Rapid declines in systolic blood pressure are associated with an increase in pulse transit time

  • Sebastian Grøvdal Schaanning,

    Roles Conceptualization, Data curation, Formal analysis, Investigation, Methodology, Software, Validation, Visualization, Writing – original draft, Writing – review & editing

    Affiliation Department of Circulation and Medical Imaging, Norwegian University of Science and Technology, Trondheim, Norway

  • Nils Kristian Skjaervold

    Roles Conceptualization, Funding acquisition, Investigation, Methodology, Project administration, Resources, Supervision, Validation, Writing – original draft, Writing – review & editing

    nils.k.skjervold@ntnu.no

    Affiliations Department of Circulation and Medical Imaging, Norwegian University of Science and Technology, Trondheim, Norway, Department of Anaesthesia and Intensive Care Medicine, Trondheim University Hospital, Trondheim, Norway

Abstract

Background

The correlation between pulse transit time and blood pressure has been proposed as a route to measure continuous non-invasive blood pressure. We investigated whether pulse transit time trends could model blood pressure trends during episodes of rapid declines in blood pressure.

Methods

From the Medical Information Mart for Intensive Care waveform database we identified substantial blood pressure reductions. Pulse transit time was calculated from the R-peak of the electrocardiogram to the peak of the arterial pulse waveform. The time-series were processed with a moving average filter before comparison. Averaged, continuous heart rate was also analysed as a control. The intra-individual association between variables was assessed per subject using linear regression.

Results

In the 511 patients included we found a median correlation coefficient between blood pressure and pulse transit time of -0.93 (IQR -0.98 to -0.76) with regression slopes of -1.23 mmHg/ms (IQR -1.73 to -0.81). The median correlation coefficient between blood pressure and heart rate was 0.46 (IQR -0.16 to 0.83). In supplementary analysis, results did not differ substantially when widening inclusion criteria, but the results were not always consistent within subjects across episodes of hypotension.

Conclusions

In a large cohort of critically ill patients experiencing episodes of rapid declines in systolic blood pressure, there was a moderate-strong intra-individual correlation between averaged systolic blood pressure and averaged pulse transit time. Our findings encourage further investigation into using the pulse transit time for non-invasive real-time detection of hypotension.

Introduction

Blood Pressure (BP) is an essential vital parameter that is monitored in most if not all hospitalized patients. BP measurements guide inpatient treatment of hypertension, act as a marker of hemodynamic status and severity of illness and can be used for cardiovascular risk stratification. The detection of acute hypotension is of particular interest, as this may signify the onset of circulatory shock, a feared clinical entity that requires prompt hemodynamic support and diagnostic workup [1]. Continuous BP-monitoring, however, is currently reserved for patients in whom the invasive insertion of an arterial line can be justified. In practice, this means that for most hospitalized patients, BP is monitored using intermittent cuff-based measurements. The interval between measurements may be several hours, and a recent publication found that a substantial proportion of significant BP-perturbations may go undetected with standard monitoring [2]. This implies that continuous non-invasive BP-monitoring (cNIBP) could benefit patients who are not candidates for arterial line placement.

Various techniques for cNIBP have been proposed, and some integrated systems have been developed. Perhaps most known is the vascular-unloading technique, originally proposed by Marey, and later by Shirer and Penaz [35], and employed by Finapres® devices. In this method, a finger cuff is combined with a measure of finger blood volume using photoplethysmography (PPG). By varying the finger cuff pressure to keep the PPG signal constant, a pressure wave can be obtained that approximates the arterial pressure waveform [6]. Another technique for cNIBP is arterial tonometry, which aims to reproduce the arterial waveform based on the transcutaneous displacement produced by a pulsating artery. The method employs an external pressure transducer placed over a superficial artery, typically the radial artery [7]. A third technique which has been subject to much research is utilization of pulse wave velocity (PWV) and the related distance-invariant pulse transit time (PTT). This concept can be rooted in two physio-mathematical relationships, namely the Moens-Korteweg equation [8] and the relationship between elastic modulus and pressure, as empirically demonstrated by Hughes et al. [9], among others. In summary, the combination of these models entails that for an elastic tube the PWV is proportional to the elastic modulus of the tube wall, which in turn is proportional to the pressure within the tube. A further elaboration on these relationships is outside the scope of this article.

PWV and PTT, have been studied extensively for their relationship to BP, dating back to the 1920s. In an oft-cited work from 1976, Gribbin and colleagues demonstrated excellent intra-subject correlation between PWV and transmural arterial pressure, in a setup with external pressure manipulation [10]. Several papers on the relationship between PTT and BP followed. The PTT was dissected into the pre-ejection period (PEP) and the vascular transit time (VTT), with the VTT corresponding to the PWV measured by Gribbin. Researchers generally found an inverse correlation between PTT and BP, but there were conflicting results regarding the contribution of the PEP and the VTT to this relationship [11, 12]. With increased focus on patient monitoring, interest in the subject has thrived again in the last decades. Several researchers have investigated PTT for BP-estimation, typically measuring PTT from the electrocardiogram (ECG) R-peak to the PPG signal at the finger. A review of these efforts was recently published by Ding and Zhang [13]. Researchers have also employed machine learning methods, such as deep neural networks, to find additional PPG-features that may complement or replace PTT for BP-estimation [14].

A popular source of data for PTT-research is the publicly available Medical Information Mart for Intensive Care (MIMIC) database, a collection of waveform data from thousands of patients admitted for intensive care [15, 16]. A summary of studies using MIMIC data to evaluate PTT for BP-estimation was recently presented by Liang et al., along with their own original research on the database [17]. In short, a negative correlation between PTT and BP is reported by all authors, but patient selection and mode of analysis differs substantially across studies. Liang, like others, noted issues with waveform synchronicity, limiting the suitability of the MIMIC database for analysis of PTT. One major issue is related to the sampling of the PPG-signal, which is usually employed as the endpoint for PTT calculation. Analysing MIMIC data, Zhu et al. noted that the PPG-derived PTT is afflicted by a sawtooth-shaped artefact, with a period of approximately 100 seconds [18]. A similar sawtooth artefact was later demonstrated in the PPG-derived PTT by another group of researchers using a separate database [19]. Bennis et al. observed the same artefact in data from a neonatal ICU, and consequently used simulated data to show that the artefact could be attributed to post-processing of the PPG signal in the Masimo module of the Philips patient monitor [20]. This artefact substantially limits the utility of using previously collected PPG-data to calculate PTT.

While there are many publications on the relationship between PTT and BP, it is our experience that these fall into one of two categories: (1) Small scale experimental studies in healthy volunteers or (2) Studies on database materials with unclear criteria for waveform selection and lack of reporting on the clinical characteristics of the subjects. Most investigations have included a low to moderate number of subjects. In our opinion, the primary potential of cNIBP lies in the early detection of hemodynamic instability, i.e. hypotension, in hospitalized patients without an arterial line. To our knowledge, no one has systematically investigated the relationship between PTT and BP during the onset of hypotension in hospitalized patients. We hypothesize that a moderately rapid (≤ 15 minutes) reduction in SBP from ≥120 mmHg to ≤ 90 mmHg is accompanied by a predictable change in PTT. Furthermore, where reference data from an arterial line is available, most investigators have tested the ability of PTT to model BP on a beat-to-beat basis. While cNIBP of such resolution may seem enticing, one could argue it may be counterproductive to attempt to capture beat-to-beat fluctuations in BP. Firstly, this places great demands on the precision of the model. Secondly, we believe it is far more clinically relevant to capture short term trends in BP than beat-to-beat fluctuations. For this reason, we chose to perform the primary analysis on time-averaged data to assess short-term BP-trends rather than beat-to-beat fluctuations. Due to issues associated with the PPG-signal in the MIMIC database, we chose to use the Arterial Blood Pressure (ABP) waveform as the endpoint for calculation of PTT.

Materials and methods

All data preparation and analysis was performed using the Python Programming Language, version 3.7 [21].

Ethics approval and consent to participate

All data was sourced from the freely accessible MIMIC-III waveform database and the associated clinical database. The following information is provided by representatives for the MIMIC-III project: “The project was approved by the Institutional Review Boards of Beth Israel Deaconess Medical Centre (Boston, MA) and the Massachusetts Institute of Technology (Cambridge, MA). Requirement for individual patient consent was waived because the project did not impact clinical care and all protected health information was deidentified” [15].

Subject selection

Data was obtained from the MIMIC-III waveform database, matched subset. The matched subset includes 10,282 subjects for whom waveforms have been matched to clinical records. All subjects in the matched subset were eligible for inclusion. For each subject, several waveform records exist, sometimes exceeding 1000. The records are principally continuous with one another, but were split at time points where signal modalities and/or signal gain were changed. For the purpose of this study, an automated script selected the longest record per subject, for which both ECG and ABP signals were available, for further analysis.

Event identification and inclusion

An event was defined as an instance where the averaged systolic blood pressure (SBPAV) decreased to ≤ 90 mmHg within 15 minutes of SBPAV exceeding or equaling 120 mmHg. For each record, a scoping algorithm identified candidate segments based on the event definition. Each candidate segment was then evaluated by a series of algorithms to determine if the signal quality was sufficient to determine all peaks needed for parameter extraction. A segment was excluded if one or more of the following features were present:

  1. Max SBPAV<120 mmHg or nadir SBPAV> 90 mmHg
  2. Segment length < 21 heartbeats
  3. ECG signal quality index (ESQI) <0.9*
  4. Presence of undetectable/ambiguous peaks

*A score from 0–1, primarily based on intercorrelation between QRS-complexes.

One segment was selected from each subject whose record contained one or more eligible segments. For subjects who had multiple eligible segments, the segment with the highest ESQI was included for primary analysis, and the segment with the second highest ESQI was retained for supplementary analysis.

Variable extraction from viable segments

Peak detection for ECG signals was performed using the BioSPPy library, version 0.6.1 [22]. ABP peaks were found based on the location of previously detected R-peaks, using the find_peaks algorithm from the SciPy library, version 1.3.1 [23]. PTT, SBP and HR were then extracted from the data, and moving averages were calculated. The latter was achieved using a 1st order Savitzky-Golay filter with a width of 21 samples. In short, this entails fitting a straight line through the preceding 10 and following 10 heartbeats in order to determine the value at each heartbeat. Extracted variables and their definitions are shown in Table 1. A cutout from one of the segments, providing a visualization of how the PTT-metric is calculated is shown in Fig 1.

thumbnail
Fig 1. Illustration of extracted variables.

The figure displays a cut-out of a waveform time series from an arbitrary subject. Annotations define the points and intervals that were extracted from waveforms and used for analysis.

https://doi.org/10.1371/journal.pone.0240126.g001

Data analysis

In order to evaluate the endpoint of the study, linear regression was chosen. For each included segment, two separate analyses were performed: (1) (2) Performing linear regression on averaged data is somewhat unorthodox. In order to verify that the results are not simply a result of the employed methodology, analysis (2) was included as a “negative” control. Reductions in SBP tend to cause reflexive tachycardia, but may also be secondary to a reduction in HR. Thus, the relationship between HR and SBP is not expected to be consistent across subjects. Ultimately, PTT needs to outperform HR in predicting SBP for the results to be considered meaningful.

Results

Summary of inclusion and descriptive statistics

Potential events were identified for 2729 subjects. Of these, 511 subjects experienced events without missing/ambiguous peaks. From each of these 511 subjects, one event was included for analysis. Datasets outlining the specific waveform segments that were included for the main analysis can be found in the S1 File. An overview of the inclusion process is shown in Fig 2. Summaries of subject- and event- characteristics are given in Tables 2 and 3.

thumbnail
Fig 2. Overview of included and excluded subjects.

The flow chart displays the process to assess subject eligibility for the main analysis.

https://doi.org/10.1371/journal.pone.0240126.g002

Results from regression analysis

Median correlation coefficients (r) obtained from linear regression versus SBPAV were as follows: PTT-RAAV -0.93 (IQR -0.98 to -0.76), HRAV 0.46 (IQR -0.16 to 0.83). For 36 subjects (7.0%), correlation coefficients between SBPAV and PTT-RAAV were positive. A frequency histogram of correlation coefficients is presented in Fig 3.

thumbnail
Fig 3. Frequency histogram of Pearson Correlation Coefficients versus SBPAV.

The figure displays the distribution of Pearson Correlation Coefficients for (a) SBPAV vs. PTT-RAAV and (b) SBPAV vs. HRAV. A value of -1 signifies a perfect negative correlation, a value of 1 signifies a perfect positive correlation, and a value of 0 means no correlation. n = 511.

https://doi.org/10.1371/journal.pone.0240126.g003

Regression slopes for the relationship between SBPAV and PTT-RAAV displayed a median of -1.23 mmHg/ms (IQR -1.73 mmHg/ms to -0.81 mmHg/ms). The distribution of slopes is shown in Fig 4.

thumbnail
Fig 4. Slopes from linear regression of SBPAV vs. PTT-RAAV.

The figure displays the estimated change in SBP as a function of the observed PTT-RAAV. Each dark line represents the slope of said relationship for one subject. n = 511.

https://doi.org/10.1371/journal.pone.0240126.g004

Fig 5 shows individual data for the subject whose combination of correlation coefficients was closest to the median. Fig 5A–5C shows the time series of extracted parameters, while associated regression plots are shown in Fig 5D–5F. The alignment of data points in vertical bands in Fig 5D illustrates the limited temporal resolution of the PTT metric at a sample rate of 125Hz. Fig 5E and 5F illustrate a side effect of comparing averaged time series: Since subsequent data points are conditioned upon each other, temporal changes can be inferred as a gradual walk over time, even in the absence of a temporal axis.

thumbnail
Fig 5. Event visualization and regression for subject 20936.

a-c: Time series of extracted parameters (SBP, PTT-RA and HR). d: Regression plot of PTT-RA vs. SBP. e: Regression plot of PTT-RAAV vs. SBP. f: Regression plot of HR vs. SBP.

https://doi.org/10.1371/journal.pone.0240126.g005

Supplementary analysis

In addition to the main analyses, three supplementary analyses were performed to gauge the robustness of the main findings.

  1. Linear regression was performed on the raw data for SBP and PTT-RA, without applying any prior averaging.
  2. For those subjects included in the primary analysis who experienced more than one event, comparative analysis was performed on the secondary event.
  3. The main analysis was repeated with less stringent inclusion criteria, accepting subjects who had < 5 missing/ambiguous peaks (contrary to 0 for the main analysis)

(1) Linear regression without averaging.

Regression analysis on raw data without averaging gave the following results.

  1. SBP vs. PTT-RA correlation coefficient: Median -0.76 (IQR -0.88 to -0.51)
  2. SBP vs. HR correlation coefficient: Median 0.21 (IQR -0.06 to 0.57)
  3. SBP vs. PTT-RA regression slopes: Median -0.81 mmHg/ms (IQR -1.08 to -0.56)

(2) Analysis of secondary events.

215 subjects experienced at least one secondary event that met the inclusion criteria. Regression analysis per subject for secondary events gave the following results:

  1. SBPAV vs. PTT-RAAV correlation coefficient: Median -0.92 (IQR -0.97 to -0.75)
  2. SBPAV vs. HRAV correlation coefficient: Median 0.53 (IQR -0.16 to 0.87)
  3. SBPAV vs. PTT-RAAV regression slopes: Median -1.24 mmHg/ms (IQR -1.72 to -0.93)

Regression slopes for primary and secondary segments were also compared in a paired within-subject fashion. Bland-Altman analysis revealed a mean bias of 0.07 mmHg/ms (95% LOA -2.8 to 2.94). The sign of the correlation coefficients differed between segments for 20 subjects (9.3%). The Bland-Altman plot is shown in Fig 6.

thumbnail
Fig 6. Bland-Altman plot of regression slopes for primary vs. secondary segments.

The mean value of the regression slope for two segments from the same subject is displayed on the x-axis. The bias, i.e. the difference between the regression slopes of the two segments, is displayed on the y-axis (n = 215).

https://doi.org/10.1371/journal.pone.0240126.g006

(3) Linear regression with less stringent inclusion criteria.

935 subjects experienced one or more events with less than 5 missing/ambiguous peaks, while still fulfilling the other inclusion criteria. Regression analysis per subject for one such event per subject gave the following results:

  1. SBPAV vs. PTT-RAAV correlation coefficient: Median -0.91 (IQR -0.97 to -0.72)
  2. SBPAV vs. HRAV correlation coefficient: Median 0.32 (IQR -0.26 to 0.78)
  3. SBPAV vs. PTT-RAAV regression slopes: Median -1.07 mmHg/ms (IQR -1.6 to -0.65)

Discussion

Main findings

On a per-subject basis, PTT-RAAV displayed a moderate to strong association with SBPAV during rapid declines in SBP. For the majority of subjects, a clear inverse correlation was observed, as predicted by the theoretical relationship between pressure and velocity. The distribution of correlation coefficients and regression slopes did not change substantially when including an additional 424 subjects in supplementary analysis with less stringent inclusion criteria. However, for a non-negligible proportion of subjects, there was no correlation, or even a positive correlation between SBPAV and PTT-RAAV. There was also a considerable range in the slopes of the regression lines, as evidenced in Fig 4. Additionally, regression slopes were not always consistent within subjects across different events. These limitations must be addressed for PTT-based tracking of BP-changes to be feasible.

With regard to the spread of correlation coefficients, some outliers may be ascribed to faulty peak detection. On post-hoc manual inspection it was apparent that ECG peak detection was poor for some subjects, with bipolar pacemakers being one particular challenge for the peak detection algorithm. These segments were retained for analysis. From a physiological standpoint, one possible cause of outliers may lie in the nuances between PTT and PWV. The PTT-metric as measured from the ECG R-peak actually includes two components, namely the pre-ejection period (PEP) and the vascular transit time (VTT). The PEP corresponds to the isovolumetric contraction of the left ventricle, while the VTT denotes the time elapsed from the pulse wave transverses the aortic valve until it appears at some more peripheral site. Principally then, VTT is the component of primary interest, as it is theoretically proportional to pressure through changes in the elastic modulus. The relationship between PEP and pressure appears to be more complex. In a series of experiments on denervated dog hearts, Wallace et al. showed that increasing HR or SV led to a shortening of PEP. Conversely, increasing aortic pressure by adjusting a mechanical resistance distal to the aorta led to lengthening of the PEP [24]. The findings of Wallace et al. suggest that changes in PEP and VTT should be equidirectional when changes in BP are mediated by changes in cardiac output, but that they would be opposite in situations where BP-changes are mediated by changes in peripheral resistance. The implication is that PTT incorporating PEP would be less robust at tracking SBP in vasodilatory shock, such as in sepsis or anaphylaxis. The etiologies of progression to hypotension in the current study are not known. However, the high proportion of subjects admitted for cardiac surgery, and the tendency of HR to be positively correlated with SBP, may suggest that the dataset is biased towards BP-reductions caused by reduction in cardiac output. Future research may focus specifically on the relationship between PTT and SBP in patients with confirmed vasodilatory shock.

Averaging time series data prior to performing regression analysis may be a dubious exercise, and presents a risk of detecting spurious relationships. Generally, averaged data will always be farther from the ‘truth’ than the underlying data. More specifically, high frequency fluctuations in the underlying data will be attenuated. With this limitation in mind, we still believe that the use of averaging prior to regression is justified in this case, for the following reasons: I) the purpose of the current work is to model short-term trends in SBP. Beat-to-beat fluctuations in SBP, which are attenuated by averaging, are of limited interest for this purpose. II) We have included HRAV (averaged in an identical way) as a negative control, specifically to address the question of the method’s validity. If correlations between SBPAV and PTT-RAAV were largely a consequence of averaging, similar results should be observed for the correlation between SBPAV and HRAV. III) Supplementary analyses on raw data without averaging show that while magnitudes are attenuated, median correlation coefficients as well as median regression slopes display similar trends.

Having made justifications for averaging data, one should also underline the potential advantages of such an approach. Firstly, the procedure allows for the PTT-metric to assume more unique values than without averaging. When the magnitude of PTT-variability is low in relation to the sample rate, this inherently limits the strength of correlation, as demonstrated in Fig 5. Secondly, averaging reduces the influence of respiration. It is well established that respiration induces oscillations in both HR and BP. Furthermore, these oscillations have been shown to occur with a phase offset [25]. The complex interaction between respiration, HR and BP may potentially distort the beat-to-beat relationship between PTT and SBP, a problem that is avoided through averaging. With regard to the consequence of spurious PTT-outliers, averaging presents both advantages and disadvantages. On one hand, the averaging process prevents such outliers from being interpreted as sudden extreme perturbations in SBP. On the other hand, averaging allows outliers to influence surrounding data points, causing sustained error over several heartbeats.

It is emphasized once again that the PTT metric was calculated using the peak of the ABP signal. Classically, PTT has been measured as the time interval from the R-peak of the ECG to the peak of the PPG waveform obtained at the finger. Theoretically, PTT should covary with BP through changes in the elastic modulus of the vessel, but this relationship assumes an elastic, homogenous vessel. The propagation of the pulse wave through a plethora of resistance vessels (i.e. small arteries and arterioles), in which the elastic properties of the wall are modulated by smooth muscle, is likely to confound the relationship between pressure and velocity. In light of the heterogeneity of distal perfusion status in critically ill patients, the PPG peak obtained at the finger may reasonably be inferior to that obtained from the radial artery waveform. In a practical scenario, the invasive ABP-signal is clearly not available, as that would defeat the purpose of trying to estimate SBP. As such, a non-invasive method of obtaining pulse peaks proximal to resistance vessels may be necessary for practical application of our method. For any such non-invasive method, wave morphology may be of importance, as wave rise time is included in the measured time interval.

Novelties of the current investigation

To the best of our knowledge, this is the first investigation into the dynamics of PTT during rapid declines in systolic blood pressure. Reliability in detecting such episodes is of the utmost importance if PTT-based cNIBP is to be employed in hospital wards in the future. Furthermore, we report within-subject comparison of the PTT-SBP relationship across different episodes in time. Such data is critical to appraise the stability of said relationship, and is rather scarce in the published literature. Lastly, we provide clinical characteristics of the included subjects, which is often absent in the recent literature.

Technical limitations

The MIMIC database has some known issues relating to waveform sampling. Signals have been downsampled to 125 Hz, limiting the resolution of the PTT and HR metrics. In this process, ECG-signals were sampled on a varying interval between 4–12 ms, instead of a constant 8 ms. Additionally, synchronicity between physiological waveforms cannot be guaranteed. These limitations should generally lead to underestimation rather than overestimation of the association between SBPAV and PTTAV, but they warrant caution in interpretation of results.

Generalizability and risk of bias

Based on the observed patient characteristics of our dataset, some reservations should be made about the generalizability of our findings. Perhaps most striking is the low proportion of non-white subjects, collectively amounting to only 15.3%. It is not known whether the relationship between PTT and BP differs substantially based on ethnicity. The median age in the cohort was 68 years, and 49.1% of subjects had a primary diagnosis of a disease of the circulatory system. It is therefore likely that there was a high prevalence of atherosclerosis among the subjects. Seeing as the theoretical relationship between BP and PTT depends on the elastic properties of arteries, and atherosclerosis may modify such properties, it is possible that our results would not be generalizable to younger, healthier individuals.

Segments to be included for analysis were selected using an automated process, which should reduce the risk of investigators introducing bias. The algorithms employed in this process, however, require a number of cutoffs for decision rules which had to be determined by the investigators. While all such decision rules were finalized prior to data analysis, cutoffs set to reduce risk of faulty peak detection and noise could conceivably lead to a systematic bias in choice of signals. Additionally, signals which are objectively unfit for analysis due to noise and difficulty in peak detection may be associated with pathological states such as arrhythmia or poor distal perfusion. The association between SBP and PTT may be systematically different in these patients, constituting a risk of bias and reduced generalizability.

Further research

Based on our findings, we present three potential avenues for further research:

(I) PTT-measurement must be performed non-invasively in order to have potential benefit. Using transmission-PPG limits sites to finger, toe or earlobe, but it may be desirable to obtain pulse waves from sites less affected by distal vasoconstriction. Investigators should therefore consider the use of reflectance-based PPG or other non-invasive sensor technologies to attempt to mitigate this issue.

(II) Substantial inter-individual variation in regression parameters prohibits widespread application of PTT-RAAV for detecting hypotension without some form of calibration or additional parameters. Further research should explore modifying factors that could aid patient selection or improve the generalizability of the method. Furthermore, it will be necessary to include patient populations in which invasive hemodynamic monitoring is not indicated, as these patients are likely the most suitable candidates for cNIBP.

(III) We have shown that PTT-RAAV will generally follow an inverse trajectory to that of SBPAV during rapid declines in SBP, indicating that the method could have high sensitivity for detecting progression to hypotension. However, as our research is limited to such episodes, we cannot comment on the specificity. Further research may aim to demonstrate whether prolongation of PTT-RAAV occurs during periods of stable or rising SBP.

Conclusions

We have investigated changes in pulse transit time during rapid declines in blood pressure, in a large cohort of patients admitted to an intensive care unit. There was a moderate-strong intra-individual correlation between averaged systolic blood pressure, and averaged pulse transit time as measured from ECG R-peak to the peak of the arterial blood pressure waveform. Several methodological challenges remain, but our findings suggest that real-time, non-invasive detection of hypotension in hospitalized patients may be feasible.

Supporting information

S1 File. Event parameters.

Contains all the information necessary to identify the segments that were used for the main analysis.

https://doi.org/10.1371/journal.pone.0240126.s001

(XLSX)

References

  1. 1. Vincent J-L, De Backer D. Circulatory Shock. New England Journal of Medicine. 2013;369:1726–34. pmid:24171518
  2. 2. Turan A, Chang C, Cohen B, Saasouh W, Essber H, Yang D, et al. Incidence, Severity, and Detection of Blood Pressure Perturbations after Abdominal Surgery: A Prospective Blinded Observational Study. Anesthesiology. 2019;130:550–9. pmid:30875354
  3. 3. Marey E-J. Pression et vitesse du sang. Physiologie experimentale. 1876;2:307.
  4. 4. Shirer HW. Blood Pressure Measuring Methods. IRE Transactions on Bio-Medical Electronics. 1962;9:116–25. pmid:13912106
  5. 5. Penaz J. Photoelectric measurement of blood pressure, volume and flow in the finger. Digest 10th Int. Conf Med Biol Eng. 1973.
  6. 6. Wesseling KH. Finapres, continuous noninvasive finger arterial pressure based on the method of Peñáz. In: Meyer-Sabellek W, Gotzen R, Anlauf M, Steinfeld L, editors. Blood Pressure Measurements. Heidelberg: Steinkopff; 1990. p. 161–72.
  7. 7. Drzewiecki GM, Melbin J, Noordergraaf A. Arterial tonometry: review and analysis. J Biomech. 1983;16:141–52. pmid:6863329
  8. 8. Gasser TC. Physical processes in the vessel. In: Krams R, Bäck M, editors. The ESC Textbook of Vascular Biology [Internet]. Oxford University Press; 2017 [cited 2020 Jan 7]. Available from: https://oxfordmedicine.com/view/10.1093/med/9780198755777.001.0001/med-9780198755777-chapter-3
  9. 9. Hughes DJ, Babbs CF, Geddes LA, Bourland JD. Measurements of Young’s modulus of elasticity of the canine aorta with ultrasound. Ultrason Imaging. 1979;1:356–67. pmid:575833
  10. 10. Gribbin B, Steptoe A, Sleight P. Pulse Wave Velocity as a Measure of Blood Pressure Change. Psychophysiology. 1976;13:86–90. pmid:1244636
  11. 11. Newlin DB. Relationships of pulse transmission times to pre-ejection period and blood pressure. Psychophysiology. 1981;18:316–21. pmid:7291449
  12. 12. Pollak MH, Obrist PA. Aortic-radial pulse transit time and ECG Q-wave to radial pulse wave interval as indices of beat-by-beat blood pressure change. Psychophysiology. 1983;20:21–8. pmid:6828608
  13. 13. Ding X, Zhang Y-T. Pulse transit time technique for cuffless unobtrusive blood pressure measurement: from theory to algorithm. Biomed Eng Lett. 2019;9:37–52. pmid:30956879
  14. 14. Kachuee M, Kiani MM, Mohammadzade H, Shabany M. Cuff-less high-accuracy calibration-free blood pressure estimation using pulse transit time. 2015 IEEE International Symposium on Circuits and Systems (ISCAS). 2015. p. 1006–9.
  15. 15. Johnson AEW, Pollard TJ, Shen L, Lehman LH, Feng M, Ghassemi M, et al. MIMIC-III, a freely accessible critical care database. Sci Data. 2016;3:1–9.
  16. 16. Goldberger Ary L., Amaral Luis A. N., Leon Glass, Hausdorff Jeffrey M., Ivanov Plamen Ch., Mark Roger G., et al. PhysioBank, PhysioToolkit, and PhysioNet. Circulation. 2000;101:e215–20. pmid:10851218
  17. 17. Liang Y, Abbott D, Howard N, Lim K, Ward R, Elgendi M. How Effective Is Pulse Arrival Time for Evaluating Blood Pressure? Challenges and Recommendations from a Study Using the MIMIC Database. J Clin Med [Internet]. 2019 [cited 2019 May 6];8. Available from: https://www.ncbi.nlm.nih.gov/pmc/articles/PMC6462898/
  18. 18. Zhu Y, Zhang H, Guan C. On robust extraction of pulse transit time from multimodal pulsatile signals. 2015 IEEE International Conference on Digital Signal Processing (DSP). 2015. p. 403–6.
  19. 19. Lin Y-T, Lo Y-L, Lin C-Y, Frasch MG, Wu H-T. Unexpected sawtooth artifact in beat-to-beat pulse transit time measured from patient monitor data. PLOS ONE. 2019;14:e0221319. pmid:31498802
  20. 20. Bennis FC, van Pul C, van den Bogaart JJL, Andriessen P, Kramer BW, Delhaas T. Artifacts in pulse transit time measurements using standard patient monitoring equipment. PLoS ONE. 2019;14:e0218784. pmid:31226142
  21. 21. Python Language Reference, version 3.7 [Internet]. Python Software Foundation; Available from: https://www.python.org
  22. 22. Carreiras C, Alves AP, Lourenço A, Canento F, Silva H, Fred A. BioSPPy—Biosignal Processing in Python [Internet]. 2015 [cited 2020 Jan 7]. Available from: https://github.com/PIA-Group/BioSPPy
  23. 23. Virtanen P, Gommers R, Oliphant TE, Haberland M, Reddy T, Cournapeau D, et al. SciPy 1.0—Fundamental Algorithms for Scientific Computing in Python. arXiv:190710121 [physics] [Internet]. 2019 [cited 2020 Jan 7]; Available from: http://arxiv.org/abs/1907.10121
  24. 24. Wallace Andrew G., Mitchell Jere H., Sheldon Skinner N., Sarnoff Stanley J. Duration of the Phases of Left Ventricular Systole. Circulation Research. 1963;12:611–9. pmid:13998538
  25. 25. Toska K, Eriksen M. Respiration-synchronous fluctuations in stroke volume, heart rate and arterial pressure in humans. J Physiol (Lond). 1993;472:501–12.