Abstract
Purpose
To determine the feasibility of using the Ultrasound Cardiac Output Monitor (USCOM) as an adjunct during hemodynamic assessments by a pediatric medical emergency team (PMET).
Methods
Pediatric in-patients at McMaster Children’s Hospital aged under 18 years requiring urgent PMET consultation, were eligible. Patients with known cardiac outflow valve defects, Pediatric Critical Care Unit in-patients, and those in cardiorespiratory arrest, were excluded. The primary outcome was feasibility, and the ease of USCOM transport and application as assessed by a self-administered user questionnaire. Secondary outcomes included the quality of USCOM measurements, and agreement in clinical versus USCOM-derived assessments.
Results
Forty-one patients from 85 eligible PMET consultations were enrolled between March and August 2011. A total of 55 USCOM assessments were performed on 36 of 41 (87.8%) participants. USCOM could not be completed in 5 (12.2%) participants due to patient agitation (n = 4) and emergent care (n = 1). USCOM was reported as easy to transport and apply by 97.4% and 94.7% of respondents respectively, not obstructive to patient care by 94.7%, and yielded timely measurements by 84.2% respondents. USCOM tracings were of good quality in 41 (75.9%) assessments. Agreement between clinical and USCOM-derived hemodynamic assessments by two independent raters was poor (Rater 1: κ = 0.094; Rater 2: κ = 0.146).
Conclusion
USCOM can be applied by a PMET during urgent hemodynamic assessments in children. While USCOM has been validated in stable children, its role in guiding hemodynamic resuscitation and informing therapeutic goals in a hemodynamically unstable pediatric population requires further investigation.
Citation: Zorko DJ, Choong K, Gilleland J, Agar B, Baker S, Brennan C, et al. (2013) Urgent Ultrasound Guided Hemodynamic Assessments by a Pediatric Medical Emergency Team: A Pilot Study. PLoS ONE 8(6): e66951. https://doi.org/10.1371/journal.pone.0066951
Editor: Partha Mukhopadhyay, National Institutes of Health, United States of America
Received: March 20, 2013; Accepted: May 13, 2013; Published: June 25, 2013
Copyright: © 2013 Zorko et al. This is an open-access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Funding: With respect to funding received from USCOM Limited, the authors declare that USCOM Limited provided the ultrasound cardiac output monitory device on loan to McMaster University for the duration of this specific research project. The authors did not receive any other funding from USCOM Limited or other sources to support this study. USCOM Limited has no intellectual property of the design, execution, analysis, interpretation, and reporting of the results, or manuscript preparation for this study. As such, this does not alter the authors’ adherence to all the PLOS ONE policies on sharing data and materials.
Competing interests: There are no potential conflicts of interest relevant to subject of this manuscript. This is an investigator-initiated project. USCOM Limited provided the ultrasound cardiac output monitoring device for the duration of the project and financial support for the statistical analysis, which was conducted at McMaster University. USCOM Limited has no intellectual property of the design, execution, analysis, interpretation, and reporting of the results, or manuscript preparation for this study.
Introduction
Timely recognition and early resuscitation of pediatric patients according to their hemodynamic physiology and goal-directed endpoints improves survival and functional outcomes in children with shock [1]–[4]. However, a clinician’s ability to estimate the hemodynamic status of children and infants based on the physical exam alone is poor [5]. There are currently several adjunctive methods of assessing a patient’s hemodynamic status, such as pulmonary artery catheters, pulse contour cardiac output, 2D echocardiography, and central venous oximetry [2], [6]. However, these tools are limited by their need for invasive access, inability to provide real-time measurements, the prerequisite expertise and setting required to conduct some of these measurements (i.e. an intensive care unit), and limited evidence of efficacy on clinically important outcomes [6]–[8]. Therefore, children requiring hemodynamic resuscitation outside the pediatric critical care unit (PCCU) remain primarily dependent on the clinical assessment. A reliable, non-invasive, objective method of assessing hemodynamic physiology that can be easily and rapidly applied in a broad range of children within and beyond the PCCU would be a valuable adjunct to optimizing the resuscitation of children in shock [9], [10].
The Ultrasound Cardiac Output Monitor (USCOM®; USCOM Ltd., Sydney, Australia) is a non-invasive bedside monitoring tool that utilizes continuous-wave Doppler ultrasound to measure hemodynamic parameters such as preload, systemic vascular resistance, cardiac index, and inotropy [11]. USCOM has been validated against echocardiography and cardiac catheter measurements in neonatal and pediatric normal cardiac anatomy populations, respectively [12]–[16]. This tool has been increasingly used to assess hemodynamic status in adult and pediatric critical care settings [9], [17]–[19], and continues to be a recommended method of monitoring therapeutic endpoints in the resuscitation of septic shock [9], [20]. However, its impact on patient important outcomes, and its use beyond the PCCU setting and by non-physician clinicians not been evaluated to date. The objective of this prospective observational pilot study was to evaluate the feasibility of using USCOM as an adjunctive tool during urgent hemodynamic assessments in children by members of a pediatric medical emergency team (PMET). We hypothesized that physician and non-physician PMET members could promptly apply USCOM during patient assessments and obtain USCOM measurements of similar quality. As there is currently no gold-standard method for the non-invasive measurement of hemodynamic endpoints in children, we also explored the agreement between USCOM-derived and clinician hemodynamic assessments.
Materials and Methods
Ethics Statement
The Faculty of Health Sciences/Hamilton Health Sciences Research Ethics Board approved a waived consent model for this study, given the minimal-risk and time-sensitive study design.
Study Participants
This pilot study was conducted at McMaster Children’s Hospital (Hamilton, Canada) between March and August 2011. Consecutive in-patients at the institution aged 0 to 18 years, requiring urgent assessment by the PMET, were eligible. Patients with known valvular cardiac defects, admitted to the PCCU, or requiring “code blue” activation (i.e. cardiorespiratory arrest) or elective PMET follow-up assessments, were excluded. At the institution, a “Code Blue Team,” not the PMET, attends cardiorespiratory arrests.
Intervention
The PMET at McMaster Children’s Hospital can be urgently consulted for critical patient events (Table S1) [21]. All PMET consultations are attended by the following: a registered nurse (RN), a registered respiratory therapist (RRT), a resident in training, and a critical care trained physician (PCCU fellow or attending). For the purposes of this study, 5 PMET members (2 attending physicians, 2 RRTs, and 1 RN) and the principle investigator (DZ) were trained in using USCOM [11]. The technique for obtaining USCOM measurements was adhered to as has been previously described [10].
In this study, USCOM measurements were performed at each eligible initial PMET patient consultation and in follow-up if clinically indicated. To minimize interference with patient management, USCOM was applied only after the clinical assessment by the PMET was completed. USCOM-derived measurements were available only to the USCOM operator, and were blinded to the study participants, healthcare staff, other PMET members, and the independent physician who provided a clinical impression of the patient’s hemodynamic status assessed just prior to the USCOM measurement. Demographic, clinical, and outcome data were recorded on standardized case report forms for each study participant.
Outcomes
The primary outcome was feasibility, as assessed by: (a) ability to enrol the sample size over the anticipated study duration; (b) protocol adherence, as assessed by the proportion of eligible PMET consultations in which USCOM was successfully conducted; and (c) the ease of transporting and applying the USCOM urgently in participants, as assessed on a 7-point Likert scale self-administered questionnaire with optional commentary.
Secondary outcomes included a quality assessment of USCOM traces, and agreement between clinician and USCOM-derived assessments of the patient’s hemodynamic status. USCOM trace quality was assessed independently by two investigators (DZ, KC) using a previously published 12-point scoring tool [22]. The quality of USCOM traces was categorized as either good (≥8 points) or poor (<8 points). Disagreements were resolved by consensus. USCOM-derived hemodynamic status was assessed independently by two investigators (KC, JG) using age-dependant normal range values for cardiac index, stroke volume, and systemic vascular resistance (Table S2, Table S3) [23]–[25]. Using these parameters, hemodynamic status was categorized as either euvolemic, hypovolemic, cardiogenic, hyperdynamic, or indeterminate. Each USCOM-derived assessment was then compared by an independent investigator (DZ) to the blinded physician’s clinical impression of the patient’s hemodynamic status.
Statistical Analyses
We estimated a convenience sample size of 40 patients over a period of 20 weeks, in order to demonstrate feasibility. This estimate was based on previous PMET data of approximately 37 new consultations per month, 57% of which occur during the day shift, and the availability of USCOM operators during these shifts. Descriptive statistics are presented as counts and percentages with 95% confidence interval (CI), means (standard deviation [SD]) or medians (minimum [min], maximum [max]) as appropriate. Agreement was assessed using the kappa statistic. Kappa values ≥0.40 represented moderate to good agreement. The quality of USCOM tracings between clinician groups was assessed using the Chi2 test, and an alpha = 0.05 (2-sided) level of statistical significance. All analyses were performed using SPSS version 20.0 (SPSS Inc., Chicago, IL). This study was reported in accordance with the Strengthening the Reporting of Observational Studies in Epidemiology (STROBE) statement [26].
Results
Eighty-nine PMET consultations were screened for eligibility during the 20-week study period between March and August 2011. Eighty-five patients met eligibility criteria, of which 41 (48.2%) were enrolled. Baseline characteristics of the study participants are presented in Table 1. Complete USCOM measurements were obtained during initial PMET assessments in 36 of the 41 (87.8%) participants, and in 19 of 22 (86.4%) scheduled follow-up visits (Figure 1). USCOM could not be conducted or was aborted in 5 (12.2%) participants because of patient agitation and emergency patient care. Of the 55 total USCOM assessments, 15 (27.3%) were conducted by a physician PMET member, 21 (38.2%) by RRT, 11 (20.0%) by the principal investigator, and 8 (14.5%) by RN.
aMissed patients indicates research personnel were available, but patient was not screened.
The response rate for the user questionnaire was 92.7% (38/41). Three surveys were not completed early in the study when the USCOM was aborted as described above. Respondents’ impressions of applying USCOM as part of their assessments are presented in Table 2. The majority (97.4%; 37/38) of respondents found USCOM somewhat easy or easy to transport to the bedside, 94.7% (36/38) reported it was easy to apply, and 84.2% (32/38) felt that it yielded timely measurements. Only 5.3% (2/38) of respondents found USCOM somewhat obstructive to the patient care area. Respondents attributed unfavourable scores to the following reasons: patient agitation (n = 6), body habitus (kyphoscoliosis; n = 1), and waveform artefact in a patient on non-invasive mechanical ventilation (n = 1).
The quality scores for USCOM tracings are presented in Table 2. Fifty-four tracings were available for quality assessment, of which 41 (75.9%) were of good quality. Of the 13 (24.1%) that were rated as poor quality, 9 were attributed to patient-related factors (i.e. agitation, tachycardia), 1 due to waveform artefact as described above, and 3 were operator-related. There was no statistical difference in the quality of USCOM tracings according to clinician type (p = 0.850).
The categorization of hemodynamic status from USCOM-derived measurements for each of the two independent raters is presented in Table 2. The inter-rater agreement was moderate (κ = 0.442, 95% CI [0.273, 0.610]). However, agreement between hemodynamic status according to USCOM and clinical impression was poor (Rater 1: κ = 0.094, 95% CI [−0.110, 0.298]; Rater 2: κ = 0.146, 95% CI [−0.007, 0.299]).
Discussion
Pilot studies are an important and underutilized prerequisite to larger scale interventional trials, particularly in the setting of urgent and critical care pediatrics [27]. Such studies can provide important information regarding the feasibility of executing intended trial procedures, protocol adherence, enrolment rates, appropriateness of proposed eligibility criteria and potential outcomes of interest, as well as safety data. Feasibility outcomes derived from pilot studies are crucial to inform the methods of adequately powered and rigorously designed interventional clinical trials [28]. Therefore, we conducted this pilot study to assess important feasibility outcomes prior to an anticipated future interventional study. The PMET environment is complex and assessments are time-sensitive. Hence it is important to evaluate the utility, portability, and applicability of a new monitoring device within this dynamic team environment to ensure that patient care and safety are maintained, prior to future efficacy studies.
This pilot study demonstrates the following key findings. First, USCOM can be applied during urgent hemodynamic assessments in pediatric patients outside of the PCCU environment, and the device can be easily transported and integrated in a PMET setting without interfering with patient care. Second, USCOM measurements can be acquired by trained physician and non-physician personnel. In this study, USCOM measurements obtained by physicians and non-physicians were of similarly good quality, and poor quality was most often attributed to patient-related factors. Third, there was only moderate agreement in the interpretation of hemodynamic status with USCOM-derived data, and poor correlation between the clinical impression of hemodynamic status and that derived from the USCOM.
This is the first study to our knowledge to evaluate the urgent use of USCOM for hospitalized children outside of a PCCU setting, and by non-physician clinicians. This has significant implications, given that early recognition and prompt reversal of cardiorespiratory decompensation improves clinical outcomes and survival amongst children with septic shock [29]–[31]. We hypothesized that both physician and non-physician PMET members can be trained in operating USCOM, and this study revealed that these PMET members demonstrated similar competency in acquiring USCOM measurements. This has potential importance if USCOM is ultimately shown to be a useful adjunct for monitoring therapeutic endpoints during resuscitation, as hemodynamic assessments can be conducted by physician or non-physician trained clinicians. Currently, some PMETs are not led by physician members [32], [33]. We also observed a higher proportion of good quality USCOM tracings and user satisfaction than previously reported [34].
We used previously published USCOM-derived hemodynamic variables for the classification of USCOM-derived hemodynamic status of the participants [23]–[25]. Both raters reported that assigning hemodynamic physiology based on USCOM in isolation of physical examination data was challenging. Along with a participant’s heart rate and blood pressure, additional significant physical findings were felt to be important to each rater’s overall impression, particularly if the patient was suspected to have compensated shock or other confounding factors that may influence heart rate, blood pressure, and perfusion. This likely contributed to the moderate agreement between raters in USCOM-derived classification of hemodynamic status. We used a limited number of key USCOM parameters to categorize hemodynamic status, which we felt were pragmatic [35], [36]. Including additional indices such as stroke volume responsiveness and Smith-Madigan Inotropy Index may have added supportive data to the assessments; however, the required software and pediatric reference ranges had not been validated at the time of our study.
An objective of this pilot study was to evaluate the correlation between clinical hemodynamic assessment and USCOM methods, rather than to validate which method is superior. The poor correlation between USCOM and clinical assessments may be explained by the following. First, there are limitations to USCOM. Pediatric USCOM validation studies were performed previously under more controlled conditions in healthy individuals, anesthetised or sedated patients in the operating room and PCCU [13], [14], [23]–[25]. As a result, it is unclear how reliable its measurements are in physiologically unstable or decompensating patients. Second, there are limitations to hemodynamic assessments based on clinical examination alone, which are well recognized [5], [9]. Third, both methods are limited when used in isolation, while the combination of information derived from both may increase the validity of hemodynamic assessments. Clinicians currently remain dependent on physical examination to provide “real-time” information on response to therapy. Adjunctive tools that can provide rapid, non-invasive information regarding a patient’s physiology may further optimize hemodynamic assessments and subsequent interventions. Only one published study to date suggests USCOM may be useful in decision-making regarding the choice of vasoactive drugs in children with septic shock [9]. However, as neither the clinical examination nor USCOM are established gold standards for hemodynamic measurement in children, the results of this pilot trial should be interpreted with caution. They rationalize the need for further validation and evidence to inform how such technology may be incorporated in goal-directed therapeutic targets within pediatric resuscitation guidelines [20].
The main limitation to enrolment in this pilot study was the availability of research personnel, which is not unique to this study. Nevertheless, we were able to apply the intervention to approximately half of the eligible patients and achieve our projected enrolment rate. Despite the reported ease of its use and transportability, there are potential barriers to the application of USCOM by the PMET. We observed that poor quality tracings and abandonment of USCOM assessment were most often related to patient distress or agitation, since noise (e.g. crying) and movement can alter USCOM measurements. Similar technical issues were observed in relation to patient body habitus, which can affect ideal patient positioning required to conduct an USCOM assessment. Waveform artefact from non-invasive mechanical ventilation has not been reported in the past and requires further study.
Conclusions
Current evidence suggests that effective management of shock in pediatric in-patients outside of the critical care setting is hindered by limitations in the physical exam and available adjuncts to hemodynamic measurement. Non-invasive tools that can better rationalize and monitor resuscitative decision-making may therefore yield better outcomes for these children. The results of this study indicate that USCOM is a feasible adjunct for rapid hemodynamic assessment for pediatric patients; however, our preliminary observations regarding the discordance between the clinical assessment and USCOM-derived measurements strongly support the need for further research. While USCOM is attractive as a rapid non-invasive tool to guide our clinical decision-making, it requires further evaluation of its utility and validity as an adjunct to monitoring goal-directed therapy, and its correlation with patient important outcomes.
Supporting Information
Table S1.
McMaster Children’s Hospital Pediatric Medical Emergency Team activation guidelines. SaO2 indicates oxygen saturation; FiO2, fraction of inspired oxygen.
https://doi.org/10.1371/journal.pone.0066951.s001
(DOC)
Table S2.
Categorization of Hemodynamic Status. Hemodynamic status was assessed in the context of abnormal vital signs, with the understanding that MAP may be normal in compensated shock, and heart rate and blood pressure may be elevated due to non-circulatory causes [35], [36]. MAP indicates mean arterial pressure; HR, heart rate; CI, cardiac index; SVRI, systemic vascular resistance index; SV, stroke volume.
https://doi.org/10.1371/journal.pone.0066951.s002
(DOC)
Table S3.
USCOM-derived hemodynamic reference ranges in healthy children [23]–[25]. HR indicates heart rate; SV, stroke volume; CI, cardiac index; SBP, systolic blood pressure; SVRI, systemic vascular resistance index.
https://doi.org/10.1371/journal.pone.0066951.s003
(DOC)
Acknowledgments
This study was conducted in accordance with Good Clinical Practice Guidelines. We thank study patients and families for participating in the trial in helping us address our research question. We also thank the allied health, nursing, and physician staff of the Pediatric Medical Emergency Team at McMaster Children’s Hospital for their cooperation in the successful completion of this pilot study, and Beverley Jacobson and USCOM Limited for providing staff training and technical support with the USCOM device.
Author Contributions
Conceived and designed the experiments: KC DZ JG EP. Performed the experiments: DZ KC JG BA SB CB. Analyzed the data: DZ KC JG EP. Contributed reagents/materials/analysis tools: DZ KC JG EP. Wrote the paper: DZ KC. Review and approval of final manuscript: JG EP CB BA SB DZ KC.
References
- 1. Carcillo JA, Kuch BA, Han YY, Day S, Greenwald BM, et al. (2009) Mortality and functional morbidity after use of PALS/APLS by community physicians. Pediatrics 124: 500–508.
- 2. Brierley J, Carcillo JA, Choong K, Cornell T, Decaen A, et al. (2009) Clinical practice parameters for hemodynamic support of pediatric and neonatal septic shock: 2007 update from the American College of Critical Care Medicine. Crit Care Med 37: 666–688.
- 3. Carcillo JA, Fields AI (2002) Clinical practice parameters for hemodynamic support of pediatric and neonatal patients in septic shock. Crit Care Med 30: 1365–1378.
- 4. Ceneviva G, Paschall JA, Maffei F, Carcillo JA (1998) Hemodynamic support in fluid-refractory pediatric septic shock. Pediatrics 102: e19.
- 5. Tibby SM, Hatherill M, Marsh MJ, Murdoch IA (1997) Clinicians’ abilities to estimate cardiac index in ventilated children and infants. Arch Dis Child 77: 516–519.
- 6. Vincent JL, Rhodes A, Perel A, Martin GS, Della Rocca G, et al. (2011) Clinical review: Update on hemodynamic monitoring – a consensus of 16. Critical Care 15: 229.
- 7. Oren-Grinberg A (2010) The PiCCO Monitor. Int Anesthesiol Clin 48: 57–85.
- 8. Harvey S, Harrison DA, Singer M, Ashcroft J, Jones CM, et al. (2005) Assessment of the clinical effectiveness of pulmonary artery catheters in management of patients in intensive care (PAC-Man): a randomized controlled trial. Lancet 366: 472–477.
- 9. Brierley J, Peters M (2008) Distinct Hemodynamic Patterns of Septic Shock at Presentation to Pediatric Intensive Care. Pediatrics 122: 752–759.
- 10. Jain S, Allins A, Salim A, Vafa A, Wilson MT, et al. (2008) Noninvasive Doppler ultrasonography for assessing cardiac function: can it replace the Swan-Ganz catheter? Am J Surg 196: 961–968.
- 11.
USCOM Ltd (2009) USCOM: The Basics [training module]. USCOM, Ltd.
- 12. Phillips R, Paradisis M, Evans N, Southwell D, Burstow D, et al. (2006) Cardiac output measurement in preterm neonates: validation of USCOM against echocardiography. Critical Care 10 Sup 1: P343.
- 13. Likes ML, Alejos JC (2010) A non-invasive assessment of cardiac output in pediatric heart transplant patients [abstract]. J Heart Lung Transplant 29: S172.
- 14. Dhanani S, Barrowman NJ, Ward RE, Murto KT (2011) Intra- and inter-observer reliability using a noninvasive ultrasound cardiac output monitor in healthy anesthetized children. Paediatr Anaesth 21: 858–864.
- 15. Knirsch W, Kretschmar O, Tomaske M, Stutz K, Nagdyman N, et al. (2008) Cardiac output measurement in children: comparison of the Ultrasound Cardiac Output Monitor with thermodilution cardiac output measurement. Intensive Care Med 34: 1060–1064.
- 16.
He SR, Sun X, Zhang C, Jian Z, Sun YX, et al.. (2012) Measurement of systemic oxygen delivery and inotropy in healthy term neonates with the Ultrasonic Cardiac Output Monitor (USCOM). Early Hum Dev. https://doi.org/10.1016/j.bbr.2011.03.031
- 17. Van Lelyveld-Haas LEM, van Zanten ARH, Borm GF, Tjan DHT (2008) Clinical validation of the non-invasive cardiac output monitor USCOM-1A in critically ill patients. Eur J Anaesthesiol 25: 917–924.
- 18. Tan HL, Pinder M, Parsons R, Roberts B, van Heerden PV (2005) Clinical evaluation of USCOM ultrasonic cardiac output monitor in cardiac surgical patients in intensive care unit. BJA 94: 287–291.
- 19. Chand R, Mehta Y, Trehan N (2006) Cardiac output estimation with a new Doppler device after off-pump coronary artery bypass surgery. J Cardiothorac Vasc Anesth 20: 315–319.
- 20. Dellinger RP, Levy MM, Carlet JM, Bion J, Parker MM, et al. (2008) Surviving Sepsis Campaign: International guidelines for management of severe sepsis and septic shock. Intensive Care Medicine 34: 17–60.
- 21. Tibballs J, Kinney S, Duke T, Oakley E, Hennessy M (2005) Reduction of pediatric in-patient cardiac arrest and death with a medical emergency team: preliminary results. Arch Dis Child 90: 1148–1152.
- 22.
On Tang C, Cattermole GN, Leung PYM, Mak PSK, Smith BE, et al.. (2009) A new method to score the quality of USCOM scans [poster presentation]. Scientific Symposium on Emergency Medicine, Hong Kong, China.
- 23. Cattermole GN, Leung PYM, Mak PSK, Chan SSW, Graham CA, et al. (2010) The normal ranges of cardiovascular parameters in children measured using the Ultrasonic Cardiac Output Monitor. Pediatr Crit Care Med 38: 1875–1881.
- 24.
Ho GYL, Cattermole GN, Chan SSW, Smith BE, Graham CA, et al.. (2013) Non-invasive transcutaneous Doppler ultrasound derived hemodynamic reference ranges in Chinese adolescents. Pediatr Crit Care Med: In press.
- 25.
Rainer TH, Lau PWZ, Cattermole GN, Ho GYL, Chan SSW, et al. (2013) Comparison of non-invasive transcutaneous Doppler ultrasound derived hemodynamic referene ranges in Chinese and Caucasian adults. Anesth Analg: In press.
- 26. von Elm E, Altman DG, Egger M, Pocock SJ, Gotzsche PC, et al. (2007) The Strengthening the Reporting of Observational Studies in Epidemiology (STROBE) statement: guidelines for reporting observational studies. Lancet 370: 1453–1457.
- 27. Arnold DM, Burns KE, Adhikari NK, Kho ME, Meade MO, et al. (2009) The design and interpretation of pilot trials in clinical research in critical care. Crit Care Med 37: S69–S74.
- 28. Thabane L, Ma J, Chu R, Cheng J, Ismaila A, et al. (2010) A tutorial on pilot studies: the what, why and how. BMC Medical Research Methodology 10: 1.
- 29. Carcillo JA, Davis AL, Zaritsky A (1991) Role of early fluid resuscitation in pediatric septic shock. JAMA 266: 1242–1245.
- 30. Ranjit S, Kissoon N, Jayakumar I (2005) Aggressive management of dengue shock syndrome may decrease mortality rate: a suggested protocol. Pediatr Crit Care Med 6: 412–419.
- 31. Inwald DP, Tasker RC, Peters MJ, Nadel S (2009) Paediatric Intensive Care Society Study Group (2009) Emergency management of children with severe sepsis in the United Kingdom: the results of the Paediatric Intensive Care Society sepsis audit. Arch Dis Child 94: 348–353.
- 32. Jenkins S, Lindsey P (2010) Clinical nurse specialists as leaders in rapid response. Clin Nurse Spec 24: 24–30.
- 33. Scherr K, Wilson DM, Wagner J, Haughian M (2012) Evaluating a new rapid reponse team. AACN Advanced Critical Care 23: 32–42.
- 34. Duchateau FX, Gauss T, Burnod A, Ricard-Hibon A, Juvin P, et al. (2011) Feasibility of cardiac output estimation by ultrasonic cardiac output monitoring in the prehospital setting. Eur J Emerg Med 18: 357–359.
- 35. Bonanno FG (2011) Clinical pathology of the shock syndromes. J Emerg Trauma Shock 4: 233–243.
- 36. Holmes CL, Walley KR (2003) The evaluation and management of shock. Clin Chest Med 24: 775–790.