Figures
Abstract
Background
To investigate the correlation between cerebral (SO2-transcranial), retinal arterial (SaO2-retinal) and venous (SvO2-retinal) oxygen saturation as measured by near-infrared spectroscopy (NIRS) and retinal oximetry respectively.
Methods
Paired retinal and cerebral oxygen saturation measurements were performed in healthy volunteers. Arterial and venous retinal oxygen saturation and diameter were measured using a non-invasive spectrophotometric retinal oximeter. Cerebral oxygen saturation was measured using near-infrared spectroscopy. Correlations between SO2-transcranial and retinal oxygen saturation and diameter measurements were assessed using Pearson correlation coefficients. Lin’s concordance correlation coefficient (CCC) and Bland-Altman analysis were performed to evaluate the agreement between SO2-transcranial as measured by NIRS and as estimated using a fixed arterial:venous ratio as 0.3 x SaO2-retinal + 0.7 x SvO2-retinal. The individual relative weight of SaO2-retinal and SvO2-retinal to obtain the measured SO2-transcranial was calculated for all subjects.
Results
Twenty-one healthy individuals aged 26.4 ± 2.2 years were analyzed. SO2-transcranial was positively correlated with both SaO2-retinal and SvO2-retinal (r = 0.44, p = 0.045 and r = 0.43, p = 0.049 respectively) and negatively correlated with retinal venous diameter (r = -0.51, p = 0.017). Estimated SO2-transcranial based on retinal oximetry showed a tolerance interval of (-13.70 to 14.72) and CCC of 0.46 (95% confidence interval: 0.05 to 0.73) with measured SO2-transcranial. The average relative weights of SaO2-retinal and SvO2-retinal to obtain SO2-transcranial were 0.31 ± 0.11 and 0.69 ± 0.11, respectively.
Conclusion
This is the first study to show the correlation between retinal and cerebral oxygen saturation, measured by NIRS and retinal oximetry. The average relative weight of arterial and venous retinal oxygen saturation to obtain the measured transcranial oxygen saturation as measured by NIRS, approximates the established arterial:venous ratio of 30:70 closely, but shows substantial inter-individual variation. These findings provide a proof of concept for the role of retinal oximetry in evaluating cerebral oxygenation.
Citation: Van Keer K, Van Keer J, Barbosa Breda J, Nassiri V, De Deyne C, Genbrugge C, et al. (2018) Non-invasive assessment of cerebral oxygenation: A comparison of retinal and transcranial oximetry. PLoS ONE 13(1): e0190612. https://doi.org/10.1371/journal.pone.0190612
Editor: Manabu Sakakibara, Tokai University, JAPAN
Received: August 10, 2017; Accepted: December 18, 2017; Published: January 5, 2018
Copyright: © 2018 Van Keer et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Data Availability: All data are available in the Supporting information section.
Funding: The authors received no specific funding for this work.
Competing interests: The authors have declared that no competing interests exist.
Introduction
The brain is highly dependent on aerobic metabolism to meet its energy demand. [1] Prolonged ischemia leads to neuronal injury and is associated with poor neurological outcome. [2–7] Because severe cerebral ischemia may exist in the absence of systemic hypoxia, there is a need for selective assessment of the central nervous system oxygenation. [8,9]
Transcranial near-infrared spectroscopy (NIRS) is increasingly used in clinical practice to measure cerebral saturation non-invasively. [10] NIRS relies on the differences in the absorption spectra of a near-infrared light beam between oxygenated and de-oxygenated hemoglobin. [11] NIRS has proven useful in a variety of clinical settings, especially during cardiac surgery. [4,12,13] However, current NIRS devices have several drawbacks. First, despite the use of different source-detector distances, NIRS is not entirely selective for the cerebral saturation: the measured oxygen saturation is "contaminated" by the saturation of the more superficial dura, skull and skin. [14,15] Second, the saturation measured by NIRS is in fact a composite of arterial, capillary and venous blood. [16] For the FORE-SIGHT® transcranial oximeter (CAS Medical Systems, Branford, CT, USA) the relative contribution of arterial and venous blood (arterial:venous ratio) to the measured composite value is 30:70, but the inter-individual variability of this ratio is considerable. [16–20] This is especially relevant in critically ill patients in whom altered perfusion pressures, disturbed (auto)regulation and administration of vasoactive drugs are likely to influence this ratio. [21–23]
These shortcomings could partially be addressed by looking at the central nervous system oxygen saturation from a different point of view. Thanks to the transparency of the ocular media, the human eye offers a convenient way to study the circulatory system. In recent years, several new diagnostic tools have been developed to investigate the retinal circulation. Dual-wavelength spectrophotometric retinal oximetry is an imaging technique that allows the visualization and measurement oxygen saturation in retinal blood vessels by comparing the relative reflectance of retinal blood vessels at 570nm and 600nm. [24] At 570nm light, oxidized and deoxidized hemoglobin absorb light equally (isosbestic wavelength), while at 600 nm deoxidized hemoglobin absorbs more light relative to oxidized hemoglobin (non-isosbestic wavelength). [25] (Fig 1). The retinal oximeter simultaneously acquires monochromatic retinal fundus pictures at 570nm and 600nm using a dual camera setup coupled to a conventional fundus camera. Upon acquisition, the unmodified xenon flash from the fundus camera illuminates the retina. The reflected light is focused in the funduscamera. Unlike in a conventional funduscamera, the image is not directly captured by a single digital camera. Rather, the optical pathway is split by a beam splitter, filtered by two narrow band-pass filters of respectively 570nm and 600nm and subsequently registered on two different digital cameras. [24] (Figs 1 and 2) By analyzing the ratio of the optical density of retinal blood vessels at both wavelengths, oxygen saturation inside the retinal blood vessels can be calculated based on the nearly linear inverse relationship between optical density ratio and hemoglobin oxygen saturation. [25] A more in-depth theoretical explanation has been published elsewhere. [26] The oxygen saturation inside the retinal vessels is displayed as a pseudo-color overlay on top of the retinal fundus picture and is measured manually by selecting a retinal blood vessel. Vessels can then be labeled as either arteries or veins based on saturation and continuity with feeder vessels. (Fig 2) Hence, in contrast to NIRS, retinal oximetry does provide a way to measure arterial and venous oxygen saturations separately at a high spatial resolution of less than 10μm. [24,27,28] Using the unmodified xenon flash of a conventional funduscamera, which is safe for human use according to American National Standards Institute (ANSI) standards, and requiring only pupil dilation, retinal oximetry provides a safe and convenient imaging modality. [29,30] Over the last decade, retinal oximetry has become an established tool for the assessment of oxygen saturation in different ocular diseases. [31–36]However, the impact of systemic parameters on retinal vessel oxygen saturation has been investigated less extensively so far. [37–41] As the retinal and cerebral circulation share a common embryological, anatomical and physiological background, retinal oximetry may also be a valuable tool in assessing cerebral oxygenation. [42–45]
Experimental setup showing seated subject, retinal oximeter, NIRS device, sphygmomanometer and pulse oximeter. Magnified detail of the dual-wavelength retinal oximeter consisting of a funduscamera, beam splitter, two band-pass filters and two digital cameras.
Sample monochromatic retinal image at 570nm (A) and 600nm (B) and resulting pseudo-color oxygen saturation overlay map (C), showing oxygen saturation in arteries (red) and veins (green) separately. Note the darker appearance of veins compared to arteries at 600nm (B) but not at 570nm (A).
This pilot study aims to provide a proof of concept for the use of retinal oximetry in assessing cerebral oxygen saturation by investigating the correlation between retinal arterial (SaO2-retinal) and venous oxygen saturation (SvO2-retinal) measured by spectrophotometric retinal oximetry and cerebral oxygen saturation (SO2-transcranial) measured by NIRS.
Methods
Study population
Healthy adult volunteers were recruited at the Department of Ophthalmology of the University Hospitals Leuven. Subjects taking vasoactive medication or with a history of ocular, vascular or neurological diseases were excluded. A written informed consent in accordance with the tenets of the Declaration of Helsinki was obtained from each participant prior to any examinations. The study was approved by the local ethics committee (Medical Ethical Board of the University Hospitals Leuven, Belgium ML9229).
Experimental design
Subjects were asked to refrain from caffeine, nicotine and physical exercise for 12h before testing. After instillation of phenylephrine 5% (Neosynephrine-POS, Ursapharm, Saarbrucken, Germany) and Tropicamide 0,5% (Tropicol, Théa Pharma, Wetteren, Belgium) in the left eye, a standard protocol, that has shown to result in good pupil dilation (required for the retinal oximeter that was used in this study) without influencing retinal oxygen saturation values, was performed. [46] Participants were seated in an upright position for at least 20 minutes prior to image acquisition. A pulse oximeter was placed on the left index finger (Ohmeda TruSat Pulse Oximeter, GE Healthcare, Finland) and an automated sphygmomanometer was placed on the right upper arm (Omron HEM-7001-E, Omron, Kyoto, Japan). The NIRS probe was placed on the left side of the forehead below the hairline facing the left frontal lobe (FORE-SIGHT® technology, CAS Medical Systems, Branford, CT, USA). Heart rate, pulse oxygen saturation (SO2-pulse), systolic and diastolic blood pressure were measured. SO2-transcranial was recorded once the pupil was dilated to a diameter of at least 5mm and continuously monitored during 10 minutes. Following two minutes of dark-adaptation, retinal fundus images were acquired using a Oxymap® T1 oximeter (Oxymap ehf., Reykjavik, Iceland). The experimental setup is summarized in Fig 1.
The quality of the oximetry images was evaluated using a standardized protocol. Images were subjectively rated on focus, contrast, glare and shadow using a pass/fail system. Failure to pass any of these criteria led to exclusion of the image. [46] Image analysis was performed with the Oxymap® Analyzer software version 2.4.0 using the following protocol: first-degree retinal vessels with a width of more than six pixels and a length between 50 and 200 pixels were selected from the color-coded saturation map. Second-degree vessels were only analyzed if first-degree vessels did not meet the criteria. Around the optic disc, an area of 15 pixels as well as branching vessels and their origin were manually excluded. Vessels were manually categorized as arteries or veins and subsequently averaged over the four quadrants. Vessel diameters were measured automatically using the Oxymap® Analyzer software. [36]
Estimated SO2-transcranial was calculated using the established relative arterial and venous weight of the FORE-SIGHT® transcranial oximeter of 30:70 using Eq 1. [18]
(1)
Individual arterial (wi) and venous (1 − wi) relative weight to obtain measured SO2-transcranial was calculated in all subjects using Eq 2.
(2)
For example, in participant i with a SaO2-retinal of 95%, a SvO2-retinal of 60% and a SO2-transcranial of 70%, the arterial (wi) and venous weight (1 − wi) would be 0.29 and 0.71 respectively, since 0.29 x 95% + 0.71 x 60% = 70%.
The arterial and venous oxygen saturation contributing to the measured SO2-transcranial were estimated by substituting SO2-pulse for arterial and by estimating venous oxygen saturation (estimated SvO2-transcranial) using Eq 3.
Statistical analysis
Statistical analysis was performed with R (R Statistical Software version 3.4.1; R Foundation for Statistical Computing, Vienna, Austria) using the “BlandAltmanLeh”, “epiR” and “Tolerance” package. [47–49] Normality of continuous variables was tested using the Shapiro-Wilk test. Demographics and measurements were analyzed with descriptive statistics (mean ± standard deviation). Correlations were evaluated using Pearson correlation coefficients. The agreement between measured SO2-transcranial and estimated SO2-transcranial was evaluated using Lin’s concordance correlation coefficient (CCC) and Bland-Altman analysis. [50,51] In terms of bias, mean of the difference between SO2-transcranial and estimated SO2-transcranial, its 95% confidence intervals and tolerance intervals (for 99% of future measurements to be within 95% confidence) were calculated. Statistical significance was based on two-sided p-values of <0.05. As this study was set up as a pilot study, power analysis and sample-size calculations were not performed. Individual data points and the R script that was used to generate the statistical analyses and figures can were uploaded as supporting information. (S1 File, S2 File)
Results
A total of 22 healthy individuals, 6 men and 16 women, were initially recruited. One subject was excluded because of inadequate quality of the oximetry images. Descriptive statistics of demographic characteristics of the included subjects are summarized in Table 1.
Average SO2-pulse, SaO2-retinal, SvO2-retinal and SO2-transcranial were 98.2 ± 1.1%, 95.3 ± 2.1%, 61.6 ± 4.0% and 72.2 ± 3.5%, respectively. SO2-transcranial showed a significant positive correlation with both SaO2-retinal and SvO2-retinal (r = 0.44, p = 0.045 and r = 0.43, p = 0.049, respectively) and a negative correlation with venous diameter (r = -0.51, p = 0.017). (Table 2) Correlations between retinal, peripheral and transcranial saturations per vessel type are shown in Table 3.
The estimated SO2-transcranial based on the fixed arterial:venous ratio of 30:70 showed a CCC of 0.46 (95% confidence interval: 0.05 to 0.73) with the measured SO2-transcranial. Bias between measured and estimated SO2-transcranial was 0.51 with a 95% confidence and 95% tolerance interval of -1.07 to 2.10 and -13.70 to 14.72 respectively. Bland-Altman analysis is shown in Fig 3.
Plot shows mean bias (solid line) and 95% limit of agreement (dashed lines).
The average relative weights of SaO2-retinal and SvO2-retinal to obtain the measured SO2-transcranial were 0.31 ± 0.11 and 0.69 ± 0.11, respectively, with large inter-individual variability ranging from 10 to 49 for arterial weight, as summarized in Fig 4. The weighs for each individual are computed using Eq 2.
Plot showing individual arterial weights for each study participant (Eq 2). Average arterial weight is 0.31 (solid horizontal line) but inter-individual differences are large, ranging from 0.10 to 0.49.
Discussion
The human retina provides a unique view on the cerebral vasculature. Many studies have pointed out structural and functional homologies between the retinal and the cerebral circulation. [42–45] Structural parameters in the retinal vascular tree have been correlated with a variety of neurological diseases. [45,52,53] The added value of functional retinal vascular parameters has been investigated less extensively. Denninghoff et al demonstrated in a swine model that SvO2-retinal correlates well with mixed venous oxygen saturation. [54] Mixed venous oxygen saturation is a valuable tool in the management of critically ill patients as a surrogate for cardiac output. In contrast to mixed venous oxygen saturation measured by a pulmonary artery catheter, retinal venous oxygen saturation can be measured non-invasively. This is the first study to investigate the correlation between retinal vessel oxygen saturation measured non-invasively by spectrophotometric oximetry, and cerebral oxygen saturation measured by NIRS.
First, this study shows that both SaO2-retinal and SvO2-retinal, measured by retinal oximetry, are positively correlated with SO2-transcranial, measured by NIRS. The analysis per vessel type revealed a significant correlation between measured and estimated SO2-transcranial. Where previous studies have demonstrated a significant correlation between SaO2-retinal and SO2-pulse, our study could not reproduce this finding. [55] Of interest is the relative low retinal arterial saturation compared to SO2-pulse, which has been reported also in previous studies and is thought to result from a counter-current effect happening in the narrow space between the central retinal artery and vein inside the optic nerve. [56] SvO2-retinal showed a borderline significance with estimated SvO2-transcranial. The above mentioned findings strengthen the hypothesis that the structural similarities between retinal and cerebral circulation also extend to a functional level. Apart from a common embryonic origin, the retinal and cerebral circulation have a common vascularization pattern during development, a non-fenestrated endothelial barrier, are covered by pericytes and glial cells and share the same control mechanisms. [43,45,57] Notable differences between both circulations include the lack of anastomoses and the distinct vascular branching pattern of the retinal circulation. [58] Local autoregulation also shows different responses between the cerebral and the retinal circulation with the retinal circulation being less reactive to hypercapnia and more reactive to hyperoxia. [59, 60] These differences in vascular reactivity should be taken into consideration when extrapolating cerebral oxygen saturation from retinal measurements.
Second, we found that the average relative partition of SaO2-retinal and SvO2-retinal to obtain the measured SO2-transcranial closely matches the established arterial:venous ratio of 30:70 calibration of FORE-SIGHT® NIRS technology. [18,20] Since the arteries supplying the retina and the frontal cortex are both direct branches of the internal carotid artery, it is reasonable to assume that the differences in arterial oxygen saturation between both circulations are small. However, the relative weight of SaO2-retinal and SvO2-retinal to SO2-transcranial showed large inter-individual variation ranging from 10% to 49% for the arterial and from 51% to 90% for the venous contribution. These variations are in line with studies that investigated the relative contribution of arterial and venous cerebral oxygen saturation to SO2-transcranial. [16,20] The relatively low CCC between measured SO2-transcranial and estimated SO2-transcranial could therefore imply that retinal oximetry, being able to measure arterial and venous saturation separately, measures some additional information that is averaged out by NIRS.
Third, and of particular interest, is the negative correlation between retinal venous diameter and SO2-transcranial. As explained above, the retinal and cerebral circulation share similar vascular regulatory mechanisms. As such, a possible explanation for these findings could be that dilated retinal veins are indicative of cerebral venodilatation which would in turn result in a higher relative contribution of the venous compartment in SO2-transcranial. [43,44,61] Moreover, retinal venous diameter is known to increase in case of increased intracranial pressure. [62,63] Because of the known influence of vessel diameter on retinal oximetry measurements, a correction for vessel diameter is automatically applied by the provided software. [64, 65] Combined with the fact that retinal oximetry has the advantage of allowing to separately measure arterial and venous oxygen saturation, retinal oximetry is expected to be less influenced by vessel diameter than NIRS.
Taken together, we believe these findings provide proof of concept for the potential added value of retinal oximetry in assessing cerebral oxygen saturation. The ability of retinal oximetry to easily measure arterial and venous oxygen saturation separately, without "contamination" of superficial peripheral circulation may be especially relevant in situations where the relative contribution of arterial and venous saturation might deviate from the established arterial:venous ratio, for instance during extracorporeal circulation or in the post-resuscitation setting, when the accuracy of NIRS could be limited due to therapeutic hypothermia, vasopressor medication, alterations in (auto)regulation of the cerebral and superficial circulation,… [14,66]
Current retinal oximeters have several limitations that inhibit the use in the settings mentioned above. First, retinal oximetry provides a snapshot in time but not a continuous measurement. Next, the size of the current retinal oximeters limits its bedside usability. In addition, the price is steep. However, recently developed hyperspectral image sensors at the size of a regular color image sensor will likely provide an answer to both issues. [67] By replacing the color image sensor with a hyperspectral sensor, any portable funduscamera could be modified into a portable retinal oximeter. Provided that retinal oximetry could be performed at the bedside, this technology could become an interesting tool in the hemodynamic and cerebral assessment of critically ill patients and hence also contribute to a better understanding of NIRS.
The present study must be interpreted within the context of its potential limitations. First, the lack of exact measurements of the true cerebral saturation precludes firm conclusions about the correlation between retinal and cerebral oxygen saturation. While we believe several of our findings point towards this correlation, it remains to be investigated in future studies. Second, a major limitation of this study is the facts that measurements were performed only under normoxic conditions with an oxygen concentration of 21%. Future studies measuring desaturation data during graded hypoxia are warranted. Third, whereas SO2-transcranial is measured continuously, retinal oximetry only provides a snapshot of the saturation in the retinal vessel at one particular moment. We tried to accommodate for this potential source of error by measuring the retinal oxygen saturation in steady state conditions after at least 20 minutes of adaptation in a seated position. However, this limitation might hamper future studies in non-steady state conditions. Last, we recognize the small sample size of the study. However, we believe this pilot study does provide a proof of concept for the relevant role of retinal oximetry in estimating cerebral oxygen saturation.
Conclusion
In conclusion, this is the first study to describe the relation between retinal and cerebral oxygen saturation measured by retinal oximetry and NIRS. The ability of retinal oximetry to measure arterial and venous oxygen saturation separately without "contamination" of superficial peripheral circulation may have future applications in clinical practice. Further studies are warranted to investigate the exact relation of retinal oximetry with cerebral and systemic oxygen saturation.
Supporting information
S1 File. Raw data.
Individual data points for all subjects.
https://doi.org/10.1371/journal.pone.0190612.s001
(CSV)
S2 File. Statistical analyses and figures.
R script for all analyses and figures included in the paper.
https://doi.org/10.1371/journal.pone.0190612.s002
(R)
References
- 1. Erecińska M, Silver IA (1994) Ions and energy in mammalian brain. Prog Neurobiol 43: 37–71. pmid:7972852
- 2. van Dijk D, Keizer AMA, Diephuis JC, Durand C, Vos LJ, et al. (2000) Neurocognitive dysfunction after coronary artery bypass surgery: A systematic review. J Thorac Cardiovasc Surg 120: 632–639. pmid:11003741
- 3. Hogue CW, Palin CA, Arrowsmith JE (2006) Cardiopulmonary bypass management and neurologic outcomes: an evidence-based appraisal of current practices. Anesthesia & Analgesia 103: 21–37. pmid:16790619
- 4. Murkin JM, Adams SJ, Novick RJ, Quantz M, Bainbridge D, et al. (2007) Monitoring brain oxygen saturation during coronary bypass surgery: a randomized, prospective study. Anesthesia & Analgesia 104: 51–58. pmid:17179242
- 5. Slater JP, Guarino T, Stack J, Vinod K, Bustami RT, et al. (2009) Cerebral oxygen desaturation predicts cognitive decline and longer hospital stay after cardiac surgery. Ann Thorac Surg 87: 36–44–discussion44–5. pmid:19101265
- 6. Meex I, Dens J, Jans F, Boer W, Vanhengel K, et al. (2013) Cerebral tissue oxygen saturation during therapeutic hypothermia in post-cardiac arrest patients. Resuscitation 84: 788–793. pmid:23313422
- 7. Coles JP. Regional ischemia after head injury. Curr Opin Crit Care. 2004 Apr;10(2):120–5. pmid:15075722
- 8. Yeh T, Gouldman J, Auden SM, Seremet V, Edmonds HL, et al. (2001) Mixed venous oxygen saturation does not adequately predict cerebral perfusion during pediatric cardiopulmonary bypass. J Thorac Cardiovasc Surg 122: 192–193. pmid:11436060
- 9. Hadolt I, Litscher G (2003) Noninvasive assessment of cerebral oxygenation during high altitude trekking in the Nepal Himalayas (2850–5600 m). Neurol Res 25: 183–188. pmid:12635520
- 10. Murkin JM (2009) NIRS: a standard of care for CPB vs. an evolving standard for selective cerebral perfusion? J Extra Corpor Technol 41: P11–P14.
- 11. Jöbsis FF (1977) Noninvasive, infrared monitoring of cerebral and myocardial oxygen sufficiency and circulatory parameters. Science 198: 1264–1267. pmid:929199
- 12. Edmonds HL Jr (2006) Pro: All Cardiac Surgical Patients Should Have Intraoperative Cerebral Oxygenation Monitoring. J Cardiothorac Vasc Anesth 20: 445–449. pmid:16750751
- 13. Murkin JM (2013) Is it better to shine a light, or rather to curse the darkness? Cerebral near-infrared spectroscopy and cardiac surgery. Eur J Cardiothorac Surg 43: 1081–1083. pmid:23543201
- 14. Davie SN, Grocott HP (2012) Impact of extracranial contamination on regional cerebral oxygen saturation: a comparison of three cerebral oximetry technologies. Anesthesiology 116: 834–840. pmid:22343469
- 15. Hirasawa A, Yanagisawa S, Tanaka N, Funane T, Kiguchi M, et al. (2014) Influence of skin blood flow and source-detector distance on near-infrared spectroscopy-determined cerebral oxygenation in humans. Clinical Physiology and Functional Imaging 35: 237–244. pmid:24750947
- 16. Watzman HM, Kurth CD, Montenegro LM, Rome J, Steven JM, et al. (2000) Arterial and venous contributions to near-infrared cerebral oximetry. Anesthesiology 93: 947–953. pmid:11020744
- 17. McCormick PW, Stewart M, Goetting MG, Balakrishnan G (1991) Regional cerebrovascular oxygen saturation measured by optical spectroscopy in humans. Stroke 22: 596–602. pmid:2028488
- 18. Ito H, Kanno I, Iida H, Hatazawa J, Shimosegawa E, et al. (2001) Arterial fraction of cerebral blood volume in humans measured by positron emission tomography. Ann Nucl Med 15: 111–116. pmid:11448068
- 19. Pattinson KTS, Imray CHE, Wright AD (2005) What does cerebral oximetry measure? British Journal of Anaesthesia 94: 863–authorreply863–4. pmid:15878895
- 20. Ikeda K, MacLeod DB, Grocott HP, Moretti EW, Ames W, et al. (2014) The accuracy of a near-infrared spectroscopy cerebral oximetry device and its potential value for estimating jugular venous oxygen saturation. Anesthesia & Analgesia 119: 1381–1392. pmid:25313967
- 21. Wong FY, Alexiou T, Samarasinghe T, Brodecky V, Walker AM (2010) Cerebral arterial and venous contributions to tissue oxygenation index measured using spatially resolved spectroscopy in newborn lambs. Anesthesiology 113: 1385–1391. pmid:21068664
- 22. Ogoh S, Sato K, Fisher JP, Seifert T, Overgaard M, et al. (2011) The effect of phenylephrine on arterial and venous cerebral blood flow in healthy subjects. Clinical Physiology and Functional Imaging 31: 445–451. pmid:21981455
- 23. Meng L, Gelb AW, Alexander BS, Cerussi AE, Tromberg BJ, et al. (2012) Impact of phenylephrine administration on cerebral tissue oxygen saturation and blood volume is modulated by carbon dioxide in anaesthetized patients. British Journal of Anaesthesia 108: 815–822. pmid:22391890
- 24. Hardarson SH, Harris A, Karlsson RA, Halldorsson GH, Kagemann L, et al. (2006) Automatic Retinal Oximetry. Investigative Ophthalmology & Visual Science 47: 5011–5016. pmid:17065521
- 25. Beach JM, Schwenzer KJ, Srinivas S, Kim D, Tiedeman JS (1999) Oximetry of retinal vessels by dual-wavelength imaging: calibration and influence of pigmentation. J Appl Physiol 86: 748–758. pmid:9931217
- 26. Beach J. (2014). Pathway to retinal oximetry. Translational vision science & technology, 3(5), 2–2.
- 27. Blondal R, Sturludottir MK, Hardarson SH, Halldorsson GH, Stefánsson E (2011) Reliability of vessel diameter measurements with a retinal oximeter. Graefes Arch Clin Exp Ophthalmol 249: 1311–1317. pmid:21499769
- 28. Geirsdottir A, Palsson O, Hardarson SH, Olafsdottir OB, Kristjansdottir JV, et al. (2012) Retinal Vessel Oxygen Saturation in Healthy Individuals. Investigative Ophthalmology & Visual Science 53: 5433–5442. pmid:22786895
- 29. Delori FC, Webb RH, Sliney DH (2007) Maximum permissible exposures for ocular safety (ANSI 2000), with emphasis on ophthalmic devices. J Opt Soc Am A, JOSAA 24: 1250–1265.
- 30. Lux AL, Mouriaux F, Guillois B, Fedrizzi S, Peyro-Saint-Paul L, et al. (2015) Serious adverse side effects after pupillary dilation in preterm infants. Journal francais d'ophtalmologie 38: 193–198. pmid:25726252
- 31. Hardarson SH, Stefánsson E (2010) Oxygen saturation in central retinal vein occlusion. American Journal of Ophthalmology 150: 871–875. pmid:20875633
- 32. Hardarson SH, Elfarsson A, Agnarsson BA, Stefánsson E (2013) Retinal oximetry in central retinal artery occlusion. Acta Ophthalmologica 91: 189–190. pmid:22429810
- 33. Türksever C, Valmaggia C, Orguel S, Schordert DF, Todorova MG (2013) Retinal vessel oxygen saturation in retinitis pigmentosa patients. Acta Ophthalmologica 91: 0–0.
- 34. Jørgensen CM, Hardarson SH, Bek T (2014) The oxygen saturation in retinal vessels from diabetic patients depends on the severity and type of vision-threatening retinopathy. Acta Ophthalmologica 92: 34–39. pmid:24330421
- 35. Geirsdottir A, Hardarson SH, Olafsdottir OB, Stefánsson E (2014) Retinal oxygen metabolism in exudative age-related macular degeneration. Acta Ophthalmologica 92: 27–33. pmid:24447786
- 36. Olafsdottir OB, Vandewalle E, Abegão Pinto L, Geirsdottir A, De Clerck E, et al. (2014) Retinal oxygen metabolism in healthy subjects and glaucoma patients. Br J Ophthalmol 98: 329–333. pmid:24403567
- 37. Traustason S, Jensen AS, Arvidsson HS, Munch IC, Søndergaard L, et al. (2011) Retinal oxygen saturation in patients with systemic hypoxemia. Investigative Ophthalmology & Visual Science 52: 5064–5067. pmid:21467173
- 38. Palkovits S, Lasta M, Boltz A, Schmidl D, Kaya S, et al. (2013) Measurement of retinal oxygen saturation in patients with chronic obstructive pulmonary disease. Investigative Ophthalmology & Visual Science 54: 1008–1013. pmid:23307953
- 39. Einarsdottir AB, Hardarson SH, Kristjansdottir JV, Bragason DT, Snaedal J, et al. (2016) Retinal Oximetry Imaging in Alzheimer’s Disease. Journal of Alzheimer's Disease 49: 79–83. pmid:26444785
- 40. Stefánsson E, Olafsdottir OB, Einarsdottir AB, Eliasdottir TS, Eysteinsson T, et al. (2017) Retinal Oximetry Discovers Novel Biomarkers in Retinal and Brain Diseases. Invest Ophthalmol Vis Sci 58: BIO227–BIO233. pmid:28810002
- 41. Eliasdottir TS, Bragason D, Hardarson SH, Vacchiano C, Gislason T, et al. (2017) Retinal oximetry measures systemic hypoxia in central nervous system vessels in chronic obstructive pulmonary disease. PLoS ONE 12: e0174026. pmid:28328974
- 42. Riva CE, Sinclair SH, Grunwald JE (1981) Autoregulation of retinal circulation in response to decrease of perfusion pressure. Investigative Ophthalmology & Visual Science 21: 34–38. pmid:7251300
- 43. Hardy P, Varma DR, Chemtob S (1997) Control of cerebral and ocular blood flow autoregulation in neonates. Pediatr Clin North Am 44: 137–152. pmid:9057788
- 44. Delaey C, Van De Voorde J (2000) Regulatory mechanisms in the retinal and choroidal circulation. Ophthalmic Res 32: 249–256. pmid:11015035
- 45. Patton N, Aslam T, MacGillivray T, Pattie A, Deary IJ, et al. (2005) Retinal vascular image analysis as a potential screening tool for cerebrovascular disease: a rationale based on homology between cerebral and retinal microvasculatures. Journal of Anatomy 206: 319–348. pmid:15817102
- 46. Vandewalle E, Abegão Pinto L, Olafsdottir OB, Stalmans I (2012) Phenylephrine 5% added to Tropicamide 0.5% eye drops does not influence retinal oxygen saturation values or retinal vessel diameter in glaucoma patients. Acta Ophthalmologica 91: 733–737. pmid:22998076
- 47.
Lehnert B (2014) BlandAltmanLeh: plots (slightly extended) Bland-Altman plots. R package version 01 0.
- 48.
Stevenson M, Nunes T, Sanchez J, Thornton R, Reiczigel J, et al. (2013) epiR: An R package for the analysis of epidemiological data. R package version 0.9–48. Vienna, Austria: R Foundation for Statistical Computing.
- 49. Young DS (2010) Tolerance: an R package for estimating tolerance intervals. Journal of Statistical Software 36: 1–39.
- 50. Lawrence I, Lin K (1989) A concordance correlation coefficient to evaluate reproducibility. Biometrics: 255–268. pmid:2720055
- 51. Bland JM, Altman DG (1999) Measuring agreement in method comparison studies. Stat Methods Med Res 8: 135–160. pmid:10501650
- 52. Wong TY, Mosley TH, Klein R, Klein BEK, Sharrett AR, et al. (2003) Retinal microvascular changes and MRI signs of cerebral atrophy in healthy, middle-aged people. Neurology 61: 806–811. pmid:14504325
- 53. Wong T-Y, Klein R, Couper DJ, Cooper LS, Shahar E, Hubbard LD, et al. Retinal microvascular abnormalities and incident stroke: the Atherosclerosis Risk in Communities Study. Lancet. Elsevier; 2001;358(9288):1134–40. pmid:11597667
- 54. Denninghoff KR, Smith MH, Lompado A, Hillman LW (2003) Retinal venous oxygen saturation and cardiac output during controlled hemorrhage and resuscitation. Journal of Applied Physiology 94: 891–896. pmid:12571124
- 55. Palkovits S, Told R, Schmidl D, Boltz A (2014) Regulation of retinal oxygen metabolism in humans during graded hypoxia. Am J Physiol Heart Circ Physiol 307: H1412–1418. pmid:25217648
- 56. Eliasdottir TS, Bragason D, Hardarson SH, Vacchiano C, Gislason T, et al. (2017) Retinal oximetry measures systemic hypoxia in central nervous system vessels in chronic obstructive pulmonary disease. PLoS ONE 12: e0174026. pmid:28328974
- 57. Li G, Shih Y-YI, Kiel JW, La Garza De BH, Du F, et al. (2013) MRI study of cerebral, retinal and choroidal blood flow responses to acute hypertension. Experimental Eye Research 112: 118–124. pmid:23623996
- 58. Cogan DG, Kuwabara T (1984) Comparison of retinal and cerebral vasculature in trypsin digest preparations. British Journal of Ophthalmology 68: 10–12. pmid:6689929
- 59. Kisilevsky M, Mardimae A, Slessarev M, Han J, Fisher J, et al. (2008) Retinal Arteriolar and Middle Cerebral Artery Responses to Combined Hypercarbic/Hyperoxic Stimuli. Invest Ophthalmol Vis Sci 49: 5503–5509. pmid:18502990
- 60. Luksch A, Garhofer G, Imhof A, Polak K, Polska E, et al. (2002) Effect of inhalation of different mixtures of O(2) and CO(2) on retinal blood flow. British Journal of Ophthalmology 86: 1143–1147. pmid:12234896
- 61. Lassen NA (1964) Autoregulation of cerebral blood flow. Circ Res 15: SUPPL:201–204. pmid:14206303
- 62. Rios-Montenegro EN, Anderson DR, David NJ (1973) Intracranial pressure and ocular hemodynamics. Arch Ophthalmol 89: 52–58. pmid:4630889
- 63. Moss HE, Treadwell G, Wanek J, DeLeon S, Shahidi M (2014) Retinal Vessel Diameter Assessment in Papilledema by Semi-Automated Analysis of SLO Images: Feasibility and ReliabilityRetinal Vessel Diameters in Papilledema. Invest Ophthalmol Vis Sci 55: 2049–2054. pmid:24609623
- 64. Hammer M, Vilser W, Riemer T, Schweitzer D (2008) Retinal vessel oximetry-calibration, compensation for vessel diameter and fundus pigmentation, and reproducibility. J Biomed Opt 13: 054015. pmid:19021395
- 65.
Hardarson S, Traustason S, Kristjansdottir J, Olafsdottir O, Eliasdottir T, et al. (2013) Standardization of Retinal Vessel Oximetry and the Influence of Vessel Diameter. ARVO Meeting Abstracts 54: 48
- 66. Absalom AR, Scheeren TWL (2013) NIRS during therapeutic hypothermia: cool or hot? Resuscitation 84: 720–721. pmid:23535209
- 67. Kaluzny J, Li H, Liu W, Nesper P, Park J, et al. (2017) Bayer filter snapshot hyperspectral fundus camera for human retinal imaging. Curr Eye Res 42: 629–635 pmid:27767345