Digital Microfluidic Dynamic Culture of Mammalian Embryos on an Electrowetting on Dielectric (EWOD) Chip

  • Hong-Yuan Huang,

    Affiliations Department of Obstetrics and Gynecology, Linkou Medical Center, Chang Gung Memorial Hospital, Taoyuan, Taiwan, Department of Obstetrics and Gynecology, Chang Gung University and College of Medicine, Taoyuan, Taiwan

  • Hsien-Hua Shen,

    Affiliation Institute of Nanoengineering and Microsystem, National Tsing Hua University, Hsinchu, Taiwan

  • Chang-Hung Tien,

    Affiliation Department of Mechanical Engineering, National Taiwan University, Taipei, Taiwan

  • Chin-Jung Li,

    Affiliation Department of Obstetrics and Gynecology, Linkou Medical Center, Chang Gung Memorial Hospital, Taoyuan, Taiwan

  • Shih-Kang Fan,

    Affiliation Department of Mechanical Engineering, National Taiwan University, Taipei, Taiwan

  • Cheng-Hsien Liu,

    Affiliation Department of Power Mechanical Engineering, National Tsing Hua University, Hsinchu, Taiwan

  • Wen-Syang Hsu,

    Affiliation Department of Mechanical Engineering, National Chiao Tung University, Hsinchu, Taiwan

  • Da-Jeng Yao

    djyao@mx.nthu.edu.tw

    Affiliations Institute of Nanoengineering and Microsystem, National Tsing Hua University, Hsinchu, Taiwan, Department of Power Mechanical Engineering, National Tsing Hua University, Hsinchu, Taiwan

Abstract

Current human fertilization in vitro (IVF) bypasses the female oviduct and manually inseminates, fertilizes and cultivates embryos in a static microdrop containing appropriate chemical compounds. A microfluidic microchannel system for IVF is considered to provide an improved in-vivo-mimicking environment to enhance the development in a culture system for an embryo before implantation. We demonstrate a novel digitalized microfluidic device powered with electrowetting on a dielectric (EWOD) to culture an embryo in vitro in a single droplet in a microfluidic environment to mimic the environment in vivo for development of the embryo and to culture the embryos with good development and live births. Our results show that the dynamic culture powered with EWOD can manipulate a single droplet containing one mouse embryo and culture to the blastocyst stage. The rate of embryo cleavage to a hatching blastocyst with a dynamic culture is significantly greater than that with a traditional static culture (p<0.05). The EWOD chip enhances the culture of mouse embryos in a dynamic environment. To test the reproductive outcome of the embryos collected from an EWOD chip as a culture system, we transferred embryos to pseudo-pregnant female mice and produced live births. These results demonstrate that an EWOD-based microfluidic device is capable of culturing mammalian embryos in a microfluidic biological manner, presaging future clinical application.

Introduction

Fertilization in vitro (IVF) was an exciting scientific achievement of the twentieth century that maintains a great impact on human lives [1]. With the increasing clinical utilization of assisted reproductive technology (ART), scientists and clinicians have acquired insight into the basic biology of gametes and embryos and translated that knowledge into improved rates of success following assisted reproduction. These approaches aimed to improve embryo development in vitro have involved mostly the chemical composition of the culture media. Both conventional and sequential culture-medium systems have been refined; the development of high-quality blastocysts in vitro is common in clinical practice [2, 3].

Human IVF bypasses the female oviduct and manually inseminates, fertilizes and cultivates an embryo in an oil-covered static microdrop (μL) containing appropriate chemicals on a Petri dish and manipulated several times with manual pipetting. The labor-intensive manipulations of repeated washing and culture drops might introduce loss and prospective stress of temperature, pH, osmolality and light, and mechanical stress, on the embryos [4]. Although the culture platforms offer some advantages, the gametes and embryos rest on inert synthetic polymers and bathe in a significant volume of medium, in extreme contrast to what occurs in vivo. Furthermore, a cultured embryo experiences not only chemical (hormonal) but also prospective physical requirements that might be important factors in the continuing pursuit of improved conditions in vitro before implantation[5, 6].

Only recently have biomimetics and microfluidics become an approach to increase the rate of success of IVF. Additional implemental and analytical approaches and techniques are emerging; an examination of various novel culture platforms to explore the impact of physical and mechanical modifications on the embryo might assist in improving further the development in vitro[710]. Because of the comparable scales, microfluidics appear to offer a prospective means to improve other common procedures or approaches used within the laboratory for IVF, and elsewhere, near that which gametes and embryos experience in vivo. Microfluidics is applied to IVF in various aspects [1113], including sperm processing, fertilization and embryo culture. Besides using the laminar flow characteristics of microfluidics to separate motile and non-motile sperm along a microchannel [1418], microfluidic IVF is considered an improved means to mimic dynamic environments in vivo for cultured embryos[12].

All reported microfluidic IVF work has been done in continuous microchannels, which differ substantially from a microdrop currently used in clinical IVF [19, 20]. In our work, a single droplet is manipulated with electrodes deposited on the platform using electrowetting on a dielectric (EWOD) and dielectrophoresis (DEP) [21], simplifying the repeated washing and medium changes currently effected with manual pipetting. The EWOD system was derived from the electrowetting phenomenon that alters the contact angle of a liquid, so to enable a discrete droplet to be manipulated through reconfigurable paths or channels defined by electrode actuation. Compared with a continuous microfluidic channel, an EWOD digital microfluidic system has the advantages of minute sample and reagent volume, precise control of a droplet, small risk of contamination, small consumption of energy, low cost, rapid analysis, portability for diagnosis at a point of care, flexibility in integrating with a detector, multiple-step processes and simultaneous reaction [2225]. In the present work, we describe a novel digital microfluidic device with EWOD to culture a mammalian embryo in a single droplet that provides a non-continuous microfluidic environment to mimic the environment in vivo for the development of an embryo and live births, presaging future clinical application.

Materials and Methods

Device construction

Devices were fabricated in the clean-room facilities of Institute of Nanoengineering and MicroSystems (iNEMS) in National Tsing Hua University (NTHU). The devices composed of coplanar-electrode type have a ground and five electrodes placed on the same plane, which bestows a flexibility to integrate further functionality with a top plate (Fig 1A). The gap between each electrode is 20 μm; the size of the control electrode is 0.75 mm × 1.5 mm. A digital microfluidic channel is made with an array of 15 × 2 control electrodes. The bottom plate has electrode arrays on indium tin oxide (ITO) glass patterned with photolithography and wet etching. Si3N4 (thickness 450 nm) and a spun-on thin film of glass (SOG) (thickness 200 nm) were deposited on ITO electrodes as a compound dielectric layer. Teflon was coated as a hydrophobic layer with contact angle 120°. The height of the microchannel space between the top and bottom plates was 260 μm with multi-tape as a spacer. The top plate was glass with a Teflon coating (thickness 20 nm) as a hydrophobic cap. In this research the microfluidic channel was pre-coated with pluronic F127 (0.08%) in a HTF medium for culturing an embryo to decrease biofouling and to enhance the protein stability [26].

thumbnail
Fig 1. Scheme of EWOD chip design and fabrication and principle of EWOD.

A. Description of EWOD chip fabrication and bottom plate of EWOD chips (white scale bar, 6 mm). B. EWOD is based on the variation of the wettability of liquids on a dielectric solid surface on altering the electric potential. The contact angle of the liquid is decreased from its initial contact angle when a voltage is applied. On releasing the electric potential, the contact angle of the liquid reverts to the initial contact angle. C. Scheme of droplet transport.

https://doi.org/10.1371/journal.pone.0124196.g001

Device design with electrowetting-on-dielectric

EWOD is based on the variation of the wettability of liquids on a dielectric solid surface on altering the electric potential. The contact angle of the liquid is decreased from its initial contact angle when a voltage is applied. On releasing the electric potential, the contact angle of the liquid reverts to the initial contact angle (Fig 1B).

To create a droplet, we used four controllable electrodes and a reservoir electrode [27, 28]. We first activated the four controllable electrodes; the liquid was drawn by the reservoir electrode and moved towards electrode 1. When the controllable electrodes were full of the liquid, we deactivated electrodes 2 and 3, and activated electrode 1 and the reservoir electrode; the liquid produced a necking phenomenon, so creating a droplet on electrode 1 (Fig 1C). To manipulate the EWOD device, we applied an AC potential signal through a signal generator (Agilent 33220A, Agilent Technologies, Inc. Taiwan), which was amplified and output into the control circuit with a power amplifier (A.A. Lab Systems-AA303, Ramat-Gan, Israel). We used a computer to plan the route of the droplet transport and to output the signal to switch relays that turn on or off through a PCI digital I/O control card (NI USB-6509, National Instrument, Inc. Taiwan). The DMF chip was connected with the clamp (Yokowo, Micro Tech, Inc. Taiwan) via a cable to enable it to deliver an electric signal from the system to the clamp inside the cell-culture incubator. All this system must cooperate with an ordinary culture system with an incubator. Therefore we have created a hole on the wall of the incubator, and the cable was able to pass through the hole for connecting the electrical signal of the chip and system.

Materials

The mouse experiments were approved by the Chang Gung Memorial Hospital (CGMH) Animal Care and Use Committee and conducted in accordance with the principles and procedures outlined in the CGMH Guidelines for the Care and Use of Laboratory Animals. Imprinting Control Region (ICR) mice were used in this research; imprinting is the mammalian phenomenon whereby the expression of a gene occurs only from one of two alleles, the other having been silenced. Female ICR mice were injected intra-peritoneally with PMSG (5 IU; Sigma, USA). 48 h after PMSG injection, the female mice were further injected intra-peritoneally with HCG (5 IU; Sigma, USA). Immediately after HCG injection, each female mouse was caged with a male mouse for mating (defined as day E0). 36 h post HCG injection, the female mice were sacrificed and the oviducts were dissected and placed in a Petri dish containing human tubal fluid medium (HTF) (Irvine Scientific, Santa Ana, CA, USA). Two-cell embryos for an embryonic development culture were released on tearing the oviducts (defined as day E1.5).

Embryo loading and culture

In a microfluidic culture experiment, a single mouse embryo at the two-cell stage was cultured in a droplet (1 μL) of medium and covered with mineral culture oil (4 μL) (OVOIL, Vitrolife, Göteborg, Sweden). The medium droplet was driven with EWOD electrodes by automatic programming for three days. The automatic programming served to control the period and amplitude of an applied voltage. The medium droplets moved up to 100 times, each occasion lasting a few seconds. In every dynamic experiment, two medium droplets were placed in the EWOD chip. The development of the embryo in an EWOD system was individually observed on day 1 (day E2.5), day 2 (day E3.5) and day 3 (day E4.5) with a microscope. The control group was a culture in devices in static fluidics without a potential applied; embryo development was also recorded as a control.

Embryo transfer

To test the reproductive outcome of the embryo collected from an EWOD chip embryo culture system, we transferred embryos to pseudo-pregnant female mice; these were ICR female recipient mice after animal balance at least six weeks of age that had been mated to sterile males the evening before the transfer. The females were tested for evidence of vaginal plugs after mating. Recovered blastocysts from each treatment group were transferred into the uterus of a separate pseudo-pregnant recipient using the method described previously [29, 30]. Briefly, the recipient mouse was anaesthetized (2.5% avertin per gram of body mass). The back of the recipient mouse was wiped with ethanol (70%) and then a small transverse incision (less than 1 cm) was made in the skin with line dissection scissors, about 1 cm to the left of the spinal cord, at the level of the last rib. The skin was slid to the left or right until the incision was over the ovary or fat pad, both of which were visible through the body wall. The body wall was picked up and a small incision made just over the ovary with fine dissection scissors to expose the oviduct, ovary and the upper part of the uterus. Embryos in a transfer medium were transferred into each uterine horn with a fine glass pipette. Each uterine horn was then returned to the body cavity; the skin incision was closed appropriately. At the end of the procedure, the mouse was returned to its cage and left undisturbed in a warm, quiet place. The mouse warm was kept warm with a heating pad placed over the top of its cage. The mouse recovered from the anaesthesia in approximately 20–30 minutes. All recipients were allowed to deliver and to raise pups, which were either raised to the weaning age to test their post-natal development or mated with their siblings from the same litter for future fertility testing.

Data analysis

The difference (as percentage) of blastocyst development in each group was analyzed with statistical software (SPSS Inc., Chicago, IL). Data (as percentage) were analyzed with a χ2 test. A value p < 0.05 was considered to be statistically significant.

Results

Manipulation of a droplet in a medium

To mimic the environment flow in vivo and to take into account a shear stress effect, a decreased velocity of droplet motion was essential to assist an embryo to develop. When an EWOD electrode was turned on, a droplet moved from the original electrode to the next electrode. Our definition of the droplet velocity is derived from dividing the distance between two electrodes by the duration of a droplet to reach the nearby electrode. In a dynamic cell culture, the droplet velocity was controlled in range 0.3 ~ 1 mm s-1 to mimic the movement of an cleavage embryo in an oviduct (Fig 2A). To maintain the droplet moving fluently, we applied an electric potential (60~68.5 V RMS) at 500 Hz. The Eq (1) describing the velocity dependence of the contact angle [31]:

thumbnail
Fig 2. Operating the EWOD system for in vitro culture of embryo.

A. Manipulating a droplet moving with an EWOD system. (a) The EWOD chip was placed in an incubator (b) Before motion, the droplet time = 0 s (c) The upper droplet moved 1.5 mm, right, time = 5 s (d) The lower droplet moved 1.5 mm, left, time = 10 s. B. Applied voltage versus velocity of droplets moving in air and in an oil-bath environment. C. Static embryo culture on a Petri dish and in an EWOD chip. The development of embryos in a EWOD chip was recorded as compatible relative to a traditional culture dish as control. D. Observation of an embryo in a moving droplet. Sequence images of embryo movement in a dynamic EWOD chip culture (a) before applied voltage. (b) The shape of droplet was altered with an applied voltage at the first electrode (c) The embryo was moving to the first electrode (d) The embryo stopped near the medium-oil interface (e) The embryo was moving rapidly when a applied voltage to the second electrode (f) The embryo was stopped near the medium-oil interface of the second electrode. *E day, the date after mating. Black arrow bar, 100 μm.

https://doi.org/10.1371/journal.pone.0124196.g002

(1) Where cosθE is the dynamic contact angle in the presence of the electric field. The γLV is the liquid-vapor surface energies respectively, d is the thickness of the solid insulator, εr and ε0 are the dielectric constant of the dielectric and the air, kB is Boltzmann’s constant, T is the temperature, and there are n adsorption sites per unit area. The overall speed v is determined by the frequency k0 and length λ of these displacements. Reasonably good curve-fits to the experimental data were obtained with this equation for values of λ and k0 which were independent of the electric field. The contact angle hysteresis can be modeled as a negative energy constant in the sinh argument expressed as an effective voltage threshold VT. The Eq (2) can be obtained as: (2)

In this research, the relation between an applied voltage and the droplet velocity of the digital microfluidic (DMF) system has been measured and shown in Fig 2B. In each experiments, we try to make the operational voltage as low as possible to reduce the effect of the electric field to the development of an embryo. However, the lowest droplet operation voltage depends on different DMF chips. Sometimes when the voltage was lower than 60V, the droplets manipulation would be unstable.

To demonstrate that the embryo development cultured on an EWOD chip was compatible with culture on a traditional culture dish, we cultured a single mouse embryo at the two-cell stage in a medium droplet (1 μL), covered with culture oil (4 μL) on an EWOD chip in static fluidics without an applied potential and in a traditional culture dish. The development of the embryo was observed individually on day 1 (E2.5), day 2 (E3.5) and day 3 (E4.5) with a microscope. The development of embryos in an EWOD chip was recorded as compatible with that of the control in a traditional culture dish (Fig 2C).

Embryo development in a dynamic culture

The embryo was moved with droplet manipulation powered by EWOD. Fig 2D shows the result of the embryo in a dynamic environment using an EWOD chip. The velocity and shear stress of the embryo were affected by the position in the droplet. The embryo was initially in the middle, moving slowly with the fluid when the droplet was driven. The embryo ended at the medium-oil interface after the droplet crossed to the next electrode. The position near the interface caused the embryo to move rapidly: the embryo moved approximately 1.5 mm in 2 s when the droplet was driven. The dynamic fluid environment was demonstrated with the EWOD chip of which the result showed the embryo to move on manipulating a droplet. A video clip appears in S1 Video.

The embryo development of dynamic culture is shown in Fig 3. The speed of cleavage of an embryo in a dynamic culture was greater than that in a static culture. The development of an embryo in droplet with different velocity in a EWOD system was hatching earlier than those embryos in the static control-group culture.

thumbnail
Fig 3. Dynamic embryo culture in EWOD chips.

Embryo culture development on the EWOD system, The upper panel is static culture on the EWOD chip: from 4-cell stage to expanded blastocyst stage; The middle and lower panels are dynamic culture with different velocity of droplets: from 4 or 8-cell stage, early blastocyst stage to hatching blastocyst stage. The embryo development in dynamic culture with the droplet velocity (15 sec /0.5 hr) was hatching earlier in comparison to the embryo in static culture. *E day, the date after mating. Black arrow bar, 100 μm.

https://doi.org/10.1371/journal.pone.0124196.g003

Embryo development and blastocyst quality

The statistical result of the embryo development in an EWOD dynamic culture (n = 18) on comparison with a static culture (n = 38) is shown in Fig 4. Most embryos were cleaved to a blastocyst stage (88.9±4.2% vs. 92.1±10.8%, respectively) as shown in Fig 4A. The blastocysts is further classified day 3 (E4.5) as full blastocyst (22.2±4.5% vs. 42.3±8.6%), expanded blastocyst (16.7±4.6% vs. 29.1±5.7%) and hatching blastocyst (50±9.8% vs. 21.2±5.5%, p = 0.028) as shown in Fig 4B.

thumbnail
Fig 4. The embryo development and blastocyst quality of dynamic culture on EWOD chip.

A. The statistics result of blastocyst development in EWOD dynamic culture (n = 18) in comparison to static culture (n = 38). Most embryos were successfully cleaved to a blastocyst stage. B. The blastocysts collected on day 3 (E4.5) is further classified according the morphology grading. A significantly higher rate of hatching blastocyst was demonstrated in dynamic culture in comparison to static culture (p = 0.028). *, Chi-square test, p<0.05. Values are mean ±standard error of measurements.

https://doi.org/10.1371/journal.pone.0124196.g004

Blastocyst recovery from EWOD chip and viability after embryo transfer

To test the reproductive outcome of the embryos successfully collected from an EWOD chip, we transferred the EWOD embryos to pseudo-pregnant female mice and produced live births. All blastocysts that developed from the EWOD and control groups after culture in vitro simultaneously were transferred into surrogate recipients (up to six embryos transferred individually for three experiments). The recipients became pregnant and were allowed to deliver and raise pups that were either raised to the weaning age to verify their post-natal development or mated with their siblings from the same litter for fertility testing (Fig 5). Of the blastocyst from EWOD embryos, 19 blastocysts were transferred to 3 pseudopregnant mice, of which 3 had pups. A total of 10 pups were born, representing 50.6% of the transferred blastocysts from EWOD group. All of the 10 pups born from EWOD group were grown up to 3 weeks with normal appearance.

thumbnail
Fig 5. Embryo transfer from EWOD embryos to pseudo-pregnant female mice.

Panel A, Preparation of recipient mice for embryo transfer. Panel B, Pups was born after transfer of embryo cultured from EWOD. The recipients were allowed to deliver and raise pups.

https://doi.org/10.1371/journal.pone.0124196.g005

Discussion

According to our results, the culture in vitro of a mouse embryo using a EWOD system to provide a dynamic fluid environment is capable of culturing mammalian embryos in a microfluidic biological manner. A comparison of the results of the dynamic culture with the static culture also proves that the dynamic culture in the EWOD system enhanced the cleavage and development of a mouse embryo to the preimplantation stage.

For the preimplantation embryo cleavage in vivo in the oviduct and ultimately developed in the uterus, contractions of the inner muscle of the oviduct and movement of epithelial cell cilia cause an embryo to move continually. To mimic the dynamic fluid environment in vivo, the EWOD system required a decreased velocity of the droplet. The embryo velocity in a droplet is random; the EWOD system provided a dynamic environment, but a fixed velocity of embryo movement was unavailable. The gradients of potassium, calcium and oxygen in a medium droplet were disrupted, however, through the manipulation of droplets that established a suitable environment for embryo development.

Microfluidics is a promising new technique. Increased attention is thus being devoted to microfluidics in the field of reproductive medicine. Microfluidics, based on the physical principles of fluid behavior in a microenvironment, serve widely for applications in chemistry and molecular biology [3235]. Many advances in nanotechnology and MEMS are applied to study biological systems. Microfluidic devices are being developed for biological assays with the potential to perform analysis with efficient throughput, to utilize samples or reagents in small amounts and for analysis of single cells [36, 37].

Microfluidic systems have been variously designed for IVF [3840]. Sano et al. developed a microfluidic sperm sorter (MFSS) for porcine IVF[41], designed to isolate motile human sperm with laminar flows. Brief co-culture of porcine oocytes with sperm gradually accumulated in the MFSS chamber improved the efficiency of producing monospermic fertilized embryos and blastocysts; efficiencies were significantly affected by the oocyte location within the chamber. In a novel microwell-structured microfluidic device integrating trapping of a single oocyte, IVF and embryo culture reported by Han et al.[12], a microwell array served to capture and to hold individual oocytes during the flow of oocyte and sperm loading, medium substitution and debris cleaning. Fertilization was achieved in the microfluidic devices with rates of fertilization similar to those of standard oil-covered drops in Petri dishes.

Embryos were cultured to blastocyst stages in devices with developmental status individually monitored and tracked, but the reported results of microfluidic IVF are still preliminary and limited. Some results are even contradictory. For example, Hickman et al. reported that the continuous media perfusion to embryos within a microchannel resulted in poor embryo development for flow rates across a range [42]. In general, researchers speculated that the refreshing of nutrition and removal of waste increased the rate of development. A small rate of development was attributed to the shear stress and continual removal of growth-promoting autocrine factors [43]. To achieve a more reproducible and biomimetic condition, further investigation must include the influences of the microchannel dimensions and the flow rates. A computerized microfluidic device for embryo culture and assay in real time is required that can perform automated periodic analyses of embryo metabolism [44].

The microfluidic IVF system also possesses several major advantages over a conventional culture system. With microfluidics, fewer sperm are needed because of the smaller medium volume. The optimal concentration of sperm was in range 2~8x104 sperm/mL, significantly less than the standard IVF (1×106 sperm/mL) in tested microfluidic channels [15].

Raty et al. found that embryos cultured in microchannels filled with static media exhibited a greater rate of cleavage and produced more blastocysts than control microdrops [45]. Some authors indicated that the polyspermic penetration was decreased in microchannels [46]. The development of mouse embryos is advanced within microfluidic devices with media flow relative to static devices, resulting in increased rates of blastocyst formation and hatching [47]. With microfabrication, versatile microfluidic devices are practicable for integrated and automated IVF functions [12], but the reported results of microfluidic IVF are still preliminary and limited; some are even contradictory.

EWOD (electrowetting on a dielectric) is a microfluidic strategy that exploits surface tension as a means to manipulate liquid droplets at the current stage [48, 49]. An electric voltage is applied; the electric charge alters the Gibbs energy on the dielectric surface, causing an altered wettability of the surface and contact angle of the droplet [50]. The development of an EWOD digitalized microfluidic system capable of transporting oocytes or embryos to a desired location, producing mammalian embryos in vitro and chemical or mechanical manipulations, is highly desirable.

We applied our expertise of digital microfluidics to demonstrate dynamic IVF, preserving the merit of the microdrop method but enhancing its capability with controllable and dynamic droplets. Our study of dynamic IVF on a digital microfluidic platform based on the preceding microdrop IVF studies, especially the developed methods, is the first. Several features have been demonstrated, especially the culture of the embryo to the ultimate stage (blastocyst), developed earlier than for the traditional culture group. The droplets containing the embryo are manipulated with electrodes deposited on the platform using EWOD and dielectrophoresis [21], simplifying the repeated washing and medium changes currently involved with manual pipetting. In contrast with the static microdrop method, the movable droplets generate internal flows that decrease the point depletion or accumulation of substances in the medium with possible detrimental effects on the embryo development [19]. The gentle agitation of a medium and embryos would enhance outcomes by improved mimicking of the fluid-mechanical stimulation that embryos experience in vivo from ciliary beating and oviductal contraction [44]. Furthermore, the embryos can be moved with the droplet at an appropriate speed as in the oviduct [20]. The droplets serve not only as an isolated environment for embryos but also to perform sophisticated fluidic functions for diagnostic protocols (embryo selection or metabolic profile).

In research on EWOD droplet manipulation, workers have tried to make the applied voltage minimal. Water droplets can be manipulated with a small voltage [51]; some researchers used varied dielectric layer materials to decrease the voltage for droplet manipulation [52]. Lin et al. used 17 Vrms for manipulation of a water droplet [53], but a greater voltage is necessary for manipulation of a cell culture medium because it contains abundant electrolyte and biomolecules. Cell-based EWOD experiments typically require 100–300 Vrms to manipulate droplets [5456]. In EWOD devices, the voltage is applied to induce an electric double layer in a dielectric layer and to alter the contact angle between a hydrophobic layer and a droplet. Even though electric charge cannot pass through the dielectric layer into a droplet on the EWOD chip, we decreased the voltage to 68 Vrms for a stable culture of an EWOD dynamic embryo and tried to discover the influence of EWOD manipulation of electrical droplets on the embryo development.

Some concerns are addressed on the effect of a EWOD system on embryogenesis. Embryogenesis is a complicated process according to which a fertilized egg develops into a multicellular organism, during which fundamental biological self organization occurs. The application of electric fields at small frequency to cells has a long and contentious history. All work indicates that developing embryos produce endogenous currents involved during embryonic development. Researchers also discussed a possible mechanism of a small electric field on isolated cells [5760]. The electric field has a potential as a research tool for various biotechnological experiments with mammalian gametes and embryos. A parthenogenetic development of oocytes induced with an electric field has been performed in mammalian species [61, 62]. Electric activation can facilitate the fertilization and early embryonic development after ICSI (Intracytoplasmic Sperm Injection) in human beings [63, 64]. Popova et al. used mouse and rat embryos as models to investigate the sensitivity of oocytes and embryos to electric fields [65]; these authors suggested that electrical activation might facilitate fertilization and early embryonic development after ICSI in human beings. As electric stimulation is known to enhance protein and DNA synthesis, to stimulate neuronal cells and to differentiate the migration of neural crest cells, this work was planned to observe the developmental differences of Xenopus embryos exposed to electric fields during embryogenesis. All this work might provide some explanation of the dynamic cultural result relative to a static cultural result that proved a dynamic culture in the EWOD system to enhance the cleavage and development of a mouse embryo and to accelerate the growth to the blastocyst stage. As little is known about the sensitivity to electric fields of mammalian embryos of various species and at various stages of the early pre-implantation development, further investigation is necessary, including the influences of the microchannel dimensions and the flow rates, to achieve a more reproducible and biomimetic condition.

Conclusion

We describe a digital microfluidic device that shows a feasibility to perform a dynamic embryo culture on an EWOD platform. We manipulated the function of a medium droplet containing a bio-sample in an EWOD chip with a modified complicated dielectric layer and a voltage in a culture incubator. The devolvement and the reproductive potential of the dynamic cultured embryo in an EWOD chip is compared with a traditional culture system. The device allows a capability of culturing mammalian embryos in a microfluidic biological manner. Further development might facilitate a new protocol embryo development and clinical IVF practice.

Supporting Information

S1 Video. Shows digital microfluidic dynamic culture of mouse embryo using EWOD chip.

https://doi.org/10.1371/journal.pone.0124196.s001

(WMV)

Author Contributions

Conceived and designed the experiments: HYH DJY SKF. Performed the experiments: HYH DJY CHT HHS. Analyzed the data: HYH DJY CHL. Contributed reagents/materials/analysis tools: HYH DJY CHL WSH. Wrote the paper: HYH DJY CJL.

References

  1. 1. Steptoe PC, Edwards RG (1978) Birth after the reimplantation of a human embryo. Lancet 8085:366.
  2. 2. Biggers JD, McGinnis LK, Lawitts JA (2005) One-step versus two-step culture of mouse preimplantation embryos: is there a difference? Hum Reprod 20: 3376–3384. pmid:16123096
  3. 3. Lane M, Gardner DK (2007) Embryo culture medium: which is the best? Best Pract Res Clin Obstet Gynecol 21:83–100. pmid:17090393
  4. 4. Xie Y, Wang F, Puscheck EE, Rappolee DA (2007) Pipetting Causes Shear Stress and Elevation of Phosphorylated Stress-Activated Protein Kinase/Jun Kinase in Preimplantation Embryos. Mol Reprod Develop 74:1287–1294. pmid:17492777
  5. 5. Westrom L, Mardh PA, Mecklenburg CV, Hakansson CH (1977) Studies on ciliated epithelia of the human genital tract. II. The mucociliary wave pattern of fallopian tube epithelium. Fertil Steril 28:955–961. pmid:892048
  6. 6. Fauci L, Dillon R (2006) Biofluid mechanics of reproduction. Annual Review of Fluid Mechanics 38:371–394.
  7. 7. Smith GD, Swain JE, Bormann CL (2011) Microfluidics for gametes, embryos, and embryonic stem cells. Sem Reprod Med 29:5–14.
  8. 8. Swain JE, Smith GD (2011) Advances in embryo culture platforms: novel approaches to improve preimplantation embryo development through modifications of the microenvironment. Hum Reprod Update 17:541–557. pmid:21454356
  9. 9. Smith GD, Takayama S, Swain JE (2012) Rethinking in vitro embryo culture: new developments in culture platforms and potential to improve assisted reproductive technologies. Biol Reprod 86:1–10.
  10. 10. Swain JE, Lai D, Takayama S, Smith GD (2013) Thinking big by thinking small: application of microfluidic technology to improve ART. Lab Chip 13:1213–1224 pmid:23400523
  11. 11. Wheeler MB, Walters EM, Beebe DJ (2007) Toward culture of single gametes: The development of microfluidic platforms for assisted reproduction. Theriogenology 68S:178–189.
  12. 12. Han C, Zhang Q, Ma R, Xie L, Qiu T, et al. (2010) Integration of single oocyte trapping, in vitro fertilization and embryo culture in a microwell-structured microfluidic device. Lab Chip 10:2848–2854. pmid:20844784
  13. 13. Pyne DG, Liu J, Abdelgawad M, Sun Y (2014) Digital Microfluidic Processing of Mammalian Embryos for Vitrification. PLOS ONE 9(9): e108128. pmid:25250666
  14. 14. Cho BS, Schuster TG, Zhu X, Chang D, Smith GD, et al. (2003) Passively driven integrated microfluidic system for separation of motile sperm. Analytical Chemistry 75:1671–1675. pmid:12705601
  15. 15. Suh RS, Zhu X, Phadke N, Ohl DA, Takayama S, et al. (2006) IVF within microfluidic channels requires lower total numbers and lower concentrations of sperm. Hum Reprod 21:477–483. pmid:16199424
  16. 16. Seo D, Agca Y, Feng ZC, Critser JK (2007) Development of sorting, aligning, and orienting motile sperm using microfluidic device operated by hydrostatic pressure. Microfluidics Nanofluidics 3:561–70.
  17. 17. Huang HY, Yao DJ, Huang HR, Li CJ, Soong YK, et al. (2014) Isolation of motile spermatozoa with a microfluidic chip having a surface-modified microchannel. JALA 19:91–99. pmid:23603751
  18. 18. Huang HY, Fu HT, Huang HJ, Li JR, Yao DJ (2014) An Integrated Microfluidic System for Human Sperm Motility Sorting. J Nanomed Nanotech 5:3.
  19. 19. Byatt-Smith JG, Leese HJ, Gosden RG (1991) An investigation by mathematical modeling of whether mouse and human preimplantation embryos in static culture can satisfy their demands for oxygen by diffusion. Human Reproduction 6:52–57. pmid:1874957
  20. 20. Kölle S, Dubielzig S, Reese S, Wehrend A, König P, et al. (2009) Ciliary transport, gamete interaction, and effects of the early embryo in the oviduct: ex vivo analyses using a new digital videomicroscopic system in the cow. Biol Reprod 81: 267–274. pmid:19299315
  21. 21. Fan SK, Hsieh TH, Lin DY (2009) General digital microfluidic platform manipulating dielectric and conductive droplets by dielectrophoresis and electrowetting. Lab Chip 9:1236–1242. pmid:19370242
  22. 22. Welters WJJ, Fokkink LGJ (1998) Fast Electrically Switchable Capillary Effects. Langmuir 14:1535–1538.
  23. 23. Pollack MG, Fair RB, Shenderov AD (2000) Electrowetting-based actuation of liquid droplets for microfluidic applications. Applied Physics Letters 77: 1725–1726.
  24. 24. Lee J, Moon H, Fowler J, Schoellhammer T, Kim CJ (2002) Electrowetting and electrowetting-on-dielectric for microscale liquid handling. Sensors Actuators A-Physical 95:259–268.
  25. 25. Cho SK, Moon HJ, Kim CJ (2003) Creating, transporting, cutting, and merging liquid droplets by electrowetting-based actuation for digital microfluidic circuits. J Microelectromechanical Systems 12:70–80.
  26. 26. Luk VN, Mo GCH, Wheeler AR (2008) Pluronic additives: A solution to sticky problems in digital microfluidics. Langmuir 24:6382–6389. pmid:18481875
  27. 27. Liu YJ, Yao DJ, Lin HC, Chang WY, Chang HY (2008) DNA ligation of ultramicro volume using an EWOD microfluidic system with coplanar electrodes. J Micromech Microeng 18:41.
  28. 28. Lin HC, Liu YJ, Yao DJ (2010) Core-shell droplets for parallel dna ligation of an ultra-micro volume using an EWOD microfluidic system. JALA 15:210–215.
  29. 29. Nagy A, Gersenstein M, Vintersten K, Behringer R (2002) Surgical Procedures. In Manipulating the Mouse Embryo: A Laboratory Manual, 3rd edition, 251–289. New York USA: Cold Spring Harbor Laboratory Press.
  30. 30. Wang X, Catt S, Pangestu M, Temple-Smith P (2009) Live offspring from vitrified blastocysts derived from fresh and cryopreserved ovarian tissue grafts of adult mice. Reproduction 138:527–535. pmid:19556437
  31. 31. Pollack MG, Shenderov AD, Fair RB (2002) Electrowetting-based actuation of droplets for integrated microfluidics. Lab Chip. 2:96–101. pmid:15100841
  32. 32. Tomlinson AJ, Guzman NA, Naylor S (1995) Enhancement of concentration limits of detection in CE and CE-MS: a review of on-line sample extraction, cleanup, analyte preconcentration, and microreactor technology. J Capillary Electrophoresis 2:247–266. pmid:9384784
  33. 33. Brody JP, Yager P (1997) Diffusion-based extraction in a microfabricated device. Sensors Actuators A-Physical 58:13–18.
  34. 34. Kenis PJA, Ismagilov RF, Whitesides GM (1999) Microfabrication inside capillaries using multiphase laminar flow patterning. Science 285:83–85. pmid:10390366
  35. 35. Takayama S, Ostuni E, LeDuc P, Naruse K, Ingber DE, et al. (2001) Laminar flows: subcellular positioning of small molecules. Nature 411:1016. pmid:11429594
  36. 36. Verpoorte E (2002) Microfluidic chips for clinical and forensic analysis. Electrophoresis 23 677–712. pmid:11891702
  37. 37. Huang WH, Cheng W, Zhang Z, Pang DW, Wang ZL, et al. (2004) Transport, location, and quantal release monitoring of single cells on a microfluidic device. Analytical Chemistry 76:483–488. pmid:14719902
  38. 38. Kim MS, Bae CY, Wee G, Han YM, Park JK (2009) A microfluidic in vitro cultivation system for mechanical stimulation of bovine embryos. Electrophoresis 30:3276–3282. pmid:19705355
  39. 39. Melin J, Lee A, Foygel K, Leong DE, Quake SR, et al. (2009) In Vitro Embryo Culture in Defined, Sub-microliter Volumes. Developmental Dynamics 238: 950–955. pmid:19301395
  40. 40. Sano H, Matsuura K, Naruse K, Funahashi H (2010) Application of a microfluidic sperm sorter to the in-vitro fertilization of porcine oocytes reduced the incidence of polyspermic penetration. Theriogenology 74:863–870. pmid:20537694
  41. 41. Sugimura S, Akai T, Somfai T, Hirayama M, Aikawa Y, et al. (2010) Time-lapse cinematography-compatible polystyrene-based microwell culture system: a novel tool for tracking the development of individual bovine embryos. Biol Reprod 83:970–978. pmid:20739661
  42. 42. Hickman DL, Beebe DJ, Rodriquez-Zas SL, Wheeler MB (2002) Comparison of static and dynamic medium environments for culturing of pre-implantation mouse embryos. Comparative Medicine 52:122–126. pmid:12022391
  43. 43. Paria BC, Dey SK (1990) Preimplantation embryo development in vitro: cooperative interactions among embryos and role of growth factors. Proc Natl Acad Sci USA 87:4756–4760. pmid:2352946
  44. 44. Heo YS, Cabrera LM, Bormann CL, Smith GD, Takayama S (2012) Real time culture and analysis of embryo metabolism using a microfluidic device with deformation based actuation. Lab Chip 12:2240–2246. pmid:22402469
  45. 45. Raty S, Walters EM, Davis J, Zeringue H, Beebe DJ, et al. (2004) Embryonic development in the mouse is enhanced via microchannel culture. Lab Chip 4:186–190. pmid:15159776
  46. 46. Clark SG, Haubert K, Beebe DJ, Ferguson CE, Wheeler MB (2005) Reduction of polyspermic penetration using biomimetic microfluidic technology during in vitro fertilization. Lab Chip 5:1229–1232. pmid:16234945
  47. 47. Swain JE, Pool TB, Takayama S, Smith GD (2008) Microfluidic technology in art: Is it time to go with the flow? J Clin Embryol 11:3.
  48. 48. Sharma S1, Srisa-Art M, Scott S, Asthana A, Cass A (2013) Droplet-based microfluidics. Methods Mol Biol 949:207–230. pmid:23329446
  49. 49. Malic L, Brassard D, Veres T, Tabrizian M (2010) Integration and detection of biochemical assays in digital microfluidic LOC devices. Lab Chip 10:418–431. pmid:20126681
  50. 50. Hunter HC, Ray AK (2009) On progeny droplets emitted during Coulombic fission of charged microdrops. Phys Chem Chem Phys 11:6156–6165. pmid:19606325
  51. 51. Berry S, Kedzierski J, Abedian B (2006) Low voltage electrowetting using thin fluoroploymer films. Journal of Colloid and Interface Science 303:517–524. pmid:16949597
  52. 52. Lin YY, Evans RD, Welch E, Hsu BN, Madison AC, et al. (2010) Low voltage electrowetting-on-dielectric platform using multi-layer insulators. Sensors Actuators B-Chemical 150:465–470. pmid:20953362
  53. 53. Lin YY, Welch ERF, Fair RB (2012) Low voltage picoliter droplet manipulation utilizing electrowetting-on-dielectric platforms. Sensors Actuators B-Chemical 173:338–345. pmid:23559693
  54. 54. Bogojevic D, Chamberlain MD, Barbulovic-Nad I, Wheeler AR (2012) A digital microfluidic method for multiplexed cell-based apoptosis assays. Lab Chip 12: 627–634. pmid:22159547
  55. 55. Eydelnant IA, Uddayasankar U, Li BY, Liao MW, Wheeler AR (2012) Virtual microwells for digital microfluidic reagent dispensing and cell culture. Lab Chip 12:750–757. pmid:22179581
  56. 56. Srigunapalan S, Eydelnant IA, Simmons CA, Wheele AR (2012) A digital microfluidic platform for primary cell culture and analysis. Lab Chip 12:369–375. pmid:22094822
  57. 57. Jaffe LF, Poo MM (1979) Neurites grow faster towards the cathode than the anode in a steady field. J Experimental Zoology 209:115–128. pmid:490126
  58. 58. Hinkle LC, McCaig D, Kenneth KR (1981) The direction of growth of differantiating neurons and myoblasts from frog embryos in an applied electric field. J Physiology 314:121–135. pmid:7310685
  59. 59. Patel N, Poo MM (1982) Orientation of neurite growth by extracellular electric fields. J Neuroscience 2:483–496. pmid:6279799
  60. 60. Robinson KR (1985) The Responses of Cells to Electrical Fields: A Review. J Cell Biol 101:2023–2027. pmid:3905820
  61. 61. Rickords LF, White KL (1992) Effect of electrofusion pulse in either electrolyte or nonelectrolyte fusion medium on subsequent murine embryonic development. Mol Reprod Develop 32:259–264. pmid:1497875
  62. 62. Suzuki H, Ogasawara I, Takahashi H, Imada Y, Toyokawa K (2001) Electrofusion of blastomeres of hamster 2-cell embryos and dynamic changes of the cytoskeletal distribution. J Reprod Develop 47:227–235.
  63. 63. Krivokharchenko A, Popova E, Zaitseva I, Vil’ianovich L, Ganten D, et al. (2003) Development of parthenogenetic rat embryos. Biol Reprod 68:829–836. pmid:12604632
  64. 64. Zhang J, Wang CW, Blaszcyzk A, Grifo JA, Ozil J, et al. (1999) Electrical activation and in vitro development of human oocytes that fail to fertilize after intracytoplasmic sperm injection. Fertil Steril 72:509–512. pmid:10519625
  65. 65. Popova E, Bader M, Krivokharchenko A (2011) Effects of electric field on early preimplantation development in vitro in mice and rats. Human Reproduction 26:662–670. pmid:21227946