Figures
Abstract
Objectives
Magnetic resonance imaging (MRI) of the chest has long suffered from its sensitivity to respiratory and cardiac motion with an intrinsically low signal to noise ratio and a limited spatial resolution. The purpose of this study was to perform chest MRI under an adapted non invasive pulsatile flow ventilation system (high frequency percussive ventilation, HFPV®) allowing breath hold durations 10 to 15 times longer than other existing systems.
Methods
One volunteer and one patient known for a thymic lesion underwent a chest MRI under ventilation percussion technique (VP-MR). Routinely used sequences were performed with and without the device during three sets of apnoea on inspiration.
Results
VP-MR was well tolerated in both cases. The mean duration of the thoracic stabilization was 10.5 min (range 8.5–12) and 5.8 min (range 5–6.2) for Volunteer 1 and Patient 1, respectively. An overall increased image quality was seen under VP-MR with a better delineation of the mediastinal lesion for Patient 1. Nodules discovered in Volunteer 1 were confirmed with low dose CT.
Citation: Beigelman-Aubry C, Peguret N, Stuber M, Delacoste J, Belmondo B, Lovis A, et al. (2017) Chest-MRI under pulsatile flow ventilation: A new promising technique. PLoS ONE 12(6): e0178807. https://doi.org/10.1371/journal.pone.0178807
Editor: Gayle E. Woloschak, Northwestern University Feinberg School of Medicine, UNITED STATES
Received: February 19, 2016; Accepted: May 19, 2017; Published: June 12, 2017
Copyright: © 2017 Beigelman-Aubry et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Data Availability: The data are accessible through Open Science Framework at the following DOI: 10.17605/OSF.IO/9XK75.
Funding: The authors received no specific funding for this work.
Competing interests: The authors have declared that no competing interests exist.
Introduction
Magnetic resonance imaging (MRI), a radiation-free technique, has long suffered from the low proton density and high magnetic susceptibility of the lungs [1]. These characteristics explain the intrinsically low signal to noise ratio and limited spatial resolution with a great sensitivity to respiratory and cardiac motions, that are most prominent in the lower and anterior sections of the chest [2]. Repetitive breath holds of around 20 seconds each are required for most MR acquisitions split into several blocks with possible additional artefacts in case of irreproducible breath holds [2]. Dedicated techniques for respiratory motion compensation have also been developed, including respiratory belts, navigator [3] and most recently 3D free-breathing isotropic radial sequences with motion compensation [4]. This unfortunately prolongs the acquisition time, with an imperfect compensation [2] and acquisitions usually performed at end of normal expiration [4]. Moreover, although sequences with parallel acquisition techniques have been developed, limitation in spatial resolution still remains, suggesting the need for developing novel approaches to suppress respiratory artefacts.
Of particular interest in this field is the method of high frequency percussive ventilation (HFPV®), developed by Forrest M. Bird, which is based on the administration of small volumes of air so called “percussions” with adjustable pressures and frequencies. These high-frequency percussions can be superimposed on (and replace) spontaneous ventilation. The system is pneumatic, works without electronics and is not influenced by electromagnetic forces. The security of the system is based on the fact that percussions are administered through an open circuit called Phasitron, composed of a sliding venturi which delivers percussive tidal volumes according to the current subject’s thoraco-pulmonary compliance [5–7]. At each high frequency insufflation the system delivers low volumes at high pressure when compliance is low and high volumes at low pressure when compliance is high. In case of increased pressure or reduced thoraco-pulmonary compliance, the piston moves back leaving the expiratory port open, enabling protection against overpressure. HFPV® has been shown to improve oxygenation compared to conventional ventilation and has been widely used in Medicine for various acute and chronic situations in all age groups (from newborn to the elderly) [8,9]. A modified HFPV® technique was recently tested in patients for whom chest immobilization was needed as part of their thoracic radiotherapy. This pilot clinical study was approved by our ethical committee (Protocol 225/14 CHUV-DO-PART) and showed that Percussion Assisted Radiotherapy was feasible, well tolerated, reproducible and potentially associated with dosimetric gains compared to other techniques for controlling lung motion. In a series of 50 consecutive radiotherapy fractions, prolonged apnoea-like lung stabilization could be obtained in 100% of the cases with a mean duration of chest stabilization of 8.3 minutes [10].
Based on this positive experience, it was decided to adapt this non invasive technique for chest MRI. The goal of our study was to evaluate the feasibility and potential benefit of combining this percussive ventilation with chest MRI, so called VP-MR, in order to assess the effect of prolonged thoracic stabilization at full inspiration on the quality of MR images.
Methods and materials
The realization of the pilot study for testing HFPV® was authorized by the Ethics Committee on Human Research of the State of Vaud (Protocol 225/14 CHUV-DO-PART). The technique was applied on one volunteer and one patient who gave their written informed consent to this study, in accordance with the ethics committee requirements.
Device
The HFPV® was obtained via a Transrespirator®, (Percussionaire® Corporation; Idaho, USA). This Transrespirator® was adapted to be MRI compatible, and one metallic component of the Phasitron was replaced by a MRI compatible part. Six meter long tubes were adapted for use with the HFPV® allowing the installation of the Transpirator® outside the MRI room (Fig 1). The settings were chosen as follows: airflow was set to obtain airway pressure between 20 and 40 cm H2O; frequency between 400 and 500 cycles per minute; FiO2 was set at 100%; work pressure was set between 20 and 40 PSI. The inspiratory and expiratory times were adjusted to eliminate the variations of the Pressure Controlled Ventilation (PCV) and continuous positive airway pressure (CPAP) demand was set at 0 cmH2O.
The Percussionaire® ventilator remains outside the room and is connected to the non invasive interface (mouthpiece (1)) via the Phasitron (2) using a 6 meter long circuit (3).
Conventional use of HFPV® superimposes high frequency ventilation on a PCV type mode. In this study we eliminated the PCV component and ventilated the subject only by tiny variations of the pressure using high frequency setting in order to obtain thoracic stabilization approaching apnoea. Owing to the positive pressure used, the lungs were explored at inspiratory volumes.
MRI feasibility
Volunteer 1, 38 y-old, was in good health with no previous known disorder. Patient 1, 55 y-old, had a history of a thymic lesion known to be stable at previous examinations and requiring CT or MR control. In order to avoid the use of ionizing radiation, a less invasive chest MRI was proposed instead of the CT-scan for the follow-up of the lesion. No contrast administration was used in either case. Both examinations were performed using Siemens 1.5T equipment (Aera, Siemens, Erlangen, Germany), at full inspiration and using cardiac gating in order to eliminate cardiac artefacts. The arms were positioned along the body. For each MRI exam, the same sequences were performed with and without the respiratory device (Table 1). Sequences with compensatory corrections of motion, e.g. BLADE were not used. The sequences in thoracic stabilization were split into three sets, while the overall time of MR exam was intended to be 45–60 minutes in both cases. Careful combinations of sequences were chosen in order to take optimal advantage of each stabilization period.
Participants underwent 3 preliminary training sessions with the technique before undergoing MRI with HFPV®. During MR acquisition, the stabilization period ended when the volunteer/patient decided to take a breath. Cardiac frequency and oxygen saturation were monitored throughout the test. The breathing cycle amplitude reduction was monitored by using navigator technique (Fig 2). Procedure tolerance was evaluated by the volunteers using an analog scale with the following values: 1 = inacceptable, 2 = bad, 3 = good, 4 = very good, 5 = excellent. In addition, the delineation of the lung-liver interface was quantified in the ultra-short echo time images by measuring the normalized maximum of the first derivative (MD) [11]. This measure is defined such that values greater than 1 indicate an improvement in image sharpness under prolonged thoracic stabilization relative to free breathing acquisition.
The x axis represents the time and the y axis the height in millimeters.
Results
VP-MR duration and tolerance for the 2 participants
The time between the start of procedure to the stabilization phase was 8 min and 6.5 min, respectively for volunteer 1 and patient 1. Three consecutive stabilization phases of 12, 11 and 8.5 min were registered for Volunteer 1. Four consecutive stabilization phases of 5, 6.2, 5.8 and 5.8 min were registered for Patient 1. Tolerance of the VP-MR examination was counted as 3 (good) for both participants.
VP-MR images
Image quality of all sequences was markedly improved under VP-MR. This concerned the mediastinum (Fig 3) as well as the coronary arteries (Fig 4), the pulmonary vessels (Fig 5), the bronchi, the lung parenchyma (Fig 6) and the chest wall. The quality of inspiration was linked to almost complete suppression of breathing artefacts while combined with an elongated aspect of the whole chest, in particular the heart and vessels (Figs 5 and 6). The T2 Fast spin echo sequence, which normally cannot be used in routine practice given its respiratory artefacts sensitivity, allowed to perfectly delineate the thymic lesion in Patient 1 (Fig 3). Some artefacts related to the VP technique were observed near the diaphragm. While the abnormal known mediastinal mass was perfectly assessed for Patient 1, incidental nodules were discovered in Volunteer 1 and confirmed with CT (Fig 6). Lung-liver interface delineation was improved in both subjects with MD values of 2.4±0.4 measured in Volunteer 1 and 1.6±0.4 in Patient 1.
(a). Fast spin echo T2 sequence at the same level with VP (b) and without (c). The thymic nodule is much better delineated under VP in b) than in c).
15 mm thick MIP reformat perfectly showing the right coronary artery without blurring artefacts—due to thoracic stabilization associated with decreased diaphragmatic motion.
(a). Conversely, this couldn’t be obtained without VP (b). Note the quality of the endoluminal signal although no contrast was administrated.
(a) and without VP (b). Incidental discovery of apical lung nodules that are more easily assessed in a) with sharper borders. Note that the lung parenchyma appears noisier in b). Correlation with thin CT slice in lung windowing (c) demonstrating the higher spatial resolution of CT with some additional details not detected by either MR sequence. Ultra short time echo sequence in sagittal 35 mm thick MIP reformat focused on the right lung. Vessels were much better assessed with increased sharpness of interfaces with VP (d) than without (e).
Discussion
The great sensitivity of chest MRI to respiratory motion has always been a technical challenge owing to long duration sequences required by low proton density, and high magnetic susceptibility of the lungs.
Respiratory triggering or gating techniques, such as navigator gating, can be used to suppress breathing artefacts, but at the price of increased total scan duration with acquisitions performed at end expiration, limiting the ability to analyze the lung. Breath holding is more efficient but intrinsically limits the duration of the acquisition. Multiple breath holds may be used for sequences requiring long acquisition duration. This however leads to increased total examination time with potential misregistration artefacts related to imperfect reproducibility of the breath hold volume thus impairing diagnostic quality of the images. While parallel imaging techniques can be used to shorten acquisition time, they also result in a decrease in signal to noise ratio (S/N). With Lung MRI using an MR-Compatible Active Breathing Control (MR-ABC), total imaging time could be reduced compared to navigator gating or respiratory belt use. ABC has been used to ensure reproducibility of breath hold volumes [12] but still requires segmenting of the examination in multiple steps. Our method presents the advantage of a long stability period during which multiple datasets covering the entire chest may be acquired at full inspiration.
Each thoracic stabilization had a respective average duration of about 10 min for Volunteer 1 and 5 to 6 min for Patient 1. To our knowledge, such a duration of thoracic stabilization has never been used in chest MR imaging. This could potentially shorten the total acquisition time to achieve whole-chest coverage while reducing the efforts of the patients and avoiding misregistration artefacts [13]. In addition, almost freezing the lung volume in full inspiration allowed the avoidance of blurring and ghosting artefacts by overall reducing the motion within voxels. In this way, an improvement of the S/N ratio with the use of signal averaging could be further expected by extending the time of a specific sequence. The potential to strictly compare various contrasts according to the sequences performed during the same stabilization period could also allow accurately merging/ the various signals with integration of diffusion parameters that could potentially improve diagnostic proposals and assessment of evolution under treatment.
As the acquisitions were performed at full inspiration, chest components were assessed in conditions similar to those obtained in routine chest CT. The motion suppression with improvement in spatial resolution resulted in a better delineation of all chest components.
Although some sequences such as conventional spin echo T2 have demonstrated their usefulness in the characterization of masses as cystic or solid, their use is unconceivable with classic chest MRI due to the couple of minutes acquisition time required [2]. Owing to the duration of the stabilization period approaching apnea with this new technique, the use of a T2 FSE allowed us to perfectly delineate the thymic lesion while using VP-MR technique compared with the sequence without VP technique in patient 1, which was similar to the CT appearance (Fig 3).
Another potential interest of VP-MR in full inspiration was the high signal within the blood pool obtained without contrast using the 3D radial Steady-State Free Precession (SSFP) acquisition that allowed displaying straight pulmonary vessels until distality, as well as the coronary arteries. The interest of 3D-SSFP sequence with respiratory self-navigation for whole-heart bright-blood coronary MRI with automatic segmentation of the blood pool with a slice thickness of 1.15 mm and isotropic voxels has already been demonstrated for the exploration of coronary arteries [14]. However, when applied to pulmonary vessels, the sinuosity of vessels reflecting the acquisition at a tidal volume currently precludes a complete assessment of disorders such as pulmonary emboli or pulmonary hypertension.
Regarding lung analysis, although new sequences have recently been developed [13,15], spatial resolution remains still a challenge, with a rate of detection of nodules between 35 and 96% [12,15]. The potential of acquisition at full inspiration would help to avoid their potential non recognition on expiration. Nodules detected in Volunteer 1 were confirmed by a low-dose CT (Fig 5). In the study of Heye et al. [13], despite the promising results of the new parallel imaging ultrafast SSFP sequence allowing for imaging in one breath-hold of 18–20 seconds with 1.9 mm isotropic spatial resolution, there were numerous false-positive nodules related to slightly blurred pulmonary vessels, possibly caused by motion affecting the MR signal. According to the authors, MR with long apnoea at full inspiration could overcome these limits [13].
Diffusion sequences didn’t show any abnormal signal within the detected nodules in Volunteer 1, precluding further evaluation. Such sequences, known to be penalized by motion and/or respiratory artefacts in uncooperative or dyspneic patients [16,17], could benefit from VP-MR.
An obvious limitation of our study is the low number of participants. However, this is aligned with the goal of this pilot experience which was to perform a first evaluation of HFPV® with MRI. These preliminary data suggest relatively good reproducibility of the technique which is in agreement with the recent data obtained for chest radiotherapy [10]. Nevertheless, it must be emphasized that further evaluation is required to explore how VP-MR technique could perform in a large sample size with various pulmonary disorders. In fact, the effectiveness and tolerance of the HFPV® has to be confirmed in subjects with lung disease, having a potential for overdistension in obstructive and underventilation in restrictive diseases. In addition, the acquisition still requires significant training of the subject and this technique might not be tolerated by some patients, particularly if applied in the MRI environment as described in this pilot study. However, as published by Peguret et al. [10], which is the first report of this technique outside the MRI, this high frequency percussion ventilation system could be successfully applied in nearly all the tested cases. Furthermore, as no contrast agent was used, the potential of the technique after gadolinium injection has not been evaluated. Similarly, perfusion and ventilation were not studied.
Conclusion
In this pilot experience, Pulsatile Flow Ventilation™ was successfully applied to induce prolonged stabilization of lung volumes with marked reduction of respiratory motion during MRI acquisitions at full inspiration. This resulted in the suppression of most respiratory-related artefacts while allowing the exploration of the whole chest at full inspiration phase with increased spatial resolution. These promising results open the field for new applications of thoracic MRI, in particular the follow-up of interstitial lung disease, bronchial or infectious disorders as well as the assessment of mediastinal masses, pulmonary artery or coronary artery disorders. These results call for developing studies to establish the real clinical value of VP-MR while improving the technique.
Supporting information
S1 Images. DICOM data from Volunteer 1 (1st half).
https://doi.org/10.1371/journal.pone.0178807.s001
(ZIP)
S2 Images. DICOM data from Volunteer 1 (2nd half).
https://doi.org/10.1371/journal.pone.0178807.s002
(ZIP)
S3 Images. DICOM data from Patient 1 (1st half).
https://doi.org/10.1371/journal.pone.0178807.s003
(ZIP)
S4 Images. DICOM data from Patient 1 (2nd half).
https://doi.org/10.1371/journal.pone.0178807.s004
(ZIP)
S1 Table. Used for the quantification of lung/liver interfaces.
https://doi.org/10.1371/journal.pone.0178807.s005
(PDF)
Author Contributions
- Conceptualization: CBA NP J. Schwitter MO SQ RM JB AL.
- Data curation: CBA NP MO BB AL J. Simons OL KG.
- Formal analysis: CBA NP MO JD.
- Funding acquisition: RM JB MS.
- Investigation: CBA NP MO BB AL J. Simons OL KG.
- Methodology: CBA NP MO BB AL J. Simons OL KG RM JB.
- Project administration: CBA NP RM JB MO.
- Resources: RM JB MS.
- Software: JD.
- Supervision: MS RM JB MO.
- Validation: CBA JD.
- Visualization: CBA JD G. Berchier CR G. Bonanno SC.
- Writing – original draft: CBA NP RM JB AL JD.
- Writing – review & editing: CBA NP RM JB AL MO JD SQ.
References
- 1. Kauczor HU, Ley S (2010) Thoracic magnetic resonance imaging 1985 to 2010. J Thorac Imaging 25: 34–38. pmid:20160600
- 2. Puderbach M, Hintze C, Ley S, Eichinger M, Kauczor HU, Biederer J (2007) MR imaging of the chest: a practical approach at 1.5T. Eur J Radiol 64: 345–355. pmid:17900843
- 3. Wild JM, Marshall H, Bock M, Schad LR, Jakob PM, Puderbach M et al. (2012) MRI of the lung (1/3): methods. Insights Imaging 3: 345–353. pmid:22695952
- 4. Biederer J, Reuter M, Both M, Muhle C, Grimm J, Graessner J et al. (2002) Analysis of artefacts and detail resolution of lung MRI with breath-hold T1-weighted gradient-echo and T2-weighted fast spin-echo sequences with respiratory triggering. Eur Radiol 12: 378–384. pmid:11870438
- 5. Short K, Bougatef A, Khusid F, Kenney BD, Miller K (2010) High-frequency percussive ventilation. Respir Care 55: 1762–1764; author reply 1764–1767. pmid:21268771
- 6. Lucangelo U, Fontanesi L, Antonaglia V, Pellis T, Berlot G, Liguori G et al. (2003) High frequency percussive ventilation (HFPV). Principles and technique. Minerva Anestesiol 69: 841–848, 848–851. pmid:14735024
- 7. Allan PF (2010) High-frequency percussive ventilation: pneumotachograph validation and tidal volume analysis. Respir Care 55: 734–740. pmid:20507657
- 8. Dumas De La Roque E, Bertrand C, Tandonnet O, Rebola M, Roquand E, Renesme L et al. (2011) Nasal high frequency percussive ventilation versus nasal continuous positive airway pressure in transient tachypnea of the newborn: a pilot randomized controlled trial (NCT00556738). Pediatr Pulmonol 46: 218–223. pmid:20963833
- 9. Vargas F, Bui HN, Boyer A, Salmi LR, Gbikpi-Benissan G, Guenard H et al. (2005) Intrapulmonary percussive ventilation in acute exacerbations of COPD patients with mild respiratory acidosis: a randomized controlled trial [ISRCTN17802078]. Crit Care 9: R382–389. pmid:16137351
- 10. Peguret N, Ozsahin M, Zeverino M, Belmondo B, Durham AD, Lovis A et al (2015) Apnea-like suppression of respiratory motion: First evaluation in radiotherapy. Radiother Oncol RADION 6640
- 11. Tibiletti M, Paul J, Bianchi A, Wundrak S, Rottbauer W, Stiller D et al. (2016) Multistage three-dimensional UTE lung imaging by image-based self-gating. Magn Reson Med 75: 1324–1332. pmid:25940111
- 12. Kaza E, Symonds-Tayler R, Collins DJ, McDonald F, McNair HA, Scurr E et al. (2015) First MRI application of an active breathing coordinator. Phys Med Biol 60: 1681–1696. pmid:25633183
- 13. Heye T, Sommer G, Miedinger D, Bremerich J, Bieri O (2014) Ultrafast 3D balanced steady-state free precession MRI of the lung: Assessment of anatomic details in comparison to low-dose CT. J Magn Reson Imaging.
- 14. Piccini D, Littmann A, Nielles-Vallespin S, Zenge MO (2012) Respiratory self-navigation for whole-heart bright-blood coronary MRI: methods for robust isolation and automatic segmentation of the blood pool. Magn Reson Med 68: 571–579. pmid:22213169
- 15. Lederlin M, Cremillieux Y (2014) Three-dimensional assessment of lung tissue density using a clinical ultrashort echo time at 3 tesla: a feasibility study in healthy subjects. J Magn Reson Imaging 40: 839–847. pmid:24123396
- 16. Abdel Razek AA, Khairy M, Nada N (2014) Diffusion-weighted MR imaging in thymic epithelial tumors: correlation with World Health Organization classification and clinical staging. Radiology 273: 268–275. pmid:24877982
- 17. Seki S, Koyama H, Ohno Y, Nishio M, Takenaka D, Maniwa Y et al. (2014) Diffusion-weighted MR imaging vs. multi-detector row CT: Direct comparison of capability for assessment of management needs for anterior mediastinal solitary tumors. Eur J Radiol 83: 835–842. pmid:24636535