Abstract
Infectious endocarditis is a life-threatening disease, and diagnostics are urgently needed to accurately diagnose this disease especially in the case of prosthetic valve endocarditis. We show here that maltohexaose conjugated to indocyanine green (MH-ICG) can detect Staphylococcus aureus (S. aureus) infection in a rat model of infective endocarditis. The affinity of MH-ICG to S. aureus was determined and had a Km and Vmax of 5.4 μM and 3.0 X 10−6 μmol/minutes/108 CFU, respectively. MH-ICG had no detectable toxicity to mammalian cells at concentrations as high as 100 μM. The in vivo efficiency of MH-ICG in rats was evaluated using a right heart endocarditis model, and the accumulation of MH-ICG in the bacterial vegetations was 2.5 ± 0.2 times higher than that in the control left ventricular wall. The biological half-life of MH-ICG in healthy rats was 14.0 ± 1.3 minutes, and approximately 50% of injected MH-ICG was excreted into the feces after 24 hours. These data demonstrate that MH-ICG was internalized by bacteria with high specificity and that MH-ICG specifically accumulated in bacterial vegetations in a rat model of endocarditis. These results demonstrate the potential efficacy of this agent in the detection of infective endocarditis.
Citation: Takemiya K, Røise JJ, He M, Taing C, Rodriguez AG, Murthy N, et al. (2021) Maltohexaose-indocyanine green (MH-ICG) for near infrared imaging of endocarditis. PLoS ONE 16(3): e0247673. https://doi.org/10.1371/journal.pone.0247673
Editor: Anjani Kumar Tiwari, Babasaheb Bhimrao Ambedkar University (A Central University), INDIA
Received: August 22, 2020; Accepted: February 11, 2021; Published: March 1, 2021
Copyright: © 2021 Takemiya et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Data Availability: All relevant data are within the manuscript and its Supporting Information files.
Funding: This research is funded by American Heart Association (AHA-SDG 17SDG33410912, KT) and National Institute of Health (NIH1R01 EB020008-01A1 and 1R01EB029320-01A1, WRT, NM and MMG). The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
Competing interests: Kiyoko Takemiya, W. Robert Taylor, Niren Murthy, and Mark M. Goodman have equity interest in Microbial Medical, Inc which is related to this study. Joachim J. Røise, Maomao He, Chung Taing, and Alexander G. Rodriguez have no conflict of interest. This does not alter our adherence to PLOS ONE policies on sharing data and materials.
Introduction
Infectious endocarditis (IE) is a systemic infection in which bacteria form vegetations on the heart valves. The diagnosis of IE is made based on the modified Duke criteria in which the involvement of endocardium or supporting tissue is one of the major diagnostic criteria [1]. Transthoracic echocardiography is widely accepted as an approach to assist in the diagnosis of IE. However, there are limitations to this imaging modality. Disadvantages include limited visualization due to body habitus and imaging in the setting of a prosthetic valve in which there are artifacts that limit visualization of the valve structures [2]. While transesophageal echocardiography improves the diagnostic accuracy in many cases, it has limitations related to imaging; it is also a semi-invasive procedure. Echocardiography is also of limited utility in its ability to determine whether the bacterial infection in the heart has been eliminated by antibiotic therapy. Therefore, there is an unmet need to develop more sensitive and less invasive imaging modalities to assist in the diagnosis and management of IE.
To develop novel diagnostic methods that detect the existence of bacteria in the tissue and organs, we have focused on the differences in sugar uptake by bacteria and mammalian cells [3–5]. Bacteria have maltodextrin transporters that actively internalize maltodextrins as an energy source; mammalian cells do not. Of particular importance to our approach is that the maltodextrin transporter can accept maltodextrins that bear structural modifications such as the addition of fluorescent tags [3, 6–8]. Therefore, maltodextrin-based imaging probes conjugated with fluorescent dyes and radionuclides, such as the positron emitter fluorine-18, can detect the existence of bacteria in the body with high sensitivity and specificity [3–5]. We recently showed that maltohexaose-based imaging probes—[F-18]fluoro-maltohexaose for positron emission tomography (PET) imaging and maltohexaose-IR786 for near infrared fluorescent imaging—are potentially useful to detect pacemaker pocket infections in a rat model [4]. We also observed that the accumulation of maltohexaose-based imaging probes is very low in the heart; other studies also indicated low accumulation of maltodextrin analogues in the heart [3, 4, 9–13]. These findings suggested the potential utility of maltohexaose-based imaging probes for the detection of IE. Maltohexaose-IR786 has a strong fluorescent signal that is detectable with clinical near infrared fluorescent imaging devices, but there is still a concern of potential cytotoxicity. Therefore, we developed a second generation maltohexaose-based imaging probe, maltohexaose-indocyanine green (MH-ICG). ICG is a near infrared fluorescent dye with low toxicity that is in clinical use and is excreted via the liver without being modified [14]. Near infrared wavelength light is permeable to soft tissue and blood for up to a few centimeters, and ICG is clinically used for vascular imaging and lymph node detection [15]. As a derivative of ICG, MH-ICG is detectable with current near infrared imaging devices for clinical use, and is therefore a promising candidate for determining the extent of bacterial infection. In this study, we evaluated the basic characteristics of MH-ICG in vitro and in vivo and whether MH-ICG is accumulated in bacterial vegetations in a rat model of endocarditis.
Materials and methods
All animal protocols were approved by the Emory University Institutional Animal Care and Use Committee along with the guidelines and rules of Association for Assessment and Accreditation of Laboratory Animal Care International (AAALAC International). All animal experiments were conducted in accordance with an approved protocol (PROTO201700806).
Synthesis of maltohexaose-indocyanine green (MH-ICG)
MH-ICG was synthesized through a copper-catalyzed click reaction with MH-azide [3] (25 mg, 0.023 mmol, 1 eq) and commercially available ICG-alkyne (21.5 mg, 0.028 mmol, 1.2 eq, obtained from Iris biotech GMBH) in the presence of DIPEA (1 mg, 7.6 μmol, 0.33 eq) and CuI in catalytic amounts (Fig 1A). After mixing in DMF (0.01 M to ICG-alkyne) at room temperature for 48 hours, the product was diluted in DI H2O (1:9) and filtered before purification using HPLC on a XBridge Prep C8 5 um OBD, 19 x 150 mm semi-prep column (flow rate: 10 ml/min. Solvent system: MeCN/H2O + 0.1% TFA, 0–20 min: 37% MeCN, 20–50 min: gradient to 100% MeCN; the product eluted at 27 min.). The resulting MH-ICG was analyzed using mass spectrometry (ESI+: found: 1863.7391m/z, calc: 1863.7408m/z for [C87H120O35N6SNa+], S1 Fig), and the purity was assessed by analytical HPLC (Fig 1B). The spectrofluorimetric properties of MH-ICG were assessed, and the absorption and emission maxima were observed at 782 nm and 814 nm, respectively, with emission being measured following excitation at 760 nm. (Fig 1C). We used 760 and 830 nm as the excitation and emission wavelengths, respectively.
a) Maltohexaose-azide was conjugated with ICG-alkyne through a copper-catalyzed click reaction. b) Analytical HPLC confirmed the purity of MH-ICG. Purified MH-ICG was injected onto an analytical Xbridge C18 5 μm 4.6 x 150 mm column. (Flow rate: 1 ml/min. Solvent system: MeCN/H2O + 0.1% TFA, 0–20 min: 40% MeCN, 20–40 min: gradient to 100% MeCN.) Product eluted at 27.2 min. c) Fluorescence spectrum of MH-ICG. The absorbance and fluorescence intensities are indicated as solid and broken lines, respectively. The emission spectrum was generated with an excitation wavelength of 760 nm.
Cytotoxicity assay
CHO-K1 cells (ATCC CCL-61) cultured in F-12K medium with 10% FBS were used to evaluate the cytotoxicity of MH-ICG. On day -1, the cells were seeded in 96 well plates at density of 1 X 103 cells/well. On day 0, the cells were loaded with 0, 2.5, 10, 25, 50, or 100 μM of MH-ICG, and the cells were cultured for 24 and 72 hours. On days 1 and 3, the medium with MH-ICG was changed to 100 μl/well of fresh F-12K medium followed by 20 μl/well of a tetrazolium compound-based reagent to evaluate the cell viability (3-(4,5-dimethylthiazol-2-yl)-5-(3-carboxymethoxyphenyl)-2-(4-sulfophenyl)-2H-tetrazolium, inner salt; MTS, CellTiter 96 AQueous One Solution Cell Proliferation Assay, Promega). The cells were incubated at 37°C for 60 minutes, and the formazan generated was measured at 490 nm with a microplate reader (Varioskan LUX, Thermo Fisher Scientific). The viability of MH-ICG treated cells was calculated as [absorbance of MH-ICG loaded cells]/[absorbance of control cells] (n = 4 /each concentration).
Uptake of MH-ICG by bacteria
In this study, we used Staphylococcus aureus (S aureus, ATCC 25923; Gram-positive bacteria) because S. aureus is one of the most commonly observed pathogens in patients with IE [1, 2, 16, 17]. To evaluate the affinity of MH-ICG for S. aureus, the bacteria at a density of 1 X 108 colony forming units (CFU)/ml were cultured with MH-ICG at concentrations of 0, 2.5, 5, 10, 20, and 30 μM for 1 hour at 37°C. The bacteria in 1 ml of culture were then washed in phosphate buffered saline (PBS) 3 times and re-suspended in 100 μl of PBS. The amount of MH-ICG internalized by bacteria was evaluated with a fluorescent microplate reader (Varioskan LUX, Thermo Fisher Scientific) (n = 3/each concentration).
Right heart IE model
Male Sprague Dawley rats (SD rats) weighing 250 to 275g were obtained from Charles River Laboratories. The animals were anesthetized with 1–2% isoflurane, and plastic catheters were inserted into the right ventricle via the right jugular vein. The location of the catheter was determined by the backflow of blood. The proximal end of the catheter was connected to a vascular access port (Braintree Scientific, Inc, U.S.A.) implanted on the back, and the incision was closed. On post-operative day 1 (POD 1), the rats were injected with 1 X 108 CFU/0.1 ml of S. aureus via the tail vein. The rats were used for experiments on POD 3 (two days after inoculation of S. aureus) when a vegetation had formed around the catheter in the right ventricle. Control rats did not receive catheterization nor inoculation of bacteria. (n = 5/each group). The rats were euthanized with carbon dioxide after the experiments.
Quantification of bacteria in the right ventricle
On POD 3, the right heart infectious endocarditis model rats (IE rats) were sacrificed with carbon dioxide, and the heart was extracted aseptically. The heart was cut in pieces and placed in Luria-Bertani (LB) broth. The bacteria in the heart were extracted and suspended in LB broth using a TissueLyser II (Qiagen). Serial dilutions of the suspensions were cultured on LB agar plates to determine the number of bacteria in the hearts.
Accumulation of MH-ICG in vivo
On POD 3, the IE rats were injected with 0.25 ml of 1 mM MH-ICG via the tail vein. Rats without catheterization and inoculation were used as controls. The rats were sacrificed 4 hours after injection of MH-ICG, and the hearts were sliced in three segments and scanned with an in vivo imaging device (In-Vivo Xtreme, Bruker). The fluorescent intensity in the vegetation and that in the left ventricle free wall were measured in each slice, and the mean intensity ratio defined as [the mean intensity of the vegetation]/ [the mean intensity of the left ventricle free wall] was calculated for each slice. The average of the intensity ratio in 3 slices was used as the intensity ratio of the sample. In the control samples, the mean intensity in the right ventricle free wall was used instead of that of the vegetation.
Biological half-life of MH-ICG
Male SD rats weighing 250 to 275 g were used for this experiment. The animals were anesthetized with 1–2% isoflurane and injected with 0.25 ml of 1 mM MH-ICG via the tail vein. At 15 and 30 minutes as well as 1, 2, 4, and 6 hours after injection of MH-ICG, 100 μl of blood was sampled with a heparinized syringe from the lateral tail vein opposite to the one used for injection of MH-ICG (n = 4 /each timepoint). The blood samples were centrifuged at 5000 X g for 10 minutes, and the plasma concentrations of MH-ICG were evaluated by a near infrared fluorescent microplate reader (Varioskan LUX, Thermo Fisher Scientific). The biological half-life of MH-ICG was calculated based on a two-compartment model with the following formula.
Where Cn indicates the plasma concentration of MH-ICG in the sample n, tn indicates the sampled time point, and T indicates the biological half-life. Based on this formula, the biologil half-life in each animal and each time point was calculated.
Biodistribution of MH-ICG
Male SD rats weighing 180 g to 190 g were injected with 0.25 ml of 1 mM MH-ICG via the tail vein and were housed in metabolic cages to collect feces and urine. The rats were sacrificed 24 hours after injection of MH-ICG, and the organs and tissue were harvested. The samples were homogenized with RIPA buffer using a TissueLyser II (Qiagen). The concentrations of MH-ICG in samples were evaluated with a fluorescent microplate reader (Varioskan LUX, Thermo Fisher Scientific) (n = 4).
Results
Basic characteristics of MH-ICG
We first evaluated the kinetics of MH-ICG internalization by S. aureus. The relationship between the concentrations of MH-ICG and the internalizing velocity of S. aureus is shown in Fig 2A. S. aureus internalized MH-ICG even at low concentrations, and the uptake velocity plateaued at a concentration of 20 μM. For 108 CFU of S. aureus, the Km and Vmax calculated using a Lineweaver-Burk plot analysis were 5.4 μM and 3.0 X 10−6 μmol/minutes/108CFU, respectively.
To evaluate the uptake of MH-ICG in bacteria, S. aureus, E. coli, and LamB mutant E. coli were cultured with MH-ICG at various concentrations (n = 3 /each concentration). The internalized amount of MH-ICG by S. aureus (a) and E. coli (b) followed Michaelis-Menten kinetics. The Km and Vmax calculated with a Lineweaver-Burk plot were 5.4 μM and 3.0 X 10−6 μmol/minutes/108 CFU for S aureus and 6.9 μM and 6.3 X 10−7 μmol/minutes/108 CFU for E. coli, respectively. LamB mutant E. coli internalized a small amount of MH-ICG at the level of detection. c) Cytotoxicity of MH-ICG was evaluated in CHO-K1 cells. The cell viability was evaluated for 24 hours (i) and 72 hours (ii) after loading of MH-ICG. The cells were incubated with 0 to 100 μM of MH-ICG (n = 4 /each concentration), and no reduction in cell viability was observed up to 100 μM of MH-ICG at either time points.
We also evaluated the kinetics of MH-ICG in Escherichia. coli (E. coli, ATCC 33456) and LamB mutant E. coli (JW-3996-1), which lacks maltoporin on the outer membrane [18]. As shown in Fig 2B, the Km and Vmax were 6.9 μM and 6.3 X 10−7 μmol/minutes for 108 CFU of E. coli, respectively. The amount of MH-ICG internalized by 1 X 108 CFU of LamB mutant E. coli was 25.1 ± 4.4 pmol using 30 μM MH-ICG compared to 19.7 ± 1.1 pmol internalized with 2.5 μM MH-ICG. The uptake of MH-ICG by LamB mutant E. coli did not follow Michaelis-Menten kinetics consistent with the loss of specific maltoporin-mediated uptake. These results demonstrate that MH-ICG uptake by bacteria is highly specific and occurs via the maltodextrin transport system.
The cytotoxicity of MH-ICG was evaluated in CHO-K1 cells. The MH-ICG showed no effects on cell growth or viability at concentrations of up to 100 μM for up to 3 days emphasizing the safety of MH-ICG for mammalian cells (Fig 2C).
MH-ICG accumulates in bacterial vegetations
The rats developed right heart endocarditis two days after inoculation of S. aureus (Fig 3A). Histological evaluation with Gram staining confirmed the accumulation of Gram-positive cocci in vegetations (Fig 3B). Total bacterial counts in the myocardium were (1.3 ± 0.5) X 108 CFU (n = 3).
a) Right heart endocarditis was established by catheterizing in the right ventricle followed by injection of S. aureus. The vegetation was found around the catheter in the right ventricle (the arrow). b) i) Whole image of the specimen. Arrow: vegetation, *: the catheter in right ventricle, RV: right ventricle, LV: left ventricle. Bar: 2 mm. ii) Low power field of vegetation around the catheter. In the vegetation, spreading and accumulation of Gram-positive bacteria are observed in the fibrous tissue. Bar: 250 μm. iii) High-power field of vegetation. Gram-positive cocci are observed among large inflammatory cell infiltrates. Bar: 25 μm. iv) High-power field of vegetation with accumulation of bacteria. The deep purple spot is composed of Gram-positive cocci. Bar: 25 μm. c) Near infrared imaging with MH-ICG in the right heart endocarditis model in rats. Accumulation of MH-ICG was only observed in the vegetation. The hearts from the control rats had a very low fluorescent signal. d) Accumulation of MH-ICG in vegetation was quantified as an intensity ratio defined as [fluorescent intensity in vegetation or the right ventricle]/[fluorescent intensity in the left ventricle free wall]. The intensity ratio in IE rats was 2.5 ± 0.2, which was significantly increased versus control rats (n = 5 /group).
To evaluate the accumulation of MH-ICG in the vegetations, the IE and control rats were injected with MH-ICG and sacrificed 4 hours after injection of MH-ICG. Fig 3C shows that MH-ICG accumulation was only observed only in the vegetation. The fluorescent intensity of MH-ICG accumulating in the vegetation was 2.5 ± 0.2 times higher than that in the left ventricle free wall, and this intensity ratio in the IE group was significantly increased versus the control group (Fig 3D).
Plasma half-life and biodistribution of MH-ICG
The pharmacokinetics of MH-ICG were evaluated in rats without infection. The MH-ICG rapidly cleared from the circulation with a pattern consistent with a dual mode of elimination (Fig 4A). The half-life of MH-ICG calculated from the plasma concentrations at 15 and 30 minutes was 14.0 ± 1.3 minutes; the half-life for the second phase from 15 and 60 minutes was 25.8 ± 3.8 minutes. Two hours after injection of MH-ICG, the plasma concentration was less than 5% compared with the plasma concentration of MH-ICG at 15 minutes. The plasma concentrations of MH-ICG decreased slowly after that. These results demonstrate that MH-ICG is cleared from the circulation very rapidly, which is desirable to minimize background fluorescence.
Healthy rats were injected with MH-ICG. a) The plasma concentration of MH-ICG was monitored for six hours. The plasma concentration was reduced very rapidly in two hours, and the biological half-life of MH-ICG in the initial phase was calculated as 14.0 ± 1.3 minutes (n = 4). Note that the concentration of MH-ICG is on a logarithmic scale, and some standard error bars overlap the average circles. b) Metabolism and distribution of MH-ICG were also evaluated: After 24 hours, about 50% of injected MH-ICG was excreted into feces. The MH-ICG that remained in the liver and in the kidneys were 11.7 ± 3.5% and 1.5 ± 0.2%, respectively (n = 4).
The biodistribution studies revealed that approximately 50% of the injected MH-ICG was excreted within 24 hours with the majority of the product in the feces (Fig 4B). The remaining MH-ICG was found in the liver (11.7 ± 3.5%) and kidneys (1.5 ± 0.2%). In the other organs, such as the heart, lungs, and intestine, less than 1% of the injected MH-ICG remained. The serum concentration of MH-ICG at 24 hours after injection was below the limit of detection. These results suggest that MH-ICG is primarily metabolized by the liver and excreted into the feces.
Discussion
We report a novel and fluorescence-based imaging probe that can detect bacterial endocarditis lesions in a highly specific fashion. We used an in vitro assay system to demonstrate that MH-ICG is not taken up by mammalian cells and that uptake by bacteria largely occurs specifically through the maltodextrin transport system. The in vivo data confirm that systemically administered MH-ICG is effectively taken up by bacteria in our endocarditis model and that the free MH-ICG is effectively cleared from the circulation relatively quickly. The latter is important for both minimizing potential toxicity as well as decreasing background signal. In summary, these results support the concept that MH-ICG has the appropriate in vivo characteristics for localization of bacterial endocarditis.
The model of endocarditis that we employed of these studies is somewhat unique in that it is a model of right heart endocarditis. Several previous studies of experimental endocarditis have used an aortic valve endocarditis model [19–23]. We elected not to use aortic valve endocarditis because of the higher incidence of systemic embolism and the very high rate of mortality (up to 40% within 48 hours) [20]. The right heart model of endocarditis that we employed exhibited low mortality with all of the animals surviving until the studies were completed. In the right heart endocarditis model, the vegetation is formed around the catheter in the right ventricle and involves the tricuspid valve thus limiting hemodynamic compromise. The number of bacteria in our IE model was in the range of 108 CFU, which is similar to that in the aortic valve endocarditis model [20–22].
MH-ICG is our second generation of maltohexaose-based fluorescent probe; it has ICG as the probe molecule instead of IR-786 used in our first generation of maltohexaose-based fluorescent probes [3, 4]. Both ICG and IR-786 are fluorescent cyanine dyes, but IR-786 consists of indolium and has a “rigidification ring” in the center of a polymethylene bridge. ICG consists of benz[e]indolium with sulfonate groups and has no “rigidification ring”. These differences affect the excitation and emission wavelengths as well as the stability and cytotoxicity of IR-786 and ICG [24]. Conceição et al. demonstrated that the cytotoxicity of polymethylene dyes on Hela cells is caused by the degradation products of these dyes, and the existence of the sulfonate groups and the “rigidification ring” do not affect the stability of these dyes [25]. They also report that the cytotoxicity of polymethylene dyes on Hela cells is related to the permeability of these dyes through the cell membrane [25]. Here, we showed that MH-ICG has lower cytotoxicity on CHO-K1 cells likely due to decreased permeability as a result of conjugation with maltohexaose. The half-life of ICG in aqueous solutions is about 17 hours [26]. This is sufficient for clinical use including bacterial imaging, but this shorter half-life of ICG might also contribute to a reduction in MH-ICG cytotoxicity.
The biodistribution studies were interesting in that approximately half of the injected MH-ICG was excreted into feces in 24 hours. This could reflect degradation of MH-ICG because ICG itself is known to be excreted into feces. Alternatively, the addition of ICG to MH might alter the clearance mechanism, i.e., 18F-MH is cleared primarily by the kidneys [4, 27].
The limitation of MH-ICG in diagnosis of endocarditis is the permeability of the near infrared wavelength. Similar to ICG, the penetration of excitation and emission wavelength of MH-ICG is limited to a few centimeters of tissue, and endocarditis lesions could not be imaged through the chest wall with MH-ICG. However, visualization of the extent of bacterial infiltration could be assessed during valve replacement and other surgeries for infectious lesions such as orthopedic device infection; similar fluorescence imaging is done in certain settings [15, 28–31]. Moreover, MH-ICG might be useful for bedside diagnosis of device infections near the skin such as pacemakers and left ventricular assist devices (LVADs) infections. In future, it is possible that the development of small near infrared wavelength detectors might enable the early diagnosis of endocarditis with a trans-esophageal approach. Finally, with proper gating, it is possible that PET-based imaging could provide the necessary resolution for detecting bacterial endocarditis using MH-based probe.
Conclusion
In this study, we demonstrated that MH-ICG can determine the presence and extent of bacterial infection in the setting of endocarditis in a highly specific manner. MH-ICG has low cytotoxicity and is a viable candidate for further study of potential clinical applications; it may provide a unique diagnostic tool for bacterial endocarditis.
Acknowledgments
The Murthy, Goodman, Takemiya, and Taylor laboratories contributed equally to this manuscript.
References
- 1. Cahill TJ, Prendergast BD. Infective endocarditis. Lancet. 2016;387(10021):882–93. pmid:26341945.
- 2. Nataloni M, Pergolini M, Rescigno G, Mocchegiani R. Prosthetic valve endocarditis. Journal of cardiovascular medicine. 2010;11(12):869–83. pmid:20154632.
- 3. Ning X, Lee S, Wang Z, Kim D, Stubblefield B, Gilbert E, et al. Maltodextrin-based imaging probes detect bacteria in vivo with high sensitivity and specificity. Nature Materials. 2011;10(8):602–7. pmid:21765397.
- 4. Takemiya K, Ning X, Seo W, Wang X, Mohammad R, Joseph G, et al. Novel PET and near infrared imaging probes for the specific detection of bacterial infections associated with cardiac devices. JACC Cardiovasc Imaging. 2018. pmid:29680350; PubMed Central PMCID: PMC6186207.
- 5. Ning X, Seo W, Lee S, Takemiya K, Rafi M, Feng X, et al. PET imaging of bacterial infections with fluorine-18-labeled maltohexaose. Angewandte Chemie. 2014;53(51):14096–101. pmid:25330976; PubMed Central PMCID: PMC4430476.
- 6. Gopal S, Berg D, Hagen N, Schriefer EM, Stoll R, Goebel W, et al. Maltose and maltodextrin utilization by Listeria monocytogenes depend on an inducible ABC transporter which is repressed by glucose. PloS one. 2010;5(4):e10349. pmid:20436965; PubMed Central PMCID: PMC2860498.
- 7. Schonert S, Seitz S, Krafft H, Feuerbaum EA, Andernach I, Witz G, et al. Maltose and maltodextrin utilization by Bacillus subtilis. Journal of Bacteriology. 2006;188(11):3911–22. pmid:16707683; PubMed Central PMCID: PMC1482931.
- 8. Dahl MK, Manson MD. Interspecific reconstitution of maltose transport and chemotaxis in Escherichia coli with maltose-binding protein from various enteric bacteria. Journal of Bacteriology. 1985;164(3):1057–63. pmid:3905762; PubMed Central PMCID: PMC219297.
- 9. Axer A, Hermann S, Kehr G, Clases D, Karst U, Fischer-Riepe L, et al. Harnessing the maltodextrin transport mechanism for targeted bacterial imaging: structural requirements for improved in vivo stability in tracer design. ChemMedChem. 2018;13(3):241–50. pmid:29195027.
- 10. Gowrishankar G, Hardy J, Wardak M, Namavari M, Reeves RE, Neofytou E, et al. Specific imaging of bacterial infection using 6’’-18F-fluoromaltotriose: A second-generation PET tracer targeting the maltodextrin transporter in bacteria. Journal of Nuclear Medicine. 2017;58(10):1679–84. pmid:28490473; PubMed Central PMCID: PMC5632736.
- 11. Gowrishankar G, Namavari M, Jouannot EB, Hoehne A, Reeves R, Hardy J, et al. Investigation of 6-[(1)(8)F]-fluoromaltose as a novel PET tracer for imaging bacterial infection. PloS one. 2014;9(9):e107951. pmid:25243851; PubMed Central PMCID: PMC4171493.
- 12. Auletta S, Varani M, Horvat R, Galli F, Signore A, Hess S. PET Radiopharmaceuticals for specific bacteria imaging: a systematic review. J Clin Med. 2019;8(2). Epub 2019/02/10. pmid:30736324; PubMed Central PMCID: PMC6406348.
- 13. Chen W, Dilsizian V. Molecular Imaging of cardiovascular device infection: targeting the bacteria or the host-pathogen immune response? Journal of Nuclear Medicine. 2020;61(3):319–26. Epub 2020/02/09. pmid:32034110.
- 14. Igarashi H, Kamisaka K, Maezawa H, Uesugi T. Degradation and metabolism of indocyanine green: high-pressure liquid chromatographic analysis. Bull Tokyo Med Dent Univ. 1990;37(4):59–69. Epub 1990/12/01. pmid:2091869.
- 15. Hilderbrand SA, Weissleder R. Near-infrared fluorescence: application to in vivo molecular imaging. Current Opinion in Chemical Biology. 2010;14(1):71–9. pmid:19879798.
- 16. Selton-Suty C, Celard M, Le Moing V, Doco-Lecompte T, Chirouze C, Iung B, et al. Preeminence of Staphylococcus aureus in infective endocarditis: a 1-year population-based survey. Clinical Infectious Diseases. 2012;54(9):1230–9. pmid:22492317.
- 17. Habib G, Hoen B, Tornos P, Thuny F, Prendergast B, Vilacosta I, et al. Guidelines on the prevention, diagnosis, and treatment of infective endocarditis (new version 2009): the Task Force on the Prevention, Diagnosis, and Treatment of Infective Endocarditis of the European Society of Cardiology (ESC). Endorsed by the European Society of Clinical Microbiology and Infectious Diseases (ESCMID) and the International Society of Chemotherapy (ISC) for Infection and Cancer. European Heart Journal. 2009;30(19):2369–413. pmid:19713420.
- 18. Baba T, Ara T, Hasegawa M, Takai Y, Okumura Y, Baba M, et al. Construction of Escherichia coli K-12 in-frame, single-gene knockout mutants: the Keio collection. Molecular Systems Biology. 2006;2:2006 0008. pmid:16738554; PubMed Central PMCID: PMC1681482.
- 19. Entenza JM, Caldelari I, Glauser MP, Moreillon P. Efficacy of levofloxacin in the treatment of experimental endocarditis caused by viridans group Streptococci. The Journal of Antimicrobial Chemotherapy. 1999;44(6):775–86. pmid:10590278.
- 20. Santoro J, Levison ME. Rat model of experimental endocarditis. Infection and Immunity. 1978;19(3):915–8. pmid:417032; PubMed Central PMCID: PMC422278.
- 21. Hanses F, Roux C, Dunman PM, Salzberger B, Lee JC. Staphylococcus aureus gene expression in a rat model of infective endocarditis. Genome Med. 2014;6(10):93. Epub 2014/11/14. pmid:25392717; PubMed Central PMCID: PMC4228149.
- 22. Hady WA, Bayer AS, Xiong YQ. Experimental endocarditis model of methicillin resistant Staphylococcus aureus (MRSA) in rat. Journal of Visualized Experiments: JoVE. 2012;(64):e3863. pmid:22711072; PubMed Central PMCID: PMC3471301.
- 23. Fogarasi M, Pullman J, Winnard P Jr., Hnatowich DJ, Rusckowski M. Pretargeting of bacterial endocarditis in rats with streptavidin and 111In-labeled biotin. Journal of Nuclear Medicine. 1999;40(3):484–90. pmid:10086715.
- 24. Lee H, Berezin MY, Henary M, Strekowski L, Achilefu S. Fluorescence lifetime properties of near-infrared cyanine dyes in relation to their structures. J Photochem Photobiol A Chem. 2008;200(2–3):438–44. Epub 2009/12/18. pmid:20016664; PubMed Central PMCID: PMC2630462.
- 25. Conceicao DS, Ferreira DP, Ferreira LF. Photochemistry and cytotoxicity evaluation of heptamethinecyanine near infrared (NIR) dyes. Int J Mol Sci. 2013;14(9):18557–71. pmid:24022690; PubMed Central PMCID: PMC3794795.
- 26. Saxena V, Sadoqi M, Shao J. Degradation kinetics of indocyanine green in aqueous solution. J Pharm Sci. 2003;92(10):2090–7. pmid:14502548.
- 27. Cherrick GR, Stein SW, Leevy CM, Davidson CS. Indocyanine green: observations on its physical properties, plasma decay, and hepatic extraction. J Clin Invest. 1960;39:592–600. Epub 1960/04/01. pmid:13809697; PubMed Central PMCID: PMC293343.
- 28. Terasawa M, Ishizawa T, Mise Y, Inoue Y, Ito H, Takahashi Y, et al. Applications of fusion-fluorescence imaging using indocyanine green in laparoscopic hepatectomy. Surgical endoscopy. 2017;31(12):5111–8. Epub 2017/04/30. pmid:28455774.
- 29. Vlek SL, van Dam DA, Rubinstein SM, de Lange-de Klerk ESM, Schoonmade LJ, Tuynman JB, et al. Biliary tract visualization using near-infrared imaging with indocyanine green during laparoscopic cholecystectomy: results of a systematic review. Surgical Endoscopy. 2017;31(7):2731–42. Epub 2016/11/16. pmid:27844236; PubMed Central PMCID: PMC5487840.
- 30. Morales-Conde S, Alarcon I, Yang T, Licardie E, Camacho V, Aguilar Del Castillo F, et al. Fluorescence angiography with indocyanine green (ICG) to evaluate anastomosis in colorectal surgery: where does it have more value? Surgical Endoscopy. 2019. Epub 2019/10/06. pmid:31586247.
- 31. Suami H, Heydon-White A, Mackie H, Czerniec S, Koelmeyer L, Boyages J. A new indocyanine green fluorescence lymphography protocol for identification of the lymphatic drainage pathway for patients with breast cancer-related lymphoedema. BMC Cancer. 2019;19(1):985. Epub 2019/10/24. pmid:31640623; PubMed Central PMCID: PMC6806520.