Skip to main content
Advertisement
Browse Subject Areas
?

Click through the PLOS taxonomy to find articles in your field.

For more information about PLOS Subject Areas, click here.

  • Loading metrics

Different mechanisms of oxygenator failure and high plasma von Willebrand factor antigen influence success and survival of venovenous extracorporeal membrane oxygenation

  • Tamara Steiger,

    Roles Data curation, Formal analysis, Methodology, Software, Visualization, Writing – original draft

    Affiliation Department of Cardiothoracic Surgery, University Hospital Regensburg, Regensburg, Germany

  • Alois Philipp,

    Roles Conceptualization, Data curation, Methodology, Resources, Writing – review & editing

    Affiliation Department of Cardiothoracic Surgery, University Hospital Regensburg, Regensburg, Germany

  • Karl-Anton Hiller,

    Roles Methodology, Software, Validation, Writing – original draft

    Affiliation Department of Conservative Dentistry and Periodontology, University Hospital Regensburg, Regensburg, Germany

  • Thomas Müller,

    Roles Data curation, Project administration, Resources, Supervision

    Affiliation Department of Internal Medicine II, University Hospital Regensburg, Regensburg, Germany

  • Matthias Lubnow,

    Roles Data curation, Investigation, Resources, Writing – review & editing

    Affiliation Department of Internal Medicine II, University Hospital Regensburg, Regensburg, Germany

  • Karla Lehle

    Roles Methodology, Project administration, Supervision, Writing – review & editing

    karla.lehle@ukr.de

    Affiliation Department of Cardiothoracic Surgery, University Hospital Regensburg, Regensburg, Germany

Abstract

Objective

Failure of membrane oxygenator (MO) function of venovenous extracorporeal membrane oxygenators (VV ECMO) remains problematic. The development of device-induced coagulation disorder (COD) or worsened gas transfer (WGT) necessitates a system exchange. The aim was to correlate von Willebrand factor antigen (vWF:Ag) with the predisposition to MO failure and mortality.

Methods

Laboratory parameters (inflammation, coagulation) and ECMO-related data from 31 VV ECMO patients were analyzed before and after the first MO exchange. Study groups were identified according to the exchange reasons (COD, WGT) and the extent of vWF:Ag (low, ≤425%; high, >425%).

Results

vWF:Ag remained unchanged after system exchange. High vWF:Ag was associated with systemic endothelial activation of older and obese patients with elevated SOFA score, increased norepinephrine and higher requirement of continuous renal replacement therapy without an effect on MO runtime and mortality. Including the mechanism of MO failure (COD, WGT), various patient group emerged. COD/low vWF:Ag summarized younger and less critically ill patients that benefit mainly from ECMO by a significant improvement of their inflammatory and coagulation status (CRP, D-dimers, fibrinogen) and highest survival rate (91%). Instead, WGT/high vWF:Ag presented older and more obese patients with a two-digit SOFA score, highest norepinephrine, and aggravated gas transfer. They benefited temporarily from system exchange but with worst survival (33%).

Conclusions

vWF:Ag levels alone cannot predict early MO failure and outcome in VV ECMO patients. Probably, the mechanism of clotting disorder in combination with the vWF:Ag level seems to be essential for clot formation within the MO. In addition, vWF:Ag levels allows the identification different patient populations In particular, WGT/high vWF:Ag represented a critically ill population with higher ECMO-associated mortality.

Introduction

All types of mechanical circulatory support devices, including venovenous extracorporeal membrane oxygenation (VV ECMO), bear a high risk of clot formation [1,2]. Both, thrombotic deposits on the surface of gas exchange fibers [3,4] as well as the extended clots inside the membrane oxygenators (MO) are discussed to impair the trans-membranous gas transfer and may eventually lead to MO failure [1,5,6]. An immediate exchange improved gas transfer performance and recovered critical coagulation parameters [1,4]. In general, two failure patterns were identified [1]: Worsened gas transfer (WGT) of the MO describes insufficient oxygenation and decarboxylation of patient blood. Device-induced coagulation disorder (COD) summarizes patients with critical changes of coagulation parameters (e.g. D-dimers, fibrinogen, platelets) and the development of acute clot formation within the pump head and MO. However, the underlying mechanisms remained unclear. Until now, there is no conclusive association between clot volume and time, anticoagulation regime or underlying disease of the patient [7].

The von Willebrand Factor (vWF) could be an important indicator protein for clot formation. ECMO support induced an acquired von Willebrand syndrome (AVWS) that is characterized by a loss of high molecular weight vWF (HMWvWF) due to high shear stress [8] and also a release of vWF of low activity from the endothelium in acute-phase response [9]. Furthermore, high accumulations of vWF in form of granular, cobweb-like and fiber-like structures were detected inside clotted MOs from ECMO patients with severe critical illness [7]. In these patients, it is obvious that shear stress activates vWF and accelerates platelet binding as an essential part in the process of clot formation [1014].

In this study, we hypothesized that high plasma levels of the vWF antigen (vWF:Ag) of VV ECMO patients that required a system exchange due to WGT or COD are an indicator for impending clot formation inside the MO and therefore, might act as a predictor for MO failure and survival.

Materials and methods

Study design and study population

The Regensburg ECMO database was queried for consecutive patients with VV ECMO support (11/2017 to 11/2019, n = 128) and documented time dependent vWF:Ag courses. Only patients with at least one system exchange and a support duration of more than 2 days were included (n = 31). Parameters indicating the need for an exchange of the ECMO system were defined by Lubnow et al. [1]. Worsened gas transfer (WGT, 11/31, 35%) included patients that required a new system due to a decrease in CO2 transfer accompanied by an up-regulation of sweep gas flow and an increase in the partial pressure of carbon dioxide post MO (pCO2 post MO). The values improved with a new system (S1A Fig). Device-induced coagulation disorder (COD, 20/31, 65%) summarized all patients that presented any kind of detectable clotting disorder: Hyperfibrinolysis (n = 6; 19%) was defined by an increase in D-dimers and a decrease in fibrinogen (and platelet counts) (S1B Fig). Clot formation within the pump head (pump head thrombosis (PHT; n = 7; 23%) was identified by an increase in plasma free hemoglobin (fHb) and lactate dehydrogenase (LDH) (S1C Fig). A sudden increase in the pressure drop across the MO (dpMO) was an indicator for extended and occlusive clots within the MO (acute MO thrombosis, n = 7; 23%) (S1D Fig). All parameters improved after a system exchange.

Furthermore, vWF:Ag levels before an exchange were used to subdivide the exchange groups. Since almost all detected vWF:Ag levels were elevated a threshold value was defined that is the median of all values before exchange (425%). Two vWF:Ag groups were defined (vWF:Ag ≤425%, n = 16; vWF:Ag >425%, n = 15; hereinafter referred to as low and high vWF:Ag, respectively). Finally, four study groups were identified: low vWF:Ag/COD (n = 11), high vWF:Ag/COD (n = 9), low vWF:Ag/WGT (n = 5), high vWF:Ag/WGT (n = 6). Patient characteristics and ECMO data are summarized in Table 1.

The primary endpoint was the effect of vWF:Ag levels on the clotting mechanism that predicts the need for an exchange. The secondary endpoint analyzed the effect of vWF:Ag levels on the MO run time and the outcome of the patients.

ECMO management

The management of ECMO patients and the procedure to monitor MO performance has been previously described in detail [1]. Laboratory parameters were gathered on a daily routine (Institute for Clinical Chemistry and Laboratory Medicine, University Hospital Regensburg): e.g. activated partial thromboplastin time (aPTT), antithrombin III (ATIII), fibrinogen, D-dimers, C-reactive protein (CRP), lactate dehydrogenase (LDH), content of hemoglobin (Hb) as well as free hemoglobin (fHb), leukocyte and platelet count. Anticoagulation therapy with heparin or argatroban was administered according to requirements. In case there was no increased risk for bleeding, we aimed for an aPTT of 60sec. The median initial heparin dosage (ECMO day 1) was 6.2 [4.3–14] U/h/kg]. The median initial argatroban dosage was 1.4 [0.9–1.7] mg/h. MO performance characteristics (blood flow, sweep gas flow, CO2 transfer, O2 transfer, pressure drop across the MO, dpMO), blood gas analysis (BGA) results (partial pressure of oxygen, PO2; partial pressure of carbon dioxide, PCO2; each in arterial, venous and post MO blood samples), dosage of norepinephrine and anticoagulative therapy are gathered for the Regensburg ECMO database on a daily basis. O2 and CO2 transfer over the membrane are calculated based on formerly gathered BGA values [15]. Factor VIII, factor XIII, as well as vWF antigen (vWF:Ag test kit, Siemens Healthcare Diagnostics Products, Marburg, Germany) and vWF activity (Innovance vWF:Act, Siemens) were detected before and after a system exchange. vWF:Ag standard values range between 57.6 and 174.1% and vWF:Act standard values range between 47.8 and 173.2%. High molecular weight multimers were not analyzed.

Data collection and statistical analysis

The study was approved by the Ethics Committee of the University of Regensburg (vote no. 20-1827-104). All data were collected in the Regensburg ECMO database from 11/2017 to 11/2019 (medical records from the University Hospital Regensburg) and fully anonymized for retrospective analysis (12/2019 to 6/2020). Need for informed consent was waived by the Ethics Committee as all devices are approved for clinical use, no personalized data and only routine laboratory parameters were used. Nominal and ordinal scaled data were depicted as frequencies and statistically analyzed using the χ2-test. Interval-scaled data were depicted as median and neighboring quartiles and statistically analyzed using Kruskal-Wallis, Wilcoxon, or Mann-Whitney-Test. Significance level was set to p = 0.05. For evaluation of different subgroups’ influence in general, the level of significance α was adjusted to α*(k) = 1-(1-α)1/k with the Error Rates Method (k = number of paired tests performed). The software package IBM SPSS-Statistics 25.0 (IBM Corp., Armonk, NY, USA) was used for the statistical evaluation.

Results

Study population

vWF:Ag levels pre exchange were used to define the study groups: the median vWF:Ag level was 425% (vWF:Ag ≤425%, vWF:Ag >425%, hereinafter referred to as low and high vWF:Ag). As shown in Table 1, patients with high vWF:Ag levels were significantly older (p = 0.008), had a higher BMI (p = 0.001) and higher SOFA (Sequential Organ Failure Assessment) score (p = 0.006). Furthermore, they required continuous renal replacement therapy (CRRT; p = 0.004) more often compared to patients with low vWF:Ag levels. Cannula size and model were comparable, except for the choice of the drainage vessel (p = 0.038; S1 Table). Apart from that, all other patient characteristics and ECMO relevant parameters were comparable between both vWF:Ag groups.

11/31 (35%) and 20/31 (65%) of all MOs were exchanged due to WGT and COD respectively. According to the findings of Lubnow et al. [1], COD is a heterogeneous group consisting of hyperfibrinolysis, pump head thrombosis and trans-membrane pressure drop (S1 Fig). Of special interest were patients with high vWF:Ag and WGT that highlighted with old age, highest BMI, double-digit SOFA-scores and increased need for CRRT. Instead, low vWF:Ag and COD were dominant in young patients with normal BMI and less critical illness (SOFA).

Technical complications demanding an exchange of the MO

MO complications leading to a system exchange can be assigned to two different failure patterns. WGT of the gas exchange membranes with insufficient oxygenation and decarboxylation of the patient’s blood and device-induced COD, which describes changes in coagulation parameters leading to clot formation in the MO and inducing procoagulant processes in the patient [1]. In order to analyze the varieties between WGT and COD, median laboratory and ECMO-related data over three days before the MO was exchanged were compared (Table 2).

thumbnail
Table 2. Ventilation-related data and laboratory parameters over 3 days before the MO was exchanged.

https://doi.org/10.1371/journal.pone.0248645.t002

WGT patients presented higher fibrinogen levels (p = 0.007) and required a higher gas flow (p = 0.040) with increased CO2 transfer (p = 0.029). A reduced platelet count was also striking, but without statistical significance. The content of hemoglobin, the fHb, aPTT, AT III, leukocyte count, D-dimers, the dose of norepinephrine as well as blood flow were equal in both groups.

Impact of vWF:Ag plasma levels on laboratory and ECMO-related parameters

High vWF:Ag was accompanied by high vWF:Act (p<0.001) as well as altered inflammatory parameters: higher levels of LDH (p = 0.006) and fibrinogen (p = 0.045) and a tendency to higher CRP (p = 0.054). Additionally, patients with high vWF:Ag levels required a higher dose of norepinephrine (p = 0.041). Other parameters were not different between both vWF:Ag groups.

Including the vWF:Ag levels it was shown that patients in the COD group with low vWF:Ag levels were associated with lowest median values for vWF:Act, CRP, LDH, factor VIII, factor XIII, fibrinogen and CO2 transfer compared to the other subgroups (Table 2). However, only LDH and fibrinogen reached significance levels. These patients also needed less norepinephrine. For statistical comparison within the COD group, see Table 2. Another remarkable subgroup was the WGT group with high vWF:Ag levels. These patients presented highest levels of vWF:Act, LDH, CRP, gas flow, and CO2 transfer and highest dosage of norepinephrine, while the pO2 post MO was the lowest among all subgroups.

vWF:Ag in the context of a MO exchange

A system exchange had no effect on vWF:Ag and vWF:Act independently of the exchange reason (COD or WGT) and the level of vWF:Ag (Table 3A and 3B). Statistical analysis compared the median values of laboratory and ECMO-relevant data (before, day of, post exchange) from the four subgroups. A system exchange led to significant changes in relevant parameters, particularly in COD patients with low vWF:Ag (Table 3A). Levels of LDH (p = 0.017) and fHb (p = 0.016) increased, while levels of fibrinogen (p = 0.033) and CO2 transfer (p = 0.026) were significantly lower on the exchange day. After exchange patients in this group presented lower CRP (p = 0.033), a shortened aPTT (p = 0.021), higher fibrinogen (p = 0.013), lower D-dimers (p = 0.005) and higher AT III (p = 0.021). Furthermore, fHb and pO2 post MO tended to improve after exchange (p = 0.062, each). pCO2 post MO tended to improve with the exchange (p = 0.053). In addition, the patients required less blood flow compared to the three days before exchange (p = 0.007). COD patients with high vWF:Ag showed less statistical benefit from a system exchange. Only CRP (p = 0.021), LDH (p = 0.038), D-dimers (p = 0.028) and fHb (p = 0.021) improved significantly after a system exchange. The leukocyte count increased minimally after exchange (p = 0.050) but remained approximately in the normal range. Gas exchange parameters were unremarkable.

thumbnail
Table 3.

A. Ventilation-related data and laboratory parameters pre and post MO exchange in COD patients. B. Ventilation-related data and laboratory parameters pre and post MO exchange in WGT patients.

https://doi.org/10.1371/journal.pone.0248645.t003

A system exchange due to WGT and low vWF:Ag only presented a significant reduction of D-dimers (p = 0.043) while pCO2 post MO tended to be lower after MO exchange. The latter increased significantly before MO exchange (p = 0.042). An increase in gas flow before and a decrease after exchange was shown, but without statistical significance. CO2 and O2 transfer were inconspicuous. In contrast, patients with WGT and high vWF:Ag benefited significantly from a system exchange. Initially low platelet counts normalized or even increased after the MO replacement (p = 0.028). Other WGT specific gas data improved after exchange—patients required less gas flow (p = 0.066). In addition, worsened CO2 transfer and increased pCO2 post MO improved significantly after exchange (p = 0.046 and p = 0.028, respectively).

As shown in S2 Table, there was no difference in the need for transfusion of red blood cells, fresh frozen plasma and platelet concentrates.

MO runtime and vWF:Ag plasma levels

High levels of vWF:Ag tended to shorten MO runtime (p = 0.054; Table 4). This effect was most pronounced in the COD group (pairwise comparison see Table 4). However, the differences (comparing all subgroups) were not significant due to the high scatter with small number of cases. The level of vWF:Ag had no effect on the total duration of ECMO independently of the exchange reasons (Table 4).

Outcome in case of worsened gas transfer or coagulation disorder

As shown in Table 4, patients whose first MO was changed due to COD were more often successfully weaned (p = 0.006) and significantly more patients discharged hospital (p = 0.015). In contrast, the level of vWF:Ag was not an indicator for successful weaning or discharge from hospital. Subgroup analysis disclosed highest successful weaning rates (91%, significant) and survival rates (91%, not significant) for patients with COD and low vWF:Ag. In the WGT group there was a 60% (3/5) mortality with low vWF:Ag and a 67% (4/6) mortality with high vWF:Ag.

Discussion

While the implantation of an ECMO system resulted in a decrease of vWF:Ag and vWF:Act [9,16], a system exchange had no effect on the levels of vWF:Ag or vWF:Act. Furthermore, the exchange reason (COD, WGT) had no effect on vWF:Ag or vWF:Act. However, the level of vWF:Ag combined with the exchange reason identified different patient populations. Patient populations were defined regarding state of critical illness, inflammation and coagulation status, benefit of a system exchange, MO runtime, and mortality. Low vWF:Ag levels and COD represent younger patients with minor acute-phase response, on trend prolonged MO runtime and highest survival rates. In contrast, high vWF:Ag levels and WGT define critically ill patients with the highest mortality rate and with severe respiratory failure requiring an escalation of trans-membranous gas transfer.

Elevated vWF:Ag levels in ECMO patients, a decrease within one day on ECMO and normalization after explant of ECMO have been reported by Tauber et al. [16]. vWF:Ag is elevated during acute-phase response [17] due to endothelial activation and injury in acute and chronic inflammation [18]. Almost all of the vWF:Ag levels before a system exchange were elevated. Therefore, we subdivided our patient population due to the median vWF:Ag level of 425% to get comparable group sizes. A similar approach has been described in other studies [19,20]. A threshold value of ≥450% was associated with a greater risk of developing acute lung injury (ALI) or acute respiratory distress syndrome (ARDS) in non-septic and septic patients [19,20]. It has been associated with adverse outcomes (mortality, duration of unassisted ventilation, organ failure) in patients with established ALI/ARDS [19]. Therefore, it was not surprising that all patients requiring VV ECMO presented elevated vWF:Ag and vWF:Act levels [9,16]. An adjustment to the threshold value of 450% did not change our results. We concluded that high vWF:Ag levels in VV ECMO patients indicate a severe inflammatory response and reflect the degree of critical illness of the patient.

In our study, patients with vWF:Ag >425% were of older age with elevated BMI and in a state of critical illness (increased SOFA score, need for CRRT). The elevated vWF-Ag reflected an increased acute phase response accompanied by increased levels of CRP and fibrinogen. Correlations between CRP and vWF:Ag in ECMO patients have also been described by Tauber et al. [16] but without differentiation of the vWF:Ag level. Furthermore, high vWF:Ag was associated with increased LDH levels indicating distinct organ damage. Under these conditions, the runtime of the first MO tended to be shorter without affecting the patients’ outcome. We assume that the severe critical illness with high vWF:Ag levels promote protein as well as platelet and leukocyte adhesion to the foreign surface of the MO [1014] and progress clot formation inside the MO that manifested in a reduced MO runtime and early MO failure.

The differentiation of reasons for a system exchange (WGT, COD) resulted in differences regarding inflammatory and coagulation markers, gas transfer capability, run time of the MO and mortality. Patient characteristics were comparable. Exchange criteria were identified previously [1]. Within 3 days before an exchange of the MO, the WGT group presented less platelets and required higher gas flow to meet the increased demand of CO2 elimination. The COD group is heterogeneous including all subgroups with detectable clot formation within the system: Sudden increases in fHb levels identified clots within the pump head (PHT). An increase in the dpMO indicated the growth of clots within the MO resulting in the occlusion of the MO (acute oxygenator thrombosis). The typical alterations in specific coagulation parameters (e.g. d-dimers, fibrinogen, platelet counts) described an impending clotting disorder (hyperfibrinolysis) [1]. Analysis of the subgroups failed due to less group size. The COD group presented a significant loss in fibrinogen levels, while the gas exchange capability of the MO remained unchanged. Despite these differences, the MO runtime as well as total ECMO supporting time was comparable between COD and WGT. However, WGT (vs. COD) resulted in less frequency of successful weaning (36% vs. 85%, p = 0.006) and reduced survival (36% vs. 80%, p = 0.015). We assume that the failing mechanism that ended in a system exchange was different comparing WGT and COD. Thrombotic deposits in membrane lungs are probably responsible for the deterioration in the gas exchange performance of artificial lungs [6]. Different mechanisms are possible. Already the adhesion of proteins and binding of blood cells onto the gas exchange fibers may impede its performance (represents the WGT group in our study). Furthermore, adhesive proteins initiate extensive clotting inside the MO and reduce the perfused area (represented by COD in our study) [6]. In both cases, the involvement of the vWF could be conceivable—a linker for blood cells (adhesion/ interconnection of platelets and leukocytes) or as a shear-dependent component in clot formation [7,1012]. In future studies, the relationship between the different clot mechanisms (COD, WGT) and the actual clot volume and the clot composition within the MOs is to be investigated [3,4,7].

Including the inflammatory component—verified by high and low vWF:Ag—patients from the COD/ low vWF:Ag group presented moderate critical illness (younger, lower SOFA score) without systemic inflammation (low CRP, LDH, fibrinogen). The mechanism of a device-induced COD seems to be reversible by exchanging the system. Our data show significant reduction of CRP, aPTT and D-dimers and an increase of AT III in these patients. After the system exchange, flow inside the MO normalized. Therefore, triggers of coagulation and inflammation are controllable temporarily. Surprisingly, 91% of these patients could be weaned successfully from ECMO and all of these patients discharged hospital. In contrast, in COD patients with high vWF:Ag levels a system exchange showed limited efficiency regarding inflammatory and coagulation parameters. Furthermore, the insertion of a new MO did not influence oxygenation and decarboxylation parameters. However, initially high levels of CRP and D-dimers decreased significantly after the system exchange, indicating that the local inflammation and hyperfibrinolysis could be remedied. So, clotting inside the MO in COD patients with high vWF:Ag is not only device-induced, but also patient-induced due to a pre-existing systemic inflammation. According to our data we suggest, that one has to differentiate between moderate inflammation (vWF:Ag ≤425%) due to foreign surface activation inside the MO and severe inflammation (vWF:Ag >425%) which originates mostly in a pre-existing status of systemic inflammation. Severe inflammation is only to a small amount induced by the MO itself. High vWF:Ag levels enhance the local inflammatory response to the foreign surface of the MO leading to clot formation and early MO failure. The procoagulant tendency of vWF depends on the predominant shear rate [21,22]. Therefore, the reduction in MO runtime in COD patients with high vWF:Ag levels might be due to alterations of blood flow. Moreover, increased local shear rates which favor the vWF-mediated clot formation and also local inflammation [23] might be considered as possible reasons. Furthermore, high levels of fHb on the exchange day normalized within one day. The dominance of this parameter is based on the fact that 35% of the COD patients presented a proven PHT. By replacing the clotted pump head, fHb decreased immediately [1].

In the WGT group with low vWF:Ag, the raise of gas flow before and the subsequent decrease after exchange as well as the alterations of pCO2 post MO mainly reflected worsening of gas transfer capability [1]. The additional reduction of D-dimers indicated the presence of thrombotic deposits within the used MO that disappeared after implantation of a new MO [24,25]. In contrast, a system exchange did not affect the other relevant parameters in the patient population. In particular, high levels of the inflammatory CRP remained high. The latter observation could also indicate that WGT has a different mechanism of clot formation within the system compared to COD (see above). In particular, the subgroup with WGT and high vWF:Ag identified critically ill patients with highest SOFA scores, obese BMI, elevated requirement of CRRT, high dosage of norepinephrine, and increased need for extracorporeal oxygenation. In contrast, a system exchange did not improve inflammation or coagulation within the WGT group with low vWF:Ag. However, the gas exchange situation improved and the system-induced thrombocytopenia regenerated significantly. The latter was also described by Lubnow et al. [1]. Despite increased pre-existing systemic inflammation (high CRP) MO runtime was not impaired with regard to vWF:Ag levels. We speculated that these patients, who tend to be in a severe septic status with thrombotic microangiopathy lack the ability to clot inside the MO as a result of disseminated intravascular coagulation (DIC) [26,27]. Furthermore, since extended protein adhesion to the gas exchange membrane reduces gas exchange capacity, there are still bedside coping strategies such as increasing gas flow or increasing blood flow to preserve sufficient gas transfer [28,29]. Furthermore, the gas transfer recovered or improved after a system exchange—this is a clear indication that protein or cellular deposits on the gas exchange membrane are responsible for the worsening of the gas transfer. Despite the improvement of the gas transfer rates with a new system, 33% of these patients died on ECMO.

Limitations of this study include its retrospective and single-center character and inclusion of only 31 VV ECMO patients with even smaller subgroups. Furthermore, the data evaluation was incomplete for vWF:Ag, vWF:Act, factor VIII and factor XIII that prevented a clear correlation with other parameters. The threshold of vWF:Ag (425%) used exceeds the standard value maximum of 174%. Therefore both study groups though indicated with “high” or “low” vWF:Ag levels represent patients with vWF:Ag levels high above standard values.

Conclusion

While typical coagulation parameters improve significantly when a system exchange is initiated, the vWF:Ag plasma levels of VV ECMO patients are not an indicator for impeding clot formation. Probably, the mechanism of clotting disorder in combination with the vWF:Ag level seems to be essential for the development of clots within the ECMO system. Nevertheless, the vWF:Ag level may provide information about the reversibility of local inflammatory processes inside the MO. Overall, the quantification of vWF:Ag levels alone neither can predict patient’s outcome nor can it predict early MO failure in VV ECMO patients.

Supporting information

S1 Fig. Parameters indicating need for system exchange.

https://doi.org/10.1371/journal.pone.0248645.s001

(PDF)

S1 Table. Cannulation specifications of the study cohort.

https://doi.org/10.1371/journal.pone.0248645.s002

(PDF)

S2 Table. Transfusion of red blood cell concentrates, fresh frozen plasma and platelet concentrates.

https://doi.org/10.1371/journal.pone.0248645.s003

(PDF)

References

  1. 1. Lubnow M, Philipp A, Foltan M, Bull Enger T, Lunz D, Bein T, et al. Technical complications during veno-venous extracorporeal membrane oxygenation and their relevance predicting a system-exchange—retrospective analysis of 265 cases. PLoS ONE. 2014;9(12):e112316. pmid:25464516
  2. 2. Philipp A, De Somer F, Foltan M, Bredthauer A, Krenkel L, Zeman F, et al. Life span of different extracorporeal membrane systems for severe respiratory failure in the clinical practice. PLoS ONE. 2018;13(6):e0198392. pmid:29856834
  3. 3. Birkenmaier C, Dornia C, Lehle K, Müller T, Gruber M, Philipp A, et al. Analysis of Thrombotic Deposits in Extracorporeal Membrane Oxygenators by High-resolution Microcomputed Tomography: A Feasibility Study. ASAIO J. 2020; 66(8):922–8. pmid:32740353
  4. 4. Dornia C, Philipp A, Bauer S, Lubnow M, Müller T, Lehle K, et al. Analysis of thrombotic deposits in extracorporeal membrane oxygenators by multidetector computed tomography. ASAIO J. 2014;60(6):652–6. pmid:25158889
  5. 5. MacLaren G, Combes A, Bartlett RH. Contemporary extracorporeal membrane oxygenation for adult respiratory failure: life support in the new era. Intens Care Med. 2012;38(2):210–20. pmid:22147116
  6. 6. Epis F, Belliato M. Oxygenator performance and artificial-native lung interaction. J Thorac Dis. 2018;10(S5):S596–605. pmid:29732176
  7. 7. Steiger T, Foltan M, Philipp A, Mueller T, Gruber M, Bredthauer A, et al. Accumulations of von Willebrand factor within ECMO oxygenators: Potential indicator of coagulation abnormalities in critically ill patients? Artif Organs. 2019;43(11):1065–76. pmid:31192471
  8. 8. Heilmann C, Geisen U, Beyersdorf F, Nakamura L, Benk C, Trummer G, et al. Acquired von Willebrand syndrome in patients with extracorporeal life support (ECLS). Intens Care Med. 2012;38(1):62–8.
  9. 9. Kalbhenn J, Schmidt R, Nakamura L, Schelling J, Rosenfelder S, Zieger B. Early Diagnosis of Acquired von Willebrand Syndrome (AVWS) is Elementary for Clinical Practice in Patients Treated with ECMO Therapy. J Atheroscler Thromb. 2015;22(3):265–71. pmid:25186021
  10. 10. Huck V, Schneider M, Gorzelanny C, Schneider S. The various states of von Willebrand factor and their function in physiology and pathophysiology. Thromb Haemost. 2014;111(04):598–609. pmid:24573248
  11. 11. Savage B, Saldívar E, Ruggeri ZM. Initiation of platelet adhesion by arrest onto fibrinogen or translocation on von Willebrand factor. Cell. 1996;84(2):289–97. pmid:8565074
  12. 12. Doggett TA, Girdhar G, Lawshé A, Schmidtke DW, Laurenzi IJ, Diamond SL, et al. Selectin-Like Kinetics and Biomechanics Promote Rapid Platelet Adhesion in Flow: The GPIbα-vWF Tether Bond. Biophys J. 2002;83(1):194–205. pmid:12080112
  13. 13. Millar JE, Fanning JP, McDonald CI, McAuley DF, Fraser JF. The inflammatory response to extracorporeal membrane oxygenation (ECMO): a review of the pathophysiology. Crit Care. 2016;20(1):387. pmid:27890016
  14. 14. Cheung P-Y, Sawicki G, Salas E, Etches PC, Schulz R, Radomski MW. The mechanisms of platelet dysfunction during extracorporeal membrane oxygenation in critically ill neonates: Crit Care Med. 2000;28(7):2584–90. pmid:10921599
  15. 15. Tschaut RJ. Extrakorporale Zirkulation in Theorie und Praxis. 1. Auflage. Pabst Science Publishers; 2005. 648 p.
  16. 16. Tauber H, Ott H, Streif W, Weigel G, Loacker L, Fritz J, et al. Extracorporeal Membrane Oxygenation Induces Short-Term Loss of High-Molecular-Weight von Willebrand Factor Multimers. Anesth Analg. 2015;120(4):730–6. pmid:25565317
  17. 17. Pottinger BE, Read RC, Paleolog EM, Higgins PG, Pearson JD. von Willebrand factor is an acute phase reactant in man. Thromb Res. 1989;53(4):387–94. pmid:2467404
  18. 18. Vischer UM. von Willebrand factor, endothelial dysfunction, and cardiovascular disease. J Thromb Haemost. 2006;4(6):1186–93. pmid:16706957
  19. 19. Ware LB, Eisner MD, Thompson BT, Parsons PE, Matthay MA. Significance of Von Willebrand Factor in Septic and Nonseptic Patients with Acute Lung Injury. Am J Respir Crit Care Med. 2004;170(7):766–72. pmid:15201135
  20. 20. Rubin DB, Wiener-Kronish JP, Murray JF, Green DR, Turner J, Luce JM, et al. Elevated von Willebrand factor antigen is an early plasma predictor of acute lung injury in nonpulmonary sepsis syndrome. J Clin Invest. 1990;86(2):474–80. pmid:2384595
  21. 21. Schneider SW, Nuschele S, Wixforth A, Gorzelanny C, Alexander-Katz A, Netz RR, et al. Shear-induced unfolding triggers adhesion of von Willebrand factor fibers. Proc Natl Acad Sci USA. 2007;104(19):7899–903. pmid:17470810
  22. 22. Rack K, Huck V, Hoore M, Fedosov DA, Schneider SW, Gompper G. Margination and stretching of von Willebrand factor in the blood stream enable adhesion. Sci Rep. 2017;7(1):14278. pmid:29079767
  23. 23. Zilberman-Rudenko J, Sylman JL, Lakshmanan HHS, McCarty OJT, Maddala J. Dynamics of blood flow and thrombus formation in a multi-bypass microfluidic ladder network. Cell Mol Bioeng. 2017;10(1):16–29. pmid:28580033
  24. 24. Lubnow M, Philipp A, Dornia C, Schroll S, Bein T, Creutzenberg M, et al. D-dimers as an early marker for oxygenator exchange in extracorporeal membrane oxygenation. J Crit Care. 2014;29(3):473.e1–5. pmid:24508200
  25. 25. Dornia C, Philipp A, Bauer S, Stroszczynski C, Schreyer AG, Müller T, et al. D-dimers Are a Predictor of Clot Volume Inside Membrane Oxygenators During Extracorporeal Membrane Oxygenation. Artif Organs. 2015;39(9):782–7. pmid:25845704
  26. 26. Claus RA, Bockmeyer CL, Budde U, Kentouche K, Sossdorf M, Hilberg T, et al. Variations in the ratio between von Willebrand factor and its cleaving protease during systemic inflammation and association with severity and prognosis of organ failure. Thromb Haemost. 2009;101(2):239–47. pmid:19190805
  27. 27. Ono T, Mimuro J, Madoiwa S, Soejima K, Kashiwakura Y, Ishiwata A, et al. Severe secondary deficiency of von Willebrand factor-cleaving protease (ADAMTS13) in patients with sepsis-induced disseminated intravascular coagulation: its correlation with development of renal failure. Blood. 2006;107(2):528–34. pmid:16189276
  28. 28. Park M, Costa ELV, Maciel AT, Silva DPE, Friedrich N, Barbosa EVS, et al. Determinants of oxygen and carbon dioxide transfer during extracorporeal membrane oxygenation in an experimental model of multiple organ dysfunction syndrome. PLoS ONE. 2013;8(1):e54954. pmid:23383011
  29. 29. Chauhan S, Subin S. Extracorporeal membrane oxygenation, an anesthesiologist’s perspective: Physiology and principles. Part 1. Ann Card Anaesth. 2011;14(3):218. pmid:21860197