003

Conclusion

In conclusion, a LAMP-LFB assay was designed for the identification of the colistin resistance gene mcr-1. The method is simple, rapid, sensitive, and specific. The LFB required no additional equipment and could offer a fast and objective way for readout of amplified products. Thus, the mcr-1-LAMP-LFB assay will be a worthy tool in clinical settings and resource-poor areas.

Acknowledgments

We gratefully acknowledge Juan Li and Jie Che (National Institute for Communicable Disease Control and Prevention) for supplying the strains.

References

  1. 1. Nation RL and Li J. Colistin in the 21st century. Curr Opin Infect Dis. 2009; 22:535–543. pmid:19797945
  2. 2. Liu Y, Wang Y, Walsh TR, Yi L, Zhang R, Spencer J, et al. Emergence of plasmid-mediated colistin resistance mechanism mcr-1 in animals and human beings in China: a microbiological and molecular biological study. Lancet Infect Dis. 2016; 16:161–168. pmid:26603172
  3. 3. Gao R, Hu Y, Li Z, Sun J, Wang Q, Lin J, et al. Dissemination and mechanism for the MCR-1 colistin resistance. PLoS Pathog. 2016; 12:e1005957. pmid:27893854
  4. 4. Wang Q, Li Z, Lin J, Wang X, Deng X and Feng Y. Complex dissemination of the diversified mcr-1-harbouring plasmids in Escherichia coli of different sequence types. Oncotarget. 2016; 7:82112–82122. pmid:27741523
  5. 5. Wang R, Dorp LV, Shaw LP, Bradley P, Wang Q, Wang X, et al. The global distribution and spread of the mobilized colistin resistance gene mcr-1. Nat Commun. 2018; 9:1179. pmid:29563494
  6. 6. Li X, Fang L, Jiang P, Pan D, Xia J, Liao X, et al. Emergence of the colistin resistance gene mcr-1 in Citrobacter freundii. Int J Antimicrob Agents. 2017; 49:786–787. pmid:28433744
  7. 7. Li Z, Tan C, Lin J and Feng Y. Diversified variants of the mcr-1-carrying plasmid reservoir in the swine lung microbiota. Sci China Life Sci. 2016; 59:971–973. pmid:27520829
  8. 8. Zeng K, Doi Y, Patil S, Huang X and Tian G. Emergence of plasmid-mediated mcr-1 gene in colistin-resistant Enterobacter aerogenes and Enterobacter cloacae. Antimicrob Agents Chemother. 2016; 60:3862–3863. pmid:26976876
  9. 9. Zhao F and Zong Z. Kluyvera ascorbata strain from hospital sewage carrying the mcr-1 colistin resistance gene. Antimicrob Agents Chemother. 2016; 60:7498–7501. pmid:27671069
  10. 10. Li X, Fang L, Song J, Xia J, Huo W, Fang J, et al. Clonal spread of mcr-1 in PMQR-carrying ST34 Salmonella isolates from animals in China. Sci Rep. 2016; 6:38511. pmid:27917926
  11. 11. Chen K, Chan E, Xie M, Ye L, Dong N and Chen S. Widespread distribution of mcr-1-bearing bacteria in the ecosystem, 2015 to 2016. Euro Surveill. 2017; 22:pii = 17–00206. pmid:29019312
  12. 12. McGann P, Snesrud E, Maybank R, Corey B, Ong AC, Clifford R, et al. Escherichia coli harboring mcr-1 and blaCTX-M on a novel IncF plasmid: first report of mcr-1 in the United States. Antimicrob Agents Chemother. 2016; 60:4420–4421. pmid:27230792
  13. 13. Xavier BB, Lammens C, Butaye P, Goossens H and Malhotra-Kumar S. Complete sequence of an IncFII plasmid harbouring the colistin resistance gene mcr-1 isolated from Belgian pig farms. J Antimicrob Chemother. 2016. 71:2342–2344. pmid:27261261
  14. 14. Zurfluh K, Klumpp J, Nuesch-Inderbinen M and Stephan R. Full-length nucleotide sequences of mcr-1-harboring plasmids isolated from extended spectrum- beta-lactamase-producing Escherichia coli isolates of different origins. Antimicrob Agents Chemother. 2016; 60:5589–5591. pmid:27324774
  15. 15. Stoesser N, Mathers AJ, Moore CE, Day NPJ and Crook DW. Colistin resistance gene mcr-1 and pHNSHP45 plasmid in human isolates of Escherichia coli and Klebsiella pneumoniae. Lancet Infect Dis. 2016;16:285–286. pmid:26774239
  16. 16. Zhao F, Feng Y, Lu X, McNally A and Zong Z. IncP plasmid carrying the colistin resistance gene mcr-1 in Klebsiella pneumoniae from hospital sewage. Antimicrob Agents Chemother. 2016; 61:pii = e02229–16.
  17. 17. Sun J, Yang R, Zhang Q, Feng Y, Fang LX, Xia J, et al. Co-transfer of blaNDM-5 and mcr-1 by an IncX3-X4 hybrid plasmid in Escherichia coli. Nat Microbiol. 2016; 1:16176. pmid:27668643
  18. 18. Delgado-Blas JF, Ovejero CM, Abadia-Patino L and Gonzalez-Zorn B. Coexistence of mcr-1 and blaNDM-1 in Escherichia coli from Venezuela. Antimicrob Agents Chemother. 2016; 60:6356–6358. pmid:27431212
  19. 19. Zhang H, Seward CH, Wu Z, Ye H and Feng Y. Genomic insights into the ESBL and MCR-1-producing ST648 Escherichia coli with multi-drug resistance. Sci Bull (Beijing). 2016; 61:875–878. pmid:27358749
  20. 20. Tacao M, Tavares RDS, Teixeira P, Roxo I, Ramalheira E, Ferreira S, et al. mcr-1 and blaKPC-3 in Escherichia coli Sequence Type 744 after Meropenem and Colistin Therapy, Portugal. Emerging Infectious Diseases. 2017; 23:1419–1421. pmid:28726622
  21. 21. Bontron S, Poirel L and Nordmann P. Real-time PCR for detection of plasmid-mediated polymyxin resistance (mcr-1) from cultured bacteria and stools. J Antimicrob Chemother. 2016; 71:2318–2320. pmid:27121402
  22. 22. Niu L, Zhao F, Chen J, Nong J, Wang C, Wang J, et al. Isothermal amplification and rapid detection of Klebsiella pneumoniae based on the multiple cross displacement amplification (MCDA) and gold nanoparticle lateral flow biosensor (LFB). Plos one. 2018; 13: e0204332. pmid:30273362
  23. 23. Lodh N, Mikita K, Bosompem KM, Anyan WK, Quartey JK, Otchere J, et al. Point of care diagnosis of multiple schistosome parasites: species-specific DNA detection in urine by loop-mediated isothermal amplification (LAMP). Acta Trop. 2017; 173:125–129. pmid:28619672
  24. 24. Tian B, Ma J, de la Torre TZG. and Balint A. Rapid newcastle disease virus detection based on Loop-mediated isothermal amplification and optomagnetic readout. Acs Sensors. 2016; 1:1228–1234.
  25. 25. Yuan H, Chao Y, Li S, Tang M, Huang Y, Che Y, et al. Picoinjection-enabled multitarget loop-mediated isothermal amplification for detection of foodborne pathogens. Anal Chem. 2018; 90:13173–13177. pmid:30354065
  26. 26. Jin-Hee K and Se-Wook O. Development of a filtration-based LAMP-LFA method as sensitive and rapid detection of E. coli O157:H7. J Food Sci Technol. 2019; 56:2576–2583. pmid:31168139
  27. 27. Li L, Zhang S and Zhang C. Establishment of a rapid detection method for rice blast fungus based on one-step loop-mediated isothermal amplification (LAMP). Plant Dis. 2018; 103:1967–1973.
  28. 28. Wang Y, Li H, Wang Y, Zhang H, Xu J and Ye C. Loop-mediated isothermal amplification label-based gold nanoparticles lateral flow biosensor for detection of Enterococcus faecalis and Staphylococcus aureus. Front Microbiol. 2017; 8:192. pmid:28239371
  29. 29. Wang Y, Li H, Li D, Li K, Wang Y, Xu J, et al. Multiple cross displacement amplification combined with gold nanoparticle-based lateral flow biosensor for detection of Vibrio parahaemolyticus. Front Microbiol. 2016; 7:2047. pmid:28066368
  30. 30. Gong L, Liu E, Che J, Li J, Liu X, Xu H, et al. Multiple cross displacement amplification coupled with gold nanoparticles-based lateral flow biosensor for detection of the mobilized colistin resistance gene mcr-1. Front Cell Infect Microbiol. 2019; 9:226. pmid:31316917
  31. 31. Wang Q, Sun J, Li J, Ding Y, Li X, Lin J, et al. Expanding landscapes of the diversified mcr-1-bearing plasmid reservoirs. Microbiome. 2017; 5:70. pmid:28683827
  32. 32. Dortet L, Bonnin RA, Pennisi I, Gauthier L, Jousset AB, Dabos L. et al. Rapid detection and discrimination of chromosome- and MCR-plasmid- mediated resistance to polymyxins by MALDI-TOF MS in Escherichia coli: the MALDIxin test. J Antimicrob Chemother. 2018; 73: 3359–3367. pmid:30184212