PONE-D-25-29278Smartwatch-Based Detection of Loss of PulsePLOS ONE

Dear Dr. Demirel,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Two reviewers have provided their feedback below. Please address each point in turn. In particular, the reviewers ask for clarification in a number of areas and request that you check for consistent use of specific terminology.

Please submit your revised manuscript by Sep 04 2025 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.
  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.
  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Joanna Tindall, PhD

Staff Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf.

2. Please note that PLOS One has specific guidelines on code sharing for submissions in which author-generated code underpins the findings in the manuscript. In these cases, we expect all author-generated code to be made available without restrictions upon publication of the work. Please review our guidelines at https://journals.plos.org/plosone/s/materials-and-software-sharing#loc-sharing-code and ensure that your code is shared in a way that follows best practice and facilitates reproducibility and reuse.

3. Please include a complete copy of PLOS’ questionnaire on inclusivity in global research in your revised manuscript. Our policy for research in this area aims to improve transparency in the reporting of research performed outside of researchers’ own country or community. The policy applies to researchers who have travelled to a different country to conduct research, research with Indigenous populations or their lands, and research on cultural artefacts. The questionnaire can also be requested at the journal’s discretion for any other submissions, even if these conditions are not met. Please find more information on the policy and a link to download a blank copy of the questionnaire here: https://journals.plos.org/plosone/s/best-practices-in-research-reporting. Please upload a completed version of your questionnaire as Supporting Information when you resubmit your manuscript.

4. When completing the data availability statement of the submission form, you indicated that you will make your data available on acceptance. We strongly recommend all authors decide on a data sharing plan before acceptance, as the process can be lengthy and hold up publication timelines. Please note that, though access restrictions are acceptable now, your entire data will need to be made freely accessible if your manuscript is accepted for publication. This policy applies to all data except where public deposition would breach compliance with the protocol approved by your research ethics board. If you are unable to adhere to our open data policy, please kindly revise your statement to explain your reasoning and we will seek the editor's input on an exemption. Please be assured that, once you have provided your new statement, the assessment of your exemption will not hold up the peer review process.

5. If the reviewer comments include a recommendation to cite specific previously published works, please review and evaluate these publications to determine whether they are relevant and should be cited. There is no requirement to cite these works unless the editor has indicated otherwise.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: No

Reviewer #2: No

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: Thank you for this timely and important contribution. Your work introduces a novel dataset and framework for smartwatch-based detection of pulse loss. However, I have several important concerns and questions regarding the clarity and completeness of the methods, evaluation, and claims that need to be addressed.

1. Classification Labels and Evaluation Metrics

The paper states that the task is a three-class classification problem: pulse loss, normal operation, and idle (non-usage of smartwatch). However, Figure 4 and several parts of the text seem to only reference binary classification (e.g., "pulse" vs. "no pulse").

The metrics presented (sensitivity and specificity) are defined and reported in the binary classification context. Please clarify:

Are you performing binary or multiclass classification?

If multiclass, please provide per-class sensitivity and specificity, or macro- or weighted-averaged values for all classes.

Figure 4 and its caption should clearly show all 3 classes, if that’s the intention.

2. Terminology: "Deep Learning" vs. "Neural Network"

Throughout the manuscript, you refer to your deep learning model as a "neural network" and at times suggest it's an autoencoder. The current architecture more closely resembles a dual-encoder fusion model with LSTM layers, not an autoencoder.

Please revise the terminology to reflect the actual structure (e.g., “dual-stream CNN-LSTM” or “multi-modal encoder network”) to avoid confusion.

3. Proposed Method Identification

It is unclear which model is considered your proposed method: is it the deep learning architecture or one of the ensemble ML classifiers?

Please explicitly label the "proposed model" in the tables and main text and clearly distinguish between baseline comparisons and the final model.

4. Feature Dimensions and Signal Processing

You extracted features from all 3 IMU channels. Please clarify:

What is the final shape of the input feature vector to the classifiers?

Did you compute features for each axis individually and then concatenate them?

Or did you average features across the three axes before feeding to the model?

You state that no filtering or normalization was applied to IMU signals. Please justify this, especially since IMU signals tend to be noisy and scale-dependent.

Did you apply any resampling or time-alignment between PPG and IMU signals?

5. Ambiguity in "IMU" Definition

Please clearly define what you mean by “IMU” in the context of your device:

Are you using accelerometer only, gyroscope only, or both?

If both are present, are they all used in the signal fusion?

6. Cross-Validation and Subject Separation

You mention using Leave-One-Subject-Out (LOSO) cross-validation. Please confirm:

Was this strict LOSO, with no subject data leakage during feature extraction or normalization?

Was LOSO also applied to the deep learning model, or just traditional classifiers?

7. PPG-DaLiA Evaluation and Ground Truth

You evaluate your model on the PPG-DaLiA dataset, which does not contain ground truth for pulse loss or smartwatch idle status.

How did you define the labels for this dataset?

How did you differentiate normal vs. pulse loss vs. non-usage in the absence of ground truth?

Please explain how Doppler Ultrasound, used as ground truth in your dataset, relates to the evaluation on PPG-DaLiA, where no such ground truth is available.

8. Comparison with Existing Methods

The manuscript does not provide any benchmark comparison with previous methods in pulse detection or loss-of-pulse detection.

Please compare your model with at least one existing baseline or state-of-the-art method to contextualize the reported accuracy.

Reviewer #2: The paper presents a system based on smartwatches with the aim to detect the loss of pulse. The paper is well writen and structured although some improvements can be achieved:

- The smartwatch (Samsung Gear S3 Frontier) characteristics should be listed as well as the characteristics of the smartwatch sensors (vendor, resolution, maximum range...)

- The sentence "An example setup of the data collection is shown in Figure 1" is writen twice in page 4

- The reference to Figure 3 in page 5 is not clear. I am not sure if it is really to Figure 3 or Figure 2. Check it. Consequently, Figure 2 or 3 are not referenced or explained in the text as only one is referenced.

- The paper compares the performance of some models. However they are not explained in detail. That is, the parameters of each model are not presented. For instance, what is the architecture of the employed Neural Network Model? The obtained performance using a NN is very dependant of its architecture.

- Table 4 (in page 11) should be placed after (or closer) to its reference (page 13).

- The sentence in page 15 "Second, when the subjects wore ..." seems to be incomplete.

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: Yes: Kianoosh Kazemi

Reviewer #2: Yes: Francisco J. González-Cañete

**********

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

Revision 1

Thank you for the constructive feedback. Due to character limitations, we have provided detailed point-by-point responses to all reviewer comments in the attached PDF (appended at the end). We kindly ask the reviewers to refer to the attachment for the full responses and clarifications.

Attachments
Attachment
Submitted filename: PLOS___LossOfPulse_response.pdf
Decision Letter - Diaa Ahmed Mohamed Ahmedien, Editor

PONE-D-25-29278R1Smartwatch-Based Detection of Loss of PulsePLOS ONE

Dear Dr. Demirel,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please submit your revised manuscript by Dec 07 2025 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.
  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.
  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols. Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols.

We look forward to receiving your revised manuscript.

Kind regards,

Diaa Ahmed Mohamed Ahmedien, Ph.D.

Academic Editor

PLOS ONE

Journal Requirements:

If the reviewer comments include a recommendation to cite specific previously published works, please review and evaluate these publications to determine whether they are relevant and should be cited. There is no requirement to cite these works unless the editor has indicated otherwise.

Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. If you need to cite a retracted article, indicate the article’s retracted status in the References list and also include a citation and full reference for the retraction notice.

Additional Editor Comments:

Dear authors,

Thank you for your careful revision, it would be helpful to further consider the first revierwer's comments as a final revision.

Best regards

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #1: (No Response)

Reviewer #2: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: N/A

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: No

Reviewer #2: No

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: I would encourage the authors to discuss and compare their approach with other recent work on smartwatch-based loss of pulse detection. For instance, Shah et al. (NShah, K., Wang, A., Chen, Y. et al. Automated loss of pulse detection on a consumer smartwatch. Nature 642, 174–181 (2025). https://doi.org/10.1038/s41586-025-08810-9) presented an automated loss of pulse detection system implemented on a consumer smartwatch, using a multi-stage pipeline with PPG, motion data, and a lightweight convolutional neural network trained on 528 engineered features.

Given that your work also focuses on smartwatch-based pulse loss detection with machine learning, it would strengthen the manuscript to position your framework in relation to this paper, highlighting methodological similarities (e.g., multimodal PPG + motion sensing, simulation of pulse loss via occlusion) as well as differences (e.g., dataset openness, model architectures, evaluation settings).

Reviewer #2: The requirements of the reviewers seem to be accomplished, at least those made by me in the first round of review.

For future responses to reviewers I suggest to include the full text of the reviewer, instead of a short sentence, in order to be understandable by the rest of reviewers or even to be remembered by the reviewer that is doing the second round of reviews. That way, the reviewers know exactly what was requested in the first round of reviews.

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: Yes: Kianoosh Kazemi

Reviewer #2: No

**********

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

To ensure your figures meet our technical requirements, please review our figure guidelines: https://journals.plos.org/plosone/s/figures

You may also use PLOS’s free figure tool, NAAS, to help you prepare publication quality figures: https://journals.plos.org/plosone/s/figures#loc-tools-for-figure-preparation.

NAAS will assess whether your figures meet our technical requirements by comparing each figure against our figure specifications.

Revision 2

We would like to thank the reviewers for their useful comments on our paper. We have provided point-by-point responses to their comments and questions. We have also made several changes to the paper in response to their concerns.

In the updated manuscript, all the changes are marked in Blue.

The replies to the itemized comments are included where the responses are colored Green, and the corresponding changes to the manuscript are pasted and colored Blue.

In addition to the uploaded PDF, we have included the reviewers’ comments and our point-by-point responses below.

Reviewer 1 comments/questions:

I would encourage the authors to discuss and compare their approach with other recent work on smartwatch-based loss of pulse detection. For instance, Shah et al. (NShah, K., Wang, A., Chen, Y. et al. Automated loss of pulse detection on a consumer smartwatch. Nature 642, 174–181 (2025). https://doi.org/10.1038/s41586-025-08810-9) presented an automated loss of pulse detection system implemented on a consumer smartwatch, using a multi-stage pipeline with PPG, motion data, and a lightweight convolutional neural network trained on 528 engineered features.

Given that your work also focuses on smartwatch-based pulse loss detection with machine learning, it would strengthen the manuscript to position your framework in relation to this paper, highlighting methodological similarities (e.g., multimodal PPG + motion sensing, simulation of pulse loss via occlusion) as well as differences (e.g., dataset openness, model architectures, evaluation settings).

Response:

We thank the reviewer for the suggestion.

We agree that positioning our work in relation to Shah et al. (Nature, 2025) strengthens the manuscript.

We have updated Section 2 (Related Work) to acknowledge their use of arterial occlusion while highlighting our unique contributions. Furthermore, we have added a paragraph to Section 5 (Discussion) to clarify why a direct algorithmic comparison was not feasible due to the proprietary, hardware-specific nature of their approach.

Changes to Section 2: Related Work

In a recent development, Google introduced the "Pixel Watch 3", featuring a loss of pulse detection system detailed by Shah et al.

Their work demonstrates the viability of using a commercial smartwatch for this purpose, employing a multi-stage pipeline that processes PPG and motion data through a lightweight convolutional neural network trained on 528 engineered features.

Notably, Shah et al. utilized arterial occlusion to simulate pulse loss, validating the relevance of the occlusion methodology used in our own study.

This approach is crucial for realism, as it allows for data collection from healthy, conscious participants, avoiding the confounding physiological effects of anesthesia and medication inherent to intra-operative clinical studies.

However, despite these advancements, the underlying datasets remain private, and ground truth is often established via secondary indicators like standard pulse oximetry.

Our work addresses these gaps by providing a multimodal dataset.

Unlike standard clinical labeling, our use of continuous ultrasound imaging also enables us to visually confirm and precisely annotate the exact millisecond of blood flow cessation in the radial artery, minimizing label noise and ensuring the model is trained on high-fidelity physiological endpoints.

Changes to Section 5: Discussion

Fourth, regarding the state-of-the-art comparison, we note that while Shah et al also utilized arterial occlusion to simulate pulse loss, their detection pipeline relies on proprietary, device-specific firmware logic.

Specifically, the authors use dynamic adjustment of LED currents and photodiode gain upon initial detection to verify pulselessness.

Because our dataset consists of passively recorded signals from a standard consumer smartwatch without such active hardware control, a direct application of their specific multi-stage algorithm to our data was not feasible.

Future work could address this by developing hardware-agnostic approximations of these active sensing protocols, or by using programmable research platforms to facilitate direct benchmarking across different devices.

Attachments
Attachment
Submitted filename: Plos_response_to_the_reviewers.pdf
Decision Letter - Agnese Sbrollini, Editor

Smartwatch-Based Detection of Loss of Pulse

PONE-D-25-29278R2

Dear Dr. Demirel,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice will be generated when your article is formally accepted. Please note, if your institution has a publishing partnership with PLOS and your article meets the relevant criteria, all or part of your publication costs will be covered. Please make sure your user information is up-to-date by logging into Editorial Manager at Editorial Manager® and clicking the ‘Update My Information' link at the top of the page. For questions related to billing, please contact billing support.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Agnese Sbrollini

Academic Editor

PLOS One

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #1: All comments have been addressed

Reviewer #2: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: Yes

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: No

Reviewer #2: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: The manuscript is written scientifically. The authors has addressed all the comments. and the discussion is now fair with comparing their method with other existing methods.

Reviewer #2: The requirements of the reviewers seem to be accomplished. Hence, it can be accepted as is without further modificaciones.

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: Yes: KIANOOSH KAZEMI

Reviewer #2: Yes: Francisco Javier González Cañete

**********

Formally Accepted
Acceptance Letter - Agnese Sbrollini, Editor

PONE-D-25-29278R2

PLOS One

Dear Dr. Demirel,

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS One. Congratulations! Your manuscript is now being handed over to our production team.

At this stage, our production department will prepare your paper for publication. This includes ensuring the following:

* All references, tables, and figures are properly cited

* All relevant supporting information is included in the manuscript submission,

* There are no issues that prevent the paper from being properly typeset

You will receive further instructions from the production team, including instructions on how to review your proof when it is ready. Please keep in mind that we are working through a large volume of accepted articles, so please give us a few days to review your paper and let you know the next and final steps.

Lastly, if your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

You will receive an invoice from PLOS for your publication fee after your manuscript has reached the completed accept phase. If you receive an email requesting payment before acceptance or for any other service, this may be a phishing scheme. Learn how to identify phishing emails and protect your accounts at https://explore.plos.org/phishing.

If we can help with anything else, please email us at customercare@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Agnese Sbrollini

Academic Editor

PLOS One