Peer Review History

Original SubmissionSeptember 4, 2025
Decision Letter - Ziyu Qi, Editor

Dear Dr. Techapongsatorn,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

==============================

ACADEMIC EDITOR: After careful consideration, we believe that the experimental model proposed in this manuscript demonstrates both innovation and merit. However, the content requires substantial enhancement. Please focus on addressing the deficiencies in the Engineering Description as pointed out by Reviewer 2, particularly by clarifying the Integration with the Laparoscopic System. Additionally, please respond to the reviewer’s concern regarding the generalizability of the idealized experiments to clinical practice. For the references suggested by Reviewer 3, please review and evaluate this cited literature to determine its relevance and whether it should be included. If it is not relevant, citation is not mandatory.

==============================

Please submit your revised manuscript by Dec 12 2025 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org . When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.
  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.
  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols . Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols .

We look forward to receiving your revised manuscript.

Kind regards,

Ziyu Qi, M.D., Ph.D.

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. We note that you have indicated that there are restrictions to data sharing for this study. PLOS only allows data to be available upon request if there are legal or ethical restrictions on sharing data publicly. For more information on unacceptable data access restrictions, please see http://journals.plos.org/plosone/s/data-availability#loc-unacceptable-data-access-restrictions.

Before we proceed with your manuscript, please address the following prompts:

a) If there are ethical or legal restrictions on sharing a de-identified data set, please explain them in detail (e.g., data contain potentially identifying or sensitive patient information, data are owned by a third-party organization, etc.) and who has imposed them (e.g., a Research Ethics Committee or Institutional Review Board, etc.). Please also provide contact information for a data access committee, ethics committee, or other institutional body to which data requests may be sent.

b) If there are no restrictions, please upload the minimal anonymized data set necessary to replicate your study findings to a stable, public repository and provide us with the relevant URLs, DOIs, or accession numbers. For a list of recommended repositories, please see

https://journals.plos.org/plosone/s/recommended-repositories. You also have the option of uploading the data as Supporting Information files, but we would recommend depositing data directly to a data repository if possible.

We will update your Data Availability statement on your behalf to reflect the information you provide.

3. We note that your Data Availability Statement is currently as follows: All relevant data are within the manuscript and its Supporting Information files.

Please confirm at this time whether or not your submission contains all raw data required to replicate the results of your study. Authors must share the “minimal data set” for their submission. PLOS defines the minimal data set to consist of the data required to replicate all study findings reported in the article, as well as related metadata and methods (https://journals.plos.org/plosone/s/data-availability#loc-minimal-data-set-definition).

For example, authors should submit the following data:

- The values behind the means, standard deviations and other measures reported;

- The values used to build graphs;

- The points extracted from images for analysis.

Authors do not need to submit their entire data set if only a portion of the data was used in the reported study.

If your submission does not contain these data, please either upload them as Supporting Information files or deposit them to a stable, public repository and provide us with the relevant URLs, DOIs, or accession numbers. For a list of recommended repositories, please see https://journals.plos.org/plosone/s/recommended-repositories.

If there are ethical or legal restrictions on sharing a de-identified data set, please explain them in detail (e.g., data contain potentially sensitive information, data are owned by a third-party organization, etc.) and who has imposed them (e.g., an ethics committee). Please also provide contact information for a data access committee, ethics committee, or other institutional body to which data requests may be sent. If data are owned by a third party, please indicate how others may request data access.

4. Please remove your figures from within your manuscript file, leaving only the individual TIFF/EPS image files, uploaded separately. These will be automatically included in the reviewers’ PDF.

5. If the reviewer comments include a recommendation to cite specific previously published works, please review and evaluate these publications to determine whether they are relevant and should be cited. There is no requirement to cite these works unless the editor has indicated otherwise.

Additional Editor Comments :

After careful consideration, we believe that the experimental model proposed in this manuscript demonstrates both innovation and merit. However, the content requires substantial enhancement. Please focus on addressing the deficiencies in the Engineering Description as pointed out by Reviewer 2, particularly by clarifying the Integration with the Laparoscopic System. Additionally, please respond to the reviewer’s concern regarding the generalizability of the idealized experiments to clinical practice. For the references suggested by Reviewer 3, please review and evaluate the cited literature to determine its relevance and whether it should be included. If it is not relevant, citation is not mandatory.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

Reviewer #1: Yes

Reviewer #2: Partly

Reviewer #3: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously? -->?>

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available??>

The PLOS Data policy

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English??>

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

Reviewer #1: The authors describe a prototype smoke evacuator incorporating a distilled water bubbling trap, activated carbon filter, and ULPA filter was tested in a sealed chamber simulating

laparoscopic surgery using porcine liver tissue obtained from food-grade sources.

Air quality metrics, including PM1.0, PM2.5, PM10, VOC, and CO₂, were measured continuously. Results were compared to

baseline and performance benchmarks from commercial smoke evacuation systems.

The authors concluded that the developed modular smoke evacuator offers a promising and cost-effective solution to improve air quality and enhance occupational safety in operating rooms. Further clinical validation in live surgical environments is warranted.

It is known that commercially available smoke evacuators can be prohibitively expensive or impractical for many surgical centers, especially in resource-limited settings.

I believe this article is interesting and it should be published in its current form.

Reviewer #2: Major Weaknesses

Incomplete Engineering Description

1. The manuscript omits essential information regarding the suction mechanism, including the motor or pump specifications (flow rate, static pressure, power, and airflow path).

No details are provided on fluid dynamics, pressure gradients, or airflow direction through the device.

Without these parameters, the study’s reproducibility and validity cannot be confirmed.

2. Unclear Integration with the Laparoscopic System

The connection between the smoke evacuator and laparoscopic trocars or insufflation systems is not described.

There is no schematic diagram showing tubing layout, flow direction, or valve mechanisms to maintain pneumoperitoneum.

This omission prevents assessment of the device’s clinical feasibility and compatibility with standard equipment.

3. Simplified Experimental Model

The closed acrylic box used as a simulated peritoneal chamber does not reflect real surgical airflow dynamics.

No computational fluid dynamics modeling or flow visualization is included.

Results obtained under these idealized conditions cannot be extrapolated to clinical environments.

Reviewer #3: Thank you for the opportunity to review your excellent paper.

I would recommend adding an additional reference, a study similar to yours but for open, not laparoscopic, procedures.

Mordechai G. Sadowsky, Shawn K. Horner, Jeffrey W. Clymer, Crystal D. Ricketts. Determination of Smoke Particulates and Formaldehyde Removal from the Surgical Site by a Smoke Evacuator. Open Access Journal of Nursing (Sryahwa) Volume 6. 2:15-20, 2023.

I find the results for CO2 rather curious. Why does your smoke evacuator decrease the levels of CO2 back to baseline? I suspect that some of the CO2 is captured by the water bubbling trap. After some time, this might lower the pH of the water. Hopefully this would not affect the capture of hydrophilic VOCs.

In general, I don't think you want your system to be absorbing CO2, as most pneumoperitoneum in laparoscopy is achieved with the use of CO2, and it would be difficult to absorb it all. You may want to add as a limitation that you did not use a CO2 atmosphere, but this is not expected to affect the results of the study. Using a system like yours, frequent changes of the distilled water would be encouraged.

**********

what does this mean? ). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy

Reviewer #1: Yes: Diogo Casal

Reviewer #2: No

Reviewer #3: No

**********

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

To ensure your figures meet our technical requirements, please review our figure guidelines: https://journals.plos.org/plosone/s/figures

You may also use PLOS’s free figure tool, NAAS, to help you prepare publication quality figures: https://journals.plos.org/plosone/s/figures#loc-tools-for-figure-preparation.

NAAS will assess whether your figures meet our technical requirements by comparing each figure against our figure specifications.

Revision 1

Response to Editor and Reviewers

We would like to sincerely thank the Academic Editor and all reviewers for their insightful, constructive, and encouraging comments. Their feedback has been invaluable in improving the clarity, methodological transparency, and scientific strength of our manuscript.

All revisions have been incorporated accordingly, and the corresponding changes are highlighted in yellow throughout the revised manuscript.

Response to Academic Editor

We are deeply grateful to the Editor for recognising the innovation and potential merit of our experimental model. In response to the Editorial recommendation, we have substantially enhanced the Engineering Description as pointed out by Reviewer 2, with particular focus on the Integration with the Laparoscopic System.

A detailed description of the airflow pathway, three-way valve configuration, and passive evacuation mechanism has been added in the Methods section. In addition, a schematic figure and an explanatory paragraph in the Discussion were included to emphasise system reliability and pressure stability during laparoscopic use.

The Editor also highlighted the importance of addressing the generalisability of the experimental model to clinical practice. This has been revised and clearly discussed under the Limitations section, explaining that the closed-chamber setup represents a validated pre-clinical model consistent with previous literature (e.g., Sadowsky et al., 2023) and serves as a necessary step before live surgical validation.

Finally, the new reference suggested by Reviewer 3 has been carefully reviewed and incorporated in the Discussion as comparative evidence. The manuscript now includes explicit statements bridging the bench-top results to future clinical translation.

Response to Reviewer #1

We sincerely thank Reviewer #1 for the encouraging and positive evaluation.

We are truly grateful that you found the study design interesting and publication-worthy. Your remarks reinforce the potential value of low-cost, modular smoke evacuation systems, especially for resource-limited settings. No specific revisions were required, but we have ensured that all technical and stylistic aspects have been refined to maintain scientific clarity and conciseness.

Once again, we greatly appreciate your supportive review and kind recommendation for publication.

Response to Reviewer #2

We deeply appreciate Reviewer #2’s comprehensive and highly constructive technical feedback. Your insightful comments have significantly strengthened the engineering and methodological aspects of our work. Revisions made in response to your points are highlighted in yellow throughout the revised manuscript.

1. Incomplete Engineering Description

We have expanded the Prototype Device Development section to include complete details on the suction mechanism, airflow dynamics, and system configuration.

The revised text specifies the tubing dimensions, pressure resistance, estimated flow rate (8–10 L/min), and airflow direction across the three-stage filtration circuit.

Additionally, a new paragraph in the Discussion elaborates on the passive flow design and confirms that flow testing demonstrated stable evacuation without turbulence or reverse flow.

2. Unclear Integration with the Laparoscopic System

This issue has been thoroughly addressed.

A new subsection entitled Integration with Laparoscopic System now explains the connection between the evacuator and laparoscopic ports through a three-way valve, maintaining pneumoperitoneum integrity.

A schematic figure was added for clarity. The Discussion also emphasises that the system operates without a continuous suction motor, preserving intra-abdominal pressure and workflow continuity during surgery.

3. Simplified Experimental Model and Generalisability

We fully acknowledge this important concern.

The Limitations section now explicitly states that the sealed acrylic chamber represents a simplified pre-clinical model aligned with prior studies (e.g., Sadowsky et al., 2023).

We clarified that this controlled environment allows quantitative reproducibility before clinical trials. A statement has been added noting that future studies will include live surgical validation under realistic OR airflow conditions.

We are sincerely thankful for these comments, which helped us refine our technical presentation and clarify the translational potential of the work.

Response to Reviewer #3

We are very grateful for Reviewer #3’s generous comments and thoughtful suggestions, which helped strengthen both the discussion and contextual grounding of our study.

All relevant revisions are highlighted in yellow in the revised manuscript.

1. Additional Reference

We have reviewed the cited work by Mordechai G. Sadowsky et al., 2023 and found it highly relevant. It has been incorporated into the Discussion in three locations:

(1) as supportive evidence for particulate and formaldehyde removal performance,

(2) to compare noise and practical usability aspects, and

(3) in the Limitations section to justify the use of controlled chamber simulations in early-phase studies.

2. Clarification Regarding CO₂ Levels

We appreciate your keen observation.

A sentence has been added to the Discussion explaining that the observed CO₂ stability was due to the absence of CO₂ capture by filters and the dilution effect from chamber ventilation. We also clarified that the system was not intended to absorb CO₂, consistent with its role in laparoscopic pneumoperitoneum maintenance.

We are sincerely thankful for your insightful analysis and valuable recommendations that enriched the scientific discussion of our study.

Once again, we extend our deepest gratitude to the Editor and all three reviewers for their exceptional feedback.

Your critical and encouraging comments have helped us substantially improve the manuscript’s clarity, scientific rigor, and translational focus.

We believe the revised version now addresses all concerns comprehensively and is substantially strengthened for publication consideration.

All revisions have been highlighted in yellow in the revised manuscript for your convenience.

With our highest respect and appreciation,

On behalf of all authors,

Suphakarn Techapongsatorn, MD, PhD.

Faculty of Medicine Vajira Hospital

Navamindradhiraj University

Attachments
Attachment
Submitted filename: Response to Editor and Reviewers.docx
Decision Letter - Ziyu Qi, Editor

Development and Evaluation of a Modular Smoke Evacuator for Surgical Smoke Control

PONE-D-25-47575R1

Dear Dr. Techapongsatorn,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice will be generated when your article is formally accepted. Please note, if your institution has a publishing partnership with PLOS and your article meets the relevant criteria, all or part of your publication costs will be covered. Please make sure your user information is up-to-date by logging into Editorial Manager at Editorial Manager®  and clicking the ‘Update My Information' link at the top of the page. For questions related to billing, please contact billing support .

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Ziyu Qi, M.D., Ph.D.

Academic Editor

PLOS One

Additional Editor Comments (optional):

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

Reviewer #3: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions??>

Reviewer #3: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously? -->?>

Reviewer #3: Yes

**********

4. Have the authors made all data underlying the findings in their manuscript fully available??>

The PLOS Data policy

Reviewer #3: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English??>

Reviewer #3: Yes

**********

Reviewer #3: The authors have fully addressed all my concerns. The authors have fully addressed all my concerns. The authors have fully addressed all my concerns.

**********

what does this mean? ). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy

Reviewer #3: No

**********

Formally Accepted
Acceptance Letter - Ziyu Qi, Editor

PONE-D-25-47575R1

PLOS One

Dear Dr. Techapongsatorn,

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS One. Congratulations! Your manuscript is now being handed over to our production team.

At this stage, our production department will prepare your paper for publication. This includes ensuring the following:

* All references, tables, and figures are properly cited

* All relevant supporting information is included in the manuscript submission,

* There are no issues that prevent the paper from being properly typeset

You will receive further instructions from the production team, including instructions on how to review your proof when it is ready. Please keep in mind that we are working through a large volume of accepted articles, so please give us a few days to review your paper and let you know the next and final steps.

Lastly, if your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

You will receive an invoice from PLOS for your publication fee after your manuscript has reached the completed accept phase. If you receive an email requesting payment before acceptance or for any other service, this may be a phishing scheme. Learn how to identify phishing emails and protect your accounts at https://explore.plos.org/phishing.

If we can help with anything else, please email us at customercare@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Ziyu Qi

Academic Editor

PLOS One

Open letter on the publication of peer review reports

PLOS recognizes the benefits of transparency in the peer review process. Therefore, we enable the publication of all of the content of peer review and author responses alongside final, published articles. Reviewers remain anonymous, unless they choose to reveal their names.

We encourage other journals to join us in this initiative. We hope that our action inspires the community, including researchers, research funders, and research institutions, to recognize the benefits of published peer review reports for all parts of the research system.

Learn more at ASAPbio .