Peer Review History

Original SubmissionMay 9, 2025
Decision Letter - Gauri Mankekar, Editor

PONE-D-25-24910Usability of a Hearing Test Mobile App Across GenerationsPLOS ONE

Dear Dr. Pastucha,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

==============================

ACADEMIC EDITOR:  Smart phone apps for hearing evaluation became popular during the COVID 19 pandemic. However, they continue to have several limitations. Few test bone conduction hearing and most are not user friendly especially for older patients.  The apps can be useful for surveillance or monitoring.  Authors have stated that " The app generally yielded results comparable to pure tone audiometry". Did the app test bone conduction? Table 2 mentions severity of hearing loss but does not mention whether the loss was conductive, sensorineural of mixed. How were the App vs Audiometry test results in patients with asymmetrical conductive or sensorineural hearing loss?  Please discuss. Also, please respond to Reviewer 1's comments.

==============================

Please submit your revised manuscript by Jul 23 2025 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org . When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.
  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.
  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: https://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols . Additionally, PLOS ONE offers an option for publishing peer-reviewed Lab Protocol articles, which describe protocols hosted on protocols.io. Read more information on sharing protocols at https://plos.org/protocols?utm_medium=editorial-email&utm_source=authorletters&utm_campaign=protocols .

We look forward to receiving your revised manuscript.

Kind regards,

Gauri Mankekar, MD,PhD,FACS

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. In the online submission form, you indicated that “The data underlying the results presented in the study are available from the corresponding author upon reasonable request.”

All PLOS journals now require all data underlying the findings described in their manuscript to be freely available to other researchers, either 1. In a public repository, 2. Within the manuscript itself, or 3. Uploaded as supplementary information.

This policy applies to all data except where public deposition would breach compliance with the protocol approved by your research ethics board. If your data cannot be made publicly available for ethical or legal reasons (e.g., public availability would compromise patient privacy), please explain your reasons on resubmission and your exemption request will be escalated for approval.

3. Please include captions for your Supporting Information files at the end of your manuscript, and update any in-text citations to match accordingly. Please see our Supporting Information guidelines for more information: http://journals.plos.org/plosone/s/supporting-information.

4. Please review your reference list to ensure that it is complete and correct. If you have cited papers that have been retracted, please include the rationale for doing so in the manuscript text, or remove these references and replace them with relevant current references. Any changes to the reference list should be mentioned in the rebuttal letter that accompanies your revised manuscript. If you need to cite a retracted article, indicate the article’s retracted status in the References list and also include a citation and full reference for the retraction notice.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: I Don't Know

Reviewer #2: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: No

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: Regarding the dissimilarity of results obtained using the app compared with conventional audiometry, consider that hearing loss will tend to worsen with age, and tinnitus often accompanies that hearing loss, making self-assessment more challenging (i.e. the patient will confuse the test tone with the tinnitus). You could therefore conceivably attribute the progressively less similar results as the participants age to tinnitus rather than incompetence with technology.

Reviewer #2: The study provides valuable insight into usability of mobile apps for hearing across generations and is of interest to the telemedicine, audiology, and app development communities. The authors state that the data are available "from the corresponding author upon reasonable request". It would be good if anonymized datasets could be made available.

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean? ). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy .

Reviewer #1: Yes:  Cynthia Collier

Reviewer #2: No

**********

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/ . PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org . Please note that Supporting Information files do not need this step.

Revision 1

Dear Editor,

We would like to thank you for your valuable feedback and comments on our manuscript. Your suggestions have greatly helped us to improve the clarity and quality of the article. Below are our responses to your comments:

Smart phone apps for hearing evaluation became popular during the COVID 19 pandemic. However, they continue to have several limitations. Few test bone conduction hearing and most are not user friendly especially for older patients. The apps can be useful for surveillance or monitoring.

Authors have stated that "The app generally yielded results comparable to pure tone audiometry". Did the app test bone conduction? Table 2 mentions severity of hearing loss but does not mention whether the loss was conductive, sensorineural of mixed. How were the App vs Audiometry test results in patients with asymmetrical conductive or sensorineural hearing loss? Please discuss.

Response:

1. Bone conduction testing

The mobile app used in our study assessed air conduction thresholds only. No bone conduction testing was performed, either by the app or performed by an audiologist. Mobile hearing test apps are generally similar to screening audiometry rather than comprehensive air and bone conduction testing. These apps are designed for self-testing and provide preliminary hearing information that encourages users to seek full clinical evaluation rather than replace professional testing. We would not expect users to perform bone conduction self-testing, especially since even trained audiologists sometimes encounter difficulties with bone conduction testing in complex cases.

In the manuscript, we used the term pure tone audiometry (PTA) as a commonly understood shorthand for air conduction audiometry conducted by an audiologist. We acknowledge that this terminology may be misleading, as PTA formally includes both air and bone conduction measurements. We have clarified this point in the revised manuscript to prevent any ambiguity.

We have added following sentences: “Although the “Hearing Test” app was originally designed to assess air conduction only, its developers have recently introduced a version capable of bone conduction testing using a dedicated bone vibrator [29]. However, this hardware is not yet commercially available to the general public and is currently used primarily in research and professional settings. For this reason, and in order to ensure accessibility and scalability, we selected the air-conduction-only version of the app, which remains freely available and does not require additional specialized equipment.” in lines 120-124.

And “Pure tone audiometry (air conduction only) was performed by an audiologist using the Sense Examination Platform” in line 126.

2. Lack of information on hearing loss type (conductive/sensorineural/mixed)

Since the app and the audiological examination both measured only air conduction thresholds, we were unable to determine the type of hearing loss (i.e., whether it was conductive, sensorineural, or mixed). The classification in Table 2 refers exclusively to the degree of hearing loss based on air conduction thresholds, following WHO guidelines.

We have clarified this in the limitation section: “Another limitation is the lack of bone conduction assessment in both the app and the reference test. Consequently, we were unable to determine the type of hearing loss (i.e., conductive, sensorineural, or mixed), which limits clinical interpretation. However, this limitation is inherent to the nature of mobile hearing applications. Mobile hearing test apps are generally similar to screening audiometry rather than comprehensive air and bone conduction testing. These apps are designed for self-testing and provide preliminary hearing information that encourages users to seek full clinical evaluation rather than replace professional testing. We would not expect users to perform bone conduction self-testing, especially since even trained audiologists sometimes encounter difficulties with bone conduction testing in complex cases.” in lines 363-369.

3. Performance in asymmetrical hearing loss

The app includes an automatic masking feature, which is activated when stimuli are presented at high intensity levels. In such cases, masking noise is delivered in the non-test ear, accompanied by an on-screen instruction informing the user not to respond to the noise. This feature is intended to reduce the risk of cross-hearing and improve measurement accuracy in cases of asymmetrical hearing. A description of this function is provided in lines 117–118 and 298–301 of the manuscript.

We also sincerely thank the reviewers for their insightful and constructive comments. Below we provide our detailed responses to each point raised, along with the corresponding changes made to the manuscript.

Reviewer #1: Regarding the dissimilarity of results obtained using the app compared with conventional audiometry, consider that hearing loss will tend to worsen with age, and tinnitus often accompanies that hearing loss, making self-assessment more challenging (i.e. the patient will confuse the test tone with the tinnitus). You could therefore conceivably attribute the progressively less similar results as the participants age to tinnitus rather than incompetence with technology.

Response:

We thank the reviewer for this comment. Indeed, tinnitus is commonly associated with age-related hearing loss and can interfere with tone perception during hearing tests. However, based on direct observation of participants during the app-based testing, we noted that older adults most frequently struggled with technical aspects of the procedure, including app installation, interpreting the interface, and following the testing instructions. These observations suggest that usability challenges played a more prominent role than auditory interference in this context. However, we recognize that tinnitus may have additionally impacted some participants’ ability to detect test tones, particularly in self-administered conditions.

We have included a sentence in the Discussion section: “It is also possible that tinnitus may have interfered with test performance in some participants, especially among older adults. Tinnitus is known to complicate tone detection, particularly during self-assessments, by masking or mimicking test stimuli. Although we did not specifically assess the presence of tinnitus in our participants, its higher prevalence among older adults may partly explain some of the discrepancies observed between app-based and conventional audiometry results in this age group [37,38].” in lines 272-276.

Reviewer #2: The study provides valuable insight into usability of mobile apps for hearing across generations and is of interest to the telemedicine, audiology, and app development communities. The authors state that the data are available "from the corresponding author upon reasonable request". It would be good if anonymized datasets could be made available.

Response:

We thank the reviewer for the comment. We have prepared the relevant anonymized data and included them as supplementary material with this revised submission.

Attachments
Attachment
Submitted filename: Response to Reviewers.docx
Decision Letter - Gauri Mankekar, Editor

Usability of a Hearing Test Mobile App Across Generations

PONE-D-25-24910R1

Dear Dr. Pastucha,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice will be generated when your article is formally accepted. Please note, if your institution has a publishing partnership with PLOS and your article meets the relevant criteria, all or part of your publication costs will be covered. Please make sure your user information is up-to-date by logging into Editorial Manager at Editorial Manager®  and clicking the ‘Update My Information' link at the top of the page. If you have any questions relating to publication charges, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Gauri Mankekar, MD,PhD,FACS

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Formally Accepted
Acceptance Letter - Gauri Mankekar, Editor

PONE-D-25-24910R1

PLOS ONE

Dear Dr. Pastucha,

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now being handed over to our production team.

At this stage, our production department will prepare your paper for publication. This includes ensuring the following:

* All references, tables, and figures are properly cited

* All relevant supporting information is included in the manuscript submission,

* There are no issues that prevent the paper from being properly typeset

You will receive further instructions from the production team, including instructions on how to review your proof when it is ready. Please keep in mind that we are working through a large volume of accepted articles, so please give us a few days to review your paper and let you know the next and final steps.

Lastly, if your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

If we can help with anything else, please email us at customercare@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Gauri Mankekar

Academic Editor

PLOS ONE

Open letter on the publication of peer review reports

PLOS recognizes the benefits of transparency in the peer review process. Therefore, we enable the publication of all of the content of peer review and author responses alongside final, published articles. Reviewers remain anonymous, unless they choose to reveal their names.

We encourage other journals to join us in this initiative. We hope that our action inspires the community, including researchers, research funders, and research institutions, to recognize the benefits of published peer review reports for all parts of the research system.

Learn more at ASAPbio .