Abstract
Validation of heart rate responses in wearable technology devices is generally composed of laboratory-based protocols that are steady state in nature and as a result, high accuracy measures are returned. However, there is a need to understand device validity in applied settings that include varied intensities of exercise. The purpose was to determine concurrent heart rate validity during trail running. Twenty-one healthy participants volunteered (female n = 10, [mean (SD)]: age = 31 [11] years, height = 173.0 [7] cm, mass = 75.6 [13] kg). Participants were outfitted with wearable technology devices (Garmin Fenix 5 wristwatch, Jabra Elite Sport earbuds, Motiv ring, Scosche Rhythm+ forearm band, Suunto Spartan Sport watch with accompanying chest strap) and completed a self-paced 3.22 km trail run while concurrently wearing a criterion heart rate strap (Polar H7 heart rate monitor). The trail runs were out-and-back with the first 1.61 km in an uphill direction, and the 1.61 return being downhill in nature. Validity was determined through three methods: Mean Absolute Percent Error (MAPE), Bland-Altman Limits of Agreement (LOA), and Lin’s Concordance Coefficient (rC). Validity measures overall are as follows: Garmin Fenix 5 (MAPE = 13%, LOA = -32 to 162, rC = 0.32), Jabra Elite Sport (MAPE = 23%, LOA = -464 to 503, rC = 0.38), Motiv ring (MAPE = 16%, LOA = -52 to 96, rC = 0.29), Scosche Rhythm+ (MAPE = 6%, LOA = -114 to 120, rC = 0.79), Suunto Spartan Sport (MAPE = 2%, LOA = -62 to 61, rC = 0.96). All photoplethysmography-based (PPG) devices displayed poor heart rate agreement during variable intensity trail running. Until technological advances occur in PPG-based devices allowing for acceptable agreement, heart rate in outdoor environments should be obtained using an ECG-based chest strap that can be connected to a wristwatch or other comparable receiver.
Citation: Navalta JW, Montes J, Bodell NG, Salatto RW, Manning JW, DeBeliso M (2020) Concurrent heart rate validity of wearable technology devices during trail running. PLoS ONE 15(8): e0238569. https://doi.org/10.1371/journal.pone.0238569
Editor: Daniel Boullosa, Universidade Federal de Mato Grosso do Sul, BRAZIL
Received: May 13, 2020; Accepted: August 19, 2020; Published: August 31, 2020
Copyright: © 2020 Navalta et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Data Availability: All files are available from the following URL: https://doi.org/10.7910/DVN/T88MIV.
Funding: The author(s) received no specific funding for this work.
Competing interests: The authors have declared that no competing interests exist.
Introduction
The use of wearable technology has steadily increased, and has been the top fitness trend since 2016 as determined by health and fitness professionals throughout the world [1]. Wearable devices can return a variety of physiological and health-related metrics including estimates of energy expenditure [2], step count [3], and heart rate [4]. With over 10,000 device options from all ends of the budgetary spectrum [5], determining an appropriate wearable device can be difficult. When considering data output from wearable devices, heart rate is generally the most valid, energy expenditure has been shown to be the least accurate, and step count validity falls between [6]. Focusing on heart rate validity while running, the majority of literature has utilized laboratory-based protocols that are steady state in nature.
Wearable technology devices that return heart rate include smart bras [7], earbuds [8, 9] and sensors placed around the forearm [10, 11] or wrist [12, 13]. Several investigations have utilized treadmill running protocols when evaluating the agreement of heart rate compared with electrocardiogram (ECG) [4, 11, 14, 15] or chest strap criterion measures [8, 10, 16–18], which will be taken as validity in the current investigation. These running protocols have utilized 1.5-minute [11], 3-minute [14, 15, 18, 19], 4-minute [17], 5-minute [4, 8, 10], and 6-minute time frames [16] with a constant speed. In investigations utilizing more than one speed [4, 8, 10, 14, 15, 17–19], treadmill speed at each successive stage is held constant. Reported heart rate validity measures are often combined with walking [8, 10, 11, 19] and are generally high across device types including Jabra ear buds (mean absolute percent error [MAPE] = 2.5%, intraclass correlation coefficient [ICC] = 0.94) [8], Garmin Forerunner 235 (mean difference = 0.1 bpm, ICC = 0.905) [18], Apple Watch (absolute percent difference = 5%, r = 0.93) [11], Fitbit Charge HR (bias = -3 bpm, r = 0.95) [4], and Mio Alpha (mean difference = -4, ICC = 0.91) [14]. As these investigations employed one or more intensities eliciting steady-state heart rate responses, it is likely that the observed high validity measures for heart rate during treadmill running is due to the controlled nature of this activity (sagittal plane motion within a narrow range of motion).
A limited number of studies have not utilized the treadmill, but engaged participants in unrestricted activities including walking and running [13, 20–22]. Dondzlia et al. evaluated the Fitbit Charge HR and Mio FUSE devices in activities including treadmill walking and running and during free living activities for a 24h period [20]. Unfortunately, heart rate validity was not reported for the free living time frame [20]. Reddy et al. utilized the Garmin VivoSmart HR+ and the Fitbit Charge 2 in a 28-min protocol of activities of daily living and interpreted acceptable heart rate accuracy for both devices (mean absolute error = 13%, r = 0.69; mean absolute error = 11%, 0.73 respectively) [21]. Brazendale et al. evaluated the Fitbit Charge HR during a 2h free-living segment in children and determined that it returned accurate heart rate values (absolute percent difference = 7%, r = 0.84) [13]. Montes et al. compared unrestricted walking and jogging through a flat hallway to a speed matched treadmill protocol and found that heart rate was significantly different between environments for certain devices [22]. It was suggested that the differences between environments could at least partially be due to the fact that running speed can vary with free motion activity while it is held constant on the treadmill [22].
As seen from the literature above, agreement between wearable devices and a heart rate criterion measure during running are generally higher during treadmill-based activity [4, 8, 11, 18] and tend to be lower when unrestricted activities are included [13, 21]. It has been suggested that the main source of error with heart rate obtained from photoplethysmography-reliant (PPG) wearable devices is motion artifact in the wave form that decreases accuracy as exercise intensity increases [23]. As running outdoors can be a variable and intermittent activity including high intensity exercise, it is important to determine the accuracy of heart rate obtained from now pervasive wearable technology devices in this applied setting. As our laboratory group has experience obtaining measurements in a variety of natural settings [3, 24–29], and as trail running is increasing in popularity [30, 31], we designed this investigation to determine heart rate validity of wearable monitors during a trail run. It was hypothesized that all devices would display acceptable heart rate validity when compared to a concurrently measured criterion. In order to be acceptable, all of the following criteria must be met: MAPE ≤5%, Lin’s Concordance Coefficient ≥ 0.90, and ICC ≥ 0.70.
Methods
Participants
Twenty-one healthy participants volunteered for this study (female n = 10, not presenting with cardiovascular, metabolic, or renal disease; no signs or symptoms suggestive of cardiovascular, metabolic, or renal disease) [32]. Heart rate validity literature has returned large effect sizes (0.91 to 0.95) [4, 18] however, to be conservative a moderate effect size (0.5) was utilized to determine that a total sample size of 21 would be sufficient. Participant descriptive characteristics included the following (mean±SD): age = 31±11 years, height = 173.0±6.9 cm, mass = 75.6±12.9 kg. Participants completed an informed consent document that was approved by the University of Nevada, Las Vegas Biomedical Sciences Institutional Review Board (protocol #1292334).
Protocol
Participants were outfitted with wearable technology devices and completed a self-paced 3.22 km (two-mile) trail run while concurrently wearing a criterion heart rate strap. Devices were secured to the participant by members of the research team to ensure appropriate fit, and to certify heart rate was obtained by each device. The trail runs were an out-and-back course with the first 1.61 km (1-mile) in a generally uphill direction, and the 1.61 (1-mile) return being generally downhill in nature. Three different trails were utilized which included elevation changes of 48 m (McCullough Hills Trail, Henderson, NV), 55m (Three Peaks Trail, Cedar City, UT) and 104m (Bristlecone Trail, Mt. Charleston, NV). The elevation profiles of these trails are shown in Fig 1.
Criterion heart rate: the criterion measure was obtained using the Polar H7 heart rate monitor (Polar Electro, Kempele, Finland), which contains a single flexible plastic sensor (2.4 x 27.9 cm), worn concurrently and placed on the sternum. The Polar H7 heart rate receiver has a sample rate of 1000 Hertz and has high agreement with ECG measurements during various exercise modalities (treadmill, cycle, elliptical; Lins’ concordance correlation coefficient = 0.99 each) [11].
ECG chest strap device and watch: Suunto Spartan Sport watch (Suunto Oy, Vantaa, Finland)—the watch was secured around the right wrist by the strap. Heart rate was obtained through the accompanying heart rate strap that was secured around the sternum just below the Polar H7 monitor. The dimensions are 50 x 50 x 13.8 mm.
Photoplethysmography-based devices: Garmin Fenix 5 wristwatch (Garmin Ltd, Olathe, KS)—the device was secured to the left wrist by the strap. This device utilizes the Garmin Elevate™ multi-sensor heart rate monitor. The dimensions are 47 x 47 x 15.5 mm.
Jabra Elite Sport earbuds (Jabra, Copenhagen, Denmark)—earbuds were secured into the ear canal. The dimensions are 120 x 45 x 179 mm.
Motiv ring (Motiv Inc, San Francisco, CA)—a size 10 ring was secured on the finger that provided the best fit, being able to fit over the knuckle and fit snugly on the finger between the proximal interphalangeal and metacarpophalangeal joints. The appropriate finger was self-selected by each participant. The outer housing is composed of a titanium alloy, and employs a green LED optical heart rate sensor.
Scosche Rhythm+ forearm band (Scosche Industries Inc., Oxnard, CA)—secured around the forearm by a breathable neoprene band. Sensor size is 54.4 x 48.8 x 14.7 mm.
Heart rate data for the Polar H7, Rhythm+ forearm band, Jabra Elite earbuds, and Suunto Spartan Sport were transmitted real time via Bluetooth to a synced iPad mini tablet (Apple Inc., Cupertino, CA) and captured into the PerformTek application (Valencell Inc., Raleigh, NC). The PerformTek application reports heart rate in a second-by-second fashion and allowed the Rhythm+ forearm band, Jabra Elite earbuds, and Suunto Spartan Sport to be time-synced with the Polar H7. Heart rate from the Motiv ring was transmitted via Bluetooth to the synced iPad mini and captured into the Motiv application (Motiv Inc, San Francisco, CA). Heart rate data from the Garmin Fenix 5 was captured onto the watch during the trail run and synced to the Garmin Connect application on the iPad mini immediately upon return to the trailhead. Heart rate data from the Motiv ring and Garmin Fenix 5 were plotted against the criterion measure to visually confirm time-syncing between devices.
Statistical analysis
All data are expressed as mean±SD. Except for the Motiv ring (minute-by-minute), data analysis was performed on second-by-second values expressed in beats per minute (bpm). Validity was determined through multiple methods [6]: Bland-Altman bias and Limits of Agreement (LOA) for repeated samples with accompanying 95% confidence intervals (CI) [33], Mean Absolute Error (MAE) and Mean Absolute Percent Error (MAPE), Lin’s Concordance Coefficient (rC), and Intraclass Correlations (ICC) (IBM SPSS, IBM Statistics version 24.0, Armonk, NY). Validity was considered if devices met all of the following criteria: MAPE within 5% [7], rC greater than 0.90 [8, 34], and ICC greater than 0.70 with a p-value less than 0.05 [35].
Results
The average time of the 2-mile trail run was 21:56 (5:38) min (mean [SD]). The uphill portion of the run was 11:39 (3:19) min, and the downhill portion was 10:26 (2:17) min. Validity measures for heart rate (bpm) over the entire length of trail run are shown in Table 1. Heart rate (bpm) validity for the uphill portion is shown in Table 2, and the return is displayed in Table 3. Bland Altman plots for each device for the entire trail run are displayed in Fig 2.
Discussion
The purpose of this investigation was to evaluate heart rate validity of several types of wearable technology devices during a variable intensity trail run. We hypothesized that all devices would have acceptable heart rate agreement compared to our criterion over the course of this exercise bout. The main findings are that regardless of device location (finger, wrist, ear, forearm), PPG-based devices do not provide acceptable heart rate validity during a trail run lasting longer than 20-min. We found that a heart rate strap recorded to a wrist watch (Suunto Spartan Sport), provided acceptable agreement for heart rate utilizing thresholds normally applied to laboratory-based research.
The Suunto Spartan Sport device has been evaluated with respect to step count accuracy [36], and proposed as a wearable capable of returning the cardiorespiratory fitness component of an integrated cross-modal cybernetic health status assessment [37]. The device has also been utilized in an outdoor environment to track altitude profile during a 64 km ultra-endurance race [38] and Grand Canyon rim to rim hike [39], and pacing and stride variations during a 44 km trail run performed in tropical conditions [40]. To our knowledge, concurrent heart rate validity has not been determined compared to a criterion measure. Thus, we report for the first time that when paired with accompanying heart rate strap, the Suunto Spartan Sport displays acceptable heart rate validity during variable intensity trail running. As the heart rate strap is an ECG-based wearable similar to the criterion measure (Polar H7) utilized in the current study, these findings should be expected.
Forearm PPG heart rate monitors have been utilized in a variety of applications including virtual reality ship handling simulators [41], during an interactive game of tag [42], evaluating driving during various weather conditions [43], and for remote monitoring of triathlon training [44]. Two investigations have determined heart rate validity of forearm devices during exercise [11, 22]. The Scosche Rhythm+ heart rate agreed with the ECG criterion at rest (rC = 0.93), and during cycle and treadmill exercise lasting 4.5 min (rC = 0.84, 0.92), but not when performing elliptical training (rC = 0.41 with arms, 0.27 without arm movement) [11]. Another investigation reported no difference in heart rate compared to Polar H7 measures when participants completed 5-min of walking and running both on a treadmill and in an unrestricted setting (p>0.05) [22]. The current investigation extends the literature on forearm heart rate monitors to outdoor exercise in the form of variable intensity trail running. Our results are interesting in this regard, as better agreement was observed during downhill running (MAPE = 3.8%, bias = 1.9 bpm, rC = 0.885) than when running was at a generally positive incline (MAPE = 6.2%, bias = 3.9 bpm, rC = 0.699). It is possible that greater device motion was experienced in the uphill portion leading to these results, and further investigation is warranted.
The Garmin Fenix series is relatively new PPG-based device and as such has limited available literature. In fact, we were only able to find two conference abstracts for the Fenix 3 [45, 46], and one recent laboratory-based investigation on the Fenix 5 [47]. The Garmin Fenix 3 estimation of maximal aerobic capacity was not different compared to laboratory-based metabolic analysis (p>0.05) [46]. Biomechanical running parameters of stride length and run cadence were not different compared to laboratory measurements (p>0.05), but significant differences were observed for vertical oscillation and ground contact time (p<0.05) [45]. Düking et al. reported moderate heart rate validity for the Garmin Fenix 5 during sitting and walking (standardized typical error of the estimate [sTEE] = 0.63, 0.62)to be poor with increased intensities of exercise (9.9 MET level sTEE = 1.24, 13.8 MET level = 1.44) and recommend caution due to the higher rates of error [47]. The results of the current investigation extend the literature in that poor heart rate validity measures in the Garmin Fenix 5 were observed during a variable intensity trail run (MAPE = 13.5%, bias = 15.9 bpm, rC = 0.316), and these measures are consistent regardless of the incline (either primarily uphill or downhill) of the trail.
While ring-based wearable devices have been presented in conference abstracts [48–51], we have been unable to find published literature incorporating its use during exercise. Thus, the current investigation reports concurrent heart rate validity in a PPG ring-based wearable device during exercise for the first time. We must recommend caution for this type of device, as heart rate validity during trail running exercise was the poorest of all PPG devices tested (MAPE = 15.9%, bias = 21.7 bpm, rC = 0.293). Future investigations utilizing ring-based wearables should be investigated in a controlled laboratory setting to determine heart rate validity in this environment. A limitation with respect to this device is that the current investigation utilized a single ring size. Additionally, the specific fingers utilized by participants was not recorded and future studies should take this into account.
Literature on PPG-based earbud heart rate validity during exercise is beginning to emerge. Investigations have found earbud-based heart rate to be acceptable during resistance training (MAPE = 6.24%) [9], graded exercise testing on a treadmill (bias = -0.2%, R2 = 0.98) [52], and during treadmill exercise and high intensity training exercises (bias = 0.8 bpm, MAPE = 2.48%, rC = 0.943; bias = -3.6 bpm, MAPE = 3.53%, rC = 0.861 respectively) [8]. When utilized in patients with cardiac diseases, these devices tended to have less agreement when heart rate was above 100 bpm and in participants with atrial fibrillation (average difference to true heart rate = 20.3 bpm, r2 = 0.434) [53]. One investigation utilizing cycle graded exercise testing found that heart validity decreased in earbud devices as the intensity of this exercise increased (MAPE at 50W = 6.4%, MAPE at 200W = 15.42%) [9]. The results of the current study indicate that heart rate validity in PPG-based earbud devices is poor during running in an outdoor environment at variable intensities of exercise.
The current investigation utilized a variety of commercially available wearable technology devices capable of returning heart rate measures. We observed a range of heart rate agreement in devices during a trail run and as expected found an ECG-based chest strap recording to a wrist worn watch to have the greatest concurrent validity. The PPG-based devices were positioned at various locations on the body, and the forearm strap device displayed the best validity of this class of wearable. Nevertheless, in an outdoor environment, heart rate obtained from the remaining PPG-based devices (wrist, forearm, ring-based, and earbuds) were observed to have poor agreement with the criterion measure. Motion artifact is the likely source of error in wearable devices that were found to have low heart rate validity [23], however the influence of ambient light [54] should also be considered another source of potential error, particularly with the use of photoplethysmography-reliant devices during outdoor exercise. The Consumer Technology Association recommends 5-min exercise durations when validating wearable technology devices [55], and investigations specific to running have utilized protocols of 1.5-minutes [11], 3-minutes [14, 15, 18, 19], 4-minutes [17], 5-minutes [4, 7, 8, 10], and 6-minutes [16]. The current investigation is unique in that heart rate validity was assessed over a much longer duration (average running time was near 22 min). Toward this end we recommend that future studies continue to determine validity in applied settings and over longer time periods that are more in line with durations being utilized by exercising individuals.
Laboratory-based research investigations into wearable technology devices have generally utilized specific thresholds for validity measurements, such as ICC above 0.70 and/or MAPE lower than 5% [7]. While these thresholds are acceptable for controlled environments, wearable devices are being used in a variety of applications that do not offer the same constraints. While we believe that researchers and consumers can make their own determination regarding the margin of error that is acceptable to their specific application, the findings of the current investigation indicate that PPG-based devices, regardless of location on the body, display poor heart rate validity during variable intensity running in a natural environment. Until technological advances occur in PPG-based devices allowing for acceptable agreement, heart rate in outdoor environments should be obtained using an ECG-based chest strap that can be connected to a wristwatch or other comparable receiver.
References
- 1. Thompson WR. Worldwide survey of fitness trends for 2020. ACSM’s Health & Fitness Journal. 2019;23(6):10–8.
- 2. Woodman JA, Crouter SE, Bassett DR Jr., Fitzhugh EC, Boyer WR. Accuracy of Consumer Monitors for Estimating Energy Expenditure and Activity Type. Med Sci Sports Exerc. 2017;49(2):371–7. Epub 2016/09/01. pmid:27580155.
- 3. Navalta JW, Montes J, Bodell NG, Aguilar CD, Lujan A, Guzman G, et al. Wearable Device Validity in Determining Step Count During Hiking and Trail Running. Journal for the Measurement of Physical Behaviour. 2018;1(2):86–93.
- 4. Jo E, Lewis K, Directo D, Kim MJ, Dolezal BA. Validation of Biofeedback Wearables for Photoplethysmographic Heart Rate Tracking. J Sports Sci Med. 2016;15(3):540–7. Epub 2016/11/03. pmid:27803634
- 5. Liguori G, Kennedy DJ, Navalta JW. Fitness Wearables. ACSM’s Health & Fitness Journal. 2018;22(6):6–8.
- 6. Bunn JA, Navalta JW, Fountaine CJ, Reece JD. Current State of Commercial Wearable Technology in Physical Activity Monitoring 2015–2017. Int J Exerc Sci. 2018;11(7):503–15. Epub 2018/03/16. pmid:29541338
- 7. Navalta JW, Ramirez GG, Maxwell C, Radzak KN, McGinnis GR. Validity and Reliability of Three Commercially Available Smart Sports Bras during Treadmill Walking and Running. Scientific Reports. 2020;10(1):7397. pmid:32355249
- 8. Bunn JA, Wells E, Manor J, Wenster M. Evaluation of earbud and wristwatch heart rate monitors during aerobic and resistance training. International Journal of Exercise Science. 2019;12(4):374. pmid:30899350
- 9. Boudreaux BD, Hebert EP, Hollander DB, Williams BM, Cormier CL, Naquin MR, et al. Validity of Wearable Activity Monitors during Cycling and Resistance Exercise. Med Sci Sports Exerc. 2018;50(3):624–33. Epub 2017/12/01. pmid:29189666.
- 10. Stahl SE, An HS, Dinkel DM, Noble JM, Lee JM. How accurate are the wrist-based heart rate monitors during walking and running activities? Are they accurate enough? BMJ Open Sport Exerc Med. 2016;2(1):e000106. Epub 2016/12/03. pmid:27900173
- 11. Gillinov S, Etiwy M, Wang R, Blackburn G, Phelan D, Gillinov AM, et al. Variable Accuracy of Wearable Heart Rate Monitors during Aerobic Exercise. Medicine and science in sports and exercise. 2017;49(8):1697–703. pmid:28709155
- 12. Bai Y, Hibbing P, Mantis C, Welk GJ. Comparative evaluation of heart rate-based monitors: Apple Watch vs Fitbit Charge HR. J Sports Sci. 2018;36(15):1734–41. Epub 2017/12/07. pmid:29210326.
- 13. Brazendale K, Decker L, Hunt ET, Perry MW, Brazendale AB, Weaver RG, et al. Validity and Wearability of Consumer-based Fitness Trackers in Free-living Children. Int J Exerc Sci. 2019;12(5):471–82. Epub 2019/03/23. pmid:30899354
- 14. Wallen MP, Gomersall SR, Keating SE, Wisloff U, Coombes JS. Accuracy of Heart Rate Watches: Implications for Weight Management. PLoS One. 2016;11(5):e0154420. Epub 2016/05/28. pmid:27232714
- 15. Thiebaud RS, Funk MD, Patton JC, Massey BL, Shay TE, Schmidt MG, et al. Validity of wrist-worn consumer products to measure heart rate and energy expenditure. Digit Health. 2018;4:2055207618770322. Epub 2018/06/27. pmid:29942628
- 16. Spierer DK, Rosen Z, Litman LL, Fujii K. Validation of photoplethysmography as a method to detect heart rate during rest and exercise. J Med Eng Technol. 2015;39(5):264–71. Epub 2015/06/27. pmid:26112379.
- 17. Dooley EE, Golaszewski NM, Bartholomew JB. Estimating Accuracy at Exercise Intensities: A Comparative Study of Self-Monitoring Heart Rate and Physical Activity Wearable Devices. JMIR Mhealth Uhealth. 2017;5(3):e34. Epub 2017/03/18. pmid:28302596
- 18. Stove MP, Haucke E, Nymann ML, Sigurdsson T, Larsen BT. Accuracy of the wearable activity tracker Garmin Forerunner 235 for the assessment of heart rate during rest and activity. J Sports Sci. 2019;37(8):895–901. Epub 2018/10/18. pmid:30326780.
- 19. Parak J, Uuskoski M, Machek J, Korhonen I. Estimating Heart Rate, Energy Expenditure, and Physical Performance With a Wrist Photoplethysmographic Device During Running. JMIR Mhealth Uhealth. 2017;5(7):e97. Epub 2017/07/27. pmid:28743682
- 20. Dondzila CJ, Lewis CA, Lopez JR, Parker TM. Congruent accuracy of wrist-worn activity trackers during controlled and free-living conditions. International journal of exercise science. 2018;11(7):575–84.
- 21. Reddy RK, Pooni R, Zaharieva DP, Senf B, El Youssef J, Dassau E, et al. Accuracy of Wrist-Worn Activity Monitors During Common Daily Physical Activities and Types of Structured Exercise: Evaluation Study. JMIR Mhealth Uhealth. 2018;6(12):e10338. Epub 2018/12/12. pmid:30530451
- 22. Montes J, Tandy R, Young J, Lee S-P, Navalta J. A Comparison of Multiple Wearable Technology Devices Heart Rate and Step Count Measurements During Free Motion and Treadmill Based Measurements. International Journal of Kinesiology and Sports Science. 2019;7(2):30–9.
- 23.
Estepp JR, Blackford EB, Meier CM. Recovering pulse rate during motion artifact with a multi-imager array for non-contact imaging photoplethysmography. IEEE Int Conf Syst Man Cybernetics (SMC)2014. p. 1462–9.
- 24. Manning JW, Montes J, Stone TM, Rietjens RW, Young JC, DeBeliso M, et al. Cardiovascular and Perceived Exertion Responses to Leisure Trail Hiking. Journal of Outdoor Recreation, Education, and Leadership. 2015;7(2):83–92.
- 25. Navalta JW, Bodell NG, Tanner EA, Aguilar CD, Radzak KN. Effect of exercise in a desert environment on physiological and subjective measures. Int J Environ Health Res. 2019:1–11. Epub 2019/06/27. pmid:31240953.
- 26. Navalta JW, Montes J, Bodell NG, Aguilar CD, Radzak K, Manning JW, et al. Reliability of Trail Walking and Running Tasks Using the Stryd Power Meter. Int J Sports Med. 2019;40(8):498–502. Epub 2019/07/10. pmid:31288288.
- 27. Tanner EA, Montes J, Manning JW, Taylor J, DeBeliso M, Young JC, et al. Validation of Hexoskin biometric shirt to COSMED K4 b2 metabolic unit in adults during trail running. Sports Technology. 2016;8(3–4):118–23.
- 28. Navalta JW, Radzak KN, Montes J, Tanner EA, Bodell NG, Manning JW. Prediction of 5km Trail Race Performance from a Shorter Distance Trail Run. Journal of Biology of Exercise. 2018;14(1):23–30. https://doi.org/10.4127/jbe.2018.0127.
- 29. Montes J, Stone TM, Manning JW, McCune D, Tacad DK, Young JC, et al. Using Hexoskin Wearable Technology to Obtain Body Metrics During Trail Hiking. Int J Exerc Sci. 2015;8(4):425–30. Epub 2015/01/01. pmid:27293503
- 30. Navalta JW, Montes J, Tanner EA, Bodell NG, Young JC. Sex and Age Differences in Trail Half Marathon Running. Int J Exerc Sci. 2018;11(6):281–9. Epub 2018/05/26. pmid:29795742
- 31. Hoffman MD, Wegelin JA. The Western States 100-Mile Endurance Run: participation and performance trends. Med Sci Sports Exerc. 2009;41(12):2191–8. Epub 2009/11/17. pmid:19915499
- 32. Riebe D, Franklin BA, Thompson PD, Garber CE, Whitfield GP, Magal M, et al. Updating ACSM’s Recommendations for Exercise Preparticipation Health Screening. Med Sci Sports Exerc. 2015;47(11):2473–9. Epub 2015/10/17. pmid:26473759.
- 33. Zou GY. Confidence interval estimation for the Bland-Altman limits of agreement with multiple observations per individual. Stat Methods Med Res. 2013;22(6):630–42. Epub 2011/06/28. pmid:21705434.
- 34. Lin LI. A concordance correlation coefficient to evaluate reproducibility. Biometrics. 1989;45(1):255–68. Epub 1989/03/01. pmid:2720055.
- 35.
Baumgartner TA, Jackson AS, Mahar MT. Measurement for Evaluation in Physical Education and Exercise Science. 8th ed. New York, NY: McGraw-Hill; 2007.
- 36. Bunn JA, Jones C, Oliviera A, Webster MJ. Assessment of step accuracy using the Consumer Technology Association standard. Journal of sports sciences. 2019;37(3):244–8. pmid:29958058
- 37.
Nag N, Pandey V, Putzel PJ, Bhimaraju H, Krishnan S, Jain R, editors. Cross-modal health state estimation. Proceedings of the 26th ACM international conference on Multimedia; 2018.
- 38. Fazackerley LA, Fell JW, Kitic CM. The effect of an ultra-endurance running race on heart rate variability. European journal of applied physiology. 2019;119(9):2001–9. pmid:31321510
- 39. Divis K, Anderson-Bergman C, Abbott R, Newton V, Emmanuel-Aviña G. Physiological and cognitive factors related to human performance during the grand canyon rim-to-rim hike. Journal of Human Performance in Extreme Environments. 2018;14(1):5.
- 40. Groslambert A, Baron B, Ouvrard T, Desmoulins L, Lacroix E, Gimenez P, et al. Influencing Factors of Pacing Variations and Performance in a 44-Kilometer Mountain Trail Race. Advances in Physical Education. 2020;10(2):81–96.
- 41.
Bassano C, Chessa M, Fengone L, Isgro L, Solari F, Spallarossa G, et al. Evaluation of a Virtual Reality System for Ship Handling Simulations. Proceedings of the 14th International Joint Conference on Computer Vision, Imaging and Computer Graphics Theory and Applications (VISIGRAPP 2019). 2019:62–73.
- 42. Moreno A, Poppe R, Gibson JL, Heylen D. Automated and unobtrusive measurement of physical activity in an interactive playground. International Journal of Human-Computer Studies. 2019;129:55–63.
- 43. Wu B-F, Chu Y-W, Huang P-W, Chung M-L. Neural Network Based Luminance Variation Resistant Remote-Photoplethysmography for Driver’s Heart Rate Monitoring. IEEE Access. 2019;7:57210–25.
- 44. Neufeld EV, Wadowski J, Boland DM, Dolezal BA, Cooper CB. Heart Rate Acquisition and Threshold-Based Training Increases Oxygen Uptake at Metabolic Threshold in Triathletes: A Pilot Study. International journal of exercise science. 2019;12(2):144. pmid:30761193
- 45. Carrier B, Holmes T, Williams L, Dahl S, Weber L, Creer A, et al. Validation of Garmin Fitness Tracker Biomechanics: 190: Board# 28 May 29 9: 30 AM-11: 00 AM. Med Sci Sports Exerc. 2019;51(5):S28.
- 46. Jolley B, Carrier B, Standifird T, Creer A. Validation of Garmin Fitness Tracker Metabolic Data (VO2max): 191: Board# 29 May 29 9: 30 AM-11: 00 AM. Medicine & Science in Sports & Exercise. 2019;51(6):S28.
- 47. Düking P, Giessing L, Frenkel MO, Koehler K, Holmberg HC, Sperlich B. Wrist-Worn Wearables for Monitoring Heart Rate and Energy Expenditure While Sitting or Performing Light-to-Vigorous Physical Activity: Validation Study. JMIR Mhealth Uhealth. 2020;8(5):e16716. Epub 2020/05/07. pmid:32374274
- 48.
Al Ahmad M, Ahmed S, editors. Piezologist: a novel wearable piezoelectric-based cardiorespiratory monitoring system. 2018 Innovations in Intelligent Systems and Applications (INISTA); 2018: IEEE.
- 49.
Boldu R, Dancu A, Matthies DJ, Cascón PG, Ransir S, Nanayakkara S, editors. Thumb-In-Motion: Evaluating Thumb-to-Ring Microgestures for Athletic Activity. Proceedings of the Symposium on Spatial User Interaction; 2018.
- 50. Marinescu R. Multi-criteria analysis of wearable activityfitness trackers and 3D concept models of smart jewellery. MATEC Web Conf. 2018;178:05017.
- 51.
Santarelli L, Diyakonova O, Betti S, Esposito D, Castro E, Cavallo F, editors. Development of a Novel Wearable Ring-Shaped Biosensor. 2018 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC); 2018: IEEE.
- 52. Leboeuf SF, Aumer ME, Kraus WE, Johnson JL, Duscha B. Earbud-based sensor for the assessment of energy expenditure, HR, and VO2max. Med Sci Sports Exerc. 2014;46(5):1046–52. Epub 2014/04/20. pmid:24743110
- 53. Leader N, Dorian P, Lam J, Lee C, Woo A, Chow C. Evaluation of heart rate trackers in patients with atrial fibrillation. Canadian Journal of Cardiology. 2018;34(10):S152–S3.
- 54. Trivedi NS, Ghouri AF, Shah NK, Lai E, Barker SJ. Effects of motion, ambient light, and hypoperfusion on pulse oximeter function. J Clin Anesth. 1997;9(3):179–83. Epub 1997/05/01. pmid:9172022.
- 55.
Association CT. Physical activity monitoring for heart rate. 2018. p. 1–12.