Peer Review History
| Original SubmissionFebruary 26, 2025 |
|---|
|
Dear Dr. Baliruno, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. Please submit your revised manuscript by Jun 09 2025 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.
If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter. If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at . Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at . Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at . Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.... We look forward to receiving your revised manuscript. Kind regards, Seth Agyei Domfeh, PhD Academic Editor PLOS ONE Journal requirements: 1. When submitting your revision, we need you to address these additional requirements. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and 2. In the online submission form, you indicated that [The data underlying the results presented in the study are available from the corresponding author]. All PLOS journals now require all data underlying the findings described in their manuscript to be freely available to other researchers, either 1. In a public repository, 2. Within the manuscript itself, or 3. Uploaded as supplementary information. This policy applies to all data except where public deposition would breach compliance with the protocol approved by your research ethics board. If your data cannot be made publicly available for ethical or legal reasons (e.g., public availability would compromise patient privacy), please explain your reasons on resubmission and your exemption request will be escalated for approval. Reviewers' comments: Reviewer's Responses to Questions Comments to the Author 1. Is the manuscript technically sound, and do the data support the conclusions? Reviewer #1: Yes Reviewer #2: No ********** 2. Has the statistical analysis been performed appropriately and rigorously? -->?> Reviewer #1: Yes Reviewer #2: No ********** 3. Have the authors made all data underlying the findings in their manuscript fully available??> The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.--> Reviewer #1: Yes Reviewer #2: Yes ********** 4. Is the manuscript presented in an intelligible fashion and written in standard English??> Reviewer #1: Yes Reviewer #2: Yes ********** Reviewer #1: Thank you for the effort on this well written paper addressing a very important concern on limited screening tests for hepatitis in a population which has substantive exposure. I have no concern with regards to dual publication, research ethics, or publication ethics. Please see below for few comments which you can take into consideration as you work towards a final version: In the abstract result section and the whole of the paper there is a result on sensitivity of 93% with a CI of 93 - 93%. Please confirm this reporting version. The cheaper RDTs here written as 'CE marked' compared with 'CE Marked' elsewhere in the document. Please consider for consistency Introduction Line 40 please consider deleting the word 'above' Line 51 - 53 is this referring to only the Iranian study? Why is this part separated from the paragraph below which is addressing results from other studies? Line 60 please consider a new paragraph Line 68 - 70 please clarify if this is to detect acute infection Methods and Materials Is 200 calculation for both positives and negatives? or for each? Is there any concern with multiple operators in the testing? Results In the flow chart how many participants were not included due to heamolysis for example or the flow chart starts at the ones selected for the main batch of samples? The flow chart itself may require redoing so that it is clearer and without any overlaps In the tables it is noted that one decimal place and 2 decimals places are used Appendix consider pictures of one each of the tests for clarity and submit as supplement. Add a description of the attachments Reviewer #2: This diagnostic performance evaluation tested four HBsAg CE marked and one WHO PQ RDTs in a blood transfusion service in Uganda. The authors evaluated 200 residual blood donor samples, with a 1:1 ratio of known positive to negative samples. Major comments: 1. Abstract. Please state the reference ELISA and RDT manufacturers. If additional space is required, the reference to the stats program could be removed. 2. Mention in the abstract the prevalence in population (eg. 50% 1:1 case:control matching) Introduction 3. The authors should provide a reference for this statement: "Currently, the global mortality from viral 31 hepatitis exceeds that of HIV, TB or malaria, and is likely to exceed the toll from those three 32 diseases combined by 2040." 4. The epidemiological estimates should be updated using the WHO global hepatitis report 2024. 5. "Several immunological methods are available to detect HBsAg, including Enzyme Immuno Assays (EIA), Radio Immuno Assays (RIA), Immuno Chromatographic Assays (ICA), and Haemo Agglutination Assays (HAA)." Both RIA and HAA are archaic methods and should not be refered to. Chemiluminescent immunoassays and EIAs are the leading methods. Immuno Chromatographic assays are commonly referred to as lateral flow or rapid diagnostic tests. 6. "In Iran, Acon, Atlas, Intec, Blue Cross, Dima and Cortez had sensitivities ranging from 97.5% -99.0 % and specificities of 97.5%- 99.2% (8). Varied performance characteristics were however recorded in East and West African populations." This study in Iran should not be referred to in the introduction and is not clearly relevant to the current study. There are more relevant data from Africa. The statement about data from East and West African populations is more relevant but belongs as an introduction to the following paragraph. I would recommend referring to a meta-analysis or review of HBsAg RDT performance. 7. When mentioning test kits, please refer to the manufacturer and city/country of origin, not just the brand name (there are many similar brands of kits). 8. "The WHO 62 prequalified diagnostic kits are recommended internationally, however, standards for rapid 63 diagnostic test prequalification by this organization are quite stringent, making them expensive. 64 In Sub-Saharan Africa, developing countries such as Uganda, use standards for validating 65 diagnostic tools set by international regulatory bodies. These bodies include: the United States 66 Food and Drug Administration and the Conformité Européenne (CE) that are less costly as a 67 precondition for considering a diagnostic test" I am not aware of any evidence that FDA or CE evaluation is less rigorous than the WHO PQ (or less costly). The reference provided is "ASLM. Lab culture, The ASLM Magazine. 2015;", which does not refer to a specific article (no URL or page number) or provide evidence for this assertion. 8. "However, for acute hepatitis B, RDTs may not be suitable since they can provide a false sense of security in case of negativity (14)." This is not the case, HBsAg is positive during acute hepatitis, the reference 14 refers to national testing guidelines and not any empirical evidence to back this assertion. 9. "We therefore, aimed to calculate the sensitivity, specificity, positive and negative predictive values of 4 brands of RDTs in Uganda." I believe you tested 5 brands? 10. The sample size formula refers to a national prevalence of 4.1% but this has not been used to calculate the sample size, instead a prevalence of 50% has been used in keeping with the study design. 11. For the Fortress diagnostic ELISA reference test, please state any available data on diagnostic/analytical sensitivity, validation data, and stringent regulatory approvals for this assay. 12. The details on methods for ELISA are not necessary, they can be moved to an appendix and just state according to manufacturers instructions. 13. Was the result of the ELISA test available to the operator doing the RDT assessments? If so, state this and list as a limitation (unblinded and not independent assessment weakens the study quality). 14. The method to retest multiple times in case of discordance with the ELISA result has resulted in an artificially (improved) assessment of test performance in relation to real conditions. This is a concern, this renders the resulting performance characteristics essentially invalid. "Retesting was done where results were discordant, this was done to improve the accuracy of the study results" This is inappropriate. If available, present the results before retesting multiple times ie. the first test result. 15. The flow chart is not interpretable, the lines are crossed and do not flow properly. Since the flowchart does not show any study exclusions, it is better summarised in a table (as shown in table 3). 16. Stratify table 1 by HBsAg test result, as well as showing overall characteristics. 17. A major consideration is that by taking a 1:1 case:control ratio, the sample prevalnece is 50% vs 4% nationally. Therefore all predictive values presented (PPV and NPV) are invalid, they are wildly different from a scenario when the prevalence is 4%. I would not present any predictive values. If these are to be shown, it should be shown as simulation data, weighting for a hypothetical scenario with sample prevalence of 4% (and where test performance in positive and negative samples is applied to the hypothetical scenario). ********** what does this mean?). If published, this will include your full peer review and any attached files.). If published, this will include your full peer review and any attached files.). If published, this will include your full peer review and any attached files.). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our For information about this choice, including consent withdrawal, please see our For information about this choice, including consent withdrawal, please see our For information about this choice, including consent withdrawal, please see our Privacy Policy..--> Reviewer #1: No Reviewer #2: No ********** [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at . PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at . PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at . PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.. Please note that Supporting Information files do not need this step.. Please note that Supporting Information files do not need this step.. Please note that Supporting Information files do not need this step. |
| Revision 1 |
|
Dear Dr. Baliruno, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. Please submit your revised manuscript by Nov 23 2025 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.
If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter. If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at . Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at . Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at . Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.... We look forward to receiving your revised manuscript. Kind regards, Seth Agyei Domfeh, PhD Academic Editor PLOS ONE Journal Requirements: If the reviewer comments include a recommendation to cite specific previously published works, please review and evaluate these publications to determine whether they are relevant and should be cited. There is no requirement to cite these works unless the editor has indicated otherwise. [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author Reviewer #2: (No Response) ********** 2. Is the manuscript technically sound, and do the data support the conclusions??> Reviewer #2: No ********** 3. Has the statistical analysis been performed appropriately and rigorously? -->?> Reviewer #2: No ********** 4. Have the authors made all data underlying the findings in their manuscript fully available??> The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.--> Reviewer #2: Yes ********** 5. Is the manuscript presented in an intelligible fashion and written in standard English??> Reviewer #2: Yes ********** Reviewer #2: There are some unresolved issues that require revision: 1. The abstract continues to refer to 296 million people with HBV, while the introduction has been updated with Global Hepatitis report data from WHO (254 million). Please be consistent. 2. The abstract refers to SD bioline as having superior specificity relative to the other assays but the confidence intervals are entirely overlapping. I would recommend recharacterizing this as all assays having good specificity, as the finding for SD bioline may be a chance finding. 3. Introduction: "An estimated 254 million people are living with hepatitis B. Only 13% 35 of people living with chronic hepatitis B infection had been diagnosed, presenting a big gap in 36 testing for HBV, and close to 3% had received antiviral therapy at the end of 2022. The WHO 37 African Region accounts for 63% of new hepatitis B infections (3)." I would recommend to report the statistics for Africa from this report (4% diagnosed and 0.2% treated). 4. "Several immunological methods are available" This isn't strictly an immunological method since we are detecting viral antigens, not antibodies. I would recharacterize as "serological methods" instead. Here I would also mention chemiluminescent immunoassays (CLIA) as this is the most commonly used method globally. 5. "The WHO prequalified diagnostic kits are recommended 63 internationally, however, standards for rapid diagnostic test prequalification by this organization 64 are quite stringent, making them expensive." Again, I am making the same point here: I do not think that the WHO PQ is more stringent than CE marking or FDA approval. I believe it is the opposite, fees for FDA or CE marking can exceed 200k USD whereas fees for WHO PQ are substantially lower and may be abridged if the product already has stringent regulatory approval. The WHO PQ process may be slower and requires an independent performance evaluation but to my knowledge there is no evidence it is more stringent. The cited article does not provide any evidence in support of the assertion made here. 6. Again I am making the same point here: The sample size calculation again makes reference to the Uganda local prevalence, which is not relevant, since a 50% prevalence was used in the study design. The authors used a study which showed a prevalence of 45% as the source estimate- surely this cannot be considered as the expected sensitivity. I would recalculate the sample size using a more realistic sensitivity estimate (eg. from the cited meta-analyiss) and include the prevalence of 50% as per the study design, as the current sample size calculation is not credible. 7. Again I am making the same point here: again, the study design refers to retesting if there was discrepancy, and then taking only the final result from 3 tests to evaluate at the final result in the event of discrepancy. At the risk of repeating myself, this is not an appropriate study design since it artificially inflates the diagnostic performance. Instead, take the result from the first RDT result only (disregard the outcome of repeated testing) and use this to evaluate performance. You may later specify the result of retesting in a separate section, but do not include this in the diagnostic test evaluation. 8. Is there any room to state what is the HBsAg prevalence in blood donors locally? 9. For the 7 samples which were negative with all 5 RDTs, was there a systematic difference observed in S/CO in the ELISA result eg. close to 1? Were ELISA results checked in duplicate? 10. Discussion: "SD Bioline, the WHO prequalified kit, had the best diagnostic performance" Again see point above about the lack of statistical difference in specificity for this test vs the others. 11. "False-negative results are a potential vehicle of silent transmit 186 ssion and spread of disease infection among people (21)." Check spelling of "transmission" and grammar of "disease infection" 12. Add to the limitations that the study design used 1:1 matching meaning that the predictive values were not interpretable for this population. 13. I would add to the interpretation that the RDTs were found not to be suitable for a blood donor screening population as the sensitivity was insufficient. ********** what does this mean?). If published, this will include your full peer review and any attached files.). If published, this will include your full peer review and any attached files.). If published, this will include your full peer review and any attached files.). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our For information about this choice, including consent withdrawal, please see our For information about this choice, including consent withdrawal, please see our For information about this choice, including consent withdrawal, please see our Privacy Policy..--> Reviewer #2: No ********** [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at . PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at . PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at . PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.. Please note that Supporting Information files do not need this step.. Please note that Supporting Information files do not need this step.. Please note that Supporting Information files do not need this step. |
| Revision 2 |
|
Dear Dr. Baliruno, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. Please submit your revised manuscript by Mar 27 2026 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.
If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at . Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at . Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at . Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.... We look forward to receiving your revised manuscript. Kind regards, Seth Agyei Domfeh, PhD Academic Editor PLOS One Journal Requirements: If the reviewer comments include a recommendation to cite specific previously published works, please review and evaluate these publications to determine whether they are relevant and should be cited. There is no requirement to cite these works unless the editor has indicated otherwise. Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. If you need to cite a retracted article, indicate the article’s retracted status in the References list and also include a citation and full reference for the retraction notice. [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions Comments to the Author Reviewer #2: (No Response) ********** 2. Is the manuscript technically sound, and do the data support the conclusions??> Reviewer #2: Partly ********** 3. Has the statistical analysis been performed appropriately and rigorously? -->?> Reviewer #2: Yes ********** 4. Have the authors made all data underlying the findings in their manuscript fully available??> The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.--> Reviewer #2: Yes ********** 5. Is the manuscript presented in an intelligible fashion and written in standard English??> Reviewer #2: Yes ********** Reviewer #2: The authors have addressed most of my comments from the previous revision. However, there are two outstanding issues: 1. In section 2.2 it continues to reference a prevalence of between 2.8 to 4.1% for calculating the sample size. This is the third time I am making the same point. The study is a case:control design, the design prevalence is 50%, you cannot use the expected population prevalence (3-4%) to estimate your sample size, when the study prevalence is 50%. The rationale given in the comments is effectively that the result was unwanted, which is unacceptable: "Thank you for your comment. I have used the average sensitivity 98.9% and average specificity of 96.7% from a recent meta-analysis conducted and the local prevalence of hepatitis among local blood donors in Uganda. Using 50% the sample size was too low for diagnostic accuracy." If I apply a sample size calculation in R: library(SampleSizeDiagnostics) # Example assumptions Se0 <- 0.95 # Round to more conservatives estimates Sp0 <- 0.95 p <- 0.50 # Case control prevalence= 50% CI <- 0.95 w <- 0.05 # Half-width of 5% SampleSizeDiagnostics(sn = Se0, sp = Sp0, p = p, w=w, CI=CI) Result: Sample size = 146 2. From the previous review: "13. I would add to the interpretation that the RDTs were found not to be suitable for a blood donor screening population, as the sensitivity was insufficient." "Thank you for this comment. I have also as part of the limitations." I cannot see this point added in the limitations. You mention the "window period" but the discussion does not specifically mention this important point that the observed sensitivity would be insufficient for blood donor screening according to WHO criteria, and low level HBsAg may been seen most commonly in people approaching spontaneous seroclearance/ functional cure. ********** what does this mean?). If published, this will include your full peer review and any attached files.). If published, this will include your full peer review and any attached files.). If published, this will include your full peer review and any attached files.). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our For information about this choice, including consent withdrawal, please see our For information about this choice, including consent withdrawal, please see our For information about this choice, including consent withdrawal, please see our Privacy Policy..--> Reviewer #2: No ********** [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] To ensure your figures meet our technical requirements, please review our figure guidelines: https://journals.plos.org/plosone/s/figures You may also use PLOS’s free figure tool, NAAS, to help you prepare publication quality figures: https://journals.plos.org/plosone/s/figures#loc-tools-for-figure-preparation. NAAS will assess whether your figures meet our technical requirements by comparing each figure against our figure specifications. |
| Revision 3 |
|
<p>Diagnostic Performance of Four Hepatitis-B Surface Antigen Conformité Européenne (CE) Marked and One WHO Prequalified Rapid Diagnostic Tests in Uganda PONE-D-25-07381R3 Dear Dr. Baliruno, We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements. Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication. An invoice will be generated when your article is formally accepted. Please note, if your institution has a publishing partnership with PLOS and your article meets the relevant criteria, all or part of your publication costs will be covered. Please make sure your user information is up-to-date by logging into Editorial Manager at Editorial Manager® and clicking the ‘Update My Information' link at the top of the page. For questions related to billing, please contact and clicking the ‘Update My Information' link at the top of the page. For questions related to billing, please contact and clicking the ‘Update My Information' link at the top of the page. For questions related to billing, please contact and clicking the ‘Update My Information' link at the top of the page. For questions related to billing, please contact billing support.... If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org. Kind regards, Seth Agyei Domfeh, PhD Academic Editor PLOS One Additional Editor Comments (optional): Reviewers' comments: |
| Formally Accepted |
|
PONE-D-25-07381R3 PLOS One Dear Dr. Baliruno, I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS One. Congratulations! Your manuscript is now being handed over to our production team. At this stage, our production department will prepare your paper for publication. This includes ensuring the following: * All references, tables, and figures are properly cited * All relevant supporting information is included in the manuscript submission, * There are no issues that prevent the paper from being properly typeset You will receive further instructions from the production team, including instructions on how to review your proof when it is ready. Please keep in mind that we are working through a large volume of accepted articles, so please give us a few days to review your paper and let you know the next and final steps. Lastly, if your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org. You will receive an invoice from PLOS for your publication fee after your manuscript has reached the completed accept phase. If you receive an email requesting payment before acceptance or for any other service, this may be a phishing scheme. Learn how to identify phishing emails and protect your accounts at https://explore.plos.org/phishing. If we can help with anything else, please email us at customercare@plos.org. Thank you for submitting your work to PLOS ONE and supporting open access. Kind regards, PLOS ONE Editorial Office Staff on behalf of Dr. Seth Agyei Domfeh Academic Editor PLOS One |