Optical Characterization of Two-Layered Turbid Media for Non-Invasive, Absolute Oximetry in Cerebral and Extracerebral Tissue

Abstract

We introduce a multi-distance, frequency-domain, near-infrared spectroscopy (NIRS) method to measure the optical coefficients of two-layered media and the thickness of the top layer from diffuse reflectance measurements. This method features a direct solution based on diffusion theory and an inversion procedure based on the Levenberg-Marquardt algorithm. We have validated our method through Monte Carlo simulations, experiments on tissue-like phantoms, and measurements on the forehead of three human subjects. The Monte Carlo simulations and phantom measurements have shown that, in ideal two-layered samples, our method accurately recovers the top layer thickness (L), the absorption coefficient (µa) and the reduced scattering coefficient (µs) of both layers with deviations that are typically less than 10% for all parameters. Our method is aimed at absolute measurements of hemoglobin concentration and saturation in cerebral and extracerebral tissue of adult human subjects, where the top layer (layer 1) represents extracerebral tissue (scalp, skull, dura mater, subarachnoid space, etc.) and the bottom layer (layer 2) represents cerebral tissue. Human subject measurements have shown a significantly greater total hemoglobin concentration in cerebral tissue (82±14 µM) with respect to extracerebral tissue (30±7 µM). By contrast, there was no significant difference between the hemoglobin saturation measured in cerebral tissue (56%±10%) and extracerebral tissue (62%±6%). To our knowledge, this is the first time that an inversion procedure in the frequency domain with six unknown parameters with no other prior knowledge is used for the retrieval of the optical coefficients and top layer thickness with high accuracy on two-layered media. Our absolute measurements of cerebral hemoglobin concentration and saturation are based on the discrimination of extracerebral and cerebral tissue layers, and they can enhance the impact of NIRS for cerebral hemodynamics and oxygenation assessment both in the research arena and clinical practice.

Introduction

The ability to noninvasively measure the concentrations of oxidized and reduced hemoglobin (i.e. oxy-hemoglobin and deoxy-hemoglobin, respectively) in biological tissues with high temporal resolution (in the order of milliseconds) is a prominent feature of near-infrared spectroscopy (NIRS) technology. This in turn allows for real-time monitoring and spatial mapping of underlying physiological parameters such as blood volume, blood perfusion, metabolic rate of oxygen, and oxygen delivery. Of current interest is the absolute measurement of the concentration and oxygen saturation of hemoglobin in the human brain, particularly by taking into account the extracerebral tissue contributions that can confound non-invasive optical measurements.

Over the past years, NIRS has seen an increasing appeal in the clinical realm. This has given rise to commercial devices that perform non-invasive cerebral oximetry by providing an estimate of oxygen saturation of hemoglobin in the human head (variously termed tissue oxygenation index (TOI), tissue saturation index (TSI), regional oxygen saturation (rSO2) and oxygen saturation (SO2)) (for a review, see Wolf et al. 2007 [1]). Currently explored applications of clinical relevance include monitoring post traumatic brain injury [2], ischemic stroke [3], coronary bypass surgery [4], and detecting, for instance, cardiopulmonary dysfunction [5], cerebral hemorrhage [6], and impaired cerebral autoregulation [7]. Despite the strong indication that accurate real-time NIRS measurement of cerebral hemoglobin parameters could significantly improve critical patient care and routine clinical practice [2], [8], [9], concerns remain on the reliability of the current methods particularly due to extracerebral tissue contamination in the NIRS signals [10], [11].

In a common research paradigm, functional NIRS (fNIRS) approaches are employed to study hemodynamic and metabolic responses to brain activation, which involve measurements of relative changes of oxy- and deoxy-hemoglobin concentrations [12], [13], [14]. Existing fNIRS strategies to partly account for extracerebral hemodynamic trends include devising task-based experimental protocols to inhibit systemic contaminants [15], [16], [17], [18]; averaging measured hemodynamic changes over multiple events (as employed by most fNIRS studies) [19]; and performing two-distance measurements and fitting and regressing out detected near signals (i.e. near the point of illumination) from far signals (i.e. far from the point of illumination) [19], [20], [21], [22], [23]. Other strategies employ methods such as adaptive filtering [24], [25]; principal component analysis (PCA) [26], [27], [28]; and independent component analysis (ICA) [28], [29]. Furthermore, computationally more sophisticated diffuse optical tomography (DOT) methods feature volumetric mesh based modeling of the human head with the aim of localizing measured hemodynamic changes under the assumption that underlying tissue is homogenous [30], [31], [32], [33] or layered [34].

While addressing the extracerebral contamination to some degree, fNIRS studies do not perform a baseline correction of measured dynamic changes (i.e. measured changes in optical coefficients divided by baseline/resting state optical coefficients), which has been shown to be an essential step for the determination of the true inter-individual variability in a cohort in rat [35]. DOT studies recognize the importance of baseline values and employ absolute baseline/resting optical coefficients in their forward modeling [30], [34]; however these studies rely on assumed values from the literature, thus neglecting inter-subject and intra-subject variability in these parameters. Relative measurements such as those typically employed in fNIRS lack the ability to provide information about the baseline state of the brain that result from baseline cerebral blood flow, blood volume, and metabolic rate of oxygen. Absolute NIRS measurements of the concentration and saturation of hemoglobin in brain tissue can fill this gap.

The propagation of NIR light in tissues depends on the spatio-temporal properties of two optical parameters, namely the absorption and the reduced scattering coefficients, according to the diffusion equation (DE). Therefore, absolute quantification of the concentration of hemoglobin chromophores (which are directly derived from the absorption coefficient) requires decoupling the contributions from these two optical coefficients to the optical measurements. This can be fulfilled by time-resolved approaches in the time-domain [36] or in the frequency-domain [37], even though specialized continuous-wave (CW) methods have also been reported [38], [39].

The importance of absolute measurements for in vivo interrogation of cerebrovascular health has been shown in animal models [35], [40], [41], [42]. Briefly, Saxena et al. performed in vivo measurements of absolute hemoglobin concentrations [i.e. oxygenated ([HbO2]), deoxygenated ([Hb]), and total ([HbT] = [HbO2]+[Hb])] and hemoglobin oxygen saturation (StO2 = [HbO2]/[HbT]) in mouse brain using a CW approach and reported a strong correlation between brain tumor size and [Hb]. Their longitudinal study that lasted 46 days suggested a hypoxic trend in the tumor region during baseline conditions provided by resting state absolute NIRS measurements [40]. Hallacoglu et al. measured the same parameters ([HbO2], [Hb], [HbT], and StO2) in the frequency-domain in the rat brain during resting state, and hypoxia and hypercapnia challenges. A high correlation of vascular cognitive impairment (VCI) with resting state values but not with relative changes induced by hypoxia or hypercapnia was reported. Absolute values measured in that study also suggested hypoxia-driven angiogenesis on VCI rat model on the basis of intra-individual differences that were measured 10 weeks apart [35]. In other frequency-domain approaches, Kurth et al. measured [HbO2], [Hb], [HbT] and StO2 in piglets, and proposed a threshold for the parameter StO2 to prevent cerebral hypoxia–ischemia and associated functional impairment [42]. Finally, Choe et al. demonstrated that absolute values of [HbO2] and [Hb] can be used to measure hypoxic stress in fetal sheep brain in utero and emphasized the potential implications of such measurements for the imaging of the human fetus [41]. Such diagnostic and/or long-term monitoring approaches presented above could only be made possible by quantification of concentration and saturation of hemoglobin in absolute terms.

In humans, absolute NIRS measurements found prominent applications in the longitudinal monitoring of infant brain development [43], [44] as well as in the detection of anomalies, for instance, led by brain injury [45] and respiratory distress [46]. Absolute brain oximetry (namely, measurement of StO2 in absolute terms) was predicted to become a way for clinicians to more quickly and noninvasively identify infants and children with altered levels of cerebral and/or somatic tissue oxygenation [47].

Despite the success in animal and infant human studies, translation of such methods to adult human brain remains a challenge. Relatively large extracerebral scalp-cortex distance in adult humans (10–23 mm) [48], [49], [50] in comparison to rats (∼1 mm) [35], piglets (∼4 mm) [51], or infants (5–11 mm) [52] creates a problem for traditional light propagation models (i.e. homogenous models) to analyze the NIRS data. While the assumption of tissue homogeneity in the probed volume may be adequate for small animal or infant human brain imaging [53], and may yield robust and reproducible measurements in the adult human brain [37], dependence of the optical coefficients to source detector separations have been reported in several studies [37], [53], [54], [55], [56], [57], indicating a non-negligible depth dependence of the tissue composition. To overcome the limitation of homogenous models, two-layered models have been developed in both the time domain [58] and the frequency domain [59], with experimental validations on tissue-mimicking layered phantoms [60], [61], [62], [63], [64], [65]. In vivo application of these approaches to noninvasive brain measurements of absolute optical coefficients have also been investigated in the frequency domain [41], [54] and the time domain [56], and differences with respect to homogenous assumption have been measured in experimental [41] and simulation based [56] data. Briefly, Gagnon et al. showed that the homogeneous model underestimated the absolute hemoglobin concentrations in the brain by about 30% (Fig. 1 in their paper) [56]. Choe et al. measured absolute changes in StO2 during hypoxia (i.e. baseline minus hypoxic state) using homogenous and two-layered models and reported that homogenous model underestimated this quantity by up to 90% (true values were determined using a hemoximeter) [41]. The advantages of using two-layer models over homogenous models in brain imaging have been demonstrated through the works discussed above; however widespread adoption of such approaches has not occurred due to limitations. These limitation include requirement of complementary MRI measurements [54], [56] or invasive means [41] to measure top layer thickness (cortex depth) and inability to measure first and second layer optical properties simultaneously.

thumbnail
Figure 1. Illustration of the two-layered liquid phantom (Del Bianco et al. 2004), where larger and smaller optical windows denote detector and source fiber positions, respectively [81].

https://doi.org/10.1371/journal.pone.0064095.g001

In this article, we report absolute measurements of the optical coefficients of two-layered media that expand on our previous work on multi-distance absolute measurements on tissue-like phantoms [66], animal brain [35], [51], and human brain [37]. We introduce an inversion procedure based on the solution of the frequency-domain diffusion equation for a two-layered medium to translate multi-distance optical data into measurements of the absolute values of the optical coefficients in both layers, and of the top layer thickness. When applied to optical measurements on the human head, our method yields hemoglobin concentrations and saturation in the extracerebral tissue (layer 1) and in the cerebral tissue (layer 2), as well as the cortical depth (scalp-cortex distance). Our method, which is based on our own implementation of Levenberg-Marquardt inversion procedure, has been tested by: 1) inverting simulated data generated by two-layered Monte Carlo simulations; 2) inverting experimental data collected on two-layered liquid phantoms; and 3) inverting experimental data collected on the forehead of human subjects in vivo. To the best of our knowledge, this is the first report of concurrent measurements of hemoglobin concentration in extracerebral and cerebral tissues combined with a measure of cortical depth using standalone NIRS.

Methods

Photon Migration in Turbid Media

Starting from the radiative transfer equation (RTE) in transport theory, which describes light propagation in random media, photon migration in highly scattering media (such as biological tissues) can be approximated by the diffusion equation (DE). Letting µa(r) and µs(r) represent the absorption and the reduced scattering coefficients, respectively, and D(r) = 1/(3 µs) the diffusion coefficient in a medium as a function of position vector r, the frequency-domain DE for the photon fluence rate due to an intensity modulated point source, is given by [67]:(1)where c is the speed of light in the media, is the Dirac delta function that represents a point-like photon source, and P is the frequency-dependent source power (photons/sec). Equation (1) reduces to the continuous wave (CW) diffusion equation by setting the angular modulation frequency (ω) to zero. Analytical solutions of the diffusion equation have been developed and reported in the literature in the time domain [58], [59], [68], [69], [70], frequency domain [59], [67], [68], and CW [69], [71] for homogenous [67], two-layered [58], [59], [69], [70] and N-layered [68] media. Other solutions of the DE for regularly bounded geometries have also been reported [72]. In this work, we have used solutions for both homogenous media (as given in Fishkin and Gratton, 1993) [67] and two-layered media (as given in Kienle et al., 1998 [59] for planar geometry, and Liemert and Kienle, 2010 [73] for cylindrical geometry), therefore we shall briefly summarize these solutions in the subsections below.

Diffusion model for homogenous media.

For a source term described by: in the frequency-domain (FD), where r and t are the spatial and temporal variables, respectively, the complex fluence rate at the point of observation in an unbounded (i.e. infinite) homogeneous medium is given by [67]:(2)where D = 1/(3 µs) is the diffusion coefficient, and r is the distance between the point source and the point of observation. Note that the second term of Eq. (2) is the solution of Eq. (1) for a homogeneous unbounded medium. Three components of the fluence rate can be extracted from Eq. (2), namely the steady-state, direct current (DC or CW) term, the alternating current (AC) amplitude, and the phase (PH) of the oscillatory term, given by:

(3)(4)(5)

In the infinite geometry, the following expressions are linearly dependent on r in the following form [74]:(6)(7)(8)

This linear dependence leads to an easy inversion procedure to evaluate µa and µs using, for instance, the slopes of the straight lines of ln(r AC) and phase ( and , respectively) as functions of r [67], [74]:(9)(10)

Diffusion model for two-layered media.

We have implemented two separate solutions of the DE for two-layered media: First, the solution using the Fourier transform formalism for the planar geometry (i.e. semi-infinite regime), given in [59], and second, the improved solution, which was developed in the cylindrical geometry, given in [73]. We used the latter solution [73] for the analysis presented in this article, because we found it to be computationally faster and more robust at large source-detector separations (r>30 mm) and high reduced scattering coefficients (µs >1 mm−1). We provide a brief description of this solution in the following.

For a point source that is incident onto the center of a layered cylindrical medium, the general solution of the two-layered DE in cylindrical coordinates is given by [73]:(11)where is the photon fluence in the kth layer of the medium (i.e. k = 1 or 2 for a two-layered medium), sn are the positive roots of the 0-order Bessel function of first kind divided by a′ = a+zb, (where a is the radius of the cylinder), Jm is the Bessel function of first kind and order m, zb = 2D1 (1+Reff)/(1−Reff), and Reff is the fraction of photons that are internally diffusely reflected at the cylinder boundary based on [75]. Here, G1 is defined in the following form:(12)where , is the distance between the isotropic point source and the point of observation (i.e. location of the detector), and z0 = 1/µ′s1. For the calculation of the diffusely reflected intensity (R), we have considered two expressions: the first, which uses only the flux at the boundary according to Fick’s Law [75]:(13)and the second, which uses a combination of fluence rate and flux, given by [76]:(14)where r represents the source-detector separation, and coefficients η1 and η2 are chosen based on the refractive index mismatch between first layer and surrounding medium [76]. We have used Eq. (13) (i.e. Fick’s law) for the calculation of R(r) throughout this study, as it showed better agreement with Monte Carlo simulations in this work. In the frequency domain, AC and PH terms are given by:

(15)(16)

The DC two-layered solution is obtained as a particular case of the FD solution when ω = 0. In this case, the imaginary part of the complex expression for reflectance reduces to zero. In the two-layered case (in comparison to the homogenous case) the inverse problem becomes much more complicated due to the increased number of unknowns. These unknowns are the absorption and reduced scattering coefficients of layer 1 (µa1, µs1) and layer 2 (µa2, µs2), the thickness of layer 1 (L), and a multiplicative AC factor (AF) and an additive phase term introduced by the measurement apparatus. The phase term can be neglected by considering the differences at different source-detector separations [for instance, PH(r2)–PH(r1)]. The solution of this problem requires nonlinear fitting procedures on the measured quantities (PH and either AC or DC as a function of r).

Inversion Procedure

To solve the inverse problem, we implemented a Levenberg-Marquard optimization routine in Matlab (Mathworks Inc, Natick, MA) based on [77] of the form:(17)

Here, x* is the result of the optimization, i.e. the best estimates of the unknown parameter vector x = [µa1, µs1, L, µa2, µs2, AF ]T, which minimizes the cost function:(18)where, ε is the error vector composed of the difference between theoretical values and experimental (or simulated) values as:

(19)Here, ACMod and PHMod are the theoretical values yielded from the two-layered model, ACExp and PHExp are the experimental values either from Monte Carlo simulations or measurements, and represents the Euclidean norm. In order to employ the Levenberg-Marquardt algorithm, the calculation of the Jacobian matrix J of the error vector with respect to the parameter vector is required:(20)

The solution is then obtained by updating x at each iteration as: , where h is obtained by solving the following linear system:(21)where ψ is a damping parameter affecting the size and direction of h and found via an appropriate line search algorithm [77]. Once the optimal fit parameters were determined, parameter statistics were computed for the converged solution using weight values equal to the mean square measurement error (namely, errors on AC and PH). The asymptotic standard parameter errors were computed by considering: , where is the covariance matrix, J is the Jacobian matrix, and W is the weighting matrix, which is constructed as a diagonal matrix with elements representing the inverse square of the measurement errors [78]. Here, diag represents a vector constructed with the diagonal elements of the covariance matrix. We considered the measurement errors to be 0.1° and 2% on each PH and AC datum, respectively, which are typical error values of the measurement apparatus used in this work [79].

Monte Carlo Simulations

Monte Carlo (MC) simulations were performed based on our previous work [80] for the validation of the forward model and testing of the inversion procedure. Briefly, each injected photon was “followed” in its path through the medium until it was either detected, or lost through the medium’s boundary. Moreover the photon was abandoned if its total pathlength (travelled in the medium) exceeded a fixed threshold (chosen differently for each source-detector distance). This corresponded to about three-four decades of the temporal point spread function (Green’s function in time domain). We note that the “late” photons, even if detected, would not contribute to the calculated parameters (DC, AC, and PH). The simulations were run until 10000 photons were detected in each receiver. This number guarantees an accuracy in the DC estimated values of 1% for non-absorbing media (as we chose for our simulations). Scaling relationships based on the microscopic Beer-Lambert law (mbll) were applied in order to scale the results to an arbitrary value of the absorption coefficient. Ten independent simulations were run in order to estimate the errors in the DC, AC and PH.

Simulations were run with refractive index pairs of 1 and 1.4 (for air and two-layered scattering media, respectively), and 1.56 and 1.33 [for surrounding medium (plastic scattering cell) and two-layered medium used in the phantom experiments (Intralipid 20%, Fresenius Kabi, Germany)], according to Mie theory with anisotropy coefficient g = 5.15×10−8 and a Henyey-Greenstein phase function with g = 0.8.

Experimental Approaches

Near infrared instrumentation.

All experiments were carried out using a commercial frequency-domain tissue oximeter (OxiplexTS, ISS Inc, Champaign, IL). Two laser diode light sources (emitting at wavelengths of 690 and 830 nm) were intensity modulated at a frequency of 110 MHz and their emitted light delivered to the medium of interest by means of optical fibers. The optical power at the end of the illumination optical fibers was ∼2 mW. On the detection end, another optical fiber was used to deliver the light to a photomultiplier tube, whose gain function was modulated at a slightly different frequency (110 MHz+5 kHz) than that of the intensity modulation of the light sources, giving rise to a beating signal of 5 kHz, which was sent to the computer for analysis. The built in Fast-Fourier-Transform algorithm computed the phase (relative to a reference signal) (PH), the average intensity (DC), and the amplitude of the intensity oscillations (AC) of the detected light. Further details of such frequency-domain system can be found in [79].

Two-layered phantom experiments.

We used a two-layered liquid phantom [81], where the layers were separated by means of a thin (23 µm) Mylar membrane. Effects of Mylar to diffuse reflectance measurements in the near-infrared region have been shown to be negligible [60]. An illustration of the phantom that was used in this work is provided in Figure 1. The phantom featured thirteen transparent optical windows at the center of its façade including one detection window (3 mm diameter) and twelve illumination windows (1 mm diameter each). Separation of the windows ranged from 6 to 50 mm. The overall dimensions of the phantom to ensure the validity of semi-infinite boundary conditions for layer 2 for a given thicknesses of layer 1 (in z-direction) were determined through Monte Carlo simulations. These dimensions are presented in Figure 1. The two layers were filled with controlled suspensions of deionized water, Intralipid 20% (Manufacturer Fresenius Kabi, Germany) and diluted India Ink (weight fraction of ink in water = 0.0074) for background dilution, scattering and absorption characteristics, respectively. Intralipid is an intravenous fat emulsion that has been widely used as a multiply scattering tissue-like phantom [82], [83], [84], [85], [86]. Scattering and absorption coefficients of the two layers were controlled by changing the amount of Intralipid and ink in each layer. We performed multi-distance frequency-domain diffuse reflectance measurements on twelve different two-layered phantom preparations for three different first layer thicknesses (8, 10, and 15 mm). Apart from 8 mm, such first layer thicknesses match true anatomical values of scalp-cortex distance in the adult human forehead (∼12–16 mm) [50]. Data from the two-layered phantoms were analyzed using the matching model and the six-parameter inversion routine described in the previous sections. Even though the data was collected between 6–50 mm source-detector distances, the range considered for the analysis was 14–50 mm (i.e. starting from the third optical window) to comply with the diffusion conditions.

Homogenous phantom experiments.

Homogenous phantom measurements were conducted in a large container in order to obtain an effectively infinite geometry arrangement, in which optical detector and source fibers (both 400 µm in diameter) were deeply immersed into the medium (at a depth of ∼60 mm) and parallel to each other. This is a typical experimental arrangement for absolute optical characterization of liquid phantoms using NIRS, where the diffused intensity is generated and detected only within the diffusing medium avoiding any complications due to the boundary conditions. We followed a common experimental protocol [74], where we employed an illumination fiber (400 µm diameter) delivering 830 nm wavelength of light and linearly scanned it away from the collection fiber by means of a programmable mechanical linear stage scanning system (Velmex Inc., Bloomfield, NY, USA) over the source–detector distances of ∼20 to 35 mm. This procedure was repeated twenty times for eight different homogenous phantom preparations. The acquisition was automated and controlled by a program written in LabVIEW software (National Instruments, Austin, TX, USA). By exploiting the linear dependence of µa and µs on the concentration of ink and Intralipid, respectively, we obtained: 1) intrinsic optical coefficients of pure Intralipid and ink; and 2) linear expressions that provided an independent means to characterize each medium that was used for the two-layered measurements.

Human subject experiments.

Participants were 3 healthy subjects (all males, mean±standard deviation age of 29±2 years). The protocol for the human subject measurements was approved by the Tufts University Institutional Review Board and written informed consent was obtained from all subjects prior to the measurements. We used an approach based on a preliminary calibration of the optical probe on a known phantom as described in [37]. The optical probe was made of polyurethane silicon and featured two detector optical fiber bundles (3 mm in core diameter) separated by 11 mm, and seven pairs of illumination optical fibers (0.4 mm in core diameter) that guided light at 690 and 830 nm, located at distances in the range 8–48 mm from the detector fiber bundles. Figure 2 illustrates the probe layout and placement on the subject’s forehead. Source fibers were located 8 to 38 mm from the first detector bundle and 18 to 48 mm from the second detector bundle in 5 mm increments. The optical probe was placed on the left side of the subjects’ forehead and held in place using a commercial sports band to exert light pressure for comfort, while guaranteeing good contact between the optical fibers and the subjects’ scalp. For the analysis, recordings from the two detector bundles were combined, providing an effective source-detector range of 8 to 48 mm. Here, the range considered for the analysis was 13–48 mm to comply with the diffusion conditions. Under the assumption that oxy-hemoglobin (HbO2), deoxy-hemoglobin (Hb) and water are the major absorbers in the probed tissue volume at the two wavelengths considered, we calculated tissue concentrations of HbO2 and Hb in each layer using the following expressions [37]:(25)(26)where εHb and εHbO2 are the molar extinction coefficients of Hb and HbO2 [87], respectively, µa (H2O) is the absorption coefficient of water [88], and CH2O-vf is the volume fraction of water content in the probed tissue volume for i = 1, 2 (i.e. extracerebral and cerebral tissue, respectively). Here, we assume a value of 0.7 (i.e. 70% water content) for CH2O-vf. [37], [56], [57] Finally, we calculated the total hemoglobin concentration and hemoglobin saturation in each layer using:

thumbnail
Figure 2. Schematic diagram of the experimental setup for the human subject measurements.

A) The polystyrene optical probe used in the in vivo measurements, and (B) its positioning on the subject’s forehead during measurements. Larger and smaller circles in (A) denote detector and source fibers positions, respectively.

https://doi.org/10.1371/journal.pone.0064095.g002

(27)(28)

Results

Results of the Monte Carlo Simulations

Recovered values from all eight Monte Carlo (MC) data sets were in good agreement with the actual values for all five parameters (µa1, µs1, L, µa2, µs2). We report these results in Figure 3 in terms of MC data set (from 1 through 8), where the error bars represent uncertainties associated with the fitting procedure. All five parameters for most MC sets were recovered within 10% of the actual values. In two of the simulations (namely, MC sets: 3 and 4), where the absorption coefficient of layer 1 (µa1 = 0.005 mm−1) was significantly different than that of layer 2 (µa2 = 0.02 mm−1 for No. 3; and µa2 = 0.025 mm−1 for No. 4), µa1 values were recovered with slightly larger errors (i.e. 12% for No. 3; and 14% for No. 4). However, deviations from the actual values for all other parameters in these simulations were still <10%. We note that the discrepancy between MC and DE would mostly account for these errors. To investigate this point, we have also carried out separate measurements on synthetic data generated using our diffusion based forward model with additive noise (typically 1% random noise). We found a more accurate recovery of the true values in these data sets, with deviations never exceeding a few percent and always smaller (as it is expected) than those found in the MC data (data not shown).

thumbnail
Figure 3. Six-parameter fitting procedure results on eight sets of two-layered Monte Carlo (MC) simulations reported in terms of MC set.

True values represent values used for µa (column 1), µs (column 2), and first layer thickness (column 3) in each MC simulation. Retrieved values for Layer 1 (row 1) and Layer 2 (row 2) are represented by open and filled circles, respectively. The error bars represent uncertainties determined by the fitting procedure.

https://doi.org/10.1371/journal.pone.0064095.g003

Our sensitivity analysis showed negligible effects of initial guesses for the parameter values in the inversion procedure. We demonstrate the method’s lack of sensitivity to initial guesses in Table 1, where two sets of initial guesses (i.e. “good” set and “bad” set) and retrieved values on the same representative data set is reported. As demonstrated in the Table 1, deviations between the retrieved and true values were less than 7% in both cases. This is a particularly important result demonstrating the method’s lack of sensitivity to local minima.

thumbnail
Table 1. Demonstration of the insensitivity of the inversion procedure to initialization.

https://doi.org/10.1371/journal.pone.0064095.t001

Results of the Phantom Measurements

Homogeneous phantom results.

In our infinite medium measurements on eight liquid homogenous phantoms, we found a linear dependence of µa and µs on India ink and Intralipid concentrations, respectively. These results are shown in Figure 4, whereby we report measured µa and µs at 830 nm as a function of volume fractions of diluted-ink and Intralipid. These are typical results for such experiments in the infinite geometry [74], [89], [90], which in fact confirm that the infinite boundary conditions were met during our measurements. We performed linear regression analysis on the experimental data and computed the slopes of µa and µs to be 2.1±0.1 (mm−1/%diluted-ink) and 19.2±1.2 (mm−1/% Intralipid), respectively. We note that these slopes represent the intrinsic µa and µs per unit concentration of diluted-ink and pure-intralipid, respectively, which were used for the accurate optical characterization of the suspensions used in the two-layered phantom measurements. We evaluated the intrinsic µa of pure-ink to be 281.8±13.6 (mm−1/% ink). This led to infinite medium measurement uncertainties of ∼6% for both µa and µs. The y-intercepts of diluted-ink and Intralipid regression lines (0.004±0.001 mm−1 and 0.03±0.04 mm−1, respectively), which represent our measured µa and µs of pure water (Figure 4). Although these values should be equal to the actual µa and µs of water at 830 nm wavelength (µa = ∼0.003 mm−1 [88] and µs = 0 mm−1, respectively), the actual values were within the measurement uncertainties.

thumbnail
Figure 4. Measurement of µa (left panel) and µ′s (right panel) at 830 nm wavelength on eight homogenous phantoms in the infinite geometry, presented as a function of volume fractions of ink dilution and intralipid, respectively.

The symbols and error bars represent mean and standard deviation of 20 independent measurements in each phantom. The slopes of the linear curves were used to characterize the precise quantities used in the two-layered phantom measurements.

https://doi.org/10.1371/journal.pone.0064095.g004

Two-layered phantom results.

In Figure S1 (supplementary material), we report the results of a representative two-layered phantom measurement including the resulting fit (solid lines) on measurements (circles) of AC (left panel) and phase (right panel) as a function of source-detector distance. Dotted lines illustrate the results yielded by the first iteration of the inversion procedure (using the initial guesses). In Figure S2, we present the evolution of the six fitted parameters [µa1, µ′s1, L, µa2, µ′s2, AF] for the same phantom considered in Figure S1 during all iterations involved in the fitting procedure, which demonstrate the robustness of the inversion procedure and once again its insensitivity to initial guesses. We are presenting results every other iteration for clarity. Horizontal solid and dashed lines represent the true values, measured independently in the infinite geometry using a homogenous model as described above in the homogenous phantom experiments subsection.

Results of the two-layered phantom measurements for all twelve phantoms are reported in Figure 5, where the error bars represent uncertainties associated with the fitting procedure. Considering the complexity of liquid phantom measurements, we have found excellent agreement between the true values and the retrieved values (i.e. measured from two-layered phantoms and retrieved simultaneously using our two-layered inversion routine). In fact, taking into account all two-layered phantom measurements [i.e. 12 phantoms ×5 key parameters (µa1, µ′s1, L, µa2, µ′s2,) = 60 parameters in total] more than 75% of the retrieved parameters were within 10% of the true values, half of which were within 5% of the true values. More specifically, the most accurately retrieved parameters in the two-layered phantom measurements were L and µa2, with ranges of absolute % deviations of 1%–8% and 3%–13%, respectively, from the true values. Subsequently, ranges of absolute % deviations in µa1, µ′s1, µ′s2 from their true values were 1%–14%, 3%–15%, and 5%–20%, respectively. Therefore, the least accurately retrieved parameter was µ′s2, where in four of the twelve phantoms we observed absolute % deviations within 15%–20% (Nos. 4, 7, 8 and 9) from the true values. The complete summary of the phantom experiments is reported in Table 2, where we also include retrieved values using a semi-infinite homogenous model for comparison. Considering the measurement uncertainties in the true values (∼6%) and in the retrieved values (as shown by the error bars in Fig. 5), as well as mixture calibration errors during the liquid phantom preparation (not quantified), the two-layered approach employed here was able to retrieve all parameters with good accuracy. We present the absolute % deviation in the measured parameters averaged across 12 phantoms in Figure 6, where the error bars represent standard errors of the % deviations. We underscore the fact that clinically most relevant parameters, L and µa2, which are representative of scalp-cortex distance and cerebral absorption coefficient, respectively, were recovered with high accuracy (absolute % error <7%).

thumbnail
Figure 5. Six-parameter fitting procedure results on twelve two-layered phantoms with a range of optical coefficients and top-layer thicknesses that are representative of the human brain.

True values represent quantities measured in the infinite medium geometry using a homogenous model. Retrieved values for layer 1 (row 1) and layer 2 (row 2) are represented by open and filled circles, respectively. The error bars represent uncertainties determined by the fitting procedure.

https://doi.org/10.1371/journal.pone.0064095.g005

thumbnail
Figure 6. Average of the absolute % errors in the measured parameters with respect to true values across twelve two-layered phantoms.

The error bars represent standard errors.

https://doi.org/10.1371/journal.pone.0064095.g006

Results of the Human Subject Measurements

We report retrieved values for µa1, µ′s1, L, µa2, µ′s2 for wavelengths 690 nm and 830 nm in Table 3. A first striking result is the reproducibility of L, which was measured independently using the two wavelengths (Table 3). The measured range for this parameter across all subjects (11.4–14.5 mm) are within anatomical values reported in the literature for the adult human forehead (∼12–16 mm) [50]. For a comparison with Table 3, in Table 4, we report optical coefficients that were retrieved using a semi-infinite homogenous model. In Tables 5 and 6, we report corresponding hemoglobin values. Other common characteristics across subjects include significantly higher µa2 values with respect to µa1 (Table 3), which corresponds to higher total hemoglobin concentrations in the cerebral tissues (82±14 µM) in comparison to superficial tissues (26±7 µM) (Table 6). We also found slightly higher hemoglobin saturation in cerebral tissues (62%±10%) in comparison to superficial tissues (56%±6%); however this difference was not significant (Table 6).

thumbnail
Table 3. Retrieved optical coefficients and first layer thickness on the forehead of three human subjects using the two-layered model.

https://doi.org/10.1371/journal.pone.0064095.t003

thumbnail
Table 4. Retrieved optical coefficients on the forehead of three human subjects using a homogenous model as a comparison with Table 3.

https://doi.org/10.1371/journal.pone.0064095.t004

thumbnail
Table 5. Retrieved concentrations of oxy-hemoglobin ([HbO2]) and deoxy-hemoglobin ([Hb]) on the forehead of three human subjects using two-layered (indicated by ‘Superficial’ and ‘Cerebral’) and homogeneous models.

https://doi.org/10.1371/journal.pone.0064095.t005

thumbnail
Table 6. Retrieved total hemoglobin concentration ([HbT]) and hemoglobin oxygen saturation (StO2) on the forehead of three human subjects using two-layered (indicated by ‘Superficial’ and ‘Cerebral’) and homogeneous models.

https://doi.org/10.1371/journal.pone.0064095.t006

Discussion

Monte Carlo Simulations

Monte Carlo methods offer a flexible and accurate approach toward simulating photon transport in turbid media. In this work, MC simulations allowed us to verify the validity of our implementation of the analytical solution of the DE for two-layered media. The optical coefficients and top layer thickness in all MC sets were recovered with good accuracy, with deviations in the order of a few percent in most cases, which are typical uncertainties found also in other studies [59], [63]. To test the performance of our inversion routine, we have also carried out separate measurements on synthetic data generated using our diffusion based forward model with additive noise (typically 1% random noise). We found accurate recovery of the true values in these data sets, with deviations never exceeding a few percent and always smaller (as it is expected) than those found in the MC data (data not shown).

Phantom Measurements

The two-layered phantom experiments presented here were of key importance in the validation of our approach, which featured an unconstrained inversion procedure of six unknown parameters with no prior knowledge. Previous experiments on two-layered phantoms, featuring two-layered models have been reported for the representation of extracerebral tissue overlaying cerebral tissue [61] (top layer thickness typically more than ∼10) and skin tissue overlaying fat/muscle tissue [60], [62], [63], [64], [65] (top layer thickness typically less than ∼6 mm). In just two of these studies optical coefficients of the two layers and the top layer thicknesses were measured simultaneously [60], [63]. In the frequency domain, Alexandrakis et al. used a simplex search algorithm and recovered the second layer optical coefficients with good accuracy (absolute % deviations ∼10%) in almost all phantoms [63]. However their measurement accuracy was lower for the first layer parameters. In the time domain, Martelli et al. used a Levenberg-Marquardt routine and reported excellent recovery of the first layer optical coefficients and second layer absorption coefficient (with absolute % deviations of less than 5%) [60]. For the first layer thickness and the second layer scattering coefficients their accuracy was lower (absolute % deviations of 11% and 33%, respectively). In the current study, using a frequency domain acquisition system and a Levenberg-Marquardt inversion routine we were able to measure all parameters with relatively high accuracy (with absolute % deviations of less than 12% for all parameters) (as seen in Figure 6). To our knowledge, such high accuracy measurements on two-layered phantoms in the frequency domain are unprecedented. Furthermore, the experimental tests (where the reference true values were obtained with diffusion theory in an infinite geometry and with source-detector separations >20 mm, thus under conditions well within the limits of applicability of diffusion theory) indicate the accuracy of our proposed diffusion-based method for the optical characterization of two-layered media. In comparison with the two-layered analysis, homogenous analysis performed: i) reasonably well in phantoms with 8 mm first layer thicknesses (Nos. 2 and 3); ii) worse in phantoms with 10 mm first layer thicknesses (Nos. 5 and 6); and iii) much worse in phantoms with 15 mm first layer thicknesses (Nos. 8,9,11 and 12) in retrieving the second layer absorption coefficient (µa2) (Table 2). We note that such results are indicative of the adequacy of homogeneous analysis for thin first layer thicknesses (here, <8 mm) in the source-detector distances that were considered (14–50 mm). Furthermore, the same comparison in the homogenous phantoms (Nos. 1, 4, 7 and 10) highlights the robustness of the two-layered analysis in homogeneous media.

Human Subject Measurements

Although we have not performed independent validation measurements (for example an MRI measurement of scalp-cortex distance), two arguments can be made in support of the reliability of the results of our method. The first argument is that the reduced scattering coefficient measured at 690 nm was larger than that of 830 nm for both layers and for all subjects, following the expected wavelength dependence relationship [91]. The second argument is the consistency of independently measured scalp-cortex distance at the two wavelengths used in all subjects (Table 3) and that all measured values are within anatomical ranges [50]. We emphasize the importance of such results given that the inversion procedure implemented by us was unconstrained, allowing the measured parameters to take any possible value during the inversion procedure. This point is demonstrated in Figure S2, where the first layer µaa1) varies from as high as 0.045 mm−1 (around 20th iteration) to as low as ∼0.005 mm−1 (around 50th iteration) in the course of the inversion procedure prior to converging to the correct value.

Concentration and saturation of hemoglobin measurements in all three subjects were within previously reported values in animal [35], [51] and human [37], [43], [56] brain. The first and the second layer measurement of some of these parameters (Tables 5 and 6) were significantly different. Such differences have also been observed by other researchers [54], [56]. The advantage of our approach over these earlier studies was the derivation of the hemoglobin parameters and the scalp-cortex distances without any a priori knowledge (these studies used MRI to measure the scalp-cortex distance). Interestingly, the total hemoglobin concentration measured in young subjects using a homogenous model in our earlier study (mean±standard deviation: 52±13 µM) [37] compares well with the current study (mean±standard deviation: 47±1 µM) when a homogenous model is used (Table 6). Moreover, these values are in between the first (30±7 µM) and the second layer (82±13 µM) values measured in the current study (Table 6). This is an indication that homogenous assumption was influenced by the extracerebral tissue physiology. Cerebral saturation in the earlier study (mean±standard deviation: 58%±13% ) [37] compares well with both the first (56%±10%) and the second (62%±6%) layer values, as well as the homogenous values (64%±8%) reported here.

Intersubject variability in baseline human head hemoglobin concentrations has been reported in several studies including those that used homogenous [37], [55] or two layered [54], [56] models. Although we have studied only a small group of subjects, in this study we have observed a similar intersubject variability in the baseline concentrations of hemoglobin, both in the extracerebral and cerebral tissue layers (Table 3). These results emphasize the fact that individualized characterization of the baseline optical coefficients of the human head is important and should be a part of routine practice in NIRS and fNIRS studies.

We point out that second layer reduced scattering coefficients measured in the subjects (0.3±0.1 mm−1 and 0.2±0.1 mm−1 at 690 nm and 830 nm, respectively) appear to be low, even though this result is consistent with some reported values in the literature (such as 0.3 mm−1 and 0.5 mm−1 as reported in Gagnon et al., (Fig. 3A) [56]). Low µ′s2 values such as these may have a physiological origin in that they may be representative of contributions from the clear cerebrospinal fluid (CSF) (featuring low scattering) to the layer 2 measurements. In this case, absorption coefficient in layer 2 (µa2) would also be underestimated due to the CSF (µa(CSF) <0.005 mm−1), which would further enhance the differences in hemoglobin concentrations between the two layers. Another possibility is that this parameter may have a low level of information content inherent to the two-layered diffuse reflectance measurements, which is an interpretation also shared by a similar study in the time domain [60]. In fact, µ′s2 was the parameter that we were least sensitive to in our phantom experiments as seen in the absolute % deviation plot (Figure 6).

Conclusions

We have presented a multi-distance frequency-domain NIRS approach that relies on a two-layered solution of the diffusion equation and an accompanying six-parameter inversion routine to simultaneously measure the absolute optical coefficients of two-layered turbid media. When applied to non-invasive NIRS measurements on the human head, our approach yields the concentration and oxygen saturation of hemoglobin in the extracerebral (layer 1) and cerebral (layer 2) tissue layers, as well as the scalp-cortex distance. This is the first frequency-domain study that presents high accuracy measurement of these parameters simultaneously using standalone NIRS. We have reported results of Monte Carlo (MC) simulations and two-layered phantom measurements. We have also reported measurements in human subjects to explore the in vivo applicability of our approach and found physiologically reasonable values for all measured parameters. Looking forward, further studies on a larger subject group would be required to gain confidence in the reproducibility of the absolute values in human subjects. Other potential in vivo experiments include studies on animal models featuring measurements under controlled physiological challenges, which would allow for a direct validation of our approach in vivo. Moreover, utilizing numerical forward models for a realistic human head geometry would be highly relevant (for instance, using NIRFAST software) [92] to test the performance of our approach in more realistic geometries.

This study has tackled one of the major issues faced by non-invasive optical measurements of the brain, namely the effect of superficial, extracerebral tissue layers. The capability of achieving depth discrimination is highly significant in functional brain studies and in absolute brain oximetry. Our results represent a step toward the goals of performing depth resolved NIRS and absolute measurements of the concentration and saturation of hemoglobin in cortical tissue.

Supporting Information

Figure S1.

A representative case illustrating the final fit (solid lines) on measurements (circles) of AC attenuation (left panel) and phase shift (right panel) as a function of source-detector distance from the two-layer phantom. Dotted lines illustrate the results yielded by the first iteration of the inversion procedure (using the initial guesses). Final fit was obtained at the 102nd iteration.

https://doi.org/10.1371/journal.pone.0064095.s001

(TIF)

Figure S2.

A representative case illustrating the evolution of the six parameters (µa1, µ′s1, L, µa2, µ′s2 and AF) during the fitting procedure on two-layered phantom data, demonstrating the robustness of the inversion procedure and its insensitivity to initial guesses. Solid and dashed lines represent true values, measured independently in the infinite geometry.

https://doi.org/10.1371/journal.pone.0064095.s002

(TIF)

Acknowledgments

We thank Dr. Alwin Kienle and Dr. Andrea Liemert for discussions over the solution of the diffusion equation for a two-layered cylinder. We thank Dr. Andrea Liemert for providing us his code in the cylindrical geometry to compare with our planar geometry implementation. The inversion procedures presented in this work employed his code for the forward calculation. We are thankful to Dr. Fabrizio Martelli for the fruitful discussions throughout this work and providing us the two-compartment phantom. We also acknowledge Dr. Eric Miller and Dr. Oguz Semerci for useful discussions on the inversion procedure.

Author Contributions

Conceived and designed the experiments: BH AS. Performed the experiments: BH. Analyzed the data: BH. Contributed reagents/materials/analysis tools: BH AS SF. Wrote the paper: BH AS SF. Conceived and designed the experiments: BH AS. Acquired and analyzed the data: BH. Interpreted the data: BH AS SF. Drafted, revised, and approved the article for publication: BH AS SF.

References

  1. 1. Wolf M, Ferrari M, Quaresima V (2007) Progress of near-infrared spectroscopy and topography for brain and muscle clinical applications. J Biomed Opt 12: 062104.
  2. 2. Sekhon MS, McLean N, Henderson WR, Chittock DR, Griesdale DE (2012) Association of hemoglobin concentration and mortality in critically ill patients with severe traumatic brain injury. Crit Care 16: R128.
  3. 3. Kellert L, Martin E, Sykora M, Bauer H, Gussmann P, et al. (2011) Cerebral oxygen transport failure?: decreasing hemoglobin and hematocrit levels after ischemic stroke predict poor outcome and mortality: STroke: RelevAnt Impact of hemoGlobin, Hematocrit and Transfusion (STRAIGHT)–an observational study. Stroke 42: 2832–2837.
  4. 4. Murkin JM, Adams SJ, Novick RJ, Quantz M, Bainbridge D, et al. (2007) Monitoring brain oxygen saturation during coronary bypass surgery: a randomized, prospective study. Anesth Analg 104: 51–58.
  5. 5. Heringlake M, Garbers C, Kabler JH, Anderson I, Heinze H, et al. (2011) Preoperative cerebral oxygen saturation and clinical outcomes in cardiac surgery. Anesthesiology 114: 58–69.
  6. 6. Hintz SR, Cheong WF, van Houten JP, Stevenson DK, Benaron DA (1999) Bedside imaging of intracranial hemorrhage in the neonate using light: comparison with ultrasound, computed tomography, and magnetic resonance imaging. Pediatr Res 45: 54–59.
  7. 7. Wong FY, Silas R, Hew S, Samarasinghe T, Walker AM (2012) Cerebral Oxygenation Is Highly Sensitive to Blood Pressure Variability in Sick Preterm Infants. PLoS ONE 7: e43165.
  8. 8. Parnia S, Nasir A, Shah C, Patel R, Mani A, et al. (2012) A feasibility study evaluating the role of cerebral oximetry in predicting return of spontaneous circulation in cardiac arrest. Resuscitation 83: 982–985.
  9. 9. Tweddell JS, Ghanayem NS, Hoffman GM (2010) Pro: NIRS is “standard of care” for postoperative management. Semin Thorac Cardiovasc Surg Pediatr Card Surg Annu 13: 44–50.
  10. 10. Davie SN, Grocott HP (2012) Impact of extracranial contamination on regional cerebral oxygen saturation: a comparison of three cerebral oximetry technologies. Anesthesiology 116: 834–840.
  11. 11. Hirsch JC, Charpie JR, Ohye RG, Gurney JG (2010) Near Infrared Spectroscopy (NIRS) Should Not Be Standard of Care for Postoperative Management. Seminars in Thoracic and Cardiovascular Surgery: Pediatric Cardiac Surgery Annual 13: 51–54.
  12. 12. Leff DR, Orihuela-Espina F, Elwell CE, Athanasiou T, Delpy DT, et al. (2011) Assessment of the cerebral cortex during motor task behaviours in adults: a systematic review of functional near infrared spectroscopy (fNIRS) studies. Neuroimage 54: 2922–2936.
  13. 13. Ferrari M, Quaresima V (2012) A brief review on the history of human functional near-infrared spectroscopy (fNIRS) development and fields of application. Neuroimage 63: 921–935.
  14. 14. Wolf M, Morren G, Haensse D, Karen T, Wolf U, et al. (2008) Near infrared spectroscopy to study the brain: an overview. Opto-Electronics Review 16: 413–419.
  15. 15. Boden S, Obrig H, Kohncke C, Benav H, Koch SP, et al. (2007) The oxygenation response to functional stimulation: is there a physiological meaning to the lag between parameters? Neuroimage 36: 100–107.
  16. 16. Franceschini MA, Fantini S, Thompson JH, Culver JP, Boas DA (2003) Hemodynamic evoked response of the sensorimotor cortex measured noninvasively with near-infrared optical imaging. Psychophysiology 40: 548–560.
  17. 17. Jasdzewski G, Strangman G, Wagner J, Kwong KK, Poldrack RA, et al. (2003) Differences in the hemodynamic response to event-related motor and visual paradigms as measured by near-infrared spectroscopy. Neuroimage 20: 479–488.
  18. 18. Yamada T, Umeyama S, Matsuda K (2012) Separation of fNIRS signals into functional and systemic components based on differences in hemodynamic modalities. PLoS ONE 7: e50271.
  19. 19. Saager RB, Telleri NL, Berger AJ (2011) Two-detector Corrected Near Infrared Spectroscopy (C-NIRS) detects hemodynamic activation responses more robustly than single-detector NIRS. Neuroimage 55: 1679–1685.
  20. 20. Gregg NM, White BR, Zeff BW, Berger AJ, Culver JP (2010) Brain specificity of diffuse optical imaging: improvements from superficial signal regression and tomography. Front Neuroenergetics 2.
  21. 21. Saager R, Berger A (2008) Measurement of layer-like hemodynamic trends in scalp and cortex: implications for physiological baseline suppression in functional near-infrared spectroscopy. J Biomed Opt 13: 034017.
  22. 22. Saager RB, Berger AJ (2005) Direct characterization and removal of interfering absorption trends in two-layer turbid media. J Opt Soc Am A Opt Image Sci Vis 22: 1874–1882.
  23. 23. Yamada T, Umeyama S, Matsuda K (2009) Multidistance probe arrangement to eliminate artifacts in functional near-infrared spectroscopy. J Biomed Opt 14: 064034.
  24. 24. Zhang Q, Brown EN, Strangman GE (2007) Adaptive filtering for global interference cancellation and real-time recovery of evoked brain activity: a Monte Carlo simulation study. J Biomed Opt 12: 044014.
  25. 25. Zhang Q, Strangman GE, Ganis G (2009) Adaptive filtering to reduce global interference in non-invasive NIRS measures of brain activation: how well and when does it work? Neuroimage 45: 788–794.
  26. 26. Franceschini MA, Joseph DK, Huppert TJ, Diamond SG, Boas DA (2006) Diffuse optical imaging of the whole head. J Biomed Opt 11: 054007.
  27. 27. Zhang Y, Brooks DH, Franceschini MA, Boas DA (2005) Eigenvector-based spatial filtering for reduction of physiological interference in diffuse optical imaging. J Biomed Opt 10: 11014.
  28. 28. Virtanen J, Noponen T, Merilainen P (2009) Comparison of principal and independent component analysis in removing extracerebral interference from near-infrared spectroscopy signals. J Biomed Opt 14: 054032.
  29. 29. Markham J, White BR, Zeff BW, Culver JP (2009) Blind identification of evoked human brain activity with independent component analysis of optical data. Hum Brain Mapp 30: 2382–2392.
  30. 30. Bluestone A, Abdoulaev G, Schmitz C, Barbour R, Hielscher A (2001) Three-dimensional optical tomography of hemodynamics in the human head. Opt Express 9: 272–286.
  31. 31. Habermehl C, Holtze S, Steinbrink J, Koch SP, Obrig H, et al. (2012) Somatosensory activation of two fingers can be discriminated with ultrahigh-density diffuse optical tomography. Neuroimage 59: 3201–3211.
  32. 32. Koch SP, Habermehl C, Mehnert J, Schmitz CH, Holtze S, et al. (2010) High-resolution optical functional mapping of the human somatosensory cortex. Front Neuroenergetics 2: 12.
  33. 33. Gibson A, Dehghani H (2009) Diffuse optical imaging. Philos Transact A Math Phys Eng Sci 367: 3055–3072.
  34. 34. Zeff BW, White BR, Dehghani H, Schlaggar BL, Culver JP (2007) Retinotopic mapping of adult human visual cortex with high-density diffuse optical tomography. Proc Natl Acad Sci U S A 104: 12169–12174.
  35. 35. Hallacoglu B, Sassaroli A, Fantini S, Troen AM (2011) Cerebral perfusion and oxygenation are impaired by folate deficiency in rat: absolute measurements with noninvasive near-infrared spectroscopy. J Cereb Blood Flow Metab 31: 1482–1492.
  36. 36. Torricelli A, Contini D, Spinelli L, Caffini M, Pifferi A, et al.. (2011) Advanced Optical Methods for Functional Brain Imaging: Time-Domain Functional Near-Infrared Spectroscopy. Advances in Optical Imaging for Clinical Medicine: John Wiley & Sons, Inc. 287–305.
  37. 37. Hallacoglu B, Sassaroli A, Wysocki M, Guerrero-Berroa E, Schnaider Beeri M, et al. (2012) Absolute measurement of cerebral optical coefficients, hemoglobin concentration and oxygen saturation in old and young adults with near-infrared spectroscopy. J Biomed Opt 17: 81406–81401.
  38. 38. Yeganeh HZ, Toronov V, Elliott JT, Diop M, Lee T-Y, et al. (2012) Broadband continuous-wave technique to measure baseline values and changes in the tissue chromophore concentrations. Biomed Opt Express 3: 2761–2770.
  39. 39. Grabtchak S, Whelan WM (2012) Separation of absorption and scattering properties of turbid media using relative spectrally resolved cw radiance measurements. Biomed Opt Express 3: 2371–2380.
  40. 40. Saxena V, Gonzalez-Gomez I, Laug WE (2007) A noninvasive multimodal technique to monitor brain tumor vascularization. Phys Med Biol 52: 5295–5308.
  41. 41. Choe R, Durduran T, Yu G, Nijland MJ, Chance B, et al. (2003) Transabdominal near infrared oximetry of hypoxic stress in fetal sheep brain in utero. Proc Natl Acad Sci U S A 100: 12950–12954.
  42. 42. Kurth CD, Levy WJ, McCann J (2002) Near-infrared spectroscopy cerebral oxygen saturation thresholds for hypoxia-ischemia in piglets. J Cereb Blood Flow Metab 22: 335–341.
  43. 43. Franceschini MA, Thaker S, Themelis G, Krishnamoorthy KK, Bortfeld H, et al. (2007) Assessment of infant brain development with frequency-domain near-infrared spectroscopy. Pediatr Res 61: 546–551.
  44. 44. Tina LG, Frigiola A, Abella R, Artale B, Puleo G, et al. (2009) Near Infrared Spectroscopy in healthy preterm and term newborns: correlation with gestational age and standard monitoring parameters. Curr Neurovasc Res 6: 148–154.
  45. 45. Grant PE, Roche-Labarbe N, Surova A, Themelis G, Selb J, et al. (2009) Increased cerebral blood volume and oxygen consumption in neonatal brain injury. J Cereb Blood Flow Metab 29: 1704–1713.
  46. 46. Lemmers PM, Toet M, van Schelven LJ, van Bel F (2006) Cerebral oxygenation and cerebral oxygen extraction in the preterm infant: the impact of respiratory distress syndrome. Exp Brain Res 173: 458–467.
  47. 47. Drayna PC, Abramo TJ, Estrada C (2011) Near-infrared spectroscopy in the critical setting. Pediatr Emerg Care 27: 432–439; quiz 440–432.
  48. 48. Okamoto M, Dan H, Sakamoto K, Takeo K, Shimizu K, et al. (2004) Three-dimensional probabilistic anatomical cranio-cerebral correlation via the international 10–20 system oriented for transcranial functional brain mapping. Neuroimage 21: 99–111.
  49. 49. Stokes MG, Chambers CD, Gould IC, Henderson TR, Janko NE, et al. (2005) Simple metric for scaling motor threshold based on scalp-cortex distance: application to studies using transcranial magnetic stimulation. J Neurophysiol 94: 4520–4527.
  50. 50. Haeussinger FB, Heinzel S, Hahn T, Schecklmann M, Ehlis AC, et al. (2011) Simulation of near-infrared light absorption considering individual head and prefrontal cortex anatomy: implications for optical neuroimaging. PLoS ONE 6: e26377.
  51. 51. Fantini S, Hueber D, Franceschini MA, Gratton E, Rosenfeld W, et al. (1999) Non-invasive optical monitoring of the newborn piglet brain using continuous-wave and frequency-domain spectroscopy. Phys Med Biol 44: 1543–1563.
  52. 52. Beauchamp MS, Beurlot MR, Fava E, Nath AR, Parikh NA, et al. (2011) The Developmental Trajectory of Brain-Scalp Distance from Birth through Childhood: Implications for Functional Neuroimaging. PLoS ONE 6: e24981.
  53. 53. Dehaes M, Grant PE, Sliva DD, Roche-Labarbe N, Pienaar R, et al. (2011) Assessment of the frequency-domain multi-distance method to evaluate the brain optical properties: Monte Carlo simulations from neonate to adult. Biomed Opt Express 2: 552–567.
  54. 54. Choi J, Wolf M, Toronov V, Wolf U, Polzonetti C, et al. (2004) Noninvasive determination of the optical properties of adult brain: near-infrared spectroscopy approach. J Biomed Opt 9: 221–229.
  55. 55. Comelli D, Bassi A, Pifferi A, Taroni P, Torricelli A, et al. (2007) In vivo time-resolved reflectance spectroscopy of the human forehead. Appl Opt 46: 1717–1725.
  56. 56. Gagnon L, Gauthier C, Hoge RD, Lesage F, Selb J, et al. (2008) Double-layer estimation of intra- and extracerebral hemoglobin concentration with a time-resolved system. J Biomed Opt 13: 054019.
  57. 57. Ohmae E, Ouchi Y, Oda M, Suzuki T, Nobesawa S, et al. (2006) Cerebral hemodynamics evaluation by near-infrared time-resolved spectroscopy: correlation with simultaneous positron emission tomography measurements. Neuroimage 29: 697–705.
  58. 58. Martelli F, Sassaroli A, Del Bianco S, Yamada Y, Zaccanti G (2003) Solution of the time-dependent diffusion equation for layered diffusive media by the eigenfunction method. Phys Rev E Stat Nonlin Soft Matter Phys 67: 056623.
  59. 59. Kienle A, Patterson MS, Dognitz N, Bays R, Wagninures G, et al. (1998) Noninvasive determination of the optical properties of two-layered turbid media. Appl Opt 37: 779–791.
  60. 60. Martelli F, Del Bianco S, Zaccanti G, Pifferi A, Torricelli A, et al. (2004) Phantom validation and in vivo application of an inversion procedure for retrieving the optical properties of diffusive layered media from time-resolved reflectance measurements. Opt Lett 29: 2037–2039.
  61. 61. Pucci O, Toronov V, St Lawrence K (2010) Measurement of the optical properties of a two-layer model of the human head using broadband near-infrared spectroscopy. Appl Opt 49: 6324–6332.
  62. 62. Li A, Kwong R, Cerussi A, Merritt S, Hayakawa C, et al. (2007) Method for recovering quantitative broadband diffuse optical spectra from layered media. Appl Opt 46: 4828–4833.
  63. 63. Alexandrakis G, Busch DR, Faris GW, Patterson MS (2001) Determination of the optical properties of two-layer turbid media by use of a frequency-domain hybrid monte carlo diffusion model. Appl Opt 40: 3810–3821.
  64. 64. Pham TH, Spott T, Svaasand LO, Tromberg BJ (2000) Quantifying the properties of two-layer turbid media with frequency-domain diffuse reflectance. Appl Opt 39: 4733–4745.
  65. 65. Ripoll J, Ntziachristos V, Culver JP, Pattanayak DN, Yodh AG, et al. (2001) Recovery of optical parameters in multiple-layered diffusive media: theory and experiments. J Opt Soc Am A Opt Image Sci Vis 18: 821–830.
  66. 66. Fantini S, Franceschini MA, Fishkin JB, Barbieri B, Gratton E (1994) Quantitative determination of the absorption spectra of chromophores in strongly scattering media: a light-emitting-diode based technique. Appl Opt 33: 5204–5213.
  67. 67. Fishkin JB, Gratton E (1993) Propagation of photon-density waves in strongly scattering media containing an absorbing semi-infinite plane bounded by a straight edge. J Opt Soc Am A 10: 127–140.
  68. 68. Liemert A, Kienle A (2010) Light diffusion in N-layered turbid media: frequency and time domains. J Biomed Opt 15: 025002.
  69. 69. Dayan I, Havlin S, Weiss GH (1992) Photon Migration in a Two-layer Turbid Medium a Diffusion Analysis. Journal of Modern Optics 39: 1567–1582.
  70. 70. Tualle JM, Prat J, Tinet E, Avrillier S (2000) Real-space Green’s function calculation for the solution of the diffusion equation in stratified turbid media. J Opt Soc Am A Opt Image Sci Vis 17: 2046–2055.
  71. 71. Liemert A, Kienle A (2010) Light diffusion in N-layered turbid media: steady-state domain. J Biomed Opt 15: 025003.
  72. 72. Arridge SR, Cope M, Delpy DT (1992) The theoretical basis for the determination of optical pathlengths in tissue: temporal and frequency analysis. Phys Med Biol 37: 1531–1560.
  73. 73. Liemert A, Kienle A (2010) Light diffusion in a turbid cylinder. II. Layered case. Opt Express 18: 9266–9279.
  74. 74. Fantini S, Franceschini MA, Gratton E (1994) Semi-infinite-geometry boundary problem for light migration in highly scattering media: a frequency-domain study in the diffusion approximation. J Opt Soc Am B 11: 2128–2138.
  75. 75. Haskell RC, Svaasand LO, Tsay TT, Feng TC, McAdams MS, et al. (1994) Boundary conditions for the diffusion equation in radiative transfer. J Opt Soc Am A Opt Image Sci Vis 11: 2727–2741.
  76. 76. Kienle A, Patterson MS (1997) Improved solutions of the steady-state and the time-resolved diffusion equations for reflectance from a semi-infinite turbid medium. J Opt Soc Am A Opt Image Sci Vis 14: 246–254.
  77. 77. Madsen K, Nielsen HB, Tingleff O (2004) Methods for non-linear least squares problems, Technical Report. Informatics and Mathematical Modelling Technical University of Denmark.
  78. 78. Press WH, Teukolsky SA, Vetterling WT, Flannery BP (1992) Numerical Recipes in FORTRAN 77, 2nd ed., The Art of Scientific Computing,. 2 ed. Cambridge, U.K.: Cambridge University Press.
  79. 79. Fantini S, Franceschini MA (2002) Frequency-domain techniques for tissue spectroscopy and imaging. In: Tuchin VV, editor. Handbook of Optical Biomedical Diagnostics. Bellingham, WA: SPIE Press.
  80. 80. Sassaroli A, Martelli F, Fantini S (2010) Perturbation theory for the diffusion equation by use of the moments of the generalized temporal point-spread function. III. Frequency-domain and time-domain results. J Opt Soc Am A Opt Image Sci Vis 27: 1723–1742.
  81. 81. Del Bianco S, Martelli F, Cignini F, Zaccanti G, Pifferi A, et al. (2004) Liquid phantom for investigating light propagation through layered diffusive media. Opt Express 12: 2102–2111.
  82. 82. Di Ninni P, Martelli F, Zaccanti G (2011) Intralipid: towards a diffusive reference standard for optical tissue phantoms. Phys Med Biol 56: N21–28.
  83. 83. Flock ST, Jacques SL, Wilson BC, Star WM, van Gemert MJ (1992) Optical properties of Intralipid: a phantom medium for light propagation studies. Lasers Surg Med 12: 510–519.
  84. 84. Bevilacqua F, Piguet D, Marquet P, Gross JD, Tromberg BJ, et al. (1999) In vivo local determination of tissue optical properties: applications to human brain. Appl Opt 38: 4939–4950.
  85. 85. van Staveren HJ, Moes CJ, van Marie J, Prahl SA, van Gemert MJ (1991) Light scattering in Intralipid-10% in the wavelength range of 400–1100 nm. Appl Opt 30: 4507–4514.
  86. 86. Di Ninni P, Berube-Lauziere Y, Mercatelli L, Sani E, Martelli F (2012) Fat emulsions as diffusive reference standards for tissue simulating phantoms? Appl Opt 51: 7176–7182.
  87. 87. Wray S, Cope M, Delpy DT, Wyatt JS, Reynolds EO (1988) Characterization of the near infrared absorption spectra of cytochrome aa3 and haemoglobin for the non-invasive monitoring of cerebral oxygenation. Biochim Biophys Acta 933: 184–192.
  88. 88. Hale GM, Querry MR (1973) Optical Constants of Water in the 200-nm to 200-microm Wavelength Region. Appl Opt 12: 555–563.
  89. 89. Martelli F, Zaccanti G (2007) Calibration of scattering and absorption properties of a liquid diffusive medium at NIR wavelengths. CW method. Opt Express 15: 486–500.
  90. 90. Spinelli L, Martelli F, Farina A, Pifferi A, Torricelli A, et al. (2007) Calibration of scattering and absorption properties of a liquid diffusive medium at NIR wavelengths. Time-resolved method. Opt Express 15: 6589–6604.
  91. 91. Fantini S, Sassaroli A (2012) Near-infrared optical mammography for breast cancer detection with intrinsic contrast. Ann Biomed Eng 40: 398–407.
  92. 92. Dehghani H, Eames ME, Yalavarthy PK, Davis SC, Srinivasan S, et al. (2008) Near infrared optical tomography using NIRFAST: Algorithm for numerical model and image reconstruction. Commun Numer Methods Eng 25: 711–732.