Figures
Abstract
To investigate the undesired irradiations outside of the treatment field by electron streams in air (air-electron-stream) during magnetic-resonance image-guided radiation therapy (MR-IGRT). A custom-made support phantom adjusting angles between the beam central axis (CAX) and the phantom surface (termed phantom-angles), were used. Using the ViewRay system, a rectangular parallelepiped phantom placed on the support phantom, was irradiated with field sizes of 6.3 cm × 6.3 cm (FS6.3) and 12.6 cm × 12.6 cm (FS12.6) at gantry angles of 0°, 30°, and 330°, and phantom-angles of 10°, 20°, and 30°. For each beam delivery, the isocenter was located at the center of mass of the phantom and 3 Gy was delivered to the isocenter (prescription dose = 3 Gy). The doses given by the air-electron-streams were measured using the EBT3 films on the panels placed orthogonal to the direction of the magnetic field at distances of 10 and 17 cm from CAX. Two dose distributions per irradiation were measured on the panel facing the phantom surface of the incident beam (front panel) and on the panel facing the phantom surface of the beam exit (end panel). We investigated the doses by the air-electron-streams by calculating the average doses inside the circles drawn around a point of the maximum dose with radii of x cm (DRx) from the dose distributions on the panels (x = 1–5 cm). The largest value of DRx was DR1 (1.64 Gy, 55% of the prescription dose) at 10 cm distance from CAX, with FS12.6, at 30° phantom-angle and 330° gantry angle. The average difference of the DR1 at the end panels (FS12.6) between the calculations and measurements was 1.36 Gy. The average global gamma passing rate with 3%/3 mm on the dose distributions at the end panels (FS12.6) was 40.3%. The calculated dose distributions on both panels were not coincident with the measured dose distributions. The Spearman’s rank correlation coefficients between the projected areas and the DRx values were always higher than 0.75 (all with p < 0.001). The doses by the air-electron-streams increased with the projected areas of the cross-sections of the treatment beams on the panels.
Citation: An HJ, Kim J-i, Park JM (2019) Electron streams in air during magnetic-resonance image-guided radiation therapy. PLoS ONE 14(5): e0216965. https://doi.org/10.1371/journal.pone.0216965
Editor: Gayle E. Woloschak, Northwestern University Feinberg School of Medicine, UNITED STATES
Received: March 25, 2019; Accepted: May 1, 2019; Published: May 15, 2019
Copyright: © 2019 An et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Data Availability: All relevant data are within the manuscript and its Supporting Information files.
Funding: JIK received a National Research Foundation of Korea (NRF) grant funded by the Korea government (MSIP) (No. 2017M2A2A7A02020643). JMP received a National Research Foundation of Korea (NRF) grant funded by the Korea government (MSIP) (No. 2017M2A2A7A02020640 and No. 2017M2A2A7A02020641). URL of funder: http://www.nrf.re.kr/index. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.
Competing interests: The authors have declared that no competing interests exist.
Introduction
Magnetic-resonance image-guided radiation therapy (MR-IGRT) became clinically available with the release of the first commercial MR-IGRT device, the ViewRay system (ViewRay Inc., Cleveland, OH), which is a combination of an on-board 0.35-T MR imaging system and a radiation therapy system with Co-60 sources [1]. Since this system does not require any extra imaging dose, the daily 3D MRI for patient setup can be acquired [2]. This enables daily adaptive radiation therapy (ART) combined with the rapid optimization algorithm and the rapid dose calculation capability of the ViewRay system [3]. Moreover, near-real-time cine planar MRI can be acquired during treatment with the ViewRay system (a single cine image at 4 frames/s or three cine images at 2 frames/s). Therefore, respiratory gated radiation therapy based on the actual near-real-time tumor motion can be performed without any external surrogate [3]. The ART capability, as well as the gating capability based on the actual tumor motion of the ViewRay system has the potential to minimize the target margins, which is beneficial for sparing doses to a nearby normal tissue around the target volume [4, 5]. This potentially results in reduction in the complications induced by radiation therapy. Furthermore, reduction in the doses given for organs at risk (OARs) by the margin reduction capability of the ViewRay system has the potential to escalate the prescription doses to the target volumes, which potentially increases the efficacy of radiation therapy.
The treatment planning system (TPS) of the ViewRay system is the MRIdian system of which dose calculation algorithm is based on Monte Carlo simulation [6]. Since the Monte Carlo dose calculation algorithm generally calculates doses at the voxels in the region of interest (ROI), i.e., voxel phantoms, the MRIdian system calculates dose distributions in the ROI, including a patient body as well as air around the patient body. This is a difference of the MRIdian system from other commercial TPSs, which generally calculate dose distributions only inside the body structure. Another feature of the MRIdian system is that it is possible to calculate dose distributions with or without magnetic field (0.35 T magnetic field) [7]. When we generated treatment plans for accelerated partial breast irradiation (APBI) with the magnetic field, we observed low dose streams in air in the direction or opposite direction of the magnetic field (termed air-electron stream) [7]. When we calculated dose distributions without the magnetic field, the directional nature of the low dose stream in air disappeared. Therefore, we concluded that it was generated by the secondary electrons scattered in air, i.e., charged particles. This phenomenon was observed frequently at the treatment plans for APBI since the target volumes of APBI were generally located close to the surface (sometimes including the patient surface owing to its margin) [7]. Because the energies of the secondary electrons of the ViewRay system are small (the gamma ray energies of Co-60 are 1.17 and 1.33 MeV), when the target volume is located deep in the patient body, the secondary electrons are absorbed in the patient body. However, if the target volume is located close to the surface or includes the patient surface (as in the situation of APBI treatment), the secondary electrons escape the patient body and scatter in air. In this situation, if there is a magnetic field, the scattered secondary electrons in air exactly orthogonal to the direction of magnetic field rotate on the direction of magnetic field due to Lorentz force [8, 9]. If the scattering directions of the secondary electrons have vectors along the magnetic field, the secondary electrons would show helical movements in the direction (or opposite direction) of the magnetic field [10]. Because the energies of the secondary electrons are small, less than 1.33 MeV, the radii of the helix would be small, and therefore, a bunch of these secondary electrons would form the air-electron stream. According to the direction of the vectors of the secondary electrons along the magnetic field, the air-electron stream can be formed in the direction, or opposite direction, of magnetic field. When the air-electron stream was formed in the direction of magnetic field of the ViewRay system (from the couch to the bore) during APBI with a patient position of the head-first supine, the air-electron stream sometimes could reach the patient’s jaw, neck, armpit, or arm [7]. This results in undesired normal tissue irradiation outside of the treatment field.
In this study, we calculated and measured the doses outside of the treatment field by the air-electron streams under various conditions by utilizing a custom-made phantom. We investigated the doses by the air-electron streams as well as the particular conditions to enhance doses by the air-electron streams.
Materials and methods
Phantoms and experimental setup
To investigate doses by the air-electron stream in the magnetic field, we used a custom-made acrylic phantom with dimensions of 15 cm × 15 cm × 10 cm (density of 1.18 g/cm3), as shown in Fig 1(A).
To vary the angles between the incident beam and the acrylic phantom surface, we designed and fabricated a support phantom as shown in Fig 1(B) and Fig 2. Whole parts of the support phantom were made of acrylic to be compatible with the magnetic field. The support for the acrylic phantom (termed phantom support) was laid over the base of the support phantom, and they were connected to each other by a rotation axis, as shown in Fig 1(C). We cut grooves at the base to fix the supports for angle adjustments, termed angle support (Fig 1(D)), to the base. The angle support can be put into the corresponding groove at the base to form a particular angle between the phantom support and the base. By combination of the angle support and its corresponding groove, angles ranging from 5° to 30° at intervals of 5° can be generated between the phantom support and the base. This means that the support phantom can adjust the angles between the couch surface and the surface of the acrylic phantom placed on the phantom support from 5° to 30° at 5° intervals since the acrylic phantom surface is parallel to the surface of the phantom support. Since the angle between the central axis (CAX) and the vector orthogonal to the acrylic phantom surface (termed phantom angle) is same as that between the phantom surface and the couch surface, the phantom angles can be adjusted from 5° to 30° at 5° intervals. To measure doses by the air-electron stream, two panels for the attachment of Gafchromic EBT3 films (Ashland ISP Advanced Materials, Wayne, NJ) were setup vertically to the base, as shown in Fig 1(B). Experimental setup with the phantom and the support phantom is shown in Fig 2.
The panels for the attachment of EBT3 films were located parallel to CAX and vertical to the direction of magnetic field. We set the panels to be parallel to each other and to be located at the same distances from CAX, which were 17 cm (long-distance setup) and 10 cm (short-distance setup). The panel in front of the incident beam cross-section was termed front panel and that facing the exit beam cross-section was termed end panel.
Conditions of beam irradiation
All beam irradiations in the present study were performed with the ViewRay system. For every beam irradiation, the isocenter located at 105 cm distance from the radiation source (source to axis distance, SAD = 105 cm) was always located at the center of mass of the acrylic phantom [1]. To investigate the effect of the phantom angles on the doses by the air-electron stream, the phantom angles tested in this study were 10°, 20°, and 30° (three phantom angles). At each phantom angle, Co-60 beams with square field sizes of 6.3 cm × 6.3 cm and 12.6 cm × 12.6 cm were delivered to the acrylic phantom (two field sizes) to investigate the field size effect on the doses by the air-electron stream. For each field size, gantry angles of 0°, 30°, and 330° (three gantry angles) were chosen. For each beam delivery, doses delivered to the surfaces of the front panel and end panel were measured with EBT3 films (two measurements for a single beam delivery). The surface doses of the front panel and end panel were measured at distances of 10 and 17 cm from CAX (two distances) under identical beam delivery condition. Therefore, 36 beams were delivered to the acrylic phantom (3 phantom angles × 2 field sizes × 3 gantry angles × 2 distances from CAX) and 72 dose distributions (× front and end panels) were measured. For the short-distance setup (panel setup at 10 cm distance from CAX) while adjusting the phantom angles from 10° to 30°, we chose the acrylic phantom (15 cm × 15 cm × 10 cm) for this study, of which dimension was smaller than those of the commercial solid water phantoms (30 cm × 30 cm × various thicknesses). For each irradiation in this study, 3 Gy was delivered to the isocenter located at the center of mass of the acrylic phantom; that is, the prescription dose was 3 Gy.
Dose calculations
We acquired CT image sets of the phantom with the support phantom to calculate dose distributions at the surfaces of the front and end panels. Since there were six experimental geometries in the present study (3 phantom angles × 2 distances between the panels and CAX), we acquired six CT image sets of the phantom with the support phantom by using the Brilliance CT Big Bore (Phillips, Amsterdam, Netherlands) with a slice thickness of 1 mm. With the CT images, dose distributions were calculated with the MRIdian system under the identical conditions as the measurements. Treatment plans were generated to deliver 3 Gy to the isocenter under each beam delivery condition. Dose distributions were calculated with a dose calculation grid size of 3 mm, which is recommended by the manufacturer for an optimal dose calculation [4, 6]. To maintain the dose calculation accuracy with the Monte Carlo dose calculation algorithm of the MRIdian system while maintaining the dose calculation speeds to be appropriate for on-table ART, the ViewRay Inc. recommends to use a dose calculation grid size of 3 mm in the clinical setting.
Gafchromic EBT3 film measurements
The dose response of the EBT3 films was calibrated with the ViewRay system under the magnetic field to eliminate the magnetic field effect on the EBT3 films [11]. The dual channel method for red and blue corrections was applied (spatial resolution = 75 dpi) [12]. According to the previous studies, uniformity correction for the scanner was applied [13, 14]. A flatbed scanner, Epson 10000XL scanner (Epson Canada Ltd., Toronto, Ontario, Canada), was used for scanning the EBT3 films. The films were scanned in 48-bit color mode, i.e., RGB mode, after 20 h of irradiation. The scanned dose distributions were analyzed with the RIT 113 software (Radiological Imaging Technology, Inc., Colorado Springs, CO).
Data analysis
At both the measured and calculated dose distributions on the surfaces of the front and end panels, the points of the maximum doses were found. After that, circles were drawn around the point of the maximum dose with radii extending from 1 to 4 cm with an interval of 1 cm. For each circle, average doses inside the circles were calculated. Therefore, the circle with a radius of 1 cm was the highest dose region in the dose distributions because there was a single inflection point in the dose distributions of the present study. The average dose inside the circle with a radius of x cm was termed DRx. Naturally, the value of DR1 was the highest and that of DR4 was the lowest. In addition, we calculated areas of isodose lines of 30% (0.9 Gy), 50% (1.5 Gy), 70% (2.1 Gy), 90% (2.7 Gy), and 100% of the prescription dose (3 Gy) in the dose distributions. The area of an isodose line of y% of 3 Gy in cm2 was termed Ay%.
To examine the differences between the calculated and measured doses, we performed global gamma evaluation with absolute doses. A gamma criterion of 3%/3 mm was used and points with doses equal to or less than 10% of the maximum dose in the dose distribution were not evaluated, i.e., the threshold dose was 10%. In addition, we calculated percent differences in the values of DRx between the calculated and measured dose distributions. For Ay%, we acquired just the differences by subtracting the Ay% value of the measured dose from that of the calculated dose.
To investigate the tendency of doses by the air-electron stream, we acquired percent differences in DRx as well as in the Ay% between the front and end panels, between the large and small field sizes (6.3 cm × 6.3 cm vs. 12.6 cm × 12.6 cm), and between the long and short distances from CAX (10 cm vs. 17 cm).
To analyze doses by the air-electron streams combined with the projected areas of the cross-sections of the treatment beams at the phantom surface on the panels, we mathematically calculated the projected areas on both the front and end panels under various conditions of the present study. For the end panels, the cross-sections of the exit beam at the support phantom beneath the acrylic phantom were calculated since the acrylic phantom was placed on the support phantom and the beam cross-section at the support phantom was where the secondary electrons were scattered in air. After that, correlations between the doses by the air-electron streams and the projected areas on the panels were analyzed by calculating Spearman’s rank correlation coefficients (r) with the corresponding p values.
Results
Doses on panel surfaces (irradiation outside of the treatment field)
End panel surface dose with a field size of 6.3 cm × 6.3 cm.
The values of DRx and Ay% with a field size of 6.3 cm × 6.3 cm acquired from the calculated and measured dose distributions on the surfaces of the end panels are shown in Tables 1 and 2, respectively. DR1 with a field size of 6.3 cm × 6.3 cm acquired from the calculated and measured dose distributions on the surfaces of the end panels are plotted according to the phantom angles shown in Fig 3.
Calculated values at the 10 cm (a) and 17 cm (c) distances from the central axis as well as measured values at the 10 cm (b) and 17 cm (d) distances. G0, G30, and G330 indicate gantry angles of 0°, 30°, and 330°, respectively.
An example of dose distributions calculated in TPS, as well as measured with the EBT3 films, is shown in Fig 4.
The dose distributions were acquired on the end panel with a field size of 6.3 cm × 6.3 cm, at a phantom angle of 30°, and at 17 cm distance from CAX. The unit of each legend is cGy and the numbers of x and y axes are pixel numbers (pixel size = 1.36 mm × 1.36 mm).
According to its definition, the values of DRx decreased with increasing radii of the circles and the values of Ay% decreased with increasing value of y.
The values of DRx as well as Ay% with a gantry angle of 0° were always significantly smaller than those with gantry angles of 30° and 330°. There were huge discrepancies in the values of both DRx and Ay% between the calculated and measured dose distributions. With increasing phantom angles, the values of DRx also increased. For the calculated and measured doses, the maximum DRx values were the values of DR1 with a phantom angle of 30°, gantry angle of 330°, and 10 cm distance from CAX, which were 1.41 Gy (47% of the prescription dose) and 1.02 Gy (34% of the prescription dose), respectively.
In the case of Ay%, isodose lines equal to or larger than 50% of the prescription dose were not observed for both the calculation and measurements. The values of Ay% with a gantry angle of 0° were always lower than those with gantry angles of 30° and 330°. The maximum value of Ay% from the measured dose distributions was A30% with a phantom angle of 30°, gantry angle of 330°, and 10 cm distance from CAX, which was 9.2 cm2.
End panel surface dose with a field size of 12.6 cm × 12.6 cm.
The values of DRx and Ay% with a field size of 12.6 cm × 12.6 cm acquired from the calculated and measured dose distributions on the surfaces of the end panels are shown in Tables 3 and 4, respectively. DR1 with a field size of 12.6 cm × 12.6 cm acquired from the calculated and measured dose distributions on the surfaces of the end panels are plotted according to the phantom angles in Fig 5.
Calculated values at the 10 cm (a) and 17 cm (c) distances from CAX as well as measured values at the 10 cm (b) and 17 cm (d) distances. G0, G30, and G330 indicate gantry angles of 0°, 30°, and 330°, respectively.
For the doses with a field size of 12.6 cm × 12.6 cm, the same tendency as that with a field size of 6.3 cm × 6.3 cm was observed, but the absolute values of DRx and Ay% with a field size of 12.6 cm × 12.6 cm were always higher than those with a field size of 6.3 cm × 6.3 cm. The maximum DRx value was DR1 with a phantom angle of 30°, gantry angle of 330°, and the 10 cm distance (2.43 Gy and 1.64 Gy for the calculated and measured values, respectively). The maximum value of Ay% from the calculated dose distributions was A70% (26 cm2) at the phantom angle of 30°, gantry angle of 330°, and 10 cm distance from CAX; however, the values of A70% from the measured dose distributions were always zero. From the measurements, A50% at a phantom angle of 30°, gantry angle of 330°, and 10 cm distance from CAX was the largest (18.5 cm2).
Front panel surface dose with a field size of 6.3 cm × 6.3 cm.
The values of DRx and Ay% with a field size of 6.3 cm × 6.3 cm acquired from the calculated and measured dose distributions on the surfaces of the front panels are shown in Tables 5 and 6, respectively. The DR1 with a field size of 6.3 cm × 6.3 cm acquired from the calculated and measured dose distributions on the surfaces of the front panels are plotted according to the phantom angles in Fig 6.
Calculated values at the 10 cm (a) and 17 cm (c) distances from CAX as well as measured values at the 10 cm (b) and 17 cm (d) distances. G0, G30, and G330 indicate gantry angles of 0°, 30°, and 330°, respectively.
For the doses at the front panel with a field size of 6.3 cm × 6.3 cm, the same tendency as that for the end panel was observed, but the absolute values of DRx and Ay% of the front panel were always smaller than those of the end panel. The maximum DRx value from the calculated and measured dose distributions were DR1 at the phantom angle of 30°, gantry angle of 330°, and 10 cm distance, which were 1.16 Gy (39% of the prescription dose) and 0.36 Gy (12% of the prescription dose). For the calculated dose distributions, the A30% at the phantom angle of 30°, gantry angle of 30°, and 10 cm distance was the largest (20.2 cm2), but the values of A30% from the measured dose distributions were always zero.
Front panel surface dose with a field size of 12.6 cm × 12.6 cm.
The values of DRx and Ay% with a field size of 12.6 cm × 12.6 cm acquired from the calculated and measured dose distributions on the surfaces of the front panels are shown in Tables 7 and 8, respectively. The DR1 with a field size of 12.6 cm × 12.6 cm acquired from the calculated and measured dose distributions on the surfaces of the front panels are plotted according to the phantom angles in Fig 7.
Calculated values at the 10 cm (a) and 17 cm (c) distances from CAX as well as measured values at the 10 cm (b) and 17 cm (d) distances. G0, G30, and G330 indicate gantry angles of 0°, 30°, and 330°, respectively.
For the doses at the front panel with a field size of 12.6 cm × 12.6 cm, the same tendency as those of the other results was observed. The absolute values of DRx and Ay% of the front panel with a field size of 12.6 cm × 12.6 cm were always larger than those with a field size of 6.3 cm × 6.3 cm. However, the absolute values of DRx and Ay% of the front panel with a field size of 12.6 cm × 12.6 cm were always smaller than those of the end panel with the same field size. The maximum DRx value from the calculated dose distributions was DR1 with a phantom angle of 30°, gantry angle of 330°, and 10 cm distance, which was 2.34 Gy (78% of the prescription dose). The maximum DRx value from the measured dose distributions was DR1 with a phantom angle of 30°, gantry angle of 30°, and 10 cm distance, which was 0.61 Gy (20% of the prescription dose). For the Ay% from the calculated dose distributions, A70% at a phantom angle of 30°, gantry angle of 30°, and 10 cm distance was the largest (28.1 cm2); however, even A30% was always zero in the case of measurements.
Differences between the calculation and measurement
Results of gamma evaluation.
The average global gamma passing rates with 3%/3 mm on the dose distributions at the end panels with field sizes of 6.3 cm × 6.3 cm and 12.6 cm × 12.6 cm were 42.0% ± 23.0% (ranging from 6.8% to 80.3) and 40.3% ± 9.8% (ranging from 17.0% to 63.1%), respectively. Those on the dose distributions of the front panels were 27.6% ± 12.9% (ranging from 10.0% to 70.4%) and 26.1% ± 13.3% (ranging from 7.9% to 52.9%), respectively. The calculated dose distributions on both panels were not coincident with the measured dose distributions.
Differences between the values of DRx from the calculated and measured dose distributions.
The average percent differences between the values of DRx from the calculated and measured dose distributions are plotted in Fig 8.
The differences with a field size of 6.3 cm × 6.3 cm at 17 cm (a) and 10 cm (c) distances from CAX as well as those with a field size of 12.6 cm × 12.6 cm at 17 cm (b) and 10 cm (d) distances from CAX.
As shown in Fig 7, huge percent differences were observed at both the end and front panels. The percent differences between the calculations and measurements at the front panels were much larger than those at the end panels. This resulted from the smaller DRx values at the normalization points (maximum doses) of the front panels than those at the end panels. The average absolute differences in DR1 of the front panels with field sizes of 6.3 cm × 6.3 cm and 12.6 cm × 12.6 cm were 26.4 ± 13.0 cGy and 46.4 ± 17.5 cGy, respectively, while those of the end panels were 61.1 ± 23.0 cGy and 136.1 ± 41.7 cGy. Although the absolute differences at the end panels were larger than those at the front panels, the percent differences normalized to the maximum dose in the dose distributions of the front panels appeared to be larger than those of the end panels because the maximum doses at the front panels were smaller than those at the end panels. In general, the calculated doses were larger than those of the measured doses, i.e., doses calculated by the air-electron streams were generally overestimated compared to the measurements.
Measured dose differences between the front and end panels
The average percent differences in the values of DRx from the measured dose distributions between the front and end panels are plotted in S1 Fig, which can be found in supporting information file. Doses at the end panels were always larger than those at the front panels.
Measured dose differences between the large and small field sizes
The average percent differences in the values of DRx from the measured dose distributions between the field sizes of 12.6 cm × 12.6 cm and 6.3 cm × 6.3 cm are plotted in S2 Fig, which can be found in supporting information file. Doses with a large field size (12.6 cm × 12.6 cm) were always larger than those with a small field size (6.3 cm × 6.3 cm).
Measured dose differences between the long and small distances from central axis
The average percent differences in DRx values from the measured dose distributions between the distances of 17 cm and 10 cm from CAX are plotted in S3 Fig, which can be found in supporting information file.
Doses at the long distance from CAX (17 cm) were always smaller than those at the short distance (10 cm).
Correlations between the projected areas and doses on the panels
The calculated values of the projected areas of the beam cross-sections at the phantom surface on the panels are shown under various conditions of the present study in S1 Table, which can be found in supporting information file.
With an increase in the phantom angles and gantry angles, the projected areas increased. The projected areas on the end panels were larger than those on the front panels. The smallest and largest areas projected on the panels were 6.2 cm2 (gantry angle = 0°, phantom angle = 10°, field size = 6.3 cm × 6.3 cm, and on the front panel) and 109.4 cm2 (gantry angle = 30° and 330°, phantom angle = 30°, field size = 12.6 cm × 12.6 cm, and on the end panel), respectively.
The values of DRx according to the projected areas on the front and end panels are plotted in Figs 9 and 10, respectively.
Plots of values of DR1 (a), DR2 (b), DR3 (c), and DR4 (d) on the front panels at the 10 cm distance from CAX according to the projected areas of the cross-section of the treatment beam at the phantom on the front panels. Plots of the values of DR1 (e), DR2 (f), DR3 (g), and DR4 (h) at the 17 cm distance from CAX according to the projected areas.
Plots of values of DR1 (a), DR2 (b), DR3 (c), and DR4 (d) on the end panels at 10 cm distance from CAX according to the projected areas of the cross-section of the treatment beam at the phantom on the end panels. Plots of the values of DR1 (e), DR2 (f), DR3 (g), and DR4 (h) at the 17 cm distance from CAX according to the projected areas.
The r values of the projected areas on the panels to the values of DRx are summarized in Table 9 with the corresponding p values.
Discussion
In the present study, we investigated undesired irradiations outside of the treatment field by the air-electron stream during MR-IGRT for tumors located at superficial regions in the body when generating a slope between the patient’s body surface and the direction of the magnetic field in cases such as MR-IGRT for APBI. This phenomenon occurs owing to the same reason as the electron return effect during MR-IGRT, which is the Lorentz force [8, 9]. Although the air-electron streams were generated with the same reason as the electron return effect, its effect on the treatment is different from that of the electron return effect because the air-electron streams result in an undesired irradiation outside of the treatment field, while the electron return effect results in an increased dose deposition at the tissue−air interface [9]. Under various conditions, we investigated which factor increased the doses outside of the treatment field by the air-electron stream. We found that the doses by the air-electron streams increased with increasing phantom angles, increasing field sizes, and decreasing distances from the treatment field. We also found that the doses by the air-electron stream increased at gantry angles of 30° and 330° compared to those at a gantry angle of 0°. In addition, the doses on the end panels were always lager than those on the front panels. The conditions of the large phantom angles, large field sizes, measurement at the end panels, and oblique gantry angles result in an increase in the projected area of the treatment beam cross-section at the phantom on the panels. It is obvious that the increases in the phantom angles and field sizes, as well as oblique beams, to the phantom increase the projected area on the panels. The projected areas on the end panels are also larger than those on the front panels under an identical condition because the treatment beam diverges. Therefore, comprehensively reviewing the results, the increase in the projected area plays a major role in increasing the out-of-field doses by the air-electron streams. This is clearly identified in the correlations between the projected areas and the values of DRx. We found very strong correlations, showing r values of up to 0.938, between the projected area and the values of DRx (all with p < 0.001). Accordingly, the largest dose in the present study was observed on the end panel with a field size of 12.6 cm × 12.6 cm at a gantry angle of 330° and phantom angle of 30°, which was 164.1 cGy (DR1).
In the case of exit beams, not only the increases in the projected areas owing to the beam divergence but also the vectors of the electrons scattered from the phantom to air might affect the increases in the doses outside of the treatment field by the air-electron stream. Since the electrons generated by the photon beams should escape the phantom to form the air-electron streams, the electrons scattered from the phantom at the beam exit would be more than those backscattered at the beam entrance considering the gamma ray direction into the phantom. This could increase doses on the end panels by the air-electron stream. However, the gamma rays at the exit were attenuated more than those at the entrance, which results in a decrease in the number of electrons scattered from the phantom. This could decrease doses on the end panels by the air-electron stream. Therefore, for the doses on the end panels, there were causes to both increase and decrease the doses on the end panels, simultaneously. Reviewing the results, the doses on the end panels were higher than those on the front panels. For further investigation, Monte Carlo simulation should be performed, and therefore, we will investigate the doses at the end panels with Monte Carlo simulation in the future.
The doses at a short distance from CAX were always larger than those at a large distance, which is irrelevant to the projected area sizes. This seems to be owing to the low energies of the scattered electrons in this study. The air-electron streams in the present study were generated from the Co-60 gamma rays (maximum energy of 1.33 MeV), and therefore, the electron energies should be lower than those of the maximum energy of the gamma ray [1]. With these low energies, most electrons in the air-electron streams might not propagate far away in air. Therefore, as increasing the distance from the treatment beam, doses by the air-electron stream decreased as shown in the results. If high-energy photon beams were used for MR-IGRT utilizing the MR-linac, such as MRIdian Linac (ViewRay Inc., Cleveland, OH), with a 6 MV photon beam or the Elekta MR-linac with a 7 MV photon beam (Elekta, Strockholm, Sweden), the air-electron streams could propagate farther than those presented in this study; therefore, it is necessary to be highly cautious for the air-electron streams when using MR-linac [15, 16].
Hackett et al. investigated the spiraling contaminant electrons utilizing the Elekta MR-linac, which increase surface doses outside of the treatment field. The spiraling contaminant electrons is quite similar to the air-electron stream in the present study since both are electrons with directional nature by the magnetic field and deposit doses outside of the treatment field. However, the spiraling contaminant electrons are generated in the components of the linac head, shielding, and in the air column though which the incident beam passes while the air-electron stream is generated by the secondary electrons scattered in air from a patient. In other words, the spiraling contaminant electrons are generated with the contaminant electrons while the air-electron streams are generated with the secondary electrons scattered in air from a patient. Therefore, the doses outside of the treatment field by the spiraling contaminant electrons were measured without a phantom in the study by Hackett et al. while the air-electron streams were measured with a phantom in the present study. The surface dose outside of the treatment field (field size = 10 cm × 10 cm) by the spiraling contaminant electrons was approximately 5% of the maximum dose at 5 cm distance from the field edge, which is much smaller than those of the present study although the photon beam energy of the Elekta MR-linac was larger than that of the ViewRay system (7 MV vs. 1.17 MeV and 1.33 MeV).
The calculated doses outside of the treatment field by the air-electron streams were not coincident with the measured ones. One of the reasons for this discrepancy might be the large dose calculation grid size of the MRIdian system in the present study, which was 3 mm. As mentioned above, to maintain the dose calculation accuracy based on the Monte Carlo algorithm, i.e., to reduce uncertainties in the calculated doses in the voxels due to small numbers of histories, as well as to enable fast dose calculation for the on-Table ART in the clinic, the dose calculation grid size in the present study was set as 3 mm following the manufacturer’s recommendation. However, the doses by the air-electron streams were surface doses, which should be assessed with a fine dose calculation resolution. The doses by the air-electron streams measured with the EBT3 films were doses at the depths of approximately 0.14 mm since the thickness of the EBT3 film is approximately 0.27 mm. Therefore, the doses on the panel surfaces calculated with a dose calculation grid size of 3 mm would be different from those measured with the films. Besides the dose calculation resolution, there might be other reasons for the discrepancy between the calculations and the measurements in this study. This will be investigated further in the future.
In an extreme case, the largest average dose inside a circle with a radius of 1 cm (area of approximately 3.14 cm2) by the air-electron stream at 10 cm distance from CAX was as large as 54.7% of the prescription dose and the area irradiated by equal to or larger than 50% of the prescription dose was 18.5 cm2. These are clinically significant undesired irradiations considering that the irradiated regions would be normal tissue far from the treatment volume [17]. However, these high dose irradiations outside of the treatment field would hardly occur during an actual treatment in the clinic since multiple beams are generally utilized for the actual treatment. In addition, because a patient’s body is generally thicker than the phantom used in this study, the exit dose would not be as high as those observed on the end panels in this study. In a previous study, we performed in vivo measurements of the doses by the air-electron streams during APBI, and found the average value of DR1 to be approximately 4% of the prescription dose, which is much smaller than those in the present study [7]. Owing to the low energies of the air-electron streams, we could easily shield these doses with only 1-cm-thick commercial build-up bolus (Superflab Bolus, Radiation Products Design Inc., Alvertville, MN). For the MR-linacs, materials thicker than 1-cm-thick bolus would be required to shield the air-electron streams owing to the higher energies of the photon beams than that of the Co-60 source.
Conclusions
The undesired irradiation outside of the treatment field owing to the magnetic field when treating tumors located close to the patient’s surface is a unique feature of MR-IGRT, which requires careful caution. As shown in the results, the calculated doses by the air-electron streams are inaccurate compared to the measurements. We found that the doses by the air-electron streams increased with the projected area of the cross-section of the treatment beam on the irradiated surface. In this situation, shielding must be considered to prevent undesirable out-of-field irradiation. The undesired irradiation outside of the treatment field would be more problematic for MR-linac, which uses photons with higher energies than that of Co-60 source in the present study because the ranges in air as well as the penetrating powers of the air-electron streams would be larger than those presented in this study.
Supporting information
S1 Fig.
Average percent differences in the values of DRx from the measured dose distributions between the front and end panels at the 17 cm (a) and 10 cm (b) distances from the central axis.
https://doi.org/10.1371/journal.pone.0216965.s001
(DOCX)
S2 Fig.
Average percent differences in the values of DRx from the measured dose distributions between the field sizes of 12.6 cm × 12.6 cm and 6.3 cm × 6.3 cm plotted at 17 cm (a) and 10 cm (b) distances from the central axis.
https://doi.org/10.1371/journal.pone.0216965.s002
(DOCX)
S3 Fig.
Plots of average percent differences in the values of DRx from the measured dose distributions between the distances of 17 cm and 10 cm from the central axis from the doses on the end panel (a) and front panel (b).
https://doi.org/10.1371/journal.pone.0216965.s003
(DOCX)
S1 Table. Areas of projection of the beam cross-section at the phantom surface on the panels.
https://doi.org/10.1371/journal.pone.0216965.s004
(DOCX)
Acknowledgments
This work was supported by the National Research Foundation of Korea (NRF) grant funded by the Korea government (MSIP) (No. 2017M2A2A7A02020640, No. 2017M2A2A7A02020641, and No. 2017M2A2A7A02020643).
References
- 1. Mutic S, Dempsey JF. The ViewRay system: magnetic resonance-guided and controlled radiotherapy. Semin Radiat Oncol. 2014;24(3):196–9. Epub 2014/06/17. pmid:24931092.
- 2. Son J, An HJ, Choi CH, Chie EK, Kim JH, Park JM, et al. Assessment of dose distributions according to low magnetic field effect for prostate SABR. J Radiat Prot Res. 2019;44(1):26–31.
- 3. Kim JI, Lee H, Wu HG, Chie EK, Kang HC, Park JM. Development of patient-controlled respiratory gating system based on visual guidance for magnetic-resonance image-guided radiation therapy. Med Phys. 2017;44(9):4838–46. Epub 2017/07/05. pmid:28675492.
- 4. Park JM, Park SY, Kim HJ, Wu HG, Carlson J, Kim JI. A comparative planning study for lung SABR between tri-Co-60 magnetic resonance image guided radiation therapy system and volumetric modulated arc therapy. Radiother Oncol. 2016;120(2):279–85. Epub 2016/07/13. pmid:27401404.
- 5. Park JM, Park SY, Kim JI, Kang HC, Choi CH. A comparison of treatment plan quality between Tri-Co-60 intensity modulated radiation therapy and volumetric modulated arc therapy for cervical cancer. Phys Med. 2017;40:11–6. Epub 2017/08/02. pmid:28760506.
- 6. Wooten HO, Green O, Yang M, DeWees T, Kashani R, Olsen J, et al. Quality of Intensity Modulated Radiation Therapy Treatment Plans Using a 60Co Magnetic Resonance Image Guidance Radiation Therapy System. Int J Radiat Oncol Biol Phys. 2015;92(4):771–8. Epub 2015/06/25. pmid:26104932.
- 7. Park JM, Shin KH, Kim JI, Park SY, Jeon SH, Choi N, et al. Air-electron stream interactions during magnetic resonance IGRT: Skin irradiation outside the treatment field during accelerated partial breast irradiation. Strahlenther Onkol. 2018;194(1):50–9. Epub 2017/09/17. pmid:28916952.
- 8. Chen X, Prior P, Chen GP, Schultz CJ, Li XA. Technical Note: Dose effects of 1.5 T transverse magnetic field on tissue interfaces in MRI-guided radiotherapy. Med Phys. 2016;43(8):4797. Epub 2016/08/05. pmid:27487897.
- 9. van Heijst TC, den Hartogh MD, Lagendijk JJ, van den Bongard HJ, van Asselen B. MR-guided breast radiotherapy: feasibility and magnetic-field impact on skin dose. Phys Med Biol. 2013;58(17):5917–30. Epub 2013/08/08. pmid:23920343.
- 10. Hackett SL, van Asselen B, Wolthaus JWH, Bluemink JJ, Ishakoglu K, Kok J, et al. Spiraling contaminant electrons increase doses to surfaces outside the photon beam of an MRI-linac with a perpendicular magnetic field. Phys Med Biol. 2018;63(9):095001. Epub 2018/03/30. pmid:29595150.
- 11. Reyhan ML, Chen T, Zhang M. Characterization of the effect of MRI on Gafchromic film dosimetry. J Appl Clin Med Phys. 2015;16(6):325–32. Epub 2015/12/25. pmid:26699587; PubMed Central PMCID: PMCPMC5690986.
- 12. Micke A, Lewis DF, Yu X. Multichannel film dosimetry with nonuniformity correction. Med Phys. 2011;38(5):2523–34. Epub 2011/07/23. pmid:21776787.
- 13. Lewis D, Chan MF. Correcting lateral response artifacts from flatbed scanners for radiochromic film dosimetry. Med Phys. 2015;42(1):416–29. Epub 2015/01/08. pmid:25563282; PubMed Central PMCID: PMCPMC5148133.
- 14. Ferreira BC, Lopes MC, Capela M. Evaluation of an Epson flatbed scanner to read Gafchromic EBT films for radiation dosimetry. Phys Med Biol. 2009;54(4):1073–85. Epub 2009/01/27. pmid:19168937.
- 15. Ramey SJ, Padgett KR, Lamichhane N, Neboori HJ, Kwon D, Mellon EA, et al. Dosimetric analysis of stereotactic body radiation therapy for pancreatic cancer using MR-guided Tri-60Co unit, MR-guided LINAC, and conventional LINAC-based plans. Pract Radiat Oncol. 2018;8(5):e312–e21. Epub 2018/04/29. pmid:29703704.
- 16. Al-Ward S, Wronski M, Ahmad SB, Myrehaug S, Chu W, Sahgal A, et al. The radiobiological impact of motion tracking of liver, pancreas and kidney SBRT tumors in a MR-linac. Phys Med Biol. 2018;63(21):215022. Epub 2018/10/31. pmid:30375365.
- 17. ICRP. The 2007 Recommendations of the International Commission on Radiological Protection. ICRP publication 103. Ann ICRP. 2007;37(2–4):1–332. Epub 2007/12/18. pmid:18082557.