A reliable, selective and sensitive stability-indicating RP-HPLC assay was established for the quantitation of bromazepam (BMZ) and one of the degradant and stated potential impurities; 2-(2-amino-5-bromobenzoyl) pyridine (ABP). The assay was accomplished on a C18 column (250 mm × 4.6 mm i.d., 5 μm particle size), and utilizing methanol-water (70: 30, v/v) as the mobile phase, at a flow rate of 1.0 ml min-1. HPLC detection of elute was obtained by a photodiode array detector (DAD) which was set at 230 nm. ICH guidelines were adhered for validation of proposed method regarding specificity, sensitivity, precision, linearity, accuracy, system suitability and robustness. Calibration curves of BMZ and ABP were created in the range of 1–16 μg mL-1 with mean recovery percentage of 100.02 ± 1.245 and 99.74 ± 1.124, and detection limit of 0.20 μg mL-1 and 0.24 μg mL-1 respectively. BMZ stability was inspected under various ICH forced degradation conditions and it was found to be easily degraded in acidic and alkaline conditions. The results revealed the suitability of the described methodology for the quantitation of the impurity (ABP) in a BMZ pure sample. The determination of BMZ in pharmaceutical dosage forms was conducted with the described method and showed mean percentage recovery of 99.39 ± 1.401 and 98.72 ± 1.795 (n = 6), respectively. When comparing the described procedure to a reference HPLC method statistically, no significant differences between the two methods in regard to both accuracy and precision were found.
Citation: Darwish HW, Ali NA, Naguib IA, El Ghobashy MR, Al-Hossaini AM, Abdelrahman MM (2021) Development and validation of a stability indicating RP-HPLC-DAD method for the determination of bromazepam. PLoS ONE 16(3): e0244951. https://doi.org/10.1371/journal.pone.0244951
Editor: Girish Sailor, Bhagwan Mahvir College of Pharmacy, INDIA
Received: March 12, 2020; Accepted: December 19, 2020; Published: March 10, 2021
Copyright: © 2021 Darwish et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Data Availability: All relevant data are within the paper.
Funding: The authors would like to extend their sincere appreciation to the Deanship of Scientific Research at King Saud University for its funding of this research through the Research Group Project to HD (RGP-322), and to Taif University Researchers Supporting Project, Taif University, Taif, Saudi Arabia to IN (TURSP-2020/56).
Competing interests: The authors have declared that no competing interests exist.
Introduction
One of the extensively used derivatives of the 1,4-benzodiazepine series is the compound bromazepam (7-bromo-5-(pyridyl-2-yl)-1,3-dihydro-2H-1,4-benzodiazepin-2-one) [1]. BMZ is reported to have similar properties to those of diazepam. The compound was used in the short-term treatment for some disorders including anxiety disorders that occur alone or associated with insomnia [2]. Benzodiazepines are a class of compounds known to have sedative, hypnotic effects. Besides, benzodiazepines show muscle relaxant, anticonvulsant and amnesic properties relatively safe when compared with other sedative drugs [3]. The inhibitory action of benzodiazepines on the CNS is a result of interaction of these compounds to the gamma-aminobutyric acid (GABA) receptors within the brain [4].
Bromazepam is among one of the widely recommended benzodiazepine derivatives for treating insomnia and anxiety [5,6]. The British Pharmacopoeial (BP) assay of BMZ in bulk powder is by anhydrous titration with perchloric acid in acetic anhydride, using potentiometry for the detection of the endpoint [7].
For the detection and determination of benzodiazepines in pharmaceutical formulations and related biological matrices, various analytical techniques have been reported [8–10]. Many publications can easily be found for BMZ analysis to study and measure therapeutic or toxic blood levels of the drug in serum [11,12]. These reported methods also reveal many analytical techniques for BMZ determination either alone or together with other benzodiazepines, in both pharmaceutical or biological matrices. These techniques include HPLC [13–24], LC-MS [25,26], GC-MS [27], TLC [28], capillary electrophoresis [29], spectrophotometry [30–32], ion-selective electrode [33–35], potentiometric [36] and voltammetric [37,38] methods.
Similar to almost all of the benzodiazepine derivatives, BMZ can be subjected to hydrolysis in acidic aqueous solutions forming degradation products. Due to the widespread use of BMZ, a study of the kinetics and mechanism of its hydrolysis is considered a critical issue. Over the past decades, several publications showed an interest in stability studies of BMZ, especially degradation of the drug due to hydrolysis, which cause the formation of the 2-(2-amino-5-bromobenzoyl) pyridine (ABP), Fig 1 [18,39,40]. Beside the BP [7] defines 2-(2-amino-5-bromobenzoyl) pyridine (ABP) as a possible impurity in pure bromazepam powder, the main metabolic pathway of bromazepam includes C3 hydroxylation and heterocyclic ring hydrolysis producing two metabolites 3-hydroxy bromazepam and ABP metabolites that excreted in the urine [38,41,42].
The main interest of presented paper is to evaluate BMZ stability in line with the applicable guidelines of ICH utilizing a full validated stability-indicating method after subjecting the investigated drug to a wide range of stress conditions [43,44]. Moreover, the development of a validated stability-indicating HPLC method for BMZ quantitation in raw material and tablets is the second goal. It is worth mentioning that recently, our research team published stability-indicating spectrophotometric methods for assaying BMZ in addaition to its degradant [45]. But the previous work was suitable only for resolving such a binary mixture of BMZ and its degradation product. Consequently, were directed this work for developing RP-HPLC with Photo-Diode Array detection because of its well known higher selectivity when compared to spectrophotometric methods. Additionally, The described HPLC methodology is favorably applied in quality control laboratories.
This method isn’t only for BMZ quantitation in the existence of its degradant ABP, which is considered one of its potential impurities [7] and its main benzophenone metabolite, but also for ABP determination with excellent accurateness even though the existence of great similarity in their chemical structures, Fig 1.
Experimental
Instrument
The chromatographic assay was established on an “Agilent 1260 Infinity HPLC system (Agilent Technologies, Germany)” that was fitted with an “Agilent 1260 Infinity preparative pump (G1361A)”, temperature of column was kept at 30°C by an “Agilent 1260 Infinity Thermostated column compartment (G1316A)”. Samples were introduced using an “Agilent 1260 Infinity preparative Autosampler (G2260A)”. The HPLC instrument was connected to an Agilent 1260 Infinity Diode array detector VL (G131SD). A reverse-phase “ZORBAX Hypersil BDS” C18 column 250 mm × 4.6 mm i.d. and 5 μm particle size (ThermoElectron Corporation, USA) was used. LC-solution software was utilized for recording and analyzing the chromatograms. For pH adjustment and sonication, “Jenway 3505 pH meter (Staffordshire, UK)” and “Sonix TV ss-series ultrasonicator (USA)”, were utilized respectively.
The IR spectra were determined on “Shimadzu IR 435 spectrophotometer (Shimadzu Corp., Kyoto, Japan)”. The mass determination was done on “Triple quadrupoles mass spectrometer” with API source, Agilent 6400, MassHunter software operated by Pentium 3 (40 MHz) processor (Hp, USA).
Material and reagents
Bromazepam standard (99.91% purity) was generously provided by the “Egyptian International Pharmaceutical Industries Company (Cairo, Egypt)”. Lexotanil® tablets (3 mg batch No. M1139B01) produced by “La Roche S.p.A. Milan, Italy”, and Calmepam® tablets (batch No. A506716) produced by “GlaxoSmithKline SAE. (Cairo, Egypt)”, were procured from the Egyptian market. HPLC grade methanol was purchased from Sigma-Aldrich Chemical (Germany). Analytical grade chloroform, sodium hydroxide, acetone, 30% hydrogen peroxide and hydrochloric acid solutions were all purchased by “El-Nasr Pharmaceutical Chemicals Co. (Cairo, Egypt)”. Water for injection B.P. 2003 was received from Egypt “Otsuka Pharmaceutical Co. (10th of Ramadan city, Egypt)”. Solvents were filtered by a “Sartorius Stedium Biotec. GmbH. membrane filters (0.45 μm) (Goettingen, Germany)”. Cellulose acetate syringe filters with 0.45 μm pore size purchased from “Gemma Medical (Barcelona, Spain)”.
Degradation studies
Different forced degradation studies were performed on BMZ including acidic, alkaline and oxidative degradation studies. Literature [46] stated that BMZ is stable in light, so light-degradation studies were not performed in our work.
Acidic hydrolyis and preparation of degradant (ABP).
As stated previously [18], the production of the degradation product (ABP) by acid hydrolysis was achieved by dissolving 1.00 gm of pure BMZ in the least volume of methanol. The second step was the addition of 50 mL of 1N HCL that were all refluxed and protected from light for 3 h. The reaction was followed using TLC, with a developing system that contained a 4:1 chloroform-acetone solution mixture, to monitor the full disappreance of BMZ spot and emergence of a new spot of ABP. After complete degradation, the solution was neutralized by dropwise adding 10 N NaOH solution until the production of a yellow precipitate of ABP, which was kept in the refrigerator overnight. The ABP precipitate was then filtered and washed with 0.01 N HCL to dissolve any BMZ still not degraded. ABP was then washed with distilled water several times then dried in oven at 70°C. Confirmation was acquired by subjecting both BMZ and ABP to IR and Mass spectrometry.
Alkaline degradation.
A weighted amount of pure BMZ (0.50 gm) was to be dissolved in the least volume of methanol, then refluxed away from light with 50 mL of 1N NaOH solution for 3 hours. The alkaline degradation reaction was monitored via TLC, similar to acidic degradation procedure. Once again, the same degradation product (ABP) was formed. However, to a lesser amount than what was produced by the acidic degradation method.
Oxidative degradation.
Compare to both the acid and alkaline degradation methods above, a very limited quantity of ABP was produced by oxidative degradation of BMZ. This was achieved by putting 0.50 gm of BMZ in a 10 mL conical flask containing 30% w/v hydrogen peroxide. Then the mixture was located over a water bath (thermostatically controlled) at temperature of 80°C for 12 h. The oxidative degradation reaction was monitored by runningTLC as before.
Procedure
In our study, chromatographic separations were effected using an isocratic mode with a C18 column. The mobile phase composed of (70:30 v/v) mixture of methanol (phase A) and water (phase B), at a constant flow rate of 1 mL/min. The eluate was scanned at a set wavelength of 230 nm at RT. Each injection run was replicated three times, with an injection volume of 20 μL. Total run time of each sample injection was approximately 6 min, and quantification of the components under invesigationwas achieved using the total peak areas of the investigated components.
Linearity and calibration curves’ construction
Accurate aliquots equal to 10–160 μg of both BMZ and its degradant ABP have been transferred from their respective working standard solutions (100 μg mL-1) to different sets of 10 mL measuring flasks and then diluted with mobile phase to the mark. For each concentration triplicate injections were performed. The relative peak areas were determined by dividing the peak areas of each component at the corresponding peak area at 8 μg mL-1 as an external standard. Thereby each component’s calibration curve was constructed by using its relative peak areas, and later regression equation was created.
Application to the pharmaceutical formulation
A fine powder has been made by first weighing and then grinding 30 Lexotanil® and Calmepam® tablets individually. Precisely weighed portion containing 50 mg of BMZ has been transferred into separate two volumetric flasks, each containing 30 mL of methanol. The solution mixtures were then sonicated for 30 min, cooled and then completed with methanol to a final volume of 50 mL. The prepared solutions were filtered and then further diluted with the HPLC mobile phase to reach a standard solution concentration of 100 μg mL-1. The detailed procedure under “linearity and calibration curve construction” was preceded and BMZ concentrations were determined by applying the regression equation, and the recovery % were computed. For implementing the standard addition step, the fine powdered Lexotanil® and Calmepam® tablets and pure BMZ were separately well mixed together before continuing the procedure referred to above.
Results and discussion
Structure elucidation of BMZ degradant (ABP)
The degradation pathway of BMZ was reported in the literature [18]. BMZ was hydrolised in acidic condition by refluxing with 1 N HCL solution for 3 h giving its degradant, due to cleavege of the 4,5-azomethine bond, followed by breakage of the 1,2-amide bond to give the benzophenone derivative; “2-(2-amino-5-bromobenzoyl) pyridine” (ABP) in addition to glycine Fig 1. On the other hand, when BMZ was exposed to hydrolysis under alkaline-stress condition, by mixing BMZ with 1N NaOH for 3 h under reflux, or by oxidative degradation of BMZ when added to an aqueous solution of 30% H2O2 for 3 h, this lead to the same degradation product (ABP) being produced. However, at a lower amount than what was produced by acidic hydrolysis. The degradant was separated as mentioned before and well characterized by IR and MS spectrometry [45].
HPLC development and optimization
Several parameters that can affect chromatographic separation were evaluated and optimized. these parameters included scanning of different wavelengths, experimenting with various types of the aqueous phase, changes of the mobile phase pH, different types and ratio of organic modifier added and flow rate.
Optimal scanning wavelength.
The ultraviolet spectra of BMZ and ABP at concentrations of 10 μg mL-1 each, was found to show a maximum absorption at a wavelength of 232 nm and 236 nm for BMZ and ABP respectively. Therefore, 230 nm was of choice after comparing the two spectra, to offer the highest sensitivity with minimal detected noise.
Type of the aqueous phase.
Various aqueous phases (water, 0.1% glacial acetic acid, 0.5% phosphoric acid and 0.01% triethylamine) were evaluated in combination with an organic modifier. Water adjusted with phosphoric acid and/or glacial acetic acid was found to show no change on the retention time or peak symmetry of both BMZ and ABP. Also, the addition of 0.01% triethylamine to water gave poor baseline. Thereby, water was selected, having the added advantage of the low cost, having minimal effect on the column, and at the same time gave a better result.
Type and ratio of organic modifier.
Various organic modifiers that included acetonitrile and methanol were evaluated in an attempt to increase the performance of the chromatographic conditions. It was found that by using methanol to the mobile phase, the resolution of components and peak symmetry was noticeably improved. The ratio of methanol also was studied (60–75%, v/v). The retention times of both BMZ and ABP were shortened as the ratio of methanol was increased. We found that concentrations of methanol of 70% (v/v) resulted in the optimal separation of BMZ and ABP with the highest validation parameters, Fig 2. However, a further increase in the amount of methanol above 75% (v/v) resulted in poor resolution and overlapped BMZ and ABP peaks.
Mobile phase: Methanol: Water (70:30 v/v). Detection: DAD at 230 nm.
HPLC results
The developed isocratic HPLC method for analysis of BMZ and ABP showed to be sensitive, accurate and highly selective. A mobile phase composed of methanol: water (70:30 v/v) was used, with retention times of 4.397, and 5.317 min for BMZ and ABP respectively, Fig 2. The total run time required for the analyses of all the compounds of interest was determined to be approximately 6 min. The required short analysis time may have the added advantage of providing a more rapid method for routine analysis of the main investigated drug, with a great level of accuratness and selectivity when compared to previously reported methods [18,19]. Calibration curves for BMZ and ABP were created by drawing the relative peak area against their correspondant concentration. The regression equations were then calculated as follow:
Where Y is known as relative peak area, while X is the concentration in units of μg mL-1, and r corresponds to the correlation coefficient, Table 1. This proposed method was further applied for the assay of BMZ in Lexotanil® and Calmepam® tablets (labeled to contain 3 mg of BMZ per tablet). For confirmation of the method validity, standard addition technique was applied, Table 2. The results obtained show no interference of additives (included in the dosage forms) with the investigated mixture.
Method validation
The method’s validation was achieved following the guidelines of the ICH [47].
Linearity.
The method’s linearity was evaluated after the optimized chromatographic conditions were achieved. This was done by collecting the area of the integrated peak of each compound at various concentrations followed by drawing the calibration graphs utilizing the relative peak area of each compound against the compounds corresponding concentrations from which regression equations were constructed. The linearity of the calibration graphs was confirmed by the high value of the correlation coefficient and the low value of the intercept. Additionally, the small values of the residual standard error confirm the fitness and linearity of the proposed method. Table 1 shows the linearity and range parameters.
Accuracy.
The percentage recoveries of pure blind samples of the studied compound were used to compute the accuracy of analytical method. In Table 3, corresponding regression equations were used to calculate relevant concentrations of compounds. Mean recoveries were almost 100% with standard deviation of less than 1.2% in all cases.
A standard addition technique was also performed to both Lexotanil® and Calmepam® tablets, by addition of certain amounts of authentic BMZ, to ensure accuracy. Here we obtained good recoveries (about 99%) and small standard deviation values (less than 2.3% in all cases) suggesting good accuracy and also revealing minimal interference from any excipients within the dosage form; Table 2.
Precision.
Repeatability of the results for three different concentrations (2, 6 and 10 μg mL-1) of both BMZ and ABP were performed through replicate analysis (n = 3), within the same day, to estimate intra-day variation. In order to estimate any inter-day variation, triplicate injections were performed on three consecutive days. Table 1 shows the calculated coefficient of variation (< .2.1% in all cases) at the selected concentration levels.
Specificity.
The specificity of the proposed method was manifested through a sufficiently good separation of the two compounds, BMZ and ABP at the retention times of 4.397, and 5.317 min, respectively, Fig 2.
LOD and LOQ.
Limits of both detection and quantitation (LOD and LOQ) values for BMZ and ABP by the HPLC method are determined via calculations [48]. The LOD values were equal to 0.2 and 0.24 μg mL-1 for BMZ and ABP respectively. The LOQs were also found to be of low values (equal to 0.6 and 0.72 μg mL-1 for BMZ and ABP respectively). These low values for LOD and LOQ thereby demonstrating the excellent sensitivity of the suggested chromatographic method; Table 1.
Robustness.
The capability to remain unaffected by minor intentional changes in some parameters of a chromatographic method is frequently used to assess the robustness of the analytical procedure, and indicates its its efficiency during regular use [47]. For the proposed methodology, its robustness was checked, by introducing small changes in the HPLC method. These changes include minor alteration of the flow rate (± 0.2 mL min-1), organic strength (± 2%) and performing the HPLC method by different analysts. The values of the relative standard deviation of recovery (%RSD) are found in Table 4. The small values of RSD % (< 2%) indicated reliable results concerning the area under the curve.
System suitability.
Tests for system suitability were carried out, aiming to check the suitability of the chromatographic system for analysis and its reproducibility. This was conducted by calculating various parameters [49] such as resolution, capacity factor, and peak symmetry, where acceptable results are shown in Table 5.
Statistical comparison to a reference method
The obtained results by the established HPLC method were compared statistically to the reference HPLC method [15] for Lexotanil® and Calmepam® tablets, where both the Student’s t-test and the variance ratio F-test were used at a 95% confidence level. It was clear from Table 6 that calculated t-test and F-test values are less than that of the tabulated ones. This may indicate that there is no significant difference, regarding both accuracy and precision, between the proposed HPLC method and the reference method.
The proposed stability-indicating RP-HPLC provided the advantage of using a Photo-Diode array detector that is sensitive and accurate due to its ability to provide additional spectral information that is required to identify eluting peaks [50]. The developed HPLC method was depended on isocratic elution and a reversed phase column utilizing an eluent without any electrolyte buffer. As the added buffer solutions to HPLC mobile phase were mostly reported to create several problems, including damage to the pump seals as well as crystal formation in the detector cell and connecting tubing [21].
Conclusion
The presented isocratic HPLC-DAD method may help in providing high selective, sensitive and reproducible quantitative stability-indicating method, for the analysis of BMZ and ABP simultaneously, at a single wavelength and within a short analysis time. The simple mobile phase composition, compared to previously published methods, with high resolution of both components, adds the advantages of saving time, cost, effort, and protection of the column. Moreover, the method shows low LOD and LOQ values and as a result, can be applied for the detecting low concentrations of BMZ and ABP.
The obtained results indicated that the suggested HPLC method can be included in some of the highly selective and sensitive methods reported for the BMZ and ABP analysis. These merits suggested the usage of the suggested method in analytical quality control (QC) that are routinely performed by regulatory agencies and QC laboratory, without the interference of some of the commonly used dosage form additives.
References
- 1.
Index Merck, Merck Research Laboratories Division of MERCK & CO. thirteenth ed. Inc, Whitehouse Station: NJ, USA 2001.
- 2.
Sweetman S.C., Martindale: The Complete Drug Reference, 36th ed. London: The Pharmaceutical Press 1039. 2009.
- 3.
Laurence, D., Bennett, P., Clinical Pharmacology, fifth ed., 1980, 274–275.
- 4.
Williams D.A., Lemke T.L. in: Foye’s Principle of Medicinal Chemistry, 6th ed., Lippincott williams & wilkins, Ny, 2008. https://doi.org/10.2174/157340608786242089 pmid:18991733
- 5.
Trevor A.J., Katzung B.G., Masters S.B., in: Katzung and Trevor’s Pharmacology, 8th 351 ed., Mc Graw-Hill, Boston, 2007.
- 6. Rang HP, Ritter JM, Flower RJ, Henderson G. Rang & Dale’s Pharmacology E-Book: Elsevier Health Sciences; 2007.
- 7.
British Pharmacopoeia, the British Pharmacopoeia Secretariat of the Medicines and Healthcare Products Regulatory Agency, UK, 2009.
- 8.
Schütz H. Benzodiazepines II: A Handbook. Basic Data, Analytical Methods, Pharmacokinetics, and Comprehensive Literature: Springer Science & Business Media; 1982.
- 9. Sioufi A, Dubois J. Chromatography of benzodiazepines. J Chromatogr B. 1990;531:459–80. pmid:1979564
- 10. Berrueta L, Gallo B, Vicente F. Biopharmacological data and high-performance liquid chromatographic analysis of 1, 4-benzodiazepines in biological fluids: a review. J pharm biomed anal. 1992;10(2–3):109–36. pmid:1356446
- 11. Michaud K, Romain N, Giroud C, Brandt C, Mangin P. Hypothermia and undressing associated with non-fatal bromazepam intoxication. Forensic sci int.2001;124(2–3):112–4. pmid:11792498
- 12. Tanaka E, Terada M, Misawa S, Wakasugi C. Simultaneous determination of twelve benzodiazepines in human serum using a new reversed-phase chromatographic column on a 2-μm porous microspherical silica gel. J Chromatogr B 1996;682(1):173–8.
- 13. Samanidou VF, Pechlivanidou AP, Papadoyannis IN. Development of a validated HPLC method for the determination of four 1, 4‐benzodiazepines in human biological fluids. Journal of separation science. 2007;30(5):679–87. pmid:17461106
- 14. Gonçalves JCS, Monteiro TM, de Miranda Neves CS, da Silva Gram KR, Volpato NM, Silva VA, et al. On-line solid-phase extraction coupled with high-performance liquid chromatography and tandem mass spectrometry (SPE-HPLC-MS-MS) for quantification of bromazepam in human plasma: an automated method for bioequivalence studies. Ther Drug Monit. 2005;27(5):601–7. pmid:16175133
- 15. Al-Hawasli H, al-Khayat MA, al-Mardini MA. Development of a validated HPLC method for the separation and analysis of a Bromazepam, Medazepam and Midazolam mixture. J. pharm anal. 2012;2(6):484–91. pmid:29403788
- 16. Podilsky G, Berger‐Gryllaki M, Testa B, Pannatier A. Development and validation of an HPLC method for the simultaneous monitoring of bromazepam and omeprazole. J Liq Chromatogr Relat Technol 2008;31(6):878–90.
- 17. Laurito TL, Mendes GD, Santagada V, Caliendo G, de Moraes MEA, De Nucci G. Bromazepam determination in human plasma by high‐performance liquid chromatography coupled to tandem mass spectrometry: a highly sensitive and specific tool for bioequivalence studies. J mass spectrum.2004;39(2):168–76. pmid:14991686
- 18. Panderi I, Archontaki H, Gikas E, Parissi-Poulou M. Acidic hydrolysis of bromazepam studied by high performance liquid chromatography. Isolation and identification of its degradation products. J Pharm Biomed Anal.1998;17(2):327–35. pmid:9638586
- 19. Hirayama H, Kasuya Y, Suga T. High-performance liquid chromatographic determination of bromazepam in human plasma. J Chromatogr B. 1983;277:414–8. pmid:6139385
- 20. Musshoff F, Daldrup T. A rapid solid-phase extraction and HPLC/DAD procedure for the simultaneous determination and quantification of different benzodiazepines in serum, blood and post-mortem blood. Int J Legal Med. 1992;105(2):105–9. pmid:1520633
- 21. He W, Parissis N, Kiratzidis T. Determination of benzodiazepines in forensic samples by HPLC with photodiode array detection. J Forensic Sci. 1998;43(5):1061–7. pmid:9729826
- 22. Boukhabza A, Lugnier AAJ, Kintz P, Tracqui A, Mangin P, Chaumont AJ. High-performance liquid chromatographic determination of bromazepam in human plasma. Analyst. 1989;114(5):639–41. pmid:2569276
- 23. Andraus MH, Wong A, Silva OA, Wada CY, Toffleto O, Azevedo CP, et al. Determination of bromazepam in human plasma by high‐performance liquid chromatography with electrospray ionization tandem mass spectrometric detection: application to a bioequivalence study. J mass spectrom. 2004;39(11):1348–55. pmid:15532067
- 24. Uddin MN, Samanidou VF, Papadoyannis IN. Development and validation of an HPLC method for the determination of six 1, 4‐benzodiazepines in pharmaceuticals and human biological fluids. J Liq Chromatogr Relat Technol. 2008;31(9):1258–82.
- 25. Chèze M, Villain M, Pépin G. Determination of bromazepam, clonazepam and metabolites after a single intake in urine and hair by LC-MS/MS: Application to forensic cases of drug facilitated crimes. Forensic Sci Int. 2004;145(2–3):123–30. pmid:15451083
- 26.
Moore C, Coulter C, Crompton K, Zumwalt M. Determination of Benzodiazepines in Urine and Blood Using Rapid Resolution Liquid Chromatography/Triple Quadrupole Mass Spectrometry. Agilent Technologies, Immunalysis Corporation, Pomona, CA. 2007:5989–7201.
- 27. Aebi B, Sturny-Jungo R, Bernhard W, Blanke R, Hirsch R. Quantitation using GC–TOF-MS: example of bromazepam. Forensic science international. 2002;128(1–2):84–9. pmid:12208027
- 28. Haefelfinger P. Determination of bromazepam in plasma by quantitative thin-layer chromatography. Chromatographia. 1978;11(1):10–3.
- 29. Hansen SH, Sheribah ZA. Comparison of CZE, MEKC, MEEKC and non-aqueous capillary electrophoresis for the determination of impurities in bromazepam. J Pharm Biomed Anal. 2005;39(1–2):322–7. pmid:15935599
- 30. Salem A, Barsoum B, Izake E. Spectrophotometric and fluorimetric determination of diazepam, bromazepam and clonazepam in pharmaceutical and urine samples. Spectrochim Acta A Mol Biomol Spectrosc. 2004;60(4):771–80. pmid:15036087
- 31. Attia M. Spectrofluorimetric quantification of bromazepam using a highly selective optical probe based on Eu3+–bromazepam complex in pharmaceutical and serum samples. Spectrochim Acta A Mol Biomol Spectrosc.2009;74(4):972–6. pmid:19781983
- 32. Salem A, Barsoum B, Izake E. Determination of bromazepam and clonazepam in pure and pharmaceutical dosage forms using chloranil as a charge transfer complexing agent. Anal letters. 2002;35(10):1631–48.
- 33. Salem A, Barsoum B, Izake E. Potentiometric determination of diazepam, bromazepam and clonazepam using solid contact ion-selective electrodes. Anal chim acta. 2003;498(1–2):79–91.
- 34. Al Attas AS. Construction and analytical application of ion selective bromazepam sensor. Int J Electrochem Sci. 2009;4:20–9.
- 35. Ali NA, Abdelrahman MM, Naguib IA, El Ghobashy MR. Development of Membrane Electrode for the Selective determination of Bromazepam in Tablets and Plasma, Anal Bioanal Electrochem. 2015; 7: 242–253.
- 36. Salem AEA, Barsoum BN, Saad GR, Izake EL. Potentiometric determination of some 1, 4-benzodiazepines in pharmaceutical preparations and biological samples. J Electroanal Chem. 2002;536(1–2):1–9.
- 37. Hernández L, Zapardiel A, López JAP, Bermejo E. Determination of camazepam and bromazepam in human serum by adsorptive stripping voltammetry. Analyst. 1987;112(8):1149–53. pmid:2889399
- 38. Valdeón JL, Escribano MTS, Hernandez LH. Determination of bromazepam and its urinary metabolites, with a previous hydrolysis reaction, by voltammetric and spectrophotometric techniques. Analyst. 1987;112(10):1365–8. pmid:2893562
- 39. De Bruyne M, Sinnema A, Verweij A. Hydrochloric acid hydrolysis of bromazepam. Identification and relevance of some additional products. Pharm Weekbl. 1982;4(1):12–5. pmid:6122197
- 40. El-Khateeb S, Amer S, Abdel-Razek S, Amer M. Stability-Indicating Method For The Determination of Bromazepam And Delorazepam Via Proton Magnetic Resonance Spectroscopy. Spectrosc lett. 1997;30(5):915–32.
- 41. Sawada H, Hara A. Studies on metabolism of bromazepam. VI. Reduction of 2-(2-amino-5-bromobenzoyl) pyridine, a metabolite of bromazepam, in the rabbit, rat, and guinea pig. Drug Metab Dispos. 1978;6(3):205–12. pmid:26538
- 42. Schwartz MA, Postma E, Kolis SJ, Leon AS. Metabolites of bromazepam, a benzodiazepine, in the human, dog, rat, and mouse. J pharm sci. 1973;62(11):1776–9. pmid:4758074
- 43.
ICH. Harmonised tripartite guideline, stability testing of new drug substances and products Q1A (R2), USA, 2003. http://www.ich.org/fileadmin/Public_Web_Site/ICH_Products/Guidelines/Quality/Q1A_R2/Step4/Q1A_R2__Guideline.pdf (accessed 20 December 2018).
- 44.
ICH. Harmonised tripartite guideline, stability testing: photostability testing of new drug substances and products Q1B, USA, 1996. http://www.ich.org/fileadmin/Public_Web_Site/ICH_Products/Guidelines/Quality/Q1B/Step4/Q1B_Guideline.pdf (accessed 20 December 2018).
- 45. Darwish HW, Ali NA, Abdelrahman MM, Naguib IA, El Ghobashy MR, Al-Hossaini AM. Stability Indicating Spectrophotometric Methods for Quantitative Determination of Bromazepam and its Degradation Product, Spectrochim Acta A Mol Biomol Spectrosc. 2020;238: 118433–11841. pmid:32403074
- 46. Le Solleu H, Demotes-Mainard F, Vincon G, Bannwarth B. The determination of bromazepam in plasma by reversed-phase high-performance liquid chromatography. J pharm biomed anal. 1993;11(8):771–5. pmid:8257743
- 47.
ICH, Q2 (R1). (2005). Validation of Analytical Procedures: Text and Methodology, ICH Harmonized Tripartite Guideline.