Characterization of a reversible thermally-actuated polymer-valve: A potential dynamic treatment for congenital diaphragmatic hernia

  • Justin S. Baba ,

    Roles Conceptualization, Data curation, Investigation, Methodology, Resources, Validation, Visualization, Writing – review & editing

    babajs@ornl.gov

    Affiliations Electrical and Electronics Systems Research Division, Oak Ridge National Laboratory, Oak Ridge, Tennessee, United States of America, Biophotonics Center, Department of Biomedical Engineering, Vanderbilt University, Nashville, Tennessee, United States of America

  • Timothy E. McKnight,

    Roles Conceptualization, Data curation, Formal analysis, Investigation, Methodology, Resources, Supervision, Validation, Visualization, Writing – original draft

    Affiliation Electrical and Electronics Systems Research Division, Oak Ridge National Laboratory, Oak Ridge, Tennessee, United States of America

  • M. Nance Ericson,

    Roles Conceptualization, Data curation, Investigation, Methodology, Resources, Validation

    Affiliation Electrical and Electronics Systems Research Division, Oak Ridge National Laboratory, Oak Ridge, Tennessee, United States of America

  • Anthony Johnson,

    Roles Conceptualization, Investigation, Methodology, Resources, Validation

    Current address: Department of Obstetrics, Gynecology, and Reproductive Sciences, University of Texas, McGovern Medical School at Houston, Houston, Texas, United States of America

    Affiliation Department of Obstetrics, Gynecology, Baylor College of Medicine, Houston, Texas, United States of America

  • Kenneth J. Moise Jr.,

    Roles Conceptualization, Investigation, Methodology, Resources, Validation

    Current address: Department of Obstetrics, Gynecology, and Reproductive Sciences, University of Texas, McGovern Medical School at Houston, Houston, Texas, United States of America

    Affiliation Department of Obstetrics, Gynecology, Baylor College of Medicine, Houston, Texas, United States of America

  • Boyd M. Evans III

    Roles Conceptualization, Data curation, Funding acquisition, Investigation, Methodology, Project administration, Resources, Supervision, Validation

    Affiliation Electrical and Electronics Systems Research Division, Oak Ridge National Laboratory, Oak Ridge, Tennessee, United States of America

Abstract

Background

Congenital diaphragmatic hernia (CDH) is a fetal defect comprising an incomplete diaphragm and the herniation of abdominal organs into the chest cavity that interfere with fetal pulmonary development. Though the most promising treatment for CDH is via interventional fetoscopic tracheal occlusion (TO) surgery in-utero, it has produced mixed results due to the static nature of the inserted occlusion. We hypothesize that a suitable noninvasively-actuatable, cyclic-release tracheal occlusion device can be developed to enable dynamic tracheal occlusion (dTO) implementation.

Objective

To conduct an in-vitro proof-of-concept investigation of the construction of thermo-responsive polymer valves designed for targeted activation within a physiologically realizable temperature range as a first step towards potential development of a noninvasively-actuatable implantable device to facilitate dynamic tracheal occlusion (dTO) therapy.

Methods

Six thermo-responsive polymer valves, with a critical solution temperature slightly higher than normal physiological body temperature of 37°C, were fabricated using a copolymer of n-isopropylacrylamide (NIPAM) and dimethylacrylamide (DMAA). Three of the valves underwent ethylene oxide (EtO) sterilization while the other three served as controls for EtO-processing compatibility testing. Thermal response actuation of the valves and their steady-state flow performances were evaluated using water and caprine amniotic fluid.

Results

All six valves consisting of 0.3-mole fraction of DMAA were tested for thermal actuation of caprine amniotic fluid flow at temperatures ranging from 30–44°C. They all exhibited initiation of valve actuation opening at ~40°C with full completion at ~44°C. The overall average coefficient of variation (CV) for the day-to-day flow performance of the valves tested was less than 12%. Based on a Student t-test, there was no significant difference in the operational characteristics for the EtO processed versus the non-EtO processed valves tested.

Conclusions

We successfully fabricated and demonstrated physiological realizable temperature range operation of thermo-responsive polymer valves in-vitro and their suitability for standard EtO sterilization processing, a prerequisite for future in-vivo surgical implantation testing.

Introduction

Congenital diaphragmatic hernia (CDH) is a fetal defect involving absence of portions of the diaphragm and herniation of abdominal organs into the chest that interfere with fetal pulmonary development. The incidence of this congenital anomaly is approximately 1 in 1754–5882 births with regional, ethnic, and gender variations [18]. A reported range of 40–80% of infants diagnosed with severe CDH die at birth or shortly thereafter [810]. In a review of the Healthcare Cost and Utilization of Project 2003 Kids’ Inpatient Database (KID), CDH ranked among the most costly and lethal of congenital malformations. CDH was sixth for average length of stay at 25 days (95% CI: 23.1–26.8 days) with total annual patient care costs of $179,470,456 (95% CI: $156,501,285 –$202,439,627). Excluding anencephaly and trisomies 13 and 18, CDH was associated with the highest risk of in-hospital neonatal death (34.4%) [11]. More recent adjusted estimates peg CDH costs at $250,000 per case with a US annual cost that exceeds $230 million, and the data reveal that earlier interventions lead to reduced overall patient treatment costs [12].

Initial treatments for CDH were based on attempts to repair herniated fetal diaphragms pre-birth via open fetoscopic surgery. Those efforts were fraught with significant risk to both mother and child, thus yielding a perinatal mortality rate in excess of 70% [13]. Subsequently, an alternative treatment method emerged based on clinical observations that non-CDH infants born with a condition of tracheal blockage–such as laryngeal or tracheal atresia–manifested enlarged hyperplastic lungs [14]. Consequently, intentional fetoscopic tracheal occlusion (TO) was evaluated as an in-utero bridging procedure for facilitating proper lung development. The interventive procedure yields an accumulation of pulmonary fluid that facilitates lung enlargement, which in-turn serves to minimize or reverse abdominal organ advancement into the pulmonary cavity [15,16]. Following those reports, there was a “PLUG” (plug-lung-until-it-grows) odyssey, with many different experimental manipulations to restrict egress of fetal lung liquid that included the use of cuffs, polymeric foams, magnetic valves, umbrellas, surgical dissection and detachable vascular occlusive balloons [17].

The standard technique for fetal endoluminal tracheal occlusion (FETO) utilizes a single port for fetoscopic access to the trachea at 26–30 weeks gestation. Under direct ultrasound visualization, a detachable balloon is advanced just above the fetal carina and inflated with isotonic solution. Once secured in place, the static obstruction remains until surgical removal at approximately 34 weeks gestation. Though now obsolete, the primary device utilized, the GVB16 Goldvalve detachable balloon, “was originally designed for interventional radiological procedures to occlude anomalous vessels” [18]. It had a maximum volume of 0.8-ml, diameter of 8.0-mm and length of 21.0-mm and was implemented to provide complete, static obstruction in the fetal trachea until its removal. Currently, the Goldballoon is the detachable balloon device being utilized in ongoing research evaluating the potential benefits of FETO for the treatment of CDH [19]. This particular device, from BALT Extrusion, was designed for endovascular treatment by interventional neuroradiology for intracranial vascular pathologies [20,21]. At present, the device is not commercially available in the United States. Clearly, none of the aforementioned FETO devices were specifically designed for in-utero treatment of CDH.

Initial studies of TO have revealed provisional benefits in CDH that include both an increase in lung tissue stretch and accelerated lung growth volume. However, prolonged static TO has also revealed deleterious effects. They include an increase in alveolar wall thickness, a reduction in surfactant synthesis due to increased differentiation of type II into type I pneumocytes, and ultimately respiratory failure after birth [2226]. Early release of static TO has been attempted to improve surfactant production in fetal lambs with and without surgically-induced CDH. It has produced varied responses in the preservation of type II pneumocytes. Adversely, early release has resulted in loss of the positive gains in lung maturation, growth, and vascular remodeling associated with TO [2628]. In recent studies of catheterized fetal lambs, Nelson et al. [29] have demonstrated that cyclic release of TO, late in gestation, induced optimal lung growth with morphological maturation of hypoplastic lung parenchyma whilst maintaining a population of type II pneumocytes. Overall, these findings indicate that TO continues to have promising therapeutic potential for CDH.

Before the benefits of TO can be fully realized, it is critically important to develop physiologic methodologies whereby the TO is not a static blockage, but rather a dynamic one that can be clinically manipulated cyclically over the entire gestational duration [30]. Recently, Jelin et al. [31] reported on the construction and demonstration of a novel dynamic tracheal occlusion (dTO) device that operates based on a built-in pressure-sensitive valve. The valve actuated at a preset pressure differential of 80-mmH2O to allow for dynamic lung fluid efflux. In a fetal lamb CDH model, the fetoscopic implanted dTO device facilitated improvements in lung morphometrics and in functional outcomes. Their design, though promising, is disadvantaged by its pre-implantation, fixed pressure set-point that may not be optimal for achieving ideal lung morphometrics and functional outcomes over the entire gestation period. Consequently, we report on the development of a more flexible-control dTO implant device that can be manipulated non-invasively to provide for remote, clinical control over fetal lung volume efflux based on adjustable time-period-modulation of valve actuation. This is a significant improvement over the aforementioned pre-implantation, fixed pressure set-point actuation dTO design.

Accordingly, the focus of this communication is the in-vitro proof-of-concept demonstration of the construction of thermo-responsive polymer valves designed for targeted activation within a physiologically realizable temperature range. Accomplishing this objective is the first step towards achieving the aforementioned dTO implant device. The ensuing summary is provided to help contextualize future clinical implementation of the fully envisioned dTO implant device system that is still under development.

Our novel dTO system is comprised of a balloon occlusion, styled after the Goldvalve device [32,33], that also features an internal lumen as a path for pulmonary fluid flow. A wireless, RF-responsive microvalve element is inserted into this lumen. The microvalve element consists of a thermally-actuated polymer-valve body with an embedded resistive heater and a RF coil-driving unit. The RF unit provides clinically directed inductive power to the heating element at very specific RF frequencies. At normal physiological temperature (37°C), the polymer-valve body remains in a swollen state that fills up the balloon lumen, thus occluding pulmonary fluid flow and providing for accumulation of fluid to yield the desired expansion of the fetal lung. Under specific RF excitation, the embedded valve-heater increases the temperature of the polymer by 3–5°C. This results in a reversible, thermal collapse of the polymer-valve body yielding a principle annular flow of pulmonary fluid through the balloon lumen via around its shrunken cross-section. Continuous flow is maintained during RF excitation and for a short period afterwards. Thereafter, the polymer-valve body returns to normal physiological temperature accompanied by reversible swelling that re-occludes the balloon lumen cross-section [34,35].

Materials and methods

Materials

N-isopropylacrylamide (NIPAM), dimethylacrylamide (DMAA), 3-methacryloxypropyltri-methoxysilane (MAPTOS), N-N’-methylenebisacrylamide (MBAA), ammonium persulfate (APS), ethanol, acetic acid, and tetramethylethylenediamine (TEMED) were purchased from Sigma-Aldrich (St. Louis, MO, USA) and used as received. Though the native monomer constituents of these chemicals are cytotoxic, their bioavailability and cytotoxicity are substantially reduced upon polymerization [36,37]. Pertinently, the US Food and Drug Administration (FDA) has approved some formulations of polyacrylamide for human and animal consumption, and also for human applications that include denture adhesives [37].

Phosphate Buffered Saline (PBS) was obtained from ATCC (Manassas, VA, USA) and used as received. Borosilicate glass capillaries were obtained from Sutter Instruments (Novato, CA, USA) and ultrasonically cleaned, using a Branson Ultrasonic Bath Cleaner: Model 1510 (Branson Ultrasonics Corporation, Danbury, CT, USA), with detergent and water prior to use.

Polymer-valve fabrication

Polymer valves were grafted onto the surface of a heater core comprising a glass capillary tube treated with a silane-coupling agent, as described in the following details.

First, glass capillary tube heater cores of 0.5-mm inner diameter (i.d.) and 1.0-mm outer diameter (o.d.) were cut to size (15-mm length). Next, they were ultrasonically cleaned (Branson: Model 1510) using a strong soap solution followed by copious rinsing in distilled water. The heater cores were then dried and submersed in a solution comprised of 1 mL of MAPTOS in 200 mL of ethanol, into which 6 mL of dilute acetic acid (1:10 glacial acetic acid in water) was added just before use. The glass heater cores were placed in this solution for 2 minutes, removed and rinsed extensively with ethanol, then air-dried. All subsequent fabrication steps were then conducted under a nitrogen-purged environment.

The MAPTOS-treated glass cores were placed individually within 10-mm long, cleaned glass capillary tubes (1.78-mm i.d., 2.0-mm o.d.) that had not been MAPTOS-treated, such that a short section of the internal core emerged from each end of the untreated tubes. The outer tube served as a mold for subsequent polymerization reactions. A polymerization solution comprising 200-μL of NIPAM in phosphate buffered saline (50-mg/mL), 2-μL of MBAA (50 mg/mL in water) and various amounts of DMAA (4, 6, 8, or 10-μL) was then mixed with 20-μL of APS (50-mg/mL in water) and 20-μL of TEMED (25 μL/mL in water). A portion of this reaction mixture was then quickly dispensed onto an emergent end of each glass core assembly. The amount used was enough to fill the annular gap between the MAPTOS-treated heater core and the inner diameter of the surrounding untreated glass tube serving as the mold for the subsequent polymerization. The filled molds were reacted overnight under nitrogen purge within a 100% relative humidity environment; typically, this was for 17 hrs.

Next, the polymerized valve molds were removed from nitrogen purge and iteratively immersed into hot (~95°C) and then cool (~25°C) water baths to induce, respectively, shrinkage and swelling of the polymer valves. The thermal cycling process resulted in detachment of the polymer from the mold, thus enabling the complete release of the heater-core-grafted-polymer-valve (henceforth referenced as polymer-valve) unit from its exterior mold. Following this, the released polymer valves were extensively rinsed in a continuously stirred water bath to wash out residual reagents. The thermal cycling and post-wash treatment was extended and iterated to flush any residual monomers from the fabricated devices. Afterwards, the polymer valves were either tested immediately or inserted into the lumen of a balloon implant for later testing.

Results and discussion

Responsive polymer micro-valves have been reported extensively in the literature. They enable reversible valving of molecular transport and fluid flow using polymer response to temperature, pH, or electrochemical redox [3848]. Though thermo-responsive polymer valves are not new, the application to the development of a clinically actuatable dynamic tracheal occlusion (dTO) valve is indeed novel. This is substantiated by two issued patents [34,35]. Our approach, utilizing thermo-responsive polymer microvalving, is depicted in Fig 1. The framework for the system is a static balloon occlusion, manufactured by NuMED, Inc. (Hopkinton, NY, USA), featuring a 1.6-mm i.d. lumen flow path. This detachable balloon may be catheter deployed using conventional surgical tools for fetoscopic surgery and immobilized in place by inflation of the outer balloon component via infusion of physiological saline.

thumbnail
Fig 1. Schematic representation of the dynamic tracheal occlusion valve.

It comprises a saline filled detachable balloon that obstructs the fetal trachea and provides an internal flow path via an RF actuatable thermo-responsive polymer-valve.

https://doi.org/10.1371/journal.pone.0209855.g001

As described, the 1-mm o.d. glass heater core element inserted into the polymer micro-valve unit was covalently grafted with the thermo-responsive copolymer of NIPAM and DMAA. Variations of this polymer have been used to provide extremely high backpressures (>2500 psi) in the closed (<32°C) state of thermally-actuated, polymer-monolith valves [38]. Additionally, they have been used in sidewall functionalization of Si and SiO2 based microfluidic devices (sub 100-μm dimensions) for valving of localized flows and generation of 5-psi backpressure between the closed and open states [39]. For the current application, the polymer is designed to have a transition temperature slightly above normal physiological temperature in the range of 40–43°C. Operatively, the polymer remains in a swollen state (valve closed) at normal body temperature and shrinks to open the valve at safe elevated temperatures under the envisioned, clinically directed RF stimulus.

Sterilization of polymer-valve units is a prerequisite for in-vivo implantation studies. Therefore, the dTO system is designed to be compatible with ethylene oxide (EtO) sterilization. Because EtO sterilization could potentially impact the microvalving functionality of the polymer, its impact was investigated and the results included as part of this proof-of-concept demonstration. Essentially, the hypothesis tested was that EtO sterilization can be conducted on dehydrated valve assemblies and that the performance of the resultant sterilized polymer-valve units is maintained when they rehydrate, such as upon fetal implantation.

Covalent grafting of copolymers onto a glass heater core

Saitoh et al. [49] have previously reported on the construction of thermo-responsive flow control elements in microfluidic devices fabricated via covalent grafting of poly-n-isopropylacrylamide (pNIPAM) onto glass capillary tubes using MAPTOS (under the trade name of Bind-Silane) as a coupling reagent during APS-initiated and TEMED- accelerated polymerization. Similarly, our polymer-valve bodies were constructed by grafting the polymerization onto the outer surface of a MAPTOS-treated glass capillary and using a larger, untreated glass capillary as a mold (hereby referred to as glass mold). Note that the inner treated glass capillary also serves as the encapsulation for the RF-driven heating element, as depicted in Fig 1. Size control of the polymer grafted onto this element was achieved by first adjusting the length of the glass mold to the desired length of the polymer-valve body, as described. Subsequently, the heater core element was extended from the ends of the glass mold such that dispensed polymerization solution could be drawn in to fill the annular gap between the heater core and the outer glass mold via capillary action (Step 1, Fig 2).

thumbnail
Fig 2. Fabrication steps of the thermally actuated polymeric valve.

https://doi.org/10.1371/journal.pone.0209855.g002

As previously described, post overnight polymerization, separation of polymer-valve units from the glass mold assemblies was achieved by iterative immersion of the coupled parts into 95°C and 25°C water baths (Step 2, Fig 2). Likewise, thermal polymer-valve-body-collapse at elevated temperatures was utilized to enable the insertion of combined polymer-valve and heater-core elements into various assemblies for testing. Those included insertion into the lumen of the NuMed tracheal balloon (Step 3, Fig 2), or into clear glass or polymer-based capillary tubes for volumetric flow and valve actuation visualization studies. Note that heater-core system testing data is not included in this report but reserved for a future proof-of concept demonstration of the implementation of an externally applied RF signal to inductively drive the incorporated resistive heater for thermal microvalving actuation of the fabricated polymer valves.

Optimization of thermal transition temperature of polymer-valve body

Shen et al. [50] have previously reported on the variation of thermal phase behavior of various copolymers of NIPAM formulated with acrylamide, DMAA, and hydroxyethyl-methacrylate. In their study, polymerizations were conducted at 60°C under a nitrogen purge using free radical polymerization with 2,2′-azobis (isobutyronitrile) (AIBN) and 3-mercaptopropionic acid (MPA) as an initiator and chain transfer reagent, respectively. Lower critical solution temperatures (LCSTs) of the various polymer formulations were determined in water based on the optical transmittance of pNIPAM-DMAA. LCSTs ranging from 36.9°C to 46.0°C were dependent upon the molar fraction of NIPAM that ranged from 0.686 to 0.893. Our approach differed from the aforementioned in employing the following: (1) ambient temperature polymerization, (2) APS as initiator, and (3) TEMED as accelerator.

Utilizing our approach, we first conducted a preliminary study to establish the thermal transition temperature profile of the pNIPAM-DMAA polymer-valve body formulation as a function of DMAA mole fraction. Resultant LCSTs were estimated by observations of copolymer opalescence at steady state temperatures achieved via temperature regulated water bath immersion. As presented in Fig 3, the LCSTs ranged from 32°C, for 0-mole·fraction DMAA, to 48°C, for 0.52-mole·fraction DMAA. As the primary goal was to achieve a transition temperature slightly above physiological, these initial results directed our further studies to polymer compositions of either 0.30 or 0.40-mole·fraction DMAA that provided a LCST of ~40°C and ~42°C, respectively.

thumbnail
Fig 3. Dimethylacrylamide (DMAA)-dependence of the lower critical solution temperature (LCST) of the copolymer.

Datapoints are based on visual determination of solution steady-state opalescence at fixed temperatures. A data connecting line is provided for trend visualization.

https://doi.org/10.1371/journal.pone.0209855.g003

Compatibility with ethylene oxide (EtO) sterilization

Heater cores with grafted pNIPAM-DMAA of 0.30-mole·fraction DMAA, as described, were extensively dialyzed against distilled water and dehydrated at ambient temperature (~21°C) in a laminar-flow hood. This resulted in collapse of the polymer-valve body against the internal glass core. Collapsed valves were then inserted into individual 1.6 mL Eppendorf tubes and shipped to facilities at Baylor College of Medicine for EtO sterilization. At Baylor, Eppendorf tubes were opened and placed in standard EtO sterilization pouches. A set of three packaged valves then underwent standard EtO processing, while the other set of three packaged valves were handled similarly but without EtO processing.

Following EtO processing, the valves (3 EtO-processed and 3 controls) were placed under caprine amniotic fluid at 30°C within a microscope dish heating assembly on a Nikon Diaphot 200 microscope. Upon fluid immersion, the collapsed polymer-valve body began to swell and rehydration was complete within ~60 minutes based on microscopic inspection of the polymer-valve diameter. Next, the diameter of each valve was measured from optical micrographs taken at various steady state temperatures ranging from 30–44°C. Fig 4 (left axis) presents the temperature dependence of the valve diameter, comprising a 1-mm diameter glass heater core and its grafted polymer coating, for both the control and EtO treated groups. When installed within a 1.6-mm diameter lumen, this valve formulation would be expected to provide a closed condition under physiological temperatures and to begin opening within the lumen at ~40°C and to complete opening by ~44°C. Student t-tests indicated no statistically significant difference in overall diameter and in the thermal dependence of valve unit operation following EtO sterilization as compared to the untreated controls.

thumbnail
Fig 4. Thermal responsive performance of 30% DMAA pNIPAM valves in caprine amniotic fluid following dehydration, ethylene oxide (EtO) sterilization, then rehydration.

P-values of control vs. exposed to EtO sterilization valves indicate that the sterilization procedure does not impact valve performance. ‘Flow Rate’ data (teal color: +) present measured flow rates, based on 80-mmHg hydrostatic head pressure, at various temperatures for a representative valve. Valve diameter data are plotted as black symbols: ■, ☐, ◆, Δ, etc.

https://doi.org/10.1371/journal.pone.0209855.g004

Steady state flow evaluations

Steady state flow and repeatability of thermo-responsive polymer valves was evaluated by placing valve units within a stirred water bath and measuring the resultant volumetric flow at various bath temperatures under a fixed hydrostatic head pressure. Heater cores with grafted pNIPAM-DMAA, as described, were immersed into a hot water bath (60°C) to collapse the polymer-valve body against the internal glass core. Collapsed valve units were then inserted into and immobilized at the midpoint of 1.78-mm i.d. glass capillaries of 10-cm length by means of a wire tethered through the hollow interior of the heater core. These assemblies were then installed into a flow setup which provided hydrostatic fluid head pressure by variation of the height of a water reservoir above the outlet of the system. The entire assembly was contained within a water-jacketed incubator maintained at 37°C, with the stirred water bath providing additional heating to the capillary element containing the thermo-reactive polymer-valve unit. Prior to flow testing, the manifold was first charged with water by immersing the capillary containing the polymer-valve unit into a 60°C water bath and forcing water through the system with a syringe. The fluid reservoir was then positioned to provide a head pressure of 80-mmH2O to simulate sufficient lung growth functional pulmonary fluid pressure of the fetal lamb [31]. Flow measurements were conducted by first setting the water bath at fixed temperatures and allowing the capillary assembly to equilibrate to this temperature over a 30-minute period. Subsequently, measurements were taken at each fixed temperature point by weighing the fluid discharge accumulated over a 2-minute period. This approach was adopted instead of implementation of direct joule heating using the embedded heater element or via external RF signal induction coupling because the thermostatic bath ensured reliable reproducibility and maintenance of valve unit temperature for the rigorous collection of temperature-set-point flow rate datasets.

Fig 4 (right axis) presents flow data from an EtO sterilized valve unit comprised of 0.30-mole·fraction DMAA. At temperatures less than approximately 40°C, the polymer-valve body is swollen and the annular valve unit blocks the principle passage of fluid through the outer valve-containment capillary housing. As temperature is increased above 40°C, the polymer-valve body begins to collapse against its internal glass core allowing for principle flow through the resultant annular gap between the polymer-valve body and the outer valve-containment capillary housing (see Fig 1).

Figs 5 and 6 present the temperature dependent flow responses of grafted pNIPAM-DMAA valves with DMAA content of 0.3 and 0.4-mole·fractions, respectively. Measurements were taken at 1°C increments over a temperature range of 37–50°C and repeated over a period of several days. The plotted values represent the average values measured with corresponding standard deviation bars included. At physiological temperature of 37°C, each valve has a fixed leakage flow rate. This contribution is due to the incomplete and non-uniform sealing of the polymer-valve body against the inner diameter of the outer valve-containment capillary housing in addition to a minor trickle flow through the 0.5-mm i.d. hollow heater-core-capillary upon which the polymer-valve body was grafted. Note that the latter flow component is a consequence of the method by which the valve unit was immobilized within the flow channel. As temperature is increased, the 0.3-mole·fraction DMAA valve unit exhibits increasing flow rate in a sigmoidal fashion with a maximum slope at ~45°C. This flow characteristic is attributed to the temperature-dependent incremental collapse of the polymer-valve body and the ensuing opening of an annular flow path between the polymer-valve body and the surrounding outer valve-containment capillary housing, as depicted in Fig 1. Evidently, the 0.4-mole·fraction DMAA valve unit thermal responses in Fig 6 are a shift to higher temperatures of the flow profile of 0.3-mole·fraction DMAA valve units in Fig 5.

thumbnail
Fig 5. Average temperature dependent flow response performance of 30% DMAA fraction valves with standard deviation bars included and a data connecting line added for trend visualization.

These were characterized in caprine amniotic fluid following dehydration, ethylene oxide (EtO) sterilization, then rehydration.

https://doi.org/10.1371/journal.pone.0209855.g005

thumbnail
Fig 6. Average temperature dependent flow response performance of 40% DMAA fraction valves with standard deviation bars included and a data connecting line added for trend visualization.

These were characterized in caprine amniotic fluid following dehydration, ethylene oxide (EtO) sterilization, then rehydration.

https://doi.org/10.1371/journal.pone.0209855.g006

Day to day reproducibility of valve unit flow rates was assessed using the coefficient of variation (CV = sigma/mu). The average percent-CV was 11.24% and 6.46% for the data presented in Figs 5 and 6, respectively. The notable variation outliers observed in the initial flow rates at 37°C and 38°C in Fig 5, are predominantly attributed to the accumulation and/or release of air bubbles (i.e., vapor occlusions). These originate from the polymer-valve body itself and/or from within the internal minor flow path through the hollow 0.5-mm i.d. capillary support structure.

Conclusions

Polymer-valve units comprised of a copolymer of NIPAM and DMAA were constructed that featured phase transition temperatures slightly above physiological body temperature of 37°C. Optimized pNIPAM-DMAA formulation valve units comprising 0.3-mole·fraction DMAA, exhibited phase transitions of ~40°C, maintained their temperature-dependent operation post ethylene oxide sterilization, and performed reproducibly in flow studies conducted over three separate days. These results potentiate further development of this thermo-responsive polymer formulation for clinically-controllable, RF-heating actuatable, implantable polymer valves suitable for applications that include dynamic fetal tracheal occlusion therapy.

Supporting information

S1 Table. Data for LCST temperature and DMAA mole-fraction (Fig 3_data).

https://doi.org/10.1371/journal.pone.0209855.s001

(DOCX)

S2 Table. Data for diameter of valves as function of caprine amniotic fluid flow temperature for 3 EtO-processed and 3 controls (Fig 4A_data).

https://doi.org/10.1371/journal.pone.0209855.s002

(DOCX)

Acknowledgments

This manuscript has been authored by UT-Battelle, LLC under Contract No. DE-AC05-00OR22725 with the U.S. Department of Energy. The United States Government retains and the publisher, by accepting the article for publication, acknowledges that the United States Government retains a non-exclusive, paid-up, irrevocable, world-wide license to publish or reproduce the published form of this manuscript, or allow others to do so, for United States Government purposes. The Department of Energy will provide public access to these results of federally sponsored research in accordance with the DOE Public Access Plan (http://energy.gov/downloads/doe-public-access-plan).

References

  1. 1. Davis CF, Sabharwal AJ. Management of congenital diaphragmatic hernia. Arch Dis Child—Fetal Neonatal Ed. 1998 Jul 1;79(1):F1–3. pmid:9797617
  2. 2. Wenstrom KD, Weiner CP, Hanson JW. A five-year statewide experience with congenital diaphragmatic hernia. Am J Obstet Gynecol. 1991 Oct 1;165(4, Part 1):838–42.
  3. 3. Seaver LH. Human Malformations and Related Anomalies. Part II—Cardiorespiratory Organs: The Lower Respiratory Organs. 2nd ed. Stevenson RE, editor. Oxford: Oxford University Press; 2006. (Oxford monographs on medical genetics).
  4. 4. Parker SE, Mai CT, Canfield MA, Rickard R, Wang Y, Meyer RE, et al. Updated national birth prevalence estimates for selected birth defects in the United States, 2004–2006. Birt Defects Res A Clin Mol Teratol. 2010 Dec;88(12):1008–16.
  5. 5. Louis AMS, Kim K, Browne ML, Liu G, Liberman RF, Nembhard WN, et al. Prevalence trends of selected major birth defects: A multi‐state population‐based retrospective study, United States, 1999 to 2007. Birth Defects Res. 2017 Nov 1;109(18):1442–50. pmid:28905502
  6. 6. 2017 Congenital Malformations Surveillance Report [Internet]. National Birth Defects Prevention Network; 2017 Nov [cited 2018 Aug 20]. (Kirby RS, Browne ML, editors. A Report from the National Birth Defects Prevention Network). Available from: https://www.nbdpn.org/docs/bdr21145-sup-0003-suppinfo03.pdf
  7. 7. Chandrasekharan PK, Rawat M, Madappa R, Rothstein DH, Lakshminrusimha S. Congenital Diaphragmatic hernia–a review. Matern Health Neonatol Perinatol [Internet]. 2017 Dec [cited 2018 Aug 20];3(1). Available from: http://mhnpjournal.biomedcentral.com/articles/10.1186/s40748-017-0045-1
  8. 8. Balayla J, Abenhaim HA. Incidence, predictors and outcomes of congenital diaphragmatic hernia: a population-based study of 32 million births in the United States. J Matern Fetal Neonatal Med. 2014 Sep;27(14):1438–44. pmid:24156638
  9. 9. Harrison MR, Adzick NS, Estes JM, Howell LJ. A Prospective Study of the Outcome for Fetuses With Diaphragmatic Hernia. JAMA. 1994 Feb 2;271(5):382–4. pmid:8054005
  10. 10. Cohen-Katan S, Newman-Heiman N, Staretz-Chacham O, Cohen Z, Neumann L, Shany E. Congenital Diaphragmatic Hernia: Short-Term Outcome. 2009;11:6.
  11. 11. Centers for Disease Control and Prevention (CDC). Hospital stays, hospital charges, and in-hospital deaths among infants with selected birth defects—United States, 2003. MMWR Morb Mortal Wkly Rep. 2007 Jan 19;56(2):25–9. pmid:17230142
  12. 12. Raval MV, Wang X, Reynolds M, Fischer AC. Costs of congenital diaphragmatic hernia repair in the United States—extracorporeal membrane oxygenation foots the bill. J Pediatr Surg. 2011 Apr 1;46(4):617–24. pmid:21496527
  13. 13. Harrison MR, Adzick NS, Flake AW, Jennings RW, Estes JM, MacGillivray TE, et al. Correction of congenital diaphragmatic hernia in utero: VI. hard-earned lessons. J Pediatr Surg. 1993 Oct 1;28(10):1411–8. pmid:8263712
  14. 14. Scurry JP, Adamson TM, Cussen LJ. Fetal lung growth in laryngeal atresia and tracheal agenesis. J Paediatr Child Health. 1989 Feb 1;25(1):47–51.
  15. 15. Wilson JM, DiFiore JW, Peters CA. Experimental fetal tracheal ligation prevents the pulmonary hypoplasia associated with fetal nephrectomy: Possible application for congenital diaphragmatic hernia. J Pediatr Surg. 1993 Nov 1;28(11):1433–40. pmid:8301455
  16. 16. DiFiore JW, Fauza DO, Slavin R, Peters CA, Fackler JC, Wilson JM. Experimental fetal tracheal ligation reverses the structural and physiological effects of pulmonary hypoplasia in congenital diaphragmatic hernia. J Pediatr Surg. 1994 Feb;29(2):248–57. pmid:8176601
  17. 17. Deprest JA, Flake AW, Gratacos E, Ville Y, Hecher K, Nicolaides K, et al. The making of fetal surgery. Prenat Diagn. 2010 Jul 1;30(7):653–67. pmid:20572114
  18. 18. Deprest J, Gratacos E, Nicolaides KH. Fetoscopic tracheal occlusion (FETO) for severe congenital diaphragmatic hernia: evolution of a technique and preliminary results. Ultrasound Obstet Gynecol. 2004 Aug 1;24(2):121–6. pmid:15287047
  19. 19. EuroSTEC. TOTALtrial-Home [Internet]. [cited 2018 Sep 12]. Available from: https://totaltrial.eu/
  20. 20. Balt. Balt Neuroradiology GOLDBALLOON [Internet]. [cited 2018 Sep 12]. Available from: http://www.avan.net.ua/catalog/catalogue_balt_neiro/10_GOLDBALLOON.pdf
  21. 21. Balt [Internet]. [cited 2018 Sep 12]. Available from: http://www.balt.fr/en
  22. 22. O’Toole SJ, Sharma A, Karamanoukian HL, Holm B, Azizkhan RG, Glick PL. Tracheal ligation does not correct the surfactant deficiency associated with congenital diaphragmatic hernia. J Pediatr Surg. 1996 Apr 1;31(4):546–50. pmid:8801310
  23. 23. Papadakis K, De Paepe ME, Tackett LD, Piasecki GJ, Luks FI. Temporary tracheal occlusion causes catch-up lung maturation in a fetal model of diaphragmatic hernia. J Pediatr Surg. 1998 Jul 1;33(7):1030–7. pmid:9694089
  24. 24. Benachi A, Chailley-Heu B, Delezoide A-L, Dommergues M, Brunelle F, Dumez Y, et al. Lung Growth and Maturation after Tracheal Occlusion in Diaphragmatic Hernia. Am J Respir Crit Care Med. 1998 Mar 1;157(3):921–7.
  25. 25. Flageole H, Evrard VA, Piedboeuf B, Laberge J-M, Lerut TE, Deprest JAM. The plug-unplug sequence: An important step to achieve type II pneumocyte maturation in the fetal lamb model. J Pediatr Surg. 1998 Feb 1;33(2):299–303. pmid:9498406
  26. 26. Davey MG, Hedrick HL, Bouchard S, Mendoza JM, Schwarz U, Adzick NS, et al. Temporary tracheal occlusion in fetal sheep with lung hypoplasia does not improve postnatal lung function. J Appl Physiol. 2003 Mar;94(3):1054–62. pmid:12571135
  27. 27. Wild YK, Piasecki GJ, De Paepe ME, Luks FI. Short-term tracheal occlusion in fetal lambs with diaphragmatic hernia improves lung function, even in the absence of lung growth. J Pediatr Surg. 2000 May 1;35(5):775–9. pmid:10813348
  28. 28. Khan PA, Cloutier M, Piedboeuf B. Tracheal occlusion: A review of obstructing fetal lungs to make them grow and mature. Am J Med Genet C Semin Med Genet. 2007 May 15;145C(2):125–38. pmid:17436297
  29. 29. Nelson SM, Hajivassiliou CA, Haddock G, Cameron AD, Robertson L, Olver RE, et al. Rescue of the Hypoplastic Lung by Prenatal Cyclical Strain. Am J Respir Crit Care Med. 2005 Jun 15;171(12):1395–402. pmid:15778486
  30. 30. Deprest JA, Nicolaides K, Gratacos E. Fetal Surgery for Congenital Diaphragmatic Hernia Is Back from Never Gone. Fetal Diagn Ther Basel. 2011 Feb;29(1):6–17.
  31. 31. Jelin EB, Etemadi M, Encinas J, Schecter SC, Chapin C, Wu J, et al. Dynamic tracheal occlusion improves lung morphometrics and function in the fetal lamb model of congenital diaphragmatic hernia. J Pediatr Surg. 2011 Jun 1;46(6):1150–7. pmid:21683214
  32. 32. Berenstein A, Lasjaunias P, brugge KG. Surgical Neuroangiography: Vol.2: Clinical and Endovascular Treatment Aspects in Adults. Springer Science & Business Media; 2012. 1321 p.
  33. 33. Alaraj A, Wallace A, Dashti R, Patel P, Aletich V. Balloons in Endovascular Neurosurgery: History and Current Applications. Neurosurgery. 2014 Feb 1;74(suppl_1):S163–90.
  34. 34. McKnight TE, Johnson A, Moise KJ, Ericson MN, Baba JS, Wilgen JB, et al. Remote Actuated Valve Implant. USPTO. 2014 Feb 25;USOO8657883B2:22.
  35. 35. McKnight TE, Johnson A, Moise KJ, Ericson MN, Baba JS, Wilgen JB, et al. Remote Actuated Valve Implant. USPTO. 2016 May 10;USOO9333067B2:23.
  36. 36. Cooperstein MA, Canavan HE. Assessment of cytotoxicity of (N -isopropyl acrylamide) and Poly (N -isopropyl acrylamide)-coated surfaces. Biointerphases. 2013 Dec;8(1):19. pmid:24706136
  37. 37. Andersen FA. Amended Final Report on the Safety Assessment of Polyacrylamide and Acrylamide Residues in Cosmetics. Int J Toxicol. 2005 Mar 1;24(2_suppl):21–50.
  38. 38. Peters EC, Svec F, Fréchet JMJ. Thermally responsive rigid polymer monoliths. Adv Mater. 1997 Jul 1;9(8):630–3.
  39. 39. Mutlu S, Yu C, Svec F, Mastrangelo CH, Frechet JMJ, Gianchandani YB. A thermally responsive polymer microvalve without mechanical parts photo-patterned in a parylene channel. In: TRANSDUCERS ‘03 12th International Conference on Solid-State Sensors, Actuators and Microsystems Digest of Technical Papers (Cat No03TH8664). 2003. p. 802–5 vol.1.
  40. 40. Yu C, Mutlu S, Selvaganapathy P, Mastrangelo CH, Svec F, Fréchet JMJ. Flow Control Valves for Analytical Microfluidic Chips without Mechanical Parts Based on Thermally Responsive Monolithic Polymers. Anal Chem. 2003 Apr;75(8):1958–61. pmid:12713057
  41. 41. Bae YH, Okano T, Kim SW. Temperature dependence of swelling of crosslinked poly(N,N′-alkyl substituted acrylamides) in water. J Polym Sci Part B Polym Phys. 1990 May 1;28(6):923–36.
  42. 42. Lokuge I, Wang X, Bohn PW. Temperature-Controlled Flow Switching in Nanocapillary Array Membranes Mediated by Poly (N -isopropylacrylamide) Polymer Brushes Grafted by Atom Transfer Radical Polymerization . Langmuir. 2007 Jan;23(1):305–11. pmid:17190519
  43. 43. Fletcher BL, McKnight TE, Fowlkes JD, Allison DP, Simpson ML, Doktycz MJ. Controlling the dimensions of carbon nanofiber structures through the electropolymerization of pyrrole. Synth Met. 2007 Apr 1;157(6):282–9.
  44. 44. Fletcher BL, Retterer ST, McKnight TE, Melechko AV, Fowlkes JD, Simpson ML, et al. Actuatable Membranes Based on Polypyrrole-Coated Vertically Aligned Carbon Nanofibers. ACS Nano. 2008 Feb 26;2(2):247–54. pmid:19206624
  45. 45. Argentiere S, Gigli G, Blasi MMIG and L. Smart Microfluidics: The Role of Stimuli- Responsive Polymers in Microfluidic Devices. Adv Microfluid [Internet]. 2012 [cited 2018 Aug 30]; Available from: https://www.intechopen.com/books/advances-in-microfluidics/smart-microfluidics-the-role-of-stimuli-responsive-polymers-in-microfluidic-devices
  46. 46. Ionov L. Hydrogel-based actuators: possibilities and limitations. Mater Today. 2014 Dec;17(10):494–503.
  47. 47. Ghosh A, Yoon C, Ongaro F, Scheggi S, Selaru FM, Misra S, et al. Stimuli-Responsive Soft Untethered Grippers for Drug Delivery and Robotic Surgery. Front Mech Eng [Internet]. 2017 [cited 2018 Nov 14];3. Available from: https://www.frontiersin.org/articles/10.3389/fmech.2017.00007/full
  48. 48. Beebe DJ, Moore JS, Bauer JM, Yu Q, Liu RH, Devadoss C, et al. Functional hydrogel structures for autonomous flow control inside microfluidic channels. Nature. 2000 Apr;404(6778):588–90. pmid:10766238
  49. 49. Tohru Saitoh, Yuichi Suzuki, Masataka Hiraide. Preparation of Poly(N-isopropylacrylamide)-Modified Glass Surface for Flow Control in Microfluidics. Anal Sci. 2002;18(2):203–5. pmid:11874128
  50. 50. Shen Z, Terao K, Maki Y, Dobashi T, Ma G, Yamamoto T. Synthesis and phase behavior of aqueous poly(N-isopropylacrylamide-co-acrylamide), poly(N-isopropylacrylamide-co-N, N-dimethylacrylamide) and poly (N-isopropylacrylamide-co-2-hydroxyethylmethacrylate). Colloid Polym Sci. 2006 Jun 1;284(9):1001–7.