Peer Review History

Original SubmissionDecember 4, 2024
Decision Letter - Abbas Farmany, Editor

-->PONE-D-24-56032-->-->Interactive Graphical Visualization For Electrochemical Voltammetry Techniques of Differential Pulse Voltammetry and Cyclic Voltammetry-->-->PLOS ONE

Dear Dr. Sunindyo,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.-->--> -->-->Please submit your revised manuscript by Mar 05 2025 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.-->-->

Please include the following items when submitting your revised manuscript:-->

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.
  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.
  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Abbas Farmany

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1.Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. Please note that funding information should not appear in any section or other areas of your manuscript. We will only publish funding information present in the Funding Statement section of the online submission form. Please remove any funding-related text from the manuscript.

3. We note that the grant information you provided in the ‘Funding Information’ and ‘Financial Disclosure’ sections do not match.

When you resubmit, please ensure that you provide the correct grant numbers for the awards you received for your study in the ‘Funding Information’ section.

4. Thank you for stating the following financial disclosure:

“The funding is from P2MI 2024 Program of School of Electrical Engineering and Informatics Institut Teknologi Bandung”

Please state what role the funders took in the study.  If the funders had no role, please state: "The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript."

If this statement is not correct you must amend it as needed.

Please include this amended Role of Funder statement in your cover letter; we will change the online submission form on your behalf.

5. When completing the data availability statement of the submission form, you indicated that you will make your data available on acceptance. We strongly recommend all authors decide on a data sharing plan before acceptance, as the process can be lengthy and hold up publication timelines. Please note that, though access restrictions are acceptable now, your entire data will need to be made freely accessible if your manuscript is accepted for publication. This policy applies to all data except where public deposition would breach compliance with the protocol approved by your research ethics board. If you are unable to adhere to our open data policy, please kindly revise your statement to explain your reasoning and we will seek the editor's input on an exemption. Please be assured that, once you have provided your new statement, the assessment of your exemption will not hold up the peer review process.

6. We note that Figure 11 and 12 in your submission contain copyrighted images. All PLOS content is published under the Creative Commons Attribution License (CC BY 4.0), which means that the manuscript, images, and Supporting Information files will be freely available online, and any third party is permitted to access, download, copy, distribute, and use these materials in any way, even commercially, with proper attribution. For more information, see our copyright guidelines: http://journals.plos.org/plosone/s/licenses-and-copyright.

We require you to either (a) present written permission from the copyright holder to publish these figures specifically under the CC BY 4.0 license, or (b) remove the figures from your submission:

a. You may seek permission from the original copyright holder of Figure 11 and 12 to publish the content specifically under the CC BY 4.0 license.

We recommend that you contact the original copyright holder with the Content Permission Form (http://journals.plos.org/plosone/s/file?id=7c09/content-permission-form.pdf) and the following text:

“I request permission for the open-access journal PLOS ONE to publish XXX under the Creative Commons Attribution License (CCAL) CC BY 4.0 (http://creativecommons.org/licenses/by/4.0/). Please be aware that this license allows unrestricted use and distribution, even commercially, by third parties. Please reply and provide explicit written permission to publish XXX under a CC BY license and complete the attached form.”

Please upload the completed Content Permission Form or other proof of granted permissions as an "Other" file with your submission.

In the figure caption of the copyrighted figure, please include the following text: “Reprinted from [ref] under a CC BY license, with permission from [name of publisher], original copyright [original copyright year].”

b. If you are unable to obtain permission from the original copyright holder to publish these figures under the CC BY 4.0 license or if the copyright holder’s requirements are incompatible with the CC BY 4.0 license, please either i) remove the figure or ii) supply a replacement figure that complies with the CC BY 4.0 license. Please check copyright information on all replacement figures and update the figure caption with source information. If applicable, please specify in the figure caption text when a figure is similar but not identical to the original image and is therefore for illustrative purposes only.

7. We note you have included a table to which you do not refer in the text of your manuscript. Please ensure that you refer to Table 1-4 in your text; if accepted, production will need this reference to link the reader to the Table.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

-->Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. -->

Reviewer #1: Yes

Reviewer #2: Partly

**********

-->2. Has the statistical analysis been performed appropriately and rigorously? -->

Reviewer #1: Yes

Reviewer #2: N/A

**********

-->3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.-->

Reviewer #1: Yes

Reviewer #2: Yes

**********

-->4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.-->

Reviewer #1: Yes

Reviewer #2: Yes

**********

-->5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)-->

Reviewer #1: The manuscript presents a comprehensive study on the development of an interactive graphical visualization system for heart disease detection using electrochemical voltammetry techniques, specifically Differential Pulse Voltammetry (DPV) and Cyclic Voltammetry (CV). Overall, the paper is well organized. Therefore, I would like to recommend the publication of this work in PLOS ONE after a minor revision. The details are as follows:

Comment 1: Introduction: Suggest to a more detailed discussion of existing solutions and their limitations.

Comment 2: The description of the sensor reader and data analysis process is thorough, but more technical details on the algorithms used for data processing would be beneficial.

Comment 3: Raw data screenshots do not add much value or insight. Suggest removing these figures if no additional insight can be provided using them.

Comment 4: The interpretation of the CV data and correlation to heart disease detection is unclear. Please add more details.

Reviewer #2: Manuscript: PONE-D-24-56032 entitled “Interactive Graphical Visualization for Electrochemical Voltammetry Techniques of Differential Pulse Voltammetry and Cyclic Voltammetry”

This study provided a thorough understanding of the rapid identification of crucial biomarkers, such as cardiac troponin, using electrochemical voltammetry techniques. The electrochemical testing data revealed reliability and effectiveness in developing these biomarkers, as the author(s) claimed in this study. The authors tried to explain their works scientifically. However, the following quarries cleared before acceptance for publication of this manuscript in your esteemed journal.

1. In the abstract section, what new findings (results) are reported by the authors from this study? What novelty of study? It should elaborate.

2. The introduction section lacks sufficient references, with only a few cited. Are there no previous studies on this topic or related fields? Clarify the rationale, objectives, and novelty of this study.

3. In the Materials & Methods section, ….. oxidation reaction occurs on the auxiliary electrode…. Was the electrochemical sensor a three-electrode system or a two-electrode system? It should be a three-electrode system (as the authors mentioned, used auxiliary electrodes). If so, what was the reference electrode?

4. This section lacks clarity and requires revision with proper references. Shorten the description of CV and DPV techniques, which are well-defined.

5. In the Results & Discussion section, why did the author(s) not describe the results of the CV and DPV graphs? Additionally, are the data presented in the CV and DPV raw? This reviewer believes these electrochemical techniques offer quantitative information, which should not be as raw data.

6. Also, the discussion section is poor and should be elaborated with proper citations.

7. Needed to correct the some grammatical errors as indicated in revised manuscript.

**********

-->6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.-->

Reviewer #1: No

Reviewer #2: No

**********

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

Attachments
Attachment
Submitted filename: PONE-D-24-56032-reviewed.pdf
Revision 1

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1.Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

Response:

We have revised the manuscript to conform to the PLOS ONE style requirements, including formatting and file naming, as outlined in the provided templates.

2. Please note that funding information should not appear in any section or other areas of your manuscript. We will only publish funding information present in the Funding Statement section of the online submission form. Please remove any funding-related text from the manuscript.

Response:

Thank you for the clarification. We have removed the funding-related text from the manuscript as requested. The funding information is now provided solely in the Funding Statement section of the online submission form.

3. We note that the grant information you provided in the ‘Funding Information’ and ‘Financial Disclosure’ sections do not match.

When you resubmit, please ensure that you provide the correct grant numbers for the awards you received for your study in the ‘Funding Information’ section.

Response:

Thank you for bringing this to our attention. We have reviewed and corrected the grant information to ensure consistency between the 'Funding Information' and 'Financial Disclosure' sections. The correct grant number has now been entered in the ‘Funding Information’ section of the online submission form.

4. Thank you for stating the following financial disclosure:

“The funding is from P2MI 2024 Program of School of Electrical Engineering and Informatics Institut Teknologi Bandung”

Please state what role the funders took in the study. If the funders had no role, please state: "The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript."

If this statement is not correct you must amend it as needed.

Please include this amended Role of Funder statement in your cover letter; we will change the online submission form on your behalf.

Response:

Thank you for the clarification. We confirm that the funders had no involvement in the study. The following statement has been added to the cover letter as requested:

"The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript."

5. When completing the data availability statement of the submission form, you indicated that you will make your data available on acceptance. We strongly recommend all authors decide on a data sharing plan before acceptance, as the process can be lengthy and hold up publication timelines. Please note that, though access restrictions are acceptable now, your entire data will need to be made freely accessible if your manuscript is accepted for publication. This policy applies to all data except where public deposition would breach compliance with the protocol approved by your research ethics board. If you are unable to adhere to our open data policy, please kindly revise your statement to explain your reasoning and we will seek the editor's input on an exemption. Please be assured that, once you have provided your new statement, the assessment of your exemption will not hold up the peer review process.

Response:

Thank you for the important reminder. The link to the repository is https://drive.google.com/drive/folders/1XJ-pjEaOa-UFj499yzcPW7GTk--q6ilt?usp=sharing

We fully support PLOS ONE’s open data policy and will ensure compliance before final publication.

6. We note that Figure 11 and 12 in your submission contain copyrighted images. All PLOS content is published under the Creative Commons Attribution License (CC BY 4.0), which means that the manuscript, images, and Supporting Information files will be freely available online, and any third party is permitted to access, download, copy, distribute, and use these materials in any way, even commercially, with proper attribution. For more information, see our copyright guidelines: http://journals.plos.org/plosone/s/licenses-and-copyright.

We require you to either (a) present written permission from the copyright holder to publish these figures specifically under the CC BY 4.0 license, or (b) remove the figures from your submission:

a. You may seek permission from the original copyright holder of Figure 11 and 12 to publish the content specifically under the CC BY 4.0 license.

We recommend that you contact the original copyright holder with the Content Permission Form (http://journals.plos.org/plosone/s/file?id=7c09/content-permission-form.pdf) and the following text:

“I request permission for the open-access journal PLOS ONE to publish XXX under the Creative Commons Attribution License (CCAL) CC BY 4.0 (http://creativecommons.org/licenses/by/4.0/). Please be aware that this license allows unrestricted use and distribution, even commercially, by third parties. Please reply and provide explicit written permission to publish XXX under a CC BY license and complete the attached form.”

Please upload the completed Content Permission Form or other proof of granted permissions as an "Other" file with your submission.

In the figure caption of the copyrighted figure, please include the following text: “Reprinted from [ref] under a CC BY license, with permission from [name of publisher], original copyright [original copyright year].”

b. If you are unable to obtain permission from the original copyright holder to publish these figures under the CC BY 4.0 license or if the copyright holder’s requirements are incompatible with the CC BY 4.0 license, please either i) remove the figure or ii) supply a replacement figure that complies with the CC BY 4.0 license. Please check copyright information on all replacement figures and update the figure caption with source information. If applicable, please specify in the figure caption text when a figure is similar but not identical to the original image and is therefore for illustrative purposes only.

Response:

Thank you for bringing this to our attention. We have reviewed Figures 11 and 12 and acknowledge that they include copyrighted content. In response, we have taken the following action:

We have removed Figures 11 and 12 from the revised manuscript to comply with the CC BY 4.0 license requirements. These figures have been replaced with original illustrations that we created ourselves for the purpose of this publication.

7. We note you have included a table to which you do not refer in the text of your manuscript. Please ensure that you refer to Table 1-4 in your text; if accepted, production will need this reference to link the reader to the Table.

Response:

Thank you for pointing this out. We have revised the manuscript to ensure that Tables 1 through 4 are explicitly referenced in the main text at appropriate locations. These references now guide the reader to the relevant data and support the corresponding discussions.

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Partly

Response:

We appreciate the reviewers’ evaluations. In response to Reviewer #2's note, we have reviewed and revised relevant sections of the manuscript to strengthen the connection between the data and the conclusions. We have clarified the explanation of the algorithms used for data analysis, added further justification for using DPV over CV in the context of sensitivity, and elaborated on how the backend system ensures reliable prediction outcomes. These improvements are reflected particularly in the revised Results & Discussion section (Section 3), as well as the Conclusion.

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: N/A

Response:

We thank the reviewer for the positive assessment. We confirm that the data analysis procedures, including peak and baseline identification as well as threshold-based decision-making, were conducted rigorously following standard practices in electrochemical signal processing. No additional statistical testing was required beyond these signal analysis methods.

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g., participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

Response:

We appreciate the reviewers’ confirmation. We have ensured that all relevant data underlying our findings—including raw voltammetry measurement files, analysis outputs, and visualizations—are made fully available as part of the manuscript's supplementary information and have also been deposited in a public repository as noted in the Data Availability Statement.

The research data is available in https://drive.google.com/drive/folders/1XJ-pjEaOa-UFj499yzcPW7GTk--q6ilt?usp=sharing

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

Response:

We thank the reviewers for their positive feedback. We have carefully proofread the manuscript to ensure clarity, correctness, and consistency in standard English. Minor typographical and grammatical issues, where identified, have been corrected in this revision.

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: The manuscript presents a comprehensive study on the development of an interactive graphical visualization system for heart disease detection using electrochemical voltammetry techniques, specifically Differential Pulse Voltammetry (DPV) and Cyclic Voltammetry (CV). Overall, the paper is well organized. Therefore, I would like to recommend the publication of this work in PLOS ONE after a minor revision. The details are as follows:

Comment 1: Introduction: Suggest to a more detailed discussion of existing solutions and their limitations.

Response:

Thank you for the constructive suggestion. In response, we have expanded the Introduction section to include a more detailed discussion of existing solutions for heart disease detection using electrochemical techniques and their limitations. Specifically, we have highlighted the challenges in interpreting raw voltammetry data manually, the lack of user-friendly interfaces in current systems, and the limitations in real-time data accessibility and automated analysis. These revisions help to better contextualize the motivation and contribution of our proposed system.

Comment 2: The description of the sensor reader and data analysis process is thorough, but more technical details on the algorithms used for data processing would be beneficial.

Response:

Thank you for the insightful feedback. In response, we have added more technical details regarding the algorithms used for baseline correction and peak detection in Section 3.2 of the manuscript. Specifically, we elaborated on the use of polynomial fitting for cyclic voltammetry (CV) and asymmetric least squares (ALS) for differential pulse voltammetry (DPV), including the rationale for choosing these methods and how they handle signal noise and baseline drift. We also clarified how these algorithms integrate with the backend system to produce reliable current response measurements for diagnostic purposes.

Comment 3: Raw data screenshots do not add much value or insight. Suggest removing these figures if no additional insight can be provided using them.

Response:

We appreciate the reviewer’s observation. In response, we have removed the raw data screenshots (Figures 3 and 4) from the revised manuscript to maintain focus and clarity. Additionally, we have ensured that the key characteristics of the raw data are now clearly described in the text to preserve the necessary context without relying on those visual elements.

Comment 4: The interpretation of the CV data and correlation to heart disease detection is unclear. Please add more details.

Response:

Thank you for the valuable feedback. In the revised manuscript, we have clarified the role of cyclic voltammetry (CV) in the context of our system. Although DPV is the primary method used for heart disease detection due to its higher sensitivity, we included CV to support system functionality testing and sensor behaviour analysis. CV helps identify redox behaviour and validate the presence of electrochemical reactions, which supports calibration and comparison. However, the final diagnostic decision regarding heart disease is based on DPV data, as CV lacks sufficient resolution for reliable quantitative assessment of cardiac troponin levels. These clarifications have been added to Section 3.2.1 and the Discussion section.

Reviewer #2: Manuscript: PONE-D-24-56032 entitled “Interactive Graphical Visualization for Electrochemical Voltammetry Techniques of Differential Pulse Voltammetry and Cyclic Voltammetry”

This study provided a thorough understanding of the rapid identification of crucial biomarkers, such as cardiac troponin, using electrochemical voltammetry techniques. The electrochemical testing data revealed reliability and effectiveness in developing these biomarkers, as the author(s) claimed in this study. The authors tried to explain their works scientifically. However, the following quarries cleared before acceptance for publication of this manuscript in your esteemed journal.

1. In the abstract section, what new findings (results) are reported by the authors from this study? What novelty of study? It should elaborate.

Response:

Thank you for highlighting this important point. In response, we have revised the abstract to clearly emphasize the novelty and key findings of our study. Specifically, the novelty lies in the integration of an automated backend system using FastAPI and cloud-based storage that processes raw voltammetry data (CV and DPV) and converts it into diagnostic predictions via a user-friendly web interface. Additionally, we highlight the implementation of specific algorithms for baseline correction (polynomial fit for CV and ALS for DPV) and their role in accurate cardiac troponin detection.

The revised abstract now explicitly states:

“A key novelty of this research lies in the integration of automated baseline correction and peak detection algorithms—specifically polynomial fitting for cyclic voltammetry (CV) and asymmetric least squares (ALS) for differential pulse voltammetry (DPV)—within a user-friendly interface accessible even to non-expert users.”

These updates clarify both the scientific contribution and the practical innovation of the study.

2. The introduction section lacks sufficient references, w

Attachments
Attachment
Submitted filename: Response to Reviewers.docx
Decision Letter - Abbas Farmany, Editor

-->PONE-D-24-56032R1-->-->Interactive Graphical Visualization For Electrochemical Voltammetry Techniques of Differential Pulse Voltammetry and Cyclic Voltammetry-->-->PLOS ONE

Dear Dr. Sunindyo,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.-->-->

Please submit your revised manuscript by Sep 14 2025 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:-->

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.
  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.
  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Abbas Farmany

Academic Editor

PLOS ONE

Journal Requirements:

If the reviewer comments include a recommendation to cite specific previously published works, please review and evaluate these publications to determine whether they are relevant and should be cited. There is no requirement to cite these works unless the editor has indicated otherwise.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

-->Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.-->

Reviewer #2: All comments have been addressed

Reviewer #3: All comments have been addressed

**********

-->2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. -->

Reviewer #2: Yes

Reviewer #3: Partly

**********

-->3. Has the statistical analysis been performed appropriately and rigorously? -->

Reviewer #2: Yes

Reviewer #3: No

**********

-->4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.-->

Reviewer #2: Yes

Reviewer #3: Yes

**********

-->5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.-->

Reviewer #2: Yes

Reviewer #3: No

**********

-->6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)-->

Reviewer #2: I read the revised manuscript and found that it addresses previous comments well and incorporates the revisions. I, however, am not sure whether the style of citing the reference in the text is ..... [1]. or ...... . [1]. Both styles are used in the manuscript; it should be corrected.

Reviewer #3: English of manuscript is below the required standards and needs a careful polish.

Section Potentiostat and voltammetry needs a careful revision. Why SPCE electrode is used? Ag/AgCl paste!!!

what is the reference electrode? non-useful parts eg. definitions needs to be removed.

the title of Anshori paper in page 5 must be deleted. what is the baseline correction system? Conclusion section must be re-prepared as more informative. Statistical analysis section or methods?

**********

-->7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.-->

Reviewer #2: Yes: Jagadeesh Bhattarai

Reviewer #3: No

**********

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

Revision 2

Response to Reviewers

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #2: All comments have been addressed

Reviewer #3: All comments have been addressed

Response

We sincerely thank Reviewer #2 and Reviewer #3 for confirming that all comments have been adequately addressed. In this revision, we have additionally improved consistency in citation style and refined the manuscript for clarity and readability as part of editorial polishing.

________________________________________

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #2: Yes

Reviewer #3: Partly

Response

We thank the reviewers for their assessment. In response to Reviewer #3’s comment, we have further clarified the methodological description and strengthened the linkage between the presented data and the conclusions. Specifically, we refined the explanation of the experimental setup, data processing workflow, and interpretation of the CV and DPV results, emphasizing that the conclusions are directly derived from the analyzed data. These revisions improve clarity and technical transparency without altering the underlying results or conclusions of the study.

________________________________________

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #2: Yes

Reviewer #3: No

Response

We thank the reviewers for their evaluation. In this study, the data analysis is based on established electrochemical signal processing methods rather than inferential statistical testing. We have clarified this distinction in the revised manuscript by explicitly describing the analytical procedures used for baseline correction, peak detection, and signal interpretation for CV and DPV data. These methods were applied consistently across all measurements and provide an appropriate and rigorous basis for the conclusions drawn

________________________________________

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #2: Yes

Reviewer #3: Yes

Response

We thank the reviewers for their confirmation. All data underlying the findings described in the manuscript, including raw voltammetry measurement files and processed analysis outputs, are fully available through the repository indicated in the Data Availability Statement.

________________________________________

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #2: Yes

Reviewer #3: No

Response

We thank the reviewers for their feedback. In response to Reviewer #3’s comment, we have carefully revised and polished the manuscript to improve clarity, consistency, and readability in standard scientific English. Typographical and grammatical errors have been corrected, and several sentences have been rephrased to enhance precision and flow, without changing the scientific content of the manuscript.

________________________________________

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #2: I read the revised manuscript and found that it addresses previous comments well and incorporates the revisions. I, however, am not sure whether the style of citing the reference in the text is ..... [1]. or ...... . [1]. Both styles are used in the manuscript; it should be corrected.

Reviewer #3: English of manuscript is below the required standards and needs a careful polish.

Section Potentiostat and voltammetry needs a careful revision. Why SPCE electrode is used? Ag/AgCl paste!!!

what is the reference electrode? non-useful parts eg. definitions needs to be removed.

the title of Anshori paper in page 5 must be deleted. what is the baseline correction system? Conclusion section must be re-prepared as more informative. Statistical analysis section or methods?

Response

We thank the reviewers for their constructive comments.

In response to Reviewer #2, we have standardized the in-text citation format throughout the manuscript to ensure consistency with the journal’s reference style.

In response to Reviewer #3, we have carefully polished the English language to meet the journal’s standards. The Potentiostat and Voltammetry section has been revised to improve clarity, explicitly justify the use of the SPCE, and clearly describe the reference electrode configuration. Non-essential definitions have been removed, and the title of the cited work by Anshori has been deleted from the main text. We have also clarified the baseline correction workflow, revised the Conclusion section to be more informative, and explicitly positioned the analytical procedures within the Methods section to address concerns regarding statistical analysis. These revisions improve clarity and presentation without altering the scientific content of the manuscript.

________________________________________

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #2: Yes: Jagadeesh Bhattarai

Reviewer #3: No

Response

We acknowledge the reviewers’ choices regarding the publication of peer review history and comply with the journal’s policy on peer review transparency and anonymity.

Attachments
Attachment
Submitted filename: Response_to_Reviewers_auresp_2.docx
Decision Letter - Abbas Farmany, Editor

-->PONE-D-24-56032R2-->-->Interactive Graphical Visualization For Electrochemical Voltammetry Techniques of Differential Pulse Voltammetry and Cyclic Voltammetry-->-->PLOS One

Dear Dr. Sunindyo,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please submit your revised manuscript by Mar 22 2026 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:-->

  • A letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.
  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.
  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

-->If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Abbas Farmany

Academic Editor

PLOS One

Journal Requirements:

1. If the reviewer comments include a recommendation to cite specific previously published works, please review and evaluate these publications to determine whether they are relevant and should be cited. There is no requirement to cite these works unless the editor has indicated otherwise.

2. Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. If you need to cite a retracted article, indicate the article’s retracted status in the References list and also include a citation and full reference for the retraction notice.

Additional Editor Comments:

- Title of manuscript must be informative and reflect the content of the work closely.

- English of manuscript needs to be improved carefully. The abstract must be prepared as more informative. For example, "The potential for heart disease can be detected using a sensor reader that utilizes electrochemical voltammetry techniques, enabling the rapid identification of critical biomarkers such as cardiac troponin." must be improved as "The risk of heart disease can be identified through a sensor that employs electrochemical voltammetry, a technique that allows for the rapid and precise detection of critical biomarkers, including cardiac troponin." etc...

Please Improve the text carefully and submit the revised version.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

-->Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.-->

Reviewer #3: All comments have been addressed

**********

-->2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. -->

Reviewer #3: Yes

**********

-->3. Has the statistical analysis been performed appropriately and rigorously? -->

Reviewer #3: I Don't Know

**********

-->4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.-->

Reviewer #3: Yes

**********

-->5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.-->

Reviewer #3: Yes

**********

-->6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)-->

Reviewer #3: (No Response)

**********

-->7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.-->

Reviewer #3: No

**********

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

To ensure your figures meet our technical requirements, please review our figure guidelines: https://journals.plos.org/plosone/s/figures

You may also use PLOS’s free figure tool, NAAS, to help you prepare publication quality figures: https://journals.plos.org/plosone/s/figures#loc-tools-for-figure-preparation.

NAAS will assess whether your figures meet our technical requirements by comparing each figure against our figure specifications.

-->

Revision 3

Response to Reviewers

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #2: All comments have been addressed

Reviewer #3: All comments have been addressed

Response

We sincerely thank Reviewer #2 and Reviewer #3 for confirming that all comments have been adequately addressed. In this revision, we have additionally improved consistency in citation style and refined the manuscript for clarity and readability as part of editorial polishing.

________________________________________

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #2: Yes

Reviewer #3: Partly

Response

We thank the reviewers for their assessment. In response to Reviewer #3’s comment, we have further clarified the methodological description and strengthened the linkage between the presented data and the conclusions. Specifically, we refined the explanation of the experimental setup, data processing workflow, and interpretation of the CV and DPV results, emphasizing that the conclusions are directly derived from the analyzed data. These revisions improve clarity and technical transparency without altering the underlying results or conclusions of the study.

________________________________________

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #2: Yes

Reviewer #3: No

Response

We thank the reviewers for their evaluation. In this study, the data analysis is based on established electrochemical signal processing methods rather than inferential statistical testing. We have clarified this distinction in the revised manuscript by explicitly describing the analytical procedures used for baseline correction, peak detection, and signal interpretation for CV and DPV data. These methods were applied consistently across all measurements and provide an appropriate and rigorous basis for the conclusions drawn

________________________________________

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #2: Yes

Reviewer #3: Yes

Response

We thank the reviewers for their confirmation. All data underlying the findings described in the manuscript, including raw voltammetry measurement files and processed analysis outputs, are fully available through the repository indicated in the Data Availability Statement.

________________________________________

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #2: Yes

Reviewer #3: No

Response

We thank the reviewers for their feedback. In response to Reviewer #3’s comment, we have carefully revised and polished the manuscript to improve clarity, consistency, and readability in standard scientific English. Typographical and grammatical errors have been corrected, and several sentences have been rephrased to enhance precision and flow, without changing the scientific content of the manuscript.

________________________________________

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #2: I read the revised manuscript and found that it addresses previous comments well and incorporates the revisions. I, however, am not sure whether the style of citing the reference in the text is ..... [1]. or ...... . [1]. Both styles are used in the manuscript; it should be corrected.

Reviewer #3: English of manuscript is below the required standards and needs a careful polish.

Section Potentiostat and voltammetry needs a careful revision. Why SPCE electrode is used? Ag/AgCl paste!!!

what is the reference electrode? non-useful parts eg. definitions needs to be removed.

the title of Anshori paper in page 5 must be deleted. what is the baseline correction system? Conclusion section must be re-prepared as more informative. Statistical analysis section or methods?

Response

We thank the reviewers for their constructive comments.

In response to Reviewer #2, we have standardized the in-text citation format throughout the manuscript to ensure consistency with the journal’s reference style.

In response to Reviewer #3, we have carefully polished the English language to meet the journal’s standards. The Potentiostat and Voltammetry section has been revised to improve clarity, explicitly justify the use of the SPCE, and clearly describe the reference electrode configuration. Non-essential definitions have been removed, and the title of the cited work by Anshori has been deleted from the main text. We have also clarified the baseline correction workflow, revised the Conclusion section to be more informative, and explicitly positioned the analytical procedures within the Methods section to address concerns regarding statistical analysis. These revisions improve clarity and presentation without altering the scientific content of the manuscript.

________________________________________

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #2: Yes: Jagadeesh Bhattarai

Reviewer #3: No

Response

We acknowledge the reviewers’ choices regarding the publication of peer review history and comply with the journal’s policy on peer review transparency and anonymity.

Attachments
Attachment
Submitted filename: Response_to_Reviewers_auresp_3.docx
Decision Letter - Abbas Farmany, Editor

Algorithm-Assisted Interpretation of Cyclic and Differential Pulse Voltammetry for Cardiac Troponin Detection

PONE-D-24-56032R3

Dear Dr. Sunindyo,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice will be generated when your article is formally accepted. Please note, if your institution has a publishing partnership with PLOS and your article meets the relevant criteria, all or part of your publication costs will be covered. Please make sure your user information is up-to-date by logging into Editorial Manager at Editorial Manager® and clicking the ‘Update My Information' link at the top of the page. For questions related to billing, please contact billing support.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Abbas Farmany

Academic Editor

PLOS One

Additional Editor Comments (optional):

Reviewers' comments:

Formally Accepted
Acceptance Letter - Abbas Farmany, Editor

PONE-D-24-56032R3

PLOS One

Dear Dr. Sunindyo,

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS One. Congratulations! Your manuscript is now being handed over to our production team.

At this stage, our production department will prepare your paper for publication. This includes ensuring the following:

* All references, tables, and figures are properly cited

* All relevant supporting information is included in the manuscript submission,

* There are no issues that prevent the paper from being properly typeset

You will receive further instructions from the production team, including instructions on how to review your proof when it is ready. Please keep in mind that we are working through a large volume of accepted articles, so please give us a few days to review your paper and let you know the next and final steps.

Lastly, if your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

You will receive an invoice from PLOS for your publication fee after your manuscript has reached the completed accept phase. If you receive an email requesting payment before acceptance or for any other service, this may be a phishing scheme. Learn how to identify phishing emails and protect your accounts at https://explore.plos.org/phishing.

If we can help with anything else, please email us at customercare@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Abbas Farmany

Academic Editor

PLOS One