Dynamic pulmonary CT perfusion using first-pass analysis technique with only two volume scans: Validation in a swine model

  • Yixiao Zhao,

    Roles Conceptualization, Formal analysis, Investigation, Methodology, Software, Writing – original draft, Writing – review & editing

    Affiliation Department of Radiological Sciences, University of California Irvine, Irvine, California, United States of America

  • Logan Hubbard,

    Roles Data curation, Formal analysis, Investigation, Methodology, Writing – review & editing

    Affiliation Department of Radiological Sciences, University of California Irvine, Irvine, California, United States of America

  • Shant Malkasian,

    Roles Formal analysis, Software, Writing – review & editing

    Affiliation Department of Radiological Sciences, University of California Irvine, Irvine, California, United States of America

  • Pablo Abbona,

    Roles Investigation, Validation, Writing – review & editing

    Affiliation Department of Radiological Sciences, University of California Irvine, Irvine, California, United States of America

  • Sabee Molloi

    Roles Conceptualization, Methodology, Supervision, Writing – review & editing

    symolloi@uci.edu

    Affiliation Department of Radiological Sciences, University of California Irvine, Irvine, California, United States of America

Abstract

Purpose

To evaluate the accuracy of a low-dose first-pass analysis (FPA) CT pulmonary perfusion technique in comparison to fluorescent microsphere measurement as the reference standard.

Method

The first-pass analysis CT perfusion technique was validated in six swine (41.7 ± 10.2 kg) for a total of 39 successful perfusion measurements. Different perfusion conditions were generated in each animal using serial balloon occlusions in the pulmonary artery. For each occlusion, over 20 contrast-enhanced CT images were acquired within one breath (320 x 0.5mm collimation, 100kVp, 200mA or 400mA, 350ms gantry rotation time). All volume scans were used for maximum slope model (MSM) perfusion measurement, but only two volume scans were used for the FPA measurement. Both MSM and FPA perfusion measurements were then compared to the reference fluorescent microsphere measurements.

Results

The mean lung perfusion of MSM, FPA, and microsphere measurements were 6.21 ± 3.08 (p = 0.008), 6.59 ± 3.41 (p = 0.44) and 6.68 ± 3.89 ml/min/g, respectively. The MSM (PMSM) and FPA (PFPA) perfusion measurements were related to the corresponding reference microsphere measurement (PMIC) by PMSM = 0.51PMIC + 2.78 (r = 0.64) and PFPA = 0.79PMIC + 1.32 (r = 0.90). The root-mean-square-error for the MSM and FPA techniques were 3.09 and 1.72 ml/min/g, respectively. The root-mean-square-deviation for the MSM and FPA techniques were 2.38 and 1.50 ml/min/g, respectively. The CT dose index for MSM and FPA techniques were 138.7 and 8.4mGy, respectively.

Conclusions

The first-pass analysis technique can accurately measure regional pulmonary perfusion and has the potential to reduce the radiation dose associated with dynamic CT perfusion for assessment of pulmonary disease.

Introduction

The advent of quantitative computed tomography (CT) imaging techniques has enabled better evaluation of pulmonary disease. Multidetector CT has previously been used for detection of emphysema [1, 2] and the development of the dual-energy CT has improved the differentiation between different contrast materials (i.e. iodine or xenon) and lung parenchyma for assessment of regional perfusion or ventilation [3, 4]. With the capability of imaging regional pulmonary blood volume, CT has been clinically implemented, thus contributing to assessment for pulmonary embolism, chronic obstructive pulmonary disease [5, 6] and lung cancer [79].

Despite positive correlation with microsphere perfusion measurement, existing dynamic CT perfusion techniques have been hampered by a number of limitations [1014]. First, CT perfusion techniques such as the maximum slope model (MSM), derive pulmonary blood flow by monitoring the contrast enhancement over time using several small volumes-of-interest (VOIs) [13]. However, this approach can be limited by short contrast transit time (<1 sec) through the small VOIs, which are inherently subject to contrast material loss from the small compartments over the measurement period. Second, curve-fitting is performed on these time-series data, which requires a large number of consecutive acquisitions for pulmonary blood flow measurement leading to high radiation dose and motion misregistration artifacts [79, 13, 1517]. Thus, the limitations of the existing perfusion models have hampered the widespread use of dynamic CT for assessment of pulmonary disease.

The purpose of this study was to validate a new low-dose dynamic CT perfusion technique based on first-pass analysis (FPA) for perfusion measurement in a swine model, with the MSM technique also implemented for comparison. The FPA CT perfusion technique has previously been validated for myocardial perfusion measurement in phantom and animal studies [1820] and addressed the underestimation of the flow measurement by imaging the whole heart within each cardiac cycle. The use of the whole organ as the perfusion compartment allows adequate time for image acquisition and contrast accumulation before contrast exits the perfusion compartment[21, 22]. By increasing the perfusion compartment within the spatial domain, it is possible to reduce the temporal sampling to a minimum of only two time points for FPA measurement. The central hypothesis is that pulmonary perfusion can be accurately measured with the CT FPA technique using only two volume scans, as compared to fluorescent microsphere perfusion method as the reference standard [23].

Material and methods

Ethics statement

The experimental protocol was approved by the Institutional Animal Care and Use Committee (IACUC, Protocol Number: AUP-18-191) at University of California, Irvine. All studies were performed under isoflurane anesthesia, and all efforts were made to minimize suffering.

General methods

Each swine was housed in a single cage in the vivarium and food was withdrawn but water was provided for 12 hours before the study. A total of six swine (41.7 ± 10.2 kg, male, Yorkshire) were used for validation of the FPA technique with fluorescent microsphere perfusion measurements. A standard MSM dynamic CT perfusion technique was also used for comparison. All experimental data were successfully acquired between June 2016 and July 2017 and analyzed between September 2016 and December 2018. All authors conducted the experiments and data acquisition. Three authors (Y.Z., L.H., S. Malkasian.) with more than 3 years of medical imaging research experience conducted the data analysis. One author (P.A.) with 19 years of clinical radiology experience, helped with the intubation, anesthesia, catheters placement and other interventional procedures.

First-pass analysis (FPA) technique

In this study, 20 dynamic CT volume scans were acquired for the MSM technique, but only two volume scans were systematically chosen for the FPA perfusion technique. Specifically, the first volume acquisition (V1) was chosen as the contrast started to enter the compartment with a threshold of 80 HU above the background blood pool enhancement, and the second volume scan (V2) was chosen near the peak of the pulmonary arterial input function as shown in Fig 1. The entire lung or lobe was used as the first-pass compartment in the pulmonary perfusion measurement. Assuming no contrast material outflow from the measurement compartment over the measurement period, the lung tissue compartment time attenuation curve (TAC) was used to calculate the change in the integrated contrast mass (ΔMc/Δt) between the first (V1) and second (V2) volume scans for the blood flow (Qave, ml/min) measurement, as shown in Eq 1. The average incoming contrast material concentration Cin_ave, mg/ml in the pulmonary artery was estimated using the two volume scans as the arterial input function (AIF) [20].

thumbnail
Fig 1. The first-pass analysis acquisition protocol.

The lung tissue compartment time attenuation curve (TAC) was used to calculate the change in the integrated contrast mass (ΔMc/Δt) between the first (V1) and second (V2) volume scans. The pulmonary arterial input function (AIF) was used to estimate the average input concentration (Cin_ave) between V1 and V2. The first volume scan (V1) was acquired with a threshold of 80 HU above the background blood pool enhancement.

https://doi.org/10.1371/journal.pone.0228110.g001

(1)

The details of voxel-by-voxel pulmonary perfusion measurement are shown in the S1 File.

Animal preparation

Each swine was premedicated via intramuscular injection of Telazol (4.4mg/kg), Ketamine (2.2 mg/kg), and Xylazine (2.2 mg/kg), intubated (Mallinckrodt, tube 6.0/6.5/7.0, Covidien, Mansfield, MA), then ventilated (Surgivet, Norwell, MA, and Highland Medical Equipment, Temecula, CA) with 1.5–2.5% Isoflurane (Baxter, Deerfield, IL) to maintain anesthesia. Under ultrasound guidance (9L Transducer, Vivid E9, GE 145 Healthcare), four introducer sheaths (AVANTIR, Cordis Corporation, Miami Lakes, FL) were placed in the left jugular vein (intravenous contrast injection), left femoral vein (intravenous fluid and drug administration), right femoral vein (intravenous microsphere injection), and the right femoral artery (mean arterial pressure monitoring), respectively. Two more introducer sheaths were placed in the right jugular vein for the insertion of a pulmonary arterial Swan-Ganz-balloon catheter and for reference microsphere blood withdrawal catheters. Specifically, under fluoroscopic guidance, the Swan-Ganz catheter was placed distally in the left-caudal lobe pulmonary artery branch for induction of balloon occlusion. An additional pulmonary arterial catheter was placed proximally in the left main pulmonary arterial trunk for blood sample withdrawal for microsphere perfusion measurement. Several perfusion conditions were then induced by inflating a balloon in several different locations in the left caudal lobe. ECG, mean arterial pressure (mmHg), end-tidal CO2 (mmHg), and O2 saturation (%) were monitored and recorded. After the final acquisition, each animal was euthanized with 10 cc of saturated KCL under deep anesthesia and both lungs were resected without draining the blood for tissue sample extraction.

Reference fluorescent microsphere perfusion measurement

Fig 2 shows a summary of data acquisition protocol. The protocol includes vascular access using fluoroscopy, dynamic CT acquisition, microsphere measurement, and postmortem tissue processing. Immediately before each CT perfusion acquisition, one color of fluorescent-labeled microspheres (NuFLOW™, IMT Laboratories, Lawrence, KS) was diluted and injected into the vena cava. A reference blood sample was then withdrawn using a syringe pump at 10ml/min via a pulmonary arterial catheter proximal to the balloon occlusion (GenieTouch; Kent Scientific, Torrington, Conn). The pump was turned on 5 seconds prior to the microsphere injection and was run for a total of 2 minutes. A total of 5 to 8 different color microspheres were used for each animal. Each color of microsphere was used for an independent CT perfusion acquisition. After all acquisitions were completed, the animal was euthanized and 6 to 8 tissue samples (approximately 10g per sample) were excised from different lobes of both lungs. Microsphere (15.5±1.10μm in diameter) analyses were conducted independently by IMT Laboratories [23].

thumbnail
Fig 2. Data acquisition protocol timeline.

The protocol includes vascular access using fluoroscopy, dynamic CT acquisition, microsphere measurement, and postmortem tissue processing. The contrast and microsphere injections were made before the continuous dynamic CT scan. The blood sample withdrawal was started 5 seconds prior to the microsphere injection and lasted for 2 minutes.

https://doi.org/10.1371/journal.pone.0228110.g002

CT scanning protocol

Each animal was first positioned supine, head first through the gantry (320-slice Aquilion One, Canon Medical Systems, Tustin, CA), and a localization scan was performed with a 16-cm z-coverage to include the balloon placed in the base of the lung. Balloon occlusion was then induced followed by contrast (0.5 or 1 ml/kg, Isovue 370, Bracco Diagnostics, Princeton, NJ) and saline chaser (0.25 or 0.5 ml/kg) injection at 5 ml/s (Empower CTA, Acist Medical Systems, Eden Prairie, MN). For each occlusion, the ventilator was turned off to simulate a breath hold, and ECG-gated volume acquisitions were acquired at functional residual capacity for approximately 20 seconds (100kVp, 200mA or 400mA, 0.35s rotation time, 320 x 0.5 mm collimation, FOV: 240-400mm). Detailed injection and scan parameters are summarized in Table 1. Dynamic CT perfusion acquisition was repeated approximately every 20 minutes. All volume acquisitions were reconstructed using 360 degree projection data using a soft tissue kernel (FC07) with a 0.5-mm slice thickness and an adaptive iterative dose reduction 3D (AIDR 3D) reconstruction algorithm. The CT dose index was recorded and the size-specific dose estimate (SSDE) was also calculated to account for the small effective chest diameter of the swine used in the study [24, 25]. In combination with the chest conversion factor, the effective dose for perfusion measurement with the FPA technique was also estimated [26].

thumbnail
Table 1. Acquisition protocols and dose metrics for the FPA and MSM perfusion measurement.

https://doi.org/10.1371/journal.pone.0228110.t001

FPA perfusion measurement

Dynamic volume acquisitions were first registered to a single coordinate system by applying a GPU-based affine and deformable registration algorithm [27]. The volume-of-interest was then placed in the pulmonary artery to derive the pulmonary arterial input function (AIF, Fig 1). For pulmonary flow measurement, only two volume scans (V1 and V2) were systematically and retrospectively selected from the AIF, where V1 was defined as the first volume scan at 80 HU above the baseline blood pool enhancement as the indication of contrast arrival time, and V2 was defined as the peak of the AIF (Fig 1). Lung segmentation (lung parenchyma in the FOV) was conducted based on V1. The contrast mass change (ΔMc) in the lung parenchyma and the average arterial input concentration (Cin_ave) were used to derive the pulmonary blood flow (A.1–A.3). Finally, to calculate perfusion, the average blood flow was normalized to each voxel (A.4) and the non-air tissue mass for each voxel was calculated based on V1 (A.5–A.6). The final pulmonary perfusion per voxel was calculated as the voxel flow normalized by voxel mass (A.7).

In order to derive regional pulmonary perfusion measurements for the validation with microsphere measurements, digital segmentations (ViTAL Images; Canon Medical Systems) were made by a chest radiologist (P.A.) with 19 years of clinical experience, to correspond with the excised tissue samples used for the reference microsphere perfusion measurements. Anatomical landmarks were used as a guide for correspondence between excised tissue samples and digital segmentation. Blood vessels were excluded by excluding voxels greater than 100 HU [1, 28]. The average CT perfusion within each digital segment was compared to the corresponding reference fluorescent microsphere perfusion measurement.

MSM perfusion measurement

Maximum slope model (MSM) perfusion data were generated using a standard commercial software (ViTAL Images, CT Body Perfusion 4D, Dual-Input Lung Workflow). After automatic registration, small regions-of-interest (ROIs) were placed within the pulmonary trunk and lung parenchyma to generate arterial input functions and tissue attenuation curve, respectively. A parametric pulmonary perfusion map (ml/min/100ml) was then generated. The perfusion maps were rescaled into ml/min/g by using the same mass distribution map (Mx, y, z, g/ml, A.6) as the FPA perfusion measurement. Regional perfusion measurements were also averaged using the same regional segmentations as the FPA measurement.

Statistical analyses

For all microsphere perfusion measurements (including all tissue samples under all perfusion conditions), the combined variance within each subject was compared with the combined variance of measurements between subjects with intra-cluster correlation. To evaluate the performance of the FPA and MSM techniques, linear regression (Pearson’s r) and Bland-Altman analyses were performed as compared to the reference fluorescent microsphere perfusion measurements. The root-mean-square error (RMSE) (mean prediction error between the measurement and the gold standard measurement), the root-mean-square deviation(RMSD) (mean prediction error between the measurement and the line of best fit in the regression analysis) and Lin’s concordance correlation coefficient [29] (concordance between the measurement and the gold standard measurement) were also determined. Student’s t-tests were used to assess mean differences of the FPA and MSM techniques with the reference microsphere perfusion measurements. Statistical software (SPSS, version 22, IBM, Armonk, NY) was used for all statistical analyses.

Results

Vitals and statistics

The FPA CT perfusion technique was validated in six swine (41.7 ± 10.2 kg) for a total of 44 perfusion measurements. Five acquisitions were excluded from the dataset since reference microsphere measurements were not available. Among the 39 successful perfusion measurements, 287 segments from different lobes of both left and right lungs (7 segments per animal on average) were used in the data analysis, as shown in Table 2.

thumbnail
Table 2. FPA and MSM mean pulmonary perfusion versus reference microsphere mean perfusion.

https://doi.org/10.1371/journal.pone.0228110.t002

The overall heart rate and mean arterial pressure during all the CT perfusion acquisitions were 94.7 ± 18.9 bpm and 68.8 ± 22.8 mmHg, respectively. The intra-cluster correlation was calculated to be r = 0.19, indicating negligible correlation within intra-cluster measurements and statistical independence [30]. The effective sample size was found to be 28, showing that 28 independent individual subjects were equivalently used in this study [31].

Lung mass validation

The average CT-measured lung mass and the actual ex-vivo mass were 455.4g (372.5 – 496.0g) and 434.3g (347.2 – 487.4g), respectively (p = 0.20, RMSE = 38.2g).

Qualitative validation

No significant image quality differences were found between images acquired using 400 and 200mA. Representative FPA and MSM pulmonary perfusion maps under 4 different balloon occlusion conditions acquired at 400 mA are shown in Figs 3 and 4. Additional FPA pulmonary perfusion maps with and without balloon occlusion acquired at 200 mA are shown in S1 Fig.

thumbnail
Fig 3. FPA pulmonary perfusion maps for different balloon occlusions.

Axial, coronal, and 3D posterior-lateral views in the presence of no occlusion (first row), a distal occlusion (second row), a distal-to-proximal occlusion (third row) and a proximal occlusion (fourth row) are shown. The arrows indicate the location of the balloon for each occlusion. The color bar indicates perfusion in the range of 0–20 ml/min/g. The images were acquired at 400 mA.

https://doi.org/10.1371/journal.pone.0228110.g003

thumbnail
Fig 4. MSM pulmonary perfusion maps for different balloon occlusions.

Axial, coronal, and sagittal views in the presence of no occlusion (first row), a distal occlusion (second row), a distal-to-proximal occlusion (third row) and a proximal occlusion (fourth row) are shown. The color bar indicates perfusion in the range of 0–600 ml/min/100ml. The images were acquired at 400 mA.

https://doi.org/10.1371/journal.pone.0228110.g004

Quantitative analysis

For all segments combined, the mean lung perfusion of the FPA and MSM techniques were 6.59 ± 3.41 (p = 0.44) and 6.21 ± 3.08 (p = 0.008) ml/min/g, respectively, while the mean microsphere perfusion was 6.68 ± 3.89 ml/min/g. Detailed pulmonary perfusion comparisons (n = 287) are summarized in Tables 2 and 3. Since the balloon occluded the arterial branch in the left caudal lobe, a reduction in perfusion can be found in the occluded regions in the left caudal anterior lobe (n = 39, FPA: 4.65 ± 2.98 vs. microsphere: 4.55 ± 2.95 in ml/min/g) and left caudal posterior lobe (n = 28, FPA: 2.76 ± 3.57 vs. microsphere: 2.17 ± 3.37 in ml/min/g) as compared to the other lobes. Except for the occluded regions, the mean perfusion values gradually decreased from the caudal lobe to the cranial lobe, indicating the change in the lung perfusion by gravitational effects, as shown in Table 2. Significant differences from the middle lobes, cranial lobes are mainly caused by the beam hardening artifacts by the contrast in vena cava and heart.

thumbnail
Table 3. FPA and MSM pulmonary perfusion regression versus reference microsphere perfusion.

https://doi.org/10.1371/journal.pone.0228110.t003

FPA perfusion measurements (PFPA) were related to the reference fluorescent microsphere measurements (PMIC) by PFPA = 0.79PMIC+1.32 (r = 0.90), with a concordance correlation coefficient (CCC) of 0.89, a root-mean-square deviation (RMSD) of 1.50 ml/min/g and a root-mean-square error (RMSE) of 1.72 ml/min/g (Fig 5A). MSM perfusion measurements (PMSM) were related to reference fluorescent microsphere perfusion measurements by PMSM = 0.51PMIC+2.78 (r = 0.64), with a CCC of 0.62, RMSD of 2.38 ml/min/g and RMSE of 3.09 ml/min/g (Fig 5C). Both FPA and MSM measurements had a downward trend in the Bland-Altman plot (Fig 5B and 5D), but the MSM technique showed worse correspondence at higher perfusion measurements. With respect to the two different tube currents, the overall measurement error was slightly lower for high tube currents as compared with the low tube currents. Specifically, for 400mA data, the RMSE and RMSD for the FPA technique were 1.37 and 1.29 ml/min/g, for the MSM technique were 2.69 and 2.14 ml/min/g, respectively. For 200 mA data, the RMSE and RMSD for the FPA technique were 2.34 and 1.92 ml/min/g, for the MSM technique were 3.87 and 2.87 ml/min/g, respectively.

thumbnail
Fig 5. Quantitative analysis of FPA and MSM perfusion measurements.

(a,c) Regression analysis comparing the result of the FPA and MSM perfusion measurements to the corresponding reference fluorescent microsphere perfusion measurements. (b,d) Bland-Altman analysis was performed for both techniques, with the limits of agreement. PFPA: FPA perfusion measurements, PMIC: Reference fluorescent microsphere measurements, PMSM: MSM perfusion measurements. RMSD = root-mean-square deviation, RMSE = root-mean-square error, CCC = concordance correlation coefficient.

https://doi.org/10.1371/journal.pone.0228110.g005

The repeatability analysis of the FPA and MSM perfusion measurements are shown in Fig 6. Two repeated contrast injections were made in two animals with the same occlusion. The interval between the two acquisitions, the average change in heart rate and blood pressure were 15 minutes, 2.18% and 12.75%, respectively. The results of the first (PT1) and the second (PT1) measurements were related by PT1_FPA = 0.94 PT2_FPA + 0.32 (r = 0.99, RMSE: 0.60 ml/min/g, RMSD: 0.17 ml/min/g) for the FPA technique and PT1_MSM = 1.08 PT2_MSM− 0.54 (r = 0.97, RMSE: 0.86 ml/min/g, RMSD: 0.23 ml/min/g) for the MSM technique (Fig6A and 6C). The repeatability for FPA and MSM measurements had negligible bias and lower variability (Fig 6B and 6D) as compared to the accuracy measurements (Fig 5).

thumbnail
Fig 6. Repeatability analysis of FPA and MSM perfusion measurments.

Two repeated contrast injections were made in two animals under the same perfusion condition. (a,c) Regression analysis comparing the result of the Trial #1 to Trial #2 using the FPA and MSM perfusion measurements. (b,d) Bland-Altman analysis was performed for both techniques, with the limits of agreement. PT1_FPA, PT2_FPA: FPA perfusion measurements, PT1_MSM, PT2_MSM: MSM perfusion measurements. RMSD = root-mean-square deviation, RMSE = root-mean-square error.

https://doi.org/10.1371/journal.pone.0228110.g006

Dose estimation

For the retrospectively acquired perfusion images at 200mA, the CT dose index for the MSM techniques was 138.7mGy. For the FPA technique, if the two volume scans were acquired prospectively, the CTDI would have been 8.4mGy. All other dose metrics are reported in Table 1.

Discussion

The results indicate that the FPA technique can accurately measure pulmonary perfusion on a voxel-by-voxel basis as indicated by the good agreement with reference fluorescent microsphere perfusion measurement. The FPA technique also has high precision for repeated perfusion measurements. Furthermore, the pulmonary perfusion measurement can be made using only two volume scans, indicating significant potential for dose reduction with its prospective implementation. Moreover, the second volume acquisition for the pulmonary perfusion measurement could potentially be used for CT pulmonary angiography; hence, the FPA technique has the potential to provide anatomical and physiological assessment of lung disease following a single contrast injection, while also reducing the radiation dose, although further validation is still necessary.

Previously reported MSM techniques [1214] use small regions-of-interest (ROIs) for perfusion measurements, resulting in underestimation of the perfusion due to the problem of contrast loss during the measurement time. Our results agree with previous reports showing the underestimation of the MSM technique, especially at higher perfusion rates. Deconvolution-based techniques have been used to resolve the underestimation issue by estimating the venous contrast outflow over the measurement time [32, 33]. However, these techniques require the entire contrast-pass curve and involve high radiation doses. On the other hand, for the proposed FPA technique, the expansion of the compartment size to encompass the whole lobe enables the possibility of measuring blood flow using only two volume scans and minimizes the problem of contrast loss over the measurement time, improving the accuracy of the perfusion measurement.

Apart from the measurement accuracy, the prospective implementation of the FPA technique with two-volume scan is the most critical issue to be solved in reducing the radiation dose. A prospective timing protocol for FPA technique has recently been reported for the optimal capture of the two-volume scans at the base and peak of the aortic enhancement [34]. It was shown that V1 can be acquired based on a bolus-tracking technique and V2 can be acquired after a pre-defined time-to-peak delay based on contrast injection time interval. A robust relation between the contrast injection time and the aortic enhancement time-to-peak delay was investigated to predict V2 timing, which can be further studied for the pulmonary system. A previous report has shown that accurate perfusion measurements can be made using the FPA technique for different timing protocols of the first (V1) and second (V2) volume scans [34]. Furthermore, a diluted test-bolus technique [35] was also reported to prospectively determine the timing for the two-volume scans, however, with slightly increased contrast and radiation dose. By injecting the same amount of diluted contrast as the actual contrast bolus, the contrast arrival timing (V1) and maximal enhancement timing (V2) can be monitored and used for the prospective FPA volume scans [35].

The breath-hold difficulty is another limitation in clinical practice due to the length of time required for image acquisition. Previously reported dynamic CT perfusion techniques [4, 79, 12, 13, 15, 33, 36] require sampling most of the contrast pass curve with a large number of volume acquisitions for approximately 30 seconds. Images are often acquired during shallow breathing, resulting in major motion artifacts. Fortunately, given the unique two-volume sampling protocol of the FPA technique, its prospective implementation will only require a short breath hold of approximately 10 seconds. The relatively short breath-hold time is expected to substantially reduce the motion misregistration artifacts and make the FPA technique more feasible for patients with lung disease.

The effective radiation dose of dynamic CT perfusion techniques for volume CT scanners with equivalent 16-cm detector has been reported to be in the range of 4.55-50mSv [79, 13, 1517], using a continuous dynamic volume scanning mode. If implemented prospectively, the effective radiation dose for the two-volume FPA technique is expected to be approximately 3.98mSv at 200mA using the SSDE and a chest conversion factor of 0.014mGy∙cm [26]. Future studies are necessary to optimize the radiation dose associated with the FPA technique, including optimization of tube current, beam energy, and other scanning parameters. Nevertheless, significant potential exists for radiation dose reduction in the proposed dynamic FPA CT perfusion technique.

Limitations

Despite improvements in quantitative pulmonary perfusion, our study had limitations. Firstly, the comparison between the physical tissue samples and the corresponding digital segmentations is challenging. Potential errors were minimized using anatomical landmarks for improved registration during the regional segmentation process. Larger physical sample sizes could help to reduce the measurement variation caused by spatial misregistration and motion. Secondly, images were acquired at a relatively high tube current, which is not necessary for accurate perfusion measurement. Therefore, additional radiation dose optimization studies will be required. Thirdly, highly attenuating contrast material in the vena cava, heart, and aorta generated significant streaking artifacts that impacted all adjacent lobes. These artifacts potentially degrade the CT perfusion measurement accuracy and precision. Further contrast dose reduction and the use of beam hardening correction algorithms [37, 38] can be helpful in reducing such artifacts. Fourthly, the z-axis range was limited to 16 cm, which is smaller than the range necessary for whole lung imaging in human subjects. However, using an individual lobe as the compartment would be sufficient for the FPA model. Finally, image quality degradation is expected in the periphery of the scan range. However, this effect was difficult to quantify since all the measurements were made at approximately the same distance from the iso-center.

Conclusions

The proposed FPA technique can provide pulmonary perfusion measurement using two volume scans. In summary, the first-pass analysis technique for pulmonary perfusion measurement was validated in a swine model and has the potential to reduce the radiation dose associated with dynamic CT perfusion for assessment of pulmonary disease.

Supporting information

S1 Fig. FPA pulmonary perfusion maps for different balloon occlusions.

Axial, coronal, and 3D posterior views in the presence of no occlusion (first row) and a distal occlusion (second row) are shown. The color bar indicates perfusion in the range of 0–20 ml/min/g. Images were acquired at 200 mA.

https://doi.org/10.1371/journal.pone.0228110.s001

(TIF)

S1 File. Appendix for first-pass analysis perfusion technique derivation.

https://doi.org/10.1371/journal.pone.0228110.s002

(DOCX)

S2 File. Raw data spreadsheet for all perfusion measurements.

https://doi.org/10.1371/journal.pone.0228110.s003

(PDF)

Acknowledgments

The authors thank Drs. Erin Angel and Pascal Salazar from Canon America Medical Systems for their technical support in this work.

References

  1. 1. Hoffman EA, Simon BA, McLennan G. State of the Art. A structural and functional assessment of the lung via multidetector-row computed tomography: phenotyping chronic obstructive pulmonary disease. Proceedings of the American Thoracic Society. 2006;3(6):519–32. pmid:16921136
  2. 2. Hoffman EA, Reinhardt JM, Sonka M, Simon BA, Guo J, Saba O, et al. Characterization of the interstitial lung diseases via density-based and texture-based analysis of computed tomography images of lung structure and function. Acad Radiol. 2003;10(10):1104–18. pmid:14587629
  3. 3. Hwang HJ, Hoffman EA, Lee CH, Goo JM, Levin DL, Kauczor H-U, et al. The role of dual-energy computed tomography in the assessment of pulmonary function. Eur J Rfadiol. 2017;86:320–34.
  4. 4. Ferda J, Ferdova E, Mirka H, Baxa J, Bednarova A, Flohr T, et al. Pulmonary imaging using dual-energy CT, a role of the assessment of iodine and air distribution. Eur J Radiol. 2011;77(2):287–93. pmid:19717260
  5. 5. Lee CW, Seo JB, Song JW, Kim MY, Lee HY, Park YS, et al. Evaluation of computer-aided detection and dual energy software in detection of peripheral pulmonary embolism on dual-energy pulmonary CT angiography. Eur Radiol. 2011;21(1):54–62. pmid:20680290
  6. 6. Pontana F, Faivre JB, Remy-Jardin M, Flohr T, Schmidt B, Tacelli N, et al. Lung perfusion with dual-energy multidetector-row CT (MDCT): feasibility for the evaluation of acute pulmonary embolism in 117 consecutive patients. Acad Radiol. 2008;15(12):1494–504. pmid:19000866
  7. 7. Ohno Y, Koyama H, Matsumoto K, Onishi Y, Takenaka D, Fujisawa Y, et al. Differentiation of Malignant and Benign Pulmonary Nodules with Quantitative First-Pass 320-Detector Row Perfusion CT versus FDG PET/CT. Radiology. 2011;258(2):599–609. pmid:21273522
  8. 8. Yuan X, Zhang J, Quan C, Cao J, Ao G, Tian Y, et al. Differentiation of malignant and benign pulmonary nodules with first-pass dual-input perfusion CT. Eur Radiol. 2013;23(9):2469–74. pmid:23793548
  9. 9. Shan F, Zhang Z, Xing W, Qiu J, Yang S, Wang J, et al. Differentiation between malignant and benign solitary pulmonary nodules: use of volume first-pass perfusion and combined with routine computed tomography. Eur J Radiol. 2012;81(11):3598–605. pmid:22608062
  10. 10. Hopkins SR, Wielputz MO, Kauczor HU. Imaging lung perfusion. Journal of Applied Physiology. 2012;113(2):328–39. pmid:22604884
  11. 11. Chon D, Beck KC, Larsen RL, Shikata H, Hoffman EA. Regional pulmonary blood flow in dogs by 4D-X-ray CT. J Appl Physiol (1985). 2006;101(5):1451–65.
  12. 12. Dakin JH, Evans TW, Hansell DM, Hoffman EA. Regional Pulmonary Blood Flow in Humans and Dogs by 4D Computed Tomography. Academic Radiology. 2008;15(7):844–52. pmid:18572119
  13. 13. Fuld MK, Halaweish AF, Haynes SE, Divekar AA, Guo J, Hoffman EA. Pulmonary perfused blood volume with dual-energy CT as surrogate for pulmonary perfusion assessed with dynamic multidetector CT. Radiology. 2013;267(3):747–56. pmid:23192773
  14. 14. Larici AR, Calandriello L, Amato M, Silvestri R, del Ciello A, Molinari F, et al. First-pass perfusion of non-small-cell lung cancer (NSCLC) with 64-detector-row CT: a study of technique repeatability and intra- and interobserver variability. La radiologia medica. 2014;119(1):4–12. pmid:24272480
  15. 15. Ohno Y, Nishio M, Koyama H, Seki S, Tsubakimoto M, Fujisawa Y, et al. Solitary pulmonary nodules: Comparison of dynamic first-pass contrast-enhanced perfusion area-detector CT, dynamic first-pass contrast-enhanced MR imaging, and FDG PET/CT. Radiology. 2015;274(2):563–75. pmid:25203128
  16. 16. Lossnitzer D, Baumann S, Henzler T. Dynamic four-dimensional CT angiography for the assessment of pulmonary perfusion in an adult patient with pulmonary artery occlusion and major aortopulmonary collateral after multistage repair of Fallot's Pentalogy. Cardiology in the young. 2017;27(6):1212–3. pmid:28606197
  17. 17. Nguyen-Kim TD, Frauenfelder T, Strobel K, Veit-Haibach P, Huellner MW. Assessment of bronchial and pulmonary blood supply in non-small cell lung cancer subtypes using computed tomography perfusion. Invest Radiol. 2015;50(3):179–86. pmid:25500892
  18. 18. Ziemer BP, Hubbard L, Lipinski J, Molloi S. Dynamic CT perfusion measurement in a cardiac phantom. Int J Cardiovasc Imaging. 2015;31(7):1451–9. pmid:26156231
  19. 19. Hubbard L, Ziemer B, Lipinski J, Sadeghi B, Javan H, Groves EM, et al. Functional Assessment of Coronary Artery Disease Using Whole-Heart Dynamic Computed Tomographic Perfusion. Circ Cardiovasc Imaging. 2016;9(12).
  20. 20. Hubbard L, Lipinski J, Ziemer B, Malkasian S, Sadeghi B, Javan H, et al. Comprehensive Assessment of Coronary Artery Disease by Using First-Pass Analysis Dynamic CT Perfusion: Validation in a Swine Model. Radiology. 2018;286(1):93–102. pmid:29059038
  21. 21. Molloi S, Bednarz G, Tang J, Zhou Y, Mathur T. Absolute volumetric coronary blood flow measurement with digital subtraction angiography. Int J Cardiovasc Imaging. 1998;14(3):137–45.
  22. 22. Molloi S, Zhou Y, Kassab GS. Regional volumetric coronary blood flow measurement by digital angiography: in vivo validation. Acad Radiol. 2004;11(7):757–66. pmid:15217593
  23. 23. Glenny R, Bernard S, Brinkley M. Validation of fluorescent-labeled microspheres for measurement of regional organ perfusion. Journal of Applied Physiology. 1993;74(5):2585–97. pmid:8335595
  24. 24. Boone J, Strauss K, Cody D, McCollough C, McNitt-Gray M, Toth T, et al. Size-specific dose estimates (SSDE) in pediatric and adult body CT examinations: report of AAPM task group 204. College Park, Md: American Association of Physicists in Medicine. 2011.
  25. 25. McCollough C, Bakalyar DM, Bostani M, Brady S, Boedeker K, Boone JM, et al. Use of Water Equivalent Diameter for Calculating Patient Size and Size-Specific Dose Estimates (SSDE) in CT: The Report of AAPM Task Group 220. AAPM Rep. 2014;2014:6–23.
  26. 26. McCollough C CD, Edyvean S, Geise R, et al. The Measurement, Reporting, and Management of Radiation Dose in CT. 2008.
  27. 27. Modat M, Ridgway GR, Taylor ZA, Lehmann M, Barnes J, Hawkes DJ, et al. Fast free-form deformation using graphics processing units. Comput Methods Programs Biomed. 2010;98(3):278–84. pmid:19818524
  28. 28. Shikata H, Hoffman EA, Sonka M, editors. Automated segmentation of pulmonary vascular tree from 3D CT images. Medical Imaging 2004; 2004: SPIE.
  29. 29. Lin LI. A concordance correlation coefficient to evaluate reproducibility. Biometrics. 1989;45(1):255–68. pmid:2720055
  30. 30. Eldridge SM, Ukoumunne OC, Carlin JB. The intra-cluster correlation coefficient in cluster randomized trials: a review of definitions. Int Stat Rev. 2009;77(3):378–94.
  31. 31. Kerry SM, Bland JM. Sample size in cluster randomisation. BMJ. 1998;316(7130):549. pmid:9501723
  32. 32. Nabavi DG, Cenic A, Dool J, Smith RM, Espinosa F, Craen RA, et al. Quantitative assessment of cerebral hemodynamics using CT: stability, accuracy, and precision studies in dogs. J Comput Assist Tomogr. 1999;23(4):506–15. pmid:10433275
  33. 33. van Elmpt W, Das M, Hullner M, Sharifi H, Zegers K, Reymen B, et al. Characterization of tumor heterogeneity using dynamic contrast enhanced CT and FDG-PET in non-small cell lung cancer. Radiother Oncol. 2013;109(1):65–70. pmid:24044795
  34. 34. Hubbard L, Malkasian S, Zhao Y, Abbona P, Molloi S. Timing optimization of low-dose first-pass analysis dynamic CT myocardial perfusion measurement: validation in a swine model. Eur Radiol Exp. 2019;3(1):16. pmid:30945100
  35. 35. Hubbard L, Malkasian S, Zhao Y, Abbona P, Molloi S. Contrast-to-Noise Ratio Optimization in Coronary Computed Tomography Angiography: Validation in a Swine Model. Acad Radiol. 2018.
  36. 36. Venkat B, Sharma S, Sharma D, Sood S, Aggarwal N, Sarkar M, et al. CT perfusion in non-small cell lung cancers for assessing treatment response, monitoring treatment and predicting prognosis. The Egyptian Journal of Radiology and Nuclear Medicine. 2018;49(2):338–45.
  37. 37. Stenner P, Schmidt B, Allmendinger T, Flohr T, Kachelrie M. Dynamic iterative beam hardening correction (DIBHC) in myocardial perfusion imaging using contrast-enhanced computed tomography. Invest Radiol. 2010;45(6):314–23. pmid:20440212
  38. 38. Kitagawa K, George RT, Arbab-Zadeh A, Lima JA, Lardo AC. Characterization and correction of beam-hardening artifacts during dynamic volume CT assessment of myocardial perfusion. Radiology. 2010;256(1):111–8. pmid:20574089