Peer Review History
| Original SubmissionMarch 5, 2026 |
|---|
|
-->PONE-D-26-05253-->-->Effect of Cold Ischemia Time and Fixative preparation on Breast Cancer Biomarker Expression: Implications for Resource-Limited Settings-->-->PLOS One Dear Dr. NDENGUE, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. Please submit your revised manuscript by Jun 03 2026 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. Please include the following items when submitting your revised manuscript:-->
If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter. If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols. As the corresponding author, your ORCID iD is verified in the submission system and will appear in the published article. PLOS supports the use of ORCID, and we encourage all coauthors to register for an ORCID iD and use it as well. Please encourage your coauthors to verify their ORCID iD within the submission system before final acceptance, as unverified ORCID iDs will not appear in the published article. Only the individual author can complete the verification step; PLOS staff cannot verify ORCID iDs on behalf of authors. We look forward to receiving your revised manuscript. Kind regards, Tomasz W. Kaminski Academic Editor PLOS One Journal Requirements: When submitting your revision, we need you to address these additional requirements. 1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and 2. We note that one of the authors of the present submission (Emile Telesphore MBOUDOU) is listed as a member of the ethics committee on the approval document provided. Could you please explain what your ethical committee's policies are when an Author is also a member of the committee, and how this was handled in this specific case? Thank you very much for your response to this request. 3. If the reviewer comments include a recommendation to cite specific previously published works, please review and evaluate these publications to determine whether they are relevant and should be cited. There is no requirement to cite these works unless the editor has indicated otherwise. Additional Editor Comments: Dear Authors, Thank you for submitting your manuscript. The study addresses an important and practical issue, and the manuscript is overall well written and clearly structured. The topic is relevant, particularly in the area of pre-analytical standardization in pathology workflows. After review, we obtained one major and two minor evaluations, and based on the combined feedback, the decision is major revision. The primary concerns relate to the study design and interpretation of the results as pointed out by two reviewers. Overall, the study has value, but substantial revision is needed, particularly in framing, interpretation, and discussion of limitations. We look forward to receiving a revised version. Best regards, Tomasz W Kaminski [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions -->Comments to the Author 1. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. --> Reviewer #1: Yes Reviewer #2: Yes Reviewer #3: Yes ********** -->2. Has the statistical analysis been performed appropriately and rigorously? --> Reviewer #1: Yes Reviewer #2: No Reviewer #3: I Don't Know ********** -->3. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.--> Reviewer #1: Yes Reviewer #2: Yes Reviewer #3: Yes ********** -->4. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.--> Reviewer #1: Yes Reviewer #2: Yes Reviewer #3: Yes ********** -->5. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)--> Reviewer #1: This manuscript evaluates on the effect of cold ischemia time and fixative preparation on breast cancer biomarker expression. This manuscript is well-written and impactful for methods standardization in pathology workflows. Reviewer #2: 1) Good manuscript? 2) Only one specimen has been used in this study although it has been divided into 50 sections. It might be useful to determine if there is variability between different tissue sources. 3) Statistical significance for the fixation time is also not provided. Reviewer #3: This study examines the effects of fixative preparation (4% neutral buffered versus non‑buffered formaldehyde) and cold ischemia time on ER, PR, and Ki67 immunohistochemical assessment in invasive breast cancer, with a stated focus on relevance to resource‑limited settings. The work addresses an important pre‑analytical issue; however, the study appears to be a single‑specimen experimental model, which substantially limits generalisability and the strength of the conclusions. The findings are largely confirmatory, with modest incremental value arising from the combined assessment of ischemia time and fixative preparation. Major Comments Single‑patient design and pseudoreplication: The study is based on tissue from a single patient, with multiple microsamples treated as experimental units. While this may be acceptable for a pilot or proof‑of‑concept study, it raises concerns regarding biological independence and pseudoreplication, particularly given the use of statistical testing. The study should be explicitly framed as exploratory, and language implying population‑level inference should be avoided. Estrogen receptor expression is known to show spatial heterogeneity within invasive breast carcinomas, as well as variability across different areas of the breast tissue more broadly. Sampling multiple microspecimens from a single tumour does not fully account for this biological variability and may bias results toward regions with uniformly high ER expression. This limitation is particularly relevant given the high baseline ER positivity in the index case, where small fixation‑related changes are unlikely to affect categorical scoring. The authors should explicitly acknowledge that heterogeneous or low‑ER tumours may be more vulnerable to pre‑analytical variation, and that findings from this single‑tumour model may not extrapolate to the full spectrum of ER expression patterns seen in clinical practice. This could also be true for Ki67 Limited novelty relative to existing literature. Previous studies have independently examined: The impact of cold ischemia time on breast cancer biomarkers Buffered versus unbuffered formalin fixation for ER and Ki67 evaluation The principal novelty appears to lie in the combined, controlled evaluation of fixation delay and fixative preparation. This distinction should be stated more clearly to avoid the impression that the individual findings are novel. Interpretation of statistically significant but small differences The manuscript reports statistically significant differences in ER and PR positivity percentages between fixatives, yet Allred scores remain unchanged. This raises questions regarding biological and clinical relevance. The conclusion that non‑buffered formaldehyde preserves biomarkers “comparably” should be tempered to acknowledge that measurable differences, albeit small, were observed. Why include fixation times of up to 144 hours as they exceed typical diagnostic workflows. The clinical relevance of these extremes should be explained or contextualised. Overgeneralisation to resource‑limited settings: While the study is motivated by challenges faced by laboratories in low‑resource settings, extrapolating from a highly controlled, single‑tumour experiment to real‑world practice may be premature. Although the need for validation is noted, this limitation should be more prominently foregrounded. MInor comments: Missing key breast‑specific literature: Multiple prior studies and ASCO/CAP guideline evidence demonstrate that ER, PR, and HER2 expression declines with prolonged cold ischemia, with ≤1 hour considered optimal and deterioration becoming detectable beyond 1–2 hours. These foundational studies should be explicitly acknowledged. Clarity of sample description: The reported microsample dimensions (“5–15 mm in length, 130 mm in width”) appear unusual and may reflect a typographical or unit error. ********** -->6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.--> Reviewer #1: No Reviewer #2: No Reviewer #3: No ********** [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] To ensure your figures meet our technical requirements, please review our figure guidelines: https://journals.plos.org/plosone/s/figures You may also use PLOS’s free figure tool, NAAS, to help you prepare publication quality figures: https://journals.plos.org/plosone/s/figures#loc-tools-for-figure-preparation. NAAS will assess whether your figures meet our technical requirements by comparing each figure against our figure specifications. |
| Revision 1 |
|
-->PONE-D-26-05253R1-->-->Pre-analytical variables affecting breast cancer biomarker expression: a controlled single-specimen study of fixation duration, cold ischemia time, and fixative preparation in a low-resource setting-->-->PLOS One Dear Dr. NDENGUE, Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process. Please submit your revised manuscript by Jun 28 2026 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file. Please include the following items when submitting your revised manuscript:-->
--> If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter. If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols. As the corresponding author, your ORCID iD is verified in the submission system and will appear in the published article. PLOS supports the use of ORCID, and we encourage all coauthors to register for an ORCID iD and use it as well. Please encourage your coauthors to verify their ORCID iD within the submission system before final acceptance, as unverified ORCID iDs will not appear in the published article. Only the individual author can complete the verification step; PLOS staff cannot verify ORCID iDs on behalf of authors. We look forward to receiving your revised manuscript. Kind regards, Tomasz W. Kaminski Academic Editor PLOS One Journal Requirements: If the reviewer comments include a recommendation to cite specific previously published works, please review and evaluate these publications to determine whether they are relevant and should be cited. There is no requirement to cite these works unless the editor has indicated otherwise. Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. If you need to cite a retracted article, indicate the article’s retracted status in the References list and also include a citation and full reference for the retraction notice. Additional Editor Comments: Dear Authors, Thank you for your careful revision and for thoroughly addressing the reviewers’ comments. The manuscript has improved considerably, and the responses satisfactorily resolved the major concerns raised during peer review. At this stage, only a few minor points remain before the manuscript can proceed toward acceptance: a) Please provide additional antibody information, including catalog numbers and concentrations/dilutions used. b) Please include a supplementary figure showing representative staining images observed in the study. These are relatively minor additions intended to improve methodological clarity and transparency. Thank you again for your efforts, and I look forward to receiving the revised version. Best regards, Tomasz W Kaminski [Note: HTML markup is below. Please do not edit.] Reviewers' comments: Reviewer's Responses to Questions -->Comments to the Author 1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.--> Reviewer #2: All comments have been addressed Reviewer #3: All comments have been addressed ********** -->2. Is the manuscript technically sound, and do the data support the conclusions? The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented. --> Reviewer #2: Yes Reviewer #3: Yes ********** -->3. Has the statistical analysis been performed appropriately and rigorously? --> Reviewer #2: Yes Reviewer #3: Yes ********** -->4. Have the authors made all data underlying the findings in their manuscript fully available? The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.--> Reviewer #2: Yes Reviewer #3: Yes ********** -->5. Is the manuscript presented in an intelligible fashion and written in standard English? PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.--> Reviewer #2: Yes Reviewer #3: No ********** -->6. Review Comments to the Author Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)--> Reviewer #2: The authors have ably respond to the comments raised. No additional comments based on the responses that have been provided. Reviewer #3: Thank you for addressing all the comments. Just added in some more antibody details (cat numbers etc) and concentrations please. Please include a supplemnetary figure showing representative staining observed. ********** -->7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files. If you choose “no”, your identity will remain anonymous but your review may still be made public. Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.--> Reviewer #2: No Reviewer #3: No ********** [NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.] To ensure your figures meet our technical requirements, please review our figure guidelines: https://journals.plos.org/plosone/s/figures You may also use PLOS’s free figure tool, NAAS, to help you prepare publication quality figures: https://journals.plos.org/plosone/s/figures#loc-tools-for-figure-preparation. NAAS will assess whether your figures meet our technical requirements by comparing each figure against our figure specifications. --> |
| Revision 2 |
|
Pre-analytical variables affecting breast cancer biomarker expression: a controlled single-specimen study of fixation duration, cold ischemia time, and fixative preparation in a low-resource setting PONE-D-26-05253R2 Dear Dr. NDENGUE, We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements. Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication. An invoice will be generated when your article is formally accepted. Please note, if your institution has a publishing partnership with PLOS and your article meets the relevant criteria, all or part of your publication costs will be covered. Please make sure your user information is up-to-date by logging into Editorial Manager at Editorial Manager® and clicking the ‘Update My Information' link at the top of the page. For questions related to billing, please contact billing support. If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org. Kind regards, Tomasz W. Kaminski Academic Editor PLOS One Additional Editor Comments: Dear Authors, Thank you for submitting the revised version of your manuscript and for carefully addressing the comments raised during the review process. The reviewers were satisfied with the revisions provided and agreed that the manuscript has been substantially improved and is technically sound. The remaining issues identified during the final evaluation were considered minor and largely editorial in nature. The additional methodological clarifications, expanded antibody information, and transparent discussion regarding the representative staining images were appreciated and adequately addressed the outstanding concerns. In light of the positive reviewer assessments and the satisfactory revisions, I am pleased to inform you that your manuscript is accepted for publication. Congratulations, and thank you for choosing the journal for the publication of your work. Sincerely, Tomasz W Kaminski Reviewers' comments: N/A |
| Formally Accepted |
|
PONE-D-26-05253R2 PLOS One Dear Dr. NDENGUE, I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS One. Congratulations! Your manuscript is now being handed over to our production team. At this stage, our production department will prepare your paper for publication. This includes ensuring the following: * All references, tables, and figures are properly cited * All relevant supporting information is included in the manuscript submission, * There are no issues that prevent the paper from being properly typeset You will receive further instructions from the production team, including instructions on how to review your proof when it is ready. Please keep in mind that we are working through a large volume of accepted articles, so please give us a few days to review your paper and let you know the next and final steps. Lastly, if your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org. You will receive an invoice from PLOS for your publication fee after your manuscript has reached the completed accept phase. If you receive an email requesting payment before acceptance or for any other service, this may be a phishing scheme. Learn how to identify phishing emails and protect your accounts at https://explore.plos.org/phishing. If we can help with anything else, please email us at customercare@plos.org. Thank you for submitting your work to PLOS ONE and supporting open access. Kind regards, PLOS ONE Editorial Office Staff on behalf of Dr. Tomasz W. Kaminski Academic Editor PLOS One |